Elecsys BRAHMS PCT: Procalcitonin
Elecsys BRAHMS PCT: Procalcitonin
Elecsys 2010
MODULAR ANALYTICS E170
05056888 003 100 cobas e 411
cobas e 601
cobas e 602
Note: Do not combine bottles from different lots. Use only control bottles out Materials required (but not provided)
of one lot with each other. ▪ 11776576322, CalSet Vials, 2 x 56 empty snap‑cap bottles
All information required for correct operation is read in from the respective
reagent barcodes. ▪ 03142949122, ControlSet Vials, 2 x 56 empty snap‑cap bottles
Please note: Both the vial labels, and the additional labels (if available) ▪ General laboratory equipment
contain 2 different barcodes. The barcode between the yellow markers is ▪ Elecsys 2010, MODULAR ANALYTICS E170 or cobas e analyzer
for cobas 8000 systems only. If using a cobas 8000 system, please turn Accessories for Elecsys 2010 and cobas e 411 analyzers:
the vial cap 180° into the correct position so the barcode can be read by the
system. Place the vial on the instrument as usual. ▪ 11662988122, ProCell, 6 x 380 mL system buffer
Storage and stability ▪ 11662970122, CleanCell, 6 x 380 mL measuring cell cleaning
Store at 2‑8 °C. solution
Do not freeze. ▪ 11930346122, Elecsys SysWash, 1 x 500 mL washwater additive
Store the Elecsys reagent kit upright in order to ensure complete ▪ 11933159001, Adapter for SysClean
availability of the microparticles during automatic mixing prior to use. ▪ 11706802001, Elecsys 2010 AssayCup, 60 x 60 reaction vessels
Stability of the reagent rackpack ▪ 11706799001, Elecsys 2010 AssayTip, 30 x 120 pipette tips
Accessories for MODULAR ANALYTICS E170, cobas e 601 and
unopened at 2-8 °C up to the stated expiration date cobas e 602 analyzers:
after opening at 2-8 °C 12 weeks ▪ 04880340190, ProCell M, 2 x 2 L system buffer
on the analyzers 4 weeks ▪ 04880293190, CleanCell M, 2 x 2 L measuring cell cleaning
solution
Stability of the calibrators and controls
▪ 03023141001, PC/CC‑Cups, 12 cups to prewarm ProCell M and
lyophilized calibrators/controls up to the stated expiration date CleanCell M before use
reconstituted calibrators/controls on 2 hours (use only once) ▪ 03005712190, ProbeWash M, 12 x 70 mL cleaning solution for run
the analyzers finalization and rinsing during reagent change
reconstituted calibrators/controls at 3 months (freeze only once) ▪ 12102137001, AssayTip/AssayCup Combimagazine M,
48 magazines x 84 reaction vessels or pipette tips, waste bags
-20 °C
▪ 03023150001, WasteLiner, waste bags
Store the calibrators and controls upright in order to prevent the solution
from adhering to the snap‑cap. ▪ 03027651001, SysClean Adapter M
Accessories for all analyzers:
Specimen collection and preparation
Only the specimens listed below were tested in a sufficient number and ▪ 11298500316, Elecsys SysClean, 5 x 100 mL system cleaning
found acceptable. solution
Serum collected using standard sampling tubes or tubes containing Assay
separating gel. For optimum performance of the assay follow the directions given in this
Li‑heparin and K3‑EDTA plasma. document for the analyzer concerned. Refer to the appropriate operator’s
manual for analyzer‑specific assay instructions.
Criterion: Slope 0.9-1.1 + intercept within < ± 2 x analytical sensitivity (LDL)
+ coefficient of correlation > 0.95. Resuspension of the microparticles takes place automatically prior to use.
Read in the test‑specific parameters via the reagent barcode. If in
Stable for 24 hours at 2‑8 °C, 3 months at -20 °C. Freeze only once. exceptional cases the barcode cannot be read, enter the 15‑digit sequence
After drawing the blood, measure samples within 24 hours or freeze at of numbers.
