Manuel Du Spectrophotomètre Jenway 7310 PDF
Manuel Du Spectrophotomètre Jenway 7310 PDF
Operating Manual
r:
Bibby Scientific 731 005 REV Ef12-10
2
Safety
Please read this information carefully prior to installing or using this equipment.
1. The unit described in this manual is designed be operated only by trained personnel. Any adjustments,
maintenance and repair must be carried out as defined in this manual, by a person qualified to be aware
of the hazards involved.
2. It is essential that both operating and service personnel employ a safe system of work, in addition to
the detailed instructions specified in this manual.
3. Other than for those items defined in the maintenance procedures herein there are no user serviceable
items in this instrument. Removal of covers and attempted adjustment or service by unqualified personnel
will invalidate the warranty and may incur additional charges for repair.
4. References should always be made to the Health and Safety data supplied with any chemicals used.
Generally accepted laboratory procedures for safe handling of chemicals should be employed.
s. If it is suspected that safety protection has been impaired in any way, the unit must be made inoperative
and secured against any intended operation. The fault condition should immediately be reported to the
appropriate servicing authority.
3
4
Contents
Page
Safety 3
SECTION 1 - Introduction 8
1.1 Instrument description 8
1.2 Instrument specification 8
SECTION 2 - Installation 12
2.1 Unpacking 12
2.2 Installation 12
2.3 Display 13
2.4 Controls 14
2.5 Rear panel 15
2.6 Front panel 15
SECTION 5 - Photometrics 25
5.1 Mode specific parameters 25
5.2 Method set up 25
5.2.1 Selecting a wavelength 26
5.3 Calibration 26
5.4 Sample measurment 26
SECTION 6 - Concentration 27
6.1 Mode specific parameters 27
6.2 Method setup 27
6.2.1 Selecting a wavelength 27
6.2.2 Settings 28
6.2.2.1 Selecting concentration units 28
6.2.2.2 Changing the resolution 29
6.2.2.3 Using a standard 29
5
6.2.2.4 Using a factor 29
6.3 Calibration 29
6.3.1 Calibrating to a standard 29
6.3.2 Calibrating to a factor 30
6.4 Sample measurement 30
6.4.1 Measuring a sample after calibrating to a standard 30
6.4.2 Measuring a sample after calibrating to a factor 30
SECTION 7 - Spectrum 31
7.1 Mode specific parameters 31
7.2 Method setup 32
7.2.1 Scan settings 32
7.2.1.1 Selecting absorbance or % transmittance 32
7.2.1.2 Setting start and end wavelengths 32
7.2.1.3 Setting the scan interval 33
7.2.1.4 Y-axis scaling 34
7.3 Calibration 34
7.4 Sample measurement 34
7.5 Data analysis 35
7.5.1 Peaks and valleys threshold 35
7.5.2 Peaks and valleys table 36
7.5.3 Spectral points analysis 36
SECTION 8 - Quantitation 38
8.1 Mode specific parameters 38
8.2 Method setup 39
8.2.1 Selecting a wavelength 39
8.2.2 Quantitation table 39
8.2.2.1 Selecting number of standards 39
8.2.2.2 Selecting concentration units 39
8.2.2.3 Changing the resolution 40
8.2.2.4 Selecting absorbance or % transmittance 40
8.2.2.5 Adding standards 40
8.2.3 Standard curve 40
8.2.3.1 Creating a new standard curve 41
8.3 Calibration 42
8.4 Sample measurement 42
8.5 Data analysis 43
SECTION 9 - Kinetics 44
9.1 Mode specific parameters 44
9.2 Method set up 45
9.2.1 Kinetics settings 45
9.2.1.1 Y-axis scaling 45
9.2.1.2 Setting lag time or start on level 46
9.2.1.3 Selecting absorbance or % transmittance 46
9.2.1.4 Changing the resolution 47
9.2.1.5 Selecting concentration units 47
9.2.1.6 Using a standard 47
9.2.1.7 Using a factor 47
9.2.1.8 Selecting a wavelength 48
6
9.2.1.9 Setting the kinetics measurement time 48
9.3 Calibration 48
9.4 Sample measurement 48
9.5 Data analysis 49
7
12.2.2 Xenon lamp module replacement 71
12 .3 Service 72
SECTION 13 - Troubleshooting 73
13.1 Error codes 73
13.2 Troubleshooting guide 75
13.3 Technical support 75
Index 85
8
9
SECTION 1 - Introduction
The 7310 and 7315 spectrophotometers are suited to a wide range of applications in education,
quality control, environmental and clinical analysis. The 7310 is a visible spectrophotometer covering
a wavelength range from 320nm to lOOOnm. The 7315 is a UVNisible spectrophotometer with a
wavelength range from 198nm to 1 OOOnm. Both models have five measurement modes: photometrics,
concentration, spectrum scanning, quantitation and kinetics. These instruments use icon driven software
and have an improved navigation system for easy and intuitive usability.
7310 7315
Wavelength
Range 320 to 1 OOOnm 198 to 1 OOOnm
Resolution 1nm
Accuracy ±2nm
Repeatability ± O.Snm
Spectral bandwidth Snm
Photometrics
Transmittance o to 199.9%
Absorbance -0.300 to 2.SOOA
Accuracy ± 1 %T. ±0. O 1 Abs at 1 . 000 Absorbance
Resolution 0.1 % T. 0.001A
Stray light <0.5% at 340nm <0.5% at 340nm and 220nm
Concentration
Range -300 to 9999
Resolution Selectable 110.110.0110.001
Calibration Blank with a single standard or factor
Units no units, %, ppm, EBC, SRM, mEq/1, mEq, M, mM, µM, nM, U, U/1, U/ml. g/l, mg/I.
µg/l, ng/1, g/dl, mg/di, µg/dl, mg/ml, µg/ml, ng/ml, µg/µI, ng/µI, mot/I, mmol/1
Factor 0.001 to 10000
Standard 0.001 to 1000
Quantitation
Range -300 to 9999
Resolution Selectable 1/0.1/0.01/0.001
Calibration Blank with up to 6 standards
Units no units, %, ppm, EBC, SRM, mEq/1, mEq, M, mM, µM, nM, U, U/1, U/ml, g/l,
mg/I, µg/1, ng/1, g/dl, mg/di, µg/dl, mg/ml, µg/ml, ng/ml, µg/µI, ng/µI, mot/I, mmol/1
Curve fit algorithms Quadratic, quadratic through zero, linear, linear through zero, interpolate
Kinetics
Measurement Time 2 to 9999 seconds
Calibration Blank with a single standard or factor
Display Concentration, rate of change, initial and final absorbance/% T
Resolution Selectable 1/0.1/0.01/0.001
10
7310 7315
Spectrum
Range 320 to 1 OOOnm 198 to 1000nm
Scan interval Selectable 1, 2 or Snm
Analysis Absorbance or % transmittance and peak and valley wavelengths
Other
Beam height 15mm
Light source Tungsten halogen lamp Xenon lamp
Lamp save Yes Not applicable
GLP Current time and date, user ID, settings lock and method lock
Number of users 999
Methods memory 48 in each measurement mode
Results memory Limited by attached mass storage device
Removable media USB (supplied)
Outputs USB, Analogue, RS232, Internal printer
Power 24V
Size (w x d x h) 275 x 400 x 220mm
Weight 6kg
11
SECTION 2 - Installation
2.1 UNPACKING
Remove the 731 O or 7315 from the packaging and ensure the following items are included:
1. Model 7310 spectrophotometer (731 001 ). or Model 7315 spectrophotometer (731 501)
5. Jenway 73 series PC software (735 100) and interface cable (013 203)
2.2 INSTALLATION
The unit should be placed on a clean flat surface which is free from drafts and vibrations. The units are
designed for operation on 90V to 264V AC input at 47 to 63Hz. Select the correct plug attachment and
attach to the power supply unit as shown below:
Connect the power supply unit to the power inlet socket on the rear panel of the instrument and
connect to the mains socket Turn the power on at the mains and switch the instrument on using the
power switch on the rear of the instrument.
The instrument will perform several power on tests before displaying the main menu:
12
1
JENWAY 3
../
7310 ../
'
2
../ 4
09•02
5
Fig 2.2.2 - All Power On Tests Complete
2.3 DISPLAY
These spectrophotometers have a dot matrix display which enables icons and graphs to be displayed
clearly. Following successful completion of the power on tests the main menu screen will be displayed:
1 JENWAY 8
7310
2 7
4 5 6
13
2.4 CONTROLS
The keypad used for these models enables an easy and effective way of navigating the different
measurement modes, entering numbers. saving and analysing results. The soft keys are active when an
icon is displayed above or adjacent to the key. The only exception to this is the back key which is always
active.
The main menu screen and surrounding keypad is displayed below.
1 JENWAY 8
1 7310 8
2 7
2 7
4 5 5 5 6
14
2.5 REAR PANEL
3 OlKt,inact paw11r~
l'fflll;'fll'l9CICl'41H'.
~t-,.,.tQI~
-11.
@) An•log
5 ----~-----k>l output 1
15
SECTION 3 - Theory and Practice of Spectroscopy Measurements
A= EI c
A is the absorbance
E is the molar absorption coefficient (I rnot'cm')
c is the concentration (mol 1·1)
is the path length (cm)
This law shows that absorbance is linear to concentration but this is only true for low concentrations. For
absorbance levels above 3 the concentration starts to move away from the linear relationship.
Transmittance is the proportion of the light which passes through the sample:
Where:
Therefore: T = 11
A = log 1
16
The optical layout of the 7310 and 7315 spectrophotometers is shown below:
/ Entrance slit
Exit Slit
Sample
Collecting Lens
The light from the pre-focused tungsten halogen (731 O) or pre-aligned xenon (7315) lamp is focused
onto the grating, with 1200 lines per millimeter, which separates the light into discreet wavelengths.
The diffracted spectrum of light then passes through a further slit and lens arrangement before passing
through the sample in the sample chamber from left to right. The light which is not absorbed by the
sample is transmitted through a collecting lens and onto the signal detector. The photo-diode detector
used is mounted directly onto the detector PCB and the output is used to calculate the % transmittance.
The result is displayed either as % transmittance or absorbance on the instrument display.
3. The sample chamber lid must be fully closed during measurement and before any readings are
recorded or printed.
4. The correct selection of sample containers is imperative for accurate and reproducible results:
a) Check that the material of the sample container is compatible with the wavelengths to be used
for measurement. In general glass can only be used down to 360nm or 320nm depending on quality.
Standard plastic cuvettes can be used down to 320nm. Special UV versions can be used down to
260nm. Below this level quartz cuvettes must be used.
b) Plastic disposable cuvettes should only be used ONCE.
c) Glass cuvettes should be thoroughly cleaned after use. Discard when scratches become evident on
optical surfaces.
d) Care should be taken when selecting semi-micro or micro cuvettes. The cuvette window on the
17
inner chamber (the area filled with sample) must be wider than the aperture in the sample holder or
light will reach the detector without passing through the sample. In this case. semi-micro or micro
cuvettes with self-screening black surrounds must be used or, alternative holders for these cuvettes
should be used.
e) Glass test tubes and other sample tubes should be used with care. Where possible, matched tubes
should be used and any index mark set to the correct position before measurements are made.
f} Ensure any sample containers used are compatible with the constituents of both the samples and
standards they are to hold. Plastic cuvettes are not compatible with organic solvents.
g) All sample containers must be handled with care; by the top, bottom and non-optical surfaces only.
Any finger marks evident must be removed by a suitable cleaning process.
h) Flow-through cuvettes must be selected with care and consideration for the sample type, sample
volume. pumping system. rinse, sample and waste handling to be used.
5. Samples and standards should not be stored in open cuvettes or sample containers as evaporation will
change the value and lead to staining of the walls which may be irreversible. If stored in stoppered and
sealed cuvettes, they should be filled with little or no air space and the values regularly checked against
a reference standard or quality control material.
6. Samples should be allowed to equilibrate to ambient temperature before measurement (unless a
suitable temperature controlled sample holder is in use). Temperaturechange during measurement may
cause air bubbles to form on the walls of the sample holder. This is a common cause of drift during
measurement.
7. In the preparation of samples and standards high grade borosilicate glass and AR grade chemicals
and reagents must be used. Good quality deionised water or other suitable solvents must be used for
dissolving or diluting samples, chemicals and reagents.
8. All measurements require calibration to a blank, for maximum accuracy this should be prepared
with care using the same deionised water or solvent used for dissolving or diluting the sample. Where
reagents are added to the sample to produce a colour proportional to its concentration a 'sample based'
blank should be used. In this case the blank should consist of all reagents or chemicals to be used,
except the sample which will produce the colour to be measured.
9. Deviations from the Beer-Lambert Law may occur at high and low concentrations giving non-linear
response during sample concentration measurements. For all new methods a linear range should be
defined by the preparation of a calibration curve. The quantitation mode may be used to construct such
a curve against which sample results are automatically measured.
10. Cuvettes and sample holders must be filled to a minimum level which covers the light path. All
Jenway spectrophotometers have a beam height of 1 Smm.
11. The instrument must be calibrated to zero absorbance/100% transmittance prior to taking readings.
In the spectrum measurement mode a baseline scan must be performed before performing a sample
scan.
18
SECTION 4 - Instrument Setup
Photometrics
measurement
mode Instrument
settings menu
To navigate around the spectrophotometer screen press the soft keys adjacent to icons displayed on the
screen. In the main menu either of the two soft keys adjacent to the measurement mode icon can be
pressed to access the mode. There is a back key which returns to the previous menu without saving any
changes.
The main menu screen provides access to all five measurement modes, the time and date menu and
the instrument settings menu. The measurement modes are spectrum, photometrics, quantitation,
concentration and kinetics. The instrument settings menu enables access to settings lock, security codes,
method lock, mode selection, user ID, screen contrast and lamp save menus.
--
measurement parameters can be accessed through the
·11"(
100.0Y.T J4L
settings menu which is displayed in the top right hand
400rm corner of the screen. The only mode where this function
09·02 D0 D is not available is the photometrics mode; instead a
toggle icon is displayed which is used to change the
19
The measurement settings can be accessed through the utility toolbar displayed on the left hand side
of the expanded operating menu. This toolbar provides the same functions in all of the measurement
modes. The utility toolbar enables access to printing, print setup options, opening, saving and deleting
results and methods and autologging options. For more details on the different functions of the utility
tool bar refer to section 10.
The time and date menu enables the current time and
JENWAY date to be set. This information will be saved on all results
7310 and displayed on printouts. The time and date menu can
be accessed from the main menu by holding the key
below the time and date icon for 2 seconds. Pressing
0902 x the key once cycles the display between time and date.
In the time and date menu to set the time press the key
II adjacent to the clock icon. Select the digit to be changed
using the keys at the bottom of the screen. Use the keys
E adjacent to the arrow icons to increase or decrease the
number. The clock function uses a 24 hour format.
D 4 5 3
In the time and date menu to set the date press the
·-·~.
key adjacent to the calendar icon. Select the digit to
be changed using the keys at the bottom of the screen.
The instrument settings menu is accessed by pressing the key below the instrument settings icon in
the main menu. This menu enables access to settings lock, security code, method lock, mode selection,
diagnostics, user ID, screen contrast and lamp save menus. The tick icon saves any changes made and
returns to the main menu.
~~~~~~~~~---
Method lock ------1-1..
contrast
Mode selection-----·~ [i+I ,-.
.....~~ .... t~. . . --~---- Lamp save
Diagnostics -------~
20
4.4 SECURITY AND SETTING PASSWORDS
6 m o press the key adjacent to the tick icon to save the security
code.
The settings lock function enables the instrument and measurement mode settings to be locked to
prevent any changes to the measurement parameters or instrument settings. The only exceptions to this
are that the user ID and contrast can be changed when the settings lock is active.
:,.,.:. 0 instantly. To unlock the settings press the key again. This
will open the security code menu as detailed in section
~~
~
<;>,.,
',~
4.4. 1. The previously set security code must be entered
to unlock the settings. When the settings lock is active
methods can still be opened, deleted and saved but the
method parameters cannot be changed.
m
been entered press the key adjacent to the tick icon. The
o 0 settings are now unlocked.
21
One press will lock the methods instantly. To unlock the methods press the key adjacent to the method
lock icon again. The methods are now unlocked. If the settings lock is active this must be disabled before
the method lock can be activated or deactivated.
In all the measurement modes if a user tries to save changes to a method when the method lock is active
the padlock icon flashes on the screen and changes cannot be saved.
-~
~ instrument settings menu by pressing the key adjacent
to the mode selection icon.
0902 x
4.6 DIAGNOSTICS
In addition to the time and date settings this instrument also has a user ID function. This function
enables an individual three digit ID number to be set. This will be displayed on all printouts and saved
results.
22
The user ID function can be accessed through the
instrument settings menu by pressing the key adjacent
to the user ID icon. Use the keys at the bottom of the
screen to select the digit to be changed. Use the keys
adjacent to the arrow icons to increase or decrease the
m a 1 number. Once the preferred user ID has been set press
the key adjacent to the tick icon to save and return to the
instrument settings menu.
This function is only available on the 7310 visible spectrophotometer which uses a tungsten halogen
lamp.
• • •
,.'j- •
m 3 0 II
icon to save and return to the instrument settings menu.
23
- The time set will begin to count down when there is
.;-.. ti~
~r 0.000~
·:·· no lamp activity. When the count down is complete the
--
·,
lamp and the fan will be turned off and the lamp save
.. ioo.o" 400rm
"11'r
J,LL
icon is shown in all the measurement modes. To bring
the instrument out of the lamp save in order to perform
,-.. ... a measurement pressthe key below the lamp save icon .
C'J·GZ -·.~
The lamp and fan will be turned back on and the lamp
will begin to warm up.
-
.;-..
The lamp cold icon is displayed adjacent to the calibrate
ti~
0.000~
·:··
.,
to zero icon in the measurement mode. The time needed
1-
-- a0 ·ioo.c-
for the lamp to warm up is five minutes .
"ITr
,. J,LL
400rm
~~-
0902
... - .. To view the remaining warm up time press the key below
c- ~~ the lamp cold icon. Calibration and measurements can
£@2011
..
~r·.. be performed whilst the lamp is warming up but these
--
. . ·,tr- results may not be accurate .
,· .J!L
0902 a0. ,;
~-
-
.;-..
Once the warm up time of five minutes is complete the
ti~
·:·· 0.000~ lamp cold icon disappears.
-
.,
~1·-
,.
ioo.o" 400rm
"11'r
J,LL
6902 00
24
SECTION 5 - Photometrics
The photometrics measurement mode enables simple measurements of absorbance and % transmittance
to be performed. The sample is measured at one wavelength and at one point in time. There are no post
measurement calculations available in this measurement mode.
0902
~r
·!"·
., toolbar on the left hand side of the screen enables access
-- ..
O.OOOABS to printing, print setup options, results, methods and
400nm autologging options. For more details on the different
ExpandedOperating Menu
100.0;{T
--L
PrinVprint settings e Toggle
Results selection menu .,
--
»-«:
Method selection menu ..... O.OOOABS -!- Increase wavelength
Autolog menu (
400rm ~ Decrease wavelength
09:02 D0 m
--
·.~ -!- and the display format.
..... 100.0?.T
~ 400rm ~
0s·02
[] 0 D
25
The toggle icon enables the large primary display to be set to show the absorbance or % transmittance.
To change the primary and secondary displays press the key adjacent to the toggle icon. Repeat presses
will cycle the displays between absorbance or% transmittance.
