OP 3D User and Installation Manual
OP 3D User and Installation Manual
ENGLISH
212972 rev. 6
0.805.4913
1 Disclaimer................................................................... 6
2 Introduction................................................................ 7
2.1 ORTHOPANTOMOGRAPH™ OP 3D.............................................................. 7
2.2 Intended use......................................................................................... 7
2.3 Intended user profile.............................................................................. 7
2.4 Associated documentation....................................................................... 7
2.5 Abbreviations......................................................................................... 8
2.6 Signal words.......................................................................................... 8
2.7 Disposal and recycling............................................................................ 8
2.8 Warnings and requirements..................................................................... 9
2.8.1 Warnings and precautions for use..................................................9
2.8.2 Warnings and precautions for installation........................................9
2.8.3 Connection requirements............................................................ 10
2.8.4 Device modifications.................................................................. 10
3 Overview................................................................... 11
3.1 Main parts........................................................................................... 11
3.2 Patient positioning lights........................................................................12
3.2.1 Positioning light locations............................................................12
3.2.2 Panoramic lights........................................................................ 12
3.2.3 3D lights.................................................................................. 13
3.3 Accessories.......................................................................................... 13
3.4 Other detachable parts..........................................................................14
3.5 Emergency stop switch..........................................................................15
4 User Interfaces......................................................... 16
4.1 Graphical User Interface (GUI)............................................................... 16
4.1.1 GUI Overview............................................................................16
4.1.2 Panoramic view......................................................................... 18
4.1.3 3D view....................................................................................20
4.1.4 Settings menu...........................................................................25
4.2 Positioning panel.................................................................................. 28
4.3 Status indicator light.............................................................................28
5 Imaging programs.................................................... 29
5.1 Panoramic programs............................................................................. 29
5.2 3D programs........................................................................................31
5.2.1 FOV Sizes (H x D)..................................................................... 31
5.2.2 3D Resolutions.......................................................................... 32
ORTHOPANTOMOGRAPH™ OP 3D 3
7 Maintenance.............................................................. 43
7.1 Cleaning and decontamination................................................................43
7.2 Quality Control..................................................................................... 44
7.2.1 PAN QC.................................................................................... 44
7.2.2 3D QC...................................................................................... 45
7.3 Calibrations for the user........................................................................ 48
7.3.1 When to calibrate the device.......................................................48
7.3.2 Preparation for Calibration.......................................................... 48
7.3.3 Calibration procedure................................................................. 50
7.4 Annual maintenance..............................................................................52
8 Troubleshooting........................................................ 53
9 Installation................................................................54
9.1 Device package.................................................................................... 54
9.2 Pre-installation requirements.................................................................. 54
9.2.1 Installation location requirements................................................ 54
9.2.2 Space requirements................................................................... 55
9.2.3 Fixing hardware and tools...........................................................56
9.3 Back panel connectors...........................................................................57
9.4 Installing the device..............................................................................57
9.5 Attaching the covers............................................................................. 65
9.6 Configuring the communication link to the workstation.............................. 70
9.6.1 Connection options.................................................................... 70
9.7 Finishing the installation........................................................................ 72
9.8 Complete device calibration................................................................... 73
9.8.1 Accessing the calibrations........................................................... 73
9.8.2 Device Movement Calibration...................................................... 74
9.8.3 Z-Movement Limits.................................................................... 74
9.8.4 X-Ray Generator Calibration........................................................ 75
9.8.5 3D Pixel Calibration....................................................................75
9.8.6 Device Geometry Calibration....................................................... 76
9.8.7 Collimator Calibration................................................................. 77
9.8.8 PAN Pixel Calibration.................................................................. 77
9.8.9 3D Geometry Calibrations........................................................... 78
9.8.10 PAN QC (optional)....................................................................79
9.8.11 3D QC.................................................................................... 81
9.8.12 Panoramic beam size verification (optional)................................. 83
9.9 Additional installation procedures............................................................84
9.9.1 Remote exposure switch installation............................................. 84
9.9.2 Mains power configuration.......................................................... 85
9.9.3 Changing the fuses.................................................................... 86
9.9.4 Additional wall bracket installation............................................... 86
9.9.5 Exhibition stand installation.........................................................88
9.9.6 Collimator adjustment................................................................ 88
9.9.7 Sensor adjustment.....................................................................90
9.9.8 Patient positioning light adjustment............................................. 91
9.9.9 Device configuration...................................................................93
9.9.10 Verification.............................................................................. 96
9.9.11 Repacking the device................................................................96
10 Technical data......................................................... 98
10.1 Technical specifications........................................................................ 98
4 ORTHOPANTOMOGRAPH™ OP 3D
ORTHOPANTOMOGRAPH™ OP 3D 5
1 Disclaimer
ORTHOPANTOMOGRAPH™ OP 3D User and Installation manual, 212972 r6.
Copyright © 2017 by PaloDEx Group Oy. All rights reserved.
ORTHOPANTOMOGRAPH™, OP™, ORTHOselect™, QUICKcompose™, SMARTview™,
ORTHOfocus™ and Low Dose Technology™ are either registered trademarks or
trademarks of KaVo Kerr Group Finland in the United States and/or other countries.
KaVo™ is either registered trademark or trademark of Kaltenbach & Voigt GmbH in the
United States and/or other countries.
All other trademarks are property of their respective owners.
Documentation, trademark and the software are copyrighted with all rights reserved.
Under the copyright laws the documentation may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form
in whole or part, without the prior written permission of PaloDEx Group Oy.
The original language of this manual is English.
PaloDEx Group Oy reserves the right to make changes in specification and features
shown herein, or discontinue the product described at any time without notice or
obligation. Contact your PaloDEx Group Oy representative for the most current
information.
The manufacturer has no liability for consequential damage, personal injury, loss,
damage or expense directly or indirectly caused by the use of the product. No agent,
distributor or other party is authorized to give warranty or other liability on behalf of the
manufacturer with respect to its products.
Manufacturer:
PaloDEx Group Oy
Nahkelantie 160
FI-04300 Tuusula
FINLAND
www.kavo.com/downloads
6 ORTHOPANTOMOGRAPH™ OP 3D
2 Introduction
2.1 ORTHOPANTOMOGRAPH™ OP 3D
The ORTHOPANTOMOGRAPH™ OP 3D (later called device) is a dental X-ray device
producing high quality digital images of dentition, TM-joints and skull. To take images,
you need a suitable workstation connected to the device and a dental imaging software
to capture and manage the images.
3D CBCT Imaging
Volume Height x Diameter (H x D)
• 5 x 5 cm; implant and other single tooth imaging
• 6 x 9 cm; single jaw imaging
• 9 x 11 cm; whole dental area imaging (both jaws)
• 9 x 14 cm (optional); both jaws including TMJ
• FOV position and height fine-tuning is possible according to scout images
CAUTION! USA only: Federal law restricts this device to sale by or on the order
of a dentist or other qualified professional.
ORTHOPANTOMOGRAPH™ OP 3D 7
2.5 Abbreviations
3D Three dimensional
ALARA As Low As Reasonably Achievable
CBCT Cone Beam Computed Tomography
D Diameter
DAP Dose Area Product
FH Frankfort-Horizontal
FOV Field Of View. The cylindrical 3D volume that is reconstructed by the
device.
GUI Graphical User Interface
H Height
LDT Low Dose Technology™
MAR Metal Artifact Reduction. Reduces the effect of metal and other dense
radiopaque objects, that typically create artifacts which are seen as
stripes and shadows.
