Manual Cs844completo
Manual Cs844completo
Carbon/Sulfur Analyzer
with Cornerstone® Brand Software
Instruction Manual
CS844/CS844ES/C844/S844/S844ES
Version 2.9x
Part Number 200-784
February 2020
Original Instructions
Quality at LECO Corporation means understanding our customer’s
requirements and establishing specifications that clearly define those
requirements. The essence of our quality philosophy is the commitment to
quality objectives, aimed at never-ending improvement and complete
customer satisfaction.
Safety Symbols
These symbols may be found on LECO equipment or their components.
Some of the symbols shown may not appear in this manual. These
symbols indicate the use of specific safety guidelines. Important safety
information is highlighted in this manual by one of the following symbols,
as well as by WARNING and CAUTION statements. Operator and service
personnel must follow these instructions for personal safety and to
prevent damage to the instrument.
CAUTION
EXPLOSION HAZARD
PINCH HAZARD
PINCH HAZARD
PINCH HAZARD
PINCH HAZARD
FIRE HAZARD
ROTATING FAN BLADES
EXPLOSIVE
FLAMMABLE
OXIDIZING
COMPRESSED GAS
CORROSIVE
TOXIC
HARMFUL
HEALTH HAZARD
ENVIRONMENTAL HAZARD
PROTECTIVE GLOVES
PROTECTIVE EYEWEAR
EAR PROTECTION
NOT ALLOWED
Cautions, Warnings, and Residual Risks
CAUTIONS
· This instrument should be operated only by technically qualified and trained
individuals who have fully read and understood these instructions.
· It is highly recommended that users attend training courses provided by LECO
Corp. prior to operating this equipment.
· This instrument should be operated only in accordance with these instructions.
· The operator should follow all of the warnings and cautions set forth in this
manual and employ all applicable standard laboratory safety procedures.
· If this equipment is used in a manner not specified by the manufacturer, the
protection measures provided by the equipment may be impaired.
· Maintenance is to be completed by a Responsible Body: a person who has the
proper training and knowledge to perform the task safely.
· Consumables and replacement parts other than those specified and provided by
the manufacturer may cause damage to the instrument as well as impair the
protection measures provided by the equipment.
· Instruments can be very heavy. Proper equipment must be used to safely
transport, position, and lift equipment (if necessary).
· Front panels and other cosmetic parts of this instrument are not designed to be
weight bearing. DO NOT use such parts as lifting points, or damage may result.
ALWAYS lift near the feet at the sides or the rear of the instrument base.
Because sound pressure levels can be accumulative and dependent on many factors,
including other equipment operating in the vicinity of a workstation, it is
recommended by some authorities that a Time Weighted Sound Pressure Level be
measured or calculated at the operator's position after installation to assess the
possibility of the need for Personal Protective Equipment (PPE) such as ear
protection.
INSTRUMENTS WITH ACCESSIBLE HIGH TEMPERATURES
RISK OF FIRE
· Empty spent crucible bucket into a nonflammable, heat-resistant container
only.
· Dispose of hot materials into a fireproof container only.
· Do not attempt to permanently dispose of spent crucibles or any hot materials
until they have cooled to a safe temperature.
INSTRUMENTS WITH GLASS TUBES
RISK OF INJURY
· Wear protective gloves and eye protection whenever handling glass tubes to
prevent injury. Other Personal Protection Equipment (PPE) may also be
required.
RISK OF INJURY
· Visually inspect all glass tubing before installation or repacking. Do not use, and
properly dispose of, any glass tubing that has cracks, chips, or scarring.
Damaged tubes could possibly rupture under pressure.
· Depressurization of the Incoming Carrier Gas may be necessary before removal
of glass tubes to prevent injury or damage to the tubes during the removal
process. Turn Off the gas in the software and wait approximately 1 minute
before attempting to remove any tubes.
The gaseous and solid productions of this analytical instrument may be toxic relative
to the sample material analyzed. Normal laboratory procedures and regulations for
handling such materials should be followed. Please refer to the Safety Data Sheet
(SDS) of the sample material for further information regarding these potential
hazards, and follow the environmental protection measures and regulations
applicable to your region.
· Be aware of all hazardous chemicals, solvents, accelerants, oxidizers, cleaners,
reagent materials, etc., used by this instrument and its methods. Not all
instruments use hazardous chemicals or materials.
· Obtain Safety Data Sheets (SDS) for all hazardous chemicals and materials
used, and store them near where they are used. Users must familiarize
themselves with all warnings, precautions, and required Safety Procedures for
the handling, use, and disposal of such materials.
· Safety information and warnings may also appear on the chemical bottles or
containers.
· Wear proper Personal Protective Equipment (PPE) if required.
· Avoid vapors and direct contact with chemicals.
1 Introduction
The Introduction chapter contains general information about the
CS844, C844, and S844 Carbon/Sulfur Analyzers, including safety
guidelines and warranty terms. Reference this chapter for replacement
parts, operating supplies, and optional accessories. To place an order
by phone, call our customer service department toll-free in the United
States at 1-800-292-6141 or 269-985-5496. Orders may also be sent
by fax to 269-982-8977.
The screenshot illustrations used throughout the manual are examples
only and may or may not be applicable to specific procedures.
1 Introduction.................................................................................... 1–9
Illustrations ................................................................................ 1–13
Software License ......................................................................... 1–19
Warranty.................................................................................... 1–23
WEEE......................................................................................... 1–24
LECO-Supplied Hewlett-Packard® Computers .................................. 1–29
LECO-Supplied Balances............................................................... 1–30
Declaration of Conformity Statement ............................................. 1–31
Product Safety ............................................................................ 1–32
Electromagnetic Compatibility Notices............................................ 1–33
Radio-Frequency Disturbance ....................................................... 1–35
Flicker and Harmonic Emissions .................................................... 1–36
About this Manual ....................................................................... 1–37
Equipment Packages.................................................................... 1–38
Options ...................................................................................... 1–43
Components and Accessories ........................................................ 1–48
Specifications ............................................................................. 1–54
2 Installations.................................................................................... 2–1
Illustrations .................................................................................. 2–3
Lifting and Moving the Instrument................................................... 2–5
Instrument Front Panel Overview CS844/C844 ................................. 2–6
Instrument Front Panel Overview S844 ............................................ 2–8
I. Definitions
"Licensed Software" shall mean the software supplied with this
Software License, which is software developed and distributed by
LECO.
"Instruction Manual" shall mean a guide describing the use and
operation of the Licensed Software, together with any related
supporting documentation.
"Licensee" shall mean and include any of the Licensee's divisions and
departments within Licensee's organization, but shall not include
wholly or partially owned subsidiaries, affiliates, successors or assigns,
or independent third parties.
VII. Confidentiality
All information and materials regarding the software, provided to
Licensee by LECO which are not generally known in the trade are
confidential, are proprietary, and such information and materials are
furnished solely for the use of Licensee under this Software License.
Licensee agrees not to use or disclose any such information or material
otherwise without prior written consent from LECO and Licensee
further agrees to take all reasonable steps requested by LECO, or
otherwise necessary to protect the confidentiality of all such
information and materials. The foregoing obligations will survive
termination of this Software License, and Licensee agrees to return all
such information and materials to LECO upon request or any non-
renewal or termination of this Software License.
VIII. Miscellaneous
a) This Software License shall be binding upon and inure to the
benefit of the parties hereto and their respective successors and
assigns, provided that Licensee shall not have any right to assign
its rights or obligations under this Software License (whether by
operation of law or otherwise) without the prior written consent
of LECO.
b) All notices hereunder shall be in writing and shall be sent by
registered or certified mail postage prepaid to LECO at the
addresses first written above. The date of postmark shall be
deemed the date on which such notice is given.
Notice
The warranty on LECO equipment remains valid only when genuine
LECO replacement parts are employed. Since LECO has no control over
the quality or purity of consumable products not manufactured by LECO, the
specifications for accuracy of results using LECO instruments are not
guaranteed unless genuine LECO consumables are employed in conjunction
with LECO instruments. If purchaser defaults in making payment for any parts
or equipment, this warranty shall be void and shall not apply to such parts
and equipment. No late payment or cure of default in payment shall extend
the warranty period provided herein.
LECO Corporation is not responsible for damage to any associated
instruments, equipment, or apparatus nor will LECO be held liable for loss of
profit or other special damages resulting from abuse, neglect, or use in
disregard of instructions. The Buyer, their employees, agents, and successors
in interest assume all risks and liabilities for the operation, use, and/or
misuse of the product(s) described herein and agree to indemnify, hold
harmless, and defend the seller from any and all claims and actions arising
from any cause whatsoever, including seller's negligence for personal injury
incurred in connection with the use of said product(s) and any and all
damages proximately resulting therefrom.
For the above reasons, LECO expects end users to dispose of the
material in an environmentally friendly way, being separate collection
and treatment.
Hewlett-Packard Support
All LECO-supplied Hewlett-Packard computers include HP® customer
technical support and warranty claim information.
The inclusion of HP customer technical support and product warranty
with LECO-supplied HP Computers ensures that any computer-related
service issues are handled directly by the experts at HP. This
eliminates the unnecessary step of working through the LECO service
professionals for a resolution from HP.
For more information, call 1-866-625-1175, or access the HP website
by following the link below and selecting the appropriate country and
support language:
http://www8.hp.com/us/en/contact-hp/ww-contact-us.html
To expedite service, please have your HP computer serial number and
model number available when contacting HP. These can be found on
the back or side of the computer tower.
Sartorius® Support
All LECO-supplied Sartorius balances are supported and covered by
warranty through Sartorius.
The direct support and warranty coverage by Sartorius with LECO-
supplied Sartorius balances ensures that any balance-related service
issues are handled directly by the experts at Sartorius. This eliminates
the unnecessary step of working through the LECO service
professionals for a resolution from Sartorius.
For more information, access the Sartorius website by following the
link below and selecting the appropriate country:
http://www.sartorius.com/en/contact/local-contact/
To expedite service, please have your Sartorius balance serial number
and model number available when contacting Sartorius.
The following information sets out the content of the EU-DoC, including
a list of EU Directives, harmonized standards, supporting standards
and other applicable documents.
International
IEC 61010-1 Safety requirements for electrical equipment for
measurement, control, and laboratory use.
USA
UL 61010-1 Safety requirements for electrical equipment for
measurement, control, and laboratory use.
Canada
CAN/CSA-C22.2 No. 61010 Safety requirements for electrical
equipment for measurement, control, and laboratory use.
USA
Canada
European Union
Definitions
This is a Class A Group 1 Product.
The screenshots used throughout the manual are examples only and
may or may not be applicable to specific procedures.
Items listed below are repeated throughout this manual and are
subject to revision. Consult the packing slip received with the
instrument.
* Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping restrictions. These items are essential for operation.
Contact your local LECO distributor for quotation and delivery.
* Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping restrictions. These items are essential for operation.
Contact your local LECO distributor for quotation and delivery.
* Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping restrictions. These items are essential for operation.
Contact your local LECO distributor for quotation and delivery.
* Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping restrictions. These items are essential for operation.
Contact your local LECO distributor for quotation and delivery.
* Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping restrictions. These items are essential for operation.
Contact your local LECO distributor for quotation and delivery.
S844ES-MCR PKG S844ES TSCREEN/CR
1 259-078 SOFTWARE KIT C/CS/S SERIES
1 501-171-HAZ* ANHYDRONE 10-16 MESH 1/LB
1 618-300-500 ASSY S844ES SUL DET T/SCRN 230V
1 619-591-921 ASSY KEY REGISTRATION
1 619-593-359 PACK ACCESSORY S844ES
*Packed and shipped separately in the U.S.A. Not included with international
shipments due to shipping restrictions. These items are essential for operation.
Contact your local LECO distributor for quotation and delivery.
Precision†
Carbon:..............................................0.0003 mg or 0.5% RSD, whichever is
greater
Sulfur: ...............................................Standard Mode:0.0003 mg or 0.5% RSD,
whichever is greater
ES Mode: 0.00005 mg or 0.5% RSD,
whichever is greater
Calibration ..........................................Standards (single or multipoint); manual;
gas dose
Chemical Reagents
Anhydrous Magnesium Perchlorate (MgClO4)
Sodium Hydroxide on an Inert Base
Platinized Silica Gel
Rare Earth Copper Oxide
Cellulose
Gas Requirements
Carrier ...............................................Oxygen, 99.5% pure, 35 psi (2.41 bar)
±10%
Pneumatic ..........................................Compressed Air (oil, water free), 40 psi
(2.76 bar) ±10%
Dosing: CS/C844 .................................Carbon Dioxide, 99.99% pure, 20 psi
(1.38 bar) ±10%
Operating Conditions
Operating Temp ..................................15 to 35 °C (59 to 95 °F)
Rel. Humidity ......................................20 to 80% (non-condensing)
Refer to the Safety Data Sheet (SDS) for the specific chemical
for additional information.
CAUTION
Refer to Specifications, page 1–54, for the approximate
shipping and instrument weight. To lift or move the
instrument, use equipment capable of safely lifting this
weight.
CAUTION
Front panels and other cosmetic parts of this instrument are
not designed to be weight bearing. DO NOT use such parts as
lifting points, or damage may result. ALWAYS lift near the feet
at the sides or the rear of the instrument base.