5.3 CALIBRATION
--
~(
..
O.OOOABS chamber and close the instrument lid. Press the key below
400nm the calibrate to zero absorbance icon. This sets the
0902 D0 D instrument to zero absorbance and 100% transmittance.
Once the calibration is complete the measure sample icon appears and the sample can be measured.
If the wavelength is adjusted before a sample is measured the measure sample icon will disappear and
the instrument must be calibrated again at the new wavelength.
-
;,,:...
It is not possible to measure a sample before the instrument
·.~
81.7r.r has been calibrated at the selected wavelength. Once the
-
calibration has been performed the measure sample
..... 0.088,:es icon is displayed and a sample can be measured. Remove
< 400nm the cuvette containing the blank solution and place a
0982
D0 0 cuvette containing the sample to be measured in the
sample holder. Close the instrument lid and press the key
below the measure sample icon.
Once the measurement is complete the photometric result will be shown on the screen. Subsequent
samplescan be measured in the same way. If the wavelength is adjusted between sample measurements
then the instrument must be calibrated again before more samples can be measured.
26
SECTION 6 - Concentration
The concentration measurement mode enables simple measurements of absorbance and concentration
to be performed. In this measurement mode it is possible to calibrate against a standard of a known
concentration or use a known factor. The sample is measured at one wavelength at one point in time.
There are no post measurement calculations available in this measurement mode.
o.ooo-
The concentration minimal operating menu enables
JW calibration to zero absorbance and simple readings to be
taken without changing any measurement parameters.
O.OOOABS Pressing the key adjacent to the JW icon opens the
expanded operating menu.
400nm
0902
--
~f
•, toolbar on the left hand side of the screen enables access
..
.. O.OOOABS to printing, print setup options, results, methods and
400nm autologging options. For more details on the different
functions of the utility toolbar refer to section 1 o. The
0902 00 ~.F settings icon enables the wavelength, units, resolution,
standard or factor to be set.
Expanded Operating Menu
PrinVprint settings
Results selection menu
-
-;,-.,
,,·,
•, 0.000~ x Settings
-
......,,..
Method selection menu ......
~
o.ooo= Increase wavelength
Decrease wavelength
Autolog menu 400rrn
09 02 00 E.F
--
wavelength in the expanded operating menu use the
09.02
27
The settings menu is accessed through the expanded
./ operating menu by pressing the key adjacent to the
settings icon. In the settings menu press the key below
A
2.008 .._ the wavelength icon.
1.000 xF
ppm 400nm
This will open a number entry screen. Use the keys at the
bottom of the screen to select the digit to be adjusted.
Use the keys adjacent to the arrow icons to increase or
,t- decrease the wavelength to the required number. Press
-!L the key adjacent to the tick icon to save the changes and
6.2.2 Settings
The settings menu enables the wavelength, units, resolution, standard or factor to be set and is
accessed from the expanded operating menu by pressing the key adjacent to the settings icon. Once
all of the required settings have been entered press the key adjacent to the tick icon to save and return
to the expanded operating menu.
./ Tick
Selecting resolution
A
2.008 .._ Standard menu
1.000 xF Factor menu
Selecting ~
concentration units - ~ppm o.ooe 400nm Selecting a
wavelength
When setting the method parameters either the standard or the factor should be selected. The standard
should be used if the factor is not known as selecting this option will calculate the factor. If the factor is
known it is not necessary to measure a known standard's concentration. When the standard or factor is
not selected the value should be set to 1.00.
The units of concentration can be selected from a number of options: no units, %, ppm, EBC, SRM,
mEq/1, mEq, M, mM, µM, nM, U, U/1, U/ml, g/1, mg/I, µg/1, ng/1, g/dl, mg/di, µg/dl, mg/ml, µg/ml, ng/ml,
µg/µI, ng/µI, mol/1, mmol/1.
In the settings menu press the key below the units icon .
---- ?. IBii'I EBC SRM mEq/1 ./ This opens the unit selection screen which displays all
mEq M mM µM nM u U/1
U/ml g/1 mg/I µgll ng/1 g/dl the different units. Use the keys adjacent to the arrow
mg/di
µglµI
µgldl
nglµI
mg/ml µglml ng/ml
mol/1 mmolll. -
.{f'.:. icons to navigate around the screen to select the required
units. Once the required units have been highlighted
·l·
-~
A.. ...}',.
-=!-'
press the key adjacent to the tick icon to save and return
to the settings menu. The selected unit will be displayed
in the minimal and expanded operating menu along with
absorbance and selected wavelength.
28
6.2.2.2 Changing the Resolution
The resolution that the concentration is displayed as can be selected from 1, 0.1, 0.01 or 0.001 by repeat
presses of the key below the resolution icon in the settings menu.
m oo 1.0 00
the digit must be pressed twice to select the adjacent
digit. For example 00 the first press of the key alters 10,
the second press alters 01.
Use the keys adjacent to the arrow icons to increase or decrease the selected number. Standard values
from 0.001 to 1000 can be entered. The standard value can be reset to zero by pressing the key adjacent
to the 000 icon. Once the standard value has been entered press the key adjacent to the tick icon to
save and return to the settings menu. The entered value is displayed in the settings menu adjacent to the
standard icon. A standard value should only be entered if the factor is not known. If the factor is known
the standard value should be set to 1.000.
o m oo 1.0 00
pressed twice to select the adjacent digit. For example
00 the first press of the key alters 10, the second press
alters 01.
Use the keys adjacent to the arrow icons to increase or decrease the selected number. Factor values of
0.001 to 10,000 can be entered. The factor value can be reset to zero by pressing the key adjacent to the
000 icon. Once the factor has been entered press the key adjacent to the tick icon to save and return to
the settings menu. The entered value is displayed in the settings menu adjacent to the factor icon. If the
factor is not known a standard should be measured in order to calculate the factor. If a standard is used
the factor value should be set to 1.000.
6.3 CAUBRATION
o.ooo- x
In the concentration measurement mode calibrations
against a standard or a factor can be performed following
·:·· .,
~1,- a zero calibration. If the factor is not known calibration
--
..
•. 0.000~ against a known standard is performed in order to
400rm calculate the factor. However if the factor is known there
is no need to calibrate using a standard. The calibration
0$02
must be performed at the same wavelength at which the
sample will be measured.
Insert a cuvette containing the blank solution into the sample chamber and close the instrument lid.
Press the key below the calibrate to zero absorbance icon. The instrument will calibrate to zero
absorbance. Insert a cuvette containing the standard concentration sample solution into the sample
chamber and close the instrument lid.
29
Press the key below the calibrate to zero absorbance
or standardicon. this will open another menu with the
--..
option to re-calibrate to zero absorbance or to calibrate
to the previously entered standard value. Press the key
adjacent to the calibrate to standard icon .
--
~( 0.121 ABS
standard icon.
,·
400rm
8902
(,)09 n~
6.3.2 Calibrating to a Factor
-
;,:.,
Insert a cuvette containing the blank solution into the
0.000~
sample chamber and close the instrument lid. Press the
-:·-
~r
key below the calibrate to zero absorbance icon. The
--
.,
It is not possible to perform sample measurements before the instrument has been calibrated at the
selected wavelength. In this operating mode the type of sample measurement performed depends on
the calibration which has been carried out.
--..
0.121 Fes
the key below the measure to standard icon. Once
the measurement is complete the concentration and
400rrn
r,!"
absorbance values are displayed.
8902 l} -
-
;,:.,
Remove the cuvette containing the blank solution and
~r..
·!'·
., in the sample chamber. Close the instrument lid and
--..
o.ooo" 400rrn
press the key below the measure to factor icon. Once
the measurement is complete the concentration and
absorbance values are displayed.
0902
(,)09 0.F In order to measure a sample based on a known factor
the value for the factor must be entered in the settings
menu before commencing measurement of the sample.
30
SECTION 7 - Spectrum
·:=l~
7.1 MODE SPECIFIC PARAMETERS
The spectrum minimal operating menu enables baseline
J~
scans and spectrum scans to be performed without
changing any measurement parameters or settings.
Pressing the key adjacent to the JW icon opens the
-r--1--,-,--.-.n-' ~ expanded operating menu.
8902
-
:~
¥]~
measurement parameters to be changed. The utility
: ....... toolbar on the left hand side of the screen enables access
:......:'
.... to printing, print setup options, results, methods and
: .""
- ... autologging options. For more details on the different
• 'I
' """
functions of the utility tool bar refer to section 10.
8902
The scan settings icon enables the graph y-axis, absorbance or % transmittance operating mode, start
and end wavelengths and scan interval to be set. The peaks and valleys threshold icon enables the
peaks and valleys threshold to be set. The peaks and valleys table icon enables the peaks and valleys of
the scan to be viewed in tabular form. The spectral points analysis icon enables points to be selected
from the scan for post measurement analysis.
31
7.2 METHOD SETUP
0902
This function enables the graph y-axis, absorbance or % transmittance operating mode, start and end
wavelengths and scan interval to be set. To access this function press the key adjacent to the scan
settings icon in the expanded operating menu.
Setting end
.__ wavelength
Setting start
------' Setting scan interval
wavelength
The operating mode is displayed on the left hand side of the menu. The default parameter is absorbance
and pressing the key adjacent to the ABS icon cycles the operating mode between absorbance and %
transmittance.
This function enables the start and end wavelengths of the spectrum scan to be set. Model 731 O has a
spectrum range from 320nm to 1000nm; model 7315 has a spectrum range from 198 to lOOOnm. To
set the start wavelength in the scan settings menu press the key adjacent to the wavelength on the far
left of the x-axis. This opens the number entry screen.
m 4 0 0 500nm
scan settings menu.
