ANDERSON NEGELE TFP-41-44-51-54-61-161-164-181-184 - 50008 - 3.3 - en - Eu

The document provides information on temperature sensors and transmitters for food and biopharma applications. It describes: 1) Temperature sensors with hygienic G1/2" connections that can be used for monitoring sterilization processes and vessel temperatures in the food industry. 2) The sensors have hygienic, easy to sterilize designs with gap-free seals and FDA compliant materials. 3) Optional transmitters can output temperature measurements on a 4-20mA current loop. Accuracy classes down to 1/10 B and operating temperatures from -50°C to 250°C are supported.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
465 views8 pages

ANDERSON NEGELE TFP-41-44-51-54-61-161-164-181-184 - 50008 - 3.3 - en - Eu

The document provides information on temperature sensors and transmitters for food and biopharma applications. It describes: 1) Temperature sensors with hygienic G1/2" connections that can be used for monitoring sterilization processes and vessel temperatures in the food industry. 2) The sensors have hygienic, easy to sterilize designs with gap-free seals and FDA compliant materials. 3) Optional transmitters can output temperature measurements on a 4-20mA current loop. Accuracy classes down to 1/10 B and operating temperatures from -50°C to 250°C are supported.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 8

SENSORS FOR FOOD AND BIOPHARMA.

Product Information TFP-41, -44, -51, -54, -61, -161, -164, -181, -184 FOOD

Temperature Sensor with G1/2" hygienic


Application/Specified usage Authorizations
·· Temperature Measurement in vessels and pipes
·· Front flush temperature measurement available

74-06
Application examples
·· Monitoring of CIP-/ SIP-process Temperature sensor
·· Measurement in vessels with agitators with front flush version TFP-164 / ... / MPU-M
·· Temperature monitoring in milk vessels

Hygienic design/Process connection


·· Flow optimized, hygienic and easy sterilizable installation by using Negele
weld-in sleeve, e.g. EMZ-132 or build-in system, e.g. EHG-... / 1/2"
·· Additional process connections: adapters for Tri-Clamp, dairy flange (DIN
11851), Varivent, DRD, APV et al
·· Sealing system free of elastomers, the connection will be without gaps and
crevices
·· Product contacting materials compliant to FDA
·· Sensor completely made of stainless steel resp. PEEK (front flush sensor)
·· Conforming to 3-A Sanitary Standard 74-06 for front flush sensors

Features/Advantages
·· Front flush mounting possible
·· Integrated transmitter optional
·· Different electrical connections available
Temperature sensor TFP-41

Options/Accessories
·· 2 x Pt100 (not retrofittable)
·· 2 x Pt100 with two transmitters (not retrofittable)
·· Programmable transmitters MPU-4 as well as MPU-M with output
4...20 mA, 2-wire
·· Integrated transmitters for HART-protocol
·· Programming adapter MPU-P 9701
·· Integrated transmitter MPU-LCD with display in connecting head
·· Pt100 chip with other classes of accuracy (1/3B, 1/10B)
·· Fast response sensor tip 3 mm and 4 mm
·· Spacer for high temperature up to 250 °C
permanent temperature up to 450 °C (on request)
·· Pre-assembled connecting cable for M12 plug
·· Fixed cable in other lengths and other material available

Accessories PVC-cable with M12-connection

PVC-cable with M12-connection made of 1.4305, IP 69 K, unshielded

M12-PVC / 4-5 m PVC-cable 4-pin, length 5 m


M12-PVC / 4-10 m PVC-cable 4-pin, length 10 m
M12-PVC / 4-25 m PVC-cable 4-pin, length 25 m
FOOD Specification 2

