HEYER Pasithec - Upgraded Version - User Manual Rev 2.3 EN 0716
HEYER Pasithec - Upgraded Version - User Manual Rev 2.3 EN 0716
3 – 07/16
1
PASITHEC User Manual
2
PASITHEC User Manual
Contents
Contents .................................................................................................................................................... 3
Statement .................................................................................................................................................. 7
1. Product Description ............................................................................................................................... 9
1.1. General Information ........................................................................................................................... 9
1.2. Symbols........................................................................................................................................... 9
1.3. Abbreviations ............................................................................................................................... 12
1.4. Warnings and Cautions ................................................................................................................ 15
2. Components ........................................................................................................................................ 21
2.1. Configuration ................................................................................................................................... 21
2.5. Breathing System.......................................................................................................................... 27
2.5.1. APL Valve ............................................................................................................................... 29
2.7. Anesthetic Gas Scavenging System-AP1000 (Optional) ............................................................... 31
2.8. SUCTION(Optional) ................................................................................................................. 32
2.9. Accessories ................................................................................................................................... 33
3. User interface ...................................................................................................................................... 34
3.1. Main interface display layout ....................................................................................................... 36
3.1.3. The Interface layout with CO2 and anesthesia gas monitoring ............................................. 39
3.2.1. Ventilation mode set and display area .................................................................................. 39
3.2.2. Parameter set and display area............................................................................................. 40
3.2.4. Machine information area..................................................................................................... 43
3.2.5. Waveform display area ......................................................................................................... 44
3.2.9. Flowmeter area ..................................................................................................................... 47
3.2.11. Shortcut key area ................................................................................................................ 48
Name description and Diagram....................................................................................................... 50
4. Pre-operative Preparation ................................................................................................................... 51
4.1. Installation and Connection ......................................................................................................... 51
4.1.1. Connection of Tubes and Cables ........................................................................................... 51
4.1.2. Installation of Breathing Bag ................................................................................................. 53
4.1.3. Connection of test plug ......................................................................................................... 54
4.1.4. Connection of O2 sensor ........................................................................................................ 55
3
PASITHEC User Manual
4
PASITHEC User Manual
5
PASITHEC User Manual
6
PASITHEC User Manual
Statement
HEYER Medical AG holds the copy rights to this manual, which is non-public, published, and reserves
the rights to keep it as a secure document. Refer to this manual when operating, maintaining and
repairing products only. Anyone other than HEYER may not make it known to others.
Proprietary materials protected by the copyright law are included in this manual. Any section of it
cannot be reproduced, copied, or translated into other languages without any prior written approval
from HEYER who reserves the copyright.
Everything written in this manual is considered to be correct. HEYER is not legally responsible for any
mistakes printed within and any damages caused by incorrect installation and operation. HEYER does
not supply privileges endowed by the patent law to any other parties. HEYER is not legally responsible
for the results caused by patent law breaking and any rights of the third party violating.
Refer to this manual before the product is used. The manual includes operating procedures which must
be performed with cautiously, operations that may result in non-normal working conditions and the
dangers which may damage equipment or cause bodily harm. HEYER is not responsible for the security,
reliability and function of the equipments in case that the dangers, damages and non-normal
phenomenon mentioned in this manual happen. Free repairs for these malfunctions will not be
provided by HEYER.
HEYER has the rights to replace any content in this manual without notice.
CAUTION: Malfunctioning equipment may become invalid and cause bodily injury if a set of
effective and approving repairing proposals cannot be submitted by the institution which is
responsible for using this equipment.
The paid theoretical framework diagram will be supplied according to customer requirements by HEYER,
plus calibrating method and other information to help the customer, under the assistance of qualified
technicians, repair the equipment parts where can be done by customer himself based on the
stipulation by HEYER.
7
PASITHEC User Manual
Use Notice
To use this product correctly and effectively, the user must read this user manual carefully before using
this product. When the user is using this product, this user manual must be understood totally and
observe strictly.
This product is only applied to the use this user manual described.
Only trained and professional service persons are allowed to repair and examine this product. If there
is any case happened in the process of use, please call us and intense service will be provided.
If there is any change in product specification, another informs will not be given.
8
PASITHEC User Manual
1. Product Description
HEYER Pasithec Anesthesia Machine is a continuous flow anesthesia system which offers manual or
automatic ventilation, easily adjustable fresh gas delivery, anesthetic agent delivery, ventilation
monitoring, convenient ergonomics, and state-of-the-art safety systems.
In terms of theory of operation, HEYER Pasithec Anesthesia Machine, driven by pneumatics and
controlled by electricity, is a device used to administer to a patient continuously or intermittently, a
general inhalation anesthetic and to maintain a patient’s ventilation. The integrated breathing system
delivers the mixture of O2, N2O, AIR and anesthetic agents for patient breathing control, while electrical
control system functions as monitoring patient parameters.
HEYER Pasithec Anesthesia Machine is intended for use by licensed clinicians, for patients requiring
anesthesia within a hospital. This product is suitable for infants, children and adults.
1.2. Symbols
The following table provides descriptions of symbols that are used on the device and/or within this
manual.
9
PASITHEC User Manual
dangerous voltage
Earth (ground)
equipotentiality
date of manufacture
manufacturer
Catalogue number
serial number
10
PASITHEC User Manual
lock
unlock
Insp. flow
Exp. flow
O2 flush O2+
fuse
Recycle
11
PASITHEC User Manual
This mark means that pneumatic and electric device rubbish is not
allowed to be dealt as unclassified city rubbish and should be
collected separately.
alarm icon
1.3. Abbreviations
Abbreviation Definition
A
AGSS anesthetic gas scavenging system
APL adjustable pressure limiting
B
BDU basic data unit
C
C compliance
CGO common gas outlet
12
PASITHEC User Manual
M
MAC Minimum Alveolar Concentration
Manual manual ventilation
PMEAN mean pressure
13
PASITHEC User Manual
mL Mill Liter
MRI Magnetic resonance imaging
MV minute volume
N
N2O nitro oxide
O
O2 oxygen
P
Paw airway pressure
Paw-t pressure-time wave
Pb plumbum
PIP Peak Inspiratory Pressure
PLAT plat pressure
PCV Pressure Control Ventilation
PEEP Positive End Expiratory Pressure
PLIMIT limit pressure
PMEAN mean pressure
PPEAK Peak pressure
PS Pressure Support Ventilation
PTARGET Target Pressure
S
SIMV Synchronized Intermittent Mandatory Ventilation
Sev. Sevoflurane
T
TINSP Inspiratory Time
TP Inspiratory Pause Time
Trigger flow trigger
TSLOPE Inspiratory Slope Time
V
VT tidal volume
W
WDT watch dog timer
14
PASITHEC User Manual
ΔP Differential Pressure
Please read and adhere to all warnings, CAUTIONs and NOTICEs listed here and in the appropriate areas
throughout this manual.
A WARNING statement gives important information that, if ignored, could lead directly to personal
injury.
A CAUTION statement gives important information that, if ignored, could lead directly to equipment
damage and indirectly to personal injury.
WARNING : HEYER Pasithec Anesthesia Machine must and can only be serviced by HEYER’s
WARNING:For a full understanding of the performance characteristics of this equipment, the user
15
PASITHEC User Manual
WARNING :Do not use the machine if any test failures. Contact HEYER’s authorized service
HEYER Pasithec is not suitable for use in a magnetic resonance imaging (MRI)environment.
WARNING:To avoid the risk of electric shock, this equipment must only be connected to a supply
WARNING :The pipeline inlet pressure of HEYER Pasithec anesthesia machine must be 280-
600kPa.
WARNING:Breathing tube, mask, SUCTION filter and so on, adapter of mainstream CO2 module,
water trap of sidestream CO2 module etc. are disposable accessories. Before using check the pipeline
to ensure the disposal subassembly is out of using .Check the tube which will be used before use and
avoid reusing.
WARNING:When the suction flow is insufficient, the suction port filter should be checked or
replaced.
WARNING:Do not use anti-static and/or electric breathing tubes or masks; otherwise, they will be
16
PASITHEC User Manual
WARNING:Do not incline the machine at the angle of more than 10 degree.
WARNING:Remove all equipments on the top cover before moving the machine. Use the handle
WARNING:Ensure that the machine does not incline when it is upgrading, downgrading, turning
around or getting across a threshold. Do not push the machine across pipelines, lines or any barrier on
the ground.
Remove all the equipments attached at the flank of the machine before
WARNING: Do not put the item over 25 kg on the top board of HEYER Pasithec anesthesia machine.
WARNING: Do not put the item over 12 kg on the work table of HEYER Pasithec anesthesia machine.
WARNING: The whole width of machine should be less than 0.8m when hanging some stuff.
WARNING:In order to protect patients and users from electrical hazards, it is imperative that all
systems consisting of electrical medical devices and other electrical devices, such as but not limited to
17
PASITHEC User Manual
WARNING: The medical gas pipeline system malfunction could cause one or more devices
connected to the system stopping their operation; this is not applicable to the anesthetic machine
WARNING: The parts of machime in contact with gas to be inhaled by patients don’t contain
WARNING:Possible explosion hazard. Do not operate machine near flammable anesthetic agents
or other flammable substances. Do not use flammable anesthetic agents (i.e. either or cyclopropane).
isoflurane, which are compatible to other gas are allowed to use in anesthetic system.
WARNING:Possible fire hazard. Fuses (i.e., additional sockets) must only be replaced by fuses of
Possible electric shock hazard. The machine may only be opened by HEYER
WARNING:Electric shock and fire hazard. Do not clean the machine while it is on and /or plugged
in.
WARNING:Disconnect the power plug from the mains supply before removing the rear panels or
18
PASITHEC User Manual
WARNING:Malfunction of the central gas supply system may cause more than one or even all
WARNING:Use a cleaning and sterilizing schedule that confirms to your institution’s sterilization
• A damaged oxygen sensor may lead to leakage of its highly caustic contents that contain
potassium chloride. Wear safety gloves and safety googles per manufacturer’s
recommendations.