MDD Medical Device Directive 93/42/EEC
PAN Panoramic
QC Quality Control
ROI Region Of Interest. The anatomical area or region that you are interested
to examine.
TMJ Temporomandibular joint
At least the following parts of the device should be re-cycled according to local and
national regulations regarding disposal of non-environmentally friendly materials:
• Tubehead (oil)
8 ORTHOPANTOMOGRAPH™ OP 3D
ORTHOPANTOMOGRAPH™ OP 3D 9
• To avoid the risk of electric shock, the device must only be connected to a mains
power outlet with protective earth.
• The device shall not be connected to an extension cord.
• Ensure the proper tightness and attachment of all of the screws and bolts when
installing the device.
• All device covers must be properly installed before handing the device to the user.
• Ensure the electrical safety of the device by inspecting the proper grounding of user
touchable metallic parts before handing the device to the user.
10 ORTHOPANTOMOGRAPH™ OP 3D
3 Overview
ORTHOPANTOMOGRAPH™ OP 3D 11
12 ORTHOPANTOMOGRAPH™ OP 3D
3.2.3 3D lights
B. Midsagittal light
C. Horizontal light, top of FOV
D. Horizontal light, bottom of FOV
3.3 Accessories
Chin Rest Bite Block Lip Support (3D,
TMJ and edentulous
patients)
ORTHOPANTOMOGRAPH™ OP 3D 13
14 ORTHOPANTOMOGRAPH™ OP 3D
Pressing the emergency stop switch immediately terminates the imaging and all device
movements.
NOTICE! An interrupted imaging process cannot be resumed. A new image needs
to be taken.
To release the emergency stop switch, rotate it clockwise.
NOTICE! Make sure that the emergency stop switch is not pressed down when
you start the imaging process.
The Graphical User Interface (GUI) has indication if the emergency stop is active and the
device status indication lights are blue.
ORTHOPANTOMOGRAPH™ OP 3D 15
4 User Interfaces
16 ORTHOPANTOMOGRAPH™ OP 3D
Test mode
Indicates if Test mode is active.
Press Test mode icon to disable/enable device radiation production.
This mode can be used for example to demonstrate the device
movements.
You can also use the positioning panel to activate the Test mode.
Patient size
Indicates the currently selected patient size preset.
Press the Patient size selection icon to open a list of patient sizes;
Small, Medium & Large.
End Study
Press End Study icon to finalize the imaging procedure and return to
the main view.
ORTHOPANTOMOGRAPH™ OP 3D 17
18 ORTHOPANTOMOGRAPH™ OP 3D
NOTICE! Panoramic image previews are not shown in full resolution on GUI.
Zoom slider
You can zoom the preview image in/out using the zoom slider.
While zoomed in, you can click and drag to scroll the image.
Brightness slider
You can adjust the brightness of the shown preview image using the
brightness slider.
Contrast slider
You can adjust the contrast of the shown preview image using the
contrast slider.
OK button
Press OK button to close the image preview.
NOTICE! None of the adjustments done to the preview image are saved to the
image.
ORTHOPANTOMOGRAPH™ OP 3D 19
4.1.3 3D view
3D imaging program selection
Dental chart is used to select the Region of Interest (ROI) for the 3D
scan.
The FOV changes automatically according to the selections.
Press teeth, jaw and TMJ icons to select which parts of the dentition
are studied:
• Single tooth (1)
• Range of teeth (2)
• One jaw and all teeth in it (3 or 4)
• Both jaws and all teeth in them (3 and 4)
• TMJ (5 or 6)
• Both TMJ's (5 and 6)
• Whole dentition (3, 4, 5 and 6)
NOTICE! Both TMJs are selectable only if 9x14 FOV size is
available.
You can select the whole dentition by double-clicking any single
selection (tooth, jaw or TMJ). Undo the "select all" by double-clicking
the selection to return to the previous state.
Clear all selections by double clicking the gray area on the lower
corners of the dental chart. Undo the "clear all" by double-clicking
the gray area to return to the previous state.
20 ORTHOPANTOMOGRAPH™ OP 3D
FOV size
Indicates the currently selected FOV size (H x D).
Press the FOV size icon to open a list of the available 3D FOV sizes;
5 x 5, 6 x 9, 9 x 11 and 9 x 14 (optional).
Press on the listed FOV size icon to activate it. The skull on the
bottom of the GUI also illustrates an approximate coverage of the
selected FOV.
Image resolution
Indicates the currently selected resolution.
Press the image resolution selection icon to open a list of available
image resolutions;
ORTHOPANTOMOGRAPH™ OP 3D 21
Brightness slider
You can adjust the brightness of the shown Scout image using the
brightness slider.
NOTICE! Adjustment is not saved to the image.
Contrast slider
You can adjust the contrast of the shown Scout image using the
contrast slider.
NOTICE! Adjustment is not saved to the image.
22 ORTHOPANTOMOGRAPH™ OP 3D
OK button
Press OK button to approve the changes made to the FOV location
and size using the Scout image and to proceed to the 3D exposure.
NOTICE! The image has indications from which direction the image is shown; A/
P (Anterior/Posterior), L/R (Left/Right) and H/F (Head/Foot).
Preview projection
Indicates which preview projection is selected.
3D preview image is shown as a slice preview (left) and a rendered
volume (right) on the GUI.
Press the preview projection icons to change the projection of the
shown preview.
Slice slider
You can scroll through the previewed 3D slices using the slice slider.
The previewed 3D volume also changes according to the currently
shown slice.
Brightness slider
You can adjust the brightness of the shown preview image using the
brightness slider.
NOTICE! Adjustment is not saved to the image.
ORTHOPANTOMOGRAPH™ OP 3D 23
Contrast slider
You can adjust the contrast of the shown preview image using the
contrast slider.
NOTICE! Adjustment is not saved to the image.
OK button
Press OK button to close the image preview.
24 ORTHOPANTOMOGRAPH™ OP 3D
1. CALIBRATIONS
Shows a list of user performable device calibrations, their completion status and last
completion date. The calibrations are performed through this menu.
See chapters Calibrations for the user and Complete device calibration for more
detailed information on calibration programs and how to perform them.
2. QUALITY CONTROL
Shows a list of Quality Control (QC) programs, their completion status and last
completion date. See chapter Quality Control for more information on how to take
Quality Control images.
ORTHOPANTOMOGRAPH™ OP 3D 25
3. SYSTEM INFO
4. PREFERENCES
Configurations Friendly name Set a name for the device, which is shown
next to the settings icon and on the
imaging software.
26 ORTHOPANTOMOGRAPH™ OP 3D
Patient Size Adjust the default mA and kV values for imaging programs.
Configuration
5. SERVICE MENU
NOTICE! This menu is intended for authorized service personnel only.
ORTHOPANTOMOGRAPH™ OP 3D 27
1. Carriage up/down slider. The carriage can be driven up and down by sliding a finger
on the carriage up/down slider. The speed of the carriage movement depends on the
distance the finger is moved.
2. Test mode button. Disables radiation production of the device. You can also use the
GUI to activate the Test mode.
3. Patient positioning lights button. Turns the patient positioning lights on/off.
4. HOME button. Drives the rotating unit to HOME (Patient-In) and scan start positions.
5. Exposure indicator (YELLOW). Illuminates when the device is radiating.
6. Status indicator (BLUE). Illuminates when the device is in Error state. The error
details are shown on the GUI.
7. Ready indicator (GREEN). Illuminates when the device is ready for imaging.
NOTICE! If carriage up/down slider backlight is not illuminated, the carriage up/
down slider is in safe mode and cannot be used. Wait until the light comes back on
or wipe the panel clean if it is dirty.
