25 ACHDLSManual 2012
25 ACHDLSManual 2012
CLIA #01D0665512
Prepared by:
Bureau of Clinical Laboratories
Quality Management Division
February 2012
ALABAMA
COUNTY HEALTH DEPARTMENT
LABORATORY SYSTEMS
(ACHDLS)
Table of Contents
Quality Management….…….…………………………..…….…………….……………….…….3
Laboratory Safety………………………………….……………………………………………..15
Miscellaneous Equipment………………………………………………………………………..19
Waived Tests……………………………………………………………………………………..25
Brightfield Microscopy……………………………………………………………………….….29
Darkfield Microscopy…..…………….….…..……………………….………….…….………...62
2. ACCURACY - State of quality of being accurate; closeness of test results to the true value
and implies freedom from error; also referred to as bias.
7. KIT - All components of a testing system (or unit) that are packaged together.
9. MEAN - A number that represents an entire set of numbers, determined for the set in any
number of ways; average.
14. PROFICIENCY SURVEY - A program in which specimens of quality control material are
periodically sent to members of a group of laboratories for analysis and comparison of each
laboratory’s results with those of other laboratories in the group through some central
organization. Participation in a proficiency survey does not replace the day-to-day quality
control process of an individual laboratory.
15. QUALITY MANAGEMENT - Sum of all those activities in which the laboratory is
engaged to ensure that information generated by the laboratory is correct. Quality
management is not restricted to the development and retention of quality control charts but
rather includes all aspects of laboratory activities that affect the results produced, from the
choices of methods, to the monitoring of instruments, to the education of personnel, to the
handling of specimens, and to the reporting of results. The true purpose of quality
management activities is to determine how correct, or incorrect, the results emanating from
the laboratory are, and to allow those managing the laboratory to determine whether or not
the lab is fulfilling its function satisfactorily.
17. REPORTABLE RANGE - The range of test values expected for a designated population of
individuals.
18. SAMPLE - In proficiency testing, means the material contained in a vial, on a slide, or other
unit that contains material to be tested by proficiency testing program participants. When
possible, samples are of human origin.
19. STANDARD DEVIATION - The most common measure of statistics, measuring how
widely spread the values in a data set are dispersed.
20. TARGET VALUE - For quantitative tests, means either the mean of all participant
responses after the removal of outliers (those responses greater than three (3) standard
deviations from the mean) or the mean established by definitive or reference methods
acceptable for use by the National Reference System for the Clinical Laboratory (NRSCL)
by the National Committee for Clinical Laboratory Standards (NCCLS).
B. All laboratory personnel must be trained properly, commensurate with their positions,
duties, and responsibilities.
C. The Alabama County Health Department Laboratory Systems (ACHDLS) will maintain
a quality control system to assure continued precision and accuracy of laboratory
results.
D. The ACHDLS will participate in the Centers for Medicare & Medicaid Services (CMS)
approved proficiency testing (PT) program.
The policies and procedures of the quality management program will be approved by the
laboratory director when first written, with notation of approval by signature and date. The
technical consultants will review the policies and procedures on a regular basis. If a policy
or procedure requires a change, a new policy or procedure will be written, approved by the
laboratory director, and distributed to testing sites. The old policy or procedure will be
retained in a file at the Bureau of Clinical Laboratories, County Assistance Section, for a
minimum of two years.
B. Procedure Manuals
A written procedure manual containing procedures for all activities of the ACHDLS
will be maintained and readily available at all times to personnel in each testing site.
E. Proficiency Testing
ACHDLS will participate in a CMS-approved PT program. PT results will be reviewed
by testing personnel and technical consultants, verifying that all information is correct
and complete before they are mailed, faxed, or e-mailed to the PT program.
Investigations of unsatisfactory PT results (performance that does not result in 100% of
acceptable responses for each analyte, including graded, ungraded, and unregulated
analytes) will be documented by testing personnel and reviewed by technical
consultants and laboratory director.
H. Personnel Assessment
An ongoing evaluation of all testing personnel will be conducted through use of
proficiency testing results, review of quality control records, Complete Health Records,
observation and annual competency evaluations for nonwaived testing procedures.
These include wet preps, RPRs, and darkfield microscopy. If during competency
evaluations an employee is found to be incompetent for one or more procedures, the
employee will not be able to perform the test(s). If (s)he is still not competent after
reevaluation, the employee cannot perform that testing until (s)he has attended training
for that procedure and deemed competent.
I. Communications
Problems that occur as a result of breakdowns in communication between testing
personnel and the authorized individual who orders or receives the results of tests will
be documented. In addition, corrective actions taken to resolve problems and minimize
communication breakdowns will be documented.
J. Complaint Investigations
Investigation of complaints will be made and corrective actions, when necessary, will
be taken. Documentation will be maintained.
M. Laboratory Log
1. The Laboratory Log is designed to assure follow-up on laboratory tests referred
from the county health department to other laboratories. The Laboratory Log is
also a method to track when a specimen leaves the county health department and
the date the report is received. This tracking system for referred specimens is
required by the Clinical Laboratory Improvement Amendments of 1988 (CLIA’88),
which is a federal regulatory program governing all laboratory testing.
2. This tracking system will include patient name, identification number, date of
service, tests referred, and date the report was received. See “Alabama County
Health Department Laboratory Systems Laboratory Log”.
3. Instructions for the Laboratory Log:
a. Place the page number in the space provided.
b. Enter the date in the space provided.
c. Place the label (PHALCON) indicating the patient identity on the sheet. Be
sure to correct the service date if labels were preprinted.
d. Indicate the referred tests on the lines provided. Use 1 line per test (i.e., For
multiple testing such as TV, indicate numbers 1, 5 and 18). If you have more
than 6 tests per patient, use the next 6 lines and mark out the space for the label.
A list of numbers corresponding to tests most commonly referred is printed at
the bottom of the log. Any other tests should be written.
e. Indicate the date the referred laboratory test report is received in the CHD.
f. Have a system devised to locate any test results not received in a timely manner.
A. Classes
1. Rapid Plasma Reagin (RPR): One-day class for the macroscopic screen for syphilis.
2. Darkfield Microscopy for the Detection and Identification of Treponema pallidum:
Two-day class to include Brightfield Microscopy, Darkfield Microscopy and
Identification of Treponema pallidum.
3. Vaginal Wet Mount: Two-day class to include Brightfield Microscopy and Vaginal
Wet Mount.
C. Faculty
1. Jane Duke
2. Charlene Thomas
D. Training Schedule
1.GC 2.PAP 3.Syphilis EIA 4.HIV 5.Chlamydia 6.MCH-FP 7.Urine C&S 8.PKU 9.Galactosemia 10.CAH 11.T4 12.TSH 13.Abn Hgb 14.Lead 15.SGOT 16.CBC 17.TB Profile 18.TV
ACHDLS - Quality Management Any other referred test must be written in. Revised 4/2012
Laboratory Personnel
in County Health
Departments
LABORATORY PERSONNEL IN COUNTY HEALTH DEPARTMENTS
The Clinical Laboratory Improvement Amendments of 1988 (CLIA’88) require the designations
of laboratory director, clinical consultant, technical consultant, and testing personnel for the
Alabama County Health Department Laboratory Systems. A current list of individuals
functioning as the laboratory director, technical consultant, and clinical consultants is included at
the end of this section. Testing personnel are persons in the local county health departments who
perform any laboratory testing (e.g., nurses, clinic nurse aides, nurse practioners, physicians,
disease intervention specialists, laboratory technicians, and nutritionists).
The laboratory director is responsible for the overall operation and administration of the
laboratory, including the employment of personnel who are competent to perform test
procedures, record and report test results promptly, accurately, and proficiently and for
assuring compliance with the applicable regulations. Some responsibilities may be
reapportioned to the technical consultants; however, the laboratory director remains
responsible for assuring that all duties are properly performed.
A. Ensure testing systems used for each of the tests performed provide quality laboratory
services for all aspects of test performance, which include preanalytic, analytic, and
postanalytic phases of testing.
B. Ensure that the physical plant and environmental conditions of the laboratory are
appropriate for the testing performed and provide a safe environment in which
employees are protected from physical, chemical, and biological hazards.
C. Ensure that test methodologies selected have the capability of providing the quality of
results required for patient care.
D. Verification procedures are adequate to determine the accuracy, precision, and other
pertinent performance characteristics of the method.
E. Laboratory personnel are performing test methods as required for accurate and
reproducible results. (This responsibility has been delegated to the technical
consultants.)
F. Ensure that the laboratory is enrolled in a CMS approved PT program for the testing
performed and that:
1. Proficiency testing (PT) samples are tested as required by CLIA’88 regulations.
2. Results are returned within the time frames established by the PT program.
G. Ensure that quality control and quality management programs are established and
maintained to assure quality of laboratory services and to identify failures in quality as
they occur. (This responsibility has been delegated to the technical consultants.)
I. Ensure that reports of test results include pertinent information required for
interpretation. (This responsibility has been delegated to the technical consultants.)
