Proline XC + Ceph Film - User Manual
Proline XC + Ceph Film - User Manual
cephalostat
user's manual
10010940_3
TABLE OF CONTENTS
1 INTRODUCTION ..............................................................................................1
2 WARNINGS AND PRECAUTIONS ..................................................................2
3 CHECKLIST - BEFORE USING THE CEPHALOSTAT ...................................3
4 PLANMECA PROLINE XC PAN/CEPH X-RAY UNIT .....................................4
4.1 The XC Cephalostat CM ................................................................................................ 4
4.2 Main parts of the Proline XC Cephalostat ...................................................................... 4
4.3 Removing and replacing the ear posts and the nasial positioner .................................. 5
4.4 Emergency stop button .................................................................................................. 5
The manufacturer, assembler, and importer are responsible for the safety, re-
liability and performance of the unit only if:
1 INTRODUCTION
The Planmeca Proline XC Cephalostat CM uses
cephalometric techniques to produce X-ray images for the
diagnosis of dentomaxillofacial anatomy. The unit is
allowed to be used only under supervision of a dental/
health care professional.
This guide describes how to use the Planmeca Proline XC
cephalostat CM. Version of the CM CEPHALOSTAT is
available for both 18cm x 24cm and 8” X 10” film sizes.
Please read these instructions carefully before using the
cephalostat. This manual should be used in conjunction
with the Planmeca Proline Panoramic X-ray unit’s user’s
manual.
The Proline X-ray unit fulfills the requirements of Directive
93/42/EEC. The Proline XC X-ray unit fulfills the
0537
requirements of standard EN 55011, class A.
NOTE Federal law restricts this device to sale by or on the
order of a dentist.
CAUTION This X-ray unit may be dangerous to both patient
and operator unless safe exposure values are
used and correct operating procedures are
observed.
NOTE The X-ray unit’s software revisions are shown on the
main display when the X-ray unit is switched on. This
manual is valid for software revisions: PG 6.54 or later
and PK 6.52 or later.
The display values shown in this guide are only examples
and should not be interpreted as recommended values
unless otherwise stated.
The exposure values given in these instructions are
based on Kodak Ektavision green sensitive film and
Kodak Ektavision screens. If you are using a different film
and screen combination you may have to use different
exposure values.
The exposure values required to produce good X-ray
images will vary considerably according to the build and
age of the patient, the film processor used and processing
procedures being followed. Therefore, the exposure
values given in this guide are average values and are only
meant to guide the user. Users are encouraged to
develop their own radiographic techniques based on
these values.
Make sure that you are fully acquainted with the
appropriate radiation protection measures and these
instructions before using the unit.
LASER RADIATION -
DO NOT STARE (21 CFR § 1040.10).
INTO BEAM
1mW CLASS II
635nm LASER
PRODUCT
AVOID
EXPOSURE-
Laser radiation
is emitted from
this aperture.
LBL-X-100
• Make sure that the film you are going to use is fresh.
Do not use old film.
• Make sure that the film and intensifying screen you are
using are compatible. Do not mix films and screens of
different color sensitivity.
• Make sure that the film and screen are the right type
for the technique you plan to use.
Cephalostat
Rotation scale
Magnification scale
Cassette holder
Release lever
Cassette
Nasial positioner
Ear posts
4.3 Removing and replacing the ear posts and the nasial positioner
Holder
Lift to release
Note
When replacing the ear posts push them firmly
to their position.
seconds.
XC_on_of
When you have selected the program type Ceph the main
display will be shown again.
Chin rest
Cassette Cassette
carriage
Cassette
Rotate the head support to the 90° position, the angle can
be seen on the rotation scale. Press the release lever on
the side of the left-hand ear post holder and slide the ear
posts out as far as they will go. Also slide the nasial
positioner out as far as it will go.
Head support
Rotation scale
Release lever
4
5
6
Nasial positioner
Ear posts
Nasial positioner
Release lever
Positioning cones
Nasial positioner
Frankfort plane
0 10 20 30 40 50 60
Holder
Nasial positioner
Line
Touch either of the soft tissue filter arrows to drive the soft
tissue filter (it is attached to the cassette carriage) to the
same position that appears on the soft tissue filter scale.
The position will appear on the main display.
The left arrow will decrease the area filtered (the soft
tissue filter will move away from the patient) and the right
arrow will increase the area filtered (the soft tissue filter
will move towards the patient).
Magnification scale
Cephscale1298.eps
Cassette holder
Cassette holder Orientate the cassette so that it is vertical and then slide
the cassette into the cassette holder, either from the left or
Cassette clip right, and position it so that the right-hand side of the
cassette is flush with the right-hand side of the cassette
holder.
Cassette
Head support
Nasial positioner
Rotation scale
4
5
6
Ear posts
Position the patient between the two ear posts. If you are
taking an anterior-posterior exposure position the patient
so that they are looking away from the cassette and if you
are taking a posterior-anterior exposure position the
patient so that they are facing the cassette.
Frankfort plane
Anterior-posterior position
Magnification scale
Cephscale1298.eps
Cassette holder
Cassette holder Orientate the cassette so that it is vertical and then slide
the cassette into the cassette holder, either from the left or
Cassette clip right, and position it so that the right-hand side of the
cassette is flush with the right-hand side of the cassette
holder.
Cassette
Head support
Rotation scale
4
Nasial positioner
5
6
Ear posts
Position the patient between the two ear posts and then
adjust the tilt of the patient’s head until the alar-tragal line
Magnification scale
Alar-tragal line
Child 76 12 1.2
Adult 78 12 1.6
Cassette holder Orientate the cassette so that it is vertical and then slide
the cassette into the cassette holder, either from the left or
Cassette clip right, and position it so that the right-hand side of the
cassette is flush with the right-hand side of the cassette
holder.
Cassette
4
5
6
Nasial positioner
Ear posts
Adjust the height of the cephalostat until the positioning
cones at the ends of the ear posts are level with the
patient’s ears.
Adjust the tilt of the patient’s head until the alar-tragal line
is 35° to 40° from the horizontal.
Press the release lever on the side of the left-hand ear
post holder and VERY carefully slide the positioning
cones into the patient’s ears.
Magnification scale
200 190 180 170 160 150 140 130
35-40°
Alar-tragal line
Child 74 12 1.0
Adult 78 12 1.2
Cassette holder Orientate the cassette so that it is vertical and then slide
the cassette into the cassette holder, either from the left or
Cassette clip right, and position it so that the right-hand side of the
cassette is flush with the right-hand side of the cassette
holder.
Cassette
4
5
6
Nasial positioner
Ear posts
Adjust the height of the cephalostat until the positioning
cones at the ends of the ear posts are level with the
patient’s ears.
Adjust the tilt of the patient’s head until the Frankfort plane
is tilted down 30°. Ask the patient to open his or her mouth
as wide as possible. Press the release lever on the side of
the left-hand ear post holder and VERY carefully slide the
positioning cones into the patient’s ears.
Magnification scale
30°
Frankfort plane
Child 74 12 1.0
Adult 78 12 1.2
Hazardous
Waste
Main materials Recyclable waste
Part disposal
for disposal material (separate
site
collection)
Frame and covers
- metal
Aluminium, X
galvanized steel, X
lead X
- plastic
PUR, X
other plastics X
Motors (X)
Component (X)
boards
Cables, Copper, X
transformers steel, X
transformer oil X
X-ray tube X
Packing Wood, X
cardboard, X
paper X
Other parts X