2011 E 0141 Upgrading Software For V2.1 B

This engineering change notice announces the upgrading of software to version 2.1(B) for several medical imaging systems, including the CR-VW 674, CR-IR 355V CL, CR-IR 391CL, and CR-IR 391V CL. The update includes new functions, error corrections, and installation procedures. Details are provided on the specific software packages released for each system, functional enhancements, added optional functions, corrected errors, and step-by-step installation instructions.

Uploaded by

Oscar Pelcastre
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
43 views

2011 E 0141 Upgrading Software For V2.1 B

This engineering change notice announces the upgrading of software to version 2.1(B) for several medical imaging systems, including the CR-VW 674, CR-IR 355V CL, CR-IR 391CL, and CR-IR 391V CL. The update includes new functions, error corrections, and installation procedures. Details are provided on the specific software packages released for each system, functional enhancements, added optional functions, corrected errors, and step-by-step installation instructions.

Uploaded by

Oscar Pelcastre
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 195

ECN No.

2011-E-0141

ENGINEERING CHANGE NOTICE (FIELD CHANGE NOTICE)

SUBJECT: Upgrading Software to V2.1(B) for CR-VW674

Design Change Reason Action Time frame


Specification/Function Change Mandatory Immediately
Pro-active Improvement Recommended (by )
Serviceability Not required (Information only) Next maintenance
Part Replacement Other ( ) Upon failure
Corrective Action At your convenience
Preventive Action Not Applicable
Regulatory Compliance Other( )
Other ( )

DATE: Apr. 28, 2011 Dept.: Medical Systems Business Div.


ORIGINATOR: H. Ichikawa EDITED BY: A. Takahashi APPROVED BY: Y. Koyanagi

AFFECTED PRODUCTS: CR-VW674


TOTAL PAGE (S): 195 ATTACHMENT: YES NO
RELATED INFORMATION: None
***********************************************************************************************************************************

DESCRIPTION OF CHANGE:
CL Application V2.1(B) for CR-VW 674, including new functions and error corrections, is ready to be released.
New Functions/Enhancement
Automatic import of endoscopic studies
Linkage of study information with DICOM MWM/PPS gateway
NAS data storage converted from DVD
Receiving Images from Console Advance V5.0, etc

- APPLICABLE TO (Serial Nos. etc.) : CR-VW 674 Software V5.0 or later


- ESTIMATED WORK TIME: 1 HOUR
- SPECIAL TOOLS / INSTRUMENTS: YES NO

- SUPPLY OF THE ARTICLES: YES NO


IF YES, WHEN: with this ECN
NAME / CODE: CR-VW 674 Application Software-B V2.1 / 114Y5097001B02 (DVD-R)
CR-VW 674Client Application Software-B V2.1 / 114Y5097006B02 (CD-R)
QUANTITY: 2disks / distributor

- RETURN OF THE REPLACED PARTS TO FUJI: YES NO

DOCUMENTS AFFECTED: YES NO


AFFECTED CURRENT NEW
DOCUMENTS REV ISSUED DATE REV ISSUED DATE

DETAILS:

See the attachment for details.

- NOTE
The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.

PAGE 1 of 2 K12N0028-6-J
ECN No. 2011-E-0141
1) Important information about SUPPLY OF THE ARTICLES:
Charge:
FREE OF CHARGE UNTIL (mmddyyyy):
WITH YOUR CHARGE

The contents described or shown in our ECN documents and/or any information provided from FUJIFILM in verbal or written communication is intended for the
addressee and/or its organization. Transfer and/or copy distribution of the information without written consent of FUJIFILM is not permitted.

PAGE 2 of 2 K12N0028-6-J
ECN 2011-E-0141

Upgrading Software to V2.1(B) for CR-VW 674, CR-IR 355V CL and

CR-IR391CL/391V CL

1. Overview 2

2. Released Items 3
2.1. Released items for CR-VW 674 3
2.2. Released items for CR-IR 355V CL 3
2.3. Released items for CR-IR 391 CL 3
2.4. Released items for CR-IR 391V CL 4

3. Functional Enhancement 5
3.1. Functional Enhancement 5
3.2. Added optional functions 6
3.3. Error corrections 6
3.3.1. Error Correction (failures found at user’s site) 7
3.3.2. Error corrections (Failures found by design and QA groups) 11

4. Installation Procedures 13
4.1. Procedure for upgrading software to V2.1(B) 13
4.1.1. Software Versions 13
4.1.2. Upgrading Procedure 14
4.1.3. Procedure for reinstallation from OS 15
4.1.4. Other extended functions 15

Attachments
Appendix 1: Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)
Appendix2: FCRView V2.1(B) Connection Setup for Windows 7 Client PC
Appendix 3: FCRView V2.1(B) Endoscopic Device Setting Procedure
Appendix 4: FCRView V2.1(B) OS Reinstallation Procedure for Windows 7
Appendix 5-1: FCRView V2.1(B) NAS Quick guide
Appendix 5-2: FCRView V2.1(B) Restoration Procedure for the Case Images Cannot be Stored into NAS
Appendix 5-3: Procedure of Moving Study Output from Storage Disk to NAS
Appendix 6: FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
Appendix7: Installation Support Software Exposure Menu Import/Export Tool for Hobbit Operation
Manual
Appendix 8: Setting Procedure for Receiving Images from Console Advance V5.0
Appendix 9: FCRView V2.1(B) Operation of Database Restoration Tool
Appendix 10: FCRView V2.1(B) Setup Procedure for Replacing Patient Information at the Time of
Receiving DICOM Images
Appendix 11: Error Code Table

1/16
ECN 2011-E-0141
1. Overview

This ECN provides the announcement about upgrading software to V2.1(B) and related information.

- The application version is changed to V2.1(B). On the CD label, it is described as Application


software-B.
- The product names for each destination are shown in the following table.

Model Type Destination V5.1 V1.0(B) / V1.1(B) and V2.0(B)


CR-IR 355CL Human Japan FCR CAPSULA VIEW C@RNACORE Station
Vet. Japan FCR CAPSULA V VIEW V Station
CR-IR 355V CL Vet. USA FCRView FCRView VETERINARY
Vet. Worldwide FCR CAPSULA V VIEW FCR CAPSULA V VIEW

Model Type Destination V2.0(B) V5.0 / V1.1(B) and V2.0(B)


CC-WS 674 Human Japan C@RNACORE
CR-VW 674 Human Worldwide FCRView

Model Type Destination V1.0 / V1.1.(B) and V2.0(B)


CR-IR 391CL Human Worldwide FCR PRIMA Console
CR-IR 391V CL Vet. Worldwide FCR PRIMA V Console

◆Caution◆
Unless specified, the following 4 models are described as FCRView to represent in this document.
CR-VW 674
CR-IR 355V CL
CR-IR 391CL
CR-IR 391V CL

2/16
ECN 2011-E-0141
2. Released Items

2.1. Released items for CR-VW 674

Distrib
No. Product name Description Part No. Media
ute
CR-VW 674 Installer of Application DVD-
1 114Y5097001B02 ✓
Application Software–B V2.1 Software R
CR-VW 674 Installer of Client
2 114Y5097006B02 CD-R ✓
Client Application Software–B V2.1 Application Software
CR-VW 674 Reference guide for
3 114Y5097051A03 CD-R ✓
Reference Guidance–B V4.0 human
CR-VW 674 Key software for
5 DICOM ORDER MWM/MPPS Key activating DICOM 114Y5097127A00 CD-R ✓
Software V1.0 MWM/MPPS

2.2. Released items for CR-IR 355V CL

Distrib
No. Product name Description Part No. Media
ute
CR-IR 355CL/355V CL Installer of Application
1 114Y5154001B02 DVD-R ✓
Application Software–B V2.1 Software
CR-IR 355CL/355V CL Installer of Client
2 114Y5154006B02 CD-R ✓
Client Application Software–B V2.1 Application Software
CR-IR 355V CL Reference guide for
3 114Y5154063A03 CD-R ✓
Reference Guidance–B V4.0 animals(WW)
CR-IR 355V CL Reference guide for
4 114Y5154064A03 CD-R ✓
Reference Guidance–B V4.0 animals(US)
CC-WS674
Key software for
CR-IR 355CL/355V CL
5 activating DICOM 114Y5154127A00 CD-R ✓
DICOM ORDER MWM/MPPS Key
MWM/MPPS
Software V1.0

2.3. Released items for CR-IR 391 CL

Distrib
No. Product name Description Part No. Media
ute
CR-IR 391CL Installer of Application DVD-
1 114Y2129001B02 ✓
Application Software–B V2.1 Software R
CR-IR 391CL Reference guide for
2 114Y2129052A03 CD-R ✓
Reference Guidance–B V4.0 human
CR-IR 391CL Reference guide for
3 114Y2129053A03 CD-R ✓
Reference Guidance–B V4.0 human in Chinese
CR-IR 391CL Key software for
4 DICOM ORDER MWM/MPPS Key activating DICOM 114Y2129127A00 CD-R ✓
Software V1.0 MWM/MPPS

3/16
ECN 2011-E-0141

2.4. Released items for CR-IR 391V CL

Distrib
No. Product name Description Part No. Media
ute
CR-IR 391V CL Installer of Application DVD-
1 114Y2129201B02 ✓
Application Software–B V2.1 Software R
CR-IR 391V CL Reference guide for
2 114Y2129061A03 CD-R ✓
Reference Guidance–B V4.0 animal
CR-IR 391CL Key software for
3 DICOM ORDER MWM/MPPS Key activating DICOM 114Y2129325A00 CD-R ✓
Software V1.0 MWM/MPPS

2.5. Service Tool

Distrib
No. Product name Description Part No. Media
ute
Menu Import Export Tool For Hobbit
1 Menu Import Export Tool 114Y5154709A00 CD-R ✓
V1.0

4/16
ECN 2011-E-0141
3. Functional Enhancement

3.1. Functional Enhancement


VW674 355V 391CL 391V
No. Item Description
CL CL

By using patient ID as a key, patient


information that is stored in FCRView is
Patient
transferred to the medical inspection
1 information ✓ ✓
equipments connected with FCRView. The
transfer
transferred patient information is used to start
a study in the destination.
The following NAS functions have been
added.
・Data writing
・Expanded automatic backup
2 NAS data storage ✓ ✓ ✓ ✓
・Data reading
・Acquisition / notification of the status
・Registration / Utility
・Convert from DVD storage
DICOM CR Applicable to the DICOM Network Service
Storage For below:
3 ✓ ✓
(Standard) SCP ・Computed Radiography Image Storage
(SCP)
DICOM DX Applicable to the DICOM Network Service
Storage below:
4 For Presentation ・ Digital X-Ray Image Storage - For ✓
SCP Presentation (SCP)

Linkage of study Applicable to DICOM MWM/MPPS Service to


information with execute the linkage of study information.
5 ✓ ✓ ✓ ✓
DICOM MWM/
MPPS gateway
Software for For demonstration use, installation and
6 demonstration activation of the application become possible ✓ ✓ ✓ ✓
use without LAN connection.
Improved The following functions have been added to
operational the viewer.
performance of ・ Mode change with right-click while
the viewer left-clicking
Gradation→Magnification/reduction→Panni
ng→Select→Gradation
7 ✓ ✓ ✓ ✓
・Perfect circle annotation
・ Windows Level manipulation by using
numeric keypad on non-FCR images
・Simplified print
・Rotation of the processed CR/DX images
・Displaying DICOM tag of the images (not

5/16
ECN 2011-E-0141
editable)
Supported Supported languages by Windows 7 are
languages by added.
8 Windows 7 ・Simplified Chinese ✓ ✓ ✓
・Traditional Chinese
・Korean
DB restoration tool The DB restoration tool, which was released in
9 a CD, has been mounted in the system folder ✓ ✓ ✓ ✓
of FCRView.

3.2. Added optional functions and integrated functions


VW674 355V 391CL 391V
No. Item Description
CL CL

The endoscopic images imported from


Automatic
endoscopic equipment are registered
1 endoscopic ✓ ✓
automatically without opening the Examination
image import
Data Importing window.
Multiple
Image import from multiple endoscope
2 endoscope ✓ ✓
equipments becomes possible.
linkage
When DICOM CT/MR and ultrasound images
DICOM patient
are imported, the study information is replaced
3 information ✓ ✓
in the corresponding patient information in
replacement
FCRView on the basis of patient ID.
The following functions and GUI are added to
the importing general-purpose image window.
・ Delete/not delete the image after the
Specification
import has completed.(display/not display
changes in the
the message when deleting) → select the
4 importing ✓ ✓
checkbox.
general-purpose
・ Thumbnails of the images are displayed in
image window
chronological / reverse chronological
order.
→select the radio button

3.3. Error corrections

3.3.1. Error correction (HotFix)

The error corrections which have been made by the following HotFix are included in V2.1(B).
V2.0(B) HF0001
V2.0(B) HF0002
V2.0(B) HF0004
V2.0(B) HF0010

6/16
ECN 2011-E-0141

3.3.2. Error Correction (failures found at user’s site)

1) DICOM data created by CT equipment cannot be imported


Phenomenon [30308: Image acquisition failed] error occurs when the DICOM data crated by CT equipment
becomes unrecognized, which disables the data import.
Causes Without the information of 0002 group, it does not satisfy the DICOM standard.

Measures Enable FCRView to import Type2 tag information necessary for DB registration. Leaving
alert logs alone, no exception error occurs if tag information does not exist.
Number QCN-002146/FMS2-10-073

2) The data stored in NAS cannot be PDI output


Phenomenon PDI cannot be created when the study data transferred to NAS is output to PDI.
Causes Unnecessary letters are attached to the correct path of the PDI output tag (7FE0,0010) in the
study data file (DDO file) in NAS.
Measures The data setting of the (7FE0,0010) tag and the function used for reading the data have been
changed to read out the correct path even when unnecessary data is attached.
Number QCN-005245/FMS2-11-213

3) Output status stays in [Undelivered]


Phenomenon After the installation of V2.0(B), output status stays in [Undelivered] although the study data
is output to the storage disk after it is verified then FCR image processing is done and then it
is verified again.
Causes During the process to update the output status, the values of the image version of DDO file
(the value that (0009,1008) and (0009,1009) are linked with a period in the DDO file) and
DB(T_TIMAGE:IMAGE_VERSION_NO) are compared.
Because two values are inconsistent when the image processing is performed, the output
status stays in [Undelivered] instead of [Delivered] even after the output is completed.
Measures The image version of the image data does not contain the value of the image version number
after the FCR image processing.
ODD file is read to update the image version number of the DB.
Number QCN-002056/FMS-10-A64
QCN-002086/FMS-10-A97
QCN-005246/FMS2-11-215

4)Error #20188 is displayed during startup.


Phenomenon Error #20188 (Download of files has failed) is displayed during startup, making the server
application fail in activation.
Causes The version check done by a Request of the process in the server PC fails, causing a
version inconsistency. The version check is unnecessary unless the client PC does it.
Measures The program has been revised to holt the version check when a Request made by the
process in the server PC

7/16
ECN 2011-E-0141
Number QCN-005083/FMS2-10-223
QCN-001762/FMS-10-543
QCN-005013/FMS2-10-121
QCN-005152/FMS2-11-101
SCN-000098/FMS-09-662

5)Even after the filmless function is set up in the user utility, film is output after the study is verified again
Phenomenon Film is output when the study is verified after the modification regardless of the filmless
function is currently set when the patient information, which has the film output setup history,
has been changed.
Causes Film is output when a study is verified again with the filmless function set.
When the [Automatic image output] of the [Printer] in the user utility is not checked, the value
of [Upon verification of a study] of the [Timing of output] is on.
Measures The program has been revised to enable the setup of [Upon verification of a study] of the
[Timing of output] in the [Printer] separately regardless of the settings of [Automatic image
output].
According to this change, the indent of the checkbox in the [Upon verification of a study] is
modified along with the [Automatic image output].
Number QCN-001866/FMS-10-678
QCN-005077/FMS2-10-198
QCN-005179/FMS2-11-131

6) 30900 error occurs when endoscopic images are imported.


Phenomenon [Start import] button is clicked when multiple numbers of studies are listed on the
“Examination data importing” window. 30900 error occurs when the import button is clicked
again after a while.
Causes When there is no other “not [Complete]” studies left while the study data is being read, [Start
Import] /[Image Viewer] /[Delete Study] buttons are activated. 30900 error occurs when
clicking [start importing] button in this status, the system starts retrieving the same study
again but there is no file to retrieve.
Measures Setup the selection and each command button on the study list as follows:
・when the “Import Status” is not [Complete]…[Delete Study], [Start Import] and [Image
Viewer] are activated.
・when “Import Status” is [Complete] …[Delete Study], [Start Import] and [Image Viewer] are
not activated.
・when no list, no selection column …[Delete Study], [Start Import] and [Image Viewer] are
not activated.

Number QCN-005049/FMS2-10-174

7) Electrocardiogram and another study cannot be displayed simultaneously


Phenomenon When an electrocardiogram study and another study are displayed simultaneously on the
viewer, “A study is being performed” error occurs on one of them then the study which has
the error cannot be opened (exclusively in Japan)
Causes

Measures The program has been revised.

8/16
ECN 2011-E-0141
Number QCN-002069/FMS-10-A86
QCN-005168/FMS2-11-107
QCN-005344/FMS2-11-326

8) Particular studies fail in image display linkage with electric chart


Phenomenon The patient information is displayed but no image on the viewer when the image display
linkage with the electric chart is performed on particular studies. The image display linkage
fails after that (exclusively in Japan)
Causes

Measures The program has been revised.


Number QCN-002145/FMS2-10-077
QCN-005042/FMS2-10-169
QCN-005061/FMS2-10-190
QCN-005187/FMS2-11-127

9) The status stays [Delivering] when the queue is initialized when outputting the data to the work disk
Phenomenon The status stays [Delivering] when the queue is initialized while the data is being output to
the work disk.
Causes Clear Output Queue process in the Service Utilities does not contain the deleting the output
queue to the work disk.
Measures The Clear Output Queue process has been revised to delete the output queue to the work
disk.
Number TIMS9586 /QCN-001893

10) The message box hides behind the “Read Image” window when selecting
Phenomenon Because the exposure menu button is clicked when the “Read Image” window is opened, the
warning dialog box hides behind it, which disables to click the [OK] button as if it freezes up.
Causes The message dialog, which does not contain particular settings to be displayed as a general
dialog, hides behind the “Read Image” window, which has been setup as the foremost
window.
Measures The display setting of the “Read Image” window has been changed to display the message
dialog on the top of all the other windows.
Number QCN-005048/FMS2-10-168

11) FCRView cannot output PDI to particular equipment


Phenomenon Lowercase letters are included in "DCMFile" folder in PDI created inWindows7 、
WindwsVista, causing DICO MDIR unreadable. The equipment below cannot read the data:
● I-PACS(Konica Minolta)
Causes "DCMFILE" is mistakenly typed as “DCMFile" manually.

Measures "DCMFile" is changed in "DCMFILE" in the folder in


"C:¥ ProgramData¥FujiFilm¥Hobbit¥Data¥PDITemplate".
Number QCN-005322/FMS2-11-296

12) Color of ultrasound images stored in storage DVD is unusually changed in display.
Phenomenon The moving images turn bluish when the data, which has been transferred from FAZONE to

9/16
ECN 2011-E-0141
the server, is imported from DVD.
Causes 1 study sometimes consists of multiple images. Usually, Raw (RGB) data is written in the
storage DVD in JPEG Lossless (color space RGB). In case, however, JpegLossy(color
space YCbCr) images are included in the same study, they are stored in JPEG Lossy.
Because the Photometric Interpretation of DICOM header stays in RGB, the images in
YcbCr are recognized as RGB, changing their tone of color.
Measures In case of the images whose the compression format has been changed (e.g. JPEG
Lossless→JPEG Lossy), adjust the Photometric Interpretation of DICOM header to the
color space of the compression format .
It is revised that Lossless→Lossy and Lossy→Lossless compression format changes have
no problem.
Number QCN-005278/FMS2-11-269
QCN-005290/FMS2-11-238

13) Patient list is not updated automatically


Phenomenon Automatic update of the patient list is not performed although it has been setup.
Causes A deadlock error occurs while the automatic update is being executed, causing the process
aborted.
Measures The program has been revised to enable the automatic update.
Number QCN-005359/FMS2-11-314
KCN-005017/FMS2-11-317

14) PDI created in other facilities cannot be read


Phenomenon A reading error occurs when the data from a particular PDI is imported. The unreadable PDIs
are created with the equipments of the following companies:
GE
Hitachi Medical
Causes Under the condition that the codes that are the same as Fuji uses for the private tags are
used for creating PDIs with the third–party equipments, VR information is mistakenly
translated when the reading is proceeding, causing an error. The conditions are as follows:
(0027,1060) is included
VR=PN tag is attached
Measures In case of the Private information of the third-party equipments, translate VR as [UN].
In case of VR=PN, the processing is skipped to prevent an error
Number QCN-001872/FMS-10-714
QCN-001926/FMS-10-769
QCN-005012/FMS2-10-122
QCN-005230/FMS2-11-197
QCN-001861/FMS-10-694
QCN-005016/FMS2-10-117
QCN-005108/FMS2-11-025
QCN-005186/FMS2-11-116

15) When activated from the main window, white-tinged images are displayed on the viewer
Phenomenon The images turn white-tinged when the viewer is activated from the main window to display
them. The images are whiter than the ones displayed on the viewer activated from the study
utility.