Device status indicator light, located on the top of the carriage, illuminates according the
device status:
• YELLOW: Device is radiating.
• BLUE: Device is in error state or pending user action. Check the GUI for details.
• GREEN: Device is ready for imaging.
28 ORTHOPANTOMOGRAPH™ OP 3D
5 Imaging programs
NOTICE! The segment widths and height differences shown here are for
illustrative purposes only.
NOTICE! The segment widths and height differences shown here are for
illustrative purposes only.
ORTHOPANTOMOGRAPH™ OP 3D 29
Bitewing program
NOTICE! The segment widths and height differences shown here are for
illustrative purposes only.
Lateral TMJ program provides a lateral view of the patient's left and right
temporomandibular joints.
NOTICE! The segment widths and height differences shown here are for
illustrative purposes only.
30 ORTHOPANTOMOGRAPH™ OP 3D
5.2 3D programs
Always select the smallest feasible FOV size, resolution and imaging parameters for the
3D image in order to follow the ALARA (As Low As Reasonably Achievable) principle.
NOTICE! It is always up to the dental professional to select the appropriate FOV,
resolution and imaging parameters.
FOV 5 x 5
Optimized for single site implants or localized diagnostics, for example, 3rd
molar extractions, impacted teeth, periodontal cases, root fractures, single
TMJ analysis, endodontics and for pediatric imaging.
Available resolutions:
High resolution
Standard resolution
Endo program
FOV 6 x 9
Optimized for multiple implant placement using surgical guides, imaging the
whole dental arc of one jaw, 3rd molar visualization, pathology (bilateral
analysis) and periodontal cases.
Available resolutions:
High resolution
Standard resolution
LDT resolution
ORTHOPANTOMOGRAPH™ OP 3D 31
FOV 9 x 11
Optimized for imaging the entire dentition, both mandibula and maxilla,
maxillary sinuses, jaws with bilateral joints and jaws with airways.
Available resolutions:
High resolution
Standard resolution
LDT resolution
FOV 9 x 14 (optional)
Optimized for imaging the entire dentition, both mandibula and maxilla,
including airway and upper cervical spine or the sinus, maxillary sinuses,
jaws with bilateral joints, jaws with airways, analyses of both TMJs and for
maxillofacial surgeries.
Available resolutions:
High resolution
Standard resolution
LDT resolution
5.2.2 3D Resolutions
The resolution affects the image quality and radiation dose to the patient. For example,
high resolution will produce more details than standard resolution, but the radiation dose
is also higher. The unit offers a Low Dose Technology™ (LDT) resolution which can be
used for example in treatment follow-up cases. The LDT resolution will result in images
proportional to the low dose and it is up to the healthcare professional to decide when it
is sufficient to be utilized.
32 ORTHOPANTOMOGRAPH™ OP 3D
95 kV
_ 8.00 mA +
Adjust FOV
ORTHOPANTOMOGRAPH™ OP 3D 33
3. Complete the device initialization by pressing the HOME button on the positioning
panel when indicator light starts blinking in blue. The device GUI also indicates when
HOME button can be pressed.
NOTICE! The device moves up or down when HOME button is pressed.
NOTICE! The device enters a power save mode after 30 minutes of inactivity.
While in the power save mode, the device cooling fan is not active, but the
positioning panel lights are lit. The device wakes up from the power save
mode when connection to it is established or when the positioning panel is
used.
5. The device GUI opens on the workstation when the connection to the device is
established.
34 ORTHOPANTOMOGRAPH™ OP 3D
6. GUI: Select or ensure the correct settings; imaging modality, imaging program
and patient size. If the preset selections are not suitable for the patient, adjust the
imaging parameters manually.
Imaging
Modality
PAN 3D
Imaging
Program
Patient size
Imaging
parameters
1. Press HOME button on the patient positioning panel. The device moves to HOME
(Patient-In) position.
NOTICE! Ensure from the GUI that the device has been set up according to
the required examination.
2. Ask the patient to remove their glasses, hearing aids, removable dentures, jewellery,
hair clips and all other things that may cause artifacts to the image.
3. Protect the patient from radiation according to the local regulations, for example
using a lead apron.
ORTHOPANTOMOGRAPH™ OP 3D 35
4. Select the patient positioning accessory according to the image to be taken, attach it
to the chin rest and place them on the lower shelf of the device, as shown below.
Panoramic Imaging:
Standard and Pediatric Standard and Pediatric TMJ imaging
panoramic and panoramic and
Bitewing imaging Bitewing imaging for
for dentate patients edentulous patients
Bite Block + Chin Rest Lip Support + Chin Rest Lip Support + Chin Rest
3D Imaging:
Lower jaw or both jaws Upper jaw
NOTICE! The chin rest needs to be positioned correctly to the device for the
selected imaging program before the device allows images to be taken.
36 ORTHOPANTOMOGRAPH™ OP 3D
7. Open the head support locking lever on the lower shelf and push the head support
towards the mirror.
8. Guide the patient to the device and instruct to stand as straight and as tall as
possible.
9. Ask the patient to grab the patient handles and place their chin on the Chin Rest and
press their lip against the Lip Support or if using the Bite Block, to bite the notches.
CAUTION! Evaluate the condition of the patient’s dentition before using
the Bite Block. If the patient's teeth are more fragile than a normal healthy
patient's, use the Lip Support and a cotton roll between the front teeth
instead.
NOTICE! If the patient has wide shoulders that could collide with the rotating
unit, ask the patient to cross their arms when holding the handles in order to
contract the soulders.
10. Ask the patient to take a step forward to straighten their spine.
ORTHOPANTOMOGRAPH™ OP 3D 37
11. Fine adjust the device's height and adjust patient's head position/orientation using
the patient positioning lights as guides.
NOTICE! To ensure optimal image quality, pay attention to the correct
patient positioning.
NOTICE! You can open the mirror and use it to help in positioning the
patient. Close the mirror before you start the imaging.
STANDARD AND PEDIATRIC PANORAMIC IMAGING:
• Align the patient so that the tilt light is approximately on the incisors and the root
tips of lower and upper incisors are parallel to the tilt light. The ORTHOfocus™
feature automatically determines the optimum panoramic image layer, enabling
forgiving patient positioning.
NOTICE! The patient’s Frankfort-Horizontal plane should be close to
parallel to the FH light.
BITEWING IMAGING:
• The patient’s occlusal plane should be horizontal and parallel to the FH light.
TMJ IMAGING:
• The patient’s Frankfort-Horizontal plane should be close to parallel to the
horizontal light. The horizontal light indicates the center of the field. You can take
a TMJ image with the patient's mouth closed or open.
38 ORTHOPANTOMOGRAPH™ OP 3D
3D IMAGING:
• The patient's occlusal plane should be horizontal and the ROI is between the top
and bottom FOV lights.
NOTICE! You may place a cotton roll or similar between the patient's front
teeth to separate the jaws.
12. Check that the patient’s head is straight and not rotated and the patient’s midline
coincides with the midsagittal light.
14. Lock the head support in place by closing the temple supports and close the locking
lever.
ORTHOPANTOMOGRAPH™ OP 3D 39
15. If you take a 3D image, place the head support strap behind patient's head and
tighten it to minimize the patient movement.
2. Ask the patient to close their lips and press their tongue against the palate if
possible.
5. The device starts to move and an audible exposure warning signal is played.
6. Hold the exposure button down until all movements have stopped.
NOTICE! The device also plays a program end tone when the imaging
process is complete.