L. Ensure that prior to testing patients’ specimens, all personnel have the appropriate
education and experience, receive the appropriate training for the type and complexity
of the services offered, and have demonstrated that they can perform all tests reliably to
provide and report accurate results. (This responsibility has been delegated to the
technical consultants.)
M. Ensure that policies and procedures are established for monitoring testing personnel
who conduct preanalytical, analytical, and postanalytical phases of testing to assure that
they are competent and maintain their competency to perform test procedures and
report test results promptly and proficiently, and whenever necessary, identify needs for
remedial training or continuing education to improve skills.
N. Ensure that an approved procedure manual is available to all personnel responsible for
any aspect of the testing process. (This responsibility has been delegated to the
technical consultants.)
O. Specify in writing the responsibilities and duties of each consultant and person
engaged in the performance of the preanalytic, analytic, and postanalytic phases of
testing that identifies which examinations and procedures each individual is authorized
to perform, whether supervision is required for specimen processing, test performance,
The technical consultant is responsible for the technical and scientific oversight of the
laboratory. The technical consultant is not required to be onsite at all times testing is
performed; however, (s)he must be available to the laboratory on an as needed basis to
provide consultation via onsite, telephone, or electronically. In addition to the laboratory
director responsibilities delegated to the technical consultant, other responsibilities are:
A. Selection of test methodology appropriate for clinical use of the test results.
B. Evaluation of testing facilities at any time to ensure testing personnel and procedures,
including the precision and accuracy of each test and test system, are in compliance
with CLIA ’88 Regulations.
D. Establishing a quality control program appropriate for the testing performed, establish
parameters for acceptable levels of analytic performance and ensuring that these levels
are maintained throughout the testing process from the initial receipt of the specimen
through sample analysis and reporting of test results.
E. Resolving technical problems and ensure that remedial actions are taken when test
systems deviate from the laboratory’s established performance specifications and
ensuring that test results are not reported until all corrective actions have been taken
and the test system is functioning properly.
F. Identifying training needs and assuring that each individual performing tests receives
regular in-service training and education appropriate for the type and complexity of the
laboratory services performed.
G. Evaluating the competency of all testing personnel and assuring that the staff maintains
its competency to perform test procedures and report results promptly, accurately, and
proficiently. The procedures for evaluation of the competency of the staff must
include, but are not limited to:
1. Direct observation of routine test performance, including patient preparation, if
applicable, specimen handling, processing and testing.
2. Monitoring the recording and reporting of test results.
3. Reviewing intermediate test results or worksheets, quality control records, PT
results, preventive maintenance and function checks.
C. Ensure that reports of test results include pertinent information for specific patient
interpretation.
Testing personnel performing waived and nonwaived tests are responsible for specimen
processing, test performance, and for reporting test results. Employees perform only those
nonwaived tests that are authorized by the laboratory director (or his/her designee) and
require a degree of skill commensurate with the individual’s education, training or
experience and technical abilities.
A. Follow the laboratory’s procedures for specimen handling and processing, test analyses,
reporting and maintaining records of patient results.
B. Maintain records demonstrating that PT samples are tested in the same manner as
patient samples.
C. Adhere to the laboratory’s quality control policies and document all quality control
activities (e.g., instrument and procedural calibrations and maintenance).
D. Follow the laboratory’s established corrective action policies and procedures whenever
test systems are not within the laboratory’s established acceptable levels of
performance.
F. Document all corrective actions taken when test systems deviate from the laboratory’s
established performance specifications.
The Laboratory Supply Evaluations are used to ensure that laboratory reagents, media, test
kits, and other laboratory supplies are maintained in adequate volumes and are used before
the expiration dates. See “Laboratory Supply Evaluation”.
Monthly Laboratory Reviews are used to monitor and investigate potential noncompliance
in the clinic’s Quality Management System. See “Monthly Laboratory Equipment and
Supply Review”.
D. Technical consults will examine laboratory monthly review documentation during visits
or may request copies be sent to their offices periodically.
VII. Competency Evaluations
The competency evaluation process will also be used to assess training needs. As needs are
identified, training will be provided to testing personnel.
A. Criteria have been defined for evaluating the competency of testing personnel.
CLIA’88 Regulations require that individuals perform only those nonwaived tests that
require a degree of skill commensurate with the individual’s education, training or
experience, and technical abilities. Elements included in training must ensure that each
individual has the skills required for:
G. Assessing and verifying the validity of patient results through the evaluation of QC
sample values prior to reporting patient test results.
X. Delegation of Responsibilities
By signing and initialing this form, the staff member attests that he/she has reviewed all equipment according to the
requirements stated in the Alabama County Health Department Laboratory Systems Policies and Procedures Manual.
Laboratory Supply Evaluation: Laboratory Supply Evaluation has been started and all laboratory testing kits and supplies are
within acceptable dates. Monthly date and initials below denote compliance to the above statement. Make use of the back of this form to document
problems, noncompliance issues, and comments.
Lab Supply JAN FEB MAR APR MA
AY JU
UNE
Evaluation
Date / Initials
JULY AUGUST SEPT OCT NOV DEC
Date / Initials
Refrigerators: Annual Temperature Chart has been started for the current calendar year with temperatures documented every day the
clinic is open. Temperatures are within acceptable range, corrective action performed/documented, and no food is present. Monthly date and
initials below denote compliance to the above statement. Make use of the back of this form to document problems, noncompliance issues, and
comments.
Refrigerators JAN FEB MAR APR MA
AY JU
UNE
Date / Initials
JULY AUGUST SEPT OCT NOV DEC
Date / Initials
HemoCue: HemoCue Maintenance Chart and Annual Temperature Chart have been started for the current calendar year with maintenance/
temperatures documented each day of use. Daily/monthly cleanings have been documented. Corrective action performed/documented when
instrument malfunctions. Monthly date and initials below denote compliance to the above statement. Make use of the back of this form to
document problems, noncompliance issues, and comments.
HemoCue JAN FEB MAR APR MA
AY JU
UNE
Date / Initials
JULY AUGUST SEPT OCT NOV DEC
Date / Initials
Centrifuge: Centrifuge Preventive Maintenance has been started for the current calendar year, a cover is provided, and monthly cleanings
have been documented. Monthly date and initials below denote compliance to the above statement. Make use of the back of this form to
document problems, noncompliance issues, and comments.
Centrifuge JAN FEB MAR APR MA
AY JU
UNE
Date / Initials
JULY AUGUST SEPT OCT NOV DEC
Date / Initials
Wet Mount Microscope: Microscope Maintenance Chart has been started for the current calendar year, a cover is provided, and daily
/ monthly cleanings have been documented. Wet Mount supplies (KOH, saline, ph paper, lens paper and cleaner) are present and not
contaminated. The last annual cleaning date is also documented. Monthly date and initials below denote compliance to the above statement.
Make use of the back of this form to document problems, noncompliance issues, and comments.
WM Microscope JAN FEB MAR APR MA
AY JU
UNE
Date / Initials
JULY AUGUST SEPT OCT NOV DEC
Date / Initials
RPR Rotator: RPR Rotator Preventive Maintenance has been started for the current calendar year. All daily, monthly, and
quarterly cleanings/checks have been documented. Monthly date and initials below denote compliance to the above statement. Make use of the
back of this form to document problems, noncompliance issues, and comments.
RPR Rotator JAN FEB MAR APR MA
AY JU
UNE
Date / Initials
JULY AUGUST SEPT OCT NOV DEC
Date / Initials
Darkfield Microscope: Microscope Maintenance Chart has been started for the current calendar year, a cover is provided, and daily/
monthly cleanings have been documented. Darkfield supplies (immersion oil, lens paper and cleaner) are present and not contaminated. The
last annual cleaning date is also documented. Monthly date and initials below denote compliance to the above statement. Make use of the back
of this form to document problems, noncompliance issues, and comments.
Darkfield JAN FEB MAR APR MA
AY JU
UNE
Microscope
Date / Initials
JULY AUGUST SEPT OCT NOV DEC
Date / Initials
Incubator: Incubator Preventive Maintenance and Annual Temperature Chart have been started for the current calendar year with
maintenance/temperatures documented each day of use. Temperatures are within acceptable range and corrective action performed/
documented. Climate control bags are present. Monthly date and initials below denote compliance to the above statement.
Make use of the back of this form to document problems, noncompliance issues, and comments.
Incubator JAN FEB MAR APR MAY JUNE
Date / Initials
JULY AUGUST SEPT OCT NOV DEC
Date / Initials
Date: _____________________________
Contact: ________________________________________
List ALL personnel performing laboratory testing, including nurses, clinic aids, nutritionist, and disease intervention
specialist. This form should be completed whenever there is staff change. When changes occur, please update the form
and send a copy to the Quality Management Office, Bureau of Clinical Laboratories in Montgomery.
Name Title
Laboratory workers will be aware of safety at all times and protect themselves and others.
Further information may be found in the recent publications of the Alabama Department of
Public Health’s Infection Control Guidelines Manual and Bloodborne Pathogens Exposure
Control Plan.