10/16
ECN 2011-E-0141
Causes LUT is obtained each time when displaying images on the viewer. LUT is not obtained when
the network condition is unstable, making the images turn white-tinged.
Measures The program has been revised to obtain LUT only for the first time to display the image.
Internally cached LUT is used later. In case LUT is not obtained for the first activation, the
default LUT is used instead。
Number QCN-005219/FMS2-11-169

16) Not to output Ultrasound images to the storage disk


Details of the The function, which does not automatically output the studies of specified modalities to the
change storage disk at the time of verification, is implemented.
Reason for the Large numbers of the ultrasound studies (as large as approximately 3000 images) are
change frequently performed in a facility. Automatic image output exhausts the storage disk with this
much large size of data within three days, causing a hindrance with slow processing speed
of the PC during operation.
Measures
Number QCN-002165/FMS2-10-083

17) Error occurs when connected to Hitachi ultrasound equipment


Phenomenon Receiving error occurs when the following Hitachi ultrasound equipments are connected
during the validation process.
HIVISION Preirus / Avius
HIVISION900/HIVISION7500(EUB7500)
Causes In case SOP Instance UID of transferred DICOM data uses whole 64 bit, extra one-bit data is
attached when the data is received, making processing error.
Measures The program has been revised to prevent to add one bit data.
Number

18) Study data import error occurs immediately after the activation
Phenomenon After the study data is imported from the Ultrasound equipment of Toshiba, the error
message (#49572) describing the failure of study data import is displayed each time
FCRView is activated.
Causes Xml file remains in the temporary data folder when the DICOM data fails in importing. The
data in this file is read when the application is activated, causing the error and error
message.
Measures Xml file is transferred from the temporary data folder to the other folder when the error
occurs.
Number QCN-001965/FMS-10-835
QCN-001621/FMS-10-340
QCN-001623/FMS-10-355

3.3.3. Error corrections (Failures found by design and QA groups)

CQ555 Phenomenon It takes long to activate [Completed] button after the image import is completed
when connected to XG1.
The program has been revised to activate the button immediately after the
Measures
exposure is completed.

11/16
ECN 2011-E-0141
CQ556 Phenomenon ID Input window does not reflect the changes on the date of birth made on the
Import window.
Measures The program has been revised.
CQ557 Phenomenon The following phenomenon occurred on the patient whose setting of the neutered
CQ558 state is [Unknown] or [UnAltered]
CQ559 Although the neutered state setting is specified as [Unknown] when the patient
information is changed, only the viewer can reflect it.
The progress bar remains to be displayed when the viewer is activated after the
study data is imported
The study list does not reflect the neutered state setting specified as [Unknown]
when the study data is imported.
Measures The program has been revised to reflect the correct information.
CQ568 Phenomenon The images, which have been stored in the storage disk after the verification, are
verified again after they are rotated by 90 degrees. Although the rotated images
are imported for reference in the study utility, the images are not rotated when
imported via DICOM offline for reference.
Measures The program has been changed to import the images after rotated.
CQ578 Phenomenon Two [0002,0010] tags are written when DICOM file output is executed.
Measures The program has been revised.
CQ591 Phenomenon 39 image frames out of 40 are imported under the high-load DICOM transfer
environment.
Measures The program has been revised.
CQ596 Phenomenon Referenced Transfer Syntax UID in File does not exist in DICOMDIR of created
PDI.
Measures The program has been revised to register the corresponding tag.
CQ657 Phenomenon [Failed] is displayed on the thumbnail image when a series of 1000 images are
dropped in the image cart.
Measures The software has been revised to display the thumbnail images normally.
CQ658 Phenomenon Transferring thumbnail images after the thumbnail image output does not put the
fail names in a sequential order.
Measures The program has been revised.
CQ708 Phenomenon 20091 error occurs in case there is no images left in local HDD during the study
data transfer to NAS, requiring restart of the system to make the process of the
data transfer continue.
The software has been revised to make the process continue without a restart.
Measures Output of the 20091 error to the indicator and to the log remains for the purpose of
a notice for the users.
CQ803 Phenomenon Thumbnails change their color unusually when particular PDIs (DICOM file) are
imported.
Measures The program has been revised.

12/16
ECN 2011-E-0141

4. Installation Procedures

Procedure Reference
Upgrading application Refer to 4.1.2 in this document
software Appendix 1: Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR
391CL/CR-IR 391V CL V2.1(B)
Appendix2: FCRView V2.1(B) Connection Setup for Windows 7 Client
PC
Appendix 3: FCRView V2.1(B) Endoscopic Device Setting Procedure

OS reinstallation procedure Refer to 4.1.3 in this document.


(Windows7) Appendix 4: FCRView V2.1(B) OS Reinstallation Procedure for
Windows 7
Alternative use of NAS Refer to 4.1.4 in this document.
instead of DVD storage and Appendix 5-1: FCRView V2.1(B) NAS Quick guide
installation procedure Appendix 5-2: FCRView V2.1(B) Restoration Procedure for the Case
Images Cannot be Stored into NAS
Appendix 5-3: Procedure of Moving Study Output from Storage Disk
to NAS
Linkage of study information Refer to 4.1.4 in this document.
with DICOM MWM/ MPPS Appendix 6: FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
gateway Appendix7: Installation Support Software Exposure Menu
Import/Export Tool for Hobbit Operation Manual
Installation procedure for Refer to 4.1.4 in this document.
standard CR storage (SCP)
Connecting to Console Refer to 4.1.4 in this document.
Advance V5.0 Appendix 8: Setting Procedure for Receiving Images from Console
Advance V5.0
Changes in operation of Refer to 4.1.4 in this document.
database restoration tool Appendix 9: FCRView V2.1(B) Operation of Database Restoration Tool
Replacing patient Refer to 4.1.4 in this document.
information when receiving Appendix 10: FCRView V2.1(B) Setup Procedure for Replacing Patient
DICOM images Information at the Time of Receiving DICOM Images
Table for new and revised Refer to 4.1.5 in this document.
error codes Appendix 11: Error Code Table

4.1. Procedure for upgrading software to V2.1(B)

4.1.1. Software Versions

<CR-VW 674>
Software version Upgrade procedure
Upgrade to V5.0, and then upgrade to V2.1(B).
V1.0 (+ HotFix)
* See ECN 2009-E-0166 for the upgrade procedure to V5.0.
Upgrade to V5.0, and then upgrade to V2.1(B).
V4.0
* See ECN 2009-E-0166 for the upgrade procedure to V5.0.

13/16
ECN 2011-E-0141
Upgrade to V5.0, and then upgrade to V2.1(B).
V4.1
* See ECN 2009-E-0166 for the upgrade procedure to V5.0.
V5.0 (+ HotFix) Direct upgrade is available.
V1.1(B) (+ HotFix) Direct upgrade is available.
V2.0(B) (+HotFix) Direct upgrade is available.

<CR-IR 355V CL>


Software version Upgrade procedure
Upgrade to V3.2 or later, and upgrade to V5.0 or later, and then upgrade to V2.1(B).
V2.0 (+ HotFix) * See ECN 2008-E-0202 for the upgrade procedure to V3.2.
* See ECN 2009-E-0165 for the upgrade procedure to V5.0.
Upgrade to V3.2 or later, and upgrade to V5.0 or later, and then upgrade to V2.1(B).
V3.0 * See ECN 2008-E-0202 for the upgrade procedure to V3.2.
* See ECN 2009-E-0165 for the upgrade procedure to V5.0.
Upgrade to V3.2 or later, and upgrade to V5.0 or later, and then upgrade to V2.1(B).
V3.1 * See ECN 2008-E-0202 for the upgrade procedure to V3.2.
* See ECN 2009-E-0165 for the upgrade procedure to V5.0.
Upgrade to V5.0 or later, and then upgrade to V2.1(B).
V3.2 (+ HotFix)
* See ECN 2009-E-0165 for the upgrade procedure to V5.0.
Upgrade to V5.0 or later, and then upgrade to V2.1(B).
V4.0
* See ECN 2009-E-0165 for the upgrade procedure to V5.0.
Upgrade to V5.0 or later, and then upgrade to V2.1(B).
V4.1 (+ HotFix)
* See ECN 2009-E-0165 for the upgrade procedure to V5.0.
V5.0 (+ HotFix) Direct upgrade is available.
V5.1 (+ HotFix) Direct upgrade is available.
V1.1(B) (+ HotFix) Direct upgrade is available.
V2.0(B) (+HotFix) Direct upgrade is available.

<CR-IR 391CL/391V CL>


Software version Upgrade procedure
V1.0 (+ HotFix) Direct upgrade is available.
V1.1(B) (+ HotFix) Direct upgrade is available.
V2.0(B) (+HotFix) Direct upgrade is available.

4.1.2. Upgrading Procedure

Upgrading Procedure for application software


Refer to Appendix 1: Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V
CL V2.1(B) for the procedure.
*Perform [.Net Framework3.5] installation. (「.Net Framework2.0」→「.Net Framework3.5」)
*In case of connection setup to Client PC running on Windows7, the procedure below also needs to
be executed along with Appendix 1: Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR
391CL/CR-IR 391V CL V2.1(B)
Appendix2: FCRView V2.1(B) Connection Setup for Windows 7 Client PC
Appendix 3: FCRView V2.1(B) Endoscopic Device Setting Procedure

Upgrading to V2.1(B) when installing the unit with the earlier application software version

14/16
ECN 2011-E-0141
Because the default settings have been changed from the earlier application software versions, refer
to the service manual to perform the procedure for upgrading application software. Note that the
following operations need to be added to the procedure.
Upgrading to V2.1(B) is required in the case of the factory setting application software of the unit is
earlier version.
Perform “Initialize configuration files” in the InstallLauncher when the upgrading installation has
been completed.

4.1.3. OS Reinstallation Procedure (Windows 7)

Refer to Appendix 4: OS Reinstallation Procedure (Windows 7)

4.1.4. Other extended functions

Alternative use of NAS instead of DVD storage and installation procedure


Refer to Appendix 5-1: FCRView V2.1(B) NAS Quick guide for the details of the alternative use of NAS
instead of DVD storage and its installation procedure.
Refer to Appendix 5-2: FCRView V2.1(B) Restoration Procedure for the Case Images Cannot be
Stored into NAS for the procedure to transfer the data stored in storage disks until V2.0(B) to NAS.
Refer also to Appendix 5-3: Procedure of Moving Study Output from Storage Disk to NAS for trouble
shooting procedures when application malfunction or LAN disconnection occurs.

◆Information◆

FCRView V2.1(B) satisfies a market requirement to use NAS as an alternative storage device
instead of DVD.

Setup procedure for linkage of study information with DICOM MWM/ MPPS gateway
Refer to Appendix 6: FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure for further information.
Refer to Appendix7: Installation Support Software Exposure Menu Import/Export Tool for Hobbit
Operation Manual for the instruction for use of the tool.

◆Information◆

To avoid the manual operation to edit one exposure menu at a time in the User Utility, the tool,
which exports the CSV of the exposure menu and imports the edited menu, is provided for service
engineers.
[Menu Import Export Tool For Hobbit V1.0 (114Y5154709A00)]

Installation procedure for Standard CR Storage (SCP)


The installation procedure for SCP is the same as the procedures set forth in the following pages in
the service manual. Because it is a standard function, the installation of the option key and application
key is not necessary.
Service manual Appendix IN 5 Settings for Import of Ultrasound
Service manual Appendix IN 6 Settings for Importing/Displaying
CT/MR Images

This functions supports the following SOP Class and Transfer Syntax.
SOP Class: -Computed Radiography Image Storage
Transfer Syntax: -Implicit VR Little Endian

15/16
ECN 2011-E-0141
-Explicit VR Little Endian
-JPEG Baseline (Process 1): Default Transfer Syntax for Lossy JPEG 8 Bit
Image Compression
-JPEG Lossless, Non-Hierarchical, First-Order Prediction (Process 14
[Selection Value 1]): Default Transfer Syntax for Lossless JPEG Image
Compression

Setup Procedure for connecting to Console Advance V5.0


Refer to Appendix 8: Setting Procedure for Receiving Images from Console Advance V5.0 for setup
procedure to transfer DX images from Console Advance V5 to FCRView.

Changes in operation of database restoration tool


The DB restoration tool, which was released in a CD as a service part, has been mounted in the
installer of FCRView.
The activation procedure has been changed. Refer to Appendix 9: Operation of Database Restoration
Tool for the details

Setup Procedure for Replacing Patient Information at the Time of Receiving DICOM Images
The function replacing the patient information when importing DICOM has been added in FCRView
V2.1(B). The study information output from the equipment with DICOM connection is replaced in the
corresponding patient information in FCRView on the basis of patient ID.
Refer to Appendix10: Setup Procedure for Replacing Patient Information at the Time of Receiving
DICOM Images for further information.

4.1.5. Cautions

Table for new and revised error codes


Refer to Appendix 11: Error Code Table

Time required when terminating the service utility


The settings need to be reflected to the Windows services when terminating the service utility
because it utilizes the functions of the Windows services independently of FCRView.
It takes longer (for approximately 50 seconds) than it did to terminate the service utility as the
Windows services are shut down and then restart.

4.1.6. Restrictions

NAS for data backup


Set up a dedicated NAS when it is used for a data backup (for the use of data storage), which is
newly implemented function in V2.1(B).Do not set up this function on the NAS which has already
been set up for the original NAS functions such as mirroring and expansion of the HDD*. This is to
avoid duplication of data in one NAS.
The original functions of mirroring and expansion of the HDD can be operated together on one NAS
as they could.

*: The destination of the automatic data transfer when the local HDD is exhausted.

16/16
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V


CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

1. Outline .......................................................................................................................................... 3
1.1. Applicable model ..................................................................................................................... 3
2. Upgrading Procedure.................................................................................................................. 3
2.1. Upgrading procedure for server PC ........................................................................................ 3
2.1.1. Backing up the report format ............................................................................................. 3
2.1.2. Backing up the setting file for free drawing function .......................................................... 4
2.1.3. Noting down the setting contents for CT/MR..................................................................... 5
2.1.4. In the institutes where NAS has been installed for the purpose of the expantion of HDD . 5
2.1.5. Uninstalling the previous version of CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR
391V CL ............................................................................................................................ 5
2.1.6. Deleting Internet Explorer cache ....................................................................................... 6
2.1.7. Inactivating the UAC function (Windows Vista/ 7) ........................................................... 13
2.1.8. Installing .NET Framework 3.5, NET Framework 3.5Language Pack, and Microsoft Chart
Control ............................................................................................................................ 17
2.1.9. Installing the Server application ...................................................................................... 21
2.1.10. Restoring the backup of report format............................................................................. 23
2.1.11. Overwriting the setting file for free drawing function........................................................ 23
2.1.12. Changing Wallpaper (Windows XP only)......................................................................... 24
2.1.13. Setting up endoscopic device ......................................................................................... 24
2.1.14. Additional procedures when upgrading PCs, to which "DICOM US Storage SCP Key
Software" or "DICOM CT/MR Storage SCP Key Software" has been installed, to V2.1(B)25
2.1.15. Activating the UAC function (Windows Vista/7)............................................................... 25
2.1.16. Operation confirmation of the server PC ......................................................................... 26
2.2. Upgrading procedure for client PC ........................................................................................ 28
2.2.1. Backing up the setting file for free drawing function ........................................................ 28
2.2.2. Uninstalling the previous version of CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR
391V CL .......................................................................................................................... 28
2.2.3. Deleting Internet Explorer cache ..................................................................................... 28
2.2.4. Turning off the UAC function (for Windows Vista) ........................................................... 28
2.2.5. Installing Microsoft .NET Framework3.5 [Windows XP and Vista only] ........................... 28
2.2.6. Installing Microsoft Chart Control [all the OS (Windows XP, Vista, 7)]............................. 30
2.2.7. Installing the Client application........................................................................................ 31
2.2.8. Connection setup for Windows 7 client PC ..................................................................... 32
2.2.9. Overwriting the setting file for free drawing function........................................................ 32
2.2.10. Setting the link target (Client PC) .................................................................................... 32
2.2.11. Turning on the UAC function (Windows Vista/7) ............................................................. 33

1/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.2.12. Changing Wallpaper (Windows XP only)......................................................................... 34


2.2.13. Operation confirmation of the client PC........................................................................... 34
3. Replacing and Placing the Reference Guide .......................................................................... 34
3.1. Replacing the electronic data of the Reference Guide .......................................................... 34
3.2. Placing the Reference Guide ................................................................................................ 34
4. Precautions during Service Operation .................................................................................... 35
4.1. Precautions for configuring settings ...................................................................................... 35

2/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

1. Outline
Since "Net Framework3.5" has to be installed in V2.1(B), upgrading procedure has been changed.
The procedure is described in this appendix.

1.1. Applicable model


Model Applicable version Remarks
V5.0 HF included
CR-VW674 V1.1(B) HF included
V2.0(B) HF included
V5.0 HF included
V5.1 HF included
CR-IR 355V CL
V1.1(B) HF included
V2.0(B) HF included
V1.0 HF included
CR-IR 391 CL/CR-IR 391V
V1.1(B) HF included
CL
V2.0(B) HF included
* "HF included" means that all the minor versions within the version are applicable.

2. Upgrading Procedure
2.1. Upgrading procedure for server PC
2.1.1. Backing up the report format
‘Reference‘
If the report format has been changed, since the settings before upgrading cannot be taken over, the
setting file needs to be backed up before uninstalling the application.
Terminate the application if running.

At the time of upgrading, double-click


"Tool\BackupSystemReportFormat\BackupSystemReportFormat.bat" in the installation media. Report
format of the system will be backed up.

The following backup file will be created according to the OS:


[Windows XP]
C:\Documents and Settings\All Users\Application Data\FujiFilm\Hobbit\SystemReportFormat.bak
[Windows Vista, Windows 7]
C:\ProgramData\ FujiFilm\Hobbit\SystemReportFormat.bak

3/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.1.2. Backing up the setting file for free drawing function


‘Reference‘
When upgrading from V5.0, V5.1, or V1.0(B), since the settings before upgrading cannot be taken over,
the setting file needs to be backed up beforehand. (Backing up is not necessary when upgrading from
V2.0(B).)
Terminate the application if running.

Copy the file "C:\Program Files\FujiFilm\Hobbit\System\ FDDSettingInfo.ini" onto the desktop, for
example.

4/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.1.3. Noting down the setting contents for CT/MR


The noted contents are going to be used in "2.1.15 Additional procedures when upgrading PCs, to
which 'DICOM US Storage SCP Key Software' or 'DICOM CT/MR Storage SCP Key Software' has been
installed, to V2.1(B)" in this appendix.

1) Launch the Service Utility, and click [Connected equipment] - [Server (this equipment)] - [DICOM
setting] to open the "Set up DICOM." dialog.
→ Note down the setting contents.

2.1.4. In the institutes where NAS has been installed for the purpose of the expantion of HDD
The service settings for the expansion of HDD need to be configured again after upgrading since
the previois settings cannot be taken over. Note down the setting contents before upgrading.
Refer to “2.3 Setting in the Service Utility” under Appendix 10 “Settings for connection of the NAS”
in the Service Manual (014-227-05E) for the setup procesure.

2.1.5. Uninstalling the previous version of CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR
391V CL
Caution
Confirm that there is no unconfirmed study or study being output.

1) Start up the application, and terminate the server/client application and the installation wizard.

2) Insert the installation CD into the DVD drive. Select [My Computer] from the [Start] menu at the left
bottom of the screen. Double-click [DVD RW drive (D:)], and then [INSTALLLAUNCHER].

5/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

3) Click.

4) When the “Uninstsll or Change program” of "Programs and features" dialog opens, select [FCR
Application] and click [Uninstall].
(Ex. Vista)

1)
5) Remove the installation CD from the DVD drive and restart the PC.

Caution
The PC must be restarted.
When the PC has restarted, enter "fujicr" for Username and "system" for Password, and log in.
If the PC is not restarted, the application installation may fail.
*Multiple users may be displayed in some cases. Select also “fujicr” and enter “system” for Password.
*Since application has been uninstalled, the screen has no wallpaper in its background (turned in black).

2.1.6. Deleting Internet Explorer cache


1) Start up Internet Explorer.
‘Information‘
How to start up Internet Explorer:
Windows XP:

From the menu, select [Run]. The [Run] dialog opens.

Input “iexplore” and click [OK].

6/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

The Internet Explorer starts.

Windows Vista / 7:

From the , input “iexplore” to “Start Search”.

