LBBBIAN Lab Procedure
LBBBIAN Lab Procedure
The demands on the analytical chemists have increased over the recent years. It is expected that the
results must have a high level of confidence, high precision, update on the recent developments in the
analytical techniques, instrumentation and automation and an increase in the number of samples to be
analyzed.
The analyst always refers to the OFFICIAL METHODS OF ANALYSIS as a major reference for
procedures to be used in every analytical work. The ASSOCIATION OF OFFICIAL ANALYTICAL
CHEMISTS major contribution to analytical science has been to bring the collaborative study technique
for the validation of analytical methods to a high degree of perfection.
It aims to establish the performance characteristics of the process in terms of accuracy, precision,
sensitivity, range, specificity, limit of detection, limit of reliable measurement, selectivity and practicality.
The analysis of very carefully prepared samples aims to establish not only the performance
characteristics of the method but of the analyst or the lab as well.
FOR ANALYTICAL WORK, it is important that the analyst must be able to develop and refine the basic
skills and lab techniques learned from general chemistry labs.
Chemicals are classified based on the purity or the results of the assay as reported by the manufacturer
Reagent Grade Chemicals (AR Grade)
Conform to the minimum standards or specifications set by the Reagent Chemical Committee of
the American Chemical Society (ACS)
Used for analytical work
Container Label contains information regarding the maximum limits of impurity
a. Measurement of Mass
Types of Balances
An analytical balance is an instrument used for determining the mass with a maximum
capacity that ranges from 1 g to a few kg and a precision at maximum capacity of at
least 1 part in 105 to 1 part in 106
Analytical balance is used for highly accurate measurements; 1 part in 105 of its
maximum capacity or even 1 part 106 of its full capacity.
A laboratory balance or auxiliary balance is rugged and with less sensitivity than an
analytical balance. Ex. platform balance
Temperature Effects
Convection currents within the balance case exert a buoyant effect on the pan and object
Warm air trapped in a closed container weighs less than the same volume at a lower
temperature
Both effects cause the apparent mass of the object to be low and can amount to as much as 10
to 15 mg.
Heated objects must always be cooled to RT before weighing
Drying or Ignition to Constant Mass a process in which a solid is subjected to several cycles of
heating, cooling and weighing steps until its mass becomes constant to within 0.2 to 0.5 mg.
Hot objects look the same at cold ones.
Weighing Bottles are used for drying, storing and weighing solids
Weighing by Difference.
The weighing bottle with the sample is weighed initially. The desired quantity of the
sample is obtained and transferred into a clean, dry container. The weighing bottle
and the remaining contents are weighed again. The decrease in the mass of the two
measurements represents the mass of the obtained sample.
Weighing by Addition
The mass of the empty weighing bottle is measured. The desired quantity of the
sample is added to the weighing bottle. The weighing bottle and the sample are
weighed together. The increase in the mass of the two measurements represents the
mass of the obtained sample.
b. Measurement of Volume
SI unit for volume is cubic decimeter or dm3.
1 liter (L) = 1 dm3 1000 mL = 1 L
1 mL = 1 x 10-3 L 1 L = 1 x 10-6 L = 1 x 10-3 mL
Characteristics of an Experiment
1. Must have a clear and cognitive purpose
2. It is selective wherein a portion of the Universe is selected for study (system). The variables of
the system must be defined.
3. It must have a beginning, middle and an end. And the change in the system is deliberately
provoked or carried out.
4. The change must be capable of being
(b) observed using either direct perception (or direct method) using the eyes or indirect
measurement involves the use of an instrument
(c) interpreted by the observer and recognize the change as a significant difference in a
variable property of the system
(d) described by observer and recorded in such as way that it can be communicated to
another scientist for testing or reproduction
5. The system and the change should be reproducible and testable by other scientists
Collection of Data
Best done in a sequential manner that is from lowest to highest or vice versa
Establish the relationship between the dependent and independent variables
If the relationship is linear then collect data such that there is equal spacing between
values
If the relationship exhibits an inflection point then y is slowly changing with x. the x
values may be more widely spaced However near the inflection point the x values must
be chose close together so that the changes in y are kept more or less constant between
all changes
If the data shows a maximum then the data can be more widely spread near the base of
the peak where y is changing very slowly with x. Then it is more closely spaced on the
sides where y is changing rapidly. This allows the location of the maximum more
precisely.
