Service Manual - AE-600N-new Version
Service Manual - AE-600N-new Version
ANALYZER
AE-600N
SERVICE MANUAL
3. Testing Procedure 2
4. MCB Board 3
7. Photometer Assembly 6 to 7
8. Trouble Shooting 8 to 13
2
SAFETY NOTES TO USERS
Be sure to read "Safety notes to users" before you use this instrument for operation.
Symbols and their meaning are as follows:
Read the statement following the symbol. The statement is alerting you to a
BIOHAZARD
potentially bio-hazardous condition.
WARNING
About diagnostics
Clinical diagnosis based on the measurement result and its treatment must to be
determined together with clinical symptoms and other checkup results by a doctor in
charge of the patient.
Take ground
Unless it is done ,it may cause an electrical shock when breakdown and electrical power
leaking occure.
3
Power unit outlet should be independent
Place instrument where independent power receptacle can be used. use receptacle
different from the one used by a device that easily generate noise such as a centrifuge
etc.
Never plug or unplug with wet hands.It may cause electrical shock.
Pull a plug out by holding a plug with two fingers , or it may cause electrical shock or
short to bring about fire.
Never use other cords and cables than the designated ones.
Cords and cables with less electrical capacity may cause fire.Do not step on,twist or pull
cords.Never use damaged power cords.
4
BIOHAZARD
Mishandling of samples
5
Disposal of waste
Disposal of the waste in accordance with as and on applicable rules for Bio-hazard waste
disposal and consulting the manufacturer or distributor of the reagents for details.
CAUTION
Use and operate the unit in accordance with enclosed "User's guide"
Warranty:
Warranty does not cover any defect, malfunctioning or damage due to: a) Usage under
the condition prohibited by the instruction. b) Customer does not do proper maintenance
required by the instruction manual. c) Customer disassembles the instrument without
permission from manufacturer/selling agencies d) Usage of generic reagents. e)
Instrument opened or serviced other than authorized person.
Note: Instrument can not be installed without proper Stabilized voltage, Earthing and
Proper environmental conditions. If fail to fulfil or providing false information will be
harmful for the instrument.
6
Steps for Installation
1.Check groundings of the power supply:
Before Installing the Analyzer, it is necessary to check for good mains supply.
It is important to have a good earthing since poor earthing causes variation in the
results which will affect the performance of the machine.
Phase (Left side in the socket) to Neutral (Right side in the socket) = 110/220Vac (±
10Vac) without fluctuation.
Phase (Left side in the socket) to Earth (Top side in the socket) = 110/220Vac (± 5Vac)
without fluctuation.
Neutral (Right side in the socket) to Earth (Top side in the socket) = not more than
± 5Vac.
Caution:
Do not use strong Acidic or Alkaline solutions which can cause irrepairable damage to
the instrument.
Do not leave the Flowcell dry. Always leave it filled with distilled / deionised water.
7
3.Testing procedure :-
I. SMPS:-
Input Voltage :- 100 to 240Vac, 50/60Hz, ±10%
Output Voltage :- 12.3Vdc, +0.2Vdc
Mains fuse :- 2A, 250Vac, (Slow blow)
II. Main Controller Board (AGDSAMCB-0C/1D/1E)
Test point voltages
Sr. No. TP # (w.r.t. GND) Voltage range Condition
1. 5V 4.9 to 5.3Vdc Power ON, LED D21 glow.
2. 3V3 3.1 to 3.3Vdc Power ON, LED D20 glow.
After booting the Instrument. LED D27
3. LV (Lamp) 5.5 to 6.1Vdc
glow.
2.9 to 3.0Vdc “Check absorbance” / “Analyze” mode,
4. PV (Peltier) (Toggling state) LED D26 & D15/D16 glow.
After booting the Instrument or during
5. 8V 7.5 to 8.2Vdc Printout, LED D22 glow.
6. Coin Cell 3Vdc Always
During stepper motor locked/rotate. LED
7. Across R2/R3 0.1 to 0.4
D1/D3 & D5/D8 LED glow.
