SurveyorPlus User LCDevices
SurveyorPlus User LCDevices
Surveyor and Xcalibur are registered trademarks of Thermo Fisher Scientific Inc. in the United States.
The following are registered trademarks in the United States and other countries:
Teflon is a registered trademark of E. I. du Pont de Nemours and Company. Windows, Excel, and XP are
registered trademarks of Microsoft Corporation.
All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
Software version: Thermo Fisher Scientific provides the Surveyor device drivers on the LC Devices.
software CD. This manual describes Surveyor devices available on the LC Devices 2.2.0 software CD or later.
These drivers are compatible with Xcalibur 2.0 or later and Thermo Foundation 1.0.x or later.
Firmware versions: Installing the Surveyor device drivers also installs the Surveyor Firmware Upgrade Utility
and the firmware files for the Surveyor devices. For the Surveyor PDA Plus Detector, ensure that the firmware
version is 2.02 or later.
Revision history: Revision A, January 2009; Revision B, April 2010; Revision C, March 2011
Changes that you make to your system may void compliance with one or more of these EMC and safety standards.
Changes to your system include replacing a part or adding components, options, or peripherals not specifically
authorized and qualified by Thermo Fisher Scientific. To ensure continued compliance with EMC and safety standards,
replacement parts and additional components, options, and peripherals must be ordered from Thermo Fisher Scientific
or one of its authorized representatives.
This section contains regulatory compliance information for the following devices of the Surveyor Plus family of LC
instruments:
THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES. OPERATION IS SUBJECT TO
THE FOLLOWING TWO CONDITIONS: (1) THIS DEVICE MAY NOT CAUSE HARMFUL
INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED,
INCLUDING INTERFERENCE THAT MAY CAUSE UNDESIRED OPERATION.
CAUTION Read and understand the various precautionary notes, signs, and symbols contained inside
this manual pertaining to the safe use and operation of this product before using the device.
Notice on Lifting and Handling of
Thermo Scientific Instruments
For your safety, and in compliance with international regulations, the physical handling of this Thermo Fisher Scientific
instrument requires a team effort to lift and/or move the instrument. This instrument is too heavy and/or bulky for one
person alone to handle safely.
Thermo Fisher Scientific has contracted with one or more recycling or disposal companies in each European Union
(EU) Member State, and these companies should dispose of or recycle this product. See www.thermo.com/
WEEERoHS for further information on Thermo Fisher Scientific’s compliance with these Directives and the
recyclers in your country.
WEEE Konformität
Dieses Produkt muss die EU Waste Electrical & Electronic Equipment (WEEE) Richtlinie 2002/96/EC erfüllen.
Das Produkt ist durch folgendes Symbol gekennzeichnet:
Thermo Fisher Scientific s'est associé avec une ou plusieurs compagnies de recyclage dans chaque état membre de
l’union européenne et ce produit devrait être collecté ou recyclé par celles-ci. Davantage d'informations sur la
conformité de Thermo Fisher Scientific à ces directives, les recycleurs dans votre pays et les informations sur les
produits Thermo Fisher Scientific qui peuvent aider la détection des substances sujettes à la directive RoHS sont
disponibles sur www.thermo.com/WEEERoHS.
C
Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xv
Related Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xvi
Safety and Special Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xvii
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .xviii
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .345
Preface
This guide describes how to use your Thermo Scientific mass spectrometry application to
control the Surveyor™ Plus high-performance liquid chromatography (HPLC) system.
Contents
• Related Documentation
• Safety and Special Notices
• Contacting Us
Tip Use the Lamp Startup Time feature to preserve the useful lifetime of the deuterium
lamp. The default lamp startup time for the PDA detector is 4:55 PM. For optimal system
performance, reset the lamp startup time to approximately two hours before you plan to
start data acquisition. For information about changing the lamp startup time, see “Setting
the Lamp Startup Time” on page 166.
Tip The default setting for the autosampler maintenance counters is 0. If you turn on the
autosampler maintenance log (see “Communication Page” on page 52) when you specify
the configuration settings for the autosampler and leave the counters at the default setting,
the autosampler status displays Maintenance Due and the data system prevents you from
starting a run. For information about changing the counter settings on the Maintenance
Information page, see “Autosampler Maintenance Information” on page 309.
Related Documentation
In addition to Help that you can access from the data system, these manuals are provided with
LC Devices as PDF files for the Surveyor Plus family of high-performance LC instruments:
• Surveyor Plus User Guide for LC Devices (formerly entitled Surveyor Plus User Guide for the
Xcalibur Data System)
• Surveyor Plus Preinstallation Requirements Guide
• Surveyor Plus Getting Connected Guide
• Surveyor Autosampler Plus Hardware Manual
• Surveyor LC Pump Plus Hardware Manual
• Surveyor MS Pump Plus Hardware Manual
• Surveyor PDA Plus Detector Hardware Manual
• Surveyor UV/Vis Plus Detector Hardware Manual
Go to Start > Programs > Thermo Instruments > Manuals > LC Devices > Surveyor >.
To open Help
• From the Instrument Setup window, choose Help >product name Help.
• If available for a specific window or dialog box, click Help or press F1 for
information about setting parameters.
Contacting Us
There are several ways to contact Thermo Fisher Scientific for the information you need.
Phone 800-532-4752
Fax 561-688-8736
E-mail [email protected]
Knowledge base www.thermokb.com
Phone 800-532-4752
Fax 561-688-8731
E-mail [email protected]
Web site www.thermo.com/ms
Go to www.thermoscientific.com/wps/portal/ts/contactus.
Detector
Autosampler
Analytical
pump
Contents
• Analytical Pump
• Autosampler
• Detector
• Solvent Platform
• Communication with the Data System Computer
• Turning Off the Computer’s Energy Saving Features
• Synchronization of the LC Devices
• Status LEDs
Analytical Pump
Thermo Scientific, part of Thermo Fisher Scientific Inc., offers two analytical pumps in the
Surveyor Plus product line: the Surveyor MS Pump Plus, which provides optimum
performance in the lower flow rate ranges needed for mass spectrometry; and the Surveyor LC
Pump Plus, which provides optimum performance in the flow rate ranges needed for
chromatography.
Both analytical pumps contain a built-in solvent degassing system that consists of four
independent chambers maintained at a constant vacuum of approximately 27 in. Hg. Each
chamber contains an 18 in. length of 0.045 in. ID Teflon™ AF tubing. This translates to a
volume of less than 500 μL per channel, adding very little to the quantity of solvent required
to purge the lines when you replace an eluent.
The pulse dampening assembly consists of a low volume T-connector that the mobile phase
passes through. Attached to the side leg of the T-connector is a 2 mL loop of stainless steel
tubing. The loop terminates with a priming valve (pulse dampener flush valve). When the
priming valve is open, you can flush or fill the dampening loop with an appropriate solvent
such as methanol or isopropanol. When the priming valve is closed, the loop is shut off from
the flow path and absorbs pump pulsations. Because the 2 mL loop is shut off from the flow
path, it adds no gradient delay volume to the LC system.
The priming solvent in the loop does not interfere with the purity of the mobile phase, and its
composition does not need to match the mobile phase.
The data system remotely controls the pump with an RS-232 communication link from the
data system computer. The only manual control is the power switch located on the front of
the pump in the lower-left corner below the door.
The Surveyor LC Pump Plus is a bench-top unit for inclusion in the Surveyor Plus high
performance liquid chromatograph (HPLC). The Xcalibur™ data system controls the pump
through an Ethernet communication link with the data system computer. The only manual
controls are the power switch located on the front of the pump in the lower-left corner below
the door and the drain valve.
The Surveyor LC Pump Plus has five major components: the vacuum degassing assembly, the
solvent proportioning assembly, the pump head assemblies, the purge manifold assembly, and
the pulse dampening assembly (see Figure 2). The pulse dampening assembly adds 400 μL of
gradient delay volume to the system. The Surveyor LC Pump Plus has status LEDs on the
front door and a low-voltage power supply.
Figure 2. Front view of the Surveyor LC Pump Plus
Solvent lines entering Pump head Line body containing
and exiting the assemblies filter frit
degassing assembly
IN OUT
A
B Left Right
C
N
LO
PE
SE
DRAIN
O
You can use the Purge command from the data system to draw solvent at a rate of 10 mL/min
through the solvent lines connected to the pump after you open the purge valve. To open the
purge valve, turn the knob on the front of the purge manifold assembly counterclockwise
(see Figure 3). When the valve is in the open position, look for the word DRAIN to appear
upside down on the knob.
Figure 3. Drain valve positions
Closed position Open position
DRAIN
SE
O
PE
LO
N
C
C
N
LO
PE
SE
DRAIN
O
Figure 4 shows the line body fitting removed from the top of the purge manifold assembly.
The line body fitting contains an inline filter frit that captures particulate matter shed by the
pump seals. This prevents particulate matter from contaminating the tubing that connects the
pump to the autosampler. As this filter frit becomes clogged, the system backpressure rises.
Replace this filter frit when the system backpressure rises above its typical operating limits.
Figure 4. Line body and filter frit
Autosampler
Thermo Fisher Scientific offers two autosampler models in the Surveyor Plus product line:
the full-featured Surveyor Autosampler Plus and the Surveyor Autosampler Plus Lite. The
full-featured autosampler includes a built-in column oven (5 to 95 °C) and a temperature
controlled tray/sample compartment (0 to 60 °C). Both models let you inject samples and
perform sample preparation tasks automatically.
Tray Compartment
The tray compartment (see Figure 5) can hold up to five conventional sample trays or one
carrier that holds three microwell plates. The tray compartment also holds up to four 16 mL
capacity reservoir vials that you can use to hold solvent, reagent, or diluent. The reservoir vials
are located behind the wash station and are designated RV1, RV2, RV3, and RV4.
Figure 5. Tray compartment
XYZ arm
mechanism RV4
RV3
RV2
RV1
Wash station
Injection port
The five conventional sample trays, from the left side to the right side of the tray
compartment, are designated A, B, C, D, and E. Each sample tray holds up to 40 standard
1.8 mL vials, for a total capacity of 200 samples. Overlays allow the sample trays to
accommodate different vial sizes. The microwell carrier can hold up to three low-density
96-well microplates or up to three high-density 384-well microplates. The microplates are
designated A, B, and C.
The tray compartment door contains a magnetic switch. The magnet is located in the door
and the switch is attached to the chassis. When you open the door, the switch signals the
autosampler that the door is open.
When you configure the autosampler (from the data system) to verify whether the tray
compartment door is open or closed, the XYZ arm automatically moves to the back of the
tray compartment when you open the door, allowing you to remove trays or replace vials.
Opening the tray compartment door while the autosampler is making an injection does not
interrupt the current run. The XYZ arm moves to the back of the compartment after the
current injection is complete. The programmed sequence of injections then halts. When you
close the tray compartment door, the sequence resumes.
These topics describe the notation for specifying vial and well locations:
• Vial Notation
• 96-Well Microplate Notation
• 384-Well Microplate Notation
Vial Notation
The notation for the location of vials is as follows:
Tray location in tray compartment:Vial location in tray
Where:
Tray location in tray compartment = A, B, C, D, or E
Vial location in tray = 1 through 40
A colon separates the tray location in the tray compartment from the vial location in the tray.
20 40 E:40
19 39
18 38
17 37
16 36
15 35
14 34
13 33
12 32
11 31
10 30
9 29
8 28
7 27
6 26
5 25
4 24
3 23
2 22
A:1 1 21
Where:
Plate = A, B, or C
Row = A through H
Column = 1 through 12
Figure 7 shows the notation for 96-well microplates when you select the top left orientation.
Figure 7. Notation for 96-well plates—Top left orientation
Plate C
A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12
C:A1
B1 B2 B3 B4 B5 B6 B7 B8 B9 B10 B11 B12
Plate A
A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12
A:A1
B1 B2 B3 B4 B5 B6 B7 B8 B9 B10 B11 B12
Figure 8 shows the notation for 96-well microplates when you select the bottom right
orientation.
Figure 8. Notation for 96-well plates—Bottom right orientation
Where:
Plate = A, B, or C
Row = A through P
Column = 1 through 24
P A C:A 1
C:A1
1 24 24 1
B:A1
B:A1 A P
P A B:A1
B:A1
1 24 24 1
A:A 1
A:A1 A P
P A A:A 1
A:A1
1 24 24 1
Injection System
The main components of the injection system include the following:
• Syringe Drive Assembly and Syringe Valve
• Wash Bottle Reservoir and Tubing
• Interchangeable Syringe
• XYZ Arm Mechanism
• Needle Assembly and Needle Tubing Assembly
• Injection Port and Transfer Tubing Assembly
• Injection Valve and Sample Loop
Figure 10. Syringe drive assembly for the Surveyor Autosampler Plus, showing the front and the back of assembly
Two-way
syringe valve
10-pin
connector
Wash bottle
tubing
Needle
tubing
Syringe
fitting
Mounting hole (with grommet)
to vertical mounting stud
The syringe valve is a two-position rotary valve. In the wash bottle position, the syringe draws
wash solvent into its syringe barrel as its plunger descends. In the needle position, the syringe
draws sample solution into the needle tubing as its plunger descends and pushes sample
solution out of the needle tubing as its plunger ascends. The syringe never draws sample
solution into its barrel.
Note The needle tubing holds approximately 560 μL of solvent. If you installed the
2500 μL standard syringe (which must be special ordered) and you plan to make large
volume injections, install the 1 mL needle tubing extension that comes with this syringe.
Two-way
syringe valve
Needle tubing
assembly
Wash
tubing
Wash
bottle
Wash solvent
Interchangeable Syringe
The standard configuration for the autosampler consists of a 250 μL dual-concentric syringe.
The dual-concentric syringe consists of a small, inner plunger, and a larger, outer plunger
(see Figure 12).
Figure 12. Dual-concentric syringe
Inner plunger
Outer plunger
The syringe uses its inner plunger to draw and deliver sample amounts equal to or less than its
maximum capacity, which is 265 μL for the 250 μL concentric syringe. The syringe uses its
outer plunger to draw and expel large volumes of solvent, for example, during a flush or wash
cycle.
Dual-concentric syringes are available in 100, 250, and 500 μL sizes. The volume of the outer
plunger region, 565 μL, is the same for all three dual-concentric syringes. The size of the inner
plunger and the injection mode determine the available injection volume range. In addition
to offering three sizes of concentric syringes, Thermo Scientific also offers a 2500 μL standard
(single plunger) syringe.
Figure 13 shows the needle descending to a depth of 2 mm from the bottom of a standard
1.8 mL vial.
Needle height
from bottom
Bottom of vial
2 mm
Wash solvent
Sample
After the autosampler withdraws the sample from the sample vial or well into the needle
tubing, the XYZ arm mechanism moves along the x-y plane, back to the home position,
which is above the injection port of autosampler, and then lowers the needle into the injection
port where it expels the sample. The sample travels through transfer tubing and into the
sample loop of the injection valve.
IMPORTANT Because the XYZ arm moves to the sample position to withdraw sample,
do not place objects taller than 1.8 inches into the tray compartment, as they will stall the
XYZ arm.
Do not move the XYZ arm manually. Instead, use the following commands and options
provided by the data system to control the position of the XYZ arm:
• To make the XYZ arm automatically move to the back of the tray compartment when you
open the tray door, select the Verify Door is Closed option, which is available when you
configure the autosampler device driver (see “Communication Page” on page 52).
• To move the XYZ arm to the back of the tray compartment, use the Position Arm to
Access Tray direct command (see “Removing and Installing Sample Trays” on page 199).
• To move the XYZ arm mechanism to its home position above the injection port, use the
Set Arm to Home Position direct command.
• To move the XYZ arm mechanism to the middle front of the tray compartment, allowing
easy access to the needle, use the Remove Needle direct command.
The needle tube assembly connects the solvent path between the needle and the syringe valve
and consists of low-pressure tubing, an internally threaded fitting that connects to the needle
assembly fitting, a sleeve, a black or red PVC cap, a needle tube guide that attaches to the
x-axis positioning frame, and an externally threaded fitting with a flangeless ferrule that
connects to the right side of the syringe valve.
IMPORTANT To prevent damage to the needle tubing, take care when you connect the
needle tubing guide to the back of the syringe drive assembly. Pinched tubing causes
performance problems.
Figure 14 shows the needle assembly and the needle tubing assembly.
Figure 14. Needle assembly and needle tubing assembly
Needle assembly
Latch nut
Compression spring
Flag Needle tubing assembly
10-32
fitting Externally threaded nut and
flangeless ferrule that connect to
the right side of the syringe valve
Blunt-tip
needle
Injection port
Figure 16 shows the 0.012 in. ID transfer tube that connects the autosampler injection port
to port 2 of the injection valve.
IMPORTANT The label attached to the transfer tube assembly specifies its internal
volume. You must enter this value when you specify the configuration settings for the
autosampler.
Transfer
tube
The sample loop is a section of stainless steel tubing with end fittings. It is an interchangeable
part that is attached to ports 1 and 4 of the injection valve. The autosampler comes with a
25 μL sample loop.
There are two positions for the six-port injection valve: fill and inject.
In the fill position (see Figure 17), the sample loop is isolated from the mobile phase stream.
As the mobile phase bypasses the sample loop, the upward movement of the autosampler
syringe plunger pushes sample into the front of the sample loop, connected to port 1 of the
injection valve. Once the sample loop is filled, excess solution exits the injection valve through
port 3 to waste.
Figure 17. Injection valve in the fill position
Transfer tubing
2 1 To
column
6
3
To
4 5
waste
Mobile phase From
Wash solvent pump
In the inject position (see Figure 18), mobile phase enters the sample loop from the back,
backflushing the contents of the sample loop onto the column. Excess sample left in the
transfer tube is expelled directly to waste. To allow ample rinsing of the sample loop with
mobile phase, the injection valve remains in the inject position during the entire run.
Figure 18. Injection valve in the inject position
Transfer tubing
2 1 To
column
6
3
To
4 5
waste
Mobile phase From
Wash solvent pump
Injection Modes
The autosampler can operate in any of the following three modes:
• No Waste Injection Mode
• Partial Loop Injection Mode
• Full Loop Injection Mode
The optimum injection mode depends on the amount of sample that you have and the degree
of precision that your application requires.
Approximately 0.25 μL of the sample is lost as it travels from the injection port, through the
transfer tubing, and into the injection valve. Because of this loss, the minimum recommended
injection volume is 1.0 μL.
The quantity of lost sample depends on the syringe rate. Decreasing the syringe rate decreases
the sample loss. For best results, use a syringe rate no greater than 4 μL/s for the no waste
injection mode.
Partial loop injections are useful when you have a limited volume of sample. Using the partial
loop injection mode, you can inject variable amounts of sample, ranging from a minimum of
0.1 μL to a working maximum of one-half the volume of your sample loop. This maximum
volume limitation is caused by the laminar flow of fluid within the stainless steel sample loop.
IMPORTANT To make precise partial loop injections, use a sample loop that is at least
twice the size of the injection volume. The accuracy of the nominal sample loop volume
is ±20%. Because the actual volume of the 25 μL sample loop (provided with the
autosampler) is anywhere from 20 to 30 μL, limit the maximum injection volume with
this loop to 10 μL. To inject more sample, use a larger sample loop.
Full loop injection is useful when you want maximum precision and have unlimited sample.
To change the injection volume, you must change the sample loop size.
Note Full loop injections are limited to the size of the configured sample loop.
In the full loop injection mode, the autosampler withdraws a large excess of solution from the
sample vial according to the following equation:
Amtw = 3 × Vinj + Vdead + 7.5 μL
Where:
Amtw = sample volume withdrawn by the autosampler
Vinj = user-specified injection volume
Vdead = dead volume held by the transfer tubing, injection port, and rotor slot
This equation is valid until the syringe reaches its maximum capacity, when only the
maximum capacity of the syringe is withdrawn. The maximum capacity of the 250 μL
concentric syringe is 265 μL.
Temperature Control
The full-featured Surveyor Autosampler Plus has two built-in temperature control features:
• Tray Temperature Control
• Column Oven Control
CAUTION The column oven can reach temperatures as high as 95 °C (203 °F).
To prevent personal injury, before installing the LC column or connecting the pump to
the autosampler, make sure that the column oven compartment is at room temperature.
Between the analytical pump and the autosampler injection valve, the mobile phase is diverted
through a heat exchanger located behind the column oven. As it passes through the heat
exchanger, the mobile phase equilibrates to the temperature of the column oven before it
reaches the injection valve. The heat exchanger adds only 3 μL of gradient delay volume to the
LC system.
Detector
The Surveyor Plus product line includes two detectors that you can control from your
Thermo Scientific data system: a PDA detector and a UV/Vis detector. The PDA detector, in
combination with the 5 or 1 cm LightPipe flowcell, provides the highest level of sensitivity
available in photodiode array detection for HPLC. The UV/Vis detector can monitor two
wavelength channels. Both detectors have a dual-lamp optical bench that covers the
UV-visible spectrum from 190 to 800 nm.
For more information about the Surveyor Plus detectors, see these topics:
• PDA Detector
• UV/Vis Detector
PDA Detector
The PDA detector is a full-featured, time-programmable, photodiode array detector that can
scan the full ultraviolet-visible range from 190 to 800 nm. The detector can acquire data at a
rate of up to 20 Hz with a 20 bit digital conversion.
The optimal location for the PDA detector is above the autosampler and below the solvent
platform. The data system computer controls the PDA detector through an Ethernet
communication link. The PDA detector consists of a dual-light source, an optical bench, a
photodiode array, a low voltage power supply, several printed circuit boards (PCBs), and
four status light-emitting diodes (LEDs).
The dual-light source includes a deuterium lamp for detection in the ultraviolet
wavelength range (190 to 360 nm) and a tungsten-halogen lamp for detection in the
visible wavelength range (360 to 800 nm). The light emitted by the two lamps overlaps in
the 300 to 500 nm range. A pair of attenuators that you can manually adjust controls
the intensity of light reaching the photodiode array.
The optical bench contains a beam combiner, focusing lens, filter wheel, flow cell, beam
shaper, folding mirror, and grating (see Figure 19). The beam combiner reflects the light
coming from the tungsten-halogen lamp so that it is parallel to and coincident with the light
from the deuterium lamp. The combined beam is then focused on the inlet window of the
flow cell through the filter wheel. The standard filter wheel has two positions. Use
position 1 (Open), which places a clear, quartz window in the optical path, for normal
operation. Use position 2 (Holmium Oxide), which places a sealed, quartz cuvette filled with
a holmium oxide/perchloric acid solution (NIST™ traceable) in the optical path, for
wavelength accuracy verification and calibration.
Figure 19. Optical bench of the PDA detector
Deuterium Beam combiner Flow cell
lamp Focusing lens Beam shaper
Filter wheel Folding mirror
1 × 512
photo diode
array
Fixed grating
Attenuators
Tungsten-halogen lamp
The PDA detector does not have independent controls, such as a keypad, to create data
acquisition methods. Instead, you create instrument methods for data acquisition with the
data system. The only manual controls for the PDA detector are the On/Off switch that
controls line power, the attenuators that control the light throughput to the diode array, and
the holmium oxide wheel that controls the position of the wavelength calibration solution.
UV/Vis Detector
The UV/Vis detector is a full-featured, time-programmable, variable-wavelength UV/Vis
(ultraviolet/visible) absorbance detector. It operates in either the single wavelength mode, the
dual wavelength UV wavelength mode, or the dual wavelength Visible mode. The wavelength
range in the single wavelength mode is 190 to 800 nm. In the dual wavelength UV mode, the
range is 190 to 450 nm and in the dual wavelength Visible mode, the range is 366 to 700 nm.
The wavelength time table is available in all three modes. The time table can contain up to
10 lines. If you enable the Zero On Wavelength Change feature, the baseline absorbance
returns to zero between each line in the time table, even if the wavelengths remain the same.
