User's Manual
Please read this manual before using the product.
Please keep this manual for future references.
Proper usage of the product relies on fully understanding of the manual.
Head Office Address: Room 408, Building A, Huahan Innovation Park, Langshan Rd,
Production Address: Building 2, 1st, 2nd and 5th floor of Building 3, Liu Xian Er Rd, No.1,
Website: http://www.szangell.com
Edition:V1.0
1
Edition:V1.0
INDEX
REMINDER!.......................................................................................................................................4
NOTE:...................................................................................................................................................4
DISCLAIMER.........................................................................................................................................5
1 INTRODUCTION............................................................................................................................. 18
5.1 INSPECTION................................................................................................................................ 29
5.2 CLEANING.................................................................................................................................. 30
5.3 CHANGING BATTERIES.................................................................................................................. 31
2
Edition:V1.0
6.1 TRANSPORTATION........................................................................................................................ 32
6.2 MOVEMENT............................................................................................................................... 32
6.3 STORAGE....................................................................................................................................32
7 SERVICE INFORMATION.................................................................................................................32
8 SPECIFICATIONS.............................................................................................................................33
9 COMPONENT................................................................................................................................. 36
10 SIZE...............................................................................................................................................36
12 SERVICE GUIDE............................................................................................................................ 37
3
Edition:V1.0
Reminder!
This manual is the official documentation for the installation, configuration, operation and
maintenance of the Flat Panel Detector Angell-CELV-101C( 17"x14"). Please read this manual before
Note:
1. The product can only be operated by trained, professional medical or engineering personnel.
3. The product is to be used in connection with the computer and monitor which complies with
4. The product can only use the designated medical power adapter and dedicated detector cables.
5. The plug of the power adapter, as a disconnecting device for the network power supply, should
6. If this equipment is modified, appropriate inspection and testing must be conducted to ensure
7. Devices with E&M emitting capabilities, such as mobile phones, radio transceivers, or remote
controllers may affect the functionality of the product. Therefore it is strongly suggested that
these devices are not used in the same room of the product.
8. Please dispose the device material or components according to the local waste disposal
Copyright
SHENZHEN ANGELL TECHNOLOGY CO., LTD. reserves all rights.
TECHNOLOGY CO., LTD., this manual, in whole or part of, cannot be reproduced.
4
Edition:V1.0
Disclaimer
1. Malfunction or damage caused by usage not described by the manual.
product.
6. X-ray irradiation, image processing, image transferring and data archiving of this product shall
comply with the local regulations of the country or area where the product is installed. The user
7. The user shall be responsible for any inaccurate treatment of patients caused by inaccurate
8. SHENZHEN ANGELL TECHNOLOGY CO., LTD. reserves the right to modify the product
5
Edition:V1.0
labels. Ignorance of these warnings may result in harmful incidents. Please read through this manual
Safety Classification
Classification of protection against electric shock: Class I.
Safety Signs
This product, its labels and related documents using the following signs
Caution
RefertoInstructionManual/Booklet
Manufacturer
CE Mark
Serial Number
Alternating Current
Direct Current
6
Edition:V1.0
Storage Enveironment
Do not store the product to any equipment that may have liquid or chemical leakage or overflow, which may
cause liquid infiltration into the product and internal electric shortage.
. Caution
Avoid contact with any sharp object, which may cause damage to housing or paint.
Avoid moving or holding the product with wet hands or tools, which may cause electric shock.
Avoid violent impact or dropping of the product. Impact exceeding the limit may cause
Storage Temperature: -20℃- 55℃, Relative Humidity: 10%-90%(Non Condensing), Aire Pressure:
700hPa -1060hPa.
Electrical Safety
(in wireless mode). It shall not be powered by other sources, otherwise it may cause fire or electric
shortage.
Caution
Do not connect more than one product to a same AC outlet, otherwise it may cause fire or electric
shock.
Do not connect a portable multi-outlet socket into the system, otherwise it may cause fire or electric
shock.
The power cord plug must be firmly and safely inserted into the network power socket! If the contact fails,
or if a metal object comes into contact with the exposed metal plug, it may cause a fire or electric shock.
Before unplugging the power plug, make sure that the power switch of the product is in the "off"
(that is, in "○") state! Otherwise, the operator may suffer an electric shock and thereby endanger
7
Edition:V1.0
Be sure to hold the plug or connector to disconnect the power cord! If you pull the power cord, the
Do not occupy the space near the power line plug, so that in case of emergency the power plug can
be unplugged promptly.