32
If the start wavelength entered is the same as the end wavelength the end wavelength will automatically
be set to be one times the scan interval. For example, if the start wavelength is entered as 500nm but
the end wavelength is already set to 500nm and the scan interval is 2nm, the end wavelength will be
automatically adjusted to 502nm. If the end wavelength entered is the same as the start wavelength
the start wavelength will automatically be set to be one times the scan interval. For example, if the end
wavelength is entered as SOOnm but the start wavelength is already set to soonm and the scan interval
is 2nm, the start wavelength will be automatically adjusted to 498nm.
400nm
I
500nm
Automatic Scaling
~ 500nm
Manual Scaling
Use the keys at the bottom of the screen to select the digit
II to be changed and use the keys adjacent to the arrow
icons to increase or decrease the number. To change the
~+~ sign press the key below the + or - icon. Repeat presses
;+:. will cycle between+ and-. Once the required absorbance
m.3 0 0
or % transmittance value has been entered press the key
adjacent to the tick icon.
33
To change the maximum absorbance or % transmittance
II value press the key adjacent to the maximum y-axis
value, this opens a number entry screen. Use the keys
~+:- adjacent to the arrow icons to increase or decrease the
~+:- number. To change the sign press the key below the + or
+ 11 . 3 0 0
- icon. Repeat presses will cycle between + and -. Once
the required absorbance or % transmittance value has
been entered press the key adjacent to the tick icon.
7.3 CALIBRATION
~ ~-j 1~.~~~'
a baseline scan which is performed across the selected
~.' ! 00
•,II
wavelength range at the selected scan interval. Insert a
cuvette containing the blank solution into the sample
chamber and close the instrument lid. Press the key below
I•' ~
0902 I::! the baseline scan icon to initiate the baseline scan. The
baseline icon will change to show baseline scan in
progress icon and a progress bar will be displayed.
,, .
measured.
.
If the wavelength range, or the scan interval, is changed
before a sample scan is performed a new baseline scan
must be performed across the new wavelength range, at
the new scan interval, before a sample can be measured.
34
To stop the scan before completion press the key below
the scan in progress icon. Confirmation will be needed
to stop the sample scan. Press the key adjacent to the
cross icon to continue with the scan of the sample
or press the key adjacent to the tick icon to confirm
stopping the scan.
Depending on how many data points have been measured either a partial scan and all the post
measurement tools will be displayed, or the instrument will return to the expanded operating menu
with no measurements saved.
-
: -""
• • I
~ ....
~
')902 ~ ·.· \·
-..
.. ...
: ..~
• • I
valleys table and spectral points analysis. The peaks
and valleys table displays all the detected peaks and
valleys above the selected threshold value. The spectral
0902
points analysis function enables points to be selected
from the scan to analyse absorbance or % transmittance
at selected wavelengths.
/
Wavelength range Threshold
Abs or % T range Percentage
35
7.5.2 Peaks and Valleys Table
': '.· ~
To display the valleys only press the key below the valley
450 •..• 1000 only icon. To redisplay the peaks and valley press the
same key again.
The spectral points analysis function enables points to be selected from the scan to analyse the
absorbance or% transmittance at a selected wavelength. To access this function press the key adjacent
to the spectral points analysis icon and this opens the selection screen.
<< < /\ > » icons increase or decrease the wavelength by ten times
the scan interval.
As the vertical line moves along the scan the wavelength and absorbance or % transmittance value is
displayed at the top of the screen. To add the selected point from the spectrum to the spectral points
analysis table press the key adjacent to the add points to spectral points analysis table icon.
<< < "A. > must be deleted from the table before another point can
be added.
36
The wavelength to be analysed can also be entered
IIJ ../ manually. This can be done in two ways; firstly by pressing
the key below the wavelength icon in the spectral
"It(' points analysis menu. This opens a number entry screen
JJ_L where the wavelength can be input manually. Use the
m 5 0 0
keys at the bottom of the screen to select the digit to be
changed and use the keys adjacent to the arrow icons to
increase or decrease the number. Once the wavelength
has been entered press the key adjacent to the tick icon
to save and return to the spectral points analysis menu.
II
key adjacent to the location in the table where the result
uffl] "It('
JJ_L
is to be stored. This opens a number entry screen where
the analysis point can also be deleted by pressing the key
m 4 4 2
adjacent to the delete icon.
Use the keys at the bottom of the screen to select the digit to be changed and use the keys adjacent to
the arrow icons to increase or decrease the number. Once the required wavelength has been entered or
the analysis point deleted press the key adjacent to the tick icon to save and return to the spectral points
analysis table. If a wavelength entered is outside the scan range the checknumber warning icon will
be displayed before the wavelength is automatically adjusted to within the range.
It is possible to access spectral points analysis before a scan is performed. Points to be analysed can be
pre-selected from the blank axis, or entered manually in the spectral points analysis table. When the
scan has been performed the points for analysis will already be in the table with the corresponding
absorbance or % transmittance values.
37
SECTION 8 - Quantitation
The quantitation measurement mode enables sample concentrations to be calculated using a standard
curve. In this mode a number of standard solutions covering a range of known concentrations are
measured at a set wavelength. The absorbance or% transmittance of these solutions is plotted to create
a standard curve. Once the standard curve has been created a sample of unknown concentration can be
measured and the concentration calculated using the standard curve.
0'302
-. o.ooo-
11=-
..
-
to printing, print setup options, results, methods and
400rm ~ autologging options. For more details on the different
functions of the utility tool bar refer to section 10.
0902
The standard curve icon enables the current standard curve and curve statistics to be viewed or a new
standard curve to be created. The quantitation table icon enables the quantitation standards to be
viewed.
Print/print settings
-
-;,,:., I
Standard curve
0.986~
l:_
ltl
·!"·
Results selection menu -~ Quantitation table
--
·.~
Method selection menu ....
,. O.OOOMS ·1-
.:j-.
Increase wavelenth
Calibrate Measure
to zero sample
38
8.2 METHOD SETUP
-o.ooo-
The wavelength can be changed in the expanded
L operating menu by using the keys adjacent to the
~,~
lUt arrow icons. Use the keys to increase or decrease the
--
O.OOOABS wavelength until the required wavelength has been
*
400rm selected. The wavelength selected should be the same
wavelength that the standards and unknown samples are
0302
to be measured at. The wavelength can also be adjusted
in the standard curve screen. Refer to section 8.2.3 for
more details.
The quantitation table enables the quantitation standards to be viewed. This function also enables the
number of standards, units, resolution and operating mode to be set. To access the quantitation table
screen press the key adjacent to the quantitation table icon in the expanded operating menu.
ppm
~
0.01): 4', ,l'(:, ABS
available the concentration measurement mode should
be used.
The units of concentration can be selected from a number of options: no units, %, ppm, EBC, SRM,
mEq/1, mEq, M, mM, µM, nM, u, U/1, U/ml, g/1, mg/I, µg/1, ng/1, g/dl, mg/di, µg/dl, mg/ml, µg/ml, ng/ml,
µg/µI, ng/µI, mol/1, mmol/1.
-~
l'L ..A
-=!-'
quantitation table screen.
39
8.2.2.3 Changing the Resolution
The resolution of the concentration can be selected from 1, 0.1, 0.01 or 0.001 by repeat presses of the
key below the resolution icon.
The operating mode can be changed between absorbance or % transmittance by pressing the key below
the Abs or % T icon. Repeat presses will cycle between absorbance or % transmittance.
The positive sign can be changed to a negative sign by pressing the key below the + icon. Once the
required number has been entered press the key adjacent to the tick icon to save. When the required
resolution, units, number of standards, operating mode and standard concentrations have been set
press the key adjacent to the tick icon to save and return to the expanded operating menu.
-o.ooo-
This function enables the current standard curve to be
L viewed, a new standard curve to be created, the curve
lt1t
~,~ fit to be adjusted and the curve statistics to be displayed.
--
0.000~ To access this function press the key adjacent to the
40Qrm ~ standard curve icon in the expanded operating menu.
09:02
40
When the current standard curve is displayed the
./ concentration and photometric value of the last sample
The wavelength which the standards were measured at is displayed on the right hand side of the screen.
To adjust the wavelength press the key adjacent to the wavelength icon.
This opens the number entry screen. Use the keys at the
bottom of the screen to select the digit to be changed
and use the keys adjacent to the arrow icons to increase
--:tr or decrease the number. Press the key adjacent to the
tick icon to save and return to the standard curve screen.
The wavelength selected should be the same wavelength
4 0 0 that the new standards and unknown samples are to be
measured at.
. ... displayed will be the gradient of the line (rn), constant (c)
and correlation coefficient (r2). To exit this screen press
(• ·· .
the key adjacent to the tick icon. The quantitation table
can also be viewed from this screen by pressing the key
adjacent to the quantltatlon table icon.
x sample chamber and close the instrument lid. Press the key
adjacent to the tick icon to perform an initial calibration
to zero absorbance.
41
Once this measurement has been performed the standard concentration samples can be measured.
Remove the cuvette containing the blank solution from the sample chamber and insert the cuvette
containing the first standardised solution to be measured.
1:0J0ppm
x pressing the key adjacent to the back icon.
To measure the next standard remove the current standard from the sample chamber and insert the
cuvette containing the second standardised solution. Press the key adjacent to the tick icon to take a
reading of the standard. Repeat this procedure until all the standards (up to a maximum of six) have been
measured. Once all the standards have been measured the instrument returns to the standard curve
screen and the new curve can be viewed. The photometric values for all of the newly added standards
can be viewed in the quantitation table.
To abort creation of the new standard curve before all the standards have been measured press the key
adjacent to the cross icon at any time and the instrument will return to the previous standard curve.
Once the curve has been created press the key adjacent to the tick icon and the screen will return to
the expanded operating menu where unknown samples can be measured. When this operating mode is
exited the current standard curve will be saved until it is replaced by a new curve.
The automatic 8 cell turret can be used to assist creating a new curve. Refer to section 11.3.1 .1 for more
details.