I. Biological Hazards
A. Use protective coverings when handling blood or other body fluids which may contain
blood or mucous membranes. For example:
1. Cover street clothes with a lab coat.
2. Protect eyes, nose, and mouth with a full face shield or safety goggles and mask.
3. Wear gloves.
4. Wear closed toed shoes.
B. Wash hands thoroughly after removing gloves, handling laboratory specimens, and
before leaving the laboratory area.
C. Remove protective clothing and gloves before leaving the lab area to prevent
contamination of other work areas.
D. Do not eat, drink, smoke, apply cosmetics or handle contact lenses in the laboratory area.
E. Avoid hand contact to mouth, nose and eyes while working in the laboratory area.
Keep pencils and instruments away from mouth.
F. Keep immunizations and skin tests up to date: Hepatitis B vaccine, tuberculin skin test
(chest x-ray as indicated), etc.
J. Clean up biological spills as soon as possible. Follow the steps listed below.
1. Use puncture resistant gloves for sharp objects involved in the spill.
2. Absorb the spill with paper towels or other appropriate absorbent material. Using
two pieces of stiff cardboard as shovels, carefully scoop the soaked towels and
debris into an appropriate biohazard waste container.
3. Clean the site with soap and water until all visible material has been removed.
4. Disinfect the area with a freshly prepared 1:10 bleach solution or other appropriate
disinfectant. This may be done by covering the area with absorbent paper towels
and flooding the towels until they are “glistening wet.” Allow the disinfectant to set
briefly, then blot up the disinfectant soaked towels and place into an appropriate
waste container.
Note: Bleach solution should be prepared fresh each day of use because it loses its
germicidal effect upon prolonged sitting.
5. Rinse the area well with water.
A. Know the chemical with which you will be working. Consult the Material Safety Data
Sheet (MSDS) from the manufacturer before you begin to use the chemical to
determine the physical and health hazards associated with it. For example:
1. Does it contain hazardous ingredients?
2. What are its physical/chemical characteristics?
3. Are there fire and explosion hazards?
4. Are there health hazards?
5. What are the first aid procedures?
6. What should be done if the chemical is spilled?
7. How should the chemical be stored?
C. Large spill example: A gallon jug of bleach was dropped and the side split, spilling the
entire contents on the floor.
1. This spill could create enough toxic vapors to possibly damage the lungs of the
individual exposed long enough to clean up the spill.
2. Evacuate the area.
D. Small spill example: A small puddle of bleach landed on the counter while measuring it
out for a 1:10 solution.
1. The spill is very small and easy to manage. Possible skin irritation would be the
hazard most likely encountered.
2. Maintain the spill within the area where the accident occurred.
3. Wear protective clothing as applicable.
4. Use absorbent towels to contain, carefully wipe up, and properly dispose of the
debris.
5. Rinse the area well to remove all traces of the chemical. (In some instances, such
as strong acids or bases, this would involve a neutralizing agent).
A. Use proper lifting techniques, putting the weight on the leg muscles to avoid back
strain.
B. Avoid frayed electrical cords and overloaded outlets to prevent an electrical fire.
C. Any electrical equipment that produces a “tingle” when touched should be disconnected
and sent for repair.
D. Keep electrical cords coiled and away from sink areas to prevent instrument
displacement and electrical shock.
F. Keep walkways free of articles such as pencils and paper clips that could cause unsure
footing.
G. Avoid exposed sharp corners such as open drawers and cabinet doors.
IV. References
B. Baron, Ellen Jo and Finegold, Sydney M. Bailey and Scott’s Diagnostic Microbiology,
8th Edition, Mosby Co., St. Louis Co.
A. In general, centrifuges are used as filtration and/or packing devices for the separation of
components within a liquid or from a liquid medium, e.g., separating plasma or serum
from the cellular components of blood. Adequate and proper separation without
damage to the components depends upon proper spin (centrifugal force produced by the
speed and length of the radial arm and the time the material is spun). Proper balance,
lubrication, and rotor function are also important in producing the desired result.
Ensuring proper instrument function with regular maintenance is absolutely essential.
B. General purpose centrifuges place a centrifugal force upon fluids to separate out
cellular components. The centrifugal forces increase with the radial arm length and
increasing speed. Usually the radial arm length is fixed, so most centrifuges increase
centrifugal forces by increasing the speed. Speeds of 5,000 revolutions per minute
(RPMs) are generally adequate for most laboratory purposes. Normograms for
calculation of relative centrifugal force (RCF) are available. The RCF can be
calculated by using the following formula:
RCF = 28.38 (r) [n/1000]2
Where r = rotor radius in inches
n = revolutions per minute (RPM)
RCF is expressed in gravities (g)
The RCF is important to know because different centrifuges with different rotor radii
will result in different RCFs when the RPM is equal.
The centrifuge may also have a braking device, speed controls, a timer, or a fuse. Some
larger centrifuges may be equipped with a tachometer.
When tubes of blood are subjected to the proper centrifugal force, the serum and clot
separate. If a serum “separator” is in the tube, it will come to reside between the serum
and the cells and cover the cells completely to prevent further cell-serum contact. This
must be performed without damage to the cellular components to prevent release of
materials such as potassium from the cells. For routine serum separations, clotted
blood is centrifuged for 10 minutes at an RCF of 850 to 1000g.
E. Maintenance
Always unplug the centrifuge from its electrical source before conducting preventive
maintenance, cleaning, and/or inspection. Use only recommended cleaners.
1. Spillage – Every spill will be cleaned up at the time of its occurrence to avoid
materials getting into the mechanism. Biological substances will be handled with
universal precautions. Be particularly cautious when cleaning up broken glass.
2. Unusual previously unnoted noises and/or vibrations – Noises and vibrations should
be listened/looked for with each use of a centrifuge. Appropriate steps will be taken
to determine the cause if they are noted. Damage to the centrifuge and/or the user
may result if vibrations or unusual noises are allowed to continue.
3. Cleaning the exterior and interior – The exterior should be cleaned each day of use.
Any spills will be cleaned up immediately. The chamber should be cleaned
monthly with soap and water followed by an adequate rinse. An appropriate
germicidal and virucidal (with minimal residual) disinfectant-type solution should
be used as the rinse. Gaskets must be washed and checked for wear and/or defects.
When tube shields and/or cups are removed and cleaned, make certain that all
rubber cushions and shields are replaced to maintain proper balance.
4. Inspection of the head, head shaft, and coupling – Inspect these components for
evidence of wear, cracks in fitting, corrosion, uneven wear, and/or signs of fatigue.
Immediately replace any part found to be unacceptable.
5. Rotor balance – Follow the manufacturer’s instructions to check the rotor balance
when any unusual vibrations occur.
6. Brushes, bearings, and commutators – Brushes will be checked as indicated in the
manufacturer’s manual. Replacement is generally indicated if they have worn
down by 5/16 to ½ inch or to within ¼ inch of spring depending upon the motor.
New brushes must slide freely in the holder but yet maintain spring tension to keep
them in good contact with the commutator.
II. Thermometers
Two basic types of thermometers are generally used in the clinical laboratory: bimetallic
and mercury column.
1. Periodically inspect each thermometer carefully for cracks in the capillary or bulb.
Check the mercury column for separations. If separations are found, the column
may be reunited by using one of the following methods:
a. Immerse the thermometer bulb in an ice-salt mixture until all mercury is drawn
into the bulb. Hold the thermometer in a vertical position, tap it gently to
dislodge gas bubbles, and allow it to warm at room temperature.
b. Hold the thermometer vertically in your fist so that the bulb is in the center of
the palm of your hand. If the mercury begins to move, continue until column is
reunited.
C. Calibration of thermometers
1. When placed in service, the accuracy of each thermometer will be checked against
an NIST-traceable thermometer. Mercury column thermometers should keep their
calibration unless the mercury separates; after reuniting the mercury column, the
thermometer should be calibrated. Bimetallic thermometers will need calibration
periodically. Correction factors should be noted on the thermometer and on the
temperature chart. Recorded temperatures will be those obtained after the
correction factor has been used. Thermometers with a correction factor greater than
1°C are not acceptable and will not be used.
2. Because of the cost of an NIST thermometer, thermometers checked against an
NIST will be used to calibrate laboratory thermometers. NIST-traceable
thermometers may be purchased from suppliers of laboratory equipment.
a. Obtain an NIST-traceable thermometer (formerly referred to as an NBS-
traceable thermometer).
b. Place the thermometer being calibrated and the NIST-traceable in the
refrigerator or incubator in which the thermometer will be routinely used. If the
thermometer is to be used for obtaining room temperatures, simply place the
NIST-traceable thermometer next to the thermometer to be calibrated. Allow
time for the temperature to stabilize.
c. Read the temperature on both thermometers.
d. Record readings on the “Thermometer Calibration Log”, and make note of the
correction factor, if any.
e. If the correction factor is greater than +1°C, the thermometer will be replaced
with one that reads within +1°C of the NIST-traceable.
f. Correction factors will be noted on the thermometer and on temperature charts.
g. When recording temperatures, the true temperature (thermometer reading + the
correction factor) will be recorded.
