Autoclave Guide
Autoclave Guide
Benchtop Steam
Sterilizers –
Guidance on
Purchase,
Operation and
Maintenance
BULLETIN
MDA DB2002(06)
October 2002
The Medical Devices Agency helps safeguard public health by working with users, manufacturers and
lawmakers to ensure that medical devices meet appropriate standards of safety, quality and performance
and that they comply with the relevant Directives of the European Union.
Our primary responsibility is to ensure that medical devices achieve their fullest potential to help
healthcare professionals give patients and other users the high standard of care they have a right to
expect.
2. INTRODUCTION ................................................................................. 4
2.1 Classification of benchtop steam sterilizers ......................................... 6
3. PURCHASING...................................................................................... 8
3.1 Service and maintenance.................................................................... 10
4. OPERATION....................................................................................... 11
4.1 Installation, validation and periodic testing ....................................... 11
4.2 Use of sterilizers................................................................................. 11
4.3 Training.............................................................................................. 11
4.4 Records............................................................................................... 12
4.5 Storage of devices after sterilization .................................................. 13
4.6 Health and safety................................................................................ 14
4.6.1 Hazards...................................................................................... 15
4.6.2 The circumvention of sterilizer safety features ......................... 15
4.6.3 Legal and insurance considerations........................................... 15
7. GLOSSARY ........................................................................................ 26
8. REFERENCES .................................................................................... 28
1
Appendix 1 Factors to consider in choosing between benchtop
sterilizers, single-use devices or SSD ................................................. 31
Flow chart for decision making.................................................................... 32
List of tables
2
1. EXECUTIVE SUMMARY
This Bulletin provides guidance on the purchase, operation and maintenance of
all types of benchtop steam sterilizers. It is a revised compilation of the
guidance previously issued in DB9605 and DB2000(05).
Who this document This guidance is intended for potential purchasers, and all current owners
is for and users of benchtop steam sterilizers and should be of particular interest
to:
3
2. INTRODUCTION
Minor surgical procedures are increasingly being performed in primary
healthcare facilities. There are a number of choices for obtaining the
sterilized devices for these procedures:
Note: The term ‘devices’ is used throughout this bulletin to encompass devices,
instruments and medical equipment.
If you choose to use reusable devices, whenever practicable you should obtain
them from a Sterile Services Department (SSD). These have the equipment and
expertise to clean and sterilize reusable medical devices effectively and
consistently, combined with economy of scale.
4
Table 1 Sterilization temperature bands, holding times and pressures for
steam sterilization.
Sterilizing temperature
range (oC) Approximate Minimum hold
pressure (bar) time (minutes)
Minimum Maximum
The effectiveness of the process depends upon direct contact between the
steam and all surfaces of the load. Therefore all items that you intend to
sterilize must be clean and dry before you place them in the sterilizer
chamber. Contamination with residual tissue, body fluids, oil or other
deposits:
• will prevent contact between the steam and the surfaces of the load;
• might become fixed to the load items and be difficult to subsequently
remove;
• might also contaminate the water and encourage bacterial growth.
Air retained within the chamber and load can also prevent steam contact.
Consequently it is important that the operator should load the sterilizer
according to the manufacturer’s instructions and take care not to overload it,
otherwise air removal might be impaired and sterilizing conditions might not
be attained.
‘…all parts of the load to be sterilized must be free of air, so that it may
be permeated by the steam.’
Sterilisation by steam under increased pressure.
The Lancet 28 February 1959.
5
2.1 Classification of A European standard for benchtop steam sterilizers is currently being
benchtop steam prepared*. It classifies benchtop steam sterilizer cycles according to the types
sterilizers of load they are intended to process, which are summarised in Table 2:
Note: Ideally, only the highest specification cycle should be available to the
operator. Other cycles should be disabled, until specifically needed.
*
‘Small steam sterilizers’ prEN13060 : 2002 (E)
6
Traditional (gravity displacement) benchtop steam sterilizers displace air
passively from the chamber and load by steam generated within the sterilizer
chamber or in a separate boiler within the sterilizer’s casing. This is known as a
‘Type N’ cycle.
Note: Type S sterilizers should be used to process only the types of loads
specified by the sterilizer manufacturer.
Type N cycles are intended to be used to sterilize solid devices that are not
wrapped. Devices that are wrapped (the term ‘wrapped’ includes sterilization
pouches) and devices that are hollow or have lumens cannot be sterilized in this
type of sterilizer. These types of loads should ideally be sterilized in a SSD but
alternatively may be sterilized in a properly functioning vacuum steam
sterilizer that has been validated for the intended load (see section 4.1).
Type B cycles are intended for wrapped solid items (e.g. forceps, dental
probes), hollow items (e.g. cannulae, tubing), whether or not they are
wrapped, and for porous loads e.g. fabrics, swabs and dressings. They are
necessary for items that cannot be processed using a Type N cycle (or a
Type S cycle, unless it is intended specifically for these load types). Type B
cycles must have a drying stage to ensure that the load is dry before the door
is opened, which can increase the total cycle time considerably.