Keeping records.
Resist temptation to perceive observations to support the hypothesis or distort unsupportive
observation to minimize disagreement. Exercise absolute objectivity and avoid prejudice,
bias and fraud.
Bias interferes with the testing of the hypothesis.
A chemist seeks a relationship between a property and what is observed. It is necessary to report not
only the quantitative relationship and the value of a calculated property but also some measure of the
reliability of the measurements and subsequent calculations.
Therefore one must report the calculated value and the error d .
Measurements are repeated in order to test the reproducibility of the experiment and gain confidence in
the accuracy of the results.
The spread or the range is used to describe the precision of a set of replicate measurements is the
difference between the highest and the lowest result. It results directly from an accumulation of all
random uncertainties in the experiment.
The cumulative influence of these variables is responsible for the observed scatter of results around
the mean.
3. GROSS ERROR
Occurs occasionally, often large and may cause a result to be either high or low.
It is often a product of human errors
For example, loss of precipitate during weighing will cause the results to decrease.
It leads to outliers. An outlier is an occasional result in replicate measurements that differs
significantly from the rest of the results.
errors during sample treatment is the major source of error in chemical analysis.
Personal errors due to judgment of color changes during visual titration causes overstepping of
end point
Estimation of values between scale divisions
b. Proportional Errors
Increase or decrease according to the size of the sample taken for analysis. The absolute
error varies with sample size but the relative error stays constant with changing sample size.
Due to the presence of interfering contaminants in the sample
Systematic instrument errors can be found and corrected by calibration. Periodic calibration of the
equipment is desirable because the response of the instrument changes with time, overuse, etc.
Error is the difference between a measured value and the true value (the known or accepted value). It
is also the estimated uncertainty in a measurement or experiment
Errors are caused by faulty calibration or standardization, random variations and uncertainties in results
Measurements are always influenced by uncertainties and these cause the results to scatter.
Precision is the closeness of results among one another obtained in exactly the same way;
reproducibility of measurements
Deviation, d = xi - x
2
xi x
Standard deviation, s =
n 1
s
% Coefficient of Variation, CV = 100 (Note: The CV is also referred to as %RSD)
x
s
% Relative Standard Deviation, RSD =
100
x
Introduction:
Chemicals and reagents are graded and classified based on the relative amounts of the analyte and
the impurities present in the sample. The composition of the representative sample of the reagent is
expressed in terms of its % analyte present. Old and unlabelled reagent bottles must be analyzed prior to
use in order to determine its purity.
The sample to be analyzed contains Na2CO3 and an inert impurity. For this kind of sample, the
direct titration of the sample with a standard solution of hydrochloric acid until the methyl orange end
point is sufficient. At the methyl orange end point, the titration reaction is shown below:
Na2CO3 + 2HCl 2NaCl + 2 H2O + CO2
Experimental Procedure
1. Preparation of 0.1M HCl
Pipet 8.5 ml of concentrated HCl to a clean glass-stoppered bottle containing approximately
900 mL of distilled water. Stopper the bottle and mix the solution thoroughly. Label the bottle.
2. Standardization of 0.1M HCl against the Primary Standard Sodium Carbonate, Na2CO3
Label three dry and clean 250-mL Erlenmeyer flasks as trial 1, trial 2 and trial 3. A reagent
bottle that contains a previously dried primary standard grade Na2CO3 (M. M. = 105.99 g/mole)
will be provided for the class. From this reagent bottle, obtain approximately 1 g of the Na2CO3 and
transfer it into a clean and dry 50-mL beaker.
