Green Line Transformer

The document summarizes an ES710-GL series isolating transformer for single-phase medical IT systems. Key features include a 30% reduction in heat dissipation loss and inrush current less than 8 times the rated current compared to standard models. Models are available for horizontal or vertical installation and include built-in temperature sensors according to standards.

Uploaded by

wakas
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
128 views

Green Line Transformer

The document summarizes an ES710-GL series isolating transformer for single-phase medical IT systems. Key features include a 30% reduction in heat dissipation loss and inrush current less than 8 times the rated current compared to standard models. Models are available for horizontal or vertical installation and include built-in temperature sensors according to standards.

Uploaded by

wakas
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 4

ES710-GL Series

Isolating transformer
for single phase medical IT systems
with -30%* heat dissipation loss and inrush current IE < 8 x In

Energy efficient Green Line (GL) Series

ES710_D00109_03_D_XXEN/10.2016
(*) compared to the Standard Series
Isolating transformer Single-phase isolating transformers
for the design of medical IT systems
ES710-GL

Application and description


The transformers of the ES710 series have reinforced insulation and comply with the
requirements of IEC 61558-1/DIN EN 61558-1 (VDE 0570-1) and IEC 61558-2-15/DIN EN
61558-2-15 (VDE 0570-2-15).
In addition, the transformers comply with the requirements of IEC 60364-7-710/DIN VDE
0100-710 (VDE 0100-710) for IT systems in medical locations. The windings are galvanically
isolated. In order to minimize electrical interferences, an electrostatic screen is installed
between the primary and secondary winding the lead out of which is connected to an
insulated terminal for connection to the equipotential bonding.
The fixing angles are isolated from the transformer core in order to guarantee an isolated
installation to comply with the requirements of DIN VDE 0100-710 (VDE 0100-710), para.
710.512.1.6.2).
The transformers are available for horizontal and vertical installation. Protection against
corrosion is guaranteed by a complete resin impregnation.
ES710-GL The transformers are designed for use in dry locations. Transformers of the SK2 series are
completely encapsulated and fulfil the requirements of protection class II.
Device features
Frequency/ratings
• Built-in temperature sensors acc. to The transformers are designed for rated frequencies of 50…60 Hz. The values specified in
DIN 44081 (120 °C) the chapter “Technical data” refer to a maximum ambient temperature of 40 °C and a rated
• Screen winding with brought-out frequency of 50 Hz.
insulated connection terminal
• Insulated mounting angles Temperature rise
Free air circulation must be ensured. An ambient temperature exceeding 40 °C will reduce
• Degree of protection, IP00 (open design)
the rated power. For temperature monitoring, a PTC thermistor is placed on each transformer
• Degree of protection, IP23 leg and the leads are connected to the terminals.
(with enclosure)
• Protection class I Enclosure
• Protection class II Appropriate steel sheet enclosures, degree of protection IP 23, are available for all standard
(option: encapsulated version) types of isolating transformers.
• Reinforced insulation Rated power
• Classification of insulation: ta40/B According to DIN VDE 0100-710 (VDE 0100-710), the rated power of the transformer shall
• Connections: screw terminals not be less than 3.15 kVA and shall not exceed 8 kVA.
• Noise level < 35 dB (A) Overload protection
(no-load and nominal load)
When isolating transformers are used to form a medical IT system in accordance with
• Vector group: IiO para. 710.512.1.6.2 of DIN VDE 0100-710 (VDE 0100-710), overload protection is not
• Inrush current IE < 8 x In permitted. In this case, short-circuit protection is required. That means, emphasis is
focused on the availability of the power supply; it is therefore essential to avoid discon-
Approvals nection on the occurrence of transient overload. The protection of isolating transformers
against overload and overtemperature can be realized by using monitoring devices in
accordance with para. 710.531.3.1. The appropriate fuses for short-circuit protection can
be selected from the table “Technical data”.

Standards
VDE test mark for ES…S-GL types,
ES710 isolating transformers comply with the device standards and the regulations for
SN-GL types are not VDE certified,
installation: DIN EN 61558-1 (VDE 570-1), IEC 61558-1, DIN VDE 0100-710 (VDE 0100-710),
DIN EN 61558-2-15 (VDE 0570-2-15), IEC 61558-2-15, IEC 60364-7-710.

Hazard warning:
When performing installation work in the environment of the transformer, it has to
be ensured that the insulation coordination of the transformer is not influenced in a
negative way.
For example, no ferromagnetic and conductive metal swarf may fall down close to the
transformer. This can interfere with the function and the dielectric properties, especially
after being turned on. The environment of the transformer must be kept free from
such particles during the entire operating time and controls must be carried out at
regular intervals.

2 ES710_D00109_03_D_XXEN/10.2016
Isolating transformer ES710

Nameplate

ES: single-phase isolating transformer


DS: three-phase isolating transformer
Nominal power
ES710/3150SN-GL
Type series S-GL = terminals front side (energy efficient)
SN-GL = terminals front side, low type of construction
(energy efficient)

Wiring diagram Terminal diagram


S-GL series, SN-GL series

FE

Dimension diagram/type of construction


S-GL series SN-GL series
Dimension E is the depth incl. terminal

Ordering information

Dimensions (mm) Cu weight Weight Core Type Art. No.