D. Placement of thermometers
Refrigerators and incubators may have “cold spots” or “hot spots”. Windows and air
vents may cause these same conditions in rooms. Therefore, it is important to place
thermometers in locations which reflect the most consistent temperature of the area.
Periodically changing locations of thermometers for short periods of time is useful in
locating “cold spots” and “hot spots” so that these areas can be avoided.
When performing RPR testing, the room temperature should be 23°C - 29°C (74°F - 84°F).
The Hemocue requires a room temperature of 18°C to 30°C (64°F to 86°F). The room
temperature of areas where reagents are stored will be maintained with the temperature
range defined by the reagent manufacturer. This range is noted on the package insert.
Biological or chemical materials in conjunction with food and drinks designed for human
consumption will not be stored in the same refrigerator.
Note: Frost-free refrigerators will not be used if reagents or materials are temperature
sensitive.
A. Function Checks
1. Read and record internal temperature – The desired temperature range is 2°C to 8°C
(36°F-46°F).
a. A thermometer should be placed inside the refrigerator in a location where it is
protected and easily read without disturbing.
b. A temperature chart will be attached to the front or side of each refrigerator.
c. Temperatures will be checked and recorded on the appropriate chart each day
the clinic is open.
d. If the temperature is not within the range of 2°C to 8°C, actions will be taken to
remedy the problem (e.g., adjust the thermostat).
i. If the thermostat is adjusted, recheck the temperature.
ii. If the temperature is still not within the acceptable range and adjusting the
thermostat has not solved the problem, move the contents of the refrigerator
to another refrigerator and contact a repairman.
iii. Dispose of materials which could have been damaged (e.g., items which were
frozen were stored for a long period of time at temperatures above 8°C).
iv. Record corrective actions on the temperature chart.
2. Check door gasket – Ensure that the door gasket seals properly when the door is
closed. If the door does not seal properly and gasket appears to be adequate, the
refrigerator may not be level.
B. Maintenance
1. Defrost unit – If the refrigerator is not frost-free, defrost the unit approximately
every three months. If frost buildup is a problem, defrosting may be required more
frequently. Follow the manufacturer’s instructions for defrosting.
2. Periodic cleaning – The interior should be decontaminated periodically. The
exterior will be cleaned, when necessary, with a damp cloth and mild soap.
V. Microbiology Incubators
B. Maintenance
Clean interior of incubator – The interior of the unit will be cleaned at least monthly.
1. Empty the incubator, and clean the interior with a warm solution of water and
appropriate disinfectant.
2. Rinse the interior with clean water and dry thoroughly.
3. Removable parts may be washed in warm, sudsy water. Avoid soap-filled pads or
metal scouring pads.
4. Return parts and contents to the incubator.
VI. References
Date Jan Intls Feb Intls Mar Intls Apr Intls May Intls June Intls July Intls Aug Intls Sept Intls Oct Intls Nov Intls Dec Intls
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
ADPH/BCL/QM 10-09
Incubator Preventive Maintenance
County Year
Site
MONTHLY JAN FEB MAR APR MAY JUNE JULY AUG SEPT OCT NOV DEC
Clean Exterior and Interior
Check Door Gasket for Proper Seal
Insure Latch Secures Properly
Date
Initials
Climate control bags must be properly sealed when used in the incubator. Please note on the
“Corrective Action Log” below if problems occur. No response recorded indicates that there
have been no climate control bag issues.
B. Specimen Collection
1. To perform a test using capillary blood, the cuvette holder should be in its loading
position. The display will show three flashing dashes and the HemoCue symbol.
2. Make sure the patient’s hand is warm and relaxed. Use only the middle or ring
finger for sampling. Avoid fingers on which rings are present.
3. Clean the finger with alcohol or a suitable disinfectant and allow it to dry or wipe it
off with a dry, lint-free wipe.
4. Using your thumb, lightly press the finger from the top of the knuckle towards the
tip. This stimulates the blood flow towards the sampling point.
5. For best blood flow and least pain, sample at the side of the fingertip, not in the
center.
6. While applying light pressure towards the fingertip, puncture the finger using a
lancet.
7. Wipe away the first 2 or 3 drops of blood.
8. Re-apply light pressure towards the fingertip until another drop of blood appears.
12. Place the filled microcuvette in the cuvette holder. This must be performed within
10 minutes after filling the microcuvette!
13. Gently slide the cuvette holder to the measuring position.
14. During the measurement, an hourglass and 3 fixed dashes will be shown on the
display.
15. After 15-60 seconds, the hemoglobin value of the sample is displayed. The result
will remain on the display as long as the cuvette holder is in the measuring position.
When operating on battery power, the analyzer will automatically turn off after
approximately 5 minutes.
16. Although the reagents are present in the microcuvette in extremely low quantities,
consult local environment authorities for proper disposal. Always handle blood
specimens with care; they might be infectious.
A. This product is a guaiac slide test for the qualitative detection of fecal occult blood. It
is useful in the diagnosis of a number of gastrointestinal disorders and is recommended
for use in:
1. Routine physical examinations.
2. Routine hospital testing.
3. Mass screening for colorectal cancer.
B. Please refer to the package inserts of the specific kit you are using.
Note: A package insert of the test kit should be placed in a sheet protector and placed in
this manual immediately after this section.
B. Please refer to the package inserts of the specific kit you are using.
Note: A package insert of the test kit should be placed in a sheet protector and placed in
this manual immediately after this section.
A. The chemical analysis of urine is performed using one of many different urine reagent
strips (dipsticks) available from a variety of manufacturers. Reagent strips for
urinalysis are firm plastic strips to which are affixed several separate reagent pads. A
color reaction develops upon contact of the urine with the reagent pads. The following
are commonly available on urine dipsticks:
1. Protein.
2. Glucose.
3. Ketone bodies.
4. Bilirubin.
5. Leukocyte esterase.
6. pH.
7. Occult blood/hemoglobin.
8. Urobilinogen.
9. Nitrite.
10. Specific gravity.
B. Refer to the package inserts for the specific reagent areas on the product you are using.
Note: A package insert of the test kit should be placed in a sheet protector and placed in
this manual immediately after this section.
A. Urine pregnancy test measures the presence of human chorionic gonadotrophin (HCG)
in urine for the early detection of pregnancy.
B. Refer to the package inserts of the specific kit you are using.
Note: A package insert of the test kit should be placed in a sheet protector and placed in
this manual immediately after this section.
HemoCue Controls (Lot number, range and values will be documented by Technical Consultant):
Level 1 -
Level 2 -
Level 3 -
Date Jan Intls Feb Intls Mar Intls Apr Intls May Intls June Intls July Intls Aug Intls Sept Intls Oct Intls Nov Intls Dec Intls
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
Correction Action Log on Back County _______________ Year ____________
FOOD OR BEVERAGE IS PROHIBITED FROM STORAGE IN CLINIC REFRIGERATORS/FREEZERS AT ALL TIMES!
ADPH/BCL//QM 10-09
Temperature Corrective Action Log
ADPH/BCL//QM 10-09
HemoCue Hb 201+ Trouble Shooting Guide
If you are unable to resolve the problem by following this Trouble Shooting Guide, please
contact HemoCue, Inc. The analyzer has no serviceable parts.
A. The microscope is perhaps the piece of equipment that receives the most use and,
unfortunately, misuse in the clinical laboratory. Microscopy is necessary to see
microbial cells and to determine morphological characteristics. It is a basic part of
work in many areas of the laboratory such as hematology, urinalysis, and microbiology.
Because the microscope is such an important piece of equipment and a precise
instrument, it must be kept in excellent condition, optically and mechanically. It must
be kept clean and aligned.
C. The total magnification observed is the product of the magnifications of these two
lenses. In other words, the magnification of the objective times the magnification of
the ocular equals the total magnification. The magnitude of magnification is inscribed
on each objective as a number. These magnification units are in terms of diameters
(10x, 40x, and 100x). Thus, 10x means that the diameter of an object is magnified 10
times its original size.
D. Because of the manner in which light travels through the brightfield microscope, the
image that is seen is upside down and reversed. The right side appears as the left, the
top as the bottom, and vice versa. This should be kept in mind when moving the slide
(or object) being observed.
A. Aperture diaphragm - either a rotating disk or an iris diaphragm on the condenser used
to direct a cone of light to the specimen and objective. It should never be used to
regulate brightness. Resolution, control, and depth of field depend on the correct
setting of the aperture diaphragm.
E. Condenser - the lens system beneath the microscope stage positioned to concentrate
light correctly on the specimen and direct light rays into the objective. When the
condenser is used at a lowered position, the resolving power is reduced.
J. Immersion oil – oil with the same refractive index as glass, 1.515; used between the
cover glass and an oil immersion objective to prevent scattering of light in air.
K. Interpupillary distance - the distance between the eyes. The eyepieces of a binocular
scope must be adjusted so that left and right images merge into one.