Type S cycles are intended for types of loads specified by the manufacturer
of the sterilizer. They have a forced air removal system. (Forced air removal
can be achieved using a vacuum pump, or superatmospheric pulsing or
steam injection through the lumens of devices).
The effectiveness of the air removal stage determines the types of load they
are designed to process. Some models have a drying stage, which will
prolong the cycle time.
7
3. PURCHASING
Before purchasing a benchtop steam sterilizer, you should consider the
numerous factors involved in obtaining sterilized medical devices. These are
summarised in Appendix 1.
Benchtop sterilizers are regulated as medical devices and must comply with
the Medical Devices Regulations 2002, which require all medical devices to
carry the CE marking. Therefore, the CE marking does not help in choosing
between makes and models.
Buyer beware
y The manufacturer should state clearly the types of load for which the
sterilizer is suitable.
y You should only use the type of sterilizer that is suitable for the types of
loads that you intend to process.
y Some sterilizers have more than one type of cycle. Ensure you use the
correct cycle for the load.
y The presence of a vacuum stage does not guarantee that the sterilizer
is capable of processing every type of load.
This will form part of the purchasing contract and places the onus on the
supplier to provide equipment that is fit for the purpose that you have
specified. The supplier will also be under an obligation to draw your
attention to any limitations to the use of the sterilizer. The contract should
also specify who has the responsibility for installing the sterilizer and
performing installation checks and tests to ensure that the sterilizer will
perform to its design specification. These tests are likely to require
specialist knowledge and equipment.
8
Many sterilizers are bought on verbal assurances of their suitability but
are later found to be unsuitable for the intended purpose. You should ask
for the assurances in writing to reduce the possibility of
misunderstandings.
It is also important to try to involve everyone who might use the sterilizer. If
possible obtain one for a trial period before committing yourself to purchase
as this will help you to assess other important aspects e.g.:
9
3.1 Service and Efficient and effective service support is vital and should be an important factor
maintenance in your purchasing decision.
If the supplier is unable to provide any of the above, ask if they have a
servicing agent, or if there is an organisation they can recommend.
Vacuum benchtop steam sterilizers are much more complicated than their
traditional counterparts and, in general, complicated devices tend to break
down more frequently, and cost more to repair.
10
4. OPERATION
4.1 Installation, After you have bought your sterilizer it has to be:
validation and
periodic testing (i) Installed and then validated.
A new sterilizer has to be installed, commissioned and validated before you
use it. You should retain all records of these activities in the sterilizer
logbook for future reference (see section 4.4). An AP will be able to provide
advice about the validation of a new sterilizer and a qualified Test Person
(sterilizers) (TP) should carry out the validation tests.
4.2 Use of sterilizers Before any loads are processed you should ensure that all the relevant checks
and tests are performed to provide assurance that the sterilizer is safe to use,
that it is functioning correctly and that it will sterilize loads consistently. Check
that there is water in the reservoir before you attempt to operate the
sterilizer. (You will find information about the appropriate water quality in
section 6).
11
Well-trained staff using well-maintained equipment minimises risks both to
themselves and to the patient; failure to sterilize a device has implications
for both. Litigation involving cross infection is particularly difficult to
defend and in the event of an adverse incident, discovery of failures in
training or maintenance may lead to a finding of liability by the courts, and
charges of professional misconduct.
You should keep a permanent record for each sterilizer, to provide evidence
4.4 Records
that it was/is functioning correctly and achieving sterilizing conditions
consistently.
This permanent record can take any convenient form e.g. a book, a loose-leaf
folder, or an electronic device (provided that it will give a printout on
demand). Appendix 2 gives examples of logbooks.
It should be kept close to the sterilizer so that records can easily be kept up
to date. It should provide a complete history of the sterilizer and should
include:
12
4.5 Storage of devices After the end of the sterilizing stage, the steam condenses in the sterilizer
after sterilization chamber, so the load will be wet unless there is a subsequent effective drying
stage whilst in the sterilizer.
Controlled atmospheres are seldom found in primary care. The sterilized load
items will be contaminated immediately the sterilizer door is opened. The
devices cannot therefore be regarded as being sterile but will be in the same
condition as the atmosphere in the treatment area.
The microbiological condition of these devices should differ little from those of
non-wrapped devices that are used direct from the sterilizer.
Sterile use
If you wish to store sterilized devices for future use in sterile conditions you
must:
• process them in suitable wrapping material, in a suitable sterilizer that has
an effective post-sterilization drying cycle;
• ensure that the packaging material is thoroughly dried before the sterilizer
door is opened.
13
Items within packaging can only be regarded as sterile if they have been
subjected to a validated sterilization process and they are dry when they
are removed from the sterilizer. Loads packaged after sterilization, for
later use, cannot be considered sterile.
Storage time
There is no specified shelf life for sterilized items. Products will remain
sterile indefinitely provided the packaging remains intact, clean and dry.
However, you should be aware that some devices and/or packaging might
deteriorate over time. You should therefore:
y set a shelf life by consultation with the manufacturers of the device and
wrapping;
y have an effective stock management system to ensure sterile items are
either used within that shelf life, or are re-sterilized.