8.3 CALIBRATION
0902
00 calibrate to zero absorbance icon. The instrument will
calibrate to zero absorbance.
Once the calibration is complete the measure sample icon appears and the sample can be measured.
If the wavelength is adjusted before a sample is measured the measure sample icon will disappear and
the instrument must be calibrated again at the new wavelength.
--
11=-
..
..
O.OOOAES -!-
.......
carrying out a measurement of an unknown sample.
Place the cuvette containing the sample in the sample
.
40Qnrn
0902 B..:,_
chamber and close the instrument lid.
Press the key below the measure sample icon and the instrument will take a reading and the
concentration and photometric values will be displayed on the screen. To view the position of this sample
on the standard curve open the standard curve screen.
42
8.5 DATA ANALYSIS
43
SECTION 9 - Kinetics
The kinetics measurement mode enables the absorbance or % transmittance of an active molecule to be
measured over a period of time; for example enzyme analysis of horseradish peroxidase. The absorbance
or % transmittance is measured at regular time intervals at a set wavelength over a period of time.
The results are plotted on a graph to show the change in absorbance or % transmittance over time.
Following sample measurement statistical analysis of all or part of the experiment can be performed.
I,. I
.....
I
parameters or settings. Pressing the key adjacent to the
JW icon opens the expanded operating menu.
0902 00
MinimalOperating Menu
The settingsicon enables the wavelength, units, resolution, graph y-axis, absorbance or% transmittance
operating mode, standard, factor, measurement time, lag time and start on level to be set. Following
the completion of a set of kinetics measurements the statisticsfor all or part of the kinetics run can
be analysed. The kinetics moving line function enables the initial and final points of the curve to be
selected, so that analysis of the area between these two points can be performed.
44
9.2 METHOD SET UP
0902
The settings menu enables the wavelength, units, resolution, graph y-axis. operating mode, standard,
factor, measurement time, lag time and start on level to be set. Once all of the required settings have
been entered press the key adjacent to the tick icon to save and return to the expanded operating menu.
Maximum ABS/%T
../-
2.~00
Tick icon
:f·.
Manual/automatic scaling ... 400nm
n
Select wavelength
Abs/% transmittance Hbs
1eoe ,: •
Standard menu
Minimum Abs/%T ·O ~00
tee xE Factor menu
I I 0e1h ( . .,
~
PPM 9020s
This function enables the scale of the y-axis of the kinetics graph to be adjusted either manually or
automatically. The automatic scaling is represented by the hand icon with a cross through it and the
manual scaling by the hand icon without the cross. To select manual or automatic scaling press the key
adjacent to the hand icon. Repeat pressing of the key cycles between manual and automatic scaling.
In automatic y-axis scaling the absorbance or %
- transmittance values are cleared from the settings screen.
When the kinetics measurement is complete the y-axis is
..n automatically re-scaled.
1&00 -:_"
I 1eF
I I auae~
~
(.:, ppm
1u~
Automatic Scaling
n
change the absorbance or % transmittance values press
L000 ·~·
the key adjacent to the minimum y-axisvalue, this opens
a number entry screen.
·8.560
I 100 xF
I I ee10s (
~
. ., pj)m 0020s
Manual Scaling
45
• ~+~
~,.:.
Use the keys at the bottom of the screen to select the digit
to be changed and use the keys adjacent to the arrow
icons to increase or decrease the number. To change the
sign press the key below the + or - icon. Repeat presses
will cycle between+ and-. Once the required absorbance
In this measurement mode starting the kinetics measurements can be delayed by setting a lag time or
a start on level. The lag time is the amount of time that the instrument will wait before starting the
kinetics measurements. The start on level is the absorbance level that must be reached before starting
the kinetics measurements. Both options can be set using the set lag time/start on menu. This function
is accessed by pressing the key below the start time pause icon on the far left side of the x-axis. This
opens the lag time/start on level menu.
To set the lag time press the key adjacent to the clock
../ icon. Use the keys at the bottom of the screen to select
the digit to be changed and the keys adjacent to the
··.f:· arrow icons to increase or decrease the number. The lag
000 .:j,. time can be reset to zero by pressing the key adjacent to
m1 0
the ooo icon.
+ PJ. 0 0 0
pressing the key below the symbol + icon.
On the left hand side of the screen the symbol can be changed to read greater than(>) or less than
(<) the absorbance level that has been entered. The start on level can also be turned off (<f>). Once
the required absorbance or % transmittance level (depending on the operating mode selected) has been
entered press the key adjacent to the tick icon to save and return to the settings menu.
The operating mode is displayed on the left hand side of the screen. The default parameter is absorbance
and pressing the key adjacent to the ABS icon cycles the operating mode between the absorbance or
% transmittance.
46
9.2.1.4 Changing the Resolution
The resolution of the concentration can be selected from 1, 0.1, 0.01 or 0.001 by repeat presses of the
key below the resolution icon.
The concentration units can be selected from a number of options: no units, %, ppm, EBC, SRM, mEq/1,
mEq, M. mM. µM, nM. u, U/1, U/ml. g/1, mg/I, µg/1, ng/1, g/dl, mg/di, µg/dl, mg/ml, µg/ml, ng/ml, µg/µI,
ng/µI, mol/l, mmol/l.
In the settings menu press the key below the units icon
---- x Im EBC SRM mEq/1 ._/
to open the unit selection screen which displays all the
mEq M mM µM nM U U/1
different unit options. Use the keys adjacent to the arrow
:.t'..
U/ml g/1 mg/I 11g/l ng/1 g/dl
mg/di 11g/dl mg/ml 11g/ml ng/ml icons to navigate around the screen to select the required
11glµI ng/111 mol/1 mmol/1. units. Once the required units have been highlighted
press the key adjacent to the tick icon to save and return
to the settings menu.
A standard value should only be entered if the factor is not known. If the factor is known the standard
value should be set to 1.000.
mo 00 1.0 00
pressed twice to select the adjacent digit. For example
00, the first press of the key alters 10, the second press
alters O 1 . Use the keys adjacent to the arrow icons to
increase or decrease the selected number. Factor values of 0.001 to 10,000 can be entered. The factor
value can be reset to zero by pressing the key adjacent to the 000 icon. Once the factor has been entered
press the key adjacent to the tick icon to save and return to the settings menu screen. The entered value
is now displayed in the settings menu adjacent to the factor icon.
If the factor is not known a standard should be measured in order to calculate the factor. If a standard is
used the factor value should be set to 1.000.
47
9.2.1.8 Selecting a Wavelength
9.3 CALIBRATION
·=i
0 wavelength at which the sample will be measured. Insert
,n ·, a cuvette containing the blank solution into the sample
-\ . +-~,-• .-,........-.-,-..-..-.-,-,,
...
~ ... ... llflfWI
chamber and close the instrument lid. Press the key below
the calibrate to zero absorbance icon. The instrument
will calibrate to 0.00 absorbance.
0902 D0
Once the calibration is complete the start kinetics measurement icon will be displayed and a sample
can be measured. If the wavelength is adjusted before the sample is measured the start kinetics
measurements icon will disappear and the instrument must be calibrated again at the new wavelength.
i'.
cz: start kineticsmeasurements icon. If a start on level or
lag time has not been set the instrument begins taking
7'
..... ..
__ ~. -.~~~. ~. ~. -.~.
·- a photometric reading every second. As each reading is
taken it is plotted on the graph.
0902 D0 I"!,.
LJ ·-·
- 0902 00
48
~l-
f',
~( []I
.c
/ \.•
If a start on level has been set the Abs/%T icon is
displayed in the centre of the screen. The instrument
will take photometric readings until the start on level is
-.... a0
reached. Once the absorbance/% transmittance level has
.... been reached the readings will be plotted on the graph .
I
..:.. I I I
....! ~
0902 ~
Once the measurement has commenced it can be aborted by pressing the key below the kinetics
measurement in progress icon.
Following the completion of the kinetics measurements the post measurement icons are displayed on
the screen. These include the statistics for the kinetics experiment and the kinetics moving line. The
post measurement statistics include rate of change, start and end absorbance levels and concentration
lc::::
of the sample. The kinetics moving line function enables the initial and final points of the experiment to
be set and the statistics for this part of the experiment to be analysed.
To access the statistics for the whole of the kinetics
,....J.. experiment press the key adjacent to the I icon which
~ opens the statistics menu over the top of the expanded
-
J
]J .... t •....
operating menu.
.....'· . .. ..
0902 00 r.!,,
L:·
l. I
The information displayed includes concentration,
01)00 ppm .,/ absorbance per minute (rate of change), correlation
xF
Hf1 rn 11 i.-u1(1
coefficient (r2), start time absorbance and end time
:-.
. :,1
r , 11,(111
To exit the statistics menu press the key adjacent to the tick icon.
The kinetics moving line screen can also be accessedfrom this menu by pressing the key adjacent to
the kinetics moving line icon. The kinetics moving line function enables the initial and final points of
the kinetics experiment to be set, allowing the statistics for this part of the experiment to be analysed
specifically. This function can be accessed from the statistics menu or from the expanded operating
menu by pressing the key adjacent to the kinetics moving line icon.
49
In the kinetics moving line screen the initial and final
./ time lines appear as two vertical lines. The line which is
1 -·~
,-; .. I
selected is represented by a dashed line -- and can be
moved using the keys below the greater than(>) or less
L than (<) icons at the bottom of the screen. Move the
<< < > >> dashed line to the point on the x-axis which is to be the
initial time.
]ttI.1f1
To select the final time press the key adjacent to the
toggle icon. This will change the initial line from a
dashed line to a solid line. The final time line will change
from a solid line to a dashed line. Use the keys below the
< > ~
to the final time on the x-axis by reducing or increasing in
<< 1 second time intervals.
,. , -. xF
H(~ in1u~,
the statics menu for that portion of the experiment. The
H_ 11}H information displayed includes concentration. absorbance
....