Make sure the power plug and outlet socket are both well grounded. Otherwise it may cause electric
shock.
The product power supply should have a good indoor grounding socket! Otherwise, it may cause
Do not power on the product when it has condensation to prevent electric shock.
Turn off the product when it won’t be used for long time.
Installation environment
Caution
Do not install the product in any of the following places. Otherwise, it may cause product failure,
b) Exposed to sunshine;
f) Dusty environment;
The power and communication cables are fairly long. Please place these cables properly to
Non-medical equipments and the accessories of the product, such as the PC, monitor and power
8
Edition:V1.0
adapters, shall be placed at least 1.5m from the patient, as the following:
The patient shall not touch any of the buttons or connectors of the product when being used.
Caution
Before Exposure
Do not place any unrelated items on top of the product, which may fall and cause harm. Prevent
metal objects such as a needle or metal fragments, or liquid splashes from infiltrating into the
Do not violently impact or drop the product. The product may be damaged from violent impact or
drop. Damage to the external housing of the product may cause electric shock or fire.
Warming up in a cold and humid environment may cause condensation on the surface of the product. In this
case, the user shall wait until the product reaches thermal equilibrium and the condensation disappears.
Otherwise, the condensation may affect image quality. If any air conditioner is installed in the environment,
please adjust the temperature so that condensation does not appear on the product.
Do not exceed the weight load limit of the flat panel. Figure 1 shows the maximum limits of weight
9
Edition:V1.0
a b c
Figure 1
1. Uniformly distributed weight load on the top of the flat panel shall not exceed 135 kg, as shown in
Figure 1a.
2. Local static load, the bearing capacity is on the geometric center of the slab Ø40cm and the back
3. Pressure on the central Ø40cm area while the flat panel is only supported on two sides, shall not
Do not bend or twist the flat panel. Otherwise the internal imaging sensor may be damaged. When
placing the flat panel in vertical position, make sure to reliably fasten the flat panel. Otherwise the
flat panel may fall down and cause personnel injury and/or product damage.
During Exposure
During exposure, the patient being examined shall stay in a fixed pose. Any movement may affect the
During exposure, do not move or touch any cable of the flat panel to avoid causing image noise,
In the following situations, immediately turn off the power, unplug the power line and contact your
Cleaning
cause fire.
10
Edition:V1.0
Caution
Generally, the surface of the product does not directly contact patients’ skin. When it does, after
exposure, use a neutral, dry cloth to clean the surface that has contacted the patient.
Do not spray disinfectant or detergent to the front side of the flat panel.
Do not use diluents, benzene, acid, alkali, alcohol, etc., which may damage the product surface.
In case that the patient being examined is injured (with blood or body fluids), it is highly recommended to
Caution
Turn off the power and unplug the power line when the product need to be maintained or inspected.
Please clean the power line and other cables regularly. Disconnect all the cables, then use a dry cloth to
If the cables accumulates dust or dirt for a long time in a humid environment, the connectors of the cables
may corrode and have contact problem, which causes the product malfunction.
For safety concerns, all regular inspection in this manual shall be performed after the product is powered off.
Possible risks
Caution
Over limit weight load, improper connection,violent shock may cause risks such as severe image
artifact or ghost.
Not calibrating regularly may result in misdiagnosis due to defect pixels or defect lines.
Improper interface with X-ray equipment may cause functional fault and risks! Loose contact may
Incorrect relative position between X-ray field and the flat panel may result in error of the exposed
Improper x-ray loading factor may cause unsatisfactory image or harm to the patient. Exposure too
low may result in insufficient image quality; exposure too high may harm the patient being
examined.
Risk of incorrect X-ray dose. Different patient shall be assigned with different x-ray dose, otherwise
11
Edition:V1.0
Do not change the working principle of the X-ray machine. Otherwise, it will cause image quality
Medical electrical equipment requires special reminders about EMC and should be used according to the
This product requires special reminders about electromagnetic compatibility (EMC). Please use it
Portable and radio frequency (RF) communication equipment may affect this product.
Do not use products other than special accessories, otherwise it may increase radiation and reduce
deflection resistance.
Basic performance
The X-ray flat panel detector (4335A) is used in conjunction with the X-ray machine for medical institutions
to perform digital X-ray static imaging of the human body. The basic performance of the product is:
1. Realize the wired communication connection and command transmission between the flat panel
2. Realize the image acquisition of the flat panel detector (actually through the static acquisition of dark
forallEQUIPMENTandSYSTEMS
customer or the user of theAngell-CELV-101C Flat Panel Detector should as sure that they are used in suchand environment.
