Transducers
Transducers
Data Sheet
System NL description Clamp-on Flow Transducers for Liquids
temperature
connection
transducer
frequency
extension
Ex certifi-
system
cation
cable
type
FS set of ultrasonic flow transducers for liquids, shear wave Non-Invasive Flow Measurement of Liquids with FLUXUS® F
S- 8 MHz, for diameter range see Technical Data
Q- 4 MHz, for diameter range see Technical Data
Clamp-on flow transducers are mounted onto the pipe
System AS M- 1 MHz, for diameter range see Technical Data
instead of inserted into it. Since they do not come in contact
K- 0.5 MHz, for diameter range see Technical Data
with the medium, they are not subjected to wear. Clamp-on
G- 0.2 MHz, for diameter range see Technical Data flow measurement is applicable even for chemically
N normal temperature range, max. 130 °C aggressive, corrosive or ultra pure substances. The retrofit-
E increased temperature range, max. 200 °C ting of existing installations can be done without cutting into
NN no explosion protection the pipe and without interruption of the process.
A2 explosion protection certification for ATEX zone 2 All FLEXIM clamp-on transducers are watertight and espe-
F2 explosion protection certification for FM Class I Div. 2 (TS only) cially made for use in harsh industrial environments. The
System TS transducer housings and the transducer cable armored
A1 explosion protection certification for ATEX zone 1 (TS only)
direct connection: direct connection of explo- connection via junction box: connection via junction box of explo- sheaths are made of stainless steel. The cables of the
NL- connection system NL with Lemo connector explosion-proof transducers are sheathed with a special
sion-proof transducers: sion-proof transducers:
AS- connection system AS with Amphenol connector plastic material. The robust construction and the integrated
FLUXUS FLUXUS TS- connection system TS: direct connection or connection via junction box transducer cables guarantee good measurement results
XXX length of extension cable: the max. length depends on the transducer over a long period of extensive usage.
FSM-NA2 transducers in the Variofix mounting fixture
frequency (see table on page 5)
connection system TS: a junction box (JB02 for ATEX zone 2 or JB01 Features
Example: for ATEX zone 1) is included.
• Flow calibrated transducers
FS G- N A1 TS- 030 shear wave transducer 0.2 MHz normal temperature for ATEX zone 1,
connection system TS with junction box JB01 with extension cable 30 m • Automatic transducer detection
• Watertight stainless steel housing
• No contact with the medium, no corrosion,
hygienic measurement, suitable for ultra pure liquids
• Low stocking costs since only 2 types of transducers are
X Product Code of Lamb Wave Transducers needed to cover the main pipe diameter range
• Measurement is independent of fluid conductivity and
description pressure
temperature
connection
transducer
frequency
extension
Ex certifi-
X
• No pressure loss, no risk of leakage
system
FSQ-NA2 transducers fixed by chains
cation
cable
L L
type
X
X Technical Data
FL set of ultrasonic flow transducers for liquids, Lamb wave
K- 0.5 MHz, for diameter range see Technical Data
H- 0.3 MHz, for diameter range see Technical Data Shear wave transducers type FSQ-NA2, FSQ-NF2, Shear wave transducers FSM-NA2, FSM-NF2,
FSQ-NNN, FSQ-EA2, FSQ-ENN (corr. to type CDQ1NZ7/CDQ1N52, FSM-NNN, FSM-EA2, FSM-ENN (corr. to type CDM1NZ7/
G- 0.2 MHz, for diameter range see Technical Data
FLUXUS FLUXUS CDQ1N52, CDQ1NZ7/CDQ1N52, CDQ1EZ7, CDQ1EZ7) CDM1N52, CDM1N52, CDM1NZ7/CDM1N52, CDM1EZ7, CDM1EZ)
N normal temperature range
rated (possible) rated (possible)
NN no explosion protection diameter range1: (10) 25…400mm diameter range1: (50) 100…2500mm
Transducer cable lengths X and Y and maximum length L of the extension cable (in m) A2 explosion protection certification for ATEX zone 2 dimensions: (42.5 x 18 x 21.5) mm dimensions: (60 x 30 x 33.5)mm
F2 explosion protection certification for FM Class I Div. 2 (TS only) material: housing: stainless steel material: housing: stainless steel
transducer type connection type NL connection type AS connection type TS contact surface PEEK (FSQ-N**) contact surface PEEK (FSM-N**)
A1 explosion protection certification for ATEX zone 1 (TS only)
-*A1 - - Q: X=3m; L=90m or Polyimid (FSQ-E**) or Polyimid (FSM-E**)
M: X=4m; L=300m NL- connection system NL with Lemo connector degree of protection: IP65 acc. to EN60529 degree of protection: IP65 acc. to EN60529
K,H,G: X=5m; L=300m AS- connection system AS with Amphenol connector operating temperature: process and ambient: FSM-N**: IP68 on request
-*A2 und -*NN S: X=1m; Y=1m; L=20m S: X=1m; Y=1m; L=20m S: X = 2m; L = 40m TS- connection system TS: direct connection or connection via junction box FSQ-N**: -40°C…130°C; operating temperature: process and ambient:
Q: X=2m; Y=1m; L=50m Q: X=2 m; Y=1m; L=50m Q: X=3m; L=90m FSQ-E**: -40°C…200°C FSM-N**: -40°C…130°C;
XXX length of extension cable: the max. length depends on the transducer
M: X =2m; Y=2m; L=100m M: X =2m; Y=2m; L=100m M: X=4m; L=300m explosion hazard zone: FSQ-NA2:ATEX zone 2, marking: FSM-E**: -40°C…200°C
frequency (see table on page 5)
K,H,G: X =2m; Y=3m; L=100m K,H,G: X =2m; Y=3m; L=100m K,H,G: X=5m; L=300m II3G T6…T4 Ta -30°C…130°C explosion hazard zone: FSM-NA2:ATEX zone 2, marking:
connection system TS: a junction box (JB02 for ATEX zone 2 or JB01
-*F2 - - S: X = 2m; L = 40m FSQ-EA2:ATEX zone 2, marking: II3G T6…T4 Ta -30°C…130°C
Example: for ATEX zone 1) is included.
