RANDOX Procedure For Glucose GOD-PAP Assay Without Deproteinisation
RANDOX Procedure For Glucose GOD-PAP Assay Without Deproteinisation
CE
PROCEDURE FOR GLUCOSE GOD-PAP ASSAY NORMAL VALUES0,4)
wITHOUT DEPROTEINISATION
Serum/Plasma 4.2-6.4 mmol/l (75-115 mg/di)
Wavelength: 500 nm, Hg 546 nm
Cuvette: cm path length Random urine 0. 0.8 mmol/l ( I5 mg/dl)
Temperature: I5-25°C or 37°C
Measurement against reagent blank 24 hour urine <2.78 mmol/d (0.5 g/d)
Pipette into test tubes It is recommended that each laboratory ce_tablish its own
reference range to reflect the age, sex, diet and geographical
Macro Semi Micro location of the population.
SENSITIVITY
Mix, incubate for 25 min at 15-25°C or 10 min at 37°C. Measure lt is recommended that each laboratory establishes its own range
the absorbance of the standard (A sandard) and the sample (A of sensitivity as this is limited by the sensicivicy of the
sample) against the reagent blank within spectrophotometer used. Under the conditions of the assay a
60 minutes. Although the absorbance may be read at any time change of 0.01 absorbance units is equivalent to
up to 60 minutes after the specified incubation time, the time 0.013 mmol/l (0.23 mg/dl).
interval from sample addition to read time must be exactly the
same for Standard/Control and Sample. REFERENCES
I.Barham, D, and Trinder, P. Analyst 1972; 97: I42.
CALCULATION 2. Clinical Chemistry Principles and Techniques, Second Edition,
RJ. Henry. D.C. Cannon and J.W. Winkelman Editors, Harper
and Row, Maryland, USA, I288, I974.
Asample Standard conc.
Glucose 3. Teuscher, A. and Richterich, P. Schweiz Med. Wschr., 1971;
(mmol/l) (mmol/l)
Concentration Asandard 101:345 and 390.
4. Tietz. N.W. Clinical Guide to laboratory tests. 2nd edition.
Philadelphia. Pa: WB Saunders Co.; I990: 246-250
Asample
x Standard conc.
(mg/dl)
Astandard (mg/d)
QUALITY CONTROL
Randox Assayed Multi-sera, Level 2 and Level 3 are recommended
for daily quality control. Two levels of controls should be assayed
at least once a day. Values obtained should fall within a specified
range. If these values fall outside the range and repetition excludes
error, the following steps should be taken:
. Check instrument settings and light source
2. Check cleanliness of all equipment in use.
3. Check water, contaminants i.e. bacterial growth may contribute
to inaccurate results.
4. Check reaction temperature.
5. Check expiry date of kit and contents.
6. Contact Randox Laboratories Customier Technical
Support, Northern Ireland (028) 94422413.
INTERFERENCES
The analytes below were tested up to the following levels and
were found not to interfere: