AccuVueMED Manual ENG
AccuVueMED Manual ENG
Released By 2019-03-19
About This Manual
The AccuVueMED User Manual provides instructions for the configuration and use of
AccuVueMED radiological software, developed by RadmediX.
Please contact us if there are any problems during the use of the product.
WebSite http://www.radmedix.com
Customer Email [email protected]
Support Team Tel +1-844-723-6334 ex. 2
Fax +1-513-278-0150
AccuVueMED_UM_COM_EN_01
About This Manual
Copyright Manual
Information contained in this manual is confidential and is the property of RADMEDIX Inc.,
and is provided exclusively for authorized licensees of RADMEDIX Inc.
The information contained in this manual may not be altered or distributed without prior
written consent of RADMEDIX Inc., and may not be disclosed to unauthorized personnel.
AccuVueMED_UM_COM_EN_01
Table of Contents
1 INTRODUCTION ................................................................................................................. 6
2 SYSTEM REQUIREMENTS .................................................................................................... 7
2.1 SUPPORTED DIGITAL X-RAY DETECTOR ..................................................................................................... 7
2.2 SYSTEM REQUIREMENTS FOR ACCUVUEMED ........................................................................................... 8
3 OVERVIEW .......................................................................................................................... 9
3.1 PROGRAM START-UP ................................................................................................................................. 9
3.2 MAIN SCREEN OF PROGRAM...................................................................................................................... 11
3.2.1 Function Tabs .................................................................................................................................... 12
3.2.2 Status Display .................................................................................................................................... 13
4 LIST .................................................................................................................................... 16
4.1 WORK LIST .............................................................................................................................................. 17
4.1.1 Main Screen of Worklist .................................................................................................................... 17
4.1.2 Add Examination Window ............................................................................................................... 22
4.1.3 Edit Projection Window ....................................................................................................................25
4.1.4 Edit Window ..................................................................................................................................... 27
4.1.5 Excel Import Window ....................................................................................................................... 29
4.2 STUDYLIST ............................................................................................................................................... 30
4.2.1 Main Screen of the Studylist ............................................................................................................ 30
4.2.2 Merge Window ...................................................................................................................................35
4.2.3 Export Window ................................................................................................................................. 37
4.2.4 Edit Window ..................................................................................................................................... 44
4.2.5 Report Window ................................................................................................................................. 46
4.2.6 Analysis ............................................................................................................................................. 53
4.2.7 Bodypart Editor ................................................................................................................................ 55
5 EXAM .................................................................................................................................. 56
5.1 TOOLBAR ................................................................................................................................................ 59
5.1.1 Toolbar Overview .............................................................................................................................. 59
5.1.2 Edit Toolbar ....................................................................................................................................... 60
5.1.3 Toolbar Icons .................................................................................................................................... 62
5.2 PROJECTION LIST & IMAGE PROCESSING TAB........................................................................................... 66
5.2.1 Projection List ................................................................................................................................... 67
5.2.2 Image Processing Tab ....................................................................................................................... 69
5.3 TRUVIEW ART ......................................................................................................................................... 71
6 CONFIGURATION .............................................................................................................. 73
6.1 DISPLAY .................................................................................................................................................. 75
6.2 SYSTEM ................................................................................................................................................... 75
6.2.1 Language ........................................................................................................................................... 76
6.2.2 Format ............................................................................................................................................... 77
6.2.3 Institution ......................................................................................................................................... 79
6.2.4 Options .............................................................................................................................................. 80
6.2.5 Workflow ............................................................................................................................................ 81
6.3 IMAGE ..................................................................................................................................................... 85
6.3.1 Image Folder ..................................................................................................................................... 85
6.3.2 Delete Option .................................................................................................................................... 86
6.3.3 Overlay .............................................................................................................................................. 88
6.3.4 Import................................................................................................................................................. 91
6.3.5 Truview ART ...................................................................................................................................... 92
6.4 STORAGE ................................................................................................................................................. 93
6.4.1 Delete................................................................................................................................................. 93
AccuVueMED_UM_COM_EN_01
6.4.2 Backup ............................................................................................................................................... 95
6.5 ACCOUNT ................................................................................................................................................ 97
6.6 BLOCK LIST ............................................................................................................................................. 98
6.7 DATASET ................................................................................................................................................. 99
6.8 NETWORK ..............................................................................................................................................100
6.8.1 Local ..................................................................................................................................................101
6.8.2 MWL................................................................................................................................................. 102
6.8.3 PACS ................................................................................................................................................. 104
6.8.4 Printer .............................................................................................................................................. 105
6.8.5 MPPS ................................................................................................................................................ 108
6.9 DETECTOR .............................................................................................................................................109
6.9.1 General ..............................................................................................................................................110
6.9.2 Detailed Configuration by Panel ..................................................................................................... 111
6.10 RIS CODE................................................................................................................................................ 113
6.10.1 Edit RIS Code Window ..................................................................................................................... 114
6.10.2 Add Marker Window ........................................................................................................................118
6.11 PROJECTION EDITOR ...............................................................................................................................119
6.11.1 Projection Editor Window ............................................................................................................... 120
6.12 RECYCLE BIN .......................................................................................................................................... 124
APPENDIX A. GLOSSARY .......................................................................................................................... 126
APPENDIX B. UPDATE .............................................................................................................................. 129
APPENDIX C. STITCHING .......................................................................................................................... 132
APPENDIX D. PORTABLE FUNCTION .......................................................................................................136
APPENDIX E. HOW TO USE ROTATION TOOLS ....................................................................................... 141
APPENDIX F. GAIN CALIBRATION ........................................................................................................... 144
AccuVueMED_UM_COM_EN_01
1 Introduction
AccuVueMED is a complete digital image processing console software specialized for the
digital X-ray detector FLAATZ and EVS series developed by DRTECH Co, Ltd.