8. A preview image appears on the GUI for a quick review (no visible artifacts etc.).
40 ORTHOPANTOMOGRAPH™ OP 3D
10. Continue to take the next image, if multiple images need to be taken.
11. Press "End Study" to end the study or return to imaging program.
12. Remove all disposable covers and decontaminate the device and the patient
positioning accessories.
1. Ensure the correct patient positioning and ask the patient to stay still during the
Scout imaging process.
4. The device starts to move and an audible exposure warning signal is played.
5. Hold the exposure button down until all movements have stopped.
NOTICE! The device also plays a program end tone when the imaging
process is complete.
7. GUI: Adjust the FOV height, diameter and location using the Scout image, if needed.
NOTICE! If you wish to verify the adjustments, you can take another scout
image by pressing the exposure button.
8. GUI: Approve the scout image and proceed to the 3D imaging phase by pressing OK
button.
12. The device starts to move and an audible exposure warning signal is played.
13. Hold the exposure button down until all movements have stopped.
NOTICE! The device also plays a program end tone when the imaging
process is complete.
ORTHOPANTOMOGRAPH™ OP 3D 41
15. Image preview appears on the GUI for a quick review (no visible artifacts etc.).
17. Continue to take the next image, if multiple images need to be taken.
18. Press "End Study" to end the study or return to imaging program.
19. Remove all disposable covers and decontaminate the device and the patient
positioning accessories.
42 ORTHOPANTOMOGRAPH™ OP 3D
7 Maintenance
The maintenance and calibration procedure intervals described here are minimum
requirements and can be made more frequent and stringent to comply with local
regulations regarding the use and maintenance of dental X-ray devices.
Autoclave
Lip Supports and Bite Blocks may be autoclaved in 121°C (250F) temperature for max. 3
minutes. Other parts are not autoclavable.
NOTICE! Autoclaving is not recommended as cleaning method, and the parts are
not intended to be sterilized.
ORTHOPANTOMOGRAPH™ OP 3D 43
7.2.1 PAN QC
NOTICE! The PAN QC is an optional, yet recommendable, procedure and is
mandatory to be performed where local regulations require it.
5. Attach the copper filter in front of the radiation window on the tube head. The filter
attaches in place with magnets.
44 ORTHOPANTOMOGRAPH™ OP 3D
7. Press and hold down the exposure switch to take the QC image.
10. Workstation: Visually evaluate the image using the dental imaging software:
11. If the image fails on any of the previously listed criteria, redo the QC program. If it
fails again, recalibrate the device or contact service.
7.2.2 3D QC
1. GUI: Go to Settings menu.
ORTHOPANTOMOGRAPH™ OP 3D 45
6. Press and hold down the exposure switch to take the QC image.
9. The QC image appears on the workstation, showing the result of the check.
10. Workstation: Visually evaluate the image using the dental imaging software for
visual defects such as artifacts.
NOTICE! The device determines if the QC image is PASSED or FAILED
according to measured data, not based on the visible image quality.
NOTICE! You should also compare the new QC image to the reference image
taken during the installation or the latest service. Doing this helps you ensure
that the image quality has remained constant.
46 ORTHOPANTOMOGRAPH™ OP 3D
11. If the image fails or any visual defects are apparent, redo the QC program. If it fails
again, recalibrate the device or contact service.
ORTHOPANTOMOGRAPH™ OP 3D 47
The device must be calibrated after the first installation, see chapter Complete device
calibration, or when parts are replaced, as described in the device service manual.
QC images have to be taken always after calibration.
48 ORTHOPANTOMOGRAPH™ OP 3D
2. A list of available device calibrations is shown with the status of the calibration.
NOTICE! Performed calibrations are indicated with green colour and the
completion date of the calibration. All calibrations need to be successfully
performed before using the device.
NOTICE! Calibration images are shown on the GUI as viewed from the
sensor. All adjustment instructions in the calibration images also refer to this
perspective.
ORTHOPANTOMOGRAPH™ OP 3D 49
3D Pixel Calibration
This program calibrates the sensor for 3D imaging.
No calibration tools are required when performing this calibration.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
3. Press and hold down the exposure button to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
3. Press and hold down the exposure button to take the calibration image.
50 ORTHOPANTOMOGRAPH™ OP 3D
4. When the exposure warning stops and the program end tone is played, the program
is complete.
3D Geometry Calibrations
These programs create the calibration data for reconstructing 3D images.
NOTICE! This calibration produces X-rays. Protect yourself from radiation.
4. Press and hold down the exposure button to take the calibration image.
ORTHOPANTOMOGRAPH™ OP 3D 51
52 ORTHOPANTOMOGRAPH™ OP 3D
8 Troubleshooting
Problem Possible cause Solution
Taken image is not Local network connection The device stores the latest
transferred to the modality is disrupted, which causes image until a confirmation
workstation. loss of data. of a succesfull transfer to
the workstation has been
received. Re-establish the
local network connection
and the image data is
transferred automatically.
Do not power the device off
or the image will be erased.
Device cannot be driven 1. Up/down slider is on 1. Clean the positioning
upwards or downwards. safe mode. panel and wait for the
2. Mechanical safety safe mode to clear out.
device on the driving 2. Contact service.
motor has activated.
3. Positioning panel not
working.
No connection to the device. 1. The device is not 1. Power the device on.
powered on. 2. Check the Ethernet
2. Problem with the local cable connections on
area network (LAN). the workstation and the
3. Check the device device.
configuration. 3. Restart the device and
the workstation.
ORTHOPANTOMOGRAPH™ OP 3D 53
9 Installation
NOTICE! This section of the manual is only for authorized service technicians.
NOTICE! It is recommended not to discard the device package and the packaging
supports as they can be used if the device needs to be repacked.
54 ORTHOPANTOMOGRAPH™ OP 3D
ORTHOPANTOMOGRAPH™ OP 3D 55
Recommended
minimum:
1630 mm / 64.2 in
Width Minimum:
1375 mm / 54 in
Recommended
minimum:
1610 mm / 63.4 in
Height Minimum:
2080 mm / 81.9 in
Recommended
minimum:
2467 mm / 97.1 in
Fixing hardware
NOTICE! Mounting bolts for floor and wall fastening are not included in the
delivery. The fixing hardware used to fix the device to the wall must be of correct
type for the wall.
The type and length of hardware to be used depend on the wall material and floor
material where the device is to be fixed. The fixing hardware must be selected so that
they withstand at least 1500 N pull-out force.
WARNING! Ensure that the mounting bolts, the wall and the floor the device is
attached to, can withstand at least 1500 N pull-out force.
Installation tools
• Electric drill
• Spanners size 13 & 17 mm
• Allen keys 3, 5 & 6 mm
• Flat head screwdriver
• Spirit level
• Pliers and wire cutter
• Scissors/knife
• Torque wrench
56 ORTHOPANTOMOGRAPH™ OP 3D
1. Power cord
2. Ethernet cable for workstation connection
3. Exposure button connector
4. Remote exposure switch connector (optional)
5. External warning light connector (optional)
6. Main fuses F1+F2 (10A fuses installed as default)
7. External warning light (optional) fuse F3
8. Control interface
NOTICE! All cables, except the power cord, must be routed through a strain relief
attached to the back panel plate.
The Control interface can be used to connect optional device controls such as external
exposure switch and door switch.
PIN 1 2 3 4 5 6 7 8
USE Exposure switch LOW signal when LOW signal when Door switch
connection device is ready exposure is ON connection
ORTHOPANTOMOGRAPH™ OP 3D 57
5. Open the lowest box to find the device column and the wall bracket.
NOTICE! Use spirit level to ensure that the wall and floor of the chosen
installation location are straight enough to allow the device to be installed
vertically straight.