Temperature sensor
Process connection thread G1/2" combined with Negele weld-in sleeves,
build-in systems, adapter sleeves
Tightening torque sensor sealing PEEK 10 Nm
sensor sealing stainless steel 20 Nm
Insertion length EL TFP-41, -51, -61, -161, -181 20...500 mm
TFP-44, -54, -164, -184 front flush
Materials connecting head stainless steel 1.4301 (AISI 304)
thermowell stainless steel 1.4404 (AISI 316L)
at TFP-44, -54, -164, -184 PEEK
Operating pressure TFP-41, -51, -61, -161, -181 50 bar maximum
TFP-44, -54, -164, -184 10 bar maximum
Temperature ranges ambient -50...+80 °C
sensor tip TFP-xx1 -50...+250 °C
sensor tip TFP-xx4 -50...+140 °C
Sensing resistor acc. to DIN EN 60751 Pt100
Electrical connection cable gland cable M16 x 1,5
connection M12 plug 1.4301 (AISI 304), 4-pins
fixed cable 2.5 m LIYY 4 x 0,25 mm²
fixed cable 2.5 m (≥ 90 °C) PTFE 4 x 0,14 mm²
Protection class IP 69 K (with electrical connection M12 plug)

Transmitter MPU-4, MPU-H, MPU-M


Temperature ranges ambient -40...+85 °C
storage -55...+90 °C
Measuring ranges MPU-4, MPU-H, MPU-M standard: -10...40 °C, 0...50 / 100 / 150 / 200 °C
special ranges free programable
Accuracy input < ±0.25 °C
Temperature drift zero, span < 0.01 % / K
Supply MPU-4, MPU-H, MPU-M 8...35 V DC
accuracy 0.01 % / V (reference: 12 V DC)
Output signal analog 4...20 mA
accuracy < ±0.1 % of measurement range
burden < 600 Ω (at U B = 24 V)
Humidity without condensation 0...98 %

Accuracy classes of temperature sensors | Tolerances for Pt100 acc. to DIN EN 60751
Pt100 A 1/3 B 1/10 B
0 °C / 100 Ω ±0,15 K / ±0,06 Ω ±0,10 K / ±0,04 Ω ±0,03 K / ±0,01 Ω
100 °C / 138,5 Ω ±0,35 K / ±0,13 Ω ±0,27 K / ±0,10 Ω ±0,08 K / ±0,03 Ω
3 Electrical Connection FOOD

Elektrischer
Electrical connection
Anschlusswithout
ohne Kopftransmitter
transmitter Elektrischer
Electrical connection
Anschlusswith
mit transmitter
Kopftransmitter

With 1 x M12 plug M12-Stecker


With M12 plug

Configuration
Belegung 1st M12 plug
1. M12-Stecker Configuration M12 plug
Belegung M12-Stecker

1 2 3 4 1: + supply
2: - supply 4...20 mA
3: not connected
4: not connected

With 2 x M12 plug Mit Kabelverschraubung


With cable gland

Configuration
Belegung 2nd M12 plug
2. M12-Stecker 4...20 mA current loop
4...20 mA Stromschleife
1 2 3 4

With cable gland

Configuration
Belegung strip terminal
Klemmen

1 2 3 4
Electrical connection with two transmitter (TFP-61)

With 1 x M12 plug (sensor 1 + sensor 2)


Klemmen
clamps forfür
1st1.Pt100
Pt100
Configuration M12 plug
Klemmen
clamps forfür
2nd2. Pt100
Pt100
(bei
(at version
Ausführung
2 x Pt100)
2 x Pt100) 1: + supply (sensor 1)
2: - supply 4...20 mA (sensor 1)
3: - supply 4...20 mA (sensor 2)
4: + supply (sensor 2)
With fixed cable

Fixed cable
Belegung connection
Kabel mit With 2 x M12 plug (sensor 1)
with
1 1 x Pt100
x Pt100
Configuration M12 plug

1: + supply (sensor 1)
2: - supply 4...20 mA (sensor 1)
3: not connected
4: not connected

Fixed cable
Belegung connection
Kabel mit Fixed cable
Belegung connection
Kabel mit With 2 x M12 plug (sensor 2)
with 2 x Pt100
2 x Pt100 (LIYY)(LIYY) with 2 x Pt100
2 x Pt100 (PTFE)(PTFE)
Configuration M12 plug

1: + supply (sensor 2)
2: - supply 4...20 mA (sensor 2)
3: not connected
4: not connected
FOOD Dimensional Drawings 4

TFP-41 | TFP-41.2 TFP-44 TFP-51 | TFP-51.2

52
52
52

Important advice for TFP-51, -51.2 and TFP-54

Tighten the sensor only at the lower, marked in yellow spanner flat
(BE = 22 mm)!

