Epa Protocol 2012
Epa Protocol 2012
gov/ord
Robert S. Wright
Air Pollution Prevention and Control Division
National Risk Management Research Laboratory
Research Triangle Park, NC 27711
Notice
Cover photo: Gas cylinders stored at the National Institute of Standards and
Technology laboratory in Gaithersburg, Maryland (photo courtesy of
National Institute of Standards and Technology)
ii
EPA Traceability Protocol for Gaseous Calibration Standards
Contents
Section 4 - EPA Traceability Protocol for Assay and Certification of Dynamic Gas
Dilution Systems ................................................................................................. 97
4.1 Procedure D1: Assay and Certification of a Dynamic Gas Dilution
System ............................................................................................................ 97
iii
EPA Traceability Protocol for Gaseous Calibration Standards
Figures
Number Page
2-3 Two designs for assay apparatus with dilution (Procedure G2) 49
4-1 One possible design of a gas dilution system for the generation 99
of pollutant gas mixtures using dynamic dilution methods
Tables
Number Page
2-2 NIST and VSL Primary Gas Mixture Suites that are declared to 9
be Equivalent
iv
EPA Traceability Protocol for Gaseous Calibration Standards
Acknowledgments
This traceability protocol is a revision of an earlier protocol1 that was itself
revised in 1997. The original protocol was prepared in 1978. This revision was
reviewed by personnel of U.S. EPA, the National Institute of Standards and
Technology (NIST), the South Coast Air Quality Management District (AQMD),
and the Van Swinden Laboratorium (VSL, the national metrology institute of the
Netherlands) as it was being prepared. The author thanks the following
individuals for their assistance in providing technical information, reviewing this
revision of the protocol, and providing comments:
Airgas
Air Liquide
Applied Gas
Bios International
Environics
Linde
Matheson
NorLab
Praxair
Scott-Marrin
Takachiho
The spreadsheet modifications for this revision were made by Dr. Brian
Lopes and Dr. Kevin Ross, who were then working in the Statistics Department
of the University of North Carolina at Chapel Hill, and by Avraham Teitz of EPA
Region 2. The spreadsheets were first constructed by Dr. Michael Messner, who
was then working at RTI International.
v
EPA Traceability Protocol for Gaseous Calibration Standards
vi
EPA Traceability Protocol for Gaseous Calibration Standards
This revision of the protocol has several major and minor changes from
the 1997 revision, including those listed below:
1
EPA Traceability Protocol for Gaseous Calibration Standards
9. If the time period between the initial and second assays is at least 6
months and if the mean concentrations from the two assays are
demonstrated to be within 1.0 percent of each other using TOST,
candidate standards whose concentrations are lower than the
concentration range given in Table 2-3 may be certified for the full
certification period given in Table 2-3 rather than only six months as
was specified in earlier versions of the protocol (see Subsection
2.1.9);
10. Standards may be recertified if the cylinder pressure after the assays
have been completed is greater than 100 psig (see Subsection
2.1.11);
11. The minimum cylinder pressure has been changed to 100 psig (see
Subsection 2.1.12);
12. The PGVP is the primary accuracy assessment for EPA Protocol
Gases. EPA air pollution monitoring regulations specify that the
regulated community must use EPA Protocol Gases from PGVP
participants (see Subsection 2.1.13);
2
EPA Traceability Protocol for Gaseous Calibration Standards
15. A new procedure has been written and a new spreadsheet has been
prepared for the assay and certification of dynamic gas dilution
systems (see Section 4). At this time, EPA does not require the
regulated community to use NIST-traceable dynamic gas dilution
systems for the calibration of ambient air or continuous emission
monitors that are required by 40 CFR Parts 50, 58, 60, and 75.
However, end users may elect to use these systems for calibrations.
3
EPA Traceability Protocol for Gaseous Calibration Standards
Use of this protocol is mandatory for certifying the calibration gases being
used for the calibration and audit of ambient air quality analyzers and continuous
emission monitors that are required by 40 CFR Parts 50, 58, 60, and 75 2-5. This
protocol describes three procedures for assaying candidate standards (i.e., not-
yet-certified standards) and for certifying that their concentrations are traceable
to a National Institute of Standards and Technology (NIST) compressed gas
reference standards [i.e., Standard Reference Materials (SRMs), NIST-Traceable
Reference Materials3 [NTRMs, including NTRM-Primes (NTRM*s) for which NIST
analyzes every sample in the production lot], and NIST Research Gas Materials
(RGMs)]. It may be used to assay and certify only candidate standards that have
the same components and concentration ranges as NIST-traceable reference
standards. A multiple-component candidate standard may be assayed and
certified under this protocol if NIST-traceable reference standards that contain
the same individual components as are in the candidate standard exist.
This protocol may be used by specialty gas producers, end users, or other
laboratories to assay candidate standards. The assay involves the direct
comparison of the candidate standards to NIST-traceable reference standards
without dilution (i.e., Procedure G1) or the indirect comparison of the candidate
standards to reference standards with dilution (i.e., Procedure G2). A candidate
standard having a concentration that is lower or higher than that of the reference
standard may be certified under this protocol if both standards' concentrations (or
diluted concentrations) fall within the well-characterized region of the analyzer's
calibration curve. This protocol places no restrictions on cylinder sizes and the
same analytical procedures must be used in assays of all cylinder sizes.
Standards that are certified under this protocol must remain in the
cylinders in which they were originally assayed. Unassayed standards in
cylinders that have been transfilled from cylinders assayed under this protocol
cannot be certified as being EPA Protocol Gases
This protocol may be used to assay and certify zero air materials.
Certified impurity concentrations can be certified using Procedures G1 or G2.
Procedure G3 can be used to certify that zero air materials conform to the purity
specifications of 40 CFR Part 72.2 (i.e., concentrations of SO2, NOX, and THC
are not greater than 0.1 ppm; whose concentration of CO is not greater than1
ppm; and whose concentration of CO2 is not greater than 400 ppm)7.
5
EPA Traceability Protocol for Gaseous Calibration Standards
6
EPA Traceability Protocol for Gaseous Calibration Standards
7
EPA Traceability Protocol for Gaseous Calibration Standards
TABLE 2-1. Summary of Compressed Gas SRMs, NTRMs, and RGMs that are
available from NIST and their Concentration Rangesa
Certified Balance SRM b range NTRMb range RGMb range
Certification
component gas period
(years)
Ammonia Nitrogen Not available Not available 5 to 50 ppm 1
Carbon dioxide Air 390 to 400 ppm 360 to 420 ppm 360 to 420 ppm 8
Carbon dioxide Nitrogen 500 ppm to 16% 5 ppm to 20% Contact NIST 8
Carbon
Air 60 to 160 ppb 50 to 500 ppb 40 to 500 ppb TBDc
monoxide
Carbon
Air 10 to 45 ppm 10 to 45 ppm Contact NIST 8
monoxide
Carbon
Nitrogen 10 ppm to 13% 101 ppm to 15% Contact NIST 8
monoxide
Formaldehyde Nitrogen Not available Not available 0.5 to 10 ppm 1
Hydrogen
Nitrogen Not available Not available Contact NIST TBD
chloride
Hydrogen
Nitrogen 5 to 20 ppm 1 to 400 ppm Contact NIST 3
sulfide
Methane Air 1 to 100 ppm 1 to 1000 ppm 1.7 to 2.2 ppm 8
Methane Nitrogen Not available 0.5 ppm to 10% Contact NIST 8
Methanol or Nitrogen
Not available Not available 75 to 500 ppm 4
ethanol or Air
Natural gas Natural
Not Available Contact NIST Contact NIST 4
componentsd gas
Nitric oxide Nitrogen 0.5 to 50 ppm 0.5 to 50 ppm Contact NIST 3
Nitric oxide Nitrogen 50 to 3000 ppm 50 to 3000 ppm Contact NIST 8
Nitrous oxide Air TBD 300 ppb to 5% 300 to 350 ppb 8
Total oxides of 3
Air 100 ppm 10 to 100 ppm 3 to 100 ppm
nitrogen (NOX)e (SRM 6)
Oxygen Nitrogen 2 to 21 % 0.4 to 25 % Contact NIST 8
Propane Air 0.1 to 500 ppm 0.1 to 500 ppm Contact NIST 8
Propane Nitrogen 100 to 2000 ppm 5 ppb to 2% Contact NIST 8
Sulfur dioxide Nitrogen 5 ppm 5 to 50 ppm Contact NIST 4
Sulfur dioxide Nitrogen 50 to 3500 ppm 50 to 5000 ppm Contact NIST 8
Volatile organics Nitrogen 5 ppb Contact NIST Contact NIST 4
Zero air
Air Not available Not available Contact NIST TBD
materialf
a
All SRMs may not be available at all times. Other SRMs may be developed in the future and
could be used as reference standards. Contact NIST for information about SRM availability.
b
Concentrations are by mole; ppb = parts per billion; ppm = parts per million.
c
To be determined.
d Natural gas components are methane, ethane, propane, n-butane, iso-butane, n-pentane, iso-
pentane, helium, nitrogen, and carbon dioxide.
e
NIST defines its total NOX standards as containing nitrogen dioxide plus contaminant nitric acid.
f
Concentrations of SO2, NOX, and total hydrocarbons (THC) are not >0.1 ppm; concentration of
CO is not >1 ppm; and concentration of CO2 is not >400 ppm as per 40 CFR Part 72.2.
8
EPA Traceability Protocol for Gaseous Calibration Standards
TABLE 2-2. NIST and VSLa Primary Gas Mixture Suites that are declared to be
Equivalent
Certified Balance gas Concentrationb Maximum Stability period
component range for gas allowable (years)
mixture suite differencec
Carbon
Nitrogen 10 ppm to 20 % 0.3 % relative 3
dioxide
Carbon
Air 100 to 500 ppm 0.5 % relative 3
dioxide
Carbon Nitrogen or
1 ppm to 10 % 0.3 % relative 3
monoxide Air
Nitrogen or
Ethanol 75 to 500 ppm 0.5 % relative 3
Air
Hydrogen
Nitrogen 10 to 1000 ppm 1.0 % relative 2 or 3f
sulfide
0.5 % relative
d
Natural gas Nitrogen Typical (0.3 % for 3
CH4)
Nitric oxide Nitrogen 0.5 ppm to 1 % 0.5 % relative 2 or 3f
Nitrogen Nitrogen or
10 ppm to 1 % 0.5 % relative 2
dioxide Air
Oxygen Nitrogen 10 ppm to 25 % 0.2 % relative 3
Nitrogen or
Propane 1 ppm to 1 % 0.3 % relative 1, 2 or 3f
Air
Sulfur dioxide Nitrogen 1 ppm to 1 % 0.5 % relative 2 or 3f
Volatile
organic Nitrogen 1 ppb to 1 ppm 2 % relative 2
compoundse
a
Information about Reference Gas Mixtures can be obtained from:
VSL [i.e., the Van Swinden Laboratorium, the National Metrology Institute of the Netherlands]
Thijsseweg 11, 2629 JA Delft, NL P.O. Box 654, 2600 AR Delft NL
Telephone: 31 (0) 15 269 1550 FAX: 31 (0) 15 261 2971
E-mail: [email protected] Website: www.vsl.nl
b
Within the listed ranges, any concentration is available. PRMs are prepared individually in 5-L
cylinders according to ISO Standard 6142 (Gas Analysis–Preparation of calibration gas mixtures-
weighing methods). After preparation, the composition is verified against VSL Primary Standard
Gas Mixtures. CRMs are available in larger size cylinders and are gravimetrically prepared by an
accredited supplier. They are certified by VSL against VSL Primary Standard Gas Mixtures.
c
Maximum allowable difference between NIST and VSL primary standard gas mixture suites.
d
Methane, ethane, propane, n-butane, iso-butane, n-pentane, iso-pentane, 2,2-dimethylpropane
(neopentane), n-hexane, carbon dioxide, and helium.
e
Ethane, ethene, propane, propene, iso-butane, iso-butene, 1-butene, n-butane, 2-methyl
butane, iso-pentane, n-pentane, 1-pentene, 1,3-butadiene, trans-2-pentene, 2-methyl pentane,
2,2,4-trimethyl pentane, n-hexane, n-heptane, benzene, toluene, n-octane, and o-xylene.
f
Stability period is dependent on the concentration of the PRM/CRM.
9
EPA Traceability Protocol for Gaseous Calibration Standards
1. A candidate GMIS must be assayed on at least three separate dates that are
uniformly spaced over at least a 3-month period. During each of these
assays, the candidate GMIS must be measured at least three times. All these
assays must use the same SRM, NTRM, RGM, PRM or CRM as the
reference standard to avoid errors associated with the use of different
reference standards for different assays.
2. For each assay, the analyst must calculate the mean concentration and U for
the three or more measurements of the candidate GMIS according to the
statistical procedures described in Appendix A or an EPA-approved
equivalent statistical technique (see Subsection 2.1.6). The value of U must
be less than or equal to 1.0 percent of the mean concentration.
3. After the three or more assays have been completed, the analyst must
calculate the overall mean concentration and U for the candidate GMIS using
the spreadsheet described in Appendix C or an EPA-approved equivalent
statistical technique (see Subsection 2.1.6). If an EPA-approved equivalent
statistical technique is used, it must be identified as such in the certificate
documentation and it must be described in media readily accessible to end
users.
5. The certification period for a GMIS is the same as for an EPA Protocol Gas of
the same composition and concentration.
6. A GMIS must be recertified after its original certification period ends. Three
or more discrete measurements of the candidate standard must be made
during a single recertification assay. Use the spreadsheet described in
Appendix C or an EPA-approved equivalent statistical technique (see
10
EPA Traceability Protocol for Gaseous Calibration Standards
2.1.3.5 Zero Gas for Multipoint Calibrations and Zero Gas Checks—
Zero gas used for multipoint calibrations, zero gas checks or for dilution of any
candidate or reference standard must be clean, dry, zero-grade air or nitrogen
containing no detectable concentration of the pollutant of interest. It may come
from compressed gas cylinders or from zero gas generators. The use of NIST-
traceable zero air material is recommended, but not required. The zero gas must
match the balance gas in the candidate standard and the reference standard,
unless it has been demonstrated that the pollutant gas analyzer is insensitive to
11
EPA Traceability Protocol for Gaseous Calibration Standards
For analyzers such as gas chromatographs, the analyst may suspect that
the zero gas reading may not accurately represent the zero-intercept of the
calibration equation. The analyst may substitute a low-concentration, NIST-
traceable reference standard for the zero gas, providing that the concentration of
this standard is less than the concentration of the candidate standard.
2.1.3.6 Zero Gas for Assay of Candidate Standards and Zero Air Material—
The zero gas used for the assay of candidate standards need not be the
same zero gas as used for the multipoint calibrations and zero gas checks, but it
must conform to the specifications in Subsection 2.1.3.5. The zero gas used for
the assay of zero air material must be an NIST-traceable zero air material.
Information concerning the zero gas (e.g., cylinder identification number, cylinder
pressure, etc.) must be recorded in the laboratory's records.
12
EPA Traceability Protocol for Gaseous Calibration Standards
The analyzer's zero and span controls may be adjusted before the start of
the multipoint calibration. If a zero or span adjustment is made, allow the
analyzer to stabilize for at least one hour before beginning the multipoint
calibration. The waiting period is necessary because some analyzers'
calibrations drift for a period of time following a zero or span control adjustment.
13
EPA Traceability Protocol for Gaseous Calibration Standards
If the relative differences for the pollutant and diluent flow rates are less than 1.0
percent, the calibrations of the gas dilution system has not drifted significantly
since its last calibration or recertification and the pollutant gas analyzer
calibration can proceed. If the relative differences for either the pollutant flow
rate or the diluent flow rate exceed 1.0 percent, significant drift has occurred and
the device must be recertified before the assay is conducted.
14
EPA Traceability Protocol for Gaseous Calibration Standards
Plot the values from the multipoint calibration and the regression curve
with confidence bands as shown in Figure 2-1. These plots will provide a
graphical representation of the calibration and will permit a qualitative
assessment of the uncertainty associated with the calibration. Record the
regression coefficients, their expanded uncertainties, and other statistical results
from the spreadsheet in the laboratory's records. Enough information should be
recorded so that the analyst can calculate the expanded uncertainty of the
certified concentration of a candidate standard that is assayed at some later date
and to reconstruct these calculations at some later date should questions arise
about the calibration data or the spreadsheet calculations.
Figure 2-1. Example regression curve and confidence bands for multipoint calibration
15
EPA Traceability Protocol for Gaseous Calibration Standards
16
EPA Traceability Protocol for Gaseous Calibration Standards
Note that possibly a more restrictive uncertainty criterion applies for the
assay of the candidate standard. The value of U for the estimated concentration
of the candidate standard must be ≤±1 percent of the concentration of the
reference standard (see Subsections 2.2.2 and 2.3.2). For example, assume
that (1) an analyzer's calibration equation is well characterized from 0 to 100 ppm
and (2) a 70-ppm candidate standard is being assayed at a later date using a 50-
ppm reference standard. Then, U must be ≤±0.5 ppm, rather than ≤±1.0 ppm
which is derived from the multipoint calibration.
