OSL Dosimetry
OSL Dosimetry
R = 0.9888
The KL function was determined from multiple
4. It was observed that the characteristics of individual dosimeters
were independent of the number of irradiation/anneal cycles as
The following properties weredetermined as 1
long as the cumulative dose was less than 1000 cGy.
irradiations of a single dosimeter and was reproducible
part of the characterization of the system:
as long as the cumulative dose to the dosimeters did not 5. The need to apply ECF corrections and the limit on cumulative
0.8 exceed 1000 cGy. Consequently, 1000 cGy was identified
Normalized reading
READER dose for each dosimeter required tracking the dose received by
as the limit of cumulative dose. each dosimeter.
• Stability
• Reading cycle 0.6
FADING 6. Uniformity of the dosimeter response and depletion rate of the
DOSIMETER 3
Polynomial
2
y = -4E-08x + 2E-05x - 0.0044x + 1.0017
Exponential
y = 0.9828e
-0.0035x
signal can be affected by the characteristics of each reader. The
• Depletion rate 2
R = 0.9839 2
R = 0.9816 Dosimeters were irradiated at different dates to a dose of
0.4
methodology used here achieved a distribution in the
• Dependence of depletion rate with 100 cGy and read in one session. Up to 4% of the signal
measurements of less than 1% (SD) after the necessary
reader was was lost over a period of 120 days with the decrease
0.2 corrections. This may be a criterion for acceptance of a reader.
• Cumulative dose limit being more pronounced in the first few days. In practice,
Reader 1 Reader 2
• Number of readings per dosimeter dosimeters will be read approximately 7 days after
• Relative dose response or element 0
0 20 40 60 80 100 120 140
irradiation. The graph below shows KF values normalized References
correction factor (ECF) Number of readings
at two days. The tests reported here used a delay of 2 or [1] Aguirre JF, Tailor R, Ibbott G, Stovall M and Hanson, W. Thermoluminescence
• Variability of ECF with reader more days after irradiation. dosimetry as a tool for the remote verification of output for radiotherapy beams: 25 years
• Variability of ECF with dose of experience. In: Proceedings of the International Symposium on Standards and Codes
ELEMENT CORRECTION FACTORS (ECF) FADING CORRECTION Normalized 2 days of Practice in Medical Radiation Dosimetry IAEA-CN-96/82, Vienna: IAEA. 2002; 191–
• Dose linearity correction
A set of dosimeters went through a series of 100 cGy 99.
• Signal fading correction 1.015
[2] Kirby TH, Hanson WF and Johnston DA. Uncertainty analysis of absorbed dose
• Energy/block correction irradiations, reading and annealing cycles and the relative 1.010
1.005 -0.0072 calculations from thermoluminescence dosimeters. Med Phys 1992; 19:1427–1433.
ANNEALING signal factors were calculated and compared with the y = 1.005x
Fa ding C orr e c tion
1.000
2 [3] R. C. Yoder and M. Salasky. Optically stimulated luminescense dosimeters – an
• Optimal annealing time and average signal. The factor for each dosimeter was constant 0.995 R =1
alternative to radiological monitoring films. Proceedings of the Optical Engineering
0.990
recommended instrumentation. within a standard deviation of 0.3 to 1.3%. See Table 1. Midwest 95, Illinois Institute of Technology, 18 May 1995.
0.985
• Variability of ECF with annealing The exercise was also performed with doses of 25 and 300 0.980 [4] P.A. Jursinic. “Characterization of optically stimulated luminescent dosimeters,
0.975 OSLDs, for clinical dosimetry measurements”. Med Phys 2007; 34:4594–4604.
cGy. In addition, readings were performed on the second 0.970
[5] J. Homnick. G. Ibbott, A. Springer, J. Aguirre. “TH-D-352-05:Optically Stimulated
reader without any significant change. 0.965
Luminescence (OSL) Dosimeters Can Be Used for Remote Dosimetry Services”. Med
0.0 20.0 40.0 60.0 80.0 100.0 120.0 140.0
Days Elapsed Phys 2008; 35:2994–2995.
The investigation was supported by PHS grants CA10953 and CA81647
awarded by the NCI, DHHS.