• Do not inhale fumes that may result from any sterilization process.
WARNING:Use care in lifting and manipulating vaporizers during the mounting process as their
weight may be greater than expected, based on their size and shape.
WARNING:All gas supplies should be of medical grade. To avoid endangering a patient, do not
perform testing or maintenance when the machine is in use.
WARNING:All gas volume, flow and leakage specifications have been tested under STPD, except
those associated with the Anaesthetic Breathing System, which be tested under BTPS.
WARNING:After the device replacement is finished, all calibrations must be done again and only
19
PASITHEC User Manual
WARNING: If the anesthetic machine is equipped with Anesthetic Gas Delivery System,
Monitoring Equipment, Alarm System and Production Device, no matter the machine is supplied by
separate device or system assembly, the anesthetic machine checklist must be provided.
WARNING: The medical device connected to the anesthesia system must conform to IEC
CAUTION:For your safety and that of your patients, strictly follow this user manual.
CAUTION:Perform the tests specified on the Pre-use Preparation and in case of a fault, do not
CAUTION:Before starting the machine, users must be familiar with the information contained in
CAUTION:If the machine does not function as described, it must be examined and repaired as
CAUTION : Ensure that the gas supply of the machine always complies with the technical
specifications.
CAUTION:Before clinical use, the machine must be correctly calibrated and the respective tests
CAUTION: After servicing, all calibrations must be performed again and only the professional service
20
PASITHEC User Manual
CAUTION: The anesthesia ventilator used together with anesthesia system conforms to ISO 80601-
2-13:2011.
CAUTION: The anesthesia system is used together with the following monitoring device, alarm
2. Components
2.1. Configuration
Touch screen
E-Flowmeter
2 drawers
21
PASITHEC User Manual
Heater function
Battery
Vaporizer
Optional
External lamp
configuration
YOKE
22
PASITHEC User Manual
Counterclockwise rotation of
flowmeter rotary knob increases the
flow; clockwise rotation of
flowmeter rotary knob decreases the
flow.
When flow meter is connected
Electrical with gas supply, rotate O2 or Air
4
flowmeter
knob, the digital screen displays
data; there is O2 or Air flowing
out.
23
PASITHEC User Manual
increase of N 2O flow is
controlled by O 2 flow, the
increase ratio of O 2 and N 2O is
approximate to 1:3.
It locates on the front of the
machine. It can provide oxygen
from 35L/min to 75 L/min to
6 O2 flush breathing system and breathing
bag.
CAUTION:When central gas supply or
backup cylinder is connected, pressing
O2 flush will deliver oxygen to breathing
system and breathing bag, even if the
machine is powered off.
WARNING:When adjusting
24
PASITHEC User Manual
Auxiliary O2 output
3
port WARNING: The requirement flow of the
device connected to the auxiliary drive gas
outlet interface cannot exceed 90L/min
25
PASITHEC User Manual
F2AL 250VAC,
F10AL 250VAC.
26
PASITHEC User Manual
10 RS-232 The RS-232 interface can be used to transmit patient’s data to other
interface device.
CAUTION:Breathing system used with HEYER Pasithec complies with ISO 80601-2-13:2011:
2011.
Breathing system is mainly used to store fresh gas including anesthetic gas, oxygen, and absorb waste
gas. It directly connects to airway to support patient’s respiration.
27
PASITHEC User Manual
8 CO2
Absorber
It is used to absorb CO2 patient having exhaled.
Canisters
28
PASITHEC User Manual
NOTE:
Inspiratory and expiratory valve guarantee the single direction of gas flow in the breathing system.
When the anesthesia machine is working normally:
In the inspiratory phase, the inspiratory valve is acting and the expiratory valve is at rest; In the
expiratory phase, expiratory valve is acting and the inspiratory valve is at rest.
2. Applied Part of HEYER Pasithec anesthesia machine is the part between mask and Y-piece.
WARNING:Keep all inlets/cables away from the APL valve, and do not put lines/ cables underneath
29
PASITHEC User Manual
CAUTION : The APL valve is automatically excluded from the breathing system whenever an
APL valve has two functions; one is that it limits the maximum pressure during manual ventilation; the
other is that the airway pressure can be exhausted quickly by lifting the APL valve.
The APL valve has a labeled knob for selecting between 0cmH2O to 70cmH2O and for indicating
approximate pressure settings. Click sound will be heard when adjusting the APL valve. In manual
mode, the APL valve knob can be rotated to change the pressure threshold at which gas will flow
through the pressure threshold at which gas will flow through the valve and into AGSS. Clockwise
rotation of the APL valve knob increases the pressure threshold, and counterclockwise rotation of the
APL valve knob decreases the pressure threshold. Lifting the top of the APL valve knob will temporarily
relieve pressure. The leakage from the anesthetic breathing in all operational modes when tested at
pressure of 3.0kPa (30cmH2O) is 40ml.
2.6. Vaporizer
The vaporizer delivers anesthetic agent of accurate concentration to breathing system. Anesthetic
agent of accurate concentration will be acquired by adjusting control knob on the vaporizer. It has
temperature compensation, flow compensation and pressure compensation. The vaporizer cannot be
used for patient with respiration because of the strong interior resistance. The vaporizer can only use
one specific anesthetic agent which can be identified by the label.
30
PASITHEC User Manual
WARNING:Only dry gas of medical degree can be used on the vaporizer. Any improper operation
CAUTION:The vaporizer used with HEYER Pasithec Anesthesia Machine complies with ISO 80601-
2-13:2011.
CAUTION:For more operating and maintaining information of the vaporizer, refer to User Manual.
Anesthetic gas scavenging system absorbs waste gas exhausted from breathing system and delivers it
to processing system.
31
PASITHEC User Manual
WARNING: Do not block the pressure relieve hole on top of the gas tank when using AGSS.
CAUTION:The anesthetic gas scavenging system used with HEYER Pasithec Anesthetic Machine
CAUTION: If the AGSS used together with anesthesia system is produced by other manufacturer,
CAUTION:For more operating and maintaining information of AGSS, refer to AP1000 User Manual.
2.8. SUCTION(Optional)
An optional suction device is available to remove secretions, blood or other semi-liquid fluid from the
patient’s respiratory tract.
Before performing suction, adjust the negative pressure on the Suction device to an appropriate value.
When using the Suction device, connect the gas inlet to the HEYER Pasithec auxiliary gas supply port.
If the Suction device is connected with other gas supplies, the pressure of gas supply should be less
than 0.4MPa.
32
PASITHEC User Manual
CAUTION: The negative pressure suction device used together with anesthesia system conforms to
EN ISO 10079-3.
CAUTION: For more information about operation and maintenance of the negative pressure
suction device, please refer to the Instruction for Use provided with the negative pressure suction
device.
2.9. Accessories
33
PASITHEC User Manual
3. User interface
WARNING: The anesthesia breathing machine used together with anesthesia system conforms to
ISO 80601-2-13:2011.
The panel of the anesthetic machine is composed of user interface, alternating current indicator light,
and function button and knob. See figure 3-1
34
PASITHEC User Manual
No Name Description
1 Touch screen Touch screen.
The user interface can display real-time pressure and flow waveforms,
respiratory loop, monitoring data, Electronic Flowmeter settings and
anesthesia System and provides both sound and visual alarms to assist
the user.
2 MANUAL/ In Standby mode, first press MANUAL/AUTO knob, the system switch to
AUTO Manual mode, then secondly switch to IPPV mode.
In automatic mode, first press MANUAL/AUTO knob, the system switch
to Manual mode, then secondly switch to the original mode. When
entering or exiting the Manual mode, the user information area displays
according to the original model.
3 MUTE If there is an alarm, press MUTE and the silence will keep for 110
seconds; press the key again, and the silence is cancelled. If the alarm
has a change but the silence less than 120s, the silence will be cancelled.
4 Knob The user can set the parameter values by rotating the knob and confirm
set values by pressing the knob. Rotating the knob clockwise increases
values and rotating the knob counter-clockwise decreases values.
CAUTION: The required circumstance of the monitoring of the breathing system as follows:
Circumstance temperature: 25 ℃;
Gas temperature: 25 ℃;
O2.
CAUTION: The anesthesia ventilator which is used with the anesthetic system complies with ISO
80601-2-13:2011.
35
PASITHEC User Manual
The user interface is divided into eleven areas, see Figure 3-2.
Because of the gas module configuration and the system setting, there are 3 layouts of the main
interface.
36
PASITHEC User Manual
37
PASITHEC User Manual
38
PASITHEC User Manual
3.1.3. The Interface layout with CO2 and anesthesia gas monitoring
Name description
Ventilation mode display: ventilation mode area shows current ventilation mode. The mode setup menu
will be open by clicking the ventilation mode area via touch screen.
3.2.1.2 During mode setting process, there are five states for mode button as shown below:
Name description Diagram
39
PASITHEC User Manual
Focusing state - when rotating the knob to the purpose mode, the corresponding
button will display this state.
White focus state - this state has been selected and then rotate knob through
the button.
Parameter setting area is located at the bottom of the screen, with green/blue background and round
white buttons, which are applied for real-time respiratory parameter adjustment, including six setting
buttons.
40
PASITHEC User Manual
The ventilation parameters of all models are arranged from left to right, as shown in the chart below:
Model Parameter Parameter Parameter Parameter Parameter Parameter 6
1 2 3 4 5
PCV-VG
VT Freq. I:E Tslope PEEP Pmax
(Optional)
ALARM
Manual on/off Blank Blank Blank Blank Blank
41
PASITHEC User Manual
2. The second status displays when the button has been selected by the
rotation of the knob;
3. The third status displays when the button has been pressed.
Alarm indication: when an alarm appears, an alarm bell icon shows in the warning area. For a senior-
level alarm, the alarm bell is red with white lettering. An intermediate-level alarm is yellow with black
lettering. A low-level alarm is blue-green with black lettering. At most, four alarms can display on the
screen at one time
By pressing MUTE button, an “X” shows on the alarm icon. Below the alarm bell a countdown timer
appears. The “X” and countdown second display the same color as the alarm bell.