17
EPA Traceability Protocol for Gaseous Calibration Standards
The measure of this uncertainty that is most directly useful to the analyst
is the value of U for a regression-predicted concentration given one or more
measurements of the candidate standard. This value may be calculated using
the spreadsheet described in Appendix A or using an EPA-approved equivalent
statistical technique (see Subsection 2.1.6). Several points should be noted
about this value. First, its magnitude decreases as n increases where n is the
number of measurements in the multipoint calibration. Second, its magnitude
decreases as n increases, where n is the number of measurements of the
candidate standard. Third, its magnitude increases as the mean measured
analyzer response ( Y ' ) for the candidate standard diverges from the overall
mean measured analyzer response ( Y ) for the multipoint calibration. These
points mean that it becomes easier to satisfy the uncertainty criterion as one
increases the number of measurements in the multipoint calibration and in the
assay of the candidate standard. Additionally, the absolute uncertainty of the
regression predicted concentration is larger at the extremes of the calibrated
concentration range than at the middle of the range.
18
EPA Traceability Protocol for Gaseous Calibration Standards
If the relative differences for the zero and span gas checks are each less
than or equal to 5.0 percent, the analyzer's current calibration is considered to be
approximately the same as during the multipoint calibration and the assay may
be conducted. The zero and span controls do not have to be adjusted following
the zero and span checks because the data reduction technique used in this
protocol does not depend on the absolute accuracy of the analyzer calibration
equation at the time of the assay.
If the relative differences for the zero or span gas checks are greater than
5.0 percent, the analyzer is considered to be out of calibration. A new multipoint
calibration may be conducted before the candidate standard is assayed or the
analyzer's zero and span controls may be adjusted to return the analyzer's
response to the original calibration levels. For some analyzers such as
nondispersive infrared instruments, daily changes in environmental variables
such as barometric pressure may shift the calibration. After any adjustment of
controls, the analyst should repeat the zero and span gas checks and recalculate
the relative differences to verify that the analyzer is in calibration.
The zero gas and reference standard measurements that are performed
for the assay of the candidate standard may also be used for the zero and span
gas checks.
Between the time of the multipoint calibration and the time of the zero and
span gas checks, the analyst may adjust the analyzer’s zero and span controls
for assays that will not be certified according to this protocol. However, these
controls must be returned to their settings at the multipoint calibration before the
zero and span gas checks or assays under this protocol.
See Subsections 2.2, 2.3, and 2.4 for Procedures G1, G2, and G3,
respectively
19
EPA Traceability Protocol for Gaseous Calibration Standards
Reassay the candidate standard at least 7 days after the first assay.
Make three or more discrete measurements of the candidate standard during the
second assay. Record these measurements in the laboratory's records.
20
EPA Traceability Protocol for Gaseous Calibration Standards
more important than not being able to find that the concentration is unstable.
See Appendix D for a further discussion of TOST calculations with examples.
If the data from the two assays are found to be statistically equivalent and
yet the analyst is concerned that these data suggest potential instability, the
analyst may elect to conduct a third assay to confirm that stability has been
achieved. The analyst must wait an additional 7 days or more and conduct the
third assay and must add the additional data to the Appendix C spreadsheet.
Record the additional data in the laboratory’s records.
21
EPA Traceability Protocol for Gaseous Calibration Standards
The following steps give a simplified procedure for determining the trace
contaminant concentration:
1. Perform monthly multipoint calibrations of the pollutant gas analyzer for the
main component and the trace contaminant using reference standards for
both components and using either Procedure G1 or Procedure G2.
2. On the day of the assay of the candidate standard, perform zero and span
gas checks for the main component as described in Subsection 2.2.6.4 or in
Subsection 2.3.8.4. If the calibration drift for the main component is within the
acceptance criterion (i.e., relative differences for the zero and span gas
checks less than or equal to 5.0 percent), then the analyst may assume that
22
EPA Traceability Protocol for Gaseous Calibration Standards
excessive calibration drift has not occurred for the trace contaminant.
However, if the calibration drift is excessive, then the analyzer is considered
to be out of calibration. Follow the procedures for out-of-calibration analyzers
that are described in Subsection 2.2.6.4 or in Subsection 2.3.8.4.
3. Perform the assays for the trace contaminant during the same assay session
as the assays of the main component of the candidate standard.
Measurement of the trace contaminant reference standard is not necessary.
4. Calculate the concentration of the trace contaminant and its uncertainty using
the spreadsheet described in Appendix A (or an EPA-approved equivalent
statistical technique, see Subsection 2.1.6). Enter data from the trace
contaminant measurements in the spreadsheet cells in Step 5 where
candidate standard data for the same day as the calibration normally would
be entered. Record these values in the laboratory's records. The trace
contaminant concentration cannot be greater than 2.0 percent of the main
component concentration.
23
EPA Traceability Protocol for Gaseous Calibration Standards
Although this protocol does not specify an acceptance criterion for U for
candidate standard concentrations, EPA's Acid Rain Program has specified in 40
CFR Part 75 that an EPA Protocol Gas must have a specialty gas producer-
certified uncertainty (95- percent confidence interval) that must not be greater
than 2.0 percent of the certified concentration (tag value) of the gas mixture5. In
general, an end user's purchase specifications for an EPA Protocol Gas should
include a specification for U that meets its needs and any regulatory
requirements concerning the uncertainty of the certified concentration.
The certified value of U for SRMs, NTRMs, RGMs, PRMs, and CRMs is
expressed on their certificates of analysis as the combined standard uncertainty
(uc) multiplied by a coverage factor (k), which is equal to 2.8,9 The combined
standard uncertainty includes the uncertainties of known sources of systematic
error as well as the random error of measurement. A value of one-half of U for
the reference standard should be used in calculating U of candidate standards
that are certified under this protocol (see Appendix C).
For those cases when the candidate standard is assayed on the same day
as the multipoint calibration, the candidate standard's concentration is
determined directly from the calibration curve. The combined standard
24
EPA Traceability Protocol for Gaseous Calibration Standards
= uc (uMAIN )2 + (uTRACE )2
If an interference-correction equation has been used to obtain a corrected
concentration for the candidate standard, the standard uncertainty for the
corrected concentration (uCORRECTION) must be included in the assessment of the
total analytical uncertainty of the candidate standard's concentration using the
following equation:
=
UC (USTANDARD )2 + (UASSAY )2 + (UCORRECTION )2
If dilution has been used in the assay of the candidate standard, the
standard uncertainty for the dilution (uDILUTION) must be included in the
assessment of the total analytical uncertainty of the candidate standard's
concentration using the following equation:
=
UC (USTANDARD )2 + (UASSAY )2 + (UCORRECTION )2 + (UDILUTION )2
25
EPA Traceability Protocol for Gaseous Calibration Standards
2.1.7 Certification Documentation for Certified Standards and Zero Air Materials
If a zero air material was assayed using Procedure G1 or G2, the certification
documentation must include the impurity concentrations of the assayed
components and a statement to the effect that it conforms to the purity
specifications of 40 CFR Part 72.2. Use the spreadsheet described in
Appendix A or an EPA-approved equivalent statistical technique to calculate
the impurity concentrations, which are the means of all measurements of the
zero air material. The significant figures in the certified impurity
concentrations generally should be the same as U (e.g., 0.01 ppm +/- 0.03
ppm, not 0.012 ppm +/- 0.30 ppm) as is discussed below.
3. A calculated estimate of U for the standard (see Section 2.1.6). The use of
an uncalculated, blanket value for U (e.g., +/- 1 percent) is not acceptable.
The estimate is expressed as a 95-percent confidence interval, which is the
26
EPA Traceability Protocol for Gaseous Calibration Standards
If a zero air material was assayed using Procedure G1 or G2, the certification
documentation must include a calculated estimate of U for the impurity
concentrations in the zero air material using the statistical procedures
described above.
4. The assayed component(s) and balance gas in the gas mixture. If the
composition or origin of the balance gas (e.g, oxygen percentage or synthetic
vs. scrubbed ambient air) has a measurable effect on the end user's pollutant
gas analyzers, it must described in the certification documentation.
5. Cylinder pressure at certification and the statement that the standard should
not be used when the pressure is below 0.7 megapascals (i.e., 100 psig).
6. Dates of the assays and certification. The certification date is the date of the
last assay.
7. Certification expiration date (i.e., the certification date plus the certification
period plus one day) (see Subsection 2.1.9).
8. Information about the reference standard used in the assay: NIST SRM
number, NIST sample number, cylinder identification number, certified
concentration, expanded uncertainty, and certification expiration date for an
SRM; cylinder identification number, certified concentration, expanded
uncertainty, and certification expiration date for an NTRM, an RGM, a PRM, a
CRM or a GMIS. The certification documentation must identify the reference
standard as being an SRM, an NTRM, an RGM, a PRM, a CRM or a GMIS.
For a candidate standard that is assayed using a GMIS, the certification
documentation must include information about the reference standard that
was used for the GMIS' assay (see above).
27
EPA Traceability Protocol for Gaseous Calibration Standards
10. The analytical method that was used in the assay and the date of the most
recent multipoint calibration of the instrument.
11. Identification of the specialty gas producer or other laboratory (i.e., producer's
or laboratory's name, city, and state, and PGVP vendor ID issued by EPA for
the production location) where the candidate standard was assayed, even if
another organization prepared the candidate standard or will sell it to an end
user. This identification must be given in the same or larger font as the other
required information in the report.
13. If applicable, a statement that the certified concentration for one component
has been corrected for interferences from other specified components.
2. If a zero air material was assayed using Procedure G3, use the spreadsheet
described in Appendix E or an EPA-approved equivalent statistical technique
(see Subsection 2.1.6) to calculate whether the purity specifications of 40
CFR Part 72.2 have been attained. If the specifications have been attained,
the certification documentation must include a statement to this effect.
28
EPA Traceability Protocol for Gaseous Calibration Standards
3. If the zero air material was assayed using Procedure G3, no uncertainty
estimate need be calculated or documented.
5. Information about the NIST-traceable zero air material used in the assay:
cylinder identification number, certified concentration, expanded uncertainty,
and certification expiration date. The certification documentation must identify
the NIST-traceable zero air material as being an SRM, an NTRM, an RGM, a
PRM, a CRM or a GMIS. For a GMIS, the NIST-traceable zero air material
that was used for its assay must be identified in the documentation for the
zero air material being certified.
8. The analytical method that was used in the assay and the date of the most
recent multipoint calibration of the instrument.
10. If the composition or origin of the balance gas (e.g, oxygen percentage or
synthetic vs. scrubbed ambient air) has a measurable effect on the end user's
pollutant gas analyzers, it must described in the certification documentation.
This certification documentation must be given to the end user of the zero
air material. The specialty gas producer or other laboratory that assayed the
standard must maintain laboratory records and certification documentation for at
least the standard's certification period or until such time as the standard is blown
down after being returned by the end user to the producer. A specialty gas
producer or other vendor may redocument a zero air material that it has
purchased from another specialty gas producer and that it wishes to sell to a third
party. However, the new certification documentation must clearly identify the
29
EPA Traceability Protocol for Gaseous Calibration Standards
specialty gas producer or other laboratory (i.e., name and location) where the
zero air material was assayed and present all the information that is contained in
the original report.
Specialty gas producers may elect to certify candidate standards for less
than the maximum certification period in Table 2-3 if they believe that they cannot
prepare standards whose stability attains the maximum certification period. The
default certification period is that given in Table 2-3. Each producer has
discretion in this matter.
30
EPA Traceability Protocol for Gaseous Calibration Standards
31
EPA Traceability Protocol for Gaseous Calibration Standards
32
EPA Traceability Protocol for Gaseous Calibration Standards
If the TOST acceptance criterion is still not attained after the second
recertification assay, the analyst must either disqualify the standard for further
use under this protocol or investigate why there is an apparent difference
between the original assays and the recertification assay. This difference may
be due to an actual instability of the gas mixture, to a reference standard
problem, to an analytical instrumentation problem, or to some other problem. If
the analyst can find a reasonable explanation for the difference and if this reason
is not instability, then the standard can be recertified. The analyst must append a
brief report on the results of the investigation to the recertification documentation
and to the laboratory's records.
Standards having certified concentrations that are lower than those given
in Table 2-3 may be recertified for the period given in Table 2-3 provided at least
6 months have elapsed between the initial certification and the recertification.
For example, a 0.5-ppm sulfur dioxide in nitrogen standard will have an initial
certification period of 6 months. After a successful recertification, this standard
will have a recertification period of 4 years. The certification date is the date of
the last assay.
33
EPA Traceability Protocol for Gaseous Calibration Standards
A certified standard should not be used when its gas pressure is below
0.7 megapascals (i.e., 100 psig). NIST has found that some gas mixtures (e.g.,
nitric oxide in nitrogen) have exhibited a concentration change when the cylinder
pressure fell below this value. There is no minimum cylinder pressure
specification for zero air materials.
Periodically, the U.S. EPA will assess the accuracy of a blind sample of
compressed gas calibration standards that have been assayed and certified
according to this protocol. The analytical results, identifying the specialty gas
producers or other analytical laboratories that assayed and certified the
standards, will be published as public information.
Either directly or through third parties, EPA procured EPA Protocol Gases
from the producers, assessed the accuracy of the gases' certified concentrations
through independent analyses, and inspected the accompanying certificates of
analysis for completeness and accuracy. The producers were not aware that
EPA had procured the gases for these audits.
In 2009, the Office of the Inspector General (OIG) published the report
EPA Needs an Oversight Program for Protocol Gases35. OIG recommended
that the Office of Air and Radiation (OAR) implement oversight programs to
34
EPA Traceability Protocol for Gaseous Calibration Standards
assure the quality of EPA Protocol Gases used to calibrate continuous emissions
monitoring systems and ambient air monitors. It also recommended that EPA's
Office of Research and Development (ORD) update and maintain this traceability
protocol to ensure that the monitoring programs' objectives are met.
To address the OIG findings for ambient air monitoring, EPA's Office of
Air Quality Planning and Standards (OAQPS), in cooperation with EPA Regions
2 and 7 developed an Ambient Air Protocol Gas Verification Program (AA-
PGVP). This program establishes gas metrology laboratories in Regions 2 and 7
to verify the certified concentrations of EPA Protocol Gases used to calibrate
ambient air quality monitors. It is expected to:
• ensure that producers selling EPA Protocol Gases participate in the AA-
PGVP, and
• provide end users with information about participating producers and
verification results.
"2.6 Gaseous and Flow Rate Audit Standards. Gaseous pollutant concentration standards
(permeation devices or cylinders of compressed gas) used to obtain test concentrations for
CO, SO2, NO, and NO2 must be traceable to either a National Institute of Standards and
Technology (NIST) Traceable Reference Material (NTRM), NIST Standard Reference
Materials (SRM) and Netherlands Measurement Institute (NMi) Primary Reference Materials
(valid as covered by Joint Declaration of Equivalence) or a NIST-certified Gas Manufacturer's
Internal Standard (GMIS), certified in accordance with one of the procedures given in
reference 4 of this appendix. Vendors advertising certification with the procedures provided in
reference 4 of this appendix and distributing gases as 'EPA Protocol Gas' must participate in
the EPA Protocol Gas Verification Program or not use 'EPA' in any form of advertising."
The results from the first two years of the AA-PGVP have been
published36. More information about the AA-PGVP can be found at OAQPS' AA-
PGVP website, www.epa.gov/ttnamti1/aapgvp.html.
In a parallel effort, EPA's Office of Air and Radiation (OAR) Clean Air
Markets Division (CAMD) conducted assessments in 2003 and 2010 of a blind
sample of EPA Protocol Gases containing CO2, NO, and SO2 that are used to
calibrate continuous emission monitors under the Acid Rain Program37. EPA has
35
EPA Traceability Protocol for Gaseous Calibration Standards
"On and after May 27, 2011 for each unit subject to this part that uses EPA Protocol gases,
the owner or operator must obtain such gases from either an EPA Protocol gas production
site that is on the EPA list of sites participating in the PGVP on the date the owner or operator
procures such gases or from a reseller that sells to the owner or operator unaltered EPA
Protocol gases produced by an EPA Protocol gas production site that was on the EPA list of
participating sites on the date the reseller procured such gases."
36
EPA Traceability Protocol for Gaseous Calibration Standards
This procedure may be used for the assay of more than one candidate
standard during the same assay session. Criteria that apply to the assay of one
candidate standard apply to the assay of multiple candidate standards.
2.2.2 Limitations
The balance gas must be the same in both the candidate standard and the
reference standard, unless it has been demonstrated that the analyzer's
response is insensitive to differences in the balance gas composition.
Figure 2-2 illustrates one possible design of the apparatus for the assay of
compressed gas calibration standards without dilution. This apparatus is
designed to allow the convenient routing of the gas mixtures to the pollutant gas
analyzer. Inert materials (e.g., Teflon®, stainless steel, borosilicate glass or
37
EPA Traceability Protocol for Gaseous Calibration Standards
Figure 2-2. Apparatus for the assay of calibration standards without dilution
The gas flow controllers may be needle values, capillary tubes, thermal mass
flow controllers, or other flow control devices. The gas mixtures are routed to the
analyzer through a union tee tube fitting. Gas in excess of the analyzer's
demand is vented, which helps to ensure that the gas entering the analyzer is at
near-ambient pressure. Normally, the excess gas is vented to the atmosphere
without any obstructions in the tubing. However, the excess gas can be routed
through an uncalibrated rotameter by rotation of a three-way valve (i.e., V3). The
38
EPA Traceability Protocol for Gaseous Calibration Standards
rotameter is used to demonstrate that the total gas flow rate exceeds the sample
flow rate of the analyzer and that no room air is being drawn in through the vent
line.