Other startup procedure:


(1) Press the “Windows” key + "R" keys to open the "Run" dialog.
(2) Input "iexplore" in the "Run" dialog and click [OK].

2) Clear cache with the procedures according to the version of Internet Explorer.

(a) Internet Explorer 6


1. Select [Internet Options] from [Tools] of the Internet Explorer menu, and then select the
[General] tab.

7/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2. Click [Delete Files...] under [Temporary Internet files] on the middle of the dialog.
→The "Delete Files" dialog opens.
3. Click the [OK] button without placing a checkmark to "Delete all offline content" to clear
cache.

4. Click the [Delete Cookies...] button.


→The "Delete Cookies" dialog opens.
5. Click the [OK] button to clear cookies.

8/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

6. Click the [OK] button on the bottom frame of the "Internet Options" window to return to the
Internet Explorer window, and then terminate the Internet Explorer.

(b) Internet Explorer 7:


1. Select [Internet Options] from [Tools] of the Internet Explorer menu, and then select the
[General] tab.

2. Click [Delete...] under [Browsing history] on the middle of the dialog.

9/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

3. Click [Delete files...] in the "Delete Browsing History" window.

→The "Delete Files" dialog opens.

4. Click [Yes].

5. Click [Delete cookies...] in the "Delete Browsing History" window.


→ The "Delete Cookies" dialog opens.
6. Click [Yes].

7. Click the [Close] button at the bottom of the "Delete Browsing History" window to close the
"Delete Browsing History" window.
8. Click the [OK] button on the bottom frame of "Internet Options" window to return to the
Internet Explorer window.

10/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

(c) Internet Explorer 8:


1. Select [Internet Options] from [Tools] of the Internet Explorer menu, and then select the
[General] tab.

2. Click [Delete...] under [Browsing history] on the middle of the dialog.

11/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

3. In the "Delete Browsing History" window, place a checkmark to the following and remove the
checkmarks from the others.
- Temporary Internet files
- Cookies
- History

4. Click the [Delete] button.

5. Click the [OK] button on the bottom frame of the "Internet Options" window to return to the
Internet Explorer window.

6. Close the Internet Explorer window.

12/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.1.7. Inactivating the UAC function (Windows Vista/ 7)


With Windows Vista/7, UAC function needs to be turned off when upgrading the software. (If not, a lot of
dialogs asking for your permission will open and the installation may not be done properly.)
When the installation of the software has completed, UAC function needs to be activate as well.

Windows Vista
1) Open [Control Panel] - [User Accounts].

2) Click [Turn User Account Control on or off], remove the checkmark from "Use User Account Control
(UAC) to help protect your computer", and click [OK].

3) When the "Windows needs your permission to continue" dialog opens, click [Continue].

4) If you are requested to restart the PC, click [Restart Now].


5) When the initial window opens after the PC has restarted, sequentially click its upper left and upper right
corners within a period of 5 seconds, and then click "Cancel" in the opened
"ServiceUtilityAuthentication" dialog.

Windows 7
1) Select “Control Panel” from the [start] menu.
The “Control Panel” window will open.

13/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2) Click “View by: Category” on the right of the window and Select “Large icons”.
The Control Panel will turn to “Large icons” view.

3) Double click “User Accounts”.


The “Make changes to your user account” window will open.

4) Select “Change User Account Control settings”.


The “User Account Control Settings” Window will open.

14/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

5) Adjust the slide bar to “Never notify”.

6) Check to see that the following message displayed on the right of the window.
“Never notify me when:
• Programs try to install software or make changes to my computer.
• I make changes to Windows settings.

7) Click [OK].
User Account Control confirmation window will open.

8) Click [Yes].
Return to the Control Panel window.

15/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

9) Double-click “Action Center”.


”Action Center” window will open.

10) Click [Restart this computer].


The PC will restart.

16/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.1.8. Installing .NET Framework 3.5, NET Framework 3.5Language Pack, and Microsoft Chart
Control

‘Information‘
In V2.0(B), ".Net Framework2.0" was used, but in V2.1(B), it has been changed to ".Net Framework3.5".
Moreover, as "Microsoft Chart Controls for Microsoft .NET Framework 3.5" is used in measurement
summary form this version, perform the installation for this as well.
Software to be installed Windows XP Windows Vista Windows 7
.NET Framework 3.5*1 √ √ -
Microsoft Chart Control √ √ √
* √ : OS to which the software needs to be installed.
- : OS to which the software is equipped by default; and thus, does not need to be installed.
1
* In Windows 7, "NET Framework 3.5 is installed by default; therefore, only "Microsoft Chart Control"
needs to be installed.

1) Insert the application installation media into the DVD drive.


2) Open the Explorer.
(a) [Windows XP, Windows Vista]
Open [My Computer] or [Computer] from the [Start] menu.
(b) [Windows 7]

Open [Computer] from the menu.

3) Double-click [DVD-RAM Drive(D:)] or [DVD-RW Drive(D:)].


→ The installation launcher starts up.
→ If a folder opens, double-click InstallLauncher to start up the installation launcher.

4) Click .
* The figure here shows "CR-VW 674"; however, the name of the model to which the installation is
performed will be displayed.

5) Click .

6) Click .

17/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

7) Click [Next].

* The installed software is displayed in green and the software which has not yet been installed is
displayed in gray. (The figure here is the one when none of the software has been installed.)

(a) Installing ".NET Framework 3.5" (only for Windows XP and Windows Vista)
1. The "Microsoft .NET Framework 3.5 SP1 Setup" window opens.

→ Select "I have read and ACCEPT the terms of the License Agreement",
and click [Install >].

18/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2. Installation starts. When the installation finishes normally, the "Setup Complete" window opens.

→ Click [Exit].
If you are requested to restarted the PC, restart the PC. And then, resume the procedure.

‘Information‘
When the PC has restarted, enter "fujicr" for Username and "system" for Password, and log
in.
Multiple users may be displayed in some cases. Select also “fujicr” and enter “system” for
Password.

(b) Installing "Microsoft Chart Control" (for Windows XP, Windows Vista, and Windows 7)

1. Perform the steps from step 2) to 7) again.

2. The "Welcome to the Microsoft Chart Controls for Microsoft .NET Framework 3.5 software
update" window opens.

→ Click [Next >].

19/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

3. The "Microsoft Software License Terms" window opens.


4. Place a checkmark to "I have read and accept the license terms", and click [Next >].

→ Installation starts. When the installation finishes normally, the "The software update has
been installed successfully" window opens. Then, click [Finish].

20/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

8) The system returns to the installation launcher window. Confirm that a checkmark is on ".Net Framework
has already been installed.".

* The figure here shows "CR-VW 674"; however, the name of the model to which the installation
is performed will be displayed.

2.1.9. Installing the Server application


1) Insert the installation CD into the DVD drive. Select [Computer] from the [Start] menu at the left bottom
of the screen. Double-click [DVD RW drive (D:)], and then [INSTALLLAUNCHER].

2) Click .
* The figure here shows "CR-VW 674"; however, the name of the model to which the installation is
performed will be displayed.

3) Click .
* The figure here shows "CR-VW 674"; however, the name of the model to which the installation is
performed will be displayed.
→ After a while, the “Welcome to the InstallShield Wizard for FCR Application” window opens.

21/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

4) Click .
→ The “Ready to Install the Program” window opens.

5) Click .
→ Installation starts. When the installation finishes normally, the "Installation Wizard Completed"
window opens.

6) Click .

→ The confirmation dialog for restarting the OS opens.

7) Remove the installation CD from the DVD drive, and click [Yes].
→ The PC restarts.

22/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.1.10. Restoring the backup of report format


Overwrite the corresponding file with the backup file, which has been backed up in "2.1.1. Backing up the
report format" in this appendix.

1) Restoring report format of the system


Manually copy the backup file onto the corresponding file.
[Windows XP]
- Copy source
C:\Documents and Settings\All Users\Application Data\FujiFilm\Hobbit\SystemReportFormat.bak
- Copy destination
C:\Documents and Settings\All Users\Application Data\FujiFilm\Hobbit\SystemReportFormat

[Windows Vista and Windows 7]


- Copy source
C:\ProgramData\ FujiFilm\Hobbit\SystemReportFormat.bak
- Copy destination
C:\ProgramData\ FujiFilm\Hobbit\SystemReportFormat

2.1.11. Overwriting the setting file for free drawing function


Overwrite “C:\Program File\FujiFilm\Hobbit\System\FDDSettingInfo.ini” with “FDDSettingInfo.ini” file, which
has been backed up in “2.1.2 Backing up the setting file for free drawing function” in this appendix.

‘Reference‘
If the settings for free drawing function had not been changed, and the setting file was not backed up
in 2.1.2, this step can be skipped.

23/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.1.12. Changing Wallpaper (Windows XP only)


With Windows XP, the wallpaper has to be changed manually.
‘Reference‘
To reduce the file size of the installer, wallpapers have been changed from BMP format to JPEG format
from 2.0(B).
If the OS is Windows XP, JPEG file cannot be set to wallpaper by API, so it has to be set manually. Setting
can be performed anytime after V2.1(B) has been installed.

After upgrading, the previous wallpaper is not taken over and the background turns into blue.

2) Click [Control Panel] – [Appearance and Themes] – [Change the desktop background] - [Display
Properties] dialog - [Desktop] tab.

1)
2)
3) Select [FCR_WallPaper], and click [Apply(A)].
1)
2)
2.1.13. Setting up endoscopic device
In applicable institutes, settings of Service Utility must be configured again even if endoscopic images are
to be imported in the conventional manner.

For details, see " [Appendix 3] V2.1(B) Endoscopic Device Setting Procedure".
3) ‘Reference‘
With upgrading to V2.1(B), endoscopic function has been extended, and settings have been greatly
changed:
- Supports automatic import of endoscopic studies.
- Using the patient ID as a key, provides the patient information in FCRView to automatically imported
endoscopic studies.
- Supports connection with multiple endoscopic devices (up to 4).

24/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.1.14. Additional procedures when upgrading PCs, to which "DICOM US Storage SCP Key
Software" or "DICOM CT/MR Storage SCP Key Software" has been installed, to V2.1(B)
1) Launch the Service Utility.
2) Click [Connected equipment] -> [Server (this equipment)] -> [DICOM setting] -> [Storage SCP settings].
→ If settings have not been changed, leave them as they are.
3) Click [OK], click [OK] again in the "Server equipment settings" window, and close the Service Utility.

4) ‘Reference‘
If the above procedure is not performed, multi-frame data cannot be imported. Performing the procedure
enables to import multi-frame data corresponding to each option.
5)

2.1.15. Activating the UAC function (Windows Vista/7)


With Windows Vista/7, the UAC function needs to be turned back to on after installing the software.
Terminate the application if running.

Windows Vista
1) Open [Control Panel] - [User Accounts].
2) Click [Turn User Account Control on or off], place a checkmark to "Use User Account Control (UAC) to
help protect your computer", and click [OK].

3) If you are requested to restart the PC, click [Restart Now].

Windows 7
1) Open [Control Panel] - [User Account]‐[User Account Control Setting]
2) Adjust the slide bar to “Default-Notify me when programs try to make changes to my computer”,
and then click [OK].
3) Restart the computer.

25/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.1.16. Operation confirmation of the server PC

1) Confirm that the version display on the bottom right is correct.

2) Confirm that the following are displayed after logging in.


- "Main" window
- Separated "Indicator"
- "Viewer" window
3) Confirmation of the version information and veterinary image processing software.
Select [Version] from [Utility] on the bottom left of the "Main" window.

26/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

Version: Shall be displayed as "-B V2.01.0000".

Human: shall not be dispalyed.


Veterinary: shall be displayed.

27/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.2. Upgrading procedure for client PC


2.2.1. Backing up the setting file for free drawing function
‘Reference‘
When upgrading from V5.0, V5.1, or V1.0(B), since the settings before upgrading cannot be taken over,
the setting file needs to be backed up beforehand. (Backing up is not necessary when upgrading from
V2.0(B).)

Copy the file "C:\Program Files\FujiFilm\Hobbit\System\ FDDSettingInfo.ini" onto the desktop, for example.

2.2.2. Uninstalling the previous version of CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V
CL
Uninstall the application.
1) Double-click [Add/Remove Programs] in the [Start] menu - [Settings] - [Control Panel]. (with Windows
XP)
* Windows Vista: Double-click [Programs and Features] in the [Start] menu - [Control Panel] -
[Programs].
* Windows 7: Click [Programs and Features] in the [Start] menu - [Control Panel]
(- [Programs]). (Item in parenthesis is displayed when "Category" is selected for "View by".)
2) After the "Add or Remove Programs" dialog opens, select [FCR Client Application] and click [Uninstall].
3) After the uninstallation completes, restart the PC.

Caution
The PC must be restarted.
When the PC has restarted, enter "fujicr" for Username and "system" for Password, and log in.
If the PC is not restarted, the application installation may fail.

2.2.3. Deleting Internet Explorer cache


Details are same as "2.1.6. Deleting Internet Explorer cache".

2.2.4. Turning off the UAC function (for Windows Vista)


Details are same as "2.1.7. Turning off the UAC function (for Windows Vista)".

2.2.5. Installing Microsoft .NET Framework3.5 [Windows XP and Vista only]

1) Insert the FCR Client Application media into the DVD drive of the client PC.
→ When the "InstallShield Wizard (InstallShield Wizard)" dialog opens, click [Cancel].

2) From the Windows menu, select [My Computer] or [Computer].

3) Right-click [DVD-RAM Drive (D:)] or [DVD-RW Drive (D:)], and select [Open].

28/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

4) Open the "D:\dotNet\DNET3.5" folder, and double-click the "dotnetfx35" icon.


→ The "Microsoft .NET Framework 3.5 SP1 Setup" window opens.
5) Select "I have read and ACCEPT the terms of the License Agreement", and click [Install >].

→ Installation starts. When the installation finishes normally, the "Setup Complete" window opens.

6) Click [Exit].
→ The system returns to the desktop screen.
→ If you are requested to restart the PC, restart it.

29/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.2.6. Installing Microsoft Chart Control [all the OS (Windows XP, Vista, 7)]

1) Open the "D:\dotNET\DNET3.5CHART" folder, and double-click the "MSChart" icon.


→ The "Welcome to the Microsoft Chart Controls for Microsoft .NET Framework 3.5 software update."
window opens.

2) Click [Next>].
→ The "Microsoft Software License Terms" window opens.
3) Place a checkmark to "I have read and accept the license terms.", and click [Next>].

→ Installation starts. When the installation finishes normally, the message "The software update
has been installed successfully." appears. Then, click [Finish].

30/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.2.7. Installing the Client application


1) Insert the "Client Application SoftWare" CD-ROM into the DVD drive.
→ The upgrading confirmation window of "FCR Client Application" opens.

2) Click [Yes].
→ After the InstallShield Wizard starts up, the "Welcome to the InstallShield Wizard for FCR Application"
window opens.

31/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

Click [Next].
→ The "Ready to Install the Program" window opens.

3) Click [Install].
→ Installation starts.
After a while, the installation completion window opens.

4) Click [Finish].
→ The installation completion window closes.

5) Remove the "Client Application Software" CD-ROM from the DVD drive.

6) Confirm that the server application icon is displayed on the desktop.

2.2.8. Connection setup for Windows 7 client PC


With V2.1(B), startup procedure of the application in the Windows 7 environment has been changed;
therefore, connection setup for client PCs needs to be changed as well.
For details, see appendix, "[ECN Appendix 2] Connection Setup for Windows 7 Client PC".

2.2.9. Overwriting the setting file for free drawing function


Overwrite “C:\Program File\FujiFilm\Hobbit\System\FDDSettingInfo.ini” with “FDDSettingInfo.ini” file,
which has been backed up in “2.2.1 Backing up the setting file for free drawing function” of this
Appendix.

‘Reference‘
If the settings for free drawing function had not been changed, and the setting file was not backed up in
2.2.1, this step can be skipped.

2.2.10. Setting the link target (Client PC)


When upgrading from V5.0/V5.1, the host name of CR-VW674 /CR-IR 355V needs to be
registered in the shortcut of the Client PC.
1. Right-click the shortcut on the desktop of the Client PC. Select [Properties]

32/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2. Input “C:/Program Files/FujiFilm/Hobbit/System/LogUpload.exe” “CVXXXXXXXX” in [Target]


of the [shortcut] tab.
NOTE:
“CVXXXXXXXX” represents the host name of the Server.

3. Click [OK]. → “Properties” closes.

2.2.11. Turning on the UAC function (Windows Vista/7)


Details are same as "2.1.15.

33/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

2.2.12. Changing Wallpaper (Windows XP only)


Details are same as "2.1.12.

2.2.13. Operation confirmation of the client PC


Confirm that the application can be started up.

3. Replacing and Placing the Reference Guide


3.1. Replacing the electronic data of the Reference Guide
If the electronic data of the Reference Guide of the previous version is on the desktop, delete this, and replace
to the latest electronic data of the Reference Guide in the Reference Guide CD. If not, install Adobe Reader
and place the latest electronic data of the Reference Guide onto the desktop. Use the following Reference
Guide CD.

3.2. Placing the Reference Guide


1) If Reference Guide is on the desktop, delete it.
2) Insert the Reference Guide CD into the DVD drive.
3) Open "D:\ReferenceGuide\English".
4) Copy necessary Reference Guide onto the desktop.
5) Remove the Reference Guide CD from the DVD drive.

34/35
[ECN Appendix 1] Upgrading Software to CR-VW 674/CR-IR 355V CL/CR-IR 391CL/CR-IR 391V CL V2.1(B)

4. Precautions during Service Operation


4.1. Precautions for configuring settings

1) Enabling the auto-verifying function


Considering the operability when the "Main" window and the "Viewer" window are horizontally aligned,
the Verify button is not displayed in the Viewer window by default. Therefore, enable the auto-verifying
at the time of installation using the Installation Wizard or the User Utility.

2) Settings for image incidental information in the User Utility


Settings for image incidental information in the Viewer window are changed from [Customize] in the
Viewer window.
Settings for image incidental information in the User Utility are not functioning. Even if the settings are
changed, the changes are not reflected to the Viewer window. The button for this setting is preserved for
the purpose of referring the previous settings when upgrading from V1.0(B) or before.

35/35
[ECN Appendix 2] Connection Setup for Windows 7 Client PC

[ECN Appendix 2] Connection Setup for Windows 7 Client PC


1. Outline
With V2.1(B), startup procedure of the application in the Windows 7 environment has been changed;
therefore, connection setup for client PCs needs to be changed as well.

2. Applicable model
Client PCs for the following applications:
- CR-IR 355CL/355V CL V2.1(B)
- CR-VW 674 V2.1(B)

3. Connection setup for client PCs

1. Press the "Windows" key, and select "Control Panel" → "System and Security" → "Administrative
Tools" → "Task Scheduler".
→ The "Task Scheduler" window opens.

2. Select "Task Scheduler Library" in the "Task Scheduler" window. Right-click displayed "FCR_Client"
and select "Properties".

→ The "FCR_Client Properties (Local Computer)" window opens.

1/3
[ECN Appendix 2] Connection Setup for Windows 7 Client PC

3. Click the "Actions" tab in the "FCR_Client Properties (Local Computer)" window, select "Start a
program" from the following option, and click "Edit...".

→ The "Edit Action" window opens.

2/3
[ECN Appendix 2] Connection Setup for Windows 7 Client PC

4. Enter the host name of the server in the text box for "Add arguments", and click "OK".

5. Click the "OK" button to close the "FCR_Client Properties (Local Computer)" window.

3/3
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting


Procedure 355V CL VW674

1. Outline
Endoscope function has been extended in FCRView V2.1(B) as follows:
- Supports automatic import of endoscopic studies.
- Using the patient ID as a key, provides the patient information in FCRView to automatically imported
endoscopic studies.
- Supports connection with multiple endoscopic devices (up to 4).

This appendix describes setting procedure for connection with endoscopic devices in FCRView V2.1(B),
including the functions mentioned above.

2. Applicable device
FCRView
New installation site of V2.1(B) which is equipped with endoscopic devices

Endoscope
FUJINON VP-4450HD
FUJINON VP-4450
FUJINON VP-4400
FUJINON EPX-3010

1/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

3. Procedure to Set up FCRView


3.1. Network Connection
Connect the FCRView and the PC of endoscopic device through networking as illustrated below.
100BASE-TX Cable

TCP/IP

CR-IR 355CL
FCRView FUJINON VP-4400

‹Note‹
- Do not connect a telephone line to the network connector on FCRView and the endoscopic device.
- Cables used for connection to FCRView and the endoscopic device should be of the type compliant
with the IEC950/UL1950 standards.

3.2. Installing the Option Key


The "Endoscope Image Importer" option key is required.

‹Note‹
When the option key and/or the application key is to be installed, it is required that the FCRView is
connected to the network.
Make sure that network connection has been established before performing the installation.

