SG 003a Multi Functional Signal Generator Manual

This document provides an instruction manual for the FNIRSI SG-003A Multi Functional Signal Generator. The manual describes the safety instructions, product description including specifications and functions, button panel and display interfaces, and signal descriptions for the generator. It allows users to output different signal types including current, voltage, passive current, and pulse signals and switch between input and output modes.

Uploaded by

bryanvelesaca
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
493 views17 pages

SG 003a Multi Functional Signal Generator Manual

This document provides an instruction manual for the FNIRSI SG-003A Multi Functional Signal Generator. The manual describes the safety instructions, product description including specifications and functions, button panel and display interfaces, and signal descriptions for the generator. It allows users to output different signal types including current, voltage, passive current, and pulse signals and switch between input and output modes.

Uploaded by

bryanvelesaca
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 17

Manuals+ — User Manuals Simplified.

FNIRSI SG 003A Multi Functional Signal Generator Instruction


Manual

Home » FNIRSI » FNIRSI SG 003A Multi Functional Signal Generator Instruction Manual

SG-003A
OPERATING MANUAL
Multi-functional Signal Generator
Contents
1 SAFETY INSTRUCTIONS
2 PRODUCT DESCRIPTION
3 BUTTON PANEL INTRODUCTION
4 DISPLAY & SIGNAL DESCRIPTION
5 EXTENSIONS
6 WIRING INSTRUCTIONS
7 FAULT RESOLUTION & EQUIPMENT
MAINTENANCE
8 PRODUCTION INFORMATION
9 Documents / Resources
9.1 References
10 Related Posts

SAFETY INSTRUCTIONS

Plug the meter into the PC, and turn it on. Wait for the PC to recognize the USB drive named “Bootloader”
Don’t use the instrument in a flammable and explosive environment
Used batteries for instrument replacement and discarded instruments cannot be disposed of with household
waste. Please handle in accordance with relevant national or local laws and regulations
When there is any quality problem with the instrument or when there is a question about using the instrument.
You can contact “Finiris” online customer service or the manufacturer, we will solve it for you in the first time

PRODUCT DESCRIPTION

2.1 Specifications

S
i
g Range Precision Resolution Max load External power supply
n
al
A
ct
iv
e
c
u
rr
e
nt 0∼24mA ± (0.1%+0.005) 0.01mA 7500Ω
(
m
A
)
o
ut
p
ut

P
a
s
si
v
e
c
u
rr
e
0∼ 24mA ± (0.1%+0.005) 0.01mA 0∼ 30V
nt
(
X
M
T
)
o
ut
p
ut

V
ol
ta
g
e
(
0∼ 24V ± (0.1%+0.005) 0.01V
V
)
o
ut
p
ut

2
4
V
0∼ 24mA ± (0.1%+0.005) 0.01mA
lo
o
p
F
r
e
q
u
e
n
c
y 0∼ 9999Hz ±2% 5 digits
(
H
z)
o
ut
p
ut

C ± (0.1%+0.005)
u
rr
e
nt
(
0∼ 24mA 0.01mA
m
A
)i
n
p
ut

V
ol
ta
g
e
( 0∼ 30V ± (0.1%+0.005) 0.01V
V
)i
n
p
ut

2.2 Function Introduction

Conversion Of Engineering Quantity Display : Signals are displayed as physical or engineering (ie pure
quantity) values, The physical quantity corresponds linearly with the engineering quantity.For example, the
current signal of 4~20mA corresponds to 0~100, that is, 4mA corresponds to 0, 12mA corresponds to 50, and
20mA corresponds to 100. This range can be modified by parameters
Preset Value : 4 groups of commonly used signal values can be preset to facilitate quick adjustment of
output
Signal Conversion : The input signal is converted into an output signal, which can be converted between
different physical quantities; for example the input frequency signal is converted into a current signal, and the
input range is 0-1000hz into 4-20mA
Programming Output : It has the output function of automatically and continuously changing the size
according to the set parameters, and can choose three modes: monotonous rise, monotonous fall, and cycle
Real-time Curve : the changing trend of input and output signals can be displayed by drawing a curve, and
the sampling time of the curve can be set. It can also automatically adjust the display curve within the range of
the maximum and minimum values within a certain period of time
Modbus Slave : FNIRSI SG-003A can communicate with configuration software or PLC via RS485, set the
instrument
Firmware Upgrade : USB emulated U disk for firmware upgrade

