Eups 70
Eups 70
EUP-S70
INSTRUCTION MANUAL
Tokyo , Japan
Q1E-EP1152-6
0123
Manufacturer: Hitachi,Ltd
2-16-1, Higashi-Ueno,Taito-ku,
Tokyo,110-0015,Japan
+81-3-6284-3668
http://www.hitachi.com/businesses/healthcare/index.html
European
Representative: Hitachi Medical Systems GmbH
Otto-von-Guericke-Ring 3 D-65205 Wiesbaden,
Germany
Local Distributor:
( 1 ) Q1E-EP1152
About this manual
( 2 ) Q1E-EP1152
Graphical Symbols for Use in Labeling of Hitachi Ultrasound Probes
Some graphical symbols that are used in labeling of Hitachi Ultrasound Probes are
compliant with EN980:2008 standard. Refer to the following table about the meanings
of them.
Hitachi,Ltd
2-16-1,Higashi-Ueno,Taito-ku,
Manufacturer Company
Name and Address Tokyo,110-0015,Japan
+81-3-6284-3668
http://www.hitachi.com/businesses
/healthcare/index.html
( 3 ) Q1E-EP1152
Definition of symbol
The following symbol is also used for HITACHI Ultrasound Probes.
( 4 ) Q1E-EP1152
CONTENTS
Page
1. Introduction ................................................ 1
1.1 Features ........................................................ 1
1.2 Principles of operation ......................................... 1
1.3 Intended Use .................................................... 1
1.4 Composition ..................................................... 2
1.5 Construction .................................................... 2
2. Inspection before Use ....................................... 3
2.1 Inspection for Appropriate Connection ........................... 3
2.2 Inspection for Material Surface ................................. 3
3. Operation Procedure ......................................... 4
4. Cleaning, Disinfection and Sterilization .................... 5
4.1 Point of use (Pre-cleaning) ..................................... 8
4.2 Containment and transportation .................................. 8
4.3 Manual Cleaning and disinfection ................................ 8
4.4 Drying ......................................................... 11
4.5 Inspection ..................................................... 11
4.6 Packaging ...................................................... 11
4.7 Sterilization .................................................. 12
4.8 Storage ........................................................ 14
5. Maintenance and Safety Inspection .......................... 14
6. Safety Precautions ......................................... 15
7. Specifications ............................................. 16
7.1 Probe .......................................................... 16
7.2 Suppliers List ................................................. 17
8. Disposal of the probe ...................................... 18
( 5 ) Q1E-EP1152
1. Introduction
1.1 Features
The probe of model EUP-S70 is a phased array sector scanning type.
The acoustic output of this probe when connected to ultrasound
scanner was measured according to the IEC 61157 standard. The table
of measured acoustic output data is contained in the operation manual
of each ultrasound scanner.
This probe is categorized in class IIa according to Directive
93/42/EEC.
According to IEC 60601-1 the probe is classified as type BF.
- 1 - Q1E-EP1152
1.4 Composition
The probe components of the EUP-S70 are as follows:
1) Probe EUP-S70 ···················· 1 piece
2) Instruction Manual ··············· 1 copy
1.5 Construction
Immersible part (IPX7)
Head Cable
Applied part
Connector
Un-immersible
Immersible part:
This part can be immersed in disinfectant
solution and also can be cleaned by water.
Un-immersible part:
This part should not be immersed in
disinfectant solution and also can not be
cleaned by water.
- 2 - Q1E-EP1152
2. Inspection before Use
Prior to use, the probe must be carefully inspected that it is
appropriate for use. If not, do not use the probe and immediately
contact a service support.
- 3 - Q1E-EP1152
3. Operation Procedure
1) Confirm that the probe is cleaned, disinfected and/or sterilized.
2) Connect the probe, operate the main unit, and adjust the image
according to the instructions given in the operation manual for
the main unit.
3) Relationship between direction of the probe and the image is shown
in Fig. 2. The right-left orientation mark on the image indicates
the direction of the index mark on the probe.