1. Insert an option key CD into the DVD drive.

2. Select "Computer" from the menu.


3. Double-click the DVD drive "(D:)".
→ The installation start window opens.

4. Click [Continue].
→ A window opens to indicate the completion of installation.

2/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

5. Click [OK].
→ The system returns to the desktop screen.
6. Remove the CD from the DVD drive.

3.3. Reinstalling the AP Key


Once you have installed the option key, be sure to reinstall the AP key.
Reinstall the AP key using the AP key CD. The installation procedure is as follows:

1. Insert the AP key CD into the DVD drive.

2. Select "Computer" from the menu.

3. Double-click the DVD drive "(D:)".


→ The installation start window opens.

4. Click [Continue].
→ A window opens to indicate the completion of installation.

5. Click [OK].
→ The system returns to the desktop screen.
6. Remove the CD from the DVD drive.

3/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

3.4. Adding/editing settings for/deleting endoscopic devices

(1) Launch the Service Utility.


(2) Click the "Linkage function" button.

(3) Click the "Endoscope Image Import" button.

(4) In the "Endoscope Image Import" window, perform the following in accordance with the changed
contents.

4/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

- Set the first unit of endoscopic device (including the case that the application was upgraded to
V2.1(B) in an institute where an endoscopic device has been used with V2.0(B) or before)
→Select a setting shown in the list, and click "Change". Proceed to step (5)
- Add an endoscopic device
→Create an arbitrary receiving folder under the C drive, and click "Add". Proceed to step (5)
* Up to 4 devices can be added.
- Change a setting for existing endoscopic device
→Select a setting you wish to change from the list, and click "Change". Proceed to step (5).
- Delete a setting for existing endoscopic device
→Select a setting you wish to delete from the list, and click "Delete". Proceed to step (7).

5/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

(5) Configure each of the following settings in the "Adding (Changing) endoscopic image sending
devices" window.

No. Item Contents


1 Supported. Set up validity of the endoscopic device setting itself.
(Uncheck only when temporarily disable the setting.)
2 Equipment name Set up an arbitrary device name (name for distinguishing settings)
with English single byte characters.
The device name must not duplicate with others.
3 Receive Folder Select a folder for receiving images from the endoscopic device in
the folder selecting dialog opened by the clicking the "Browse"
button.
4 Connection method Select [FTP].
5 Auto import flag Set up validity of the automatic import.
ON : Supports the automatic import
OFF : Imports images only manually, not automatically (the
same operation as the conventional endoscopic image import)
6 Flag for patient information Using the patient ID as a key, replaces the study information (patient
replacement information within it) with the patient information in FCRView.
ON : Use the patient information in the FCRView side. If a Recept
computer/MWM is connected, patient information in them is
used in according to the priority setting for the search object.
OFF : Use the patient information sent from the endoscopic device.

6/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

7 Modality The string set for modality (0008, 0060) when performing automatic
import.
Generally, the default string, "ES", is used.
8 FPT connected user name Select the user name of the FTP server.
When setting up multiple endoscopic devices connected to the FTP
server, select different users (duplicate settings are invalid). The FTP
user name in the endoscopic device has to be set in accordance with
the user name set here. See "4. Procedure to set up endoscopic
device".
* In V2.1(B), when the endoscopic image import option is enabled, the following endoscopic device will
be set up.
However, you have to open the "Changing endoscopic image sending device information"
window once and click "Change" to enable this setting. If this setting has not been enabled, an
error will occur at the startup and FCRView will not start up.

■Default settings
Equipment name : FujinonES
Receive Folder : C:\FUJINON_Data
(The folder above will be created when V2.1(B) is installed.)
Connection method : FTP
Auto import flag : OFF
Flag for patient information replacement : ON
Modality : ES
FPT connected user name : FujinonES

(6) After inputting each item, click the "Add (Change)" button. To revoke the setting and go back to the
previous window, click the "Cancel" button.
(7) To continuously change settings, repeat operations from step (4).
(8) Click "Finish" in the "Endoscope Image Import" window.

7/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

3.5. Common settings for endoscopic image import system

Settings in "System" - ”Settings for endoscopic image import” in the Service Utility have been
changed as follows:

Change Number Setup item name Default setup value


Deleted 20-1 Support availability on endoscopic image import (Deleted as the setting
and the receiving system procedure has been changed to
the one described in 3.1)
Deleted 20-2 Receiving folder for endoscopic image import (Deleted as the setting
procedure has been changed to
the one described in 3.1)
Added 20-4 Polling interval for checking receive folder (sec) 10
Added 20-5 Interval of study receiving completion when 3600
connecting with M-Receive (sec)
Added 20-7 Temporary folder for endoscopic image import C:\FUJINON_Auto

20-4: Polling interval for checking receive folder (sec)


Polling interval for monitoring endoscopic image reception folder and performing import
■Setting range: [1 - 120]
20-5: Interval of study receiving completion when connecting with M-Receive (sec)
No need to set up this item as this is only for Japan.
20-7: Temporary folder for endoscopic image import
Import control folder used in functional structure

8/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

3.6. Create a Temporary Folder


Create a folder for ” 20-7 Temporary folder for endoscopic image import” setup by “3.5. Common
settings for endoscopic image import system “. Default setup is C:\FUJINON_Auto.

Caution:
Unless creating the folder, Error 20167 occurs when starting the server application.

4. Endoscopic Device Setting Procedure

4.1. VP-4450HD / VP-4450 / VP-4400

When connecting FCRVIEW with Advancia (VP-4450), perform the following procedures.

Connecting Procedures
1. Confirm the settings on Advancia are as follows.

‹Note‹
If settings are different from the following, change them to the following.

Item Setup Value


“Login Name” The " FPT connected user name " set in Service
Utility of Hobbit (Default: “FujinonES”)
“Password” VP-4400
“CF Card Backup” Yes
2Compression Rage” under “CF Card Setup” 1/5(JPEG) or 1/10(JPEG) or 1/20(JPEG)
“Wait Time for Image Capture” 5 [s]
“Display Info.” under “Display Setting” Patient ID

Precautions
● Setting the Image Compression Ratio
‹Note‹
Be sure to select items other than "1/1(TIFF)".
The compression ratio for images taken with Advancia (VP-4450) can be selected from following.
- 1/1(TIFF)
- 1/5(JPEG)
- 1/10(JPEG)
- 1/20(JPEG)
When connecting to the FCRView, only "JPEG" can be used.
Note that, if set to "TIFF", images cannot be imported to FCRView.

● Setting “Display Info.” under “Display Setting”"

9/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

‹Note‹
Be sure to select "Patient ID".
The “Display Info.” under “Display Setting” can be selected from the following.
- Patient ID
- Study number
When connecting to the FCRView, only "Patient ID" can be used.
Do not set to "Study number". Note that, if set to "Study number", studies cannot be imported to
FCRVIEW.

● Backing up to CF Card
If you are to set "Yes" for “CF Card Backup”, the CF card must be inserted.
Prepare a CF card (sold separately) supported by Advancia and insert into the CF card slot
beforehand.

● Confirming Version Information


Confirm that the versions on the Advancia side are the following or the later.
Press the <Shift> key and "COMMENT" on Advancia to confirm.

Software Ver1.101-1215
MainCPU 0
MainDSP 0x00150001
MainBoardFPGA 237-0x8B05
OS 305
Driver 02.00.02
Image Processing Parameters 0.0009
FrontCPU Ver0.330
RearCPU Ver0.505
MD2503 Ver0.023
HIST Ver4.020
AFE Ver1.02
PATIENT PCB Ver0.00

10/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

4.2. EPX-3010
Change the Login Name to ” FPT connected user name” setup by the service UTL of FCRView.

When connecting FCRView with Justia (EPX-3010), perform the following procedures.

Connecting Procedures
1. In the "Adding endoscopic image sending devices" window of the Service Utility, change the following
item from the default settings.
Receive Folder: DCIM
2. Press <F9> key on the keyboard connecting to the endoscopic device.
→ The following window opens.

3. Select "2. Network", and press the <Enter> key.


4. Select "Yes" for "Login Setup", and press the <Enter> key.
5. Select "Yes" for "Login Name Change", and press the <Enter> key.
6. Change the Login Name to "FujinonES".
7. Select "Yes" for "End", and press the <Enter> key.
8. Select "Yes" for "Login Password Change", and press the <Enter> key.
9. Select "Input New Password", and enter "VP-4400".
10. Select "Retype New Password", and enter the same password as in step 9.
11.Select "Yes" for "End", and press the <Enter> key.
12.Select "Yes" for "End", and press the <Enter> key.
13.Press the "Esc" key.
14.Following the message displayed on the screen, restart the endoscopic device.

Precautions
● Setting Image Output Folder
1. Press the <Ctrl> key and the <F1> key simultaneously.
→ The "Network Setup" window opens.
2. Set the "Root Folder" (the folder to which the images taken will be output) to "DCIM".

‹Note‹
11/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

Be sure to select "DCIM" for "Root Folder".


With Justia (EPX-3010), the following can be selected.
- "DCIM"
- "DCIM + Serial" (Serial means the serial number of the endoscopic device.)
When connecting to the FCRView, only "DCIM" can be used.
If you select "DCIM + Serial" as a Root Folder, FCRView does not recognize the folder even if
you create the folder in FCRView.

● Confirming Version Information


1. Press the <Shift> key and the <F5> key simultaneously.
→ The version information display menu opens.

2. Confirm that the versions on the Justia side are the following or the later.

MainCPU 1.100 or later


MainPLD 0.017 or later
FrontCPU 1.003 or later
FrontSCP 0.001 or later
DspCPU 1.002 or later
MD2503 0.015 or later
LsCPU 0.053 or later
LanCPU 1.04 or later
LanPLD 0120 or later

12/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

5. Confirmation procedure
Set and confirm automatic output.
* For details on setup procedure, see "Appendix 4 Settings for Automatic Output" under "MU:
Maintenance Utility" of the service manual of CR-VW 674/CR-IR 391CL.

5.1. When automatic import is OFF


5.1.1. Starting up FCRView

1. Double-click on the desktop.


→ The FCRView opening window appears.
In about one minute, the FCRView login window opens.
2. Log in to the system.
→ The FCRVIEW application starts.
*For information on how to log in to the FCRView system, see "■ Preparatory work (login)” in “5.3
Checking Every Step from the Inputting of Images to the Saving of Images onto DVD” under “IN:
Installation.” of the Service Manual of CR-VW 674/CR-IR 391CL.

5.1.2. Importing Images


1. Select "Import" tab from the "Main" window, and click "ES".

→ The "Examination data importing window" window opens.

‹Note‹
After receiving the image, do not access the receive folder from explorer.

‘Reference‘
If the receive folder is opened using the explorer with its View set to Thumbnails, the message "34925:
Unsupported images exist in the endoscopic image receiving folder." is displayed.
This is due to the "Thumbs.db"(hidden file) file which is automatically created in the receive folder by
Windows system.
13/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

2. Select the study to be imported and select [Start Import].

3. Click .
→ The system returns to the "Main" window.

4. Click "Refresh".

5. Double-click the imported endoscopic study, and confirm that the images are displayed in the
"Viewer" window.

14/15
[ECN Appendix 3] FCRView V2.1(B) Endoscopic Device Setting Procedure

5.1.3. Outputting Images


5.1.3.1. Saving images to Storage Disk
Confirm that the endoscopic images are stored into the storage disk.
*For details on outputting images, see the Reference Guide.

5.1.3.2. Outputting General-Purpose Images


Confirm that the endoscopic images can be saved in the following format.

Format Content
JPEG Lossy JPEG lossy compression images
DICOM Stream Abstract syntax: Secondary Capture Image Storage

*For details on setting or output confirmation, see "Appendix 1 Settings for Output of Thumbnail Images"
under "Appx IN: Installation – Appendix" of the Service Manual of CR-VW 674/CR-IR 391CL.

5.2. When automatic import is ON


i. Transfer images from the endoscopic device.
ii. Update the Patient List in the "Main" window. Confirm that the transferred endoscopic study is
displayed.
* It may take some time to import depending on the timing, but it will be done within one minute at the
latest if it has been set properly.
iii. Confirm that the study in the Patient List is displayed in the Viewer when it is double-clicked and that
the number of images transferred and the number of images displayed are the same.

6. Backing up the Setup File


Back up the FCRView setup file.
*For details of the backup procedure, see “6.1 Backing Up the FCRView Setup File” under “IN:
Installation” of the Service Manual of CR-VW 674/CR-IR 391CL.

7. Precaution
None

15/15
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

[ECN Appendix 4] FCRView V2.1(B)


OS Reinstallation Procedure for Windows 7

1. Setting of the PC ................................................................................... 3

1.1. Initializing HDD ...............................................................................................3


1.2. Reinstalling Windows 7................................................................................7
1.3. Setting the User Account Control (Inactivate) ..................................11
1.4. Installing the Driver Software .................................................................14
1.5. Installing the Chipset drivers ..................................................................15
1.6. Installing the video driver .........................................................................15
1.7. Installing the network driver ...................................................................15
1.8. Installing the audio driver .........................................................................15
1.9. Check the Device Manager .......................................................................16
1.10. Installing the.NET Framework 1.1..........................................................16
1.11. Installing Adobe Reader ............................................................................19
1.12. Defining the date format and character to be used as a decimal
point .................................................................................................................19
1.13. Setting up the taskbar ...............................................................................22
1.14. Display settings ............................................................................................24
1.15. Setting up the power option and password requirement on wakeup
...........................................................................................................................24

1.16. Setting up Security Center: automatic updating setting ................26


1.17. Setting up Action Center ...........................................................................27
1.18. Setting up Defragmentation .....................................................................28
1.19. Additional Settings ......................................................................................30
1.20. Setting Adobe Reader ................................................................................31
1.21. Setting the IP address................................................................................31
1.22. Setting the Network Discovery ...............................................................32
1.23. Setting to share DVD drive (for Client PC only) .................................33
1.24. Adding Components ....................................................................................34

1/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

1.25. Copying Reference Guide..........................................................................36


1.26. Installation of RU PC Tool .........................................................................37
1.27. Installation of MSChart ..............................................................................37

2. Installing Application ............................................................................. 39

2.1. Application Installation (Server PC) ......................................................39


2.2. Application Installation (Client PC) .......................................................49
2.3. Setting the link target (Client PC) ..........................................................50
2.4. Installing Veterinary Image Processing Software (For CR-IR 355V
CL (Server PC)/CR-IR 391V CL) ...............................................................52
2.5. Installing Standard Kit for overseas (For Server PC only)..............52
2.6. Setting the User Account Control (Activate) ......................................52
2.7. Installing Application Key Software (for Server PC only) ...............52
2.8. Setting procedures for Veterinary..........................................................53

2/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

1. Setting of the PC

NOTE:
The procedures described in this chapter are for installing applications on CR-VW 674
(FCRView). In the case of the models other than CR-VW674 (FCRView), please apply the
appropriate model names/codes accordingly when following the procedures.

CR-IR 355V CL FCR CAPSULA V VIEW


CR-IR 391CL FCR PRIMA Console
CR-IR 391V CL FCR PRIMA V Console

1.1. Initializing HDD


1) Turn ON the monitor and PC power switches. The PC starts running.
2) Press the <F12> key when appears “F12 Boot Options” at the upper right corner of the
window.

The “Boot Device Menu” window will open.


3) Insert the “Windows 7 installation DVD” into the DVD drive.
4) Select “Onboard or USB CD-ROM Drive” using arrow keys, and press the <Enter> key.
The “Press any key to boot from CD...” message will appear on the bottom of the window.
5) Press the <Enter> key within three seconds.
The language selection window will open.

3/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

6) Select “My language is English”.


The “Enter your language and other preferences” window will open.
7) Click [Next].
The “Install now” window will open.

*In case a dialogue “ Windows found problems with your computer’s startup options. Do
you want to apply repairs and restart your computer?” is displayed, click [repair and
restart], and then follow the procedure below.
In case the dialogue above is not displayed, skip to 8)
a) Press <F12> when “DELL” screen appears.
b) Select [Onboard or USB CD-ROM Drive] from the Boot Menu and press [Enter]
key.
c) “Press any key to boot from CD or DVD” will be displayed. Press [Enter] key.
d) “Installing Windows “ dialogue will be displayed. Select [My language is English]
e) A dialog to select languages will be displayed. Click [Next] without selecting any
language.
f) The “Install now” window will open. Click [Repair my computer] on lower left of

4/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

the dialogue.

8) Select “Repair your computer” at the bottom left of the window.


The “System Recovery Options” window will open.

9) Click [Next].
The “Choose a recovery tool” window will open.

5/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

10) Select “Command Prompt”.


The command prompt window will open.

11) Input “diskpart” from the keyboard and press the <Enter> key.

12) Input “select△disk△0” (“△” replace a single-byte space) and press the <Enter> key.
A message saying “Disk 0 is now the selected disk” will be displayed in the command prompt
window.

13) Input “clean” from the keyboard and press the <Enter> key.

14) Confirm a massage saying “Disk Part succeeded in cleaning the disk.” and close the
command prompt window by clicking its [x] mark.
The system will return to the “Choose a recovery tool” window.

6/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

15) Click the [Restart].


The PC will restart.

1.2. Reinstalling Windows 7


1) Press the <F12> key when “F12 Boot Options” appear at the upper right corner of the window.
2) Select “Onboard or USB CD-ROM Drive” using arrow keys, and press the <Enter> key.
The “Press any key to boot from CD...” dialogue appears on the bottom of the window.
3) Press the <Enter> key within three seconds.
The language selection window will open.

4) Select “My language is English”.


The “Enter your language and other preferences” window will open.

7/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

5) Confirm language and other preferences and Click [Next].


The “Install now” window will open.

6) Select “Install now”.


The “Please read the license terms” window will open.

7) Place a checkmark to “I accept the license terms” and click [Next].


The “Which type of installation do you want?” window will open.

8/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

8) Select “Custom(advanced)”. Then “Where do you want to install Windows?” dialogue box will
open.

.
9) Select “Disk 0 Unallocated Space” and click [Next], and then the OS installation processing
will start.
Upon completion of installation, the “Choose a user name for your account and …” window
will open.

10) Make the following settings and click [Next].

9/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

・ Type a user name (for example, John): fujicr


・ Type a computer name: CVXXXXXXXX (“CV” in capital letters; enter the last
eight digits of Serial Number for “xxxxxxxx”)
The “Set a password for your account” window will open.

11) Make the following settings and click [Next].


• Type a password (recommended) : system
• Retype your password: system
• Type a password hint (required) : fujicr
12) The “Help protect your computer and improve Windows automatically” window will open.
• Select “Ask me later”.
The “Review your time and date settings” window will open.

10/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

13) Check the “Time zone”, “Date” and “Time” and click [Next]
The “Select your computer’s current location” window will open.

14) Select “Public network”.


The desktop window will open.

1.3. Setting the User Account Control (Inactivate)


Turn off and disable “User Account Control” from prompting for credentials to install applications,
and changing the elevation prompt behavior.

1) Select “Control Panel” from the [start] menu.


The “Control Panel” window will open.

2) Click “View by: Category” on the right of the window and Select “Large icons”.
The Control Panel will turn to “Large icons” view.

11/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

3) Double click “User Accounts”.


The “Make changes to your user account” window will open.

4) Select “Change User Account Control settings”.


The “User Account Control Settings” Window will open.

5) Adjust the slide bar to “Never notify”.

12/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

6) Check to see that the following message displayed on the right of the window.
“Never notify me when:
• Programs try to install software or make changes to my computer.
• I make changes to Windows settings.

7) Click [OK].
User Account Control confirmation window will open.

8) Click [Yes].
Return to the Control Panel window.

13/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

9) Double-click “Action Center”.


”Action Center” window will open.

10) Click [Restart this computer].


The PC will restart.

1.4. Installing the Driver Software


■ Installing the Dell Resource DVD utility software
1) Insert “DRIVERS AND UTILITIES ALREADY INSTALLED ON YOUR COMPUTER” DVD into
the PC.
The “AutoPlay” window will open.

2) Place a checkmark to “Always do this for software and games” and click “Run autoRcd.exe”.
The “Welcome Dell System Owner “ window will open.
3) Place checkmark to “Click here to prevent this information from being displayed again” and
click [Next].
The Dell Resource CD utility software will start running.
4) Click [Drivers]

14/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

”OptiPlex 780 Windows 7 32-bit Driver”

1.5. Installing the Chipset drivers


CR-IR 391CL/CR-IR 391V CL:
See [1.3.2 Installing the Driver Software] - [● Installing the Controller drivers] under “MC:
Checks, Replacement and Adjustment of Parts” of the “CR-IR 391CL Service Manual”.
CR-VW 674/CR-IR 355V CL:
See [1.4.2 Installing the Driver Software] - [● Installing the Chipset drivers] under “MC:
Checks,Replacement and Adjustment of Parts” of the “CR-VW 674 Service Manual”.

NOTE:
Select appropriate version at “Rev.****” of each driver.

