Extech EC150 Manual
Extech EC150 Manual
Conductivity Meter
Pen Style Water Quality Meter
Model EC100
Introduction
Congratulations on your purchase of the Extech Pen Style Water Quality instrument; the Model
EC100 measures Conductivity and Temperature. The instrument is housed in an IP65 Water-proof
enclosure for safety. This instrument is shipped fully tested and calibrated and, with proper use, will
provide years of reliable service.
Features
Automatic Ranging and Manual Ranging capability
Dual Display with ATC (automatic temperature control)
Data hold for freezing displayed readings
Low battery indicator
Automatic power-off for maximum battery efficiency
Switchable temperature units of measure (C/F)
Multi-point and one-touch calibration features
Powered by four (4) LR44 batteries
Supplied Materials
EC100 meter
Four (4) LR44 button batteries
Operation manual (hard copy, mini-disk, and on-line availability at www.extech.com)
Battery Installation
The meter is shipped with the four (4) LR44 button batteries
removed. The user must install the batteries before the meter
can be used. Refer to accompanying diagram.
1. Unscrew the battery compartment cover (top of
meter) in a counter-clockwise direction. Please do
not discard the black washer.
2. Install the four (4) LR44 button batteries, carefully
orienting the batteries and observing polarity.
3. Replace the battery compartment cover.
4. Please remove the batteries while the meter is not in
use for long periods.
Display Description
1. Data Hold icon
2. Micro- and milli-Siemens units
3. Primary measurement reading
4. Temperature reading
5. Temperature units of measure
6. Battery strength indicator
7. Calibration icon
Keypad Description
1. Power ON-OFF and SET button
2. Down Arrow, Data Hold, and Calibration button
3. Up arrow
Conductivity Ranges
Range 1 0 to 1999 µS
Range 2 0 to 19.99 mS
Setup Mode
Parameter P1: Temperature Units, Ambient Temperature, and Temperature Coefficient settings
1. From the normal operating mode, press and hold the SET button for at least 2 seconds until the
‘Px’ icon appears on the meter display (x = setup parameter number).
2. Use the arrow buttons to scroll to the P1 icon if necessary.
3. Press the SET button momentarily, the ‘C’ or ‘F’ icon should now be flashing and the ‘t.ut’ icon
(abbreviation for temperature units) will be visible above the flashing unit.
4. Use the arrow keys to select the desired unit of measure.
5. Press the SET button momentarily to confirm the selection.
6. A temperature value will be flashing at the bottom of the display and the ‘t.nr’ icon
(normalization temperature i.e., reference temperature) will be visible above the flashing
temperature. See Appendix C (Temperature Effects) for more information on reference
temperature.
7. Use the up arrow button to toggle between 68°F and 77°F (the default setting is 77°F).
8. Press the SET button to confirm the setting.
9. The temperature coefficient value should now be flashing on the bottom of the display with the
‘t.Co’ icon visible above it. Refer to Appendix C for details on Temperature Coefficient.
10. Use the arrow buttons to select the coefficient temperature (default is 23.1°C).
11. Press the SET button momentarily to confirm the selection.
12. The meter display should return to the P1 level starting point.
13. Press and hold the SET button for at least 2 seconds to return to the normal operation mode or
use the arrow buttons to move to Parameter P3 (see below).
Parameter P3: Meter Reset
This parameter can be used to restore all settings to their factory default state.
1. If continuing from Parameter P1 skip directly to step 2 below. If starting from the normal
operating mode, press and hold the SET button for at least 2 seconds until the ‘Px’ icon appears
on the meter display (x = setup parameter number).
2. Use the arrow buttons to scroll to the P3 icon if necessary. The ‘rSt’ display icon will be visible
above the P3 icon.
3. Press the SET button momentarily; a ‘y’ or an ‘n’ will be flashing.
4. Use the arrow buttons to select ‘y’ for YES RESET or ‘n’ for NO RESET.
5. Press the SET button momentarily to confirm the setting.
6. Press and hold the SET button for at least 2 seconds to return to the normal operation mode or
use the arrow buttons to move to Parameter P4 (see below).
All EU users are legally bound by the battery ordinance to return all used batteries to
collection points in your community or wherever batteries / accumulators are sold!
Disposal in the household garbage is prohibited!
Calibration Schedule
Calibration is necessary and should be performed regularly.
If measuring in the mid-ranges calibrate the meter at least once per month and soak the probe
for 15 minutes before each use.
If measuring in extreme temperature environments, or at the low end of the measurement
range, calibrate the meter at least once per week.
Troubleshooting
Power ON is attempted but there is no display
Be sure to press the ON-OFF but for at least 100mS to switch the meter ON.
Check that batteries are positioned correctly, making good contact, and follow correct polarity.
Replace the batteries if necessary.
Remove and replace the existing batteries.
Display switches OFF
This is normal when Auto Power OFF is activated.
Replace the batteries if necessary.
Air Bubbles adhered to Electrode
Stir the electrode completely and be sure to dip the electrode into a solution at an oblique angle.
Vertical dipping can cause many air bubbles to adhere.
Gently tap the bottom of the solution container while stirring the electrode in the solution.
Air can be blown across the electrode before dipping it into the solution.
Error Codes
Refer to the Table below for details on Error Codes displayed by the meter.
*442: 40% sodium sulfate, 40% sodium bicarbonate, and 20% chloride
Where:
KTref = Conductivity at Tref
KT = Conductivity at T
Tref = Reference temperature
T = Sample temperature
= Temperature coefficient
Note: The correction is accurate only within a limited temperature range near T1 and T2; the greater the
difference between T and Tref, the higher the risk of error.
T2 should be selected as a typical sample temperature and should be approximately 10°C different
from T1. The temperature coefficients of the following electrolytes generally fall into the ranges
shown below:
Acids: 1.0 – 1.6%/°C
Bases: 1.8 – 2.2%/°C
Salts: 2.2 – 3.0%/°C
Drinking water: 2.0%/°C
Ultrapure water: 5.2%/°C
Copyright © 2013 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in any form
ISO‐9001 Certified
www.extech.com
11 EC100-EU-EN v2.2 06/13