The apparatus may be modified in several ways that will not diminish its
performance. For example, the two three-way valves could be replaced by
solenoid valves, by a single four-way valve with three input ports and one output
port or by a rotating multiposition step valve. See also Subsection 2.1.2
concerning the acceptability of alternative designs for the assay apparatus.
If the analyzer has not been in continuous operation, turn it on and allow it
to stabilize (e.g., for at least 12 hours) before beginning the measurements.
39
EPA Traceability Protocol for Gaseous Calibration Standards
2.2.6.3 Linearity—
The data reduction technique used in this procedure requires that the
analyzer have a well-characterized, but not necessarily linear, calibration curve
(see Subsection 2.1.4.2 concerning the well-characterized calibration curve).
High-concentration-range analyzers of the type that are required for this
procedure may not be inherently linear, but in such cases they usually have a
predictable, non-linear calibration curve that can be described by a polynomial
equation or can be mathematically transformed to produce a straight-line
calibration curve that is suitable for use in this procedure (see Subsection 2.1.4.3
40
EPA Traceability Protocol for Gaseous Calibration Standards
The gas mixtures to be used during the zero and span gas checks need
not be, but can be, the reference standards used for the assay of the candidate
standard or for the multipoint calibration. The reference standard for the span
gas check must be traceable to an SRM, an RGM, a PRM, a CRM, or an NTRM.
Information concerning this standard (e.g., cylinder identification number,
certified concentration, expanded uncertainty, certification expiration date, etc.)
must be recorded in the laboratory's records. The zero gas must meet the
requirements in Subsection 2.1.3.4. The use of NIST-traceable zero air material
for the zero gas check is recommended, but not required. The analyst may
substitute a low-concentration, NIST-traceable reference standard for the zero
gas if there is reason to believe that the zero gas reading may not accurately
represent the zero-intercept of the calibration equation.
Make three or more discrete measurements of the zero gas and three or
more discrete measurements of the span gas reference standard. "Discrete"
means that the analyst must change the gas mixture being sampled by the
analyzer between measurements. For example, the analyst might alternate
between measurements of the reference standard and measurements of the zero
gas. Record these measurements in the laboratory's records.
s R
≤ RS
n 100
41
EPA Traceability Protocol for Gaseous Calibration Standards
where
s = standard deviation of the analyzer's response;
n = the number of measurements of the gas mixture; and
RRS = the mean analyzer response to the reference standard.
The value of the standard error of the mean can be made smaller by
increasing the number of measurements. This calculation will enable the analyst
to determine how many replicate measurements are needed during the assay of
the candidate standard to obtain acceptable precision. The analyst may wish to
use a data logger or data acquisition system with averaging capability to obtain
more precise measurements. If the value of the standard error is not acceptable,
then the analyzer must be repaired or another analyzer must be used.
Next, verify that excessive zero drift has not occurred since the multipoint
calibration. For the zero gas measurements, calculate the relative difference (in
percent) between the current mean analyzer response during the zero gas check
and the corresponding response that is predicted from the multipoint calibration
regression equation. That is,
Repeat this relative difference calculation for the span gas reference
standard measurements. That is,
Then, if the relative differences for the zero and span gas checks are less
than or equal to 5.0 percent, the analyzer's current calibration is considered to be
approximately the same as during the multipoint calibration and the assay may
be conducted. The zero and span controls need not be adjusted and the assay
may be conducted. The data reduction technique does not require that some
calibration drift has not occurred since the multipoint calibration. Some minor
calibration drift is acceptable because the effect of any drift will be corrected
during the reduction of the assay data.
42
EPA Traceability Protocol for Gaseous Calibration Standards
However, if the relative difference for either the zero or the span gas
check is greater than 5.0 percent, then the analyzer is considered to be out of
calibration. A new multipoint calibration may be conducted before the candidate
standard is assayed or the analyzer's zero and span controls may be adjusted to
return the analyzer's response to the original calibration levels. For some
analyzers such as nondispersive infrared instruments, daily changes in
environmental variables such as barometric pressure may shift the calibration.
After any adjustment of controls, the analyst should repeat the zero and span gas
checks and recalculate the relative differences to verify that the analyzer is
sufficiently in calibration. The analyzer will be considered to be out of calibration
if the relative differences remain greater than 5.0 percent.
2. Verify that a multipoint calibration has been performed within 1 month prior to
the assay date and that it demonstrates acceptable instrument performance
(see Subsection 2.1.4.2). Additionally, verify that the zero and span gas
checks indicate that the analyzer is in calibration (see Subsection 2.2.6.4).
Finally, verify that the concentrations of the candidate and reference
standards fall within the well-characterized region of the analyzer's calibration
curve (see Subsection 2.1.4.2).
3. Measure and adjust the flow rates of the gas mixtures (i.e., reference
standard(s), candidate standard, and zero gas) to approximately the same
value that will provide enough flow for the analyzer and sufficient excess to
assure that ambient air will not be drawn into the vent line.
4. In succession, measure the zero gas, the reference standard(s), and the
candidate standard(s) using the analyzer. Use valves V1 and V2 to select
each of the gas mixtures for measurement. Allow sufficient time between
measurements to completely purge the analyzer of the previous gas and for
the analyzer to achieve a stable reading. If the response for each
measurement is not stable, the precision of the measurements will decline
and the candidate standard may not be certifiable under this protocol. Record
the analyzer response for each measurement in the laboratory's records,
using the same response units (e.g., volts, area counts, etc.) as was used for
the multipoint calibration. At this point, do not convert these data into
concentration values using the calibration equation or perform any
mathematical transformations (see Subsection 2.1.4.3) if the data were so
transformed. These steps will be done later. Also, do not make any zero
43
EPA Traceability Protocol for Gaseous Calibration Standards
The analyst may assay multiple candidate standards during the same assay
session. For example, a single set of measurements may involve a zero gas,
a reference standard and three candidate standards. Criteria that apply to the
assay of one candidate standard apply to the assay of multiple candidate
standards. The analyst should be aware that the effect of any short-term
calibration drift will be greater when multiple candidate standards are
assayed. This greater effect is due to the longer period of time between
reference standard measurements. Unacceptable uncertainties of the
estimated concentrations for the candidate standards may occur as a result of
the longer assay session.
If the assay date is different from the calibration date and if there is no
statistical difference between the two sets of data (i.e., the calibration data
44
EPA Traceability Protocol for Gaseous Calibration Standards
and the zero and span gas check data), entering both sets of data in the Xi
and Yi columns in Step 1 of the Appendix A spreadsheet will produce a
smaller expanded uncertainty for the estimated concentration than if only the
calibration data were entered in these columns. However, a larger expanded
uncertainty will be produced if there is a statistical difference between the two
sets of data and if both are entered in these columns. The analyst should
check which approach yields the smaller value for the expanded uncertainty.
8. If the multipoint calibration data and the assay data underwent any
mathematical transformations before their statistical analysis, perform the
reverse transformations for the estimated concentration and the expanded
uncertainty. Record the transformed values in the laboratory’s records.
45
EPA Traceability Protocol for Gaseous Calibration Standards
Dilution of the candidate and reference standards with zero gas allows
greater flexibility in the range of concentrations of both the candidate and
reference standards that can be assayed. The procedure employs a low-
concentration-range (i.e., ambient air quality level) pollutant gas analyzer to
compare quantitatively diluted gas samples of both the candidate and reference
standards. This analyzer is more likely to have an inherently linear response
than a high-concentration-range analyzer. However, dilution introduces
additional error into the assay which would not be present if the standards were
assayed without dilution. This additional error is measured by an accuracy check
of the assay apparatus which is performed as part of the multipoint calibration.
This procedure may be used for the assay of multiple candidate standards
at the same time. Criteria that apply to the assay of one candidate standard
apply to the assay of multiple candidate standards.
2.3.2 Limitations
This procedure is not intended to be used for the assay and certification of
a GMIS that could be used as an analytical reference material in the assay of
candidate standards. Procedure G1 should be used for GMISs.
47
EPA Traceability Protocol for Gaseous Calibration Standards
ppm. The expanded uncertainty for the example candidate standard’s estimated
concentration must be ≤±0.5 ppm.
The balance gas in both the candidate and reference standards must be
identical, unless either a high dilution flow rate ratio (i.e., at least 50 parts zero
gas to 1 part standard) is used for the assay or it has been demonstrated that the
analyzer is insensitive to differences in the balance gas.
48
EPA Traceability Protocol for Gaseous Calibration Standards
Figure 2-3 Two designs for assay apparatus with dilution (Procedure G2)
49
EPA Traceability Protocol for Gaseous Calibration Standards
In the upper diagram in Figure 2-3, either zero gas or a diluted standard
can be routed to the analyzer by rotation of three three-way values (i.e., V1, V2,
and V3). One gas flow controller (i.e., C1) regulates the flow rates of the
reference and candidate standards. These flow rates can be measured by a
single flowmeter connected to an outlet port on valve V2 or by a flowmeter built
into C1. Another gas flow controller (i.e., C2) regulates the flow rate of the zero
gas. This flow rate can be measured by a flowmeter connected to an outlet port
on valve V3 or by a flowmeter built into C2. The gas flow controllers may be
needle valves, capillary tubes, thermal mass flow controllers, or other suitable
devices (see Subsection 2.3.4). If different flow rates are used for the reference
and candidate standards during the assay (see Subsection 2.3.5), separate gas
flow controllers may be used for the two standards. However, the same
flowmeter must be used to measure the two flow rates to minimize error in the
measurement (see Subsection 2.3.4). Flow rates should be controlled and
measured with a relative uncertainty of 1 percent or less. For large dilutions of
the standards, the reference and candidate standard flow rates may be quite
small. Therefore, the internal volume of the tubing and components should be
kept small to minimize the flushing time when valve V1 is rotated. The mixing
chamber combines the two gas streams and should be designed to produce
turbulent flow to ensure thorough mixing of the gas streams. The diluted gas
mixtures are routed to the analyzer through a union tee tube fitting, which vents
excess gas flow. Normally, the excess gas is vented to the atmosphere without
any obstructions in the tubing and the gas entering the analyzer is at near-
atmospheric pressure. However, the excess gas can be routed through an
uncalibrated rotameter by rotation of a three-way valve (i.e., V4). The rotameter
is used to demonstrate that the total gas flow rate exceeds the sample flow rate
of the analyzer and that no room air is being drawn in through the vent line.
The assay apparatus may be modified in several ways that will not
diminish its performance. For example, the three-way valves could be replaced
by solenoid valves. Alternatively, valve V1 could be replaced by a single length
of tubing that is connected manually to the two standards' pressure regulators in
succession (see also Subsection 2.1.2 concerning the acceptability of alternative
designs for the assay apparatus.).
In the lower diagram in Figure 2-3, the reference and candidate standards
are diluted with a gas dilution system. This gas dilution system may use capillary
tubes, positive-displacement pumps, thermal mass flow controllers, or other
suitable devices to dilute the standards. If a gas dilution system is used, it must
have a specified accuracy of not greater than 1.0 percent of the undiluted
reference standard concentration. The analyst must check the accuracy of the
gas dilution system during the multipoint calibration by measurement of a check
standard (see Subsections 2.1.4.2 and 2.3.8.1).
50
EPA Traceability Protocol for Gaseous Calibration Standards
The gas dilution system used in this procedure need not be certified as
NIST-traceable using the procedures given in Section 4 because this procedure
is adequate for establishing the traceability of candidate standards using dilution.
2.3.4 Flowmeters and Gas Dilution Systems for Dilution of Gas Mixtures
The reference standards' certifications must cover the range of flow rates
that will be used by the gas dilution systems and their uncertainties of the flow
rates must be documented in a calibration certificate. Their uncertainties must
be evaluated and expressed according to accepted metrological practice (see
Reference 9) and must not be greater than plus or minus 1 percent of set-point
for the flow rates that will be used for gas dilution. Uncertainties that are much
less than plus or minus 1 percent of reading are desirable so that the
uncertainties of the gas mixture concentrations are decreased. They must be
stable, repeatable, and linear and have good resolution.
Figure 2-2 shows flow measurement ports on valves V2 and V3. In this
configuration, a single flowmeter can be used to measure both the standard flow
rate and the zero gas flow rate. Such an approach would reduce measurement
errors arising from differences in the calibration of multiple flowmeters.
Alternatively, the reference standard and zero gas flow rates can be measured
separately at the outlet of the mixing chamber or the outlet of the gas dilution
system. In either case, a NIST-traceable volumetric flowmeter such as a wet test
meter, a thermal mass flowmeter, or a soap bubble flowmeter can be used (see
Subsection 2.1.3). Each flow rate must be measured separately while the other
51
EPA Traceability Protocol for Gaseous Calibration Standards
flow rates are set to zero. Care must be exercised to ensure that each measured
flow rate remains constant when combined with the other flow rate(s) and
between the time of measurement and the time of the assay. Additionally, care
must be taken to ensure that the flowmeter does not cause any back pressure in
the gas stream and any resulting change in the flow rate through the flow
controller.
Volume PM − PWV TS
Flow rate =
Time PS TM
where
If using a flowmeter that does not use water or a soap bubble for measurement
the term PWV can be ignored.
52
EPA Traceability Protocol for Gaseous Calibration Standards
same dilution ratio can be used for both the reference standard and candidate
standard measurements.
The flow rates or settings used for the zero gas, reference standard, and
candidate standard should be selected carefully to provide diluted concentrations
for both the candidate and reference standards that fall in the well-characterized
region of the analyzer's calibration curve. The diluted concentration of the
candidate standard may be greater than or lesser than the diluted concentration
of the reference standard. Any assay error due to the dilution process will be
reduced if the same dilution flow-rate ratio or concentration setting can be used
for both the candidate and reference standards. Select the diluted
concentrations of the reference and candidate standards, and select flow rates or
concentration settings that will produce the highest analyzer responses within the
well-characterized region of the analyzer's calibration curve. If possible, select
flow rates or a flowmeter range such that the flow rates to be measured fall in the
upper half of the range.
53
EPA Traceability Protocol for Gaseous Calibration Standards
candidate and reference standards contain dissimilar balance gases (e.g., air
versus nitrogen or different proportions of oxygen in the balance air), either a
high dilution flow-rate ratio (i.e., at least 50 parts zero gas to 1 part standard)
should be used or it must have been demonstrated that the analyzer's response
is not sensitive to differences in the balance gas composition. This
demonstration may be accomplished by showing that no difference exists in an
analyzer's response when measuring a compressed gas calibration standard that
has been diluted with identical flow rates of different balance gases.
If the analyzer has not been in continuous operation, turn it on and allow it
to stabilize (e.g., for at least 12 hours) before beginning any measurements.
If the multipoint calibration data have been fitted to a linear (i.e., straight-
line) model, then only a single reference standard need be measured during the
assay of the candidate standard. If these data have been fitted to a quadratic or
higher-order polynomial model, then at least two reference standards must be
measured. One reference standard is adequate to determine the slope of a
linear equation, but additional reference standards are needed to determine the
curvature of quadratic or higher-order polynomial equations. The concentrations
of the additional reference standards should be located at the maximum
54
EPA Traceability Protocol for Gaseous Calibration Standards
difference between the polynomial curve and the corresponding straight line
between the zero gas and the highest-concentration reference standard.
Next the analyst must verify that the dilution error is not excessive. For
the check standard measurements, calculate the relative difference (in percent)
between the mean analyzer response to the undiluted or diluted check standard
and the corresponding response that is predicted from the multipoint calibration
55
EPA Traceability Protocol for Gaseous Calibration Standards
If the relative difference is greater than 1.0 percent, the dilution error is
considered to be excessive. The analyst must investigate why the relative
difference is excessive. The problem may be due to errors in the reference
standard and check standard concentrations, errors in assay apparatus or to
some other source. Assays may not be conducted until the relative difference for
a subsequent accuracy check is less than or equal to 1.0 percent.
2.3.8.3 Linearity—
The data reduction technique used in this procedure requires that the
analyzer have a well-characterized, but not necessarily linear, calibration curve
(see Subsection 2.1.4.2). Many lower-concentration analyzers of the type that
may be used for this procedure have straight-line calibration curves. If not, they
usually have a predictable nonlinear calibration curve that can be described by a
polynomial equation or can be mathematically transformed to produce a straight-
line calibration curve suitable for use in this procedure. Any such polynomial
equation or mathematical transformation should be verified during the multipoint
calibration. Caution should be exercised in using a transformed calibration curve
because zero or span control adjustments to the analyzer may produce
unexpected effects in the transformed calibration curve.
56
EPA Traceability Protocol for Gaseous Calibration Standards
The gas mixtures to be used during the zero and span gas checks need
not be the same as any of the reference standards used for the assay of the
diluted candidate standard or for the multipoint calibration. The reference
standard for the span gas check must be traceable to an SRM, an RGM, a PRM,
a CRM, or an NTRM. The zero gas must meet the requirements in Subsection
2.1.3.4. The use of NIST-traceable zero air material for the zero gas check is
recommended, but not required. Information concerning these standards (e.g.,
cylinder identification number, certified concentration, uncertainty, certification
expiration date, etc.) must be recorded in the laboratory's records.