> 3V Filter wheel plate slot not in the opto path.
8. FWO (J3) Pin 2 Filter wheel plate slot in the opto path,
< 0.5V LED D36 glow
> 3V Aspiration switch without press position.
9. OPT (J15) Pin 2 Aspiration switch press position, LED D37
< 0.5V glow
During peristaltic dc pump on, LED D9
10. Across V1 (J4) > 11V glow.
< 0.5V Aspiration switch press position.
± 2.9 to 3.1V
(AGDSAMCB-0C) “Check absorbance” / “Analyze” mode.
± 2.9 to 7V Cooling action - LED D13 & D16 glow.
Across V4/PLTR
(AGDSAMCB- Hot action - LED D15 glow
11. (J8) 1D/E)
Exit from “Check absorbance” / “Analyze”
< ±0.2V mode.
LED D26 & D15/D16 off.
Thermal PRT. (J21)
12. 4.9 to 5.3Vdc After booting the Instrument.
Pin 2
Thermal PRT. (J21) After booting the Instrument or during
13. Pin 3 7.5 to 8.2Vdc Printout.
Thermal PRT. (J21) After booting the Instrument or during
14. -5.0 to -12.0Vdc
Pin 4 Printout.
15. PHM (J23) Pin 1 3.1 to 3.3Vdc Power ON.
16. PHM (J23) Pin 2 3.1 to 3.3Vdc Power ON.
17. PHM (J23) Pin 3 3.1 to 3.3Vdc Power ON.
18. USB (J22) pin 1 & 2 4.9 to 5.3Vdc Power ON.
19. USB (J22) pin 3 & 4 < 0.5V Device (Keyboard / Mouse) not connected.
8
> 2.7V Device (Keyboard / Mouse) connected.
9
4.Replacement of tubings:
Replace the aspiration and peristaltic pump tubing if it shows signs of wear and tear.
10
5.Replacement of peristaltic Pump:
Note: After replacement of the Pump, it is essential to recalibrate the peristaltic pump.
11
6.Photometer Assembly:
Lamp Assembly :
NOTE:
Lamp voltage should be 5.5 – 5.8 Volts.(More voltage will affect the lamp life.)
Lens adjustment should be disturbed only if the counts are not adjusted by lamp
adjustment.
12
Trouble Shooting for Instrument:
If machine crashes :- Internal action not completed as expected during sample analysis
or while saving the desired changes.
Could not Save Date & Time :- Date & Time Could not be set to RTC.
NOTE: If the above things happen then attach a pendrive, go to 'Set Up' option, Copy
Log and send to HO.
13
Troubleshooting
Problem Cause Action
Instrument doesn’t Mains supply not proper. Loose connection at Mains
ON or ON-OFF switch/socket/mains cord.
frequently. Loose connection.
Fuse absent/blown.
Display touch not Display over tightened to Slightly loosen the display fitting.
working. the instrument body.
Perform touchscreen calibration using
FPC cable connection not a mouse.
proper.
Refit/ change FPC.
Display not calibrated.
Display working not FPC cable connection not Slightly tighten the display.
proper. proper.
Refit / change FPC.
Due to vibrations
developed because of Don't place any other device
filter wheel rotation or generating high frequency waves near
some other high the analyzer.
frequency device placed
on the same platform. Change display
Date / Time changes Coin battery life may be Change the coin battery.
after power ON over.
Clean the clip contact / re-solder socket
Loose connection in battery pins.
socket clip/solder pins.
Fluctuation in test Noise in Mains Earth. Provide noise free Earth mains supply.
results.
Bubble in liquid path. Clean flowcell window externally.
14
Problem Cause Action
Liquid not aspirated Micro switch not working Check if the peristaltic pump tubing is
when the connected properly, if not reconnect
ASPIRATION correctly.
SWITCH is pressed
& no buzzer sound. Perform Pump Calibration.
In Self test / Bubble in Flowcell. Check the liquid path and flowcell
Photometer tubing fitting for bubble.