Figure 2-1
Front Panel CS844/C844
Figure 2-2
Front Panel S844
Figure 2-3
Rear Panel
11
Ethernet
12 Dose B Pressure Relief
13 Dose A Pressure Relief
14 Pressure Monitor
15 Dose B In
16 Dose B Exhaust
17 Dose A In
18 Dose A Exhaust
19 Oxygen In
20 Oxygen Pressure Relief
21 Pneumatic Pressure Relief
22 Pneumatic Pressure In
23 Carrier Flow Exhaust
24 Touch Screen Power
25 Touch Screen
26 Touch Screen Video
CAUTION
Many substances that do not normally burn in air, as well as
other substances that are combustible in oxygen, may burn
violently when a higher percentage of oxygen is present.
OXIDIZING/PROTECTIVE EYEWEAR/PROTECTIVE
GLOVES
Oxygen is an oxidizer and is pressurized. It can become
volatile when subjected to heat or high temperatures.
Protective eyewear and gloves should be worn when handling
oxygen.
8. Connect the gas dose tubing assembly from the gas dose tank to Dose
A In and Dose B In on the rear of the instrument. A “T” assembly is
provided to connect both inlets if desired. Refer to Figure 2-3,
page 2–10.
A. Turn On the dose gas at the tank, and then set the regulator to
20 psi (1.4 bar).
B. If required, connect the Dose A Exhaust, Dose B Exhaust, and
pressure monitor, on the rear of the instrument, to the facility
exhaust.
The PRESSURE MONITOR should be used whenever the DOSE A
EXHAUST and DOSE B EXHAUST outlets are routed to an external vent.
This ensures that the pressure measured at the PRESSURE MONITOR
port is the same as the DOSE A EXHAUST and the DOSE B EXHAUST.
9. Install the oscillator tube and fill the coolant reservoir. Refer to
Installing the Oscillator Tube, page 2–16.
10. Install the bracket for the vacuum cleaner.
11. Install the vacuum cleaner. Refer to Installing the Vacuum Cleaner,
page 2–20.
12. Remove and pack the following reagent tubes according to the
procedures described in the Maintenance section.
· Catalyst Tube (heated), page 6–16: (CS/C844).
· Cellulose Tube (nonheated), page 6–19: (CS/C844)
· Incoming Purifier Tube (heated), page 6–12: (CS/C844).
· Incoming Scrubber Tube (nonheated), page 6–9: (CS/C844).
· Flow Scrubber Tube (nonheated), page 6–21: (CS/C/S844).
13. Reinstall the reagent tube(s) if not already completed.
14. Plug the instrument into facility AC power. Refer to Installing the
Electrical, page 2–18.
15. Turn On the power switch located on the left-side of the instrument.
Allow 1 to 2 minutes for the instrument to start.
16. Turn On the monitor.
17. Turn On the computer. If using a non-LECO PC, refer to Setting Up PC
for Use, page 2–31.
Figure 2-4
Pedestal Assembly
Analyzer
Power Input 230 VAC +10/-15% at max load, 50/60 Hz,
single phase, 25 A maximum
Line Cord (LECO P/N 619-590-705) NEMA L14-30 captive
line cord, 30 A/250 V, 10 AWG, 3 conductor,
8 ft 2 in (2.5 m)
Receptacle (LECO P/N 765-978) NEMA L14-30R, included
Outlet in instrument component pack
Receptacle (LECO P/N 608-898) 1- gang, included in
Cover instrument component pack
Figure 2-5
End View of Plug: NEMA L14-30P (30 A)
Computer (optional)
Power Input 115 VAC/230 VAC, 50/60 Hz, 240 W
Line Cord detachable with NEMA 5-15 plug and
C13 connector, included with computer
Monitor (optional LCD)
Power Input 100 VAC to 240 VAC, 50/60 Hz, 1.3 A,
C14 inlet connector
Line Cord detachable with NEMA 5-15 plug and
C13 connector, included with monitor
9. Attach one end of the second vacuum hose to the open port on the
vacuum manifold. Route the hose beneath the instrument and attach
the remaining end to the hose block located on the mechanical flow
controller access panel inside the left door.
10. Reinstall the lower right casting and stainless steel work surface.
11. Slide the vacuum manifold into the bracket on the rear of the
unit.
5. Locate the Ethernet cable and connect one end to the network
interface card on the rear of the computer. Connect the other end
CAUTION
For instruments with connectors on the back panel only: To
prevent damage to the device, use only the power supply that
was included with the touch screen monitor.
1. Unpack the monitor.
2. Remove the three screws that secure the cable access panel on
the back of the monitor. Refer to Figure 2-13, page 2–30.
Make sure that the spring-loaded stop pin is positioned toward the
top of the analyzer. Refer to Figure 2-12, page 2–29.
Figure 2-6
Monitor Post and Base
3. Using the 7/32-inch Allen wrench (hex wrench), tighten the
setscrew in the base to secure the post.
Figure 2-7
Post Collar/Tilt Mechanism Assembly
5. Attach the mounting plate to the tilt mechanism with the four
10-32 3/8-inch Phillips head screws supplied with the kit. Refer to
Figure 2-8, following.
Figure 2-8
Attach Mounting Plate
Figure 2-9
Vertical Monitor Installation
Figure 2-10
Horizontal Monitor Installation
Figure 2-11
Monitor Assembly
Figure 2-12
Touch Monitor Installation
Antivirus Software
LECO recommends that antivirus checking be turned off for the
following files to avoid interference with data collection or storage:
use the appropriate drive letter, typically C:\ProgramData\LECO
(hidden OS folder).
Windows Settings
To use LECO software, certain Windows settings are required to ensure
performance of the instrument. These settings are set up automatically
during software installation; however, problems may be encountered if
they are ever changed. The required Windows settings are listed in the
following table.
Power Options Required Value
Display Turn Off display after Never
Sleep Sleep after Never
Hard Disk Turn Off hard disk after 0 minutes
· Do not insert the Registration Key into the USB port until
prompted by the LECO software installer.
4. Select Uninstall.
Refer to the Safety Data Sheet (SDS) for the specific chemical
for additional information.
Figure 3-1
Software Overview Portrait Mode
—Tasks in progress
1. Open Task Center by selecting either icon from the right corner of
the title bar. The Task Center progress screen opens.
Ambient Gauges
The ambient gauges provide a quick status check on some of the most
critical system parameters.
Grid Current
Plate Current
horizontally .
The Plot Layout button is available only on instrument configurations
that analyze more than one analyte.
Exit
Exit the software and set the gas
Conserve
to a low flow conserve mode.
Gas
Exit
Exit the software and turn Off
Conserve
the gas.
Gas
Keyboard Shortcuts
Function key labels can be enabled or disabled from the Display
Settings screen of the software. Function keys are functional even
when the labels are not displayed in the software. Refer to Setting
Display Settings, page 4–5.
CAUTION
Aborting some sequences can leave the instrument in an
unsafe state.
The option box to configure the plot will vary depending on the type
of display and the instrument configuration.
· Select Curve to change the plot to a signal curve.
· Select Advanced to plot the signal curve with extra information,
such as y-axis values or furnace behavior. To view x- and y-axis
coordinates, select and hold on a trace. Refer to Configuring the
Display, page 4–5.
· Select Scatter and the results will appear as a scatter plot. Scatter
mode requires a minimum of two replicates to be displayed.
· To Zoom and magnify the plot, use one of the following methods.
o On single-touch displays: Select Zoom. Select and hold a
corner of the area of interest, and then slowly drag diagonally
toward the opposite corner of the area.
Importing
Import allows the user to restore exported data to the Cornerstone
brand software.
Copy Output
1. From the Analysis screen, select the set(s) to copy.
Export Output
1. From the Analysis screen, select the set(s) to export.
Cloning
1. Select the information to be cloned.
Using Help
· Use the outline to select a specific section of the manual. Select the
arrows to expand or collapse sections with subheadings.
and To fields.
Select a desired section of the manual from the outline, and the From
and To fields automatically populate with the correct page range for
that section.
2. Select Print.
Multi Field is a feature available on the Action bar that will edit
fields from multiple sets or replicates at once if the same edit is
desired. For example, the same description can be added to multiple
sets.
2. Check the flow scrubber and replace when the Anhydrone has
caked. Refer to Replacing the Flow Scrubber Tube, page 6–21.
3. Check the particle filter and replace when the filter has darkened
significantly. Refer to Replacing the Particle Filter, page 6–27.
4. Inspect the incoming scrubber and replace when the Anhydrone
has caked or when the LECOSORB® has turned color. Refer to
Replacing the Incoming Scrubber Tube, page 6–9.
5. Check the cellulose trap and replace when 1-inch of the cellulose
has turned black or if the cellulose has turned brown. Refer to
Replacing the Cellulose Trap Tube, page 6–19.
CAUTION
Do not touch the crucibles with bare hands after they have
been pre-baked in the tube furnace, even when they are cool.
Natural oils from your hands will contaminate the clean
crucibles.
Logging in a Blank
1. Enter the number of replicates for the blank set by adjusting the
Logging in a Standard
1. Enter the number of replicates for the sample adjusting the + or
Logging in a Sample
1. Enter the number of replicates for the sample by adjusting the +
2. Select Sample .
3. Enter a name for the sample.
4. For each rep, enter the mass of the standard, select Default: [X],
select Read Mass from Balance, or press the Print button on the
balance, and then select Enter.
5. Select Method, and then select a method name from the list.
Splitting
A replicate can be split off from a set and made into a new set of its
own. The Name, Method, Comments, and Description fields will be
copied into the new set. The comments associated with each sample
replicate will be retained.
1. Select the replicate(s) to be split.
Merging
To merge two or more sets into one, the Name and Method must
match. The Description and Comment fields do not have to match.
When sets are merged, sample replicate comments are retained. The
merged sample sets will have the same description as is associated
with the first sample set.
1. Select the sets to be merged.
select Delete .
Any unanalyzed sample with a mass, except for blanks, must be
excluded before it can be deleted.
Figure 3-3
.csv File Created in Excel
Figure 3-4
.csv File Created in Notepad
6. Select Exclude .
7. Select Save to save changes.
Include .
select Edit .
Include .
2. Select Calibrate .
3. Select the tab for the desired analyte.
4. Choose one of the following options as desired.
· To replace the existing calibration with a new
calibration: Select Don’t Use next to Existing Standards and
Use next to New Standards.
· To keep the existing calibration without adding any
new sets: Select Use next to Existing Standards and Don’t
Use next to New Standards.
· To add new sets to the existing calibration: Select Use
next to both Existing Standards and New Standards.
5. Sets and replicates can be included or excluded as desired. Select
the set or replicate, and then select Include or Exclude on the
Action bar. Refer to Including and Excluding, page 3–24.
6. To change the regression type and regression order complete the
following steps.
A. Select Edit. The Editing screen appears.
B. Under New Calibration, select the Regression Type drop-
down arrow, and then select the desired regression type.
C. Select Close to close the Editing screen.
7. Select Save to save changes.
CAUTION/PROTECTIVE EYEWEAR/PROTECTIVE
GLOVES
Carbon dioxide is a gas under pressure. It can become volatile
when subjected to heat or high temperatures. It is an
asphyxiant and may displace the oxygen in the room.
Protective gloves and eyewear should be worn at all times
when handling this material.
1. Attach the gas dose lines on the instrument. Pressurize the dose
gas regulator to 20 psi.
2. In order to purge air from the dose gas lines:
A. Select Diagnostics, and then select Digital I/O.
B. Activate the dose pneumatic valve that corresponds with the
attached lines.
C. Activate the gas dose that corresponds with the
attached lines.
D. Allow the lines to purge for at least 2 minutes.
E. Deactivate the gas dose that corresponds with the
attached lines.
F. Deactivate the dose pneumatic valve that corresponds with
the attached lines.
3. In the software, select Settings, and then select Standards.
The Gas Doser is available only on the following models: CS844 and
C844.
Gas dose loop volumes can be determined by performing a series of gas
doses using pure gases on a calibrated instrument. The analyte results, given
in % or ppm, are used to back calculate the dose loop volume through the
following equations.
1. Convert Analyte Element Composition ( ) to grams of Analyte Element
( ):
, or
where is the number of moles of dose gas per mole of analyte element.
For example, 1 mol of CO2 contains 2 moles of O2 and 1 mol of C.
1.
2.
3.
4.
Example 2.
Ten CO2 gas doses were performed using Gas Dose B and an average Oxygen
result of 410 ppm was obtained.
1.
2.
3.
4.
, where
2. Calculate the molar fraction (MF) of Analyte Gas in the Cylinder:
or
5. Convert to :
6. Convert to :
7. Convert to :
1.
2.
3.
4.
5.
6.
7.
Example 2.
A mixed gas cylinder of 1000 ppm CO2 in Argon is connected to the gas doser
and used for gas dosing. The expected Oxygen composition in % and ppm is
calculated as follows:
1. l
2.
3.
4.
5.
6.
7.
.
6. The newly added drift standard appears in the Standards Added
For Drift table.