1.6. Installing the video driver


CR-IR 391CL/CR-IR 391V CL:
See [1.3.2 Installing the Driver Software] - [● Installing the Video drivers] under “MC: Checks,
Replacement and Adjustment of Parts” of the “CR-IR 391CL Service Manual”.
CR-VW 674/CR-IR 355V CL:
See [1.4.2 Installing the Driver Software] - [● Installing the video driver] under “MC: Checks,
Replacement and Adjustment of Parts” of the “CR-VW 674 Service Manual”.

NOTE:
Select appropriate version at “Rev.****” of each driver.

1.7. Installing the network driver


CR-IR 391CL/CR-IR 391V CL:
See [1.3.2 Installing the Driver Software] - [● Installing the Network drivers] under “MC: Checks,
Replacement and Adjustment of Parts” of the “CR-IR 391CL Service Manual”.
CR-VW 674/CR-IR 355V CL:
See [1.4.2 Installing the Driver Software] - [● Installing the network driver] under “MC: Checks,
Replacement and Adjustment of Parts” of the “CR-VW 674 Service Manual”.

NOTE:
Select appropriate version at “Rev.****” of each driver.

1.8. Installing the audio driver

15/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

CR-IR 391CL/CR-IR 391V CL:


See [1.3.2 Installing the Driver Software] - [● Installing the Audio drivers] under “MC: Checks,
Replacement and Adjustment of Parts” of the “CR-IR 391CL Service Manual”.
CR-VW 674/CR-IR 355V CL:
See [1.4.2 Installing the Driver Software] - [● Installing the audio driver] under “MC: Checks,
Replacement and Adjustment of Parts” of the “CR-VW 674 Service Manual”

NOTE:
Select appropriate version at “Rev.****” of each driver.

1.9. Check the Device Manager


To determine what driver software needs to be installed, you can use Device Manager.
To open Device Manager, perform the following steps:
1) Right-click “Computer” from the menu and select “Properties” from the pull-down menu that
opens.
The “System” window will open.
2) Select “Device Manager”.
The “Device Manager” window will open.
3) The software for items marked “!” in Device Manager is not installed. Install the driver software
for such items.

1.10. Installing the.NET Framework 1.1


1) Insert the CR-VW 674 application CD into the DVD drive.
NOTE:
Click [Finish] to close “InstallLauncher” dialogue box in case the dialogue box appears.
The installation start confirmation window will open.

NOTE:
For Client PC, the “InstallShield Wizard ” dialogue will be displayed. Click [Cancel]. In case
[Program Compatibility Assistant] dialog is displayed after clicking [Cancel], click [This program
installed correctly].
For Client, "Welcome to the InstallShield Wizard for FCR Client Application” dialogue is
displayed when [Cancel] is not selected. Then click [Cancel] and perform the following
procedure;
a) ”InstallShield Wizard Completed” dialogue is displayed. Then click [Finish].
b) ”Program Compatibility Assistant” is displayed. Then click [This program installed

16/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

correctly].

2) Select “Computer” from the [Windows] menu.


The “Computer” window will open.
3) Right-click “D drive” and select “open”.
4) Open folders in the following sequence: “dotNet”, “DNET1.1”, and “DNET1.1”.
NOTE:
* For the Client PC, open [D:¥dotNet¥DNET1.1] folder and double-click [dotnetfx.exe].

5) Double-click “dotnetfx”.
The installation start confirmation window will open.

6) Click [Yes].
In a while, the “License Agreement” window will open.

7) Select the “I agree” radio button, and click [Install].


Installation processing will start. When the installation completes, the completion window will
open.

8) Click [OK].
9) Open folders in the following sequence: “D drive”, “dotNet”, “DNET1.1”, and “DNET1.1SP1”.
NOTE:
*For the Client PC, open [D:¥dotNet¥DNET1.1SP1].

17/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

10) Double-click “NDP1.1sp1-KB867460-X86”.


The confirmation window for starting installation of “.NET Framework 1.1 Service Pack 1” will
open.

11) Click [OK].


The supplemental end user license agreement window will open.

12) Click [I accept].


Installation processing will start. When the installation completes, the completion window will
open.

18/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

13) Click [OK].

1.11. Installing Adobe Reader


Refer to [1.9.4 Installing Adobe Reader and Copying the Reference Guide] under “MC: Checks,
Replacement and Adjustment of Parts” of the CR-VW 674/CR-IR 391CL Service Manual
(014-227-05E).

INFORMATION:
Use following updated parameters to install Adobe Reader:
Step 2:
[“D:/Adobe Reader/English/AdbeRdr910_en_US_Std.exe”]
→[“D:/Adobe Reader/English/AdbeRdr940_en_US.exe”]
Step 3:
“Adobe Reader 9” → ”Adobe Reader 9. 4.0”/
Step 5‐Ⅰ:
[Remove the checkmark from “General” - “Check for updates”.]
→[Remove the checkmark from “Updater” - “Check for updates”.]

1.12. Defining the date format and character to be used as a decimal point
1) Select [Control Panel] from the menu.
2) Double-click “Regional and Language”
The “Regional and Language” window will open.
3) Click the “Formats” tab to select a language (region) to use from the pull-down menu in
“Current format”.
The “Customize Regional Options” window will open.

19/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

4) Click [Additional Setting].


The [Customize Format] window will open.

5) Select the “Number” tab.


▪ [Decimal symbol]: “.” (period)
▪ [Display leading zeros]: “0.7”

20/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

6) Select the “Date” tab.


▪ [Date formats]‐[Short date] : “M/d/yyyy”

7) Select the “time” tab.


▪ [Time formats]‐[Short Time] : “HH:mm” (24h)

21/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

▪ [Time formats]‐[Long Time] : “HH:mm:ss” (24h)

1.13. Setting up the taskbar


Change the taskbar appearance settings in the following way.
1) Select [Control Panel] from the [Windows] menu.
The Control Panel will be displayed.
2) Double click [Taskbar and Start Menu].
The “Taskbar and Start Menu Properties” window will open.
3) Place a checkmark to “Auto-hide the taskbar”
“Taskbar buttons”: Select “Never combine”

22/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

4) Click [Customize...] in “Notification area”. The “Select which icons and notifications appear on
the taskber” window will open.
5) Place a checkmark to “Always show all icons and notifications on the taskbar”

23/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

6) Click [OK].
7) Click [OK].
The system will return to the Control Panel.
8) Close the Control Panel
The system will then return to the desktop.

1.14. Display settings


1) [Control Panel]‐[Display] ‐ [Adjust resolution]‐[Advanced Settings]‐[“Adapter” tab]‐
[List All Modes]
The “List All Modes” window will open.

1.15. Setting up the power option and password requirement on wakeup


1) Select [Control Panel] from the [Windows] menu.
The Control Panel will be displayed.

24/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2) Double-click [Power Options].


“Power Options” window will open.
3) Select “Requires a password on wakeup.”

“System settings” window will open.


4) Make the following settings.

Server PC* Client PC


When I press the power button Sleep Shut down

5) Click [Save change].


The system will return to the “Power Options” window.
6) Select [Preferred plan] - [Balanced (recommended)].

25/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

Select “Change plan setting.”


The “Edit Plan Settings” window will open.
7) Set “Turn off the display” to “20 minutes” and “Put the computer to sleep” to “Never”.

8) Click [Save change].


The system will return to the “Power Options” window.
9) Close the “Power Options” window.
The system will return to the desktop.

1.16. Setting up Security Center: automatic updating setting


1) Select [Control Panel] from the [Windows] menu.
The Control Panel will be displayed.
2) Double-click [Windows Update].
”Windows Update” window will open.
3) Double-click [Change settings].

26/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

The “Choose an automatic updating option” window will open.


4) Select “Never check for updates (not recommended)” and click “OK”.

The system will return to the “Windows Update” window.


5) Click [Close].
The Control Panel will be displayed.

1.17. Setting up Action Center

1) In the “Control Panel” window, double-click “Action Center”.


The “Windows Action Center” window will open.
2) Click [Action Center].
The “Action Center” window will open.

27/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

3) Click [Change Action Center settings].

The “Change Action Center settings” window will open.

4) Remove the checkmark of check boxes of [Security messages] and [Maintenance messages]
and click [OK].

5) Close all windows.

1.18. Setting up Defragmentation


1) Select “Computer” from the [Windows] menu.
The “Computer” window will open.

28/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2) Right-click “Local Disk (C:)” and select “Properties”.

.
The “Local Disk (C:) Properties” window will open.
3) Select “Tools” tab and select [Defragmenter Now].

The “Disk Defragmenter” window will open.


4) Select [Configure schedule].

29/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

The “Disk Defragmenter” window will open.


5) Remove the checkmark from “Run on a schedule (recommended)” and click [OK] .

The “Disk Defragmenter” window will open.


6) Close the “Disk Defragmenter” window.
The “Local Disk (C:) Properties” window will open.
7) Click [OK].
The “Computer” window will open.
8) Close the “Computer” window.
The system will return to the desktop screen.

1.19. Additional Settings

1.19.1. Displaying PC Lock

Setting procedures
(1) Click [Start]- [Search programs and files]. Enter “C:/ProgramData/FujiFilm/Hobbit/UserTemp”.
(2) Right-click “PCLock” and click [Pin to Taskbar”

30/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

1.20. Setting Adobe Reader


1) Double-click “Adobe Reader 9” shortcut icon on the desktop.
2) “Adobe Reader - License agreement” dialogue box will be displayed. Click [Accept].
3) “Adobe Reader” will be displayed. Select [Preferences] from [Edit] menu. Make the following
settings and click [OK].
[Updater]‐[Check for updates]: Select ”Do not download or install updates
automatically”
[Internet]‐[Display PDF in browser]: Uncheck the checkbox.
[Internet]‐[Allow speculative downloading in the background]: Uncheck the checkbox
[Trust Manager] ‐ [Load trusted root certificates from an Adobe server]: Uncheck the
checkbox.
4) Close “Adobe Reader”

1.21. Setting the IP address


1) [Control Panel] - [Network and Sharing Center] - [Change adapter settings] - [Local Area
Connection] - [Properties]
Remove the checkmark from [Internet Protocol Version 6 (TCP/IPv6)]

31/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2) [Control Panel] - [Network and Sharing Center] - [Change adapter settings]-[Local Area
Connection] - [Properties] - [Internet Protocol Version 4 (TCP/IPv4)] - [Properties]
・ Select [Use the following IP address] then set the following
IP address :Server PC [172.16.1.20]
:Client PC [172.16.1.21]*
NOTE:
*Client setting does not apply to CR-IR 391CL/CR-IR 391V CL.

Subnet mask :[255.255.0.0]


Default gateway :Blank (no input)
・ [Use the following DNS server addresses]
Preferred DNS server: Blank (no input)
Alternate DNS server :Blank (no input)

1.22. Setting the Network Discovery


1) [Control Panel] - [Network and Internet] - [Change advanced sharing settings]
・ [Public(current profile)] - [Network discovery], select [Turn on network discovery] and then
[Save changes].

32/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2) [Control Panel]-[Network and Internet]-[Network and Sharing Center]


Make sure that the following setting has been made.
・ Unidentified network: Public network

1.23. Setting to share DVD drive (for Client PC only)

NOTE:
This setting does not apply to CR-IR 391CL/CR-IR 391V CL.

1) [Control Panel] - [Folder Options]-[“View”] tub


・ Remove the checkmark Use sharing Wizard (Recommended)
* Make sure that no media has been inserted into the DVD drive.

33/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2) [Computer] - [DVD RW Drive]: right-click - [properties] - [“Sharing”] tub - [Advanced Sharing]


・ Check the checkbox for [Share this folder], then click [OK]

1.24. Adding Components


1) [Control Panel] - [Programs and Features] - [Turn Windows features on or off]
・ Check the checkboxes for the following items, and then click [OK]

34/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

only for Server

35/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

only for Server

only for Server

* After the installation, restart the PC if requested.

1.25. Copying Reference Guide


1) Insert “Reference Guide CD” into the DVD drive.

For CR-VW 674: Reference Guide CD (114Y5097051AXX)


For CR-IR 355V CL: Reference Guide CD (114Y5154063AXX)

36/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

For CR-IR 391CL: Reference Guide CD (114Y2129052AXX)


For CR-IR 391V CL: Reference Guide CD (114Y2129061AXX)

2) Open “D:¥ReferrenceGuide¥English¥WW” folder and select all the displayed PDF files in it,
then right-click to select [Copy] from the menu appeared.
3) Right-click on the desktop blank area and select [Paste] from the menu appeared.
4) Close “WW” window and remove the “Reference Guide CD” from the DVD drive.

1.26. Installation of RU PC Tool


Refer to the corresponding ECN regarding each RU.

1.27. Installation of MSChart


1) Insert FCRView Application Software into the DVD drive.
2) “CR-VW 674 Setup”(for 355CL: “CR-IR 355CL, for 355V CL: “CR-IR 355V CL Setup”, for
391:”CR-IR 391CL Setup”)dialogue is displayed. Click [Finish]. For Client, “InstallShield Wizard”
dialogue is displayed. Click [Cancel].
* In case [Program Compatibility Assistant] dialogue is displayed after clicking [Cancel] on
Client, click “This program installed” (for 355: “This program installed correctly”. For 391:
“This program installed correctly”.
* For Client, "Welcome to the InstallShield Wizard for FCR Client Application” dialogue is
displayed when [Cancel] is not selected. Then click [Cancel] and perform the following
procedure;
a) “Are you sure you want to cancel FCR Client Application setup wizard?” dialogue is
displayed. Then click [Yes].
b) “InstallShield Wizard Completed” dialogue is displayed. Then click [Finish].
3) Open [D:¥dotNet¥DNET3.5¥DNETCHART] folder for Server / [D:¥dotNet¥DNET3.5CHART]
folder for Client to double-click [MSChart.exe].
4) “Software Update Microsoft~” dialogue is displayed. Then click [Next >].

37/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

5) “Microsoft Software License Terms” dialogue is displayed. Place a checkmark to ”I have read and
accept the license terms” and click [Next >].

6) “The software update has been installed successfully” is displayed. Then click [Finish].
7) Remove the FCRView Application Software from the DVD drive.
8) Restart the PC.

38/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2. Installing Application

NOTE:
The procedures described in this chapter are for installing applications on CR-VW 674 (FCRView).
In the case of the models other than CR-VW 674 (FCRView), please apply the appropriate model
names/codes accordingly when following the procedures.

CR-IR 355V CL FCR CAPSULA V VIEW


CR-IR 391CL FCR PRIMA Console
CR-IR 391V CL FCR PRIMA V Console

2.1. Application Installation (Server PC)


1) Insert “CR-VW 674 Installation” (for 355: “CR-IR 355CL Installation”, for 391: “CR-IR 391CL
Installation”) media into the DVD drive.
2) “CR-VW 674 Setup” (for 355: “CR-IR 355CL Setup”, for 391: “CR-IR 391CL Setup”) dialogue
box will be displayed. Click [CR-VW 674 Application setup] (for 355: [CR-IR 355 CL Application
Setup], for 391: [CR-IR 391 CL Application Setup]).

3) An application selection dialogue box will be displayed. Click [Install .Net Framework]

39/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

4) Click [Install .Net Framework 1.1], and then click [Next]. (.Net Framework1.1 Activation will be
executed)

40/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

5) When returned to the Application selection dialogue box, click [Install SQL Server]

6) An installation language selection dialogue box will be displayed. Click [Other] from the list, and
the click [Next].

7) Installation of SQL Server will be executed. Press [Enter] key when “Press any key to
continue…” appears.

41/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

8) A confirmation dialogue box for restart will be displayed. Remove the installation media from the
DVD drive and click [Yes].

9) The PC is restarted.
10) The logon screen will be displayed. Make sure that the user name is “fujicr” and enter “system”
for password, and then click “->”.
11) Insert “CR-VW 674 Installation” (for 355: “CR-IR 355CL Installation”, for 391: “CR-IR 391CL
Installation”) media into the DVD drive.
12) “CR-VW 674 Setup” (for 355: “CR-IR 355CL Setup”, for 391: “CR-IR 391CL Setup”) dialogue will
be displayed. Click [CR-VW 674 Application setup] (for 355: [CR-IR 355 CL Application Setup],
for 391: [CR-IR 391 CL Application Setup]).

42/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

13) An application selection dialogue box will be displayed. Click [Install CR-VW 674 V x. x] (for 355:
[Install CR-IR 355 CL V x. x], for 391: [Install CR-IR 391 CL V x. x]).
(“x.x” refers to the version)

14) After a while, “Welcome to the InstallShield Wizard for FCR Application” dialogue box will be
displayed. Click [Next].

43/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

15) “Ready to Install the Program” dialogue box will be displayed. Click [Install].

16) After a while, “InstallShield Wizard Completed” dialogue box will be displayed. Click [Finish]. (If
the generation of DB failed during installation, installation is aborted and rollback is performed.)

44/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

17) “To continue installation, you need to restart the system” dialog box will be displayed. Click
[Yes].

18) After the PC is restarted, “[20102] Setup incompleted. Application key software is not installed”
dialogue box is displayed. Click [OK].

19) Remove the installation media from the DVD drive, and then insert it again.
20) Click [Start] - [Computer], then double-click [DVD Drive (D:)].
21) Double-click “InstallLauncher.exe “ located right under CDROOT.

22) “CR-VW 674 Setup” (for 355: “CR-IR 355CL Setup”, for 391: “CR-IR 391CL Setup”) dialogue
box will be displayed. Click [CR-VW 674 Application setup] (for 355: [CR-IR 355CL Application
setup], for 391: [CR-IR 391CL Application setup]).

45/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

23) An application selection dialogue box will be displayed. Click [Install Configuration files].

24) The next dialogue box to select the type of the facility will be displayed. Select [Hospital /
Medical Clinic] and click [Next].

NOTE:
* In the case of CR-IR 355V CL/CR-IR 391V CL, Select [Veterinary / Hospital].

46/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

25) The next dialogue box to select a language will be displayed. Select [English(Not in USA)] from
the language list and click [Next].

26) The next dialogue box will be displayed. Select [Color] from the list of the monitor and click
[Next].

27) Installation will be completed and the application selection dialogue box will return. Click [Install
Menu / Parameter files].

47/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

28) The Next dialogue box to select a language will be displayed. Select [English(Not in USA)] from
the language list and click [Next].

29) A dialogue box inquiring whether to use “FNC” will be displayed. Select [ON] then click [Next].

30) Installation will be completed and the application selection dialogue box will return. Click

48/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

[Finish].

31) Remove the media from the DVD drive.

2.2. Application Installation (Client PC)

NOTE:
This setting does not apply to CR-IR 391CL/CR-IR 391V CL.

1) Insert the Client Installation media into the DVD drive.


2) After a while, “Welcome to the InstallShield Wizard for FCR Client Application” dialogue box will
be displayed. Click [Next].

3) “Ready to Install the Program” will be displayed. Click [Install].

49/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

4) After a while, “Install Wizard Completed” will be displayed. Click [Finish].


5) From Explorer, open the “Tool” folder located right under the Client CDROOT, and then
double-click [SetupForClient.vbs].
6) A command prompt will appear and ask for the type of the facility. Enter “1” (Human Hospital)
and press [Enter] key (for 355: enter “2” (Veterinary Hospital) and press [Enter] key).
7) As “Please Select Language:” appears, enter “3” (English (Other))and press [Enter] key.
8) As “Select a type of product model” appears, enter “2” (Workstation) and press [Enter] key (for
355: enter “1” (Console) and press [Enter] key).
9) As “Select a type of monitor type” appears, enter “1” (Color) and press [Enter] key.
10) Make sure that “Success!!” is displayed in the command prompt when the settings are
successfully completed. Press [Enter] to close the command prompt.
11) Remove the media from the DVD drive.

2.3. Setting the link target (Client PC)


When upgrading from V5.0/V5.1, the host name of CR-VW674 /CR-IR 355V needs to be
registered in the shortcut of the Client PC.
1. Right-click the shortcut on the desktop of the Client PC. Select [Properties]

50/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2. Input “C:/Program Files/FujiFilm/Hobbit/System/LogUpload.exe” “CVXXXXXXXX” in [Target]


of the [shortcut] tab.
NOTE:
“CVXXXXXXXX” represents the host name of the Server.

3. Click [OK]. → “Properties” closes.

51/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2.4. Installing Veterinary Image Processing Software (For CR-IR 355V CL (Server
PC)/CR-IR 391V CL)
1) Insert the following CD attached to the product respectively into the DVD drive.
・ [CR-IR 391V Image Processing Software (114Y2129009AXX)]CD
・ [CR-IR 346V Image Processing Software (114Y5342705AXX)]CD
2) Click [Start] ‐[Computer], and then double-click [D drive].
3) Double-click “OptionVeterinary.exe “ located right below CDROOT.
4) “CR-IR ***V Image Processing Software” dialogue will be displayed. Click [Continue].
5) “Successful Application key installation” dialogue will be displayed. Click [OK].
6) Remove “V Image Processing Software” CD from the DVD drive.