The flickering frequency of higher priority alarm among all alarms indicates all alarms, alarm bell and
countdown seconds (if any). The alarm information text keeps own color respectively (if any).
42
PASITHEC User Manual
Power status includes AC power and backup battery state, which will be always
shown. When AC power is normal, the white plug icon shows consistently
without flashing. When net power is disconnected or fails, the white plug icon
disappears.
43
PASITHEC User Manual
CAUTION: Date format and value can be set in system setup menu.
Current time indicates at the time display area, either 12 hour or 24 hour format.
WARNING: Time format and value can be set in system setup menu.
Waveform display area is located in the center of the screen, including waveform and breathing loop.
Waveform name Diagram
44
PASITHEC User Manual
(CO 2-t)
Breathing Loops
The patient message area is to the right of the alarm zone, and to
the left of the information area. In this area, there are two icons:
patient type and trigger type.
45
PASITHEC User Manual
When patient triggers, the trigger logo will show in the patient
information area, and disappear after 250ms. The background color will
restore the original color.
46
PASITHEC User Manual
The electronic flow meter displays real-time N2O, air and O2 flow values.
Every flowmeter has 2 virtual flow tubes that display flow. Each flow tube
has the same icon, scale, structure and color (grey). The units and
numerical readings are displayed in white. The fill color of virtual flow
tubes is the same with gas icon color.
47
PASITHEC User Manual
Message prompt area is in the upper left corner of the parameter settings area. Message prompt area is
used to display critical information or to provide real-time help to the user.
The shortcut key area includes shortcuts to alarm settings, system settings, calibration, trend, breathing
loop, self test, configuration and main interface.
Alarm settings: when user presses alarm setting key, alarm menu appears. If
this button is pressed in open state, alarm setting menu closes, and returns
to main interface.
48
PASITHEC User Manual
System settings: by pressing system setup key, the system setup menu
appears. If this menu is pressed in open state, system settings menu
disappears, and returns to main interface.
Trend: by pressing trend key, the trend menu appears. If this button is pressed
in open state, trend menu disappears and returns to main interface.
Loops: by pressing Loops key, the loops menu appears. If this button is
pressed in open state, the functional menu disappears and returns to
main interface.
49
PASITHEC User Manual
The background of the overall menu is gray with the menu title shown centered, at the top of the
window. Below the menu title there are a number of tabs containing a variety of content within the
menu. In the top, right corner of the window is an "x" button which closes the menu when pressed.
The menu elements include option key, button, label, indicative text and pictures, etc.
50
PASITHEC User Manual
4. Pre-operative Preparation
Fig.4-1
Second step:
Connect one end of heating cable with the
interface below the breathing system; the other
end connected with interface marked “Heater”
in the outlet module.
Fig.4-2
Third step:
Connect one end of drive gas tube with the
interface below the breathing system.
Fig.4-3
51
PASITHEC User Manual
Fourth step:
Connect the other end of drive gas tube with
interface marked “Ventilator” in the outlet
module.
Fig.4-4
52
PASITHEC User Manual
Step 1:
Hold breathing bag to bag arm directly.
Fig.4-5
Second step:
Connect manual bag to manual bag arm with an
upper force.
Fig.4-6
53
PASITHEC User Manual
Fig.4-7
Second step:
Connect Y-piece to the Y-Piece seal.
Fig.4-8
54
PASITHEC User Manual
First step:
Pull out O 2 cell plug from inspiratory valve
cover.
Fig.4-9
Fig.4-10
Second step:
Connect the transparent joint which connects
O 2 sensor by cable to the top port of O 2 sensor.
Fig.4-11
Fig.4-12
55
PASITHEC User Manual
Third step:
Connect O 2 sensor with inspiratory valve cover.
Fig.4-13
Fourth step:
Connect the other end of O 2 sensor connecting
cable to the port marked “Oxygen” of the outlet
module.
Fig.4-14
CAUTION:The anesthetic gas/CO2 monitoring module has mainstream and sidestream. The two
56
PASITHEC User Manual
Step 1 :
Snap the probe on top of the airway adapter. It
will click into place when properly seated.
Fig.4-15
Fig.4-16
Step 2 :
Connect the RS232 interface of the adapter to
the RS232 interface on the back of the
machine.
Fig.4-17
57
PASITHEC User Manual
Step 3:
Fig.4-19
Step 4:
Connect the other end of the adapter to the
face mask or breathing bag.
Fig.4-20
Step 1 :
58
PASITHEC User Manual
Step 2:
Connect one end of the elbow to the Y-piece
Fig.4-22
Step 3:
Connect the other end of the elbow to the
breathing bag, or to patients through tube.
Fig.4-23
Step 4:
Connect the other end of the elbow to the
sampling tube, and make a connection between
the sampling tube and the CO 2 or anesthesia
gas water trap.
Fig.4-24
CAUTION: Make sure AGSS installs at a place where operator can get a clear view of it.
CAUTION: Connect AGSS according to the practical situation if no GCX rail is available on the
anesthesia machine.
59
PASITHEC User Manual
CAUTION: Make sure good connection in case of anesthesia gas leaks to operating room.
Step 1 :
A wide slot at the left side of anesthesia
machine specially designed for putting AGSS;
put AGSS to U-bracket.
Fig.4-25
Step 2 :
Hold AGSS with both hands ensuring the fixity
faces the wide slot. Slide AGSS horizontally to
rail till its limiting position.
Fig.4-26
60
PASITHEC User Manual
Step 3 :
Connect the two ends of screw thread to
port beneath breathing system and
AGSS respectively.
Fig.4-27
Step 4 :
Connect exhaust pipeline to exhaust port of
AGSS, screw it clockwise tightly.
Fig.4-28
WARNING:For a full understanding of the performance characteristics of this equipment, the user
61
PASITHEC User Manual
WARNING:Do not use the machine if any test failures. Contacts authorized service representative
WARNING:Make sure that the breathing system is in good connection and with no damage.
2. Make sure the breathing system is correctly connected and has no damage; the absorber canister
is fully filled with soda lime.
3. Make sure the vaporizer is well locked and is filled with adequate anesthetic agent.
4. Make sure the pipeline gas supply system is correctly connected and is at proper pressure.
7. Make sure the castors are not loose and the forefront two castors are locked.
8. Make sure the anesthesia machine is connected to power socket and power indicator is light.
There is no power supply if power indicator does not light.
1. Turn on power switch; user interface is light, STANDBY interface displays after self-test is finished.
62
PASITHEC User Manual
2. After the anesthesia machine operates under STANDBY mode for five minutes, pull out power cord
3. Make sure power failure alarm occurs (alarm bell sounds or user interface displays POWER
FAILUREURE message).
CAUTION:Operator should make sure gas supplies are correctly connected during operation,
there is no leak,wrong connection and damage in breathing circuit。 If something abnormal occurs,
WARNING: The malfunction of the center gas supply system may cause one or even all connection
1. Cut off pipeline gas supply. Perform the following steps if pipeline gauge does not read zero:
2. Connect O2 supply.
3. Set flow to the middle level.
6. Make sure O2 cylinder gauge reading restores to zero. Low gas source alarm occurs when O 2
pressure is decreasing.
63
PASITHEC User Manual
Ventilator Vt:700mL
Freq.:20
I:E:1:2
PAW-High:40cmH2O
Anesthesia Machine All gas flow:OFF
Push O2 flush to inflate bellows.
4. Make sure:
6. Make sure:
• Disassemble O2 sensor, make sure that O2 takes up 21% of room air according to the test result.
64
PASITHEC User Manual
• Put O2 sensor in pure oxygen; make sure that O2 takes up 100% according to the test result.
• Set low minute volume alarm limit to 10.0L/min, make sure low minute volume alarm is
eliminated.
• Mechanical ventilation stops when the machine operates at Manual ventilation mode.
• Make sure continuous PAW alarm occurs when continuous pressure is at 10cmH2O for 15
seconds.
65
PASITHEC User Manual
5. Operating guide
5.1. Opening
WARNING: Make sure the assembly of patient breathing tube and controlling setting is correct
before ventilating.
Fig.5-1
66
PASITHEC User Manual
Fig.5-2
After the system self-test:
1) Select and confirm “Continue”, the
Automatic Test begins.
Fig.5-3
2) Select and confirm “Skip”, the
“Manual Test” interface appears.
Fig.5-4
67
PASITHEC User Manual
Fig.5-5
Select “Continue”, the system
enter into the Automatic Test
interface.
“Automatic Test … ” appears in the
user interface.
Fig.5-6
2. Select “Continue” and press the knob ,the system begins to re-test,
The interface displays shown as above picture.
68
PASITHEC User Manual
Fig.5-7
2. If the bellows is not full, the
message “Push “O 2+ ” flush to
completely fill the Bellows”
appears on the user interface.
1) Select “Continue” to perform
Manual Test again.
2) Select “Skip” to proceed to
Manual Test.
Fig.5-8
69
PASITHEC User Manual
Fig.5-10
70
PASITHEC User Manual
71
PASITHEC User Manual
Fig.5-12
5. If the leakage is at the range of
100 — 400 ml/min, the interface
shows as the picture on the right.
Fig.5-13
6. If the safety valve is valid and
the leakage is less than
100 ml/min, the automatic test is
passed and then the system
proceeds to the Manual Test.
Fig.5-14
72
PASITHEC User Manual
Fig.5-17
73
PASITHEC User Manual
Fig.5-18
3. If the Y-piece port and manual bag
arm connector are not occluded, the
interface after test is shown as the
picture on the right.