Triplicate samples of accurately weighed primary standard grade or pure Na2CO3 must be
obtained for the standardization of the HCl solution. Read and understand the procedure described
below.
With the use of an analytical balance, accurately weigh the beaker containing the pure Na2CO3.
Record the value on the laboratory notebook under the heading trial 1 across the description, mass
of beaker plus Na2CO3. Obtain a portion of the Na2CO3 and transfer it into the first Erlenmeyer
flask. Reweigh the beaker and the remaining contents. Record the value under the heading mass
of beaker less Na2CO3. The difference must lie within the range of 0.10 and 0.15 g.
Repeat the same procedure until all three flasks contain pure sodium carbonate, Na2CO3.
Calculate the mass (g) Na2CO3 obtained for each trial.
Dissolve the Na2CO3 in about 40-60-ml distilled water and add 2 to 3 drops of methyl orange
indicator solution. With a white board under the flask, the solution must appear yellow in color.
Rinse the buret with about 20 mL of the prepared HCl solution and discard the rinsing down the
sink. Pour HCl solution into the buret and fill in the air gap. You may need to pour more HCl into
the buret. Wipe the buret to remove any excess acid solution.
Record the initial buret reading and titrate the solution containing the sodium carbonate with the
HCl solution until the methyl orange end point. This is indicated by the appearance of a permanent
faint pinkish orange cololr that will persist for about 30 seconds.
Note: Prepare 1.0 g KHP in 100.0 mL distilled water and add 2 to 3 drops of the methyl orange
indicator solution. This is the color that must be observed at the end point of the titration.
Repeat the titration with the other two samples, recording all data in the Lab Journal.
Introduction
The total acid content of a sample of vinegar is readily determined by the direct titration with a
standard solution of a strong base. The acidity of vinegar is usually expressed in terms of the percentage
of acetic acid, CH3COOH, present in the sample. The analyte, acetic acid, is a monoprotic weak acid.
Although the chemical formula, CH3COOH, indicates that there are four (4) hydrogen atoms, only one is
considered acidic or ionizable. The acidic H is the one contained in the carboxylic acid group, COOH.
This hydrogen is referred to as the titratable H because it is the one that will react with the titrant
solution, NaOH.
The composition of most liquid commercial products is expressed in terms of mass to volume
percent. This concentration unit is differentiated from the mass percent by writing the symbol (w/v) or
(wt/vol) after the value of the composition. Commercially available vinegar products have a
concentration of 4% to 6% (w/v) CH3COOH. This means that the vinegar contains 4-6 g acetic acid per
100 mL of the vinegar sample.
The vinegar sample will be analyzed using titration of aliquot portions. An aliquot is a portion
taken from the sample that is a representative of the bulk. The aliquot is assumed to have the same
composition as the bulk and is homogeneous. This means that the concentration of the aliquot is the
same as the concentration of the solution from which the aliquot was obtained.
An aliquot portion of the vinegar (not the entire vinegar present in the bottle) will be obtained and
diluted to an accurate volume with the use of the volumetric flask. From this dilute vinegar solution,
triplicate portions of equal volumes (aliquots) will be measured and titrated with the standard solution of
the strong base. Observe that what is being titrated is just a fraction (portion) of the entire bottle.
However, after incorporating the dilution factor, the concentration of the aliquot portion will be the same
as what will be obtained if the entire vinegar in the bottle was titrated.
Experimental Procedure
1. Preparation of 0.1 M NaOH
Boil approximately 1L of distilled H2O. Be sure to protect the water from the atmosphere as it
cools. Measure 4 to 5 mL of a previously prepared 50% NaOH and add it to the distilled H2O.
Securely cover the bottle with a polyethylene cap and mix the solution thoroughly. Label the bottle.
Protect the solution from unnecessary exposure to the atmosphere.
Important! The physical appearance of the dilute NaOH solution is identical with
distilled water. LABEL the beaker with 0.1 M NaOH.