Q: X=3m; L=90m II3G T6…T4 Ta -30°C…200°C FSM-EA2:ATEX zone 2, marking:
M: X=4m; L=300m FL H- N A1 TS- 030 lamb wave transducer 0.3 MHz normal temperature for ATEX zone 1,
FSQ-NF2: Div. 2 FM-C, marking: II3G T6…T4 Ta -30°C…200°C
K,H,G: X=5m; L=300m connection system TS with junction box JB01 with extension cable 30 m
NI/Cl. I, II, III / Div. 2 / FSM-NF2: Div. 2 FM-C, marking:
Gp A-G / T4 Ta = 125°C NI/Cl. I, II, III / Div. 2 /
Gp A-G / T4 Ta = 125°C
FLEXIM GmbH
Wolfener Str. 36
D-12681 Berlin
Subject to change without notification. GSCLAMPLIQUIDV2-2XEN internet: www.flexim.de Fon:+49 (30) 93 66 76 60 Subject to change without notification. GSCLAMPLIQUIDV2-2YEN
FLUXUS® is a protected trademark of FLEXIM GmbH. page 5 of 6 e-mail: [email protected] Fax:+49 (30) 93 66 76 80 FLUXUS® is a protected trademark of FLEXIM GmbH. page 1 of 6
Technical Data (continued) Dimensions of the Transducers (in mm)
FSM-NA2, FSM-NF2, FSM-NNN, FSM-EA2, FSM-ENN FSQ-NA2, FSQ-NNN, FSQ-EA2, FSQ-ENN FLG-NA1, FLH-NA1, FLK-NA1 FLG-NA2, FLG-NF2, FLG-NNN, FLH-NA2, FLH-NF2,
Shear wave transducers FSM-NA1, FSQ-NA1 Shear wave transducers FSS-NF2, FSS-NNN
(corr. to type CDM1N31, CDQ1N31) (corr. to type CDS1NZ7/CDS1N52, CDS1N52, CDS1NZ7/CDS1N52) FLH-NNN, FLK-NA2, FLK-NF2, FLK-NNN
rated (possible) FSM: (50) 100…2500mm rated (possible)
diameter range1: FSQ: (10) 25…400mm diameter range1: (6) 10…70mm
30
18
dimensions: (79.5 x 30 x 39)mm dimensions: (26 x 13 x 17)mm
50
material: housing: stainless steel material: housing: stainless steel
47
contact surface: PEEK contact surface: Polyetherimid 60 ###1N33
degree of protection: IP65 acc. to EN60529 degree of protection: IP65 acc. to EN60529
21.5
operating temperature: process: -40°C...130°C operating temperature: process: -40°C...130°C
ambient: -40°C...130°C ambient: -40°C...130°C 128.5
42.5
33.5
use in explosive atmosphere explosion hazard zone: FSS-NF2: Div. 2 FM-C, marking:
- hazard zone: zone 1 and 2 NI/Cl. I, II, III / Div. 2 /
- marking: 0044; II2G Gp A-G / T4 Ta = 125°C
67.5
70
EEx m II T6…T4 Ta -20°C…120°C Lamb wave transducers FLG-NA2, FLG-NF2,
- certification: IBExU 98 ATEX 1012 X FLG-NNN, FLH-NA2, FLH-NF2, FLH-NNN, FSM-NA1, FSQ-NA1 FSS-NNN FSS-NF2 mit
- type of protection: encapsulation FLK-NA2, FLK-NF2, FLK-NNN Anschlusssystem NL
(corr. to type CRG1NC3, CRG1N52, CRG1N52/CRG1NC3,
128.5
Shear wave transducers FSK-NA2, FSK-NF2, FSK-NNN, CRH1NC3, CRH1N52, CRH1N52/CRH1NC3,
FSG-NA2, FSG-NF2, FSG-NNN CRK1NC3, CRK1N52, CRK1N52/CRK1NC3)
(corr. to type CDK1NZ7/CDK1N52, CDK1N52, CDK1NZ7/CDK1N52, rated (possible) range FLG:
CDG1NZ7/CDG1N52, CDG1N52, CDG1NZ7/CDG1N52)
of application1: Junction Box JB0X
30
rated diameters
rated (possible) FSK: (100) 200…3600mm (400) 500 to 6500 mm for wall
diameter range1: FSG: (400) 500… 6500mm thickness (14 to 27) mm Dimensions (in mm)
dimensions: FSK: (126.5 x 47 x 56)mm
24.50
FLH:
13.2
FSG: (129.5 x 47 x 66.5)mm rated diameters (250) 330 to 5000 mm
38.9
30.10
material: PEEK with stainless steel cap for wall thickness (9 to 18) mm
degree of protection: IP65 acc. to EN60529
15
FLK:
operating temperature: process: -40°C…130°C rated diameters (100) 200 to 3600 mm
60 26 32.70
ambient: -40°C…130°C for wall thickness (5 to 11) mm
explosion hazard zone: FSK-NA2, FSG-NA2: ATEX zone 2 dimensions: (128.5 x 50 x 67.5)mm 79.5
marking: material: PPSU with stainless steel cap
II3G T6…T4 Ta -40°C…130°C degree of protection: IP65 acc. to EN60529
FSG-NA2, FSG-NNN, FSG-NF2 FSK-NA2, FSK-NNN, FSK-NF2
156
FSK-NF2, FSG-NF2: Div. 2 FM-C operating temperature: process: -40°C…170°C
174
marking: ambient: -40°C…170°C