2.5. Installing Standard Kit for overseas (For Server PC only)


1) Insert “Standard Kit CD” into the DVD drive.
2) Click [Start]‐[Computer], and then double-click [DVD Drive (D:)]
3) Double-click ElevateSetup.vbs located right under the CDROOT.
4) A command prompt will open and display “Press any key to continue…” Press [Enter] key.
5) As “Press Any Key” appears, press [Enter] key.
6) As “Press Any Key to continue…” appears, press [Enter] key.
7) Remove the CD from the DVD drive.

2.6. Setting the User Account Control (Activate)


1) 「Control Panel」‐「User Accounts and Family Safety」‐「User Accounts」‐「Change User
Account Control settings」
2) Choose when to be notified about changes to your computer. Adjust the slide bar to [Default
-Notify me only when programs try to make changes to my computer] and the click
[OK].
3) Restart the PC

2.7. Installing Application Key Software (for Server PC only)


1) Insert “CR-VW674 Application Key Software CD” into the DVD drive.
2) Click [Start] – [Computer], and then double-click [DVD Drive (D:)].
3) Double-click “ApplicationKey.exe” in CDROOT.
4) “Application Key” dialogue box will appear. Click [Continue].
5) A dialogue to inform of successful installation of the application key will be displayed. Click [OK].
6) Remove the CD from the DVD drive.

52/53
[ECN Appendix 4] FCRView V2.1(B) OS Reinstallation Procedure for Windows 7

2.8. Setting procedures for Veterinary


Please refer to “Various Setting Procedures for Veterinary” in CR-IR 355V CL/CR-IR 391V CL service
manual.
* Make the settings after all the installations have been completed.

[Various Setting Procedures for Veterinary] – [CR-IR 391V CL Service Manual]


[Various Setting Procedures for Veterinary] – [CR-IR 355V CL Service Manual]

53/53
[ECN Appendix 5-1] Procedure of Moving Study Output to Storage Disk to NAS

[ECN Appendix 5-1] FCRView V2.1(B) NAS Quick guide

Setting up the NAS


Connect the NAS to the FCRView, make the following procedure.
1. FCRView and the NAS connect to the same network(Service Manual Appendix 19 Setting for Connection of
the NAS).
2. Start FCRView-ServiceUtility ,make the following settings.
I. Click [Connected equipment]. Open [Setting connection devices] window.

II. Click [Backup NAS]. Open [Image save settings] window.

III. Check [Store images to NAS].

1/4
[ECN Appendix 5-1] Procedure of Moving Study Output to Storage Disk to NAS

IV. Click [Add]. Setting network of the NAS. The NAS network setting is the same to [HDD](Service
Manual 2.3 Setting in the Service Utility).
V. Change the auto backup destination folder.
Click [DBAutoBackup]. Open the [Image save setting] window(Ⅵ).

VI. Click the [Browse]. Select the auto backup destination folder.
* It is necessary for a folder of NAS to be set as a network drive. The setting of the network
drive refers to a service manual.

3. Start FCRView ,make sure the “Indicator Disc Icon” lights.

Output to the NAS


“Output to the NAS” is the same “Output to the DVD”(Reference Guide 7.3 Saving to Storage Disks). Saves image
automatically to the NAS when the study is verified. If the same study already exists on the NAS, it is overwritten.

2/4
[ECN Appendix 5-1] Procedure of Moving Study Output to Storage Disk to NAS

Loading images from the NAS


The study exists in database, but it does not exist in HDD.
Output image is associated with the study database. Thus the user can get to images from the NAS without
a special operation.

1. Starting up the “Image Viewer” to view images.

2. Loading an image from NAS into the PC.

Import from the NAS


The study does not exist in database and HDD.
Use “Import DICOM Off-line”(Reference Guide 3.5 Import Various Images and Refer).

1. Select the folder that it’s name is output date to import the study.
The NAS manages the study as unit of output date.

3/4
[ECN Appendix 5-1] Procedure of Moving Study Output to Storage Disk to NAS

2. Select the study to import from “Study list”.


3. Click [Start retrieving].

Indicate the NAS states


Connection
The Disk icon of “Indicator” represents NAS connection status.
On(Available)

Off(Unavailable)

Free space
The Disk icon of “Indicator” represents free space of NAS.
The 1st warning < Free space

The 2nd warning < Free space ≦ the 1st warning

Free space ≦ the 2nd warning

4/4
[ECN Appendix 5-2] FCRView V2.1(B) Restoration Procedure for the Case Images Cannot be Stored in NAS

[ECN Appendix 5-2] FCRView V2.1(B)


Restoration Procedure for the Case Images Cannot be Stored into NAS

1. Introduction
A trouble sometimes occurs in the operation of storing images into NAS (as a substitution for storage
disk), and images cannot be stored. This appendix describes the processes of handling this trouble.

2. Cause of the trouble


The trouble may occur if the FCRView server application is forcibly terminated or the LAN cable is
pulled out while images are being stored into NAS.
Generally, the connection can be restored by restarting the PC or reconnecting the LAN cable.

3. Restoration procedure

(1) Images cannot be stored into NAS Case #1


Perform the following procedure:
(1) Restart the FCRView server application.
(2) Check the on/off state of the NAS icon in the FCRView indicator to check whether
FCRView recognizes the NAS or not.
(3) Launch the Service Utility and select an unrecognized NAS from [Connected
equipment] → [BackupNAS]. Then, open a dialog by clicking [Change], close the dialog
by clicking [OK], and click [OK] to close the NAS setting window.
(4) Terminate the Service Utility.

* When images could not be stored into NAS even after performing the steps (1) through (4)
described above.
(1) Start up the FCRView server application.
(2) Delete the output cue to be output to NAS.
(3) Open the NAS utility window, click [Internal information reconfiguration], and take a
note of the save destination path of the present month's label name in the [Internal
information reconfiguration] dialog.
(4) Terminate the FCRView server application and restart the PC.
(5) Terminate the FCRView server application.
(6) Open the save destination path, which was noted down in step (3), using the Explorer,
and delete "DICOMDIR".
(7) Start up the FCRView server application.
(8) Open the NAS utility window and click [Internal information reconfiguration]. Then,
select the present month's label name in the [Internal information reconfiguration]
dialog and click the [Start] button.

1/2
[ECN Appendix 5-2] FCRView V2.1(B) Restoration Procedure for the Case Images Cannot be Stored in NAS

(2) Images cannot be stored into NAS Case #2


As images cannot be stored into NAS after the study database has been initialized, perform the
following. However, there is no need to perform the procedure on the new NAS as images can be
stored into it.
(1) Restart the FCRView server application.
(2) Open the NAS utility window and click [Internal information reconfiguration]. Confirm
that there is nothing in the list of [Internal information reconfiguration] dialog, and click
[Cancel].
(3) Open the NAS utility window and click [Database registration of internal information].
(4) Open the NAS utility window and click [Internal information reconfiguration]. Confirm
that there is one or more lists in the [Internal information reconfiguration] dialog, and
click [Cancel].

2/2
[ECN Appendix 5-3] Procedure of Moving Study Output to Storage Disk to NAS

[ECN Appendix 5-3] Procedure of Moving Study Output from Storage


Disk to NAS

1. Introduction
The procedure of moving study information which has been output to the storage disk with V2.0(B) or
before to NAS is described in this appendix.
However, the study information which has been output to the work disk cannot be moved to NAS.

2. Work procedure
[Preparation]
- Prepare the storage disk to be moved to NAS.
- Specify NAS as the output destination under [Connected equipment]->[BackupNAS] in the Service
Utility.

[Procedure]
1. Finish the Hobbit server application.
2. Start C:\Program Files\FujiFilm\Hobbit\System\DVDMoveToNASTool.exe.
3. Select the drive for the storage disk to be stored from the combo box named as DVD Drive
Select.
4. Insert the storage disk into the selected drive.
5. Select the moving destination of NAS drive from the combo box named as NAS Drive Select.
6. Click the [START] button.
7. The progress is displayed in the progress bar. Confirm that "Process Complete. Please check
the C:\DVDMoveToNASToolForm.log." is displayed in the message box when the process
finishes.
8. Open C:\DVDMoveToNASToolForm.log and confirm that there is no error between [Date]
ProcessSTART and [Date] ProcessEND in the log.

[Items to be checked when an error is displayed]


If the following errors are displayed while performing [Procedure], perform each corresponding
confirmation and procedure.

1. When "Please check your setting. Restart Tool." is displayed


→The settings on Hobbit side are not correct. Reconfirm the settings, and then restart and
re-execute the tool.

2. When "Storage Media is not in the DVD Drive." is displayed


→The DVD media inserted into the DVD drive is not a storage disk. Insert a storage disk and
re-execute the tool.

3. When "NAS dose not have Enough Free Space." is displayed


→The capacity of the copy destination of NAS drive is insufficient. Add a new NAS and specify it.
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure

1. Outline
This appendix describes settings for linkage of study information with DICOM MWM gateway and
settings for linkage of study result information with DICOM MPPS gateway for FCRView V2.1(B).

A part of the settings for MWM study information linkage will be common to the settings for MWM patient
information linkage. Moreover, as the DICOM AE title of FCRView for MWM service can be set from
V2.1(B), this appendix also describes about that.

2. Setting up the MWM Study Information Linkage


2.1. Setting up connection in the Service Utility

(1) Launch the Service Utility.

(2) Click the “Linkage function” button.

1/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(3) Click the “Linkage of MWM study information” button.

(4) Configure settings in the following table in the “Linkage setting of MWM study information” window.

2/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure

No. Item Content


1 Support the linkage of MWM study information Place a checkmark to "Supported".
2 Equipment name Set up the name of the equipment to be
connected.
3 Host name Set up the name of the equipment to be
connected on the network .
4 DICOM AE title Set up the DICOM AE title.
5 IP address Set up the network address of the
equipment to be connected.
6 Port number Set up the port number of the equipment to
be connected.

* Equipment name, Host name, DICOM AE title, IP address, and Port number will be common to the
settings for MWM patient information linkage.

(5) After configuring the settings, click the "OK" button.

(6) The system returns to the linkage function setting window. Click the “Back” button to return to the
Service Utility top window.

3/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(7) Click the “Connected equipment” button.

(8) Click the “Server (this equipment)” button.

4/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure

5/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(9) Click the “DICOM setting” button.

6/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(10) Configure settings in the following table under the “MWM SCU setting” tab.

No. Item Content


1 AE title Set up the DICOM AE title.
2 Timeout value Set up a timeout value.

Caution) AE title and Timeout value will be common to the settings for MWM patient information linkage.

7/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
3. Setting up the MPPS Linkage
3.1. Setting up connection in the Service Utility

(1) Launch the Service Utility.

(2) Click the “Linkage function” button.

8/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(3) Click the “MPPS linkage” button.

(4) Configure settings in the following table in the “MPPS linkage setting” window.

9/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure

No. Item Content


1 MPPS linkage support Place a checkmark to "Supported".
2 Character sets of menu name to be used as If you are in a multi-byte country and
protocol name multi-byte is supported by the equipment to
be connected, select "MultiByte". Other than
that, select "SingleByte".
3 Equipment name Set up the name of the equipment to be
connected.
4 Host name Set up the name of the equipment to be
connected on the network .
5 DICOM AE title Set up the DICOM AE title.
6 IP address Set up the network address of the
equipment to be connected.
7 Port number Set up the port number of the equipment to
be connected.

Caution) "Supported" of MPPS linkage support is checked by default. For institutes where MPPS linkage is
not used, be sure to remove the checkmark.

(5) After configuring the settings, click the "OK" button.

(6) The system returns to the linkage function setting window. Click the "Back" button to return to the
Service Utility top window.

(7) Click the "Connected equipment" button.

10/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure

11/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(8) Click the "Server (this equipment)" button.

(9) Click the "DICOM setting" button.

12/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
(10) Configure settings in the following table under the "MPPS SCU setting" tab.

No. Item Content


1 AE title Set up the DICOM AE title.
2 Timeout value Set up a timeout value.

13/14
[ECN Appendix 6] FCRView V2.1(B) DICOM MWM/MPPS Setup Procedure
4. Operation Confirmation

(1) Confirming MWM linkage:


Open the "Work List" tab in the "Main" window while CR study has been registered to the set DICOM
MWM gateway. Confirm that the acquired study is displayed under the Work List tab.

(2) Confirming MPPS linkage:


Double-click a study in the Work List to open the "Read Image" window. Confirm that the status of the
study changes on the set DICOM MPPS gateway.

(3) Confirming MPPS linkage:


Finish an exposure. Confirm that the status of the study changes.

14/14
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

Installatin Support Software


Exposure Menu Import/Export
Tool for Hobbit
Operation Manual

-1–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

Change history
Edition Changed contents
1st Create 1st edition

-2–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

Contents
1. OUTLINE ..........................................................................................................................................4

1.1. CONTENT ....................................................................................................................................4


1.2. OPERATING CONDITIONS (SYSTEM REQUIREMENTS) .......................................................................4
1.3. ESSENTIAL SOFTWARE .................................................................................................................4
1.4. TARGET HOBBIT ...........................................................................................................................4

2. SETUP PROCEDURE ......................................................................................................................5

2.1. REQUIREMENTS FOR SETUP ..........................................................................................................5


2.2. SETUP PROCEDURE......................................................................................................................5

3. OPERATION DESCRIPTIONS .........................................................................................................7

3.1. STARTUP PROCEDURE ..................................................................................................................7


3.2. WINDOWS/SCREENS ....................................................................................................................7
3.3. OPERATING PROCEDURE ..............................................................................................................9

4. IMPORT SOURCE FILE .................................................................................................................11

4.1. FILE RESTRICTIONS ....................................................................................................................11


4.2. PRECAUTIONS FOR CREATING/EDITING EXCEL FILE.......................................................................11

5. IMPORT OPERATION OVERVIEW................................................................................................13

6. EXPORT OPERATION OVERVIEW ...............................................................................................14

-3–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

1. Outline
1.1. Content
There are an enormous number of X-ray exposure menus at institutions where the ordering equipment is
installed. Some institutions may register over 10,000 menus and it would take a considerable amount of
time and labor to register a large number of menus one by one at such institutions.

The exposure menu import/export tool is a support tool to reduce such burden. The tool extracts the
exposure/study menu data from the exposure menu database (DB) and exports them to the Microsoft
Excel file, making it possible to display/edit the menu list. And the tool imports the edited exposure/study
menus.

This tool is an installation support tool for service personnel and will not be released to users.

1.2. Operating conditions (system requirements)


It has been confirmed that this tool operates under the following conditions:
- PC PC/AT-compatible PC or equivalent
- OS Microsoft Windows XP SP3 or
Microsoft Windows VISTA Business SP2 or
Microsoft Windows 7

1.3. Essential software


One of the following listed software must be installed for the tool:
- Microsoft Excel 2000 (Office 2000) * Windows XP only
- Microsoft Excel 2003 (Office 2003)
* We do not guarantee the operation of the tool with Excel versions other than listed above.

1.4. Target Hobbit


The tool is compatible with the following Hobbit exposure menu DB:
- CR-IR355CL/355V CL V2.1(B)
- CR-VW674 V2.1(B)
- CR-IR391CL/391V CL V2.1(B)

* Other than the Japanese environment, the operation under the English environment has been
confirmed. However, there is a restriction for the use (see “3.3.1. Import process”).
* Make sure the file format of the Excel file to be used can be opened with the essential software.
For example, the excel format file which is not compatible with Excel 2003 cannot be opened under the
environment in which Excel 2003 has been installed. Therefore, please confirm that before operating
the file.
* Operation under the Chinese, Taiwanese, and Korean environments has been confirmed using
Windows 7 and Excel 2003. We do not guarantee the operation with other versions.

-4–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

2. Setup procedure
2.1. Requirements for setup
- PC which fulfills conditions in 1.2 Operating conditions and in 1.3 Essential software
- Menu Import Export Tool for Hobbit application software

2.2. Setup procedure


* The CD-ROM drive is represented by E:\ here.
The screen display may slightly differ depending on the OS.

1. Insert the Menu Import Export Tool CD-R into the CD-ROM drive.

2. Execute the Setup.exe file from a browser such as Windows Explorer.

3. Click [OK] in the "MenuImportExportTool Setup" dialog.

4. Click in the "Specify Destination Directory" dialog.


If the setup destination folder has to be changed, click the [Change Directory] button and change the
directory.

5. Click [Continue] in the "Choose Program Group" dialog.


If you want to change the registration destination, change Program Group and Existing Group.

6. Progress bar is displayed and the installation starts.


In some OS or environments, a dialog like the following may be displayed.
If the dialog is displayed, click [Yes] not to overwrite the existing file.
Be careful not to overwrite the existing files by mistake. Or, other applications can become inoperable.

7. When the Setup completion dialog is displayed, click [OK] to complete the setup.

Setup is now complete.


-5–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
If any error occurs during setup, perform the procedure from the beginning.

Do not install the tool into the PC where the Hobbit has been installed.
Be sure to take the backup of the database (DB) files (MenuData.mdb, XconParam.mdb,
Display.mdb and Selector.mdb) whenever you perform conversion, considering the
possible failure of the conversion with the tool and creation of defective files.
Operating target DB files of the tool are as follows:
MenuData.mdb
XconParam.mdb
(DisplayData.mdb)
These DB files are located in the following folder:
C:\ProgramData\FujiFilm\Hobbit\Param
* If the OS is Windows XP, mdb files are located in the following folder:
C:\Documents and Settings\All Users\Application Data\ FujiFilm\Hobbit\Param
Do not replace the database files while the application is running.

-6–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

3. Operation descriptions

3.1. Startup procedure


Select [MenuImportExportTool.exe] from the registered program group.

3.2. Windows/Screens

Figure 1 Main window


The following window is displayed when the tools starts up. The window is used for various setup
operations, executions, and log display.

1 11

3 4

12 13
7 8

9 10 15
14

■Description of each item


(1) Export source folder setup button
Specify the folder where the export source database file (.mdb file) exists. When you click this button,
the folder selection dialog box [Figure 2] opens.
* You cannot directly input the path into the folder path display text box.

(2) Export file setup button


-7–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
Specify the Excel file to be exported. When you click this button, the save file selection dialog
(Windows common dialog) opens.
* You cannot directly input the path into the file path display text box.

(3) Export execution button


Executes the export process. This button is enabled by specifying the export source folder, export file,
and the log folder.
Perform the export process after closing the export file (if an existing file is specified as the
export file).

(4) Export log file display button


Displays the log file where the errors on the export process are output. This function is available when
the log file (Export.log) exists in the log folder specified in (11). It gives an error if the log file does not
exist.

(5) Import file setup button


Specify the import source Excel file. When you click this button, the file selection dialog (Windows
common dialog) opens.
* You cannot directly input the path into the file path display text box.

(6) Import destination folder setup button


Specify the folder where the import destination database file (.mdb file) exists. When you click this
button, the folder selection dialog box [Figure 2] opens.
* You cannot directly input the path into the folder path display text box.

(7) Import option setup area


(9) Specify the operation to be performed at the click of the import execution button. Choose one of
the following options:
Check Only Checks the contents of the import source file. Outputs the checking result to the log file when the
setting value is abnormal. The actual import operation is not performed.
Import Reflects the menu data registered in the import source file to the Hobbit menu database. When
an abnormal value is found in the data to be imported, the import process aborts. The menus
until abnormality detected are imported. Abnormal menus will not be registered. The menu data
with abnormality is output to the log file.

(8) Language setup area


Specify the PC environment for Console use. Choose one of the following options:
Multi Byte Used in multi byte language environment (Japanese, Chinese, Korean, and Taiwanese
environments).
Single Byte Used in single byte language environment (environments other than Japanese, Chinese,
Korean, and Taiwanese ones).

(9) Import execution button


Executes the import process. This button is enabled by specifying the import source file, import
destination folder, and log folder.

(10) Import log file display button


Displays the log file where the abnormal record values specified in the import source file or the errors
-8–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
on the import process are output. This function is available when the log file (Import.log) exists in the
log folder specified in (11). It gives an error if the log file does not exist.

(11) Log folder setup button


Specify the folder where the log file of the import/export process is output. When you click this button,
the folder selection dialog [Figure 2] opens.
* You cannot directly input the path into the folder path display text box.

(12) Progress bar


Shows the progress of the import/export process.

(13) Cancel button


Aborts the import/export process.

(14) Message display area


Displays the result of the import/export process.

(15) Exit button


Terminates the application.

Figure 2 Folder selection dialog box


This dialog box appears when you are to specify the import destination folder, export source folder, or
log folder.

3.3. Operating procedure


You can operate this tool with the following procedures.