AccuVueMED provides the best image processing solution for fine tuned quality images
by integrating with the X-ray generator and the digital detector. AccuVueMED not only
processes the acquired images but also complies with DICOM standards which allow the
user to transmit and receive data with the PACS system and print images through the
DICOM printer.
AccuVueMED_UM_COM_EN_01 6 / 149
2 System Requirements
FLAATZ 330N
FLAATZ 500
FLAATZ 560
FLAATZ 760
EVS 4343 Series
EVS 4343 A Series
AccuVueMED_UM_COM_EN_01 7 / 149
2.2 System Requirements for AccuVueMED
The table provides the minimum recommended specifications for the operating system
that runs AccuVueMED. All specifications are subject to change without notice.
Items Recommended
CPU Intel i5-3470
RAM 8 GB
HDD 500 GB
NVIDIA GeForce
VGA
GT630 1GB
ODD DVD Recorder
Windows 7 Pro
OS 32 / 64bit
Win8, Win10
Display Size 23 inch
Display Resolution 1920 x 1080
* Optimal performance is not guaranteed for PC that does not comply with the recommended specifications.
AccuVueMED_UM_COM_EN_01 8 / 149
3 Overview
In order to run the program, double click the [AccuVueMED] icon on the desktop or select
‘AccuVueMED’ from ‘Windows Menu’ – ‘All Programs’ – ‘AccuVueMED’.
A login window as seen in [Figure 4] will appear if the program is properly initiated.
AccuVueMED_UM_COM_EN_01 9 / 149
If you enter an invalid ID or Password, an ‘Error Window’ will appear as shown in [Figure
5] and you will not be able to log in.
AccuVueMED_UM_COM_EN_01 10 / 149
3.2 Main Screen of Program
You will see the AccuVueMED main screen as in [Figure 6] after logging in.
a) Function Tabs
b) Status Display
AccuVueMED_UM_COM_EN_01 11 / 149
3.2.1 Function Tabs
Two tab icons can be found on the upper left corner in the main screen as in [Figure 7]
AccuVueMED_UM_COM_EN_01 12 / 149
3.2.2 Status Display
Status display bar can be found on the bottom in the main screen as in [Figure 8].
Display the status of hardware and the connection status between equipment.
[FLAATZ-600]
- AED mode
- Switch box mode
Click this icon to change the mode. You can change the
default setting using ECali1. (ECali1 - Configuration -
Detector X - Exposure mode)
AccuVueMED_UM_COM_EN_01 13 / 149
Display the connection status and the available space left on
the currently connected HDD.
The displayed number indicates the remaining HDD
capacity.
Display the connection status (On/Off) with the X-ray
detector.
[EVS Series]
Display the temperature.
Display the connection status with the Storage Server.
When you click this icon, a pop-up window will appear. You
can change the PACS settings or conduct a test in the new
window. The UI is very similar to the PACS menu of
Configuration. .
Refer to 6.8.3 PACS for more details.
AccuVueMED_UM_COM_EN_01 14 / 149
② Current Time & Log-in Information Display
③ Configuration
AccuVueMED_UM_COM_EN_01 15 / 149
4 List
The List tab displays the lists of patients and studies, which can be divided into two
categories as described below.
Please Note!
The information on the Worklist and Studylist is stored on the Local Storage
(HDD) as default settings.
Every time AccuVueMED runs, a new Worklist is retrieved from the Worklist
server and stored on the Local Storage.
Only the contents on the list which are edited and deleted in the AccuVueMED
Vet affect the Local Storage as default. However, edited records may be
transmitted to the PACS server using the ‘Send’ command.
AccuVueMED_UM_COM_EN_01 16 / 149
4.1Work List
Display the list of patients. The list is displayed by clicking the [Work List] icon on the
upper left side of the List tab.
AccuVueMED_UM_COM_EN_01 17 / 149
① List of Patients
Type Description
ID Patient ID
Name Patient name
Study Desc. Study Description
Accession Number.
Acc. No.
X-ray study registration number
Date/Time Time of Registration
D.O.B. Date of Birth
Sex Sex of the patient
Order Date Time of order
Operator Radiographer Name
Exam requesting physician’s
Requesting Physician
name
Projection can be performed in the Exam tab by double clicking on the selected
patient.
You can also click [Exam] icon after selecting a specific patient to perform
projection. Refer to 5. Exam for more details.
