You are on page 1/ 2

Blood Gas Analyzer

Specifications
Operating Parameters
Reportable Range Calculated Parameters Sample Type
pH : 6.500~7.800 pH(T), pCO2(T), pO2(T), tCO2, Heparinized whole blood
pCO2 : 5.0~150.0 mmHg HCO3-, HCO3-(std), BE(ecf),
pO2 : 10~680 mmHg BE(B), Anion gap(K), tHb, sO2 Sample Volume
Na+ : 80~200 mmol/L
K+ : 0.1~20.0 mmol/L
Ca2+(7.4), pO2(A-a) 100 µL Blood Gas Analyzer
Ca2+ : 0.25~5.00 mmol/L Sample Analysis Time
Cl- : 50~150 mmol/L 50 seconds
Glu : 5~500 mg/dL
Lac : 0.2~15.0 mmol/L
Hct : 10~70%

Cartridge
Test Panel Options Test Volume Options Storage Condition
・CarePak 106 50 tests / 2 or 3 weeks 15~25 ℃
pH, Na+, K+, Ca2+, Cl-, Hct 100 tests / 2 or 3 weeks

・CarePak 108
200 tests / 2 or 3 weeks
300 tests / 2 or 3 weeks
Components
・Micro-sensor card
One-Touch Care
・Cal1, Cal2, Cal3
pH, pCO2, pO2, Na+, K+, Ca2+,
・Reference solution
Cl-, Hct
・Sampler & waste bag
・Valve and tubing
・CarePak 110
・EEPROM
pH, pCO2, pO2, Na+, K+, Ca2+,
Cl-, Hct, Glu, Lac Dimensions / Weight
182(W) x 148(H) x 100(D) mm
1.7 kg

Instrument & Connectivity


Computer Display Dimensions / Weight
1.1 GHz dual core processor 9 inch TFT LCD touch screen 269(W) x 391(H) x 255(D) mm
SSD 32GB storage wide view angle 8.4 kg (accessories excluded)

Operating System Printer System Connectivity


Microsoft® Windows® IoT 10 Enterprise 2 inch thermal printer (built-in) Bidirectional LIS/HIS communication

Power Interface
Voltage Power Adaptor Interface
100~240 Va.c. Input: 100~240 Va.c., 1.5 A, ・650nm Laser Diode scanner (built-in)
50/60 Hz, Max. 1.5A ・2 x USB 3.0
Frequency Output: +24 Vd.c., Max. 2.7A ・Serial port (RS-232)
50/60 Hz ・Ethernet port (RJ45)
Battery Operation ・HDMI interface (external display)
Max. 2 hours

i-SENS, Inc.
43, Banpo-daero 28-gil, Seocho-gu, Seoul 06646, Korea TEL +82-2-910-0722 FAX +82-2-941-0868 www.i-sens.com
AD138 REV2 07/2019
Blood Gas Analyzer

01 Maintenance Free
• All-in-One multi test cartridge contains all reagents
& consumables Intuitive Touch Screen Interface
• Self-cleaning sample probe helps maintain sample Built-In Printer
integrity • Microsoft® Windows® 10 operation
• 9 inch TFT LCD touch screen • Hard copy printout of patient and QC
02 Fast, Safe & Reliable Results • Wide view angle calibration results
• EEPROM chip enables automatic cartridge
information loading
• Fast patient test results: 50 seconds
Intelligent Sample Probe
• Fully automated calibration and system External Input / Output Ports
performance monitoring • Syringe or capillary position depending
on sample type selection
• Serial port (RS-232)
03 Easy to Use & Simple Operation • Rinses out septum and probe after each
• Ethernet port (RJ45)
• 2 x USB 3.0
• Syringe and capillary samples without adapters sample to avoid carry-over
• HDMI Interface (external display)
• Illuminated sampling area for low-light
• Built-in sample barcode scanner & printer
• Internal battery provides up to 2 hours of full
testing environments
• Exposed probe only during sampling
operation without AC power Built-In Barcode Scanner
• One-touch design
(automatically retracts for safety purposes)
• Internal barcode scanner enters
04 Connectivity Management samples and operator ID automatically
• Analyzer monitoring including QC status
• Real-time data transfer to HIS/LIS

• Middleware keep strict quality control and strict


Built-In EEPROM Chip
data security & integrity
• Automatic cartridge information
05 Efficiency Rechargeable Internal Battery recognition
Up to 2 hours of sample measurement • Prevention of copy reagents
• Wide range of available cartridges (by analyte •
• Ensures valid cartridges and quality of
menu and volumes) allows for best-fit to testing needs
• Light weight and small foot-print
reagents are used

Care Connect®
Management Function
• Analyzer, Material Registration Operator,
• Micro Sensor Technology
Patient and Ward Management Micro sensors fabricated on i-SENS’ unique
Convenient screen-printed electrodes allow for accurate
POC Devices Integration Device Management analysis of samples at small volumes
• CareSens Expert Plus,
i-SmartCare 10, A1Care

Analyte Consolidation • All-in-One Multi Test Cartridge Technology


DM • Glucose, Ketone, HbA1c, Providing Differentiated
Customer Value Each cartridge contains all reagents, sensors,
Blood Gas, Metabolites, Electrolytes
calibration solutions and complete fluidic
components from sampler to waste bag required
Quality Control for optimal testing
• Variety of QC Export Options and
Statistics Reports
Improved
Work Efficiency
Data Connectivity
• HIS/LIS/EMR
• POCT1-A and HL7 Protocols

You might also like