-20 °C. Bring the cooled reagents to approximately 20 °C and place on the reagent
Frozen samples can lead to a lower recovery of up to 8 %. disk (20 °C) of the analyzer. Avoid foam formation. The system
The sample types listed were tested with a selection of sample collection automatically regulates the temperature of the reagents and the
tubes that were commercially available at the time of testing, i.e. not all opening/closing of the bottles.
available tubes of all manufacturers were tested. Sample collection systems Place the reconstituted calibrators (in the system-compatible bottles with
from various manufacturers may contain differing materials which could barcoded labels) in the sample zone.
affect the test results in some cases. When processing samples in primary All the information necessary for calibrating the assay is automatically read
tubes (sample collection systems), follow the instructions of the tube into the analyzer.
manufacturer.
After calibration has been performed, discard the calibrators.
Centrifuge samples containing precipitates before performing the assay.
Analyze the controls PC PCT1 and PC PCT2. The information on the
Do not use samples and controls stabilized with azide. barcoded label of the control serum bottle is read in automatically. After the
Ensure the samples, calibrators and controls are at 20‑25 °C prior to control procedure has been performed, discard the controls.
measurement.
Calibration
Due to possible evaporation effects, samples, calibrators and controls on Traceability: This method has been standardized against the BRAHMS PCT
the analyzers should be analyzed/measured within 2 hours. LIA assay.
Materials provided Every Elecsys BRAHMS PCT reagent set has a barcoded label containing
See “Reagents – working solutions” section for reagents. specific information for calibration of the particular reagent lot. The
predefined master curve is adapted to the analyzer using PCT Cal1 and
▪ 2 barcode cards PCT Cal2.
▪ control barcode sheet Calibrator sequence on all systems: Always measure PCT Cal2 before
▪ 2 x 8 bottle labels (calibrators) PCT Cal1.
▪ 2 x 14 bottle labels (controls) Calibration frequency: Calibration must be performed once per reagent lot
using fresh reagent (i.e. not more than 24 hours since the reagent kit was
▪ 6 empty labeled snap-cap bottles registered on the analyzer). Renewed calibration is recommended as
follows:
3 Müller B, White JC, Nylén ES, et al. Ubiquitous Expression of the This product may not be used by purchaser to conduct any Point-of-Care
Calcitonin-I Gene in Multiple Tissues in Response to Sepsis. J Clin testing including but not limited to any near patient testing on hospital wards
Endocrinol Metab 2001;86(1):396-404. and/or emergency departments and/or in doctor’s offices and/or in any
4 Weglöhner W, Struck J, Fischer-Schulz C, et al. Isolation and other locations outside private or public clinical laboratories. No general
characterization of serum procalcitonin from patients with sepsis. patent or other license of any kind other than this specific right of use from
Peptides 2001;22:2099-2103. purchase is granted hereby.
For further information, please refer to the appropriate operator’s manual for
5 Gaïni S, Koldkjær OG, Mǿller HJ, et al. A comparison of high-mobility the analyzer concerned, the respective application sheets, the product
group-box 1 protein, lipopolysaccharide-binding protein and information and the Method Sheets of all necessary components (if
procalcitonin in severe community-aquired infections and bacteraemia: available in your country).
a prospective study. Crit Care 2007;11(4):77-87.
A point (period/stop) is always used in this Method Sheet as the decimal
6 Castelli GP, Pognani C, Cita M, et al. Procalcitonin, C-reactive protein, separator to mark the border between the integral and the fractional parts of
white blood cells and SOFA score in ICU: diagnosis and monitoring of a decimal numeral. Separators for thousands are not used.
sepsis. Minerva Anestesiol 2006;72:69-80.
Reagent developed in collaboration with B·R·A·H·M·S.
7 Gaïni S, Koldkjær OG, Pedersen C, et al. Procalcitonin,
lipopolysaccharide-binding protein, interleukin-6, and C-reactive protein B·R·A·H·M·S PCT is a registered trademark of BRAHMS Aktiengesellschaft.
in community-acquired infections and sepsis: a prospective study. Crit
Care 2006;10(2):53-63.
