Proveo 8: User Manual
Proveo 8: User Manual
PROVEO 8
User Manual
10 733 910 – Version 02
4 Design 12
4.1 F42 floor stand 12
4.2 CT42 telescope mount 13
4.3 PROVEO 8 Optics carrier modules 14
5 Functions 15
5.1 Balancing system 15
5.2 Brakes 15
5.3 Illumination 17
5.4 Fusion Optics 17
6 Controls 18
6.1 PROVEO 8 optics carrier including focus, tilt and XY
with IVC 18
6.2 PROVEO 8 optics carrier including focus, tilt and XY
with IVA 19
6.3 Surgeon panel 20
6.4 Balancing 20
6.5 Control unit 20
6.6 F42 stand 20
Legal disclaimer 6.7 F42 Terminals 21
All specifications are subject to change without notice. 6.8 CT42 Terminals on the rack controller 22
The information provided by this manual is directly related to the operation of the 6.9 CT42 horizontal arm 23
equipment. Medical decision remains the responsibility of the clinician. 6.10 Footswitch and handles 23
Leica Microsystems has made every effort to provide a complete and clear user
manual highlighting the key areas of product use. Should additional information 7 Preparation before surgery 26
regarding the use of the product be required, please contact your local Leica 7.1 Transportation (F42) 26
representative. 7.2 Installing a binocular tube 27
You should never use a medical product of Leica Microsystems without the full 7.3 Setting the binocular tube 28
understanding of the use and the performance of the product. 7.4 Adjusting the eyepiece 29
7.5 Change the 0° assistant to another position 29
Liability 7.6 Changing the camera 30
For our liability, please see our standard sales terms and conditions. Nothing in this
7.7 Balancing and locking the parallelogram 32
disclaimer will limit any of our liabilities in any way that is not permitted under
applicable law, or exclude any of our liabilities that may not be excluded under 7.8 Positioning on the operating table 33
applicable law.
10 Menu 57
10.1 Menu – "Maintenance" 57
10.2 Menu – "How to..." 58
10.3 Menu – "Service" 58
11 Accessories 59
11.1 Devices and accessories manufactured by Leica
Microsystems (Schweiz) AG 59
11.2 Accessories list for advanced configurations 60
11.3 Drapes 60
13 Disposal 65
14 What to do if ...? 65
14.1 Malfunctions 65
14.2 Malfunctions documentation accessories 66
14.2 Error messages on the control unit 66
15 Specifications 68
15.1 PROVEO 8 68
15.2 F42 floor stand 69
15.3 CT42 telescope mount 69
15.4 CT42 rack controller 70
15.5 Ambient conditions 70
15.6 Standards fulfilled 70
15.7 Limitations of use (only F42) 70
15.8 Dimensional drawings 72
The symbols used in this user manual have the following meaning:
3 Safety notes X When instructing new users, do so thoroughly and explain the
meanings of the warning signs and messages.
The PROVEO 8 surgical microscope is state-of-the-art technology. X Allocate responsibilities for commissioning, operation and
Nevertheless, hazards can arise during operation. maintenance. Monitor compliance with this.
X Always follow the instructions in this user manual, and in X The PROVEO 8 is intended for professional use only.
particular the safety notes.
WARNING
3.1 Intended use
Risk of electric shock!
• The PROVEO 8 surgical microscope is an optical instrument for X Connect this equipment only to a supply mains with
improving the visibility of objects through magnification and protective earth.
illumination. It can be applied for observation and
documentation and for human and veterinary medical X Only use the PROVEO 8 surgical microscope if it is free of
treatment. defects.
• The major field of use is Ophtalmology. X Inform your Leica representative or Leica Microsystems
• The microscope device is not for use in microsurgery (Neuro,
(Schweiz) AG, Medical Division, 9435 Heerbrugg, Switzerland,
Plastic/Reconstructive, Ear, Nose and Throat surgeries). immediately about any product defect that could potentially
• The PROVEO 8 surgical microscope may be used only in closed
cause injury or harm.
rooms and must be placed on a solid floor or mounted to the X If you use accessories from other manufacturers with the
ceiling. PROVEO 8 surgical microscope, make sure that these
• It is available on the floor stand F42 and on the telescope mount
manufacturers confirm that the combination is safe to use.
CT42. Follow the instructions in the user manual for those accessories.
The F42 is a floor stand for three-dimensional positioning of the • Modifications, installations to or service on the PROVEO 8
PROVEO 8 in the room. surgical microscope may be carried out only by technicians who
The CT42 is a ceiling mounted electrical linear actuator for are explicitly authorized by Leica to do so.
lifting and lowering the PROVEO 8. It is operated by a hand For further details refer to the installation manual.
switch. • Only original Leica replacement parts may be used in servicing
• The PROVEO 8 surgical microscope is subject to special
the product.
precautionary measures for electromagnetic compatibility. It • After service work or technical modifications, the device must
must be installed and commissioned in accordance with the be readjusted in accordance with our technical specifications.
guidelines and manufacturer's declarations and recommended • If the instrument is modified or serviced by unauthorized
safety distances (according to EMC tables based on persons, is improperly maintained (as long as maintenance was
EN60601‑1‑2). not carried out by a qualified trained service engineer), or is
• Portable and mobile as well as stationary RF communications
handled improperly, Leica Microsystems will not accept any
equipment can have a negative effect on the reliability of the liability.
