Leybold Quick-Start Guide For IO Interface

This quick start guide provides instructions for mounting and connecting an I/O1000 I/O module. The guide instructs the user to mount the module on a DIN-TS35 top hat rail and connect it to a mass spectrometer module via a data cable. It also describes how to connect the I/O module to an external controller via various interface options like RS-232, RS-485, analog inputs/outputs, and digital inputs/outputs. The rear view of the I/O module is shown, identifying the connection ports.

Uploaded by

Selva Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
74 views

Leybold Quick-Start Guide For IO Interface

This quick start guide provides instructions for mounting and connecting an I/O1000 I/O module. The guide instructs the user to mount the module on a DIN-TS35 top hat rail and connect it to a mass spectrometer module via a data cable. It also describes how to connect the I/O module to an external controller via various interface options like RS-232, RS-485, analog inputs/outputs, and digital inputs/outputs. The rear view of the I/O module is shown, identifying the connection ports.

Uploaded by

Selva Kumar
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Quick-start installation guide for I/O1000 I/O module

Safety instructions
This quick-start guide does not replace the – Generally applicable safety and accident
installation manual for the I/O1000 I/O module. prevention regulations
Safe use of the module is ensured only by reading – International, national and local standards and
the installation manual. guidelines
Additional important information is provided in
– Additional provisions and regulations that are
the installation manual for the I/O1000 I/O module.
specific to the unit
The manual can be found either on the USB
memory stick or on the INFICON website. The USB ► Use only original parts or parts approved by the
memory stick is included with the shipment of the manufacturer.
I/O module. ► Keep the installation manual available at the
operating site.

Intended use ► All work must be performed only by technical


specialists who have been trained on the unit.
The I/O module is a device interface between the MSB
box of the mass spectrometer module LDS3000 and ► Allow personnel in training to work with the unit
an external controller, for example. only under the supervision of technical specialists.

► Install, operate and service the unit only in ► Make sure that the authorized personnel have read
compliance with the installation manual. and understood this manual and all other
applicable documents, especially the information
► Comply with the limits of application. on safety, maintenance and repairs, before starting
work.

User requirements ► Define responsibilities, authorizations and


supervision of personnel.
► Operate and install the unit only if it is in perfect
working order and as intended, in a safety-
conscious manner and fully aware of dangers, in User requirements
compliance with this manual.
► Read, observe and follow the information in the
► Fulfill and ensure compliance with the following installation manual and the working instructions
regulations: created by the owner, especially the safety
– Intended use instructions and warnings.
► Perform all work based on the complete manual.

Document: jima10en1-03 (1503)


Mounting of I/O module  Data cables from INFICON

1 Connect I/O module (connection LD) via data cable


with leak detector.
2 Connect I/O module via desired interfaces with
external controller:
• RS232 (RS-232 interface)
• RS485 (RS-485 interface)
• ANALOG IN (analog input)
• ANALOG OUT (analog outputs)
• PLC IN (digital inputs)
• PLC OUT (digital outputs)

Description of the layout


Fig: 1 Mount I/O module on DIN-TS35 top hat rail see Installation Manual for I/O module, jiqc10.

 DIN-TS35 top hat rail


FUSE and DIP switch S1, S2 f
1 Hook unit on top hat rail at bottom. Fuses for digital outputs and DIP switches (under the
2 Press unit onto top hat rail at top. cover)
Fuses for digital outputs 1 ... 4 and 5 ... 8:
2 x 0.75 A (Schurter: 7010.9800.xx)
Connecting I/O module with leak detector
LDS HLD Contact
DIP switch S1
3000 6000 4 3 2 1
Factory setting (default 0 0 0 0
value of the interface
 
protocol by the leak
detector or control unit)
ASCII protocol   0 0 1 0
LD protocol   0 0 1 1
Binary protocol  0 1 0 1
LDS1000 protocol  0 1 1 0
Normal protocol  0 0 0 1
Simple protocol  0 1 0 0

1 = ON, 0 = OFF
Fig: 2 Rear view of I/O module
Contact
1 PLC IN 6 S1, S2 and FUSE DIP switch S2
2 RS232 7 LD 4 3 2 1
3 STATUS 8 ANALOG IN Activate boot mode for software update X + 0 0
4 ANALOG OUT 9 RS485
Disable bus terminator 120 Ω for RS-485 1 X 0 0
5 PLC OUT 10 24V OUT
The I/O module communicates via data cable with the 1 = ON, 0 = OFF, + = switching from OFF to ON during
leak detectorand is supplied with voltage by the data operation,
cable. X = random

INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany

Vi sit o ur we bsi t e fo r con t act i nfo rm ati on an d o t her sal es o ffice s wo rl dw id e. w w w . i n f i c o n . c o m