2.5.2 HZ-2000H Transformer Tan Delta-User Manual
2.5.2 HZ-2000H Transformer Tan Delta-User Manual
USER MANUAL
I.Summary of Instrument.................................................................................................................1
II.Security Measures........................................................................................................................ 1
III.Performance Feature.................................................................................................................. 2
IV.Technical Indicators..................................................................................................................... 4
effectively find out the whole damp deterioration and local defects of insulation of electrical
handover and preventive test. The measurement of dielectric loss of transformer, mutual
inductor, reactor, capacitor, bushing and lightning arrester is the most basic method to
measure its insulation performance. The dielectric loss tester has broken through the
technology, using single chip microcomputer and modern electronic technology to carry
out automatic frequency conversion, mode/number conversion and data operation. Strong
anti-interference ability, fast test speed, high precision, automatic digitization, easy
operation; Power supply adopts high-power switching power supply, output 45Hz and
Automatic filter 50Hz interference, suitable for substation and other electromagnetic
interference large field test. Widely used in the power industry transformer, mutual
inductor , casing, capacitor, lightning arrester and other equipment dielectric loss
measurement.
II.Security Measures
1. Read this manual carefully before using this instrument.
2. The operator of the instrument should have general knowledge of the use of
3. This instrument can be used both inside and outside, but it should be kept away
from places such as rain, corrosive gas, excessive concentration of dust, high
5. The maintenance, care and adjustment of the instrument should be carried out by
professionals.
6. Before any connection, the grounding terminal of the instrument must be reliably
1
connected with the earth with an earth cable.
7. Due to the high voltage generated by the test equipment, the tester must strictly
follow the safe operation procedures and prevent others from contacting the high-voltage
components and circuits. Direct test personnel must fully understand the high-voltage test
circuit, and instrument operation points. Non-engaged testers must stay away from the
high-pressure test area, which must be clearly marked with fences or ropes, warning signs,
etc.
8. The adjustment, repair and maintenance of the instrument must be carried out
without electricity. If it must be live, the operator must be very familiar with the high voltage
9. When the fuse is damaged, the same insurance must be replaced. It is forbidden to
10. When the instrument fails, turn off the power switch and wait for one minute
before checking.
III.Performance Feature
1. The instrument adopts Fourier transform digital filtering technology to measure
capacitance, dielectric loss and other parameters. The precision of the test results is
frequency interference in the field, and reliable data can be measured even in the
3. Use full-touch super large LCD, easy to operate. Full touch LCD screen, large full
graphical operation interface, each process is clear, operators do not need additional
professional training to use. The whole process can be measured with a touch.
4. Data storage: it is equipped with calendar chip and large-capacity memory, which can
save the test results at any time, check the history record at any time, and print out. The
current time and storage time can be displayed and printed at any time.
2
5. Scientific and advanced data management: instrument data can be exported through U
disk. Data can be viewed and managed by special software on any PC.
6. The instrument is easy to operate, and the measurement process is controlled by the
7. Integrated model, with standard capacitor and high-voltage power supply, facilitates
8. The measurement accuracy of the instrument is high, which can meet the requirements
of oil dielectric loss measurement. Therefore, the measurement of oil dielectric loss can
be achieved only by equipping standard oil cup and special test line.
9. It has reverse wiring low-voltage shielding function. When the 220kV CVT bus is
grounded, 10kV reverse wiring dielectric loss measurement can be performed on the
10. Able to do ac withstand test. It is convenient for PT, CT secondary side ac withstand
11. It can identify the frequency of external high voltage power supply (40Hz~70Hz), and
allow the use of frequency power supply or serial resonant power supply to conduct
12. With CVT test function, CVT self-excitation test can be realized, and 4 protection limits
of high voltage/current and low voltage/current can be set to ensure the safety of
13. When testing CVT, not only the capacitance and dielectric loss values of C1 and C2
can be automatically tested, but also the total capacitance and dielectric loss values
14. Equipped with thermal printer for printing and output, with calendar clock, convenient
15. With a computer interface. A computer can control 32 instruments, which can be
integrated into the comprehensive high-voltage test vehicle to achieve measurement, data
16. Grounding protection function, when the instrument is not grounded or the grounding
3
is bad, the instrument does not enter the normal program, does not output high voltage.