PROVEO 8 surgical microscope's functionality. • The effect of the surgical microscope on other instruments has
• Always release the brakes to move or relocate the PROVEO 8
been tested as specified in EN 60601-1-2. The system passed the
surgical microscope. emission and immunity test. Comply with the usual pre
• The essential performance of the PROVEO 8 is to provide
cautionary and safety measures relating to electromagnetic and
illumination and mechanical stability of the optics carrier in any other forms of radiation.
position. • The electric installation in the building must conform to the
national standard, e.g., current-operated ground leakage
3.2 Directions for the person protection (fault-current protection) is suggested.
responsible for the instrument • Like any other instrument in the operating theater, this system
X Ensure that the PROVEO 8 surgical microscope is used only by may fail. Leica Microsystems (Schweiz) AG therefore
persons qualified to do so. recommends that a backup system be kept available during the
X Ensure that this user manual is always available at the place operation.
where the PROVEO 8 surgical microscope is in use.
X Carry out regular inspections to make certain that the
authorized users are adhering to safety requirements.
Coaxial Red Reflex illumination Decisions about brightness must be made case by case. In any
event, the surgeon must evaluate the risks and benefits of the used
Light setting Recommended maximum exposure time according to 1) [min.]
light intensity. Despite all efforts to minimize the risk of retinal
25% 10
injury by surgical microscopes, damage may still occur.
50 % 4.5
Photochemical retinal damage is a possible complication of the
75% 3 necessity to use bright light to make eye structures visible during
100 % 2 difficult ophthalmological processes.
Retina Protection
14
Function Activated
In addition, the Retina Protection function can be activated during
the surgery to reduce the main light intensity below 10% and
RedReflex below 20%.
If the integrated slit illumination motor fails, the slit • when transporting the microscope
illumination can be manually activated/deactivated using • when re-equipping
the rotary knob (2).
WARNING CAUTION
Risk of injury from parts falling down! Damage of the touch panel!
X Before the operation, make sure that the optical X Operate the touch panel using your fingers only.
components and accessories are sufficiently secured and Never use hard, sharp or pointed objects made out of
cannot move. wood, metal or plastic.
X Never clean the touch panel using cleaners that contain
WARNING abrasive substances. These substances can scratch the
Risk of infection! surface and cause it to be become dull.
X Always use the PROVEO 8 surgical with sterile controls
and knobs. CAUTION
Surgical microscope can move without warning!
WARNING X Always lock the footbrake when you are not moving the
Danger of fatal electrical shock! system.
X Operate the system only with all equipment in its proper
position (all covers fitted, doors closed).
CAUTION
Damage to the PROVEO 8 surgical microscope due to
WARNING uncontrolled tilting!
Danger of fatal electrical shock! X Hold the handle when releasing the brake.
X The PROVEO 8 surgical microscope may be connected to
a grounded socket only (F42).
CAUTION
Damage to the PROVEO 8 surgical microscope during
WARNING transportation!
Motors return to their park position! X Never move the stand in the extended condition.
X Before switching on the microscope, ensure that the X Never roll over cables lying on the floor.
travel paths of XY-unit and focus are free of X Do not transport or store the system in areas with an
obstructions. elevation angle bigger than 10°.
WARNING CAUTION
Light that is too intense can damage the retina! Damage to the housing of the stand or to the touch panel of
X Observe the warning messages in the chapter on "Safety the control unit!
notes". If the optics carrier is moved into the transport position or
from the transport position to the operating position:
X Ensure that the transport lock must be locked.
WARNING
Danger to the patient due to failure of the magnification
motor! CAUTION
If the magnification motor fails, the magnification can be There is a risk of damage to the surgical microscope from
manually adjusted using the rotary knob. uncontrolled tilting!
X Firmly hold the handles before triggering the "All
CAUTION
Risk of infection!
X Leave sufficient space around the stand to ensure that
the sterile drape does not come into contact with non-
sterile components.
CAUTION
Risk of overheating of the system!
Covering the air inlet can result in a controlled shutdown of
the system due to overheating.
X Make sure that there is always some space between the
air inlet and the drape.
CAUTION
Damage of the magnification motor!
X Only adjust the magnification manually if the
magnification motor is defective.
CAUTION
Danger to the patient due to changes in the user settings!
X Never change the configuration settings or edit the user
list during an operation.
CAUTION
Risk of injury!
X Pay special attention to the required safety distances if
you use the Combination Mode function together with
accessories from other manufacturers that can reduce
the working distance to less than 140 mm (non-contact
wide-angle observation systems), since focus together
with Combination Mode is a semi-automated function.
CAUTION
Danger of collision!
The surgical microscope can collide with surrounding
components, the ceiling or lamps.
X Check the danger area before moving the swing arm.
X Carefully move the ceiling mount upwards, and observe
ceiling and lamps.
