Time 2190 Product Data Sheet
Time 2190 Product Data Sheet
●Main unit 1
●5MHz double element transducer 1
●Couplant 1
●TIME certificate 1
●Warranty card 1
●Instruction manual 1
Optional Accessory
●Standard block
●Optional transducers (see next page)
●Dataview software
Technical Specification
NEW
®
TIME 2190
Power three AA battery or NiMH batteries
Measuring Range
Transducer Type Frequency Picture Remark
in Steel(mm)
5.0MHz
DK537EE 1-508 Standard delivery
Double element
10MHz
Field application case:
DLK1025 0.5- 12.7 Single element with
Measure 0.8mm with coating
delay block
1MHz
DCC110 7.62-508 Measure FRP material
Single element
12MHz
DLK1225 0.5-5.08 Measure very small workpiece
Single element
5MHz
DHT-537 1-508 High temperature up to 500℃
Double element
15MHz
DEFM2-SE 0.15-20 Measure thin workpiece
Single element
2.5MHz
SZ2.5P 3-300
Double element
2MHz
TSTU32 5-40(cast iron) Cast iron
Single element
5MHz
ZW5P 4-80 High temperature up to 300℃
Double element
Detecting Modes
●The standard echo detection mode measures the thickness based on the time interval between the excitation pulse and the first back
wall echo. User can measure uncoated materials in this mode.
●Automatic echo-echo detection mode allows thickness measurement of materials with paint or coating because the time interval
between two successive back-wall echoes eliminate paint or coating thickness.
●Paint thickness measurement can simultaneously display layer thickness and substrate thickness.
●The instrument includes three detection modes (Mode 1, Mode 2, and Mode 3)
Mode 1: Measures the time interval between the main pulse signal and the first back-wall echo with direct contact transducer.
Mode 2: Measure the time interval between the interface echo (or delay line echo) and the first back-wall echo with a delay line or
immersion transducer.
Mode 3: Measure the time interval between two successive back-wall echoes with a delay line or a immersion transducer.
When measuring materials with highimpedance The back-wall echo is typicallythe negative
Mode 2 uses a delay line such as metals andceramics, the interface echo electrode unless it isfrom an interface between
transducer or a immersion is usuallypositive, while when measuring low- a lowacoustic impedance material anda high
transducer impedance materials like most plastics,the echo acoustic impedancematerial that are bonded
is negative. together.
Applications
Pipe φ3 Thin Tube Grey Cast Iron Thinning of Tetrafluoroethylene Casting Valve
Material Stamping Parts
Casting LLDPE664 LPG tank Oil Pipeline FRP Valve FRP Oil Tank