Electron Spin Resonance
Electron Spin Resonance
Physics Department
Measurement of the Landé factor using the Electron Spin Resonance − Lab 6∗
+ 3/2
Learning Objectives: Measurement of the Landé fac-
tor using AC and DC methods. Understand how the 2P
+ 1/2
3/2
coil size is essential for the experiment. Understand the − 1/2
purpose of a calibration of an instrument and how it is − 3/2
applied to data (arc calibration). Estimation of the sys-
tematics. + 1/2
2P
1/2
− 1/2
I. PHYSICAL BACKGROUND
III. ELECTRON SPIN RESONANCE • The potential difference, V , across the 1-Ω mea-
EXPERIMENTS suring resistor is proportional to the coil current,
I, and therefore to the magnetic flux density B, of
In order to determine the Landé factor g for the free the almost homogeneous magnetic field B,~ between
electron using Electron Spin Resonance, one uses the the Helmholtz coils.
resonance formula (9) which is satisfied when resonance
occurs. The correlation between the magnetic field B,
which causes the splitting of the electron energy levels, 1. Resistor Calibration
and the frequency f of the photon to be absorbed by
the electron, which causes a transition between the two The 1-Omega-resistor is the crucial element of both AC-
energy levels, is linear. Thus, if we vary B (or f ), and and DC- experiments. Using a simple circuit from Fig. 5,
4
MHz
Digital Counter
ESR basic
unit
ESR Adapter
Connections of the Helmholtz coils
f
1000
+12V
0
−12V Y
Z A
VARIAC Oscilloscope
+ transformer 1Ω
Ch 1 Ch2 EXT
DVM
FIG. 4: Schematic diagram of the setup for AC-method measurement of the Electron Spin Resonance. The blue solid lines
show connecting cables. The Voltage-signal across the measuring 1-Ω resistor can be connected either to the EXT, or to Ch2,
input of the oscilloscope. Connection to Ch2 is a better choice since it gives a possibility to observe the Voltage signal on the
scope.
measure it resistance. Do several (5 - 7) measurements along this axis the magnetic field is homogeneous.
vs. different Voltage or different well known resistance of
the high quality resistor. 3. Turn the Variac knob to 0 V and disconnect it from
the power network, if found connected.
a 200 Ω c b 4. Connect the apparatus, as shown in Fig. 4.
R Rx 5. Call the instructor to check if the setup is connected
correctly.
9. Turn the frequency on the ESR basic unit to min- 16. Continue increasing the frequency f in steps of
imum (this is approximately 20 MHz) and keep it 5 MHz till you reach the maximum possible value
constant. (approximately 80 MHz), and for each f repeat the
procedure from the previous item.
10. Decrease slowly the Variac Voltage, i.e., the AC
current through the Helmholtz coils, till resonant 17. Insert the small coil with the DPPH sample in, and
signals are observed on the oscilloscope. repeat the measurements as for the medium coil
starting from the lowest available frequency and
11. Tune finely the amplitude of the AC Voltage, so going to the highest available frequency in steps
that the minima (resonances) of the oscillating- of 5 MHz.
circuit signal are equally placed with respect to
18. Insert the large coil with the DPPH sample in, and
each other (see Fig. 7). When this is done, the Volt- repeat the measurements as for the medium coil
age shown by the DVM is directly proportional to
starting from the lowest available frequency and
the resonant magnetic field intensity B which sat- going to the highest available frequency in steps
isfies the resonant formula (9) for the chosen fre-
of 5 MHz.
quency f .
VRF
0
V 1Ω
Vres
V (a.u)
−Vres
0
−V 1Ω
V 1Ω
0 0
V 1Ω = V 1Ωsin2πνt = V 1Ωsinφ , ν=60 Hz
t (a.u), φ (a.u)
FIG. 6: Diagram showing the Voltage signals on the oscilloscope when resonance takes place. Black is the Voltage signal
from the oscillating circuit, VRF ; red is the alternating magnetic field (the Voltage across the measuring resistor, V1Ω ). Note:
resonance is observed no matter if the resonant minima are equally spaced.
the almost homogeneous magnetic field B,~ between 10. Switch the power on on these components which
the Helmholtz coils. The power to all the electronic require it (do not forget to turn on the ESR basic
modules should be OFF. For the Variac power sup- unit).
ply this means the knob turned to 0. Begin with
the medium coil and the DPPH sample placed in 11. Set the frequency on the ESR basic unit to ∼ 40
the magnetic field of the Helmholtz coils. MHz. Increase the AC amplitude by few small
divisions.
