Application Book
Application Book
Index
Summary
The following application notes are suitable only for the configuration described in each notes
Application Note SK12-AG-01
Pine Leaves
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten by hands.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-AG-02
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-AG-03
Ground Vegetable
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-AG-04
Maize
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-AG-05
Hay
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-AG-06
Beans
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-AG-07
Lentils
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-AG-08
Roots
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
9 ml of HNO3 65%, 1 ml of HF 40%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-AG-09
Tea Leaves
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-01
Acid Slag
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
7 ml of HNO3 65%, 3 ml of HF 40%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-02
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-03
Grass
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-04
Leaves
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-05
Municipal Sludge
Summary
This method provides the acid digestion of the
sample in a closed vessel device using 7. Cool the rotor by air or by water until the
temperature control microwave heating for the solution reaches room temperature.
8. Open the vessel and transfer the solution
metal determination by spectroscopic methods.
to a marked flask.
Instrumentation
Microwave program
Microwave acid digestion apparatus (1) (2)
Step Time T1 T2 P Power
Milestone Ethos or Start labstation with internal 1 00:20:00 180°C 120°C 30 bar Max power*
temperature sensor, 640-260 terminal with 2 00:15:00 180°C 120°C 30 bar Max power*
easyCONTROL software installed and Pro 24/16 (1), (2) Optional sensors
high throughput rotor. Temperature profile
Sample weight
Up to 0.2 g
Reagents
6 ml of HNO3 65%, 2 ml of H2O2 30%,
2 ml of HF 40%
Procedure
1. Place a TFM vessel on the balance plate, tare
Note
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-06
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
6 ml of HNO3 65%, 4 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-07
Pollen
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
10 ml of HNO3 65%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-08
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of HF 40%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-09
Wastewater
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 2 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-10
Wood chips
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.2 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-11
Industrial Wastewater
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 1 g
Reagents
6 ml of HNO3 65%, 2 ml of H2O2 30%,
2 ml of HF 40%.
Procedure Note
1. Place a TFM vessel on the balance plate, tare *Max power: 1500W for Ethos and 1200W for Start units.
it and weigh of the sample. Use up to 500 Watt for operations with 5 or less vessels
2. Introduce the TFM vessel into the HTC safety simultaneously.
shield.
This procedure is only a guideline and it may need to be
3. Add the acids; if part of the sample stays on
modified or changed to obtain the required results on your
the inner wall of the TFM vessel, wet it by
sample.
adding acids drop by drop, then gently swirl
the solution to homogenize the sample with Milestone suggest to use pressure sensor (P) only for
the acids. method developments or unknown samples. See the
4. Close the vessel and tighten the cover. pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity Always use hand, eye and body protection when operating
and connect the temperature sensor. with the microwave system.
6. Run the microwave program to completion.
-Rev 0/2011-
Application Note SK12-EN-12
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.25 g
Reagents
6 ml of HNO3 65%, 4 ml of HF 40%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK-EN-13
EPA 3051
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
10 ml of HNO3 65%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-14
Green Algae
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-15
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-EN-16
Raw Waste
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
7 ml of HNO3 65%, 2 ml of H2O2 30%
1 ml of HF 40%.
Procedure Note
1. Place a TFM vessel on the balance plate, tare *Max power: 1500W for Ethos and 1200W for Start units.
it and weigh of the sample. Use up to 500 Watt for operations with 5 or less vessels
2. Introduce the TFM vessel into the HTC safety simultaneously.
shield.
This procedure is only a guideline and it may need to be
3. Add the acids; if part of the sample stays on
modified or changed to obtain the required results on your
the inner wall of the TFM vessel, wet it by
sample.
adding acids drop by drop, then gently swirl
the solution to homogenize the sample with Milestone suggest to use pressure sensor (P) only for
the acids. method developments or unknown samples. See the
4. Close the vessel and tighten the cover. pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity Always use hand, eye and body protection when operating
and connect the temperature sensor. with the microwave system.
6. Run the microwave program to completion.
-Rev 0/2011-
Application Note SK12-EN-17
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
6 ml of HNO3 65%, 1 ml of H2O2 30%
3 ml of HF 40%.
Procedure Note
1. Place a TFM vessel on the balance plate, tare *Max power: 1500W for Ethos and 1200W for Start units.
it and weigh of the sample. Use up to 500 Watt for operations with 5 or less vessels
2. Introduce the TFM vessel into the HTC safety simultaneously.
shield.
This procedure is only a guideline and it may need to be
3. Add the acids; if part of the sample stays on
modified or changed to obtain the required results on your
the inner wall of the TFM vessel, wet it by
sample.
adding acids drop by drop, then gently swirl
the solution to homogenize the sample with Milestone suggest to use pressure sensor (P) only for
the acids. method developments or unknown samples. See the
4. Close the vessel and tighten the cover. pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity Always use hand, eye and body protection when operating
and connect the temperature sensor. with the microwave system.
6. Run the microwave program to completion.
-Rev 0/2011-
Application Note SK12-EN-18
Urban Wastewater
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of HF 40%.
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-01
Ham
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-02
Agarose
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-03
Baby food
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-04
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-05
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.5 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%
Procedure
1. Place a TFM vessel on the balance plate, tare Note
it and weigh of the sample.
*Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety
shield. Use up to 500 Watt for operations with 5 or less vessels
3. Add the acids; if part of the sample stays on simultaneously.
the inner wall of the TFM vessel, wet it by This procedure is only a guideline and it may need to be
adding acids drop by drop, then gently swirl modified or changed to obtain the required results on your
the solution to homogenize the sample with sample.
the acids.
Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover.
method developments or unknown samples. See the
5. Insert the segment into the microwave cavity
pressure sensor user manual for more information.
and connect the temperature sensor.
6. Run the microwave program to completion. Always use hand, eye and body protection when operating
with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-06
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.5 g
Reagents
15 ml of HNO3 65%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-07
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
10 ml of HNO3 65%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-08
Bovine Milk
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-09
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare *Max power: 1500W for Ethos and 1200W for Start units.
it and weigh of the sample.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
simultaneously.
shield.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with Milestone suggest to use pressure sensor (P) only for
the acids. method developments or unknown samples. See the
4. Close the vessel and tighten the cover. pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
Always use hand, eye and body protection when operating
and connect the temperature sensor.
with the microwave system.
6. Run the microwave program to completion.
-Rev 0/2011-
Application Note SK12-FO-10
Casein
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-11
Cheese Ashes
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.5 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-12
Chewing Gum
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.2 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-13
Chocolate
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-14
Cocoa Powder
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-15
Corn Starch
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-16
Dried Fish
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-17
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-18
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-19
Fish in Oil
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-20
Fresh Kidney
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-21
Fresh Liver
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-22
Lecithin
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011
Application Note SK12-FO-23
Lyophilized Meat
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011
Application Note SK12-FO-24
Margarine
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.2 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011
Application Note SK12-FO-25
Milk Powder
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011
Application Note SK12-FO-26
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-27
Mint Oil
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-28
Monogliceride
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.1 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-29
Mushroom
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-30
Noodles
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-31
Olives
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-32
Oyster Tissue
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-33
Palm Oil
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.1 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-34
Pet Food
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-35
Raw Sugar
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note PRO-FO-36
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 4 ml
Reagents
8 ml of HNO3 65%, 2ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 04_09-
Application Note SK12-FO-37
Rice Flour
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
6 ml of HNO3 65%, 1 ml of H2O2 30%
3 ml of H2O.
Note
Procedure
*Max power: 1500W for Ethos and 1200W for Start units.
1. Place a TFM vessel on the balance plate, tare Use up to 500 Watt for operations with 5 or less vessels
it and weigh of the sample. simultaneously.
2. Introduce the TFM vessel into the HTC safety
shield. This procedure is only a guideline and it may need to be
3. Add the acids; if part of the sample stays on modified or changed to obtain the required results on your
sample.
the inner wall of the TFM vessel, wet it by
adding acids drop by drop, then gently swirl Milestone suggest to use pressure sensor (P) only for
the solution to homogenize the sample with method developments or unknown samples. See the
the acids. pressure sensor user manual for more information.
4. Close the vessel and tighten the cover. Always use hand, eye and body protection when operating
5. Insert the segment into the microwave cavity with the microwave system.
and connect the temperature sensor.
6. Run the microwave program to completion.
-Rev 0/2011-
Application Note SK12-FO-38
Salami
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-39
Sheep Milk
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-40
Soya Fat
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.1 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-41
Soya Oil
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-42
Soybean
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-43
Sunflower Seeds
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-44
Tea Leaves
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-45
Tinned Beef
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-46
Tinned Pork
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-47
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-48
Vegetable Oil
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-49
Wheat Flour
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011 -
Application Note SK12-FO-50
Tobacco Leaves
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-51
Fertilizer
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-52
Lichen
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-53
Barley
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on
This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by
modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl
sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
4. Close the vessel and tighten the cover. method developments or unknown samples. See the
5. Insert the segment into the microwave cavity pressure sensor user manual for more information.
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-54
Hop
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-55
Grass
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-56
Grape Leaves
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-57
Dried Graminacee
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-58
Cereals
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-59
Spinache
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-60
Tomatoes
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-61
Wine
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.2 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-62
Coffee
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.2 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-63
Butter
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.2 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-64
Fresh Meat
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.5 g (0.3 g of dry sample max)
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-0/2011-
Application Note SK12-FO-65
Salad
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-66
Vinegar
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-67
Yeast
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-
Application Note SK12-FO-68
Corn Flakes
Summary
This method provides the acid digestion of the 7. Cool the rotor by air or by water until the
sample in a closed vessel device using solution reaches room temperature.
temperature control microwave heating for the 8. Open the vessel and transfer the solution
metal determination by spectroscopic methods. to a marked flask.
Sample weight
Up to 0.3 g
Reagents
8 ml of HNO3 65%, 2 ml of H2O2 30%.
Procedure
Note
1. Place a TFM vessel on the balance plate, tare
it and weigh of the sample. *Max power: 1500W for Ethos and 1200W for Start units.
2. Introduce the TFM vessel into the HTC safety Use up to 500 Watt for operations with 5 or less vessels
shield. simultaneously.
3. Add the acids; if part of the sample stays on This procedure is only a guideline and it may need to be
the inner wall of the TFM vessel, wet it by modified or changed to obtain the required results on your
adding acids drop by drop, then gently swirl sample.
the solution to homogenize the sample with
the acids. Milestone suggest to use pressure sensor (P) only for
method developments or unknown samples. See the
4. Close the vessel and tighten the cover.
pressure sensor user manual for more information.
5. Insert the segment into the microwave cavity
and connect the temperature sensor. Always use hand, eye and body protection when operating
6. Run the microwave program to completion. with the microwave system.
-Rev 0/2011-