DD+DIS110.09E - Installation Instructions Digitizer - Software - C25 - 3102
DD+DIS110.09E - Installation Instructions Digitizer - Software - C25 - 3102
Imaging Services
CR 25.0
Type 5156/100/105
► Referenced Documents
Document Title
DD+DIS112.09E CR 25.0 Type 5156/100 Up to SN < 6000
Service Bulletin:
Announcement of Digitizer Software C25_3102
DD+DIS111.09E CR 35-X Type 5158/100 and
CR 25.0 Type 5156/105 As of SN ≥ 6000
Service Bulletin:
Announcement of Digitizer Software C25_3102
NOTE:
Dispose the enclosure document after finishing your task
(observe the confidentiality note below).
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Revision 0
05-2009 printed in Germany Document Node ID: 27300476
Enclosure_e_template_v07
Agfa Company Confidential Copyright © 2009 Agfa HealthCare N.V.
DD+DIS110.09E Enclosure
► Manufacturer
Agfa HealthCare N.V.
Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright © 2009 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
LIST OF CONTENTS
4 INSTALLATION VIA FLOPPY DISKS (CR 25.0 TYPE 5156 / 100 / 105 UP TO SN < 6500,
CR 35-X TYPE 5158 / 100 UP TO SN < 4000) ......................................................................13
4.1 Creating Installation Floppy Disks...........................................................................................13
NOTE:
The new software is upgradeable from every previous version of C25_xxxx* software.
XXXX = Version number of the software
REQUIRED TOOLS:
• For installation via floppy disk
Applicable for:
CR 25.0 Type 5156 / 100 / 105 up to SN < 6500,
CR 35-X Type 5158 / 100 up to SN < 4000:
- 1 floppy disk for the backup of site specific data
- 7 floppy disks for the installation files
- Service PC with floppy disk drive
9 pins 25 pins
Figure 1
REQUIRED TIME:
• Approximately 30 minutes for the software upgrade via floppy discs
• Approximately 60 minutes for the installation via floppy disks on a
new hard disk or when corrupted files detected
• Approximately 10 minutes for the installation via USB Memory Stick
(software upgrade/new hard disk/reformatted hard disk)
NOTE:
Register this software installation in the local service management system as
modification, e.g. in SMS as “MOD”.
Following steps are necessary to install software via USB Memory stick:
(1) Download the Digitizer Software C25_3102 (File size approx. 8 MB) from:
• MEDNET GSO => Computed Radiography => CR Digitizer => CR 25.0 =>
Freeware => CR 25.0 - Software - Device Software - C25_3102
or
• MEDNET GSO => Computed Radiography => CR Digitizer => CR 35-X =>
Freeware => CR 35-X - Software - Device Software - C25_3102
(2) Extract all files from the downloaded zip file into a folder on the Service PC.
Among the extracted files is the zip file c25_3102.zip.
(3) Extract all files from c25_3102.zip with extraction mode <Extract to here>.
Folder C25_3102 including seven sub-folders is created automatically.
(4) Copy the folder C25_3102 to the USB Memory Stick. The created structure
shown in Figure 1 must be maintained (“Root” is the USB Memory Stick).
NOTE:
C25_<SerialNumber>.bac is optional, for the case, that a
backup has been made previously.
IMPORTANT:
The digitizer checks light incidence.
Following error message may appear on the screen of the digitizer:
Digitizer locked
RIGHT SIDE PANEL NOT CLOSED.
Close right side panel.
Code 24791
Figure 2
If this error message appears, lean the cover on the side of the device.
Then continue with next steps.
(3) Enter the Service menu on the digitizer keypad and select
<3 Save data> Æ <2 Machine specific data>.
(4) Follow the instructions on the digitizer display.
NOTE:
If the installation fails, repeat installation with installation method like
“Installing Software C25_3102 on new Hard Disk” in section 3.3.3.
NOTE:
If the installation fails, repeat installation with installation method like
“Installing Software C25_3102 on new Hard Disk” in section 3.3.3.
3.3.3 Installing Software C25_3102 on new Hard Disk or when corrupted Files detected
NOTE:
Without restoring the device-specific parameters, an operation of the digitizer is
impossible.
(1) Enter the Service menu on the digitizer keypad and select
<5 INSTALL data / 2 Machine specific data>.
(2) Follow the guidelines on the digitizer display.
Use the backup that was created with the previous software version before the
Hard Disk was replaced by a new Hard Disk or if corrupted files detected.