AccuVueMED_UM_COM_EN_01 18 / 149
② Search Bar & Refresh Icon
Search Bar
Search the entered value such as Patient, ID, Name, and Access Number in the
Worklist (when ‘Modality Query Base’ is selected in Configuration options)
Search the entered value in the Worklist server that has selected the entered
value as Base. (when ‘Patient Query Base’ is selected in Configuration options)
If you click the [Search] icon after entering the value to be searched, only
the relevant works will be displayed on the list.
AccuVueMED_UM_COM_EN_01 19 / 149
Advanced Search Window
Refresh icon
Retrieve recently registered patients from the PACS server.
AccuVueMED_UM_COM_EN_01 20 / 149
③ Add / Edit / Delete Icon
Icon Description
④ Exam
AccuVueMED_UM_COM_EN_01 21 / 149
4.1.2 Add Examination Window
AccuVueMED_UM_COM_EN_01 22 / 149
① Patient Information
Category Description
ID Patient ID
Accession Number Registered identification number
Name Patient Name
Operator Radiographer Name
Gender Gender of the patient
Birth Date Date of birth of the patient
Age Age of the patient (Year / Month selection available)
RIS Code RIS Code to be used for examination
Requesting Physician Exam requesting physician’s name
② Emergency
When clicking this icon, the default values are automatically entered as ID and
Accession Number and it switches to the Exam tab.
③ Projection Information
Save the patient with the entered information and close the window.
⑥ Exam
AccuVueMED_UM_COM_EN_01 23 / 149
Save the patient with the entered information and directly switch to the Exam tab for
examination.
Refer to 5. Exam for more details.
⑦ Cancel
* Please Note!
If multiple projections are registered for a single patient, each projection receives a
different Accession Number [Figure 12].
Click the [OK] icon to register each projection for two separate studies in the Worklist
as seen in [Figure 13]
AccuVueMED_UM_COM_EN_01 24 / 149
4.1.3 Edit Projection Window
The ‘Edit Projection Window’ will appear when clicking the [Edit] icon.
① Anatomy Information
AccuVueMED_UM_COM_EN_01 25 / 149
② Projection & Position List
③ Add
Register the currently set Projection, Position, and Receptor Type information.
The currently set Projection, Position and Receptor Type are also registered
when clicking the selected Position.
Up Down
⑥ Delete
⑦ OK
⑧ Cancel
AccuVueMED_UM_COM_EN_01 26 / 149
4.1.4 Edit Window
The ‘Edit Window’ will appear when selecting a study from the Worklist and then clicking
the [Edit] icon.
Note: The ‘Edit Window’ is not activated if the study information is retrieved from the
PACS server.
AccuVueMED_UM_COM_EN_01 27 / 149
① Study Information
Category Detail
ID Patient ID
Name Patient Name
Accession Number Registered Identification Number
Operator Radiographer Name
Study Description Study Description
Sex Sex of the Patient
Birth Date Date of Birth of the Patient
Age Age of the Patient (Year/Month Selection Available)
Schedule date & time Scheduled date and time of imaging
Requesting Physician Exam requesting physician’s name
② OK
③ Cancel
AccuVueMED_UM_COM_EN_01 28 / 149
4.1.5 Excel Import Window
② Limitation
You can select the patient’s sex among the 'M', 'F', and 'U'.
The format of Birth Date and Study Date is 'MM/DD/YYYY'. Any other format is not
provided.
The format of Birth Time and Study Time is 'HH:MM:SS'. Any other format is not
provided.
③ Excel import
AccuVueMED_UM_COM_EN_01 29 / 149
4.2 Studylist
Display the examined study list. When clicking the [Study List] icon on the upper side of
the screen, the study list will be activated.
AccuVueMED_UM_COM_EN_01 30 / 149
Height of the list is adjustable using [Ctrl key] + Mouse wheel or using [Ctrl key] +
[Arrow key].
Types of information are as noted in [Table 6].
Category Description
ID Patient ID
Patient Selection
Patient can be selected by left-clicking. .
The selected patient is highlighted in blue. In the ‘① Examined Patient List’ and
‘② Examined Study List per Patient’, an activated study is highlighted in darker
color, while an inactivated study is highlighted in lighter color.
The selected patients can be edited or deleted.
Display detailed study information of the selected patient on ‘① List of the Examined
Patients’.
Types of information are as noted in [Table 7].
AccuVueMED_UM_COM_EN_01 31 / 149
Table 7. Information Types of Examined Study List
Items Description
Acc. No. Accession Number. Registered Identification Number
Study ID RIS code
Study Desc. Study Description
Status [Number of images] / [Number of projections]
Date/Time Date / Time for examined study
PACS [Number of images transmitted] / [Number of projections]
Printer [Number of images printed] / [Number of projections]
Report Report documentation
Operator Radiographer Name
Requesting Physician Exam requesting physician’s name
③ Thumbnail
Search Bar
Search the values of Patient ID, Name, and Access Number within the Studylist.
If you click the search icon after entering the value to be searched, only
the relevant works will be displayed on the list.