Over-current protection, the instrument will not be damaged in short circuit or breakdown.
17. Electric shock protection function: when the operator of the instrument accidentally
touches the electric shock, the instrument will immediately cut off the high voltage to
IV.Technical Indicators
Accuracy: Cx: ±(readings×1%+1pF)
tgδ: ±(readings×1%+0.00040)
Capacitance range:
Boost and drop voltage mode: the voltage is set at will. Such as 5123 v.
Test frequency: 40-70Hz single frequency can be set at will. For example, 48.7 Hz.
External high voltage: maximum test current 5A, power frequency or frequency
4
CVT self-excitation method low voltage output: output voltage 3~50V, output current
3~30A
storage data).
Weight: 28kg
measurement methods, namely, the forward connection measurement method and the
Figure 1
At the 10kV side of the high-voltage power supply, the high-voltage power supply is
divided into two channels, and the first channel feeds the standard capacitor CN inside the
machine. The dielectric loss of this capacitor is very small, which can be considered as
5
zero, i.e. pure capacitive current. This current ICN can be used as the capacitive current
reference. On the Cx test side, the test current Icx is collected into the machine through
the sampling resistance R. The Icx can be decomposed into horizontal and vertical
components, as shown in figure 2. tgδ value can be obtained by calculating the ratio of the
In FIG. 1 (a), Cx is the ungrounded sample, and the sample current Icx enters the
sampling resistance R from the sample end to obtain the full current value. In FIG. 1 (b),
Cx is the grounding sample, the Cx end of the machine is directly grounded, and the
current Icx obtains the full current value from the high-voltage end of the sample to the
Figure 2
instrument.
—— the grounding end of the instrument is reliably connected with the earth
during use
2.Wiring reference
6
2.2 Reverse Wiring, Internal Standard Capacitance, Internal High Voltage (Conventional
Reverse Wiring)
7
2.5 Positive Wiring, Internal Standard Capacitance, External High Voltage:
2.7 Positive Wiring, External Standard Capacitance, External High Voltage (High Voltage
Dielectric Loss):
8
2.9 CVT self-excitation measurement:
is a single saving capacity, the bus cannot be grounded; If C1 is multi-saving capacity, the
bus can be grounded, C11 and C12 can be measured by conventional positive and
In CVT self-excitation measurement, the instrument first measures C1, then automatically
It should be noted that the high-voltage wire should be suspended and not touch the
ground, otherwise the additional dielectric loss to the ground will cause errors. The
high-voltage socket can be connected with the CVT test product with a thin cable and
hoisted.
9
2.10 Reverse wire low voltage shielding test
Reverse wiring low voltage shielding function, the primary wiring can measure both the
By selecting the reverse wiring low-voltage shield, 10kV reverse wiring dielectric loss
measurement can be carried out on the upper capacitor C1 without removing the busbar.
As shown in the figure below: the busbar is grounded, the upper end of C1 is not
disassembled, the lower end of C1 is connected to the high-voltage core wire, and the end
of C2 is connected to the Cx core wire. The instrument adopts the reverse wiring /10kV/M
measurement mode, which can measure the capacitance and dielectric loss of C11 and
10
VII.Brief Introduction to Instrument Functions
The Instrument Panel is Shown in Figure 9 Below
2. RS232 interface - used to connect computer, control instrument and upload data.
connected by a special low-voltage cable during the positive connection. This cable
is single-layer shielded with a special crocodile clip, which is 8m long and connected
to the low-voltage end of the test product. This end is empty when reverse wiring.
4. Color touch screen display -640*480 resolution, control and display menu and
10. High voltage switch - responsible for the starting and closing of high voltage
power inside the instrument. Only when using external high pressure to close, the
11. Power socket -- AC 220V±10%, 50±1Hz power input port, with 5A fuse.
11
1 2 3 4 5 6 7 8 9 10 11 12
12
13.High voltage wiring diagram of instrument side
High voltage grounding shielded socket (the interior is already grounded, just plug it in))
The small black clip is high voltage shield, generally suspended (with high voltage)
The big red clip is connected to the high voltage part under test
13
display the starting interface in FIG. 10.