(1) Restore the test image files from CD to the directory E:\testimg on the
digitizer. For current version please refer to the service documentation,
chapter 3.2 Tools and auxiliary means.
(2) Press the <Reset> switch on the CPU-Board.
NOTE:
Between restoring the backup and installing the CPF, a reset is necessary!
(1) Get the CPF file either from a previous backup or from the Processing Station.
Otherwise generate the CPF file with the QS, the VIPS or the CCM tool.
Please refer to the respective service documentation, chapter 3.2, for details.
(2) Enter the Service menu on the digitizer keypad and select
<5 INSTALL data> / <6 CPF-File>.
(3) Follow the instructions displayed on the digitizer display.
(4) Reconnect the digitizer to the network.
(5) Disconnect the service PC.
(6) Switch off and switch on the digitizer.
(7) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
You can create installation floppy disks either using the wizard (see 4.1.1) or manually
(see 4.1.2).
(1) Download the Digitizer Software C25_3102 (File size approx. 8 MB) from:
MEDNET GSO => Computed Radiography => CR Digitizer => CR 25.0 =>
Freeware => CR 25.0 - Software - Device Software - C25_3102
(2) Extract all files from the downloaded zip file into a folder on the Service PC.
Among the extracted files is the zip file c25_3102.zip.
(3) Extract all files from c25_3102.zip with extraction mode <Extract to here>.
Folder C25_3102 including seven sub-folders is created automatically.
(4) Take 7 empty, formatted diskettes.
(5) Double-click on <Start.bat>. The copy process starts automatically.
(6) Insert an empty floppy disk and follow the instructions on the screen.
(7) When requested, remove floppy disk from the floppy disk drive and label it with the
corresponding floppy name, e.g. C253102_1_3 for the first disk of installation files
(for a filename overview refer to table 1 on next page).
(8) Follow the instructions and insert the next floppy disks.
When all 7 floppy disks were copied, the prompt will be closed automatically.
NOTE:
Check the number of floppies after finishing the copy process.
There must be 7 floppy disks.
Result The set of floppy disks for installation are labeled with the correct name and contain the
correct files for the installation.
(1) Download the Digitizer Software C25_3102 (File size approx. 8 MB) from:
MEDNET GSO => Computed Radiography => CR Digitizer => CR 25.0 =>
Freeware => CR 25.0 - Software - Device Software - C25_3102
(2) Extract all files from the downloaded zip file into a folder on the Service PC.
Among the extracted files is the zip file C25_3102.zip.
(3) Extract all files from c25_3102.zip with extraction mode <Extract to here>.
Folder C25_3102 including seven sub-folders is created automatically.
NOTE:
For each of these folders, a floppy disk with the corresponding floppy
name must be created. Follow the instructions below.
Result The set of floppy disks for installation are labeled with the correct name and contain the
correct files for the installation.
IMPORTANT:
The digitizer checks light incidence.
Following error message may appear on the screen of the digitizer:
Digitizer locked
RIGHT SIDE PANEL NOT CLOSED.
Close right side panel.
Code 24791
Figure 3
If this error message appears, lean the cover on the side of the device.
Then continue with next steps.
(3) Enter the Service menu on the digitizer keypad and select
<3 Save data> Æ <2 Machine specific data>.
(4) Follow the instructions on the digitizer display.
NOTE:
During the installation from floppy disks the digitizer shows no status report.
It is necessary to wait after inserting of a floppy, until the LED of the digitizers
floppy disk drive stops flashing.
If you want to see the installation status during the software installation, connect
the Service PC via serial connection cable (RS232) to the digitizer. By using the
Hyper Terminal the several installation steps of the digitizer can be followed.
(1) Insert the floppy disk labeled PORTEX1 into the floppy disk drive of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of the floppy disk drive is off.
(4) Remove the floppy disk.
(1) Insert the floppy disk labeled PORTEX2 into the floppy disk drive of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of the floppy disk drive is off.
(4) Wait additionally 1 minute, to be sure that all files are unzipped.
(5) Remove the floppy disk.
NOTE:
Do not insert any other floppy disk!
(1) Insert the floppy disk labeled C253102_1_3 into the floppy disk drive of the
digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of floppy disk drive is off.
(4) Remove the floppy disk.
(1) Insert the floppy disk labeled C253102_2_3 into the floppy disk drive of the
digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of floppy disk drive is off.