2.3 Interface Terminal

There are 4 inner diameter 4mm wiring ports on the top surface, which are represented by different color rings
The blue port (IN-) is an alternate function, and it is also an input-, which can be switched to a 24V independent
test power supply. The (IN-) port is internally shorted with the black (COM) port in the input state. When the (IN-
) port is an independent power supply, there is a 24V voltage between the (COM) port
The Red (OUT) port and black (COM) port form an output port
Yellow (IN+) and blue (IN-) form the input port
When the (IN-) port is an independent power supply, it can be called a (24V) port, and it forms an output port
pair with the (COM) port

NOTE
The safety limit voltage of the input terminal cannot be greater than 30Vpp, and the over limit will damage the
equipment

2.4 Power & Charge


Built-in battery: 3000mA large capacity lithium battery
Battery display: Displays the current battery level in the upper right corner of the display
Charging port: Type-C port (DC 5V)
State of charge: When charging, the power button will display a red light.When fully charged, it will
automatically stop charging and display a blue light

BUTTON PANEL INTRODUCTION

All buttons are made of white translucent silicone material. The color LED backlight is set under the Button,
which can make the key surface change color with the change of function and mode. Give a more intuitive
status indication
The keys are mainly divided into 3-Key area and 5-Key area , and each key has multiple functions. Among
them, the FN Button and the 5-Key area key have the up/down function and the function marked on the
key, which will be changed according to the state of the FN key and the interface mode. Start-up defaults to
the upshift function
FN key state is divided into the default white light state and red light state
5-Key area The status is divided into Button white light (Manual setting mode); Button green light (Preset
output mode); Button blue light (Programming mode); No
button light (Conversion mode) Button Function Operation Description:

1. RUN Light: The power on/on output is white light


2. IN Input key: Used to switch the input signal mode in the main interface; When FN red light is on, click IN
to enter the input setting
3. OUT Output key: used to switch the output signal mode on the main interface. When the FN red light is on,
click FN to enter the output setting FN Exit Return key:In the main interface, click
4. FN , it will turn into red light to activate the downshift function of other keys (enter different modes, etc.).In
the setting interface, the function of FN is to return
5. Main Interface : Main display interface RCVK Up function key: When the key is white, it is to add value or
select upward (when the output pulse is quantitative and the cursor is not in, it is the same as the programming
mode).When the FN red light is on (the button is green), click
6. RCVK to enter/exit the preset output mode. At this time, the up, down, left and right buttons can quickly recall
the preset signal value
7. AUTO Left function key: when the key is white, it is to select to the left (when the output pulse is quantitative
and the cursor is not present, it is the same as the programming mode).When the FN red light, click AUTO
to enter/exit the programming mode (the button blue light), at this time, the up key-single output, the down key-
reset, the left key-end the output, the right key-pause/ continue
8. ON/OFF Confirmation button: when the main interface (Turn on/off the output); when setting the
interface(Click for “OK”)
9. CURVE Right function key: When the key is white light (select to the right ~ when the output pulse is
quantitative and the cursor is not in, it is the same as the programming mode).When FN is red light (click to
switch curve interface/character interface)
10. CONVRET Down function key: When the key is white, it is to decrement the value or select down (when the
output pulse is quantitative and the cursor is not in, it is the same as the programming mode; When FN is red
light (click to enter/exit the conversion mode)
DISPLAY & SIGNAL DESCRIPTION

4.1 Switching Of Signal Types

Press FN on the main interface. After the FN button turns red, if you want to switch the input signal type,
press the IN button to enter the input setting
If you want to switch the output signal type, press OUT key to enter the output setting. After entering the
setting interface, you can directly select the desired signal
type from left to right.Press FN to exit to the main interface
There are also a variety of signal modes available for switching under various signal types. Various signal types
also have different parameters that can be set

4.2 mA Current Signal

Mode Switch : When the main interface FN is white, press IN/OUT to switch the display of input/output
actual value and engineering quantity
Output Open Circuit Detection : When the output is turned on, if the interface displays an open circuit and
flashes, it means that the output signal is not connected
or the output is close to zero
Output Circuit Voltage FNIRSI SG-003A can set current output no-load voltage. The purpose of 12V is to
save power. Some devices require a 24V drive capability
Output Range : The limit range of the current output can be set

4.3 V Voltage Signal

Mode Switch : When the main interface FN is white.Press IN/OUT to switch the display of actual value
and engineering quantity of input/output
Overcurrent Protection : If the output voltage is shorted, the output will automatically shut down
Output Range : The limit range of the voltage output can be set

4.4 XMT Passive Current Signal


Mode Switch :When the main interface FN is white.Press IN/OUT to switch the display of actual value
and engineering quantity of input/output
Output Range : The limit range of passive current output can be set