74
PASITHEC User Manual
Fig.5-20
5. If no operation is done during the
first two 60s count down after the
Manual test interface appears, the
interface displays shown as the
picture on the right.
Fig.5-21
6.Manual test success
If the manual test succeeds , the
interface displays as shown on the
right.
The system will enter into Standby
mode automatically after 5s count
time.
Fig.5-22
75
PASITHEC User Manual
The patient type of the HEYER Pasithec anesthesia system has 2 type options: Child and Adult. The user
can select the needed.
CAUTION : The changing of the patient type is only allowed in STANDBY mode
Fig.5-23
Press the knob to select the patient
type button in the bottom, left corner
of the screen. This button will show
yellow when selected.
Fig.5-24
76
PASITHEC User Manual
Fig.5-25
Fig.5-26
77
PASITHEC User Manual
In addition, Standby mode is available. In Standby mode, mechanical ventilation cannot be carried
out and there are no ventilation waveforms or monitoring values.
In Manual mode, ventilation can be carried out by manual operation; there are ventilation waveforms
and monitoring values at present.
All the breathings are provided by the anesthesia machine, the patient cannot switch them by self.
The ventilation is managed by the predetermined ventilation volume, breathing frequency and ratio
of inspiratory time to expiratory time to keep the tidal volume stable, thereby assure the stable MV.
78
PASITHEC User Manual
Ventilation is managed according to a set goal pressure. Ventilation capacity (and velocity) is changed
accordingly, and related to respiratory system compliance and airway resistance. Tidal volume
changes with pulmonary compliance and airway resistance. Flow is slow-moving wave. Pulmonary
alveoli are full at the beginning of inspiratory which is helpful for the gas exchange in the lung.
79
PASITHEC User Manual
SIMV is a ventilation mode which combines spontaneous breathing with control ventilation. Within
the trigger window, the patient can trigger the command positive pressure ventilation synchronizing
with the spontaneous breathing. Between the two command ventilation cycle, the patient is allowed
to breath spontaneously. The command breathing follows the presets volume. By setting frequency
80
PASITHEC User Manual
and tidal volume to assure the minimum MV, cooperating with the patient’s spontaneous breathing to
reduce the rivalry between patient and machine and negative effect of blood dynamics, and prevents
the potential syndrome, such as air pressure injury and so on.
By changing the presets frequency to change the breathing support level, in other words, from entire
support to partial support, the machine can be used for long-term patient to retreat from machine.
81
PASITHEC User Manual
PS belongs to a partial ventilation support model, and is an auxiliary mechanical ventilation mode for
patient trigger, pressure goal and flow switch. Patient triggers ventilation and control respiratory
frequency and tidal volume. When airway pressure reaches default pressure support level and inhale
velocity reduces to the level lower than threshold, inhale phase switches to exhale phase. For
appropriate setup level, there will be few human-machine confrontation, it will effectively reduce
work of breathing and increase the effectiveness of effort to breathe. Some studies believe that 5-8
cmH2O PS can overcome the resistance of endotracheal tube and breathing machine circuit, so PS can
be used for machine withdrawal process. PS tidal volume is determined by breathing system
compliance and resistance.
• Positive End-Expiratory Pressure (PEEP): OFF, 3-30cmH2O; • Inspiratory Slope Time (TSLOPE): 0-2s.
Fig.5-30 Waveform of PS
82
PASITHEC User Manual
In PCV-VG, breaths are controlled by the anesthesia machine (mandatory) and will be delivered at a
set rate and set volume (VT). Inspiratory pressure will be regulated to achieve the operator set volume
(Tidal Volume). The pressure range that the anesthesia machine will use is between the PEEP + 2
cmH2O level on the low end and 5 cmH2O below Pmax on the high end. The inspiratory pressure
change between breaths is a maximum of +/- 3 cmH2O. PCV-VG begins by first delivering a volume
breath at the set tidal volume. The patient’s compliance is determined from this volume breath and
the inspiratory pressure level is then established for the next PCV-VG breath. Inspiratory pressure
from the last 3 breaths will be averaged to determine the pressure needed to maintain the set tidal
volume. Breath to breath inspiratory pressure change will be in increments of 3 cmH2O up to PMAX - 5
cmH2O. When PMAX is reached the pressure is maintained at the PMAX value until the end of inhalation.
The inspiratory and expiratory valves will adjust to maintain the PMAX value. When PMAX is reached, the
next breath cycle will be normal PCV-VG control and should not maintain PMAX.
83
PASITHEC User Manual
5.6.1.7. Synchronous intermittent mandatory ventilation with Pressure Control (SIMV-PC) ----
Optional
In SIMV - PC breaths will be controlled by the anesthesia machine (mandatory) or triggered by the
patient (spontaneous). When controlled by the anesthesia machine, breaths will be pressure limited
and time cycled, resulting in an operator set pressure (PTARGET) being delivered for an operator set
period (TINSP). The anesthesia machine delivers the mandatory breath in synchrony with the patient’s
inspiratory effort. If no inspiratory effort is detected, the anesthesia machine will deliver a
mandatory breath at the scheduled time. Mandatory breaths will be PCV-type breaths with
decelerating flow waveform. Between mandatory breaths the patient will be able to breathe
spontaneously. These breaths can be pressure supported. When PMAX is reached the pressure is
maintained at the PMAX value until the end of inhalation. The inspiratory and expiratory valves should
adjust to maintain the PMAX value. SIMV-PC mode setup parameters:
84
PASITHEC User Manual
The user can select the ventilation mode according to the patient’s actual situation.
The ventilation mode menus locate on
the left corner of screen. Click on the
ventilation menu to select the
corresponding ventilation mode.
Fig.5-31
85
PASITHEC User Manual
Fig.5-32
Click the PS button again to confirm
your selection.
Fig.5-33
If the patient type and ventilation mode have been selected, the system will display default parameters
according to the ventilation mode. The user could set the needed breathing parameter.
ΔP of PS mode is used below as an example of how this is done. These instructions also apply for other modes
and parameters.
86
PASITHEC User Manual
Fig.5-34
Press the knob to enable the
adjustment of the parameter setting.
The button will be highlighted yellow
when adjustment is allowed.
.
Fig.5-35
87
PASITHEC User Manual
Fig.5-36
Fig.5-37
CAUTION: The menu window covers the second and third waveform.
CAUTION: When selected, a short cut key will display the selected menu. When Normal is
selected all menus will be removed.
88
PASITHEC User Manual
89
PASITHEC User Manual
The alarm settings can be opened by touching the Alarm shortcut key. The alarm setting menu includes
four sub-options: Vent, Gas, Agent, and Log. The default alarm sub-option on start-up is
Vent Alarm limits.
Vent:
In the middle of the vent menu, the Fig.5-39 low and high alarm limits of MV, Pressure, and FreqMIN
can be set.
90
PASITHEC User Manual
Gas:
Fig.5-40
Agent:
Fig.5-41
Log:
Fig.5-42
Fig.5-43
91
PASITHEC User Manual
System setting menu includes three sub-menus: Display, Info and Setup. Display submenu is default
menu.
Fig.5-44
WAVE 3: three options: Flow, CO2 and VT, which displays in WAVE 3 area. Touch Sound: two options:
ON and OFF.
CO 2 UNIT: two options: % or mmHg.
The View Info. includes the follow
items.
EFM RT Sens Mon: view the real-time monitoring value of the oxygen flow meter sensor, the N2O flow
meter sensor, the air flow meter sensor, and the pressure sensor.
RT Vent Mon: view the real-time monitoring value of the inspiratory volume, the expiratory volume,
the system resistant, the system compliance, and the tube compliance.
SW Versions: the version of GUI, BDU, SBD, PSU, KBD, and EFM.
92
PASITHEC User Manual
HW Versions: the version of MCB, ADB, BCB, SBD, APB, GSPB_O2, GSPB_AIR, KBD, ISB, and EFMB.
5.7.4. Calibration
The Calibration sub-menu includes a User viewable item and Service only items. The user calibratable
item is the oxygen sensor.
CAUTION: Remove the oxygen sensor from the breathing system and expose the sensor to room air
Note: A plug is provided on a chain attached to the side of the breathing system for plugging the O2
CAUTION: The alarm “O2 Cal Due” will be displayed when the oxygen sensor has not been calibrated
93
PASITHEC User Manual
Fig.5-48
When the calibration is successful, the
interface will display “Calibration
Successful” and the running bar will be
completely green.
Fig.5-49
94
PASITHEC User Manual
Fig.5-50
5.7.5. Trend
When the trend shortcut is selected, the trend data can be viewed. The machine will record the last 8
hours of trend data at intervals of 30 seconds.
The trend time will appear on the left side of the menu, and the trend data will appear on the right
side of the menu, including Ppeak, Pplat, Pmean, PEEP, Freq, Vt, MV, FiO2, etCO2, FiCO2, Agent1,
Agent2, N2O, MAC, FG-O2, FG-AIR and FG-N2O.
Fig.5-51
The next page includes data from 7
parameters, as shown in Figure.
Fig.5-52
95
PASITHEC User Manual
5.7.6. Check
In standby mode, when the “Check” button is selected, the system test will be performed including Leak
test and Compliance Test.
Fig.5-54
If the Leak Test is success, the
interface is shown as the picture on
the right.
Select “Finish” to perform
Compliance Leak
Fig.5-55
96
PASITHEC User Manual
Fig.5-56
Compliance Test
Fig.5-58
97
PASITHEC User Manual
Fig.5-59
If the Compliance Test is success,
the interface is shown as the picture
on the right.
Select “Return” to finish the test.
Fig.5-60
Fig.5-61
98
PASITHEC User Manual
5.7.7. Configuration
99
PASITHEC User Manual
Fig.5-65
Enter the right password and click on “
”, the interface shown as the picture
on the right appears
Fig.5-66
100
PASITHEC User Manual
Fig.5-67
5.8. Shutdown
Turn OFF the gas supplies, until the gas in the system has gone out completely;
Turn the power switch to OFF to shut down the anesthesia system.