7. Detach the preinstalled installation bolts, washers and locking nuts from the column
side bracket.
8. Take the wall side bracket of the wall mount from the column box.
58 ORTHOPANTOMOGRAPH™ OP 3D
9. Attach the wall side bracket to the column side bracket using the same bolts,
washers and locking nuts that were previously detached. Do not tighten the bolts
fully yet.
10. Adjust the angle and the depth of the wall bracket and tighten the angle adjustment
screws (2 pcs) to lock the wall bracket into the desired position. Screw the depth
adjustment screw in to hand tightness.
11. Mark the installation screw locations to the wall (2 pcs) and to the floor (4 pcs) using
the holes in the wall bracket and the column base plate as a template.
NOTICE! The top surface of the wall side bracket should be approximately
790 mm (31.1") from the floor when attached to the wall.
CAUTION! If the device cannot be bolted to the floor, an additional wall
bracket or exhibition stand is required. See chapter Additional wall bracket
installation and Exhibition stand installation for more instructions.
12. Lift the column aside and drill appropriate holes for the used fixing hardware.
13. Attach the column to the wall with the wall bracket using appropriate bolts and
washers. Use a spirit level on the top of the column to ensure the column is vertically
aligned in both directions. The wall bracket allows the device to be tilted in all
directions, if needed.
NOTICE! The wall bracket cannot be used in adjustment if the device is
bolted to the floor.
NOTICE! If the angle/position of the wall bracket was required to be
changed, re-tighten the wall bracket angle and depth adjustment screws.
14. Install the wall bracket cover plate with two (2) screws.
ORTHOPANTOMOGRAPH™ OP 3D 59
15. Attach the column to the floor using appropriate screws and washers.
NOTICE! Ensure that the column is vertically straight before fully tightening
the floor screws.
16. Remove the bubble wrap between the outer and inner column covers. Ensure that
the outer column cover plate is properly in place and doesn't scratch the inner
column cover.
18. Pull the carriage to the edge of the pallet so that the diagonal sections of the
plywood supports are outside of the pallet.
19. Flip the carriage 90° to the floor so that it stands in upright position using the
cutouts in the plywood as handles.
60 ORTHOPANTOMOGRAPH™ OP 3D
20. Pull the protective plastic aside from the carriage to reveal the base plate and back
panel connectors.
21. Check that the device's mains power configuration is applicable to the installation
location.
NOTICE! The device is delivered with 10A fuses installed. If installing to other
than 230V mains voltage the fuses need to be changed. See chapters Mains
power configuration and Changing the fuses for more information.
NOTICE! If mains connection type needs to be changed, see chapter Mains
power configuration for more information.
23. Loosen the three carriage locking bolts on top of the column and ensure that they
are facing outwards.
24. Push the carriage installation rail into the slot on the column. The rail locks itself into
the slot.
ORTHOPANTOMOGRAPH™ OP 3D 61
25. Lift the carriage onto the column using the installation rail as a hinge and by lifting
from the cutouts in the plywood support.
26. Lift the carriage locking bolts (3 pcs) up and tighten them to 25 Nm torque to secure
the carriage.
CAUTION! The carriage may fall down during use, if the carriage locking
bolts are not properly installed and tightened.
27. Detach the plywood cover from one side first by detaching all the seven (7) indicated
screws and pulling it off.
62 ORTHOPANTOMOGRAPH™ OP 3D
28. Detach the upper and rear screws (4 pcs) of the plywood cover on the other side and
carefully pull it off with the metal supports attached.
NOTICE! Hold the plywood cover in place when detaching the last screw to
prevent it from falling down.
NOTICE! Be careful not to scratch the unit with the metal supports attached
to the plywood cover.
29. Detach the plywood top cover and the metal supports on top of the carriage (4
screws).
30. Remove the metal support from the rear of the carriage (4 screws).
31. Remove the protective plastic from the carriage carefully so that it doesn't catch and
break any circuit boards.
ORTHOPANTOMOGRAPH™ OP 3D 63
32. Detach the two screws holding the front support bar in place, push it inside the
upper shelf frame and reattach it.
33. Detach the linear movement support from the top shelf (2 screws).
34. Detach the rotating unit support from the carriage (4 screws), push the rotating unit
outwards a bit and lift the support off from the rotating unit .
NOTICE! The rotating unit can now move freely, ensure that it does not hit
the lower shelf.
35. Connect the Column control and Emergency switch connector to the connectors in
the column.
64 ORTHOPANTOMOGRAPH™ OP 3D
36. Connect the Mains cable (1), Ethernet cable (2) and Exposure button cable (3) to
the rear of the carriage.
NOTICE! Route the Ethernet cable and exposure button cable through the
strain relief.
NOTICE! You can also connect other additional cables to the carriage at this
point. See chapter Back panel connectors.
NOTICE! The cables will move when the unit is driven upwards. Ensure that
the cables connected to the device are long enough so that the unit can be
driven to upmost position without causing any strain to the cables.
1. Attach upper shelf covers. Place the right cover onto the upper shelf and then push
the left cover against it.
2. Secure the covers with two screws on the top of the covers and then one on the
bottom.
ORTHOPANTOMOGRAPH™ OP 3D 65
3. Attach the right carriage side cover. Insert the claws on the cover into the slots on
the rear of the column and turn the cover to correct position. Secure the cover in
place with one screw under the status indicator light.
4. Press the side covers together and ensure that they sit properly.
5. Secure the side covers in place with the remaining four (4) screws.
NOTICE! When securing the left cover, ensure that the horizontal lasers sit in
the middle of the groove so that the cover does not block the positioning light
beam.
6. Attach the lower shelf tray by pressing the upper edges down and turning the lower
part against the side covers until it snaps in place.
66 ORTHOPANTOMOGRAPH™ OP 3D
7. Attach the mirror by guiding the pins on the carriage plate (4pcs) through the
keyholes on the mirror plate and sliding the mirror down to lock it in place.
8. Route the mains cord, exposure button cable, Ethernet cable and other optional
cables through the cable clamp on the rear of the device.
9. Attach the back panel cover by sliding it to its place and secure it with one screw,
located on the left side of the unit.
ORTHOPANTOMOGRAPH™ OP 3D 67
10. Attach the carriage back plate by pressing the hooks on the lower side of the plate to
the gaps in the column and turning the upper side in place.
11. Attach the filler strips over the gaps under the carriage back plate.
12. Attach the patient handles with four screws. Tighten the screws to 10 Nm torque.
13. Attach the tube head plate in place by pressing it in one side at a time.
68 ORTHOPANTOMOGRAPH™ OP 3D
14. Install the head support to the lower shelf by screwing the rods in. Slide the head
support into the rods.
15. Attach the exposure button holder to the column or other metallic surface using its
built in magnets or screw it to a non-magnetic surface.
16. Remove all remaining protective film from the device covers and the positioning
panel.
ORTHOPANTOMOGRAPH™ OP 3D 69
2. Connect the device to the local area network (LAN) or directly to the workstation
with an Ethernet cable. See chapter Connection options for more detailed
information.
NOTICE! The unit requires 1Gb Ethernet connection between the unit and the
reconstruction workstation. The workstation must be connected to the unit in
accordance with IEC 60601-1-1.
3. Configure the imaging software to connect to the device. See software instructions
for more information.
4. When communication link has been established, the device GUI appears on the
screen.