7. Select Add, and then select Save. The newly added drift standard
will now be listed in the Manage Drift Standards table with the
Drift standards included in the original calibration. Refer to
Managing Drift Standards, following.
Using Next
The Next button is not available on all instrument configurations.
Aborting an Analysis
During sample analysis, the analysis can be stopped or aborted. To
stop an analysis, select Abort.
3. To delay the start of the analysis until a predefined time, select the
time/date button and enter the desired time. A green and black
Pause symbol will be displayed in the replicate orb to indicate that
it is a delayed paused.
Deleting a Sample
Refer to Excluding or Deleting a Sample, page 3–33, for more
information.
desired, select Multi on the Action bar, and then select the
set orbs next to the desired sets.
Filtering by Methods
1. Select the Method tab.
2. Select inside the Method box. A list of methods appears.
3. Select the desired method from the list, and then select Enter.
Filtering by Excluded
1. Select the Excluded tab.
Filtering by Description
1. Select the Description tab.
2. Select inside the Description box. The keyboard appears.
Filtering by Comments
1. Select the Comments tab.
2. Select inside the Comments box. The keyboard appears.
Filtering by Operator
1. Select the Operator tab.
2. Select inside the Operator box. The keyboard appears.
Use the following steps to determine how the analyte units display in
the software. If all uses of an analyte should use the same unit and
formatting, it is easiest to set this using the Analyte Units tab.
1. Select the desired sample display, and then select Edit. Analyte
Units will be automatically highlighted.
2. For units, select the Mass Ratio field. A keyboard appears.
3. Select the desired unit.
4. Select Enter.
5. Select the desired Rounding Method, which controls the method
used to determine the number of significant digits to display or
report. Refer to Cornerstone Brand Software Rounding, page
7–14, for more information.
6. Select Enter.
· Numeric settings will be displayed only if all numeric entries
selected are of the same type. Mass and Mass Ratio fields cannot
be mixed.
· Numeric settings set in the Analyte Units tab apply to all fields
unless they are overridden by a Style setting. To change the
Style refer to Editing Styles, page 4–11.
are gray .
Removing
1. Select any row, column, or field.
2. Select Remove.
Spanning Fields
A field may be stretched across multiple columns or rows using the
span feature. A field cannot be spanned without adjacent empty fields.
Span can be used in Single- and Multi-selection modes. Fields cannot
be spanned across a freeze bar.
1. Double-tap a field to select and enable Span.
Editing Styles
A style is a collection of settings that controls how information is
displayed. Styles can be edited for fields in Display, Reports, and
Transports. The style for each field is unique to each Display, Report or
Transport configuration.
· Repeatedly selecting Select All will first select all available fields,
then will select all fields for a single analyte.
4. Select Save .
Applying a Display
1. Select Settings, and then select Display.
2. For Current Sample Display, select the desired display.
3. Select Enter.
select Edit .
2. Select the Name field. A keyboard appears.
3. Enter a name for the method.
4. Enter a description if desired.
5. Enter the method parameters. Refer to Method Parameters, page
4–15.
Cloning a Method
1. Select the desired method.
Deleting Methods
The following section explains how to delete a method from the
software.
When a method is deleted, all associated items such as samples,
calibrations, and blanks that are referenced by that method will also
be permanently deleted.
General Parameters
Purge Time—The time before an analysis when the furnace is closed and
atmospheric impurities are purged from the combustion area with oxygen.
Delay Time—The amount of time between the purge time and analysis in
which the instrument establishes a stable baseline for detection.
Sample Cool Time—The time to wait after the analysis has completed
before lowering the furnace pedestal.
Furnace Mode—Selects the Furnace Mode during an analysis. Selections:
Constant or Ramped.
Analysis Mode is available for CS844ES and S844ES models only.
Calculations
Use this section to view, edit, or manage existing calibrations, drifts,
and blanks.
1. To view, select Calibration, Drift, Manage Drift Standards,
or Blank.
2. If changes are necessary, select Edit.
· For more information on calculating calibrations, refer to
Calculation of the Calibration Equation, page 7–10.
· For more information on calculating drifts, refer to Drift
Correction, page 7–9.
· For more information on managing drift standards, refer to
Managing Drift Standards, page 3–51.
· For more information on calculating blanks, refer to Blank,
page 7–7.
A user can select the set they would like to run “in control” and then
General
Select Enable to define standard blanks as part of the QC Method, and
then enter the desired number of Blanks to Analyze and Blanks to
Calculate.
Blanks to Calculate—This value indicates the number of replicates
from which to calculate the standard deviation. The replicates used will
be the last of the analyzed set, so this value should be less than the
number of Blanks to Analyze.
Conditioners
Enter the number of conditioning samples to run before and after the
QC blanks. Some instruments do not support pre-blank and/or post-
blank conditioning samples.
Schedule
Blanks can be required after a user defined time has elapsed.
Select Yes for After Time Lapse, and enter a time in the box to
schedule the blank analysis after a certain length of time.
QC Checks
The parameters controlling the analysis of check standards are found
in the method’s Quality Control Checks screen.
General
Only standards that have one or more elements with Check Standard
set to Yes in the Standards screen will be viewable within QC checks.
Select Enable in the include check standards as part of the QC routine.
Choose the minimum number of check replicates required to be run in
order for QC to be satisfied. Refer to the example in Check Standards,
page 4–22, for more information.
Schedule
Check standards can be required after different specified events to
confirm that the instrument is still providing expected results.
Select Yes for After Blanks to schedule a check after running blank
analyses.
Select Yes for After Calibrating to schedule a check after calibration.
Select Yes for After Drifting to schedule a check after running a drift
standard.
Check Standards
The specific standards to be analyzed are selected in this section and
the number of replicates to analyze applies to each selected standard.
Check marks to the right of the certified and uncertainty values
indicate which elements are set as Check Standards in the Standards
screen.
Example: If three check standards are selected and the number of
replicates to analyze is set to 2, then a total of six analyses will need
to be performed: two for each standard.
Once analyzed on the Analysis screen, select all the check standards
sets. Select the Check button on the action bar to reset the QC check
standards setting to be in control.
General
The inclusion of calibration standards into QC can be enabled or
disabled from this screen.
Choose the minimum number of calibration replicates required in order
for QC to be satisfied.
Schedule
Calibration standards can be required following the analysis of blanks,
or if a check standard failed to be within range.
Calibration Standards
Similarly to the QC Checks screen, the standards to use for the
calibration are specified this screen, and the number of replicates to
analyze applies to each selected standard. Refer to Check Standards,
page 4–22.
Check marks to the right of the calibration standards indicate that the
standard is required to be within the analyte’s upper and lower limits
for the calibration to be saved.
Make sure that the appropriate range is selected within Standards to
obtain good calibration curves.
On the Analysis screen, select all the calibration standards sets. Save
the new calibration in order to reset the QC calibration standards
setting to be in control. Refer to Performing Calibrations, page 3–38.
The Quality Control button on the action bar can be selected for
the software to automatically insert the necessary blank, check
standard, and calibration standard replicates as specified on the QC
method screens, before the user sample.
These QC replicates will be analyzed and at their completion, QC will
again determine if it is in control.
· If QC is in control, then the user’s sample analysis will be
performed.
· If QC is not in control, the necessary QC steps will continue to be
requested to be performed until QC is in control. Instrument
maintenance may be required.
Creating a Field
1. Select Settings, and then select Fields.
3. Enter a name for the new field, and enter a description if desired.
4. Select the Type.
Number—Provides a number.
Editing a Field
1. Select Settings, and then select Fields.
2. Select a field.
3. Select Edit, and then refer to Creating a Field, page 4–25, for
instructions on editing field parameters.
2. Add the multiply function from the Functions tab of the calculation
options.
3. Add Carbon from the Fields tab. Carbon is available only when
Replicate and Carbon are selected.
Functions
Numeric Functions
Input—Expression that evaluates to a number
Output—Numeric results of operation
Function Definition In Use
+ Plus
- Minus
x Multiply
÷ Divide
Pi (constant value of
π
3.14…)
Ln Natural log
Parentheses Function
The parentheses function is used to group elements of an equation
together to clarify which parts are evaluated first.
Function Definition In Use
() Parentheses
“If/then/else”
If/then/else
statement
Comparison Functions
Input—Expression that evaluates to a number
Output—Logical result of the operation (true or false)
Function Definition In Use
= Equal to
≠ Not equal to
Greater than or
>=
equal to
“And” logic
And
statement
“Not” logic
Not
statement
Text Function
This function lets the operator enter text into the equation editor.
Function Definition In Use
“If/then/else”
If/then/else
statement
Creating Reports
1. Select Settings, and then select Reports.
Editing Reports
1. Select Settings, and then select Reports.
2. Select a Report.
3. Select Edit, and then refer to Creating Reports, page 4–33, for
instructions to edit report parameters.
Select Fields:
A. The new field will appear at the end of the display area.
B. Drag the field to the desired place in the report.
C. To add a return arrow, select the existing return arrow (found
at the end of the line in the display area and move it to the
desired area.
D. To remove a return arrow, drag it to the end of the line or
section in the display area where there is already a return
arrow, and then release it The duplicate symbol appears.
There will always be a return arrow at the end of the section (Set
Begin, Rep, Set End).
Plots Tab
1. Select Plots.
2. Use adjustment tools inside the lower-right of the rectangles to
resize the rectangle. This will not affect the alignment. The
nature of each plot can also be adjusted by selecting it and
selecting the Configuration Options underneath the Setup
window.
3. Drag plots to change desired order.
If no “Available Fields” are selected for a plot, that plot will not be
printed.
Editing Styles
For more information on editing styles, refer to Editing Styles,
page 4–11.
select Edit .
2. Enter a name for the standard. The part number and lot number
of the standard is recommended to be used as the name.
4. If the standard is a Gas Dose, select which dose gas to use (Gas
Dose A, Gas Dose B, Gas Dose AB). If it is not a Gas Dose, go to
step 6, following.
Last Used and Last Modified display information about the standard.
These fields cannot be edited.
6. Select the tab for the desired analyte. The parameters that
appear depend on the selected analyte.
select Edit .
· To import an existing template, select Import. Refer to
Importing and Exporting, page 3–22.
Before importing a template, import its referenced method in order
to avoid a blank method field in the template.
2. Enter a name for the template.
3. Add a description for the template, if desired.
4. Select the template type: Single Sample or Multi Sample. The
Template Type option is available only when editing Single-
Sample types that were created prior to Cornerstone brand
software version 2.8.2. LECO recommends changing any such
templates to Multi-Sample types so that they behave in the same
way as new sample templates. Refer to Using Sample Templates,
page 4–42.
5. Select the method to be used for the template.
6. Use the login bar to add blanks, standards, drifts, samples,
and/or gas doses to the template as described in Logging in a
Sample, page 3–32, and Logging in a Standard, page 3–31. Fill in
all relevant fields to properly run an analysis.
7. Select Save to keep the information, or select Cancel and the
template information will not be saved.
3. Select the desired template from the list at the bottom of the
software screen.
Creating Transports
1. Select Settings, and then select Transports.
Transmit Option
Transmit Option determines the format for the transport (File,
Ethernet, Serial Port).
· XXX is the model number and #### is the serial number of the
analyzer.
Ethernet
Port—The COM port is used for this transport format (for example:
COM1, COM2).
Baud Rate—Bits per second data transmission rate used to configure
the COM port.
Parity—Data transmission parity used to configure the COM port.
Data Bits—Number of data bits used to configure the COM port.
Stop Bits—Number of stop bits used to configure the COM port.
Transport Automatically
Transport Automatically determines if results are automatically
transported at the end of the analysis.
Transport Only When All Replicates Within Set Analyzed—
Determines whether the transport is delayed until all replicates within
a set are analyzed.
Transport Headers
Transport Headers determines when to transport set and rep headers.
Select Never, Once per Transport, or Always.
Transport Format
Transport Format determines the format of the output created: CSV
(RFC 4180), Delimited, XML Format.
Character Encoding
Select the type of character encoding desired for transmitting data:
ASCII, UTF-8, UTF-8 without BOM, Unicode, Unicode without BOM.
Editing Transports
1. Select Settings, and then select Transports.
2. Select a Transport.
Samples Tab
For more information on setting the transport layout, refer to Samples
Tab, page 4–35.
Transport Output
1. Select the set(s) to transport.
User permissions allows the user to edit, add, include, exclude, delete,
and clone user accounts to limit what different users can change within
the software. Once user accounts are created, users can log in and out
of the software by selecting the Session button at the top of the
screen.
There are two user accounts set on the instrument by default:
Administrator and Guest.
• Administrator has complete control over user permissions. The
Administrator account permissions cannot be changed. The
purpose of the Administrator account is to set up and edit any
additional user accounts. The Administrator account is also used to
enable/disable the User Security in the software.
• The Guest account allows only read-only permission to the
software. The Guest account permissions cannot be edited.
When user permissions are enabled in the Administrator account, a
new user account can be created only by the Administrator or other
users that are set up with permission to edit user accounts. In order to
add, edit, delete or clone a user, the currently logged in user must
have permission to edit user accounts and enter their password.
The Remote Query (RQ) option must be enabled on the security key
in order to allow mobile application access to user accounts. Refer to
Adding Options to the Registration Key, page 2–36, for more
information.