Packages must be inspected for damage before they are opened. If there is
any sign of damage to the packaging, the contents must be re-sterilized
before they are used.
4.6 Health and safety Users will go some way towards meeting their obligations under the Consumer
Protection Act and Health and Safety at Work etc Act by ensuring that
equipment:
Users should also ensure that there is a written scheme of examination for
the pressure system under the PSSR. The pressure system must be examined
periodically in accordance with this scheme.
14
4.6.1 Hazards
Some common hazards associated with benchtop steam sterilizers are:
To prevent serious injury to the operator and others, the sterilizer door must
prevent access to the chamber while it is under pressure.
y You should not be able to open the door until the ‘cycle complete’ signal
is indicated and there is no residual pressure in the chamber.
y If the chamber pressure is indicated electronically, there should also be a
mechanical indication in case the electronics or power fails.
y Opening the sterilizer door when a fault message has been cancelled
may result in hot water being spilt and possible injury to the operator.
Sterilizers that are not maintained correctly and are not tested periodically
can be dangerous. The force on a benchtop sterilizer door can be about ¾
tonne.
15
Pressurised steam is hazardous. Sterilizers should be insured to cover the
particular risks associated with steam sterilizers. Some insurers specialise
in this type of risk and might be the most economical choice.
The PSSR covers the installation and use of this type of equipment and
amongst other things require:
Their actions might also invalidate insurance cover taken out to indemnify
users and their employers against legal liability.
16
5. MAINTENANCE, PERIODIC TESTING, ROUTINE
MONITORING AND DOCUMENTATION
Maintenance, periodic testing, routine monitoring and documentation are an
5.1 Introduction
essential combination to ensure that a sterilizer is functioning correctly and
that it will produce sterilized loads consistently. This is because the
effectiveness of the sterilization process cannot be verified retrospectively by
inspection or testing of the product, and can only be guaranteed if sterilizing
conditions are created throughout the sterilizer chamber and the load during
every cycle. Guidance on these activities is provided in HTM2010 (NHS
Estates) and DB9804 (MDA 1998).
5.2 Maintenance A Test Person (TP) should draw up a schedule for periodic testing. It is the
responsibility of the TP and the owner or user to ensure that these tests are
performed.
5.3 Periodic testing Periodic testing consists of a programme of tests that are intended to
demonstrate that the sterilizer’s performance is satisfactory. The tests are
carried out at daily, weekly, quarterly and yearly intervals, with the user and the
TP sharing the responsibility for performing them. These tests are preceded by
safety checks which are intended to ensure the sterilizer is both safe to use and
to test.
If the sterilizer fails any safety check you should not attempt to test it until
the faults have been corrected and the sterilizer passes all safety checks.
y The user does the daily tests. After suitable training and with the agreement
of the Authorised Person (AP) the user may also do the weekly tests.
y Some of the weekly tests for vacuum benchtop steam sterilizers normally
require the services of a TP and use of specialised equipment, and therefore
cannot be performed by the user. However, some manufacturers have
designed an automated test facility into the sterilizer so that the sterilizer
can perform some of the specialised tests itself.
17
y The quarterly and annual tests require specialised equipment and skills, and
should be carried out only by a properly qualified TP. Each cycle available
to the user should be tested.
If the sterilizer is not tested periodically you will not know if it is working
correctly.
(i) The steam penetration test is intended to show that steam will penetrate
rapidly and evenly into a test device that is at least as difficult to sterilize as
the intended load. The test device contains an indicator that responds
(usually it changes colour – and should do so completely) only when steam
penetration is adequate. It is essential to use both the steam penetration test
device and the indicator specified by the sterilizer manufacturer, otherwise
the test results may be dangerously misleading (see section 5.4). The test
piece and the indicator should be as specified in BS EN867, or an alternative
provided that it is equivalent.
If you do not use the test device and indicator combination specified by
the sterilizer manufacturer, the results might be dangerously misleading.
(ii) The automatic control test may be done at the same time as the steam
penetration test, but is not required if the sterilizer is equipped with a
recorder that provides a permanent record of the temperature, pressure and
elapsed time during all sterilizing cycles. However, it is essential to compare
the printed record of every cycle with one obtained when the sterilizer was
known to be functioning correctly (e.g. during the periodic testing performed
by the TP).
18
You should check with the manufacturer whether you have to pre-heat the
sterilizer chamber before performing these tests, as this can extend the test
time.
Appendix
Type N Type B & S DB9804
Weekly test 4
(Traditional) (Vacuum) reference
reference
a) Air leakage test (automatic) 9 2.2.1 A.3.1
b) Automatic air detection 9 2.3
system function test
c)* Automatic control test 9 9 1.1 A.2
d)* Steam penetration test 9 1.2 A.1
The air leakage test is intended to check that air will not leak into the sterilizer
during periods of vacuum, at a rate that is greater than that specified by the
sterilizer manufacturer. Air leaking into the chamber can:
19
5.3.4 Quarterly and annual checks and tests
These require specialised test equipment and only a person (e.g. a TP) who has
the necessary training, experience, skills and equipment should perform them.