NI/Cl. I, II, III / Div. 2 / explosion hazard zone: FLG-NA2, FLH-NA2, FLK-NA2:
Gp A-G / T4 Ta = 125°C ATEX zone 2, marking:
II3G T6…T4 Ta -40°C…130°C
Shear wave transducers FSK-NA1, FSG-NA1
47
47
(corr. to type CDK1N31, CDG1N31) FLG-NA2, FLH-NA2, FLK-NA2:
rated (possible) FSK: (100) 200…3600mm Div. 2 FM-C, marking:
diameter range1: FSG: (400) 500…6500mm NI/Cl. I, II, III / Div. 2 /
dimensions: FSK-NA1: (126.5 x 50 x 53.5)mm Gp A-G / T4 Ta = 125°C
129.5 126.5
FSG-NA1: (129.5 x 50 x 64)mm Lamb wave transducers FLG-NA1, FLH-NA1, Ø
material: PEEK with stainless steel cap 70 9
FLK-NA1 (corr. to type CRG1N33, CRH1N33, CRK1N33)
degree of protection: IP65 acc. to EN60529 2
rated (possible) range FLG: 119 163.5
operating temperature: process: -40°C…130°C of application1: rated diameters (400) 500 to 6500 mm
66.5
56
ambient: -40°C…130°C for wall thickness (14 to 27) mm
use in explosive atmosphere FLH:
hazard zone: ATEX zone 1 and 2 rated diameters (250) 330 to 5000 mm
marking: 0044; II2G for wall thickness (9 to 18) mm
Ex q II T6…T3 Ta -40°C…180°C FLK: Technical Data
certification: IBExU 04 ATEX 1011 X rated diameters (100) 200 to 3600 mm FSK-NA1 FSG-NA1
type of protection: powder filling for wall thickness (5 to 11) mm Junction box JBXX
dimensions: (128.5 x 50 x 67.5)mm dimensions: see above
material: PPSU with stainless steel cap possible materials: JB01S4E3M, JB01S4E3P: stainless steel SS316 1.4404
degree of protection: IP65 acc. to EN60529 all others: stainless steel SS304 1.4301
50
50
operating temperature: process: -40°C…170°C gasket: - zone 1: silicone
ambient: -40°C…170°C - zone 2: EPDM
transducers FLG-NA1, FLH-NA1, FLK-NA1: degree of protection: IP67 acc. to EN60529
126.5 129.5 operating temperature: -40°C to 80°C
explosion hazard zone: ATEX zone 1 and 2
marking: 0044; II2G Additionally for the types JB01S4E3M, JB01S4E3P
Ex q II T6…T3 Ta -40°C…140°C explosion hazard zone: zone 1 (ATEX)
certification: IBExU 04 ATEX 1011 X marking: 0044; II2G
64
type of protection: powder filling 53.5 EEx e mb II T4...T6 Ta -40°C...80°C
certification: IBExU 01 ATEX 1161
type of protection: junction box: increased safety
decoupled network: encapsulation
1 The range in parentheses is possible under good conditions, but is not specified.
Subject to change without notification. GSCLAMPLIQUIDV2-2YEN Subject to change without notification. GSCLAMPLIQUIDV2-2XEN Subject to change without notification. GSCLAMPLIQUIDV2-2XEN
FLUXUS® is a protected trademark of FLEXIM GmbH. page 2 of 6 FLUXUS® is a protected trademark of FLEXIM GmbH. page 3 of 6 FLUXUS® is a protected trademark of FLEXIM GmbH. page 4 of 6
Technical Data (continued) Dimensions of the Transducers (in mm)
FSM-NA2, FSM-NF2, FSM-NNN, FSM-EA2, FSM-ENN FSQ-NA2, FSQ-NNN, FSQ-EA2, FSQ-ENN FLG-NA1, FLH-NA1, FLK-NA1 FLG-NA2, FLG-NF2, FLG-NNN, FLH-NA2, FLH-NF2,
Shear wave transducers FSM-NA1, FSQ-NA1 Shear wave transducers FSS-NF2, FSS-NNN
(corr. to type CDM1N31, CDQ1N31) (corr. to type CDS1NZ7/CDS1N52, CDS1N52, CDS1NZ7/CDS1N52) FLH-NNN, FLK-NA2, FLK-NF2, FLK-NNN
rated (possible) FSM: (50) 100…2500mm rated (possible)
diameter range1: FSQ: (10) 25…400mm diameter range1: (6) 10…70mm
30
18
dimensions: (79.5 x 30 x 39)mm dimensions: (26 x 13 x 17)mm
50
material: housing: stainless steel material: housing: stainless steel
47
contact surface: PEEK contact surface: Polyetherimid 60 ###1N33
degree of protection: IP65 acc. to EN60529 degree of protection: IP65 acc. to EN60529
21.5
operating temperature: process: -40°C...130°C operating temperature: process: -40°C...130°C
ambient: -40°C...130°C ambient: -40°C...130°C 128.5
42.5
33.5
use in explosive atmosphere explosion hazard zone: FSS-NF2: Div. 2 FM-C, marking:
- hazard zone: zone 1 and 2 NI/Cl. I, II, III / Div. 2 /
- marking: 0044; II2G Gp A-G / T4 Ta = 125°C
67.5
70
EEx m II T6…T4 Ta -20°C…120°C Lamb wave transducers FLG-NA2, FLG-NF2,
- certification: IBExU 98 ATEX 1012 X FLG-NNN, FLH-NA2, FLH-NF2, FLH-NNN, FSM-NA1, FSQ-NA1 FSS-NNN FSS-NF2 with
- type of protection: encapsulation FLK-NA2, FLK-NF2, FLK-NNN connection system NL
(corr. to type CRG1NC3, CRG1N52, CRG1N52/CRG1NC3,
128.5
Shear wave transducers FSK-NA2, FSK-NF2, FSK-NNN, CRH1NC3, CRH1N52, CRH1N52/CRH1NC3,
FSG-NA2, FSG-NF2, FSG-NNN CRK1NC3, CRK1N52, CRK1N52/CRK1NC3)
(corr. to type CDK1NZ7/CDK1N52, CDK1N52, CDK1NZ7/CDK1N52, rated (possible) range FLG:
CDG1NZ7/CDG1N52, CDG1N52, CDG1NZ7/CDG1N52)
of application1: Junction Box JB0X
30
rated diameters
rated (possible) FSK: (100) 200…3600mm (400) 500 to 6500 mm for wall
diameter range1: FSG: (400) 500… 6500mm thickness (14 to 27) mm Dimensions (in mm)
dimensions: FSK: (126.5 x 47 x 56)mm
24.50
FLH:
13.2
FSG: (129.5 x 47 x 66.5)mm rated diameters (250) 330 to 5000 mm
38.9
30.10
material: PEEK with stainless steel cap for wall thickness (9 to 18) mm
degree of protection: IP65 acc. to EN60529
15
FLK:
operating temperature: process: -40°C…130°C rated diameters (100) 200 to 3600 mm
60 26 32.70
ambient: -40°C…130°C for wall thickness (5 to 11) mm
explosion hazard zone: FSK-NA2, FSG-NA2: ATEX zone 2 dimensions: (128.5 x 50 x 67.5)mm 79.5
marking: material: PPSU with stainless steel cap
II3G T6…T4 Ta -40°C…130°C degree of protection: IP65 acc. to EN60529
FSG-NA2, FSG-NNN, FSG-NF2 FSK-NA2, FSK-NNN, FSK-NF2
156
FSK-NF2, FSG-NF2: Div. 2 FM-C operating temperature: process: -40°C…170°C
174
marking: ambient: -40°C…170°C
NI/Cl. I, II, III / Div. 2 / explosion hazard zone: FLG-NA2, FLH-NA2, FLK-NA2:
Gp A-G / T4 Ta = 125°C ATEX zone 2, marking:
II3G T6…T4 Ta -40°C…130°C
Shear wave transducers FSK-NA1, FSG-NA1
47
47
(corr. to type CDK1N31, CDG1N31) FLG-NA2, FLH-NA2, FLK-NA2:
rated (possible) FSK: (100) 200…3600mm Div. 2 FM-C, marking:
diameter range1: FSG: (400) 500…6500mm NI/Cl. I, II, III / Div. 2 /
dimensions: FSK-NA1: (126.5 x 50 x 53.5)mm Gp A-G / T4 Ta = 125°C
129.5 126.5
FSG-NA1: (129.5 x 50 x 64)mm Lamb wave transducers FLG-NA1, FLH-NA1, Ø
material: PEEK with stainless steel cap 70 9
FLK-NA1 (corr. to type CRG1N33, CRH1N33, CRK1N33)
degree of protection: IP65 acc. to EN60529 2
rated (possible) range FLG: 119 163.5
operating temperature: process: -40°C…130°C of application1: rated diameters (400) 500 to 6500 mm
66.5
56
ambient: -40°C…130°C for wall thickness (14 to 27) mm
use in explosive atmosphere FLH:
hazard zone: ATEX zone 1 and 2 rated diameters (250) 330 to 5000 mm
marking: 0044; II2G for wall thickness (9 to 18) mm
EEx q II T6…T3 Ta -40°C…180°C FLK: Technical Data
certification: IBExU 04 ATEX 1011 X rated diameters (100) 200 to 3600 mm FSK-NA1 FSG-NA1
type of protection: powder filling for wall thickness (5 to 11) mm Junction box JBXX
dimensions: (128.5 x 50 x 67.5)mm dimensions: see above
material: PPSU with stainless steel cap possible materials: JB01S4E3M, JB01S4E3P: stainless steel SS316 1.4404
degree of protection: IP65 acc. to EN60529 all others: stainless steel SS304 1.4301
50
50
operating temperature: process: -40°C…170°C gasket: - zone 1: silicone
ambient: -40°C…170°C - zone 2: EPDM
transducers FLG-NA1, FLH-NA1, FLK-NA1: degree of protection: IP67 acc. to EN60529
126.5 129.5 operating temperature: -40°C to 80°C
explosion hazard zone: ATEX zone 1 and 2
marking: 0044; II2G Additionally for the types JB01S4E3M, JB01S4E3P
EEx q II T6…T3 Ta -40°C…140°C explosion hazard zone: zone 1 (ATEX)
