18, 2009 SERVICE NOTE Page 1 of 2

FILE TITLE LOGIQ e/Vivid e/LOGIQ i SN 71092


TAB 310
LCD Cable Selection

APPLIES TO:
All LOGIQ e/Vivid e/LOGIQ i units

PROBLEM:
There are two kinds of LCD cables (GPN: 5125944-3&5125944-4) in LCD Cable Kit (GPN: 5366398).
Refer to Figure 1. Field engineer needs to check and select the right one for the customer unit.

5125944-4

5125944-3

Figure 1 Two LCD Cables in FRU 5366398

ROOT CAUSE:
For the Le/Ve/Li units manufactured after Sep 2008, new LCD cable with lock mechanism (GPN:
5125944-4) is released to improve the connection between cable and LCD panel. But for the
previous units, the original LCD cable (GPN:5125944-3) is still required.

Copyright© 2009 by General Electric Company Inc. All Rights Reserved


Sep.18, 2009 SERVICE NOTE Page 2 of 2

FILE TITLE LOGIQ e/Vivid e/LOGIQ i SN 71092


SOLUTION:
After disassembling the LCD, Please see the connector on the back of the LCD panel.
1. If the connector is same as Figure 2 shows, pls select 5125944-3 LCD cable without locking
mechanism.

Figure 2 Connector 1 on the back of LCD panel

2. If the connector is same as Figure 3 shows, pls select 5125944-4 LCD cable with locking
mechanism.

Figure 3 Connector 2 on the back of LCD panel

Submitted by: Da Qi Roger


FOR ANY FURTHER INFORMATION, PLEASE CONTACT:

Technical Support
GEHC - Americas GEHC - Europe GEHC - Asia
OLC-EMEA
Online Services Ultrasound Asia
Phone: +49 (0) 212-2802-652 (-OLC) Australia +(61) 1-800-647-855
Online Services Ultrasound
+33 1 3083 1300 China +(86) 800-810-8188
Phone: 1-800-321-7937 (USA/Canada)
(English/German all segments India +(91) 1-800-11-4567
1-262-524-5300 (Latin America)
incl.training): Japan +(81) 42-648-2924
Fax: 1-414-647-4125
Fax: + 49 (0) 212-2802-431 Korea +(82) 2620-13585
Singapore +(65) 6277-3444