1) test frequency: 50.0± 5.0hz means 45/55hz dual frequency conversion test. Generally,
the field test is 50±5Hz or 60±5Hz. You can choose something else, of course. If you want
a single frequency test such as 50Hz, you can choose 50.0± 0.0hz. Here the frequency
2) test mode: reverse wiring means using reverse wiring mode to test products. Here is
the drop-down menu. You can choose forward connection, reverse connection, CVT
3) test voltage: 10000V represents the value of test voltage. Minimum 100V, maximum
10000V. The voltage value here can be set at will, such as 7892V.
4) high voltage selection: internal high voltage means the use of internal high voltage
External high voltage is only used when external high voltage power supply is connected.
5) standard selection: internal standard means the use of internal standard capacitance.
External standard is only used when external standard capacitors are used.
Note: external standard, external high voltage are used for large capacity or high voltage
dielectric loss. The general equipment of substations can be tested with the internal
14
standard and internal high voltage.
1)To test the modification of the frequency, click the "45.0±0.0Hz" scarlet letter with the
Enter the frequency value 50 directly from the keyboard and press OK. If you press OK, it
will be 50.0±0.0Hz.
If "45.0±0.0Hz" "±0.0" is modified, use a stylus or finger to click on the scarlet letter 0.0 to
appear the keypad, and then input 5.0 from the keypad and press OK.
2)The modification of test voltage is the same as the modification frequency. Use a stylus
or finger to click on the red letters in "test voltage 5000 V". The little keyboard pops up.
Enter the test voltage value of 10000 directly on the keypad, and press OK, the keypad
disappears. The value of the test voltage will become the value just entered from the
keyboard.
3 ) To modify the test mode, click "Reverse wiring" with a stylus or finger. Figure 12 is
displayed
15
Figure 12 mode selection modification diagram
All test modes are on the drop-down menu. Click on the mode you want to select.
Positive wiring, reverse wiring, CVT dielectric loss, low voltage shielding of reverse wiring.
4) The modification of high-voltage selection is the same as the modification of test mode.
With a touch-screen pen or finger click on the "internal high pressure", the drop-down
menu appears in the high pressure, external high pressure, click on the high pressure you
want to choose.
5) The modification of standard selection is the same as the modification of test mode.
With a touch-screen pen or finger, click on "Internal Standard" to display the "Internal
Standard" and "External Standard" in the drop-down menu. Just click on the standard you
want to select.
1) One end of the grounding wire is connected with the grounding column of the
instrument, and the other end is connected with the earth reliably to ensure that the
2) Positive wiring: insert the high-voltage cable plug into the HV socket of the back door,
clamp the red pliers at the other end to the high-end lead of the tested product, and hang
the black pliers or clamp them on the red pliers. Insert the CX low voltage cable into the
CX socket, and the red clamp at the other end, the Black clamp suspended or connected
16
with the shielding device at the lower end of the sample.
3) Reverse wiring: insert the high voltage cable plug into the HV socket of the back door,
clamp the red pliers at the other end to the high-end lead of the tested product, and
suspend the red pliers or connect the shielding device. Cx socket is not used.
5.Testing process:
Select the test items, test frequency, test voltage, etc., as shown in FIG. 13.
After making sure the wiring is correct, click "start instrument test" with a stylus or finger.
The instrument adopts dual frequency conversion test. Figure 14 is 55Hz test and figure
15 is 45Hz test.
17
FIG. 14 instrument 55Hz test
6.Test results:
18
Fig. 16 Test results of reverse wiring
The capacitance value and dielectric loss value of C1 and C2 are the values of C1 and C2
of CVT equipment.
The capacitance and dielectric loss of CZ are the total capacitance and dielectric loss of
CVT equipment.
1) Test data can be stored after testing. Click the "store" button with the stylus or finger,
19
and figure 18 appears
Click "1" in the data number box with a stylus or finger, and a keypad will appear. Figure
Enter the storage number and press OK. The data number 5 appears in the data number
box. Press "store", the instrument will beep, and then return to the interface of test results,
2 ) Test data printing. After the test, click the "print" button with the touch pen or finger
20
under the test result interface to print the test data. Figure 20 below is a data printing
diagram.