13
11
12
13
13
14
10
11 12
13
9 8 8
2 15 3 4 5 4 6 6
MD Medical device
2 Fabrication label
7 System weight label
8 Only to be operated
Trained personnel
by trained personnel
Utilisation réservée à
un personnel formé
Carrier
f MM = month (2 digits)
g TT = day (2 digits)
11 Open
h Production starting date
3 Follow the User Manual
12 Closed
La fiabilité de la mise à la
14 Monitor weight label
A
B
100 –240 V 50 - 60 Hz
500 VA
7 Equipotential
2 Fabrication label bonding
8 Max. load for optics
Max. 8.0 kg (17.64 lb)
carrier
9 Open
10 Closed
a Prefix number
b Leica system article no.
c Serial number 11 Crushing hazard label
d Incremental number starting at
1 for each batch
e JJ = year (2 digits)
f MM = month (2 digits) 12 Monitor weight label
g TT = day (2 digits) max. 16 kg (35.27 lb)
La fiabilité de la mise à la
terre n'est assurée que si 14 MODEL PROVEO 8 Type label
l'équipement est connecté à une STAND C42 / CT42
prise équivalente, marquée
"Hôpital seulement" ou "Qualité MODULE Swing Arm
hôpital".
15 UDI Label
5 MET label
Device Identifier (DI)
GS1 Data Matrix Code
Production Identifier (PI)
Serial number
Date of manufacturing
4 Design
4.1 F42 floor stand
1 PROVEO 8 optics carrier
2 Parallelogram
3 Monitor arm (optional)
4 Video monitor (optional)
5 CT42 telescope mount
6 Horizontal arm
7 Remote control telescope
8 Up
9 Down
7
8
5.2.2 Selected brakes at the F42 floor stand 5.2.3 Brakes at the CT42 telescope mount
With the function "Selected brakes" the user can release any
individual brake.
1 Rotation of tower
2 Rotation of parallelogram
3 Up/Down movement of parallelogram
4 Rotation of microscope carrier
5.3 Illumination
The illumination of the surgical microscope PROVEO 8 consists of
two LED modules and are located in the optics carrier.
There are two lamps, main lamp and Red Reflex lamp.
6 Controls
6.1 PROVEO 8 optics carrier including focus, tilt and XY with IVC
19
18 17
1 XY Coupling 11 0° assistant
2 Rotary knob for changing the 0° assistant position 12 Surgeon panel
3 Fine focus for external camera 13 2× CAN sockets - only for Leica accessories
4 Rotary knob "Magnification" (emergency operation only) 14 C-mount adapter
5 Rotary knob "Slit illumination" (emergency operation only) 15 Cooling slots
6 Slot for filter slide 16 Objective
7 Handle 17 Mounting threads for accessories
8 Rotary knob for Red Reflex illumination diameter 18 Socket for BIOM
9 PROVEO 8 optics carrier
10 Fine focus for assistant
1 Surgeon panel
2 Infrared receiver for remote control (Recorder) 5
3 Touchpanel
1
6.4 Balancing 4
2
3
WARNING
Risk of human errors due to loud disturbing noises during
surgery!
X Don't plug/unplug external devices at the "Line In" input
during surgery.
X Make sure that the volume level on audio device is set to
the minimum before connecting to the "Line In" input.
1 Mains switch
2 Mains fuses
3 Equipotential bonding socket
For connecting the PROVEO 8 to an equipotential bonding
device. This is part of the customer's building installation.
Observe the requirements of EN 60601-1 (§ 8.6.7).
Handles
Footswitches and handles can be assigned individually for Footswitches and handles can be assigned individually for
each user in the configuration menu. each user in the configuration menu.
OCT Up
All Lights On/Off VR Mode On/Off OCT Live Mode/Stop OCT Mode On/Off
OCT Down
Handles Handles
OCT Up
OCT Z – OCT Z+
OCT Down
Handles
7 Preparation before surgery X Ensure that the PROVEO 8 is in the transport position.
CAUTION
Surgical microscope can move without warning!
X Always lock the footbrake when you are not moving the
system.
CAUTION
Damage to the PROVEO 8 surgical microscope due to
In case the PROVEO 8 is not in transport position, refer to
uncontrolled tilting!
section 8.6.
X Hold the handle when releasing the brake.
CAUTION
Damage to the housing of the stand or to the touch panel of
the control unit!
If the optics carrier is moved into the transport position or
from the transport position to the operating position:
X Ensure that the transport lock must be locked.
WARNING
Risk of injury from parts falling down!
X Before the operation, make sure that the optical
components and accessories are sufficiently secured and
7.2 Installing a binocular tube cannot move.
WARNING
Risk of injury due to downward movement of the surgical
microscope!
X Complete all preparations and adjustments to the stand
before the operation.
X Never change the accessories or attempt to rebalance
the microscope while it is above the field of operation.
X Before changing accessories, always lock the
parallelogram (see chapter 7.7.2).
X Balance the PROVEO 8 after re-equipping it.
X Do not release the brakes when the instrument is in an
unbalanced state.
X Before re-equipping during the operation, first swing
the microscope away from the operating field.
X Make sure that the optical accessories are clean and free of dust
and dirt.
X Loosen clamping screw (1).
X Insert the accessories into the dovetail ring.
X Tighten the clamping screw (1).
1
1
A parfocally adjusted microscope ensures that assistant's X Rotate the eyecups up or down until the desired distance is set.
X When the desired position is reached, push the top down and
lock the knob (1) again.
7.6.2 Laserfilter for IVC X Insert the laserfilter IVC (4) into the laserfilter slot of the optics
The plane of filter is sharply imaged in the same plane as the object.
