Absorbance 96 App
Version 1.2
4 Operation............................................................................................................................................ 5
II
General information
1 General information
The Absorbance 96 App is to be used only in combination with the Absorbance 96 reader.
The Absorbance 96 App is not a stand-alone product. Please also refer to the separate op-
erating instructions for the Absorbance 96 reader.
The Absorbance 96 App is only compatible with the Absorbance 96 reader. The safeties,
precautions and validations included in the Absorbance 96 App have only been tested for
the Absorbance 96 and shall not be used for any other purpose on any other device.
Warning:
This is a warning that indicates an immediately or potentially dangerous situa-
tion, which, if not avoided, will result in death or serious injury.
Caution:
Caution indicates a potentially hazardous situation, which, if not avoided, could
result in minor or moderate injury.
Information:
This is a piece of information indicating certain properties that must be ob-
served.
This user manual may only be used according to its intended purpose. It may not be repro-
duced, changed, or translated in another language without the prior written consent of
Byonoy GmbH.
Ensure that the manual is always accessible and is read and understood by all
persons operating the instrument.
The latest version of this manual is available via the Absorbance 96 App.
1
App installation procedure and updates
The Absorbance 96 App is compatible with Microsoft Windows (Windows 7 or above) and
Mac OS (High Sierra 10.13. or above).
To install the Absorbance 96 App, insert the USB stick into the computer and follow the steps
below:
Modern operating systems offer plug and play functionality. Older operating systems may
require manual installation of the software drivers. The drivers for Windows are available on
the USB stick.
To install the proper drivers and software on the computer, the user must have
administrator rights or obtain such permissions.
The Absorbance 96 App can be uninstalled using the computer’s standard uninstall proce-
dure.
Updates
To check for new updates, click the update button in the lower left corner. If an update is
available, this will be shown in a bar at the top of the screen.
2
Absorbance 96 App overview
The Absorbance 96 App is used to operate the Absorbance 96 reader and process the re-
sults. After the app is started and the reader is connected, the main window appears auto-
matically.
3
Absorbance 96 App overview
File Types
Description Extension
Protocol .byop
Assay .byoa
Export .csv/.pdf
The protocol file type .byop allows to save settings and to load predefined protocols for re-
peat assays.
The assay file type .byoa is used to save completed assays. In addition to the settings, the
results of the assay are also saved.
4
Operation
4 Operation
Select “New” from the menu bar. Your experiment conditions can now be defined.
Workflow Parameters
1. Wavelengths
Under “Wavelengths”, a signal and reference wavelength (both in nanometers) may be cho-
sen.
The signal wavelength is the wavelength at which the principal measurement will be run.
The reference wavelength is an additional wavelength at which the device will measure and
whose results will be subtracted from the values measured at the signal wavelength. If no
reference wavelength is desired, click the “None” button beside the reference wavelength.
2. Readout Method
Select your readout method, either Endpoint or Kinetic.
The kinetic readout method is not available for every reader type.
To perform kinetic measurements and avoid misinterpretation of the results, observe the
following instructions:
Acclimatize the connected reader and microtiter plate to the ambient tempera-
ture for at least 1 hour.
3. Evaluation
The evaluation section allows to choose which evaluative method will be used for the
readout.
No Evaluation
Quantitative
Upon selection of the quantitative evaluation method, a drop-down menu will appear. Here,
a fitting method can be chosen. The selected fitting method will be used for the standard
curve and the data analysis.
5
Operation
In a quantitative assay, a predefined standard curve can be used to determine the concen-
trations of a new set of samples with identical experiment parameters. After selecting the
checkbox “Use predefined standard curve”, you may enter the curve parameters manually
or load a standard curve from a previous assay.
After selecting the checkbox “Use predefined standard curve”, you cannot as-
sign any new standards.
Qualitative
Upon selection of qualitative evaluation, a formula for the cut-off definition will appear.
The terms for cut-off definition from left to right according to Figure 2:
The formula setup for the cut-off definition covers multiple formula types. These are listed
below.
6
Operation
Example:
To replace the minus function, the user must simply enter a negative value. For example, if
the cut-off is
In case the Cut-Off is known or provided by the kit manufacturer, and therefore does not
need to be calculated, select 0 OD and enter the value into the middle field:
The examples shown in this manual do not represent actual data, rather func-
tion to clarify software features, and may not be referenced for experimental
methods or procedure.
Plate Layout
In the plate layout section, samples, standards, blanks, and controls can be defined. Select
the type from the options below the plate layout; click and drag to define. Duplication pat-
terns allow the user to define multiple replicates for samples and standards.
The format of the plate layout and the corresponding buttons are defined by
the evaluation method. Quantitative and qualitative evaluations have different
fields and buttons.
When replicates are defined, the mean value is used for the calculations.
In the plate layout for a quantitative assay, a field is provided where the concentrations of
the standards and the samples can be entered. The unit can be entered manually or chosen
from the dropdown menu. Using the dilution factor, diluted samples can be used, and the
software will calculate the undiluted concentration.
B0 control
7
Operation
Create a protocol
Once the parameters of a given experiment have been set, these settings may be saved as
a protocol. All parameters, such as wavelength and plate layout, will be saved.
Once all settings have been completed, the read out can be started. The app will automati-
cally guide the user through the readout process.
4.3 Results
After a successful measurement, the results section will open automatically. The results are
displayed in different views:
Plate view: Provides an overview of the different OD values in their respective wells. Unde-
fined wells are not shown. Click on a given well to see the OD value, label, and, if applicable,
concentration.
Graph view: A representation of the data will be shown on the graph. The standard curve is
dependent on the fitting method chosen. By choosing the right fitting method, the distance
between the curve and each point is minimized. Using the standard curve, the concentration
of each sample is calculated. The equation below the graph reflects the fitting method ap-
plied and offers further data.
Tabular view: Provides an overview of the various samples, wells, OD values, and calculations.
Parameter Description
Export
8
EC-Conformity Declaration
5 EC-Conformity Declaration