12
Edition:V1.0
Harmonicemissions
IEC61000-3-2 ClassA
Voltagefluctuations/flicker
emissions Complies
IEC61000-3-3
ortheuserofAngell-CELV-101CFlatPanelDetectorshouldassurethatitisusedinsuchanenvironment.
Electromagneticenvironment-
Immunitytest IEC60601testlevel Compliancelevel
guidance
Electrostaticdischarge :t6kVcontact :t6kVcontact Floorsshouldbewood,concreteor
(ESD) :t8kVair :t8kV air ceramictile.If floorarecovered
IEC61000-4-2 withsyntheticmaterial, therelative
humidityshouldbeatleast30%.
Electricalfast :t2kVforpowersupply :t2kV forpowersupply Mainspowerqualityshouldbethat
transient/burst lines lines ofatypicalcommercialor hospital
IEC61000-4-4 :t1kVforinput/output lines :t1kVfor input/outputlines environment.
Surge :t1kVdifferentialmode :t1kVdifferentialmode Mainspowerqualityshouldbethat
IEC61000-4-5 :t2kVcommonmode :t2kV commonmode ofatypicalcommercialor hospital
environment.
Voltagedips,short <5%UT <5%UT Mainspowerqualityshouldbethat
interruptionsand (>95%dipinUT) (>95%dipinUT) ofatypicalcommercialor hospital
voltagevariationson for0.5cycle for0.5cycle environment.Iftheuserofthe
powersupplyinput 40%UT 40%UT Angell-CELV-101CFlatPanelDetect
lines (60%dip inUT) (60%dipinUT) or requirescontinuedoperation
IEC61000-4-11 for5cycles for 5cycles
duringpowermainsinterruptions,it is
70%UT 70%UT recommendedthat the
(30%dip inUT) (30%dipinUT) Angell-CELV-101C
for25cycles for25cycles
FlatPanelDetectorbepowered
<5%UT <5%UT fromanuninterruptiblepower
(>95%dipinUT) (>95%dipinUT)
supplyorabattery.
for5sec for 5sec
Powerfrequency 3A/m 3A/m Powerfrequencymagneticfields
(50Hz/60Hz)magnetic shouldbeatlevelscharacteristicof
field
IEC61000-4-8 atypicallocationinatypical
commercialorhospital environment.
NOTE UTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.
13
Edition:V1.0
TheAngell-CELV-101CFlatPanelDetectorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomeror
Compliance
theuserofAngell-CELV-101C/FlatPanelDetectorshouldassurethatitisusedinsuchanenvironment.
Immunitytest IEC60601testlevel Electromagneticenvironment-guidance
level
Portable andmobileRFcommunicationsequipment
shouldbeusednoclosertoanypart
oftheAngell-CELV-101C
FlatPanelDetectorincludingcables,thanthe
recommendedseparationdistancecalculatedfrom the
Recommendedseparationdistance 푑=
ConductedRF 3Vrms 3V 3.5
[푉 ] 푝
IEC61000-4-6 150kHzto80MHz 1
3.5
푑=[퐸 ] 푝 80MHzto800MHz 푑 =
Radiated RF 3V/m 3V/m 1
IEC61000-4-3 80MHzto2.5GHz 7
퐸1
푝800MHzto2.5GHz
WherePisthemaximumoutputpowerratingofthe
transmitterinwatts(W)according tothetransmitter
manufactureranddistherecommendedseparation
distanceinmetres(m).
FieldstrengthsfromfixedRFtransmitters,as
a
determinedbyanelectromagneticsitesurvey, shoul
dbelessthanthecompliancelevelineach
b
frequencyrange.
Interferencemayoccur inthe vicinityof equipment
marked withthefollowingsymbol:
NOTE1 At80MHzand800MHz,thehigherfrequencyrangeapplies.
reflectionfromstructures,objectsandpeople.
a
Fieldstrengthsfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandland
mobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywith
accuracy.ToassesstheelectromagneticenvironmentduetofixedRFtransmitters,anelectromagneticsitesurvey
shouldbeconsidered.IfthemeasuredfieldstrengthinthelocationinwhichtheAngell-CELV-101CFlatPanelDetectorisused
exceedstheapplicableRFcompliancelevelabove,theAngell-CELV-101CFlatPanelDetectorshouldbeobservedtoverify
normaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasreorienting
orrelocating the Angell-CELV-101CFlat PanelDetector.
b
Overthefrequencyrange150kHzto80MHz,fieldstrengthsshouldbelessthan3V/m.