Make three or more discrete measurements of the zero gas and three or
more independent measurements of the diluted reference standard. Record
these measurements in the laboratory's records.
Next, the analyst must verify that the analyzer's precision is acceptable.
Calculate the mean and standard deviation of the analyzer's response to the zero
gas. Repeat these calculations for the diluted reference standard
measurements. These calculations are performed in Step 6 of the spreadsheet
described in Appendix A. Record these calculations in the laboratory's records.
The standard error of the mean for each set of measurements must be less than
or equal to 1.0 percent of the mean response to the diluted reference standard.
That is,
S R
≤ DRS
n 100
where
The value of the standard error of the mean can be made smaller by
increasing the number of measurements. This calculation will enable the analyst
to determine how many replicate measurements are needed during the assay of
the diluted candidate standard to obtain acceptable precision. The analyst may
wish to use a data logger or data acquisition system with data averaging
capability to obtain more precise measurements. If the value of the standard
error of the mean is not acceptable, then the analyzer must be repaired or
another analyzer must be used for the assay.
Next the analyst must verify that excessive calibration drift has not
occurred since the multipoint calibration. For the zero gas measurements,
calculate the relative difference (in percent) between the mean analyzer
response during the zero gas check and the corresponding response that is
predicted from the multipoint calibration regression equation. That is,
57
EPA Traceability Protocol for Gaseous Calibration Standards
Then, if the relative differences for the zero and span checks are less than
or equal to 5.0 percent, the analyzer's current calibration is considered to be
approximately the same as during the multipoint calibration and the assay may
be conducted. The zero and span controls need not be adjusted and the assay
may be conducted. The data reduction technique does not require that some
calibration drift has not occurred since the multipoint calibration. Some minor
calibration drift is acceptable because the drift will be corrected for during the
reduction of the assay data.
However, if the relative difference for either the zero or the span gas
checks is greater than 5.0 percent, then the analyzer is considered to be out of
calibration. A new multipoint calibration may be conducted before the candidate
standard is assayed or the analyzer's zero and span controls may be adjusted to
return the analyzer's response to the original calibration levels. For some
analyzers such as nondispersive infrared instruments, daily changes in
environmental variables such as barometric pressure may shift the calibration.
After any adjustment of the controls, the analyst should repeat the zero and span
gas checks and recalculate the relative differences to verify that the analyzer is
sufficiently in calibration. The analyzer will be considered to be out of calibration
if the relative differences remain greater than 5.0 percent.
The zero gas and diluted reference standard measurements that are
performed for the assay of the diluted candidate standard may also be used for
the zero gas and span gas checks.
58
EPA Traceability Protocol for Gaseous Calibration Standards
1. Verify that the assay apparatus is properly configured as shown in Figure 2-3
or Figure 2-4 and as described in Subsection 2.3.3. Inspect the analyzer to
verify that it appears to be operating normally and that all controls are set to
their expected values. Record these values in the laboratory's records.
2. Verify that the calibrations of the flow rate/control reference standards or gas
dilution systems have not drifted significantly since their last calibration or
recertification. This verification need not be done more frequently than
monthly, but that it must have been done within a month before the assay.
Use an NIST-traceable flow rate/control reference standard to check at least
one flow rate setting for each pollutant and dilution gas stream in the assay
apparatus. Record the indicated and actual flow rates in the laboratory's
records. Calculate the relative difference (in percent) between the indicated
and actual volumes or flow rates. That is,
If the relative differences for the pollutant and diluent flow rates are less than
1.0 percent, the calibrations of the flow rate/control reference standards or
gas dilution systems have not drifted significantly since their last calibration or
recertification. The assay can proceed. If the relative differences for either
the pollutant flow rate or the diluent flow rate exceed than 1.0 percent, then
significant drift has occurred and the device should be recertified before the
assay is conducted.
3. Verify that a multipoint calibration of the analyzer has been performed within 1
month prior to the assay date and that it demonstrates acceptable instrument
performance and that the dilution error that is not excessive (see Subsections
2.1.4.2, 2.3.6, and 2.3.8.1). Additionally, verify that the zero and span gas
checks indicate that the analyzer is in calibration (see Subsection 2.3.8.4).
Finally, verify that the concentrations of the diluted reference and candidate
standards fall within the well-characterized region of the analyzer's calibration
curve (see Subsection 2.1.4.2).
4. Determine and establish the flow rates or concentration settings of the gas
mixtures (i.e., reference standard(s), candidate standard, and zero gas) that
will be used for the assay (see Subsections 2.3.4, 2.3.5, and 2.3.8.2). Also
check that the total flow rate coming from the mixing chamber will provide
enough flow for the analyzer and sufficient excess to ensure that no ambient
air will be drawn into the vent line. Changes in the sample pressure may
change the calibration curve. When using the same flow rates for both
candidate and reference standards, carefully set the delivery pressures of the
59
EPA Traceability Protocol for Gaseous Calibration Standards
Record the measured flow rates and the undiluted and diluted reference
standard concentrations in the laboratory's records.
5. In succession, measure the zero gas, the diluted reference standard(s) and
the diluted candidate standard using the analyzer. For each measurement,
adjust the flow rates, if necessary, to those determined in Step 4, and allow
sufficient time between measurements to completely purge the analyzer of
the previous gas and for the analyzer to achieve a stable reading. If the
reading is not stable, the precision of the measurements will decline and the
candidate standard might not be certifiable under this protocol. Record the
analyzer response for each measurement, using the same response units
(e.g., volts, area counts, etc.) as was used for the multipoint calibration. At
this point, do not convert the data into concentration values using the
calibration equation. Do not perform any mathematical transformations of the
data. These steps will be done later. Do not make any zero control, span
control, or other adjustments to the analyzer during this set of measurements.
Record these analyzer responses in the laboratory's records.
The analyst may assay multiple candidate standards during the same assay
session. For example, a single set of measurements may involve a zero gas,
a diluted reference standard, and three diluted candidate standards. Criteria
that apply to the assay of one candidate standard apply to the assay of
multiple candidate standards. The analyst should be aware that the effect of
any short-term calibration drift will be greater when multiple candidate
standards are assayed. This greater effect is due to the longer period of time
between reference standard measurements. Unacceptable uncertainties of
the estimated concentrations for the diluted candidate standards may occur
as a result of the longer assay session.
60
EPA Traceability Protocol for Gaseous Calibration Standards
analysis of the multipoint calibration data such that the expanded uncertainty
for the regression-predicted concentration of the candidate standard is ≤1
percent of the concentration of the reference standard.
If the assay date is different from the calibration date and if there is no
statistical difference between the two sets of data (i.e., the calibration data
and the zero and span gas check data), entering both sets of data in the Xi
and Yi columns in Step 1 of the Appendix A spreadsheet will produce a
smaller expanded uncertainty for the estimated concentration than if only the
calibration data were entered in these columns. However, a larger expanded
uncertainty will be produced if there is a statistical difference between the two
sets of data and if both are entered in these columns. The analyst should
check which approach yields the smaller value for the expanded uncertainty.
The uncertainty associated with the dilution of the reference standard and the
candidate standard needs to be included in the calculation of the expanded
uncertainty in those cases in which the gas dilution apparatus' settings are
different for the reference standard than for the candidate standard.
However, this uncertainty component does not have to be included in the
calculation of the expanded uncertainty if exactly the same settings for the
gas dilution apparatus are used for the analysis of the reference standard and
the candidate standard. In this case, no variability in the results is introduced
by using a constant dilution setting.
61
EPA Traceability Protocol for Gaseous Calibration Standards
9. If the multipoint calibration data and the assay data underwent any
mathematical transformations before their statistical analysis, perform the
reverse transformations for the estimated concentration and the expanded
uncertainty. Record the transformed values in the laboratory’s records.
Certified Undiluted Conc. = (Mean Diluted Conc) (Total Gas Flow Rate)
(Standard Gas Flow Rate)
where Total Gas Flow Rate = Standard Gas Flow Rate + Zero Gas Flow Rate
62
EPA Traceability Protocol for Gaseous Calibration Standards
63
EPA Traceability Protocol for Gaseous Calibration Standards
This procedure may be used for the assay of more than one candidate
during the same assay session. Criteria that apply to the assay of one candidate
apply to the assay of multiple candidates.
At this time, EPA does not require the regulated community to use NIST-
traceable zero air materials for the calibration of ambient air or continuous
emission monitors that are required by 40 CFR Parts 50, 58, 60, and 75.
However, end users may elect to purchase these standards from specialty gas
producers who elect to assay and certify them.
2.4.2 Limitations
This procedure is not intended to be used for the assay and certification of
the impurity concentrations of bulk zero air material such as could be used to fill
individual cylinders that might be sold as NIST-traceable zero air materials.
Rather, it is intended to be used for the assay of the impurity concentrations of
candidate zero air materials in individual cylinders because the concentrations in
these cylinders may differ from those in the bulk gas because of residual
impurities that were in the cylinders before filling.
This procedure is not intended to be used for the assay and certification of
a zero air GMIS that could be used as a reference standard in the assay of
candidate zero air materials. Procedure G1 should be used for the assay and
certification of a candidate zero air GMIS.
Figure 2-2 illustrates one possible design of apparatus for the assay of
candidate zero air materials without dilution. See Subsections 2.1.3 and 2.2.3 for
a discussion of this apparatus. The apparatus may be modified in several ways
that will not diminish its performance.
See Section 2.2.4 for the requirements for the pollution gas analyzer. Use
the most sensitive operating range that allows measurement of the zero gas, the
candidate zero air material, the upscale reference standard, and the reference
standards used in the multipoint calibration. The detection limit for the analyzer
64
EPA Traceability Protocol for Gaseous Calibration Standards
must be less than the go/no-go acceptance criterion. If the analyzer has not
been in continuous operation, turn it on and allow it to stabilize (e.g., for at least
12 hours) before beginning any measurements.
Note that the uncertainty of the calibration curve may comprise a larger
fraction of the expanded uncertainty for a candidate zero air material than for a
candidate standard whose concentration is near the upper end of the analyzer's
range. The inclusion of more measurements or more concentrations in the
multipoint calibration will reduce the uncertainty of the calibration curve, which
may allow a clearer determination of whether a candidate zero air material
conforms to the impurity specifications of Part 72.2.
2.4.5.3 Linearity—
See Subsection 2.3.5.3 for a discussion of the linearity of the pollution gas
analyzer's calibration curve.
65
EPA Traceability Protocol for Gaseous Calibration Standards
with an NIST-traceable zero air material for the assay of the candidate zero air
material. Information concerning the standard (e.g., cylinder identification
number, certified concentration, uncertainty, certification expiration date, etc.)
must be recorded in the laboratory's records.
If the multipoint calibration data have been fitted to a linear (i.e., straight-
line) regression model, then only a single upscale reference standard need be
measured during the assay of the candidate zero air material. If these data have
been fitted to a quadratic or higher-order regression model, then at least two
upscale reference standards must be measured during the assay. One upscale
reference standard is adequate to determine the slope of a linear equation, but
additional upscale reference standards are needed to determine the curvature of
quadratic and higher-order polynomial equations. The concentrations of the
additional upscale reference standards should be located at the maximum
difference between the polynomial curve and the corresponding straight line
between the zero gas and the highest-concentration upscale reference standard.
2. Verify that a multipoint calibration has been performed within 1 month prior to
the assay date and that demonstrates acceptable instrument performance
(see Subsections 2.1.4.2 and 2.1.4.3). Additionally, verify that the
concentrations of the candidate zero air material and the reference standards
fall within the well-characterized region of the analyzer's calibration curve.
3. Measure and adjust the flow rates of the gas mixtures (i.e., upscale reference
standard(s), candidate zero air material, and zero gas) to approximately the
same value that will provide enough flow for the analyzer and sufficient
excess to assure that no ambient air will be drawn into the vent line.
4. Make three or more discrete measurements of the zero gas, three or more
discrete measurements of the upscale reference standard, and at least one
measurement of the candidate zero air material(s). Use Valves V1 and V2 to
select each of the gases for measurement. "Discrete" means that the analyst
66
EPA Traceability Protocol for Gaseous Calibration Standards
must change the gas being sampled by the analyzer between measurements.
The analyst may assay multiple zero air materials during the same assay
session. For example, the analyst might cycle through measurements of the
upscale reference standard, the zero gas, the first candidate zero air material,
and then repeat this cycle substituting in second and third candidate zero air
materials. Allow sufficient time between measurements to completely purge
the analyzer of the previous gas and for the analyzer to achieve a stable
reading. Changing the order that the gas mixtures are measured helps the
analyst to discover any effect of that one measurement may have on
subsequent measurements. For each discrete measurement, allow ample
time for the analyzer to achieve a stable response. If the response for each
measurement is not stable, the precision of the measurements will decline
and the candidate standard may not be certifiable under this protocol.
Criteria that apply to the assay of one candidate standard apply to the assay
of multiple candidate standards. The analyst should be aware that the effect
of any short-term calibration drift will be greater when multiple candidate zero
air materials are assayed. This greater effect is due to the longer period of
time between reference standard measurements.
6. If any one or more of the measurements is invalid or abnormal for any reason,
discard all three measurements and repeat the measurements. Such
measurements may be discarded if the analyst can demonstrate that the
experimental conditions were inappropriate during these measurements.
Data cannot be discarded just because they appear to be outliers but may be
discarded if they satisfy statistical criteria for testing outliers.39 The analyst
must record any discarded data and a brief explanation as to why these data
were discarded in the laboratory's records.
7. Verify that the analyzer's precision is acceptable. Calculate the mean and
standard deviation of the analyzer's responses to the zero gas. Repeat the
calculations for the analyzer's responses to upscale reference standard.
These calculations are performed in Step 5 of the spreadsheet described in
Appendix E. Record these calculations in the laboratory's records. The
standard error of the mean for each set of measurements must be less than
or equal to 1.0 percent of the mean response to the upscale reference
standard. That is,
67
EPA Traceability Protocol for Gaseous Calibration Standards
S RURS
≤
n 100
where
s = standard deviation of the analyzer's response;
n = the number of measurements of the zero gas; and
RURS = the zero-corrected mean analyzer response to the upscale
reference standard.
If the value of the standard error of the mean is not acceptable, then the
analyzer must be repaired or another analyzer must be used for the assay.
8. Verify that excessive calibration drift has not occurred since the multipoint
calibration. For the zero gas measurements, calculate the relative difference
(in percent) between the current mean analyzer response during the assay
and the corresponding response that is predicted from the multipoint
calibration regression equation. That is,
Repeat this relative difference calculation for the upscale reference standard
measurements. Record these calculations in the laboratory's records.
Then, if the relative differences for the zero gas and upscale reference
standard checks are less than or equal to 5.0 percent, the analyzer's current
calibration is considered to be approximately the same as during the
multipoint calibration and the multipoint calibration does not have to be
repeated. The data reduction technique does not require that no calibration
drift has occurred since the multipoint calibration. Some minor calibration drift
is acceptable because the effect of any drift will be corrected during the
reduction of the assay data.
However, if the relative difference for either the zero gas or the upscale
reference standard check is greater than 5.0 percent, then the analyzer is out
of calibration. Either a new multipoint calibration must be conducted or the
analyzer's zero and span controls must be adjusted to return the analyzer's
response to the original calibration levels. For some analyzers such as
68
EPA Traceability Protocol for Gaseous Calibration Standards
After any new multipoint calibration, repeat the measurements of the zero
gas, the candidate zero air material, and the upscale reference standard.
Record these measurements in the laboratory's records.
After any adjustment of controls, repeat the measurements of the zero gas,
the candidate zero air material, and the upscale reference standard. Record
these measurements in the laboratory's records. Recalculate the relative
differences to verify that the analyzer has returned to being in calibration.
The analyzer will be considered to be still out of calibration if the relative
differences remain greater than 5.0 percent. In such case, the analyzer may
need to be repaired before the assay is repeated.
The spreadsheet also calculates the percentage of the uncertainty that is due
to the multipoint calibration. This percentage is needed for the total
uncertainty calculations when two or more assays fall under the same
multipoint calibration. Record this value in the laboratory’s records.
10. If the multipoint calibration data and the assay data underwent any
mathematical transformations before their statistical analysis, perform the
reverse transformations for the estimated concentration and the expanded
uncertainty. Record the transformed values in the laboratory’s records.
11. The specialty gas producer or other laboratory that assayed the candidate
zero air material must maintain laboratory records until such time as the
cylinder is blown down after being returned by the end user to the producer.
69
EPA Traceability Protocol for Gaseous Calibration Standards
This protocol describes two procedures for assaying the permeation rate
of a candidate permeation device calibration standard and for certifying that the
assayed permeation rate is traceable to National Institute of Standards and
Technology (NIST) reference standards. This protocol is mandatory for certifying
the permeation device calibration standards used for the pollutant monitoring that
is required by the regulations of the Code of Federal Regulations, Chapter 40,
Parts 50 and 582,3 for the calibration and audit of ambient air quality analyzers.
This protocol covers the assay and certification of sulfur dioxide (SO2) and
nitrogen dioxide (NO2) permeation device calibration standards. It may be used
by permeation device producers, standard users, or other analytical laboratories.
The assay procedure may involve the comparison of the candidate standards to
compressed gas reference standards (i.e., Procedure P2), or to mass reference
standards (i.e., Procedure P3).