The GG420 UV protection filter is built-in. In addition, the "GG475
UV protection filter" and the "Protection Filter 5×" are available.
X Remove the filter cover.
X Push in the filter slide, inclined slightly upwards, until it
engages.
WARNING
Risk of injury due to downward movement of the surgical
microscope!
X Never change the accessories or attempt to rebalance
the microscope while it is above the field of operation.
X After re-equipping, always rebalance the microscope on
the parallelogram.
CAUTION
There is a risk of damage to the surgical microscope from
uncontrolled tilting!
X Firmly hold the handles before triggering the
X Release the parallelogram (see chapter 7.7.3).
"All Brakes" function.
X Hold the microscope by the handles.
X Turn one handle to release the brakes (All Brakes). X Grip and turn one handle to release the brakes.
X Check whether the microscope drifts up or down. X At the same time, pull the transport lock (1) and bring it into a
Microscope drifts downwards: vertical position.
X Turn balancing knob (2) clockwise.
Microscope drifts upwards:
X Turn balancing rotary knob (2) counterclockwise.
WARNING
Risk of injury due to downward movement of the surgical
microscope!
X Always lock the parallelogram:
• when transporting the microscope
• when re-equipping
CAUTION
Danger of damage!
X Before lifting the optics carrier make sure that the area
above the parallelogram is clear to avoid collisions with
OR lamps, ceiling etc.
Forwards
X Using the handrail, carefully push the surgical microscope to the
operating table and position it as desired.
• All positions are also possible as the mirror image
Backwards position.
• The instrument must be positioned such that the range
Unless they are individually configured for the current user, the of movement is large enough for the expected tasks.
brakes are released by turning the handles as follows:
X Turn backwards and hold: Selected brakes are released X Press the footbrake at the front end (2) (lock closed) until it
max. 135° S
angle
WARNING
Danger to the patient due to downward movement of the
CT42 telescope mount!
X Do not use the up/down movement when the
microscope is over the patient.
Raising and lowering the CT42 telescope mount 7.9 Attaching sterile controls and
The CT42 telescope mount can be raised and lowered electrically.
These functions can be controlled via buttons on the remote control
drape
unit. WARNING
Risk of infection!
Move telescopic arm to the desired height: X Always use the PROVEO 8 surgical with sterile controls
• "Up" key (1): Raise telescopic arm. and knobs.
• "Down" key (2): Lower telescopic arm.
CAUTION
Danger of collision!
The surgical microscope can collide with surrounding
components, the ceiling or lamps.
X Check the danger area before moving the swing arm.
X Attach steam-sterilizable covers to accessories as well (if
X Carefully move the ceiling mount upwards, and observe
present).
ceiling and lamps.
X Swing the microscope away.
X Remove the sterile components.
X Lock the swing arm.
X Switch off the power switch on the swing arm.
X Press the "Up" key on the remote control and raise the stand.
CAUTION
Risk of infection!
X Leave sufficient space around the stand to ensure that
the sterile drape does not come into contact with non-
sterile components.
X Activate the "All Brakes" function on the handle and extend the
parallelogram.
WARNING
Motors return to their park position!
X Before switching on the microscope, ensure that the
travel paths of XY-unit and focus are free of
obstructions.
WARNING
Check basic functions of the system:
Danger of fatal electrical shock! • Check the Main and Red Reflex illumination.
X The PROVEO 8 surgical microscope may be connected to
• Check the footswitch functions.
a grounded socket only (F42). • Check the rotary handles.
X Switch on the microscope using the power switch (1) on the • Check if the brakes work reliable.
stand. The main screen is displayed on the control unit.
The system is starting an initialization procedure.
Retina
Protect
The icons have the following meaning (only for information): X Grasp the optics carrier by both handles (1).
X Turn one handle to release the brakes (All Brakes).
1 Status of XY-Unit X Position the optics carrier and release the handle.
2 Status of OCT
3 Status of Inverter Also refer to the chapter 7.7.4.
4 IOL Mode/Retina Protect
5 Status of Recorder
CAUTION
6 Status of VR Mode
Damage to the PROVEO 8 surgical microscope due to
7 Value of main illumination
uncontrolled movement!
8 Value of Red Reflex illumination X Hold the handle when releasing the brake.
9 Value of magnification
Keys with touch function:
10 Adjusting the tilt angle
11 Reset button for focus (line shows position of focus)
8.4.2 Adjusting the Red Reflex illumination 8.4.5 Adjusting the magnification (zoom)
diameter You can adjust the magnification using a footswitch/handle or the
You can adjust the Red Reflex illumination diameter using the "Magnification" adjustment bar on the "Main" menu screen of the
rotary knob (1) or the footswitch/handle. control unit.
X Turn rotary knob (1) and set the Red Reflex illumination
On the touch panel monitor in the "Main" menu screen
diameter as required.
The user can activate the Retina Protection function during the The magnification changes.
surgery via the footswitch or the touch panel of the control unit.
When Retina Protection function is activated, the main light • Clicking the or button changes the
intensity is reduced to 10% and the Red Reflex intensity is reduced magnification value in increments of 1. Holding down
to 20%. The user can still adjust the light intensity below the the button with your finger changes the value in
threshold. When the user deactivates the Retina Protection increments of 5.
function, the light intensity will go back to previous intensity. • You can adjust the magnification motor speed in the
"Speed" menu.