8 Clec’h C, Ferriere F, Karoubi P, et al. Diagnostic and prognostic value
of procalcitonin in patients with septic shock. Crit Care Med
2004;32(5):1166-1169. Symbols
9 Rey C, Los Arcos M, Concha A, et al. Procalcitonin and C-reactive Roche Diagnostics uses the following symbols and signs in addition to
protein as markers of systemic inflammatory response syndrome in those listed in the ISO 15223‑1 standard.
critically ill children. Intensive Care Med 2007;33:477-484.
10 Andreola B, Bressan S, Callegaro S, et al. Procalcitonin and C- Contents of kit
Reactive Protein as Diagnsotic Markers of Severe Bacterial Infections Analyzers/Instruments on which reagents can be used
in Febrile Infants and Children in the Emergency Department. Pediatr
Infect Dis J 2007;26(8):672-677. Reagent
11 Novotny A, Emmanuel K, Matevossian E, et al. Use of procalcitonin for Calibrator
early prediction of lethal outcome of postoperative sepsis. The
American Journal of Surgery 2007;194:35-39. Volume after reconstitution or mixing
12 Hausfater P, Juillien G, Madonna-Py B, et al. Serum procalcitonin COBAS, COBAS E, ELECSYS, MODULAR and PRECICONTROL are trademarks of Roche. INTRALIPID is a
measurement as diagnostic and prognostic marker in febrile adult trademark of Fresenius Kabi AB.
patients presenting to the emergency department. Crit Care All other product names and trademarks are the property of their respective owners.
2007;11(3):60-69. Significant additions or changes are indicated by a change bar in the margin.
13 Dahaba AA, Hagara B, Fall A, et al. Procalcitonin for early prediction of © 2014, Roche Diagnostics
survival outcome in postoperative critically ill patients with severe
sepsis. Br J Anaesth 2006;97:503-508.
14 Rau B, Schilling MK, Beger HG. Laboratory Markers of Severe Acute
Pancreatitis. Dig Dis 2004;22:247-257. Roche Diagnostics GmbH, Sandhofer Strasse 116, D-68305 Mannheim
www.roche.com
15 Sato N, Endo S, Kasai T, et al. Relationship of the serum procalcitonin
level with the severity of acute pancreatitis. Research Communications
in Molecular Pathology and Pharmacology 2004;115,116:243-249.
16 Stolz D, Christ-Crain M, Gencay MM, et al. Diagnostic value of signs,
symptoms and laboratory values in lower respiratory tract infection.
Swiss Med Wkly 2006;136:434-440.
17 Christ-Crain M, Müller B. Biomarkers in respiratory tract infections:
diagnostic guides to antibiotic prescription, prognostic markers and
mediators. Eur Respir J 2007;30:556-573.
18 Occupational Safety and Health Standards: bloodborne pathogens.
(29 CFR Part 1910.1030). Fed. Register.
19 Directive 2000/54/EC of the European Parliament and Council of
18 September 2000 on the protection of workers from risks related to
exposure to biological agents at work.
20 Meisner M. Procalcitonin (PCT) – A new innovative infection parameter.
Biochemical and clinical aspects. Thieme Stuttgart, New York 2000,
ISBN: 3-13-105503-0.
21 Chiesa C, Panero A, Rossi N, et al. Reliability of Procalcitonin
Concentrations for the Diagnosis of Sepsis in Critically ill neonates. Clin
Infect Dis 1998;26:664-672.
22 American College of Chest Physicians/Society of Critical Care Medicine
Definitions for sepsis and organ failure and guidelines for the use of
innovative therapies in sepsis. Crit Care Med 1992;20:864-874.
23 Bablok W, Passing H, Bender R, et al. A general regression procedure
for method transformation. Application of linear regression procedures
for method comparison studies in clinical chemistry, Part III.
J Clin Chem Clin Biochem 1988 Nov;26(11):783-790.