4. Select the box next to Enable User Security. Select Yes to enable
user security.
5. Select the box next to User Policy Domain and choose the users
group to be affected by the user policy settings: No Users,
Mobile Users Only, or All Users.
3. Enter a password.
11. All or None may be selected from the list of Menu Bar items to
choose what the user may access. Certain functions within these
Menu Bar items, such as Editable or Read Only, may also be
selected for the user account by selecting the expand/collapse
arrow next to the Menu Bar item’s name. Some is selectable only
if the menu bar items have a mixture of Editable and Read-Only
functions selected.
12. Select Save on the Action Bar when done. The new user will now
appear in the list of users.
2. To edit a user account, select the user from the list by selecting
the orb next to their name and then selecting Edit on the Action
Bar. Enter the password if prompted to do so.
2. From the list of users, select the user account to clone for a new
user account.
3. Select Clone on the Action Bar. A copy of the selected user account will
appear in the list of users.
4. Select the new user account and edit any necessary information by
selecting Edit on the Action Bar.
2. A window pops up that gives you the option to select a different user
account, log out, go back, or exit the software.
4. Select the desired user account from the list of users and
select Enter.
4. Contact LECO Service and relay the challenge key found on the Reset
Password screen.
Security Considerations
Each enabled Cornerstone instrument transmits a laboratory name and
laboratory key to the server and remains connected to the server
using a “long-polling” technique. When activated by the user, the
mobile device transmits the user-supplied laboratory name, laboratory
key, user id, and user password to the server. For each Cornerstone
instrument connected to the server with a matching laboratory name
and key, the server forwards the user id and password to the
instrument for authentication.
The internet-based server passes all authentication and instrument
data between the mobile devices and the Cornerstone instruments.
User ids, passwords, and instrument data are not stored on the
internet-based server. Data is only retained by the Cornerstone
instrument itself. As soon as the Cornerstone instrument is
disconnected from the server, its data is no longer accessible to any
mobile device.
Setting Up Notifications
The RQ option can be used to notify users of instrument events.
1. In the software, navigate to Settings Ø Networking Ø Email
Server Parameters.
2. Select the desired notification server by completing the following
steps.
· To send notifications using LECO's Cornerstone Cloud
Server:
a) Select the parameter field next to Notification Server.
b) Select Cornerstone Cloud Server in the Notification
Server window at the bottom of the screen, and then
select Enter.
If the Cornerstone Health Management (CHM) participation level is
Not Specified, a notice will appear. CHM-0 or CHM-1 must be chosen
to use the Cornerstone Cloud Server. Refer to Cornerstone® Health
Management, page 5–36.
· To send notifications using another server:
a) Select the parameter field next to Notification Server.
b) Select Manual in the Notification Server window at the
bottom of the screen, and then select Enter.
c) Enter information for applicable fields in the Email
Notification Information area. These settings are used to
set up the instrument to send emails. The information
for these fields is typically provided through an IT
department within the user’s organization.
The text message address can be found in the documentation for the
phone service; it can also be retrieved by sending a new SMS (not
MMS) text message to an email address from the phone. Contact the
phone service to find out if your phone is set up for SMS or MMS text
messages.
3. When the replicate to which Assign Notify has been added completes
analysis, will be sent via the channel selected in the Notifications settings
screen. Refer to Setting Up Notifications, page 4–62.
Auto Cleaner
1. Select Instrument, and then select Automation.
2. Select General Parameters under Auto Cleaner to determine the
parameters for the auto cleaner operation.
3. For Auto Cleaner State, select:
· Enabled.
· Disabled.
4. For Number of Clean Cycles, enter the number of brush strokes to
run each time the combustion tube is cleaned.
5. For Analysis Clean Interval, enter the number of analyses to
count before an auto clean is performed. Enter a value between
1 and 200. For example, entering 1 means the auto clean cycle
will run after every analysis. It is recommended to run auto clean
after every analysis.
6. Analysis Count displays the number of analyses that have
been run.
Calibration
Each of the motors has a set of defined positions that need to be
calibrated. This section describes the steps for performing calibration
of these positions using Cornerstone brand software.
It does not address the various mechanical adjustments that are also
available on the shuttle loader hardware for properly placing the
mechanism.
Calibration Prerequisites
In order for the shuttle loader to be calibrated, the motors must be
powered On and have had their home position set. This is indicated in
the software location: Instrument Ø Automation Ø Shuttle Loader Ø
General Parameters.
Select Initialize Shuttle Loader Motors to power On or initiate the
homing process. This process may take several minutes depending
upon the version of firmware present on each of the motors.
Pedestal
Use the Pedestal subsection to calibrate the following:
Rotary Cold to Pedestal Position Offset—The position of the cold
grippers over the pedestal.
Horizontal Position Offset at Pedestal—The position of the
horizontal motor when at the pedestal.
Lift Position Offset When Raised—The height the lift motor will rise
when lifting crucibles off of the pedestal.
Carousel
Use the Carousel subsection to calibrate the following:
Carousel Home Position—The rotation of the carousel assembly.
Rotary Cold to Carousel Position Offset—The position of the cold
grippers over the carousel.
Horizontal Position Offset at Carousel—The position of the horizontal
motor when at the carousel.
Vertical Home Position—The height of the carousel assembly.
Lift Position Offset When Raised—The height the lift motor will rise
when lifting crucibles off of the carousel. Note that changes to this value
here will change the same value on the Pedestal subsection.
The following automation controls may be helpful in determining the
calibrations:
Toggle State of Cold Gripper—Open or closed.
Toggle State of Lift Motor—Up or down.
The life cycle sequence in this subsection will repeatedly grab a
crucible that is placed on the pedestal and place it on carousel location
A1. It will then return it to the pedestal and repeat the pattern. Place
crucibles in the A2 and A10 positions of the carousel to ensure that the
opened gripper arms do not touch the crucibles next to the pickup
position.
The system is properly aligned when the pedestal, rotary motor, and
pick-up position on the carousel are co-linear.
Disposal
Use the Disposal subsection to calibrate the following:
Rotary Hot To Dump Position Offset—The position of the hot
grippers over the dump position.
The following automation control may be helpful in determining the
calibrations:
Toggle State of Hot Gripper—Open or closed.
Full
The Full subsection provides access to all the calibration parameters in
the Pedestal and Carousel subsections. It is intended as an end-to-end
test of the parameters. The life cycle tests in this subsection are
helpful for verifying the consistency of placing and retrieving crucibles
from different carousel locations. For the SL60, the life cycle
performed in this subsection allows for the selection of various
carousel positions.
· A1, F1—the life cycle will take a crucible from the pedestal and
place it on carousel location A1. It will then move that crucible
back to the pedestal and then place it on carousel location F1. This
pattern will be repeated until the life cycle is stopped.
· A1, B1, C1, D1, E1, F1—this life cycle is similar to the previous
one, with the exception that the crucible will be placed in the
number 1 position on all platters.
Reasons to Calibrate
Once the calibration process is completed, the calibration parameters
should remain in effect, even if the Cornerstone brand software is shut
down and restarted. However, a new calibration is required to ensure
proper and accurate crucible retrieval, placement, and disposal in the
following circumstances:
· Maintenance is performed in which a motor is removed or replaced.
· The loader hardware position is physically adjusted.
Maintenance
For information on maintaining the shuttle loader, refer to Shuttle
Loader, page 6–56.
Secura Balance
Complete the instructions in this section for the Secura balance only.
For the Entris II balance, complete the following steps in Entris II
Balance, page 5–18.
6. Select the appropriate Windows version for your PC, and then
select the folder.
10. If the software does not recognize the balance upon installation of
the driver and selection of the correct COM port, complete the
following steps:
A. Change to a different COM port.
B. Turn Off the LECO software
C. Restart the LECO software.
D. Select the new COM port.
11. Continue to Configuring the Software, page 5–19.
(Menu).
Database
The database is an organized collection of information from each
analysis. The database keeps track of setup parameters, calibration
settings, and other information, and it is vital to the software and
instrument's performance.
Database Information
In the software, navigate to Instrument Ø Data Management Ø
Database Information to view the location of the database and the
number of sets and reps that have been analyzed.
Automatic
1. Turn On automatic backups by selecting On next to Automatic.
Select Off to disable automatic backups.
2. Select the box next to Interval, and then enter the desired
backup interval (in days).
3. Select the box next to Time, and then enter the time of day for
the backup to occur. It is recommended that you choose a time
when the software will not be analyzing samples and no one will
be using the instrument.
Once saved, the box next to Next Backup displays when the next
scheduled backup will occur.
4. Select Save to save the settings and exit this screen.
Use the Start and End tabs to create custom date filters. Refer to
History, page 5–25, for instructions on using the calendar.
Use the Start and End tabs to create custom date filters. Refer to
History, page 5–25, for instructions on using the calendar.
History
History is available only with the AT option. When the AT option is
activated within the software, the feature is enabled automatically and
cannot be disabled.
1. In the software, navigate to Instrument Ø Data Management Ø
History to access the history. Any existing history appears on the
History screen and can be exported to a file.
2. Select the box next to Time. The calendar appears.
3. Select the desired date and time for the starting date to view the
history from by using one of the following methods.
· Use the keyboard to enter the desired date and time in the
text box.
· Use the calendar to select the desired date and time.
· Each day, the software will check the IR cell function and, if
necessary, tune the emitter to improve cell performance.
Counters
This feature tracks the periodic maintenance performed on selected
instrument components and alerts the operator when periodic
inspection or maintenance is necessary. Inspection counters can be
ignored, inspected, or performed. However, Stop counters require that
the specified maintenance be performed before running another
analysis.
LECO provides a general guideline for periodic maintenance in the
Frequency column, which may need to be adjusted depending on a
particular analysis. Refer to Periodic Maintenance Schedule, page 6–5.
The list of maintenance counters can be sorted by selecting the name
tab, description tab, expiration tab, and last modified tab.
Default Maintenance counters cannot be deleted, but they may be
excluded. Excluded counters will not prompt the user for an action
Refer to Including and Excluding, page 3–24.
Log Entry
When the operator selects the desired maintenance item, additional
comments appear in the log entry. To edit the comments, select the
Comments field; the keyboard appears. Enter the desired text, and
then select Enter.
Comments are logged when Perform is selected on the Action bar.
Navigate to Instrument Ø Data Management Ø Maintenance Counters
History to view the log.
General Maintenance
The General Maintenance screen provides quick access to frequently
used tasks. Refer to Maintenance, page 6–1 for step-by-step
maintenance procedures.
Heaters—Turns On or Off the incoming purifier heater, catalyst
heater, or dust filter heater.
Digital Outputs—Turns On or Off the coolant pump.
Incoming Pressure—Changes the valve state to read the
incoming pressure. The pressure should read 35 psi.
Combustion Tube—Follow the steps to perform maintenance
on the combustion tube if automated combustion removal
feature is available.
Oscillator Tube —Notifies when the oscillator tube was last
replaced and to log in a new tube.
The following items are not available on all instruments.
Hardware Calibrations
Refer to Calibrating the Hardware, page 9–5.
Error Reports
Error reports provides an error report package, which can then be sent
to LECO for diagnosis.
1. In the software, select Instrument, and then select Maintenance.
2. Select Error Reports.
3. Select Execute next to Create Error Report to create and save the
error report.
The Error Report Path provides the user with the location that the
error report was saved. It is not an editable field.
Remote Access includes the Remote Query (RQ) option, the Remote
Sample Login (RSL) option, the Remote Control (RC) option, and Data
Transmit. These software options are found on select LECO products to
facilitate their use in an automated environment. They can be used to
control the instrument from a remote location via network interface
and/or to receive data from the instrument.
RQ allows a remote computer to request data and settings from the
instrument. RQ does not allow control of the instrument, logging in
samples, or modifying any data on the instrument.
RSL allows a remote computer to log in unanalyzed samples while the
local computer retains control. RC allows a remote application running
on another computer to control the instrument without interaction
from the operator. Data Transmit allows sample data to be transmitted
to a remote computer. For additional information and instructions,
refer to Remote Sample Login, page 5–34; Remote Control, page
5–35; Data Transmit, page 5–35; and Analysis Results, page 5–35.
Data Transmit needs at least uni-directional communication from the
local computer to the remote computer.
RSL and RC communicate with the local computer via a bi-directional
communication protocol using XML commands. The remote computer
exchanges commands and data with a remote computer via network
connection. The network connection is a TCP/IP stream with a
configurable port.
The remote computer sends commands to the local computer, and the
local computer acknowledges the commands or returns the requested
data. Return data will be in XML format; however, sample results are
returned in the format specified by the current sample display layout
for remote clients defined within Cornerstone. For a list of commands
that can be sent to the local computer and what will be returned to the
remote computer, refer to XML Command Table, page 5–35.
The text of the commands and data can be in ASCII or Unicode and
are formatted in XML. The XML format is a way of tagging the
commands and data fields inside of angle brackets (less-than and
greater-than signs). The use of spaces, tabs, carriage returns, and line
feeds are optional.
Figure 5-2
Remote Access Screen
Data Transmit
Data transmit mode allows sample data to be sent to a remote
computer by transmitting sample results in a pre-defined (but
configurable) format to a remote computer. Refer to Creating and
Editing Transports, page 4–44.