3)The test. After printing or storing data, click the "exit" button with the touch pen or
finger, and the device will return to the initial startup interface.
following aspects:
When the red clip is used to connect the sample in the field measurement, the red clip
must be in good contact with the sample, otherwise the discharge of the contact point will
cause serious fluctuation of the data. Especially the oxide layer of the drain line is too thick.
Poor grounding can cause serious fluctuation of instrument protection or data. The paint
and rust on the grounding point should be scraped off, and zero resistance grounding
must be guaranteed!
Electromagnetic PT
Direct measurement of the lower coupling capacitance of CVT will result in negative
21
When measuring electromagnetic PT with end shielding method, negative dielectric loss
occurs due to "T-shaped network interference" caused by dampness. Blow-dry the three
skirts of ceramic sleeve and terminal disc below. It can also be measured by conventional
Air humidity makes dielectric loss measurement abnormally increased (or decreased or
even negative) and unstable, and shielding rings may be added when necessary. This
method is controversial because the electric field distribution of the test sample is
The output voltage of the generator is unstable when it is supplied with power, and there
are high voltage burrs, which may lead to the wrong connection of 380V protection to start.
Several power filters can be connected in series at the output end of the generator to
6. Test Line
Because of long-term use, it is easy to cause hidden circuit break, short circuit of core wire
and shield, or bad contact of plug. Users should always maintain the test line.
When testing standard capacitors, full shielded plugs should be used to eliminate the
influence of additional stray capacitance, otherwise the accuracy of the instrument can not
be reflected.
should be suspended, otherwise additional stray capacitance and dielectric loss to the
After wiring, please choose the correct measurement mode (positive, negative and CVT),
22
distinguished between test method error and instrument error. When there is a problem,
the wiring can be checked first, and then whether it is a fault of the instrument.
9. Instrument failure
Use a multimeter to measure whether the test line is open, or whether the core line and
shield are short-circuit; 220V input power supply is too high or too low; and whether the
grounding is good.
dielectric loss by positive and negative wiring. If the results are correct, it can be judged
When the interference is very serious, the frequency conversion measurement can
get accurate and reliable results. For example, when measuring with 55Hz, the measuring
system only allows 55Hz signal to pass, and 50Hz interference signal is effectively
suppressed. The reason is that the measuring system can easily distinguish different
frequencies. The following simple calculation can illustrate the effect of frequency
selection measurement:
The sinusoidal waves with a difference of 1 times of the two frequencies are
superimposed together, and the high-frequency ones are interference with a amplitude 10
Y=1.234sin(x+5.678°)+12.34sin(2x+87.65°)
Calculate A=Y1-Y3=2.4559,B=Y0-Y2=0.2442,then:
φ=tg-1(B/A)=5.678° V= A2+B2/2=1.234
This is just the phase and amplitude of the low frequency part, and the interference is
suppressed. The measurement points of the actual waveform are up to tens of thousands,
23
2、The relationship between frequency and dielectric loss
There are two ideal models of RC series and parallel dielectric losses:series model
tgδ=2πfRC , Parallel model tgδ=1/(2πfRC) , tgδis proportional and inverse with the
frequency f.As shown in the figure, f has a great influence on the two models of perfect
direct ratio and perfect inverse ratio. However, the actual capacitor is a hybrid model with
Actual sample
Parallel model
f f
Low frequency dielectric loss High frequency dielectric loss curve or low
The instrument adopts automatic frequency conversion to measure one point at each side
of the interference frequency of 50Hz (45Hz and 55Hz), and then calculate the data at the
circuits, it is impossible for the medium in a single sample to cause the energy absorption
peak at low frequency, and the dielectric loss near the power frequency always changes
monotonously with the frequency. Therefore, this measurement method will not bring
obvious errors. In fact, the average of the first two dielectric loss values is very close, even
instrument has become a conventional instrument for dielectric loss measurement, and its
24
XI.Packing List of Instrument
1 Main engine 1
2 Power line 1
3 Ground lead 1
4 Fuse pipe 2
7 Print paper 1
25