Fill the buret with the prepared NaOH solution. Record the initial buret reading and titrate the
solution containing the dissolved potassium hydrogen phthalate with the NaOH solution until the
phenolphthalein end point. This is indicated by the appearance of a permanent faint pink color
that will persist for about 30 seconds. Recorde the Repeat the titration with the other two
samples, recording all data in the Lab Journal.
Report the results of the analysis in terms of % (w/v) CH3COOH in the vinegar sample
Notes:
1. Do not handle NaOH pellets with fingers. This substance is corrosive.
2. Never use a glass-stoppered bottle to store the prepared NaOH solution. The NaOH will
cause the glass to freeze making it difficult, or sometimes impossible to open the bottle. It is
for the same reason that one should not use a buret with a glass-stopcock for titration
involving NaOH solution.
3. Store the remaining standard NaOH solution for succeeding experiments.
4. Dry KHP in the oven at 105 110oC for two hours or more.
5. The mass (g) of the weighed primary standard must lie within the same value until the first 2
significant figures.
6. To avoid spattering, be sure to have the tip of the buret near the surface of the solution being
titrated then add NaOH solution. Avoid splashing drops upon the sides of the flask.
7. Always use a volumetric or transfer pipet when measuring aliquot portions.
Introduction:
When choosing the best visual indicator to use for a titrimetric analysis, one must be
knowledgeable of the changes in the properties of the titration mixture as the reaction progresses. In an
acid-base reaction, this is determined by monitoring the change in the pH of the reaction mixture as the
titrant is added. The visual indicator is chosen so that the change in color of the indicator occurs at the
stage very close to the equivalence point of the reaction. In the absence of a satisfactory visual
indicator, graphical determination of the end point is the best way to determine the equivalence point.
During the titration of the acid or base, the potential between two electrodes dipped in the analyte
solution is monitored as the titrant is added. One of the electrodes is an indicating electrode, which is a
glass electrode, normally the standard calomel electrode (SCE), with an invariant potential. Nowadays,
combination electrodes are widely used, where both indicating and reference electrodes are combined in
one compartment.
The equivalence point in the titration is determined by plotting, pH vs the volume of titrant added.
A sharp break in the titration curve is observed in the vicinity of the equivalence point. The equivalence
point in the titration curve can be determined by extending 2 straight lines (A and B) along the lower
and upper curve that comes before and after the sharp rise or inflection curve. Two parallel lines are
drawn along y-axis before and after the inflection curve. A third line is drawn through the midpoint of
the two parallel lines. The equivalence point is the point where the middle line intersects the titration
curve. The corresponding volume is the equivalence point volume.
14.000
12.000
10.000
8.000
pH
6.000
4.000
2.000
0.000
1.00
6.00
11.00
16.00
21.00
26.00
31.00
36.00
41.00
46.00
51.00
56.00
61.00
66.00
71.00
Experimental Procedure:
1. Use of the pH meter.
(a) Wash the electrode with distilled water each time it is to be used.
(b) Squirt the electrode with distilled water with the use of a wash bottle. Always wipe the
electrode with tissue paper using light strokes.
(c) Before using, calibrate the pH meter in a buffer solution of pH 7 for upper calibration or pH 4
for lower calibration.
Prepare a potentiometric titration curve by plotting pH against volume of base titrant. Identify the
end point of the titration and the volume (mL) of the titrant needed to reach the end point. Shade the
area where the phenolphthalein end point occurred.
Report the molar concentration of the unknown solution of the strong acid for each trial and make a
comparison of the concentration for the 2 methods.
Experimental Procedure:
4. Determination of the Acid Dissociation Constant, Ka of a Weak Acid
Note: Each group will be assigned a specific weak acid as unknown.
B. Titration of the Dilute Acid Unknown Using Phenolphthalein as the Visual Indicator
a. Pipet a 25 ml aliquot portion from the dilute acid unknown and transfer this into a 100ml
beaker. Measure the initial pH reading of the solution. Add 2-3 drops of phenolphthalein
indicator solution and measure the pH reading. Do you observe a change in the pH
reading?
b. Titrate the solution until the phenolphthalein end point. Record the buret reading.