The compressed gas reference standards that may be used under this
protocol are NIST Standard Reference Materials (SRMs), Van Swinden
Laboratorium (VSL, the National Metrology Institute of the Netherlands) Primary
Reference Materials (PRMs) that are equivalent to SRMs, NIST-Traceable
Reference Materials (NTRMs), or gas manufacturer's intermediate standards
(GMISs). These standards are described in Subsection 2.1.2 of this report.
71
EPA Traceability Protocol for Gaseous Calibration Standards
72
EPA Traceability Protocol for Gaseous Calibration Standards
6. The dilution gas (air or nitrogen) used in the assay (for Procedure P2);
73
EPA Traceability Protocol for Gaseous Calibration Standards
12. Chronological record of all certifications for the standard by the laboratory;
13. Statement that the standard will retain its certification only as long as 5
percent of the original liquid weight or a visible amount of liquid remains in
it; and
74
EPA Traceability Protocol for Gaseous Calibration Standards
A series of NIST SO2 permeation tubes SRMs were studied at NIST for
their long-term stability at temperatures of 20, 25 and 30° C 46.47. Contrary to
previous reports, their steady-state permeation rates were not constant, but
showed a very slight downward trend with time. The amount of decay in the
permeation rate is less than 2% of its initial rate over the life of the tube. In
normal day-to-day use, permeation rate changes occurring over the entire
steady-state life of the tube would not be discernible. The NIST S02 permeation
tube SRMs had been certified to an uncertainty of ± 2% which is well beyond the
variation in permeation rate usually observed during short-term use.
NO2 permeation devices must be stored in and used in dry dilution gas.
One study showed that NO2 permeation rates were significantly lower in
moderately humid air (i.e., 30 to 40 percent relative humidity) than in dry air on
the preceding day.47 Furthermore, the permeation rates did not return to the
original levels after dry air had passed over the device for 24 hours. Another
study found that NO2 concentrations from a permeation device declined by about
one-third as relative humidity levels increased from 0 to 100 percent.50
75
EPA Traceability Protocol for Gaseous Calibration Standards
76
EPA Traceability Protocol for Gaseous Calibration Standards
77
EPA Traceability Protocol for Gaseous Calibration Standards
3.2.2 Limitations
79
EPA Traceability Protocol for Gaseous Calibration Standards
The mixing chamber combines the gas streams and should provide
turbulence to ensure thorough mixing of the two gas streams. The diluted gas
mixtures are routed to the analyzer through a union tee tube fitting, which vents
excess gas flow. Normally, the excess gas is vented to the atmosphere without
any obstructions in the tubing and the gas entering the analyzer is at near-
atmospheric pressure. However, the excess gas can be routed through an
uncalibrated rotameter by rotation of a three-way valve (i.e., V4). The rotameter
is used to demonstrate that the total gas flow rate exceeds the sample flow rate
of the analyzer and that no room air is being drawn in through the vent line.
Check the apparatus carefully for leaks and correct all leaks before use.
80
EPA Traceability Protocol for Gaseous Calibration Standards
Figure 3-1. One possible design of the apparatus for the assay of permeation
device calibration standards referenced to a compressed gas reference standard
See Subsection 2.2.4. The pollutant gas analyzer must have a well-
characterized calibration curve and a range capable of measuring the diluted
concentrations of both the candidate and reference standards. It must have
good resolution, good precision, a stable response, and low output signal noise.
It must have good specificity for the pollutant of interest so that it has no
detectable response to any contaminant that may be contained in the standards.
A suitable analyzer with acceptable performance specifications may be selected
from the list of EPA-designated reference and equivalent method analyzers37. If
81
EPA Traceability Protocol for Gaseous Calibration Standards
the balance gas of the reference standard must be different from the zero gas
used for dilution (e.g., air versus nitrogen or different proportions of oxygen),
either a high dilution ratio (i.e., at least 50 parts zero gas to 1 part standard)
should be used, or the analyzer must be proven to be not sensitive to differences
in the balance gas composition. The latter may be demonstrated by showing no
difference in an analyzer's response when measuring a calibration standard that
has been diluted with identical flow rates of the different balance gases.
If the analyzer has not been in continuous operation, turn it on and allow it
to stabilize (e.g., for at least 12 hours) before beginning any measurements.
Next the analyst must verify that the dilution error is not excessive. For
the check standard measurements, calculate the relative difference (in percent)
between the mean analyzer response and the corresponding response that is
predicted from the multipoint calibration regression equation and the undiluted or
diluted check standard concentration. That is,
82
EPA Traceability Protocol for Gaseous Calibration Standards
3.2.5.3 Linearity—
See Subsection 2.2.6.3 for information about linearity.
The dilution flow rates used for the reference standard and the candidate
standard should be selected carefully to provide diluted concentrations for both
standards that fall in the well-characterized region of the analyzer's calibration
curve. Additionally, the magnitude of U for the estimated concentration of the
diluted candidate standard must be ≤ ±1.0 percent of the concentration of the
diluted reference standard. It is desirable to measure both dilution flow rates with
the same flowmeter (i.e., M1). This practice reduces measurement errors
associated with the use of multiple flowmeters.
Flowmeters M1 and M3, shown in Figure 3-1, measure in-line flow rates
and do not operate at atmospheric pressure because of backpressure from
downstream components. Also, this backpressure is variable, depending on the
total dilution and reference standard flow rates. Thus, the flowmeters must
compensate for the variable in-line pressure. Thermal mass flowmeters do not
need to be corrected for pressure effects. Measurements from pressure-
sensitive flowmeters such as rotameters or wet test meters must be carefully
corrected for the actual in-line pressure during the flow rate measurements.
Alternatively, the flow rates can be measured at the outlet of the dilution
apparatus, with the excess gas flow vent temporarily plugged. In this case, a
volumetric meter such as a piston-type positive-displacement flowmeter, a wet
test meter or a soap-film flowmeter can be used, and flow measurements may be
conveniently referenced to atmospheric pressure. Each flow rate must be
measured independently while the other flow rate is set to zero. Great care must
83
EPA Traceability Protocol for Gaseous Calibration Standards
be exercised to ensure that each measured flow rate remains constant when
combined with the other flow rate and between the time of measurement and the
time of the assay.
The flowmeters used should be stable, repeatable, linear, and have good
resolution. The flowmeters must not contaminate or react with the gas mixture
passing through them. If possible, select flow rates or flowmeter ranges such
that the measured flow rates fall in the upper half of the flowmeters' ranges. The
flowmeters should be carefully calibrated at several flow rates to prove linearity.
The calibration of the zero gas flowmeter should be accurate to ≤ 1.0 percent,
referenced to an accurate flow or volume standard traceable to a NIST primary
standard (see Subsection 3.1.2). This flowmeter calibration should be checked
and recertified on a regular schedule (e.g., yearly). The recertification frequency
is to be determined from stability information such as a chronological control
chart of calibration data.
where
84
EPA Traceability Protocol for Gaseous Calibration Standards
See Subsections 2.1.3 and 3.1.2 for information about the reference
standard.
See Subsection 2.1.3.6 for information about the zero gas. If possible, the
zero gas should be the same as the balance gas of the reference standard.
3. Verify that the flowmeters are properly calibrated (see Subsection 3.2.7).
5. Determine and establish the reference standard flow rate and the dilution
flow rates and diluted concentrations for the reference standard and the
candidate standard that will be used for the assay (see Subsection 3.2.7).
Ensure that the diluted reference standard and diluted candidate
standards concentrations are within the well-characterized region of the
analyzer's calibration curve (see Subsection 2.1.4.2). Also check that both
dilution flow rates will provide enough flow for the analyzer, with sufficient
excess to ensure that no ambient air will be drawn into the vent line. Also
adjust the flow rate of the portion of the dilution flow that passes over the
candidate standard (i.e., flow controller C3), and adjust the purge flow rate
(i.e., flow controller C2) to approximately the same value.
85
EPA Traceability Protocol for Gaseous Calibration Standards
Calculate the diluted candidate standard concentration (in ppm) using the
following equation:
where
Record the measured flow rates and the undiluted and diluted reference
standard concentrations in the laboratory's records.
6. In succession, measure the zero gas, the diluted reference standard and
the diluted candidate standard using the analyzer. Use valves V1, V2,
and V3 to select each of the three gas mixtures for measurement. For
each measurement, adjust the flow rates, if necessary, to those
determined in Step 5, and allow ample time for the analyzer to achieve a
stable reading. If the reading for each measurement is not stable, the
precision of the measurements will decline and the candidate standard
might not be certifiable under this protocol. Record the analyzer response
for each measurement, using the same response units (e.g., volts, area
counts, etc.) as was used for the multipoint calibration. At this point, do
not convert the data into concentration values using the calibration
equation. Do not perform any mathematical transformations of the data.
These steps will be done later. Do not make any zero control, span
control, or other adjustments to the analyzer during this set of
measurements. Record the responses in the laboratory's records.
The analyst may assay multiple candidate standards during the same
assay session. For example, a single set of measurements may involve a
zero gas, a diluted reference standard, and three diluted candidate
standards. Criteria that apply to the assay of one candidate standard
apply to the assay of multiple candidate standards. The analyst should be
aware that the effect of any short-term calibration drift will be greater when
multiple candidate standards are assayed. This greater effect is due to
the longer period of time between reference standard measurements.
Unacceptable uncertainties of the estimated concentrations for the diluted
candidate standards may occur as a result of the longer assay session.
86
EPA Traceability Protocol for Gaseous Calibration Standards
10. If the multipoint calibration data and the assay data underwent any
mathematical transformations before their statistical analysis, the analyst
must perform to reverse transformations for the estimated concentration
and U. Record the transformed values in the laboratory’s records.
11. Finally, calculate the certified permeation rate (in nanograms/minute) and
uncertainty for the candidate standard using the equations below:
87
EPA Traceability Protocol for Gaseous Calibration Standards
A permeation device that has not been previously assayed must be tested
for a stable permeation rate as follows: Reassay the permeation rate at least 24
hours after the first assay and compare the two assayed concentrations. The
Appendix C spreadsheet or an EPA-approved equivalent statistical technique
(see Subsection 2.1.6) must be used to evaluate the stability of the permeation
rate by comparison of the two assays. If the mean estimated concentrations
from the assays differ by less than 1.0 percent from the overall mean estimated
concentration using Schuirmann's two one-sided test (TOST)13-18, the permeation
rate can be considered to be stable and the candidate standard may be certified
for use. Otherwise, equilibrate the device longer at the operating temperature
and repeat the test, using the second and third assays as if they were the first
and second. This process may be repeated until the rate stabilizes. Permeation
devices that are not stable may not be used for calibration or audit purposes.
Candidate standards that fail the initial and the repeat stability tests are unstable
and are disqualified for further use under this protocol.
88
EPA Traceability Protocol for Gaseous Calibration Standards
3.3.2 Limitations
89
EPA Traceability Protocol for Gaseous Calibration Standards
- Locate the balance away from potential sources of drafts such as doors,
windows, aisles with frequent traffic, ventilation ducts, and equipment with
fans or moving parts.
- Locate the balance out of direct sunlight and away from local heating or
cooling sources such as open flames, hot plates, water baths, ventilation
ducts, windows, and heat-producing lamps.
- Locate the balance on a sturdy base (ideally, a stone weighing table) and
away from any equipment that produces vibrations. If this is not possible,
isolate the balance from such equipment by placing a stabilizing slab
under the balance or composite damping-pads under the balance legs.
90
EPA Traceability Protocol for Gaseous Calibration Standards
91
EPA Traceability Protocol for Gaseous Calibration Standards
100(readability)
Minimum weighing interval =
(expected permeation rate)
92
EPA Traceability Protocol for Gaseous Calibration Standards
where the weighing interval is in minutes; the readability is in grams; and the
expected permeation rate is in grams per minute.
1. Turn on the balance and allow it to warm up for the period specified in the
operator's manual. To ensure maximum stability, it is recommended to
keep the balance turned on at all times. This procedure enables the
balance to be operational at all times and eliminates the need for a
warmup period before analyses. Newer balances are always turned on
(except for their displays) when they are plugged in.
2. Check the balance level and, if necessary, adjust the level according to
the manufacturer's instructions.
4. Zero (i.e., tare) and calibrate the balance according to the manufacturer’s
directions. Record the tare reading in the laboratory’s records. Many
newer balances calibrate themselves automatically or only require a key to
be touched to calibrate themselves.
93
EPA Traceability Protocol for Gaseous Calibration Standards
and maximum temperatures must not have deviated from the setpoint by
more than 0.1 °C. If these temperatures deviate by more than this
amount, the current assay and all previous assays are invalidated.
10. Record the date, time and the candidate standard's identification number
and current weight in the laboratory's records.
The candidate standard must be weighed at least six times after its
permeation rate has stabilized at the certification temperature. After the six or
more weighings, the analyst may assess the stability and uncertainty of the
permeation rate by using the spreadsheet described in Appendix B or an EPA-
approved equivalent statistical technique (see Subsection 2.1.6). The analyst
may calculate a provisional permeation rate from the measured weights and the
time between weighings using the following equation:
Based on this data analysis, the analyst may perform additional weighings to
reduce the uncertainty or to replace data that are discarded because they were
obtained before the permeation rate stabilized.
The certified permeation rate for the candidate standard is the slope of the
least squares regression line for data from at least six weighings of the candidate
94
EPA Traceability Protocol for Gaseous Calibration Standards
standard after the permeation rate has stabilized. This statistical analysis
technique produces permeation rate estimates that are more precise than those
calculated from weight differences between individual weighings. Although the
minimum number of weighings is six, more precise estimates will be obtained for
more weighings.
Calculate the certified permeation rate and its uncertainty using the
spreadsheet given in Appendix B or using an EPA-approved equivalent statistical
technique (see Subsection 2.1.6). Figure 3-3 presents an example of the
spreadsheet's graphic output.
3.420
3.415
3.410
3.405
Weight (grams)
3.400
3.395
3.390
3.385
3.380
3.375
3.370
314 316 318 320 322 324 326
Julian Date
The permeation rate is the slope (b1) of the least squares regression line
where the x-values are the elapsed times from the initial weighing and the y-
values are the measured weights of the candidate standard. The spreadsheet
also calculates the predicted initial weight (b0) of the permeation device and 95-
percent confidence limits (CLs) for b0 and b1.
After the data from the six or more weighings have been entered in the
spreadsheet, examine the 95-percent CLs for b0. If the measured weight from
the initial weighing falls outside of these limits, the candidate standard may not
have been completely equilibrated at the initial weighing. The analyst may elect
to discard the data from the initial weighing to reduce the uncertainty of the
certified permeation rate.
95
EPA Traceability Protocol for Gaseous Calibration Standards
Examine the upper and lower 95-percent CLs for b1. They should differ
from b1 by no more than 1 percent of its value. That is,
If these two criteria are met, the candidate standard can be certified with a
permeation rate equal to b1 and an uncertainty equal to the larger of the two
values. If the criteria are not met, the analyst must make additional weighings of
the candidate standard as described in Subsection 3.3.5. These additional
measurements will be pooled with the previously collected measurements. The
pooled data will be used to obtain new estimates of the b1 and its uncertainty.
When an acceptable value for the uncertainty is obtained, record it and the slope
in the laboratory's records. If an acceptable value is not obtained, the candidate
standard cannot be certified under this protocol.
The total analytical uncertainty of the certified permeation rate includes the
uncertainty of regression slope and the uncertainty of the mass reference
standard that was used to verify the balance’s calibration. The two components
are combined using the following equation for the propagation of errors:
1/2
Uncertainty (Total) Uncertainty (Slope) 2 Uncertainty (Mass) 2
= +
Permeation Rate Slope Mass
96
EPA Traceability Protocol for Gaseous Calibration Standards
This protocol describes a procedure for assaying gas dilution systems and
for certifying that the concentrations in the pollutant gas mixtures that they
generate are traceable to NIST-certified gaseous, flow rate, volume, mass or
time reference standards. In general, a pollutant gas mixture containing a known
concentration is prepared by diluting an NIST-traceable gaseous reference
standard with a zero gas. The flow rates of the reference standard and the zero
gas must be traceable to NIST-traceable flow rate, volume, mass or time
reference standards. Dynamic gas dilution systems employ various dilution
methods to prepare pollutant gas mixtures. Examples of these methods include
capillary tubes, critical orifices, volumetric piston pumps, and thermal mass-flow
controllers. This protocol may be used by the manufacturers of gas dilution
systems, other laboratories, and end users.
The gas mixtures generated by gas dilution systems certified under this
protocol may be used for calibrations and audits of ambient air quality analyzers,
including both field and laboratory based instrumentation, and continuous
emission monitors for pollutant monitoring that is required by the regulations in
40 CFR Parts 50, 58, 60, and 753-6.
Note that EPA Method 20552 may also be used for the verification of gas
dilution systems for field instrument calibrations of EPA Methods 3A, 6C, 7E, 10,
97
EPA Traceability Protocol for Gaseous Calibration Standards
15, 16, 20, 25A and 25B that are required by the regulations in 40 CFR Part 60
as well as performance specifications according to 40 CFR Part 60, Appendix F.