Refresh Icon
AccuVueMED_UM_COM_EN_01 32 / 149
⑤ Tool Icons
AccuVueMED_UM_COM_EN_01 33 / 149
⑥ Send
AccuVueMED_UM_COM_EN_01 34 / 149
4.2.2 Merge Window
Study lists appear on each side. (One page only shows 100 items.)
Required items can be easily searched by using the search tool.
Identical items selected in ① will be also located in ② if the same list is checked.
You can view the list of taken jobs, detailed list of shooting, and thumbnails at a
glance.
If ‘Show worklist’ is selected, patient list information is displayed in ①. You can merge
the images by selecting the patient on the patient list.
AccuVueMED_UM_COM_EN_01 35 / 149
③ Merge Icon
④ Close
AccuVueMED_UM_COM_EN_01 36 / 149
4.2.3 Export Window
Save the images of the selected study from the Studylist in a USB, CD, DVD, or PDF and
print it through the DICOM Printer.
4.2.3.1 Drive
AccuVueMED_UM_COM_EN_01 37 / 149
4.2.3.2 Print
① Print Setting
Printer Selection
Select a DICOM Printer.
Register the Printer using 'Config' - 'Network' - 'Printer' menu.
Film Orientation
Select a film orientation between portrait / landscape.
Film Size
Select the size of the film.
Layout
Set the print layout.
AccuVueMED_UM_COM_EN_01 38 / 149
The [Divide] icon must be clicked in order to apply any changes made in the
Layout.
Zoom
Perform zoom in/out of the image.
② Preview
Thumbnail
The image to be printed appears in the thumbnail list on the right.
Preview
If an image is selected in the thumbnail, a large image appears in the center as a
preview screen.
Merge
Merge the divided screen which has been separated by the ‘Divide’ function.
Only adjacent screens of the same sizes can be merged
The merged shape must be a square.
Divide
Divide the screen by the value set in ‘Layout’.
Rotation
Rotate the screen 90 degrees clock-wise.
Flip
Reverse the up and down sides.
Print
Print out using general printer.
100%
Perform paper printing in the actual object size.
Fit to Screen
Perform paper printing according to the screen size.
③ DICOM Print
AccuVueMED_UM_COM_EN_01 39 / 149
4.2.3.3 CD / DVD
AccuVueMED_UM_COM_EN_01 40 / 149
Eject
Eject the tray of the CD / DVD Writer.
③ Export Process
Estimated Time
Present the expected time needed for Export.
Time Left
Display the estimated time left to the completion of Export.
④ Task Icons
Burn
Perform the Export task.
Cancel
Cancel the Export task.
AccuVueMED_UM_COM_EN_01 41 / 149
4.2.3.4 E-Mail
① Address Book
Select E-mail address and double-click to populate the “To:” field automatically.
User can add or edit or delete addresses using add, edit, delete button.
② To
③ Subject
④ Body
AccuVueMED_UM_COM_EN_01 42 / 149
⑤ Account
⑥ Sending status
⑧ Send or Cancel
AccuVueMED_UM_COM_EN_01 43 / 149
4.2.4 Edit Window
Only the patient information (ID, Name, Sex, Birth Date, and Age) can be edited when
clicking the [Edit] icon in the patient list of the Studylist (upper part of the list).
OK
Save and close the window.
Cancel
Cancel and close the window.
AccuVueMED_UM_COM_EN_01 44 / 149
4.2.4.2 Edit Study Information
AccuVueMED_UM_COM_EN_01 45 / 149
4.2.5 Report Window
4.2.5.1 Report
Create a report using the Report button in the StudyList.
The User can attach up to 4 images using the 'Image load' button.
AccuVueMED_UM_COM_EN_01 46 / 149
③ Write opinion
Layout button
Select 1X1, 1X2, 2X2 layout.
AccuVueMED_UM_COM_EN_01 47 / 149
Image load button
Select an image
OK / Cancel button
Attach the selected image to the report / Cancel and close image load window.
4.2.5.3 Select the image
You can select the image when clocking "Image load" button
Search area
Search the patient
Patient/Series/Image list
Select an image.
OK button
Load the selected patient`s image.
Cancel button
Cancel to load the image.
AccuVueMED_UM_COM_EN_01 48 / 149
4.2.5.4 Input the opinion
AccuVueMED_UM_COM_EN_01 49 / 149
4.2.5.5 Set report form
The User can set the report form using the 'setting' menu.
Address/ Extra Info
Input the Hospital address or further information.
Session
Input the titles of each Session.
Visible setting
Select the contents of the report to be printed.
All items are selected as default.
Save as default button
Save the current setting as default.
OK button
Use the selected form.
Cancel button
Cancel the selected form.
AccuVueMED_UM_COM_EN_01 50 / 149
4.2.5.6 Print
Click the [Save PDF] button to save the path and file name as a PDF file.
AccuVueMED_UM_COM_EN_01 51 / 149
4.2.5.8 Export of DCM report
DCM report is created when the send button is clicked. The created DCM report
is then sent to PACS.
The report can be found and viewed on the PACS.
