H79IQ001EN-A ServiceReportAbbemat Eco 3000 3100 3200 Fields

This document is a maintenance and calibration report for an Anton Paar OptoTec device. It details checks performed on the device's environmental conditions, equipment condition, functions, measurements, adjustments, and final checks. Any replacement parts or deviations from specifications are noted. The report is signed by both the service engineer and customer to confirm the work completed and device status.

Uploaded by

Diego Bou
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
17 views6 pages

H79IQ001EN-A ServiceReportAbbemat Eco 3000 3100 3200 Fields

This document is a maintenance and calibration report for an Anton Paar OptoTec device. It details checks performed on the device's environmental conditions, equipment condition, functions, measurements, adjustments, and final checks. Any replacement parts or deviations from specifications are noted. The report is signed by both the service engineer and customer to confirm the work completed and device status.

Uploaded by

Diego Bou
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 6

Anton Paar OptoTec

Maintenance and Calibration Report

Device / System:

Serial Number:

Software Version:

Customer Details
Name:

Person in charge:

Address (street) 1:

Address (street) 2:

City:

Country:

Additional information:

Anton Paar Certified Service

Your Service Partner:


Working for You.

Anton Paar High-precision Instruments H79IQ001EN-A


www.anton-paar.com page 1 of 6
Anton Paar OptoTec

Condition of Equipment before Maintenance

Environmental Condition
Ambient temperature according to specification (10 °C to 35 °C or 50 °F to 95 °F)  Yes  No
Ambient relative humidity according to specification (10 % to 90 %, non condensing)  Yes  No

Comment:

Equipment Condition
Is the instrument adjusted?  Yes  No Last adjustment:
Outside  Excellent  Good  OK  Bad
Prism/well condition  Clean, transparent  Slightly dirty  Dirty/corroded

Equipment Condition Abbemat 3100/3200


Condition of drying cartridge is good  Yes  No
Prism/sample set temperature  Yes  No  N/A
20 °C/68 °F stable?
User defined prism/sample
set temperature:
User defined prism/sample  Yes  No  N/A
set temperature stable?

Comment:

Function Test
Yes No N/A Remarks
Touchscreen/membrane keyboard   
RS-232 Interface   
Ethernet Interface   
USB Ports   
Serial USB   
Fan   

Anton Paar High-precision Instruments H79IQ001EN-A


www.anton-paar.com page 2 of 6
Anton Paar OptoTec

Measurement and Performance Check

Incoming Checks
Are the refractive index measurements repeatable?  Yes  No

Comment:

Use „Other“ for Abbemat 3000


Temperature 20.00 °C 25.00 °C Other:
Refractive Index of pure water nD nD nD
Measurement value nD nD nD
Deviation nD nD nD

Is the measured refractive index within the customers specifications?  Yes  No

Comment:

Anton Paar High-precision Instruments H79IQ001EN-A


www.anton-paar.com page 3 of 6
Anton Paar OptoTec

Adjustments

Temperature Adjustment and Calibration


Abbemat T-Check works only with Abbemat 3200 and a „Helper-Refractometer“ e.g. Abbemat 350

Temperature adjustment required?  Yes  No  N/A


Signature of customer

Temperature calibration performed (Abbemat 3200 only)?  Yes  No  N/A


Report No.:

Adjustment with Pure Water


Set value temperature Set value nD
Adjustment performed  Yes  No  N/A

Certificate available  Yes  No


Provided by: Certificate No.: Valid until:

Comment:

Adjustment with Refractive Index Standard 1 (Abbemat 3100/3200 only)


Substance
Set value temperature Set value nD
Adjustment performed  Yes  No  N/A

Certificate available  Yes  No


Provided by: Certificate No.: Valid until:

Comment:

Adjustment with Refractive Index Standard 2 (Abbemat 3100/3200 only)


Substance
Set value temperature Set value nD
Adjustment performed  Yes  No  N/A

Certificate available  Yes  No


Provided by: Certificate No.: Valid until:

Comment:

Anton Paar High-precision Instruments H79IQ001EN-A


www.anton-paar.com page 4 of 6
Anton Paar OptoTec

Final Checks

Check Results
Are the refractive index measurements repeatable?  Yes  No

Comment:

Use „Other“ for Abbemat 3000


Temperature 20.00 °C 25.00 °C Other:
Refractive Index of pure water nD nD nD
Measurement value nD nD nD
Deviation nD nD nD

Is the measured refractive index within the customers specifications?  Yes  No

Comment:

Anton Paar High-precision Instruments H79IQ001EN-A


www.anton-paar.com page 5 of 6
Anton Paar OptoTec

Conclusion

Replaced Parts
Spare Part Material Number Quantity

General Information
Comments:

At the time of the maintenance and calibration service no deviations to


equipment relevant specifications were found:
 Yes  No

Signatures
Service Engineer Customer
Name Name
Date Date
Company Company

Signature Signature

Anton Paar High-precision Instruments H79IQ001EN-A


www.anton-paar.com page 6 of 6