2011-E-0141
DESCRIPTION OF CHANGE:
CL Application V2.1(B) for CR-VW 674, including new functions and error corrections, is ready to be released.
New Functions/Enhancement
Automatic import of endoscopic studies
Linkage of study information with DICOM MWM/PPS gateway
NAS data storage converted from DVD
Receiving Images from Console Advance V5.0, etc
DETAILS:
- NOTE
The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.
PAGE 1 of 2 K12N0028-6-J
ECN No. 2011-E-0141
1) Important information about SUPPLY OF THE ARTICLES:
Charge:
FREE OF CHARGE UNTIL (mmddyyyy):
WITH YOUR CHARGE
The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.
PAGE 2 of 2 K12N0028-6-J
ECN 2011-E-0141
CR-IR391CL/391V CL
1. Overview 2
2. Released Items 3
2.1. Released items for CR-VW 674 3
2.2. Released items for CR-IR 355V CL 3
2.3. Released items for CR-IR 391 CL 3
2.4. Released items for CR-IR 391V CL 4
3. Functional Enhancement 5
3.1. Functional Enhancement 5
3.2. Added optional functions 6
3.3. Error corrections 6
3.3.1. Error Correction (failures found at user’s site) 7
3.3.2. Error corrections (Failures found by design and QA groups) 11
4. Installation Procedures 13
4.1. Procedure for upgrading software to V2.1(B) 13
4.1.1. Software Versions 13
4.1.2. Upgrading Procedure 14
4.1.3. Procedure for reinstallation from OS 15
4.1.4. Other extended functions 15
Attachments
Appendix 1: Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
Appendix2: FCRView V2.1(B) Connection Setup for Windows 7 Client PC
Appendix 3: FCRView V2.1(B) Endoscopic Device Setting Procedure
Appendix 4: FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
Appendix 5-1: FCRView V2.1(B) NAS Quick guide
Appendix 5-2: FCRView V2.1(B) Restoration Procedure for the Case Images Cannot be Stored into NAS
Appendix 5-3: Procedure of Moving Study Output from Storage Disk to NAS
Appendix 6: FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
Appendix7: Installation Support Software Exposure Menu Import/Export Tool for Hobbit Operation
Manual
Appendix 8: Setting Procedure for Receiving Images from Console Advance V5.0
Appendix 9: FCRView V2.1(B) Operation of Database Restoration Tool
Appendix 10: FCRView V2.1(B) Setup Procedure for Replacing Patient Information at the Time of
Receiving DICOM Images
Appendix 11: Error Code Table
1/16
ECN 2011-E-0141
1. Overview
This ECN provides the announcement about upgrading software to V2.1(B) and related information.
◆Caution◆
Unless specified, the following 4 models are described as FCRView to represent in this document.
CR-VW 674
CR-IR 355V CL
CR-IR 391CL
CR-IR 391V CL
2/16
ECN 2011-E-0141
2. Released Items
Distrib
No. Product name Description Part No. Media
ute
CR-VW 674 Installer of Application DVD-
1 114Y5097001B02 ✓
Application Software–B V2.1 Software R
CR-VW 674 Installer of Client
2 114Y5097006B02 CD-R ✓
Client Application Software–B V2.1 Application Software
CR-VW 674 Reference guide for
3 114Y5097051A03 CD-R ✓
Reference Guidance–B V4.0 human
CR-VW 674 Key software for
5 DICOM ORDER MWM/MPPS Key activating DICOM 114Y5097127A00 CD-R ✓
Software V1.0 MWM/MPPS
Distrib
No. Product name Description Part No. Media
ute
CR-IR 355CL/355V CL Installer of Application
1 114Y5154001B02 DVD-R ✓
Application Software–B V2.1 Software
CR-IR 355CL/355V CL Installer of Client
2 114Y5154006B02 CD-R ✓
Client Application Software–B V2.1 Application Software
CR-IR 355V CL Reference guide for
3 114Y5154063A03 CD-R ✓
Reference Guidance–B V4.0 animals(WW)
CR-IR 355V CL Reference guide for
4 114Y5154064A03 CD-R ✓
Reference Guidance–B V4.0 animals(US)
CC-WS674
Key software for
CR-IR 355CL/355V CL
5 activating DICOM 114Y5154127A00 CD-R ✓
DICOM ORDER MWM/MPPS Key
MWM/MPPS
Software V1.0
Distrib
No. Product name Description Part No. Media
ute
CR-IR 391CL Installer of Application DVD-
1 114Y2129001B02 ✓
Application Software–B V2.1 Software R
CR-IR 391CL Reference guide for
2 114Y2129052A03 CD-R ✓
Reference Guidance–B V4.0 human
CR-IR 391CL Reference guide for
3 114Y2129053A03 CD-R ✓
Reference Guidance–B V4.0 human in Chinese
CR-IR 391CL Key software for
4 DICOM ORDER MWM/MPPS Key activating DICOM 114Y2129127A00 CD-R ✓
Software V1.0 MWM/MPPS
3/16
ECN 2011-E-0141
Distrib
No. Product name Description Part No. Media
ute
CR-IR 391V CL Installer of Application DVD-
1 114Y2129201B02 ✓
Application Software–B V2.1 Software R
CR-IR 391V CL Reference guide for
2 114Y2129061A03 CD-R ✓
Reference Guidance–B V4.0 animal
CR-IR 391CL Key software for
3 DICOM ORDER MWM/MPPS Key activating DICOM 114Y2129325A00 CD-R ✓
Software V1.0 MWM/MPPS
Distrib
No. Product name Description Part No. Media
ute
Menu Import Export Tool For Hobbit
1 Menu Import Export Tool 114Y5154709A00 CD-R ✓
V1.0
4/16
ECN 2011-E-0141
3. Functional Enhancement
5/16
ECN 2011-E-0141
editable)
Supported Supported languages by Windows 7 are
languages by added.
8 Windows 7 ・Simplified Chinese ✓ ✓ ✓
・Traditional Chinese
・Korean
DB restoration tool The DB restoration tool, which was released in
9 a CD, has been mounted in the system folder ✓ ✓ ✓ ✓
of FCRView.
The error corrections which have been made by the following HotFix are included in V2.1(B).
V2.0(B) HF0001
V2.0(B) HF0002
V2.0(B) HF0004
V2.0(B) HF0010
6/16
ECN 2011-E-0141
Measures Enable FCRView to import Type2 tag information necessary for DB registration. Leaving
alert logs alone, no exception error occurs if tag information does not exist.
Number QCN-002146/FMS2-10-073
7/16
ECN 2011-E-0141
Number QCN-005083/FMS2-10-223
QCN-001762/FMS-10-543
QCN-005013/FMS2-10-121
QCN-005152/FMS2-11-101
SCN-000098/FMS-09-662
5)Even after the filmless function is set up in the user utility, film is output after the study is verified again
Phenomenon Film is output when the study is verified after the modification regardless of the filmless
function is currently set when the patient information, which has the film output setup history,
has been changed.
Causes Film is output when a study is verified again with the filmless function set.
When the [Automatic image output] of the [Printer] in the user utility is not checked, the value
of [Upon verification of a study] of the [Timing of output] is on.
Measures The program has been revised to enable the setup of [Upon verification of a study] of the
[Timing of output] in the [Printer] separately regardless of the settings of [Automatic image
output].
According to this change, the indent of the checkbox in the [Upon verification of a study] is
modified along with the [Automatic image output].
Number QCN-001866/FMS-10-678
QCN-005077/FMS2-10-198
QCN-005179/FMS2-11-131
Number QCN-005049/FMS2-10-174
8/16
ECN 2011-E-0141
Number QCN-002069/FMS-10-A86
QCN-005168/FMS2-11-107
QCN-005344/FMS2-11-326
9) The status stays [Delivering] when the queue is initialized when outputting the data to the work disk
Phenomenon The status stays [Delivering] when the queue is initialized while the data is being output to
the work disk.
Causes Clear Output Queue process in the Service Utilities does not contain the deleting the output
queue to the work disk.
Measures The Clear Output Queue process has been revised to delete the output queue to the work
disk.
Number TIMS9586 /QCN-001893
10) The message box hides behind the “Read Image” window when selecting
Phenomenon Because the exposure menu button is clicked when the “Read Image” window is opened, the
warning dialog box hides behind it, which disables to click the [OK] button as if it freezes up.
Causes The message dialog, which does not contain particular settings to be displayed as a general
dialog, hides behind the “Read Image” window, which has been setup as the foremost
window.
Measures The display setting of the “Read Image” window has been changed to display the message
dialog on the top of all the other windows.
Number QCN-005048/FMS2-10-168
12) Color of ultrasound images stored in storage DVD is unusually changed in display.
Phenomenon The moving images turn bluish when the data, which has been transferred from FAZONE to
9/16
ECN 2011-E-0141
the server, is imported from DVD.
Causes 1 study sometimes consists of multiple images. Usually, Raw (RGB) data is written in the
storage DVD in JPEG Lossless (color space RGB). In case, however, JpegLossy(color
space YCbCr) images are included in the same study, they are stored in JPEG Lossy.
Because the Photometric Interpretation of DICOM header stays in RGB, the images in
YcbCr are recognized as RGB, changing their tone of color.
Measures In case of the images whose the compression format has been changed (e.g. JPEG
Lossless→JPEG Lossy), adjust the Photometric Interpretation of DICOM header to the
color space of the compression format .
It is revised that Lossless→Lossy and Lossy→Lossless compression format changes have
no problem.
Number QCN-005278/FMS2-11-269
QCN-005290/FMS2-11-238
15) When activated from the main window, white-tinged images are displayed on the viewer
Phenomenon The images turn white-tinged when the viewer is activated from the main window to display
them. The images are whiter than the ones displayed on the viewer activated from the study
utility.
10/16
ECN 2011-E-0141
Causes LUT is obtained each time when displaying images on the viewer. LUT is not obtained when
the network condition is unstable, making the images turn white-tinged.
Measures The program has been revised to obtain LUT only for the first time to display the image.
Internally cached LUT is used later. In case LUT is not obtained for the first activation, the
default LUT is used instead。
Number QCN-005219/FMS2-11-169
18) Study data import error occurs immediately after the activation
Phenomenon After the study data is imported from the Ultrasound equipment of Toshiba, the error
message (#49572) describing the failure of study data import is displayed each time
FCRView is activated.
Causes Xml file remains in the temporary data folder when the DICOM data fails in importing. The
data in this file is read when the application is activated, causing the error and error
message.
Measures Xml file is transferred from the temporary data folder to the other folder when the error
occurs.