AccuVueMED_UM_COM_EN_01 52 / 149
4.2.6 Analysis
You can check the shooting statistics (number of captured patients, number of shots,
number of regas, number of PACS transmissions, number of prints, etc.) during the set
period.
① Statistical results
AccuVueMED_UM_COM_EN_01 53 / 149
② Period setting
Search
Show the statistics according to the set period.
Export
Save the statistics according to the set period as an Excel file.
⑤ Analysis by projection
AccuVueMED_UM_COM_EN_01 54 / 149
4.2.7 Bodypart Editor
① Patient information
Select image of patient and select body part from ③Body part position for change
body part
③ Body part position
Click OK button to apply changes and Click Cancel button to cancel changes.
If click OK, changed information, Body part and image parameter will apply to select.
AccuVueMED_UM_COM_EN_01 55 / 149
5 Exam
The Exam tab processes the exam and edits images.
* Review mode is same as Exam tab bun cannot process exposure.
① Patient Information
: Pauses the current patient’s imaging session and the screen is returned to
AccuVueMED_UM_COM_EN_01 56 / 149
the patient list. The patient information currently in session is highlighted in orange on
the patient list.
④ Toolbar
View the list for images, or set the post-processing parameters which will be applied
to the images.
Refer to 5.2 Projection List & Image Processing Tab for more details.
Some settings are only available when integrated with X-ray generator.
Tube Voltage (kVp), Tube Current (mA), Tube Current Amount (mAs) setting.
When clicking the tube current amount icon (mAs), the unit can be changed to
exposure time (seconds).
Patient Size
Select the volume of the patient‘s body.
The values of the tube current, tube voltage and tube current amount are
renewed subsequent to the patient’s volume.
AEC
Set the AEC to be used.
Density
AccuVueMED_UM_COM_EN_01 57 / 149
Configure the imaging density.
Focus
Configure the Focal Size.
Grid
Configure the applied Grid.
Save
Store the images on the Local Storage.
Finish
Save the captured (or edited) image to Local Storage and automatically move to
the list screen.
Send
Transmit the images to the PACS server.
If long-clicking the icon, all the images that have been acquired or edited are
transmitted to the PACS server.
Indicator Icon
AccuVueMED_UM_COM_EN_01 58 / 149
5.1 Toolbar
② Tools List
③ Edit
AccuVueMED_UM_COM_EN_01 59 / 149
5.1.2 Edit Toolbar
The currently applied icons in the Exam tab are displayed in this box.
Icons in this box can be deleted by dragging & dropping the icon into ‘② Tools
Collection’ on the right.
Icons can be added into this box by dragging icons in ‘② Tools Collection’ and
dropping it on any empty location.
When the horizontal width of the Toolbar changes from 1 to 4, the column will be
visible from the left.
AccuVueMED_UM_COM_EN_01 60 / 149
② Available Icons
Add the icons with ROI Shutter Sizes specified by the User.
The following window appears when clicking the icon.
Enter the size (Width and Height) and click [OK] to add a new ROI.
⑤ OK
⑥ Cancel
AccuVueMED_UM_COM_EN_01 61 / 149
5.1.3 Toolbar Icons
The functions of all the icons in the toolbar are described in [Table 9].
For more details about certain tools, please refer Appendix E.
Icon Description
AccuVueMED_UM_COM_EN_01 62 / 149
Display the "OBL" marker on the screen
Zoom-in and out of the image by using left mouse button drag.
AccuVueMED_UM_COM_EN_01 63 / 149
Measure the length of the image
Stitch images
Ruler calibration
AccuVueMED_UM_COM_EN_01 64 / 149
Select annotation
Free angle image rotate tool (Draw angle will rotate image)
Free area image crop and rotate (Draw rectangle will crop image)
AccuVueMED_UM_COM_EN_01 65 / 149
5.2 Projection List & Image Processing Tab
The Projection List & Image Processing Tab can be used through the tab icon on the
upper right corner.
Figure 37. Projection List (Left) & Image Processing Tab (Right)
AccuVueMED_UM_COM_EN_01 66 / 149
5.2.1 Projection List
POP-UP function
AccuVueMED_UM_COM_EN_01 67 / 149
default when receiving the image from the detector.
Find auto position
∙ All projection: Find unexposed position among the all projections when you save or
send an image.
∙ One projection: Find unexposed position in the selected projection when you save
or send an image.
∙ Do not apply: do not find unexposed position when you save or send an image.
Auto Save
∙ Use : Enable auto save. If enabled, image will auto saved according to time interval
when exposure end.
∙ Time setting : Set the time interval for image auto save after exposure.
RIS Code Save/Update : Update RIS code to current exposure list. If no RIS code,
create new RIS code using current exposure list.
Position info. Update : Update position information according to current position`s
information. (Marker / rotation / flip / bucky type)
② Projection Guide
AccuVueMED_UM_COM_EN_01 68 / 149
5.2.2 Image Processing Tab
① Histogram
AccuVueMED_UM_COM_EN_01 69 / 149
② Image Enhancing Parameters
③ Parameter Functions
Wizard
Open the multi-image viewer window.
8 types of images are shown based on the currently set parameter.
Apply
Apply the currently set parameters to the image.