4.5 Hz Pulse Signal

Mode : The pulse signal has four modes: frequency, PWM duty cycle, speed frequency, and pulse
quantitative/counting. When the main interface FN is white, long press OUT to switch mode
Range : Select the range of frequency output: 0~99.999Hz, 0~999.99Hz, 0~9999Hz, 0~200kHz
Method : Select level/open circuit according to the presence or absence of pull-up
Output Peak Value : The level height of the output waveform, the highest is 24V
Frequency Mode Operation Instructions : The frequency range can be switched by pressing OUT when the
main interface FN is white
Operating Instructions In PWM Mode : PWM displays two parameters, the frequency value (that is, the period
value) and the duty cycle. The frequency value will be displayed in the original engineering quantity position
(no engineering quantity display). Note that the maximum frequency does not exceed 9999Hz. The frequency
value/duty cycle parameters can be set when outputting. When the main interface FN is white, press OUT
to switch, and there is a yellow cursor indication
Operating Instructions For Speed Mode

1. The speed mode is used to calculate the speed of the motor or gear. One revolution of the motor or gear
produces 1 pulse, one minute is 3000 revolutions, so 1 second is 50 pulses
2. Time Unit : Frequency=RPS(per second)*a number of pulses per revolution, RPM(per minute)=RPS*60,
RPH(per hour)=RPS*3600.In the speed setting selection, note that the maximum frequency does not exceed
9999Hz
3. Number Of Revolution Pulses : The number of pulses per revolution of the motor or gear, which can be
selected in the speed setting
4. Tooth Missing : It is mainly used to simulate the engine speed, select it in the speed setting

Quantitative/Counting Mode Operation Instructions : The main purpose of pulse quantitative is to simulate
the quantitative control sensor of flow.the frequency value will be displayed in the original quantity position. The
cycle is displayed at the original engineering position, and the unit is milliseconds. When the main interface
FN is white, press OUT to switch between cycle and quantitative values, which are distinguished by cursors
The period value can be modified during quantitative output.In this mode, when the cursor disappears after the
output is turned on, the backlight of the up, down, left, and right keys will turn blue. The function of the up,
down, left, and right buttons will also change, that is, the up is a single pulse sent. Down is program stop and
reset, left is no function, right is start/pause

4.6 24V Loop Detection


Select the 24V label left and right in the OUT signal type interface to enter this function state. In this state,
displays the output 24V loop current on the main interface. Unlimited input signal detection type

1. Loop 1: After starting the output, 24V voltage is output between the (OUT) port and (COM) port, and the loop
current is detected at the same time, and the current is displayed in the “output 24V loop” area. This function
can be used to test, for example, pneumatic control valves or two-wire transmitters, supply them with a 24V
power supply and detect current
2. Loop 2: After activating the independent 24V power supply. The (IN-) port outputs 24V voltage between the
(24V) port and the (COM) port.It does not affect the detection type of the input signal by the (IN+) port. In this
case, select mA for the input signal type. Then the (IN-) port can be combined with the (IN+) port to form a 24V
output, And at the same time detect the current loop. The function is similar to the circuit one
3. Port (IN-) can be switched to an independent 24V power supply function. It outputs a single DC24V voltage,
which is the (24V) port at this time.(24V) port is used with (IN+) to test two-wire transmitters

Notice
①The difference between 24V loop detection and standalone 24V is:24V loop detection output signal port to
measure feedback current.
②Independent 24V port and output port can form a 24V power supply,It can also be combined with the input (+) to
form a transmitter to supply power and measure the feedback current of the transmitter

Mode switch : When the main interface FN is white.Press OUT to switch the display position of output
actual loop current value and engineering quantity
Activate independent 24V : After this item is checked, the independent 24V power supply starts to output
from the (24V) port (IN-) port and the (COM) port, and is not affected by input and output signal type switching
Activate Hold : After this item is checked, each time the machine is turned on, the 24V state will remain in the
state of the last turn-on. Otherwise, every time you turn on the 24V, you need to turn it on manually

EXTENSIONS

5.1 Engineering

When the signal value of an analog quantity is used in the industrial field, it is generally necessary to convert it
into engineering quantity value. Therefore, in the input and output settings of each signal, the instrument can
set the upper and lower limits of its engineering quantity and the upper and lower limits corresponding to the
signal value

5.2 Preset Value

Parameter Setting : You can set the signal value corresponding to the up, down, left, and right buttons
Operation Instructions : Start and Exit: When the main interface FN is red, press the up key RCVK to
Start/Exit the preset
Up, Down, Left, And Right Keys : After starting the preset, the backlight of the up, down, left and right keys
will turn green, and the key function will also change
to “output the set value”