At this time, EPA does not require the regulated community to use NIST-
traceable dynamic gas dilution systems for the calibration of ambient air or
continuous emission monitors that are required by 40 CFR Parts 50, 58, 60, and
75. However, end users may elect to use them for calibrations.
4.1.2 Limitations
This protocol is not intended for the blanket certification of multiple gas
dilution systems that may be identical in operating principle, construction, and
materials. It is intended for the certification of individual gas dilution systems.
This protocol is not intended for the certification of static gas dilution systems.
Figure 4-1 illustrates one possible design of a gas dilution system for the
generation of pollutant gas mixtures using dynamic dilution methods. This
apparatus is designed to allow the quantitative mixing of a pollutant gas with a
zero gas using gas flow control devices. It also is designed to allow the
measurement of the pollutant gas flow rate and the zero gas flow rate. Inert
materials (e.g., Teflon®, stainless steel, borosilicate glass or silanized glass) and
clean, noncontaminating components are to be used in those portions of the
apparatus that are in contact with the pollutant gas and the gas mixture being
generated. The pathways of the gas streams through the gas dilution system are
controlled by rotation of two three-way valves. Pressure regulators, shutoff
valves, and gas flow control devices regulate the flow from the cylinders. The
gas flow control devices may be needle values, capillary tubes, volumetric piston
pumps, thermal mass flow controllers, or other flow control devices. The
pollutant gas and zero gas flow rates are measured at ports that are connected
to the three-way valves. The pollutant gas stream and the zero gas stream must
be mixed in a chamber whose design ensures that the resulting gas mixture is
thoroughly homogenized. The gas mixture output is connected to the sample
line of the pollutant gas analyzer.
Figure 4-1 shows separate flow measurement ports for the pollutant gas
stream and the zero gas stream. In this configuration, a single flow rate
reference standard can be used to measure either the pollutant gas flow rate or
the zero gas flow rate. Such an approach reduces measurement errors arising
from differences in the calibration of multiple reference standards. Alternatively,
the flow rates can be measured at the outlet port of the gas dilution system with
one of the gas streams shut off. The gas dilution system's design must ensure
that the flow rate in one gas stream remains constant when combined with the
other gas stream.
98
EPA Traceability Protocol for Gaseous Calibration Standards
Pollutant Flow
3-Way Measurement
Gas Valve Port Pollutant Flowrate
Flow Reference
Control Standard
Shutoff
Valve
Pollutant
Gas
Cylinder Diluted Gas
Output Port
Pollutant Gas
Mixing
Analyzer to be
Chamber
Calibrated
Dilution Flow
Measurement
Gas Port Dilution Flowrate
Flow Reference
Control Standard
Shutoff 3-Way
Valve Valve
Zero
Gas
Cylinder Gas Dilution System
Figure 4-1. One possible design of a gas dilution system for the generation
of pollutant gas mixtures using dynamic dilution methods
99
EPA Traceability Protocol for Gaseous Calibration Standards
The apparatus may be modified in several ways that will not diminish its
performance. For example, the calibration could be performed using a flow rate
reference standard that is mounted in-line upstream of the flow control device.
If the analyzer has not been in continuous operation, turn it on and allow it
to stabilize (e.g., for at least 12 hours) before beginning the measurements.
100
EPA Traceability Protocol for Gaseous Calibration Standards
Zero gas used in gas dilution systems should be clean, dry, zero-grade air
or nitrogen containing no detectable concentration of the pollutant of interest. It
may come from compressed gas cylinders or from zero gas generators. The use
of NIST-traceable zero air material is recommended, but is not required. The
zero gas should match the balance gas in the gaseous reference standard,
unless it has been demonstrated that the pollutant gas analyzer being calibrated
is insensitive to differences in the balance gas composition. It also should
contain no contaminant that causes a detectable response in the analyzer or that
suppresses or enhances the analyzer's response to the pollutant. The oxygen
content of zero air should be approximately that of ambient air, unless it has
been demonstrated that varying the oxygen content does not suppress or
enhance the analyzer's response. The water vapor concentration in the zero gas
should be less than 5 ppm. Information concerning the zero gas (e.g., cylinder
identification number, impurity levels, cylinder pressure, etc.) must be recorded in
the laboratory's records.
The flow rate reference standards for calibrating gas dilution systems
must be traceable to NIST primary flow rate, volume, mass or time reference
standards 54 by original calibration by their manufacturer and by annual
recertification at a NIST-accredited state weights and measures laboratory or at a
calibration laboratory that is accredited by the National Voluntary Laboratory
Accreditation Program (NVLAP), the American Association for Laboratory
accreditation (A2LA) or by the International Laboratory Accreditation Conference
(ILAC) under ISO/IEC 17025 (General Requirements for the Competence of
Testing and Calibration Laboratories).10-12 Their certifications must cover the
range of flow rates that will be used by the gas dilution systems. The expanded
uncertainties (U) of the standards must be documented in a calibration certificate
or they must be available through the manufacturer or calibration facility. Their
expanded uncertainties must not be greater than plus or minus 1 percent of
reading for the flow rates that will be used for gas dilution. Expanded
uncertainties that are much less than plus or minus 1 percent of reading are
desirable so that the uncertainties of the gas mixture concentrations are
decreased. They must be stable, precise, linear, and have good resolution.
101
EPA Traceability Protocol for Gaseous Calibration Standards
If the gas dilution system has multiple flow rate ranges or multiple flow
control devices, a multipoint calibration must be done for all ranges and flow
control devices that will be used later for generating pollutant gas mixtures.
102
EPA Traceability Protocol for Gaseous Calibration Standards
Volume PM − PWV TS
Flow rate =
Time PS TM
where
103
EPA Traceability Protocol for Gaseous Calibration Standards
the cubic and quartic models and believes that better fits of the data can be
obtained by performing multipoint calibrations over more limited concentration
ranges and by using straight-line or quadratic models. Additionally, a multipoint
calibration should not change regression model orders from one year to the next.
Two points should be noted about the third uncertainty component. First,
its magnitude decreases as n increases where n is the number of measurements
104
EPA Traceability Protocol for Gaseous Calibration Standards
in the multipoint flow rate calibration. Second, its magnitude increases as the
mean measured response ( Y ' ) for the flow-rate reference standard diverges from
the overall mean measured response ( Y ) for the multipoint calibration. That is,
the absolute uncertainty of the regression predicted concentration is larger at the
extremes of the calibrated concentration range than at the middle of the range.
Note that Cells D29, H27, B44, and H44 of the day-of-use worksheet in
the Appendix F spreadsheet allows the analyst to enter pollutant gas and zero
gas flow controller settings and to obtain the regression-predicted flow rates and
the corresponding relative standard uncertainties (i.e., uc/flow rate) associated
solely with the calibration. To determine whether these predicted flow rates are
in the well-characterized region of the calibration curve, multiply these
uncertainties by the quantity 2 x (largest calibration flow rate/predicted flow rate)
and compare the resulting value to 0.01, which is the acceptance criterion for the
well-characterized region.
105
EPA Traceability Protocol for Gaseous Calibration Standards
The reference standard and a zero gas are to be connected to the inlet of
the gas dilution system. The pollutant and zero gas flow controllers' settings
must be adjusted such that the concentration of the diluted reference standard is
within +/- 1.0 percent of the concentration of the undiluted check standard. Note
that the annual gas check worksheet in the Appendix F spreadsheet or an EPA-
approved equivalent statistical technique (see Subsection 2.1.6) allows the
analyst to enter the reference standard concentration, the check standard
concentration, and various pollutant and zero gas flow rates to match the two
concentrations. The worksheet also determines the correct settings for the
pollutant and zero gas flow controllers to match the two concentrations and the
relative concentration difference (RCD) between the two concentrations. That is
106
EPA Traceability Protocol for Gaseous Calibration Standards
Calculate the mean and standard deviation of both sets of measurements using
the annual gas check worksheet in the Appendix F spreadsheet.
Next the analyst must verify that the systematic error (bias) of the dilution
is not excessive by comparing the mean analyzer responses for the diluted
reference standard and the undiluted check standard. Calculate the relative
agreement between the two responses using the annual gas check worksheet in
the Appendix F spreadsheet. That is,
The acceptance criterion for the relative agreement includes the relative
standard uncertainties for the concentrations of the reference and check
standards, the uncertainties of the pollutant and zero gas flow rates for the
reference standard, the relative standard deviations for the individual
measurements of the diluted reference standard and undiluted check standard,
and the RCD. That is,
2 2 2
uref ucheck u flow1 u flow 2 sconc1 sconc 2
2 2 2
Criterion =
RCD + 2 +
+ +
+
+
ref check flow1 flow2 meanconc1 meanconc 2
This calculation is performed using the annual gas check worksheet in the
Appendix F spreadsheet.
Document the RCD, the relative agreement, and the acceptance criterion
in the laboratory's records. If the agreement is greater than the acceptance
criterion, the dilution error is considered to be excessive. The analyst must
investigate why the relative difference is excessive. The problem may be due to
errors in the reference standard and check standard concentrations, errors in
assay apparatus or to some other source. Pollutant gas analyzer calibrations
may not be conducted until the agreement for a subsequent gas check is less
than or equal to the criterion for that gas check.
107
EPA Traceability Protocol for Gaseous Calibration Standards
balance gas for both reactive gas standards. The undiluted reactive check
standard must have a certified concentration that is at least an order of
magnitude less (i.e., a factor of 10) than the concentration of the reactive
reference standard that is diluted. Information concerning both standards (e.g.,
cylinder identification number, certified concentration, uncertainty, traceability to
NIST) and the diluent gas must be recorded in the laboratory's records.
The reference standard and a zero gas are to be connected to the inlet of
the gas dilution system. The pollutant and zero gas flow controllers' settings
must be adjusted such that the concentration of the diluted reference standard is
within +/- 1.0 percent of the concentration of the undiluted check standard. Note
that the annual gas check worksheet in the Appendix F spreadsheet allows the
analyst to enter the reference standard concentration, the check standard
concentration, and various pollutant and zero gas flow rates to match the two
concentrations. The worksheet also determines the correct settings for the
pollutant and zero gas flow controllers to match the two concentrations and the
relative concentration difference (RCD) between the two concentrations. That is
Next the analyst must verify that the dilution-plus-wall-loss error is not
excessive by comparing the mean analyzer responses for the diluted reference
standard and the undiluted check standard. Calculate the relative agreement
between the two responses using the annual gas check worksheet in the
Appendix F spreadsheet or using an EPA-approved equivalent statistical
technique (see Subsection 2.1.6). That is,
The acceptance criterion for the relative agreement includes the relative
standard uncertainties for the concentrations of the reference and check
standards, the uncertainties of the pollutant and zero gas flow rates for the
reference standard, the relative standard deviations for the individual
measurements of the diluted reference standard and undiluted check standard,
and the RCD. That is,
2 2 2
uref ucheck u flow1 u flow 2 sconc1 sconc 2
2 2 2
Criterion =
RCD + 2 +
+ +
+
+
ref check flow1 flow2 meanconc1 meanconc 2
108
EPA Traceability Protocol for Gaseous Calibration Standards
This calculation is performed using the annual gas check worksheet in the
Appendix F spreadsheet.
Document the RCD, the relative agreement, and the acceptance criterion
in the laboratory's records. If the agreement is greater than the acceptance
criterion, the dilution error is considered to be excessive. The analyst must
investigate why the relative difference is excessive. The problem may be due to
errors in the reference standard and check standard concentrations, errors in
assay apparatus or to some other source. Pollutant gas analyzer calibrations
may not be conducted until the agreement for a subsequent gas check is less
than or equal to the criterion for that gas check.
Next the analyst must verify that the flow-rate shift is not excessive. For
the annual multipoint calibration measurements and the monthly QC check
measurements, calculate the relative difference (in percent) between the mean
QC check flow rate and the corresponding flow rate that is predicted from the
annual multipoint calibration regression equation. That is,
109
EPA Traceability Protocol for Gaseous Calibration Standards
Note that Cells D29, H27, B44, and H44 of the day-of-use worksheet in
the Appendix F spreadsheet allows the analyst to enter pollutant gas and zero
gas flow controller settings and to obtain the regression-predicted flow rates and
the corresponding relative standard uncertainties (i.e., uc/flow rate) associated
solely with the calibration. To determine whether these predicted flow rates are
in the well-characterized region of the calibration curve, multiply these
uncertainties by the quantity 2 x (largest calibration flow rate/predicted flow rate)
and compare the resulting value to 0.01, which is the acceptance criterion for the
well-characterized region.
110
EPA Traceability Protocol for Gaseous Calibration Standards
are produced when implementing this protocol. The calculated value of U must
be given in the certification documentation. The use of an uncalculated or
blanket estimate of U (e.g. +/- 1 percent) is not acceptable.
For each dynamic gas dilution system that is assayed using this protocol,
the assay results must be documented in a written report, which contains at least
the following information:
2. The manufacturer, the model, and the serial number of the dynamic gas
dilution system;
3. The flow-rate ranges for which the dynamic gas dilution system is certified
based on the annual multipoint calibration;
4. The dynamic gas dilution system's calibration equation based on the annual
multipoint calibration. The calibration equation can be supplemented by a
graphical representation of the calibration curve if desirable;
111
EPA Traceability Protocol for Gaseous Calibration Standards
5. A calculated estimate of U for the dynamic gas dilution system. The use of an
uncalculated, blanket value for U (e.g., +/- 1 percent) is not acceptable. The
estimate is expressed as a 95-percent confidence interval, which is the
combined standard uncertainty (uc) multiplied by a coverage factor (k), which
is equal to 2.8,9 It must include the uncertainty in the flow-rate reference
standard(s), uncertainty in the annual multipoint calibration, uncertainty in the
annual nonreactive or reactive gas calibration check, uncertainty in monthly
flow-rate QC checks, and random measurement error. Use the spreadsheet
described in Appendix F or an EPA-approved equivalent statistical technique
to calculate U. If an EPA-approved equivalent statistical technique is used
rather that the Appendix F spreadsheet, it must be identified as such in the
certificate documentation and must be described in media readily accessible
to end users. As a rule-of-thumb, U should be rounded to one significant
figure unless the leading figure is a 1 in which case two significant figures
should be reported12. The estimate of U (e.g., +/- 20 milliliters per minute)
may be supplemented by a statement of the equivalent fractional uncertainty
(e.g., +/- 1.0 percent) if desired (see Section 4.1.5);
7. Certification expiration date (i.e., the multipoint calibration date plus one year
plus one day);
9. The relative difference(s) that was calculated from the annual nonreactive
(and reactive) gas calibration check(s);
10. Information about the undiluted reference standard and the diluted check
standard that were used in the annual nonreactive (and reactive) gas
calibration check(s): NIST SRM number, NIST sample number, cylinder
identification number, certified concentration, expanded uncertainty, and
certification expiration date for an SRM; cylinder identification number,
certified concentration, expanded uncertainty, and certification expiration date
for an NTRM, an RGM, a PRM, a CRM or a GMIS. The certification
documentation must identify the reference standard as being an SRM, an
NTRM, an RGM, a PRM, a CRM or a GMIS. For a GMIS, the reference
standard that was used for its assay must be identified in the documentation
for the candidate standard being certified.
112
EPA Traceability Protocol for Gaseous Calibration Standards
issued by EPA for the production location) where the dynamic gas dilution
system was calibrated, even if another organization will sell it to an end user.
This identification must be given in the same or larger font as the other
required information in the report.
The certification of a dynamic gas dilution is valid for only one year
following its annual multipoint calibration.
113
EPA Traceability Protocol for Gaseous Calibration Standards
Section 5 - References
6. F.R. Guenther, W.D. Dorko, W.R. Miller, and G.C. Rhoderick. The NIST
Traceable Reference Program for Gas Standards. National Institute of
Standards and Technology. Special Publication 260-126. 1996. 40 pp.
9. B.N. Taylor and C.E. Kuyatt. Guidelines for Evaluating and Expressing the
Uncertainty of NIST Measurement Results: 1994 Edition. National
Institute of Standards and Technology. Technical Note 1297. 1994.
(http://physics.nist.gov/pubs/guidelines/TN1297/tn1297.pdf, accessed
April18, 2011).
115
EPA Traceability Protocol for Gaseous Calibration Standards
10. G.L. Harris. State Weights and Measures Laboratories: State Standards
Program Description. National Institute of Standards and Technology.
Special Publication 791. 1994. 130 pp.
13. D.J. Schuirmann. "A Comparison of the Two One-Sided Tests Procedure
and the Power Approach for Assessing the EQuivalence of Average
Bioavailability", Journal of Pharmokinetics and Biopharmaceutics. 15(6):
657-680. 1987.
14. S.J. Richter and C. Richter. "A Method for Determining Equivalence in
Industrial Applications", Quality Engineering, 14(3):375-380. 2002.
16. G.B. Limentani et al. "Beyond the t-Test: Statistical Equivalence Testing",
Analytical Chemistry. 77(11):221A-226A. 2005.
116
EPA Traceability Protocol for Gaseous Calibration Standards
22. R.C. Shores, F. Smith, and D.J. von Lehmden. "Stability Evaluation of
Sulfur Dioxide, Nitric Oxide and Carbon Monoxide Gases in Cylinders."
EPA/600/4-86, U.S. Environmental Protection Agency, Research Triangle
Park, NC. 1984. 52 pp.
24. National Council of the Paper Industry for Air and Stream Improvement,
Inc. "An Investigation of the Stability of H2S in Air Cylinder Gases."