Index mark
Right-left
orientation mark
- 4 - Q1E-EP1152
4. Cleaning, Disinfection and Sterilization
The probe must be reprocessed after each use. Refer to the reprocessing
instruction in this chapter.
- 5 - Q1E-EP1152
Levels of reprocessing requirements:
Depending on the application of the product and with regard to risk
evaluation, the user has to classify the medical device according to
the current Medical Device Directive for processing of medical devices
as uncritical, semi-critical or critical. Supporting information
concerning this topic is listed in the table below. The user is responsible
for correct classification of the medical device.
Cleaning
Application part contacts Disinfection
intracorporeal tissue
critical (Disinfectant with
directly virucidal effect - minimum)
(operative application)
Sterilization
- 6 - Q1E-EP1152
The flowchart of the reprocessing process of this probe is as follows.
Point of use
(Pre-cleaning)
Containment and
transportation
Manual Disinfection
Drying
Packing
Sterilization
- 7 - Q1E-EP1152
4.1 Point of use (Pre-cleaning)
- 8 - Q1E-EP1152
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
1) The temperature of the detergent solution should be between 15-30°C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer’s instruction. If a differing detergent
is used, please also note the approved material compatibility for
the medical device.
2) Immerse the immersible part of the probe without connector into the
diluted detergent solution (see Fig. 4). Wipe the immersible part
of the probe under the surface of the detergent solution with a soft
cloth to remove all visible soil. Be sure that all grooves of the
probe are implemented during the cleaning process.
3) The immersible part of the probe should be left in the detergent solution
according to the specified contact time of the detergent manufacturer.
4) Wipe the un-immersible parts of the probe with a soft cloth dipped
with the detergent solution.
5) Rinse the probe with running tap water for 1 minute. (alternatively:
immerse the immersible part of the probe in a tray filled with deionized
water/tap water (see Fig. 4) for 5 min.)
6) Visually check the outer surface of the probe for cleanness. If necessary,
use magnifying glass for visually check. If there is still soil visible,
repeat all above steps.
- 9 - Q1E-EP1152
Manual disinfection:
1) Prepare the disinfectant solution in a tank with cold water (please
follow the instructions of the disinfectant manufacturer regarding
application, concentration, microbiological efficiency, service life
and contact time).
2) Confirm the concentration of the disinfectant before immersing the
probe. Although Cidex® OPA does not need to be diluted, it is recommended
to use test strips to verify the concentration. The test strips can
indicate whether or not the concentration is above the Minimum
Effective Concentration (MEC). Please also note the expiration date
of the test stripes. Temperature of disinfectant solution should be
minimum 20°C. The minimum contact time is 5 minutes. If a different
disinfectant is used, follow the manufacturer’s instructions. Please
also consider the material compatibility for the medical device.
3) Immerse the immersible part of the probe into the disinfectant (see
Fig. 4). Set a clock to insure the recommended contact time which
is 5 minutes.
4) Rinse the immersible part of the probe with deionized water for 1
minute. (alternatively: immerse the immersible part of the probe in
a tray filled with deionized water (see Fig. 4) for 5 min.)
5) Visually check the outer surface of the probe for leavings of the
disinfectant. If necessary, repeat the rinsing.
Water
Detergent
Disinfectant
- 10 - Q1E-EP1152
4.4 Drying
1) Wipe the probe with a single-use, fluff-free wipe or towel to
remove moisture from the surface of the probe.
2) Dry the probe naturally in an ambient temperature between 15-30°C
for a minimum of 4 hours. Alternatively the equipment can be
dried using a drying heater at a temperature of less than 60°C.
4.5 Inspection
Inspect the equipment for any damage such as crack, scratch or
deformation. Do not use it if any damage is found.
- 11 - Q1E-EP1152
4.7 Sterilization
The probe can be sterilized using either ethylen oxide gas (EtO)
sterilization or plasma sterilization (see table below).