• These values can be saved individually for each user (see
chapter 9.3.3).
CAUTION
Damage to the magnification motor!
X Only adjust the magnification manually if the
magnification motor is defective.
WARNING
Danger to the patient due to failure of the integrated slit
illumination!
If the integrated slit illumination motor fails, the slit
illumination can be manually activated/deactivated using
the rotary knob (2).
NOTE
X Make sure that the video monitor does not collide with the
parallelogram of the stand.
X If the Enfocus OCT scanhead is mounted below the optics carrier,
rotate the optics carrier by 90 degree to avoid colliding the
scanhead with the microscope front panel.
panel function
CAUTION
The quick access line is displayed at the top of each screen to enable
Damage of the touch panel! access to the following menus at any time.
X Operate the touch panel using your fingers only.
1 Main menu
Never use hard, sharp or pointed objects made out of 2 Speed menu (see chapter 9.3.3)
wood, metal or plastic. 3 Menu (see chapter 9.3)
X Never clean the touch panel using cleaners that contain
4 DIC menu
abrasive substances. These substances can scratch the 5 Warning messages
surface and cause it to be become dull. 6 Change Monitor Display View (activated when OCT is enabled in
the accessory list)
9.1 Menu structure 7 Display recording in progress (green blinking), no recording (no
color)
1
9.2 Selecting users
In the "Main" and "Speed" menu screens, the three buttons
2
"Presets", "User List" and "Show Settings" appear in the dynamic
button bar at all times.
9.2.1 Presets
You can find a list of default users preset by Leica for the most
common types of operation under "Presets".
Click the "1-15" or "16-30" button to switch between screens.
X Select a user.
The "Select" button is shown.
X Click "Select".
The user settings are loaded.
• When the user list is open, it can be edited at any time.
• Before each operation, make sure that your desired user
is selected and familiarize yourself with the assignment
of the handles and the optional footswitch (if used).
User Settings can be protected via a password. To change the user
settings, please refer to chapter 9.3.
9.2.3 Show Settings
X Press the "Show Settings" button in the dynamic button bar to
see an overview of the user settings of the current user.
X Click one of the default users, then click "Select".
The PROVEO 8 surgical microscope is ready to use.
9.3 Menu – User Settings • You can also add a user from the operational menu.
• If you want to keep the current settings, you can save
You can configure user settings in this menu.
them by clicking the "Save" button (which appears as
X Click the "Menu" button and select "USER SETTINGS". soon as the basic settings of the current users have been
changed), either for the current user ("Save") or under a
new username ("Save as New").
by a lock icon in front of the user setting name in the Select User
page.
9.3.2 Setting the "Main" start values Fast Slitlamp When activated, the slitlamp moves with
For the selected user you can set the start values for the main light, fast travel speed.
Red Reflex light and magnification on this screen. Focus Linked to Zoom When activated, the focus speed depends
on the magnification:
• low magnifaction – fast focus speed
• high magnification – slow focus speed
XY Linked to Zoom When activated, the XY speed depends on
the magnification:
• low magnifaction – fast XY speed
• high magnification – slow XY speed
9.3.6 Combination Mode X Enter the total number of steps by clicking the or key in
With the Combination Mode you can create an individual procedure "Number of Steps".
for each user. You can save the following parameters for various X Select the appropriate parameters for each step in "Step".
frequently recurring phases (max. 5 steps) of the operation: X Set the valid focus mode "Relative" or "Absolute" for all
• Main Light brightness parameters.
• Red Reflex brightness • Relative range: –75 mm … +75 mm
• Magnification • Absolute range: –37.5 mm … +37.5 mm
• Focus Mode
Combination Mode parameters
• Inverter
• ADF1 (Additional function 1) • Magnification for activating (on) and deactivating (off). When
• ADF2 (Additional function 2) set to "off" the magnification stays at the actual value.
• Inverter signal; e.g. for triggering internal or external inverter
from SDI Oculus
• ADF1, ADF2 pulse signal for triggering external systems, e.g.
Room Light on/off, external slit lamp …
The focus can take on two states:
• "Absolute": the learned, absolute position is accurately
approached.
Range: –37.5 mm … +37.5 mm
• "Relative": the learned distance between 2 dots, e.g. for defined
contact lenses for retina operations.
Range: –75 mm … +75 mm
CAUTION
Risk of injury!
X Pay special attention to the required safety distances if
In this screen, the user can enable or disable the desired you use the Combination Mode function together with
Combination Mode parameters for the individual users. accessories from other manufacturers that can reduce
the working distance to less than 140 mm (non-contact
wide-angle observation systems), since focus together
When switching through the Combination Mode function,
with Combination Mode is a semi-automated function.
only the actively set parameters for the individual user are
activated.
For the Combination Mode function to be available, you
must first assign it to a key on your footswitch.
Running Combination Mode
X Activate the key of your footswitch to which the Combination
Mode function is assigned by clicking it.
You run through a continuous loop of the stored steps.
9.3.7 VR Mode
On this page , you can store user-specific settings for the VR mode
(vitreoretinal mode).
These functions are activated at the factory. If VR mode is activated, no user settings can be configured.
You can deactivate the functions using the "ON" key. To do so, first deactivate VR mode.