certification: IBExU 04 ATEX 1011 X marking: 0044; II2G
64
type of protection: powder filling 53.5 EEx e mb II T4...T6 Ta -40°C...80°C
certification: IBExU 01 ATEX 1161
type of protection: junction box: increased safety
decoupled network: encapsulation
1 The range in parentheses is possible under good conditions, but is not specified.
Subject to change without notification. GSCLAMPLIQUIDV2-2XEN Subject to change without notification. GSCLAMPLIQUIDV2-2YEN Subject to change without notification. GSCLAMPLIQUIDV2-2XEN
FLUXUS® is a protected trademark of FLEXIM GmbH. page 2 of 6 FLUXUS® is a protected trademark of FLEXIM GmbH. page 3 of 6 FLUXUS® is a protected trademark of FLEXIM GmbH. page 4 of 6
Technical Data (continued) Dimensions of the Transducers (in mm)
FSM-NA2, FSM-NF2, FSM-NNN, FSM-EA2, FSM-ENN FSQ-NA2, FSQ-NNN, FSQ-EA2, FSQ-ENN FLG-NA1, FLH-NA1, FLK-NA1 FLG-NA2, FLG-NF2, FLG-NNN, FLH-NA2, FLH-NF2,
Shear wave transducers FSM-NA1, FSQ-NA1 Shear wave transducers FSS-NF2, FSS-NNN
(corr. to type CDM1N31, CDQ1N31) (corr. to type CDS1NZ7/CDS1N52, CDS1N52, CDS1NZ7/CDS1N52) FLH-NNN, FLK-NA2, FLK-NF2, FLK-NNN
rated (possible) FSM: (50) 100…2500mm rated (possible)
diameter range1: FSQ: (10) 25…400mm diameter range1: (6) 10…70mm
30
18
dimensions: (79.5 x 30 x 39)mm dimensions: (26 x 13 x 17)mm
50
material: housing: stainless steel material: housing: stainless steel
47
contact surface: PEEK contact surface: Polyetherimid 60 ###1N33
degree of protection: IP65 acc. to EN60529 degree of protection: IP65 acc. to EN60529
21.5
operating temperature: process: -40°C...130°C operating temperature: process: -40°C...130°C
ambient: -40°C...130°C ambient: -40°C...130°C 128.5
42.5
33.5
use in explosive atmosphere explosion hazard zone: FSS-NF2: Div. 2 FM-C, marking:
- hazard zone: zone 1 and 2 NI/Cl. I, II, III / Div. 2 /
- marking: 0044; II2G Gp A-G / T4 Ta = 125°C
67.5
70
EEx m II T6…T4 Ta -20°C…120°C Lamb wave transducers FLG-NA2, FLG-NF2,
- certification: IBExU 98 ATEX 1012 X FLG-NNN, FLH-NA2, FLH-NF2, FLH-NNN, FSM-NA1, FSQ-NA1 FSS-NNN FSS-NF2 mit
- type of protection: encapsulation FLK-NA2, FLK-NF2, FLK-NNN Anschlusssystem NL
(corr. to type CRG1NC3, CRG1N52, CRG1N52/CRG1NC3,
128.5
Shear wave transducers FSK-NA2, FSK-NF2, FSK-NNN, CRH1NC3, CRH1N52, CRH1N52/CRH1NC3,
FSG-NA2, FSG-NF2, FSG-NNN CRK1NC3, CRK1N52, CRK1N52/CRK1NC3)
(corr. to type CDK1NZ7/CDK1N52, CDK1N52, CDK1NZ7/CDK1N52, rated (possible) range FLG:
CDG1NZ7/CDG1N52, CDG1N52, CDG1NZ7/CDG1N52)
of application1: Junction Box JB0X
30
rated diameters
rated (possible) FSK: (100) 200…3600mm (400) 500 to 6500 mm for wall
diameter range1: FSG: (400) 500… 6500mm thickness (14 to 27) mm Dimensions (in mm)
dimensions: FSK: (126.5 x 47 x 56)mm
24.50
FLH:
13.2
FSG: (129.5 x 47 x 66.5)mm rated diameters (250) 330 to 5000 mm
38.9
30.10
material: PEEK with stainless steel cap for wall thickness (9 to 18) mm
degree of protection: IP65 acc. to EN60529
15
FLK:
operating temperature: process: -40°C…130°C rated diameters (100) 200 to 3600 mm
60 26 32.70
ambient: -40°C…130°C for wall thickness (5 to 11) mm
explosion hazard zone: FSK-NA2, FSG-NA2: ATEX zone 2 dimensions: (128.5 x 50 x 67.5)mm 79.5
marking: material: PPSU with stainless steel cap
II3G T6…T4 Ta -40°C…130°C degree of protection: IP65 acc. to EN60529
FSG-NA2, FSG-NNN, FSG-NF2 FSK-NA2, FSK-NNN, FSK-NF2
156
FSK-NF2, FSG-NF2: Div. 2 FM-C operating temperature: process: -40°C…170°C
174
marking: ambient: -40°C…170°C
NI/Cl. I, II, III / Div. 2 / explosion hazard zone: FLG-NA2, FLH-NA2, FLK-NA2:
Gp A-G / T4 Ta = 125°C ATEX zone 2, marking:
II3G T6…T4 Ta -40°C…130°C
Shear wave transducers FSK-NA1, FSG-NA1
47
47
(corr. to type CDK1N31, CDG1N31) FLG-NA2, FLH-NA2, FLK-NA2:
rated (possible) FSK: (100) 200…3600mm Div. 2 FM-C, marking:
diameter range1: FSG: (400) 500…6500mm NI/Cl. I, II, III / Div. 2 /
dimensions: FSK-NA1: (126.5 x 50 x 53.5)mm Gp A-G / T4 Ta = 125°C
129.5 126.5
FSG-NA1: (129.5 x 50 x 64)mm Lamb wave transducers FLG-NA1, FLH-NA1, Ø
material: PEEK with stainless steel cap 70 9
FLK-NA1 (corr. to type CRG1N33, CRH1N33, CRK1N33)
degree of protection: IP65 acc. to EN60529 2
rated (possible) range FLG: 119 163.5
operating temperature: process: -40°C…130°C of application1: rated diameters (400) 500 to 6500 mm
66.5
56
ambient: -40°C…130°C for wall thickness (14 to 27) mm
use in explosive atmosphere FLH:
hazard zone: ATEX zone 1 and 2 rated diameters (250) 330 to 5000 mm
marking: 0044; II2G for wall thickness (9 to 18) mm
EEx q II T6…T3 Ta -40°C…180°C FLK: Technical Data
certification: IBExU 04 ATEX 1011 X rated diameters (100) 200 to 3600 mm FSK-NA1 FSG-NA1
type of protection: powder filling for wall thickness (5 to 11) mm Junction box JBXX
dimensions: (128.5 x 50 x 67.5)mm dimensions: see above
material: PPSU with stainless steel cap possible materials: JB01S4E3M, JB01S4E3P: stainless steel SS316 1.4404
degree of protection: IP65 acc. to EN60529 all others: stainless steel SS304 1.4301
50
50
operating temperature: process: -40°C…170°C gasket: silicone
ambient: -40°C…170°C
transducers FLG-NA1, FLH-NA1, FLK-NA1: degree of protection: IP67 acc. to EN60529
126.5 129.5 operating temperature: -40°C to 80°C
explosion hazard zone: ATEX zone 1 and 2
marking: 0044; II2G Additionally for the types JB01S4E3M, JB01S4E3P
EEx q II T6…T3 Ta -40°C…140°C explosion hazard zone: zone 1 (ATEX)