Number QCN-001965/FMS-10-835
QCN-001621/FMS-10-340
QCN-001623/FMS-10-355
CQ555 Phenomenon It takes long to activate [Completed] button after the image import is completed
when connected to XG1.
The program has been revised to activate the button immediately after the
Measures
exposure is completed.
11/16
ECN 2011-E-0141
CQ556 Phenomenon ID Input window does not reflect the changes on the date of birth made on the
Import window.
Measures The program has been revised.
CQ557 Phenomenon The following phenomenon occurred on the patient whose setting of the neutered
CQ558 state is [Unknown] or [UnAltered]
CQ559 Although the neutered state setting is specified as [Unknown] when the patient
information is changed, only the viewer can reflect it.
The progress bar remains to be displayed when the viewer is activated after the
study data is imported
The study list does not reflect the neutered state setting specified as [Unknown]
when the study data is imported.
Measures The program has been revised to reflect the correct information.
CQ568 Phenomenon The images, which have been stored in the storage disk after the verification, are
verified again after they are rotated by 90 degrees. Although the rotated images
are imported for reference in the study utility, the images are not rotated when
imported via DICOM offline for reference.
Measures The program has been changed to import the images after rotated.
CQ578 Phenomenon Two [0002,0010] tags are written when DICOM file output is executed.
Measures The program has been revised.
CQ591 Phenomenon 39 image frames out of 40 are imported under the high-load DICOM transfer
environment.
Measures The program has been revised.
CQ596 Phenomenon Referenced Transfer Syntax UID in File does not exist in DICOMDIR of created
PDI.
Measures The program has been revised to register the corresponding tag.
CQ657 Phenomenon [Failed] is displayed on the thumbnail image when a series of 1000 images are
dropped in the image cart.
Measures The software has been revised to display the thumbnail images normally.
CQ658 Phenomenon Transferring thumbnail images after the thumbnail image output does not put the
fail names in a sequential order.
Measures The program has been revised.
CQ708 Phenomenon 20091 error occurs in case there is no images left in local HDD during the study
data transfer to NAS, requiring restart of the system to make the process of the
data transfer continue.
The software has been revised to make the process continue without a restart.
Measures Output of the 20091 error to the indicator and to the log remains for the purpose of
a notice for the users.
CQ803 Phenomenon Thumbnails change their color unusually when particular PDIs (DICOM file) are
imported.
Measures The program has been revised.
12/16
ECN 2011-E-0141
4. Installation Procedures
Procedure Reference
Upgrading application Refer to 4.1.2 in this document
software Appendix 1: Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR
391CL/CR-IR 391V CL V2.1(B)
Appendix2: FCRView V2.1(B) Connection Setup for Windows 7 Client
PC
Appendix 3: FCRView V2.1(B) Endoscopic Device Setting Procedure
<CR-VW 674>
Software version Upgrade procedure
Upgrade to V5.0, and then upgrade to V2.1(B).
V1.0 (+ HotFix)
* See ECN 2009-E-0166 for the upgrade procedure to V5.0.
Upgrade to V5.0, and then upgrade to V2.1(B).
V4.0
* See ECN 2009-E-0166 for the upgrade procedure to V5.0.
13/16
ECN 2011-E-0141
Upgrade to V5.0, and then upgrade to V2.1(B).
V4.1
* See ECN 2009-E-0166 for the upgrade procedure to V5.0.
V5.0 (+ HotFix) Direct upgrade is available.
V1.1(B) (+ HotFix) Direct upgrade is available.
V2.0(B) (+HotFix) Direct upgrade is available.
Upgrading to V2.1(B) when installing the unit with the earlier application software version
14/16
ECN 2011-E-0141
Because the default settings have been changed from the earlier application software versions, refer
to the service manual to perform the procedure for upgrading application software. Note that the
following operations need to be added to the procedure.
Upgrading to V2.1(B) is required in the case of the factory setting application software of the unit is
earlier version.
Perform “Initialize configuration files” in the InstallLauncher when the upgrading installation has
been completed.
◆Information◆
FCRView V2.1(B) satisfies a market requirement to use NAS as an alternative storage device
instead of DVD.
Setup procedure for linkage of study information with DICOM MWM/ MPPS gateway
Refer to Appendix 6: FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure for further information.
Refer to Appendix7: Installation Support Software Exposure Menu Import/Export Tool for Hobbit
Operation Manual for the instruction for use of the tool.
◆Information◆
To avoid the manual operation to edit one exposure menu at a time in the User Utility, the tool,
which exports the CSV of the exposure menu and imports the edited menu, is provided for service
engineers.
[Menu Import Export Tool For Hobbit V1.0 (114Y5154709A00)]
This functions supports the following SOP Class and Transfer Syntax.
SOP Class: -Computed Radiography Image Storage
Transfer Syntax: -Implicit VR Little Endian
15/16
ECN 2011-E-0141
-Explicit VR Little Endian
-JPEG Baseline (Process 1): Default Transfer Syntax for Lossy JPEG 8 Bit
Image Compression
-JPEG Lossless, Non-Hierarchical, First-Order Prediction (Process 14
[Selection Value 1]): Default Transfer Syntax for Lossless JPEG Image
Compression
Setup Procedure for Replacing Patient Information at the Time of Receiving DICOM Images
The function replacing the patient information when importing DICOM has been added in FCRView
V2.1(B). The study information output from the equipment with DICOM connection is replaced in the
corresponding patient information in FCRView on the basis of patient ID.
Refer to Appendix10: Setup Procedure for Replacing Patient Information at the Time of Receiving
DICOM Images for further information.
4.1.5. Cautions
4.1.6. Restrictions
*: The destination of the automatic data transfer when the local HDD is exhausted.
16/16
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
1. Outline .......................................................................................................................................... 3
1.1. Applicable model ..................................................................................................................... 3
2. Upgrading Procedure.................................................................................................................. 3
2.1. Upgrading procedure for server PC ........................................................................................ 3
2.1.1. Backing up the report format ............................................................................................. 3
2.1.2. Backing up the setting file for free drawing function .......................................................... 4
2.1.3. Noting down the setting contents for CT/MR..................................................................... 5
2.1.4. In the institutes where NAS has been installed for the purpose of the expantion of HDD . 5
2.1.5. Uninstalling the previous version of CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR
391V CL ............................................................................................................................ 5
2.1.6. Deleting Internet Explorer cache ....................................................................................... 6
2.1.7. Inactivating the UAC function (Windows Vista/ 7) ........................................................... 13
2.1.8. Installing .NET Framework 3.5, NET Framework 3.5Language Pack, and Microsoft Chart
Control ............................................................................................................................ 17
2.1.9. Installing the Server application ...................................................................................... 21
2.1.10. Restoring the backup of report format............................................................................. 23
2.1.11. Overwriting the setting file for free drawing function........................................................ 23
2.1.12. Changing Wallpaper (Windows XP only)......................................................................... 24
2.1.13. Setting up endoscopic device ......................................................................................... 24
2.1.14. Additional procedures when upgrading PCs, to which "DICOM US Storage SCP Key
Software" or "DICOM CT/MR Storage SCP Key Software" has been installed, to V2.1(B)25
2.1.15. Activating the UAC function (Windows Vista/7)............................................................... 25
2.1.16. Operation confirmation of the server PC ......................................................................... 26
2.2. Upgrading procedure for client PC ........................................................................................ 28
2.2.1. Backing up the setting file for free drawing function ........................................................ 28
2.2.2. Uninstalling the previous version of CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR
391V CL .......................................................................................................................... 28
2.2.3. Deleting Internet Explorer cache ..................................................................................... 28
2.2.4. Turning off the UAC function (for Windows Vista) ........................................................... 28
2.2.5. Installing Microsoft .NET Framework3.5 [Windows XP and Vista only] ........................... 28
2.2.6. Installing Microsoft Chart Control [all the OS (Windows XP, Vista, 7)]............................. 30
2.2.7. Installing the Client application........................................................................................ 31
2.2.8. Connection setup for Windows 7 client PC ..................................................................... 32
2.2.9. Overwriting the setting file for free drawing function........................................................ 32
2.2.10. Setting the link target (Client PC) .................................................................................... 32
2.2.11. Turning on the UAC function (Windows Vista/7) ............................................................. 33
1/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
2/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
1. Outline
Since "Net Framework3.5" has to be installed in V2.1(B), upgrading procedure has been changed.
The procedure is described in this appendix.
2. Upgrading Procedure
2.1. Upgrading procedure for server PC
2.1.1. Backing up the report format
Reference
If the report format has been changed, since the settings before upgrading cannot be taken over, the
setting file needs to be backed up before uninstalling the application.
Terminate the application if running.
3/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
Copy the file "C:\Program Files\FujiFilm\Hobbit\System\ FDDSettingInfo.ini" onto the desktop, for
example.
4/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
1) Launch the Service Utility, and click [Connected equipment] - [Server (this equipment)] - [DICOM
setting] to open the "Set up DICOM." dialog.
→ Note down the setting contents.
2.1.4. In the institutes where NAS has been installed for the purpose of the expantion of HDD
The service settings for the expansion of HDD need to be configured again after upgrading since
the previois settings cannot be taken over. Note down the setting contents before upgrading.
Refer to “2.3 Setting in the Service Utility” under Appendix 10 “Settings for connection of the NAS”
in the Service Manual (014-227-05E) for the setup procesure.
2.1.5. Uninstalling the previous version of CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR
391V CL
Caution
Confirm that there is no unconfirmed study or study being output.
1) Start up the application, and terminate the server/client application and the installation wizard.
2) Insert the installation CD into the DVD drive. Select [My Computer] from the [Start] menu at the left
bottom of the screen. Double-click [DVD RW drive (D:)], and then [INSTALLLAUNCHER].
5/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
3) Click.
4) When the “Uninstsll or Change program” of "Programs and features" dialog opens, select [FCR
Application] and click [Uninstall].
(Ex. Vista)
1)
5) Remove the installation CD from the DVD drive and restart the PC.
Caution
The PC must be restarted.
When the PC has restarted, enter "fujicr" for Username and "system" for Password, and log in.
If the PC is not restarted, the application installation may fail.
*Multiple users may be displayed in some cases. Select also “fujicr” and enter “system” for Password.
*Since application has been uninstalled, the screen has no wallpaper in its background (turned in black).
6/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
Windows Vista / 7:
2) Clear cache with the procedures according to the version of Internet Explorer.
7/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
2. Click [Delete Files...] under [Temporary Internet files] on the middle of the dialog.
→The "Delete Files" dialog opens.
3. Click the [OK] button without placing a checkmark to "Delete all offline content" to clear
cache.