TFP-161 TFP-161 / ... / MPU-M TFP-61

59
5 Dimensional Drawings FOOD

TFP-164 TFP-164 / ... / MPU-M TFP-181 | 181.2 TFP-184

Sensor tip diameter and response time


All temperature sensors are available with smaller sensor tips, to ensure a shorter response time. The below-mentioned
times were measured by emersing a temperature sensor from room temperature into boiling water.

Sensor tip Ø 6 mm Sensor tip Ø 4 mm Sensor tip Ø 3 mm Sensor front flush

50 %-time: t 50 ≤ 3.0 s 50 %-time: t 50 ≤ 2.4 s 50 %-time: t 50 ≤ 0.5 s 50 %-time: t 50 ≤ 5.7 s


90 %-time: t90 ≤ 8.0 s 90 %-time: t90 ≤ 6.5 s 90 %-time: t90 ≤ 1.5 s 90 %-time: t90 ≤ 33.2 s

Ø6 Ø6

Ø6

Ø4 Ø3
FOOD Installation | Warnings 6

Conditions for a measuring point according to 3-A Sanitary Standard 74-06

·· The sensors TFP-44, -54, -164, -184 conforming to the 3-A Sanitary Standard.

·· The sensors are designed for CIP-/ SIP-cleaning. Maximum 140 °C / 120 minutes.

·· Only with the build-in system CLEANadapt (EMZ, EMK, EHG with tube ≥ DN25, ISO 20 and G1", Adapter AMC and AMV)
allowed.

·· Using the weld in sleeve EMZ, EMK the weld must comply to the requirements of the current 3-A Sanitary Standard.

·· Mounting position, self draining and the position of the leackage hole must be in accordance to current 3-A Sanitary
Standard.

Mechanical connection/Installation Conventional usage

·· Use only Negele CLEANadapt system for safe operation ·· Not suitable for applications in explosive areas.
of measuring point! ·· Not suitable for applications in safety-relevant system
parts (SIL).

Transport/Storage
Standards and guidelines
·· Do not store outside
·· Store in an area that is dry and dust-free ·· Compliance with the applicable regulations and direc-
·· Do not expose to corrosive media tives is mandatory.
·· Protect against solar radiation
·· Avoid mechanical shock and vibration
·· Storage temperature -55...+90 °C
·· Relative humidity max. 98% Note on CE

·· Applicable directives:
Electromagnetic Compatibility Directive 2014/30/EU
Cleaning/Maintenance ·· Compliance with the applicable EU directives is identified
by the CE label on the product.
·· When using a pressure washer, do not point the nozzle ·· The operating company is responsible for complying
directly at the electrical connections. with the guidelines applicable to the entire
installation.

Reshipment
Disposal
·· Sensors shall be clean and free of media or heat-
conductive paste and must not be contaminated with ·· Electrical devices should not be disposed of with
dangerous media! household trash. They must be recycled in accordance
·· Use suitable transport packaging only to avoid with national laws and regulations.
damage of the equipment! ·· Take the device directly to a specialized recycling com-
pany and do not use municipal collection points.
7 Order Code FOOD

Order code for version with 1 x Pt100

TFP-41 (connecting head Ø 55 mm)