4. Connect the medium plug-in coil to the ESR basic 12. Start to increase slowly the amplitude of the
unit, and insert the DPPH sample inside it. DC Voltage by turning its Voltage-regulating knob
clockwise. Observe on the scope the Voltage signal
5. Mount the ESR basic unit in such a way that the
from the oscillating circuit and look for the reso-
DPPH sample is in middle of the coils and is per-
nance signal.
pendicular to their common axis. This is because
along this axis the magnetic field is homogeneous. 13. Once the resonance signal is observed, adjust the
Variac Voltage to some small amplitude so that the
6. Turn the Variac knob to 0 V and disconnect it from edges of the resonant signal have small, but finite
the power network, if found connected. lengths. Adjust the scope so that the edges of the
resonant signal are completely symmetric with re-
7. Connect the apparatus, as shown in Fig. 4.
spect to the central vertical line of the screen.
8. Call the instructor to check if the setup is connected 14. Turn the frequency on the ESR basic unit to min-
correctly. imum (this is approximately 20 MHz) and keep it
constant.
9. Check that all the components which need it
are connected to the power network (oscilloscope, 15. Decrease slowly the DC Voltage, i.e., the DC cur-
power supplies, DVM, frequency counter. rent through the Helmholtz coils, till the resonant
7
V RF
V res
V (a.u)
−V res
rms V 1Ω
V 1Ω = V res
t (a.u.), φ (a.u.)
FIG. 7: Diagram showing the Voltage signals on the oscilloscope when resonance takes place and the Voltage value shown
rms
by the voltmeter, V1Ω , corresponds to the intensity of the resonant magnetic field B. Black is the Voltage signal from the
oscillating curcuit, VRF ; red is the alternating magnetic field (the Voltage across the measuring resistor, V1Ω ).
signal is observed on the oscilloscope. going to the highest available frequency in steps
of 5 MHz.
16. Tune finely the amplitude of the DC Voltage, so
that the minimum of the oscillating-circuit signal 22. Insert the large coil with the DPPH sample in, and
is centered at the central vertical line of the oscil- repeat the measurements as for the medium coil
loscope’s screen. When this is done, the Voltage starting from the lowest available frequency and
shown by the DVM is directly proportional to the going to the highest available frequency in steps
resonant magnetic field intensity B which satisfies of 5 MHz.
the resonant formula (9) for the chosen frequency
f.
17. Record in a table the values of f (MHz) and V (V). IV. DATA ANALYSIS
18. Record the instrumental uncertainties σf (MHz) • For each method (AC and DC), for each coil, cal-
and σV (V) in the same table. culate the intensity of the resonant magnetic field
B ± σB from the recorded values of V. The func-
19. Increase the frequency by 5 MHz, and repeat the
tional dependence of B vs. V is linear:
last 3 steps above.
20. Continue increasing the frequency f in steps of B = a + bV, (12)
5 MHz till you reach the maximum possible value
(approximately 80 MHz), and for each f repeat the where
procedure from the previous item.
a ± σa = (−0.2563 ± 0.2088) × 10−4 T, (13)
21. Insert the small coil with the DPPH sample in, and
repeat the measurements as for the medium coil
starting from the lowest available frequency and b ± σb = 0.4109 × 10−2 ± 0.39895 × 10−4 T /V, (14)
8
MHz
Digital Counter
ESR basic
unit
FIG. 8: Schematic diagram of the setup for DC-method measurement of the Electron Spin Resonance. The blue solid lines show
connecting cables. The Voltage-signal across the measuring 1-Ω should be connected to Chennel 1 input of the oscilloscope.
The oscillating-circuit Voltage signal should be connected to Chennel 2. The oscilloscope should be set to operate in XY-mode.
Observe the Voltage signal on the oscilloscope.
cov(a, b) = −7.05628 × 10−10 T 2 /V. (15) • For each calculated g-factor value above gi , calcu-
late its uncertainty σgi .
σB is calculated using error propagation – σa and
σb are propagated, V is treated as a constant. Since
2
a and b are correlated the expression for σB should • Present graphically your results: plot gi ± σgi as a
∂B ∂B
contain a term 2 ∂a ∂b cov(a, b). function i.
• For each method (AC and DC), for each coil, de- • Using weighted mean, calculate a final value g ± σg .
termine the parameters of the linear dependence B
vs. f by fitting lines to the corresponding data [2]. • Compare a canonical (standard) Landé factor g
Why does one fit the dependence B vs. f and not (ge = −2.0023193043622 ± 0.0000000000015) [3]
f vs. B? with your determination.
• From the slope of each linear fit B vs. f calculate
[1] E.N. Hristea et al., Int. J. Mol. Sci. 7, 130 (2006). [3] http://physics.nist.gov/cgibin/cuu/Category?view//
[2] Lecture 9, How to use Physica to make Analysis and =html&All+values.x=80&All+values.y=11
present Results.