The setting of "Electronic inverter active" depends on the connected
inverter: Ending VR mode
• Oculus SDI 4c with BIOM: Electronic Inverter active = inactive X Press the button with the "VR Mode on/off" assignment again.
• all others: Electronic Inverter active = active The microscope undoes all actions again.
Footswitch assignment VR When an electrical BIOM is in use the VR mode is
automatically activated/deactivated.
You can store a special footswitch assignment for VR (vitreoretinal)
mode here.
The user can set desired values for quick focus and quick tilt
according to his preferences.
X Activate the footswitch button assigned to the quick focus or
quick tilt function by clicking it.
Distance for quick Distance to move upwards from current
focus position after activating the assigned key.
X Click the "OK" button to save the user at the desired location
under the name you have entered.
Select Eyepiece
In this field you can select the type of the eyepieces being used by
the surgeon.
X Scroll forwards or backwards in the list by clicking the
arrowheads.
Select Objective
In this field you can enter the objective currently being used by the
surgeon.
X Scroll forwards or backwards in the list by clicking the
arrowheads. X Press the button for the topic desired.
Detailed information "How to ..." is displayed.
10.2 Menu – "How to..." 10.3 Menu – "Service"
X Press the Menu button and select "HOW TO" (1).
X Press the Menu button and select "SERVICE MENU" (1).
1 1
This screen displays, in short form, user instructions for operating This area is password-protected.
your surgical microscope.
Before starting the service menu, end the recording
procedure on the documentation system. Otherwise data
could be lost.
11.1.2 Patient-side
Picture Devices and accessories
Binocular tube, 30° to 150°,
T, Typ II L Objective APO WD175 OCT ready
Objective APO WD200 OCT ready
Objective APO WD225 OCT ready
RUV800 WD175
RUV800 WD200
Binocular tube, 10° to 50°, Typ II,
UltraLow ™III
12.5.2 Instructions
Workplace
X Remove surface contamination with a disposable cloth/paper cloth.
Cleaning: manually
• Equipment: water, detergent, alcohols, microfiber cloth
Procedure
X Rinse surface contamination off of the product (temp. < 40 °C). Use some rinsing agent depending upon degree of contamination.
X Alcohol may also be used to clean the optics if heavy contamination such as fingerprints, grease streaks etc. is present.
X Dry off products, except for optical components, with a disposable cloth/paper cloth. Dry off optical surfaces with a micro-fiber cloth.
Cleaning: automatically
• Equipment: cleaning/disinfecting device
It is not recommended to clean products with optical components in a cleaning/disinfecting device. In addition, optical components must
not be cleaned in ultrasonic baths in order to prevent damage.
Disinfection
The alcohol disinfection solution "Mikrozid, Liquid" may be used in accordance with the instructions on the label.
Please note that after disinfection, the optical surfaces must be rinsed thoroughly with fresh drinking water, followed by fresh
demineralized water. The products must be dried thoroughly before the subsequent sterilization.
Maintenance
No special requirements.
Control and functional test
Check the snap-on behavior of rotary knobs and handles.
Packaging
Individual: A standard PE bag may be used. The bag must be large enough for the product so that the closure is not under tension.
Sterilization
See Sterilization table on chapter 12.5.3.
Storage
No special requirements.
Additional information
None
Contact information of manufacturer
Address of local agent
Leica Microsystems (Schweiz) AG verified that the aforementioned instructions for the preparation of a product are suitable for its reuse.
The processing person is responsible for reprocessing with the equipment, materials and personnel and for achieving the desired results in
the reprocessing installation. In general, this requires validations and routine monitoring of the process. Every deviation from the supplied
instructions should also be examined carefully by the processing person to determine effectiveness and possible detrimental consequences.
1) This medical device falls within the validated sertility claims of the STERRAD®100S / STERRAD® 100NX™ / STERRAD®50 / STERRAD®200
Systems.Please follow the instructions for use of your STERRAD® System Umser's Guide prior to sterilizing devices in STERRAD® Systems.
13 Disposal
The respective applicable national laws must be observed for
disposal of the products, with the involvement of corresponding
disposal companies. The unit packaging is to be recycled.
14 What to do if ...?
If your instrument has a malfunction that is not described
here, please contact your Leica representative.
14.1 Malfunctions
Malfunction Cause Remedy
The image remains unfocussed. Eyepieces are not mounted correctly. X Screw the eyepieces all the way on.
Diopters not set correctly. X Perform dioptric correction exactly according to the
instructions (see chapter 7.4.1).
Magnification cannot be adjusted Failure of magnification motor. X Press the magnification rotary knob.
electrically. X Set magnification by turning (see chapter 8.4.4).
Parallelogram cannot be moved. Parallelogram locked in position. X Release the locking mechanism (see chapter 7.7.3).
PROVEO 8 is not correctly balanced. Position of accessory was changed after X Balance out PROVEO 8 (see chapter 7.7.1).
balancing.
The image appears shaded through the Accessories not installed properly. X Install the accessories exactly in the holders (see
microscope at the edges and the chapter 7.2).
illumination field is outside the field of
vision.
Functions cannot be activated using the Incorrect assignment entered at control X Change the assignment using the control unit.
foot switch. unit.