Reset
Reset the parameter to its initial setting.
Save
Save the currently set parameters and apply them when performing the projection
of the relevant projection-position.
AccuVueMED_UM_COM_EN_01 70 / 149
5.3 Truview ART
② Magnified screen
It magnifies selected position in the image viewer.
AccuVueMED_UM_COM_EN_01 71 / 149
You can magnify the certain area by double clicking the area.
③ Compare
Compare the two images with and without applying Truview ART.
2 images will appear. Left image is not adjusted by TRUVIEW ART, and Right image
is adjusted by TRUVIEW ART.
④ Save
Save TruviewART enable / disable status to parameter.
AccuVueMED_UM_COM_EN_01 72 / 149
6 Configuration
Configure the user settings of AccuVueMED.
Category Description
AccuVueMED_UM_COM_EN_01 73 / 149
Configure unnecessary RIS code filtering subsequent to
Blocked
ultrasound or other devices.
List
Refer to 6.6 Block List for more details.
AccuVueMED_UM_COM_EN_01 74 / 149
6.1 Display
6.2 System
Subcategory Description
Language Select system language.
Format Select date and name format.
Institution Enter the information of the institution.
Enable and disable automatic running of AccuVueMED
Options
at startup of Windows.
Workflow Set miscellaneous function.
AccuVueMED_UM_COM_EN_01 75 / 149
6.2.1 Language
AccuVueMED_UM_COM_EN_01 76 / 149
6.2.2 Format
AccuVueMED_UM_COM_EN_01 77 / 149
Enable Name Format
Name format configuration is available when this checkbox is checked.
Name Order
Alphabetic
Ideographic
Phonetic
Name Use Option: Select the type of name to use. At least one option should be
selected.
Last Name
First Name
Middle Name
AccuVueMED_UM_COM_EN_01 78 / 149
6.2.3 Institution
AccuVueMED_UM_COM_EN_01 79 / 149
6.2.4 Options
Auto Start
If checked, AccuVueMED will automatically start when Windows starts.
Not available on Win8.
Virtual Keyboard
If checked, virtual keyboard will be available for use.
AccuVueMED_UM_COM_EN_01 80 / 149
6.2.5 Workflow
AccuVueMED_UM_COM_EN_01 81 / 149
Example) Patient ID Generation on May 31st 2012 at 23:45
Generation
Identifier ID Generation Example
Rule
YYMMDD Time 120531112345
MMDDYY Automatic Sequencing 120531000001
DDMMYY Time 310512112345
YYYYMMDD Automatic Sequencing 20120531000001
MMDDYYYY Time 05312012112345
DDMMYYYY Automatic Sequencing 31052012000001
AccuVueMED_UM_COM_EN_01 82 / 149
6.2.5.1 Quick exposure mode (Add patient)
Exposure mode and exam sequence will changed when activate quick exposure
mode.
① Quick Exam
Originally located EMERGENCY button will replaced to QuickExam button.
Click QuickExam button to enter Quick exposure. Quick exposure has no position
and patient information.(Added patient`s information, position and other information
will filled “UNKNOWN”)
② EMERGENCY
Originally EMERGENCY button moved to work list.
For exposure EMERGENCY patient, no need to enter add patient window.
AccuVueMED_UM_COM_EN_01 83 / 149
6.2.5.2 Quick exposure mode (exam)
More quickly exposure will possible when entering exam window with QuickExam.
After using quick exposure mode, PACS sending function will not work till change
patient`s information. (correct position and patient name will need to send image to
PACS)
To change information of quick exposure study, recommend 4.2.7 Bodypart Editor to
change information. 4.2.7 Bodypart Editor is more easily and fast.
AccuVueMED_UM_COM_EN_01 84 / 149
6.3 Image
Subcategories Description
Image Folder Set the location to save the images.
Delete Option Set delete options for image files.
Overlay Set overlay on the images.
Import Import images to the Studylist.
Truview ART Enable or disable the Truview ART feature
AccuVueMED_UM_COM_EN_01 85 / 149
Open the folder selection window.
Select the folder where the images are to be stored.
(The folder path to save the images must be set to Drive: \Images\)
Save Raw DICOM automatically
If checked, Raw DICOM is created automatically when acquired images are saved.
Use Password
If checked, deletion of image requires a password.
Use Reject Code
If checked, a reject code (reason for deletion) is required whenever attempting to
delete an image file.
AccuVueMED_UM_COM_EN_01 86 / 149
Use Recycle Bin
If checked, recycle bin function is activated.
② Reject Code
AccuVueMED_UM_COM_EN_01 87 / 149
6.3.3 Overlay
Edit types of overlay to be displayed on the top left, top right, bottom left and bottom
right of the images.
The editing method is the same while the displaying locations differ.
AccuVueMED_UM_COM_EN_01 88 / 149
① Registered Overlay Type & Font Size
② Edit Icon
Add a new overlay to the list.
The following window as shown in [Figure 50] appears when clicking the icon.
The selected overlay is added after entering the size and clicking [OK].
The Texts that have already been registered will be displayed in black.
