Asm 142
Asm 142
© Axess Technology
For further information, please refer to
Operating manual supplied with your unit. Helium signal analog scale display . . . . 7
Audio alarm . . . . . . . . . . . . . . . . . . . . . . . . 7
Zero function . . . . . . . . . . . . . . . . . . . . . . . 8
P/N: 107768 Ed.04 1
Detector connections
Left side Right side
14
7 Control keys (4 keys) 24 Menu selection access key (4 keys)
LEAK RATE COR Torr.l/s
F4 F4
Starting a cycle CYCLE
EVAC TEST
(1) Read
STDBY leak SNIFFER
STDBY
EVAC
value
TEST
SNIFFER
Ending Sniffing test
F1 READY FOR CYCLE
F2
Ending a cycle mode F3 INLET VENT : off
CYCLE
(1) As soon as the inlet pressure reaches 10 mbar (7.5 F4
torr), the unit goes in gross leak test mode, or when the
TEST STDBY
pressure has reached 5 . 10-1 mbar (0.37 torr), the unit
goes in fine leak test mode. Selecting a test mode C 210 6
He signal analog scale display Audio alarm
How to read the He signal analog scale? Leak detector in hard vacuum or The audio alarm offers 2 modes of operation. They are both linked to the zero function.
reject point is visualized by a blinking led. sniffing test mode and zero function zero function not activated: the audio alarm starts when the He signal exceeds a fixed
if the leak value exceeds the reject point, the leds not activated. set point: this set point is programmable.
will turned red (the blinking led will turn orange). zero function activated: the audio alarm is modulated with respect to the position of
if the leak value remains under the reject point, the leds will remain green. the helium background.
Pa.m3/s
the digital display becomes 0.0E-00. On and after this time, it will display only 2 leds flash
Operator
ZERO could find an example in the Operating manual C 540 1 . 10-7 1 . 10-8
hPa .
1.1 10-7 STDBY 9 . 10EVAC
Torr
-8 . 10-7
2TEST
8
INLET PRESSURE
NEXT - + RESET
Air inlet vent
Opening / closing air inlet valve
Purpose: the air inlet valve vents the inlet of the detector back to atmosphere at the end (user level 2 , 3 or 4 )
of the test.
The indicator “inlet valve = off” indicates that the venting valve is not activated
(= closed) at the end of cycle. F3 F3
The setting by default is «off» (= valve closed). To open/close air inlet valve with user level 1 C 500
9
Calibration
F1 AUTOCAL IN PROGRESS