0% found this document useful (0 votes)
125 views

Lifesaver 108 - Specification

This document summarizes the specifications of the Lifesaver-108 defibrillator in 12 sections. It provides details on the display, dimensions, controls, power supply, environmental tolerances, defibrillation capabilities, ECG functionality, printer specifications, SpO2, NIBP, AED mode, and PACER settings. Key specifications include a 7" color display, 8 kg weight, 3 hour battery backup, biphasic defibrillation up to 200J, 12-lead ECG acquisition, and monitoring of vital signs like oxygen saturation and blood pressure.

Uploaded by

Luis Anzola
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
125 views

Lifesaver 108 - Specification

This document summarizes the specifications of the Lifesaver-108 defibrillator in 12 sections. It provides details on the display, dimensions, controls, power supply, environmental tolerances, defibrillation capabilities, ECG functionality, printer specifications, SpO2, NIBP, AED mode, and PACER settings. Key specifications include a 7" color display, 8 kg weight, 3 hour battery backup, biphasic defibrillation up to 200J, 12-lead ECG acquisition, and monitoring of vital signs like oxygen saturation and blood pressure.

Uploaded by

Luis Anzola
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

Specifications

Pages
1/4 Product: Lifesaver - 108

1. General

Display : 7”Colour TFT


Display area : 150 mm x 90 mm
Dimensions : H280 x W 330 x D220 mm (Max) without Paddle
Weight : Approx.8 Kgs.
Unit ON-OFF : By front panel push button
Operation : By front Panel keyboard, Encoder
Alarm : User adjustable high & low limit, Audible and visual
alarming
Trend : 24 hrs Graphical and Tabular trend of ECG, SpO2, NIBP
EtCO2
Charge/ Discharge : Charge and Shock buttons on front panel and paddles.
Control

2. Power Supply

Input : 150 - 276 VAC, 50 Hz, 1 Ø


Battery : 12 V 4.5Ah, Rechargeable
Back-up time : 3 hours min. on full charge in ECG mode
70--100 shock of 150 Joules deliveries (Full charge).
Power consumption : 30 Watts (max.) in ECG mode

3. Environmental

Operating Temp : -5ºC to +50 ºC


Humidity : Up to 95 % non-condensing

4. Defib (External Paddles)

Operating Mode : Manual


Waveform : Biphasic Truncated Exponential
Energy : 2 to 200 joules in 12 steps selection
Energy display : Displays selected and charged/delivered energy
Synchronous Cardio Version : Energy delivery begins within 60 ms of the QRS Peak
Charging Time : 6 seconds on Mains
Paddle Assembly : External paddle assembly with coil retractable high
voltage cable. Adults & Pediatric paddle integrated to
same handle
Check Mode : Manual check mode
Patient Impedance range : 35 ohm to 250 ohm
Shock Delivery : Through paddles or pads

Rev 3
Specifications
Pages
2/4 Product: Lifesaver - 108

5. Defib (Internal Paddles)(Optional)

Operating Mode : Manual


Waveform : Biphasic Truncated Exponential
Energy : 2 to 50 joules in 8 steps selection
Synchronous Cardio Version : Energy delivery begins within 60 ms of the QRS Peak
Charging Time : 6 seconds on Mains
Paddle Assembly : Internal paddle assembly with paddle extension cord

6. ECG

Frequency response : 0.5 Hz to 35 Hz


Patient isolation : Optical isolation
Lead selection : Lead I, II and III
Gain setting : 0.5, 1.0 and 2.0
CMRR : > 90 db
Sweep Speed : 12.5, 25.0, 50.0 mm/sec
Digital filter : Digital signal processing for good quality of ECG
ECG traces viewing : 4 seconds of ECG traces can be viewed
Range : 10 – 250 BPM
Accuracy : +/- 2 BPM
Resolution : 1 BPM
Tone : Beep on every QRS detected
Alarm limit setting : 10 – 250 (step of 1)

7. 12 Channel ECG

Frequency response : 0.05 Hz to 150 Hz with notch at 50Hz


Filter : DSP
Input impedance : >10 MΩ
Time constant : > 3.2 sec
A/D conversion : 12 bit
Noise : < 20µVp-p
Leakage current : < 10 µA
Patient isolation : Optical isolation.
Gain : 5, 10 & 20 mm/mV selectable.
CMRR : > 120 dB
Sweep speed : 25.0 mm/S.
Accuracy of HR : +/- 2 BPM.
Sampling Frequency : 250 samples per second simultaneous
Storage : 12 lead resting ECG,
Printout : Manual printing with grid On/Off facility, print Colour
reports

Rev 3
Specifications
Pages
3/4 Product: Lifesaver - 108

Features of Tab:

 Acquisition
o Choice of ’12 Lead ECG’ and ‘Remote Monitoring’
o Simultaneous acquisition of 12 Leads of ECG @ 250 samples per second per
channel with ADC resolution of 12-bit.
o Frequency Response – DSP filter of 0.05 to 150Hz with notch at 50Hz
o Facility to store 12 lead resting ECG
o Printout on Wi-Fi printer
o Facility to send email
o Facility to view patients vitals in central hospital

 Settings
o 12 lead ECG or Remote monitoring
o ECG configuration

8. Printer

Paper Type : Thermal paper 50/55mm width, 20 meters length for


2” printer
Event Printing : Prints 5 Seconds pre & post ECG with Date, Time,
Energy and mode, Shock Status
ECG in Defib Mode : Prints last 10sec ECG through paddle
Print Annotates in Defib Mode : Time, Date, Mode, lead name, gain, sweep speed
ECG in Monitor Mode : Prints last 10sec ECG through cable
Print Annotates in Monitor Mode: Time, Date, selected Lead name, Gain, Sweep speed
Manual Defib Check Mode : Prints the status of Shock, Energy, Date and Time

9. SpO2
Range : 0 - 100%
Accuracy : 70 - 100% +2 digits, 50 - 69% +3 digits
Pulse rate : Displayed in ECG window, when PR is selected
Pulse beep : Yes, with volume control
PR Range : 10 to 250 bpm (+/- 2 bpm)
Pitch variation with change in oxygen saturation.

10. NIBP

Systolic, Diastolic and Mean arterial pressure


Range : 10-250mmHg
Method : Oscillometric
NIBP interval : User selectable - 2 min. to 250 min
Modes : Manual, auto
Cuff deflation : Automatic
Safety features : Automatic deflation if cuff pressure exceeds 280 mmHg
NIBP status display as:
“Ready”, when not taking reading

Rev 3
Specifications
Pages
4/4 Product: Lifesaver - 108

“Measuring”, when taking reading and


“Error”, if reading could not be taken due to some error while measuring
Patient Type : Adult, Pediatric, Neonatal

11. AED Mode

Wave form : Biphasic truncated exponential


AED Shock Series : Energy level 200J
Charge Command : Automatic after identifying shockable rhythm
Shock Command : Front panel button, shock
Controls : Analyze
Defibrillation Electrode : Self adhesive, adult pads
Voice and Text prompts : Extensive text and audible messages to guide user

12. PACER

Wave form : Monophasic


Modes : Demand or Non Demand
Pulse Current : From 10mA to 150 mA (resolution of 10 mA),
Pulse Duration : 40 ms (+/- 1ms)
Pacing rate : From 20 ppm to 250 ppm (increments of 1 ppm)
Refractory period : 340 ms (from 30 to 80 ppm); 240 ms(from 90 to 180
ppm).
Maximum output voltage: 150 V.
Impedance range : 200 ohm to 1 k ohm

*
Due to our continuous product development program, the specification may change without prior notice.