Guidance on quarterly and annual testing should be sought from an AP.
5.4 Use of chemical and If the sterilizer’s controller indicates a failed operating cycle, the cycle must be
biological indicators regarded as unsatisfactory, regardless of the results obtained from any chemical
or biological indicators. Chemical and biological indicators do not indicate
that the load is sterile.
• process indicators e.g. autoclave tape and indicators printed onto bags and
pouches. These indicators serve only to distinguish processed items from
unprocessed items, and should not be used for any other purpose;
• performance indicators for specific tests e.g. the indicators used to check
the effectiveness of steam penetration into a test pack or a process
challenge device;
• integrating indicators (emulating integrators) are available for
monitoring steam sterilizers. They are designed to monitor the
attainment of two or more critical variables in the sterilization process,
either by a graduated response or a defined end point reaction.
20
5.5 Procedure on failure A failure of a test implies that the sterilizer is not working to specification. The
of a test user should have a written procedure for handling test failures but, in all cases,
the sterilizer must be withdrawn from service, the failure investigated, the cause
rectified, and the sterilizer re-tested successfully before being used.
Note: The user has the ultimate responsibility for certifying that the
sterilizer is fit for use.
5.6 Monitoring and Records of maintenance, testing and operating cycles provide evidence that the
documentation process will deliver sterile product consistently. HTM2010: Part 4 (NHS
Estates 1997) provides guidance on the testing documentation that should be
kept. Records of checks, tests and maintenance performed on the sterilizer’s
chamber must be documented and kept securely as specified in the PSSR (see
section 4.6.3 and Appendix 5).
Routine monitoring
For each production cycle you should:
Cycle records
Every production cycle must be fully documented and the records kept securely
for the time specified by management. The information recorded should
include:
• the date and cycle number;
• the type of load (e.g. whether porous materials, solid instruments, hollow
instruments or a mixture etc.);
• the sterilization cycle selected;
• whether the cycle was a pass or a fail;
• the chart record for the cycle;
• the identity of the operator.
21
Recorders
It is not a requirement to have a recorder fitted, or attached, to the sterilizer
but it is desirable because it:
The printout should be kept securely in the sterilizer logbook (see section
4.4). Electronic data storage can replace printed records.
Note: Some types of printouts fade quickly (e.g. from thermal printers) and
you might therefore need to take special action to preserve these records
(e.g. photocopying).
• the user should compare production cycle records to verify that sterilizing
conditions have been achieved for each load;
• the results of the weekly tests should be compared to establish whether the
sterilizer is functioning correctly and achieving sterilizing conditions;
• the results of periodic tests and performance re-qualification tests can be
compared.
22
6. MAINTENANCE OF RESERVOIRS AND
STERILIZER CHAMBERS
It is important that the sterilizing process should not contaminate the load
6.1 Introduction
items. It is possible for benchtop sterilizer loads to be contaminated by
impurities in the water used to generate the steam. Contamination can be
minimised by appropriate maintenance of the sterilizer chamber and
reservoir, and by the use of suitable quality water.
Water supply
6.2 Sources of
Benchtop steam sterilizers generate their own steam either within the
contamination
chamber or in an adjacent boiler within the casing. Water droplets are
present in steam, therefore it will contain the same contaminants as the
water. These include minerals, toxic metals, and micro-organisms and their
toxic products. When the steam condenses on the load during sterilization,
contaminants will be transferred to the surfaces of the load items where they
will be concentrated when the load dries. The quality of the water in the
sterilizer reservoir, and chamber/boiler, is therefore crucial.
Reservoir water
Benchtop sterilizers usually have a reservoir for storing water to supply the
chamber/boiler. Many benchtop steam sterilizers discharge condensate and
residual water back into the reservoir at the end of each cycle Water left
standing in the reservoir, and residual water or moisture in the
chamber/boiler following a sterilization cycle, will quickly become
colonised with micro-organisms which can be harmful to the patient.
Endotoxins
Although the micro-organisms in the water will be killed during the
sterilization cycle, a heat-stable toxic substance (endotoxin) in the cell wall
of many of them will remain intact. Endotoxins are resistant to steam
sterilization and are only inactivated by heating for several hours at
temperatures above 180oC. Endotoxins will continually accumulate in the
water and increasingly contaminate the steam until the water is changed. There
are also benchtop steam sterilizers that discharge residual water and condensate
into a separate container or direct to a drain. This minimises the accumulation
of endotoxins in the water in the reservoir, but it does not remove the need to
drain and clean it frequently.
Debris
Microbial growth may be assisted by contamination of the water, for
example with debris from poorly cleaned instruments, or oil from dental
handpieces.
23
6.3 Cleaning The water reservoir should be cleaned regularly, however few sterilizer
reservoirs are designed to be easily cleaned. The MDA recommends that you
follow the manufacturer’s guidance on how to clean the reservoir. There are
anecdotal reports that some maintenance organisations have suggested the use
of disinfectants to clean reservoirs, the internal surfaces of pipework and
chambers/boilers. These disinfectants might cause corrosion of the chamber or
other components of the pressure containment system and cause failure, with
the associated high risk of injury to persons nearby (see section 4.6.1). If you
are considering using disinfectants for this purpose, you should follow the
advice of the sterilizer manufacturer.