8/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
6. Click the [OK] button on the bottom frame of the "Internet Options" window to return to the
Internet Explorer window, and then terminate the Internet Explorer.
9/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
4. Click [Yes].
7. Click the [Close] button at the bottom of the "Delete Browsing History" window to close the
"Delete Browsing History" window.
8. Click the [OK] button on the bottom frame of "Internet Options" window to return to the
Internet Explorer window.
10/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
11/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
3. In the "Delete Browsing History" window, place a checkmark to the following and remove the
checkmarks from the others.
- Temporary Internet files
- Cookies
- History
5. Click the [OK] button on the bottom frame of the "Internet Options" window to return to the
Internet Explorer window.
12/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
Windows Vista
1) Open [Control Panel] - [User Accounts].
2) Click [Turn User Account Control on or off], remove the checkmark from "Use User Account Control
(UAC) to help protect your computer", and click [OK].
3) When the "Windows needs your permission to continue" dialog opens, click [Continue].
Windows 7
1) Select “Control Panel” from the [start] menu.
The “Control Panel” window will open.
13/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
2) Click “View by: Category” on the right of the window and Select “Large icons”.
The Control Panel will turn to “Large icons” view.
14/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
6) Check to see that the following message displayed on the right of the window.
“Never notify me when:
• Programs try to install software or make changes to my computer.
• I make changes to Windows settings.
7) Click [OK].
User Account Control confirmation window will open.
8) Click [Yes].
Return to the Control Panel window.
15/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
16/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
2.1.8. Installing .NET Framework 3.5, NET Framework 3.5Language Pack, and Microsoft Chart
Control
Information
In V2.0(B), ".Net Framework2.0" was used, but in V2.1(B), it has been changed to ".Net Framework3.5".
Moreover, as "Microsoft Chart Controls for Microsoft .NET Framework 3.5" is used in measurement
summary form this version, perform the installation for this as well.
Software to be installed Windows XP Windows Vista Windows 7
.NET Framework 3.5*1 √ √ -
Microsoft Chart Control √ √ √
* √ : OS to which the software needs to be installed.
- : OS to which the software is equipped by default; and thus, does not need to be installed.
1
* In Windows 7, "NET Framework 3.5 is installed by default; therefore, only "Microsoft Chart Control"
needs to be installed.
4) Click .
* The figure here shows "CR-VW 674"; however, the name of the model to which the installation is
performed will be displayed.
5) Click .
6) Click .
17/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
7) Click [Next].
* The installed software is displayed in green and the software which has not yet been installed is
displayed in gray. (The figure here is the one when none of the software has been installed.)
(a) Installing ".NET Framework 3.5" (only for Windows XP and Windows Vista)
1. The "Microsoft .NET Framework 3.5 SP1 Setup" window opens.
→ Select "I have read and ACCEPT the terms of the License Agreement",
and click [Install >].
18/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
2. Installation starts. When the installation finishes normally, the "Setup Complete" window opens.
→ Click [Exit].
If you are requested to restarted the PC, restart the PC. And then, resume the procedure.
Information
When the PC has restarted, enter "fujicr" for Username and "system" for Password, and log
in.
Multiple users may be displayed in some cases. Select also “fujicr” and enter “system” for
Password.
(b) Installing "Microsoft Chart Control" (for Windows XP, Windows Vista, and Windows 7)
2. The "Welcome to the Microsoft Chart Controls for Microsoft .NET Framework 3.5 software
update" window opens.
19/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
→ Installation starts. When the installation finishes normally, the "The software update has
been installed successfully" window opens. Then, click [Finish].
20/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
8) The system returns to the installation launcher window. Confirm that a checkmark is on ".Net Framework
has already been installed.".
* The figure here shows "CR-VW 674"; however, the name of the model to which the installation
is performed will be displayed.
2) Click .
* The figure here shows "CR-VW 674"; however, the name of the model to which the installation is
performed will be displayed.
3) Click .
* The figure here shows "CR-VW 674"; however, the name of the model to which the installation is
performed will be displayed.
→ After a while, the “Welcome to the InstallShield Wizard for FCR Application” window opens.
21/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
4) Click .
→ The “Ready to Install the Program” window opens.
5) Click .
→ Installation starts. When the installation finishes normally, the "Installation Wizard Completed"
window opens.
6) Click .
7) Remove the installation CD from the DVD drive, and click [Yes].
→ The PC restarts.
22/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
Reference
If the settings for free drawing function had not been changed, and the setting file was not backed up
in 2.1.2, this step can be skipped.
23/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
After upgrading, the previous wallpaper is not taken over and the background turns into blue.
2) Click [Control Panel] – [Appearance and Themes] – [Change the desktop background] - [Display
Properties] dialog - [Desktop] tab.
1)
2)
3) Select [FCR_WallPaper], and click [Apply(A)].
1)
2)
2.1.13. Setting up endoscopic device
In applicable institutes, settings of Service Utility must be configured again even if endoscopic images are
to be imported in the conventional manner.
For details, see " [Appendix 3] V2.1(B) Endoscopic Device Setting Procedure".
3) Reference
With upgrading to V2.1(B), endoscopic function has been extended, and settings have been greatly
changed:
- Supports automatic import of endoscopic studies.
- Using the patient ID as a key, provides the patient information in FCRView to automatically imported
endoscopic studies.
- Supports connection with multiple endoscopic devices (up to 4).
24/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
2.1.14. Additional procedures when upgrading PCs, to which "DICOM US Storage SCP Key
Software" or "DICOM CT/MR Storage SCP Key Software" has been installed, to V2.1(B)
1) Launch the Service Utility.
2) Click [Connected equipment] -> [Server (this equipment)] -> [DICOM setting] -> [Storage SCP settings].
→ If settings have not been changed, leave them as they are.
3) Click [OK], click [OK] again in the "Server equipment settings" window, and close the Service Utility.
4) Reference
If the above procedure is not performed, multi-frame data cannot be imported. Performing the procedure
enables to import multi-frame data corresponding to each option.
5)
Windows Vista
1) Open [Control Panel] - [User Accounts].
2) Click [Turn User Account Control on or off], place a checkmark to "Use User Account Control (UAC) to
help protect your computer", and click [OK].
Windows 7
1) Open [Control Panel] - [User Account]‐[User Account Control Setting]
2) Adjust the slide bar to “Default-Notify me when programs try to make changes to my computer”,
and then click [OK].
3) Restart the computer.
25/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
26/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
27/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
Copy the file "C:\Program Files\FujiFilm\Hobbit\System\ FDDSettingInfo.ini" onto the desktop, for example.
2.2.2. Uninstalling the previous version of CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V
CL
Uninstall the application.
1) Double-click [Add/Remove Programs] in the [Start] menu - [Settings] - [Control Panel]. (with Windows
XP)
* Windows Vista: Double-click [Programs and Features] in the [Start] menu - [Control Panel] -
[Programs].
* Windows 7: Click [Programs and Features] in the [Start] menu - [Control Panel]
(- [Programs]). (Item in parenthesis is displayed when "Category" is selected for "View by".)
2) After the "Add or Remove Programs" dialog opens, select [FCR Client Application] and click [Uninstall].
3) After the uninstallation completes, restart the PC.
Caution
The PC must be restarted.
When the PC has restarted, enter "fujicr" for Username and "system" for Password, and log in.
If the PC is not restarted, the application installation may fail.
1) Insert the FCR Client Application media into the DVD drive of the client PC.
→ When the "InstallShield Wizard (InstallShield Wizard)" dialog opens, click [Cancel].
3) Right-click [DVD-RAM Drive (D:)] or [DVD-RW Drive (D:)], and select [Open].
28/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
→ Installation starts. When the installation finishes normally, the "Setup Complete" window opens.
6) Click [Exit].
→ The system returns to the desktop screen.
→ If you are requested to restart the PC, restart it.
29/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
2.2.6. Installing Microsoft Chart Control [all the OS (Windows XP, Vista, 7)]
2) Click [Next>].
→ The "Microsoft Software License Terms" window opens.
3) Place a checkmark to "I have read and accept the license terms.", and click [Next>].
→ Installation starts. When the installation finishes normally, the message "The software update
has been installed successfully." appears. Then, click [Finish].
30/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
2) Click [Yes].
→ After the InstallShield Wizard starts up, the "Welcome to the InstallShield Wizard for FCR Application"
window opens.
31/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
Click [Next].
→ The "Ready to Install the Program" window opens.
3) Click [Install].
→ Installation starts.
After a while, the installation completion window opens.
4) Click [Finish].
→ The installation completion window closes.
5) Remove the "Client Application Software" CD-ROM from the DVD drive.
Reference
If the settings for free drawing function had not been changed, and the setting file was not backed up in
2.2.1, this step can be skipped.
32/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
33/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
34/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
35/35
[ECN Appendix 2] Connection Setup for Windows 7 Client PC
2. Applicable model
Client PCs for the following applications:
- CR-IR 355CL/355V CL V2.1(B)
- CR-VW 674 V2.1(B)
1. Press the "Windows" key, and select "Control Panel" → "System and Security" → "Administrative
Tools" → "Task Scheduler".
→ The "Task Scheduler" window opens.
2. Select "Task Scheduler Library" in the "Task Scheduler" window. Right-click displayed "FCR_Client"
and select "Properties".
1/3
[ECN Appendix 2] Connection Setup for Windows 7 Client PC
3. Click the "Actions" tab in the "FCR_Client Properties (Local Computer)" window, select "Start a
program" from the following option, and click "Edit...".
2/3
[ECN Appendix 2] Connection Setup for Windows 7 Client PC
4. Enter the host name of the server in the text box for "Add arguments", and click "OK".
5. Click the "OK" button to close the "FCR_Client Properties (Local Computer)" window.
3/3
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
1. Outline
Endoscope function has been extended in FCRView V2.1(B) as follows:
- Supports automatic import of endoscopic studies.
- Using the patient ID as a key, provides the patient information in FCRView to automatically imported
endoscopic studies.
- Supports connection with multiple endoscopic devices (up to 4).
This appendix describes setting procedure for connection with endoscopic devices in FCRView V2.1(B),
including the functions mentioned above.
2. Applicable device
FCRView
New installation site of V2.1(B) which is equipped with endoscopic devices
Endoscope
FUJINON VP-4450HD
FUJINON VP-4450
FUJINON VP-4400
FUJINON EPX-3010
1/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
TCP/IP
CR-IR 355CL
FCRView FUJINON VP-4400
Note
- Do not connect a telephone line to the network connector on FCRView and the endoscopic device.