Standalone Configuration
In standalone configuration, the unit is connected directly to a workstation and the unit
then reserves the whole network interface card. The unit and the workstation function
as a single standalone system with no access to other networks. If it is necessary to be
connected to another network or to a server, the workstation will require two network
interface cards.
In standalone configuration, both the ETHERNET interface of the workstation and
the device IP address need to be set to have IP addresses from the same subnet, for
70 ORTHOPANTOMOGRAPH™ OP 3D
example 192.168.1.1 for the PC and 192.168.1.2 for the device. The recommended
Subnet mask is 255.255.255.0.
A static IP address can be configured to the device through the device driver. See
software documentation for more information on how to configure the imaging software
for device connection required for the static IP address configuration.
NOTICE! When configuring a static IP address it must be set for both the device
and the PC. The device may have initial local host IP Address (169.xxx.xxx.xxx)
but this is not static unless set. It is not recommended to use addresses from the
local host address range as static IP addresses.
ORTHOPANTOMOGRAPH™ OP 3D 71
3. Ensure that the wall mount screws match the wall material.
4. Ensure that the wall mount and floor screws withstand 1500N pull-out force.
5. Check that the power cord and the Ethernet cable, along with all the other cables
connected, are installed so that the up/down movement of the device doesn't cause
stress to them.
8. Check the electrical safety of the device by inspecting the proper grounding of user
touchable metallic parts.
9. Check that the IEC 60601-1 leakage current levels are not exceeded.
11. Check that the patient positioning lights work properly and are aligned correctly. The
positioning light can be checked during the Device Geometry Calibration.
12. Check that the device has been configured correctly for the installation location. See
chapter Device configuration for more information.
13. Carry out the device calibrations described in chapter Complete device calibration.
17. Gather the screws left over from the installation for storage.
72 ORTHOPANTOMOGRAPH™ OP 3D
1. Turn the device on and establish a connection to the workstation. See chapters
Powering on the device and Preparing the device for imaging for more information.
4. The GUI informs that additional options have been added under the Service menu.
5. Select Calibrations under the Service menu to see the list of all calibration
programs.
ORTHOPANTOMOGRAPH™ OP 3D 73
3. Press and hold the exposure button. The device starts to move.
4. Keep the exposure button pressed down during the whole calibration.
5. When the device's movements stop and the program end tone is played, the
program is complete.
2. Press and hold down the exposure button. The carriage starts to move upwards.
3. Keep the exposure button pressed down until the device is in the maximum allowed
height and release the button.
4. Wait a minimum of 2 seconds and then Press and hold the exposure button down to
drive the device downwards.
5. When the device's movements stop and the program end tone is played, the
program is complete.
74 ORTHOPANTOMOGRAPH™ OP 3D
3. Press and hold down the exposure button to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
3. Press and hold down the exposure button to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
ORTHOPANTOMOGRAPH™ OP 3D 75
NOTICE! This calibration is iterative and may need to be run a couple of times
before it passes.
3D Geometry Calibration Phantom is required when performing this calibration.
3. All patient positioning lights come on. Check that the lights hit the grooves on the
phantom to ensure that the positioning lights are correctly aligned. If the lights do
not hit the grooves, they need to be re-aligned. See chapter Patient positioning light
adjustment.
5. Press and hold down the exposure button to take the calibration image.
6. When the exposure warning stops and the program end tone is played, the program
is complete.
NOTICE! If the device geometry calibration fails and the GUI informs you
to adjust the height of the sensor. See chapter Sensor adjustment for more
information.
NOTICE! Repeat the 3D Pixel Calibration if any mechanical adjustment was
needed.
76 ORTHOPANTOMOGRAPH™ OP 3D
3. Press and hold down the exposure button to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
NOTICE! If the collimator calibration fails and the GUI informs you to adjust
the height or tilt of the collimator, adjust the collimator according to chapter
Collimator adjustment.
NOTICE! Repeat the 3D Pixel Calibration and Device Geometry Calibration if any
mechanical adjustment was required.
ORTHOPANTOMOGRAPH™ OP 3D 77
3. Press and hold down the exposure button to take the calibration image.
4. When the exposure warning stops and the program end tone is played, the program
is complete.
78 ORTHOPANTOMOGRAPH™ OP 3D
4. Press and hold down the exposure button to take the calibration image.
ORTHOPANTOMOGRAPH™ OP 3D 79
5. Attach the copper filter in front of the radiation window on the tube head. The filter
attaches in place with magnets.
7. Press and hold down the exposure switch to take the QC image.
10. Workstation: Visually evaluate the image using the dental imaging software:
11. If the image fails on any of the previously listed criteria, redo the QC program. If it
fails again, recalibrate the device or contact service.
80 ORTHOPANTOMOGRAPH™ OP 3D
12. Save the passed QC image as a reference image for the user and let the user know
the reference image location.
9.8.11 3D QC
NOTICE! The device's country selection needs to be configured per local
requirements before performing the QC programs. See chapter Device
configuration.
6. Press and hold down the exposure switch to take the QC image.
ORTHOPANTOMOGRAPH™ OP 3D 81
9. The QC image appears on the workstation, showing the result of the check.
10. Workstation: Visually evaluate the image using the dental imaging software for
visual defects such as artifacts.
NOTICE! The device determines if the QC image is PASSED or FAILED
according to measured data, not based on the visible image quality.
NOTICE! You should also compare the new QC image to the reference image
taken during the installation or the latest service. Doing this helps you ensure
that the image quality has remained constant.
11. If the image fails or any visual defects are apparent, redo the QC program. If it fails
again, recalibrate the device or contact service.
12. Save the passed QC image as a reference image for the user and let the user know
the reference image location.
82 ORTHOPANTOMOGRAPH™ OP 3D
3. GUI: Select Panoramic Beam Size Verification program from the list.
NOTICE! The Panoramic Beam Size Verification program appears under
Quality Control menu only when the Service menu has been unlocked.
5. Press and hold down the exposure switch to take the calibration.
ORTHOPANTOMOGRAPH™ OP 3D 83
1. Exposure switch
2. Exposure warning light
3. Exposure button connector
1. Connect one end of the remote exposure switch cable to the connector J2 inside the
remote exposure switch.
2. Attach the remote exposure switch to the wall with two screws.
NOTICE! Guide the remote exposure switch cable through the small cut-
out in the bottom of the remote exposure switch housing and ensure that the
cable does not press against the circuit board.
3. Guide the remote exposure switch cable to the back panel of the device, see chapter
Back panel connectors.
4. If the user wants, the exposure button can be attached also to the remote exposure
switch instead of attaching it directly to the device.
84 ORTHOPANTOMOGRAPH™ OP 3D
See chapter Changing the fuses for information on how to install the fuses.
ORTHOPANTOMOGRAPH™ OP 3D 85
If the device mains voltage needs to be changed, change the fuses accordingly. See
chapter Changing the fuses for information on how to install the fuses.
The device is supplied with pre-installed 10A fuses. The fuses may need to be replaced
as part of the installation depending on the mains voltage:
• 10A suited for 220-240 VAC mains voltage
• 15A suited for 100-120 VAC mains voltage
To release a fuse cap, push it inwards and rotate counterclockwise with a flat head
screwdriver.
Remove the fuse from the socket and replace it with a new one. Install the fuse by
pushing the fuse cap in and rotating it clockwise with a screwdriver.
The 10A and 15A fuses are physically of different sizes and have their own fuse caps. If
you need to change the fuse rating, also the fuse cap needs to be changed. The fuses
are supplied with the corresponding fuse caps.