111_ 1111-
per minute (rate of change). correlation coefficient (r2) •
50
SECTION 10 - Saving, Printing and Autologging
The utility toolbar in the expanded operating menu provides access to printing, print setup options,
opening, saving and deleting results and methods and autologging options.
If the method lock has been activated the method selection menu is disabled.
o.ooo-
PrinVprint settings ----+-.;-.,
-
Resultsselection menu ·: ·.•,
Method selection menu - ,.:_-~
o.ooo= -!-
Autolog menu -----11-~
- 09'02
400rro ~
Up to 48 methods can be saved in each of the measurement modes to the internal memory of the
instrument. Once the method has been saved to the internal memory the method can be saved directly
to a USB memory stick and opened on another instrument.
To select letters use the keys adjacent to the arrow icons to move around the menu. Once the required
letter is highlighted press the key below the pencil icon to select the letter. Up to eight characters can
be selected. To change the letter case or to use numbers in the method name press the key below the
AB icon. Repeat pressing of this key will cycle between upper case, lower case and numbers. Once the
required name for the method has been entered press the key adjacent to the tick icon to save the
method and return to the method selection menu.
x
method with the new one; press the key adjacent to the
tick icon to confirm replacement. Press the key adjacent
to the cross icon to cancel and return to the method
01: EXP_1?
naming menu. To exit the method naming menu without
saving the method press the back key.
51
10.1.2 SAVING METHODS TO USB MEMORY STICK
Irr ' ·~-.~- 1 be saved to USB memory stick. To save a single method
press the key adjacent to the single method save icon.
.:'
x
To save all 48 methods press the key adjacent to the 1
[T. ·-
1·48 ,. · ( 0
&/
the method selection screen.
[T.,·-,•
I ••·<·
• I
' . ·,x
ill' .. ,..- confirm saving the method(s). Press the key adjacent
to the cross icon to cancel and return to the option of
saving one or all of the methods to USB memory stick.
Methods can be opened from the internal memory of the instrument or from a USB memory stick.
Before a method can be opened from a USB memory stick it must be saved to the internal memory of
the instrument first.
--
.. 100.Qr.r
.. 400r,,i
0902
00
If a method hasn't been selected it is still possible to calibrate the instrument and take measurements.
The instrument will operate in a generic method, 0, where all the current operating settings will be
stored. There are separate method O's for each measurement mode.
52
10.2.2 OPENING METHODS FROM USB MEMORY STICK
x
,. icon. To open all methods press the key adjacent to the
• ..,. t·4 s- ~l~.
-
~- ~ 1 - 48 method open icon. Press the key adjacent to the
tick icon to confirm selection or press the key adjacent
to the cross icon to cancel and return to the method
selection screen.
& x
to cancel deletion and return to the method selection
menu.
Resultscan only be saved to a USB memory stick if there is a valid USB memory stick inserted into the
front of the instrument. The instruments are supplied with a USB memory stick but other types which
may be used include: the PNY Attache Premium, the Emtec C250, The Kingston Data Traveler G2, the
Sony Pocket Bit and the Integral Flexi.
53
·:·· ., l.OOOWTI L In the expanded operating menu press the key adjacent
to the USB memory stick icon to open the results
~r llli.
-- ,
selection menu.
,·
0.099ABS -!-
.: ..
40Qr,m
.'·..__
.
0902 0
If the USB memory stick doesn't have any results stored
on it only the save to USB memory stick icon will be
displayed. To save a result press the key below the save
to USB memory stick icon. This will open the results
naming menu.
To select letters use the keys adjacent to the arrow icons to move around the menu. Once the required
letter is highlighted press the key below the pencil icon to select the letter. Up to eight characters can
be selected. To use numbers in the results name press the key below the AB icon. Repeat pressing of this
key will cycle between upper case and numbers. Once the required name for the result has been entered
press the key adjacent to the tick icon to save and return to the results selection menu.
ISAMPLE_1
!O:S/07/2010 11 II
I [I] ~
The saved results are listed alphabetically with the date
and time that the result was generated.
Results can only be opened if a valid USB memory stick is inserted into the front of the instrument.
!BEER
lo:,,,0u2010 11.1s
l[IJ /
1
In the expanded operating menu press the key adjacent to
the USB memory stick icon to open the results selection
I
SAMPLL1
30/07/2810 10.021
I menu. Select the result to be opened by pressing the
key adjacent to the result. Press the key below the load
from USB memory stick icon.
54
The result will be displayed on the screen.
Operatmg mode Photometdcs
Operator Id 2
Hethod Id 0
Wavelengt:h 420 The information displayed is specific to the measurement
Sam~ tnte-rval I
Sampfe Count:
Results Oest1nat1on Internal
0 mode which the result was generated and saved in.
Cal T1me/O.ate 10 · 54 · 33 05/0712010
ID T1meOateCeU Use the keys below the arrow icons to display more
001 01 4 0. 0s110 · 54 40 0510712010
information.
-
,tr ~ ~
Results can only be deleted if a valid USB memory stick is inserted into the front of the instrument.
IBEER
I05/0 12010 II 18
7
l[IJ J
1
In the expanded operating menu press the key adjacent to
the USB memory stick icon to open the results selection
SAMPLE_l menu. Select the result to be deleted by pressing the key
adjacent to the result. Press the key below the delete
icon.
&
95/87/~10 II II
~'~
x
return to the results selection menu.
liililf
10.7 PRINTING
-
o.ooc-
The utility toolbar in the expanded operating menu
~
·:·- . enables results to be printed and print setup options to
--..
.. O.OOOABS the printouts and language of the printouts to be set.
400nm Depending on the measurement mode the printouts can
0902
00 be customized to print off post analysis statistics.
55
10.7.1 Print Setup
o.ooo- x To open the print setup menu hold the key adjacent to
the printer icon for 2 seconds in the expanded operating
menu.
--.. O.OOOIR,
400rYn
091:!2
The destination of the printouts can be the internal printer or an external serial printer. The results can
only be sent to an external serial printer if there is a serial printer connected to the instrument via the
RS232 serial port. Press the key adjacent to the computer icon to select the external serial printer. The
results can only be sent to the internal printer if there is an internal printer connected. To select the
internal printer for the printout destination press the key adjacent to the printer icon.
Once the required printout destination and language has been selected press the key adjacent to the
tick icon to save and return to the expanded operating menu.
English
56
Repeat presses will cycle between 0, 1, 2, 5, 10 or 50. Once the required options have been selected
press the key adjacent to the tick icon to save and exit print-set up.
-
·.--r
reported when the printer icon is pressed. Depending on
. ... 0.088ABS "It('
JJ_L the printout destination selected the result will be sent to
< 400nm
0902 a0 ~
the internal printer or to the external serial printer. If the
printer icon is pressed when there is not a result on the
screen the no result to printer or no result to RS232
icon (depending on results destination) will flash up on
the screen.
Operatmg mode Photometr,cs Saved results can be printed by pressing the key below
Operator Id 2
Nethod Id
Wavelength 420
0 the printer icon. The printout will include the date
Sampleinterval I
Sample Count 0 and time the result was generated, the user ID and the
Re.suits De-stmat,on lnter-nal
Cal T ,me/Date
ID
10: 54 · 33 0510712010
T1meDateCe:II
measurement parameters.
001 01 4 0. 09110 54 · 40 0510712010
--L J•L
~
57
10.8 AUTOLOGGING
·:··..,
o.ooo- the same sample to be performed with a set time period
between each measurement. This produces a batch
--
o.ooo=
»-«:
-!- of results for the same sample. The autolog function
,:
400rm ~ also enables the results to be autologged to different
destinations. The autolog menu is accessed from the
0902
[] 0
utility toolbar in the expanded operating menu by
pressing the key adjacent to the autolog icon.
Expanded Operating Menu
lil0000 00!s Z
The default batch name will be displayed. To save the
E II Jwoooo-o II ../ batch name under the default name press the key
l!IBr:DEFGHI adjacent to the tick icon. To change the name press
.J~~LMt·iOPQF: the key below the eraser icon. One press of the key
s Tu I.) IJJ>=: ..... z will delete one letter, holding the key for 2 seconds will
delete the name completely. To select letters use the keys
adjacent to the arrow icons to move around the menu.
Once the required letter is highlighted press the key below the pencil icon to select the letter. Up to
eight characters can be selected. To use numbers in the batch name press the key below the AB icon.
Repeat pressing of this key will cycle between upper case and numbers. Once the required name for the
batch has been entered press the key adjacent to the tick icon to save and return to the autolog menu.
58
The number of repetitions and time interval will be
--
autologging press the key below the measure sample
Autologging can be stopped before all the measurements have been completed by pressing the key
adjacent to the autolog icon. Confirmation will be needed to stop autologging. Press the key adjacent
to the tick icon to confirm stopping autologging or press the key adjacent to the cross icon to continue
autologging. If autologging is stopped before completion the measurements already recorded will be
saved under the previously entered batch name.
~1
again the batch append/delete screen is displayed.
10.9 CONNECTING TO A PC
Connect the interface cable to the RS232 serial port on the rear of the instrument and connect to the
serial port on the rear of the PC. Turn the PC on and load the PC software by inserting the PC software
disc into the CD drive. If the PC software does not auto run open My Computer and double click on
the Jenway 73 series software icon. Follow the instructions to install the PC software to the required
location. Refer to the PC software manual for further instructions. Once the software is installed, turn
the mains power on to the instrument.
9600 baud
8 data bits
No parity
1 stop bit
59
SECTION 11 - Accessories and Spare Parts
There are two types of accessories which can be fitted in the sample chamber - passive or active
accessories. The range of passive accessories includes 1 O x 1 Omm single cuvette holders, single water
heated cuvette holders, adjustable path length (10 to 100 mm) cuvette holders, test tube holders,
boiling tube holders, film holders and micro-cuvette holders. The range of active accessories includes an
automated 8 cell changer, sipper pump, pettier and combined peltier sipper pump. The instrument must
be turned off before any accessories are fitted.