3.3.1 Import
1. Start up the tool
2. Specify the import source Excel file (Figure 1 (5) → Select the file)
3. Specify the import destination folder (Figure 1 (6) → Figure 2)
4. Specify the log file (Figure 1 (11) → Figure 2)
5. Specify Check/Import (Figure 1 (7))
6. Setup language (Figure 1 (8))
7. Start the import process (Figure 1 (9))
Cancel (Figure 1 (13))
8. Confirm the log (Figure 1 (11))
-9–
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
9. Exit (Figure 1 (15))

Operating target files of the tool are as follows:


MenuData.mdb
XconParam.mdb
(DisplayData.mdb)
These files are located in the following folder of the PC where the Hobbit has been installed:
C:\ProgramData\FujiFilm\Hobbit\Param
* If the OS is Windows XP, mdb files are located in the following folder:
C:\Documents and Settings\All Users\Application Data\ FujiFilm\Hobbit\Param
Although only the MenuData.mdb and XconParam.mdb files are revised by the import process, the
DisplayData.mdb file must also be corresponding with these files.
Therefore, copy these three files at once, edit them by the tool, and then restore them to the Hobbit.
Separately save these files before editing them so that you can restore them to the state before editing.

When the import process is interrupted, inconsistency may occur among the database files at the import
destination.
Never use the parameter file created by interrupting.
Moreover, do not use the database file created by interrupting for re-importing process.
Backup the database files before executing the import process, and use them for re-import process.

On English environment, when a multi byte character is included in a setting value of a field in the import
source Excel file, it may not be detected as an abnormal value. When creating or editing the import source
Excel file, be sure to do so in the English environment in order to prevent multi byte characters from being
input and imported.

3.3.2 Export
1. Start up the tool
2. Specify the export source folder (Figure 1 (1) → Figure 2)
3. Specify the export file (Figure 1 (6) → Select the file)
4. Specify the log file (Figure 1 (11) → Figure 2)
5. Start the export process (Figure 1 (3))*
Cancel (Figure 1 (13))
6. Confirm the log (Figure 1 (4))
7. Exit (Figure 1 (15))

*Close the file specified in the step (3)"Specify the export file" before performing the step
(5)"Start the export process". Moreover, make sure that read-only attribution of the MDF files
existing in the export source folder has been released.

If an existing file is specified as an export destination file when performing export, the existing file
is deleted after the overwrite confirmation dialog is displayed, and then menu data are exported to
a new file with the same file name. When specifying an existing file, close the existing file before
starting the export process.

- 10 –
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

4. Import source file


4.1. File restrictions
The import source file must be created by Microsoft Excel. Use the file exported from Hobbit as a format
file.

Figure 1. Worksheet overview

The import source file consists of three sheets. The contents of each sheet are as follows:
- Sheet1: Exposure menu data
- Sheet2: Study menu data
- Sheet3: Detail study menu data

The contents of each row in the sheets are as follows:


- 1st row : Field ID…do not edit
- 2nd row : Classification (field common to CR/DR or field specific to CR)
- 3rd row : Field name
- 4th row and lower :Setting values for each menu
* The classification is for the sake of convenience. Fields are interpreted as CR field irrespective of the
classification.

4.2. Precautions for creating/editing Excel file


■Common
- Do not rearrange rows or insert new ones between 1 and 3 rows.
- Do not add a new worksheet. Use the 1st sheet for the exposure menu data, the 2nd sheet for the
- 11 –
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition
study menu data, and the 3rd sheet for the detail study data.
- Register one menu data per row.
- Register the menu data in the 4th row or lower.
- Cell format - the display format of all cells should be text.
- For configurable value of each item, see appendix, "パラメータ設定値.xls".
- Newly added exposure and study menus are not registered to the display menu group. After the import
process, register them to the display menu group as necessary in the User Utility.

■Exposure menu data


- Although it is possible to rearrange columns or insert new ones, be sure to keep the consistency
between the field IDs and their values.
- When an exposure menu is newly registered, copy the row of the base exposure menu to a blank row,
and then edit only the items which need to be changed.
- Code is a mandatory item. If its value is empty or entirely filled with spaces, its row is skipped and its
menu is not registered.
- When three or more consecutive rows are empty or entirely filled with spaces for Code, the tool
terminates the import process as it considers that the end of menu data is reached.
- On the import process, when items other than Code are empty or entirely filled with spaces, it is
considered that the cells are without data, and default values are set on registration.
- When the data type of a field in the database is Yes/No (see appendix, "パラメータ設定値.xls" - the
default value is Yes or No), 1 or 0 is set for this field (1: Yes, 0: No). If import is performed with a value
other than 1 or 0, the value is treated as an abnormal value.
- Despite deleting the row of an optional exposure menu from the Excel file exported from Hobbit, the
menu is not deleted from the exposure menu database. If you want to delete the menu from the Hobbit
exposure menu database, use the User Utility.

■Study menu
- Do not rearrange columns or insert new ones for Study menu data (Sheet 2) and Detail study menu
data (Sheet 3).
- For Detail study menu data (Sheet 3), input the study menu data in the same Code continuously. Do
not insert the data with a different Code in the middle.
- Since the study menu data in the import source file is imported without checking the values of each
field, make sure to observe the following rules when editing the fields below:
<Study menu data (Sheet 2)>
Code: Should be unique each other in the study menu data.
<Detail study menu data (Sheet 3)>
Code: Should have been defined in Study menu data (Sheet 2).
ExposureMenuCode: Should have been defined in Exposure menu data (Sheet 1).
ExposeOrder: Should be unique within the same Code in Detail study menu data (Sheet 3).
The initial value should be “0” and the number should be incremented by one.
FramePosition: Should be unique within the same Code in Detail study menu data (Sheet 3).
The initial value should be “0” and the number should be incremented by one.

- 12 –
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

5. Import Operation Overview


■When the import process is successfully completed
→ The process finishes and the following message appears on the message display area:
<When abnormal data is not detected>
"Import process is successful. Abnormal data were NOT found."
<When abnormal data is detected>
"Import process is successful. Abnormal data were found."

When a code including abnormal data is detected, its name and table/filed are output to the log.

■When the menu database (MenuData.mdb and/or XconParam.mdb) does not exist in the import
destination parameter folder
→ The process aborts and an error dialog box opens.

■When a Code field in the import source Excel file is blank, is entirely filled with spaces, or includes an
unacceptable character
→ The tool does not import the affected row (record) and proceeds to the next one.

■When the record of a duplicate menu code exists in the import destination menu database for Console
→ The following dialog box opens, asking whether you want to overwrite.

Click one of the following buttons to proceed the process:


Yes Overwrites this menu data only.
Yes All Overwrites all the subsequent duplicate items of menu data and another
confirmation dialog box will not open.
No Does not overwrite this menu data.
No All Does not overwrite any of the subsequent duplicate items of menu data and
another confirmation dialog box will not open.

■When a menu data in the import source Excel file is blank or abnormal
→ The tool gives an error and aborts the process. The content of the error is output to the log.

■When three or more consecutive rows in the import source Excel file have no menu data
→ The tool terminates the process as it considers that the end of menu data is reached.

■When an error occurs during the process


→ The tool opens an error dialog box and aborts the process. The content of the error is output to the
log. The following message appears on the message display area.
"An error has occurred. Import is not completed."

- 13 –
Installation Support Software
[ECN Appendix 7-1] Exposure Menu Import/Export Tool for Hobbit Operation Manual: First edition

6. Export Operation Overview


■When the export process is successfully completed
→ The process finishes and the following message appears on the message display area
"Export process is successful."

■When the menu database (MenuData.mdb and/or XconParam.mdb) does not exist in the export
source parameter folder
→ The process aborts and an error dialog box opens.

■When the specified export file already exists


→ A confirmation dialog box opens, asking whether you want to overwrite.

■When an error occurs during the process


→ The tool opens an error dialog box and aborts the process. The content of the error is output to the
log. The following message appears on the message display area:
"An error has occurred. Export is not completed."

Period

- 14 –
[ECN Appendix 8] Setting Procedure for Receiving Images from Console Advance V5.0

[ECN Appendix 8] Setting Procedure for Receiving Images from Console Advance
V5.0

1. Outline
This appendix describes setting procedure for receiving images from Console Advance V5.0.

2. Applicable model
FCRView
- CR-VW674 V2.1(B)
Console Advance
- Console Advance V5.0

When connecting to Console Advance V5.0, connecting configuration vary depending on the installation
environment including image input device.

When Console Advance is connected to only CR Image Reader:


→ Images are received with DICOM Private CR Storage (conventional function)
When Console Advance is connected to only FPD:
→ Images are received with DICOM Digital X-Ray Image Storage - For Presentation
When Console Advance is connected to both CR Image Reader and FPD:
→ Hobbit V2.0(B)…Image are received with DICOM Digital X-Ray Image Storage - For
Presentation
Hobbit V2.1(B)…Images are received with DICOM Standard CR Storage
or with DICOM Digital X-Ray Image Storage - For Presentation
3. Setting procedure
■Setting of Hobbit 1 (DICOM SCP setting)
1. Launch the Service Utility.
→ The "Service Utility top" window opens.

2. Click [Connected equipment].


→ The "Setting connection devices" window opens.

3. Click [Server ( equipment itself)].


→ The "Server equipment settings" window opens.

4. Click [DICOM setting].


→ The "Set up DICOM." window opens.

5. Select the [Storage SCP settings] tab and confirm the following settings (no need to change them from
the defaults).
AE Name : "FCR-SCP" (default)
Port number : "5001" (default)
Timeout value : "300" (default)

1/3
[ECN Appendix 8] Setting Procedure for Receiving Images from Console Advance V5.0

6. Click [OK].
→ The system returns to the "Server equipment settings" window.
7. Click [OK].
→ The system returns to the "Setting connection devices" window.
8. Click [Back].
→ The system returns to the "Service Utility top" window.

■Setting of FCRView 2 (Setting "Default Patient ID Value")


Same as the settings described in "2.3 Setting the 'Default Patient ID Value'" in "Appendix 8 Image
Reception from the CR Console (CR-VW 674 Only)" under Appx IN of CR-VW 674 / CR-IR 391CL Service
Manual.

■Settign of FCRView 3 (LUT Setting)


Change the “Gradation characteristic of image” of DX to P-Value. Refer " 6.6 Modality Settings'" in " 6.
Service Settings" under MU of CR-VW 674 / CR-IR 391CL Service Manual.

2/3
[ECN Appendix 8] Setting Procedure for Receiving Images from Console Advance V5.0

■Setting of Console Advance


1. When receiving images with DICOM CR Storage
See "Appendix 10 DICOM CR Storage" under Appx IN of DR-ID300CL Service Manual.

2. When receiving images with DICOM Digital X-Ray Image Storage - For Presentation
See "Appendix 11 DICOM DX Storage: Transfer of Processed Images" under Appx IN of
DR-ID300CL Service Manual.

When receiving DR images, configure settings as follows so that images are output with the gradation
process applied.

"2.1.4 Registration of Information on Other Connected Equipment for the DICOM DX Storage
Function"

In step 13, select "Apply LUT correction process to the image"*.


* Perform the settings under "2.1.9 Display Gradation Correction Setup".

In step 15, select "14" for Bit Stored.

■Reception Confirmation
1. Start up the FCRView client.

2. Click the [Patient List] tab to open the patient list.

3. Transfer a study from Console Advance.

4. The study transferred in step 3 is displayed on the patient list.


In this condition, confirm the following:
- [Patient ID] and [Patient Name] corresponds to the study transferred in step 3.
- [Modality] is "CR" or "DX" (depending on the modality transferred).
- When you double-click the study, the image is displayed in the Viewer.

3/3
[ECN Appendix 9] FCRView V2.1(B) Operation of Database Restoration Tool

[ECN Appendix 9] FCRView V2.1(B) Operation of Database Restoration


Tool

1. Outline
Since the database restoration tool released in FCRView V2.0(B) as a service part has some changes in
V2.1(B), description on those changes and supplementary on the operation are described in this
appendix.

2. Applicable items
The database restoration tools to be released in FCRView V2.1(B) or later are applicable.
The database restoration tool for V2.0(B) released in 2010-E-0356 does not work in V2.1(B) or
later.

3. Changed contents
Database Restoration Tool is installed together with the application software of V2.1(B) or later.
Some changes in the startup procedure and functions have been made from the one provided for
V2.0(B).

(1) Startup procedure


From the [Start] menu, input "RestorationTool" in the Start Search field and press the "Enter"
key.

(2) Startup process 1


When the restoration tool starts up, the following message (Yes/No) appears to confirm with user.
The tool shows a dialog if the reply is Yes, and ends if the reply is No.

“Before start this software, check that following applications are not running.
- Server Application
- ServiceUtility
- UserUtility
- BackupUtility”

(3) Startup process 2


The Windows service, added in V2.1(B), is terminated.

4. Operation
After operating the restoration tool, perform the following:

- After applying the restoration tool, restart the PC.


- Confirm all the settings after the restoration is executed.
This is because the restoration tool only restores the FCRView configuration file but does not
create folders.

1/1
[ECN Appendix 10] FCRView V2.1(B) Procedure to set up for Replacing Patient Information at the Time of Receiving DICOM
Images

[ECN Appendix 10] FCRView V2.1(B) Setup Procedure for Replacing


Patient Information at the Time of Receiving DICOM Images

1. Outline
In FCRView V2.1(B), a function to replace patient information at the time of receiving DICOM images
has been added. This function replaces patient information attached to various types of studies sent
from DICOM connection devices with the patient information in the FCRView, using the patient ID as a
key.
This function enables to display necessary information when confirming images in the Viewer of the
FCRView, even with images from devices which attaches little amount of patient information. This
function also enables to match patient information among devices.

This appendix describes the setup procedure for the replacement of patient information and the
procedure to confirm the settings.

2. Applicable items
Institutes of which devices* are connected with FCRView V2.1(B) by DICOM connection
*On March 2011, the following modalities are applicable:
- CR (Standard / Private)
- DX (for Presentation)
- CT
- MR
- XA
- RF
- US (including Multi-frame)
- SC (including Multi-frame)

1/4
[ECN Appendix 10] FCRView V2.1(B) Procedure to set up for Replacing Patient Information at the Time of Receiving DICOM
Images

3. Procedure to set up FCRView


3.1. Adding/editing settings for/deleting devices to be targeted for replacing patient information
(1) Start the Service Utility.
(2) Click the "Linkage function" button.

(3) Click the "DICOM Reception" button.

2/4
[ECN Appendix 10] FCRView V2.1(B) Procedure to set up for Replacing Patient Information at the Time of Receiving DICOM
Images

(4) In the "DICOM Reception" window, perform the following in accordance with the changed contents.

- Newly set up or add a device to be targeted for replacing patient information


→ Click "Add". Proceed to step (5).
- Change a setting for existing device targeted for replacing patient information
→ Select a setting you wish to change from the list, and click "Change". Proceed to step (5).
- Delete a setting for existing device targeted for replacing patient information
→ Select a setting you wish to delete from the list, and click "Delete".
(5) Configure each of the following settings in the "Adding (Changing) DICOM image sending device
information " window.

No. Item Contents


1 Equipment name Arbitrarily describe an explanation of the connected device.
This item is used to prevent mis-setting when multiple similar devices
are to be connected.
2 AE title Set up an AE title of the targeted DICOM connection device with
English single byte characters.

3/4
[ECN Appendix 10] FCRView V2.1(B) Procedure to set up for Replacing Patient Information at the Time of Receiving DICOM
Images

This item must consist with the AE title setting on the DICOM
connection device (Storage-SCU) side; therefore, be sure to check
the setting of the DICOM connection device in advance.
3 Flag for patient information Switch to enable/disable the setting.
replacement

(6) After configuring settings, click "Add". To revoke the setting and go back to the previous window,
click "Cancel".
(7) To continuously change settings, repeat operations from step (4).
(8) Click "Finish" in the "Patient information notification" window.

4. Confirmation procedure
Send a study to FCRView from the connected device.
Select the sent Patient ID in the "Main" window - "Patient List", and display the sent study in the Viewer.
Confirm that the information displayed in the "Patient List" and the study information displayed in the
Viewer window are the same.

5. Precaution
With this setting, if priority search object for patient information is set to Recept Computer or MWM in
FCRView, patient information in them is used as a priority.

4/4
[ECN Appendix 11] Error Message Table
[ECN Appendix 11] Error Message Table
Event Source Message/Description Message Causes Remedies Remarks

16000 GeneralLiblary.CustomMessag [16000] Failed in deleting template files. *Restart the server computer.

eBox

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

16001 GeneralLiblary.CustomMessag [16001] The file in the output destination cannot be

eBox written because the file is being used in another

process.

16002 GeneralLiblary.CustomMessag [16002] Failed in outputting CSV file. *Restart the server computer.

eBox

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

16003 GeneralLiblary.CustomMessag [16003] Expiry date of ACTIVE LINE has been

eBox expired. If you want continuously to use this function,

update the agreement to extend the expiry date.

16004 GeneralLiblary.CustomMessag [16004] The process has been canceled because the

eBox extension for the unsent billing logs has been

exceeded. Contact your technical support.

16005 GeneralLiblary.CustomMessag [16005] Failed in outputting the report. *Restart the server computer.

eBox

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

16006 GeneralLiblary.CustomMessag [16006] Failed in acquiring the measurement data *Restart the server computer.

eBox from the server.

>Contact your service engineer if the same symptom

1/28
[ECN Appendix 11] Error Message Table

recurs even after it has been restarted.

16007 GeneralLiblary.CustomMessag [16007] The name of legend has not been entered.

eBox

16008 GeneralLiblary.CustomMessag [16008] The information other than numerical value

eBox is entered to the lower limit value.

16009 GeneralLiblary.CustomMessag [16009] The information other than numerical value

eBox is entered to the upper limit value.

16010 GeneralLiblary.CustomMessag [16010] Study assessment has not been entered.

eBox

16011 GeneralLiblary.CustomMessag [16011] Conditional equation is incorrect. *Enter a conditional equation correctly.

eBox

16012 GeneralLiblary.CustomMessag [16012] Evaluation result of conditional equation is *Modify the conditional equation so that the

eBox not a bool value (true or false). evaluation result becomes a bool value (true or

false).

16013 GeneralLiblary.CustomMessag [16013] Template file does not exist. Report output *Select a correct template file in the system settings

eBox was canceled. for measurement summary.

16014 GeneralLiblary.CustomMessag [16014] No base line value has been registered. *Click the Add button and add a baseline value.

eBox

16015 GeneralLiblary.CustomMessag [16015] Failed in reading the DICOM tag information.

eBox

16016 GeneralLiblary.CustomMessag [16016] DICOM tag information file does not exist. *Carry it out again after a while.

eBox

16017 ImageViewer.ImageView.Shape [16017] Failed in BMD measurement. *Ensure that the image is for BMD measurement.

s.BoneMineralShape

*Use the images and perform the measurement in

2/28
[ECN Appendix 11] Error Message Table

accordance with the guideline. Contact your service

engineer if the same symptom recurs.

16018 ImageViewer.ImageView.Shape [16018] The image file required for BMD does not *Restart the server computer.

s.BoneMineralShape exist.

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

16019 ImageViewer.ImageView.Shape [16019] The image file required for BMD does not *Restart the server computer.

s.BoneMineralShape exit.

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

16020 ImageViewer.ImageView.Shape [16020] Failed in detecting the aluminum slope. *Confirm that the aluminum slope is being detected

s.BoneMineralShape normally.

*If the solid image subtraction is performed, set the

selection range again.

16021 ImageViewer.ImageView.Shape [16021] The solid image subtraction point has been *Confirm that the coordinate of solid image

s.BoneMineralShape set outside of the image area. subtraction is in the image frame.

16022 ImageViewer.ImageView.Shape [16022] The measurement result of BMD *Ensure that the measurement operations and

s.BoneMineralShape measurement has exceeded the standard range. images used are in accordance with the guideline.

*Use the images and perform the measurement in

accordance with the guideline. Contact your service

engineer if the same symptom recurs.

16023 ImageViewer.ImageView.Shape [16023] The aluminum slope detection ROI has been *Change the aluminum slope detection ROI in such a

s.BoneMineralShape set outside of the image area. way as to surround the aluminum slope.

3/28
[ECN Appendix 11] Error Message Table

16024 ImageViewer.ImageView.Shape [16024] The aluminum slope detection process was *Delete the displayed BMD measurement and newly

s.BoneMineralShape not performed. perform the BMD measurement.

16025 ImageViewer.ImageView.Shape [16025] Failed in automatically detecting the *Set the aluminum slope position manually.

s.BoneMineralShape aluminum slope.

16026 ImageViewer.ImageView.Shape [16026] The aluminum slope is not in the aluminum *Confirm that the aluminum slope is in the

s.BoneMineralShape slope area. appropriate position.

16027 ImageViewer.ImageView.Shape [16027] The center line of second metacarpal bone *Confirm that the center line position of the second

s.BoneMineralShape has been set outside of the image area. metacarpal bone has been set properly.

16028 ImageViewer.ImageView.Shape [16028] Failed in detecting the center of the second *Confirm that the center line position of the second

s.BoneMineralShape metacarpal bone. metacarpal bone has been set properly.

16029 ImageViewer.ImageView.Shape [16029] Measurement object data are not in the *Confirm that the ROI position and size of the
s.BoneMineralShape measurement ROI of the second metacarpal bone. second metacarpal bone measurement has been set

properly.