CAUTION! Use only specified fuses. See the list of fuses from chapter Technical
specifications.
86 ORTHOPANTOMOGRAPH™ OP 3D
1. Detach two screws holding the cable router on the rear of the column.
2. Detach two screws holding the wall bracket rail cover in place.
4. Guide the bottom part of the rail cover through the additional wall bracket holder.
5. Lift the wall bracket holder up to the position where the rail has openings on both
sides to fit the holder in.
ORTHOPANTOMOGRAPH™ OP 3D 87
10. Continue to install the column normally according to chapter Installing the device.
1. Transport the exhibition stand to the location, where the device is to be installed.
2. Attach the device column on the exhibition stand. See the installation instructions
supplied with the exhibition stand for more details.
1. Remove the tubehead rear cover by detaching the 3 indicated screws and pull the
cover outwards. The screw on the middle of the cover has a screw cover cap over it.
88 ORTHOPANTOMOGRAPH™ OP 3D
2. Remove the tubehead front cover by detaching 4 screws holding the cover on the
tubehead frame and pull it outwards.
NOTICE! The GUI informs you how much the collimator needs to be adjusted
and in which direction. One full turn of the screw (360°) equals about 0.7 mm.
4. To rotate the collimator LEFT/RIGHT, loosen the two upper locking screws (1), the
lower locking screw (2) and turn the adjustment screw (3):
a) Clockwise to turn the collimator left.
b) Counter clockwise to turn the collimator right.
NOTICE! You may also loosen the screws 2 and 3, turn the collimator to the
required position by hand and then tighten the screws.
5. After the adjustments, tighten the locking screws and redo the collimator calibration
to verify the adjustment.
ORTHOPANTOMOGRAPH™ OP 3D 89
1. Pull the left and right sensor collar cover halves downwards.
2. Pull the inner collar cover half downwards to unlock the teeth holding the covers
together.
4. Remove the screw that holds the sensor front and back cover halves.
90 ORTHOPANTOMOGRAPH™ OP 3D
7. Turn the adjustment screw to adjust the height of the sensor. One turn is equivalent
of 0.7 mm.
ORTHOPANTOMOGRAPH™ OP 3D 91
2. The positioning lights are held in place in a fixture with one screw. Loosen the screw
to release the light.
Horizontal lights Midsagittal light
6. Attach the 3D Geometry Calibration Phantom to the device and verify the correct
alignment of the lights with Device Geometry calibration program selected.
92 ORTHOPANTOMOGRAPH™ OP 3D
Tilt light
1. Detach the Layer laser cover from the lower shelf by detaching a screw on the
underside and pulling it out.
2. If you need to rotate the light, loosen the screw A. If you need to reposition the light
assembly, loosen the screw B and move the assembly sideways by turning the screw
C.
5. Attach the 3D Geometry Calibration Phantom to the device and verify the correct
alignment of the lights with Device Geometry calibration program selected.
ORTHOPANTOMOGRAPH™ OP 3D 93
Configurations Friendly name Set a name for the device, which is shown
next to the settings icon and on the imaging
software.
94 ORTHOPANTOMOGRAPH™ OP 3D
ORTHOPANTOMOGRAPH™ OP 3D 95
9.9.10 Verification
The device offers Panoramic and 3D verification test programs with the device
movements disabled. These programs can be used for radiation related measurements,
for example, during device installation and acceptance tests.
A separate radiation measuring device/equipment is required for taking the radiation
measurements.
1. Go to Settings menu.
4. Select the PAN testing tab for Panoramic verification test or 3D testing tab for 3D
verification test.
The PAN Testing program allows adjustment of the exposure values kV and mA for the
tests.
3. GUI: Select Service menu that appears under the Service menu.
96 ORTHOPANTOMOGRAPH™ OP 3D
5. Press and hold down the exposure switch to drive the unit to the correct height.
7. Detach the device from mains power and detach any external cables.
8. Detach the covers in reverse order to what is described in chapter Attaching the
covers.
NOTICE! It's recommended to wrap protective material around the covers
and place them in boxes to protect them against scratches.
9. Detach the column interconnection cables, install the packaging supports back to
their original place and install the plywood supports in reverse order to what is
described in chapter Installing the device.
10. Loosen the three locking screws holding the carriage on the column and turn them
outwards.
11. Carefully push the latch inside the carriage with a flat tool to release the carriage and
carefully lower it to the floor.
NOTICE! This step requires at least two persons.
12. Detach the carriage from the column, detach the column from the wall and floor and
move the carriage and the column onto the transportation pallet.
ORTHOPANTOMOGRAPH™ OP 3D 97
10 Technical data
98 ORTHOPANTOMOGRAPH™ OP 3D
Device data
Model: PCX–1
Protection against electric shock Class I
Degree of protection Type B applied with no conductive
connection to the patient
Protection against the ingress of liquids IP20
Cleaning methods • Distilled water
• Ethanol 96%
• Isopropyl alcohol
• Soapy water
• Cavicide™ disinfectant
• Metasys™ disinfectant
• Disposable plastic covers for bite block,
chin rest, lip support and head support
Tube voltage 60 – 95 kV
Max. tube current 16 mA
Target angle 5°
Focal spot 0,5 (IEC 60336/2005)
Total filtration min. 3,2mm Al @ 95kV
Leakage technical factors 2717 mAs/h @ 95kV/4mA
ORTHOPANTOMOGRAPH™ OP 3D 99
Electrical connections
Nominal mains voltage 100–240 Vac
Input power frequency 50 / 60 Hz
Nominal current 10A @ 220–240 VAC, 15A @ 100 – 120
VAC
Main fuses (F1 & F2) 220–240Vac:
Littelfuse 215 (Time–Lag) 10A
Cooper Bussman (Time Delay) S505H–10–
R
100–120Vac:
Littelfuse 326 (SlowBlow) 15A
Cooper Bussman (Time Delay) MDA–15
External warning light fuse (F3) Cooper Bussmann (Time Delay) S506–2–R
Power consumption 2,3 kVA @ 220–240 VAC, 1,65 kVA @
100–120 VAC
Maximum impedance of mains 0,2 Ω
Recommended circuit breaker max. 30 mA
Mains cord EU: 16 A / 250 VAC; 50/60 Hz H05VV–
F3G1.5
US: 15 A / 125 VAC 50/60 Hz SJT 3 x AWG
14
100 ORTHOPANTOMOGRAPH™ OP 3D
X–ray generator
Generator High frequency DC generator
Nominal power 1520 W nominal @ 95 kV, 16 mA
Max. electric output 1,7 kW
Anode voltage 60 – 95 kV (+/– 5 kV)
Anode current 2 – 16 mA (+/– 20%)
Exposure time accuracy +/– 5%
DAP display accuracy +/– 25%
Supply frequency 50 – 100 kHz
Spine compensation kV / mA compensated
Accuracy of radiation output > 95%
User interface
Program and technical factors selection, GUI, optional remote exposure switch,
exposure control exposure button with 3m cable.
Patient positioning Positioning panel, integrated
ORTHOPANTOMOGRAPH™ OP 3D 101
Device package
Package dimensions (L x W x H) 1200 x 770 x 1100 mm
47.2" x 30.3" x 43.3"
Ambient temperatures
Transportation and storage -25°…+55°C
Operation Temperature +10°...+35°C
RH 30–80%
Atmospheric pressure 70 – 106 kPa
102 ORTHOPANTOMOGRAPH™ OP 3D
ORTHOPANTOMOGRAPH™ OP 3D 103
NOTICE! Device's radiation dose production vary from unit to unit. Radiation
dose production, shown on the GUI, is calculated by scaling a measured reference
dose production value with the selected imaging program technical factors and
DAP correction factor described in chapter Device configuration. Air KERMA
production can be calculated by dividing the provided DAP value with the active
sensor area.