Analysis Results
Analysis results can be obtained in two ways for both the Remote
Sample Login and the Remote Control modes:
Remote Connection
The port used for remote communication can be configured from the
Remote Access screen in the Instrument section of the software,
shown in Figure 5-2, page 5–34.
The character encoding that is applied to the data sent to and from
remote clients can be selected. The options are UTF-16 and UTF-8.
Refer to Figure 5-2, page 5–34.
4. Select the link to the desired PIB under Available PIBs. Review
the PIB information.
Disabling an Analyte
Leak Check
The maximum pressure change allowed during a leak check can be
changed for certain leak checks, depending on instrument
configuration.
Segmented leak check is not applicable to all instrument
configurations.
1. Select Instrument, and then select System.
2. On the System screen for Leak Check, select the parameter field
next to the desired leak check.
Printer
Use this feature to select the printer for general printing purposes.
For printer installation, refer to the manual supplied with the printer
and Installing the Computer, page 2–22.
1. Select Instrument, and then select System.
2. On the System screen for Peripheral Devices, select the Printer
parameter field next to System Printer.
3. Select the desired printer name.
Refer to the Safety Data Sheet (SDS) for the specific chemical
for additional information.
Figure 6-1
Cleaning O-Ring
Figure 6-2
O-Ring Shine
5. Install the o-ring, making sure it does not twist or roll and that no
part of the o-ring is nicked or cut. Make sure not to pinch the
o-ring between the sealing surfaces.
OXIDIZING/HARMFUL/
PROTECTIVE EYEWEAR/PROTECTIVE GLOVES
Anhydrone is an oxidizer and a strong dehydrating agent.
Take the following precautions when handling Anhydrone:
CORROSIVE/HARMFUL/
PROTECTIVE EYEWEAR/PROTECTIVE GLOVES
LECOSORB is harmful in contact with skin and may cause skin
burns, corrosion, or irritation. It may also cause eye damage
or irritation. Protective gloves and eyewear should be worn at
all times when handling the product.
Refer to the Safety Data Sheet (SDS) for additional
information.
When packing the reagent tubes, tap them gently to settle the
reagents evenly. This prevents channeling and ensures efficient
operation.
FLAMMABLE/HEALTH HAZARD/HARMFUL/
PROTECTIVE GLOVES/PROTECTIVE EYEWEAR
Acetone is highly flammable and should be kept away from
hot surfaces. It may cause eye irritation and/or drowsiness.
Do not breathe fumes. Protective eyewear and gloves are
required when handling Acetone.
Figure 6-3
Incoming Scrubber Tube Packing
HEALTH HAZARD/HARMFUL/
PROTECTIVE EYEWEAR/PROTECTIVE GLOVES
Copper Oxide is harmful if swallowed and may cause skin
irritation. It may also cause eye damage or irritation. It may
cause damage to organs through prolonged exposure.
Protective gloves and eyewear should be worn at all times
when handling the product.
Refer to the Safety Data Sheet (SDS) for additional
information.
FLAMMABLE/HEALTH HAZARD/HARMFUL/
PROTECTIVE GLOVES/PROTECTIVE EYEWEAR
Acetone is highly flammable and should be kept away from
hot surfaces. It may cause eye irritation and/or drowsiness.
Do not breathe fumes. Protective eyewear and gloves are
required when handling Acetone.
FLAMMABLE/HEALTH HAZARD/HARMFUL/
PROTECTIVE GLOVES/PROTECTIVE EYEWEAR
Acetone is highly flammable and should be kept away from
hot surfaces. It may cause eye irritation and/or drowsiness.
Do not breathe fumes. Protective eyewear and gloves are
required when handling Acetone.
10. If the tube was previously packed, remove the spent packing
material and clean the tube. To clean the tube, rinse it with
water, then acetone and allow it to dry. Refer to Figure 6-5,
page 6–18.
11. Pack the tube using the appropriate steps in Packing the Catalyst
Tube, following.
The following instructions explain how to pack the cellulose trap tube.
The following instructions explain how to pack the flow scrubber tube.
OXIDIZING/HARMFUL/
PROTECTIVE EYEWEAR/PROTECTIVE GLOVES
Anhydrone is an oxidizer and a strong dehydrating agent.
Take the following precautions when handling Anhydrone:
When packing the reagent tubes, tap them gently to settle the
reagents evenly. This prevents channeling and ensures efficient
operation.
PROTECTIVE GLOVES/PROTECTIVE EYEWEAR
Wear appropriate personal protective equipment, including
safety glasses or goggles and cut-resistant gloves, when
handling quart wool.
Refer to the Safety Data Sheet (SDS) for additional hazard,
handling, storage, and personal protection information.
FLAMMABLE/HEALTH HAZARD/HARMFUL/
PROTECTIVE GLOVES/PROTECTIVE EYEWEAR
Acetone is highly flammable and should be kept away from
hot surfaces. It may cause eye irritation and/or drowsiness.
Do not breathe fumes. Protective eyewear and gloves are
required when handling Acetone.
The screen filters are located at the end of the various catalyst and
reagent tubes in the instrument. The filters should be cleaned
whenever the reagents are replaced.
The filters are easily removed using the screen removal tool from the
Accessory Pack, a threaded tool, or any #10-24 screw.
1. If one has not already been done, perform a leak check. Refer to
Leak Check, page 8–15.
2. Remove the o-rings from the screen filters.
3. Inspect, clean, and lightly grease the o-rings. Replace the o-rings
if they are cracked or worn. Set the o-rings aside. Refer to
Cleaning and Greasing the O-Rings, page 6–8.
4. Insert the screen filters in an ultrasonic cleaner.
5. Sonicate the screen filters for 10 minutes.
6. Remove the screen filters from the ultrasonic cleaner and allow
them to air dry.
7. Install the o-rings on the screen filters.
8. Insert the clean screen filters into their original location; or
replace the clean filters with spare filters, if available, and store
for future use.
9. Perform and pass a leak check. Refer to Leak Check, page 8–15.
Figure 6-8
Screen Filters
PINCH HAZARD
The upward pedestal motion causes a pinch point.
Figure 6-9
Pneumatic Gas Filter
8. Unscrew the pneumatic gas filter from the pneumatic filter holder
and inspect for residue. Replace the filter if necessary.
9. Inspect for moisture inside the pneumatic gas filter housing.
A. If moisture is present, release the valve on the bottom of the
pneumatic gas filter housing.
B. After moisture is relieved, tighten valve on pneumatic
housing until secure.
10. Reinstall the pneumatic gas filter into the housing and screw
in place.
11. Attach the pneumatic gas filter housing into its location and screw
it in place.
12. Restore pneumatic air flow at the source.
13. Reinstall the side instrument panel.
14. Turn On the instrument power.
15. Turn On gas by selecting Gas On the Action bar and then
selecting On.
16. Perform and pass a leak check. Refer to Leak Check, page 8–15.
Figure 6-10
Particle Filter Replacement
C844 Models
1. Perform a leak check. Refer to Leak Check, page 8–15.
2. Verify that the pedestal is in the lower position. If it is not, select
locked position .
CS844/S844 Models
1. Perform a leak check. Refer to Leak Check, page 8–15.
2. In the software, select Instrument and then select Maintenance.
3. Select General Maintenance.
4. Turn Off the heated dust filter.
locked position .
15. Lower the auto cleaner lock ring to secure the auto cleaner in
place.
16. Reconnect the pneumatic lines by inserting the fitting back into
the auto cleaner assembly.
17. In the software, select Instrument, Maintenance, and then
General Maintenance.
18. Turn On the heated dust filter by selecting On.
19. Perform and pass a leak check. Refer to Leak Check, page 8–15.
PINCH HAZARD
The upward pedestal motion causes a pinch point.
Manual Removal
1. Perform a leak check. Refer to Leak Check, page 8–15.
2. Verify that the pedestal is in the lower position. If it is not, select
the Pedestal button to lower it.
3. Remove the pedestal.
4. Hold the bottom of the combustion tube with one hand as you
actuate the combustion tube removal levers with the other.
5. Once the combustion tube assembly is unlocked, safely lower the
combustion tube and remove it.
Automated Removal
1. Perform a leak check. Refer to Leak Check, page 8–15.
2. In the software, select Instrument and then select Maintenance.
3. Select General Maintenance.
4. Select Execute next to Step 1 under Combustion Tube, and
remove the silicon nitride pedestal when the pedestal has
lowered.
5. Select Execute next to Step 2 under Combustion Tube to
automatically lower the combustion tube.
6. Remove the combustion tube from the cup assembly.
Replacement
1. Disassemble the combustion tube assembly. Refer to Figure
10-24, page 10–28.
2. Loosen the plastic lock nut and remove it from the combustion
tube.
3. Pull the combustion tube out of the assembly.
4. Inspect, clean, and lightly grease the combustion tube o-rings,
applying a thin coating of vacuum grease. Replace the o-rings if
they are cracked or worn. Refer to Cleaning and Greasing the
O-Rings, page 6–8.
Manual Installation
1. Verify that the combustion tube release levers are in the actuated
(angled) position.
2. Reinstall the combustion tube assembly into the unit by raising it
into place. Make sure that the combustion tube release levers
return to the locked upright position and that the tube is securely
locked into the unit.
3. Perform and pass a leak check. Refer to Leak Check, page 8–15.
Automated Installation
1. Place the combustion tube assembly on the lower cup assembly
making sure that the combustion tube assembly is seated firmly
against the work surface.
2. Select Execute next to Step 3 under Combustion Tube to raise
and lock the combustion tube into place.
3. Reinstall the silicon nitride pedestal when the cup assembly
lowers.
4. Select Execute next to Step 4 under Combustion Tube to perform
and pass a leak check. Refer to Leak Check, page 8–15.
HARMFUL/PROTECTIVE GLOVES/
PROTECTIVE EYEWEAR
Citranox® is an acid and can cause skin and eye irritation.
Protective gloves and eyewear should be worn at all times
when handling this product. The cleaner fluid will foam.
C844 Models
On occasion it may be necessary to clean the filter assembly to
remove any dust that has accumulated. Dust may accumulate if a filter
is not installed or if the filter has not been replaced as recommended
in the Periodic Maintenance Schedule, page 6–5.
Figure 6-12
Right Casting Removal
Figure 6-13
Pinch Valve Tubing
5. Slide the tubing out of the housing assembly. Remove the hose
clamps on both ends of the tubing.
6. When inspecting the tube, check for any buildup of dust and clean
it out. Check to verify there are no cracks in the tubing; replace it
if necessary.
7. Inspect the pinch tube union located on the front of the lower
pneumatic cylinder. Remove any dust that has accumulated
within the union.
8. Install the hose clamps on the tubing and attach both ends.
9. Slide the tubing back into the pinch valve, and then reinstall
the cap.
10. Install the bottom casting and work surface.
11. Turn On the pneumatic gas.
12. Perform and pass a segmented leak check. Refer to Leak Check,
page 8–15.
The coolant in the heat exchanger cools the induction coil and should
be changed every ninety days to maintain furnace efficiency.
Before replacing the coolant, locate the furnace coolant drain line
assembly, supplied in the Components and Accessories, page
1–48, and a container to accept the spent coolant.
Figure 6-15
Drain Fitting Access Door
Weekly
PINCH HAZARD
The upward pedestal motion causes a pinch point.
Monthly
1. Perform the weekly maintenance prior to performing the following
steps.
2. Remove the vacuum cleaner manifold from its location on the
back lower right position of the unit and disconnect the hoses
being careful not to spill any dust remaining in the manifold.
3. Use a brush to remove as much dust from the manifold as
possible.
4. Wipe out the inside of the manifold.
5. Make sure that the pinch tube port is clear of dust.
6. Remove all o-rings from the vacuum manifold ports. Clean, lightly
grease and reinstall the o-rings. Replace any o-rings that are
damaged. Refer to Cleaning and Greasing the O-Rings, page 6–8.
7. Wipe off the hose connection ends.
8. Reinstall the hoses and the vacuum manifold on the back of the
unit.
PINCH HAZARD
Raising the electrode/pedestal causes a pinch point.
Figure 6-16
Gas Dose Assembly
9. On the gas dose assembly, there are two endcaps. Grasp one
endcap and then rotate and pull out to remove the assembly.
Repeat for second endcap. Refer to Figure 6-17, following.
Figure 6-17
Endcap for Gas Dose Assembly
Figure 6-18
Endcap O-Ring
11. Obtain the screen/stem doser removal tool from the accessory
pack. A #6 screw may be used instead, if desired.
12. Thread the tool into the gas doser valve stem, and then gently
pull out valve stem. Repeat for other valve stem. Refer to
Figure 6-19, following.
Figure 6-19
Valve Stem in Gas Dose Manifold
13. Inspect, clean, and lightly grease the o-rings on the valve stems.
Replace the o-rings if they are cracked or worn. Apply a thin
coating of vacuum grease. Refer to Figure 6-20, following, and
Cleaning and Greasing the O-Rings, page 6–8.
Figure 6-20
Valve Stem
14. Insert one valve stem into the gas doser manifold, and then push
until seated. Repeat for the other valve stem.