The baseline absorbance does not return to zero between the last two lines in the table.
To provide a complete spectrum of ultraviolet and visible light, the detector uses a
deuterium lamp for the UV range (190 to 365 nm) and a tungsten lamp for the visible range
(366 to 800 nm). The lamps are protected by a cover with a special safety interlock to reduce
the possibility of human exposure to harmful UV light.
To control the UV/Vis Detector, you must first add the detector to the software instrument
configuration (see “Adding Devices to the Instrument Configuration” on page 44).
After you add the UV/Vis detector to the instrument configuration, you can set up the data
acquisition parameters from the UV/Vis detector view of the Thermo Xcalibur Instrument
Setup window (see “UV-Vis Detector Instrument Method Parameters” on page 106).
You can turn the detector lamps on or off and zero the detector output from the Direct
Control dialog box (see “UV-Vis Detector Direct Controls” on page 183).
Solvent Platform
The solvent platform, located on the top of the LC stack, holds four 1 liter solvent reservoir
bottles and one 1 liter wash bottle. Four 1/8 in. OD × 1/16 in. ID, FEP solvent lines carry
solvent from the reservoir bottles down to the vacuum membrane degasser, which is built into
the analytical pump. One 1/8 in. OD × 1/16 in. ID, FEP solvent line carries solvent from the
wash bottle to the two-position syringe valve of the autosampler.
The Surveyor MS Pump Plus uses an RS-232 serial link to communicate with the data system
computer. The accessory kit for the Surveyor MS Pump Plus includes a USB cable.
In addition to connecting the communication cables, you must enter the appropriate stack
addresses for the Surveyor Autosampler Plus, Surveyor PDA Plus Detector, Surveyor UV/Vis
Plus Detector, and Surveyor LC Pump Plus when you configure the device drivers. The stack
address or stack number that you enter when you add one of these devices to the
configuration for your instrument must match the unit ID setting on the back panel of the
device. For more information about configuring the Surveyor Plus device drivers, see
Chapter 2, “Thermo Foundation Instrument Configuration.”
To turn off the screen saver and energy saving features for the Windows XP operating
system
1. From the Windows™ XP™ desktop, choose Start > Control Panel.
Tip If you selected the Classic Start Menu option for the Start menu properties,
choose Start > Settings > Control Panel.
2. Double-click Display.
The Display Properties dialog box appears.
To turn off the screen saver an energy saving features for the Windows 7 operating
system
1. From the Windows 7 desktop, choose Start > Control Panel.
2. Choose Large Icons from the View By menu.
4. Click the Screen Saver icon on the bottom-right side of the window.
The Screen Saver Settings dialog box appears (see Figure 24).
Figure 25. Select a Power Plan page of the Power Options dialog box
e. Set the Turn Off Hard Disk After option to Never as follows:
i. Click the icon to the left of Hard Disk.
ii. Click the icon to the left of Turn Off Hard Disk After.
iii. If the setting is not set to Never, click the setting and select Never.
f. Set the Hibernate After option to Never as follows:
i. Click the icon to the left of Sleep.
ii. Click the icon to the left of Hibernate After.
iii. If the setting is not set to Never, click the setting and select Never.
g. Click Apply to accept these settings.
h. Click OK to close the Advanced Settings page.
7. Close the Control Panel window.
Figure 28 shows the cable assembly with seven green combicon connectors: two are labeled
PUMP; three are labeled DETECTOR; and two are unlabeled. You can identify the
connector for the autosampler by the small, A/S tag on its adjacent cable. You can identify the
M/S connector by its shape and by the small, M/S tag on its adjacent cable.
Figure 28. 7-connector system interconnect cable
DETECTOR
M/S To
detector
DETECTOR
To To
detector
DETECTOR
MS detector To
detector
To
PUMP
A/S
autosampler
To
To
PUMP
pump
pump
Using the 7-connector cable with the Surveyor MS Pump Plus requires an additional adapter
cable. Figure 29 shows the interconnect cable connections for a Surveyor LC system.
Figure 29. Back panels of Surveyor devices, showing the 7-connector system interconnect cable
MS Detector
DETECTOR
M/S Detector
DETECTOR
Detector
DETECTOR
Detector
Autosampler Pump
PUMP
A/S
Pump
PUMP
Interconnect cable
Detector Detector
(back panel) (back panel)
To To
MS detector Autosampler MS detector Autosampler
(back panel) (back panel)
Pump
connector
MS pump connection
MS PUMP
PUMP
Pump
connector of
MS pump Surveyor MS adapter cable LC pump
adapter cable Pump (back panel)
(back panel)
Surveyor MS Pump
connector
Figure 30 shows the cable assembly with five green combicon connectors. The assembly has
two pump connectors: one is labeled MS PUMP and one is labeled LC Pump. You can
identify the connectors for the autosampler, detector, and MS detector by the small tags on
their adjacent cables. To connect the MSQ Plus Mass Detector, you must modify the wiring
connections inside the MS detector connector.
To To A/S
MS Detector Autosampler
LC PUMP
To
MS PUMP
LC Pump
To
MS Pump
LC Pump
MS PUMP
MS Pump
Detector
Interconnect cable (back panel)
To
MS detector
Autosampler
(back panel)
Surveyor MS
Pump Plus
(back panel)
For more information about connecting the interconnect cable to the devices of your LC
system, refer to the Surveyor Plus LC System Getting Connected Guide.
During a run, the system issues the following sequence of run control signals:
1. The data system computer issues a request to perform an injection.
2. When the autosampler becomes ready, it issues the A/S Ready signal.
The autosampler goes into the Ready state when all the configuration and instrument
method conditions are met, including the closing of the tray compartment door, and the
sample tray and column oven temperature zone readings showing they are within
tolerance of their setpoints.
3. When the pump pressure stabilizes, the pump issues the Pump Ready signal to the
autosampler.
The autosampler injection valve switches to the fill position, and then the autosampler
pushes sample into the sample loop.
4. The autosampler issues the Gradient Start signal to the pump. This signal commands the
pump to start its gradient program.
5. When its piston cam reaches the home position, the pump issues the Inject Hold release
signal to the autosampler.
6. The autosampler injection valve switches to the inject position, allowing the mobile phase
to backflush the contents of the sample loop onto the column. The autosampler then
issues a momentary Inject Out signal to the detector.
7. The detector starts acquiring data.
Status LEDs
Each of the Surveyor Plus LC devices has a panel of four status LEDs located on the front of
the left door. All of the devices have these three LEDs: Power, Comm, and Run. In addition,
the detector has a Lamps LED; the autosampler has a Temp LED; and the analytical pump
has a Degas LED.
Tip To ensure that the controlled temperature zones are in equilibrium at the set
temperature before the autosampler makes an injection, select the Wait for Temperature
Ready check box when you add the autosampler to the data system instrument
configuration.
When you select this check box, the autosampler waits until the controlled temperature
zones are in equilibrium at the set temperature before it makes an injection. While the
temperature zones are equilibrating to the set temperature, the Temp LED remains amber.
When you do not select this check box, the autosampler does not wait for the temperature
zones to equilibrate to the set temperature before making an injection and the Temp LED
remains green.
–or–
Contents
• Checking the Communication Hardware
• Opening the Thermo Foundation Instrument Configuration Application
• Adding Devices to the Instrument Configuration
• Specifying the Configuration Settings
• Closing the Foundation Application
Before you specify the configuration settings, check the unit ID setting on the back panels of
the autosampler, LC pump, UV/Vis detector, and PDA detector. The unit ID consists of two
rotary switches that are factory set to 01 (see Figure 32).
Figure 32. Unit ID with a stack address of 01
2 3 2 3
0 1
0 1
4
4
5 6
5 6
9
9
7 8 7 8
Set to 0 Set to 1
Unlike the other Surveyor Plus devices, the Surveyor MS Pump Plus communicates with the
data system through an RS-232 serial connection.
1. Check the settings of the rotary switches on the back panels of the Surveyor Autosampler
Plus, the Surveyor UV/Vis Plus Detector or the Surveyor PDA Plus Detector, and the
Surveyor LC Pump Plus.
2. For the Surveyor MS Pump Plus, verify that the serial communication cable connecting
the MS pump to the data system computer is attached to COM port 1.
From the Windows taskbar, choose Start > Programs > Thermo Foundation 1.0 >
Instrument Configuration.
The Thermo Foundation Instrument Configuration window appears (see Figure 33).
This window lists the installed device drivers.
Figure 33. Thermo Foundation Instrument Configuration window
In the Available Devices list, double-click the icons for your LC devices.
A copy of the icon appears in the Configured Devices list (see Figure 34).
Figure 34. Thermo Foundation Instrument Configuration window with devices added to the
Configured Devices list
After specifying the configuration options for the instrument devices, you must close the
Instrument Configuration application before you can open the Thermo Scientific data
system.
Note The Accela UV/Vis Detector driver is provided with LC Devices 2.5.0 or later.
For information about specifying the configuration settings for the autosampler, see these
topics:
• Tray Page
• Communication Page
• Signal Polarity Page
• Firmware Page
The Surveyor Autosampler Configuration dialog box appears with the Tray page
displayed (see Figure 35).
Figure 35. Tray page of the Surveyor Autosampler Configuration dialog box
Tray Page
The autosampler accessory kit contains several tray types, including trays that hold 1.8 mL
vials, trays that hold 96-well microplates, and trays that hold 384-well microplates.
Use the Tray page (see Figure 36) to specify the tray type that you are using.
Note The Autosampler Accessory Kit contains the vial trays and microplate carriers.
The current version of the accessory kit contains a standard tray, a short microwell
carrier, and a tall microwell carrier with a solid metal bottom.
Previous versions of the accessory kit contained a riser plate that you could use in
combination with the short microwell carrier to load standard depth 96-well or
384-well microplates into the autosampler tray compartment. Thermo Fisher
Scientific has replaced the short microwell carrier and riser plate option with a tall
microplate carrier that has a solid metal bottom for thermal conduction.
b. In the Well Ordering area, click the button that represents your preference for the
order of sequence injections.
The conventional autosampler tray holds 40 standard 1.8 mL vials. For other types of vials,
select the custom vial setting and the appropriate tray adapter.
The autosampler accommodates a variety of microwell plates. You can use the tall or short
microwell carriers to accurately position the various microwell plates at an appropriate
location in the autosampler tray compartment. The tall microwell carrier with a solid metal
bottom provides the same height as the short microwell carrier and riser plate combination.
The distance that the needle travels to reach the bottom of the vial is pre-calibrated for the
eight standard configurations. You must perform a Well Bottom Distance calibration
(see “Well Bottom Distance Calibration” on page 305) to use one of the three custom
configurations.
For information about the height limitations for the various tray selections, see Table 8.
Note Do not use the short microwell carrier to hold the 384-well, high density microtitre
plates. Do not use the tall microwell carrier to hold PCR plates.
1 2 3 4 5 6 7 8 9 10 11 12 1 24
Bottom Specifies that the A1 well is located at the bottom right of the microplate.
Right
H P
G
F
E
D
C
B
A A
12 11 10 9 8 7 6 5 4 3 2 1 24 1
Well Ordering Use the buttons in the Well Ordering area to select the sampling path that the autosampler
(for the microplate follows during a sequence run. The sampling path you select is independent of the A1 well
trays) position.
The path options are shown below:
Graphic Description
The autosampler samples wells from bottom to top and then from left to
right, as you view the plate from the top.
The autosampler samples wells from left to right and then from bottom to
top, as you view the plate from the top.
The autosampler samples wells from top to bottom and then from left to
right, as you view the plate from the top.
The autosampler samples wells from left to right and then from top to
bottom, as you view the plate from the top.
Placing objects taller than 1.87 in. into the autosampler tray compartment will stall the
autosampler arm. For custom vials, use the appropriate tray insert to ensure correct
positioning. In addition, to trigger the vial sensor, position the vials in the tray so that the top
of each vial reaches the minimum height of 1.55 in.
Do not use microtitre plates that exceed the heights listed in Table 8.
Table 8. Autosampler tray compartment height limitations
Tray type Maximum height
Conventional (standard 1.8 mL vials) N/A
Custom Vial Setting N/A
96 Well Microplate + Tall Microwell Carrier 0.77 in.
96 Well Microplate + Tall, Solid Microwell Carrier or 0.77 in.
96 Well Microplate + Short Microwell Carrier + Riser Plate
1 mL or 2 mL Deep Well Plate + Short Microwell Carrier 1.80 in.
96 Well Microplate + Short Microwell Carrier 1.80 in.
96 Well PCR Plate + Short Microwell Carrier 1.80 in.
Custom 96 Well Setting 1.80 in.
384 Well Microplate + Tall Microwell Carrier 0.77 in.
384 Well Microplate + Tall, Solid Microwell Carrier or 0.77 in.
384 Well Microplate + Short Microwell Carrier + Riser Plate
Custom 384 Well Setting 0.77 in.
Communication Page
Use the Communication page to specify these options:
• The autosampler’s stack address
• The syringe size and the sample loop size
• The dead volume of the transfer tubing that connects the autosampler injection port to
the injection valve
• Whether the controlled temperature zones must be at their set temperature, the tray
compartment door must be closed, or both before the autosampler makes an injection
• Whether the autosampler uses the bottom distance for the configured tray type or the
stored custom value, or determines the bottom distance of sample vials or wells for each
injection or at the start of a sequence
• Whether the maintenance log is enabled
1. Open the Surveyor Autosampler Configuration dialog box (see “Surveyor Autosampler
Plus Configuration Settings” on page 45).
2. Click the Communication tab.
The Communication page appears (see Figure 37).
Figure 37. Communication page
3. In the Stack Address box, type the appropriate stack address or use the up and down
arrows to select the appropriate stack address.
The stack address must match the unit ID setting located on the back panel of the
autosampler. The rotary switches of the unit ID are factory set to 01 and the default Stack
Address is 1. The value of 00 is reserved for service functions.
4. Under Syringe, in the Type list, select the size of the syringe that is attached to the
autosampler.
The default is Concentric 250 μL, the syringe that ships with the autosampler.
5. (Optional) To prevent the autosampler from making an injection when the temperature
zones are not equilibrated at the set temperature, select the Wait for Temperature Ready
check box.
When the Wait for Temperature Ready check box is selected, the autosampler does not
trigger a run until the column oven temperature, the sample tray temperature, or both
have reached their setpoint values.
6. (Optional) To prevent the autosampler from making an injection when the tray
compartment door is open, select the Verify Door Is Closed check box.
When the Verify Door Is Closed check box is selected, the autosampler cannot trigger a
run when the tray compartment door is open. If the tray door is opened during a
sequence run, the XYZ arm moves to the back of the tray compartment at the end of the
current run, and the Xcalibur sequence halts.
7. (Optional) To activate the maintenance log, select the Enable Maintenance Log check
box.
The maintenance log keeps an internal count of the total injections, total valve cycles,
total needle usage, and total syringe cycles. When any of the counters exceed the user set
scheduled maintenance time (SMT), the autosampler cannot trigger a run until you
perform the scheduled maintenance or you clear the check box.
8. In the Vial Bottom Sensing area, select the type of vial bottom sensing that is appropriate
for your application:
• To activate vial bottom sensing for every injection in a sequence, select the On
option.
• To activate vial bottom sensing for only the first injection in a sequence, select the
Auto option.
• To deactivate vial bottom sensing, select the Off option.
Each tray type has a stored value for the distance that the needle must travel to reach the
bottom of the vial or well. When you activate vial bottom sensing, the autosampler
performs a search routine to determine the actual location of the vial or well bottom. If
the search routine determines a new value for the bottom distance, it is stored until you
modify the tray type. If you do not want the needle to touch the bottom of a vial or well,
deactivate vial bottom sensing.
9. In the Dead Volume (μL) box, type the value specified on the label attached to the
transfer tube assembly.
Tip When you use the No Waste injection mode, for best results, calibrate the transfer
tube volume, and then type the empirically determined value in the Dead Volume
box. To calibrate the transfer tube volume, refer to the hardware manual for the
autosampler.
10. In the Sample Loop Volume (μL) box, type the nominal size of the sample loop attached
to the injection valve.
The autosampler ships with a 25 μL sample loop.
Select this check box if you want the heated zones to reach their setpoint before the
autosampler makes an injection.
Verify Door Is Closed Selecting this check box specifies that the autosampler waits for the door to be closed before
starting a run and that the XYZ arm moves to the back of the tray compartment when you
open the tray compartment door.
Select this check box if you want the autosampler to pause a sequence and send the XYZ
arm to the back of the tray compartment when you open the tray compartment door.
Enable Maintenance Selecting this check box activates the maintenance log.
Log
The maintenance log keeps an internal count of the total injections, total valve cycles, total
needle usage, and total syringe cycles. When any of the counters exceed the user-set
scheduled maintenance time (SMT), the autosampler cannot trigger a run until you
perform the scheduled maintenance or you clear the check box.
For information about setting up the scheduled maintenance time for the autosampler’s
hardware components, see “Autosampler Maintenance Information” on page 309.
Tip When you select the Enable maintenance log check box, check the maintenance
schedule and make sure that the settings are appropriate for your autosampler.
Vial Bottom Sensing Specifies whether the vial bottom sensing feature is turned on or off or is set to determine
the distance to the bottom of the first vial or well in a sequence of injections.
The transfer tubing has a label that specifies its dead volume.
Sample Loop Volume Specifies the nominal size of the sample loop that is connected to the injection valve. The
autosampler cannot detect the sample loop size; it uses the sample loop size that you type in
this box.
2 3 2 3
0 1
0 1
4
4
5 6
5 6
9
7 8 7 8
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 UNIT ID ENET
GND
+ 5V @ 150mA
PUMP STOP
GRAD START
PUMP READY
INJECT HOLD
INJECT OUT
A/S READY
GND
+ 5V @ 150mA
TF 4
TF 3
TF 2
TF 1
GND
GND
Use the Signal Polarity page of the Surveyor Autosampler Configuration dialog box to specify
the signal polarities for the input and output run signals. For the Surveyor LC Pump Plus and
the Surveyor MS Pump Plus, ensure that the Pump Ready Active High and the Injection
Hold Release Active High check boxes in the Input area are selected.
1. Open the Surveyor Autosampler Configuration dialog box (see “Surveyor Autosampler
Plus Configuration Settings” on page 45).
2. Click the Signal Polarity tab.
The Signal Polarity page appears (see Figure 39).
3. In the Input area, ensure that the Pump Ready Active High and Injection Hold Release
Active High check boxes are selected (see Figure 39). These check boxes are selected in
the default instrument method.
4. In the Output area, if your instrument consists entirely of Surveyor Plus devices, do not
select the check boxes.
The Pump Ready Active High and the Injection Hold Release check boxes are selected in the default instrument
method.
These check boxes are clear in the default instrument method. For a Thermo Scientific LC/MS system, leave these
check boxes clear.
Autosampler Ready Use this check box to set the polarity of the A/S Ready output signal:
Active High
• When this check box is selected, the signal from the autosampler is LO (Closed) and
goes HI (Open) when the autosampler is ready.
• When this check box is clear, the signal from the autosampler is HI (Open) and goes
LO (Closed) when the autosampler is ready.
Note The A/S Ready terminal sends a signal that indicates that the autosampler is in the
Ready state. The autosampler is ready when it meets the selected ready conditions. The
ready conditions you select include sample tray temperature, column oven temperature,
door is closed, maintenance log, and vial bottom sensing (see “Communication Page” on
page 52).
Injection Out Active Use this check box to set the polarity of the Inject Out output signal:
High
• When this check box is selected, the signal from the autosampler is LO (Closed) and
goes HI (Open) when the autosampler injects a sample.
• When this check box is clear, the signal from the autosampler is HI (Open) and goes
LO (Closed) when the autosampler injects a sample.
Note The Injection Out terminal sends a signal to the other LC modules or peripheral
devices when the autosampler injects a sample.
Gradient Start Active Use this check box to set the polarity of the Gradient Start output signal:
High
• When the check box is selected, the signal from the autosampler is LO (Closed) and
goes HI (Open) to trigger the start of the pump’s gradient program.
• When the check box is clear, the signal from the autosampler is HI (Open) and goes LO
(Closed) to trigger the start of the pump’s gradient program.
Note The Grad Start terminal sends a signal to the pump to start the gradient program.
Firmware Page
Use the Firmware page to check the firmware version of the autosampler.
1. Open the Surveyor Autosampler Configuration dialog box (see “Surveyor Autosampler
Plus Configuration Settings” on page 45).
2. Click the Firmware tab.
The Firmware page appears (see Figure 40).
Figure 40. Firmware page
3. If you have upgraded your autosampler, upload the firmware version as follows:
a. Click Upload Firmware Version.
The firmware version of the autosampler appears next to Current Version.
b. Click OK to save the settings and close the dialog box.
If you have finished configuring all of your LC devices, go to “Closing the Foundation
Application” on page 66.
1. Ensure that the pump is connected to the data system computer and powered on.
2. For a dual-pump system, ensure that pump 2 is powered off.
3. In the Configured Devices list, double-click the Surveyor MS Pump Plus icon.
The Surveyor MS Pump Plus dialog box appears (see Figure 41).
Figure 41. Surveyor MS Pump Plus dialog box
4. If two pumps are connected to the data system computer, make sure that the appropriate
pumps are selected in the Pump 1 and Pump 2 area.
You can use two pumps of the same pump type to perform two-dimensional
chromatography or high-pressure gradient mixing.
5. Click OK to close the dialog box.
1. In the Configured Devices list, double-click the Surveyor PDA Plus icon.
The Surveyor PDA Plus Configuration dialog box appears (see Figure 43).
Figure 43. Surveyor PDA Plus Configuration dialog box
2. In the Stack Number box, type the appropriate number (unit ID).
3. Click OK to close the Surveyor PDA Plus Configuration dialog box.
When you have finished configuring the instrument devices, go to the next topic: Closing the
Foundation Application.
The Surveyor UV/Vis Configuration dialog box appears (see Figure 44).
Figure 44. Surveyor UV/Vis Configuration dialog box
2. In the Stack Address box, type the appropriate number (unit ID). The Stack Address
setting must match the rotary switches on the back panel of the detector.
3. Click OK to close the Surveyor UV/Vis Configuration dialog box.
When you have finished configuring all of your LC devices, go to the next topic: Closing
the Foundation Application.
To save the instrument configuration and close the Instrument Configuration window
Tip You must close the Instrument Configuration window before you open the
Xcalibur data system. The two applications cannot be open at the same time.
To automate control of the liquid chromatography devices, you must create an instrument
method, and then specify the instrument method to be used for each run in an acquisition
sequence from the Sequence Setup window.
Instrument methods contain the analysis wavelengths, the chromatographic conditions, and
the autosampler injection parameters, such as the injection mode, required for data
acquisition.
1. Open the Instrument Setup window (Xcalibur data system) or equivalent window for
your Thermo Scientific mass spectrometry application.
2. Specify the instrument method settings for each device.
3. Save the instrument method.
Contents
• Opening the Instrument Setup Window
• Surveyor LC Pump Plus Instrument Method Parameters
• Surveyor MS Pump Plus Instrument Method Settings
• Autosampler Instrument Method Parameters
• PDA Detector Instrument Method Parameters
• Triggering an External Device with the PDA Detector
• UV-Vis Detector Instrument Method Parameters
• Saving the Instrument Method
To set up the instrument method parameters for the Surveyor LC Pump Plus, see these topics:
• Surveyor LC Pump Plus General Page
• Surveyor LC Pump Plus Gradient Program Page
1. In the Solvent Name boxes in the Solvents area, type the solvent names.
2. In the Description box in the Column area, type a description of the LC column.
3. In the Pressure Limits area, do the following:
a. In the Min Pressure box, type the minimum operating pressure.
