RTD Measurement System Design Essentials
RTD Measurement System Design Essentials
Introduction
High-accuracy temperature measurements provide essential data for industrial automation applications to ensure both product quality and safety. Many types of
temperature sensors are available, and each one has its advantages and disadvantages. This application note focuses on resistance temperature detectors (RTD), and
explains the design essentials to optimize the measurement accuracy.
Generally, RTDs generate more stable and repeatable outputs, compared to thermocouples and thermistors. Hence, RTDs achieve higher measurement accuracy.
The goal is to accurately measure the RTD resistance and convert it to temperature using an equation or a lookup table. For ideal cases:
or
However in practice, the lead wires of the RTD have resistance. Long lead wires greatly affect the measurement accuracy. Therefore, the actual resistance measured by
the circuits shown in Figures 1 and 2 is:
RTD + (2 × R WIRE ),
where R WIRE is the resistance of the lead wires, assuming both wires have the same resistance. Although theoretically acceptable, the same R WIRE implies that both
wires are of the exact same length and made with the exact same material. Such an assumption cannot be guaranteed in critical temperature sensing applications. For
this reason, RTDs feature 3-, and 4-wire configurations to help eliminate the measurement error contributed by lead wires.
Page 1 of 8
Figure 2. 2-wire constant voltage excitation configuration.
One method to further improve the measurement accuracy is by adding an analog switch to the circuit. The ADC then measures the voltage (VX) at the output of the
excitation signal and obtains a value for R WIRE 1. By assuming R WIRE 1 is approximately the same as R WIRE 3, the R WIRE 3 can be subtracted out. Referring to Figure 3, in
current excitation configuration, R WIRE 1 resistance equals to:
and
This method to improve the measurement accuracy does require extra hardware and adds complexity to the software.
Page 2 of 8
Figure 4. 3-wire constant voltage excitation configuration.
because no current passes through R WIRE 2 or R WIRE 3. Therefore, the voltage across R WIRE 2 + RTD + R WIRE 3 is the same as the voltage across the RTD.
Unfortunately, when using a constant voltage excitation configuration, because of the voltage divider effect, R WIRE 1 and R WIRE 4 still creates errors in the RTD
measurement unless the ADC system has the ability to measure the voltage at the excitation voltage output (VX). If the voltage at VX is known, then the reference current
can be calculated by
Many other factors in the signal chain affect the accuracy of measurement. These factors include the input impedance of the ADC system, the resolution of the ADC, the
amount of current through the RTD, the stability of the voltage reference, and the stability of the excitation signals.
The inputs of the ADC system must be high impedance to avoid voltage drops across the lead wires (RWIRE 2 and R WIRE 3 in 4-wire configuration for example). If the
ADC doesn't have high impedance inputs, buffers should be added in front of the inputs of the ADC.
Page 3 of 8
Figure 6. 4-wire constant voltage excitation configuration.
Heating Error
Although an RTD is a sensor, it is also a resistor. When current passes through a resistor, there is power dissipation. The dissipated power heats up the resistor. This
self-heating effect creates error in the measurement. Excitation current must be carefully chosen to ensure the error created is within the error budget. The key formula to
calculate the self-heating error is
2
ΔT = (I REF ×RRTD ) × F,
where F is the self-heating factor of RTD, expressed in mW/°C. For example, a PT-100 platinum RTD with a 0.05°C/mW self-heating factor submerged in ice water.
When the measuring temperature is 0°C, R RTD equals to 100Ω. If the IREFis set to 10mA, the self-heating error becomes
2
((0.01A) × 100Ω) × 50°C/W = 0.5°C.
Depending on the application, this error may or may not be acceptable. For high-accuracy measurements, a lower excitation current reduces the self-heating error. For
example, if IREF is lowered to 1mA, the self-heating error becomes 0.005°C. This level of error is much more tolerable. While reducing the excitation current reduces the
self-heating error, it also reduces the voltage signal span across the RTD, thus requiring the RTD signal to be amplified so that the ADC can extract more discrete signal
levels. An alternative would be to use a higher resolution ADC.
Up to this point, all the formulas discussed involve either IREF or VREF. But, what if these excitation signals are not stable? Instability can result from short term or long-
term drift. Clearly, if the excitation signals become inaccurate, all the calculations above have errors. Consequently, periodic calibrations are required. Of course, an
engineer could use a super stable voltage reference with ultra-low temperature drift and long-term drift. However, usually such devices are very costly. Alternatively, the
ratiometric temperature measurement method eliminates errors caused by inaccurate excitation signals.
where
VIN equals the voltage across the RTD. For current excitation mode
Substituting VIN and VREF into the ADC conversion formula yields,
and subsequently,
Page 4 of 8
Similarly for voltage excitation,
In both cases, after the simplification, RRTD becomes a function of R REF and ADC code; hence, the accuracy of the RTD measurement depends on RREF . For this
reason, when selecting a reference resistor, the engineer must pick one with low temperature and long-term drift.