6.4 Water quality The MDA recommends the use of sterile water for irrigation BP in benchtop
steam sterilizers as it has specification limits for mineral, toxic metal, and
endotoxin contaminants.
Sterile water for irrigation safeguards both the sterilizer and the patient and is
the preferred quality if you perform surgically invasive procedures, in which
endotoxin might be introduced parenterally into the patient. Endotoxin
contamination of the sterilizer water might be of less concern if all the
procedures carried out in a practice present low risk of endotoxin
contamination to the patient (e.g. the instruments only make contact with
intact skin or mucous membrane).
6.5 Suggested The following guidance should help you to minimise contamination of your
maintenance sterilizer.
• Don’t leave water standing in the reservoir for more than a few hours. If
you are not sure how long the water has been there, change it.
• At the end of the working day, or whenever the sterilizer is to be left
unused for several hours, drain the chamber and water reservoir, rinse all
internal surfaces with sterile water for irrigation and leave them dry.
24
• Don’t top up the reservoir. First drain the contents and then rinse it
carefully with sterile water for irrigation before refilling to the level
recommended by the manufacturer.
• Ensure all load items are scrupulously clean and dry before placing them
in the sterilizer.
25
7. GLOSSARY
AP – see Registered Authorised Person
BP – British Pharmacopoeia
Forced air removal (active air removal) is the removal of air from the
chamber using mechanical means. (A vacuum pump, steam pulsing or steam
injection through the lumen of a device may be considered to be forced air
removal).
Manager – the person who is ultimately accountable for the operation of the
premises. Depending on the nature of the organisation, this may be the owner,
occupier, employer, general manager, chief executive, or other person of
similar authority. In small, autonomous installations the manager might also be
the user.
Operator – any person with the authority to operate a sterilizer, including the
noting of sterilizer instrument readings and simple housekeeping duties and by
agreement daily/weekly testing.
Porous material – material, or load configuration, that can hold or trap air that
will interfere with steam penetration.
26
Registered Authorised Person (sterilizers) [AP(s), abbreviated to AP] is a
person designated by management to provide independent auditing and advice
on sterilizers and sterilization and to review and witness validation and periodic
test documentation. A list of suitably qualified APs is maintained by the
Institute of Healthcare Engineering and Estates Management (IHEEM) (see
Appendix 6).
Sterilizing conditions – the ranges of the cycle variables that must prevail
throughout the chamber and load during the holding time.
Type B sterilization cycles are intended for the sterilization of wrapped and
non-wrapped solid, hollow and porous loads. They have a forced air removal
system.
Type S sterilization cycles are intended only for the sterilization of products
specified by the manufacturer of the sterilizer. They have a forced air
removal system.
27
8. REFERENCES
British Standards Institute. BS 3970-1:1990 Sterilizing and disinfecting
equipment for medical products. Specification for general requirements. BSI.
Gifford P. HSE’s Strategy for the healthcare sector. ISSM Journal Vol.3 No.1
8-10 July - September 1998.
Health and Safety at Work etc Act 1974, HMSO ISBN 01105437743.
Leigh B. The Law and Hospital Infection. Journal of Hospital Infection, vol
40, Supplement A, September 1998, ISSN 0195-6701.
Medical Devices Agency (1998) The validation and periodic testing of vacuum
benchtop steam sterilizers. DB9804.
28
Medical Devices Agency (2002) Sterilization, disinfection and cleaning of
medical equipment: guidance on decontamination from the Microbiology
Advisory Committee to Department of Health Medical Devices Agency. ISBN
1 85839 518 6.
NHS Estates (1997) Health Technical Memorandum HTM 2031 Clean steam
for sterilization. The Stationery Office.
The Pressure Systems Safety Regulations 2000 SI 2000 No. 128, The
Stationery Office.
The Provision and Use of Work Equipment Regulations 1998, SI No 2306, The
Stationary Office.
29
DB 9804. The validation and periodic testing of benchtop vacuum steam
sterilizers. June 1998.
30
Appendix 1 Factors to consider in choosing between benchtop
sterilizers, single-use devices or SSD
By using single-use devices or by having your devices reprocessed by a SSD, you reduce your legal
liability that might arise from reprocessing devices yourself.
The practicability of using a SSD will be determined by a number of factors, particularly the timely
availability of sterilized devices, which will depend on:
Safety (e.g. the assurance that the load has been sterilized) should be the prime consideration. The most
cost-effective solution should emerge from careful consideration of the respective costs of each option.
These should be weighed against the costs for the benchtop sterilizer option, including:
• the purchase price of the sterilizer;
• time required for daily testing;
• cost of maintenance and periodic testing;
• operational costs (electricity, sterile water, insurance etc.);
• cost of additional instruments to accommodate prolonged cycle times;
• cost of training operators;
• reliability factor (where assessable);
• cost of meeting legal requirements (see section 4.7.2);
• cost and operation of automated instrument cleaning equipment.