certification: IBExU 04 ATEX 1011 X marking: 0044; II2G
64
type of protection: powder filling 53.5 Ex e mb II T4...T6 Ta -40°C...80°C
certification: IBExU 01 ATEX 1161
type of protection: junction box: increased safety
decoupled network: encapsulation
1 The range in parentheses is possible under good conditions, but is not specified.
Subject to change without notification. GSCLAMPLIQUIDV2-2XEN Subject to change without notification. GSCLAMPLIQUIDV2-2XEN Subject to change without notification. GSCLAMPLIQUIDV2-2YEN
FLUXUS® is a protected trademark of FLEXIM GmbH. page 2 of 6 FLUXUS® is a protected trademark of FLEXIM GmbH. page 3 of 6 FLUXUS® is a protected trademark of FLEXIM GmbH. page 4 of 6
Connection Systems Product Code of Shear Wave Transducers
Data Sheet
System NL description Clamp-on Flow Transducers for Liquids
temperature
connection
transducer
frequency
extension
Ex certifi-
system
cation
cable
type
FS set of ultrasonic flow transducers for liquids, shear wave Non-Invasive Flow Measurement of Liquids with FLUXUS® F
S- 8 MHz, for diameter range see Technical Data
Q- 4 MHz, for diameter range see Technical Data
Clamp-on flow transducers are mounted onto the pipe
System AS M- 1 MHz, for diameter range see Technical Data
instead of inserted into it. Since they do not come in contact
K- 0.5 MHz, for diameter range see Technical Data
with the medium, they are not subjected to wear. Clamp-on
G- 0.2 MHz, for diameter range see Technical Data flow measurement is applicable even for chemically
N normal temperature range, max. 130 °C aggressive, corrosive or ultra pure substances. The retrofit-
E increased temperature range, max. 200 °C ting of existing installations can be done without cutting into
NN no explosion protection the pipe and without interruption of the process.
A2 explosion protection certification for ATEX zone 2 All FLEXIM clamp-on transducers are watertight and espe-
F2 explosion protection certification for FM Class I Div. 2 (TS only) cially made for use in harsh industrial environments. The
System TS transducer housings and the transducer cable armored
A1 explosion protection certification for ATEX zone 1 (TS only)
direct connection: direct connection of explo- connection via junction box: connection via junction box of explo- sheaths are made of stainless steel. The cables of the
NL- connection system NL with Lemo connector explosion-proof transducers are sheathed with a special
sion-proof transducers: sion-proof transducers:
AS- connection system AS with Amphenol connector plastic material. The robust construction and the integrated
FLUXUS FLUXUS TS- connection system TS: direct connection or connection via junction box transducer cables guarantee good measurement results
XXX length of extension cable: the max. length depends on the transducer over a long period of extensive usage.
FSM-NA2 transducers in the Variofix mounting fixture
frequency (see table on page 5)
connection system TS: a junction box (JB02 for ATEX zone 2 or JB01 Features
Example: for ATEX zone 1) is included.