O. Parcentric - the ability to center a specimen in the field of view for one objective and
have almost the same field in place when rotating to another objective.
P. Parfocal - the objectives are constructed so that only slight refocusing with the fine
adjustment knob is needed after rotating to another objective.
Q. Resolution - the ability of a microscope to reveal fine detail in a specimen. The better
the resolving power of a microscope, the closer two objects can be and still be
distinguished as two objects.
S. Working distance - distance between the coverslip of a slide and the tip of an objective.
The low power objective has the greatest working distance. The oil immersion
objective has a very small working distance.
A. The basic structures of compound microscopes are categorized into 4 groups (see
“Microscope Diagram”): the framework, the illumination system, the magnification
system and the adjustment system.
d. Koehler Illumination
For optimum results in light microscopy, it is crucial that the light path be set
properly before the light reaches the specimen. In 1893, German scientist
3. Magnification System – contains important parts and plays a vital role in the use of
the microscope.
The ocular (eyepiece) is a lens that magnifies the image formed by the
objective.
The usual magnification of the ocular is 10x.
Most microscopes have two oculars and are called binocular microscopes.
Some microscopes only have one ocular and are called monocular microscopes.
The magnification produced by the ocular, when multiplied by the
magnification produced by the objective, gives the total magnification of the
object being viewed.
There are usually three objectives on each microscope, with magnifying powers
of 10x, 40x, and 100x. They are mounted on the nosepiece - a pivot enabling a
quick change of the objectives. They are described or rated according to focal
length (inscribed on the outside of the objective). The focal length of a lens is
very close to the working distance. The greater the magnifying power of a lens,
the smaller the focal length or the working distance.
a. Low-Power Objective:
i. Is usually a 10x magnification lens.
4. Adjusting System
The body tube is the part of the microscope through which the light passes to
the ocular.
The tube length from the eyepiece to the objective lens is generally 160 mm.
This is the tube that actually conducts the image.
The adjustment system enables the body tube to move up or down for focusing
the objectives.
This system usually consists of two adjustments, one coarse and the other fine.
▪ Coarse adjustment - gives rapid movement over a wide range and is used to
obtain an approximate focus. This adjustment should only be used with the
10x objective.
▪ Fine adjustment - gives very slow movement over a limited range and is
used to obtain exact focus after prior coarse adjustment. This adjustment
should only be used with the 40x and 100x objectives.
A. Daily Maintenance
1. Clean all optical surfaces with lens cleaner and cotton-tipped applicators and/or lens
paper.
a. Eyepieces, objectives, and condenser should be cleaned using a cotton-tipped
applicator.
b. The light source should be cleaned using lens paper.
2. Protect the microscope with a cover (or trash can liner if other cover is unavailable)
when not in use.
3. Initial that “1” and “2” have been completed.
B. Monthly Maintenance
1. Clean all nonoptical surfaces with mild detergent, and rinse with warm water.
2. Check KOH solution for contamination.
a. If contaminated, replace with in-date KOH.
b. If transferring KOH from a stock bottle to a working bottle, label the
working bottle with name of reagent, % strength, and expiration date.
3. Check saline solution for contamination.
a. If contaminated, replace with in-date saline.
b. If transferring saline from a stock bottle to a working bottle, label the working
bottle with name of reagent, % strength, and expiration date.
4. Check the appearance of the pH paper according to the pH paper package insert.
5. Initial that all duties listed above have been completed.
A. Turn lamp voltage down and then off if the microscope has a rheostat.
C. Keep at least one extra bulb available in case the bulb in the microscope blows. Contact
your technical consultant for ordering information. Follow manufacturer’s instructions
for ordering and replacing bulbs, or check with your technical consultant.
D. If your microscope requires a halogen bulb, DO NOT TOUCH the new bulb with your
fingers or gloves. Install it with a lint-free tissue.
E. Avoid exposure to corrosive fumes, extreme heat or cold, or sudden drastic temperature
change. When moving from one temperature extreme to another, allow optical parts to
equilibrate until all moisture has evaporated.
F. If immersion oil is required, use only chemically inert, low fluorescent, PCB-free
immersion oil. Low viscosity is acceptable. DO NOT USE CEDAR WOOD OIL.
Different brands of immersion oil are incompatible and should not be mixed.
D. When moving from one objective to another, use fine adjustment knob to refocus.
F. The higher the magnification, the more light you will need.
G. Always clean oil from objective and/or condenser at the end of the day.
H. Be sure low power objective is in working position at the end of the day.
I. Cover the microscope when you have finished for the day.
J. Try not to place the microscope on the same work bench as a centrifuge to avoid the
microscope and specimen from vibrating. If they will be on the same bench, secure a
mat that absorbs vibrations to place under the microscope.
2. There are dark shadows in the field which move as the eyepiece is turned.
Problem Solution
The surface of the eyepiece has scratches. Replace the eyepiece.
The eyepiece is dirty. Clean the eyepiece.
IX. References
January 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
August 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
I. Principle
The direct microscopic examination of a saline suspension of material taken from the area of
inflammation can provide a quick presumptive diagnosis for Gardnerella vaginalis and
Candida albicans and/or a definitive diagnosis for Trichomonas vaginalis. The addition of
KOH to a portion of the suspension helps clear debris, allowing better visualization of C.
albicans (yeast).
A. Specimens from the human body may be infectious. Use universal precautions.
B. If the specimen does not meet the criteria for acceptability, collect a new sample.
C. Speculum
F. Cover slips 22 x 22 mm
G. Microscope with mechanical stage, low power (10x) and high power (40x)
objectives
B. Solutions should be clear. Observe for turbidity, and discard if cloudiness is noted.
VI. Calibration
A. Set or verify Koehler Illumination on the microscope each day of use. This may not
be possible on some microscopes due to illumination being permanently set at the
factory.
IX. Procedure
A. Preparation of slides
1. Prepare slide by placing 1 drop of normal saline on 1 end of the slide and
1 drop of KOH on the other end of the slide.
2. On the slide, mix each solution separately using a swab. Dip swab in saline
first, then in KOH, or use 2 different swabs.
X X X
X XX X
X X X
6. Move stage so the KOH specimen is under the high power objective. Examine
10 high power fields for yeast cells (other elements are usually destroyed by
X. Interpretation of Results
A. Trichomonas vaginalis
1. Characteristics of the discharge:
a. Yellowish to gray-green
b. Frothy
c. Copious
d. Foul odor
2. Microscopic description using 10x objective:
a. Tiny
b. Round
c. Transparent
d. The size of white cells
e. Rapid movements (jerks and loops)
3. Microscopic description using 40x objective:
a. Approximately 15 mm (10-20 mm)
b. Round
c. Ovoid
d. Globular
e. Oval nucleus
f. Motility: Whirls, jerks, turns, seeming to vibrate
g. Undulating membrane on 1 side only, like the fin of a fish, the main
impression is movement.
4. Flagella:
a. Four flagella at the anterior end
b. Whip-like
c. Very motile
d. Slightly larger than white cells
e. Can be easily identified by their undulating swimming motion
f. Distinguish the trichomonad from the white cell by using the 40x
objective to detect the beating flagella.
g. These organisms are called “friendly white cells” because the flagella
appear to wave.
C. Gardnerella vaginalis
1. Characteristics of the discharge:
a. Thin
b. Homogenous
c. Grayish
d. Adherent
e. Foul odor
2. Microscopic description:
a. Small
b. Thin
c. Gram-negative rod
d. Bacterium
e. Parasitizes the surface of vaginal epithelial cells
3. Microscopic description of “clue cells":
a. Squamous, vaginal epithelial cells covered with many bacteria, giving
them a stippled or granular appearance.
b. Borders are obscured or fuzzy because of the adherence of the bacteria.
c. So many organisms may attach to a vaginal epithelial cell that its entire
border is obscured.
d. Absence of a distinct border, rather than the granular appearance of the
cytoplasmic or nuclear portion, identifies the clue cell.
e. Look for sheets of epithelial cells that are studded with bacteria.
4. Other elements seen in the presence of clue cells:
a. On the saline preparation trichomonads may be seen with clue cells, but
mycelia and clue cells are almost never found together.
5. Diagnosis of Gardnerella:
a. At least 20% of the epithelial cells present must be clue cells to establish
the diagnosis of Gardnerella.
A. Reporting of Organisms
1. Trichomonas vaginalis: Observation of a single motile trichomonad
establishes a diagnosis.
a. When Trichomonas are observed, report as “Present”.
b. When no Trichomonas are observed, report as “Absent”.
Note: The use of signs (e.g. “-“, “+”, “0”) and words (e.g. “positive”,
“negative”, “zero”, “none”) must be avoided as they can be misinterpreted.
No documentation means test is not performed.
2. Yeast: Observation of yeast, pseudohyphae, hyphae or buds establishes a
diagnosis.
a. When yeast, pseudohyphae, hyphae or buds are observed, report as
“Present”.
b. When no yeast, pseudohyphae, hyphae or buds are observed, report as
“Absent”.