Error code Source Visible for user Meaning Possible cause of error
0112h PROVEO 8 Optics yes Main LED overtemperature • Main LED too hot.
• No air flow possible.
• Opitcs Carrier might be covered.
0116h PROVEO 8 Optics yes RedReflex LED overtemperature • RedReflex LED too hot.
• No air flow possible.
• Opitcs Carrier might be covered.
0402h Slitlamp adapter yes Lamp defective Lamp defective
FE01h MDC yes Optics carrier not found • CAN connection loose.
• Cable damaged.
• Controller defective.
FE02h MDC yes XY controller not found • CAN connection loose.
• Cable damaged.
• Controller defective.
FE03h MDC yes F42 controller not found • CAN connection loose.
• Cable damaged.
• Controller defective.
FE04h MDC yes Slitlamp adapter not found • CAN connection loose.
• Cable damaged.
• Controller defective.
• No slitlamp adapter installed (Device list not correctly
configured).
FE06h MDC yes DIC800 not found • CAN connection loose.
• Cable damaged.
• Controller defective.
• No DIC800 installed (Device list not correctly
configured)
FE07h MDC yes HDR not found • RS232 connection loose.
• Cable damaged.
• Controller defective.
• No IVR850 installed (Device list not correctly
configured).
Error code Source Visible for user Meaning Possible cause of error
FE08h MDC yes External inverter (SDI) not found • CAN connection loose.
• Cable damaged.
• Controller defective.
• No SDI module installed (Device list not correctly
configured).
FE09h MDC yes Camera control unit not found • RS232 connection loose.
• Cable damaged.
• Controller defective.
• No CCU module installed (Device list not correctly
configured).
15.1.3 Microscope carrier Integrated Prepared for integration of video camera system
documentation and digital recording system.
Rotation of optics 380° Open architecture
Inclination tilt –15° / +105° Connectors Numerous built-in connectors for Video and control
data transfer. Internal power supply 12 VDC,
XY speed Zoom linked XY speed
19 VDC, 24 VDC and AC terminals
Indicator LED for Video record status
Carrier for monitor 860 mm long and flexible arm with 4 axis for
IR sensor for remote control of the HDR rotation and inclination to carry optional video
monitor.
XY range 62 × 62 mm Max. monitor weight: 15.5 kg
Focus range 75 mm Materials Use of RoHS conform materials
Camera image size with respect to the field of view Surface coating The microscope is coated with a paint which is
system designed to provide an antimicrobial effect on
• Camera for surfaces.
visible light Field of view
Minimum height In park position: 1945 mm
Range Cantilever Max. 1925 mm
Camera image size Load Max. 8.0 kg from microscope dovetail ring interface
Weight Approx. 380 kg without load
Field of view
15.3 CT42 telescope mount
• TLC 12ZWDN-100 (230 V)
15.2 F42 floor stand • TLC 12ZWDN2-100 (120 V)
Electrical data
Electrical data
Power connection 600 VA Power connection 100–120 V~ 50-60 Hz 1200 W
100–240 V~ 50-60 Hz 220–230 V~ 50 Hz 890 W
Protection class Class 1 Protection class Class 1
Type Floor stand with 4 electromagnetic brakes. Ceiling attachment Max. distance from concrete ceiling to
intermediate ceiling: 1500 mm
Base 690 × 690 mm with four 360° rotating castors with Attachment to concrete shell construction ceiling:
a diameter of 150 mm each, one parking brake 440 mm hole circle
4 x M12 HSLB M12/15
Balancing Adjustable gas spring via balancing knob
Weight Swing arm 70 kg
Floor stand control The latest electronics control for the continuous
Monitor arm 15.5 kg
unit governing of all motor functions and the light
Ceiling bracket 60 kg
intensity.
Telescope 50 kg
Menu selection based on unique software for user-
specific configuration, with built-in electronic Total weight Approx. 200 kg
auto-diagnosis and user support
Brakes Swing arm:
Control unit stand Open architecture for future software Four electromagnetic brakes, operated by turning
developments the handles, one stop lever for vertical movement
Control elements 14/12 function wireless footswitch and rotary Load Swing arm: Max. 8.0 kg from microscope/dovetail
handles ring interface
max. 1231
max. 1
50°
°
. 190
max
no
nde
sto
p
max
. 190
°
max.
180°
380x380
Ø400-Ø530
170
40
H* (standard = 1140 / max. 1640)
±190°
500 (stroke)
600
2900* (max 3600)
390**
350**
1600 (recommended)
1720 (max. 2320)
350° ±150°
1420 (max. 1920)
1205**
415 985 90
1490
The PROVEO 8 surgical microscope is intended for operation in an environment as specified below.
The customer or the user of the PROVEO 8 surgical microscope should make sure that it is used in such an environment.
RF emissions in accordance with CISPR 11 Group 1 The PROVEO 8 surgical microscope uses RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
conducted emissions in accordance with CISPR 11 Class A The PROVEO 8 is suitable for use in establishments other
than domestic and those directly connected to the public
Harmonic emissions according to IEC 61000-3-2 Class A low-voltage power supply network that supplies buildings
used for residential purposes.
Emission of voltage fluctuations/flicker according to IEC 61000-3-3 Complies
The PROVEO 8 surgical microscope is intended for operation in an environment as specified below.