• Flow calibrated transducers
FS G- N A1 TS- 030 shear wave transducer 0.2 MHz normal temperature for ATEX zone 1,
connection system TS with junction box JB01 with extension cable 30 m • Automatic transducer detection
• Watertight stainless steel housing
• No contact with the medium, no corrosion,
hygienic measurement, suitable for ultra pure liquids
• Low stocking costs since only 2 types of transducers are
X Product Code of Lamb Wave Transducers needed to cover the main pipe diameter range
• Measurement is independent of fluid conductivity and
description pressure
temperature
connection
transducer
frequency
extension
Ex certifi-
X
• No pressure loss, no risk of leakage
system
FSQ-NA2 transducers fixed by chains
cation
cable
L L
type
X
X Technical Data
FL set of ultrasonic flow transducers for liquids, Lamb wave
K- 0.5 MHz, for diameter range see Technical Data
H- 0.3 MHz, for diameter range see Technical Data Shear wave transducers type FSQ-NA2, FSQ-NF2, Shear wave transducers FSM-NA2, FSM-NF2,
FSQ-NNN, FSQ-EA2, FSQ-ENN (corr. to type CDQ1NZ7/CDQ1N52, FSM-NNN, FSM-EA2, FSM-ENN (corr. to type CDM1NZ7/
G- 0.2 MHz, for diameter range see Technical Data
FLUXUS FLUXUS CDQ1N52, CDQ1NZ7/CDQ1N52, CDQ1EZ7, CDQ1EZ7) CDM1N52, CDM1N52, CDM1NZ7/CDM1N52, CDM1EZ7, CDM1EZ)
N normal temperature range
rated (possible) rated (possible)
NN no explosion protection diameter range1: (10) 25…400mm diameter range1: (50) 100…2500mm
Transducer cable lengths X and Y and maximum length L of the extension cable (in m) A2 explosion protection certification for ATEX zone 2 dimensions: (42.5 x 18 x 21.5) mm dimensions: (60 x 30 x 33.5)mm
F2 explosion protection certification for FM Class I Div. 2 (TS only) material: housing: stainless steel material: housing: stainless steel
transducer type connection type NL connection type AS connection type TS contact surface PEEK (FSQ-N**) contact surface PEEK (FSM-N**)
A1 explosion protection certification for ATEX zone 1 (TS only)
-*A1 - - Q: X=3m; L=90m or Polyimid (FSQ-E**) or Polyimid (FSM-E**)
M: X=4m; L=300m NL- connection system NL with Lemo connector degree of protection: IP65 acc. to EN60529 degree of protection: IP65 acc. to EN60529
K,H,G: X=5m; L=300m AS- connection system AS with Amphenol connector operating temperature: process and ambient: FSM-N**: IP68 on request
-*A2 und -*NN S: X=1m; Y=1m; L=20m S: X=1m; Y=1m; L=20m S: X = 2m; L = 40m TS- connection system TS: direct connection or connection via junction box FSQ-N**: -40°C…130°C; operating temperature: process and ambient:
Q: X=2m; Y=1m; L=50m Q: X=2 m; Y=1m; L=50m Q: X=3m; L=90m FSQ-E**: -40°C…200°C FSM-N**: -40°C…130°C;
XXX length of extension cable: the max. length depends on the transducer
M: X =2m; Y=2m; L=100m M: X =2m; Y=2m; L=100m M: X=4m; L=300m explosion hazard zone: FSQ-NA2:ATEX zone 2, marking: FSM-E**: -40°C…200°C
frequency (see table on page 5)
K,H,G: X =2m; Y=3m; L=100m K,H,G: X =2m; Y=3m; L=100m K,H,G: X=5m; L=300m II3G T6…T4 Ta -30°C…130°C explosion hazard zone: FSM-NA2:ATEX zone 2, marking:
connection system TS: a junction box (JB02 for ATEX zone 2 or JB01
-*F2 - - S: X = 2m; L = 40m FSQ-EA2:ATEX zone 2, marking: II3G T6…T4 Ta -30°C…130°C
Example: for ATEX zone 1) is included.
Q: X=3m; L=90m II3G T6…T4 Ta -30°C…200°C FSM-EA2:ATEX zone 2, marking:
M: X=4m; L=300m FL H- N A1 TS- 030 lamb wave transducer 0.3 MHz normal temperature for ATEX zone 1,
FSQ-NF2: Div. 2 FM-C, marking: II3G T6…T4 Ta -30°C…200°C
K,H,G: X=5m; L=300m connection system TS with junction box JB01 with extension cable 30 m
NI/Cl. I, II, III / Div. 2 / FSM-NF2: Div. 2 FM-C, marking:
Gp A-G / T4 Ta = 125°C NI/Cl. I, II, III / Div. 2 /
Gp A-G / T4 Ta = 125°C
FLEXIM GmbH
Wolfener Str. 36
D-12681 Berlin
Subject to change without notification. GSCLAMPLIQUIDV2-2YEN internet: www.flexim.de Fon:+49 (30) 93 66 76 60 Subject to change without notification. GSCLAMPLIQUIDV2-2XEN
FLUXUS® is a protected trademark of FLEXIM GmbH. page 5 of 6 e-mail: [email protected] Fax:+49 (30) 93 66 76 80 FLUXUS® is a protected trademark of FLEXIM GmbH. page 1 of 6
Connection Systems Product Code of Shear Wave Transducers
Data Sheet
System NL description Clamp-on Flow Transducers for Liquids
temperature
connection
transducer
frequency
extension
Ex certifi-
system
cation
cable
type
FS set of ultrasonic flow transducers for liquids, shear wave Non-Invasive Flow Measurement of Liquids with FLUXUS® F
S- 8 MHz, for diameter range see Technical Data
Q- 4 MHz, for diameter range see Technical Data
Clamp-on flow transducers are mounted onto the pipe
System AS M- 1 MHz, for diameter range see Technical Data
instead of inserted into it. Since they do not come in contact
K- 0.5 MHz, for diameter range see Technical Data
with the medium, they are not subjected to wear. Clamp-on
G- 0.2 MHz, for diameter range see Technical Data flow measurement is applicable even for chemically
N normal temperature range, max. 130 °C aggressive, corrosive or ultra pure substances. The retrofit-
E increased temperature range, max. 200 °C ting of existing installations can be done without cutting into
NN no explosion protection the pipe and without interruption of the process.
A2 explosion protection certification for ATEX zone 2 All FLEXIM clamp-on transducers are watertight and espe-
F2 explosion protection certification for FM Class I Div. 2 (TS only) cially made for use in harsh industrial environments. The
System TS transducer housings and the transducer cable armored
A1 explosion protection certification for ATEX zone 1 (TS only)
direct connection: direct connection of explo- connection via junction box: connection via junction box of explo- sheaths are made of stainless steel. The cables of the
NL- connection system NL with Lemo connector explosion-proof transducers are sheathed with a special
sion-proof transducers: sion-proof transducers:
AS- connection system AS with Amphenol connector plastic material. The robust construction and the integrated
FLUXUS FLUXUS TS- connection system TS: direct connection or connection via junction box transducer cables guarantee good measurement results
XXX length of extension cable: the max. length depends on the transducer over a long period of extensive usage.