Note: The use of signs (e.g. “-“, “+”, “0”) and words (e.g. “positive”,
“negative”, “zero”, “none”) must be avoided as they can be misinterpreted.
No documentation means test is not performed.
3. Clue Cells: Epithelial cells obliterated with bacteria.
a. Clue cells are significant only when symptoms are present.
b. Twenty percent of total epithelial cells must be clued to establish bacterial
vaginosis.
c. When clue cells are observed, report as “Present”.
d. When no clue cells are observed, report as “Absent”.
Note: The use of signs (e.g. “-“, “+”, “0”) and words (e.g. “positive”,
“negative”, “zero”, “none”) must be avoided as they can be misinterpreted.
No documentation means test is not performed.
4. White Blood Cells (WBCs):
a. Must be enumerated and reported by indicating low and high numbers
determined per high power field (hpf).
b. WBCs ≥ 10/hpf is significant in vaginal wet mount observation.
c. WBC quantity is determined by the following method:
i. Ten high powered fields are viewed microscopically using the
systematic pattern of exam (See Section IX.C.5).
ii. The average total of the 10 high powered fields is determined (See
Section XII) and reported using the following range:
Rare = 0-1/hpf
Few = 2-5/hpf
Moderate = 6-20/hpf
Note: The range wordage (Rare, Few, Moderate, Many, Numerous) may
be used instead of the numbers. Do not use abbreviations. No
documentation means test is not performed.
5. "Amine Test” or “whiff test”- a tool (along with other test observations and
patient symptoms) to assist the clinician in establishing bacterial vaginosis.
a. The test is performed by placing KOH on the wet mount slide and noting
the presence or absence of a “fishy odor”.
b. When a fishy odor is detected, report as “Positive”.
c. When no fishy odor is detected, report as “Negative”.
Note: The use of signs (e.g. “-“, “+”, “0”) must be avoided as they can be
misinterpreted. No documentation means test is not performed.
6. Other findings that may be documented:
1. Bacteria: Quantitate as 1+, 2+, 3+ or 4+.
2. Red Blood Cells (RBCs)
a. When RBCs are observed, report as “RBCs”.
XII. Calculations
C/10 = A where
C = (Cellular total of 10 high power fields)
A = (Average cellular total /high power field)
Example: Craig uses the systematic pattern of exam to observe a wet mount slide. After
10 high power field observations, he notes the following:
HPF One = 2, HPF Two = 3, HPF Three = 1, HPF Four = 6, HPF Five = 1,
HPF Six = 1, HPF Seven = 2, HPF Eight = 1, HPF Nine = 3, HPF Ten = 3
C= 2+3+1+6+1+1+2+1+3+3
C = 23 Craig determines the total cellular elements.
A = 23/10 Craig determines the average total by dividing by 10 hpfs.
A = 2.3/hpf Craig determines the average to be 2.3/hpf.
Based on the range chart (See Section XI.4.c.ii), Craig’s average (2.3/hpf) is in the
“Few” “2-5/hpf” range. He may report his observation as “2-5/hpf” or “Few”.
XIII. Troubleshooting
A. If patient results do not agree with clinical symptoms, an error may have occurred.
B. Review the "Sources of Error" for specific problems which may affect testing and
attempt to determine the cause of the problem.
D. Double check the specimen identification to ensure that the testing is being
performed on a properly collected and labeled specimen on the correct patient.
E. Compare the procedure as you performed it with each step in the written procedure.
Make sure the written procedure was followed exactly as it is written.
H. Does the testing personnel have adequate training and experience to properly
perform the procedure?
A. Specimen errors
B. Reagent errors
C. Equipment errors
D. Procedure Errors
1. Exam not performed using a thorough and systematic review of the slides.
A. Once the potential source of the problem is discovered, take proper actions such as:
1. Collect new specimen.
2. Use new reagents/materials.
3. Correct the equipment problems or replace the equipment.
4. Follow written procedure exactly.
5. Change testing personnel until additional training can be conducted.
B. Retest the patient specimens and reevaluate results to determine if they are
acceptable.
E. Once the problem has been resolved, report patient results on CHR-11 or CHR-12c,
including time, date, and initials of analyst.
B. A normal patient may have 1-3 WBCs and/or epithelial cells per high power field.
C. Few, if any, WBCs are usually seen with Gardnerella vaginalis. If increased
numbers of WBCs are seen with clue cells, consider other infections.
Not applicable.
Take care in interpreting apparent results: artifacts are common in KOH preps as a result
of cell degeneration, air bubbles, crystallization, and glycerol.
C. If the problem cannot be resolved, reschedule the patient to return at a later date or
send the patient to another location equipped to perform the examination.
XX. References
The rapid plasma reagin (RPR) 18-mm circle card test is a macroscopic, nontreponemal
flocculation card test used to screen for syphilis. The antigen is prepared from a
modified Venereal Disease Research Laboratory (VDRL) antigen suspension
containing choline chloride to eliminate the need to heat inactivate serum,
ethylenediaminetetraacetic acid (EDTA) to enhance the stability of the suspension, and
finely divided charcoal products as a visualizing agent. In the test, the RPR antigen is
mixed with unheated or heated serum or with unheated plasma on a plastic-coated card.
The RPR test measures IgM and IgG antibodies to lipoidal material released from
damaged host cells as well as to lipoproteinlike material and possibly cardiolipin
released from the treponemes. The antilipoidal antibodies are produced not only as a
consequence of syphilis and other treponemal diseases but also in response to
nontreponemal diseases of an acute and chronic nature in which tissue damage occurs.
If antibodies are present, they combine with the lipid particles of the antigen causing
them to agglutinate. The charcoal particles coagglutinate with the antibodies and
appear as black clumps against the white card. If antibodies are not present, the test
mixture is uniformly gray. The test can be purchased in kit form or in component parts
from many commercial sources. Without some other evidence for the diagnosis of
syphilis, a reactive nontreponemal test does not confirm Treponema pallidum infection.
A. Specimen
1. Avoid accidental infection when collecting and processing samples by
observing universal precautions.
2. Serum and plasma are suitable specimens for the qualitative test; however
serum is preferred for the quantitative test. Test plasma samples within 48
hours of collection.
3. An acceptable specimen does not contain particulate matter that would
interfere with reading test results.
Note: Hemolysis may be caused by transporting blood in freezing or
extremely hot weather without proper insulation.
B. Collection
1. Serum - Collect whole blood into a clean, dry tube without an anticoagulant.
2. Plasma - Collect blood in a tube containing EDTA as an anticoagulant.
Completely fill the tube or collect blood until the vacuum in the collection
tube has been exhausted.
3. Label each specimen with patient identifier and date.
C. Handling Serum
D. Handling Plasma
1. Centrifuge the specimen at room temperature at 1500 - 2000 rpms for at least
5 minutes to sediment cellular elements.
2. Plasma may need to be retained in the original collection tube if the test is to
be performed immediately. If not, plasma should be removed from cellular
elements.
3. Store plasma specimens at 2-8°C, and test within 48 hours. Plasma samples
must be at room temperature (23-29°C or 73-85°F) at the time of testing.
4. Do not heat plasma.
5. Do not use plasma specimens for confirmatory treponemal tests.
F. Core blood and spinal fluid are unsuitable for RPR testing.
IV. Reagent, Supplies, and Equipment (See “Health Department RPR Reagent
Requisition”)
H. Humidifying cover.
A. Needles
1. Check the calibrated needle every day of use if the needle has been dropped,
wiped, or when control pattern is not met to ensure the delivery of the correct
volume of antigen suspension (60 drops +/-2 drops per ml; 17µl per drop).
2. Place the needle on a 1-ml syringe.
3. Fill the syringe or pipette with RPR antigen suspension.
4. Holding the syringe or pipette in a vertical position, count the number of
drops delivered in 0.5 ml.
5. The needle is correctly calibrated if 30 drops +/-1 drop is delivered in 0.5 ml.
6. Replace the needle if it does not meet this specification. Be sure to test the
calibration of the replacement needle.
B. Rotator
1. Speed - For rotators without a digital readout, the speed can be estimated by
counting the number of rotations per minute.
a. To count the rotations, place a pen or pencil next to the rotator and count
the number of times the rotator touches the pen or pencil in 60 seconds.
b. If the rotator is properly adjusted, the count should be 100 +/- 2.
c. The rotator’s speed should be calibrated each day of use.
2. Time - The rotator’s timer should be checked against another laboratory
timer or stopwatch quarterly. The rotator’s timer should be +/-15 seconds of
the set time.
3. Circumference - Check quarterly to make sure the rotator is producing a
circle with a diameter of 2 cm or ¾ inch.
a. Place a blank sheet of paper underneath the edge of the bottom of the
rotator.
b. With the rotator off, secure a pen or pencil next to the side of the
rotating platform, and position it so that the writing surface is touching
the piece of paper.
c. Turn the rotator on for several seconds, and let the pen or pencil “draw”
a circle on the piece of blank paper. This circle is the circumference of
the rotating action.
d. After stopping the machine from rotating, remove the paper and measure
with a ruler the diameter of the circle drawn on the paper. If the
diameter of the circle is not 2 cm or ¾ inches, call your technical
consultant.