- Cables used for connection to FCRView and the endoscopic device should be of the type compliant
with the IEC950/UL1950 standards.
Note
When the option key and/or the application key is to be installed, it is required that the FCRView is
connected to the network.
Make sure that network connection has been established before performing the installation.
4. Click [Continue].
→ A window opens to indicate the completion of installation.
2/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
5. Click [OK].
→ The system returns to the desktop screen.
6. Remove the CD from the DVD drive.
4. Click [Continue].
→ A window opens to indicate the completion of installation.
5. Click [OK].
→ The system returns to the desktop screen.
6. Remove the CD from the DVD drive.
3/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
(4) In the "Endoscope Image Import" window, perform the following in accordance with the changed
contents.
4/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
- Set the first unit of endoscopic device (including the case that the application was upgraded to
V2.1(B) in an institute where an endoscopic device has been used with V2.0(B) or before)
→Select a setting shown in the list, and click "Change". Proceed to step (5)
- Add an endoscopic device
→Create an arbitrary receiving folder under the C drive, and click "Add". Proceed to step (5)
* Up to 4 devices can be added.
- Change a setting for existing endoscopic device
→Select a setting you wish to change from the list, and click "Change". Proceed to step (5).
- Delete a setting for existing endoscopic device
→Select a setting you wish to delete from the list, and click "Delete". Proceed to step (7).
5/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
(5) Configure each of the following settings in the "Adding (Changing) endoscopic image sending
devices" window.
6/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
7 Modality The string set for modality (0008, 0060) when performing automatic
import.
Generally, the default string, "ES", is used.
8 FPT connected user name Select the user name of the FTP server.
When setting up multiple endoscopic devices connected to the FTP
server, select different users (duplicate settings are invalid). The FTP
user name in the endoscopic device has to be set in accordance with
the user name set here. See "4. Procedure to set up endoscopic
device".
* In V2.1(B), when the endoscopic image import option is enabled, the following endoscopic device will
be set up.
However, you have to open the "Changing endoscopic image sending device information"
window once and click "Change" to enable this setting. If this setting has not been enabled, an
error will occur at the startup and FCRView will not start up.
■Default settings
Equipment name : FujinonES
Receive Folder : C:\FUJINON_Data
(The folder above will be created when V2.1(B) is installed.)
Connection method : FTP
Auto import flag : OFF
Flag for patient information replacement : ON
Modality : ES
FPT connected user name : FujinonES
(6) After inputting each item, click the "Add (Change)" button. To revoke the setting and go back to the
previous window, click the "Cancel" button.
(7) To continuously change settings, repeat operations from step (4).
(8) Click "Finish" in the "Endoscope Image Import" window.
7/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
Settings in "System" - ”Settings for endoscopic image import” in the Service Utility have been
changed as follows:
8/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
Caution:
Unless creating the folder, Error 20167 occurs when starting the server application.
When connecting FCRVIEW with Advancia (VP-4450), perform the following procedures.
Connecting Procedures
1. Confirm the settings on Advancia are as follows.
Note
If settings are different from the following, change them to the following.
Precautions
● Setting the Image Compression Ratio
Note
Be sure to select items other than "1/1(TIFF)".
The compression ratio for images taken with Advancia (VP-4450) can be selected from following.
- 1/1(TIFF)
- 1/5(JPEG)
- 1/10(JPEG)
- 1/20(JPEG)
When connecting to the FCRView, only "JPEG" can be used.
Note that, if set to "TIFF", images cannot be imported to FCRView.
9/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
Note
Be sure to select "Patient ID".
The “Display Info.” under “Display Setting” can be selected from the following.
- Patient ID
- Study number
When connecting to the FCRView, only "Patient ID" can be used.
Do not set to "Study number". Note that, if set to "Study number", studies cannot be imported to
FCRVIEW.
● Backing up to CF Card
If you are to set "Yes" for “CF Card Backup”, the CF card must be inserted.
Prepare a CF card (sold separately) supported by Advancia and insert into the CF card slot
beforehand.
Software Ver1.101-1215
MainCPU 0
MainDSP 0x00150001
MainBoardFPGA 237-0x8B05
OS 305
Driver 02.00.02
Image Processing Parameters 0.0009
FrontCPU Ver0.330
RearCPU Ver0.505
MD2503 Ver0.023
HIST Ver4.020
AFE Ver1.02
PATIENT PCB Ver0.00
10/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
4.2. EPX-3010
Change the Login Name to ” FPT connected user name” setup by the service UTL of FCRView.
When connecting FCRView with Justia (EPX-3010), perform the following procedures.
Connecting Procedures
1. In the "Adding endoscopic image sending devices" window of the Service Utility, change the following
item from the default settings.
Receive Folder: DCIM
2. Press <F9> key on the keyboard connecting to the endoscopic device.
→ The following window opens.
Precautions
● Setting Image Output Folder
1. Press the <Ctrl> key and the <F1> key simultaneously.
→ The "Network Setup" window opens.
2. Set the "Root Folder" (the folder to which the images taken will be output) to "DCIM".
Note
11/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
2. Confirm that the versions on the Justia side are the following or the later.
12/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
5. Confirmation procedure
Set and confirm automatic output.
* For details on setup procedure, see "Appendix 4 Settings for Automatic Output" under "MU:
Maintenance Utility" of the service manual of CR-VW 674/CR-IR 391CL.
Note
After receiving the image, do not access the receive folder from explorer.
Reference
If the receive folder is opened using the explorer with its View set to Thumbnails, the message "34925:
Unsupported images exist in the endoscopic image receiving folder." is displayed.
This is due to the "Thumbs.db"(hidden file) file which is automatically created in the receive folder by
Windows system.
13/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
3. Click .
→ The system returns to the "Main" window.
4. Click "Refresh".
5. Double-click the imported endoscopic study, and confirm that the images are displayed in the
"Viewer" window.
14/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure
Format Content
JPEG Lossy JPEG lossy compression images
DICOM Stream Abstract syntax: Secondary Capture Image Storage
*For details on setting or output confirmation, see "Appendix 1 Settings for Output of Thumbnail Images"
under "Appx IN: Installation – Appendix" of the Service Manual of CR-VW 674/CR-IR 391CL.
7. Precaution
None
15/15
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
1/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
2/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
1. Setting of the PC
NOTE:
The procedures described in this chapter are for installing applications on CR-VW 674
(FCRView). In the case of the models other than CR-VW674 (FCRView), please apply the
appropriate model names/codes accordingly when following the procedures.
3/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
*In case a dialogue “ Windows found problems with your computer’s startup options. Do
you want to apply repairs and restart your computer?” is displayed, click [repair and
restart], and then follow the procedure below.
In case the dialogue above is not displayed, skip to 8)
a) Press <F12> when “DELL” screen appears.
b) Select [Onboard or USB CD-ROM Drive] from the Boot Menu and press [Enter]
key.
c) “Press any key to boot from CD or DVD” will be displayed. Press [Enter] key.
d) “Installing Windows “ dialogue will be displayed. Select [My language is English]
e) A dialog to select languages will be displayed. Click [Next] without selecting any
language.
f) The “Install now” window will open. Click [Repair my computer] on lower left of
4/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
the dialogue.
9) Click [Next].
The “Choose a recovery tool” window will open.
5/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
11) Input “diskpart” from the keyboard and press the <Enter> key.
12) Input “select△disk△0” (“△” replace a single-byte space) and press the <Enter> key.
A message saying “Disk 0 is now the selected disk” will be displayed in the command prompt
window.
13) Input “clean” from the keyboard and press the <Enter> key.
14) Confirm a massage saying “Disk Part succeeded in cleaning the disk.” and close the
command prompt window by clicking its [x] mark.
The system will return to the “Choose a recovery tool” window.
6/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
7/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
8/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
8) Select “Custom(advanced)”. Then “Where do you want to install Windows?” dialogue box will
open.
.
9) Select “Disk 0 Unallocated Space” and click [Next], and then the OS installation processing
will start.
Upon completion of installation, the “Choose a user name for your account and …” window
will open.
9/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
10/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
13) Check the “Time zone”, “Date” and “Time” and click [Next]
The “Select your computer’s current location” window will open.
2) Click “View by: Category” on the right of the window and Select “Large icons”.
The Control Panel will turn to “Large icons” view.
11/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
12/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
6) Check to see that the following message displayed on the right of the window.
“Never notify me when:
• Programs try to install software or make changes to my computer.
• I make changes to Windows settings.
7) Click [OK].
User Account Control confirmation window will open.
8) Click [Yes].
Return to the Control Panel window.
13/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
2) Place a checkmark to “Always do this for software and games” and click “Run autoRcd.exe”.
The “Welcome Dell System Owner “ window will open.
3) Place checkmark to “Click here to prevent this information from being displayed again” and
click [Next].
The Dell Resource CD utility software will start running.
4) Click [Drivers]
14/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
NOTE:
Select appropriate version at “Rev.****” of each driver.
NOTE:
Select appropriate version at “Rev.****” of each driver.
NOTE:
Select appropriate version at “Rev.****” of each driver.
15/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
NOTE:
Select appropriate version at “Rev.****” of each driver.
NOTE:
For Client PC, the “InstallShield Wizard ” dialogue will be displayed. Click [Cancel]. In case
[Program Compatibility Assistant] dialog is displayed after clicking [Cancel], click [This program
installed correctly].
For Client, "Welcome to the InstallShield Wizard for FCR Client Application” dialogue is
displayed when [Cancel] is not selected. Then click [Cancel] and perform the following
procedure;
a) ”InstallShield Wizard Completed” dialogue is displayed. Then click [Finish].
b) ”Program Compatibility Assistant” is displayed. Then click [This program installed
16/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
correctly].
5) Double-click “dotnetfx”.
The installation start confirmation window will open.
6) Click [Yes].
In a while, the “License Agreement” window will open.
8) Click [OK].
9) Open folders in the following sequence: “D drive”, “dotNet”, “DNET1.1”, and “DNET1.1SP1”.
NOTE:
*For the Client PC, open [D:¥dotNet¥DNET1.1SP1].
17/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
18/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
INFORMATION:
Use following updated parameters to install Adobe Reader:
Step 2:
[“D:/Adobe Reader/English/AdbeRdr910_en_US_Std.exe”]
→[“D:/Adobe Reader/English/AdbeRdr940_en_US.exe”]
Step 3:
“Adobe Reader 9” → ”Adobe Reader 9. 4.0”/
Step 5‐Ⅰ:
[Remove the checkmark from “General” - “Check for updates”.]
→[Remove the checkmark from “Updater” - “Check for updates”.]