60
Squeeze the grey plastic clips together so that the printer
top opens. Slot the printer into the top of the instrument
and push down until it fits flush to all four sides.
Insert the paper roll into the printer - ensuring that there
is some paper sticking out of the printer before clicking
the grey plastic back into place. Switch the instrument
on. The power and error lights on the printer will flash.
Once the instrument power on tests are complete press
the feed button to check that the paper is fed correctly.
'
/ Unscrew the thumb screw to undo the passiveaccessory.
Lift out the passive accessory. To fit an active accessory
unscrew screws 1 to 4 and lift out the metal base plate.
~
1. ·-~·
. . 2.
Thumb
screw -
f)_
3. 4.
-,
This will expose the bottom of the sample chamber with
the power supply connection needed to operate the
active accessories.
61
11.2.3.1 Automatic 8 cell turret
11.2.3.2 Peltier
6.
62
11.2.3.3 Sipper pump
Bi directional flow A
(sipping)
Continuous flow
For sipping:
63
7. Cut a small length of the sipper pump tube and push
this over one end of the capillary tube. Connect this to
the inlet port of the flow-through cuvette.
For pumping:
3. Fit the other end onto the inlet port on the flow-
through cuvette.
4. Fit the second 300mm length of tubing to the outlet
port of the flow-through cuvette. Once secured, ensure
the tubing is routed into the two retaining clips located
on the base plate at the side of the pump head.
5. Fit the other end of the tubing onto the outlet port,
located on the inside of the front bulkhead.
11.2.3.4 Combined sipper peltier pump Refer to section 11.2.3.3 for more details.
64
11.3 USING THE ACCESSORIES
-.. O.OOOAB3
SOOnrr,
the 8 cell turret icon. The O position should always be used
for the zero calibration sample .
0902
To perform measurements using the automatic 8 cell turret, insert the cuvettes containing the samples
into turret positions 1 to 7. Insert the cuvette containing the blank solution into turret position 0. Enter
the required measurement mode and set up the required measurement parameters. Press the key below
the calibrate to zero icon. The instrument will automatically move the turret around to position zero to
perform the measurement. Once the calibration is complete the measure sample icon will appear and
the turret will return to its original starting position.
To measure the next sample select the next turret position and press the key below the measure sample
icon. Repeat this process until all the samples have been measured. To adjust the wavelength press the
key below the 8 cell turret icon and use the arrow icons to adjust the wavelength.
The 8 cell turret can be used to support creation of a new standard curve in the quantitation measurement
mode. Refer to section 8.2.3.1 for more details.
Use the keys adjacent to the arrow icons to increase the turret position, until the required standard
position has been selected. Press the key adjacent to the tick icon to measure the standard. The standard
concentration and photometric value will then be displayed. The standard can be re-measured by pressing
the key adjacent to the back icon.
To measure the next standard select the next turret position and press the key adjacent to tick icon.
Repeat this process until all the standards have been measured.
65
11.3.2 Peltier
--
"I·...')
·-
-
ioo.o-' ·11'(
J.LL
adjacent to the peltier icon. Below the peltier icon there
is an arrow icon to indicate if the current temperature is
40Qrm
below or above the set temperature. To adjust the set
09•02 00 19.9
-+12G..5 .. ..-
point temperature hold the key below the peltier icon
for 2 seconds.
This opens the peltier settings screen. Use the keys at the
II ../ bottom of the screen to select the digit to be changed
and use the keys adjacent to the arrow icons to increase
oc --:tr
J.:. .
or decrease the number. The temperature can be set
in °c or °F by pressing the key adjacent to the 0c icon.
Repeat presses will cycle between "C and 0F.
0 0
Once the required temperature has been selected press
../ the key adjacent to the tick icon to save and return to the
expanded operating menu. The peltier will begin to heat
--:tr or cool depending on the current temperature.
OF J.:. .
[!] 6 8 0
;-,
- When the sipper is in use the sipper pump icon is
O.OOOABS ~IJ displayed in the bottom right hand corner of the screen.
--
..
..
100.0%T "itr
JJ,L
the sipper pump icon. The sipper pump can operate
400nm in manual or timed mode, depending on the option
selected in sipper pump settings. To open the sipper
0902 00 -~
:
pump settings hold the key below the sipper pump icon
for 2 seconds.
a
(~.
-:i- tick icon to save and return to the expanded operating
$ menu.
"Ir
-
icon.
..
100.0%T 1'tr
JJ,L
40Qnm
es ez 00 -~
66
- Confirmation will be needed to start the sipper pump.
0.
~
Press the key adjacent to the tick icon to confirm and
--
~f
..
start the sipper pump. Press the key adjacent to the cross
icon to cancel and return to the expanded operating
..
. ..,.
menu.
09·02 a0 .1
:.
~r.. 0.
stop icon. Ensure that the flow through cuvette contains
--
enough sample before pressing the key below the
measure sampleicon.
a0 . ..,.
.1
0902 :.
,,&
Once the cuvette is full press the key adjacent to the stop
icon to stop the sipper pump. The time taken for sample
uptake is recorded.
. ,i
67
To fine tune the amount of sample uptake press the key
below the plus or minus icon to increase or decreasethe
amount of sample taken up. The recorded time will be
adjusted accordingly. Once the fine tuning is complete, or
if none is required, pressthe key adjacent to the tick icon
>
sequence. If an air gap is not required press the key below
,,8
:»
the 000 icon to set the air gap to zero. If a previously
programmed air gap is to be used press the key adjacent
to the double greaterthan icon to skip this stage and
____________ '110_0_..
__ retain the current air gap time.
To program an air gap remove the inlet tubing from the sample container and press the key adjacent to
the single greater than icon. The sipper pump will start and air will be pumped through the tubing to
the flow through cell.
If the original pump direction selected was forwards, selecting the forwards direction at this stage will
send the sample to waste and selecting the backwards direction will send the sample back to the sample
container. Once the required direction has been selected press the key adjacent to the tick icon to save
the calibration sequence and return to the expanded operating menu. To exit the sipper calibration
sequence without saving any changes press the back key at any point during the calibration sequence.
68
- To perform a measurement place the sipper tubing into
;-..
O.OOOAEG (":~ the sample and press the key below the sipperpump
~r
--
icon.
1Tr
.. 100.Q%T JJ,L
400r.m
ti.· ..
1!902 00 · ~:~
-
;-..
Confirmation will be needed to start the sipper pump.
--
return to the expanded operating menu. Press the key
i- adjacent to the tick icon to confirm and start the sipper
.. pump. The pump will run for the previously recorded
-
;-..
Confirmation will be needed to start the sipper pump.
--
~r..
return to the expanded operating menu. Press the key
adjacent to the tick icon to confirm and start the sipper
pump. The pump will run for the previously recorded air
• !· gap take up time .
1!902 00 <'
-
;-..
Once this stage of the calibration sequence is complete
-
..
..
00
...:,
sipper pump. Press the key adjacent to the cross icon to
cancel and return to the expanded operating menu. Press
the key adjacent to the tick icon to confirm and start the
0902 sipper pump. Depending on the disposal route previously
selected the sample will either go to drain or back to the
sample container.
-.. 100.Q%T
400nm
200: ~· I
the set point temperature adjacent to the sipper peltier
icon. Adjacent to the peltier icon is an arrow to indicate
if the current temperature is below or above the set
0902 ·28.0 C ••
temperature.
69
The pump direction is displayed by an arrow icon below the sipper pettier icon. The combined sipper
peltier pump combines the functionality of the peltier and sipper pump. To open the sipper peltier
settings hold the key below the sipper pettier icon for 2 seconds.
11.4 SPARES
70
SECTION 12 - Maintenance and Service
Ensure the external surfaces of the unit are clean and free from dust. The sample area should always
be kept clean and any accidental spillage should be wiped away immediately. To give added protection
when not in use. the unit should be disconnected from the mains supply and covered with the optional
dust cover.
The only routine maintenance which maybe required is the replacement of the light source. The
replacement lamps are available from your local distributor (refer to section 11.4 for spare part codes).
Only genuine replacement lamps should be used. Similar lamps may have different filament configurations
or be wavelength restricted for domestic or commercial use and will give errors if used.
2. Withdraw the lamp access panel and unscrew the lamp bracket fixing screw.
3. Grasp and rotate the lamp bracket to gain access to the lamp.
4. Remove the old lamp from the holder. The lamp is a plug-in fit and should be removed by gently
easing it from the holder.
5. Carefully remove the replacement lamp from the packaging. Ensure that the glass portion of the lamp
is not touched as finger marks will damage the lamp resulting in a reduced performance. If accidental
damage occurs the surface of the lamp may be cleaned using propan-2-ol.
6. Insert the lamp into the holder, ensuring it is fully pushed home.
7. Rotate the lamp bracket and put back into operational position. Replace the lamp bracket fixing screw
and tighten.
8. Replace the lamp access panel and fix in place with the two screws.
9. Reconnect the power supply, turn on the unit and ensure that the lamp is illuminated after a few
seconds.
For further instructions refer to the service manual.
This option is only valid for the 7315 spectrophotometers and must be done by an accredited service
engineer. Refer to section 12.3 for more details.
71
12.3 SERVICE
Our dedicated service staff are on hand to help in the unlikely event that your Jenway equipment
develops a fault. Please contact them by one of the following means with a clear description of the
problem:
E-mail: [email protected]
Tel: +44 {O) 1785 810475
Fax: +44 {O) 1785 810471
On occasion it may be necessary for your equipment to be sent back to our Service Department for
repair. In this case please contact the Service Department for a reference number which you should
include with your faulty equipment. Please also ensure you include a clear description of the fault and a
completed copy of our Decontamination Certificate. This is available as a downloadable pdf file at www.
jenway.com, or contact us and we will be happy to fax you a copy. Please clearly mark the package for
the attention of the Service Department and post to the following address:
All replacement parts are guaranteed for 1 year and where ever possible, returned equipment is turned
around in 5 working days.