101
PASITHEC User Manual
CAUTION: If alarm occurs, the patient’s safety must be protected firstly, then do troubleshooting
or necessary remedy.
CAUTION: When several alarms occur at the same time, alarm message only displays the front two
CAUTION: When alarm silencing, the alarm bell has dashed “X” on itself. After 120s, the
alarm bell renews to the old state; If the alarm is not dealt in time, alarm continues.
CAUTION: If alarm occurs when the anesthetic machine works normally, visible and audible signals
WARNING: Do not set alarm limit exceeding the extremes, otherwise, alarm system will failure.
WARNING: A potential hazard can exist if different alarm presets are used for the same or similar
equipment in any single area, such as ICU or cardio tic operating room.
WARNING: The operator should examine if the current alarm setting is suitable for every patient.
WARNING: When power supply is interrupted, system will still renew the last alarm settings when
opening again.
The alarm states have 3 priorities: high, medium, low. High alarm needs to be dealt in time.
102
PASITHEC User Manual
The operator can judge if the alarm system is functional by visible and audible alarm message. The
distance that the operator can hear the alarm sound or distinguish the priority of alarm is 4 meters at
least. And the distance to see the alarm message is not less than 1 meter at the front of the machine.
The top area of the user interface displays alarm message, see figure 6-1.
Table 6-1
Priority Definition Alarm Tone Mute Notice Message
Medium Reversible injury. 3 alarm tones, alarm 120s Yellow background,”!!” flickering,
cycle is 10s. displaying frequency is 0.5Hz.
The alarm method includes technical alarm and functional alarm; technical alarm includes startup test
list and alarm occurring during normal operation.
103
PASITHEC User Manual
PAW(Child) 40 8
Freq(Adult) 18 --- 8~60 ---
Freq(Child) 30 ---
FiO2 100% 18% 21%~100% 18%~99%
Hal. 1.5 OFF 0.1%~8.4% OFF,0.1%~8.3%
Iso. 2.3 OFF 0.1%~8.4% OFF,
0.1%~8.3%
Enf. 3.4 OFF 0.1%~9.9% OFF,
0.1%~9.8%
Sev. 4.2 OFF 0.1%~9.9% OFF,
0.1%~9.8%
Des. 12.0 OFF 0.1%~21.9% OFF,
0.1%~21.8%
ETCO2 6.6% OFF 0.1%~9.9% OFF,
0.1%~9.8%
50mmHg OFF 1~75mmHg OFF,
1~74mmHg
INSCO2 0.7% --- ---
0.1~1.4%
5mmHg --- 1~10 ---
Other default items:
Pressure:PLAT
Table 6-2
Alarm Name Priority Cause Remedy
Low gas supply High Drive gas pressure is less Check pipeline gas supply and
pressure than 2.9psi±15% replace pipeline.
104
PASITHEC User Manual
Monitor O2
concentration≥10%,
cancel the alarm
immediately.
105
PASITHEC User Manual
BDU communication High GUI receives and sends data Ventilator failure.
from BDU unit failure lasts
failure
for 2s Switch to manual mode and the
user interface is still available.
User interface switches to
STANDBY mode automatically.
EFM communication High GUI receives and sends data Switch to the manual mode
from EFM unit failure lasts
failure for 2s Call service Rep
Gas module High Pull out adapter probe in Call service Rep.
probe failure gas module for several
seconds.
106
PASITHEC User Manual
Interior error in gas High Any case of software, Call service Rep.
module hardware, electric
“rotational speed“ exceeds
limit, no factory calibration
or factory calibration failure
appears, the alarm will
occur.
107
PASITHEC User Manual
Table 6-3
Alarm Name
Priority Condition Solution Remedy
108
PASITHEC User Manual
109
PASITHEC User Manual
1.Reduce alarm
setting;
2.Reduce air or
N2O
compensation
1.Reduce MV
1.Alarm occurs.; alarm setting
1. MV≤lower alarm limit In
Low MV Medium MANUAL mode, this alarm
2.Ventilator 2.Check if there is
could be closed.
works normal leak in patient end.
110
PASITHEC User Manual
Low
ETCO2 concentration is lower
ETCO2 concentr
High than low alarm limit and lasts
ation
at least 2 continual breaths.
High
INCO2 concentr Alarm according to detected real-
High
ation time data
Gas module
When gas module
communication Check connection
communication is abnormal, Alarm occurs.
failure or replace gas
High such as unconnected, and lasts Alarm occurs.
module
for 5s.
111
PASITHEC User Manual
Table 6-4
BDU EEPROM data fail Incorrect calibration in EEPROM check Call service Rep.
PAW sensor fail The data of the pressure sensor is Call service Rep.
incorrect.
Ambient pressure sensor fail The data of the ambient pressure sensor Call service Rep.
is incorrect.
PEEP valve fail Incorrect PEEP valve data Call service Rep.
Inspiratory valve fail Incorrect monitor sensor ZERO state Call service Rep.
SW version fail BDU, GUI, KBD software version Call service Rep.
is not the same with the released one.
BDU Comm. fail If GUI cannot receive and send data from All breathing functions fail,
BDU unit for 0.5s. call service Rep.
KBD Comm. fail No communication between GUI and All breathing functions fail,
Keyboard. call service Rep.
112
PASITHEC User Manual
Inspiration valve fail Incorrect voltage value the The machine can continue
inspiratory valve feedbacks to work, but there is only
manual mode but
automatic mode, the
monitor function is
effective, call service Rep.
O2 supply fail The pressure of drive gas is lower Check if the pipeline
than 200kPa ± 15% . supply is normal, connect
the backup gas supply.
EFM communication GUI receives and sends data from Switch to the manual mode
EFM unit failure lasts for 2s
fail Call service Rep
113
PASITHEC User Manual
Table 6-5
No. Message Cause Remedy
114
PASITHEC User Manual
6. Gas monitoring 1. Sample line is blocked or 1.Check sample line, water trap and
unconnected.
has no Y-piece filter.
sampling signal
2.Replace if necessary.
11. CO2 sensor 1. CO2 gas measurement 1.Use external gas measuring
failure system has failed.
system.
115
PASITHEC User Manual
14. EXP.flow sensor 1. Expiratory flow sensor has 1. Replace flow sensor.
failure failed.
15. Fresh gas flow too high 1.Total fresh gas flow is above 1. Reduce fresh gas flow;
19L/min.
2. Check vaporizer setting.
16. Fresh gas low or leak 1. Fresh gas setting too low; 1.Increase fresh gas flow;
2. Leak. 2. Repair leak.
17. Gas sensor 1.Complete gas measurement 1.Use external gas measuring
failure system fails.
system;
2. Call service Rep.
116
PASITHEC User Manual
18. High airway 1. Hifh alarm limit for the 1.Check hose system and tube;
pressure
airway pressure has been
2. Correct the ventilation settings.
exceeded;
2.Ventilation hose kinked;
3. Ventilation settings are not
correct.
19. Inaccurate fresh 1. Reduced accuracy of fresh 1.Reduce fresh gas flow for each
gas flow gas flow measurement.
gas to below 12L/min;
20. Inhaled CO2 1.Soda lime in circle system 1. Increase fresh gas flow;
high
exhausted;
2.Replace breathing system;
2.Leak or fault in breathing
system; 3.Adjust alarm limits if necessary;
4.Check ventilation settings;
3.High ventilation
5. Replace flow sensor.
frequencies;
4.Dead space ventilation.
117
PASITHEC User Manual
22. Inhaled N2O 1. Inhaled N2O concentration 1. Check N2O concentration in the
concentration exceeds the high alarm limit fresh gas flow.
high of 82%.
2.Check O2 supply;
25. No air supply 1.Comm pressed air supply 1.Open optional backup air
has failed; supply;
118
PASITHEC User Manual
26. No N2O supply 1.N2O supply has failed; 1. Open N2O backup cylinder;
29. O2 flow 1. Fresh gas flow 1.Use only O2 as fresh gas and
measurement measurement for O2
observe total flow meter.
failure has failed.
30. PEEP high 1. Expiratory pressure 5cmH2O 1.In automatic ventilation modes
above PEEP for 2 breaths, or
check the ventilation parameters.
Expiratory pressure 5cmH2O
above PEEP in pressure
2.Check the anesthetic gas
support mode for more than
scavenging line.
30s.
119
PASITHEC User Manual
120
PASITHEC User Manual
WARNING: Use a cleaning and sterilizing schedule that conforms to your institution’s sterilization
Refer to the operating and maintaining manual of all the sterilizing equipments.
The O2 sensor may leak and burn (by Chlorine Potassium Oxide) if damaged.
WARNING:Do not inhale any agents generated during cleaning and disinfecting.
CAUTION:User cleaning agent sparingly. Excess fluid could enter the machine, causing damage.
CAUTION:Do not autoclave any parts of the machine unless specifically identified as autoclave
able in the User Manual. Clean the machine only as specified in the User Manual.
Never use abrasive agents (i.e. steel wool or silver pool ish) to clean components.
121
PASITHEC User Manual
or accelerated aging.
Only the components marked 134℃ are heat-resistant and pressure-resistant parts
All cleaning solutions used must have a pH between 7.0 and 10.5.
CAUTION:Never use the oxygen sensor or its connector in any type of liquid.
CAUTION:The valve disc is fragile and must, therefore, be handled with care while removing the
CAUTION:If moisture remains in the bellows after cleaning, it may become tacky.
CAUTION:Prior to use after cleaning or disinfecting, power up the machine and follow the on-
screen prompts to perform the Leak Test and the Compliance Test.
CAUTION: The PAW gauge cannot withstand immersion or the heat and pressure of autoclaving.