1.12. Defining the date format and character to be used as a decimal point
1) Select [Control Panel] from the menu.
2) Double-click “Regional and Language”
The “Regional and Language” window will open.
3) Click the “Formats” tab to select a language (region) to use from the pull-down menu in
“Current format”.
The “Customize Regional Options” window will open.
19/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
20/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
21/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
22/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
4) Click [Customize...] in “Notification area”. The “Select which icons and notifications appear on
the taskber” window will open.
5) Place a checkmark to “Always show all icons and notifications on the taskbar”
23/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
6) Click [OK].
7) Click [OK].
The system will return to the Control Panel.
8) Close the Control Panel
The system will then return to the desktop.
24/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
25/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
26/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
27/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
4) Remove the checkmark of check boxes of [Security messages] and [Maintenance messages]
and click [OK].
28/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
.
The “Local Disk (C:) Properties” window will open.
3) Select “Tools” tab and select [Defragmenter Now].
29/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
Setting procedures
(1) Click [Start]- [Search programs and files]. Enter “C:/ProgramData/FujiFilm/Hobbit/UserTemp”.
(2) Right-click “PCLock” and click [Pin to Taskbar”
30/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
31/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
2) [Control Panel] - [Network and Sharing Center] - [Change adapter settings]-[Local Area
Connection] - [Properties] - [Internet Protocol Version 4 (TCP/IPv4)] - [Properties]
・ Select [Use the following IP address] then set the following
IP address :Server PC [172.16.1.20]
:Client PC [172.16.1.21]*
NOTE:
*Client setting does not apply to CR-IR 391CL/CR-IR 391V CL.
32/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
NOTE:
This setting does not apply to CR-IR 391CL/CR-IR 391V CL.
33/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
34/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
35/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
36/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
2) Open “D:¥ReferrenceGuide¥English¥WW” folder and select all the displayed PDF files in it,
then right-click to select [Copy] from the menu appeared.
3) Right-click on the desktop blank area and select [Paste] from the menu appeared.
4) Close “WW” window and remove the “Reference Guide CD” from the DVD drive.
37/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
5) “Microsoft Software License Terms” dialogue is displayed. Place a checkmark to ”I have read and
accept the license terms” and click [Next >].
6) “The software update has been installed successfully” is displayed. Then click [Finish].
7) Remove the FCRView Application Software from the DVD drive.
8) Restart the PC.
38/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
2. Installing Application
NOTE:
The procedures described in this chapter are for installing applications on CR-VW 674 (FCRView).
In the case of the models other than CR-VW 674 (FCRView), please apply the appropriate model
names/codes accordingly when following the procedures.
3) An application selection dialogue box will be displayed. Click [Install .Net Framework]
39/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
4) Click [Install .Net Framework 1.1], and then click [Next]. (.Net Framework1.1 Activation will be
executed)
40/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
5) When returned to the Application selection dialogue box, click [Install SQL Server]
6) An installation language selection dialogue box will be displayed. Click [Other] from the list, and
the click [Next].
7) Installation of SQL Server will be executed. Press [Enter] key when “Press any key to
continue…” appears.
41/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
8) A confirmation dialogue box for restart will be displayed. Remove the installation media from the
DVD drive and click [Yes].
9) The PC is restarted.
10) The logon screen will be displayed. Make sure that the user name is “fujicr” and enter “system”
for password, and then click “->”.
11) Insert “CR-VW 674 Installation” (for 355: “CR-IR 355CL Installation”, for 391: “CR-IR 391CL
Installation”) media into the DVD drive.
12) “CR-VW 674 Setup” (for 355: “CR-IR 355CL Setup”, for 391: “CR-IR 391CL Setup”) dialogue will
be displayed. Click [CR-VW 674 Application setup] (for 355: [CR-IR 355 CL Application Setup],
for 391: [CR-IR 391 CL Application Setup]).
42/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
13) An application selection dialogue box will be displayed. Click [Install CR-VW 674 V x. x] (for 355:
[Install CR-IR 355 CL V x. x], for 391: [Install CR-IR 391 CL V x. x]).
(“x.x” refers to the version)
14) After a while, “Welcome to the InstallShield Wizard for FCR Application” dialogue box will be
displayed. Click [Next].
43/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
15) “Ready to Install the Program” dialogue box will be displayed. Click [Install].
16) After a while, “InstallShield Wizard Completed” dialogue box will be displayed. Click [Finish]. (If
the generation of DB failed during installation, installation is aborted and rollback is performed.)
44/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
17) “To continue installation, you need to restart the system” dialog box will be displayed. Click
[Yes].
18) After the PC is restarted, “[20102] Setup incompleted. Application key software is not installed”
dialogue box is displayed. Click [OK].
19) Remove the installation media from the DVD drive, and then insert it again.
20) Click [Start] - [Computer], then double-click [DVD Drive (D:)].
21) Double-click “InstallLauncher.exe “ located right under CDROOT.
22) “CR-VW 674 Setup” (for 355: “CR-IR 355CL Setup”, for 391: “CR-IR 391CL Setup”) dialogue
box will be displayed. Click [CR-VW 674 Application setup] (for 355: [CR-IR 355CL Application
setup], for 391: [CR-IR 391CL Application setup]).
45/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
23) An application selection dialogue box will be displayed. Click [Install Configuration files].
24) The next dialogue box to select the type of the facility will be displayed. Select [Hospital /
Medical Clinic] and click [Next].
NOTE:
* In the case of CR-IR 355V CL/CR-IR 391V CL, Select [Veterinary / Hospital].
46/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
25) The next dialogue box to select a language will be displayed. Select [English(Not in USA)] from
the language list and click [Next].
26) The next dialogue box will be displayed. Select [Color] from the list of the monitor and click
[Next].
27) Installation will be completed and the application selection dialogue box will return. Click [Install
Menu / Parameter files].
47/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
28) The Next dialogue box to select a language will be displayed. Select [English(Not in USA)] from
the language list and click [Next].
29) A dialogue box inquiring whether to use “FNC” will be displayed. Select [ON] then click [Next].
30) Installation will be completed and the application selection dialogue box will return. Click
48/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
[Finish].
NOTE:
This setting does not apply to CR-IR 391CL/CR-IR 391V CL.
49/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
50/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
51/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
2.4. Installing Veterinary Image Processing Software (For CR-IR 355V CL (Server
PC)/CR-IR 391V CL)
1) Insert the following CD attached to the product respectively into the DVD drive.
・ [CR-IR 391V Image Processing Software (114Y2129009AXX)]CD
・ [CR-IR 346V Image Processing Software (114Y5342705AXX)]CD
2) Click [Start] ‐[Computer], and then double-click [D drive].
3) Double-click “OptionVeterinary.exe “ located right below CDROOT.
4) “CR-IR ***V Image Processing Software” dialogue will be displayed. Click [Continue].
5) “Successful Application key installation” dialogue will be displayed. Click [OK].
6) Remove “V Image Processing Software” CD from the DVD drive.
52/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
53/53
[ECN Appendix 5-1] Procedure of Moving Study Output to Storage Disk to NAS
1/4
[ECN Appendix 5-1] Procedure of Moving Study Output to Storage Disk to NAS
IV. Click [Add]. Setting network of the NAS. The NAS network setting is the same to [HDD](Service
Manual 2.3 Setting in the Service Utility).
V. Change the auto backup destination folder.
Click [DBAutoBackup]. Open the [Image save setting] window(Ⅵ).
VI. Click the [Browse]. Select the auto backup destination folder.
* It is necessary for a folder of NAS to be set as a network drive. The setting of the network
drive refers to a service manual.
2/4
[ECN Appendix 5-1] Procedure of Moving Study Output to Storage Disk to NAS
1. Select the folder that it’s name is output date to import the study.
The NAS manages the study as unit of output date.
3/4
[ECN Appendix 5-1] Procedure of Moving Study Output to Storage Disk to NAS
Off(Unavailable)
Free space
The Disk icon of “Indicator” represents free space of NAS.
The 1st warning < Free space
4/4
[ECN Appendix 5-2] FCRView V2.1(B) Restoration Procedure for the Case Images Cannot be Stored in NAS
1. Introduction
A trouble sometimes occurs in the operation of storing images into NAS (as a substitution for storage
disk), and images cannot be stored. This appendix describes the processes of handling this trouble.
3. Restoration procedure
* When images could not be stored into NAS even after performing the steps (1) through (4)
described above.
(1) Start up the FCRView server application.
(2) Delete the output cue to be output to NAS.
(3) Open the NAS utility window, click [Internal information reconfiguration], and take a
note of the save destination path of the present month's label name in the [Internal
information reconfiguration] dialog.
(4) Terminate the FCRView server application and restart the PC.
(5) Terminate the FCRView server application.
(6) Open the save destination path, which was noted down in step (3), using the Explorer,
and delete "DICOMDIR".
(7) Start up the FCRView server application.
(8) Open the NAS utility window and click [Internal information reconfiguration]. Then,
select the present month's label name in the [Internal information reconfiguration]
dialog and click the [Start] button.
1/2
[ECN Appendix 5-2] FCRView V2.1(B) Restoration Procedure for the Case Images Cannot be Stored in NAS
2/2
[ECN Appendix 5-3] Procedure of Moving Study Output to Storage Disk to NAS
1. Introduction
The procedure of moving study information which has been output to the storage disk with V2.0(B) or
before to NAS is described in this appendix.
However, the study information which has been output to the work disk cannot be moved to NAS.
2. Work procedure
[Preparation]
- Prepare the storage disk to be moved to NAS.
- Specify NAS as the output destination under [Connected equipment]->[BackupNAS] in the Service
Utility.
[Procedure]
1. Finish the Hobbit server application.
2. Start C:\Program Files\FujiFilm\Hobbit\System\DVDMoveToNASTool.exe.
3. Select the drive for the storage disk to be stored from the combo box named as DVD Drive
Select.
4. Insert the storage disk into the selected drive.
5. Select the moving destination of NAS drive from the combo box named as NAS Drive Select.
6. Click the [START] button.
7. The progress is displayed in the progress bar. Confirm that "Process Complete. Please check
the C:\DVDMoveToNASToolForm.log." is displayed in the message box when the process
finishes.
8. Open C:\DVDMoveToNASToolForm.log and confirm that there is no error between [Date]
ProcessSTART and [Date] ProcessEND in the log.
1. Outline
This appendix describes settings for linkage of study information with DICOM MWM gateway and
settings for linkage of study result information with DICOM MPPS gateway for FCRView V2.1(B).