A B C D E F G H (kg) (kg) U/I
280 180 370 240 290 145 11 x 28 15 49 180/93 ES710/3150S-GL B 9209 0061
280 150 420 240 290 115 11 x 28 24 59 210/63 ES710/4000S-GL B 9209 0062
S-GL series

280 160 420 240 290 125 11 x 28 25 61 210/73 ES710/5000S-GL B 9209 0063
280 175 420 240 290 140 11 x 28 26 65 210/88 ES710/6300S-GL B 9209 0064
280 190 420 240 290 155 11 x 28 27 74 210/103 ES710/8000S-GL B 9209 0065
280 207 420 240 290 172 11 x 28 33 85 210/120 ES710/10000S-GL B 9209 0066
240 180 310 200 280 145 11 x 28 < 200 15 49 180/93 ES710/3150SN-GL B 9209 0121
280 150 360 240 320 115 11 x 28 < 170 24 59 210/63 ES710/4000SN-GL B 9209 0122
SN-GL series

280 160 360 240 320 125 11 x 28 < 180 25 61 210/73 ES710/5000SN-GL B 9209 0123
280 175 360 240 320 140 11 x 28 < 195 26 65 210/88 ES710/6300SN-GL B 9209 0124
280 190 360 240 320 155 11 x 28 < 210 27 74 210/103 ES710/8000SN-GL B 9209 0125
280 205 373 240 330 175 11 x 28 < 205 33 85 210/120 ES710/10000SN-GL B 9209 0126

ES710_D00109_03_D_XXEN/10.2016 3
Technical data4(-4/(-TFSJFT

Type ES710/3150 ES710/4000 ES710/5000 ES710/6300 ES710/8000 ES710/10000

Power/voltages/currents
Rated power 3150 VA 4000 VA 5000 VA 6300 VA 8000 VA 10000 VA
Rated frequency 50…60 Hz 50…60 Hz 50…60 Hz 50…60 Hz 50…60 Hz 50…60 Hz
Rated input voltage AC 230 V AC 230 V AC 230 V AC 230 V AC 230 V AC 230 V
Rated input current 14.2 A 18 A 22.5 A 28.5 A 36 A 45.3 A
Rated output voltage AC 230/115 V AC 230/115 V AC 230/115 V AC 230/115 V AC 230/115 V AC 230/115 V
Rated output current 13.7 A 17.4 A 21.7 A 27.4 A 34.7 A 43.5 A
Inrush current IE GL version < 8 x În < 8 x În < 8 x În < 8 x În < 8 x În < 8 x În
Leakage current ≤ 0.5 mA ≤ 0.5 mA ≤ 0.5 mA ≤ 0.5 mA ≤ 0.5 mA ≤ 0.5 mA
No-load input current i0 GL version ≤2% ≤2% ≤2% ≤2% ≤ 2% ≤ 2%
No-load output voltage u0 ≤ 236 V ≤ 233 V ≤ 234 V ≤ 235 V ≤ 233 V ≤ 233 V
Short-circuit voltage uk ≤3% ≤3% ≤3% ≤3% ≤3% ≤3%
Environmental conditions
Ambient temperature ≤ 40 °C ≤ 40 °C ≤ 40 °C ≤ 40 °C ≤ 40 °C ≤ 40 °C
No-load temperature rise ≤ 22 °C ≤ 22 °C ≤ 26 °C ≤ 31 °C ≤ 33 °C ≤ 36 °C

ES710_D00109_03_D_XXEN / 10.2016 / pdf / © Bender GmbH & Co. KG, Germany – Subject to change! The specified standards take into account the version that was valid at the time of printing.
Full-load temperature rise ≤ 55 °C ≤ 53 °C ≤ 62 °C ≤ 67 °C ≤ 76 °C ≤ 65 °C
Noise level (under no-load conditions and nominal load) ≤ 35 dB(A) ≤ 35 dB(A) ≤ 35 dB(A) ≤ 35 dB(A) ≤ 35 dB(A) ≤ 35 dB(A)
Other
Insulation classification ta40/B ta40/B ta40/B ta40/B ta40/B ta40/B
Degree of protection IP00 IP00 IP00 IP00 IP00 IP00
Protection class I/II* I/II* I/II* I/II* I/II* I/II*
Core U/I GL version 180/93 210/63 210/73 210/88 210/103 210/120
Recommended use when
used in accordance with DIN VDE 0100-710 25 A gL/gG 35 A gL/gG 50 A gL/gG 50 A gL/gG 63 A gL/gG 80 A gL/gG
Recommended use when used in accordance
with DIN VDE 0100-710 GL version 25 A gL/gG 25 A gL/gG 35 A gL/gG 50 A gL/gG 50 A gL/gG 63 A gL/gG
Induction 0.86 T 0.94 T 1T 1.05 T 1T 1.1 T
Rprimary 0.255 Ω 0.135 Ω 0.100 Ω 0.080 Ω 0.064 Ω 0.050 Ω
Rsecondary 0.230 Ω 0.110 Ω 0.095 Ω 0.070 Ω 0.056 Ω 0.036 Ω
Efficiency 95 % 96 % 96 % 96 % 96 % 96 %
Documentation number: D00109
Loss at 20…22 ° C ambient temperature
Fe loss (iron loss) GL version < 18 W < 19 W < 25 W < 33 W < 37 W < 42 W
Cu loss (copper loss) GL version < 95 W < 80 W < 100 W < 125 W < 160 W < 210 W
Heat dissipation loss at 40 ° C ambient temperature and 100 % continuous load
Heat dissipation loss GL version < 135 W < 115 W < 145 W < 185 W < 230 W < 295 W
* Option: completely encapsulated version

Bender GmbH & Co. KG Bender Italia Srl


P.O. Box 1161 • 35301 Gruenberg • Germany Via Piero Martinetti 6 • 20147 Milano • Italy