Density or Relative Density of Light Hydrocarbons by Pressure Hydrometer
Density or Relative Density of Light Hydrocarbons by Pressure Hydrometer
weaken plastics while aromatics also tend to attack the surface of plastics 6.2.1 The liquid inlet valve and the liquid outlet valve shall
and should similarly not be used. A light aliphatic hydrocarbon is be tightly connected to a base plate that shall be bored to give
recommended for cleaning. (Warning—Extremely flammable. Harmful if
inhaled. Vapors may cause flash fire.) both valves a common inlet to the cylinder. The vapor vent
A1. APPARATUS
A1.1 Apparatus Verification and Certification A1.1.1.2 Hydrometers shall be certified or verified at inter-
A1.1.1 Hydrometers, shall either be certified or verified. vals of no more than 24 months.
Verification shall be either by comparison with a certified A1.1.2 Thermometers, shall be verified at intervals of no
hydrometer (see 6.1.1) or by the use of a certified reference more than six months for conformance with specifications.
material (CRM) specific to the reference temperature used. Either comparison with a referenced temperature measurement
A1.1.1.1 The hydrometer scale shall be correctly located system traceable to an international standard, or a determina-
within the hydrometer stem by reference to the datum mark. If tion of ice point, is suitable.
the scale has moved, reject the hydrometer.
SUMMARY OF CHANGES
Subcommittee D02.02/COMQ has identified the location of selected changes to this standard since the last
issue (D1657–02(2007)) that may impact the use of this standard.
(1) Added 1.3–1.6 to provide additional clarification. (5) Enhanced safety note in 6.2 to provide improve guidance.
(2) Added ASTM Test Method D1298 (API MPMS Chapter (6) Revised 9.6, 11.2, 11.2.1, and 11.2.2 to update the correct
9.1) to the Referenced Documents and standard text. GPA reference.
(3) Added an additional sentence to 4.1 to provide additional (7) Added 11.4 and 12.3 to clarify to the user the reliability of
clarification on what additional corrections are needed after the precision and bias statement.
obtaining the observed hydrometer reading.
(4) Added 6.1.1.3 to clarify the differences between verified
and certified apparatus; the need for the user to ascertain that
their instrument conforms to the criteria in the standard and the
various corrections that the user may need to perform to the
observed hydrometer reading to complete the determination
procedure.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/