Measure the pH of the resulting mixture. Repeat the titration for a total of 3 trials using
25-mL aliquot portions. Results must agree to at least 0.1 mL. Calculate the average
volume of NaOH needed to reach the phenolphthalein end point.
c. Pipet a fourth 25-mL aliquot portion from the dilute acid unknown. Measure the initial pH
reading.
d. Titrate the solution until only one-half of the average volume of the NaOH solution
needed to reach the phenolphthalein end point has been added. Measure the pH of the
mixture.
e. Repeat procedures c and d for two more additional trials.
Prepare a potentiometric titration curve by plotting pH against volume of base titrant. Identify the
end point of the titration and the volume (mL) of the titrant needed to reach the end point. Shade the
area where the phenolphthalein end point occurred.
Compare the pKa obtained from the two methods. Check your experimental values with the value
obtained from the Table of the Acid Dissociation Constant.
Introduction:
Potassium permanganate has a very high oxidizing power with a reduction potential of 1.507 V.
In titrimetric analysis, it is frequently used as titrant due to its intense coloring power, which requires no
auxiliary indicator. When the titration is carried out in acidic medium, the permanganate ion is reduced
to the colorless manganous ion (Mn2+) in aqueous solution. A drop of excess permanganate turns the
solution into pink color, which could last for about 15 seconds. This color is taken as the endpoint of the
titration. The solution readily decomposes to brown MnO2 upon reaction with heat, sunlight or organic
contaminant.
Experimental Procedure
1. Preparation of a 0.02 M Potassium Permanganate.
Weigh approximately 0.8-g of a good grade potassium permanganate and place it in a 250-ml
beaker. Dissolve the salt by adding 50-ml of water and stirring. Decant the solution into a large
beaker and add 50-ml of additional water to dissolve the crystals remaining in the first beaker.
Repeat this procedure until all crystals are dissolved. Dilute the solution until the total volume of
distilled water added is about 250 mL. Transfer the solution to a glass stoppered bottle and label
properly.
Thus 1 mL of a sample of hydrogen peroxide with the label 100-volumes will yield 100 mL of O2
measured at standard temperature and pressure (STP). The label 20-volumes refers to a 6% (w/v)
H2O2, 40-volumes is also 12% (w/v) H2O2, and 100-volumes hydrogen peroxide refers to 30% (w/v)
H2O2.
Transfer a 25 mL aliquot of a 20-volume solution to a 500 mL volumetric flask and dilute to the
mark. Label this as dilute sample solution. Measure a 25 mL aliquot portion of the dilute sample
solution and transfer into an Erlenmeyer flask. Add 20 mL of distilled water, 20 mL of dilute H2SO4
solution (1:5 dilution) and titrate with 0.02 M KMnO4 to the first permanent faint pink color. Repeat
the titration to two consecutive determinations and the trials should agree within 0.1 mL.
Introduction:
Due to the relatively weak strength of I2 as an oxidizing agent, there are limited substances that
can be titrated with it. Other titration methods that involve the substance I2 is the indirect titration of I2
with sodium thiosulfate, Na2S2O3, and back-titration of I2 with sodium thiosulfate. In the indirect
titration of I2, the preliminary step involves the formation of I2 from the reaction of the oxidizing agent
with KI. The liberated I2 is then titrated with the sodium thiosulfate solution.
Experimental Procedure:
1. Preparation of 0.1 M Sodium Thiosulfate
Boil 500 mL of distilled water for 2 minutes and let cool. Weigh out 12.50 g Na2S2O3 5H2O
and dissolve the crystals in the previously boiled distilled H2O. Add 0.1 g Na2CO3 as a preservative.
Introduction:
The ability of I2 as an oxidizing agent, there are limited substances that can be titrated with it.