A part of the settings for MWM study information linkage will be common to the settings for MWM patient
information linkage. Moreover, as the DICOM AE title of FCRView for MWM service can be set from
V2.1(B), this appendix also describes about that.
1/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(3) Click the “Linkage of MWM study information” button.
(4) Configure settings in the following table in the “Linkage setting of MWM study information” window.
2/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
* Equipment name, Host name, DICOM AE title, IP address, and Port number will be common to the
settings for MWM patient information linkage.
(6) The system returns to the linkage function setting window. Click the “Back” button to return to the
Service Utility top window.
3/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(7) Click the “Connected equipment” button.
4/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
5/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(9) Click the “DICOM setting” button.
6/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(10) Configure settings in the following table under the “MWM SCU setting” tab.
Caution) AE title and Timeout value will be common to the settings for MWM patient information linkage.
7/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
3. Setting up the MPPS Linkage
3.1. Setting up connection in the Service Utility
8/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(3) Click the “MPPS linkage” button.
(4) Configure settings in the following table in the “MPPS linkage setting” window.
9/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
Caution) "Supported" of MPPS linkage support is checked by default. For institutes where MPPS linkage is
not used, be sure to remove the checkmark.
(6) The system returns to the linkage function setting window. Click the "Back" button to return to the
Service Utility top window.
10/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
11/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(8) Click the "Server (this equipment)" button.
12/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(10) Configure settings in the following table under the "MPPS SCU setting" tab.
13/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
4. Operation Confirmation
14/14
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
-1–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
Change history
Edition Changed contents
1st Create 1st edition
-2–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
Contents
1. OUTLINE ..........................................................................................................................................4
-3–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
1. Outline
1.1. Content
There are an enormous number of X-ray exposure menus at institutions where the ordering equipment is
installed. Some institutions may register over 10,000 menus and it would take a considerable amount of
time and labor to register a large number of menus one by one at such institutions.
The exposure menu import/export tool is a support tool to reduce such burden. The tool extracts the
exposure/study menu data from the exposure menu database (DB) and exports them to the Microsoft
Excel file, making it possible to display/edit the menu list. And the tool imports the edited exposure/study
menus.
This tool is an installation support tool for service personnel and will not be released to users.
* Other than the Japanese environment, the operation under the English environment has been
confirmed. However, there is a restriction for the use (see “3.3.1. Import process”).
* Make sure the file format of the Excel file to be used can be opened with the essential software.
For example, the excel format file which is not compatible with Excel 2003 cannot be opened under the
environment in which Excel 2003 has been installed. Therefore, please confirm that before operating
the file.
* Operation under the Chinese, Taiwanese, and Korean environments has been confirmed using
Windows 7 and Excel 2003. We do not guarantee the operation with other versions.
-4–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
2. Setup procedure
2.1. Requirements for setup
- PC which fulfills conditions in 1.2 Operating conditions and in 1.3 Essential software
- Menu Import Export Tool for Hobbit application software
1. Insert the Menu Import Export Tool CD-R into the CD-ROM drive.
7. When the Setup completion dialog is displayed, click [OK] to complete the setup.
Do not install the tool into the PC where the Hobbit has been installed.
Be sure to take the backup of the database (DB) files (MenuData.mdb, XconParam.mdb,
Display.mdb and Selector.mdb) whenever you perform conversion, considering the
possible failure of the conversion with the tool and creation of defective files.
Operating target DB files of the tool are as follows:
MenuData.mdb
XconParam.mdb
(DisplayData.mdb)
These DB files are located in the following folder:
C:\ProgramData\FujiFilm\Hobbit\Param
* If the OS is Windows XP, mdb files are located in the following folder:
C:\Documents and Settings\All Users\Application Data\ FujiFilm\Hobbit\Param
Do not replace the database files while the application is running.
-6–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
3. Operation descriptions
3.2. Windows/Screens
1 11
3 4
12 13
7 8
9 10 15
14
3.3.1 Import
1. Start up the tool
2. Specify the import source Excel file (Figure 1 (5) → Select the file)
3. Specify the import destination folder (Figure 1 (6) → Figure 2)
4. Specify the log file (Figure 1 (11) → Figure 2)
5. Specify Check/Import (Figure 1 (7))
6. Setup language (Figure 1 (8))
7. Start the import process (Figure 1 (9))
Cancel (Figure 1 (13))
8. Confirm the log (Figure 1 (11))
-9–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
9. Exit (Figure 1 (15))
When the import process is interrupted, inconsistency may occur among the database files at the import
destination.
Never use the parameter file created by interrupting.
Moreover, do not use the database file created by interrupting for re-importing process.
Backup the database files before executing the import process, and use them for re-import process.
On English environment, when a multi byte character is included in a setting value of a field in the import
source Excel file, it may not be detected as an abnormal value. When creating or editing the import source
Excel file, be sure to do so in the English environment in order to prevent multi byte characters from being
input and imported.
3.3.2 Export
1. Start up the tool
2. Specify the export source folder (Figure 1 (1) → Figure 2)
3. Specify the export file (Figure 1 (6) → Select the file)
4. Specify the log file (Figure 1 (11) → Figure 2)
5. Start the export process (Figure 1 (3))*
Cancel (Figure 1 (13))
6. Confirm the log (Figure 1 (4))
7. Exit (Figure 1 (15))
*Close the file specified in the step (3)"Specify the export file" before performing the step
(5)"Start the export process". Moreover, make sure that read-only attribution of the MDF files
existing in the export source folder has been released.
If an existing file is specified as an export destination file when performing export, the existing file
is deleted after the overwrite confirmation dialog is displayed, and then menu data are exported to
a new file with the same file name. When specifying an existing file, close the existing file before
starting the export process.
- 10 –
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
The import source file consists of three sheets. The contents of each sheet are as follows:
- Sheet1: Exposure menu data
- Sheet2: Study menu data
- Sheet3: Detail study menu data
■Study menu
- Do not rearrange columns or insert new ones for Study menu data (Sheet 2) and Detail study menu
data (Sheet 3).
- For Detail study menu data (Sheet 3), input the study menu data in the same Code continuously. Do
not insert the data with a different Code in the middle.
- Since the study menu data in the import source file is imported without checking the values of each
field, make sure to observe the following rules when editing the fields below:
<Study menu data (Sheet 2)>
Code: Should be unique each other in the study menu data.
<Detail study menu data (Sheet 3)>
Code: Should have been defined in Study menu data (Sheet 2).
ExposureMenuCode: Should have been defined in Exposure menu data (Sheet 1).
ExposeOrder: Should be unique within the same Code in Detail study menu data (Sheet 3).
The initial value should be “0” and the number should be incremented by one.
FramePosition: Should be unique within the same Code in Detail study menu data (Sheet 3).
The initial value should be “0” and the number should be incremented by one.
- 12 –
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
When a code including abnormal data is detected, its name and table/filed are output to the log.
■When the menu database (MenuData.mdb and/or XconParam.mdb) does not exist in the import
destination parameter folder
→ The process aborts and an error dialog box opens.
■When a Code field in the import source Excel file is blank, is entirely filled with spaces, or includes an
unacceptable character
→ The tool does not import the affected row (record) and proceeds to the next one.
■When the record of a duplicate menu code exists in the import destination menu database for Console
→ The following dialog box opens, asking whether you want to overwrite.
■When a menu data in the import source Excel file is blank or abnormal
→ The tool gives an error and aborts the process. The content of the error is output to the log.
■When three or more consecutive rows in the import source Excel file have no menu data
→ The tool terminates the process as it considers that the end of menu data is reached.
- 13 –
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
■When the menu database (MenuData.mdb and/or XconParam.mdb) does not exist in the export
source parameter folder
→ The process aborts and an error dialog box opens.
Period
- 14 –
[ECN Appendix 8] Setting Procedure for Receiving Images from Console Advance V5.0
[ECN Appendix 8] Setting Procedure for Receiving Images from Console Advance
V5.0
1. Outline
This appendix describes setting procedure for receiving images from Console Advance V5.0.
2. Applicable model
FCRView
- CR-VW674 V2.1(B)
Console Advance
- Console Advance V5.0
When connecting to Console Advance V5.0, connecting configuration vary depending on the installation
environment including image input device.
5. Select the [Storage SCP settings] tab and confirm the following settings (no need to change them from
the defaults).
AE Name : "FCR-SCP" (default)
Port number : "5001" (default)
Timeout value : "300" (default)
1/3
[ECN Appendix 8] Setting Procedure for Receiving Images from Console Advance V5.0
6. Click [OK].
→ The system returns to the "Server equipment settings" window.
7. Click [OK].
→ The system returns to the "Setting connection devices" window.
8. Click [Back].
→ The system returns to the "Service Utility top" window.
2/3
[ECN Appendix 8] Setting Procedure for Receiving Images from Console Advance V5.0
2. When receiving images with DICOM Digital X-Ray Image Storage - For Presentation
See "Appendix 11 DICOM DX Storage: Transfer of Processed Images" under Appx IN of
DR-ID300CL Service Manual.
When receiving DR images, configure settings as follows so that images are output with the gradation
process applied.
"2.1.4 Registration of Information on Other Connected Equipment for the DICOM DX Storage
Function"
■Reception Confirmation
1. Start up the FCRView client.
3/3
[ECN Appendix 9] FCRView V2.1(B) Operation of Database Restoration Tool
1. Outline
Since the database restoration tool released in FCRView V2.0(B) as a service part has some changes in
V2.1(B), description on those changes and supplementary on the operation are described in this
appendix.
2. Applicable items
The database restoration tools to be released in FCRView V2.1(B) or later are applicable.
The database restoration tool for V2.0(B) released in 2010-E-0356 does not work in V2.1(B) or
later.
3. Changed contents
Database Restoration Tool is installed together with the application software of V2.1(B) or later.
Some changes in the startup procedure and functions have been made from the one provided for
V2.0(B).
“Before start this software, check that following applications are not running.
- Server Application
- ServiceUtility
- UserUtility
- BackupUtility”
4. Operation
After operating the restoration tool, perform the following:
1/1
[ECN Appendix 10] FCRView V2.1(B) Procedure to set up for Replacing Patient Information at the Time of Receiving DICOM
Images
1. Outline
In FCRView V2.1(B), a function to replace patient information at the time of receiving DICOM images
has been added. This function replaces patient information attached to various types of studies sent
from DICOM connection devices with the patient information in the FCRView, using the patient ID as a
key.
This function enables to display necessary information when confirming images in the Viewer of the
FCRView, even with images from devices which attaches little amount of patient information. This
function also enables to match patient information among devices.
This appendix describes the setup procedure for the replacement of patient information and the
procedure to confirm the settings.