The customer or the user of the PROVEO 8 surgical microscope should make sure that it is used in such an environment.
When the PROVEO 8 surgical microscope is exposed to any of the disturbances below, you might notice one of the following effects:
• flickering/noise on the HD Monitor
• interruptions on the HD Montior
Non of the listed effects above have an impact on the essential performance or safety and effectiveness of the PROVEO 8 surgical microscope. Its no
unacceptable risk for user, patient or environment to be expected.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Discharge of static ± 8 kV contact discharge ± 8 kV contact discharge Floors should be of wood, concrete or ceramic
electricity (ESD) according tile. If floors are covered with synthetic
to IEC 61000-4-2 ± 15 kV air discharge ± 15 kV air discharge material, the relative humidity should be at
least 30 %.
Electrical fast transient/ ± 2 kV for power supply line ± 2 kV for power supply line Mains power quality should be that of a typical
burst immunity according commercial or hospital environment.
to IEC 61000‑4-4 ± 1 kV for input and output ± 1 kV for input and output
lines lines
Surges according to ± 1 kV differential mode ± 1 kV differential mode Mains power quality should be that of a typical
IEC 61000-4-5 commercial or hospital environment.
± 2 kV common mode ± 2 kV common mode
Voltage dips, short <5 % UT (>95 % dip in UT) 70% UT 25/30 cycles Mains power quality should be that of a typical
interruptions and power for ½ cycle commercial or hospital environment.
supply voltage fluctuations 40% UT 10/12 cycles When short interruptions of 5 % UT for
IEC 61000-4-11 40 % UT (60 % dip in UT) 5 seconds occur, the PROVEO 8 surgical
for 5 cycles 40% UT 5/6 cycles microscope will cease operation and restart
automatically. It can be brought back to the
70 % UT (30 % dip in UT) 0% UT 0.5/0.5 cycles state it was before with user intervention.
for 25 cycles If the user of the PROVEO 8 surgical microscope
0% UT 1/1 cycles requires that the instrument remain functional
<5 % UT (>95 % dip in UT) even after power interruptions, it is
for 5 sec 0% UT 250/300 cycles recommended that the PROVEO 8 surgical
microscope be provided with an auxiliary power
source such as an uninterruptible power supply
(UPS) or battery back-up.
The PROVEO 8 surgical microscope is intended for operation in an environment as specified below.
The customer or the user of the PROVEO 8 surgical microscope should make sure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Conducted RF - Disturbance 3/6 Veff 10 Veff d = 0.35 √P for 150 kHz to 80 MHz
variables according to 150 kHz to 80 MHz d = 0.35 √P for 80 MHz to 800 MHz
IEC 61000-4-6 d = 0.7 √P for 800 MHz to 2.7 GHz
Radiated RF - Disturbance 3 V/m 10 V/m Where P is the maximum output power rating of the transmitter in
variables according to 80 MHz to 2.7 GHz 80 MHz to 3 GHz watts (W) according to the transmitter manufacturer and d is the
IEC 61000-4-3 recommended separation distance in meters (m). Field strengths
from fixed RF transmitters, as determined by an electromagnetic site
Proximity fields survey, should be less than the compliance level in each frequency
range. Interference may occur in the vicinity of equipment marked
with the following symbol:
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation amount is affected by absorption and reflection from
structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the PROVEO 8 surgical microscope is used exceeds the applicable RF compliance level above, the PROVEO 8
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the PROVEO 8.
Recommend separation distances between portable and mobile RF telecommunications equipment and the PROVEO 8 surgical
microscope
The PROVEO 8 surgical microscope is intended for operation in an electromagnetic environment in which radiated RF interference is controlled. The
customer or user of the PROVEO 8 surgical microscope can help prevent electromagnetic interference by maintaining the minimum distance between
portable/mobile RF communication equipment (transmitters) and the PROVEO 8 surgical microscope, depending on the output power of the
communication equipment, as stated below.
Rated maximum output 150 kHz up to 80 MHz 80 MHz up to 800 MHz 800 MHz up to 2.5 GHz
power of transmitter in W d = 0.35 √P d = 0.35 √P d = 0.7 √P
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note 1 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
If use is made of accessories or cables other than those specified in this user manual or approved by the manufacturer of the
PROVEO 8 surgical microscope, this can lead to an increase in electromagnetic radiation or a reduction in EMC.
The PROVEO 8 surgical microscope must not be used directly adjacent other instruments. If it is necessary to operate it in the vicinity
of other instruments, the devices should be monitored to ensure that they function properly in this arrangement.
17 Annex
17.1 Checklist before the operation
Patient ........................................................................................................................................................................................
Surgeon ........................................................................................................................................................................................
Date ........................................................................................................................................................................................
2 Installing the accessories X Lock the PROVEO 8 in place and install all accessories on
the microscope so it is ready for use (see chapter 7.2).
X Position the handles as desired.
X Connect the mouthswitch and/or footswitch, if used.
X Check the camera image on the monitor and realign if
necessary.
X Check that all equipment is in its proper position (all
covers fitted, doors closed).
3 Checking the tube settings X Check the tube and eyepiece setting for the selected user.
7 Positioning at the OP table X Position the PROVEO 8 on the OP table as required and lock
the footbrake (see chapter 7.1) (only F42).
www.leica-microsystems.com