MDA’s advice is to use a SSD to reprocess all reusable devices, wherever possible. This has been
shown to be cost effective (Wilson et al 1999).
* Turn-round time is the time between the used instruments being despatched and their return from the SSD.
31
Flow chart for decision making
N
N Y
Is the cost of SSD etc. equal I want to process solid, I want to process
to or less than the running unwrapped instruments wrapped loads, porous
costs of own sterilizer? for immediate use or for loads and tubular devices
clean storage
Y
Use traditional (non- Use suitable vacuum
Use vacuum) benchtop steam benchtop steam sterilizer
SSD sterilizer
Appendix 2 Ten good-practice points for the use of benchtop
steam sterilizers
• Where possible, use a sterile service facility rather than processing locally.
• All items must be clean and dry before loading into the sterilizer. Do not overload
the sterilizer – the load items might not be sterilized.
• Sterilization performance must be checked frequently (daily [this includes the steam
penetration test on vacuum benchtop sterilizers] and weekly by the user; quarterly
and annually by a competent test person). Keep written records. This is in addition to
routine maintenance and cleaning.
• Drain and clean chamber and reservoir at the end of each day and leave dry.
Replenish with sterile water for irrigation from an unopened container.
• Have the sterilizer's pressure system checked for safety. Keep records of all checks
and repairs to the pressure system. This is a legal requirement. Do not circumvent
safety features.
• Keep written records all testing and maintenance carried out on every sterilizer. The
records should be kept in a logbook.
33
Appendix 3 Examples of logbook pages
Appendix 3(iv) and Appendix 3(v) are reproduced with the kind permission of Scottish Healthcare Supplies.
See Appendix 6 for contact details.
Autoclave details
Hospital/location Department Serial No.
Ref. No.
34
Appendix 3(ii) Daily test sheet
Reservoir water changes (where applicable). Drain, rinse and refill with sterilized water for irrigation. See
HTM 2031
Cycle number when water Water changed
Comments
changed by
Monday
Tuesday
Wednesday
Thursday
Friday
Saturday
Sunday
35
Appendix 3(iii) Weekly test sheet
Tests to be carried out in accordance with DB9804.
*Automatic
*Automatic Automatic Steam Weekly safety
air detector Certified fit
Week Cycle air leakage control test penetration checks
system for use
beginning number test result result test Satisfactory/
function test by user
Pass/Fail Pass/Fail Pass/Fail Unsatisfactory
Pass/Fail
P/F P/F P/F P/F S/U
P/F P/F P/F P/F S/U
P/F P/F P/F P/F S/U
P/F P/F P/F P/F S/U
P/F P/F P/F P/F S/U
P/F P/F P/F P/F S/U
P/F P/F P/F P/F S/U
P/F P/F P/F P/F S/U
* Only where the sterilizer has an in-built self-test programme. Otherwise the test should be carried out by a TP
and copies of the TP’s test sheets should be inserted.
36
Appendix 3(iv) Autoclave history record sheet
Type of autoclave
Hospital/Location Start date for this sheet
Department Ref. No Ser. No
©Scottish Healthcare Supplies 0131 551 8905 PHR1 / ISSUE 2 / Jan 2000
Appendix 3(v) Production log sheet- benchtop autoclave
Date Cycle Cycle start Cycle Description of load Cycle Printout Comments
number time selected pass checked OK and
(if operator
applicable) initials
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
38
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
Yes/No Yes/No
©Scottish Healthcare Supplies 0131 551 8905 UIUS0 / ISSUE 1 / Jan 2000
Appendix 4 User tests for benchtop steam sterilizers
1 Daily tests
You should perform the daily tests every day that the sterilizer is used, as they are your check that the
sterilizer is achieving sterilizing conditions. (The daily tests also form part of the quarterly and annual
tests carried out by the TP(s).)
The test requires the temperature and pressure profiles, and the elapsed time of the cycle to be compared
with the values obtained when the sterilizer was known to be working correctly e.g. immediately after the
TP(s) tested it using calibrated instruments. You should perform the test using the sterilizing cycle with
the highest temperature compatible with the load. If your sterilizer does not have a printer fitted, observe
and note the following during the sterilizing (holding) stage of the cycle:
A printer fitted to the sterilizers will perform this task for you, providing an accurate, permanent, record
and saving your time. Note it is the MDA's opinion that benchtop sterilizers should be equipped with
a printer. You should compare the values that you noted, or the values on the print out, with those on the
master process record (see section 5.7).
• the chamber temperature and pressure is within the limits of the appropriate band, for the duration
of the holding time, as specified in Table 1(see section 2);
• a visual display of ‘cycle complete’ is indicated;
• no mechanical or other anomaly is observed.
For traditional, gravity displacement sterilizers (Little Sister® types), the test should be performed with the
chamber empty (see HTM 2010 Part 3,) but to save time, it is often performed while sterilizing a load. If
you perform the test with a load, ideally the load composition should be the same each day). For sterilizers
with a type B or type S cycle, you can do this test at the same time as the steam penetration test (T1.2) but
the steam penetration test must be performed with the chamber empty except for the test device.