FSM-NA2 transducers in the Variofix mounting fixture
frequency (see table on page 5)
connection system TS: a junction box (JB02 for ATEX zone 2 or JB01 Features
Example: for ATEX zone 1) is included.
• Flow calibrated transducers
FS G- N A1 TS- 030 shear wave transducer 0.2 MHz normal temperature for ATEX zone 1,
connection system TS with junction box JB01 with extension cable 30 m • Automatic transducer detection
• Watertight stainless steel housing
• No contact with the medium, no corrosion,
hygienic measurement, suitable for ultra pure liquids
• Low stocking costs since only 2 types of transducers are
X Product Code of Lamb Wave Transducers needed to cover the main pipe diameter range
• Measurement is independent of fluid conductivity and
description pressure
temperature
connection
transducer
frequency
extension
Ex certifi-
X
• No pressure loss, no risk of leakage
system
FSQ-NA2 transducers fixed by chains
cation
cable
L L
type
X
X Technical Data
FL set of ultrasonic flow transducers for liquids, Lamb wave
K- 0.5 MHz, for diameter range see Technical Data
H- 0.3 MHz, for diameter range see Technical Data Shear wave transducers type FSQ-NA2, FSQ-NF2, Shear wave transducers FSM-NA2, FSM-NF2,
FSQ-NNN, FSQ-EA2, FSQ-ENN (corr. to type CDQ1NZ7/CDQ1N52, FSM-NNN, FSM-EA2, FSM-ENN (corr. to type CDM1NZ7/
G- 0.2 MHz, for diameter range see Technical Data
FLUXUS FLUXUS CDQ1N52, CDQ1NZ7/CDQ1N52, CDQ1EZ7, CDQ1EZ7) CDM1N52, CDM1N52, CDM1NZ7/CDM1N52, CDM1EZ7, CDM1EZ)
N normal temperature range
rated (possible) rated (possible)
NN no explosion protection diameter range1: (10) 25…400mm diameter range1: (50) 100…2500mm
Transducer cable lengths X and Y and maximum length L of the extension cable (in m) A2 explosion protection certification for ATEX zone 2 dimensions: (42.5 x 18 x 21.5) mm dimensions: (60 x 30 x 33.5)mm
F2 explosion protection certification for FM Class I Div. 2 (TS only) material: housing: stainless steel material: housing: stainless steel
transducer type connection type NL connection type AS connection type TS contact surface PEEK (FSQ-N**) contact surface PEEK (FSM-N**)
A1 explosion protection certification for ATEX zone 1 (TS only)
-*A1 - - Q: X=3m; L=90m or Polyimid (FSQ-E**) or Polyimid (FSM-E**)
M: X=4m; L=300m NL- connection system NL with Lemo connector degree of protection: IP65 acc. to EN60529 degree of protection: IP65 acc. to EN60529
K,H,G: X=5m; L=300m AS- connection system AS with Amphenol connector operating temperature: process and ambient: FSM-N**: IP68 on request
-*A2 und -*NN S: X=1m; Y=1m; L=20m S: X=1m; Y=1m; L=20m S: X = 2m; L = 40m TS- connection system TS: direct connection or connection via junction box FSQ-N**: -40°C…130°C; operating temperature: process and ambient:
Q: X=2m; Y=1m; L=50m Q: X=2 m; Y=1m; L=50m Q: X=3m; L=90m FSQ-E**: -40°C…200°C FSM-N**: -40°C…130°C;
XXX length of extension cable: the max. length depends on the transducer
M: X =2m; Y=2m; L=100m M: X =2m; Y=2m; L=100m M: X=4m; L=300m explosion hazard zone: FSQ-NA2:ATEX zone 2, marking: FSM-E**: -40°C…200°C
frequency (see table on page 5)
K,H,G: X =2m; Y=3m; L=100m K,H,G: X =2m; Y=3m; L=100m K,H,G: X=5m; L=300m II3G T6…T4 Ta -30°C…130°C explosion hazard zone: FSM-NA2:ATEX zone 2, marking:
connection system TS: a junction box (JB02 for ATEX zone 2 or JB01
-*F2 - - S: X = 2m; L = 40m FSQ-EA2:ATEX zone 2, marking: II3G T6…T4 Ta -30°C…130°C
Example: for ATEX zone 1) is included.
Q: X=3m; L=90m II3G T6…T4 Ta -30°C…200°C FSM-EA2:ATEX zone 2, marking:
M: X=4m; L=300m FL H- N A1 TS- 030 lamb wave transducer 0.3 MHz normal temperature for ATEX zone 1,
FSQ-NF2: Div. 2 FM-C, marking: II3G T6…T4 Ta -30°C…200°C
K,H,G: X=5m; L=300m connection system TS with junction box JB01 with extension cable 30 m
NI/Cl. I, II, III / Div. 2 / FSM-NF2: Div. 2 FM-C, marking:
Gp A-G / T4 Ta = 125°C NI/Cl. I, II, III / Div. 2 /
Gp A-G / T4 Ta = 125°C
FLEXIM GmbH
Wolfener Str. 36
D-12681 Berlin
Subject to change without notification. GSCLAMPLIQUIDV2-2XEN internet: www.flexim.de Fon:+49 (30) 93 66 76 60 Subject to change without notification. GSCLAMPLIQUIDV2-2XEN
FLUXUS® is a protected trademark of FLEXIM GmbH. page 5 of 6 e-mail: [email protected] Fax:+49 (30) 93 66 76 80 FLUXUS® is a protected trademark of FLEXIM GmbH. page 1 of 6