Note The maximum pressure setting is 43.10 MPa, 431.00 bar, or 6258.00 psi.
Reduce the maximum pressure setting for PEEK fittings to 28 MPa, 276 bar, or
4000 psi.
Placeholder line
Note If you set the limit for the maximum pressure to less than 21 MPa, the
allowable flow rate range is 0.000 to 9.999 mL/min. If you set the maximum pressure
limit to 21 MPa (215 bar, 3129 psi) or higher, the maximum allowable flow rate is
5.000 mL/min.
The first time line in the time program table is set to 0.00 minutes and is uneditable. The
second time line in the program is a placeholder line and has no effect on the pumping
conditions.
For example, the pump program shown in Figure 47 produces a proportioned mobile
phase consisting of 80% methanol and 20% water (v/v). The proportions of the solvents
remain constant throughout the run. The mobile phase flow rate remains constant
at 1 mL/min throughout the run.
3. To create a gradient pump method, enter at least two steps in the Gradient Program table.
For each additional line in the program, type these values:
• A time value in the Time column
• A flow rate in the Flow column
• The solvent percentages for the mobile phase in the solvent columns
4. Press ENTER.
The last line in the gradient table, which has an * in place of a line number, is a
placeholder and has no effect on the pump conditions.
Table 14 lists the gradient conditions used to create the gradient profile shown in
Figure 48.
Table 14. Gradient program
Time (min) Solvent composition
0.00 to 5.00 Held constant at 90:10 water/methanol
5.00 to 25.00 Linear ramp from 10% to 90% methanol
25.00 to 25.01 Stepped down to initial solvent composition of
90:10 water/methanol
25.01 to 40.00 Column is equilibrated at the initial solvent composition of
90:10 water/methanol
At 25.01 minutes, the program returns to its initial solvent composition. To equilibrate
the column prior to the next run, the solvent composition is held at the initial solvent
composition for a period of 15 minutes.
Note When you are performing a sequence of gradient runs, add an equilibration
step at the end of the Gradient Program.
Figure 48. Example of a gradient program for the Surveyor LC Pump Plus
Placeholder line
Specifies the gradient steps. The table can contain up to 80 rows. The solvent composition boxes are interactive. After
you type a value in column A and press ENTER or click another box, the value in column B changes to maintain a
total of 100%, and so on. The solvent composition changes linearly between time points.
Time (min) Specifies the time duration for each gradient step. The initial time is always 0 minutes.
Table 15. Gradient Program page parameters for the Surveyor LC Pump Plus (Sheet 2 of 2)
Parameter Description
Flow (mL/min) Specifies the flow rate.
To open the Surveyor MS Pump Plus view, click the Surveyor MS Pump Plus icon in the
view bar of the Instrument Setup window.
Tip After you specify the solvents that make up the mobile phase, you can switch back
and forth between the Pump General and Gradient Program pages.
When you connect two pumps to the data system computer, you must specify the general
parameters for both pumps. The autosampler triggers the gradient program for Pump 1 at the
start of a run. Pump 1 triggers the gradient program for Pump 2 at a user-specified time
during the run (see “Programming the Pumps in a Dual-Pump System” on page 84).
The Instrument Method report attached to the raw file contains the values that you select or
enter on this page. You can access this report in the Qual Browser view by choosing
View > Reports > Instrument Method.
Note The Idle Settings list is available for a dual-pump system in the Pump 2
area of the Pump General page.
f. In the Min Pressure (bar) box, type or select the minimum operating pressure for the
pump.
An appropriate minimum pressure setting prevents the pump from operating when
the solvent reservoirs have run dry or the system plumbing has developed a leak.
Running the pump without solvent quickly ruins the piston seals. Select a value that
is well below the typical operating pressure for your application. In the event that the
pressure falls below this limit for more than one minute, the pump automatically
stops and sends an error message to the computer.
The range is 0 to 400 bar (5800 psi) with a default value of 0 bar.
g. In the Max Pressure (bar) box, type or select the maximum operating pressure for the
pump.
An appropriate maximum pressure setting prevents the pump from operating with a
restriction on the outlet side of the pump. Excess pressure can damage the HPLC
column and any other component between the restriction and the pump. Select a
value that is well above the typical operating pressure for your application, but also
below the pressure that can damage your system. If the pressure rises above this limit,
the pump automatically stops and sends an error message to the computer.
The range is 0 to 400 bar (5800 psi) with a default value of 400 bar.
h. In the Pressure Stability (bar) box, type or select an appropriate pressure stability
value from 1 to 100.
The MS pump sends a Ready signal when the system pressure reaches this limit. A
lower value requires greater pressure stability before the pump becomes ready. A
higher value is more forgiving of pressure pulsations.
i. In the Pressure Units list, select the pressure units for the pressure readback.
Note 1 bar = 14.5 psi
Clearing the check box associated with a solvent (A, B, C, or D) makes the solvent column
unavailable on the Gradient Program page.
When you use this injection logic, set up the pump as the start
instrument.
Min Pressure Specifies the minimum operating pressure of the pump. If the system pressure falls below
this level for more than one minute, the data system triggers turns off the mobile phase flow.
Default: 0 bar
Max pressure Specifies the maximum operating pressure of the pump. If the system backpressure rises
above this level for more than one minute, the data system triggers the pump to turn off the
mobile phase flow.
Default: 10
Use the Gradient Program page to specify the solvent composition and flow rate for your
chromatographic method.
The Gradient Program page contains a gradient table with time lines for the solvent
composition and flow rate, and an interactive view that displays a graphical profile of the
solvent gradient or flow rate gradient.
Each row in the gradient table defines the solvent composition and flow rate for a specific
time point. The solvent composition columns (A%, B%, C%, and D%) work interactively
with each other, maintaining a total solvent composition of 100%.
Between time points, the solvent composition changes linearly, whereas the flow rate changes
as a discontinuous step-function.
IMPORTANT The gradient table must contain at least two time lines. For an isocratic
gradient program, the two time lines specify the same solvent composition and flow rate.
5. For an isocratic separation method, do the following on the Gradient Program page for
the Surveyor MS Pump Plus:
a. Double-click the Time column in the second row of the gradient program table, and
then type a time value in this cell.
The time range is 0.01 to 655.00 minutes.
If the time value for the second time point is longer than the data acquisition time for
the PDA detector or the MS detector, the data system waits for the gradient program
to finish before proceeding to the next row in the acquisition sequence.
b. Type the same values for the solvent composition and the flow rate that you specified
for the initial time point. Press ENTER to accept the values.
IMPORTANT You must specify the flow rate and solvent composition for at
least two time lines.
c. Verify the gradient program in the Solvent Gradient view (see Figure 51).
Figure 51. Isocratic gradient program
Tip If you do not want the detector to acquire data during the column
equilibration period, you can add an equilibration period to the end of the
gradient program and make the gradient program time for the pump longer than
the data acquisition time for the detector.
Default: 100% A
μL/min The μL/min box specifies the flow rate of the mobile phase at the time specified in the
associated Time box.
Default: 10 μL/min
P2 (Dual-pump For a dual-pump system, the time line position of the ball in the P2 column specifies the
system) time when Pump 1 triggers Pump 2.
Graph view
Type of View Select Solvent Gradient to view the solvent gradient profile created by the pump program.
Select Flow Gradient to view the flow rate changes created by the pump program.
The Gradient Profile area graphically displays the values entered in the pump program. The
y axis represents percent composition or flow rate and the x axis represents time in minutes.
Each solvent is color-coded for better visualization of the programmed gradient.
For information about the Sample Preparation page, see Chapter 4, “Sample Preparation
Routines.”
To specify the injection and temperature control settings for your instrument method
Note For more information about the injection modes, see “Injection Modes” on
page 19.
2. For the No Waste injection mode, type or select a loop loading speed in the Loop Loading
Speed box.
For most applications, keep the loop loading speed at its default value.
3. For the Partial Loop and No Waste injection modes, type or select an injection volume in
the Injection Volume box.
The injection volume range depends on the syringe size (see “Injection Volume Range”
on page 91).
4. In the Needle Height from Bottom (mm) box, type or select the height from the bottom
of the vial or well that the needle of the autosampler descends to before withdrawing
sample from the vial/well location.
The default value is 2 mm and the range is 0.1 to 18 mm. Entering a lower value causes
the needle to descend closer to the bottom of the vial or well. The standard 1.8 mL vials
supplied with the autosampler have a depth of approximately 20 mm.
Figure 13 on page 15 shows the needle descending to a depth of 2 mm from the bottom
of a standard 1.8 mL vial.
5. In the Syringe Speed (μL/s) box, keep the syringe speed at its default value (8 μL/s for the
250 μL concentric syringe) for most applications.
If you see the sample in the needle tubing break up as the autosampler withdraws sample
from a vial or well location into the needle tubing, adjust the syringe speed.
6. In the Flush Volume (μL) box, type or select a volume from 0 to 6000 μL or keep the
value at its default of 400 μL.
7. In the Flush/Wash Source list, select a flush/wash solution.
For most applications, select Bottle (wash bottle). If you select one of the reservoir vial
locations as your flush/wash source, make sure that you load the autosampler with a
reservoir vial.
For information about the location of the 16 mL reservoir vials, see “Tray Compartment”
on page 5. For information about specifying the contents of the reservoir vials and the
wash bottle, see “Reservoir Content Page” on page 92.
8. In the Wash Volume (μL) box, keep the wash volume at its default value of 0 μL.
To wash the exterior of the needle after each injection, type a value from 1 to 6000 μL in
the Wash Volume box.
9. In the Flush Speed (μL/s) box, set a flush speed appropriate for the flush solvent:
• If you are using a flush solvent of low viscosity such as methanol, keep the flush speed
at its default (100 μL/s for the 250 μL concentric syringe)
• If you are using water or a methanol/water mixture as the flush solvent, reduce the
flush speed.
Tip If you hear a grinding noise from the autosampler as it performs a flush, lower the
flush speed.
10. In the Post-Injection Valve Switch Time (min) box, keep the time value at its default of
0 minutes for most applications. This leaves the sample loop in the inject position during
the entire run, allowing ample rinsing of the sample loop between injections.
11. In the Tray Temperature Control area, to control the temperature of the tray
compartment, select the Enable Tray Temperature Control check box. Then type an
appropriate temperature from 0 to 60 °C in the Temperature box.
12. In the Column Oven Control area, to control the temperature of the LC column, select
the Enable Column Oven Control check box. Then type an appropriate temperature
from 5 to 95 °C in the Temperature box. Controlling the temperature of the LC column
increases the reproducibility of the chromatographic retention times.
The autosampler ships with a 250 μL concentric plunger syringe, but you can replace this
syringe with a 100 or 500 μL concentric plunger syringe or a 2500 μL single plunger syringe.
The injection volume range and the syringe speeds depend on the syringe size as described in
these topics:
• Injection Volume Range
• Syringe and Loop Loading Speeds for Injections
The default value and the range depend on the syringe size (see “Injection Volume Range”
on page 91).
Needle Height from Specifies the height from the bottom of the vial or well where the needle withdraws liquid
Bottom (mm) (see Figure 13 on page 15).
The default value and the range depend on the syringe size (see “Syringe and Loop Loading
Speeds for Injections” on page 91).
Flush Volume (μL) Specifies the volume of liquid that the autosampler flushes through the inside of the needle,
the injection port, and the transfer tube. During a flush cycle, the injection valve is in the
inject position to prevent the flush solvent from entering the sample loop.
To reduce carryover, increase the flush volume. When the specified flush volume exceeds the
syringe capacity, the autosampler takes multiple draws from the reservoir or bottle.
During a wash cycle, the autosampler lowers the needle into the wash port. When the
specified wash volume exceeds the syringe capacity, the autosampler takes multiple draws
from the reservoir vial or wash bottle.
A time value greater than 0.1 minutes specifies the time when the injection valve switches
from the inject position to the load position at the end of an injection cycle.
To reduce the gradient delay volume for low-flow gradient applications, consider switching
the injection valve from the inject position to the load position during the run by typing a
nonzero value in this box. Allow enough time for the mobile phase to backflush the sample
out of the sample loop.
Default: 0 minutes
Range: 0 to 999.0 minutes
Loop Loading Speed Specifies the rate at which the autosampler meters the sample into the sample loop for no
(μL/s) waste injections. During a no waste injection, the autosampler meters the sample through
the transfer tubing to the injection valve at the specified syringe speed. The injection valve
then switches to the load position, and the autosampler meters the sample into the sample
loop at the specified loop loading speed.
The default value and range depend on the syringe size (see “Syringe and Loop Loading
Speeds for Injections” on page 91).
Injection Mode
Partial Loop Select this option if you want to inject variable volumes.
IMPORTANT To make precise partial loop injections, limit the maximum injection
volume to less than half the nominal sample loop size. Because the accuracy of the
nominal volume of the sample loop is ± 20% (which means that the actual volume of the
standard 25 μL sample loop is between 20 and 30 μL), for best results, limit the maximum
injection volume to 10 μL for the standard 25 μL sample loop.
To maintain a stable temperature, set the temperature to a value that is a least 10 °C above
or below the ambient room temperature. As a safety feature, a thermostat turns off the
power to the sample tray if the temperature reaches 65 °C.
Temperature (° C) Specifies the tray temperature.
Default: 30 °C
Range: 0 to 60 °C
Column Oven Control
Enable Tray Temp Select this check box to enable temperature control for the column oven.
Control
In the default instrument method, this check box is clear; temperature control for the
column compartment is not enabled.
Temperature (° C) Specifies the column oven temperature.
To maintain a stable temperature, set the temperature to a value that is a least 10 °C above
or below the ambient room temperature. As a safety feature, a thermostat turns off the
power to the column oven if the temperature reaches 110 °C.
Default: 30 °C
Range: 5 to 95 °C
For partial loop and no waste injections, the minimum injection volume is 0.1 μL. The
maximum injection volume depends on the syringe size (see Table 20).
You can specify how fast, in microliters per second, the syringe withdraws liquid from a tray
vial or plate well, and then expels liquid into the autosampler injection port. The default
syringe speed and the syringe speed range depend on the syringe size (see Table 21).
The appropriate syringe speed depends on the sample. For viscous samples, use a syringe
speed lower than the default value to avoid stalling the syringe. For samples of low viscosity or
surface tension, use a lower syringe speed to prevent the sample bolus from breaking apart
during the transport process.
Tip Select the syringe size on the Communication page of the Surveyor Autosampler
Configuration dialog box. The autosampler ships with a 250 μL concentric syringe.
For no waste injections, you can use a higher syringe speed to push the sample from the
autosampler injection port to the injection valve and a lower loop loading speed to meter the
sample into the sample loop. The range for the loop loading speed depends on the syringe size
(see Table 22).
Table 22. Loop loading speeds
Syringe size Minimum speed Maximum speed Default speed
(μL) (μL/s) (μL/s) (μL/s)
100 0.01 13.20 3.00
250 0.04 33.00 8.00
500 0.08 66.11 8.00
2500 0.41 330.80 25.00
Figure 55 shows the relative size of the 1 liter wash bottle and the 16 mL reservoir vials.
Figure 55. Reservoir vials and wash solvent bottle
Wash bottle
Reservoir vials
2. Identify the solvents contained in the 16 mL reservoir vials and the 1 L wash bottle by
typing their names in their respective boxes.
The TF terminal output signal is LO (Closed) by default. To change the polarity of the TF
output signal to HI (Open), select the Timed Events Active High check box on the Signal
Polarity page of the Surveyor Autosampler Configuration dialog box (see “Signal Polarity
Page” on page 56).
Figure 57. Time function event terminals on the back panel of the autosampler
TF terminals
Use the Timed Events page (see Figure 58) to specify the timed events for the time function
terminals (TF1 to TF4).
Figure 58. Timed Events page for the autosampler
Default: 0
Range: 0 to 9999.0 minutes
TF1, TF2, TF3, and TF4 Specifies whether the event is on or off at the specified time.
2. In the Run Length (min) box, type a run time from 0 to 600 minutes.
3. In the Filter Rise Time (sec) list, select a rise time from 0 to 10.0 seconds.
Rise time is inversely proportional to the amount of baseline noise. If your chromatogram
contains closely eluting peaks, minimize baseline noise while retaining maximum
resolution by selecting a rise time that is approximately one-tenth of the narrowest peak’s
width at half-height (FWHM). Increasing the rise time above this level increases peak
tailing, which can reduce the resolution of closely eluting peaks.
Figure 60 shows the effect of rise time on peak tailing.
Figure 60. Effect of rise time on peak tailing
Minutes
5. To collect a wavelength scan, select the Collect Spectra Data check box, and then make
the following entries and selections:
a. In the Start Wavelength (nm) box, type a starting wavelength from 190 to 799 nm.
b. In the End Wavelength (nm) box, type an ending wavelength from 191 to 800 nm.
The ending wavelength must be greater than the starting wavelength.
c. In the Wavelength Step (nm) box, type a value for the wavelength interval.
The default value is 1 nm. The maximum step size depends on the wavelength range
defined by the starting and ending wavelengths. The range is 1 to 610 nm for a scan
from 190 to 800 nm. If you are collecting spectral data for a spectral library, use a
wavelength step of 1 nm.
d. In the Sample Rate (Hz) list, select a sample rate.
For optimal integration of the chromatographic peaks, select a sample rate that
acquires a minimum of 20 points across the baseline width of the narrowest peak of
interest. For example, if the baseline width of the narrowest peak is 20 seconds, select
a sample rate of 1 Hz or higher.
Increasing the sampling rate increases the data file size.
e. In the Filter Bandwidth (nm) box, type a bandwidth.
Depending on the wavelength range defined by the starting and ending wavelengths,
the range is the subset of odd integers from 1 to 49 nm. The maximum scan range for
a particular bandwidth is limited as follows:
• Start Wavelength (minimum) = 190 nm + (bandwidth – 1)/ 2
• End Wavelength (maximum) = 800 nm – (bandwidth – 1)/ 2
Increasing the bandwidth decreases the spectral resolution (see Figure 61).
Figure 61. Spectra of toluene, effect of bandwidth on spectral resolution
1 nm bandwidth
11 nm bandwidth
6. To collect discrete channel data, select one of the options in the Channels area, and then
make the following entries:
a. In the Sample Rate list, select a sampling rate (in data points per second). Higher data
rates for discrete channels do not add significantly to the data file size.
b. For each discrete channel (A, B, and C), type a wavelength from 190 to 800 nm in
the Wavelength (nm) box and a filter bandwidth in the Filter Bandwidth (nm) box.
Depending on the discrete wavelength, the acceptable bandwidth range is 1 to 49 nm
in odd-number increments with 1 nm meaning no filtering. Bandwidth values
outside the range of the detector are not allowed. For example, for the discrete
wavelength of 200 nm, the maximum bandwidth setting is 21 nm. At a bandwidth
setting of 21 nm, the reported absorbance value for 200 nm is a weighted average
from 190 to 210 nm. A value greater than 21 nm would be outside the lower range
limit of the detector, which is 190 nm.
For information about the Method page parameters, see these topics:
• PDA Detector Instrument Method Parameters
• Display Method Parameters
• Timed Events Table for the PDA Detector
For information about adding timed events to trigger external devices, see “Triggering an
External Device with the PDA Detector” on page 104.
At the end of the run time, the PDA detector goes to the Ready for Download state.
Filter Rise Time Specifies the detector’s response time to the signal in seconds.
The selections are 0.00, 0.10, 0.20, 0.50, 1.00, 2.00, 5.00, and 10.00 seconds.
Table 24. PDA detector Method page parameters for an instrument method (Sheet 2 of 3)
Parameter Description
Units
Wavelength/ Displays absorbance spectra with the wavelength in nanometers on the x axis and the
Absorbance absorbance in milli-absorbance units (mAU) on the x axis.
Default: Wavelength/Absorbance.
Default: 200 nm
Range: 190 to 799 nm
End Wavelength (nm) Specifies the last wavelength in nanometers of the scan range.
Default: 600 nm
Range: 191 to 800 nm
Default: 1 nm
The highest spectral resolution is 1 nm and the lowest resolution is the full width of the
scan. When you specify a spectral scan from 200 to 300 nm at a step of 10 nm, the PDA
detector sends 11 scan wavelengths to the data system computer.
Sample Rate Specifies the number of data points per second that the PDA detector sends to the data
system for each scan wavelength.
For optimal integration of the chromatographic peaks, acquire 20 points across the baseline
width of the peak.
For the Surveyor PDA Detector Plus, the sample rates are 0.5, 1.0, 2.0, 4.0, 5.0, 10.0, and
20 Hz.
Table 24. PDA detector Method page parameters for an instrument method (Sheet 3 of 3)
Parameter Description
Filter Bandwidth Specifies the bandwidth for the scan wavelengths.
Bandwidth values are limited to the subset of odd integers from 1 to 49. In addition, the
maximum scan range for a particular bandwidth is limited as follows:
Lower range = 190 nm + (bandwidth – 1) / 2
Upper range = 800 nm – (bandwidth – 1) / 2
Default: 9 nm
When you select the Diode/Intensity option, you can view the light intensity detected by
the photo diodes of the diode array. The x axis corresponds to the photo diode number
(2 to 511).
There are 512 diodes in the diode array. Diodes 0 and 1 are not used. Diode 2 corresponds
to wavelength 190 nm and diode 511 corresponds to 800 nm. This gives a spacing of
611 nm/510 diodes = 1.2 nm/diode. The PDA detector interpolates the data to give
integer values.
Spectra (with Diode/Intensity option)
Collect Spectral Data Select this check box to scan a range of diodes.
Start Diode Specifies the first diode in the scan range.
(diode num)
Default: 2
Range: 2 to 510, in increments of 1
Table 25. PDA Method page parameters for a display method (Sheet 2 of 2)
Parameter Description
Channels (with Intensity Units)
No Channels Specifies that no discrete channel measurements are to be acquired. When you select this
option, the parameters in the Channels area are unavailable.
One Channel Specifies the acquisition of one discrete channel. The data for Channel A is displayed in the
bottom view of the Display page in black.
Two Channels Specifies the acquisition of two discrete channels. The data for Channel B is displayed in the
bottom view of the Display page in red.
Three Channels Specifies the acquisition of three discrete channels. The data for Channel C is displayed in
the bottom view of the Display page in green.
Sample Rate Not applicable
[Ch. A, B, C] Specifies the diode that you want the detector to monitor.
Diode Number
Channel A default: 40
Channel B default: 55
Channel C default: 450
Use the Timed Events area to program actuation of a back panel contact closure as a function
of either time or the absorbance level from one of the discrete wavelength channels. You can
use this feature to trigger an external device, such as a fraction collector, and to zero the
absorbance level during an acquisition run.
For information about connecting an external device to the PDA detector, refer to the
Surveyor PDA Plus Detector Hardware Manual.
Table 26. Timed events table on the Surveyor PDA Plus Method page (Sheet 1 of 2)
Parameter Description
Timed Events
Use this area to add, modify, and remove timed events. You can specify the time when the event occurs, the event
type, the channel, the absorbance level (mAU), and the delay settings. Click New to open the Timed Events dialog
box.
Timed Events table This table lists all the events in chronological order. You can add, modify, and remove events
by double-clicking the event setting you want to modify, or by using the New, Delete, and
Delete All buttons.
Table 26. Timed events table on the Surveyor PDA Plus Method page (Sheet 2 of 2)
Parameter Description
Time Specifies the time in minutes when the detector triggers the event. The maximum time is
the data acquisition time specified in the Run Length box (see Figure 59 on page 95).
Use a Level Trigger event to specify the absorbance threshold level for triggering an event.
When you select a Level type event, you must specify the channel of the event, the
absorbance level that is to trigger the event, and the delay time that the event is to occur
after the absorbance level trigger event.