122
PASITHEC User Manual
7.1. Guideline
Operator should have a better understanding of WARNINGS and CAUTIONS before cleaning and
disinfecting. Different methods will be used for different parts. Parts need to be completely dried after
cleaning and disinfecting.
After cleaning and disinfecting, the peruse calibration must be done before using after the
anesthesia machine is reinstalled, refer to “Preoperative preparation” in HEYER Pasithec User
Manual.
The recommended disinfection time of the machine is 100~300, according to the use condition and
sterilization time of anesthesia machine specificially.
ISO17664 conformity
Process of autoclave sterilization of the anesthetic breathing system conforms to ISO17664:2004. The
condition conforming to ISO17664:2004 is use bacteria or virus filter to filter the flow that exhaled by
the patient and enters into the system and the flow which returns to the patient, so the installation of
the filter must be correct.
Cleaning Information
Cleaning is supposed to do before using the breathing system for the first time after installation.
If there is potential infective object in the breathing system, such as blood or secretion, please clean
it with disposable cloth and authorized sterilant instantly.
Only the components marked 134℃ are heat-resistant and pressure-resistant parts that are capable
of withstanding autoclave sterilization, all parts expect O2 sensor and PAW gauge can be washed with
moderate sterilant in the automatic washing machine.
O2 sensor and PAW gauge can be only cleaned on surface, O2 sensor, cannot be autoclave sterilized
and cleaned with automatic washing machine.
123
PASITHEC User Manual
124
PASITHEC User Manual
WARNING:Wear latex gloves during disassembly in case that operator’s hands get damaged.
Fig.7-1,7-2
Take the heater cable from the outlet module.
Fig.7-3,7-4
Take the heater cable from the heater port in
the breathing system.
Fig.7-5,7-6
125
PASITHEC User Manual
Fig.7-7,7-8
Take the screwed pipe from the breathing
system and store safely.
Fig.7-9,7-10
Take the integrated gas sampling pipe from the
outlet module.
Fig.7-11,7-12
One hand holds the bracket, the other hand
holds the other side of the breathing system ,
and lift the breathing system vertically, take the
breathing system from the anesthesia machine.
Fig.7-13
126
PASITHEC User Manual
Fig.7-14
Fig.7-15
Fig.7-16
Take PAW gauge from the breathing system
and store it safely.
Fig.7-17,7-18
Take O 2 sensor from the inspiratory valve.
127
PASITHEC User Manual
Fig.7-19,7-20,7-21
Turn the bellows counterclockwise and take it
from the breathing system.
Fig.7-22,7-23
128
PASITHEC User Manual
Fig.7-24,7-25
As shown in the right figure, autoclave sterilize
the disassembled breathing system.
Fig.7-26
129
PASITHEC User Manual
A B C D
Workstation(Outside) √√√
Vaporizer √
Connection Cables and Wires √
Breathing Bag, Y-Fitting, T-Fitting √
Reusable Screw Tube √ √ √ √
Breathing System(After √ √ √ √
Disassembly)
Insp./Exp. Valve Disc √ √ √ √
Insp./Exp. Valve Bracket √ √ √ √
Insp./Exp. Valve Dome √ √ √ √
O-Ring √ √ √ √
Apl Valve √ √ √ √
Paw Gauge √
O2 Sensor √
Flow Sensor Surface √
Flow Sampling Tube √
Flow Sampling Detector √ √ √
Absorber Canister √ √ √
Bellows Dome √ √ √
Bellows √ √ √
Bellows Gasket √ √ √ √
AGSS √ √ √ √
AGSS Transfer Tube √ √
√ √
√ √
√ √ √
130
PASITHEC User Manual
1. Breathing Circuit
After using on every patient, clean breathing circuit with 70%~90% alcohol, wash with clean water
and dry, or place it in autoclave staircase for sterilization.
After using on every patient, clean silicon screw tube and breathing bag with clean water, completely
dry and put them in autoclave staircase for sterilization.
CAUTION:Do not use ultra-violet wave to sterilize silicon tube or breathing bag in case of aging.
Either vaporing or immersion disinfection can be used in practice, in case of immersion all sterilized
parts must be dried with the high pressure air or oxygen before reuse.
4. Insp./Exp. Valve
Dismount the cover of the inspiratory and expiratory valves by rotating it counter-clockwise, then
takes away the valve patch, clean valve seat and circle together. And clean cover of the inspiratory
and expiratory valves and valve patch with the gauze soaked with water soluble sterilizing agent,
after all parts cleaned and dried recover it in original integration. Then one must check the leakage
and the movement of the inspiration and expiration valves in accordance with the required
regulation and checking procedure. Please handle all parts with care preventing any damage.
WARNING:Take care and do not damage or crash valve disc during cleaning.
WARNING: Never separate the diaphragm and the valve seat in a pressure-relief valve.
CAUTION: Do not immerse them more than 15 minutes to prevent inflation or aging.
131
PASITHEC User Manual
CAUTION: Dry by hanging while fully spread. If moisture is left in the bellows, they may become
tacky.
1) Disassembling.
2) To prevent component damage, clean them lightly. Put the recommended non-enzyme mild agent
4) Check the components if they are broken or damp, then perform the assembling and function test.
CAUTION:For more cleaning and disinfecting information of AGSS, refer to AGSS User Manual
(Material No.:130).
8. Term maintenance
Do not use lubricant containing oil or lipin, because when O2 concentration reaches a certain
degree, burning or explosion may happen.
WARNING;Please observe disinfection control and safety regulation because the devices having
132
PASITHEC User Manual
WARNING: Moving part and removable elements is in danger of jamming hands or crush.
WARNING: In the process of removing product, strike and quiver must be avoided..
WARNING: Exhausted and old instruments (such as battery and LCD screen) harmful to
This chapter provides message needed by HEYER Pasithec anesthetic machine in term maintenance.
Replace or adjust some components before doing some calibration or adjustment. If user uses
improper components such as improper specification causes malfunction, our company will not take
responsibility. Do not use malfunction device. Replacement and maintenance should be finished by
authorized service engineer or qualified, trained persons with repairmen experience. After
maintenance, the device should be tested and ensure its normal function and comply with the
manufacture’s regulation. Use components manufactured or sold by our company if replacement is
needed, then test and ensure device comply with the manufacture’s specification. If service support
is needed, please call local service engineer. In any case, the repairmen cost will include the current
price and reasonable personnel cost, but except repairmen items within HEYER company’s warranty.
CAUTION: Person without repair experience about such device must not repair this device.
133
PASITHEC User Manual
Table 8-1
Action After each repair Every 12 months Every 36 months
Check before use X X X
Check before operation √ X X
Check with eyes √ X X
Replace exhausted √ X X
component
Replace O2 sensor √ √ √
Replace/maintenance √ √ X
battery
Function test √ X X
CAUTION: According to this schedule to do maintenance timely and replace corresponding parts.
CAUTION: The recommended repair time interval is 5 years. Breathing system, vaporizer and parts
WARNING: The ineffective sodium lime has an obvious change in color when it’s ineffective.
1. Assure the appearance of HEYER Pasithec no damage, the anesthesia machine can run
normally.
2. Assure absorbent canister install correctly and fill enough soda lime. Assure enough
anesthetic in vaporizer.
3. Assure the exhaust gas absorber pipe connected with APL valve no damage and dry inside.
4. Assure O2, N2O, AIR, VAC and EVAC connection pipe no damage.
5. Assure O2, N2O, AIR, VAC and EVAC flexible pipe no damage (if they are be used).
6. Assure AC power cable no damage.
134
PASITHEC User Manual
CAUTION: The life-span of all components above is the result in normal circumstance and work
condition.
135
PASITHEC User Manual
WARNING: Exhausted and old O2 sensor must be treated according to the present local regulation.
3. Connect new O2 sensor to cable connection port and install it on inspiratory valve of breathing
system.
CAUTION: Specific parameters refer to the latest published technical data by manufacture.
O2 sensor can be used to detect partial O2 concentration of the anesthetic machine. O2 sensor
belongs to consumptive products, so attentions should be paid to its useful-life and use as the
characteristics and technical requirements provided by manufacture. Main technical
requirements of O2 sensor in HEYER Pasithec as follows:
136
PASITHEC User Manual
Performance parameters:
Interface form and definition: FCC-68 4 core telephone pin (RJ11-4), see the following figure.
Range :0-1500mBar O2
Resolution :1mBar O2
ZERO signal(N2,+20℃):<200uV
Pressure range:0.5~2.0Bar
Long-term output excursion:<5%(in 100% O2 circumstance and use more than one year)
137
PASITHEC User Manual
Complied standard:EN12598/ISO7767
ALARM: Disconnect AC power before fuse replacement, otherwise, it may cause injury even death.
WARNING: When fuse replacement, fuse of the same type and size must be used, otherwise, it
CAUTION: Fuse belongs to damageable assembly, proper force must be used when replacement.
CAUTION;Use screwdriver to swirl fuse and not use force largely and suddenly, otherwise, it
Replacement steps:
138
PASITHEC User Manual
CAUTION: Only service engineer authorized by our company can replace battery. If the anesthetic
machine is not used in a long time, please call service engineer to cut power supply connection.
CAUTION: The exhausted and old battery must be treated according to present local regulations.
Battery configuration
Charging: when the AC power is connected, the system will charge the battery automatically, the
recommended charging time is no less than 8 hours.
Discharging: when the device uses battery to supply power, normally the work time is 120
minutes.When the battery’s electric capacity is insufficient, the device will display “Low power”
alarm until power off. When the “Low power” alarm occurs, the user should connect AC power
immediately and avoid the system closes (considering safety, manual startup should be done to
renew operation after the device closes automatically.)
Do not disassembly the battery causally and short battery connection line to avoid danger.
Battery’s Preservation
If the battery is not used for a long time, it should be preserved after the battery is charged fully.
When the preservation time exceeds 3 months, charge the battery at least every 3 months.