AccuVueMED_UM_COM_EN_01 89 / 149
Delete the selected annotation from the list.
The selected annotation moves up by one row.
The selected annotation moves down by one row.
Change font size of each location.
AccuVueMED_UM_COM_EN_01 90 / 149
6.3.4 Import
AccuVueMED_UM_COM_EN_01 91 / 149
6.3.5 Truview ART
AccuVueMED_UM_COM_EN_01 92 / 149
6.4 Storage
Configurations concerning the deletion and storage of data in the operating pc where
the images will be saved.
Details Description
Configure the deletion and automatic
Delete deletion function of images stored in
Local Storage.
Compress and back-up the images stored
Backup
in Local Storage.
6.4.1 Delete
Configure the deletion and automatic deletion function of images stored in Local
Storage.
AccuVueMED_UM_COM_EN_01 93 / 149
① Storage Delete
The [Delete] Icon deletes all images stored in the Local Storage.
Deleted images cannot be restored, so please make sure to perform a backup prior to
deleting images!
② Auto Delete
Auto Delete
If checked, all images satisfying the specified conditions are automatically deleted
from the PC Storage.
Disk Size
If checked, images in the PC Storage are automatically deleted once the used
disk space exceeds the set limit.
If remaining storage is less than ‘Disk Size’, overflowed images will be
automatically deleted. (Whole disk volume except ‘Disk Size’ is allocated for
AccuVueMED.)
Max: maximum limit of disk space set for storage of images in the PC Storage.
Period
If checked, images in the PC Storage are continuously deleted automatically
during the set period below.
ex1) Set to "2", "Month": Delete all images that were stored 2 months earlier as of
today..
ex2) Set to "1" "Day": Delete all images that were stored 1 day earlier as of today.
The cycle may be set by one of the followings; Day/Week/Month/Year.
Only one of the two auto deleting methods, ‘Disk Size’ and ‘Period’, may be selected
at a time.
AccuVueMED_UM_COM_EN_01 94 / 149
6.4.2 Backup
AccuVueMED_UM_COM_EN_01 95 / 149
Back-up the currently used database in the selected folder.
② Automatic Backup
AccuVueMED_UM_COM_EN_01 96 / 149
6.5 Account
AccuVueMED_UM_COM_EN_01 97 / 149
6.6 Block List
Filter unnecessary RIS codes such as codes related to ultrasound or other devices.
AccuVueMED_UM_COM_EN_01 98 / 149
6.7 Dataset
Series Order
Set the series order to be used when transmitting the images to PASC
One series: Combine the images in the same study into one series (DR method)
Each series: Separate the images in the same study into individual series.
AccuVueMED_UM_COM_EN_01 99 / 149
6.8 Network
Subcategory Description
Network connection is required.
Configure the network information of the PC (or
Local
Work Station) currently operating
AccuVueMED.
Network connection is required.
MWL Configure the network information of the server
storing the Modality Worklist.
Network connection is required.
Storage Configure the network information of the
Storage Server, which stores the image files.
Printer connection is required.
Printer Configure the connection information of the
printer.
Network connection is required.
MPPS Configure the network information of the MPPS
server
Configure the network information of the PC (or Work Station) currently operating
AccuVueMED.
AE Title
Enter the name of the local PC (or Work Station) which will be displayed on the
Network.
IP
Enter the IPv4 Address of the PC (or Work Station) currently operating
AccuVueMED.
Port
Enter the Port number.
It should be configured in accordance with the linked server.
All entered information will be automatically saved and applied.
Time-out
Response standby time with the server.
It is regarded as connection failure if the server is unresponsive for the set period
of time.
Query
Decide whether you want to perform the refresh automatically or manually.
Interval
Refresh is performed automatically every input interval.
③ Query
Decide which item to use for the condition to load the worklist
Type
Modality Query Base : Invokes the Worklist only in terms of duration and modality.
Patient Query Base : The Worklist is loaded in addition to the items set as Base in
the Period and Modality types.
Base
Perform a Patient Query Base with the set items.
④ Modality
Specify the format of the date to be used for the condition in which the Worklist is to
be loaded.
① Storage Server
② Option
Time-out
Response standby time with the server.
It is regarded as connection failure if the server is unresponsive for the set period
of time.
6.8.4 Printer
② Option
③ Medium Type
④ Border
⑤ Smoothing
⑥ Magnification
⑦ Density
⑨ Overlay information
Instructions for use are identical with the overlay configuration of the image.
Station AE Title
Inner Name of MPPS Server
Time-out
Response standby time with the server.
It is regarded as connection failure if the server is unresponsive for the set period
of time.
6.9 Detector
Subcategory Description
Register and edit the model of panel
(detector) being used.
General
Execute detector software, ECali1 (or
Caliview) and Terminal Emulator.
Detailed configuration screen related to
Panel A, registered in ‘General’ tab.
Panel A Detailed configuration items for panel are
created in accordance with the number of
panels registered in ‘General’.
③ Terminal
① Receptor
② Model
③ Serial Number
④ Grid Type
⑤ Image Mode
⑥ Transmission Mode
Edit
Manage the information, such as detailed projection information and imaging
environment, of the RIS Code.