Default: 0 seconds
Range: 0 to 65 535, in increments of 1
Buttons
New Click this button to open the Timed Event dialog box where you can add external events to
the external events table.
Delete Click this button to remove an external event from the external events table. Select the
external event that you want to remove, and then click Delete.
Delete All Click this button to remove all external events from the external events table. To remove all
events, click Delete All.
Note You can also zero the absorbance level for a discrete channel during an acquisition
run by adding a zero event to the Timed Events table.
1. In the Timed Events area of the Surveyor PDA Plus Method page (see Figure 62), click
New.
Figure 62. Timed Events area of the Method page
3. In the Time (min) box, type the length of time after the chromatographic run starts that
the event is to occur.
The range is 0.00 to the run length time for the detector. The Event Off, Event On, and
Zero Data event types occur at this time. Because the Level Trigger event type occurs after
this time is exceeded, do not enter a time value that exceeds the expected time (for
example, the expected retention time of the analyte that you want a fraction collector to
collect) of the triggering event.
The Channel list, Level box, and Delay box are available for the Level Trigger event type.
For more information about the Level trigger type, see the next procedure.
4. To trigger an external device when the absorbance for a discrete channel reaches a specific
level, do the following:
a. In the Type list, select Level Trigger.
b. In the Channel list, select the discrete wavelength channel to be used to trigger the
event.
The available selections are A, B, or C.
c. In the Level (mAU) box, type a value for the absorbance value that triggers the event.
The default value is 0.00 mAU and the range is –2000 to 4000.00 mAU.
d. In the Delay (sec) box, type a time in seconds.
The delay time compensates for the time interval between, for example, when a
compound leaves the detector and when it reaches a fraction collector. The default
value is 0 and the range is the set of integers from 0 to 65 535.
5. To zero the absorbance level for all the wavelength channels that the detector is
monitoring at a particular time in the run, do the following:
a. In the Type list, select Zero Data.
b. In the Time box, type the time that you want the absorbance level to be zeroed.
6. Click OK to accept the new event and close the Timed Event dialog box.
1. In the Rise Time (sec) list, select a rise time from 0 to 10.0 seconds.
Rise time is the response time of the detector, in seconds, to the signal. Increasing the rise
time decreases the baseline noise; however, setting the rise time to a value greater than
one-tenth the width of the chromatographic peak at half-height results in peak
broadening. The one-second default value is appropriate for most LC applications.
Figure 65 shows the effect of rise time on peak tailing.
Minutes
2. In the Data Rate (Hz) list, select a data rate from 4 to 20 points per second.
The optimal data rate depends on the expected baseline width of your chromatographic
peaks. For optimal integration, acquire a minimum of 20 data points across the baseline
width of the narrowest peak. For example, if the narrowest chromatographic peak has a
baseline width of 10 seconds, select a data rate of 2 Hz or higher.
3. To zero the baseline at a specific time point, select the Enable Autozero check box and
type a time in the Autozero Time (min) box.
The detector resets its output voltage to zero at this time point.
4. In the Program Type area, select one of the following options:
• To collect one chromatogram during a run, select the Single Wavelength
190 to 800 nm option. You can program time wavelength changes in the program
table.
• To collect two chromatograms in the UV range during a run, select the Dual
Wavelength UV 190 to 450 nm option. You can program time wavelength changes
in the program table.
• To collect two chromatograms in the visible range during a run, select the Dual
Wavelength Vis 366 to 700 nm option. You can program time wavelength changes
in the program table.
5. To zero the baseline when a programmed wavelength change occurs, select the Zero on
Wavelength Change check box.
The detector does not zero the baseline on the first or last rows of a wavelength program.
If the wavelength program contains three or more rows, the detector zeroes the baseline
on the second row and all successive rows until it reaches the last row. The detector zeroes
its output signal even if the same wavelength is listed in the second through second to last
rows of the table.
Selections: 0.0, 0.01, 0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 seconds
Data Rate Specifies the number of points per second acquired per wavelength.
Default: 10 Hz
Default: 0 minutes
Range: 0 minutes to the run length
Program Type
Single Wavelength Specifies that the detector is to acquire data on one wavelength channel. The acquisition
UV/Vis 190–800 nm wavelength is time programmable.
Dual Wavelength UV Specifies that the detector is to acquire data on two wavelength channels in the UV range.
190–365 nm The acquisition wavelengths are time programmable.
Dual Wavelength Vis Specifies that the detector is to acquire data on two wavelength channels in the visible range.
366–700 nm The acquisition wavelengths are time programmable.
Other options
Zero on Wavelength When this check box is selected, the detector zeros the absorbance data at the times specified
Change in the Time column for rows 2 through the last time line.
Programmable wavelength table
Time Specifies the time when the detector is to start monitoring the wavelength or wavelengths
listed in the Wavelength 1 and Wavelength 2 columns.
To view the chromatogram for this wavelength in Qual Browser, select Wavelength 1 in the
Range list for the Chromatogram window.
Wavelength 2 Specifies the wavelength that the detector is to start monitoring at the time specified in the
Time column.
To view the chromatogram for this wavelength in Qual Browser, select Wavelength 2 in the
Range list for the Chromatogram window.
7. In the Comment box, type a comment concerning the changes you made to the
instrument method.
8. Click Continue to close the File Save – Audit Trail dialog box and save the instrument
method.
The ability to add 512 tasks to a sample preparation routine gives you considerable flexibility;
however, the arrangement of the tasks in the task list must follow a logical order. For example,
you cannot add a task that deposits liquid before you add a task that draws liquid. In addition,
if you are using the 250 μL dual-concentric syringe that ships with the autosampler, you must
follow an additional set of rules that maintain the proper positioning of the inner and outer
plungers of the syringe.
Contents
• Opening the Sample Preparation Page
• Sample Preparation Rules
• Building the Sample Preparation Routine
• Sample Locations for the Sample Preparation Tasks
• Draw, Eject, and Transfer Volumes for Sample Preparation Tasks
• Syringe Speed Range for Sample Preparation Tasks
• Sample Preparation Routine Example
Table 28 categorizes the sample preparation tasks into three groups based on whether the task
uses the inner bore only, the outer bore only, or either bore of the dual-plunger concentric
syringe. For information about the dual-plunger concentric syringe, see “Interchangeable
Syringe” on page 14.
Table 28. Sample preparation tasks arranged in groups according to bore usage
Inner bore only Outer bore only Inner or outer bore
Draw from Sample Mix at Sample Draw from Reservoir
Transfer from Mix at Reservoir Deposit Liquid in Sample
Sample to Sample
Transfer from Wash Needle Deposit Sample in Reservoir
Sample to Reservoir
Flush to Waste Transfer from Reservoir to Sample
Transfer from Reservoir to Reservoir
For the 250 μL concentric syringe, these rules restrict the arrangement of the tasks in the
sample preparation list:
• For tasks performed with the inner bore of the dual-concentric syringe, the maximum
volume (sample + air bubble) that the autosampler can draw, deposit, or transfer is
limited to the nominal size of the syringe, 250 μL.
• For tasks performed with either bore, if the sum of the requested volume
(liquid + air bubble) plus any volume left in the needle tubing from a previous step is
less than the nominal syringe size, the dual-plunger concentric syringe uses the small
bore. If this volume is greater than the nominal syringe size, the syringe uses the outer
bore.
• For tasks performed with the outer bore of the dual-plunger concentric syringe, the
maximum volume (sample + air bubble) that you can draw, deposit, or transfer is limited
to 500 μL, except for the Wash Needle and Flush to Waste tasks.
• Crossover between bores is not allowed. This means that you cannot add a task that uses
the outer syringe bore immediately following a task that uses the inner syringe bore.
To switch from the inner bore to the outer bore of the syringe, you must insert a
Flush to Waste step or a Wash Needle step. These tasks home the position of the syringe
plungers.
Tip The syringe position and nominal volume limit the task order and transfer
volume:
• If the previous task used the small bore of the syringe, you must add a
Flush to Waste or a Wash Needle task before you can add a task that requires
the use of the large bore of the syringe. These tasks home the concentric
syringe plungers.
• If the previous task used the small bore of the syringe, the transfer volume of
the current task is limited to the nominal syringe size or less. See “Sample
Preparation Rules” on page 115 for more information.
b. Make the appropriate selections and entries for each task parameter.
c. Click Add To Task List.
The task appears in the task list in the Method area.
Tip To save an instrument method, the sample preparation routine must conform
to the rules described in “Sample Preparation Rules” on page 115.
Use this area to select and add tasks to the task list in the Method area.
You can add up to 512 tasks to a sample preparation routine; however, the arrangement of the tasks in the task list
must follow a logical order or you will not be able to save the instrument method. For example, a draw task must
precede a deposit task. In addition, if you are using the 250 μL concentric syringe that comes with the autosampler,
you must follow an additional set of rules that allow for the proper positioning of the inner and outer plungers of the
syringe. (See “Sample Preparation Rules” on page 115).
If you are using the 2500 μL, single plunger syringe, to transfer volumes larger than 500 μL, add an optional tubing
extension assembly to the needle tubing.
Task List Description
Deposit Liquid in Deposits liquid in a sample vial or well.
Sample Parameter Reference, range, or selections
Sample Location “Sample Locations for the Sample Preparation Tasks” on page 122
Volume “Draw, Eject, and Transfer Volumes for Sample Preparation Tasks” on
page 123
Syringe Speed “Syringe Speed for Tasks that Can Use Either Plunger” on page 125
Needle Height Default: 2 mm
Range: 0 to 18 mm
Deposit Liquid in Deposit liquids in a reservoir vial.
Reservoir Parameter Reference, range, or selections
Volume “Draw, Eject, and Transfer Volumes for Sample Preparation Tasks” on
page 123
Destination RV1, RV2, RV3, or RV4
Reservoir
Syringe Speed “Syringe Speed for Tasks that Can Use Either Plunger” on page 125
When the concentric syringe plungers are in the home position and the needle tubing is
flushed, the autosampler can next perform any of the sample preparation tasks, with the
exception of the deposit tasks.
Parameter Reference, range, or selections
Volume Range: 100 to 6000 μL
Reservoir bottle, RV1, RV2, RV3, or RV4
Syringe Speed 1.65 to 661.38 μL/s for the concentric syringes
0.83 to 330.85 μL/s for the 2500 μL standard syringe
To mix a sample, the autosampler draws a specified volume of the sample into the needle
tubing at a specified speed, and then ejects the sample back into the same location at a
specified speed.
Parameter Reference, range, or selections
Sample Location “Sample Locations for the Sample Preparation Tasks” on page 122
Volume The maximum volume is 490 μL for the concentric syringes and
1490 μL for the standard 2500 μL syringe.
Draw Speed The draw or delivery speed range is from 1.65 to 661.38 μL/s for the
Delivery Speed concentric syringes, and from 0.83 to 330.85 μL/s for the standard
2500 μL syringe.
Cycles Range: 1 to 10
Needle Height Default: 2 mm
Range: 0 to 18 mm
Mix at Reservoir Mixes the contents of a reservoir vial.
Parameter Reference, range, or selections
Volume The maximum volume is 490 μL for the concentric syringes and
1490 μL for the standard 2500 μL syringe.
Destination RV1, RV2, RV3, or RV4
Reservoir
Draw Speed The draw or delivery speed range is from 1.65 to 661.38 μL/s for the
concentric syringes, and from 0.83 to 330.85 μL/s for the standard
Delivery Speed
2500 μL syringe.
Cycles Range: 1 to 10
Needle Height Default: 2 mm
Range: 0 to 18 mm
Transfer from Reservoir Transfers liquid from one reservoir to another.
to Reservoir Parameter Reference, range, or selections
Volume “Draw, Eject, and Transfer Volumes for Sample Preparation Tasks” on
page 123
Source Reservoir bottle, RV1, RV2, RV3, or RV4
Destination RV1, RV2, RV3, or RV4
Reservoir
Syringe Speed “Syringe Speed for Tasks that Can Use Either Plunger” on page 125
Use the Method area to display the parameters of a task, remove a task, clear all tasks in the task list, or import the
sample preparation routine from a stored instrument method.
The Method area consists of the (sample preparation) task list, the Remove Task button, the Clear All Tasks button,
and the File Name (read only) box and Import (method) button.
File Name box Displays the name of the selected instrument method that contains the sample preparation
routine that you want to import into the current instrument method.
Buttons
Remove Task Use this button to remove a task from the task list in the Method area.
Clear All Tasks Use this button to clear all tasks from the task list in the Method area.
Import Use this button to open the Open dialog box where you can select an instrument method
with the sample preparation routine that you want to add to the current instrument
method.
After you select an instrument method and click Import, the File Name (read only) box
displays the file name of the instrument method you want to import.
For the Relative Location option, the selections are Current, Current + 1, Current + 2, and
Current + 3. Current is the sample location in the current sequence row. Current + 1 is the
next adjacent vial or well location in the tray, and so on. For example, if the vial location in
the current sequence row is B:1, Current is B:1, then Current + 1 is B:2, and so on.
For the Absolute Location option, you specify an absolute vial or well location. The vial or
well locations depend on the tray configuration (see Table 30).
Table 31. Draw, eject, and transfer volumes for sample preparation tasks (Sheet 2 of 2)
Maximum volume (μL) Maximum volume (μL)
Task
with the inner plunger with the outer plunger
Wash Needle N/A 6000
Flush to Waste N/A 6000
Draw from Reservoir Nominal size of the concentric syringe minus the current 500 μL minus the current
contents (liquid + air bubble) of the needle tubing contents of the needle tubing
Deposit Liquid in Total contents (liquid + air bubble) of the needle tubing Total contents
Sample from previous steps (liquid + air bubble) of the
needle tubing from previous
steps
Deposit Sample in Total contents (liquid + air bubble) of the needle tubing Total contents
Reservoir from previous steps (liquid + air bubble) of the
needle tubing from previous
steps
Transfer from Nominal size of the concentric syringe minus the current 500 μL
Reservoir to Sample contents (liquid + air bubble) of the needle tubing
Transfer from Nominal size of the concentric syringe minus the current 500 μL
Reservoir to Reservoir contents (liquid + air bubble) of the needle tubing
If you enter a speed that is invalid for the volume you are transferring, an error message
appears.
For information about the syringe speed range for the sample preparation tasks, see these
topics:
• Syringe Speed for Tasks that Can Use Either Plunger
• Syringe Speed for Tasks that Use the Inner Plunger
For the tasks listed in Table 32, the autosampler uses the following:
• The outer plunger if the volume of liquid + air being transported is greater than the
nominal syringe size
• The inner plunger if the volume of liquid + air being transported is less than or equal to
the nominal syringe size
Table 32. Tasks that can use either syringe plunger
Task
Draw from Reservoir
Deposit Liquid in Sample
Deposit Sample in Reservoir
Transfer from Reservoir to Sample
Transfer from Reservoir to Reservoir
Table 33 lists the speed range for the sample preparation tasks that can use either plunger of
the concentric syringe.
Note The ToolTips for the speed parameters list a range of allowable speeds. The
minimum speed listed is valid for only the inner plunger, and the maximum speed listed is
valid for only the outer plunger of the configured syringe.
If you enter a volume that triggers the use of the outer plunger and a speed that is valid for
only the inner plunger, an error message appears when you attempt to add the task to the
sample preparation list.
Table 33. Syringe speed ranges for tasks that can use either syringe plunger
Syringe size Volume Minimum speed Maximum speed Default speed
(μL) (liquid + air) (μL/s) (μL/s) (μL/s)
100 ≤ 100 0.33 13.25 3.0
100 >100 1.65 661.38 250
250 ≤ 250 0.83 33.09 8.0
250 >250 1.65 661.38 250
500 ≤ 500 1.65 66.10 8.0
500 >500 1.65 661.38 250
2500 0 to 1500 0.83 331 25.0
(standard)
The syringe size determines the default syringe speed and the syringe speed range
(see Table 34).
Use a syringe speed lower than the default for viscous samples. Also, use a syringe speed lower
than the default for samples of very low viscosity or surface tension to prevent the sample
bolus from breaking apart during the transport process.
Note You select the size of the syringe by using the Syringe Type list on the
Communication page of the Surveyor AS Configuration dialog box.
Table 34. Syringe speeds for tasks that use the inner plunger of the concentric syringe
Syringe size Minimum speed Maximum speed Default speed
(μL) (μL/s) (μL/s) (μL/s)
100 0.33 13.25 3
250 0.33 33.10 8
500 1.65 66.11 8
Figure 70. Settings for the Transfer from Sample to Sample task
Default settings
c. Click Add To Task List to add the task to the sample preparation routine.
When the autosampler performs this task, it withdraws 50 μL of sample from the
current vial location in the sequence table and deposits it in the current + 1 vial
location.
4. Add a task that transfers diluent (with the inner syringe plunger) from a reservoir vial to
the sample location:
a. In the Prep Operations list, select Transfer from Reservoir to Sample. The
parameters for the task appear below the Add To Task List button.
b. Keep all the default parameter settings for the Transfer from Reservoir to Sample task,
except those that are shown below (see Figure 71).
Tip When accuracy is important, keep the transfer volume below the nominal
syringe size, so that the autosampler uses the inner syringe plunger. Solvent
transfers made with the inner plunger are more accurate than solvent transfers
made with the outer plunger because the stepper motor for the inner plunger
takes smaller steps and provides finer control than the stepper motor for the outer
plunger.
Figure 71. Settings for the Transfer from Reservoir to Sample task
Default settings
c. To add this task to the sample preparation routine twice, click Add To Task List
twice.
When the autosampler performs this task, it uses the inner syringe plunger to
withdraw 225 μL of diluent from the reservoir vial 1 and deposit the diluent in the
current + 1 vial location. Performing this task twice transfers a total of 450 μL of
diluent to the current + 1 vial location.
5. Add a task that homes the position of the concentric syringe plungers as follows:
Note The syringe cannot switch between bores without first homing the position of
the two syringe plungers. The Transfer from Reservoir to Sample task that you added
in step 4 uses the inner syringe bore. The Mix at Sample task that you add in step 6
uses the outer syringe bore. The Wash Needle task that you add in this step homes the
position of the two syringe plungers.
a. In the Prep Operations list, select Wash Needle. The task parameters appear below
the Add To Task List button.
b. Keep all the default parameter settings for the Wash Needle task (see Figure 72).
Figure 72. Default parameter settings for the Wash Needle task
Default settings
c. Click Add To Task List to add this task to the sample preparation routine.
6. Add a task that mixes the solution in the new sample location:
a. In the Prep Operations list, select Mix at Sample. The task parameters appear below
the Add To Task List button.
b. Keep all the default parameter settings for the Mix at Sample task, except those listed
below (see Figure 73).
c. Click Add To Task List to add this task to the sample preparation routine.
When the autosampler performs this task, it aspirates and expunges 250 μL of the
sample solution in the current + 1 vial location 10 times.
Figure 73. Settings for the Mix at Sample task
Default settings
Default setting
7. Review the task list by expanding the tasks in the Method list (see Figure 74).
Figure 74. Method area with an expanded task list
8. Save the instrument method (see “Saving the Instrument Method” on page 109).
To perform this sample routine, you must insert empty vials into your sample tray.
For example, to dilute five samples, place the samples in vial locations A1, A3, A5, A7,
and A9. Place empty vials in vial locations A2, A4, A6, A8, and A10. Fill reservoir vial 1
with an appropriate diluent that matches the sample matrix.
If you do not want to inject the diluted samples, create a five-line sequence that lists the
vial locations of the original samples (A1, A3, A5, A7, and A9). The original samples are
not injected.
If you want to inject the diluted samples, create a ten-line sequence that lists all the vial
locations (A1 to A10). Ensure that you use the method that contains the sample
preparation routine on the odd rows only (Rows 1, 3. 5. 7. and 9).
Daily Operation
During the initial installation of the Surveyor Plus LC system, a Thermo Fisher Scientific field
service engineer sets up the solvent lines, primes the pulse dampener of the Surveyor MS
Pump Plus, and calibrates the PDA detector.
Note The pulse dampener of the Surveyor LC Pump Plus does not require priming.
To prepare the system for daily operation, check the status of the devices, warm up the
detector’s lamps, and remove air from the solvent lines.
To maintain optimal performance of the system, calibrate the PDA detector on a monthly
basis or whenever you move the detector, and check the light throughput to the diode array
whenever you move the detector, observe an increase in detector noise, or replace the flowcell
or lamps (see “Verifying the Performance of the PDA Detector” on page 313).
Prime the pulse dampener of the Surveyor MS Pump Plus on a monthly basis or when you
notice an increase in pump pulsation.
Contents
• Turning On the Power to Each LC Device
• Initiating Communication with the Data System
• Checking the Status of the LC Devices
• Turning Devices On, Off, or into Standby from the Info View
• Filling the Solvent Reservoir Lines with Fresh Solvent
• Priming the Pulse Dampener of the Surveyor MS Pump Plus
• Accessing the Direct Controls
The following procedure describes how to verify instrument communications from the
Thermo Xcalibur data system.
1. From your computer desktop, choose Start > Programs > Thermo Xcalibur > Xcalibur.
The Thermo Xcalibur Roadmap window appears (see Figure 75).
Figure 75. Thermo Xcalibur Homepage window with the Roadmap and Info view displayed
Information view Roadmap
Shortly after you open the data system, the Comm LEDs on the front panels of the LC
devices turn green if the devices are powered on and connected to the data system
computer.
2. If the Comm LEDs remain amber, do the following:
• Verify that the Ethernet communication cables for the Surveyor PDA Plus Detector
or Surveyor UV/Vis Plus Detector, Surveyor Autosampler Plus, and Surveyor LC
Pump Plus are connected to the Ethernet switch and that the Ethernet switch is
connected to the data system computer. Verify that the RS-232 cable for the Surveyor
MS Pump Plus is connected to the data system computer.
• Verify that the stack addresses in the configuration for the Surveyor Autosampler
Plus, Surveyor PDA Plus Detector or Surveyor UV/Vis Plus Detector, and Surveyor
LC Pump Plus are correct.
For information about configuring your instrument devices, see Chapter 2, “Thermo
Foundation Instrument Configuration.”
Tip You can also view the system pressure on the pump page of the Inlet Direct Control
dialog box in the Tune window for your Thermo Scientific mass spectrometer.
1. In the Home Page window, open the Information view by doing one of the following:
• Choose View > Info View.
–or–
• Click the Information view icon ( ) in the toolbar.
If you have just recycled the power and have not yet downloaded a method, you see the
following status readouts on the Status page:
• Initializing is displayed while the data system attempts to connect to an instrument
module.
• Lamp Warm-up is displayed for the detector while the deuterium lamp is igniting.
• Ready to Download is displayed after the data system establishes communication
with an instrument device and after each run has ended.
Check the status of each device by clicking its name in the device tree list. The status
information for a specific device is displayed in the lower portion of the Status page.
Table 35 describes the states that might appear below the Surveyor AS listing.
AS Status Page
Table 36 describes the readbacks on the AS Status page of the autosampler status view.
Table 36. Status readbacks for the autosampler (Sheet 1 of 3)
Readback Description
Target injection
Vial Position Displays the current vial or microplate well position. For information about the vial and
well notation, see “Vial and Well Notation” on page 6.
Injection Volume Displays the injection volume for the current injection.
For information about selecting the Verify Door Is Closed check box,
see “Communication Page” on page 52.
Closed The Door State displays Closed for these conditions:
• The Verify Door Is Closed Check box is selected and the tray
compartment door is closed.
• The Verify Door Is Closed check box is clear and the door is open
or closed.
The autosampler cannot start a run until you zero the counter or clear
the Enable Maintenance Log check box.
Oven Page
Figure 78 shows the Oven page of the status view for the autosampler.