Follow the manufacturer's instructions of the sterilizer regarding
usage, temperature and sterilization-time.
The sterilization method and operating conditions are as follows.
WARNING
1) Before performing sterilization, check that the operation data of
sterilizer are in conjunction with min. and max. data applicable
for the probe.
- 12 - Q1E-EP1152
The packaging before sterilization is as follows.
1) Put the probe into TYVEK
pouch.
Probe
TYVEK Pouch
Fig. 4 Packaging in the pouch
2) Seal the TYVEK Pouch using
a heat sealer. Ensure that Sealed
the seal is complete.
Probe
TYVEK Pouch
Fig. 5 Sealing
3) Put the sealed pouch into a
tray or plastic mesh wire for
sterilization.
Tray for
sterilization
Probe in
the Pouch
- 13 - Q1E-EP1152
4.8 Storage
- 14 - Q1E-EP1152
6. Safety Precautions
WARNING
・ Never use the probe if the probe head, housing or cable are
cracked or damaged.
・ Warning is case of using probe covers which latex is contained
to. The latex may cause such allergic reactions as itching,
rubor, urticaria, swelling, fever, anhelation, wheezing,
depression of blood pressure, shock and so on. For the patients
suspected of latex allergy, do not use the latex-containing
medical devices. If you observe any of above mentioned
symptoms in your patient during the operation, stop the use
of the latex-containing medical devices immediately and take
an appropriate treatment to the patient.
CAUTION
・ By examination of an early pregnancy the exposure time shall
be as short as possible. Start examination with acoustic output
power set to L (Low).
・ The probe connector is not water proof. Do not allow liquid
to contact the connector.
・ Do not drop, hit or bent the probe.
・ Use only water, detergents and disinfectants in the suppliers
list. Between use store the probe holder of scanner.
・ Under sterile condition use appropriate protection for probe
and cable. Some Latex material may create allergic reactions.
・ The probe is not delivered disinfected or sterilized. Before
using the probe, disinfect or sterilize it.
・ The acoustic lens of the probe is manufactured very thin and
delicate to get the high resolution. Therefore, in case of
wiping off the ultrasound jelly or cleaning the surface of
the acoustic lens, please use the soft cloth or tissue paper
and handle with care.
- 15 - Q1E-EP1152
7. Specifications
7.1 Probe
Type : EUP-S70 Sector Scanning probe
Acoustic working
frequency : 3.0MHz
Technology : Phased Array Probe
Dimensions : See Fig. 7.
Weight : Approx. 0.7kg
(incl. cable and connector)
Scanning angle : 90°
Probe materials : Bio-compatible allergy free components
Acoustic output : According to IEC 61157
(See Main Unit manual.)
Applicable system : Depending on production and upgrade
status
For detailed information contact a
service support.
Classification : MDD classification IIa.
Cleaning : Applicable detergents are listed in
the suppliers list
Disinfection : Applicable disinfectants are listed
in the suppliers list
Sterilization : Plasma sterilization
Operating conditions :
Ambient temperature ; 25 – 35°C
Contact surface temperature ; max. 42°C
(temperature of examinee)
Relative humidity ; 30 – 85%
(Subject to no condensation)
Storage conditions :
Temperature ; –10 – +55°C
Relative humidity ; 10 – 95%
(subject to no condensation)
- 16 - Q1E-EP1152
7.2 Suppliers List
The products listed below are seriously tested and approved for
use with the Sector Scanning Probe EUP-S70.
- 17 - Q1E-EP1152
8. Disposal of the probe
Recycle or dispose the equipment properly in compliance with your
organizational rules and your local Law.
CAUTION
Before disposing the equipment, disinfect or take other
infection-prevention measures.
Disposal of the equipment without taking the proper preventative measures
can lead to infection.
- 18 - Q1E-EP1152
Unit: mm
Fig. 7 Dimensions
E
-19- Q1E-EP1152