NCASI Special Report No. 90-09, New York, NY. 1990. 13 pp.
25. W.R. Miller and G.C. Rhoderick. "Stability of compressed gas mixtures
containing low level volatile organic compounds in aluminum cylinders,"
Fresenius J Anal Chem. 351(2-3):221-229. 1995.
28. W.R. Miller and F.R. Guenther. "The certification of a nominal 20 μmol/mol
H2S/N2 standard reference material using two independent methods," Anal
Bioanal Chem. 387(7):2433–2439. 2007.
30. R.S. Wright et al. Performance Audits of EPA Protocol Gases and
Inspection and Maintenance Calibration Gases, Journal of the Air Pollution
Control Association, 37(4):384-385, 1987.
31. R.S. Wright et al. Accuracy Assessment of EPA Protocol Gases in 1988,
Journal of the Air and Waste Management Association, 39(9):1225-1227,
1989.
117
EPA Traceability Protocol for Gaseous Calibration Standards
36. U.S. Environmental Protection Agency. U.S. EPA Ambient Air Monitoring
Protocol Gas Verification Program, Annual Report, CY 2010, Office of Air
Quality Planning and Standards, 2011
(www.epa.gov/ttn/amtic/files/ambient/qaqc/aagvp2010report.pdf,
accessed April 18, 2011).
39. American Society for Testing and Materials. Standard Practice for Dealing
with Outlying Observations. ASTM Standard Practice E 178-80, 1980.
41. American National Standards Institute (ANSI) and American Society for
Testing and Materials (ASTM). Standard Specification for Laboratory
Weights and Precision Mass Standards. ANSI/ASTM Standard E 617-91,
1991.
118
EPA Traceability Protocol for Gaseous Calibration Standards
49. W.J. Mitchell et al. “Simple Systems for Calibrating and Auditing SO2
Monitors at Remote Sites.” Atmospheric Environment. 26A(1):191-194.
1992.
119
EPA Traceability Protocol for Gaseous Calibration Standards
55. R.S. Wright and R.W. Murdoch. "Labratory Evaluation of Gas Dilution
Systems for Analyzer Calibration and Calibration Gas Analysis", Air &
Waste, 44(4):428-430. 1994.
120
EPA Traceability Protocol for Gaseous Calibration Standards
Purpose
Linear: y = 0 + 1x+
Quadratic:y = 0 + 1x + 2 x2 +
Cubic: y = 0 + 1 x + 2 x2 + 3x3 +
Quartic: y = 0 + 1 x + 2 x2 + 3 x3 + 4x4 +
The output quantities are estimated parameters (s) and related uncertainties.
The workbook then helps the user perform the following functions:
Warning
EPA discourages the use of cubic and quartic models and believes that better fits
of the data can be obtained by performing multipoint calibrations over more
limited concentration ranges and by using straight-line or quadratic models.
Inclusion of cubic and quartic models in the spreadsheet is for experimental use
or for situations in which there is a theoretical basis for the use of such higher-
order models. Analysts should be aware that apparent higher-order calibration
curves may be caused by artifacts such as inaccurate reference standards or
leaks in a gas dilution system. They should not use higher-order regression
models to fit multipoint calibration data that have inadequate precision and that
should be fitted to lower-order regression models. If analysts suspect that
121
EPA Traceability Protocol for Gaseous Calibration Standards
Organization
The workbook consists of several worksheets, which are displayed as tabs at the
bottom of the screen. The functions of these worksheets are described below:
Conventions
The Measurement Data worksheet includes instructions that guide the user
through the steps in its use. The worksheet is also color coded to simplify use.
Shaded cells that are bordered in blue lines are for input of data. These cells are
unprotected, but all other cells of the Measurement Data worksheet are
protected. The only other unprotected cell in the workbook is cell F4 of the Chart
Data worksheet. That cell controls the width of the confidence bands plotted in
the Curves 1 and Curves 2 charts.
122
EPA Traceability Protocol for Gaseous Calibration Standards
Derived values and statements are colored red. These cells contain formulas
and are protected to prevent alteration.
Use
The Measurement Data worksheet guides the user through six steps.
In this step, the user reviews the estimates of the intercepts (0), slopes (1) and
other coefficients (2, 3 , and 4 ) for the four models, examines their confidence
intervals and the residual error variances (s2). The result of an F-test indicates
which of the models is best. The linear model is recommended unless the
quadratic or higher-order model significantly reduces the residual error.
In this step, the user reviews the charts named Curves 1, Residuals 1, Curves 2,
Residuals 2, etc. These charts help the user understand why one model
performs better than the other.
In this step, the user enters the assumed concentration of a candidate standard
and selects a replication number, r. Based on the calibration results, the
worksheet estimates the expanded uncertainty associated with an approximate
level of confidence of 95% that would result from measuring such a standard r
123
EPA Traceability Protocol for Gaseous Calibration Standards
times. The user can use this as a guide for deciding whether to proceed with
analysis, producing additional calibration points, or taking some corrective action.
In this step, the user enters the responses to a candidate standard that is tested
on the same day as the calibration of STEP 1. The worksheet provides an
estimate of the candidate's concentration and its 95% uncertainty. The
worksheet also indicates whether the variability in responses is larger than
expected (unacceptable).
In this step, the user enters the responses to a candidate standard that is tested
on a different day from the calibration of STEP 1. The worksheet first assesses
the zero and span responses. If the zero and span responses are acceptable,
the user proceeds to enter the results from testing a candidate standard. The
results include those for zero and nonzero reference standards. The table below
specifies the minimum number of different reference standard concentrations
needed for each model, in addition to the zero standard. Failure to meet these
requirements will result in serious calculation errors and invalid results.
124
EPA Traceability Protocol for Gaseous Calibration Standards
polynomial calibration curve and the corresponding straight line between the zero
gas and the highest-concentration standard.
The spreadsheet determines whether the regression curve has changed since
the initial calibration. The data are corrected for any change and the estimated
concentration of the candidate standard is provided together with its expanded
uncertainty associated with an approximate level of confidence of 95%.
Enter the calibration data in the shaded spaces below. The first column (I)
simply counts the calibration points that you enter. The second column (X) is for
the certified concentrations of the calibration gas standards. The third column
(Y) is for the instrument responses corresponding to the calibration standards.
The number of points cannot exceed 50.
i Xi Yi Color Code
1 0.000 0.2194
2 0.500 0.7141 red = derived value (protected)
3 1.000 1.2885
4 1.500 1.9132 blue = entered value (unprotected)
5 2.000 2.5910
6 2.500 3.2866 black = fixed text (protected)
7 3.000 4.1078
8 3.500 4.9446
9 4.000 5.8145
10 4.500 6.7230
11 5.000 7.7284
12 5.500 8.7566
13 6.000 9.8013
14 6.500 10.8818
15 7.000 12.0982
16 7.500 13.3122
17 8.000 14.5840
18 8.500 15.9238
19 9.000 17.3271
125
EPA Traceability Protocol for Gaseous Calibration Standards
Review the estimates of the coefficients (b0, (b1,...) for the linear and quadratic
models, their confidence, and the residual error variances (s2).
The corresponding intervals take the form estimate +/- t*u, where: estimate is the
quantity of interest, u is its uncertainty, and t is a coverage factor.
For each model, the coverage factor, t, is determined from the t-distribution with
appropriate degrees of freedom (df).
Note that the uncertainties are not displayed, but can easily be derived from the
estimate, expanded uncertainty, and the coverage factor.
F ratio = 1026.76
F critical = 2.3167 (5% significance level)
126
EPA Traceability Protocol for Gaseous Calibration Standards
F ratio = 0.9385
F critical = 2.3849 (5% significance level)
The error variances are not significantly different at the 5% level. The quadratic
model appears to be a better choice than cubic.
F ratio = 1.8954
F critical = 2.4630 (5% significance level)
The error variances are not significantly different at the 5% level. The cubic
model appears to be a better choice than quartic.
View the charts named Curves 1 and Residuals 1. Curves 1 shows bands
illustrating expanded uncertainties for the estimated regression. Compare these
bands with those of the quadratic regression, Curves 2. (Note: You can change
127
EPA Traceability Protocol for Gaseous Calibration Standards
Enter the concentration at which you would like to evaluate the uncertainty of
estimation and prediction. Also enter r, the number of assays to be performed.
Increasing r reduces the prediction uncertainty, but with diminishing effect.
Concentration = 6
r= 3
Review the estimated mean response (estimate that only takes into account the
calibration uncertainty), and the confidence intervals. Review the predicted
mean response and its confidence intervals. To satisfy the EPA protocol
requirements, the 95% confidence limits for the concentration should be 1% of
the concentration.
Expanded Uncertainty
Estimate Lower Upper
Instrument Response = 9.8006 9.7858 9.8155
Expanded Uncertainty
Prediction Lower Upper
Instrument Response = 9.8006 9.7698 9.8314
Concentration = 6.0000 5.9860 6.0140
95% uncertainty in prediction = 0.23%
128
EPA Traceability Protocol for Gaseous Calibration Standards
This step checks the accuracy of the apparatus to verify that the error associated
with dilution is not excessive. If no certified standards are diluted in performing
the multipoint calibration, skip to Step 5.
mean = 9.850
Cal. Response = 9.8006332
RD = 0.50%
Relative difference is ok
Note: This step applies only to candidate standards that are assayed on the
same day as the calibration.
Analysis Estimated
Number Response Concentration
1 4.500 3.2466
2 4.501 3.2473
3 4.499 3.2460
NOTE: For Cubic and Quartic Model estimates, view the Calculations in the
spreadsheet’s shaded regions
mean = 3.2466
standard deviation = 0.0006
df = 2
F= 0.0008
F sig? = FALSE (The sample variance is acceptable.)
Pr(>F) = 0.9992
95% Uncertainty = 0.55% (This is the relative expanded uncertainty.)
23.97% = = Portion of Variance Due to Calibration
129
EPA Traceability Protocol for Gaseous Calibration Standards
This step applies to candidate standards that are assayed on a different day than
the initial calibration. Before candidate standards are run, the measurement
system is challenged with zero and span checks. Three or more discrete checks
of the zero gas and three or more checks of the span gas are made. Enter the
results below:
Response to Response
Zero gas Span conc. to Span
0.000 9.000 16.010
0.001 9.000 16.000
-0.001 9.000 15.990
n= 3 n= 3
mean = 0.000 mean = 16.000
s= 0.001 s= 0.010
Cal. 0.196 Cal. Response = 17.301
Resp.=
130
EPA Traceability Protocol for Gaseous Calibration Standards
STEP 6 (Continued)
See table in Step 6 of the Instructions for the minimum number of different
nonzero reference standard concentrations required.
Lower Upper
2.9114 2.9560
These upper and lower limits are compared with the corresponding limits
estimated on different assay dates to establish that the candidate standard has
not drifted.
131
EPA Traceability Protocol for Gaseous Calibration Standards
132
EPA Traceability Protocol for Gaseous Calibration Standards
This workbook enables the user to estimate the rate at which the weight of a
permeation tube decreases. A linear relationship between the tube's weight and
elapsed time is established. If the estimated weight at time zero is significantly
different from the actual weight at time zero, then at least the earliest data pair
should be removed from the analysis. Uncertainty of the slope estimate (the rate
of weight loss or drift) will be determined. The traceability protocol requires that
this estimate have a relative uncertainty of less than 1%.
Each worksheet contains instructions that guide the user through the steps in
usingthe worksheet. The worksheets are also color-coded to simplify use.
Shaded cells that are bordered in blue lines are cells whose contents you can
change (i.e., enter data). In other sheets you can change the following variables:
133
EPA Traceability Protocol for Gaseous Calibration Standards
Derived values are colored red. These cells contain formulas that should not be
changed. The cells are protected to prevent alteration.
Step 1: Enter the elapsed times (all in the same units) and corresponding
tube weights in the Data worksheet. The worksheet will compute
the total weight loss for each observation.
Step 3: Examine the corresponding Curve chart. You may need to adjust
the chart's axis scaling. The points should appear to fall virtually on
top of the black line. The black line should be very close to its
confidence bands (colored red and blue).
Step 5: Print the one-page report provided in the Report sheet. The report
summarizes the assay data and indicates the uncertainty of the
estimate.
134
EPA Traceability Protocol for Gaseous Calibration Standards
Elapsed
Time Weight Enter the data in the blue-bordered spaces. The
first column (X) is for the elapsed time. The time of
I Xi Yi the first entry should be zero. The second column
(Y) is for the tube weights.
0 0 4.354206
1 8641 4.33745 n = number of weighing. This can’t exceed 50.
2 18722 4.316766 n=7
3 40322 4.273494
4 64802 4.224514
5 74882 4.20378
6 84962 4.18439 No data entry is required for derived values, which
are colored red, such as n and I. These values are
tabulated automatically and their cells are protected
from alteration.
3. Regression Worksheet
STEP 1
Review the estimates of the intercept (b0), slope of the regression line (b1), and
their confidence limits along with the estimates of variance-covariance matrix (V)
and the residual error variance (Var).
Derivation of the estimated intercept (b0) and slope (b1) of the regression line
135
EPA Traceability Protocol for Gaseous Calibration Standards
STEP 2
Examine the upper and lower limits for the intercept, b0.
STEP 3
Examine the upper and lower limits for the slope, b1. The limits should differ from
the estimate by no more than 1% of the estimated slope.
136
EPA Traceability Protocol for Gaseous Calibration Standards
portion of the data. The initial measurements may not align with subsequent
measurements if the device was in the process of stabilizing or equilibrating
during those times. If this is the case, the initial points of the Residual chart
would appear to be outliers. The residuals with the same sign (all positive or all
negative) and their magnitude will likely be greater than the magnitude of
subsequent residuals. If this is the case, consider removing the initial points from
the computations and re-enter the remaining times and weights with the times
adjusted so the first entry has time zero.
If the uncertainty is acceptable, print the Report spreadsheet and include it with
the certification documentation.
4. Report Worksheet
Test Results
137
EPA Traceability Protocol for Gaseous Calibration Standards
138
EPA Traceability Protocol for Gaseous Calibration Standards
In this sheet the results of two or three assays are entered. Calibration dates are
entered so Assays having the same calibration uncertainty may be treated
correctly. (Assays having a common calibration share the same calibration
uncertainty.)
Enter the results for up to three separate assays in chronological order below.
ASSAY 1
ASSAY 2
139
EPA Traceability Protocol for Gaseous Calibration Standards
COMPARISONS
Assay 1 vs Assay 2
Two One-Sided Equivalence
with Assay 1 Variance Components
Lower Upper Lower Test Upper Test Calibration Imprecision
Calibration Assay ConfidenceConfidence Signficance Signficance Total
1 1 ---------------------- ------------------------- 0.125 0.125 0.25
2 2 0.644989 1.355011 TRUE TRUE 0.125501 0.125501 0.251001
Assay 1 vs Assay 3
Two One-Sided Equivalence
with Assay 1 Variance Components
Lower Upper Lower Test Upper Test Calibration Imprecision
Calibration Assay ConfidenceConfidence Signficance Signficance Total
1 1 ---------------------- ------------------------- 0.125 0.125 0.25
3 3 1.644752 2.355248 TRUE TRUE 0.126002 0.126002 0.252004
Assay 2 vs Assay 3
Two One-Sided Equivalence
with Assay 1 Variance Components
Lower Upper Lower Test Upper Test Calibration Imprecision
Calibration Assay ConfidenceConfidence Signficance Signficance Total
2 2 ---------------------- ------------------------- 0.125501 0.125501 0.251001
3 3 0.563533 1.436467 TRUE TRUE 0.126002 0.126002 0.252004
OVERALL ESTIMATE
140
EPA Traceability Protocol for Gaseous Calibration Standards
*4 and 6 are cases where there is no 3rd assay. In case 4, the two assays share
a common calibration. In case 6, the two assays have different calibrations.
Variance
Expanded Wt. of Wt.
Calibration Estimate Uncert. Raw Wt. Adj. Wt. *Conc. *Est.
1 500 0.100% 100000 0.333 166.667 0.02778
2 500 0.100% 100000 0.333 167.00 0.02789
3 502 0.100% 100000 0.333 167.333 0.02800
The standard error and expanded uncertainty displayed above do not account for
uncertainty in the reference standards used to calibrate the analytical instrument,
or for the uncertainty due to the dilution flow rate. In the first space below, enter
the expanded uncertainty (typically 2 times the standard error) of the reference
standards. If different calibration standards had different uncertainties, enter the
largest.
Example: If NIST SRMs were used in the calibration and their certified
concentrations were 100 +/- 1 ppm, 200 +/- 1 ppm, 300 +/- 2 ppm, 400 +/- 3 ppm
and 500 +/- 4 ppm, then the largest 95% uncertainty is for the 100 ppm standard:
1/100 = 0.01 or 1%.
(SRM uncertainties are expressed as two-sigma limits which are 95% confidence
intervals.)
141
EPA Traceability Protocol for Gaseous Calibration Standards
1 1.002 1.0040
1 0.902 0.8136
1 0.802 0.6432
1 0.701 0.4914
X= 1 0.601 0.3612
1 0.501 0.2510
1 0.401 0.1608
1 0.301 0.0906
1 0.200 0.0400
1 0.100 0.0100
1 0.000 0.0000
Xi,j denotes the number that is found in the ith row and the jth column. X0,1 =
1.002. The first row and column are numbered zero.
A matrix that has only one column is called a column vector, and a matrix that
has only one row is called a row vector.
