0% found this document useful (0 votes)
79 views

019-Compression Load Cell

This 3-page document provides the standard calibration procedure for compressive load cells used by INAT, an international agent group that provides calibration, inspection, and other services. The procedure outlines the scope, equipment, preliminary operations, calibration process, and related documents for calibrating compressive load cells according to international standards. Load cells are calibrated by applying various loads and recording readings, which are compared to a master standard. Upon completion, a calibration certificate and sticker are issued.

Uploaded by

Manzoor Ahmad
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
79 views

019-Compression Load Cell

This 3-page document provides the standard calibration procedure for compressive load cells used by INAT, an international agent group that provides calibration, inspection, and other services. The procedure outlines the scope, equipment, preliminary operations, calibration process, and related documents for calibrating compressive load cells according to international standards. Load cells are calibrated by applying various loads and recording readings, which are compared to a master standard. Upon completion, a calibration certificate and sticker are issued.

Uploaded by

Manzoor Ahmad
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 3

International Agent Group

Calibration, NDT, Trading, Inspection, Testing, QMS


Consultancy, IT Solutions and Contracting

STANDARD CALIBRATION PROCEDURE FOR Doc. No.


Rev. 01 Feb-2021
COMPRESSIVE LOAD CELL INAT/MECH/LCP-019

STANDARD CALIBRATION PROCEDURE

FOR

COMPRESSIVE LOAD CELL

Prepared by

HAMMED KHAN Signature Date :

Reviewed by

MUHAMMAD AZAM Signature Date :

Approved by

MANZOOR AHMAD Signature Date :

Controlled Copy No: Uncontrolled Copy:

Issued to:

This document is the property of INAT and is issued only for persons or entity as written on this
manual. It is not to be retransmitted, printed or copied without prior written permission of company

Issue July, 2014 Page 1 of 3

P.O. Box: 30060, Al-Khobar 31952, Tel: 00966 3 8933250, Fax: 00966 3 8643351,
Email: [email protected], [email protected], www.inatt.com
International Agent Group
Calibration, NDT, Trading, Inspection, Testing, QMS
Consultancy, IT Solutions and Contracting

STANDARD CALIBRATION PROCEDURE FOR Doc. No.


Rev. 01 Feb-2021
COMPRESSIVE LOAD CELL INAT/MECH/LCP-019

Table of Contents

Sr. No. Title Page No(s).


1.0 Scope 03
2.0 Normative Reference 03
3.0 Standard Equipment 03
4.0 Preliminary Operations 03
5.0 Calibration Procedure 03
6.0 Related Documents 03

Issue July, 2014 Page 2 of 3

P.O. Box: 30060, Al-Khobar 31952, Tel: 00966 3 8933250, Fax: 00966 3 8643351,
Email: [email protected], [email protected], www.inatt.com
International Agent Group
Calibration, NDT, Trading, Inspection, Testing, QMS
Consultancy, IT Solutions and Contracting

STANDARD CALIBRATION PROCEDURE FOR Doc. No.


Rev. 01 Feb-2021
COMPRESSIVE LOAD CELL INAT/MECH/LCP-019

1.0 Scope

This procedure provides guidelines for calibration of Compressive Load Cell.

2.0 Normative Reference

BS EN ISO 7500-1:2018 and Acceptance criteria ± 1%

3.0 Standard Equipment

2.1 Standard Compressive Load Cells with a valid certificate of calibration traceable to
international standard in accordance with an industrial code or a regulation shall be used
as a standard for calibration.
2.2 The Master standard shall be used under controlled environmental conditions e.g. dust
free environment, Lab Temperature 22°C ± 2°C, Humidity 60%RH ± 20%RH..
4.0 Preliminary Operations

3.1 The details of Compressive Load Cells shall be entered in the Calibration Job Form.
3.2 The Load Cells shall be inspected and marked () in relevant block of equipment's
primary inspection section of Calibration Data Sheet.
5.0 Calibration Procedure

4.1 The Load Cells to be calibrated shall be maintained at same place of (standard Load
Cells) for 01 hour minimum for stabilization at ambient temperature.
4.2 Load cell under test will be installed in series with Standard Compressive load cell shall
be exercised on load for 2 to 3 times.
4.3 The Load corresponding to 20%, 40%, 60% and 80% of full range of Load Cells shall be
applied and the reading on the Load Cells Indicators shall be noted accordingly.
4.4 The relevant section of data sheet form: call/013 shall be marked.
4.5 The readings of Load Cells shall be recorded in observed value column against the
standard readings in set value column of data sheet form: call/013.
4.6 Calibration Certificate shall be issued for calibrated Load Cells as per data sheet form:
call/013 record.
4.8 A Calibration Sticker containing having following information shall be affixed on the
calibrated Load Cells.
a) Item code
b) Calibrated on
c) Next calibration date
6.0 Related Documents

All the latest formats are documented in Master List of Document / Record and placed in CQF
file for reference.
All the latest Standard Calibration Procedures (SCP) are controlled and issued to concerned
Calibration Engineers.