You should perform this test at the start of every day the sterilizer is used, as it is important to be sure that
the air removal stage is effective, and that any residual air and other non-condensable gases (NCG) will
not interfere with the sterilization process. It is essential to perform this test with only the test device in
the chamber. Anything else in the chamber will disrupt the test and produce an erroneous result.
This test is analogous to the Bowie and Dick test performed on large porous load sterilizers in hospitals’
SSD departments. A specified test pack of material, containing a special chemical indicator, is processed
through a sterilization cycle. When the level of NCG is below predetermined limits, steam will penetrate
rapidly and completely into the pack and the indicator will show a uniform colour change. When NCG is
present, it will collect towards the centre of the pack as a bubble and will impair contact between the
steam and the indicator. The temperature or moisture level (or both) will be lower in the region of the
bubble and will result in a non-uniform colour change of the indicator. When the indicator is examined
39
after processing, a uniform colour change of the indicator over its entire surface shows that sufficient air
and other NCG had been removed to allow the steam to penetrate rapidly and evenly into the test pack.
The result of the test should be recorded in the sterilizer log book, and the indicator paper should be
marked with the result and kept for reference for six months. (The relevant standard, BSEN867: 2001
requires the colour of the processed indicator to remain stable for six months after processing, when stored
according to the manufacturer's instructions.)
If the test result is unsatisfactory, repeat the test. If it is still unsatisfactory, the sterilizer is faulty.
Call an engineer to investigate and repair the fault. The sterilizer must not be used until the fault
has been rectified.
There are two types of test pack, a standard test pack that you make up yourself, and a proprietary single-
use test pack that you purchase. The specification for the standard pack is provided in BS EN 867: 2001
Part 5. You have to incorporate into the pack an indicator sheet made to a stringent specification.
Proprietary packs normally incorporate an appropriate indicator sheet during manufacture.
Note: You must use the test pack specified by the sterilizer manufacturer because if you do not, the
results could be dangerously misleading.
If you use the standard test pack, you should use it according to the instructions of the sterilizer
manufacturer but if you use a proprietary test pack, follow the instructions of the manufacturer of the test
pack.
This type of cycle is for processing the loads specified by the sterilizer manufacturer. The sterilizer
manufacturer will therefore have to specify the test to demonstrate the effectiveness of the air removal
stage. You should use only that test unless there is an alternative available that has equivalent
performance. (NB Validation data must be available to demonstrate equivalence.)
Place the test pack in the position specified by the manufacturer - which should be the position from
which it is most difficult to remove air from the load. Select a standard cycle or the cycle specified by
the sterilizer manufacturer.
Note: if there is a special cycle for the air removal test, it must have exactly the same air-removal
process as the standard processing cycles.
At the end of the test, examine the indicator sheet. The test is satisfactory if the indicator has changed
colour uniformly over its entire surface, as specified by the indicator manufacturer.
2 Weekly tests
You should perform these before starting the sequence of weekly tests.
Examine the door seal and replace it if necessary. You might need to call an engineer to do this.
40
Note: It would not be prudent to attempt to open the door while the chamber is pressurised. The
door securing mechanism (which includes hinges as well as the door lock and interlocks) should
be checked before the chamber is pressurised, or at the start of a cycle. You should consult the
sterilizer manufacturer for details of the appropriate procedure.
This test should be performed weekly. It is an important test because if air leaks into the chamber at a rate
greater than that specified by the sterilizer manufacturer:
• it might interfere with the penetration of steam into the load and prevent sterilization;
• during the drying stages, it will not have passed through the bacterial retentive filter and there is a risk
of recontaminating the load.
Air is first removed from the chamber until the pressure is the lowest achieved in all of the cycles
available on the sterilizer and then the vacuum source is isolated and all valves connected to the chamber
are closed. The absolute pressure is measured at the end of the vacuum stage. Any subsequent rise in the
chamber pressure will be caused by air leaking into it - and the rate of pressure rise in the chamber is
measured.
Ideally the sterilizer should be equipped with an automated test cycle so the user can do the test. If there is
not an automatic test facility, a TP(s) has to do the test using special, calibrated instruments.
• the absolute pressure at the end of the air removal stage is within the limits specified by the
manufacture and
• the rate of pressure rise must not be greater than 1.3 mbar per minute.
A machine that fails to meet the requirements of this test should not be used until the fault has been
rectified and the test satisfactorily completed.
You should carry out this test weekly, following the instructions of the sterilizer manufacturer
The sterilizer must indicate clearly whether the test result is a pass or fail.
The test should be considered satisfactory if the absolute pressure at the start of the 10 minute period is
within the limits specified by the manufacturer, and the rate of pressure rise during the test is not greater
than 1.3 mbar per minute.
The detailed method and pass / fail criteria are in DB 9804 Annex A.