15. Reinstall the endcap onto the valve stem by aligning the flat parts
of the endcap to the top and bottom of valve stem, and then
turning the endcap to secure in place. Repeat for the second
endcap.
CAUTION
The end cap must be secured or it will blow off the gas dose
manifold when pressure is applied.
CAUTION
DO NOT use alcohol (methyl, ethyl or isopropyl) or any strong
solvent. Do not use thinner or benzene, abrasive cleaners or
compressed air to clean the touch monitor.
The following procedure is not required and does not appear on all
touch screen models.
The LECO-supplied touch screen has been calibrated before shipment.
The following procedure applies only if the calibration has to be
repeated.
The Elo software will automatically start whenever the desktop screen
orientation is changed. Close the window because the calibration is
complete.
6. Verify the touch screen works for computer programs.
There are two air filters on the rear of the instrument. The top filter is
attached to the heat exchanger, and the lower filter is attached to the
rear panel. Air filters should be visually inspected daily and cleaned
weekly. Cleaning schedule varies according to instrument usage and
environment. Failure to regularly clean the air filters will reduce air
flow and reduce the capacity of the instrument to manage generated
heat.
1. Remove the air filters. The top filter is removed by sliding it up,
and the lower filter is removed by sliding it to the right, toward
the doser fittings.
2. Vacuum the air filters daily and wash them monthly. Wash the air
filters in a mild solution of detergent, and then allow them to air
dry.
3. Reinstall the air filters when dry.
Figure 6-21
Air Filter Location
Shuttle Loader
1. Before performing maintenance on the shuttle loader hardware,
place the motors into the maintenance position by navigating to
Instrument Ø Automation Ø Shuttle Loader Ø Maintenance.
CAUTION
POSSIBLE SHUTTLE LOADER MOTOR DAMAGE
Never remove or reinstall any of the shuttle loader motors
without performing the maintenance step in the following
software location:
2. Select the Execute button for Step 1 to place the motor into a
maintenance position and to turn Off power to the motor.
3. Perform the desired shuttle loader maintenance.
4. Once maintenance has been completed, select the Execute button
for Step 2 to return power to the motor for normal use.
Figure 6-22
Rotary Motor and Gripper Assembly
Figure 6-24
Crucible Gripper Assembly
Figure 6-25
Shuttle Loader Horizontal Housing
Calculate Area
The final output signals, whether baseline corrected, wedge rule
corrected, or uncorrected, are used to determine a raw peak area,
Arange, for each analyte range. This raw area is reported in the
expanded replicate information in the Cornerstone® brand software.
Further corrections, as follows, are applied to the raw total area before
they can be used with a calibration.
Adjusted Area
The adjusted area, Aadj, range, corrects the raw area for sensitivity
factors and blank values.
Blank
The blank value, Ab,range, is the average raw area from a set of samples
(empty crucibles, boats, etc.) determined by the user. The blank value
is used to account for analyte gases that are repeatedly introduced
through consumables or atmosphere. The default blank value is 0
1 n
Ab, range = å Arange
n 1
where n is the number of blank samples as selected by the user.
Adjusted Area
The adjusted area can be calculated as follows:
Aadj,range = (Arange - Ab ,range ) × Fsensitivity
Determine Range
This section is applicable only to dual-range instruments.
If an instrument utilizes both a high and low range for a given analyte,
the correct mass result must be used. The result may come from the
high range, low range, or a blend of both depending on the method
settings.
If a blended result is desired, the value is determined based on the
high and low range thresholds set by the user (ThH and ThL) and the
maximum peak height of the low range result for each analysis (Hp).
If:
Hp > ThH, Use high range
Hp < ThL, Use low range
ThL < Hp < ThL, Use blended results, detailed as follows:
The peak height is reported in the expanded replicate information in
the Cornerstone brand software.
Drift Correction
The drift factor, Fdrift, corrects for day-to-day changes in signal due to
changing environmental, matrix, or sample conditions. It is a simple
multiplicative factor applied to the final analyte mass and reduces the
frequency with which new calibrations need to be set. The drift factor
value is determined against a calibration and has a default value of 1.
Gdrift = G × Fdrift
æ ö
÷÷ × (G high (H p - Th L ) + Glow (Th H - H p ))
1
Gblended = çç
è Th H - Th L ø
Thorough calibrations are required at both the low and high ranges to
optimize signal blending.
Determine Composition
The determined analyte mass, G or Gblended, is used in conjunction with
the sample mass to determine the analyte composition.
æGö
C ( ppm) = 10 6 × ç ÷
èmø
Calibration Coefficient
Calculated for Each
k The best-fit coefficients determined weighted least Calculated
Method
squares
Calibration Equation
A weighted least squares algorithm is used to determine the
polynomial coefficients, k.
W × Aadj = k0 + k1 × G
By selecting the Regression Type option to Force Through Origin, the
operator can force the calibration equation through the origin:
k0 = 0
W × Aadj = k0 + k1 × G + k2 × G 2 + k3 × G3
Apply Calibration
Given the coefficients, K, the determined analyte mass, G, can be
determined for each replicate by solving the following equation:
Aadj = k 0 + k1 × G
For higher order calibrations, Cornerstone brand software finds a valid
solution for the following polynomial equation:
Aadj = k0 + k1 × G + k2 × G 2 + k3 × G 3
æGö
C (%) = 100 × ç ÷
èmø
The composition will also be determined in parts per million (ppm):
æGö
C ( ppm) = 106 × ç ÷
è mø
Weighting Factors
Weighting factors, W, are calculated as:
g × wi
Wi =
0.0001 + g i2
where
1 n
g= å gi
n 1
g is the certified analyte mass for each replicate, i, included in the
calibration and w is a user entered weighting factor. By default, the
user defined weighting factors, w, are set to 1.0. If the higher order
calibration option is enabled, the operator can view and edit the
weighting factors.
å (g - gi ,certified )
n
i ,measured
RMS Error (g) = 1
n -n
Where n is the degree of freedom as determined by the combination of
Regression Order and Regression Type as follows:
Regression Force Full
Order Through Origin Regression
Linear 1 2
*Quadratic 2 3
*Cubic 3 4
*Available only with the Higher Order Calibration Option
å (g )
n
i , measured - gi , certified
R2 = 1- 1
2
å (g )
n
i , measured -g
1
Override Conditions
When possible, Cornerstone brand software chooses a Regression
Order of “Linear” and a Regression Type of “Full Regression” whenever
a new calibration equation is created.
To prevent invalid calibration equations, the software will automatically
override user settings and defaults to ensure that a valid calibration
equation is determined. The following criteria are used:
Single Cluster
No matter how many replicates are included, calibrations that are
recognized as a single cluster of measurements will automatically be
set to a Regression Order of “Linear” and a Regression Type of “Force
Through Origin.”
The calibration data will be interpreted as a single cluster if any of the
following criteria are true:
· The range of analyte present in the calibration data is less than
three times the range of the absolute error (g) of the calibration.
max (analyte( g ) cert ) - min (analyte( g ) cert ) < 3 · [max (Error(g )) - min (Error(g ))]
Bypass
P 12 psi 0.35" PV1
Block
NC
FLTR1 R1 SV102 High J
O
SO2 P
Incoming Purifier
com NO
1 LPM
3 LPM
FS
Flow Scrubber
K
P
Heated
A Dust Filter
Catalyst
PT3
F 12 psi
E
Furnace
Bypass
Cooling 0.35"
Block
N
Network High
O2
CO2
Bypass
Block
Cooling
I
Loop
35 psi
SV107
B
Particle
Cooling
Dose B
NC
Filter
Loop
55 psig Pinch CV1
Incoming Scrubber
G
Valve
Gas PS1
Bypass
Block
Dosing M
com
4 psi / NO
Cellulose Trap
SV104 PT2
Exhaust
Dose A
P
12 psi
NC
13" 6"
1 LPM
R31
1 LPM
R32
NC
5 LPM NC com
To To
R3 SV103
Dose Dose MeFC1
B A
3 LPM
Legend
Incoming O2/Doser Leak Check
Connection to Analytical Manifold
Segment
Connection to Detector Manifold Furnace Leak Check Segment
Indicates a Path Within a Manifold Detector Leak Check Segment
(Dashed Line)
Description Purpose
A Incoming Purifier CxHy removal; CO oxidation
B Incoming Scrubber CO2 conversion; Moisture removal
C Lance Restrictor Control lance flow rate
D Proportional Valve Maintains furnace pressure
E Furnace Convert solid samples to gaseous phase
F Flow Scrubber Moisture removal
G Particle Filter Particle removal
H Mechanical Flow Controller Control system flow rate
I Low SO2 IR Cell Low SO2 composition measurement
J High SO2 IR Cell High SO2 composition measurement
K Catalyst CO and SO2 oxidation
L Cellulose Trap SO3 removal
M Low CO2 IR Cell Low CO2 composition measurement
N High CO2 IR Cell High CO2 composition measurement
O Proportional Valve Maintains IR cell pressure
P Flow Sensor Measures the gas flow rate
Figure 7-1
Flow Diagram (CS844)
SV102
FLTR1
R1
Filter
0.35"
High K
Incoming Purifier
com NO
NC
1 LPM
SO2
SV401
Flow Scrubber
NO
Heated
A Dust Filter
SV101
Catalyst
55 psig
PT3
NC
com
F
P
E 790 mmHg
Furnace N
Cooling
PV1
Network 0.35"
O2 High P
CO2
Cooling
Loop
35 psi
B PT401
Particle
Cooling
Dose B I
Filter
Loop
55 psig
P
Pinch CV1
Incoming Scrubber
3 LPM
FS
Valve
Gas PS1
Dosing
com
4 psi / NO
Cellulose Trap
SV104
Dose A PT2 M
NC
12/17 psi
1 LPM
R31
P
1 LPM
R32
CO2
SV107
SV105 H 13"
com
R2 6"
NC
NO
PV2 MeFC1 Low Low
SV311
SV312
NC
5 LPM NC com
To To
Dose Dose R3 SV103
B A Exhaust
Legend
Indicates a Path Within a Manifold (Dashed Line) Detector Leak Check Segment
Description Purpose
A Incoming Purifier CxHy removal; CO oxidation
B Incoming Scrubber CO2 conversion; Moisture removal
C Lance Restrictor Control lance flow rate
D Proportional Valve Regulate furnace pressure
E Furnace Convert solid samples to gaseous phase
F Flow Scrubber Moisture removal
G Particle Filter Particle removal
H Mechanical Flow Cont. Control system flow rate
I Low SO2 IR Cell Low SO2 composition measurement
J High SO2 IR Cell High SO2 composition measurement
K Catalyst CO and SO2 oxidation
L Cellulose Trap SO3 removal
M Low CO2 IR Cell Low CO2 composition measurement
N High CO2 IR Cell High CO2 composition measurement
O Mechanical Flow Cont. Control system flow rate for enhanced sensitivity mode
P Proportional Valve Maintains IR cell pressure above barometric
Figure 7-2
Flow Diagram (CS844ES)
Bypass
P 12 psi PV1
Block
NC
FLTR1 R1 SV102
N
I
Incoming Purifier
com NO
1 LPM
3 LPM
FS
Flow Scrubber
P
A
Catalyst
PT3
F 12 psi
E
Furnace
Bypass
Cooling 0.35"
Block
Network High L
O2
CO2
Bypass
Block
Cooling
Loop
35 psi
SV107
B
Particle
Cooling
Dose B
NC
Filter
Loop
55 psig Pinch CV1
Incoming Scrubber
G
Valve
Gas PS1
Bypass
Block
Dosing K
com
4 psi / NO
Cellulose Trap
PT2
SV104
Exhaust
Dose A
P
12 psi
NC
6"
1 LPM
R31
1 LPM
R32
CO2 Low
R2 SV105
com NO CO2
PV2
SV311
H
SV312
NC
5 LPM NC com
To To
R3 SV103
Dose Dose MeFC1
B A
3 LPM
Legend
Incoming O2/Doser Leak Check
Connection to Analytical Manifold
Segment
Connection to Detector Manifold Furnace Leak Check Segment
Indicates a Path Within a Manifold Detector Leak Check Segment
(Dashed Line)
Description Purpose
A Incoming Purifier CxHy removal; CO oxidation
B Incoming Scrubber CO2 conversion; Moisture removal
C Lance Restrictor Control lance flow rate
D Proportional Valve Maintains furnace pressure
E Furnace Convert solid samples to gaseous phase
F Flow Scrubber Moisture removal
G Particle Filter Particle removal
H Mechanical Flow Controller Control system flow rate
I Catalyst CO and SO2 oxidation
J Cellulose Trap SO3 removal
K Low CO2 IR Cell Low CO2 composition measurement
L High CO2 IR Cell High CO2 composition measurement
M Proportional Valve Maintains IR cell pressure
N Flow Sensor Measures the gas flow rate
Figure 7-3
Flow Diagram (C844)
Bypass
P 12 psi 0.35" PV1
Block
NC
SV102 High H I
FLTR1 R1
SO2 J
com NO
1 LPM
3 LPM
FS
Flow Scrubber
P
Heated
Dust Filter
PT3
D 12 psi
C
Furnace
Bypass
Cooling
Block
Network
O2
Bypass
Block
G
35 psi
SV107
Particle
NC
Filter
Pinch CV1 E
Valve
PS1
Bypass
Block
com
4 psi / NO
PT2
SV104
Exhaust
P
12 psi
NC
13"
Low
R2 SV105
com NO SO2
PV2
B 2 LPM NC com F
5 LPM
SV106
NC com
R3 SV103 MeFC1
3 LPM
Legend
Incoming O2/Doser Leak Check
Connection to Analytical Manifold
Segment
Connection to Detector Manifold Furnace Leak Check Segment
Indicates a Path Within a Manifold Detector Leak Check Segment
(Dashed Line)
Description Purpose
A Lance Restrictor Control lance flow rate
B Proportional Valve Maintains furnace pressure
C Furnace Convert solid samples to gaseous phase
D Flow Scrubber Moisture removal
E Particle Filter Particle removal
F Mechanical Flow Controller Control system flow rate
G Low SO2 IR Cell Low SO2 composition measurement
H High SO2 IR Cell High SO2 composition measurement
I Proportional Valve Maintains IR cell pressure
J Flow Sensor Measures the gas flow rate
Figure 7-4
Flow Diagram (S844)
SV102
FLTR1
R1
Filter
0.35"
com NO
High
NC
1 LPM
SO2
SV401
Flow Scrubber
NO
Heated
Dust Filter
SV101
55 psig
PT3
NC
com
D
P
C 790 mmHg
Furnace
Cooling
PV1
Network
O2 K
35 psi
PT401
Particle
G
P
Filter
Pinch CV1 E
3 LPM
FS
Valve
PS1
com
4 psi / NO
SV104
PT2
NC
12/17 psi
SV107
SV105 F 13"
com
R2
NC
NO
PV2 MeFC1 Low
B 5 LPM NC com 3 LPM
SV106 SO2
5 LPM NC com
R3 SV103 Exhaust
Legend
Indicates a Path Within a Manifold (Dashed Line) Detector Leak Check Segment
Description Purpose
A Lance Restrictor Control lance flow rate
B Proportional Valve Regulate furnace pressure
C Furnace Convert solid samples to gaseous phase
D Flow Scrubber Moisture removal
E Particle Filter Particle removal
F Mechanical Flow Cont. Control system flow rate
F Low SO2 IR Cell Low SO2 composition measurement
H High SO2 IR Cell High SO2 composition measurement
J Mechanical Flow Cont. Control system flow rate for enhanced sensitivity mode
K Proportional Valve Maintains IR cell pressure above barometric
Figure 7-5
Flow Diagram (S844ES)
The Diagnostics Flow Tab is divided into two parts: an analytical flow
diagram at the bottom and a diagnostic chart at the top. For
instructions on using the diagnostic chart to plot instrument
parameters, refer to Diagnostic Chart, page 8–4. The flow diagram is
interactive, enabling the user to actuate valves and plot ambient
parameter values by selecting the indicators within the diagram.