2. Applicable items
Institutes of which devices* are connected with FCRView V2.1(B) by DICOM connection
*On March 2011, the following modalities are applicable:
- CR (Standard / Private)
- DX (for Presentation)
- CT
- MR
- XA
- RF
- US (including Multi-frame)
- SC (including Multi-frame)
1/4
[ECN Appendix 10] FCRView V2.1(B) Procedure to set up for Replacing Patient Information at the Time of Receiving DICOM
Images
2/4
[ECN Appendix 10] FCRView V2.1(B) Procedure to set up for Replacing Patient Information at the Time of Receiving DICOM
Images
(4) In the "DICOM Reception" window, perform the following in accordance with the changed contents.
3/4
[ECN Appendix 10] FCRView V2.1(B) Procedure to set up for Replacing Patient Information at the Time of Receiving DICOM
Images
This item must consist with the AE title setting on the DICOM
connection device (Storage-SCU) side; therefore, be sure to check
the setting of the DICOM connection device in advance.
3 Flag for patient information Switch to enable/disable the setting.
replacement
(6) After configuring settings, click "Add". To revoke the setting and go back to the previous window,
click "Cancel".
(7) To continuously change settings, repeat operations from step (4).
(8) Click "Finish" in the "Patient information notification" window.
4. Confirmation procedure
Send a study to FCRView from the connected device.
Select the sent Patient ID in the "Main" window - "Patient List", and display the sent study in the Viewer.
Confirm that the information displayed in the "Patient List" and the study information displayed in the
Viewer window are the same.
5. Precaution
With this setting, if priority search object for patient information is set to Recept Computer or MWM in
FCRView, patient information in them is used as a priority.
4/4
[ECN Appendix 11] Error Message Table
[ECN Appendix 11] Error Message Table
Event Source Message/Description Message Causes Remedies Remarks
16000 GeneralLiblary.CustomMessag [16000] Failed in deleting template files. *Restart the server computer.
eBox
process.
16002 GeneralLiblary.CustomMessag [16002] Failed in outputting CSV file. *Restart the server computer.
eBox
16004 GeneralLiblary.CustomMessag [16004] The process has been canceled because the
16005 GeneralLiblary.CustomMessag [16005] Failed in outputting the report. *Restart the server computer.
eBox
16006 GeneralLiblary.CustomMessag [16006] Failed in acquiring the measurement data *Restart the server computer.
1/28
[ECN Appendix 11] Error Message Table
16007 GeneralLiblary.CustomMessag [16007] The name of legend has not been entered.
eBox
eBox
16011 GeneralLiblary.CustomMessag [16011] Conditional equation is incorrect. *Enter a conditional equation correctly.
eBox
16012 GeneralLiblary.CustomMessag [16012] Evaluation result of conditional equation is *Modify the conditional equation so that the
eBox not a bool value (true or false). evaluation result becomes a bool value (true or
false).
16013 GeneralLiblary.CustomMessag [16013] Template file does not exist. Report output *Select a correct template file in the system settings
16014 GeneralLiblary.CustomMessag [16014] No base line value has been registered. *Click the Add button and add a baseline value.
eBox
eBox
16016 GeneralLiblary.CustomMessag [16016] DICOM tag information file does not exist. *Carry it out again after a while.
eBox
16017 ImageViewer.ImageView.Shape [16017] Failed in BMD measurement. *Ensure that the image is for BMD measurement.
s.BoneMineralShape
2/28
[ECN Appendix 11] Error Message Table
16018 ImageViewer.ImageView.Shape [16018] The image file required for BMD does not *Restart the server computer.
s.BoneMineralShape exist.
16019 ImageViewer.ImageView.Shape [16019] The image file required for BMD does not *Restart the server computer.
s.BoneMineralShape exit.
16020 ImageViewer.ImageView.Shape [16020] Failed in detecting the aluminum slope. *Confirm that the aluminum slope is being detected
s.BoneMineralShape normally.
16021 ImageViewer.ImageView.Shape [16021] The solid image subtraction point has been *Confirm that the coordinate of solid image
s.BoneMineralShape set outside of the image area. subtraction is in the image frame.
16022 ImageViewer.ImageView.Shape [16022] The measurement result of BMD *Ensure that the measurement operations and
s.BoneMineralShape measurement has exceeded the standard range. images used are in accordance with the guideline.
16023 ImageViewer.ImageView.Shape [16023] The aluminum slope detection ROI has been *Change the aluminum slope detection ROI in such a
s.BoneMineralShape set outside of the image area. way as to surround the aluminum slope.
3/28
[ECN Appendix 11] Error Message Table
16024 ImageViewer.ImageView.Shape [16024] The aluminum slope detection process was *Delete the displayed BMD measurement and newly
16025 ImageViewer.ImageView.Shape [16025] Failed in automatically detecting the *Set the aluminum slope position manually.
16026 ImageViewer.ImageView.Shape [16026] The aluminum slope is not in the aluminum *Confirm that the aluminum slope is in the
16027 ImageViewer.ImageView.Shape [16027] The center line of second metacarpal bone *Confirm that the center line position of the second
s.BoneMineralShape has been set outside of the image area. metacarpal bone has been set properly.
16028 ImageViewer.ImageView.Shape [16028] Failed in detecting the center of the second *Confirm that the center line position of the second
16029 ImageViewer.ImageView.Shape [16029] Measurement object data are not in the *Confirm that the ROI position and size of the
s.BoneMineralShape measurement ROI of the second metacarpal bone. second metacarpal bone measurement has been set
properly.
16030 ImageViewer.ImageView.Shape [16030] Failed in analyzing the measurement ROI of *Expand the measurement ROI size of the second
16031 ImageViewer.ImageView.Shape [16031] Failed in BMD measurement. *Ensure that the measurement operations and
16032 ImageViewer.ImageView.Shape [16032] The center line of the second metacarpal *Confirm that the center line of the second
s.BoneMineralShape bone is too short. Measurement processing was metacarpal bone has been set properly.
canceled.
16033 ImageViewer.ImageView.Shape [16033] The vertical length of measurement ROI of *Expand the ROI size of the second metacarpal bone
4/28
[ECN Appendix 11] Error Message Table
16034 ImageViewer.ImageView.Shape [16034] The horizontal length of measurement ROI *Expand the ROI size of the second metacarpal bone
16035 ImageViewer.ImageView.Shape [16035] Failed in generating the image for *Restart the server computer.
s.BoneMineralShape measurement.
16036 ImageViewer.ImageView.Shape [16036] BMD measurement cannot be performed on *Ensure that the images are in accordance with the
16037 ImageViewer.ImageView.Shape [16037] Failed in saving files. *Confirm that the destination file is not locked.
s.BoneMineralShape
16038 ImageViewer.ImageView.Shape [16038] The aluminum slope is not in the aluminum *Confirm that the aluminum slope is in the
16039 ImageViewer.ImageView.Shape [16039] The solid image subtraction point has been *Confirm that the coordinate of solid image
s.BoneMineralShape set outside of the image area. subtraction is in the image frame.
16040 ImageViewer.ImageView.Shape [16040] Failed in detecting the horizontal line of solid *Adjust the position of the solid image subtraction
s.BoneMineralShape image subtraction. point so that the exposure subject does not overlap
subtraction point.
16041 ImageViewer.ImageView.Shape [16041] Failed in detecting the vertical line of solid *Adjust the position of the solid image subtraction
s.BoneMineralShape image subtraction. point so that the exposure subject does not overlap
subtraction point.
5/28
[ECN Appendix 11] Error Message Table
16042 ImageViewer.ImageView.Shape [16042] Note that you were charged for using BMD *You will be charged at the time BMD measurement
16044 GeneralLiblary.CustomMessag [16044] This operation can not be canceled. Do you want to continue processing?
eBox
16045 GeneralLiblary.CustomMessag [16045] Template file does not exist. Printing *Open the setup change window of simplified print
17000 GeneralLiblary.CustomMessag [17000] Failed in initializing process of work list. *Restart the server computer.
eBox
17006 GeneralLiblary.CustomMessag [17006] Failed in updating the work list. *Restart the server computer.
eBox
17009 GeneralLiblary.CustomMessag [17009] Failed in starting the study.No authority to *No patient registration authority.
17011 GeneralLiblary.CustomMessag [17011] Failed in starting the study.Exposure cannot *Exposure screen is starting up.
eBox be started.
17013 GeneralLiblary.CustomMessag [17013] Failed in starting the study.The exposure of *The displayed work list may not be the latest.
eBox the specified study has finished already. Refresh the work list.
6/28
[ECN Appendix 11] Error Message Table
17015 GeneralLiblary.CustomMessag [17015] Study canceled in the past will be started. *Ordering device and study information of the
synchronized.
study.
17017 GeneralLiblary.CustomMessag [17017] Failed in starting the study. Failed in *Restart the server computer.
17019 GeneralLiblary.CustomMessag [17019] Failed in starting the study. Study or *Unregistered study/exposure menu codes may be
recurs.
17024 GeneralLiblary.CustomMessag [17024] Failed in updating the work list. *Restart the server computer.
eBox
17028 GeneralLiblary.CustomMessag [17028] Failed in updating the work list. *Restart the server computer.
eBox
7/28
[ECN Appendix 11] Error Message Table
17033 GeneralLiblary.CustomMessag [17033] Failed in initializing process of work list. *Restart the server computer.
eBox
17036 GeneralLiblary.CustomMessag [17036] Failed in starting the study. Patient ID has *Ensure that the correct Patient ID of the
eBox not been specified. corresponding study has been specified from the
ordering device.
recurs.
17038 GeneralLiblary.CustomMessag [17038] Patient ID entered includes characters *Following characters strings are not allowed to use
sure?
17039 GeneralLiblary.CustomMessag [17039] Banned characters are used for patient *Following characters are not allowed to use for
8/28
[ECN Appendix 11] Error Message Table
sure?
17040 GeneralLiblary.CustomMessag [17040] Banned characters are used for Kanji patient *Following characters are not allowed to use for
Single-byte characters
sure?
17041 GeneralLiblary.CustomMessag [17041] The Species contains illegal character. *Following characters are not allowed to use for
eBox species.
sure?
17042 GeneralLiblary.CustomMessag [17042] The Breed contains illegal character. *Following characters are not allowed to use for
sure?
17043 GeneralLiblary.CustomMessag [17043] Banned characters are used for Kanji *Following characters are not allowed to use for
9/28
[ECN Appendix 11] Error Message Table
Single-byte characters
17044 GeneralLiblary.CustomMessag [17044] Banned characters are used for responsible *Following characters are not allowed to use for
sure?