Figure 78. Oven page
Table 37 describes the readbacks on the Oven page of the autosampler status view.
Table 37. Oven page readbacks
Readback Description
Control On Displays Yes or No.
You can turn off the oven temperature control by doing one of the following:
• Using the Turn Off Oven Temperature command (see “Controlling the Tray and
Oven Compartment Temperatures” on page 199)
• Running an instrument method that has the Enable Column Oven Control check
box cleared
Current Temperature (C) Displays the temperature monitored by the autosampler’s internal temperature sensor.
Set Temperature (C) Displays the user-specified temperature.
You can change the set temperature by doing one of the following:
• Using the Set Oven Temperature command (see “Controlling the Tray and Oven
Compartment Temperatures” on page 199)
• Running an instrument method that has a temperature setting for the column oven
Ready Status Displays one of the following:
Ready The oven compartment temperature has reached the set temperature, you
did not select the Wait for Temperature Ready check box when you
specified the configuration settings for the autosampler, or both.
For information about the Wait for Temperature Ready check box, see
“Communication Page” on page 52.
Not Ready The oven compartment temperature has not reached the set temperature.
Tray Page
Figure 79 shows the Tray page of the autosampler status view.
Figure 79. Tray page
Table 38 describes the readbacks on the Tray page of the autosampler status view.
Table 38. Tray page parameters
Readback Description
Control On Display Yes or No.
You can turn off the tray temperature control by doing one of the following:
• Using the Turn Off Tray Temperature command (see “Controlling the Tray and Oven
Compartment Temperatures” on page 199)
• Running an instrument method that has the Enable Column Oven Control check
box cleared
Current Temperature (C) Displays the temperature monitored by the autosampler’s internal temperature sensor.
Set Temperature (C) Displays the user-specified temperature.
You can change the set temperature by doing one of the following:
• Using the Set Tray Temperature command (see “Controlling the Tray and Oven
Compartment Temperatures” on page 199)
• Running an instrument method that has a temperature setting for the column oven
Ready Status Displays one of the following:
Ready The oven compartment temperature has reached the set temperature,
you did not select the Wait for Temperature Ready check box when
you specified the configuration settings for the autosampler, or both.
For information about the Wait for Temperature Ready check box,
see “Communication Page” on page 52.
Not Ready The oven compartment temperature has not reached the set
temperature.
Table 40 describes the status readbacks on the General page of the pump status view.
Table 40. General page of the pump status view (Sheet 1 of 2)
Readback Description
Status Displays the pump status as follows:
Ready The pump is ready for a run.
Running Isocratic The solvent flow is on, but the pump is not running a gradient
program.
Homing The pump pistons are moving toward the home position.
Running Gradient The pump is running the solvent conditions specified in the
gradient program of an instrument method. The gradient
program can contain isocratic or gradient solvent conditions.
Low Pressure Error The system pressure has fallen below the minimum pressure
setting for more than one minute.
If you are controlling the pump from the Direct Control dialog
box, the status returns to Ready after the pump stops pumping.
The error is posted to the pump’s event log.
High Pressure Error The system pressure has risen above the maximum pressure
setting.
Note When a pressure error occurs, one of the following occurs:
• If you are controlling the pump from the Direct Control dialog box, the status
returns to Ready after the pump stops pumping. The error is posted to the
pump’s event log.
• If you are controlling the pump from the Sequence Setup window, the
Acquisition Server dialog box appears.
Run Time Displays the elapsed time since the last instrument method downloaded, you turned the
device on from the Status page of the Information view, or you started the pump from the
Direct Control dialog box.
Flow Rate Displays the user-specified flow rate. For an instrument method, the flow rate changes
linearly between time lines.
Pressure Displays the pressure monitored by the pump.
Pressure SD Displays the standard deviation of the pressure monitored by the pump.
Start Gradient A check appears in this box when the pump receives the Start gradient signal from the
autosampler.
Pump Ready A check appears in this box when the pressure monitored by the pump reaches the stability
level specified in the instrument method.
Table 41 describes the status readbacks on the Extended page of the pump status view.
Table 41. General page of the pump status view
Readback Description
Flow Rate Displays the user-specified flow rate. Flow rate changes are linear between time lines in the
gradient program.
Tip You can change the percent composition of solvents A, B, C, and D by running an
instrument method or by downloading new solvent conditions from the (Inlet) Direct
Control dialog box.
Pressure Displays the pressure monitored by the pump.
Pressure SD Displays the standard deviation of the pressure monitored by the pump.
Solvent A, B, C, or D Displays the user-specified percent composition of solvent A, B, C, or D.
Stroke Count Displays the stroke count for the pump pistons.
When the stroke count exceeds 1 000 000, the number is displayed in red.
Table 42 describes the status readbacks on the Degasser page of the pump status view.
Table 42. General page of the pump status view
Readback Description
Degasser Status The states of the degasser are as follows:
Ready The vacuum level has reached a sufficient level to degas
the mobile phase solvents.
Not Connected The pump is not communicating with the data system
computer.
Error An error condition has occurred.
Vacuum Level Displays the vacuum level of the solvent degassing chambers.
Motor Speed Displays the motor speed of the vacuum degasser.
Figure 86 shows the Status view for the Surveyor UV/Vis Plus Detector. You can view the
absorbance level of the chromatographic baseline from this view.
Figure 86. Surveyor UV/Vis Plus Detector Status view
Table 45 describes the status readbacks for the Surveyor UV/Vis Plus Detector.
Table 45. UV/Vis detector status readbacks
Readback Description
State Displays the following states: Ready for Run and Off.
Run Time Displays the elapsed run time.
Data Type Displays the program type specified in the current instrument method: Dual UV (D2),
Dual Visible (W), or Single Wavelength.
Lambda 1 (nm) Displays the current wavelength setting for the Wavelength 1 channel.
Lambda 2 (nm) Displays the current wavelength setting for the Wavelength 2 channel.
Absorbance 1 (AU) Displays the current absorbance level for the Wavelength l channel.
Absorbance 1 (AU) Displays the current absorbance level for the Wavelength 2 channel.
Turning Devices On, Off, or into Standby from the Info View
You can turn on certain activities for a device from its device listing in the Info View. To use
this feature, the device must be powered on and communicating with the data system.
Right-click the device listing on the Status page and choose Turn Device On.
Depending on the device, the following actions occur:
• When you turn on the pump, it begins pumping solvents from the last set of
downloaded parameters.
• When you turn on the detector, its lamps turn on.
• When you turn on the autosampler, it adjusts its controlled temperature zones to the
last set of downloaded parameters.
Right-click the device listing on the Status page and choose Turn Device into Standby.
Depending on the device, the following actions occur:
• When you place the pump in Standby mode, it stops pumping.
• When you place the detector in Standby mode, nothing happens.
• When you place the autosampler in Standby mode, nothing happens.
Right-click the device listing on the Status page and choose Turn Device Off.
Depending on the device, the following actions occur:
• When you turn off the pump, it stops pumping.
• When you turn off the detector, the lamps turn off.
• When you turn off the autosampler, nothing happens.
For information about using the direct controls to start the solvent flow from the Surveyor
MS Pump Plus, see “Pump Direct Controls” on page 185.
1. For any solvent line that you want to pull solvent through, disconnect the tubing that is
connected to the corresponding degasser outlet port.
2. Connect a Luer-Lok™ fitting to the degasser outlet port.
3. Connect a 10 mL syringe to the Luer-Lok fitting, and pull solvent through the line.
4. Remove the Luer-Lok fitting from the outlet port.
5. Reconnect the solvent line to the degasser outlet port.
Closing the pulse dampener flush valve after you fill the coil shuts the dampening coil off
from the mobile phase stream. Because a permeable membrane separates the mobile phase
stream from the dampening coil, the solvent in the coil slowly diffuses into the mobile phase
stream. When the pump is in continuous or frequent use, prime the pulse dampener on a
monthly basis.
Note Do not fill the pulse dampener with an aggressive acid or a buffered solution. The
recommended filling solvents are methanol, acetonitrile, or isopropyl alcohol. If you
accidentally fill the pulse dampener with an acidic or buffered solvent, flush the loop with
a miscible solvent, and then refill it with methanol, acetonitrile, or isopropyl alcohol.
Note For information about using the Direct Control dialog box for the Surveyor
MS Pump Plus, see “Pump Direct Controls” on page 185.
–or–
• From the tune window for your Thermo Scientific mass spectrometer, choose
Setup > Inlet Direct Control. The Inlet Direct Control dialog box that contains
tabbed pages for each configured LC device appears. Click the Surveyor MS Pump
Plus tab.
Note For information about using the Inlet Direct Control dialog box, see
Chapter 12, “Making a Single Injection from the Tune Window.”
For information about using the direct controls to prepare your LC system for daily operation,
see the Chapter 6, “Direct Controls.”
• If you are controlling your instrument from a Thermo Scientific data system other
than Xcalibur, open the instrument control area.
The instrument control area opens to the first module displayed in the view bar on
the left side of the window.
Figure 89. Instrument Setup window open to the Surveyor AS Method page of the Surveyor AS view
2. To open the view for the device of interest, click its corresponding icon in the view bar.
The view for the specific device appears.
3. Access the direct controls for each device from the menu bar of the device view as follows:
Direct Controls
You can control some device features without downloading an instrument method. For
example, you can turn the lamps on or off, set the solvent conditions for the pump, start and
stop the solvent flow from the pump, and control the autosampler’s temperature zones and its
XYZ arm by using the direct controls provided by the data system.
For information about accessing the direct controls for each device, see “Accessing the Direct
Controls” on page 158.
Contents
• PDA Detector Direct Controls
• UV-Vis Detector Direct Controls
• Pump Direct Controls
• Autosampler Direct Controls
Note For instructions on loading samples into the autosampler, see “Loading the
Autosampler” on page 210.
To open the Direct Control dialog box for the PDA detector
The Surveyor PDA Plus Direct Control dialog box opens to the Display page
(see Figure 91).
Figure 91. Surveyor PDA Plus Control dialog box tabs
Table 46 lists the direct control procedures by the page where they are performed.
Table 46. Direct control procedures
Page Procedure
Display Page “Checking the Chromatographic Baseline” on page 168
“Verifying the Performance of the PDA Detector” on page 313
Configuration Page “Turning the Lamps On or Off ” on page 163
“Resetting the Lamp Usage Hours” on page 164
“Changing the Polarity of the Analog Outputs” on page 164
“Testing the Analog Outputs” on page 165
“Setting the Lamp Startup Time” on page 166
Information Page “Viewing, Exporting, and Clearing the Error Log for the PDA Detector” on page 171
“Checking the Firmware Version of the PDA Detector” on page 171
For information about using the Calibration page to determine the wavelength accuracy of
the detector and the dark current produced by the diode array, see “Calibrating the PDA
Detector” on page 319.
1. Open the Configuration page of the Surveyor PDA Plus Direct Control dialog box
(see “Configuration Page” on page 171).
2. Click Turn On for the associated lamp.
When you turn on the deuterium lamp, its Status readback reads Starting during the
30-second ignition period, and then it changes to On. If there is a problem with either
lamp, its Status readback reads Failed.
Note The intensity of the deuterium lamp falls off very slightly over a period of time
after the lamp is turned on. Plan to wait at least one hour for the lamp to stabilize
after a cold start before collecting data in the spectral range of the deuterium lamp.
1. Open the Configuration page of the Surveyor PDA Direct Control dialog box
(see “Configuration Page” on page 171).
2. Click Reset Lifetime for the associated lamp.
The stored total run time for the associated lamp resets to zero, and the Last Lifetime
Reset readback is updated to the current date and time.
Tip Avoid indiscriminately clicking the Reset buttons. Click Reset only after you
replace the associated lamp with a new one.
There are two output signal terminals on the back panel of the PDA detector: Event and
Ready (see Figure 92).
Figure 92. Event and Ready terminals on the back panel of the PDA detector
Ready terminal
Event terminal
2 3 2 3
0 1
0 1
4
4
5 6
5 6
9
7 8 7 8
1. Open the Configuration page of the Surveyor PDA Direct Control dialog box
(see “Configuration Page” on page 171).
If the output polarity is Active Low, the Set Active High button is available. If the output
polarity is Active High, the Set Active Low button is available.
1. Open the Configuration page of the Surveyor PDA Plus Direct Control dialog box
(see “Configuration Page” on page 171).
2. Under Analog Outputs (see Figure 93), do the following:
• Click Turn On or Turn Off to trigger the external device.
• Click Zero DACs to calibrate external instruments, such as an SS420X analog to
digital converter.
When you click Zero DACs, the DACs outputs are set to zero for about 20 seconds.
You can extend this calibration time by clicking the button again.
If you do not change the lamp startup time, the data system automatically turns the lamps on
at 4:55 PM on weekdays.
1. Open the Configuration page of the Surveyor PDA Plus Direct Control dialog box
(see “Configuration Page” on page 171).
2. In the Programmed Lamp Startup area, click Change (see Figure 94).
Figure 94. Programmed Lamp Startup area of the Configuration page
The Lamp Startup Time dialog box appears (see Figure 95).
Figure 95. Lamp Startup Time dialog box
1. In the Xcalibur Should Start the Lamps At box, click the hours or minutes text to
highlight it.
Tip To change both the hours and the minutes, change one time field, and then click
OK to accept the setting and close the dialog box. Then open the dialog box a second
time and change the other time field.
Default: 4:55 PM
On These Days Specifies the days that the Xcalibur data system
automatically turns the lamps on. The selections are
Never, Weekdays, and Every Day.
Default: Weekdays
Buttons
OK Applies the settings and closes the dialog box.
Cancel Cancels the applied time value or date.
To check the stability of the chromatographic baseline before starting a sequence run
8. Select the instrument method that you plan to use in your acquisition sequence.
9. Click Open to download the method to the PDA detector and close the dialog box.
10. In the Control area, click Start Data.
The graphical display begins to update. The spectral scan appears in the top pane and the
chromatographic data for the discrete channel wavelengths appears in the bottom pane.
When you click Start Data, the Start Data button changes to the SNAPSHOT button
and the Zero Data button is available (see Figure 98).
Figure 98. Display page with data displaying for an instrument method
Instrument method with
a .meth file extension
Spectral data
Viewing, Exporting, and Clearing the Error Log for the PDA Detector
You can access an error log for the PDA detector from the Xcalibur data system.
1. Open the Information page of the Surveyor PDA Direct Control dialog box (see
“Information Page” on page 181).
2. Click Request Log.
Open the Information page of the Direct Control dialog box for the PDA detector
(see “Information Page” on page 181).
The firmware version of the PDA detector is displayed in the lower-right portion of the
Information page.
Configuration Page
Use the Configuration page of the Surveyor PDA Plus Direct Control dialog box to perform
these procedures:
• Turning the Lamps On or Off
• Resetting the Lamp Usage Hours
• Changing the Polarity of the Analog Outputs
• Testing the Analog Outputs
• Setting the Lamp Startup Time
1. Open the Surveyor PDA Direct Control dialog box (see “PDA Detector Direct Controls”
on page 162).
2. Click the Configuration tab.
The Configuration page appears (see Figure 100).
Use the controls in this area to turn the lamps on and off and to reset the lamp lifetime hours after you replace a lamp.
Deuterium Lamp
Status readback Indicates the status of the deuterium lamp.
Graphic Text Meaning
Off The deuterium lamp is off.
Reset Lifetime button Resets the date and time monitored by the Last Reset readback and the Lifetime readback.
Note Click Reset Lifetime whenever you replace the deuterium lamp. This resets the Last
Reset readback to the current date and time and restarts the Lifetime readback at zero.
Table 48. Configuration page of the Direct Control dialog box for the PDA detector (Sheet 2 of 4)
Parameter Description
Tungsten Lamp
Status readback Indicates the status of the tungsten lamp, as follows:
Graphic Test Meaning
On The tungsten lamp is on.
Reset Lifetime button Resets the date and time monitored by the Last Reset readback and the Lifetime readback.
Note Click Reset Lifetime whenever you replace the tungsten lamp. This resets the Last
Reset readback to the current date and time and restarts the Lifetime readback at zero.
Analog Outputs
Use the controls in this area to configure and test the Event and Ready outputs.
Determine if the external device is triggered by contact closure or a TTL (transistor - transistor - logic) signal. If the
external device requires a TTL signal, determine the polarity of the signal: active high or active low.
Note For TTL connections:
• Logic level zero (0) is less than +0.8 Vdc (reading from Ground).
• Logic level one (5) is approximately +5 Vdc (+2.4 Vdc minimum from Ground).
Possible configurations of the Event and Ready outputs include the following:
Output state Output polarity Vdc
Off Active High +5
Table 48. Configuration page of the Direct Control dialog box for the PDA detector (Sheet 3 of 4)
Parameter Description
Ready Output
Turn On and Turn Off Indicates whether the Ready output state is On or Off. If the Ready output state is On, the
buttons Turn Off button is available. If the Ready output state is Off, the Turn On button is
available.
Ready Output Indicates whether the Ready output polarity is Active High or Active Low. If the Ready
readback output polarity is Active Low, the Set Active High button is available. If the Ready output
polarity is Active High, the Set Active Low button is available.
Set Active Low or Set Set the polarity of the Ready Output signal.
Active High buttons
Note The output polarity must match the polarity (either Active High or Active Low) of the instrument connected
to the Ready output connection on the back panel of the PDA detector. The signal terminal is the high connection
and the ground (GND) terminal is the low connection. The autosampler requires active low remote inputs.
Event Output
Use the controls in this area to specify the parameters for output events.
If the external device is triggered by a contact closure, connect the PDA EVENT terminal (pin 8) to the positive pin
on the external device input, and connect one of the PDA GND terminals (either of pins 1 or 7) to the external
device negative pin.
If the external device is triggered using a TTL signal, connect the PDA +5V output (pin 2) to the positive Input
terminal of the external device, and connect the PDA EVENT output (pin 8) to the negative Input terminal of the
external device.
Turn On and Turn Off Indicates whether the Event output state is On or Off. If the Event output state is On, the
buttons Turn Off button is available. If the Event output state is Off, the Turn On button is
available.
Event Output readback Indicates whether the Event output polarity is Active High or Active Low. If the Event
output polarity is Active Low, the Set Active High button is available. If the Event output
polarity is Active High, the Set Active Low button is available.
Set Active Low or Set Set the polarity of the Event Output signal.
Active High buttons
Note The output polarity must match the polarity (either Active High or Active Low) of the instrument connected
to the Event output connection on the back panel of the PDA detector. The signal terminal is the high connection
and the GND Terminal is the low connection.
Short DACs Output
Zero DACs button To calibrate external instruments, such as an SS420X analog to digital converter, use the
controls in this area to short the DACs outputs to zero for about 20 seconds.
Table 48. Configuration page of the Direct Control dialog box for the PDA detector (Sheet 4 of 4)
Parameter Description
Programmed Lamp Startup
Programmed Lamp Displays the current lamp startup program and provides controls to create a lamp startup
Startup program that automatically turns on the lamps for operation at a specified time and day
(never, weekdays, every day).
Change Opens the Lamp Startup Time dialog box where you can specify a lamp startup time
(see “Setting the Lamp Startup Time” on page 166).
Display Page
Use the Display page to perform these procedures:
• “Checking the Chromatographic Baseline” on page 168
• “Adjusting the Light Throughput to the Diode Array” on page 315
1. Open the Surveyor PDA Direct Control dialog box (see “PDA Detector Direct Controls”
on page 162).
2. Click the Display tab.
The Display page appears (see Figure 101).
A date and time stamp are appended to the file name. The format of the date and time
stamp is MMDDYYHHMMSS. MM is the month. DD is the day of the month. YY are the
last two digits of the year. HH is the hour in military time. MM are minutes. SS are
seconds. You can open this file with the Excel™ application.
For a wavelength/absorbance method, the file contains the absorbance values for the spectral
scan and the discrete channel wavelengths at the moment that you clicked Snapshot.
For an intensity/diode method, the file contains the intensity values for the scan of the
diode array and the intensity values for up to three individual diodes.
Note The Snapshot button is available only after you click the Start Data button and the
graphical display has started updating.
Zero Data Zeros the absorbance data.
After you click Zero Data, the Current Method status box displays the message Zeroed Data
Display.
Note This button is only available in the wavelength/absorbance mode.
Stop Data Stops the updating of the graphical display.
After you click Stop Data, the Current Method status box displays the message Stopped
Data Display.
Levels (mAU)
Channel A/B/C level This graphical display shows the intensity or absorbance level of each discrete channel, A, B,
or C, specified in the method.
Lamp Lifetimes (hours)
Deuterium Displays the cumulative time in hours that the Deuterium lamp has been in the On state
since you clicked the Reset button.
Note Click Reset Lifetime on the Configuration page whenever you replace the
deuterium lamp. This resets the Last Reset readback to the current date and time and sets
the Lifetime readback to zero.
Tungsten Displays the cumulative time in hours that the tungsten lamp has been in the On state since
you clicked Reset Lifetime.
Note Click the Reset Lifetime button on the Configuration page whenever you replace
the tungsten lamp. This resets the Last Lifetime Reset readback to the current date and
time and restarts the Lifetime readback to zero.
If this readback displays Not Connected, make sure that the PDA detector is powered on.
Check the Ethernet connections and the instrument configuration settings for the PDA
detector.
Diode array scan rate Displays the current configuration setting for the diode array scan rate. The possible diode
array scan rates are 20, 40, and 80 Hz.
PDA status Displays the current state of the PDA detector. The possible states are Not Ready for a Run
and Ready for a Run. This readback displays Not Ready for a Run if both lamps are off or if
the deuterium lamp is warming up.
Wavelength calibration Displays whether the wavelength calibration has been applied.
Dark current Displays whether the dark current calibration has been applied.
calibration
Method validity Displays whether a method has been downloaded to the PDA detector.
Filter wheel position Displays the position of the filter wheel. The filter wheel has two positions: 1 and 2. When
the filter wheel is in position 1, the readout displays Run Position. When the filter wheel is
in position 2, the readout displays Calibration Position.
Deuterium lamp Displays whether the deuterium lamp is on or off.
Tungsten lamp Displays whether the tungsten lamp is on or off.
Current method (top right)
Current method Displays the status of the current method.
Spectrum
Spectrum Plot Depending on the method type (.meth or .spda), these scan plots are displayed:
• For display methods (.spda), the scan plot shows intensity (counts) normalized to
100 percent on the y axis and diode number (2 to 511) on the x axis.
• For instrument methods (.meth), the scan plot shows absorbance (mAU) normalized to
100 percent on the y axis and wavelength (nanometers) on the x axis.
Each discrete channel is color coded. Channel A is yellow, Channel B is purple, and
Channel C is blue.
Information Page
Use the Information page to monitor the chronological run log, instrument version, and
system date and time. The error log is an ongoing record of critical and non-critical errors in
the PDA detector. The log is generated and stored in the detector. This data is useful for
troubleshooting hardware and software problems.
1. Open the Direct Control dialog box for the PDA detector (see “PDA Detector Direct
Controls” on page 162).
2. Click the Information tab.
The Information page appears (see Figure 102).
Figure 102. Information page
Click Zero.
Table 51. Direct control parameters for the UV/Vis detector (Sheet 2 of 2)
Parameter Description
Tungsten Lamp
Lamp On Turns the tungsten lamp on.
Lamp Off Turns the tungsten lamp off.
Other Options
Zero Zeros the absorbance reading from the detector.
Readbacks
Status Displays the state of the UV/Vis detector:
Ready The detector is ready to start data acquisition.
Off The detector is not ready to start data acquisition
because one or both lamps are off.