14
Edition:V1.0
Recommendedseparationdistancesbetweenportableandmobile
RFcommunicationsequipmentandtheEQUIPMENTorSYSTEM–
forEQUIPMENTorSYSTEMthat arenotLIFE-SUPPORTING
Recommendedseparationdistancesbetween
portableandmobileRFcommunicationsequipmentandtheAngell-CELV-101C Flat Panel Detector
TheAngell-CELV-101CFlatPanelDetectorisintendedfor useinanelectromagneticenvironmentinwhichradiatedRFdisturbances
arecontrolled.ThecustomerortheuseroftheAngell-CELV-101CFlatPanelDetectorcanhelppreventelectromagnetic
interferencebymaintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment
(transmitters)andtheAngell-CELV-101CFlatPanelDetectorasrecommendedbelow,accordingtothemaximumoutputpowerof
thecommunicationsequipment.
NOTE1 At80MHz and800 MHz,theseparation distance for the higher frequency range applies.
NOTE2 These guidelines maynot apply in all situations.Electromagnetic propagation is affectedby absorption and
reflection from structures,objects and people.
15
Edition:V1.0
Front type B
Back
16
Edition:V1.0
17
Edition:V1.0
1 Introduction
Angell-CELV-101C Flat Panel Detector (in the following referred to as “FP”) works
Features:
This manual covers the description and instructions of the using Angell-CELV-101C.
Flat panel detectors consist of scintillation screens and TFT panel. The scintillation
screen converts incoming x-ray energy into visible light photons. The TFT panel converts
light signal to electronic signal. Through charge amplification and A/D conversion, the final
The product shall be used together with medical x-ray equipment, for medical
The medical x-ray equipment shall have loading time at least 0.5s, and dose at least
50nGy.
After the digital x-ray imaging system installation completes, the system needs to be
tested and verified to ensure safety and functionalities. Installation, operation and
fluoroscopy.
18
Edition:V1.0
For medical facilities, for digital radiology of any human body parts.
1.2.4 Contraindication
Pregnant women, newborns, infants, critically ill patients, shall be treated according
to doctors’ instructions.
Angell-CELV-101C
Effective Size: Size code
product code
19
Edition:V1.0
Notes:
①FP Indicators.
POWER flash once every two seconds;Battery error (too hot) flash once per
FP is connected to PC through the detector cable and Ethernet cable through this
The lower left corner of the image corresponds to where the detector cable connector
20
Edition:V1.0
is.
FP PC
21
Edition:V1.0
Angell-CELV-101C has been designed for various applications of digital x-ray imaging. FP
PC with a wireless router (WR). The synchronization between image acquisition and X-ray
exposure can be realized through the internal AED function of FP. There are 3 major
(—Wired,…Wireless)
The pogo pin port of Detector cable (Figure a) connects to the detector (Figure b).
Ethernet cable (Figure c) and the power adapter cable connects to the other end of the
detector cable (Figure d).Connect the Ethernet cable to the computer and connect the
power line of the adapter to an AC outlet, and press the power switch to turn on the
detector. At this time the detector is powered by the adapter, andthe lithium battery isbeing
charged.
22
Edition:V1.0
Directly start the power switch button on the flat-panel detector, the device into
1. LAN settings
Search for the WIFI with angell_7628_FPD-XXXX (suffix) in the wireless network,
connect WIFI, The password has to be “12345678” ,the IP address of the wireless
2. If unable to connect,manualsettings
Open the browser, enter the IP address:172.16.2.1, and enter the username and
23
Edition:V1.0
3.4 PC Configurations
24
Edition:V1.0
For the connection configurations ①,and ③ ,at least one Ethernet card need to be
installed in the PC. The Ethernet card shall be connected to WR’s LAN port or FP, and its
Please locate the SDK release file SDK_Vx.x.x.zip in the SDK DVD enclosed in the
product package. Unzip the file and copy the folder to PC to use. The SDK release file
After installation of SDK, the user can start to use TestUI to connect to FP and
acquire images (Please refer to section 4). The software developers can perform
functions supplied in the SDK. Please refer to the SDK Interface Manual included
(enclosed in the SDK release file) for detailed software developing instructions.
The firewall or anti-virus software installed on the user’s PC, may keepPZMedical’s
TestUI.exe, or other programs calling PZMedical’s SDK, from accessing the network data
and cause TestUI not functioning properly. The user may need to contact the software
supplier on how to allow certain programs such as TestUI to access the network, or, if
25
Edition:V1.0
library files.
