EuroMedical Instruments EuroProbe Quick Start Guide

1. The quick start guide outlines 3 steps to set up the Gamma module: plug in the power cord, connect probes with the connector mark facing up, and switch on the device by pressing the button. 2. Connecting the Gamma module to the Fluorescence module involves placing the Gamma module on top, connecting the two with a cable, and connecting a dual Gamma Fluo probe before starting up. 3. To use both modules, switch between Gamma detection for Tc99 and Fluorescence detection by pressing the pedal for 3 seconds.

Uploaded by

Julio Vellame
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
156 views

EuroMedical Instruments EuroProbe Quick Start Guide

1. The quick start guide outlines 3 steps to set up the Gamma module: plug in the power cord, connect probes with the connector mark facing up, and switch on the device by pressing the button. 2. Connecting the Gamma module to the Fluorescence module involves placing the Gamma module on top, connecting the two with a cable, and connecting a dual Gamma Fluo probe before starting up. 3. To use both modules, switch between Gamma detection for Tc99 and Fluorescence detection by pressing the pedal for 3 seconds.

Uploaded by

Julio Vellame
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Quick start guide of the Gamma module in 3 steps

1 – Plug the power cord in the socket at the


back of the module

2 - Connect the probe(s) before switching the module


on, with the connector mark up **
Automatic recognition of probes. After switching on the
module, the green LED will indicate which probe is in operation

3 - Switch on the device by pressing the switch


button. By Default: Tc99 selected, Sensitivity scale set to
1000, integration time set to 1 second, sound set to BIP

Adjusting sound volume: Press the knob to turn the LED on, and turn to
desired level.
Setting sensitivity scale: Press the knob to turn the LED on, and turn to
desired level
Example for 50 MBq of Tc99:
Injection the day before: Set sound scale at 100 or 250
Injection in the morning: Set sound scale at 500 or 1000

Set-up choices before or during use

Isotope selection: Tc99 - Press to select


another isotope
Counting time: 10’’ - Press to select another
time variable
Integration time: 1’’ - Press to select another
integration time
Select config: Press to select the
configuration (No. 0 by default)
Sound mode: BIP - Press to select FM

WARNING: TURN OFF DEVICE BEFORE CONNECTING OR DISCONNECTING


PROBES
Warning: This Quick Start Guide* is not substitute to the user manual.

* Valid only for the latest EUROPROBE 3.2 version of the module marketed since December 2014.
Getting Started with the Fluorescence Module

1 - Place the Gamma module on the


Fluorescence module -
Fix the Velcro then connect the two modules
using the cable according to the color code

2 - Connect the DUAL Gamma Fluo probe


before start-up**

3 - Connect the pedal to the device

4 - Switch on the device by pressing the


button. Default: Tc99

Use: Switch from Gamma detection (Tc99) to Fluorescence detection (Opto) by


pressing on the pedal for 3 seconds.

Contacts
Sales contact
EuroMedical Instruments
BP 241 – 3 rue de l’arrivée
Office 1018 bis
75749 Paris Cedex 15- France
Tel : + 33 9 83 62 63 17
Fax : + 33 9 83 82 63 17
[email protected]
www.em-instruments.com

Manufacturer
Eurorad
2 rue Ettore Bugatti
67201 Eckbolsheim - France
Tel : +33 3 88 26 81 30
Fax : +33 3 88 28 45 48
www.eurorad.com
[email protected]

**All probes must be disinfected and placed in a sterile sleeve before use