These topics describe the direct controls for your Surveyor Plus pump:
• Surveyor LC Pump Plus Direct Controls
• Surveyor MS Pump Plus Direct Controls
For information on opening the Direct Control dialog box and using the direct controls for
the Surveyor LC Pump Plus, see these topics:
• Opening the Direct Control Dialog Box for the Surveyor LC Pump Plus
• Operation - Surveyor LC Pump Direct Control Parameters
• Removing Air from the Solvent Lines Connected to the Surveyor LC Pump Plus
• Downloading New Solvent Conditions to the Pump
• Starting and Stopping the Solvent Flow
• Clearing an Error Condition for the Pump
Opening the Direct Control Dialog Box for the Surveyor LC Pump Plus
In the pump view, the direct controls are on the Operation - Surveyor LC Pump Direct
Control dialog box. In the Tune window for your Thermo Scientific mass spectrometer, the
direct controls for the pump are on the pump page of the Inlet Direct Control dialog box.
To access the direct controls for the pump from the Tune window
1. Open the Tune window for your Thermo Scientific mass spectrometer.
2. Choose Setup > Inlet Direct Control.
The Inlet Direct Control dialog box that contains tabbed pages for each configured LC
device appears.
3. Click the Surveyor LC Pump tab.
Removing Air from the Solvent Lines Connected to the Surveyor LC Pump Plus
Use the purge feature to remove air bubbles from the solvent lines or the pump heads. If the
solvent line connecting the solvent reservoir bottle to the degassing assembly is completely
dry, attach a syringe to the drain line and open the drain valve. Make sure that the pump
power is on, and then draw the syringe barrel back, pulling solvent into the line.
Tip Cyclical baseline noise can be an indication of air in the pump heads.
To remove air from the solvent lines or the pump head assemblies
1. Open the drain valve by turning it counterclockwise 180° to the purge position. Look for
the word DRAIN to appear upside down on the knob (see Figure 104).
Figure 104. Drain valve knob in the closed and open positions
Closed position Open position
DRAIN
SE
O
PE
LO
N
C
C
N
LO
PE
SE
DRAIN
O
Figure 105. Operation page of the Surveyor LC Pump Direct Control dialog box
1. Open the Operation - Surveyor LC Pump Direct Control dialog box (see “Surveyor LC
Pump Plus Direct Controls” on page 185).
2. Do one of the following:
• To start the solvent flow, click Pump On.
• To stop the solvent flow, click Pump Off.
Tip To open the direct controls for the pump from the Tune window, choose Setup >
Inlet Direct Control from the menu bar. Then click the pump tab.
Tip When the power to the pump is off, the proportioning valves are closed. Before
you take the LDA apart to perform maintenance, turn off the power to the pump to
prevent solvent from leaking out of the proportioning assembly.
2. To open the Direct Control dialog box from the pump view, do the following:
a. From the menu bar of the pump view, choose Surveyor MS Pump Plus >
Direct Control.
The Direct Control dialog box for the pump appears (see Figure 106).
Figure 106. Direct Control dialog box for the Surveyor MS Pump Plus
The maximum flow rate for the Surveyor MS Pump Plus is 1000 μL/min.
c. Type percentages for the solvent lines that you want to draw solvent through in the
Inlet boxes (A, B, C, and D).
Click (Start).
Click (Stop).
2. In the Min. Pressure box, type the minimum operating pressure for the pump.
3. In the Max. Pressure box, type the maximum operating pressure for the pump.
The maximum pressure for the Surveyor MS Pump Plus is 1000 bar (100 MPa or
14 504 psi).
4. Click OK to accept the settings and close the dialog box.
You can access the direct control commands for the autosampler from the autosampler view of
your data system or the tune application for your mass spectrometer.
To open the Inlet Direct Control dialog box from the tune application
Figure 108. Inlet Direct Control dialog box for the Surveyor pump and autosampler
Tip During a flush operation, the autosampler draws wash solvent into the syringe and
then pushes wash solvent through the needle tubing and the injection port to waste.
If you want to remove residual sample from the sample loop, use the Flush Syringe at Fill
Position command.
To flush air out of the wash bottle tubing and the autosampler syringe
Concentric
syringe
To To
Waste
2 1 column 2 1 column
tube
6 6
3 3
4 5 4 5
From From
pump pump
Waste Waste
container container
Mobile phase
Wash solvent
For information on specifying the configuration settings for the Surveyor Autosampler Plus,
see “Surveyor Autosampler Plus Configuration Settings” on page 45.
Use the direct commands in the Direct Control dialog box or the Inlet Direct Control dialog
box to control the oven and tray compartment temperatures. For information about opening
the Direct Control and Inlet Direct Control dialog boxes, see “Autosampler Direct Controls”
on page 192.
IMPORTANT Do not set the oven temperature above the boiling point for the
mobile phase solvent.
3. Click Apply.
Sequence Setup
This chapter describes how to acquire and view chromatographic and PDA spectral data using
the Xcalibur data system.
Contents
• Creating a Single Sample Sequence
• Equilibrating the Chromatographic Column
• Loading the Autosampler
• Starting Data Acquisition
• Working with the Real Time Plot View
1. If it is not already open, open the New Sequence Template dialog box (see “Opening the
Sequence Setup Window” on page 201).
2. In the General area, make the following entries and selections:
a. In the Base File Name box, type a name for the raw data file.
b. Browse to the data file directory where you want to store your raw data files.
The data system adds the .raw file extension to the data files that contain the
chromatographic and spectral data.
c. Browse to the instrument method that you want to use to acquire your raw data files.
Instrument methods have a .meth file extension. The Instrument Setup view for the
Surveyor Plus devices is described in Chapter 3, “Instrument Method Setup.”
d. If you have not yet created a processing method that contains the information needed
to quantitate your unknowns, do not select a processing method.
You can create a processing method and reprocess your stored data files at a later date.
Processing methods have a .pmd file extension. For information about performing
tests to determine the suitability of your chromatographic method and on creating
calibration curves to quantitate your unknowns, refer to the Thermo Xcalibur
Quantitative Analysis User Guide.
3. In the Samples area, specify the vial locations by doing one of the following:
• To specify a contiguous set of sample vials or well locations:
– In the Number of Samples box, type the number of samples that you want to
analyze.
– In the Initial Vial Position box, type the vial position.
The vial positions for the conventional trays are shown in Figure 114.
RV4
RV3
Standard RV2
1.8 mL vial
RV1
A B C D E
Tray handle
For more information about the vial locations for conventional trays and the well
locations for microwell plates, see “Vial and Well Notation” on page 6.
• To select a non-contiguous set of sample vials or well locations, click Select Vials, and
then select the vial or well locations by using the Vial Selection dialog box.
For information about using the Vial Selection dialog box, see “Selecting the Vial
Locations Interactively” on page 206.
4. In the Injections per Sample box, type the number of injections per sample that you want
the autosampler to make per vial or well location.
5. In the Base Sample ID box, type an identifying name for the sample.
Base sample IDs are optional. If you do not enter a sample identification, the data system
automatically uses the vial position as the sample identification. If you enter a sample
identification, the data system automatically appends the vial position to your entry.
6. Click OK to display your sequence spreadsheet (see Figure 115).
The injection volume displayed in the Inj Vol column matches the injection volume
contained in your instrument setup method. You can override this injection volume
value.
7. To change the injection volume, double-click the spreadsheet cell containing the injection
volume value that you want to change, highlight the current value, and then type a new
value in the cell.
For full details of all the parameters in the New Sequence Template dialog box, refer to
the Help or the Xcalibur manual set.
Figure 115. Sequence Setup view, showing newly created one-line sequence
3. Browse to the appropriate folder where you want to save the sequence.
Vm = Πr2 × L
Where:
Vm = volume in mL
r = column radius in cm (radius = inner diameter/2)
L = column length in cm
1. From the Thermo Xcalibur Roadmap menu, choose GoTo > Instrument Setup to
display the Instrument Setup window.
2. Download the same solvent percentages and flow rate as those contained in your method
to the pump:
• For information on downloading new solvent conditions to the Surveyor LC Pump
Plus and starting the solvent flow, see “Surveyor LC Pump Plus Direct Controls” on
page 185.
• For information on downloading new solvent conditions to the Surveyor MS Pump
Plus and starting the solvent flow, see “Surveyor MS Pump Plus Direct Controls” on
page 190.
3. In the status view for your pump (available on the Status page of Info View), monitor the
readings in the Pressure Status area to ensure that the pressure is appropriate for your
application. See “Viewing the Status of Each Device” on page 137.
Before you start a sequence run, ensure that you have samples in the locations specified in
your sequence.
Note To trigger the vial sensor, make sure to position custom vials in the tray so that the
top of the vial reaches the minimum height of 1.55 inches. If you place vials that fall
below this minimum height in the tray, the vial sensor does not detect them. When the
sequence reaches a vial that is below the minimum height, the sequence halts, and the Vial
Not Found message appears.
1. Open the sequence file containing the information for the sample that you want to inject:
a. From the Sequence Setup view, choose File > Open.
The Open dialog box appears (see Figure 120).
Figure 120. Open dialog box, showing the selection of a sequence file
Figure 121. Sequence Setup view, showing the first row selected
3. Confirm that you have a vial in the position specified in the sequence row.
4. From the toolbar, click the Run Sample button.
The Run Sequence dialog box appears (see Figure 122).
The User box contains your login name, and the Run Rows box contains the row number
that you selected in the sequence spreadsheet.
Figure 122. Run Sequence dialog box
After the autosampler injection valve switches to the inject position, the autosampler
sends a signal to the detector to begin data acquisition.
6. Set up how the run is started as follows:
• To start the run automatically, select the Start When Ready check box.
After you complete the entries and selections in the Run Sequence dialog box and
click OK, the run begins after the pump sends a pump ready signal to the
autosampler. The pump does not indicate the Ready state until it monitors a stable
backpressure as defined in your instrument method.
• To start the run manually, clear the Start When Ready check box.
7. In the After Sequence Set System area, select one of these three options:
• To keep the devices in the On state, select the On option.
• To automatically turn off the solvent flow from the pump and the detector’s lamps at
the end of the sequence run, select the Off option.
• To turn off the solvent flow from the pump but keep the lamps at the end of the
sequence run, select the Standby option.
8. Keep the parameters in the other areas at their defaults.
You can use the Instrument Method and Programs areas in the Run Sequence dialog box
to specify particular acquisition or processing requirements. For full details of these
features, refer to the Xcalibur Help.
9. Start the run as follows:
• For automated runs, click OK to start the run.
• For manual runs, click OK, and then choose Actions > Start Analysis from the
Sequence Setup menu. To subsequently control the acquisition, choose Actions >
Pause Analysis or Actions > Stop Analysis.
1. Choose View > Real Time Plot View, or click the Real Time Plot View button on the
Home Page toolbar.
2. If the display is not already locked, click the Lock Display button to lock the display.
Locked
Locking the display lets you monitor the real-time progress of your run.
In the unlocked position, you cannot monitor the real-time progress of your run, but you
Unlocked can review your data. For example, you can display the spectrum for a particular peak that
has already eluted. Data collection continues off screen as you review your data.
If you are collecting PDA scan data, a view similar to that shown in Figure 124 is
displayed. The view contains three cells: a chromatogram cell, a spectrum cell, and a Total
Scan cell.
Figure 124. Real Time Plot view, showing the acquisition of PDA scan data and one discrete channel
Lock display button
1. Click the chromatogram cell to make it the active cell with a gray border.
2. Choose View > Ranges.
The Chromatogram Ranges dialog box appears (see Figure 125).
Qual Browser
This chapter provides an introduction to the Xcalibur Qual Browser functions that you can
use to review the PDA data contained in your .raw data files.
Contents
• Opening a Raw Data File in Qual Browser
• Working with the Cell Grid
• Changing the Font Size of the Display
• Viewing a Report of the Instrument Method
• Creating a Layout for PDA Data
• Viewing the Spectrum for a Specific Time Point
• Viewing the Chromatogram for a Specific Wavelength
• Determining Peak Areas
• Calculating the Purity of the Chromatographic Peaks
1. Click the Qual Browser icon in the Roadmap view of the Homepage window, or choose
GoTo > Qual Browser.
The empty Qual Browser window appears (see Figure 126).
4. Select the layout from the list at the bottom of the dialog box.
The available selections are Default Layout and Current Layout. Select Current Layout if
the current layout for the Qual Browser window is different from the default layout and
you want to apply it to your data file.
5. Click Open.
If the default layout has not been modified and your raw file contains MS data in
addition to PDA or UV data, the data file opens with the MS TIC chromatogram in the
upper cell and a mass range spectrum in the lower cell. For an explanation of cells, see
“Working with the Cell Grid” on page 224. The y axis for these cells is set to relative
absorbance (see Figure 128).
Figure 128. Qual Browser view with a chromatogram cell displaying MS TIC data and a spectrum cell displaying
mass range data
6. To apply a custom window layout to the Qual Browser view, do the following:
a. Choose File > Layout > Apply.
The Open Layout File dialog box appears (see Figure 129).
Figure 130. Qual Browser view with a chromatogram cell displaying a Total Scan from the PDA detector and a spectrum cell
displaying the spectral data from the PDA detector for time 0
Cell States
When you open a raw data file in the Qual Browser window, the information within the data
file is displayed as a grid of cells.
There are three hierarchal states for a cell within the grid:
• Inactive Cells
• Active but Unpinned Cells
• Active and Pinned Cells
The grid always contains either one active but unpinned cell or one pinned cell. If the grid
contains more than one cell, only one cell can be active and the rest of the cells are inactive.
Inactive Cells
Inactive cells are not highlighted with a gray border and the pin icon in their upper-right
corners is gray. The cell in the lower portion of Figure 131 is inactive as indicated by the
absence of a gray border. Menu commands, toolbar buttons, and cursor actions do not affect
inactive cells. To zoom in on the contents of a cell or access its shortcut menu, you must make
the cell active.
Figure 131. Qual Browser window, displaying an active chromatogram cell and an inactive PDA spectrum cell
Gray border
Active cell
Unpinned
To automatically change the range of a cell by clicking in the grid, you must pin the cell. For
example, to display the spectrum for the 1 minute time point without opening the Spectrum
Ranges dialog box, pin the spectrum cell, and then click the 1 minute time point in the
inactive chromatogram cell.
Figure 132. Qual Browser window, displaying an inactive cell and a pinned cell
Inactive cell
Pinned cell
Cursor Actions
Within the cells of the grid, you can use the cursor in three ways:
• A click selects a point on the cell.
• A line dragged parallel to any axis selects a range.
• A line dragged in any diagonal direction selects an area.
The effect of these actions depends on the state of the cell. Within an active cell, cursor
actions rescale the plot (see Table 54).
Table 54. Effect of cursor action in an active cell
Cursor action Effect
Drag parallel to x axis Rescales graph showing selected x-axis range only, same
y-axis range.
Drag parallel to y axis Rescales graph showing selected y-axis range only, same
x-axis range.
Dragged area Rescales graph showing both the selected x-axis and y-axis
ranges.
If one of the cells is pinned, the cursor action in any of the inactive cells affects the pinned cell
(see Table 55).
Table 55. Effect of cursor action in an inactive cell on the pinned cell
Pinned cell Cursor action Effect
Spectrum Click in a chromatogram cell The spectrum cell displays the
spectrum at that retention time.
Chromatogram Click in a spectrum cell The chromatogram cell displays the
chromatogram for the wavelength
selected in the spectrum cell.
In Figure 133, the spectrum cell on the bottom of the view is pinned. Clicking the 1.8 minute
time point in the unpinned chromatogram cell causes the spectrum of benzene, which elutes
at 1.8 minutes, to appear in the pinned spectrum cell.
Figure 133. Qual Browser window with a pinned spectrum cell and an unpinned chromatogram cell
In the chromatogram cell, the pointer is
set to 1.8 minutes, the peak apex for
benzene.
In Figure 134, the chromatogram cell on the top of the view is pinned. Clicking a specific
wavelength in the unpinned spectrum cell displays the chromatogram of the scan wavelength
in the pinned chromatogram cell.
Figure 134. Qual Browser window with a pinned chromatogram cell and an unpinned spectrum cell
Clicking (Full Size) in the toolbar sizes the active or pinned cell to the full width and
height of the window as shown in Figure 135.
Dragging across a region in the active or pinned cell zooms in on that region as shown in
Figure 136.
Figure 136. Full size view of the spectrum cell zoomed in on the 320 to 380 nm region
1. From the Thermo Xcalibur Roadmap window, choose Tools > Configuration.
The Thermo Xcalibur Configuration dialog box appears.
2. Click the Fonts tab.
The Fonts page appears (see Figure 137).
Figure 137. Fonts page in the Thermo Xcalibur Configuration dialog box
3. To increase the font size of the chromatogram, spectrum, and map axis labels, in the
Proportional Pitch Font area, select a larger font size from the Max list, and select a larger
font size from the Min list.
4. To increase the font size for the Spectrum List, Scan Header, Scan Filters, or Report, in
the Fixed Pitch Font area, select a larger font size from the Max list, and select a larger
font size from the Min list.
Figure 138 shows the chromatogram axes labeled with a font size of 8. Figure 139 shows
the chromatogram axes labeled with a font size of 16.
To create a layout file for PDA data, follow these procedures in order:
1. Specifying the Chromatogram and Spectrum Ranges
2. Specifying the Display Options for the Chromatogram and Spectrum Cells
3. (Optional) Inserting Cells
4. Saving the New Layout
Follow these topics to change the upper cell to a chromatogram and the lower cell to spectral
data acquired by the PDA detector:
• Specifying the Chromatogram Ranges
• Specifying the Spectrum Range
1. Open a data file (.raw) that has PDA data and MS data (see “Opening a Raw Data File in
Qual Browser” on page 219).
The data file opens with the MS TIC chromatogram in the upper cell and a mass range
spectrum in the lower cell.
2. Pin the chromatogram cell (see “Active and Pinned Cells” on page 225).
3. Right-click the chromatogram cell.
4. From the shortcut menu, choose Ranges.
The Chromatogram Ranges dialog box appears (see Figure 141).
Figure 141. Chromatogram Ranges dialog box with the default layout settings
5. In the Range area, set the displayed time range of the chromatogram:
• For a generic layout file, leave an asterisk in the Time Range box.
When you open a .raw data file, the x axis of the chromatogram is scaled to the run
time for your detector in the instrument setup method used to acquire the data.
• To specify a specific time range, type a beginning time point and an ending time
point separated by a dash in the Time Range (minutes) box.
6. In the Plot Properties area (see Figure 142), make the following selections:
a. In the Detector list, select PDA.
b. In the Peak Algorithm list, select Avalon.
c. In the Plot Type list, select a plot type:
• Select Wavelength to display the chromatogram for a specific wavelength within
your scan range or to display the averaged results from a range of wavelengths in
your scan range.
• Select Total Scan to display the average absorbance for each time point of all the
wavelengths in your scan range.
• Select Spectrum Maximum to display a plot of the maximum absorbance values
in your scan range for each time point.
d. Select a wavelength range (if you selected the Total Scan plot type, this box is
unavailable):
• To display the chromatogram for a specific scan wavelength, type the wavelength
number in the Range(s) box.
• To display a plot of the average absorbance values for a range of wavelengths, type
the beginning wavelength number and the ending wavelength number separated
by a dash in the Range(s) box. For example, type 200-300 to display a plot of the
average absorbance values for the scanned wavelengths from 200 to 300 nm.
Figure 142. Plot Properties area of the Chromatogram Ranges dialog box
7. Click OK to close the dialog box and view your scan chromatogram.
Figure 143 shows a chromatogram for a wavelength in the PDA scan.
To modify the layout for the spectrum cell, go to “Specifying the Spectrum Range” on
page 238.
Figure 143. Chromatogram view with a chromatogram for a PDA scan wavelength
220 nm
scan wavelength
1. If you have not already done so, open a data file (.raw) that has PDA data and MS data
(see “Opening a Raw Data File in Qual Browser” on page 219).
The data file opens with the MS TIC chromatogram in the upper cell and a mass range
spectrum in the lower cell.
2. Pin the chromatogram cell (see “Active and Pinned Cells” on page 225).
3. Right-click the chromatogram cell.
4. From the shortcut menu, choose Ranges.
The Chromatogram Ranges dialog box appears. Figure 141 on page 234 shows the
default settings. Figure 144 shows the settings for a discrete channel wavelength.
Figure 144. Chromatogram Ranges dialog box with the selection of a discrete channel wavelength
5. In the Range area, set the displayed time range of the chromatogram:
• For a generic layout file, leave an asterisk in the Time Range box.
When you open a .raw data file, the x axis of the chromatogram is scaled to the run
time for your detector in the instrument setup method used to acquire the data.
• To specify a specific time range, type a beginning time point and an ending time
point separated by a dash in the Time Range box.
6. In the Plot Properties area, select the following:
a. In the Detector list, select UV.
b. In the Peak Algorithm list, select Avalon.
c. In the Plot Type list, select Channel A, Channel B, or Channel C.
7. Click OK to exit the dialog box and view your discrete chromatogram.
Figure 145 shows a discrete channel chromatogram.
To modify the layout for the spectrum cell, go to “Specifying the Spectrum Range” on
page 238.
Figure 145. Chromatogram view with a chromatogram for a PDA discrete channel
Discrete Channel B
wavelength
1. If you have not already done so, open a data file (.raw) that has PDA data and MS data
(see “Opening a Raw Data File in Qual Browser” on page 219).
The data file opens with the MS TIC chromatogram in the upper cell and a mass range
spectrum in the lower cell.
2. Pin the spectrum cell.
3. Right-click the spectrum cell.
4. From the shortcut menu, choose Ranges.
The Spectrum Ranges dialog box appears (see Figure 146).
5. In the Plot Properties area, select PDA in the Detector list.
6. For a generic layout file for PDA data, keep the other settings in the Spectrum Ranges
dialog box at their defaults as shown in Figure 146.
Figure 146. Spectrum Ranges dialog box with the PDA detector selected
If you have not already done so, modify the layout for the chromatogram cell as described in
“Specifying the Chromatogram Ranges” on page 233.
If you have specified the ranges for the chromatogram cell and the spectrum cell, go to the
next procedure, “Specifying the Display Options for the Chromatogram and Spectrum Cells.”
Specifying the Display Options for the Chromatogram and Spectrum Cells
After you specify the chromatogram and spectrum ranges, specify the display options for the
chromatogram and spectrum cells as described in these topics:
• Setting the Display Options for the Chromatogram Cell
• Setting the Display Options for the Spectrum Cell
b. Select the check boxes associated with the labels that you want to display, such as
retention time, area, height, and name.
7. Click OK to accept the settings and close the Display Options dialog box.
Inserting Cells
Occasionally, you might want to add more cells to the Qual Browser window. For example,
you might want to add a cell containing a map plot (contour or 3D) to the view screen, or you
might want to display several discrete or scan wavelengths in separate cells.
Chromatogram cell
Spectrum cell
Spectrum cell
3. Change the lower cell so that it displays the Map view (see Figure 153) as follows:
a. Pin the cell by clicking its pin button.
b. Click the View Map button in the toolbar.
Figure 153. Qual Browser window with a chromatogram cell, spectrum cell, and map cell
4. Select the appropriate range options for the map cell as follows:
a. Right-click the map cell.
b. From the shortcut menu, choose Ranges.
The Map Ranges dialog box appears (see Figure 154).
Figure 154. Map Ranges dialog box
c. In the Wavelength box, type the wavelength range that you want to display.
d. In the Time box, type the time range that you want to display.
5. Select the appropriate display options for the map cell as follows:
a. Right-click the map cell.
b. From the shortcut menu, choose Display Options.
The Display Options dialog box appears.
c. Click the Axis tab.
The Axis page appears (see Figure 151 on page 244).
d. In the Units area, select the Absolute option.
e. Click the Normalization tab.
The Normalization page appears.
f. In the Normalize Method area, select the Auto Range option.
g. Click the Style tab.