A control card should be run each day patient specimens are tested. Controls will be
run before patient testing is performed. Control results must be recorded on the STD
Patient Log.
B. Daily Controls
1. Check room temperature. For accurate and reproducible test results, the RPR
card antigen suspension, controls, and test specimens must be at room
temperature (23-29°C or 73-85°F) when tests are performed.
2. At each routine test run, check the expiration date of the antigen.
3. Determine antigen suspension reactivity with control cards of graded
reactivity (reactive, minimally reactive, and nonreactive) which have been
reconstituted with distilled water.
4. Use only RPR antigen that reproduces the established reactivity pattern of
controls.
5. Date and initial.
C. Monthly Controls
1. Clean humidifier sponge.
2. Remove dust in and/or on the rotator.
3. Clean black rubber sheeting on top of oscillating plate.
4. Date and initial.
D. Quarterly Controls
1. Check rotator timer (8 min +/- 15 sec) and record data.
2. Check rotator circumference (2 cm or ¾ inch) and record data.
3. Date and initial.
IX. Procedure
B. Place a drop of serum or plasma onto an 18-mm circle of the RPR test card, using
a disposable Dispenstir.
C. Using the inverted Dispenstir (closed end), spread the serum or plasma to fill the
entire circle.
E. Holding the dispensing bottle and needle in a vertical position, dispense several
drops on the edge of the card to clear the needle of air.
I. Immediately remove the card from the rotator; briefly rotate and tilt the card by
hand (3 or 4 to-and-fro motions) to aid in differentiating nonreactive from
minimally reactive results.
J. Read the test reactions (without magnification) in the “wet” state under a high-
intensity incandescent lamp.
X. Interpretation of Results
B. A reactive RPR card test may suggest past or present infection with a pathogenic
treponeme; however, it may also be a false-positive reaction.
False-positive reactions can result from laboratory error as well as serum
antibodies unrelated to syphilis infection.
Technical errors are detected by a nonreactive RPR card test with a second
serum specimen.
False-positive RPR card tests from infections with a nontreponemal disease or
other disease conditions are identified by an accompanying nonreactive
treponemal test.
C. A nonreactive RPR card test without clinical evidence of syphilis may suggest no
current infection or an effectively treated infection.
A nonreactive RPR card test without clinical evidence of syphilis can be seen
in early primary syphilis, in secondary syphilis, as a result of the prozone
reaction, and in some cases of late syphilis.
A nonreactive card test result does not rule out an incubating syphilis
infection.
XII. Calculations
Not applicable.
XIII. Troubleshooting
A. If patient results do not agree with clinical symptoms, an error may have occurred.
C. Review the "Sources of Error" for specific problems which may affect testing, and
attempt to determine the cause of the problem.
D. If the cause of the problem is not easily determined, follow a systematic approach to
troubleshoot the situation.
E. Double check the specimen identification to ensure that the testing is being
performed on a properly collected and labeled specimen on the correct patient.
F. Compare the procedure as you performed it with each step in the written procedure.
Make sure the written procedure was followed exactly as it is written.
I. Does the testing personnel have adequate training and experience to properly
perform the procedure?
A. If the temperatures of the sera, reagents, or testing area are less than 23°C (73°F),
test reactivity decreases; if temperatures are greater than 29°C (85°F), test reactivity
increases.
B. If the speed of the mechanical rotator is too fast or slow, improper antigen-antibody
interaction will cause unpredictable results.
D. If the card is excessively rotated and tilted by hand after removal from the rotator, a
false reactive may occur.
F. If the antigen is outdated or not adequately tested for standard reactivity, the results
may be inaccurate.
G. If the serum is unevenly spread in the circle, the antigen and antibody may not mix
properly.
I. If the moistened humidifying cover is not used to cover tests as they are being
rotated, proper humidity will not be maintained, and test components may dry on
the card causing false reactive results.
A. When you think the cause of the problem is discovered, take proper actions such as:
1. Collect new specimen.
2. Use new reagents/materials.
3. Correct equipment problems or replace the equipment.
4. Follow written procedure exactly.
5. Change testing personnel until additional training can be conducted.
B. Retest the patient specimens, and reevaluate to determine if results are acceptable.
E. Document the problem and the corrective action on the appropriate laboratory
record form.
Not applicable.
C. The RPR card test may be reactive in persons from areas where yaws, pinta or
nonvenereal syphilis is endemic. Generally, residual titer from these infections will
be < 1:8.
E. Nontreponemal test titers of persons who have been treated in latent or late stages
of syphilis or who have become reinfected do not decrease as rapidly as those of
persons in the early stages of their first infection. In fact, these persons may remain
“serofast,” retaining a low-level reactive titer for life.
XX. Reference
Person Certified to
Perform RPR Testing:
Health Department::
Street Phone
Date Requested:
ANTIGEN (3/BX)
DISPENSTIRS (500/BX)
Please use this form when ordering RPR supplies. Individual components should be ordered as
needed. “Whole” kits are no longer in stock.
Please do not phone in orders unless absolutely necessary. Use of this order form will ensure
accuracy and expediency in filling of requisitions.
ADPH/BCL/QM 12-11
STD Patient Log and QC Form
County Date / /
Site Page of
ADPH/BCL/QM/REV 04-10
RPR Rotator Preventive Maintenance
CHD
Site
Instructions: Year
1. Write in actual value, if applicable
2. If no value, use a check mark.
3. Record date and initials when performed.
MONTHLY JAN FEB MAR APR MAY JUNE JULY AUG SEPT OCT NOV DEC
Clean Humidifier Sponge
Remove Dust in or on
the Rotator
Clean Black Sheeting on
Oscillating Plate
Date
Initials
QUARTERLY FIRST QUARTER SECOND QUARTER THIRD QUARTER FOURTH QUARTER
Check Rotator time
(8 min +/-- 15 sec)
Check Rotator Circumference
(2 cm or ¾ inch)
Date
Initials
I. Principle
C. The single-reflecting condenser contains one reflecting surface which does not produce
a sharp focusing of the hollow cone of rays. This characteristic makes it easier to
manipulate, but less intense illumination is produced. Thus, the single-reflecting
condenser is less desirable when high intensity of illumination is required.
D. Most darkfield condensers require the numerical aperture (NA) of the oil immersion
objective be reduced below that of the condenser. This can be accomplished by
inserting a funnel stop in the objective or by using the oil immersion objective with a
built-in iris diaphragm.
A. Aperture diaphragm - either a rotating disk or an iris diaphragm on the condenser used
to direct a cone of light to the specimen and enter the objective. It should never be used
to regulate brightness. Resolution, control, and depth of field depend on the correct
setting of the aperture diaphragm.
E. Condenser - the lens system beneath the microscope stage positioned to concentrate
light correctly on the specimen and direct light rays into the objective. When the
condenser is used at a lowered position, the resolving power is reduced.
J. Immersion oil - oil with the same refractive index as glass, 1.515; used between the
cover glass and an oil immersion objective to prevent scattering of light in air.
K. Interpupillary distance - distance between the eyes. The eyepieces of a binocular scope
must be adjusted so that left and right images merge into one.
O. Parcentric - the ability to center a specimen in the field of view for one objective and
have almost the same field in place when rotating to another objective.
P. Parfocal - the objectives are constructed so that only slight refocusing with the fine
adjustment knob is needed after rotating to another objective.
Q. Resolution - the ability of a microscope to reveal fine detail in a specimen. The better
the resolving power of a microscope, the closer two objects can be and still be
distinguished as two objects.
S. Working distance - distance between the coverslip of a slide and the tip of an objective.
The low power objective has the greatest working distance. The oil immersion
objective has a very small working distance.
Always complete the microscope adjustment and have the microscope in satisfactory
working condition BEFORE collecting a darkfield specimen for examination.
Note: To complete the microscope adjustment and to verify that the microscope is in good
working order before examining patient material, prepare a suspension of gingival
scrapings in a drop of saline on a slide of proper thickness, mount with a coverslip, and use
this as a control slide.
C. Lower the substage slightly and place 2-3 drops of immersion oil on the top of the
condenser.
D. Center the specimen over the condenser with the mechanical slide carrier.
E. Slowly raise the substage until complete oil contact between the top of the condenser
and the bottom of the slide occurs.
F. Rotate the nosepiece to center the 10x objective over the specimen.
G. Bring the specimen into focus by using the coarse adjustment knob.
H. At this point, center the light in the field by rotating the 2 centering screws at the base
of the darkfield condenser.
I. Rotate the nosepiece until the high-dry (40x) objective is in place over the specimen.
J. Bring the specimen into focus by using the fine adjustment knob only.
K. If a satisfactory image is obtained, place a small drop of immersion oil on the cover
glass.
L. Rotate the nosepiece until the oil immersion objective (100x) is in place over the
specimen and is in contact with the oil on the cover glass.