Another titration method involving the substance I2 is the indirect titration of I2 with sodium thiosulfate,
Na2S2O3. of the back-titration of I2. In the indirect titration of I2,, preliminary step involves the
formation of I2 from the reaction of the oxidizing agent with KI. The liberated I2 is then titrated with the
sodium thiosulfate.
Experimental Procedure:
1. Preparation of 0.025 M I2
Transfer 10 mL of 0.5 M I2 solution into an amber bottle (1-L capacity) and add 200 mL of
distilled H2O. Cover the bottle and mix the solution thoroughly.
2. Standardization of 0.025 M I2
a. Preparation of Ascorbic Acid Standard Solution
Weigh accurately 0.25 to 0.45 g pure ascorbic acid and dissolve in 50 mL of distilled H2O.
Transfer into a 250-mL volumetric flask and dilute to the mark. Cover the flask and mix
thoroughly.
Report the concentration of the ascorbic acid solution in molarity.
Discussion:
Complexometric titrations have been carried out successfully on nearly all-common cations. These
reactions virtually replaced the former tedious gravimetric analyses for many metals in a variety of
samples. The most popular complexing agent is the EDTA (or H4Y, ethylenediaminetetraacetic acid). It
reacts with metal ions on a 1:1 stoichiometric mole ration.
For convenience, the free acid form of EDTA is often abbreviated as H4Y and the EDTA anion as
Y4-. Eriochrome Black T (EBT) is usually represented in abbreviated form as a tri basic acid, HIn3. An
NH3-NH4Cl buffer of pH 9 to 10 is often used for the metals that form complexes with ammonia. At pH
10, the predominant form of Eriochrome Black T is the blue HIn2 form. The reaction that results in the
color change can be written as
MIn- + HY2 MY2 + HIn2
(red) (blue)
In order for the color change to occur at the proper time, the stability of MIn- complex must be less
than that of MY2 so that the metal M releases the indicator when only a slight excess of EDTA is added.
If the sample titrated does not contain Mg, some Mg salt can be added to EDTA before this
solution is standardized. The titrant then (pH 10) is a mixture of MgY2 and Y4-. As this is added to the
solution containing Ca2+, the more stable CaY2 is formed, liberating Mg2+ to react with the indicator and
form the red MgIn-. After the calcium has been used up, additional titrant convenrts MgIn- to MgY2- and
the indicator reverts to the blue HIn2 form.
The complex formed between calcium ion and Eriochrome Black T is not very stable and the color
change occurs prematurely in the titration of Ca2+ with EDTA. Standardizing the EDTA solution against
standard calcium ion solution using conditions identical to those used for the unknown can reduce the
titration error. However, magnesium forms a stronger complex with the indicator than calcium and a
proper end point is obtained in an ammonia buffer of pH 10.
The total hardness of water due to calcium and magnesium ions can be determined by direct
titration with H4Y (ethylenediaminetetraacetic acid) using Eriochrome Black T or calmagite indicator.
The structure of the H4EDTA and the anionic form at pH 10 of the ligand is
HOOCH2C CH2COOH -
OOCH2C CH2COO
-
NCH2CH2N NCH2CH2N
- -
HOOCH2C CH2COOH OOCH2C CH2COO
-4 -4
H4EDTA (or H4Y) EDTA (or Y )
The EDTA anion is potentially a hexadentate ligand, which may coordinate with a metal ion through its
two nitrogens and four carboxylate groups.
Experimental Procedure:
1. Preparation of 0.01M EDTA solution:
Transfer about 2-g of disodium dihydrogen EDTA dihydrate and 0.1-g of MgCl2·6H2O into a
clean 400-ml beaker and dissolve the solids in water. If the solution appears turbid, add NaOH
solution until the solution is clear. Transfer the solution into an amber bottle, add enough distilled
water to make about 0.50 L and mix thoroughly. Label the contents of the bottle as 0.01 M EDTA.