4.2 FP Start-up
power switch for 3 seconds to turn on the FP.After startup, Power light will turn Blue.
1) Open the DXRay TestUI folder, double click DXRay TestUI.exe to run. From TestUI's
window , After connection is successful, The status of FP is shown in the right down
After FP is properly opened, user can click on the "Exposure mode" button to
enter the Exposure mode. click on the "FP matrix" button to choose the FP matrix.Then
can Then directly fire x-ray or click on the "Gathering dark figure"buttons in the TestUI to
26
Edition:V1.0
get an image.
After receiving the image, the image is displayed on the left side. Coordinate and
Pixel value is displayed in the status display bar in the lower left corner. as the following:
27
Edition:V1.0
Select the file under the folder where DXRay TestUI.exe is located, then open the
28
Edition:V1.0
4.7 FP Shutdown
FP in the boot state , press and hold the power switch 3 seconds can enable FP
shutdown. The POWER light and the LINK lightsand the READY lights will turn off.
5.1 Inspection
To ensure safe and proper use of this equipment, please inspect it before use.
If any problem is found during the inspection, please follow the instructions in this
chapter.
Result
Inspection Resolution
Date Date Date
Check if the cable is
Contact PZM or
damaged, check if the cable OK/NG OK/NG OK/NG
Distributors
external surface is broken.
Cables
Reconnect the
Check all the connections OK/NG OK/NG OK/NG connectors and
fasten them.
Check if the surface is
Flat Contact PZM or
damaged and/or connection OK/NG OK/NG OK/NG
Panel Distributors
is loose.
29
Edition:V1.0
Result
Inspection Resolution
Date Date Date
Reconnect
Check if the power LED
OK/NG OK/NG OK/NG power cable and
lights up.
detector cable
Take note of
General
Fire x-ray, check if x-ray error message if
exposes correctly and OK/NG OK/NG OK/NG there is any.
image acquired normally Contact PZM or
Distributor.
Regular inspection shall be carried out at least once a year. In case of any problem
during the inspection or the user cannot perform the inspection, please contact PZM or
the Distributor.
Result
Inspection Resolution
Date Date Date
Expose x-ray and examine Any problems
the image quality and all OK/NG OK/NG OK/NG contact PZM or
functions. Distributor
5.1.3 Re-Calibration
When the exposure condition has significantly changed (such as change of tube, or
large variation of the distance between flat panel and tube), the flat panel need to be
As the product being used for long time, the calibration template may drift and there
may exist new defects. Therefore, the product shall be re-calibrated at least once every 6
5.2 Cleaning
Generally, the surface of the product does not directly contact patients’ .
Caution
When it does, after exposure, use a neutral, dry cloth to clean the
surface that has contacted the patient. In case that the patient being
30
Edition:V1.0
Turn off the power and unplug the power line before cleaning the
Caution
product. Do not use alcohol, benzene, diluents and any other
If the housing surface is contaminated with dirt, please take the following steps to
clean:
Soak a cloth in water-diluted neutral detergent (5%), wring it out and wipe the
month.
After cleaning with neutral detergent, make sure to wipe it again with a cloth
The product must use the dedicated Lithium battery supplied by PZM,
Caution
Usage of any other battery may cause abnormal temperature rise
recharge the battery once every half year to prevent battery leakage
31
Edition:V1.0
6.1 Transportation
6.2 Movement
Pay attention to the product when moving it. Lift and place it gentally.
Caution
The product shall be placed at the bottom of any package
containers. Do not drop the product from height. Do not place the
any other items on top of the product. Protect the product against
6.3 Storage
7 Service Information
Under normal use and regular inspection and maintenance, the product expect to
7.2 Repair
Ifit is necessary for the maintenance personnel, MANUFACTURER will provide circuit
32
Edition:V1.0
This product’s key components (components that are necessary to maintain the
product’s functionalities) will be available for two years after the product becomes
obsolete.
8 Specifications
33
Edition:V1.0
Input:100-240VAC,50/60Hz
Output:+19VDC,4.74A
Humidity:30-75%(non-condensing)
Atmospheric pressure:700hPa
-1060hPa
Humidity:10%-90%(non-condensing)
Atmospheric pressure:700hPa
-1060hPa
FP Size 460mm×384mm×15.5mm
8.2 Characteristic
8.2.1 MTF
34
Edition:V1.0
8.2.2 DQE
35
Edition:V1.0
9 Component
Standard Parts
Flat Panel Detector____________________1
compact disc________________________1
10 Size
Flat Panel Detector
Unit:mm
36
Edition:V1.0
12 Service Guide