Improper maintenance will cause battery broken, the battery should be replaced immediately to
avoid fluid leak erode device. Call manufacture when replace battery.
Battery’ removal
Use screw driver to remove the screws of the back cover board, disconnect the battery module
with device to remove battery.
139
PASITHEC User Manual
Since installation, consumptive components names and code in every time dimension as the table in
below:
9. Technical Data
CAUTION: All technical specifications should be updated based on the real state of HEYER
Pasithec Anesthesia Machine; no notification will be made if any change happens.
CAUTION:All display values are measured under dry environment and proper pressure.
WARNING: Cylinder valve should be closed during pipeline gas supply, so that gas shortage can
Pneumatic system plays the role of supplying O2, N2O, AIR and anesthetic agent to breathing system,
ensuring their pressures are within normal range respectively, so that normal operation of
anesthesia machine and patient’s safety can be guaranteed.
140
PASITHEC User Manual
141
PASITHEC User Manual
Gas Supply
Table 9-1
Pipeline Gas O2, N2O,AIR
Pipeline Connection DISS-male,DISS-female,NIST(ISO 5359).
Connectors for each of the gas cannot be interchanged.
Pipeline Input Pressure 280-600kPa(2.8-6bar)
142
PASITHEC User Manual
HEYER Pasithec Anesthesia Machine can be divided into three parts: UI, drive control and circuit. UI
consists of computer system, auxiliary control circuit, power and etc. which mainly takes charge of
display, alarm, keyboard input and interface expansion. Drive control has BDU as its core. Exterior
sensor, signal processing, amplifier, power and battery are part of drive control which mainly used
for data acquisition and ventilation mode control. UI and drive control should be connected by
cables.
143
PASITHEC User Manual
144
PASITHEC User Manual
CAUTION:All physical specifications are of approximate values, no CAUTION can be made if any
change happens.
CAUTION:Do not put heavy stuff on the top panel or in the drawers.
Table 9-3
Dimensions: 1403mm(H)×681mm(W)×855mm(D)
Weight: 100kg ( not include vaporizer and
System
backup cylinder)
Top Panel Load: 25kg
Castor 125mm,each front castor with brake
Drawer 142mm(H)×376.5mm(W)×438mm(D)
Display 12.1’ TFT LCD
Cylinder Gauge(AIR,O2, Range:0-1MPa. Stepping:0.05MPa Accuracy:±2.5% of
N2O) full range
Range:-10 - 100cmH2O. Stepping:200Pa Accuracy:±2.5% of
PAW Gauge
full range
CAUTION:The anesthesia machine should be stored in draught room where there is no caustic
Table 9-4
Temperature Operation: 10~40℃
145
PASITHEC User Manual
Storage: 70~106kPa
Transport 70~106kPa
Power AC100-240V,50Hz/60Hz
Class I device
Common device/IP21
Traveling device
Continuous running
NOTE:
The applied Part of HEYER Pasithec anesthesia machine is the part between mask and Y-piece.
146
PASITHEC User Manual
NOTE:
1. When the anesthesia system stops transferring, the pipeline supplying pressure should be 280 to
600KPa.
2. When main power fails, the system switches to battery power automatically, and all the anesthesia
3. When main power and internal battery both fails, the gas output is not affected.
9.6.1. Electrical
WARNING: IEC60601-1 is applicable to the connections between medical devices, and to the
connections between at least one medical device and one or more non-electrical devices. Even if no
functional connections between the single device, when using one auxiliary power outlet, they
forms a medical electrical system. User must be aware of the risk of increased leakage current and
temperature and other increased risk may exist, when the device is connected to the auxiliary
socket-outlet.
WARNING:When earth wire has failure, if the device connects with auxiliary power supply, it may
WARNING:If the exterior protective grounded wire has questions during wiring, interior power
Table 9-5
Supply Voltage 100-240VAC,50/60Hz
Maximum Input Current 8A
AC power fuse F10AH 250VAC,φ5X20(F)
147
PASITHEC User Manual
9.6.2. Pneumatic
Table 9-6
WARNING Using cell phone or other radio radiant equipment near this product may cause
malfunction. Closely monitor the working condition of this equipment if there is any radio
radiant supply nearby.
Using other electrical equipment in this system or nearby may cause interference. Check if the
equipment works normally in these conditions before using on a patient.
Do not put any object which is not in accordance with EN60601-1 in the 1.5M range of patients.
148
PASITHEC User Manual
An isolated transformer must be used for alternating current supply (in accordance with IEC60989),
or additional protective ground wires are equipped if all the devices (for medical or non-medical
use) are connected to HEYER Pasithec by using signal input/signal output cable.
If a portable all-purpose outlet is used as the alternating current supply, it must be in accordance
with EN60601-1-1 and cannot be put on the floor. Using another portable all-purpose outlet is not
recommended.
Do not connect the non-medical equipment directly to the alternating current outlet on the wall.
Only the alternating current supply of the isolated transformer can be used. Otherwise, the surface
leaking current may exceed the range permitted by EN60601-1 under the normal conditions, and
disoperation may cause injury to patients or operators.
HEYER Pasithec is equipped with all-purpose alternating current outlet for connecting other medical
equipments. Do not connect non-medical equipment to these outlets. Otherwise, the surface
leaking current may exceed the range permitted by EN60601-1 under normal conditions and
disoperation may be dangerous to patients or operators.
A complete system current leaking test (according to EN60601-1) must be performed after any
equipment is connected to these outlets.
operators.
149
PASITHEC User Manual
The HEYER Pasithec Anesthetic machine is intended for use in the electromagnetic environment
specified below. The customer of the user of the HEYER Pasithec Anesthetic machine should assure
that it is used in such and environment.
IEC 61000-3-2
Voltage Complies
fluctuations/
flicker emissions IEC
61000-3-3
150
PASITHEC User Manual
The HEYER Pasithec Anesthetic machine is intended for use in the electromagnetic environment
specified below. The customer or the user of HEYER Pasithec Anesthetic machine should assure that
it is used in such an environment.
Electrical fast ±2 kV for power ±2kV for power Mains power quality should be
supply lines supply lines that of a typical commercial or
transient/burst
hospital environment.
IEC 61000-4-4
±2 kV common ±2 kV common
mode mode
151
PASITHEC User Manual
70% UT 70% UT
(30% dip in UT) for 25
cycles (30% dip in UT) for
<5% UT
(>95% dip in UT) for 5
25 cycles
sec
<5% UT
CAUTION UT is the a.c. mains voltage prior to application of the test level.
The HEYER Pasithec Anesthetic machine is intended for use in the electromagnetic environment
specified below. The customer or the user of HEYER Pasithec Anesthetic machine should assure that it
is used in such an environment.
bands
10 V/m 23
10 V/m 80 MHz 𝑑 = [𝐸 ] √𝑃 800 MHz to 2.5 GHz
1
Radiated RF to 2.5 GHz
IEC
61000-4-3
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m). b
153
PASITHEC User Manual
CAUTION 1 At 80 MHz and 800 MHz, the higher frequency range applies.
CAUTION 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
The ISM(industrial, scientific and medical) bands between 150kHz and 80MHz are 6.765MHz to
6.795MHz; 13.553 MHz to 14.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70
MHz.
b
The compliance levels in the ISM frequency bands between 150kHz and 80MHz and in the
frequency range 80 MHz to 2.5GHz are intended to decrease the likelihood that mobile/portable
communications equipment could cause interference if it is inadvertently brought into patient
areas. For this reason, an additional factor of 10/3 is used in calculating the recommended
separation distance for transmitters in these frequency ranges.
c
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the HEYER Pasithec Anesthetic machine is used exceeds
the applicable RF compliance level above, the HEYER Pasithec Anesthetic machine should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the HEYER Pasithec Anesthetic machine.
d
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
154
PASITHEC User Manual
portable and mobile RF communications equipment and the HEYER Pasithec Anesthetic machine
The HEYER Pasithec Anesthetic machine is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the HEYER Pasithec
Anesthetic machine can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
HEYER Pasithec Anesthetic machine as recommended below, according to the maximum output
power of the communications equipment.
Rated (m)
maximum
150 kHz to 80 MHz 150 kHz to 80 MHz 80 MHz to 800 800 MHz to 2.5
output power outside ISM bands
in ISM bands MHz GHz
of transmitter
3.5 12 23
𝑑 = [ ] √𝑃 12 𝑑 = [ ] √𝑃 𝑑=[ ] √𝑃
𝑉1 𝑑 = [ ] √𝑃 𝐸1 𝐸1
(W) 𝑉2
155
PASITHEC User Manual
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
CAUTION 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
CAUTION 2 The ISM(industrial, scientific and medical) bands between 150kHz and 80MHz are
6.765MHz to 6.795MHz; 13.553 MHz to 14.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz
to 40.70 MHz.
CAUTION 4 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
156
PASITHEC User Manual
Table 9-7
157
PASITHEC User Manual
In all of software version, the anesthesia machine is able to carry out a compliance correction which
compensates for volume "lost" in hoses and in the breathing system.
The anesthesia machine determines the system compliance (hoses and breathing system) and
required later for compliance correction during the self test or standby leak test. The hose of
standard equipped compliance has a fixed value of 11mL/kPa.
Having the system compliance (Csys) and ΔP, Anesthesia machine can calculate which volume (ΔV)
has been "consumed" by the system, or, in other words, which volume has not been delivered to
patient.
ΔV = Csys x ΔP
The anesthesia machine increases the volume to be delivered by the value ΔV.
VT corr = VT + ΔV
The anesthesia machine repeats this procedure during each breathing phase so that the VT
delivered to the patient gradually approaches the set VT. This procedure is completed as soon as
158
PASITHEC User Manual
the plateau value stops changing (±0.5 cmH2O).The displayed flow and the resulting values (for
example, MV) are updated accordingly.