For a platinum RTD, the Callendar-Van Dusen equation describes the relationship between resistance and temperature as:
where
Page 5 of 8
-3
A = 3.908 × 10
-7
B = -5.775 × 10
-12
C = -4.183 × 10 when t < 0°C
C = 0 when t > 0°C
This equation provides the expected RTD resistance given a known temperature. If the temperature range of interest is above 0°C, then the constant C becomes 0 and
the equation becomes a quadratic formula. Solving a quadratic equation is straightforward. But, if the temperature goes below 0°C and the C constant becomes non-zero,
the equation becomes a difficult 4th order polynomial. In this case, polynomial interpolation approximation is a very useful tool. Here is a Microsoft Excel solution:
1. On a spreadsheet, create two columns of data. One column lists the temperature. The second column lists the corresponding RTD resistances calculated from the
Callendar-Van Dusen equation.
2. Create an X-Y scatter plot.
3. Add a polynomial trendline for the plot. A higher order of the polynomial gives a more accurate approximation.
4. Select "Display Equation on Chart" in the "Format Trendline" menu.
The resulting polynomial equation for a PT100 for t < 0°C is:
6 5 4 3 2
t = -1.6030e -13 × r + 2.0936e -10 × r -3.6239e -8 × r -4.2504e -6 × r + 2.5646e -3 × r + 2.2233 × r -2.4204e2
Increasing the decimal places of the polynomial coefficients reduces error. With four decimal places, as shown in the formula above, the temperature approximation error
is less than 0.005°C, tolerable for most applications.
Page 6 of 8
MAXREFDES67# reference design implements the 4-wire ratiometric configuration and polynomial approximation, as described above. In addition, design files and
firmware are available for subsequent modification and implementation. Furthermore, this reference design (Figures 9, 10, 11) is a complete universal analog input for
industrial applications. Besides RTD measurement, this unique 24-bit front-end also accepts bipolar voltage and current, and thermocouple (TC) inputs. Built in Maxim
Integrated's ultra-small Micro PLC form factor, the MAXREFDES67# performs with an effective resolution up to 22.3 bits with temperature error as low as ±0.1% across a
range of -40°C to 150°C. Figure 12 shows the temperature error measured by the MAXREFDES67# RTD input versus temperature referenced to three different
thermometers. The references are the Omega HH41 thermometer, the ETI reference thermometer, and Fluke 724 temperature calibrator, respectively. The
MAXREFDES67# connected RTD probe (Omega P-M-1/10-1/4-6-0-G-3) was placed in the Fluke 7341 calibration bath and calibrated at 20°C.
Page 7 of 8
Figure 12. MAXREFDES67# error vs temperature, using an Omega P-M-1/10-1/4-6-0-G-3, 4-wire RTD calibrated at 20°C.
Conclusion
Temperature is the most measured industrial parameters. Though precision system design using techniques like the ratiometric method and polynomial approximation
make very accurate measurement systems, with Maxim's reference design system, designers can now develop high-accuracy RTD temperature measurement or
thermocouple measurement systems faster than ever before. The MAXREFDES67# allows modification and implementation, and is a complete universal analog input for
industrial applications. Besides RTD measurement, it also accepts bipolar voltage, current and thermocouple inputs, performs with an effective resolution, and low
measurement error making it more effective than other options.
A similar version of this application note appeared February 2016 on EDN Network.
Related Parts
MAX11254 24-Bit, 6-Channel, 64ksps, 6.2nV/ÃHz PGA, Delta-Sigma ADC with SPI Interface Free Samples
MAX14759 Above- and Below-the-Rails Low On-Resistance Analog Switches Free Samples
MAX14763 Above- and Below-the-Rails Low On-Resistance Analog Switches Free Samples
MAX14930 4-Channel, 2.75kV RMS and 3.75kV RMS Digital Isolators Free Samples
MAX17498C AC-DC and DC-DC Peak-Current-Mode Converters for Flyback/Boost Applications Free Samples
MAX17552 60V, 100mA, Ultra-Small, High-Efficiency, Synchronous Step-Down DC-DC Converter with 22μA Free Samples
No-Load Supply Current
MAX17651 60V, 100mA, Ultra-Low Quiescent Current, Linear Regulator Free Samples
MAX31723 Digital Thermometers and Thermostats with SPI/3-Wire Interface Free Samples
More Information
For Technical Support: https://www.maximintegrated.com/en/support
For Samples: https://www.maximintegrated.com/en/samples
Other Questions and Comments: https://www.maximintegrated.com/en/contact
Page 8 of 8