USEPA2004 P Hinsolids
USEPA2004 P Hinsolids
:2178
Facility: Mansfield Revision 3
Department:Wet Chemistry Published Date:6/27/2012 6:22:14 AM
Title: pH in Aqueous and Soil Samples Page 1 of 6
The data report packages present the documentation of any method modification related to the
samples tested. Depending upon the nature of the modification and the extent of intended use, the
laboratory may be required to demonstrate that the modifications will produce equivalent results for
the matrix. Approval of all method modifications is by one or more of the following laboratory
personnel before performing the modification: Area Supervisor, Department Supervisor, Laboratory
Director, or Quality Assurance Officer.
This method is restricted to use by or under the supervision of analysts experienced in the
operation of the pH equipment and in the interpretation of data. Each analyst must demonstrate the
ability to generate acceptable results with this method by performing an initial demonstration of
capability, analyzing a proficiency test sample and completing the record of training.
After initial demonstration, ongoing demonstration is based on acceptable laboratory performance
of at least a quarterly laboratory control sample or acceptable performance from an annual
proficiency test sample. A major modification to this procedure requires demonstration of
performance. The identification of major method modification requiring performance demonstration
is directed by the Quality Assurance Officer and/or Laboratory Director on a case-by-case basis.
2. Summary of Method
An aliquot of a water/aqueous sample (analyzed “as is”); soil/sediment or waste sample; is mixed
with laboratory reagent water. The pH of the solution is measured electrometrically with a pH meter
using a combination electrode.
3. Reporting Limits
There is no reporting limit for this method.
4. Interferences
Samples with very low or very high pH may produce incorrect readings on the meter. For samples
with a true pH of > 10, the measured pH may be incorrectly low. This error can be minimized by
Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work,
the published version of the document should be viewed online.
Document Type: SOP-Technical Pre-Qualtrax Document ID: W-003 and W-004
Alpha Analytical, Inc. ID No.:2178
Facility: Mansfield Revision 3
Department:Wet Chemistry Published Date:6/27/2012 6:22:14 AM
Title: pH in Aqueous and Soil Samples Page 2 of 6
using a low-sodium-error electrode. Strong acid solutions, with a true pH of < 1, may give
incorrectly high pH measurements.
Temperature fluctuations may cause measurement errors. The use of an automatic temperature
compensation probe will minimize temperature errors. If this is not an option, the temperature and
pH shall be recorded at the time of the pH reading.
Measurement errors can occur when the electrodes become coated. If an electrode becomes
coated with sample material that will not rinse off, perform the following maintenance of the probe.
1) clean the electrode in an ultrasonic bath, or 2) wash the electrode with detergent, rinse several
times with reagent water, place in 1:10 HCl solution so that the lower third of the electrode is
submerged, and then thoroughly rinse with reagent water.
Solvents should be avoided as they may damage the probe operation.
9. Quality Control
The laboratory must maintain records to document the quality of data that is generated. Ongoing
data quality checks are compared with established performance criteria to determine if the results
of analyses meet the performance characteristics of the method.
9.1 Blank(s)
Not applicable to this method.
9.2 Laboratory Control Sample (LCS)
A second source pH 7.00 buffer is analyzed immediately following pH meter calibration.
Criterion for acceptable recovery is + 0.05 pH units. Corrective action: If the LCS is outside the
acceptable range, re-calibrate the pH meter. Acceptable second source LCS recovery must be
obtained prior to the analysis of samples.
9.3 Initial Calibration Verification (ICV)
Refer to Section 9.2.
9.4 Continuing Calibration Verification (CCV)
A CCV is analyzed every 10 samples and at the end of an analytical batch to monitor the
accuracy of the calibration during analysis. The CCV is a second source buffer solution (4.00,
7.00, or 10.00) in the appropriate range of the samples. Criterion for acceptable recovery is +
0.05 pH units. Corrective action: If the CCV is outside of acceptable range, re-calibrate the pH
meter and re-analyze all samples since the last acceptable CCV.
9.5 Matrix Spike
Not applicable to this method.
9.6 Laboratory Duplicate
Samples are analyzed in duplicate at a frequency of 10% or 1 per batch of 10 samples or less,
whichever is more frequent. Relative percent difference (RPD) between sample duplicate
Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work,
the published version of the document should be viewed online.
Document Type: SOP-Technical Pre-Qualtrax Document ID: W-003 and W-004
Alpha Analytical, Inc. ID No.:2178
Facility: Mansfield Revision 3
Department:Wet Chemistry Published Date:6/27/2012 6:22:14 AM
Title: pH in Aqueous and Soil Samples Page 4 of 6
results must be < 15%. Corrective action: If RPD does not meet criteria, determine if the cause
is matrix related or lab related. If matrix related, include a narrative with the client report
explaining the matrix effect on the imprecision of the pH result. If error was lab related, repeat
the analysis.
10. Procedure
10.1 Equipment Set-up
Turn on pH meter and allow to warm up.
Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work,
the published version of the document should be viewed online.
Document Type: SOP-Technical Pre-Qualtrax Document ID: W-003 and W-004
Alpha Analytical, Inc. ID No.:2178
Facility: Mansfield Revision 3
Department:Wet Chemistry Published Date:6/27/2012 6:22:14 AM
Title: pH in Aqueous and Soil Samples Page 5 of 6
Weigh 20g of soil/sediment or waste in a 50mL beaker. Add 20mL reagent water and
continuously stir the soil/water slurry on a magnetic stir plate for 5 minutes. NOTE: The analyst
may use 10g soil and 10mL reagent water, dependent upon the sample, provided the soil/water
ratio is 1:1. For sediments with high organic content: Weigh 10g to 20g in 125mL plastic bottle
with a 1/1 ratio of DI water. Shake sample for 30 minutes on a Shaker Table. Allow to settle
and then shake again for 5 minutes and let settle again.
Let a soil/water slurry suspension stand for about 1 hour to allow most of the suspended
particles to settle out. Let a waste/water slurry suspension stand for about 15 minutes to settle
particles.
Place the pH electrode in the beaker and adjust the level of the electrode such that it is just
below the suspension.
Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work,
the published version of the document should be viewed online.
Document Type: SOP-Technical Pre-Qualtrax Document ID: W-003 and W-004
Alpha Analytical, Inc. ID No.:2178
Facility: Mansfield Revision 3
Department:Wet Chemistry Published Date:6/27/2012 6:22:14 AM
Title: pH in Aqueous and Soil Samples Page 6 of 6
16. Attachments
None.
Printouts of this document may be out of date and should be considered uncontrolled. To accomplish work,
the published version of the document should be viewed online.
Document Type: SOP-Technical Pre-Qualtrax Document ID: W-003 and W-004