Table 9-8
Maximum Inspiratory 80cmH2O
Pressure
Noise of the whole Work normally(not including alarm),no more than 60dB(A)
Machine
Warm-up Time ≥5min
MV Max 20L/min
Inspiratory Flow Max 75L/min
Pressure Transmission 5-80cmH2O
Range
159
PASITHEC User Manual
Table 9-9
Item Range Stepping Default Value Remark
160
PASITHEC User Manual
Table 9-10
Gas source Anesthetic system
Gas component O2
Rating supply pressure 400kPa
Inlet pressure range 280kPa ~ 600kPa
Flow valve range 5~75L/min
1
Output Pressure range: 0 ~ 8kPa;Flow range: 0 ~ 75L/min
Label 1:Within rated outlet pressure range and under double maximum rated inlet pressure
condition.
Table 9-11
According to ISO8835-2, test under peak flow 60L/min, fresh gas 10L/min.
Table 9-12
Item Range Stepping Accuracy
161
PASITHEC User Manual
Battery 100%,75%,50%,25%,0%。
voltage When battery is used to supply power, this sign displays available remaining electric
state voltage. When the machine is connected with AC power, this sign means charging.
display
Pressure monitor range: 0 - 80cmH2O.According to different airway pressure, gain the
wave form displays is different.
Paw-t 0 ~ 10cmH2O,pressure axis gain: 5cmH2O
wave 0 ~30cmH2O,pressure axis gain: 10cmH2O
form 0 ~80cmH2O,pressure axis gain: 20cmH2O
Time axis is a fixed range.(the axises of Flow-t,Paw-t is same.):
When gas module is opened,0 ~15s;Gas module is closed, 0 ~20s.
Display range of flow rate: -90 - 90L/min,gain :45L/min. On time axis, the positive axes
Flow-t stands for inspiratory direction; under the time axis, the negative axes stands for
wave expiratory direction. Flow rate is 0L/min, which means no gas flow rate in airway.
form
CO2 CO2 display range is 0 - 76mmHg,gain is 38mmHg。Time axis is a fixed range. When
time gas module is opened,0 ~15s;Gas module is closed, 0 ~20s.
wave
form
Horizontal abscissa displays Paw:fixed range :-20 ~ 80cmH2O,gain is 20cmH2O。
P/V Loop Vertical abscissa displays tidal Volume:fixed range is 0 ~ 1500ml,gain is 250ml.
162
PASITHEC User Manual
Table 9-13
Response time No more than 15s.
Sensor type Chemical fuel battery
Antipated battery Normal operation period is 12 months.
life-span
O2 monitor monitors and displays O2 concentration in the patient loop.
O2 sensor accessory includes one O2 sensor. This sensor can produce a
voltage proportional with O2 partial pressure in its detected surface and
its metal electrode will exhaust gradually in the process of oxidation.
The gas monitoring module is used to connect other medical devices, provide monitor data, monitor
and display CO2, N2O and 5 anaesthetic gases (halothane, isoflurane, enflurane, sevoflurane,
desflurane).
163
PASITHEC User Manual
The gas monitoring module is connected to the patient’s breathing pipeline to monitor the inspiratory
and expiratory gas and can be applied to Adult/Child and Infant.
Table 9-14
IRMA CO2 / AX+
Model
Working principle of gas The gas module is a mainstream gas monitoring module, connecting to
module
patient to monitor gas concentration of breath.
The gas module can be used for adults, children and infants.
Table 9-15
Basic Configuration
38×37×34mm(1.49″×1.45″× 1.34″)
164
PASITHEC User Manual
IRMA AX+:
700-1200hPa(700 hPa corresponding to an
altitude of 3048m/10000feet )
165
PASITHEC User Manual
IRMA AX+:
4.5-5.5VDC, Max1.4W
Max 50℃/122℉
Disposable infant:
NOTE 1: After being in a condensing atmosphere, this module shall be stored for more than 24
The humidity range 50-100% is valid for the temperature range -40 - 40℃.
166
PASITHEC User Manual
Table 9-16
Data Output
Fi and ET The newest breathing value (Fi and ET) is continually displayed after one
breath.
Table 9-17
Gas Analyzer
Probe 2-9 channel NDIR type gas analyzer measuring at 4-10 μm. Pressure,
temperature and full spectral interference correction.
N2O≤300ms
167
PASITHEC User Manual
Range1)
8-12
10-15
22-25
Table 9-19
Accuracy specifications –all conditions1)
Gas Accuracy
168
PASITHEC User Manual
NOTE 1: The accuracy specification is valid for the operating temperature and humidity
conditions specified, except for interference specified in the table “Interfering gas and vapor effects”
below.
NOTE 2: The accuracy specification for IRMA AX+ is not valid if more than two agents are present
Table 9-20
Interference gas and vapor effects
Metered dose Not for use with metered dose inhaler propellants
inhaler propellants3)
169
PASITHEC User Manual
NOTE 1: Negligible interference, effect included in the specification “Accuracy, all conditions”
above.
NOTE 2: Interference at indicated gas level. For example, 50vol% helium typically decreases the
CO2 readings by 6%. This means that if measuring on a mixture containing 5.0vol% CO2 and 50vol%
helium, the measured CO2 concentration will typically be (1-0.06)*5.0vol%=4.7 vol% CO2.
170
PASITHEC User Manual
Table 9-21
ISA CO2 / AX+
Model
Working principle of gas The gas module is a sidestream gas monitoring module, connecting to
module
patient to monitor gas concentration of breath by electrical machine
inhaling trace gas.
The gas module can be used for adults, children and infants.
171
PASITHEC User Manual
Table 9-22
Basic Configuration
23×64×39mm(0.9″×2.5″× 1.5″)
Weight 70g
(95%RH at 30℃)
172
PASITHEC User Manual
ISA AX+:
NOTE 1: After condensation, the unit shall be stored for more than 24 hours in an environment
Table 9-23
Data Output
ISA AX+: CO2, N2O, primary and secondary agents(HAL, ENF, ISO, SEV, DES)
173
PASITHEC User Manual
Table 9-24
Gas Analyzer
Sensor head 2-9 channel NDIR type gas analyzer measuring at 4-10 μm.
HAL,ISO,ENF,SEV,DES≤350ms
Table 9-25
174
PASITHEC User Manual
The following accuracy specifications are valid for dry, single gases at 22±5℃ and 1013±40 hPa.
Accuracy specifications - standard conditions
15-25% Unspecified
8-25% Unspecified
10-25%
22-25%
Table 9-26
Gas Accuracy
NOTE 1: The accuracy specification is not valid if more than two agents are present in the gas
mixture. If more than two agents are present, an alarm will be set.
Table 9-27
175
PASITHEC User Manual
Metered dose Not for use with metered dose inhaler propellants
inhaler propellants4)
176
PASITHEC User Manual
NOTE 1: Negligible interference, effect included in the specification “Accuracy, all conditions”
above.
NOTE 2: Negligible interference with N2O/O2 concentrations correctly set, effect included in the
NOTE 3: Interference at indicated gas level. For example, 50vol% helium typically decreases the
CO2 readings by 6%. This means that if measuring on a mixture containing 5.0vol% CO2 and 50vol%
helium, the measured CO2 concentration will typically be (1-0.06)*5.0vol%=4.7 vol % CO2.
177
PASITHEC User Manual
10.Warranty
In addition to the legal warranty acc. to HBG §377, HEYER MEDICAL AG shall grant a warranty of 12
months for the purchase of a new apparatus from the HEYER product range. The warranty period
begins with the date of invoice and is subject to the following conditions:
1. Within the warranty period we will eliminate free of charge any defects or damages on the
device that are shown to be caused by a manufacturing or material error. The warranty does
not include easily breakable parts, e.g. glass or consumable parts.
2. Warranty services can only claimed upon submission of a delivery note (bill of delivery or
invoice); the type and method of damage remedy (repair or replacement) shall be at the
discretion of HEYER MEDICAL AG. Warranty services do not result in an extension of the
warranty period, nor do they entail a new warranty being granted. There is no independent
warranty period for installed spare parts.
3. Excluded from the warranty are: Damages caused by improper use, operating errors,
mechanical stress or non-observance of the operating instructions, as well as damages
caused by force majeure or by extraordinary environmental conditions.
4. Warranty services may only be claimed if proof is submitted to confirm that all service and
maintenance work has been carried out by authorized staff.
5. The warranty includes all faults that impair a faultless functioning of the device on the basis
of technical defects of individual components. The warranty obligation can only be
recognized by us if the device has been used properly and according to its intended use and
no repair attempts have been undertaken by the client himself or by third parties. The
warranty claim does not include faults caused by mechanical damages or if the device is being
operated with accessories originating from third parties.
6. The warranty is also void if changes, alterations or repairs are made to the device by persons
not authorized to do this.
7. The warranty claim only applies to customers of HEYER MEDICAL AG; it cannot be transferred
to third parties.
8. The rejected device is to be shipped back to our plant postage free. In case of a request by
our customer service department, the costs for shipping to the plant are to be initially
generally borne by the customer. After successful repair, we will send the device back freight
collect. If HEYER MEDICAL AG confirms the existence of a warranty claim, the customer will
receive reimbursement for the costs of delivery and/or transport of the apparatus. Repair
parts that do not fall under the warranty claim will be billed by us. The shipping of the device
to us always counts as a complete assignment to eliminate all faults and/or replace missing
178
PASITHEC User Manual
parts, unless the customer expressly excludes partial services. Additional claims to transfer
or reduce and replace damages of any kind in particular also of damages not incurred on the
object of delivery itself are excluded.
NOTES
179
PASITHEC User Manual
180
PASITHEC User Manual
181
PASITHEC User Manual
HEYER MEDICAL AG
HEYER Medical AG
Carl-Heyer-Str. 1/3
D-56130 Bad Ems / Germany
Phone: +49 2603 791 3 ◦ Fax: +49 2603 70424
[email protected] ◦ www.heyermedical.de
182