41
2.3 Air detectors
Because effective removal of NCG is so important to the achievement of sterilizing conditions, porous
load sterilizers must be fitted with a means to detect, during every cycle, residual NCG that would prevent
the attainment of sterilizing conditions. These devices are commonly called air detectors. Vacuum
benchtop steam sterilizers must be equipped with either an air detector, or another system that is capable
of being validated, for detecting the presence of NCG.
Air detector system performance tests and function tests must also be performed quarterly and annually by
the TP(s) using independent, calibrated instruments. The results of these tests shall demonstrate the correct
functioning of any automatic air detection system fitted to the sterilizer.
42
Appendix 5 Regulation of transportable steam sterilizers
The medical devices directive and the medical devices regulations
Medical devices are regulated throughout Europe by the Medical Devices Directive, Council Directive 93/42
EEC – which has been transposed into UK law as the Medical Devices Regulations 2002 SI 2002 No 618.
These pieces of legislation regard sterilizers as medical devices if their manufacturer intends them to be used
for re-processing reusable medical devices.
Since the Directive and the Regulations came into force in June 1998, medical devices may be placed on the
market (i.e. supplied) only if the manufacturer claims the equipment meets the relevant ‘essential
requirements’ of the Directive. This means that the equipment is considered to be as safe as possible and is
fit for its intended purpose. Such devices must bear the CE marking to signify the claim of conformity and
this enables the product to be freely sold throughout the EC without further control.
To help manufacturers to meet the essential requirements, the European Commission has mandated
harmonised European standards to cover specific essential requirements. Products manufactured to comply
with such standards are automatically presumed to satisfy the relevant essential requirements. A European
standard for transportable steam sterilizers is in preparation (BS prEN 13060:2002) but it will be some time
before it is finished. Until then only the following standards are considered to be relevant to vacuum
benchtop steam sterilizers:
European Standard BS EN 554: 1994: Sterilization of medical devices - validation and routine control of
sterilization by moist heat.
British Standard 3970: Part 1: 1990: Sterilizing and disinfecting equipment for medical products;
specification for general requirements.
Further information on the Medical Devices Regulations is available from the MDA.
Operation.
There is a duty on the employer to ensure that anyone using, managing or supervising work equipment has
received adequate training (PSSR). The employer must also provide:
• all procedures and information needed for the equipment to be operated safely;
• any special procedures to be followed in the event of an emergency (e.g. failure of the door gasket);
• information on the dangers of forcing doors (either open or closed);
• instructions for checking door locking mechanisms in both the open and closed positions;
• instructions for checking that the chamber is not pressurised before attempting to open the door.
Maintenance
The equipment must be properly maintained and kept in good repair to prevent danger. The type and
frequency of maintenance will depend upon a number of factors including:
• the age of the equipment;
• how much it is used;
• reports of previous maintenance or inspection;
• any repairs or modifications that have been made;
• manufacturers instructions;
• reports of examinations made under the written scheme of examination.
43
The written scheme of examination.
The PSSR require pressure equipment to be inspected periodically and the owner's responsibilities
(summarised) are:
• to define the scope of the written scheme of examination. A Competent Person (pressure vessels) (CP)
must provide the written scheme, CPs may be found via the manufacturer or insurer);
• to ensure that the parts of the sterilizer defined in the written scheme are examined by a CP;
• not to allow the sterilizer to be operated unless a written scheme has been drawn up and certified as
suitable by a CP;
• to ensure that the system is properly maintained in good repair, so as to prevent danger.
The Competent Person (Pressure Vessels) has two principal duties under the Regulations:
Information on Competent Persons (Pressure Vessels) can be obtained from United Kingdom Accreditation
Service (see Appendix 6).
An Authorised Person (sterilizers) will be able to advise on the application of these Regulations to any
particular system.
44
Appendix 6 Sources of further information
Association of British Health-Care Industries (ABHI)
St Georges House
195-203 Waterloo Road http://www.abhi.org.uk
London Email: [email protected]
SE1 8WD Tel: 020 7787 3060 Fax: 020 7787 3061
NHS Estates
The Information Centre
1 Trevelyan Square Website: http://www.nhsestates.gov.uk/
Boar Lane Email: [email protected],
Leeds LS1 6AE Tel: 0113 254 7070 Fax: 0113 254 7167
45
DISTRIBUTION
This Device Bulletin should be brought to the attention of general practitioners, community healthcare workers, dental
practitioners, chiropodists and podiatrists, practice nurses, dental nurses, operating theatre staff, infection control teams,
risk managers and environmental health officers.
TECHNICAL ENQUIRIES
Enquiries concerning the content of this Device Bulletin should be addressed to:
Tel: 020 7972 8219 Fax: 020 7972 8106 Tel: 020 7672 8273 Fax: 020 7972 8106
Email: [email protected] Email: [email protected]
Copies of this Device Bulletin are free to health and social care providers and may be obtained on written request from:
Department of Health
PO Box 777
London
SE1 6XH Fax: 01623 724 524 Email: [email protected]
Otherwise, copies of the bulletin at a charge of £25 per copy may be obtained from:
Our website lists all current Device Bulletins and safety warnings: http://www.medical-devices.gov.uk