104 ORTHOPANTOMOGRAPH™ OP 3D
400 ± 20 mm
15.7 ± 0.8 in
951 - 1706 mm / 37.4 - 67.2 in
320 ± 20 mm
12.6 ± 0.8 in
900 - 1655 (nominal 940) mm
20 mm 16 mm
35.4 - 65.2 (nominal 37) in
A= 0.05m2
188 mm 198 mm
7.4 in 93 - 128 mm 7.8 in
219 mm 3.7 - 5 in 229 mm
8.6 in 9 in
450 mm
17.7 in
ORTHOPANTOMOGRAPH™ OP 3D 105
PERPENDICULAR INSTALLATION
** 566 mm / 22.3 in
** 1385 mm / 54.5 in
* 1330 mm / 52.4 in
340 mm
13.4 in * 550 mm / 21.7 in
* 1100 mm / 43.3 in
** 660 mm / 26 in
** 1375 mm / 54 in
99 mm / 4 in 99 mm / 4 in max. 15
max. 15° °
1377 mm / 54 in
** 1375 mm / 54 in ** 1358 mm / 53 in
106 ORTHOPANTOMOGRAPH™ OP 3D
54 mm / 2 in
0°
x. 4
ma
** 1309 mm / 52 in
** 1292 mm / 51 in
NOTICE! This installation setup may require the carriage plywood supports to be
removed before lifting the carriage on the column, as they might hit the wall.
192 mm / 7.6 in
128 mm / 5 in
30 mm 17.5 mm / 0.7 in
1.2 in
16 mm / 0.6 in
260 mm / 10.2 in
300 mm / 11.8 in
340 mm / 13.4 in
380 mm / 15.0 in
420 mm / 16.5 in
ORTHOPANTOMOGRAPH™ OP 3D 107
Date of manufacture
Serial number
Caution
Radiation warning
Laser warning
Dangerous voltage
On or enabled
Off or disabled
Exposure switch
Ethernet
Focal spot
Do not reuse
Recyclable
108 ORTHOPANTOMOGRAPH™ OP 3D
Operating instructions
Refer to operating instructions for more information. The
operating instructions can be supplied electronically or in
paper format.
CE (0537) symbol
NRTL Mark
Conforms to AAMI ES60601-1:2006. Certified to CSA.
ORTHOPANTOMOGRAPH™ OP 3D 109
PCX-1 is suitable for use in the specified electromagnetic environment. The purchaser or user of
PCX-1 should assure that it is used in an electromagnetic environment as described below:
Radio-Frequency Group 1 PCX-1 uses RF energy only for its internal function. Therefore,
Emissions CISPR11 the RF emission is very low and not likely to cause any
interference in nearby electronic equipment.
110 ORTHOPANTOMOGRAPH™ OP 3D
PCX-1 is suitable for use in the specified electromagnetic environment. The purchaser or user of
PCX-1 should assure that it is used in an electromagnetic environment as described below:
Electrostatic ± 8 kV for contact ± 8 kV for contact Floors are wood, concrete, or ceramic
discharge (ESD) discharge discharge tile, or floors are covered with synthetic
IEC61000-4-2 material and the relative humidity is at
±2, 4, 8, 15 kV for ±2, 4, 8,15 kV for least 30 percent.
air discharge air discharge
Electrical fast ±2 kV for power ±2 kV for power Mains power quality is that of a typical
transient/burst IEC supply lines (100 supply lines commercial and/or hospital environment
61000-4-4 kHz)
±1 kV for input/
±1 kV for input/
output lines
output lines
(100kHz)
Voltage dips, short -0 % U mains; 0,5 -0 % U mains; 0,5 Mains power quality is that of a
interruptions and cycle At 0°, 45°, cycle At 0°, 45°, typical commercial and/or hospital
voltage variations 90°, 135°, 180°, 90°, 135°, 180°, environment. If the user of PCX-1
on power supply 225°, 270° and 225°, 270° and requires continued operation during power
input lines 315° 315° mains interruptions, it is recommended
that PCX-1 be powered from an
IEC61000-4-11 -0 % U mains; 1 -0 % U mains; 1 uninterruptible power supply.
cycle, At 0° cycle, At 0°
-70 % U mains; -70 % U mains;
25/30 cycles At 0° 25/30 cycles At 0°
-0 % U mains; -0 % U mains;
250/300 cycle At 0° 250/300 cycle At 0°
NOTICE! UT is the a.c. mains voltage prior to application of the test level.
ORTHOPANTOMOGRAPH™ OP 3D 111
PCX-1 issuitable for use in the specified electromagnetic environment. The purchaser or user of
PCX-1 should assure that it is used in an electromagnetic environment as described below:
Radiated RF IEC 10 V/m 80 MHz 10 V/m 80 MHz 800 MHz to 2,7 GHz
61000-4-3 to 2,7 GHz to 2,7 GHz Where P is the maximum output power rating of the
Immunity to Immunity to transmitter in watts (W) according to the transmitter
proximity fields proximity fields manufacturer and d is the recommended separation
from RF wireless from RF wireless distance in meters (m). Field strengths from fixed RF
communication communication transmitters, as determined by an electromagnetic
equipment, equipment, site survey,* are less than the compliance level in
levels according levels according each frequency range.**
to table to table Interference may occur in the vicinity of equipment
60601-1-2 ed4 60601-1-2 ed4 marked with the following symbol:
table 9. table 9.
*Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios,
amateur radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
performed. If the measured field strength exceeds the RF compliance level above, observe PCX-1 to verify
normal operation in each use location. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating PCX-1.
Guidance for actions taken can be found from AAMI TIR 18:2010, Guidance on electromagnetic compatibility
of medical devices in healthcare facilities.
NOTICE! Precautions to take if the use location is near (e.g. less than 1,5 km from) AM, FM or TV
broadcast antennas.
**Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m. The Recommended
Separation Distances are listed in the next table.
NOTICE! These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
112 ORTHOPANTOMOGRAPH™ OP 3D
Recommended Separation Distances for Portable and Mobile RF Communications Equipment IEC
60601-1-2
The PCX-1 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer of the user of PCX-1 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the PCX-1
as recommended below, according to the maximum output power of the communications equipment.
Frequency of Transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.7 GHz
Equation
1 2 0,6 0,6
100 20 6 6
For transmitters rated at maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTICE! At 80 MHz, the separation distance for the higher frequency range applies.
NOTICE! These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
USE LIMITATION:
External components
WARNING! Use of accessories, transducers and cables other than those specified
or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this device
and result in improper operation.
ORTHOPANTOMOGRAPH™ OP 3D 113
Interference may occur in the vicinity of equipment marked with the following symbol:
114 ORTHOPANTOMOGRAPH™ OP 3D
ORTHOPANTOMOGRAPH™ OP 3D 115
116 ORTHOPANTOMOGRAPH™ OP 3D
216453-2
Headquarters
PaloDEx Group Oy | Nahkelantie 160 | P.O. Box 20
FI-04301 Tuusula | FINLAND | Tel. +358 10 270 2000
http://www.kavo.com/en/contact
www.kavokerr.com
USA
Instrumentarium Dental | 11727 Fruehauf Drive | Charlotte, NC
28273 | U.S.A
Tel. 800-558-6120 | Fax. 877-292-6050
Email: [email protected]