Lifepak 20 PM Checklist
Lifepak 20 PM Checklist
3201896_R_PIP_Checklist_LP20.pdf Page 1 of 3
PIP Checklist
Testing
1 Power On
a. Confirm the Service indicator is off.
b. Confirm the device completes the Power On sequence.
2 Date and Time
Check/set date and time
3 ECG Lead Characteristics
a. Confirm leads off screen messages.
b. Record lead II ECG gain (tolerance 36 to 44 mm) Measured
Value _______________
c. Record lead I ECG gain (tolerance 18 to 22 mm)* Measured
Value _______________
*(Test Equipment other then 215A/217A may have different output)
d. Record lead V1/C ECG gain (tolerance 36 to 44 mm) Measured
Value _______________
(N/A if 3 Lead)
4 Printer Test
Confirm printed test strip and CHECK PRINTER message.
5 SpO2 Oximeter (if SpO2 option is installed)
Confirm SpO2 reading is between 90% and 100%.
6 Therapy Impedance Sense
a. Confirm display of PADDLES LEADS OFF message. (248 Ohms)
b. Confirm no display of PADDLES LEADS OFF message. (182 Ohms)
7 Therapy User test
Confirm Device passes User Test
8 Therapy Delivered Energy and Sync
a. Record Sync R-wave delay (maximum 60 ms): Measured
Value _______________
b. 2 J – Record delivered energy (tolerance 1.0 to 3.0 J). Measured
Value _______________
c. 70 J – Record delivered energy (tolerance 65.1 to 74.9 J). Measured
Value _______________
d. 360 J – Record delivered energy (tolerance 334.8 to 385.2 J). Measured
Value _______________
9 Therapy Paddles ECG Gain and AED mode Test
a. Record ECG gain (tolerance 36 to 44 mm) Measured
Value _______________
(Note: 38 to 50 mm with QED-6)
b. AED / Manual modes check
10 Therapy Remote Sync Test
a. Confirm Remote Sync: Sync LED is flashing
b. Confirm correct response to energy transfer