TFP-44 (connecting head Ø 55 mm, front flush)
TFP-51 (connecting head Ø 55 mm, with spacer)
TFP-54 (connecting head Ø 55 mm, with spacer, front flush)
TFP-161 (connecting head Ø 18 mm, electrical connection M12 plug)
TFP-164 (connecting head Ø 18 mm, electrical connection M12 plug, front flush)
TFP-181 (connecting head Ø 18 mm, electrical connection 2.5 m PTFE-cable, other lengths: see accessories,
no transmitter possible!)
TFP-184 (connecting head Ø 18 mm, electrical connection 2.5 m PTFE-cable, other lengths: see accessories,
front flush, no transmitter possible!)
Sensor length in mm
020...500 (in steps of 5 mm)
xxx (special length on request)
Diameter thermowell in mm (not selectable for TFP-44, -54, -164, -184)
6
8
10
12
Diameter sensor tip in mm (not selectable for TFP-44, -54, -164, -184)
X (no reduction)
3 (only for thermowell 6 mm)
4 (only for thermowell 6 mm and 8 mm)
6 (only for thermowell 8 mm and 10 mm)
8 (only for thermowell 12 mm)
Accuracy class Pt100
A
1/3B
1/10B
Electrical connection
(not selectable for TFP-161, -164, -181, -184)
PG (cable gland M16x1.5)
M12 (M12 plug, standard with MPU-LCD)
Transmitter

X (without)
only for TFP-41, -44, -51 and -54
MPU-4 (programmable)
MPU-H (HART-protocol)
MPU-LCD (with display)
only for TFP-161 and -164

MPU-M (programmable)
Measuring range MPU
(only for types with transmitter;
not at MPU-LCD)
-10...40 (range -10...40 °C)
0...50 (range 0...+50 °C)
0...100 (range 0...+100 °C)
0...150 (range 0...+150 °C)
0...200 (range 0...+200 °C)
xx...yy (special range)

TFP-41 / 100 / 6/ X/ A/ PG / MPU-4 / 0...100


FOOD Product Information TFP-41, -44, -51, -54, -61, -161, -164, -181, -184 8

Order code for version with 2 x Pt100

TFP-41.2 (connecting head Ø 55 mm, 2 x Pt100, no transmitter possible!)


TFP-51.2 (connecting head Ø 55 mm, 2 x Pt100, with spacer, no transmitter possible!)
TFP-61 (higher connecting head Ø 55 mm, 2 x Pt100, prepared for 2 x transmitter)
TFP-61-H (like TFP-61, but with spacer)
TFP-181.2 (connecting head Ø 18 mm, electrical connection 2.5 m PTFE-cable; other lengths: see at accessories)

Sensor Length in mm
020...500 (in steps of 5 mm)
xxx (special length)
Diameter thermowell in mm
6
8
10
12
Diameter sensor tip in mm
X (no reduction)
3 (only with thermowell 6 mm)
4 (only with thermowell 6 mm and 8 mm)
6 (only with thermowell 8 mm and 10 mm)
8 (only with thermowell 12 mm)
Accuracy class Pt100
A
1/3B
1/10B
Electrical connection (only for TFP-41.2 and TFP-51.2)
PG (cable gland M16x1.5)
2PG (2 x cable gland M16x1.5)
2M12 (2 x M12 plug)
Electrical connection (only for TFP-61 and TFP-61-H)
M12 (M12 plug)
2M12 (2 x M12 plug)

Continue if TFP-61 oder TFP-61-H is selected!


No further options for TFP-41.2, -51.2, -181.2!
1. Transmitter
MPU-4 (programmable)

Measuring range 1. MPU


-10...40 (measuring range -10...40 °C)
0...50 (measuring range 0...+50 °C)
0...100 (measuring range 0...+100 °C)
0...150 (measuring range 0...+150 °C)
0...200 (measuring range 0...+200 °C)
xx...yy (special range)
2. Transmitter
MPU-4 (programmable)
Measuring range 2. MPU
-10...40 (-10...40 °C)
0...50 (0...+50 °C)
0...100 (0...+100 °C)
0...150 (0...+150 °C)
0...200 (0...+200 °C)
xx...yy (special)

TFP-61 / 100 / 6/ X/ A/ M12 / MPU-4 / 0...50 / MPU-4 / 0...50

50008 / 3.3 / 2019-07-26 / SD / EU