0.999
0.915
0.828
0.738
y= 0.644 is a column vector.
0.549
0.448
0.346
0.237
0.122
0.001
Subscripts following vector names denote the row or column of the vector. For
example, y1 is the number in the second row of y, 0.915. (Remember that we
begin counting rows with zero.)
Matrix operations that come into play for calibration include multiplication,
142
EPA Traceability Protocol for Gaseous Calibration Standards
transposition, and inversion. The rules for these operations can be found in any
introduction to matrices. We will use the following notation for these operations:
X' denotes the transpose of X (the ith column of X becomes the ith row of X')
1 1 1 1 ... 1
X' = 1.002 0.902 0.802 0.701 ... 0.000
1.0040 0.8136 0.6432 0.4914 ... 0.0000
X'Y denotes multiplication of matrices X' and Y. X' must have the same number
of columns as Y has rows. For the matrix X above,
The linear regression approach is illustrated below for the simple quadratic curve.
The starting point for regression analysis will be a matrix named X. This matrix
will have 3 columns (one for each coefficient to be determined). The number of
rows will be the same as the number of calibration measurements that are
performed by the measurement system. The first column is a vector of 1s. The
second column contains the certified concentrations of the calibration standards.
The third column contains the squares of the values appearing in the second
column. When this matrix is multiplied by the vector of coefficients [b0, b1, b2],
the result is a vector of responses, so that:
yi = b0 + b1 xi + b2 xi2
143
EPA Traceability Protocol for Gaseous Calibration Standards
Now, we're interested in estimating the the coefficients b0, b1, and b2, and we're
also interested in computing how much error is involved when we use the
information to estimate the concentration in an "unknown."
Example
0.0046
b = 1.1837 b' = 0.0046 1.1837 -0.1932
-0.1932
One assumption that underlies the regression approach is that random error is
constant across the measurement range. Sometimes it may be necessary to
apply a transformation in order to achieve this characteristic, called homogeneity
of variance. An estimate of this variance is obtained using matrix operations:
This estimate's "degrees of freedom" (df) is the number of calibration points less
the number of coefficients estimated for the calibration equation.
144
EPA Traceability Protocol for Gaseous Calibration Standards
V = Var * (X'X)-1
Hypothesis tests can be performed and confidence intervals can be estimated for
each coefficient using the coefficient's estimate, the coefficient's variance
(contained in V), and the degrees of freedom, df.
Deriving the confidence intervals for this estimate requires finding two alternative
concentrations, one higher and one lower than the estimate, such that the
probability of having produced a lesser or greater average response is sufficiently
small. For a 95% confidence interval, the lower bound is a concentration whose
response would be less than the observed response with 97.5% probability; the
145
EPA Traceability Protocol for Gaseous Calibration Standards
upper bound is a concentration whose response would be less than the observed
response with 97.5% probability.
0.601 = b0 + b1 C + b2 C2
(b0 - 0.601) + b1 C + b2 C2 = 0
Now, if the concentration really had been at this value, the 95% confidence
interval for the mean response of six measurements would be symmetric about
the observed response:
As above, t = 2.306006
x = 1 0.553935 0.306843 = [1, resp, resp2]
xb = 0.601 (check)
0.550418 to 0.557456
146
EPA Traceability Protocol for Gaseous Calibration Standards
Fortunately, even with the quadratic calibration curve, with good precision, the
confidence intervals will be within a small enough region that the curve is close to
linear and the interval will be very nearly symmetric. The uncertainty criterion for
multipoint calibration requires the 95% confidence interval's half-width to be less
than 1%. The calibrated range of the analyzer extends across all concentrations
for which the criterion is satisfied.
The calibration curve's uncertainty is acceptable for concentrations > 0.21 ppm.
As discussed in Subsection 2.1.5.2, the stability test requires at least three initial
measurements of the candidate standard plus at least three additional
measurements following a period of 7 days or more. The standard's
concentration must be in the well-characterized region of the analyzer per
Subsection 2.1.4.2.
Concentrations are estimated using the calibration curve from at least three
measurements for the initial assay and at least three measurements for the
second assay. Schuirmann’s two one-sided tests (TOST) will be used to test for
equivalence with a acceptance criterion (i.e.,tolerance) of 1.0%. The calculation
follows:
147
EPA Traceability Protocol for Gaseous Calibration Standards
Note that we are using alpha, and not the usual alpha/2, in the t-value.
If this confidence limit is within the TOST acceptance region, which is calculated
as
then the null hypothesis is rejected and the two sets of results are considered
equivalent.
If the confidence limit is not within the TOST acceptance criterion, the stability
test can be repeated after an additional 7 days or more using the second and
third sets of results in the calculations, as above. If another statistical failure is
found, then the candidate standard is determined to be instable and it is
disqualified for further use.
Example:
n1 = 3 n2 = 3
s1 = 0.0020817 ppm x1 = 0.9943333 ppm
s2 = 0.0030551 ppm x2 = 0.9936667 ppm
sp2 = 0.002614065 tol = 1.0%
148
EPA Traceability Protocol for Gaseous Calibration Standards
x1 - x2 = 0.000666667
t(1-alpha,df) = 2.131847
CI lower limit = -0.00388
CI upper limit = 0.00522
TOST lower limit = -0.00873
TOST upper limit = .0.01061
The confidence interval fits in the TOST acceptance region, so the two sets of
measurements are equivalent and the candidate standard can be certified.
6. Recertification
Per Subsection 2.1.11, a standard can be recertified after the certification period
has elapsed, if the standard has been reassayed and the Appendix C
spreadsheet indicates that the TOST acceptance criterion has been attained. To
determine whether the concentration of the standard has changed since the
initial certification, new measurements are made using a pollutant gas analyzer
that has been calibrated according to Subsection 2.1.4.2. Original certification
data are used to provide initial estimates of mean (x1) and standard deviation
(s1). The recertification data are used to estimate a second mean (x2) and
second standard deviation (s2). These are used in Schuirmann’s two one-sided
test (TOST) to determine if the measurements are equivalent. As outlined in the
previous section, for a 95% confidence level a critical value for t is based on a
90% confidence interval on the difference of the means with degrees of freedom
equal to the number of initial and recertification data minus 2. A pooled estimate
of the standard deviation (sp) is derived from s1 and s2. If the confidence interval
is within the TOST acceptance region, then the two sets of measurements are
equivalent and the standard is stable. If not, the standard cannot be recertified.
Example:
n1 = 6 n2 = 3
s1 = 0.0029439 ppm x1 = 0.9956667 ppm
s2 = 0.0026458 ppm x2 = 0.991 ppm
sp = 0.002394438 tol = 1%
ppm
149
EPA Traceability Protocol for Gaseous Calibration Standards
x1 - x2 = 0.00466667
t(1-alpha/2,df) = 1.894579
CI lower limit = 0.00145891
CI upper limit = 0.00787442
TOST lower limit = -0.005266
TOST upper limit = 0.014600
The confidence interval fits within the TOST acceptance region, so the standard
may be recertified. The certified concentration of the standard is the grand mean
of the combined data set.
150
EPA Traceability Protocol for Gaseous Calibration Standards
Enter the calibration data in the shaded spaces below. The first column (I)
simply counts the calibration points that you enter. The second column (X) is for
the certified concentrations of the calibration gas standards. The third column
(Y) is for the instrument responses corresponding to the calibration standards.
The number of points cannot exceed 50.
i Xi Yi Color Code
1 0.0000 0.1900
2 0.0000 0.2000 red = derived value (protected)
3 0.0000 0.2100
blue = entered value
4 1.0000 1.1400 (unprotected)
5 1.0000 1.1500
6 1.0000 1.1600 black = fixed text (protected)
7 2.0000 1.9900
8 2.0000 2.0000
9 2.0000 2.0100
10 3.0000 2.7400
11 3.0000 2.7500
12 3.0000 2.7600
13 4.0000 3.3900
14 4.0000 3.4000
15 4.0000 3.4100
16 5.0000 3.9400
17 5.0000 3.9500
18 5.0000 3.9600
Review the estimates of the coefficients (b0, (b1,...) for the linear and quadratic
models, their confidence, and the residual error variances (s2).
151
EPA Traceability Protocol for Gaseous Calibration Standards
For each model, the coverage factor, t, is determined from the t-distribution with
appropriate degrees of freedom (df).
Note that the uncertainties are not displayed, but can easily be derived from the
estimate, expanded uncertainty, and the coverage factor.
F ratio = 219.69
F critical = 2.3849 (5% significance level)
152
EPA Traceability Protocol for Gaseous Calibration Standards
F ratio = 0.9286
F critical = 2.5536 (5% significance level)
The error variances are not significantly different at the 5% level. The quadratic
model appears to be a better choice than cubic.
F ratio = 0.9286
F critical = 2.5536 (5% significance level)
The error variances are not significantly different at the 5% level. The cubic
model appears to be a better choice than quartic.
View the charts named Curves 1 and Residuals 1. Curves 1 shows bands
illustrating expanded uncertainties for the estimated regression. Compare these
bands with those of the quadratic regression, Curves 2. (Note: You can change
the associated approximate confidence level by changing the "p-value" in cell F4
of the worksheet named Chart Data.) Residuals 1 shows how the calibration
points deviated from the calibration line. Look for a simple pattern (such as a
quadratic curve) in the chart. If such a pattern appears, the quadratic model may
be better. View Residual 2, the deviations from the best- fit quadratic curve. If
Residual 2 effectively removes the simple pattern observed in Residual 1 and if
the magnitude of the deviations has been significantly reduced (as evidenced by
a reduction in the uncertainty u2), then the quadratic model is superior. An F-test
can be run to determine if the two overall uncertainties of fit are significantly
different.
153
EPA Traceability Protocol for Gaseous Calibration Standards
This test can only be used when using the linear or quadratic model.
Enter the impurity specification for the zero air material estimation and prediction
0.2821
STEP 5 Test for Zero Material on Different Day from Initial Calibration
This step applies to zero material tests on a different day than the initial
calibration.
This test can only be used when using the linear or quadratic model. Before the
zero material limit is input, the measurement system is challenged with zero and
span checks. Three or more discrete checks of the zero gas and three or more
checks of each span gas are made. If necessary, second span should be higher
concentration than the first. Enter the results below:
154
EPA Traceability Protocol for Gaseous Calibration Standards
Coefficients are not significantly different. Consider including the new data as
part of original calibration (Step 1).
0.2916
155
EPA Traceability Protocol for Gaseous Calibration Standards
The general workflow for this workbook is from right to left. The steps involved
are:
To be done annually:
Zero Gas Annual Calibration
Pollutant Annual Calibration
Annual Gas Check
To be done monthly:
Monthly Flow Check
To be done on day of use:
Day of Use
Organization
The workbook consists of several worksheets, which are displayed as tabs at the
bottom of the screen. The functions of these worksheets are described below:
Read Me (this sheet) - describes the workbook, explaining how to use the
worksheets
Monthly Flow Check - to verify that GDS calibration has not shifted since
annual calibration
Annual Gas Check - to be performed after annual calibration to verify that the
GDS actually produces the concentrations of diluted mixtures for both
nonreactive and reactive gases that agree with theoretical concentrations
For both the Pollutant Annual Calibration and Zero Gas Annual Calibration:
Calibration - allows for user input of calibration and other analytical data
Curves 1 - displays the calibration data, the best-fit line, and expanded
uncertainties.
156
EPA Traceability Protocol for Gaseous Calibration Standards
Curves 2 - displays the calibration data, the best-fit quadratic curve, and
expanded uncertainties.
Chart Data - includes the data used to create the Curves and Residuals
charts
Linear: y = b0 + b1x+ e
Quadratic: y = b0 + b1x + b2x2 + e
It is assumed that the valve setting standards (x) have negligible uncertainties.
The output quantities are estimated parameters (bs) and related uncertainties.
The workbook then helps the user perform the following functions:
Conventions
The Pollutant Gas and Zero Gas Annual Calibration worksheets includes
instructions that guide the user through the steps in its use. Each worksheet is
also color coded to simplify use. Shaded cells that are bordered in blue text are
for input of data. These cells are unprotected, but all other cells of the
Calibration worksheet are protected.
Derived values and statements are colored red. These cells contain formulas
and are protected to prevent alteration.
157
EPA Traceability Protocol for Gaseous Calibration Standards
Enter the annual pollutant flow rate calibration data in the shaded spaced below.
The first column (i) simply counts the calibration points. The second column (X)
is for the pollutant flow controller settings. The third column (Y) is for the flow
rate reference standard readings corresponding to these settings. The number
of points cannot exceed 50.
i Xi Yi Color Code
1 78.0000 78.4529
2 70.0000 70.1984 red = derived value (protected)
3 60.0000 60.2776 blue = entered value (unprotected)
4 50.0000 50.1985 black = fixed text (protected)
5 40.0000 40.0262
Review the estimates of the parameters (b0, b1, …) for the linear and quadratic
models, their expanded uncertainties and the overall uncertainty of the fit (u2).
158
EPA Traceability Protocol for Gaseous Calibration Standards
View the charts named Curves 1 and Residuals 1. Curves 1 shows bands
illustrating expanded uncertainties for the estimated regression. Compare these
bands with those of the quadratic regression, Curves 2. (Note: You can change
the associated approximate confidence level by changing the "p-value" in cell F4
of the worksheet named Chart Data.) Residuals 1 shows how the calibration
points deviated from the calibration line. Look for a simple pattern (such as a
quadratic curve) in the chart. If such a pattern appears, the quadratic model may
be better. View Residual 2, the deviations from the best- fit quadratic curve. If
Residual 2 effectively removes the simple pattern observed in Residual 1 and if
the magnitude of the deviations has been significantly reduced (as evidenced by
a reduction in the uncertainty u2), then the quadratic model is superior. An F-test
can be run to determine if the two overall uncertainties of fit are significantly
different.
The overall uncertainties of fit are not significantly different at the 5% level.
The linear model appears to be the better choice.
This worksheet should be used for the annual nonreactive gas calibratoin check
(Sec. 4.1.6.3) and the annual reactive gas calibration check (Sec. 4.1.6.4). It
should only be used after filling out the Pollutant Annual Calibration worksheet
and the Zero Gas Calibration Worksheet.
159
EPA Traceability Protocol for Gaseous Calibration Standards
Enter the expanded uncertainty (U) from the reference standard's certificate of
analysis. This value is the coverage factor (k=2) times the combined standard
uncertainty (uc). For example, absolute values are given in ppm and relative
values are given in percent. Enter the value as a decimal value (e.g., 1 percent
should be entered as 0.01)
Enter the expanded uncertainty (U) from the check standard's certificate of
analysis. This value is the coverage factor (k=2) time the combined standard
uncertainty (uc). It is necessary to answer whether or not this is absolute or
relative. For example, absolute values are given in ppm and relative values are
given in percent Enter the value as a decimal value (e.g., 1 percent should be
entered as 0.01)
STEP 3 Enter Settings for Pollutant and Zero Gas Flow Controllers
Select settings for the gas flow controllers that fall in the same range as were
used in the annual calibrations and that produce a diluted concentration of the
reference standard that is equal to the undiluted concentration of the check
standard.
160
EPA Traceability Protocol for Gaseous Calibration Standards
Adjust the pollutant and zero gas settings so that these two concentrations match
within 1 percent. Do not proceed until the Match value is less than 1.0 percent.
161
EPA Traceability Protocol for Gaseous Calibration Standards
Mean response for reference std Mean response for check std
1.000 1.000
Relative combined std uncertainty for std concs, flow rates, and anal responses
0.016163
Lower limit for diluted conc Upper limit for diluted conc.
0.39 0.60
This worksheet should only be used after filling out the Annual Gas Calibration
Check worksheet
Enter expanded uncertainties (U) for both pollutant gas and zero gas flow rate
reference standards
Enter flow controller setting within range of the annual calibration for both
pollutant gas and zero gas
162
EPA Traceability Protocol for Gaseous Calibration Standards
Enter the flow rates measured by the flow rate reference standard
5. Day-of-Use Worksheet
This worksheet should only be used after filling out the Monthly Flow Check
worksheet
It is only valid within one month of 04/30/12
163
EPA Traceability Protocol for Gaseous Calibration Standards
Enter the relative expanded uncertainty (U/conc) from the reference standard's
certificate of analysis. This value is the coverage factor (k=2) times the
combined standard uncertainty (uc). Enter the value as a decimal value (e.g., 1
percent should be entered as 0.01)
STEP 2 Enter Desired Flow Rates for Pollutant and Zero Gas
Enter the desired flow rates for both the pollutant and zero gas to achieve your
target dilution and obtain the proper flow controller settings to achieve these flow
rates.
Pollutant Gas Flow Controller Setting Zero Gas Flow Controller Setting
59.76781 5906.997
Pollutant flow rate relative std Zero gas flow rate relative std
uncertainty from calibration uncertainty from calibration
0.001516 0.00303
Pollutant gas flow rate relative Zero gas flow rate relative
combined std uncertainty combined std uncertainty
0.005225 0.005846
164
PRESORTED STANDARD
POSTAGE & FEES PAID
EPA
PERMIT NO. G-35