17045 GeneralLiblary.CustomMessag [17045] Responsible organization contains invalid *Following characters are not allowed to use for
sure?
17046 GeneralLiblary.CustomMessag [17046] Failed in MPPS linkage. The setting *Study condition was not notified. The study
eBox necessary for communication could not be acquired. condition must be changed on other device.
10/28
[ECN Appendix 11] Error Message Table
restarted.
17048 GeneralLiblary.CustomMessag [17048] Failed in MPPS linkage. The necessary *Study condition was not notified. The study
eBox information could not be acquired from the database. condition must be changed on other device.
17050 GeneralLiblary.CustomMessag [17050] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. The study
restarted.
17052 GeneralLiblary.CustomMessag [17052] Failed in MPPS linkage. No response was *Study condition may not be notified. The study
eBox replied from other device. condition must be changed on other device.
restarted.
11/28
[ECN Appendix 11] Error Message Table
17054 GeneralLiblary.CustomMessag [17054] Failed in MPPS linkage. Unexpected *Study condition may not be notified. The study
restarted.
17056 GeneralLiblary.CustomMessag [17056] Failed in updating the study database in *Study condition was notified, but it must be
17058 GeneralLiblary.CustomMessag [17058] Failed in MPPS linkage. *Study condition may not be notified. The study
restarted.
17060 GeneralLiblary.CustomMessag [17060] Patient ID entered includes characters *Following characters are not allowed to use for
12/28
[ECN Appendix 11] Error Message Table
17061 GeneralLiblary.CustomMessag [17061] Banned characters are used for Kanji patient *Following characters strings are not allowed to use
17062 GeneralLiblary.CustomMessag [17062] Banned characters are used for patient *Following characters are not allowed to use for
or more
17063 GeneralLiblary.CustomMessag [17063] Banned characters are used for RP ID. *Following characters are not allowed to use for RP
17064 GeneralLiblary.CustomMessag [17064] Banned characters are used for AE title. *Following characters are not allowed to use for AE
\ Space
17065 GeneralLiblary.CustomMessag [17065] Failed in updating the work list. *Restart the server computer.
eBox
17102 GeneralLiblary.CustomMessag [17102] Number of search results is higher than the *Desired order may not be displayed on the list.
13/28
[ECN Appendix 11] Error Message Table
17104 GeneralLiblary.CustomMessag [17104] Display only the orders which have been *Desired order may not be displayed on the list.
worklists are being received. Change search conditions and retry searching.
17106 GeneralLiblary.CustomMessag [17106] Failed in receiving the work list. The setting *Setting may not be completed.
17108 GeneralLiblary.CustomMessag [17108] Failed in receiving the work list. The *Restart the server computer.
17110 GeneralLiblary.CustomMessag [17110] Failed in receiving the work list. Failed in *After restarting the other device, ensure the device
17112 GeneralLiblary.CustomMessag [17112] Failed in receiving the work list. *After restarting the other device, ensure the device
14/28
[ECN Appendix 11] Error Message Table
17114 GeneralLiblary.CustomMessag [17114] Failed in receiving the work list. Unexpected *After restarting the other device, ensure the device
17116 GeneralLiblary.CustomMessag [17116] Failed in receiving the work list. *After restarting the other device, ensure the device
17118 PatientPortal [17118] Failed in reading received data in the work *After restarting the other device, ensure the device
15/28
[ECN Appendix 11] Error Message Table
exceeded.
17124 StudyList [17124] Failed in MPPS linkage in some studies. *Study condition was not notified. The study
restarted.
20553 FFCustomMsgBox [20553] Failed in MPPS linkage. *Study condition may not be notified. Confirm the
restarted.
20554 FFCustomMsgBox [20554] Failed in MPPS linkage. The setting *Study condition was not notified. Change the
necessary for communication could not be acquired. condition on the other device side.
16/28
[ECN Appendix 11] Error Message Table
restarted.
20555 FFCustomMsgBox [20555] Failed in MPPS linkage. The necessary *Study condition was not notified. Change the study
information could not be acquired from the database. condition of the study on the other device side.
20556 FFCustomMsgBox [20556] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. Change the study
restarted.
20557 FFCustomMsgBox [20557] Failed in MPPS linkage. No response was *Study condition may not be notified. Confirm the
restarted.
17/28
[ECN Appendix 11] Error Message Table
20558 FFCustomMsgBox [20558] Failed in MPPS linkage. Unexpected *Study condition may not be notified. Confirm the
restarted.
20559 FFCustomMsgBox [20559] Failed in updating the study database in *Notification of study condition was succeeded.
MPPS linkage.
20569 FFCustomMsgBox [20569] Failed in MPPS linkage. The setting *Study condition was not notified. Change the
necessary for communication could not be acquired. condition on the other device side.
restarted.
20570 FFCustomMsgBox [20570] Failed in MPPS linkage. The necessary *Study condition was not notified. Change the study
information could not be acquired from the database. condition of the study on the other device side.
18/28
[ECN Appendix 11] Error Message Table
20571 FFCustomMsgBox [20571] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. Change the study
the connection. condition of the study on other device.
restarted.
20572 FFCustomMsgBox [20572] Failed in MPPS linkage. No response was *Study condition may not be notified. Confirm the
restarted.
20573 FFCustomMsgBox [20573] Failed in MPPS linkage. Unexpected *Study condition may not be notified. Confirm the
19/28
[ECN Appendix 11] Error Message Table
restarted.
20574 FFCustomMsgBox [20574] Failed in MPPS linkage. *Study condition may not be notified. Confirm the
study condition on other device.
restarted.
20575 FFCustomMsgBox [20575] Failed in updating the study database in *Notification of study condition was succeeded.
MPPS linkage.
20626 GeneralLiblary.CustomMessag [20626] Failed in launching the ASSISTA Portal Site. *Setting information was not found or was in error.
eBox
20742 FCRWeb [20742] No NAS space necessary for outputting *Add a NAS and retry.
data.
20/28
[ECN Appendix 11] Error Message Table
20743 FCRWeb [20743] No NAS for outputting data was found. *Please check NAS power and network connections
20744 FCRWeb [20744] Failed in acquiring a setting information. >Contact your service engineer.
20745 FCRWeb [20745] Failed in registration. *Request on the server computer failed. Check the
startup status of the server computer. If it fails many
20767 FFCustomMsgBox [20767] The study information was not registered *After restarting the server computer, delete the
30807 FCRWeb [30807] Failed in MPPS linkage. The setting *Study condition was not notified. Change the
necessary for communication could not be acquired. condition on the other device side.
21/28
[ECN Appendix 11] Error Message Table
restarted.
30808 FCRWeb [30808] Failed in MPPS linkage. The necessary *Study condition was not notified. Change the study
information could not be acquired from the database. condition of the study on the other device side.
30809 FCRWeb [30809] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. Change the study
restarted.
30810 FCRWeb [30810] Failed in MPPS linkage. No response was *Study condition may not be notified. Confirm the
restarted.
22/28
[ECN Appendix 11] Error Message Table
30811 FCRWeb [30811] Failed in MPPS linkage. Unexpected *Study condition may not be notified. Confirm the
restarted.
30812 FCRWeb [30812] Failed in MPPS linkage. *Study condition may not be notified. Confirm the
restarted.
30813 FCRWeb [30813] Failed in MPPS linkage. The setting *Study condition was not notified. Change the
necessary for communication could not be acquired. condition on the other device side.
restarted.
23/28
[ECN Appendix 11] Error Message Table
30814 FCRWeb [30814] Failed in MPPS linkage. The necessary *Study condition was not notified. Change the study
information could not be acquired from the database. condition of the study on the other device side.
30815 FCRWeb [30815] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. Change the study
restarted.
30816 FCRWeb [30816] Failed in MPPS linkage. No response was *Study condition may not be notified. Confirm the
restarted.
30817 FCRWeb [30817] Failed in MPPS linkage. Unexpected *Study condition may not be notified. Confirm the
24/28
[ECN Appendix 11] Error Message Table
30818 FCRWeb [30818] Failed in updating the study database in *Notification of study condition was succeeded.
MPPS linkage.
30819 FCRWeb [30819] Failed in MPPS linkage. *Study condition may not be notified. Confirm the
restarted.
30820 FCRWeb [30820] Failed in updating the study database in *Notification of study condition was succeeded.
MPPS linkage.
30917 ImportImage [30917] Failed in deleting files. *The original file being imported cannot be deleted.
25/28
[ECN Appendix 11] Error Message Table
30918 ImportImage [30918] A failure occurred in operation of Complete *Perform the operation again.
button.
34944 ImportImage [30944] Already being imported. *Operation cannot be performed because the target
34945 ImportImage [30945] Complete the study in a receiving status. +Are you sure?
will be cleared.
34947 ImportImage [30947] The original image being imported will be +Are you sure?
deleted.
34952 StudyTableRegist [34952] An error occurred in auto importing process *An error occurred in receiving the study in the
image.
34953 ImageRetrieveRequest [34953] An error occurred in auto importing process *An error occurred in importing the image in the
image.
40500 (イベントログ出力無し) [40500] A fatal error occurred. > Press the button below to close the dialog, and
40501 (イベントログ出力無し) [40501] Server computer cannot be accessed. *Check the connection of network cable and hub.
26/28
[ECN Appendix 11] Error Message Table
40502 (イベントログ出力無し) [40502] Failed in gaining access to the server *Check the connection of network cable and hub.
40503 (イベントログ出力無し) [40503] Failed in gaining access to the server *Check the connection of network cable and hub.
40504 (イベントログ出力無し) [40504] Failed in gaining access to the server *Check the connection of network cable and hub.
40505 (イベントログ出力無し) [40505] Failed in gaining access to the server *Check the connection of network cable and hub.
40506 (イベントログ出力無し) [40506] Failed in gaining access to the server *Check the connection of network cable and hub.
69601 FFServiceUtility [69601] Connected device name is duplicating. *Return to the setup window and re-input correctly.
69602 FFServiceUtility [69602] Delete the setting of selected connected +Are you sure?
device.
69603 FFServiceUtility [69603] The same receive folder path already exists. *Return to the setup window and re-input correctly.
27/28
[ECN Appendix 11] Error Message Table
69604 FFServiceUtility [69604] Specified receive folder path does not exist. *Check that the receive folder of connected device
exists.
69605 FFServiceUtility [69605] The same path for the receive folder and *Return to the setup window and re-input correctly.
69606 FFServiceUtility [69606] Set the receive folder path within 80 *Return to the setup window and re-input correctly.
characters.
Period
28/28