(4) Remove the floppy disk.
(1) Insert the floppy disk labeled C253102_3_3 into the floppy disk drive of the
digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the control LED of floppy disk drive is off.
(4) Wait additionally 1 minute, to be sure that all files are unzipped.
(5) Remove the floppy disk.
NOTE:
Do not insert any other floppy disk!
(6) Close the digitizer. The device will restart automatically.
(7) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
NOTE:
If the installation fails, repeat installation with installation method like
“Installing Software C25_3102 on new Hard Disk”.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
4.3.2 Installing Software C25_3102 on new Hard Disk or when corrupted files detected
NOTE:
This procedure must be carried out only if the hard disk of the digitizer has been
replaced by a new one or if corrupted files detected on the hard disk.
In this case the hard disk must be formatted using the installation disk HDFORMAT.
(1) Insert the floppy disk labeled HDFORMAT into the floppy disk drive of the
digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Confirm partitioning/formatting of the hard disk on your terminal with <y>.
(4) Wait until the system prompt is displayed on the terminal.
(5) Remove the floppy disk.
(1) Insert the floppy disk labeled PORTEX2 into the floppy disk drive of the digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the system prompt is displayed on the terminal.
(4) Remove the floppy disk.
NOTE:
Do not insert any other floppy disk!
(1) Insert the floppy disk labeled C25_3102_1_3 into the floppy disk drive of the
digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the system prompt is displayed on the terminal.
(4) Wait additionally 1 minute, to be sure that all files are unzipped.
(5) Remove the floppy disk.
(1) Insert the floppy disk labeled C25_3102_2_3 into the floppy disk drive of the
digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the system prompt is displayed on the terminal.
(4) Remove the floppy disk.
(1) Insert the floppy disk labeled C25_3102_3_3 into the floppy disk drive of the
digitizer.
(2) Press the <Reset> switch on the CPU-Board.
(3) Wait until the system prompt is displayed on the terminal.
Wait additionally 1 minute, to be sure that all files are unzipped.
(4) Remove the floppy disk.
(5) Press the <Reset> switch on the CPU-Board.
(6) Wait until the system prompt is displayed on the terminal.
The system extracts the C25_3102 software.
(7) Switch off and then switch on the digitizer.
(8) Wait until the digitizer is ready.
NOTE:
Without restoring the device-specific parameters, an operation of the digitizer is
impossible.
(1) Enter the Service menu on the digitizer keypad and select
<5 INSTALL data / 2 Machine specific data>.
(2) Follow the guidelines on the digitizer display.
Use the backup floppy disk that was created with the previous software version
before the Hard Disk was replaced by a new Hard Disk or if corrupted files
detected.
(1) Restore the test image files from CD to the directory E:\testimg on the
digitizer. For current version please refer to the service documentation,
chapter 3.2 Tools and auxiliary means.
(2) Press the <Reset> switch on the CPU-Board.
NOTE:
Between restoring the backup and installing the CPF, a reset is necessary!
(1) Get the CPF file either from a previous backup or from the Processing Station.
Otherwise generate the CPF file with the QS, the VIPS or the CCM tool.
Please refer to the respective service documentation, chapter 3.2, for details.
(2) Enter the Service menu on the digitizer keypad and select
<5 INSTALL data> / <6 CPF-File>.
(3) Follow the instructions displayed on the digitizer display.
Have a floppy disk ready for the backup of device-specific parameters.
(4) Reconnect the digitizer to the network.
(5) Disconnect the service PC.
(6) Switch off and switch on the digitizer.
(7) Check if the digitizer performs its self test and afterwards changes
its state to Ready.
If the digitizer does not perform the self test, the software may not be installed
successfully. This can be caused by a corruption of the file system on the Hard Disk.
In this case perform the installation according to:
Section 4.3.2 Installing Software C25_3102 on a new Hard Disk
and then section 5 Installation of C25_3102 Language Files.
NOTE:
With software version C25_3007 the term “floppy” has been exchanged with the term
“medium”.
To install C25_3102 language files and to change the language of the digitizer user
interface perform following procedure:
NOTE:
If you use the digitizer for Mammography applications, enable “Mammo” manually
in the Service Menu of the digitizer.
(1) Enter the Service menu on the digitizer keypad and select
<1 INFO> / <1 Device info>.
Result If the entry Software is C25_3102, the software installation was successful.
Otherwise repeat the installation.