Calibration / In Photometer gain count may
Analyze ( high water be less.Perform Clean the external walls of flowcell with
blank absorbance) Photometer calibration if tissue paper.
done long back.
OR Clean flowcell thoroughly with cleaning
Filter wheel position not solution followed by deionised water.
Error related to filter / proper.
clean flowcell path. Check filter wheel opto setting.
Low intensity of lamp.
Check filter wheel motor connection.
Damaged Filter
Check the Lamp pcb soldering.
Liquid not moving Liquid path may be Ensure that pump tubing and
properly. blocked. aspiration tubing are properly
OR connected.
Peristaltic pump Flowcell or pump tube
operates but liquid is may be punctured. Check tubings for blockage, crimps, or
not aspirated. leaks.
Pump rotation not proper.
Check pump motor connector
connection.
Replace tubings.
USB Mouse/ USB port may not be Check USB port fitting.
Keyboard/ Pendrive properly fitted.
not working. Check USB port connector.
Check USB PCB
Check USB dry solder or replace it.
MCB board may be
damaged. Replace MCB board.
15
16
Problem Cause Action
The graph of a 'Graph' option not Select 'Graph' option as soon as
Kinetic chemistry is selected entered in 'Sample' screen to get graph
not printed along with printed for each sample.
result.
Printer printout is Printer roll may be old/ Replace printer roll.
faint. bad quality.
OR Make sure that the paper roll is
Printer does not work. Printer roll not properly installed properly with the heat
installed. sensitive surface up.(the instrument
will not print if the paper is fed
reversed.)
Temperature attained Flowcell not inserted Ensure that the flowcell is completely
in the flowcell is not properly in its slot. inserted in its slot.
within ±0.1ºC of the
selected temperature. Reagents were not Refrigerated reagents should be
OR allowed to come to room allowed to come to room temperature
Flowcell Temp not temperature before use. before use.
proper.
Temp pcb not connected Press and fit Temp pcb & tight the
properly. fitting screw.
17
18
Poorly reproducible Aspiration of sample not Check if the aspiration volume is
or erratic readings. done properly. correctly programmed and whether the
amount of liquid aspirated is as
Contamination in programmed.
reagents.
If it is found that the aspiration is not
Contaminated test-tubes / proper, perform pump calibration.
pipette tips.
Check expiry date of the reagents and
control material used.
19
20
Error messages / Causes Corrective Action
Flags
Temperature not Getting lower or higher Check peltier.
attaining results for kinetic or fixed
time chemisties. Check temperature PCB contact with
flowcell housing
Fluctuations observed for
temperature displayed in
analyze mode / check
absorbance mode.
'SD' flag shown Non-linearity or deviation in Check for micro bubbles.
graph.
Sample may not be properly aspirated.
21
Error messages / Causes Corrective Action
Flags
RH flagging If reagent blank absorbance Problem with the reagent.
is greater than Reagent Abs.
Low in Test Parameters. Check for reagent expiry date and
reconstitution stability (if applicable).
22
AE-600N Spare Parts
Sr no. Spare Part Name Item Code Description
1 Lamp Assembly AA000084
12 Stepper Motor(Filter
Wheel) EA000009 EA STEPPER MOTOR ASSEMBLY
27
Power entery module EA000005 MAIN SWITCH AND SOCKET ASSEMBLY
28 Fan EA000003 EA FAN ASSEMBLY
29 Photometer PCB PA000013 PCB ASSEMBLY PHOTOMETER 2020
30 Photometer AA000011 PHOTOMETER ASSEMBLY
31 Temperature PCB PA000009 PCB ASSEMBLY TEMP SENSOR 2020
32 PT Adaptor (For
Coagulation Test) AP000091 Coagulation tube adaptor
33 Cuvettes – 10 Nos. MB000083 CUVETTES
23
34 VARIABLE VOLUME MICRO PIPETTE (5-
Micro Pipette 5-50 FG000031 50ΜL)
24