16030 ImageViewer.ImageView.Shape [16030] Failed in analyzing the measurement ROI of *Expand the measurement ROI size of the second

s.BoneMineralShape the second metacarpal bone. metacarpal bone horizontally or vertically.

16031 ImageViewer.ImageView.Shape [16031] Failed in BMD measurement. *Ensure that the measurement operations and

s.BoneMineralShape images used are in accordance with the guideline.

*Use the images and perform the measurement in

accordance with the guideline. Contact your service

engineer if the same symptom recurs.

16032 ImageViewer.ImageView.Shape [16032] The center line of the second metacarpal *Confirm that the center line of the second

s.BoneMineralShape bone is too short. Measurement processing was metacarpal bone has been set properly.

canceled.

16033 ImageViewer.ImageView.Shape [16033] The vertical length of measurement ROI of *Expand the ROI size of the second metacarpal bone

s.BoneMineralShape the second metacarpal bone is too short. measurement vertically.

4/28
[ECN Appendix 11] Error Message Table

Measurement processing was canceled.

16034 ImageViewer.ImageView.Shape [16034] The horizontal length of measurement ROI *Expand the ROI size of the second metacarpal bone

s.BoneMineralShape of the second metacarpal bone is too short. measurement horizontally.


Measurement processing was canceled.

16035 ImageViewer.ImageView.Shape [16035] Failed in generating the image for *Restart the server computer.

s.BoneMineralShape measurement.

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

16036 ImageViewer.ImageView.Shape [16036] BMD measurement cannot be performed on *Ensure that the images are in accordance with the

s.BoneMineralShape images other than FCR images. guideline.

16037 ImageViewer.ImageView.Shape [16037] Failed in saving files. *Confirm that the destination file is not locked.

s.BoneMineralShape

16038 ImageViewer.ImageView.Shape [16038] The aluminum slope is not in the aluminum *Confirm that the aluminum slope is in the

s.BoneMineralShape slope area. appropriate position.

16039 ImageViewer.ImageView.Shape [16039] The solid image subtraction point has been *Confirm that the coordinate of solid image

s.BoneMineralShape set outside of the image area. subtraction is in the image frame.

16040 ImageViewer.ImageView.Shape [16040] Failed in detecting the horizontal line of solid *Adjust the position of the solid image subtraction

s.BoneMineralShape image subtraction. point so that the exposure subject does not overlap

the horizontal and vertical line of the solid image

subtraction point.

16041 ImageViewer.ImageView.Shape [16041] Failed in detecting the vertical line of solid *Adjust the position of the solid image subtraction

s.BoneMineralShape image subtraction. point so that the exposure subject does not overlap

the horizontal and vertical line of the solid image

subtraction point.

5/28
[ECN Appendix 11] Error Message Table

16042 ImageViewer.ImageView.Shape [16042] Note that you were charged for using BMD *You will be charged at the time BMD measurement

s.BoneMineralShape measurement. is performed. (Up to one time a study)

16043 ImageViewer.ImageView.Shape [16043] You are charged according to the number of

s.BoneMineralShape times you use BMD measurement.

16044 GeneralLiblary.CustomMessag [16044] This operation can not be canceled. Do you want to continue processing?

eBox

16045 GeneralLiblary.CustomMessag [16045] Template file does not exist. Printing *Open the setup change window of simplified print

eBox process was canceled. and select a correct template file.

17000 GeneralLiblary.CustomMessag [17000] Failed in initializing process of work list. *Restart the server computer.

eBox

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

17006 GeneralLiblary.CustomMessag [17006] Failed in updating the work list. *Restart the server computer.

eBox

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

17009 GeneralLiblary.CustomMessag [17009] Failed in starting the study.No authority to *No patient registration authority.

eBox perform the operation.

17011 GeneralLiblary.CustomMessag [17011] Failed in starting the study.Exposure cannot *Exposure screen is starting up.

eBox be started.

*Please "Complete" or "Suspend" ongoing study,

and retry again.

17013 GeneralLiblary.CustomMessag [17013] Failed in starting the study.The exposure of *The displayed work list may not be the latest.

eBox the specified study has finished already. Refresh the work list.

6/28
[ECN Appendix 11] Error Message Table

17015 GeneralLiblary.CustomMessag [17015] Study canceled in the past will be started. *Ordering device and study information of the

eBox previously canceled study may not have been

synchronized.

*The process is continued. However, if synchronizing

is required, delete the study once and re-register the

study.

17017 GeneralLiblary.CustomMessag [17017] Failed in starting the study. Failed in *Restart the server computer.

eBox registering the patient information.

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

17019 GeneralLiblary.CustomMessag [17019] Failed in starting the study. Study or *Unregistered study/exposure menu codes may be

eBox exposure menu code could not be analyzed. specified.

*Ensure study/exposure menus specified from

ordering device are correct. Reissue the order or

register them manually.

*Contact your service engineer if the same symptom

recurs.

17024 GeneralLiblary.CustomMessag [17024] Failed in updating the work list. *Restart the server computer.

eBox

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

17028 GeneralLiblary.CustomMessag [17028] Failed in updating the work list. *Restart the server computer.

eBox

7/28
[ECN Appendix 11] Error Message Table

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

17033 GeneralLiblary.CustomMessag [17033] Failed in initializing process of work list. *Restart the server computer.

eBox

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

17036 GeneralLiblary.CustomMessag [17036] Failed in starting the study. Patient ID has *Ensure that the correct Patient ID of the

eBox not been specified. corresponding study has been specified from the

ordering device.

*Reissue the order specifying Patient ID.

*Contact your service engineer if the same symptom

recurs.

17038 GeneralLiblary.CustomMessag [17038] Patient ID entered includes characters *Following characters strings are not allowed to use

eBox prohibited to use. for Patient ID.

" # %% && ' + ? \ / * @ < blank only

*Banned characters are replaced with blank. Are you

sure?

17039 GeneralLiblary.CustomMessag [17039] Banned characters are used for patient *Following characters are not allowed to use for

eBox name. patient name.

" # %% && ' + ? \ / * @ < =

8/28
[ECN Appendix 11] Error Message Table

*Banned characters are replaced with blank.Are you

sure?

17040 GeneralLiblary.CustomMessag [17040] Banned characters are used for Kanji patient *Following characters are not allowed to use for

eBox name. Kanji patient name.

Single-byte characters

*Banned characters are replaced with blank. Are you

sure?

17041 GeneralLiblary.CustomMessag [17041] The Species contains illegal character. *Following characters are not allowed to use for
eBox species.

" %% && + ? \ * @ <

*Banned characters are replaced with blank. Are you

sure?

17042 GeneralLiblary.CustomMessag [17042] The Breed contains illegal character. *Following characters are not allowed to use for

eBox breeds. Input again

" %% && + ? \ * @ <

*Banned characters are replaced with blank. Are you

sure?

17043 GeneralLiblary.CustomMessag [17043] Banned characters are used for Kanji *Following characters are not allowed to use for

eBox Responsible person name. Responsible person name.

9/28
[ECN Appendix 11] Error Message Table

Single-byte characters

*Banned characters are replaced with blank. Are you


sure?

17044 GeneralLiblary.CustomMessag [17044] Banned characters are used for responsible *Following characters are not allowed to use for

eBox person name. Responsible person name.

" # %% && + ? \ * @ < =

*Banned characters are replaced with blank. Are you

sure?

17045 GeneralLiblary.CustomMessag [17045] Responsible organization contains invalid *Following characters are not allowed to use for

eBox character. Responsible organization name.

" %% && + ? \ * @ <

*Banned characters are replaced with blank. Are you

sure?

17046 GeneralLiblary.CustomMessag [17046] Failed in MPPS linkage. The setting *Study condition was not notified. The study

eBox necessary for communication could not be acquired. condition must be changed on other device.

*Setting may not have finished.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

10/28
[ECN Appendix 11] Error Message Table

restarted.

17048 GeneralLiblary.CustomMessag [17048] Failed in MPPS linkage. The necessary *Study condition was not notified. The study
eBox information could not be acquired from the database. condition must be changed on other device.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

17050 GeneralLiblary.CustomMessag [17050] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. The study

eBox the connection. condition must be changed on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

17052 GeneralLiblary.CustomMessag [17052] Failed in MPPS linkage. No response was *Study condition may not be notified. The study

eBox replied from other device. condition must be changed on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

11/28
[ECN Appendix 11] Error Message Table

17054 GeneralLiblary.CustomMessag [17054] Failed in MPPS linkage. Unexpected *Study condition may not be notified. The study

eBox response was received. condition must be changed on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

17056 GeneralLiblary.CustomMessag [17056] Failed in updating the study database in *Study condition was notified, but it must be

eBox MPPS linkage. changed on other device upon deletion of the

study/completion of the exposure.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

17058 GeneralLiblary.CustomMessag [17058] Failed in MPPS linkage. *Study condition may not be notified. The study

eBox condition must be changed on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

17060 GeneralLiblary.CustomMessag [17060] Patient ID entered includes characters *Following characters are not allowed to use for

eBox prohibited to use. Patient ID. Input again.

12/28
[ECN Appendix 11] Error Message Table

" # %% && ' + ? \ / * @ < space

17061 GeneralLiblary.CustomMessag [17061] Banned characters are used for Kanji patient *Following characters strings are not allowed to use

eBox name. for Kanji patient name. Input again

5 characters of space or more

17062 GeneralLiblary.CustomMessag [17062] Banned characters are used for patient *Following characters are not allowed to use for

eBox name. patient name. Input again

" # %% && ' + ? \ / * @ < = 5 characters of space

or more

17063 GeneralLiblary.CustomMessag [17063] Banned characters are used for RP ID. *Following characters are not allowed to use for RP

eBox ID. Input again

17064 GeneralLiblary.CustomMessag [17064] Banned characters are used for AE title. *Following characters are not allowed to use for AE

eBox name. Input again

\ Space

17065 GeneralLiblary.CustomMessag [17065] Failed in updating the work list. *Restart the server computer.

eBox

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

17102 GeneralLiblary.CustomMessag [17102] Number of search results is higher than the *Desired order may not be displayed on the list.

eBox maximum acquiring number of set work list.

Change search conditions and retry searching.

13/28
[ECN Appendix 11] Error Message Table

Or change the maximum acquiring number.

17104 GeneralLiblary.CustomMessag [17104] Display only the orders which have been *Desired order may not be displayed on the list.

eBox received because time out was occurred while the

worklists are being received. Change search conditions and retry searching.

17106 GeneralLiblary.CustomMessag [17106] Failed in receiving the work list. The setting *Setting may not be completed.

eBox necessary for communication could not be acquired.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

17108 GeneralLiblary.CustomMessag [17108] Failed in receiving the work list. The *Restart the server computer.

eBox necessary information could not be acquired from

the database. >Contact your service engineer if the same symptom

recurs even after it has been restarted.

17110 GeneralLiblary.CustomMessag [17110] Failed in receiving the work list. Failed in *After restarting the other device, ensure the device

eBox establishing the connection. has started properly.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

17112 GeneralLiblary.CustomMessag [17112] Failed in receiving the work list. *After restarting the other device, ensure the device

eBox has started properly.

*Restart the server computer.

14/28
[ECN Appendix 11] Error Message Table

Change search conditions and retry searching.

*Contact your service engineer if the same symptom


recurs even after it has been restarted.

17114 GeneralLiblary.CustomMessag [17114] Failed in receiving the work list. Unexpected *After restarting the other device, ensure the device

eBox response was received. has started properly.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

17116 GeneralLiblary.CustomMessag [17116] Failed in receiving the work list. *After restarting the other device, ensure the device

eBox has started properly.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

17118 PatientPortal [17118] Failed in reading received data in the work *After restarting the other device, ensure the device

list. has started properly.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

15/28
[ECN Appendix 11] Error Message Table

17123 PatientPortal [17123] The maximum digit number of Patient ID is

exceeded.

17124 StudyList [17124] Failed in MPPS linkage in some studies. *Study condition was not notified. The study

condition must be changed on other device.

*Restart and check the other device.

>Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20427 FCRWeb [20427] Studies with the same StudyInstanceUID


cannot be output at the same time. Please output

the studies one by one.

20553 FFCustomMsgBox [20553] Failed in MPPS linkage. *Study condition may not be notified. Confirm the

study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20554 FFCustomMsgBox [20554] Failed in MPPS linkage. The setting *Study condition was not notified. Change the

necessary for communication could not be acquired. condition on the other device side.

*Setting may not be completed.

16/28
[ECN Appendix 11] Error Message Table

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20555 FFCustomMsgBox [20555] Failed in MPPS linkage. The necessary *Study condition was not notified. Change the study
information could not be acquired from the database. condition of the study on the other device side.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

20556 FFCustomMsgBox [20556] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. Change the study

the connection. condition of the study on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20557 FFCustomMsgBox [20557] Failed in MPPS linkage. No response was *Study condition may not be notified. Confirm the

replied from other device. study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

17/28
[ECN Appendix 11] Error Message Table

20558 FFCustomMsgBox [20558] Failed in MPPS linkage. Unexpected *Study condition may not be notified. Confirm the

response was received. study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20559 FFCustomMsgBox [20559] Failed in updating the study database in *Notification of study condition was succeeded.

MPPS linkage.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

20569 FFCustomMsgBox [20569] Failed in MPPS linkage. The setting *Study condition was not notified. Change the

necessary for communication could not be acquired. condition on the other device side.

*Setting may not be completed.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20570 FFCustomMsgBox [20570] Failed in MPPS linkage. The necessary *Study condition was not notified. Change the study

information could not be acquired from the database. condition of the study on the other device side.

*Restart the server computer.

18/28
[ECN Appendix 11] Error Message Table

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

20571 FFCustomMsgBox [20571] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. Change the study
the connection. condition of the study on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20572 FFCustomMsgBox [20572] Failed in MPPS linkage. No response was *Study condition may not be notified. Confirm the

replied from other device. study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20573 FFCustomMsgBox [20573] Failed in MPPS linkage. Unexpected *Study condition may not be notified. Confirm the

response was received. study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

19/28
[ECN Appendix 11] Error Message Table

restarted.

20574 FFCustomMsgBox [20574] Failed in MPPS linkage. *Study condition may not be notified. Confirm the
study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

20575 FFCustomMsgBox [20575] Failed in updating the study database in *Notification of study condition was succeeded.

MPPS linkage.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

20626 GeneralLiblary.CustomMessag [20626] Failed in launching the ASSISTA Portal Site. *Setting information was not found or was in error.

eBox

*Contact your service engineer.

20742 FCRWeb [20742] No NAS space necessary for outputting *Add a NAS and retry.

data.

*Contact the maintenance service immediately.

20/28
[ECN Appendix 11] Error Message Table

20743 FCRWeb [20743] No NAS for outputting data was found. *Please check NAS power and network connections

and set up again.

>If fails to output to/import from NAS upon


disconnection, restart the server.

*Add a NAS and retry.

>Contact the maintenance service immediately.

20744 FCRWeb [20744] Failed in acquiring a setting information. >Contact your service engineer.

20745 FCRWeb [20745] Failed in registration. *Request on the server computer failed. Check the
startup status of the server computer. If it fails many

times, restart the server computer.

>Contact your service engineer if the same symptom

recurs even after it has been restarted.

20767 FFCustomMsgBox [20767] The study information was not registered *After restarting the server computer, delete the

properly. registered study once and register the study again.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

30807 FCRWeb [30807] Failed in MPPS linkage. The setting *Study condition was not notified. Change the

necessary for communication could not be acquired. condition on the other device side.

*Setting may not be completed.

21/28
[ECN Appendix 11] Error Message Table

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

30808 FCRWeb [30808] Failed in MPPS linkage. The necessary *Study condition was not notified. Change the study
information could not be acquired from the database. condition of the study on the other device side.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

30809 FCRWeb [30809] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. Change the study

the connection. condition of the study on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

30810 FCRWeb [30810] Failed in MPPS linkage. No response was *Study condition may not be notified. Confirm the

replied from other device. study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

22/28
[ECN Appendix 11] Error Message Table

30811 FCRWeb [30811] Failed in MPPS linkage. Unexpected *Study condition may not be notified. Confirm the

response was received. study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

30812 FCRWeb [30812] Failed in MPPS linkage. *Study condition may not be notified. Confirm the

study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

30813 FCRWeb [30813] Failed in MPPS linkage. The setting *Study condition was not notified. Change the

necessary for communication could not be acquired. condition on the other device side.

*Setting may not be completed.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

23/28
[ECN Appendix 11] Error Message Table

30814 FCRWeb [30814] Failed in MPPS linkage. The necessary *Study condition was not notified. Change the study

information could not be acquired from the database. condition of the study on the other device side.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

30815 FCRWeb [30815] Failed in MPPS linkage. Failed in establishing *Study condition was not notified. Change the study

the connection. condition of the study on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

30816 FCRWeb [30816] Failed in MPPS linkage. No response was *Study condition may not be notified. Confirm the

replied from other device. study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

30817 FCRWeb [30817] Failed in MPPS linkage. Unexpected *Study condition may not be notified. Confirm the

response was received. study condition on other device.

24/28
[ECN Appendix 11] Error Message Table

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been


restarted.

30818 FCRWeb [30818] Failed in updating the study database in *Notification of study condition was succeeded.

MPPS linkage.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

30819 FCRWeb [30819] Failed in MPPS linkage. *Study condition may not be notified. Confirm the

study condition on other device.

*Restart and check the other device.

*Contact your service engineer if the same symptom

recurs even after the server computer has been

restarted.

30820 FCRWeb [30820] Failed in updating the study database in *Notification of study condition was succeeded.

MPPS linkage.

*Restart the server computer.

*Contact your service engineer if the same symptom

recurs even after it has been restarted.

30917 ImportImage [30917] Failed in deleting files. *The original file being imported cannot be deleted.

25/28
[ECN Appendix 11] Error Message Table

30918 ImportImage [30918] A failure occurred in operation of Complete *Perform the operation again.

button.

34944 ImportImage [30944] Already being imported. *Operation cannot be performed because the target

study is being imported.

34945 ImportImage [30945] Complete the study in a receiving status. +Are you sure?

34946 ImportImage [30946] Checking condition and rotation of the image

will be cleared.

34947 ImportImage [30947] The original image being imported will be +Are you sure?

deleted.

34948 ImportImage [34948] There are images which could not be

ImageRetrieveRequest imported. Please select imported studies from the


patient list and confirm the study images.

34952 StudyTableRegist [34952] An error occurred in auto importing process *An error occurred in receiving the study in the

of the endoscopic image. endoscopic image auto import service.

*Import from the study import window of endoscopic

image.

34953 ImageRetrieveRequest [34953] An error occurred in auto importing process *An error occurred in importing the image in the

of the endoscopic image. endoscopic image auto import service.

*Import from the study import window of endoscopic

image.

40500 (イベントログ出力無し) [40500] A fatal error occurred. > Press the button below to close the dialog, and

then restart the same study from "Refer" tab page.

40501 (イベントログ出力無し) [40501] Server computer cannot be accessed. *Check the connection of network cable and hub.

Process time out was occurred.

26/28
[ECN Appendix 11] Error Message Table

> Press the button below to close the dialog, and

then restart the same study from "Refer" tab page.

40502 (イベントログ出力無し) [40502] Failed in gaining access to the server *Check the connection of network cable and hub.

computer. Timeout occurred while processing.

> Press the button below to close the dialog, and

then restart the same study from "Refer" tab page.

40503 (イベントログ出力無し) [40503] Failed in gaining access to the server *Check the connection of network cable and hub.

computer. Timeout occurred while processing.

> Press the button below to close the dialog, and

then restart the same study from "Refer" tab page.

40504 (イベントログ出力無し) [40504] Failed in gaining access to the server *Check the connection of network cable and hub.

computer. Timeout occurred while processing.

> Press the button below to close the dialog, and

then restart the same study from "Refer" tab page.

40505 (イベントログ出力無し) [40505] Failed in gaining access to the server *Check the connection of network cable and hub.

computer. Timeout occurred while processing.

> Press the button below to close the dialog, and

then restart the same study from "Refer" tab page.

40506 (イベントログ出力無し) [40506] Failed in gaining access to the server *Check the connection of network cable and hub.

computer. Timeout occurred while processing.

> Press the button below to close the dialog, and

then restart the same study from "Refer" tab page.

69601 FFServiceUtility [69601] Connected device name is duplicating. *Return to the setup window and re-input correctly.

69602 FFServiceUtility [69602] Delete the setting of selected connected +Are you sure?

device.

69603 FFServiceUtility [69603] The same receive folder path already exists. *Return to the setup window and re-input correctly.

27/28
[ECN Appendix 11] Error Message Table

69604 FFServiceUtility [69604] Specified receive folder path does not exist. *Check that the receive folder of connected device

exists.

69605 FFServiceUtility [69605] The same path for the receive folder and *Return to the setup window and re-input correctly.

temporary folder cannot be set.

69606 FFServiceUtility [69606] Set the receive folder path within 80 *Return to the setup window and re-input correctly.

characters.