The Style page appears (see Figure 155).
Figure 155. Style page in the Display Options dialog box for the map cell
1. From the Qual Browser window, choose File > Layout > Save As.
The Save Layout File dialog box appears (see Figure 156).
Figure 156. Save Layout File dialog box
1. Open a raw data file with PDA data and apply an appropriate layout file (see “Opening a
Raw Data File in Qual Browser” on page 219).
2. Pin the spectrum cell.
3. Click a time point in the chromatogram cell.
The spectrum for the selected time point appears in the spectrum cell (see Figure 157).
Figure 157. Qual Browser window with a chromatogram of the PDA test mixture and a
spectrum for the 7.42 minute time point
1. Open a raw data file with PDA data and apply an appropriate layout file (see “Opening a
Raw Data File in Qual Browser” on page 219).
2. Pin the chromatogram cell.
3. Click a wavelength in the spectrum cell.
The scan chromatogram for the selected wavelength appears in the chromatogram cell as
shown in Figure 158.
Figure 158. Qual Browser window, displaying a chromatogram cell and a spectrum cell
334 nm
scan wavelength
Note The Avalon Peak Detection Settings page appears only when Avalon is
selected as the peak algorithm in the Chromatogram Ranges dialog box.
b. Click Auto Calc Initial Events to force the Avalon peak integration algorithm to
determine the “best” values for the following initial events: Start Threshold,
End Threshold, Peak Threshold, P-P Threshold, Bunch Factor, Negative Peaks, and
Tension.
Figure 161 shows a chromatogram for the 239 nm scan wavelength that is integrated
with the “best” initial integration values as determined by the Avalon peak integration
algorithm. The peaks areas are shown above the apexes of the integrated peaks.
Figure 161. Avalon Peak Detection Settings page displaying a chromatogram acquired
with the PDA detector
Figure 163. Comparison of peak purity results for two different scan wavelengths
Figure 164. Chromatogram for the 225 nm scan wavelength with default integration parameters
Benzene
Impurity
Figure 165. Chromatogram for the 225 nm scan wavelength with Auto calc integration parameters
Contents
• Calibrating the Surveyor LC Pump Plus
• Recording Maintenance Information for the Surveyor LC Pump Plus
• Firmware Version Information for the Surveyor LC Pump Plus
The Surveyor LC Pump Plus is factory calibrated to produce an accurate flow rate while
pumping water at 1 mL/min. Its ability to accurately proportion binary mobile phases is also
tested at a flow rate of 1 mL/min.
The Surveyor LC Pump Plus does not require calibration upon arrival at its shipping
destination. However, over time, the pressure readings from its built-in pressure transducer
can drift and require re-zeroing. If you are running the LC pump at high backpressures, you
might notice an increase in its pressure pulsation. You can reduce this pressure pulsation by
adjusting the compressibility setting of the LC pump.
If you are pumping solvents other than water and your application is sensitive to pressure
pulsation, you might want to optimize the compressibility setting. After you optimize the
compressibility setting, check the flow rate accuracy. If adjusting the compressibility setting
has affected the flow rate accuracy, adjust the alpha setting.
In addition to the alpha setting and the compressibility setting, the pressure transducer that is
attached to the back of the purge manifold assembly is also factory calibrated. If you replace
the pressure transducer, you must update the pressure sensor adjustment setting. Even if you
never replace the pressure sensor, you must occasionally re-zero its output, as the readout
tends to drift under normal usage.
Default value
For information on determining the appropriate value for your solvent conditions, see
“Updating and Optimizing the Compressibility Setting” on page 266.
Pressure Recorder Full Specifies the full-scale range of the recorder connected to the pressure terminals on the back
Scale panel of the pump (see “Pressure Recorder Full Scale” on page 264).
For information on setting the full-scale range for the pressure recorder, see “Scaling the
Output of the Pressure Recorder Terminals” on page 268
Flow Rate Adjustment Makes fine adjustments to the flow rate (see “Flow Rate Adjustment (Alpha)” on page 265).
(Alpha)
For information on calibrating the alpha value, see “Updating the Flow Rate Adjustment
Setting” on page 268.
Table 56. Calibration parameters for the Surveyor LC Pump Plus (Sheet 2 of 2)
Parameter Description
Pressure Sensor Adjusts the pressure sensor reading (see “Pressure Sensor Adjustment” on page 265).
Adjustment
For information on adjusting the pressure sensor reading, see “Updating the Pressure Sensor
Adjustment Setting” on page 269.
Set Downloads the new setting to the pump.
Pressure Transducer
Zero Calibrates the current pressure reading as the zero point (see “Pressure Transducer Zero” on
page 265).
For information on setting the zero point for the pressure transducer, see “Zeroing the
Pressure Transducer” on page 269.
Compressibility
The compressibility of a liquid is a measure of its resistance to a decrease in volume caused by
an increase in pressure. Compared to gases, most liquids are relatively incompressible. Water, a
commonly used mobile phase solvent, is even less compressible than most organic solvents
because of its extensive hydrogen bonding and cluster structure.
For information on updating the compressibility setting, see “Updating and Optimizing the
Compressibility Setting” on page 266.
The compressibility effect is caused by the portion of the cam cycle where the pistons are
compressing the mobile phase. The compressibility portion of the cam cycle extends from the
time when the primary piston starts discharging until the time the secondary piston reaches
full intake—approximately 80° of the cam cycle. To minimize the pressure pulsation and
maintain a constant flow rate, the on-board CPU compensates for this compression by
making fine adjustments to the speed of the stepping motor. If the compressibility setting is
incorrect, the LC pump cannot effectively minimize the pressure pulsation.
Figure 169 shows the overlaid pressure traces for three runs of water pumped at a nominal
flow rate of 3 mL/min. The compressibility settings for the three runs were 0.45 GPa–1,
1.25 GPa–1, and 3.0 GPa–1, as labeled. The compressibility of water is 0.45 GPa–1.
Figure 169. Overlaid LC pump pressure traces, showing the effect of the compressibility setting
3.0 GPa-1
1.25 GPa-1
0.45 GPa-1
Table 57 lists the calculated pressure pulsation and the empirically determined flow rates for
the three different compressibility settings. The pressure pulsation as well as the actual flow
rate increased as the compressibility setting increased.
Table 57. Measured flow rate and pressure pulsation versus compressibility setting
Compressibility % Pressure Actual flow rate
% Rel diff
setting pulsation (mL/min)
0.45 GPa–1 ± 0.2% 3.01 + 0.3%
1.25 GPa–1 ± 0.4% 3.05 + 1.8%
3.0 GPa–1 ± 1% 3.18 + 5.9%
For example:
When the set value = 1, the full scale is 4.9 MPa (50 kgf/cm2)
When the set value = 10, the full scale is 49.0 MPa (500 kgf/cm2)
4.9 MPa = 50 kgf/cm2 = 49 bar = 711 psi
Note The value for the scaling factor is stored until you change it or until you turn off the
power to the Surveyor LC Pump.
Figure 170. Terminals located on the back of the Surveyor LC Pump Plus
Pressure terminals
UNIT ID ENET
+5V @ 150mA
NC
PROG. START
PUMP READY
INJECT HOLD
GND
PUMP ON
GND
NC
FW DOWNLOAD
NC
NC
GND
+5V @ 15OmA
PRESSURE+
PRESSURE -
For information on setting the scaling factor for the pressure recorder, see “Scaling the Output
of the Pressure Recorder Terminals” on page 268.
For information on setting the alpha value, see “Updating the Flow Rate Adjustment Setting”
on page 268.
For information on updating the pressure sensor adjustment, see “Updating the Pressure
Sensor Adjustment Setting” on page 269.
For instructions on opening the Calibration page, see “Accessing the Calibration Options for
the Surveyor LC Pump Plus” on page 260.
To fine-tune the performance of the Surveyor LC Pump Plus, follow these procedures:
• Updating and Optimizing the Compressibility Setting
• Scaling the Output of the Pressure Recorder Terminals
• Updating the Flow Rate Adjustment Setting
• Updating the Pressure Sensor Adjustment Setting
• Zeroing the Pressure Transducer
1. Open the Calibration page for the Surveyor LC Pump (see “Accessing the Calibration
Options for the Surveyor LC Pump Plus” on page 260).
2. Type a new value into the Compressibility setting box. Then click Set.
c. In the Method area, enter the specified flow rate in the Flow Rate box and enter the
specified mobile phase composition in the Component A through Component D
boxes.
d. Click Download to send these parameters to the onboard CPU of the Surveyor LC
Pump Plus.
e. Click Pump On to turn on the pump flow.
3. Record the pressure trace.
4. If the pressure pulsation is too high for your application, change the compressibility
setting. Then, repeat steps 2 and 3. Continue to change the compressibility setting until
the level of pump pulsation is adequate for your application.
5. After you change the compressibility setting to minimize the pressure pulsation, check the
flow rate. If necessary, adjust the flow rate adjustment setting as described in “Updating
the Flow Rate Adjustment Setting” on page 268.
1. Open the Calibration page for the Surveyor LC Pump (see “Accessing the Calibration
Options for the Surveyor LC Pump Plus” on page 260).
2. Enter a value from 1 to 10 in the Pressure Recorder Full Scale box (see Figure 168 on
page 261). Then click Set. For information on scaling the output of the pressure recorder
terminals, see “Pressure Recorder Full Scale” on page 264.
1. Open the Calibration page for the Surveyor LC Pump (see “Accessing the Calibration
Options for the Surveyor LC Pump Plus” on page 260).
2. In the Flow Rate Adjustment (alpha) box (see Figure 168 on page 261), type a new value:
• If the actual flow rate is below the set value, increase the alpha value. If you are
pumping water, increasing the alpha value by 1increases the flow rate by 1%.
For example, if the pump is set to deliver a flow rate of 1 mL/min and the actual flow
rate is 0.96 mL/min, which is 4 percent below the set rate, raise the current setting
by 4. If the current setting is 5.0, enter a new setting of 9.0.
• If the actual flow rate is above the set value, decrease the alpha value. If you are
pumping water, decreasing the alpha value by 1 decreases the flow rate by 1%.
For example, if the pump is set to deliver a flow rate of 1 mL/min and the actual flow
rate is 1.04 mL/min, which is 4 percent above the set rate, lower the current setting
by 4. If the current setting is 5.0, enter a new setting of 1.0.
3. Click Set.
1. Open the Calibration page for the Surveyor LC Pump (see “Accessing the Calibration
Options for the Surveyor LC Pump Plus” on page 260).
2. In the Pressure Sensor Adjustment box (see Figure 168 on page 261), type the value on
the data sticker “0.XXXX” multiplied by 10 000.
3. Click Set.
DRAIN
SE
O
PE
LO
N
C
C
N
LO
PE
SE
DRAIN
O
3. Open the Calibration page of the Direct Control dialog box by choosing
Surveyor LC Pump > Direct Control > Calibration.
4. Click Zero in the Pressure Transducer area (see Figure 168 on page 261).
5. When you are finished zeroing the pressure transducer, close the drain valve by gently
turning the knob clockwise until you feel resistance.
Note Applying excessive force to the drain valve knob reduces the life span of its
O-ring.
Each pump head assembly contains a piston seal. The two seals can wear at different rates,
causing leaks to occur in one or both pump heads. Before dismantling the pump head
assemblies, attempt to determine the source of the leakage. To save time and materials, replace
only the piston seal in the pump head assembly that is leaking. Or, as a preventive measure,
replace the piston seals after pumping 120 L of mobile phase.
From the menu bar of the Surveyor LC Pump Plus view, choose Surveyor LC Pump >
Direct Control > Maintenance.
The Maintenance - Surveyor LC Pump Direct Control page appears (see Figure 173).
Figure 173. Direct Control - Maintenance dialog box
To set the pistons to the maintenance position before removing the pump heads
Every time you install a new piston seal in the left pump head, click Reset in the Left Seal
Volume area to set the recorded volume of liquid pumped by the left pump head assembly
back to zero.
Right Seal Volume
Reset Zeros the recorded volume of mobile phase pumped by the right pump head.
Every time you install a new piston seal in the right pump head, click Reset in the Right
Seal Volume area to set the recorded volume of liquid pumped by the right pump head
assembly back to zero.
Pump Operating Time
Reset Resets the recorded operating time for the Surveyor LC Pump Plus to zero.
If you install a new motor, click Reset in the Pump Operating Time area to set the recorded
Pump Operating Time back to zero.
Position Pistons
Maintenance Position Sets the pistons to the maintenance position.
Set the pump pistons to the maintenance position before removing the pump heads.
Contents
• Calibrating the Autosampler
• Autosampler Maintenance Information
• Autosampler Validation Information
The autosampler does not require calibration upon arrival at its shipping destination.
However, if you use custom vials or custom microplates, you must perform the Well Bottom
Distance calibration, which determines the actual depth of a vial or microplate well (see “Well
Bottom Distance Calibration” on page 305). If problems occur with the column oven control,
the tray temperature control, or the arm positioning, contact a Thermo Fisher Scientific
service representative. This appendix includes calibration procedures for these items; however,
to calibrate the autosampler’s temperature zones and XYZ arm position, you must have the kit
that contains the necessary test fixtures.
The column oven and oven compartment temperature calibrations and the arm calibration
are typically performed by a Thermo Fisher Scientific field service representative. To perform
these calibrations yourself, you must have the service kit that contains the temperature test
fixtures. In addition, to calibrate the temperature sensors for the autosampler’s
temperature-controlled zones, your Thermo Scientific application must be capable of
providing feedback for these sensors.
Note Typically, a Thermo Fisher Scientific field service representative calibrates the
controlled temperature zones and the XYZ arm position. To perform the temperature
calibrations yourself, order the Field Service Calibration Kit (P/N 60053-62001). To
calibrate the XYZ arm, order the LED light fixture and the non-reflective target port
fixture.
If you choose to use custom vials or custom microwell plates, you must perform the
Well Bottom Distance calibration, which determines the depth of the custom vial or
microplate well. Perform this calibration you select a new custom tray configuration and each
time you use a new type of custom vial or custom microwell plate.
Note If you are controlling the autosampler from a Thermo Scientific data system other
than the Xcalibur data system, refer to the Help provided with your data system for
information about accessing the Calibration dialog box for the autosampler.
CAUTION Because the column oven compartment can reach temperatures as high as
95 °C (203 °F), ensure that the column oven is off and that the column oven
compartment is at room temperature before you install the oven sensor test fixture.
Specifying the Internal Set Target for the Column Oven Calibration Temperature
Use the Internal Set Target page of the Column Oven Air Sensor Calibration wizard to specify
the calibration temperature.
1. Set up the system to perform the column oven calibration (see “Preparing to Calibrate the
Column Oven Temperature” on page 275).
2. Select the Column Oven Air Sensor Calibration option, and then click Next.
The Diagnostic - Internal Set Target page of the Column Oven Air Sensor Calibration
wizard appears (see Figure 176). The default calibration temperature is 50.0 °C.
Figure 176. Diagnostic - Internal Set Target page of the Column Oven Air Sensor Calibration
wizard with the default temperature setting of 50.0 °C
3. In the Calibration Target Temperature (C) box, type 30 (see Figure 177).
Figure 177. Diagnostic - Internal Set Target page of the Column Oven Air Sensor Calibration
wizard with the user setting of 30 °C
The Diagnostic - External Standard page of the Column Oven Air Sensor Calibration
wizard appears (see Figure 178). The default calibration temperature for the external
standard is 50 °C.
Figure 178. Oven page and the Diagnostic - External Standard page of the Column Oven Air
Sensor Calibration wizard
1. Specify the calibration temperature and start the Column Oven Air Sensor Calibration
wizard (see “Specifying the Internal Set Target for the Column Oven Calibration
Temperature” on page 276).
2. Wait for the current temperature readback on the Oven page to reach 30.0 °C
(see Figure 179).
The Current temperature readback displays the temperature recorded by the
autosampler’s internal temperature sensor.
Figure 179. Oven page, showing that the current temperature has reached the set temperature
3. When the current temperature readback on the Oven page (autosampler view on the
Status page of the Information view) reaches exactly 30.0 °C (see Figure 179), do the
following:
a. Type the reading from the 869C thermometer in the External Standard
Temperature (C) box (see Figure 180).
For example, if the thermometer reading is 29.8 °C, type 29.8 in the External
Standard Temperature (C) box. The range is 0 to 110.0 °C.
Figure 180. Diagnostic - External Standard page of the Column Oven Air Sensor
Calibration wizard
The data system downloads the temperature calibration to the autosampler and the
Finish button becomes available (see Figure 181).
Figure 181. Diagnostic - External Standard page of the Column Oven Air Sensor
Calibration wizard
Finish button
d. Repeat step 3a through step 3c until the readings from the 869C thermometer and
the current temperature readout agree within ±0.2 °C.
4. When the temperatures on the 869C thermometer and the Current Temperature
readback are within ±0.2 °C of each other, click Finish (see Figure 181).
5. On the Oven page, verify that the Current Temperature readback is stable at 30.0 °C with
a maximum temperature drift of ±0.2 °C.
6. After you verify the stability of the column oven temperature, turn off the column oven
using the autosampler direct controls and remove the temperature probe.
For information about the autosampler direct controls, see “Controlling the Tray and
Oven Compartment Temperatures” on page 199.
To perform the vial tray metal sensor calibration procedure, you must have the following
items provided in the Field Service Calibration Kit (P/N 60053-62001):
• Calibrated Omega 869C RTD thermometer or equivalent
• A/S vial tray sensor
4. Open the Calibration dialog box for the autosampler (see “Calibrating the Autosampler”
on page 274).
5. Select the Vial Tray Metal Sensor Calibration option (see Figure 183).
Figure 183. Calibration dialog box
6. Click Next.
The Diagnostic - Internal Set Target page of the Vial Tray Metal Sensor Calibration
wizard appears (see Figure 184).
The default external standard temperature is 40.0 °C.
Specifying the Internal Set Target for the Tray Compartment Temperature
Use the Diagnostic – Internal Set Target page (see Figure 184) of the Vial Tray Metal Sensor
Calibration wizard to specify the calibration temperature.
Figure 184. Diagnostic - Internal Set Target page of the Vial Tray Metal Sensor Calibration wizard
Default temperature
To specify the target calibration temperature for the tray compartment and start the
calibration
1. Prepare the system for the temperature calibration (see “Preparing to Calibrate the Tray
Compartment Temperature” on page 281).
2. In the Calibration Target Temperature box, type 30.0, and then click Start Calibration
(see Figure 185).
Figure 185. Diagnostic -Internal Set Target page of the Vial Tray Metal Sensor Calibration
wizard
User setting
The data system turns on the tray temperature control and downloads the set temperature
of 30.0 °C. The Next button becomes available.
3. Click Next.
The Diagnostic - External Standard page appears (see Figure 186). The default external
standard temperature is 40.0 °C. The range is 0 to 110.0 °C.
Figure 186. Tray page and the Diagnostic - External Standard page of the Vial Tray Metal
Sensor Calibration wizard
1. When the current temperature readback on the Tray page reaches exactly 30.0 °C
(see Figure 187), do the following:
a. In the External Standard Temperature (C) box (see Figure 187) on the External Set
Target page, type the reading from the thermometer.
Figure 187. Tray page (status view) and the Diagnostic - External Standard page of the Vial Tray
Metal Sensor Calibration wizard
b. On the Tray page, verify that the Current Temperature readback still displays
30.0 °C.
c. On the Diagnostic - External Standard page (see Figure 187), click Adjust.
The data system downloads the calibration to the autosampler and the Finish button
becomes available (see Figure 188).
Figure 188. Diagnostic - External Standard page of the Vial Tray Metal Sensor Calibration
wizard with an available Finish button
d. Repeat step 1a through step 1c until the temperatures on the thermometer and the
current temperature readout (see Figure 187) agree within ± 0.2 °C.
2. When the temperatures on the 869C thermometer and the Current Temperature readout
are within ±0.2 °C of each other, click Finish at the bottom of the Diagnostic - External
Set Target page.
3. On the Tray status page, verify that the Current Temperature readout is stable at 30.0 °C
with a maximum temperature drift of ± 0.2 °C.
4. After you have verified the stability of the oven tray temperature, remove the temperature
probe.
Arm Calibration
During an injection, the XYZ arm moves between the specified vial or well location and the
injection port. When you move the autosampler, you can jar the XYZ arm, causing it to lose
alignment.
Performing the arm calibration procedure requires you to visually align the XYZ arm over the
injection port. To make this visual alignment, you must replace the needle with an LED light
fixture that shines a small beam of red light onto the port and replace the stainless steel
injection port with a non-reflective target fixture. These test fixtures do not come with the
autosampler; you must order them from Thermo Fisher Scientific.
After following the instructions in the Arm Calibration wizard to adjust the XYZ arm
alignment, test the alignment by making a set of injections from one or two 384-well
microplates. The microplate carrier and the 384-well microwell plates are provided in the
Autosampler Accessory Kit.
To calibrate the XY position of the XYZ arm, you must have these items:
• LED light fixture (P/N 60357-60021)
• Target port fixture (P/N 60357-20021)
• Microplate carrier
• One or two 384-well microtitre plates
• Masking tape
Figure 189 shows the LED light and the target port fixtures.
Figure 189. Arm calibration fixtures available by special order
To calibrate XY position of the XYZ arm and verify the calibration, follow these procedures:
1. Modifying the Autosampler Instrument Configuration
2. Starting the Arm Calibration Wizard
3. Moving the Arm to the Needle Removal Position
4. Installing the LED Light Fixture and the Target Port Fixture
5. Moving the XYZ Arm to the Home Position
6. Aligning the Light Beam with the Target
7. Applying the Arm Calibration Setting
8. Checking the Arm Calibration Offset Values
9. Testing the Alignment of the XYZ Arm
Figure 190. Tray page with the 384-well microplate tray type selected
Figure 191. Communication page with the Verify Door Is Closed check box cleared
When the Verify Door Is Closed check box is selected, the data system prevents you from
starting an injection sequence or using some of the direct control commands for the
autosampler when the tray compartment door is open.
6. Click OK at the bottom of the dialog box to accept the settings and close the dialog box.
7. Click Done to close the Thermo Foundation Instrument Configuration window.
For information about the LED light and non-reflective needle port fixtures that you must
have to perform the Arm Calibration wizard, see “Arm Calibration” on page 287.
1. If you have not already done so, modify the autosampler instrument configuration as
described in “Modifying the Autosampler Instrument Configuration” on page 288.
2. Open the view for the autosampler.
3. From the Surveyor AS menu, choose Calibration.
The Calibration dialog box appears (see Figure 192).
4. Select the Arm Calibration option.
Figure 192. Calibration dialog box with the Arm Calibration option selected
Note The user interface specifies the x- and y-axis offsets in microns, which is another
term for micrometers.
The diameter of the hole in the center of the target port fixture is 0.047 inches.
Go to the next procedure, “Moving the Arm to the Needle Removal Position.”
Go to the next procedure, “Installing the LED Light Fixture and the Target Port Fixture.”
Installing the LED Light Fixture and the Target Port Fixture
This page of the Arm Calibration wizard instructs you to remove the needle and install the
calibration light source (see Figure 195).
Figure 195. Request to install the LED light fixture page
The needle assembly is a welded piece of 0.012 in. ID stainless steel tubing with an externally
threaded fitting, a needle flag, a latch, and a compression spring (see Figure 196). It slides into
the needle mount on the XYZ arm and is secured with the latch. The needle tubing assembly
has an internally threaded fitting that connects to the 10-32 externally threaded fitting of the
needle assembly.
Figure 196. Needle assembly
Latch
10-32 fitting
Compression
spring
Needle flag
Needle mount on
the XYZ arm
Resistor
Figure 198. Inserting the LED holder into the needle mount
LED holder
Needle mount
d. Push the LED holder down until it meets resistance from the needle mount.
e. Make sure that the light fixture cable is routed so that it does not obstruct the
movement of the XYZ arm.