Report the concentration of the EDTA solution in terms of molarity and the Ca titer (mg Ca /mL
solution).
3. Determination of Ca in Drugs
The Lab Instructor will provide you with an OTC drug sample that contains calcium lactate only
and the some information regarding the composition of the drug. From this data, you will make the
detailed procedure for the determination of Ca in the drug sample. Based on your proposal, the
volume of the titrant that will be consumed for each analysis must lie within the range of 10 to 40
mL. Show the calculations to your Lab Instructor before proceeding.
Measure about 50 mL of distilled water into a 250-mL beaker and dissolve one or more tablets.
If the sample is not very soluble in distilled H2O, add about 5 mL of 0.5 M HCl and this will improve
its solubility. Transfer the dissolved tablet into a volumetric flask and dilute to the mark. Label the
mixture as dilute tablet solution (DTS). Determine the volume of the aliquot portion of the DTS that
must be measured so that
With the use of a volumetric pipet, obtain an aliquot portion of the DTS and transfer it into a 250-
ml Erlenmeyer flask. To the first sample add 1.0 ml of the buffer solution and 5 drops of the
indicator solution. Titrate with the standard EDTA solution to a color change of wine-red to pure-
blue. Perform the titration in triplicate measurements.
Report the composition of the sample in terms of value indicated in the label of the supplement, i.e.
as mg calcium lactate, Ca(CH3CH(OH)CO2)2, per tablet for Ca supplements or mg Ca per tablet.
The absorbance of a solution is equal to the product to the molar absorptivity, a, the cell path
length of the container and the concentration of the absorbing species. The application of Beer s law to
the quantitative analysis is broad. The information that can be obtained varies from the source of the
electromagnetic radiation, physical state of the absorbing species and the ability of the instrument to
isolate the emr absorbed or emitted by the absorbing species.
Experimental Procedure
1. Preparation of the Standard Solution
Prepare 250 mL of 0.01 M MnSO4 standard solution from MnSO4·H2O.
Transfer 10.0 mL of the 0.01 M MnSO4 solution into a a250-mL beaker and add about 100
mL of distilled water. Add 10 mL 85%H3PO4 and 0.5 g KIO4. Heat the mixture gently and allow
the solution to boil for about 3 minutes. The color of the mixture will change to purple. Allow the
solution to cool and transfer the contents quantitatively to a 250-mL volumetric flask. Rinse the
beaker with about three 20-mL portions of distilled H2O. Transfer the rinsing into the volumetric
flask. Dilute to the mark with distilled H2O and cover. Mix the solution thoroughly. Label the
solution
Calculate the concentration of permanganate standard solution (PSS). Express your answer in
terms of the moles Mn per Liter of solution.
Calculate the molar concentration of Mn present in each of the five standard solutions.
4. Beer s Law
Note: Perform this portion of the experiment only when the sample is ready.
Adjust the wavelength setting of the Spectronic 20 to the working wavelength. With distilled H2O
as the blank, set the absorbance of the blank (distilled H2O) to zero. Measure the absorbance of the
five standard solutions. Do 3 readings for every standard solution. Beer s Law is obeyed if the
absorbance reading increases with increasing concentration of the standard solution.
Calculate the average absorbance of each standard solution. Plot the absorbance (y-axis) versus the
concentration of the standard solution. Use the methods of least squares to determine the best straight
line for the data points. Report the equation of the line and the regression coeffient, r.
5. Analysis of Sample
C. Preparation of the Sample
Weigh the unknown sample and transfer into a 100 mL beaker. Add 50 mL of distilled H2O,
10 mL 85%H3PO4 and 0.5 g KIO4. Heat the mixture gently and allow the solution to boil for about
3 minutes. Allow the solution to cool and then transfer the colored solution into a 100-mL
volumetric flask. Rinse the beaker three times using 10-mL portions of distilled H2O. Transfer the
rinsing into the volumetric flask and dilute to the mark. Cover the flask and mix the solution
thoroughly. Label the solution as unknown solution (US)