Additionally, by examining the motion of the dashed lines in the flow
diagram, the user can visualize the flow path for the current
valve state.
There are two states in which the user can operate the Flow Tab:
locked and unlocked. The user can change the operation state by
toggling between the Lock and Unlock buttons on the action bar at the
top of the screen.
When the Lock button is selected, the user may plot ambient values,
valve states, and interlock states on the chart. The user cannot change
valve states by selecting valves within the flow diagram.
When the Unlock button is selected, the user may plot ambient values
and interlock states and may actuate individual valves.
CAUTION
Actuating valves in the unlocked state will cause the valves
to change and the instrument to react accordingly. Changing
the state of some valves could cause the instrument to react
in an unintended manner, causing personal injury or damage
to the instrument. It is recommended that this diagnostic be
used under the supervision of trained service engineers.
In either mode, the flow diagram shows the theoretical gas flow path
for the current valve state as well as the current parameter values. A
warning triangle will flash beside a readout if a parameter is outside its
desired range.
Parameters
Within each parameter box, the current value, as well as its minimum
and maximum values, are displayed. Select the reset button on the
action bar to reset the minimum and maximum values within the
parameter box. The minimum/maximum values automatically reset
when opening the software.
A warning triangle will flash within a parameter box if a parameter is
outside its desired range. The warning symbol will flash in the
Diagnostics button, on the Ambient tab, and in the upper left corner of
a parameter box in the Ambient screen.
Diagnostic Chart
Select the desired parameter to plot the value on the diagnostics
chart. The parameter value is displayed on a vertical axis and the time
is displayed on the horizontal axis. More than one parameter can be
plotted on the graph at a time. Each parameter has its own unique
color that is displayed when selected.
The x- and y-axes of the diagnostic chart are adjustable. The plus and
minus signs on the axes indicate which direction to slide in order to
zoom the desired direction. To zoom out, select the axis bar and slide
to the left/up. To zoom in, select the axis bar and slide to the
right/down.
To view previous data, drag or flick in the chart area. The software
automatically saves the last 30 days of ambient data. The chart's scroll
speed is proportional to the speed of the flick. To stop the chart from
scrolling, select and hold anywhere on the chart area.
The diagnostic chart can be paused by selecting the play button in the
lower right corner of the chart. When adjusting the x-axis, or scrolling
back in time, the chart automatically goes into pause mode. Selecting
the play button will return to live plotting mode.
Parameter Definition
Low CO2 The IR cell output voltage in volts. The emitter is turned Off
in Gas Conserve and Gas Off states. The output voltage is set
by the software. To determine if the cell is working properly,
refer to Detectors, page 5–27.
High CO2 The IR cell output voltage in volts. The emitter is turned Off
in Gas Conserve and Gas Off states. The output voltage is set
by software. To determine if the cell is working properly, refer
to Detectors, page 5–27.
Low SO2 The IR cell output voltage in volts. The emitter is turned Off
in Gas Conserve and Gas Off states. The output voltage is set
by software. To determine if the cell is working properly, refer
to Detectors, page 5–27.
High SO2 The IR cell output voltage in volts. The emitter is turned Off
in Gas Conserve and Gas Off states. The output voltage is set
by software. To determine if the cell is working properly, refer
to Detectors, page 5–27.
Low CO2 The temperature of the Low CO 2 IR cell. The value should be
Temperature approximately 45 °C.
High CO2 The temperature of the High CO2 IR cell. The value should be
Temperature approximately 50 °C.
Low SO2 The temperature of the Low SO2 IR cell. The value should be
Temperature approximately 45 °C.
High SO2 The temperature of the High SO2 IR cell. The value should be
Temperature approximately 50 °C.
Upstream The oxygen pressure before oxygen enters the furnace. This
Pressure (PT1) value should be 12 psi. The sensor also reads the incoming
pressure in certain valve states. The incoming pressure
should be 35 psi. This is the oxygen pressure from the
oxygen source (tank). For CS844ES and S844ES models
only: when analyzing in Enhanced Sensitivity mode, the
upstream pressure will read 17 psi.
Downstream The combustion gas pressure as it leaves the furnace and
Pressure (PT2) flows through the furnace filter. The downstream pressure
should be approximately 12 psi. For CS844ES and S844ES
models only: when analyzing in Enhanced Sensitivity mode,
the downstream pressure will read 17 psi.
Back Pressure The pressure of the exhaust gas before leaving the
(PT3) instrument.
Carrier Flow (FS) The flow rate from the oxygen supply (tank). The carrier flow
rate should be 3 LPM.
Grid Current The electrical DC current flowing from the cathode to the grid
of the oscillator vacuum tube V2. As RF power increases
during analysis the grid current rises.
Plate Current The electrical DC current flowing from the cathode to the
plate of the oscillator vacuum tube V2. As RF power increases
during analysis the plate current rises.
Cabinet The internal cabinet temperature inside the instrument.
Temperature Before analysis, the ambient temperature should be close to
room temperature.
Plate Voltage The voltage sourced by the plate supply, which feeds the
plate of the oscillator tube.
Gas Dose A The temperature of gas dose loop A, which is set at 45 °C.
Temperature
Gas Dose B The temperature of gas dose loop B, which is set at 45 °C.
Temperature
Atmospheric The atmospheric pressure around the instrument. PT4 is used
Pressure for gas dose calculations.
IR Cell Pressure For CS844ES and S844ES models only. The gas pressure
(PT401) in the IR Cells. The IR cell pressure should be between
0.5-17.5 psi.
Low Flow MeFC For CS844ES and S844ES models only. The temperature
Temperature of the low flow mechanical flow controller, which is set at
45 °C.
The Digital I/O screen allows the user to view all the digital inputs and
outputs and to actuate all the digital outputs within the instrument.
Similar to the ambient screen, it is divided into two parts, a diagnostic
chart at the top and the digital inputs and outputs at the bottom. The
current valve states are also displayed above the list of digital outputs.
For both digital inputs and outputs, a highlighted orb indicates that the
digital input or output is energized, and a gray orb indicates that it is
non-energized.
The Digital I/O tab can be operated in two states: locked (default) and
unlocked. The user can change the operation state by toggling
between the Lock and Unlock buttons at the top of the screen.
When the Lock button is selected, the state of the digital outputs
cannot be changed. The digital inputs and outputs can be selected as
they are in the ambient screen and can be plotted on the diagnostic
chart. A thin line represents a digital input or output in the non-
energized state, and a thick line represents it in the energized state. If
ambient parameters have been selected from either the Flow or
Ambient screens, they will also be displayed on the Diagnostic Chart in
the Digital I/O screen.
When the Unlock button is selected, selecting a digital output will
actuate that component.
CAUTION
Actuating valves in the unlocked state will cause the valves
to change and the instrument to react accordingly. Changing
the state of some valves could cause the instrument to react
in an unintended manner, causing personal injury or damage
to the instrument. It is recommended that this diagnostic be
used under the supervision of trained service engineers.
Gas Dose is available only on the following models: CS844 and C844.
During a leak check, the flow is turned Off and the system is sealed
by the incoming and exhaust solenoids. Following a leak check, the
flow remains Off until it is turned On by enabling the Gas On state or
by leaving Diagnostics.
Pressurization Failure
It is possible that the system will fail to pressurize when performing a
leak check or during operation. If this occurs, there is a gross leak in
the analytical or pneumatic gas systems. Although this is uncommon,
first check the auto cleaner o-rings, the combustion tube o-rings, and
the pinch tubing. If the leak occurred after maintenance, check the
assembly or area that was serviced.
Pressure Increase
If the system pressurizes and the pressure increases during a leak
check, either a valve has failed to close and is allowing gas to enter
the system or the temperature of a heater has not stabilized. In the
latter case, allow the temperatures to stabilize, and then perform
another leak check.
Pressure Decrease
If the system pressurizes and the pressure decreases during the leak
check, there is a leak in the system. In this case, check the last action
or maintenance item that was performed. Something may have been
disturbed or incorrectly installed during another procedure. If reagent
materials were changed, make sure no glass wool is interfering with
the sealing surface of the o-rings. Check to make sure fittings are
correctly tightened. Refer to Fitting Assembly and Reassembly, page
9–10 for the correct tightening procedures for the various fittings used
throughout the system. Refer to Manifold Flow Diagrams, page 9–14,
to determine which areas are affected by the leak.
The System Check is used to check all basic functions necessary for
the unit to perform an analysis. Depending on the instrument
configuration, this may include checks of the communication,
electrical, pressure, flow, temperature, motor(s), and interlock
systems. If any of the items fail, it will be necessary for the user or
service engineer to correct the problems before the unit can be used
for analysis.
Prior to performing a system check, ensure that the system has been
on for at least 30 min to allow all temperatures to stabilize.
Perform a System Check by completing the following steps.
1. In the software, select Diagnostics, and then select the System Check
tab.
2. Select Start, and then wait for the System Check sequence to finish.
3. Address any items that do not pass.
The following section explains the flow rates that can be expected from
the various gas flow states.
Gas On—The Gas On flow is 3 lpm.
For CS844ES and S844ES models only: The Gas On flow is
determined by the current method, which is defined by the
method of the next logged-in sample, or the last analyzed
method when no samples are logged in. For methods with the
Analysis Mode set to Normal, the Gas On flow is 3 lpm. For
methods with the Analysis Mode Set to Enhanced Sensitivity, the
Gas On flow is 0.8 lpm.
Conserve—With the pedestal up, the Gas Conserve flow is
approximately 1 lpm. Gas Conserve is not an option with the pedestal
down.
Gas Off— When going from Gas On or Gas Conserve to Gas Off the
instrument will depressurize and wait until the upstream pressure is
below 2 psi before stating that the gas flow is off.
Question Answer
Why are the cell To prolong the life of the emitter, the IR cells in the
voltages so low (close instrument are designed to turn Off when the carrier
to 0.00 volts)? flow is Off or in Conservation Mode.
Why am I prevented Calibration data is clustered around a single data point.
from selecting “Full In these cases, the Calibration Type will automatically be
Regression” in the changed to “Force Through Origin.”
Calibration Screen? The Cornerstone® software prevents users from selecting "Full Regressio
Why do some of the To reduce instrument noise when not in use, the RF Fan,
fans and pumps Coolant Fan, and Coolant Pump are turned Off when the
change when changing carrier flow is Off or in Conservation Mode. The Cabinet Fan
gas states? also lowers the speed but does not turn Off. These fans and
pump fully turn On again when the carrier flow is On.