M. Bring the specimen into focus by using the fine adjustment knob only. (Use of the
coarse adjustment knob with the high-dry or oil immersion objectives may cause
damage to the objectives by allowing them to come in contact with the specimen slide).
IV. Maintenance
See “Microscope Maintenance Chart”
A. Daily Maintenance
1. Clean all optical surfaces using lens cleaner and lens paper.
2. Eyepieces and objectives
a. Dampen a cotton-tipped applicator with lens cleaner.
b. Begin in the middle of the eyepiece or objective, and work in a circular motion
toward the outer edges.
c. Lightly touch the surface with lens paper to dry it.
3. Condenser and illuminator
a. Dampen lens paper with lens cleaner and gently clean the surface.
b. Use a piece of dry lens paper to dry the surface.
B. Monthly Maintenance
1. Clean all non-optical surfaces with warm water and mild detergent.
2. Please indicate N/A on the maintenance chart for the checking of KOH.
3. Check saline for cloudiness or contamination.
4. Document the maintenance on the chart.
5. Document the initials of the person performing maintenance.
A. Turn lamp voltage down and then off if the microscope has a rheostat.
B. If slideways and gears on the mechanical stage become difficult to move, lubricate with
machine oil or light grease. Document in the corrective action section of the
maintenance chart.
C. Keep at least one extra bulb available in case the bulb in the microscope blows.
Contact your technical consultant for ordering information. Follow manufacturer’s
instructions for ordering and replacing bulbs, or check with your technical consultant.
D. If the microscope requires a halogen lamp, DO NOT TOUCH the new bulb with your
fingers or gloves. Hold the lamp with a lint-free tissue to install it.
E. Avoid exposure to corrosive fumes, extreme heat or cold, or sudden drastic temperature
change. When moving from one temperature extreme to another, allow optical parts to
equilibrate until all moisture is evaporated.
D. When moving from one objective to another, use fine adjustment knob to refocus.
G. Always clean the oil from the objective and/or condenser at the end of the day.
H. Be sure low power objective is in working position at the end of the day.
I. Cover the scope when you have finished for the day.
VIII. References
January 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
August 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
A. Specimen
1. The ideal specimen for darkfield examination is a serous fluid rich in T.
pallidum but contains few blood cells (treponemes may be obscured if many
cells are present).
2. Consider every genital lesion in sexually active patients as syphilis until
subjected to a darkfield examination and proven otherwise. Other lesions on
the skin or mucous membranes should be examined when syphilis is
suspected.
3. Darkfield examination of oral lesions is not recommended. All positive
darkfield tests from mouth specimens must be confirmed by a direct
fluorescent antibody test. The indigenous flora of the oral cavity frequently
contain a spiral organism, T. denticola, which is indistinguishable from T.
pallidum.
4. If topical antimicrobial therapy has been applied to a syphilitic lesion, it may
not be possible to demonstrate motile T. pallidum even if several specimens
are examined.
B. Collection
1. Lesions
a. Remove any scab or crust covering the lesion.
b. Secondary infection exudate, if any, should be removed with a sterile
gauze sponge.
c. If necessary, compress the base of the lesion or apply a suction cup to it
C. Handling
1. Label slide with patient identifier.
2. Examine the slide within 5-20 minutes of collection, either by bringing the
patient to the microscope or the microscope to the patient.
Note: Any appreciable delay in examining a specimen may result in questionable
findings because the motility of the treponemes may be reduced or completely
lost.
A. Reagents
1. Saline, physiological (0.85%), sterile
2. Disinfectant (70% alcohol or iodine solution on swabs)
B. Equipment
1. Microscope assembly containing:
a. Condenser – darkfield, oil immersion condenser; single- or double-
reflecting type
b. Substage - a rack-and-pinion focusing substage for holding a darkfield
condenser
c. Microscope stand with course and fine adjustment knobs and revolving
nosepiece for 3 objectives
d. Body - inclined monocular or binocular type
e. Stage - plain stage with an attachable graduated or ungraduated
mechanical slide carrier
f. Objectives - parfocal
10x - low-power used to focus on specimen and center condenser;
NOT used to search specimen
40x - 45x - high-power used to search the specimen
90x – 100x – oil immersion fitted with a funnel stop or built-in iris
diaphragm used for final identification of organisms
g. Oculars - 10x
h. Illuminator - should be built into the base of the microscope
It should not be attached to the darkfield condenser because the heat
generated may cause complete loss of a critical identifying criterion
such as motility.
The built-in base illuminator should consist of a 6.0- to 6.5-volt or
equivalent, high intensity lamp with a variable transformer for
regulating light intensity.
If a separate external illuminator is used for the microscope, it should
be equipped with an iris diaphragm and a 100-watt lamp which
requires the microscope to have a flat-surface mirror for reflecting the
light into the darkfield condenser.
i. Eye shields - can be obtained for the oculars of some binocular
microscope models, thus eliminating the necessity of having to work in a
darkened room.
2. Microscope slides - 1 x 3 inches (clean and free of scratches)
a. Using slides of the correct thickness is very important.
b. The thickness required by American-made microscopes is usually
engraved on the top of the darkfield condensers.
c. For foreign-made microscopes, refer to the manufacturer’s literature.
3. Cover glass - size number 1, 22 x 22 mm (clean and free of scratches)
4. Oil - immersion, nondrying, Cargille type A (Cargille code 1248, R. P.
Cargille, Inc., Cedar Grove, N.J.) or equivalent
VI. Calibration
A. To complete the microscope adjustment and verify that the microscope is in good
working order before examining the patient material, prepare a suspension of gingival
scrapings in a drop of saline on a slide of proper thickness, then mount with a cover
glass. Examine under oil immersion in a systematic manner.
X X X
X XX X
X X X
IX. Procedure
A. The earliest and empirically most specific means of diagnosing syphilis (if yaws,
bejel, and pinta are excluded) is by darkfield microscopy. The demonstration of
treponemes with characteristic morphology and motility for T. pallidum
constitutes a positive diagnosis of syphilis in primary, secondary, or early
congenital stages, whatever the outcome of serologic testing. False-positive
darkfield tests may occur with oral specimens; therefore, such positive specimens
must be confirmed by direct fluorescent antibody tests specific for identification
of T. pallidum. When patients with untreated primary syphilis are positive by
darkfield microscopy but are serologically nonreactive, they usually become
serologically reactive within several days to several weeks. In other stages, the
patient should be seroreactive; if not, the darkfield interpretation and serologic
results should be analyzed to decide whether test results may be false positives or
false negatives, respectively.
Results Report
Organisms found that have characteristic Darkfield positive
morphology and motility of T. pallidum
XII. Calculations
XIII. Troubleshooting
A. Review the "Sources of Error" for specific problems which may affect testing, and
attempt to determine the cause of the problem.
B. If the cause of the problem is not easily determined, follow a systematic approach
to troubleshoot the situation.
C. Double check the specimen identification to ensure that the testing is being
performed on a properly collected and labeled specimen on the correct patient.
D. Compare the procedure as you performed it with each step in the written
procedure. Make sure the written procedure was followed exactly as it is written.
B. Microscopy errors
1. If immersion oil is not placed between the condenser and the slide or if
immersion oil with an incorrect refractory index is used, no light will reach
the specimen.
2. If slides of improper thickness are used, focusing on the lesion material and
illumination of the objects on the slide will be affected.
3. If, while using a paraboloid condenser, the concave side of the microscope
mirror is used with an external light source, the light intensity in the field of
view will decrease.
4. If the darkfield condenser is not properly centered, the object will be
illuminated poorly or not at all.
5. If the darkfield condenser is not properly focused, the most intense
illumination of the object will decrease.
6. If oil and dust are on the subsurface or reflecting area of the darkfield
condenser, the intensity of illumination of the object will decrease.
7. If the high numerical aperture of the oil-immersion objective is not
compensated with a funnel stop or an iris diaphragm, undiffracted direct light
will enter the oil objective.
8. If immersion oil is on the lens of the low-power (10x) or high-power (40x-
45x) objectives, the picture will be hazy without sharp definition.
9. If the specimen is inadequately illuminated, it will be impossible to
differentiate T. pallidum from other spiral organisms.
10. If the microscope is focused on the cover glass instead of on the specimen, a
false-negative report might be issued.
11. If the search of the specimen is inadequate or unmethodical, a false-negative
report might be issued.
A. When you think you have discovered the cause of the problem, take proper actions
such as:
1. Collect new specimen.
2. Use new reagents/materials.
3. Correct equipment problems or replace the equipment.
4. Follow written procedure exactly.
5. Change testing personnel until additional training can be conducted.
B. Retest the patient specimens and reevaluate to see if results are acceptable.
E. Document the problem and the corrective action on the appropriate laboratory
record form.
Not applicable.
A. Oral lesions at or near the gingival margin are unsatisfactory for darkfield
examination, as the indigenous flora in this area frequently contain a spiral
organism indistinguishable from T. pallidum.
XX. References
ADPH/BCL/QM/REV 04-10