72
SECTION 13 - Troubleshooting
If an error code is displayed it will be accompanied by a spanner icon and a symbol to indicate if the
error is a warning (caution icon) or fatal (stop icon). If the error is fatal contact your local distributor
or Jenway service department (refer to section 12.3). If the error is a warning it may be possible to
retry the test. In this case a back icon will also be displayed. The table below shows the error codes:
111!1
Err 1 System Parameters Failure
This error indicates that the essential system parameters are corrupt.
The most likely causes of this error are:
1. Fram chip failure.
Solution: Restart the unit, if the problem persist, contact a service
technician.
73
Err 5 Light Saturation Not Found
This error indicates that the peak light hasn't been found at zero.
The most likely causes of this error are:
Fatal 1. Lamp failure.
2. Deterioating lamp signal.
3. Sample or cuvette in the sample holder.
Solution: Ensurethat the sample holder is empty. Restart the unit, if the
problem persists contact a service technican.
Err 9
74
13.2 TROUBLESHOOTING GUIDE
Issue Solution
Unable to achieve zero Ensure that there is not a sample in the sample chamber.
absorbance or 100% Ensure the instrument lid is closed before and during the calibration.
transmittance when Ensure the lamp is working - if the lamp has failed replace the lamp
calibrating (7310) or lamp module (7315).
Unable to achieve a reading Ensure the correct cuvette is being used so that light isn't being
when measuring a sample absorbed by the cuvette.
Ensure the sample isn't too dense that light is not transmitted through
the sample.
Ensure the lamp is working.
Unable to add additional The table can only hold 6 points. Delete one or more points before
points to the spectral selecting additional points to add to the table.
points analysis table
Unable to print results using Ensure internal printer selected in autolog menu.
the internal printer Ensure there is paper in the unit.
Ensure there is a result displayed on screen.
Unable to save results to Ensure the USB stick is inserted into the front of the unit
the USB memory stick Ensure the USB memory stick has free capacity
Ensure there is a result displayed on the screen.
Unable to continue kinetics Due to the nature of the kinetics measurements stopping the
measurement after measurement before completion will result in some measurement
cancelling abort action time being lost. Therefore the experiment cannot be continued. It must
be restarted.
The peaks and valleys table The peaks and valleys threshold is turned off. Change this value to 1 %,
icon is not displayed 5% or 10% and the peaks and valleys table icon will be displayed.
following a spectrum scan
The measure sample icon A calibration must be performed at the new wavelength. When the
disappears after changing calibration is complete the measure sample icon will be displayed.
the wavelength
Jenway have a dedicated Technical Support team made up of experienced scientists who are on
hand to help with any applications advice and questions you may have about our products and
how to use them. If you require any technical or application assistance please contact the team at:
E-mai I: [email protected].
75
SECTION 14 - Declaration of Conformity
76
SECTION 14 - Declaration of Conformity
(Mr C. Warren)
77
SECTION 15 - Glossary of Icons
Common
0 Back key
Common
../ Tick icon - Done/yes
Common
x Cross icon - Cancel/no
Common
@) Result already saved to USB memory stick
Common ~m 1 .·· ..
~-··· Single method open icon
78
Common Calibrate to zero icon
Common
ffi Delete
Common
Common
m Method icon - opens method selection screen
Create method
~
Common Open method
Main Menu
]~
400 J-00
Opens spectrum measurement mode
79
Main Menu x Instrument settings
Auto log Timer icon - Time interval between each sample measurement repetition
030s
Autolog ~~ Batch ABC icon - Set batch name for repeated measurements
c
80
Instrument [±) Diagnostics
Settings
Photometrics m. Measure sample
Concentration
m.F Measure to a factor
Concentration
& Check standard
p~ ,•:,•:
81
Spectrum Single less than icon - decrease the wavelength by 1 times scan interval
Spectrum Double greater than icon - increase the wavelength by 1 O times scan
interval
Spectrum
> Single greater than icon - increase the wavelength by 1 times scan
interval
Quantitation
/ Curve fit - enables the curve fit to be adjusted
Quantitation
L Sigma icon - Displays curve statistics
Quantitation
0 Back icon - Re-measure previous standard
Quantitation
x Cross icon - Cancel creation of new standard curve
Quantitation
/ Curve statistics - used in print setup
Kinetics
x Settings menu
Kinetics 000 000 - Resets lag time, standard or factor values to zero
Kinetics
II 1,:T
AbS
ABS/%T icon - Set the start on level
82
Kinetics ~ . Start kinetics measurement
Kinetics 1:1 Toggle icon - switches between the initial and final point lines
Kinetics { Double less than icon - Decrease the time by 5 seconds intervals
Kinetics
< Single less than icon - Decrease the time by 1 second intervals
Kinetics Double greater than icon - Increase the time by 5 seconds intervals
~
Kinetics
Kinetics
>. Single greater than icon - Increase the time by 1 second intervals
p] fitted
Accessories
0 8 cell turret icon - the automatic 8 cell turret is in use
Accessories
(I Error finding cell position o on turret
Accessories
& No turret - Method created on a unit with a turret accessory fitted
fJ}::}
Accessories oc Degrees Celsius
Accessories
& No peltier - Method created on a unit with a peltier accessory fitted
f);/
Accessories Sipper pump in use - running in forward direction
~;
Accessories
,;, Sipper pump in use - running in reverse direction
Accessories
> Start sipper pump - enables take up time to be set
83
Accessories Skip take up time - uses the previously set take up time
~
Accessories
- Reduce sample uptake/reduce air gap
Accessories
+ Increase sample uptake/increaseair gap
Accessories
&fj•, No sipper - Method created on a unit with a sipper accessory fitted
Accessories ~J
I> 0
Sipper peltier pump in use
&
Accessories No sipper peltier - Method created on a unit with a sipper peltier
accessory fitted
fj}
IW
Accessories PC comms - shown when the instrument is connected to a PC and
communicating via the PC software
,/
Error Spanner - Refer to section 13
84
INDEX Page
ACCESSORIES 60
Accessories - connecting 60
Accessories - using 65
AUTOLOGGING 58
CONCENTRATION 27
CAUBRATION 29
METHOD SETUP 27
MODE SPECIFIC PARAMETERS 27
SAMPLE MEASUREMENT 30
Connecting to a PC 59
Controls
Display 13
Keypad 13
Rear panel 15
Declaration of conformity 76
Error codes 73
Good practice guidelines 17
ICON GLOSSARY 78
Installation 12
Unpacking 12
Instrument setup 19
Contrast 23
Expanded operating menu 19
GLP settings 22
Lamp save 23
Mode selection 22
Navigating 19
Utility toolbar 20
KINETICS 44
CALIBRATION 48
DATA ANALYSIS 49
METHOD SET UP 45
MODE SPECIFIC PARAMETERS 44
SAMPLE MEASUREMENT 48
Lamp replacement
Tungsten halogen lamp replacement 71
Xenon lamp module replacement 71
Language 56
PHOTOMETRICS 25
CALIBRATION 26
METHOD SET UP 27
MODE SPECIFIC PARAMETERS 25
SAMPLE MEASURMENT 26
PRINTING 55
Print set up 56
QUANTITATION 38
CAUBRATION 42
DATA ANALYSIS 43
METHOD SETUP 39
MODE SPECIFIC PARAMETERS 38
SAMPLE MEASUREMENT 42
85
Routine maintenance 71
Safety 3
SAVING, PRINTING AND AUTOLOGGING 51
DELfTING METHODS 53
DELETING RESULTS 55
OPENING METHODS 52
OPENING RESU LTS 54
PRINTING RESULTS 55
SAVING METHODS 51
SAVING RESULTS 53
Screen setup 19
SERVICE 71
Spare parts 70
SPECTRUM 31
CALIBRATION 34
DATA ANALYSIS 35
METHOD SETUP 32
MODE SPECIFIC PARAMETERS 31
SAMPLE MEASUREMENT 34
Technical specification 8
Technical support 75
Theory and practice of spectroscopy measurements 16
Spectroscopy measurement 16
Theory of spectroscopy measurements 16
Troubleshooting 73
Utility toolbar 20
86
87
Bibby Scientific
Bibby Scientific Ltd Bibby Scientific France SAS Bibby Scientific Italia Sri
BeaconRoad Stone Zl du Rocher Vert - BP 79 Via Alcide de Gasperi56
StaffordshireSTl 5 OSA 77793 Nemours Cedex 20077 Riozzo di Cerro al Lambro
United Kingdom France Milano Italia
Tel: +44 (0)1785 812121 Tel: +33 1 64 45 13 13 Tel: +39 (0)2 98230679
Fax: +44 (0)1785 813748 Fax: +33 16445 13 00 Fax: +39 (0)2 98230211
e-mail: [email protected] e-mail: [email protected] e-mail: [email protected]
www.bibby-scientific.com www.bibby-scientific.com www.bibby-scientific.it
Bibby Scientific Middle East Ltd Bibby Scientific US Ltd Bibby Scientific(Asia) Ltd
BPO Box 27842 3 Terri Lane Suite 10 Room 607 Yen Sheng Centre
Engomi 2433 Burlington NJ 08016 64 Hoi Yuen Road Kwun Tong
Nicosia Cyprus USA Kowloon Hong Kong
Tel: +357 22 660 423 Tel: 800-225-9243 Tel: +852 3583 1581
Fax: +357 22 660 424 Fax: 609-589-2571 Fax: +852 3583 1580
e-mail: [email protected] www.bibby-scientific.com e-mail [email protected]
www.bibby-scientific.com www.bibby-scientificasia.com