Why does the The flow path has changed compared to the previous
Upstream and generation of instruments. This behavior is considered
Downstream Pressure normal and has not been observed to decrease precision
vary during analysis or affect results.
and when raising and
lowering the pedestal?
Figure 9-1
Casting Removal
3. In the software, turn On the gas flow by selecting Gas, and then
selecting On.
PINCH HAZARD
The upward pedestal motion causes a pinch point.
Figure 9-2
Safety Pressure Valve and Adjustment
LECO Part
Description Where Used
Number
5/16", Blue
619-591-553 Oscillator Cooling outside RF Enclosure
Polyurethane Tubing
619-591-794 8 mm PTFE Tubing Oscillator Cooling inside RF Enclosure
619-591-795 8 mm PTFE Tubing Oscillator Cooling inside RF Enclosure
1/8" White
619-590-823 Pinch Valve
Polyurethane Tubing
1/8" Black
625-710-795 Automation (if installed)
Polyurethane Tubing
1/4" Orange
625-710-695 Pneumatics
Polyurethane Tubing
1/4" Yellow
619-591-384 Pedestal (Top)
Polyurethane Tubing
1/4" Red
619-591-383 Pedestal (Bottom)
Polyurethane Tubing
1/4" White
619-590-820 Brush (Up)
Polyurethane Tubing
1/4" Black
619-590-821 Brush (Down)
Polyurethane Tubing
Flow Sensor to Exhaust Manifold, Catalyst to
783-771 1/4" PFA Tubing
Back Pressure Manifold (if installed)
615-830 1/8" PFA Tubing Analytical Manifold to Lance
1/4" Urethane
807-196 Analytical Manifold to Lower Pedestal
Tubing
3/8" White Silicone
625-516 Lower Pedestal to Pedestal Bracket Union
Tubing
3/8" Orange Silicone
501-110 Pedestal Bracket Union to Vacuum Manifold
Tubing
Compression Fittings
Refer to Figure 9-3, following, for the correct installation and
orientation of the nut and ferrules.
Figure 9-3
Ferrule Installation
Assembly
1. Rotate the nut finger-tight.
Reassembly
You may disassemble and reassemble compression tube fittings many
times.
CAUTION
Always depressurize the system before disassembling a
swage tube fitting.
1. Prior to disassembly, mark the tube at the back of the nut; mark
a line along the nut and fitting body flats. Use these marks to
ensure that the nut is returned to the previously pulled-up
position.
2. Insert the tube into the fitting until the front ferrule seats against
the fitting body.
CAUTION
Do not use the compression fitting gap inspection gauge with
reassembled fittings.
SV101 FLTR1
NC
O2
Incoming/Exhaust
35 psi Manifold Incoming
55 psig
Detector Manifold Scrubber
SV107
MeFC1
Exhaust NC 3 LPM
13" Low
Bypass
SO2 IR Cell
6.0" Low
CO2 IR Cell Bypass
Figure 9-4
Manifold Flow Diagram (CS844)
PT1 P
SV102
4 PSI / NO
R1 12/17 psi COM PS1
PT2 P
NO 2 LPM SV105 12/17 psi
55 psig
Dose B 1 LPM COM
R32 SV312
Analytical Manifold COM R2 NC
COM
1 LPM NC
SV104 NC NO SV106
CO2 Dose A PV2
R31 SV311 5 LPM NC
COM
Incoming
Purifier
20 psi NC
1 LPM R3 SV103
Gas Dose Manifold
55 psig
SV101 FLTR1
NC
O2 Incoming/Exhaust
Manifold Detector Manifold
35 psi
Incoming
Scrubber
55 psig MeFC1
3 LPM
SV107
NC
Exhaust
13" Low
SO2 IR Cell 0.35" High
SO2 IR Cell
6.0" Low
CO2 IR Cell Catalyst
NC
0.35" High P
CO2 IR Cell PT401 SV401 Flow
Furnace CO2 IR Cell COM
NO MeFC
Detector P
Figure 9-5
Manifold Flow Diagram (CS844ES)
SV101 FLTR1
NC
O2
Incoming/Exhaust
35 psi Manifold
Incoming
Detector Manifold
55 psig
SV107 Scrubber
MeFC1
Exhaust NC 3 LPM
Bypass
Bypass Catalyst
6.0" Low
Bypass
CO2 IR Cell
Figure 9-6
Manifold Flow Diagram (C844)
CV1 Particle
Filter
P 4 PSI / NO
R1 SV102 PT2
COM PS1
P
PT1 NO 2 LPM SV105 12 psi
1 LPM COM
12 psi
COM R2 NC
COM
Analytical Manifold
SV104 NC NO SV106
PV2
5 LPM SV103
COM
R3 NC
SV101 FLTR1
NC
O2
Incoming/Exhaust
35 psi Manifold
55 psig
Detector Manifold
SV107
MeFC1
Exhaust NC 3 LPM
13" Low
Bypass
SO2 IR Cell
Bypass
Furnace
P
Detector
PT3
Back Pressure 790 mmHg
Pneumatics Controller PV1
FS
Connection Types
Figure 9-7
Manifold Flow Diagram (S844)
CV1 Particle
Filter
PT1 P
SV102
4 PSI / NO
R1 12/17 psi COM PS1
PT2 P
NO 2 LPM SV105 12/17 psi
1 LPM COM
SV104 NC NO SV106
PV2 5 LPM NC
COM
R3 SV103
SV101 FLTR1
NC
O2 Incoming/Exhaust
Manifold Detector Manifold
35 psi
55 psig MeFC1
3 LPM
SV107
NC
Exhaust
13" Low
SO2 IR Cell 0.35" High
SO2 IR Cell
NC
Incoming O2 & Doser
P
SV401 Flow
PT401 COM
Furnace NO MeFC
P
Detector Back Pressure PT3
790 mmHg
Pneumatics
Controller
PV1
Connection Types
FS
Tool required Electronic Flow Sensor
Back One-touch Front
O-ring 3.0/0.8 LPM
Panel Panel
Figure 9-8
Manifold Flow Diagram (S844ES)
The valve states are viewable in the software, but they cannot be
changed. In the software, navigate to Instrument Ø Diagnostics Ø
Valve States.
Figure 10-95
22 Mntd Disp Port T/Screen Monitor 1 of 2
This section includes only the pages of the schematic drawing that
are applicable for service of the instrument.
R
Relative Error: The percent relative
error versus the certified composition
of the sample. Relative
Error=100*[(xi-xt)/xt] when xi is
Calculated Value and xt is Certified
Value.
Residual Error: Plots the residual
error versus the certified composition
of the sample. Residual
Error=Calculated-Certified.
A
Accessories List .............................................................................. 1–48
Adjusting Safety Pressure Regulator.................................................... 9–6
Air Filter, Maintaining ...................................................................... 6–55
Ambient Gauges ............................................................................. 3–13
Ambient Monitor Ranges .................................................................... 8–8
Ambient Parameters .......................................................................... 8–4
Ambient Parameter Definitions ........................................................... 8–6
Analyte Analysis, Disabling............................................................... 5–40
Analyzer, Installing ......................................................................... 2–12
Assembling and Reassembling Fittings............................................... 9–10
Assign Notify .................................................................................. 4–65
Auto Increment Sample Name .......................................................... 5–43
Automation
Auto Cleaner ......................................................................... 5–5
Auto Cleaner, Maintaining ..................................................... 6–28
Shuttle Loader ....................................................................... 5–6
B
Back Pressure Control...................................................................... 5–41
Balance
Configuring ......................................................................... 5–19
Installing ............................................................................ 5–13
Resetting ............................................................................ 5–21
Setting up Communication .................................................... 5–14
Testing ............................................................................... 5–21
Barometric Pressure, Calibrating ....................................................... 5–42
D
Daily Startup Procedure ................................................................... 3–28
Database
Backup ............................................................................... 5–22
Restoring ............................................................................ 5–23
Validating ........................................................................... 5–22
Declaration of Conformity Statement ................................................. 1–31
Detectors ....................................................................................... 5–27
Infrared (IR) Cells ................................................................ 5–28
Diagnostic Chart ............................................................................... 8–4
Diagnostics ...................................................................................... 8–1
Digital I/O...................................................................................... 8–10
Digital Input Definitions ........................................................ 8–14
Digital Output Definitions ...................................................... 8–11
E
Electric Installation.......................................................................... 2–18
Equipment Packages ....................................................................... 1–38
Error Reports.................................................................................. 5–31
Ethernet Adapter, Instrument ........................................................... 2–33
F
Facility Requirements (in Specifications) ............................................ 1–54
Filter Assembly, Cleaning ................................................................. 6–37
Firewall Settings, Configuring ........................................................... 2–35
Fitting Assembling and Reassembling ................................................ 9–10
Flexible Replacement Tubing .............................................................. 9–9
Flow ................................................................................................ 8–3
Flow Scrubber Tube Replacing .......................................................... 6–21
Front Panel Overview, Instrument ................................................2–6, 2–8
Furnace Out Tube, Maintaining ......................................................... 6–41
G
Gas Flow States ................................................................................ 9–3
Gas Dose Maintenance, Performing ................................................... 6–49
Gas Doser ...................................................................................... 5–43
Gas
Pedestal Down State ............................................................ 5–39
Standby Mode ..................................................................... 5–39
Standby Time ...................................................................... 5–39
Wake Up, Scheduling............................................................ 5–40
Gas Wake Up, Scheduling ................................................................ 5–40
Greasing O-Rings, and Cleaning.......................................................... 6–8
I
Illustrations.................................................................................... 10–1
Incoming Purifier Tube, Replacing ..................................................... 6–12
Incoming Scrubber Tube, Replacing .................................................... 6–9
Infrared (IR) Cells ........................................................................... 5–28
Installing
Analyzer ............................................................................. 2–12
Balance .............................................................................. 5–13
Computer............................................................................ 2–22
Electrical............................................................................. 2–18
Oscillator Tube .................................................................... 2–16
Pedestal.............................................................................. 2–17
Software ............................................................................. 2–34
Touch Screen Monitor ........................................................... 2–23
Vacuum Cleaner .................................................................. 2–20
Instrument Ethernet Adapter............................................................ 2–33
Instrument Front Panel Overview .................................................2–6, 2–8
Instrument Rear Panel Overview ....................................................... 2–10
L
Language Selection ......................................................................... 3–18
Leak Check .................................................................................... 8–15
Leak Check Troubleshooting ............................................................. 8–16
Lifting and Moving the Instrument....................................................... 2–5
Logging in
Blank.................................................................................. 3–31
Drift ................................................................................... 3–32
Gas Dose ............................................................................ 3–32
Sample ............................................................................... 3–32
Standard............................................................................. 3–31
Template ............................................................................ 3–33
N
Notice, Warranty............................................................................. 1–23
O
Optional Equipment......................................................................... 1–43
Optional Equipment Illustrations ....................................................10–101
O-Rings, Cleaning and Greasing.......................................................... 6–8
Oscillator Tube, Installing................................................................. 2–16
Override Conditions......................................................................... 7–12
P
Packing
Catalyst Heater Tube ............................................................ 6–17
Cellulose Trap Tube.............................................................. 6–19
Flow Scrubber Tube.............................................................. 6–22
Incoming Purifier Tube.......................................................... 6–13
Incoming Scrubber Tube ....................................................... 6–10
Part Numbers, Replacement Tubing..................................................... 9–9
Particle Filter, Replacing................................................................... 6–27
Q
Quality Control (QC), Configuring...................................................... 4–17
R
Rear Panel Overview, Instrument ...................................................... 2–10
Recalculating .................................................................................. 3–52
Registration Key
Error Messages .................................................................... 2–36
Troubleshooting ................................................................... 2–35
Updating............................................................................. 5–37
Replacement Tubing Part Numbers...................................................... 9–9
Replacing
Dust Eject Tubing................................................................. 6–44
Particle Filter ....................................................................... 6–27
Pneumatic Gas Filter ............................................................ 6–25
Rounding, Cornerstone® Brand Software............................................ 7–14
S
Safety Pressure Regulator Adjustment ................................................. 9–6
Sample
Analyzing ............................................................................ 3–53
Deleting.............................................................................. 3–54
Logging in ........................................................................... 3–32
Sample Filters ................................................................................ 3–56
Applying Saved.................................................................... 3–60
Comments .......................................................................... 3–59
Date................................................................................... 3–58
Deleting.............................................................................. 3–60
Description.......................................................................... 3–58
Excluded............................................................................. 3–58
Methods.............................................................................. 3–57
U
Updating
Registration Key .................................................................. 5–37
Software ............................................................................. 5–37
User Permissions ............................................................................ 4–51
V
Vacuum Cleaner
Installing ............................................................................ 2–20
Maintaining ......................................................................... 6–48
Manifold.............................................................................. 6–48
Valve States, Viewing ...................................................................... 9–19
W
Warranty ....................................................................................... 1–23
WEEE Statements ........................................................................... 1–24
Windows® Settings.......................................................................... 2–33
Wiring Diagrams, Electrical............................................................... 2–18