Refer to 6.10.1 Edit RIS Code Window for more details.
[Deleting Sequence]
Delete 1) Select the code to be deleted from the list.
2) Click the [Delete] icon.
Extract and add the RIS code from the patient list retrieved
from the MWL server.
Icon Description
Both icons perform identical tasks.
, Add the projections relevant to the selected
Add Projection/Position in the RIS Code.
Enter the automatically configured environment value when imaging of the pertinent
RIS Code is performed.
HR / HT
Select the image transfer mode according to the imaging body part.(EVS-2430 Only)
HR : High Resolution, HT : High Transfer
Patient Size
Select the patient size.
4 levels of selections are available.
AEC
Set the AEC.
Grid
Select the Grid Type being used. (None / Fixed / Moving).
Focal Spot
Icon Description
Add a new Marker.
Refer to 6.10.2 Add Marker Window for more
Add details.
Location
Select the location for the marker to appear on the image
Text
Select the type of Marker
In addition to the texts provided by default, random texts can be used as a marker
User registered markers can be selected from the list
The Marker list window can be edited by using the [Add] Icon and [Delete] Icon
To change overall marker size, go to ‘Projection’ tab – “Marker font size” menu.
Marker Size
Change marker size.
Rotation
Set rotation selected receptor's all projection & position.
Flip V and Flip H are first applied and then Rotation is applied.
Delete
Delete the selected Projection or Position
Recent selections which will be deleted by clicking the [Delete] icon are
highlighted in yellow
Edit
Edit the Projection and Position after opening the ‘Project Editor Window’.
Refer to 6.11.1 Projection Editor Window for more details.
① Anatomy
Icon Description
Edit the name of the selected category from the list with
the entry in the left textbox
[Edit Sequence]
1) Select the category to be edited from the list
Edit
2) Enter the new information in the textbox on the left of
the icon
3) Click the [Edit] icon
Add the category with the information entered into the
textbox on the left of the icon
[Add Sequence]
Add 1) Enter the new information in the textbox on the left of
the icon
2) Click the [Add] icon
Delete the selected category from the list
[Delete Sequence]
Delete 1) Select the category to be deleted from the list
2) Click the [Delete] icon
④ Imaging Environment
If an image is deleted while the recycle bin option is activated, you can see the
information of the deleted image through “Recycle Bin” tab.
Terms Description
If you receive the update installer file, please run the provided setup file.
Support up to 4 mages.
Set SID as 180cm.
When performing the projection, please minimize the
patient's movements.
Caution
When performing the projection, please make sure that
the overlapping area is at least 10cm.
Please note there may be errors in automatic
alignment according to projection environment.
In order to perform the stitching function, you should select the projection in following two
ways.
WHOLE SPINE
LONG BONE
Thumbnail
Display an image that can be used for stitching.
Image Viewer
Display aligned the images automatically.
Toolbar
A collection of icons and buttons that let you perform various commands.
When you want to align the images in the image viewer automatically.
When you finish all the stitching works, press the [OK] button to exit.
Descriptions of the icons in the toolbar are as follows.
Icon Description
Stitching direction.
Zoom In
Zoom Out
① Click button.
② The 'Save Image List Window’ comes up.
Select a detector.
Select List on the right side and click '<' button.
The list on the left side is saved to the detector when clicking the [Save] button.
① Select a detector.
② The list will be saved in detector.
③ Show Current work list.
④ Function buttons.
< Button
Add a selected list.
Delete
Delete selected list.
Delete all
Delete all work list.
UP/Down
Change work list's order.
Refresh
Refresh work list.
Set portable
Change the detector's image acquisition mode to portable mode.
Caution: If it is set to portable mode, AccuVueMED vet and detector's
connection will be disconnected.
1) Image rotate
Image rotate tool operation procedure.
Prepare rotation
Apply rotation
① Click tool for activate free area crop and rotate mode.
Draw rectangle to crop and rotate.
Preparation
Calibration Procedure
* Check the radiation field and clear Acryl Phantom as well as Bucky before calibration.
* If the Mean values of every image have more than 300Gy difference, you have to re-
take images.
1. Run Ecali1
2. Select the detector to perform the Gain Calibration from the Detector Menu. (Tool bar
→ Detector)
Detector Selection
3. Select Gain Map Calibration from the Tool bar or select the shortcut icon.
7. The program automatically determines the appropriateness of the X-ray radiation dose
when the image is acquired.
8. If provided with a notification of low radiation dose, please select [OK] then increase
the X-ray radiation dose and re-project.
10. If the X-ray radiation dose is appropriate, please select [OK] and proceed with the
projections
11. A notification informing the requirement of 7 images will appear. The required number
of images may be set in Configuration. Select [OK] and proceed with the projections of 7
images.
.
Notification Informing 7 Images Requirement
12. Wait until the next designated projection is ready after the initial projection.
13. The next projection will take place after the projection standby timer (60 seconds)
disappears. The projection standby time may be set in Configuration.
14. The Gain Calibration results are automatically saved as Gain Map (*.GMP) when all
required images are projected and the existing files are backed-up.