3. Install the target port fixture as follows:
a. Remove the aluminum retention clip (see Figure 199) from the wash station housing,
pull it forward, and then upward.
Figure 199. Retention clip for the autosampler injection port
Retention clip
b. Pull the injection port out from the wash station (see Figure 200).
Figure 200. Autosampler injection port connected to the transfer tube nut
Injection port
4. Insert the target port fixture (see Figure 189 on page 287) into the position previously
held by the injection port.
5. Click Next to proceed to the next page of the wizard.
Go to the next procedure, “Moving the XYZ Arm to the Home Position.”
Go to the next procedure, “Aligning the Light Beam with the Target.”
Two lead screws and two stepper motors control the x-axis and y-axis position of the XYZ arm
(see Figure 203).
The Left and Right buttons move the XYZ arm along the x axis. The Forward and Backward
buttons move the XYZ arm along the y axis.
Figure 202. Alignment buttons page
x-axis
y-axis
lead screw
lead screw
y axis
x axis
To align the light beam over the hole in the target port fixture
1. In the Step list of the Arm Calibration wizard, select 25x (793.75 microns).
This selection moves the arm 793.75 micrometers (0.03125 inches) each time you click
Left, Right, Forward, or Backward.
1 micrometer = 3.937 × 10–5 inches
2. Click Left, Right, Forward, and Backward to align the position of the light beam with
the hole in the center of the target port fixture.
3. In the Step list, select a smaller step size, and continue aligning the position of the
light beam.
4. After you align the light beam with the hole in the center of the target port fixture,
click Next to proceed to the next page of the wizard.
1. Ensure that the light beam is centered over the hole in the target port fixture
(see Figure 205).
Figure 205. LED light beam centered over the hole in the target port fixture
The diameter of the hole in the center of the target port fixture is 0.047 inches.
2. Click Finish to accept the settings and close the Arm Calibration wizard.
3. Remove the LED and target port calibration fixtures.
4. Reinstall the injection port as follows:
a. Insert the injection port into the port in front of the wash station.
b. Align the retention clip with the slot located in front of the wash station.
c. Using the retention clip to push the injection port down, align the circular cutout in
the retention clip over the alignment screw.
d. Insert the end of the retention clip into the wash station slot as far as it will go.
Figure 207 shows the installation of the retention clip.
Injection port
Alignment screw
Retention clip
installed
Leave the autosampler view of the Instrument Setup view open and go to the next procedure,
“Testing the Alignment of the XYZ Arm.”
1. Cover columns 1, 2, 23, and 24 of two 384-well microplates with tape (see Figure 208).
Figure 208. 384-well microplate positions
A1 A24
A
P P1 P24
1 24
2. Place the microplates in each end of the microwell carrier, leaving the middle position
empty.
3. Load the microwell carrier into the tray compartment of the autosampler as follows:
a. From the menu bar, choose Surveyor AS > Direct Control.
The Direct Control dialog box appears (see Figure 209).
Use the Well Bottom Distance wizard to determine the distance that the needle must travel to
reach the bottom of a vial or well. The XYZ arm uses this value when you select one of the
custom tray configurations.
IMPORTANT Because the autosampler stores only one value for the custom well bottom
distance, you must perform a well bottom distance calibration each time you select a new
custom tray type configuration and each time you use a different type of custom vial or
custom microwell plate.
Figure 211. Calibration dialog box with the Well Bottom Distance option selected
5. Select the Well Bottom Distance option, and then click Next.
The Select Calibration Method page appears (see Figure 212).
Figure 212. Select Calibration Method page
Go to the next procedure, “Selecting the Calibration Method for the Well Bottom Distance”
on page 306.
Note The autosampler stores only one well bottom distance value for custom tray
configurations. The number at the top of this dialog box is the current value for
the distance.
1. In the Enter New Value (microns) box, type the new value in micrometers
(see Figure 213).
The range is 15 000 to 46 990 micrometers (15 to 46.9 mm).
Note The user interface specifies the bottom distance in microns, which is another
term for micrometers.
To determine the well bottom distance with the autosampler’s needle sensor
IMPORTANT Remove vial caps or well plate lids before performing an active well
bottom distance calibration. As the needle pierces a vial cap or a well plate lid, the
spring in the needle mechanism is compressed, which can cause premature activation
of the needle sensor.
IMPORTANT Before placing a custom tray into the autosampler tray compartment,
check the height limitations for 96-well and 384-well plates listed in the hardware
manual for your autosampler. Tall objects will stall the autosampler arm.
2. Place the vial into a tray or the well plate into a carrier, and then place the tray or carrier
into the tray compartment of the autosampler.
3. In the Calibrate at Well/Vial position box (see Figure 214), type a vial or well location.
Note If you are calibrating the bottom distance of a vial, verify that you have placed a
vial in the selected location.
1. Click the tab for the page that contains the maintenance item that you want to change the
usage limit for.
2. In the Usage Limit box, type the usage limit.
3. Click Set New Limit.
To reset the usage counter after performing the scheduled maintenance task
Note If you selected the Enable Maintenance Log check box on the Communication page
of the Surveyor AS Autosampler dialog box when you configured the Surveyor
Autosampler, you can set the scheduled maintenance times (see “Communication Page”
on page 52).
When anyone of these counters on the Maintenance Information pages exceeds the
user-specified usage limit, the Maintenance Due readback displays Yes and the data system
prevents the autosampler from making injections.
To enable the autosampler, you must zero the counter or change the usage limit. If the usage
limit is based on a scheduled maintenance plan, perform the scheduled maintenance.
Table 61. Maintenance Information dialog box pages and parameters
Page or Parameter Description
Page
Injections Use this page to set the scheduled maintenance time for the number of user-defined
injections (excluding any cancelled injections).
Valve Cycles Use this page to set the scheduled maintenance time for the number of user-defined valve
cycles for the injection valve (from Fill to Inject).
Needle Cycles Use this page to set the scheduled maintenance time (SMT) for the number of user-defined
needle cycles when the needle is inserted into a septum, injection port, or wash station.
Syringe Cycles Use this page to set the scheduled maintenance time (SMT) for the number of user-defined
syringe cycles where the syringe is put in the Ready position.
Parameter
Usage Limit Use this box to specify the new usage limit for the selected maintenance item.
Note When a counter exceeds the usage limit, the autosampler cannot start a run until
you reset the counter to zero, change the usage limit, or clear the Enable Maintenance Log
check box on the Communication page of the Surveyor Autosampler Configuration
dialog box (see “Communication Page” on page 52).
Set New Limit Downloads the new usage limit to the autosampler.
Current Counter and Displays the current count. When the counter exceeds the usage limit (scheduled
Zero Counter maintenance time), the autosampler cannot start a run until you perform the scheduled
maintenance or clear the Enable Maintenance Log check box on the Communication page
of the Autosampler Configuration dialog box.
Zero Counter Zeros the counter for the selected maintenance item.
Contents
• Verifying the Performance of the PDA Detector
• Calibrating the PDA Detector
The PDA detector is factory tested for linearity, noise, and drift. Because of the sensitivity of
its optical bench, adjust the light throughput to the diode array and recalibrate the PDA
detector after you install it, and each time you move it, change its flow cell, or replace either of
the lamps.
To adjust the light throughput to the diode array, follow these procedures:
• Creating a Display Method
• Adjusting the Light Throughput to the Diode Array
1. In the view bar of the Instrument Setup window, click the Surveyor PDA Plus button.
The view for the PDA detector appears.
2. In the Units area, select the Diode/Intensity option (see Figure 217).
Figure 217. Surveyor PDA Plus Method page with the Diode/Intensity option selected
3. Verify that the following parameters are specified on the Surveyor PDA Method page in
the Spectra area:
• Start Diode (diode num) = 2
• End Diode (diode num) = 511
• Diode Step (diode num) = 1
Note For a display method, the following parameters are not downloaded to the PDA
detector: run length, filter rise time, and sample rate.
5. Click OK.
6. From the menu bar, choose Surveyor PDA > Direct Control.
7. Click Yes.
The Save PDA Display Method dialog box appears.
8. Save the display method.
The data system saves the method with a .spda file extension.
Note The .spda file extension is a special file extension used for all method files based
on Diode/Intensity units. When you create a file with an .spda file extension, you can
only load and use it in the Direct Control dialog box.
After you save the method, the Direct Control dialog box appears with the Display page
open.
9. To adjust light throughput to the diode array, go to “Adjusting the Light Throughput to
the Diode Array.”
Adjust the light throughput to the diode array whenever you do the following:
• Observe an increase in the detector noise level.
• Move the detector.
• Replace either lamp or the flowcell.
The first time that you adjust the attenuators, you must create a display method that records
light intensities. You can identify display methods by their .spda file extension. After creating
the method for adjusting the attenuators, save it with a descriptive name, such as
diagnostics.spda, and store it for future use to simplify future adjustments of the attenuators.
Note Before you adjust the attenuators, replace the column with a flow restrictor, and set
the pump to deliver HPLC-grade water at a flow rate of 1 mL/min through the flowcell.
To adjust the light output from the lamps, follow these procedures in order:
1. “Preparing the LC System for an Attenuator Adjustment” on page 316
2. “Adjusting the Attenuators” on page 317
3. “Adjusting the Attenuators” on page 317
Tungsten lamp
attenuator tab
1. If you have not already done so, create a display method to view the light intensity from
the lamps (see “Creating a Display Method” on page 313).
2. Load the display method to the detector as follows:
a. In the Direct Control dialog box for the PDA detector, click the Display tab.
The Display page appears.
b. In the Control area, click Load Method.
The Open Display Method dialog box appears (see Figure 220).
Figure 220. Open Display Method dialog box for the PDA detector
Figure 221. Display page for the PDA detector with an intensity spectrum
For a wavelength/absorbance method, the file contains the absorbance values for the
spectral scan and the discrete channel wavelengths at the moment that you clicked
Snapshot.
For an intensity/diode method, the file contains the intensity values for the scan of
the diode array and the intensity values for up to three individual diodes.
5. After you finish adjusting the attenuators, replace the flowcell access cover, close the front
doors of the detector, and replace the flow restrictor with your LC column.
To calibrate the wavelength accuracy of the optical bench, the PDA detector uses a holmium
oxide reference solution, one of the four calibration files provided with the Xcalibur data
system, or your own custom wavelength calibration file.
IMPORTANT Before you perform a wavelength calibration, verify that the diode array is
not saturated (see “Adjusting the Light Throughput to the Diode Array” on page 315).
451
641
417
241
287
361
485
278
333
345
250
230 270 310 350 390 430 470 510 550 590 630 670
Wavelength (nm)
Before you perform a wavelength calibration, verify that the diode array is not saturated (see
“Adjusting the Light Throughput to the Diode Array” on page 315).
1. From the PDA detector view, choose Surveyor PDA Plus > Direct Controls.
The Direct Control dialog box appears.
2. Click the Calibration tab.
Note The diode array used in the PDA detector has a background count level at each
diode even when both lamps are off. This background count level is called the dark
current. This background or dark current must be subtracted from the intensity counts
before the intensity values are converted to absorbance units (AU). After you perform a
dark current calibration, the PDA detector corrects the spectrum intensity values for the
dark current.
1. Turn on both lamps and wait one hour for the lamps to equilibrate (see “Turning the
Lamps On or Off ” on page 163).
2. In the Dark Current Calibration area, click Calibrate.
The Dark Current Calibration wizard appears.
3. Follow the instructions provided by the wizard. For more information, click Help at the
bottom of the Dark Current Calibration dialog box.
1. Disconnect the LC column from the system, and then pump HPLC-grade methanol
through the flowcell to remove any contaminants.
2. Turn on both lamps and wait one hour for the lamps to equilibrate (see “Turning the
Lamps On or Off ” on page 163).
3. Verify that the array is not saturated (“Adjusting the Light Throughput to the Diode
Array” on page 315).
4. In the Wavelength Calibration area, click Calibrate.
The Wavelength Calibration wizard appears.
5. Follow the instructions provided by the wizard. For more information, click Help at the
bottom of the Wavelength Calibration dialog box.
• To reset the dark current calibration values, in the Dark Current Calibration area, click
Reset Dark Current Calibration To Default Values.
• To reset the wavelength calibration values, in the Wavelength Calibration area, click Reset
Wavelength Calibration To Default Values.
The environmental conditions of your laboratory can cause the dark current of the diode
array to increase over time. For best results, perform an array calibration (dark current) after
any of the following events occur:
• After 100 hours of use or monthly, whichever comes first
• Whenever a 5 °C change in the operating environment occurs
• After you move the detector
• After you replace the lamp
• After you download a new firmware file
Because the dark current produced by the diodes rises as the temperature within the detector
rises, warm up the lamps for one hour before you perform a dark current calibration.
Warming up the lamps for one hour allows the detector to equilibrate to its normal operating
temperature.
The lamps turn off during the dark current calibration routine. After the dark current
calibration routine finishes, the lamps turn back on.
5. Read the list of preconditions, and determine if the PDA detector is ready:
• If the detector meets all the preconditions, click Next to proceed with the calibration.
The status page appears.
• If the detector does not meet all the preconditions, click Cancel to exit the wizard.
Then prepare the PDA detector for calibration and begin this wizard again.
6. Observe the status pane as the calibration proceeds (see Figure 225).
Figure 225. Calibration status page
b. Type a name in the File Name box, and then click Save.
Once you have saved the file with a name of your choice, you can view or print the
contents of the file using any text editing program (see Figure 228).
c. Read the preconditions, and determine if the detector meets these preconditions:
• If the detector meets all of the preconditions, click Next to proceed to the next
page of the wizard where you are prompted to select a wavelength file.
• If the detector does not meet all the preconditions, click Cancel to exit the
wizard and prepare the PDA detector for calibration.
Note On any page of the wavelength calibration wizard, you can click Cancel to
exit the procedure.
4. Do the following:
a. Select a peak set from the list.
The peak set should span the wavelengths you use under normal operating
conditions.
Figure 233 shows the selection of the Holmium Oxide 5 peak set. This wavelength
list instructs the program to calibrate the detector at each of the five wavelengths
shown.
Note The data system has four calibration files to select from. For example, the
Holmium Oxide UV file contains five wavelengths in the UV region, while
the other files use sets of wavelengths from both the UV and visible wavelength
regions. The holmium oxide absorbance maxima are selected from a spectrum
published in “Holmium Oxide Solution Wavelength Standard from
240 to 640 nm - SRM 2034 (NIST Special Publication 260-54).”
Figure 233. Choose Calibration Peak List page
Click to create a
custom peak list.
Note For information about creating a custom peak list, see “Creating and
Editing a Custom Wavelength Calibration List” on page 337.
b. Click Next.
The next page of the wizard appears (see Figure 234).
5. Do the following:
a. Observe the status pane that tells you the wavelength file is being downloaded
(see Figure 234).
Figure 234. Download status for the selected peak list page
Instruction to click Next
b. After you see the message: Click the Next button to proceed with the wavelength
calibration, click Next.
The next page of the Wavelength Calibration wizard appears (see Figure 235). This
page instructs you to place the filter wheel in position 2.
Figure 235. Request to rotate the filter wheel to position 2 page
7. Click Next.
The calibration status page of the wizard appears (see Figure 237).
Figure 237. Calibration status page
8. Do the following:
a. Observe the status pane (see Figure 237).
The diagnostics program waits a few seconds for the rise time filter to equilibrate, and
then the detector takes a holmium oxide scan.
b. When you see the message: Click the Next button to proceed with calibration,
click Next.
The next page of the wizard appears (see Figure 238).
b. Type a name in the File Name box, and then click Save.
After you save the file with a name of your choice, you can view or print the contents
of the file using any text editing program (see Figure 242).
Figure 242. Wavelength Calibration file, viewed in Microsoft Notepad
13. To apply the calibration results to the detector, click Apply on the Delta Values page (see
Figure 240 on page 335).
The success confirmation page appears (see Figure 243).
Figure 243. Success confirmation page for a wavelength calibration
To open the Custom List Name dialog box and the Edit Peak List area
4. Type a name for the list in the box, and then click OK to add the name to the Choose
Calibration Peak List and close the dialog box.
The name of the custom list appears in the Choose Calibration Peak List box.
5. In the Choose Calibration Peak List box, click the custom list name.
The wavelength list becomes available.
6. In the wavelength list, click a wavelength that you want to edit.
The Edit Peak List area appears to the right of the wavelength list.
For information about setting up the LC/MS system, refer to the Getting Connected Guide
for your mass spectrometer. For information about acquiring raw data files from the tune
program for your mass spectrometer, refer to the Getting Started Guide for your mass
spectrometer.
Note You can use the PDA Direct Control dialog box to perform diagnostic tests such as
checking the performance of the deuterium lamp. To acquire PDA data for a single
injection, you must run a single row sequence from the Sequence window.
Figure 246. Surveyor LC Pump page of the Inlet Direct Control dialog box
• For the Surveyor MS Pump Plus, click the Surveyor MS Pump Plus tab. Select the
Take Pump Under Control check box (see Figure 247). Type the appropriate flow
rate in the Flow box. Type the appropriate solvent proportions in the Inlet boxes.
Select the Pump Ready check box. Click (Start) to start the solvent flow.
Figure 247. Direct Control dialog box for the pump
c. In the Vial box, type the location of the sample. Ensure that you have placed a sample
in this location.
For information about the vial and well notation, see “Vial and Well Notation” on
page 6.
d. In the Volume (μL) box, type the volume that you want to inject.
e. Select the injection mode: Full, Partial, or No Waste. For more information, see
“Injection Modes” on page 19.
f. Click Apply.
The autosampler loads the sample into the sample loop connected to the injection
valve as follows:
The XYZ arm of the autosampler moves to the sample location, and then lowers
the needle into the sample vial. The syringe plunger descends, drawing sample
into the needle tubing. The XYZ arm moves back to the injection port, and then
lowers the needle into the injection port. The syringe plunger ascends, pushing
sample into the injection port, through the transfer tubing, and into the sample loop
of the injection valve.
5. On the Surveyor MS Pump Plus page, select the Release Injection check box.
The injection valve switches to the inject position, allowing mobile phase to backflush the
sample out of the sample loop and into the LC column.
Index
Numerics B
384-well microplates, well positions 10, 302 back panels
96-well microplates, well positions 7 Ethernet port location 24
unit ID location 42
baseline
A drift, reducing 209
absorbance level noise, reducing 187
zeroing for the PDA detector during a run 105 zeroing absorbance level
zeroing for the UV/Vis detector 23 PDA detector 103
zeroing on the Display page 178 UV/Vis detector 23
accessories
reservoir vials 5
solvent lines 23 C
standard vials 6 cables
wash bottle assembly 13 Ethernet 3, 41
active and pinned cells, Qual Browser 225 RS-232 serial 2
air bubbles, loaded into the sample loop 19 system interconnect 32
alpha values for flow rate adjustment 265, 268 calibration
analog outputs arm calibration, autosampler 287
changing the polarity 164 column oven 275
testing 165 dark current 323
LC pump 260
areas of peaks, displaying 242
tray temperature 281
arm calibration, autosampler 287
wavelength 328
attenuators, PDA 22 well bottom distance, custom vials or microplates 305
autosampler calibration curves, creating 203
automated sample preparation 113
cell grid, Qual Browser
busy status message 139
adding cells to the real-time display 216
calibrating 274
cell states 224
configuration parameters 45
cursor actions 226
custom vials or microplates, using 305
inserting cells 244
Direct Inlet Control dialog box 343
channels, instrument method 100
instrument method parameters 85
loading trays 210 chemicals, filtering 210
timed events table parameters 94 chromatograms
transfer tubing 17 cursor actions, cell grid 226
vial and well notation 6 labels (retention time, area, and so on) 241
autozero function for the UV/Vis detector 107 peak detection 252
UV data, displaying 250
Avalon integration algorithm 237
column
See LC column
column arrangement, sequence spreadsheet 205
M PDASnapshot.csv 318
peaks, chromatographic
maintenance log, autosampler 54
areas, displaying 251
manuals, accessing xvi
peak detection 251
methods
purity settings 255
See instrument methods or display methods
polarity, active high or active low 164
microplates
Post-Injection Valve Switch Time box 89
384-well 10
Power LED
96-well 7
autosampler 38
description 6
LC pump 37
selecting tray options 47
MS pump 36
MS pump
PDA detector 39
instrument method parameters 74
UV/Vis detector 40
operating without solvent 77
power, turning on/off 134
pressure settings, minimum and maximum 77
pressure stability 77 precision, injection-to-injection 89
priming the pulse dampener 156 preheating tubing 21
Ready signal 77 Prep Operations list 116
pressure
minimum for LC pump 69
N recorder terminals 264
names, chromatographic peaks 242 sensor adjustment 269
needle and needle tubing assemblies, description 16 units for displaying backpressure 63
New Sequence Template dialog box 202 pressure recorder terminals, scaling the output 268
no waste injection pressure transducer
description 19 replacing 265
injection volume recommendation 91 zeroing 265, 269
instrument method parameter 90 processing method (.pmd) file type 203
noise, detector 170 pump
Normalization page, Display Options dialog box 240 See LC pump or MS pump
Pump Ready signal 35
O
On, Off, Standby shortcut menu 154 Q
On/Off switches for LC devices 135 Qual Browser
output signal terminals 164 accessing 219
Oven page readbacks, autosampler 142 cell grid 224
cursor actions 226
display options, UV chromatograms 240
P layout files, creating 233
partial loop injection
description 20
instrument method parameter 89
R
PCR plate, configuring 48 Ready signal, pump 77
PDA detector ready to download message 137
dark current calibration 323 real-time data
data acquisition parameters 95 locking 214
display method parameters 101 reviewing 215
functional description 22 regulatory compliance iii
instrument method parameters 98 removal position, needle 303
light throughput, adjusting 315 report, instrument method 232
parameters in method report 232 Reservoir Content page, autosampler solvent descriptions 93
timed events parameters 102 Reset Lifetime button 173
wavelength calibration 328
tubing wavelength
behind column oven 21 accuracy, calibrating 328
degasser 2 calibration list, editing 338
needle tubing 13 data acquisition range
solvent lines 23 PDA detector 99
wash bottle 13 UV/Vis detector 108
Turn Device On command 154 WEEE compliance ix
well bottom distance calibration 305
U well position and well ordering, selecting 49
unit ID
description 42 X
relation to communication protocols 24 x- and y -axis offset values, arm calibration 300
unpinned cells, Qual Browser 225 Xcalibur data system
UV/Vis detector Instrument Setup Window, opening 68
configuring 66 Qual Browser, opening 219
direct controls 183 Roadmap view 135
instrument method parameters 108 Sequence Setup window, opening 201
lamp controls 183 XYZ arm
parameters in method report 232 description 15
wavelength modes 23 home position 296
moving to the back of the tray compartment 303
V position, calibrating 287
valves
six-port injection valve 18 Z
syringe two-position rotary valve 13 Zero DACs button 175
variable volume injections, types 20 Zero Data
Verify Door Is Closed check box 6, 54, 290 button on Display page 170
vial bottom sensing 54 event type, PDA detector 105
Vial Not Found message 210 zeroing, absorbance level
vial trays, notation 6 PDA detector 103
viscous flush solvents 89 UV/Vis detector 23, 107
volume
flush 197
full loop injections 20
gradient delay, pulse dampener in LC pump 3
no waste injections 19, 90
partial loop injections 20
preheat tubing 21
sample loops 18
sample preparation operations 115
standard needle tubing and needle tubing extension 13
tubing in chambers of vacuum degassing unit 2
W
Wait for Temperature Ready check box 54
wash bottle location 13