NGA Analysis With ISO6974-4
NGA Analysis With ISO6974-4
This method is for determining the chemical composition of natural gases and similar gaseous
mixtures within the composition range shown in the specification sheet. This test method provides
data for calculating physical properties of the sample, such as heating value and relative density,
or for monitoring the concentrations of one or more of the components in a mixture. A total of 1
valves and 2 columns are applied in this GC system. Sample is introduced into the sample loop
for determination. Using a pre-column, C6+ components are back-flushed as a single peak. The
valve timing then allows C1-C5, N2, CO2 to be eluted to TCD through DC-200 in that order. The
final analysis time is approximately 22 minutes. The system includes LabSolution workstation
software and BTU and Specific Gravity calculation software.
Analyzer Information
Concentration Range:
Concentration Range
No. Name of Compound Detector
Low Conc. High Conc.
1 N2 0.001% 15.0% TCD-1
2 CH4 75.0% 100.0% TCD-1
3 CO2 0.001% 10.0% TCD-1
4 C2H6 0.001% 10.0% TCD-1
5 C3H8 0.001% 3.0% TCD-1
6 i-C4H10 0.001% 1.0% TCD-1
7 n-C4H10 0.001% 1.0% TCD-1
8 Neo-C5H12 0.001% 0.5% TCD-1
9 i-C5H12 0.001% 0.5% TCD-1
10 n-C5H12 0.001% 0.5% TCD-1
11 C6+ 0.001% 0.2% TCD-1
Detection limits may vary depending on the sample. Please contact us for more consultation.
System Features
Typical Chromatograms
uV(x100,000)
CH4
N2
1.7
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
C2H6
n-C4H10
n-C5H12
i-C5H12
CO2
C6+
0.2 i-C410
C3H8
0.1
0.0
2.5 5.0 7.5 10.0 12.5 15.0 17.5 20.0 min
Shimadzu Corporation
www.shimadzu.com/an/