5.3 Signal Conversion

The input signal is converted into an output signal.For example, the input frequency signal is converted into a
current signal, and the input range is 0-1000hz and converted into 4-20mA
Parameter Settings :
1. Input upper limit/input lower limit: the amplitude range of the input signal
2. Output upper limit/output lower limit: the amplitude range of the output signal
3. Overrun allowed: For example, 1-5V is converted into 4-20ma. If the user checks this option, when the input is
on, the output will follow to OmA. If the user does not
check it, then the output is 4mA and the output is locked within the set range
4. Automatic operation: After checking this option, the signal conversion will be automatically started at the next
power-on, and the signal conversion will be automatically canceled when the signal conversion is exited

Instructions :
Start and Exit : When the main interface FN is red, press the key CONVERT to start/exit the signal
conversion mode. After startup, the backlight of the up, down, left and right keys will turn off and have no
function

5.4 Program Output

The programming output can automatically complete N cycles of signal output according to the parameters set
by the user. This function is mainly used for the aging test of electric valve or pneumatic valve, Or test work
such as PLC program debugging

Parameter Settings :

1. Mode: It is divided into three modes: single rise, single fall, and cycle
2. Number of loops: 0 is infinite, 30000 is the maximum
3. Start value: The value at each start
4. End value: The value at the end of each
5. Increment step value: The value of each increment
6. Increment step time: The time between each increment
7. End stop: The time delay at the end value
8. Decrement step value: The value to decrease each time
9. Decrease step time: The time between each decrease
10. Loop stop: Time delay at the start value

Supplementary Parameter Description :


The start value and end value of single rise and single fall are selected according to
high and low, and does not automatically close
Instructions :
1 Start and Exit: When the main interface FN is red, press the left key AUTO to start/exit the programming
output mode
2 Up, Down, Left, and Right keys: After starting the programming output mode, the backlight of the up, down,
left and right keys will turn blue. The functions of the up, down, left and right buttons will also change, that is,
the top is a single run, the bottom is a program reset, the left is a single-step run, and the right is a
Stop/Continue

5.5 Real-time Curve


Instructions
When the main interface FN is red, press the right button CURVE to switch the
interface with/without curve
Parameter Description :

1. Turn on the output curve: turn on the following display curve of the output signal (Redline)
2. Automatic output scaling: automatically adjust the scale range of the output signal, and automatically adjust the
scale range of the y-axis with the maximum and minimum values of the curve
3. Input automatic scaling: automatically adjust the scale range of the input signal, and automatically adjust the
scale range of the y-axis according to the maximum and minimum values of the curve.
4. Acquisition time: the refresh interval of the curve, and the time scale will also change with it

5.6 Output Time

Output Activation Output Time : When checked, each output will automatically turn off the output after the
output time expires
Output Time : Set output time

5.7 Communication

USB connection : Because the USB analog serial port is used for communication. Note that charging will slow
down after checking the USB connection
Slave address : MODBUS slave address. Please refer to the attached tables 1, 2, and 3 on page 14 for the
address and description of the instrument register

WIRING INSTRUCTIONS

All Switched by Electronic Switches


All-round Signal Output :
Current/Voltage/Frequency
All Switched by Electronic Switches
All-round Signal Output :
Current/Voltage/Frequency

witch this input (-), the port to 24V, as the power supply for the valve feedback, the input and output are carried out
at the same time

This input (-), the port is closed to 24V, as the negative pole of the input signal, the input and output are performed
at the same time

FAULT RESOLUTION & EQUIPMENT MAINTENANCE

7.1 Fault Resolution

No response when SG-003A is turned on

1. Determine if the battery has power and if the indicator light is normal when charging
2. Firmware upgrade error operation leads to system crash
Abnormal screen display Check if the screen is connected properly.
Abnormal output value or measured value

1. Check the signal line


2. Check the interface for signal selection
3. The target parameter of the connection exceeds the range indicated by the technical indicator? It may cause
the instrument to work abnormally or even damage

Error in input and output It is normal that there may be errors in input and output.Because the input and output
are not isolated

7.2 Equipment Maintenance

The meter is powered by 3.7V rechargeable lithium battery, please use the power adapter to extend battery life
when using it for a long time
This instrument is not waterproof, please do not use it in a high humidity environment
Do not place the unit in a place that is unstable or may be subject to strong vibrations
Do not place this unit in places with high humidity, dust, direct sunlight, outdoors, or near high temperatures

7.3 Firmware Upgrade

SG-003A uses USB emulated U disk for firmware upgrade


Upgrade Steps

1. Plug the meter into the PC, and turn it on. Wait for the PC to recognize the Udrive named “Bootloader”
2. Copy the firmware to be upgraded to the U disk
3. After a few seconds, the device is powered on normally and the upgrade is successful

Notice

1. If the U disk cannot be recognized normally when connected to the PC in the power-on state. At this point,
please shut down and restart!
2. USB firmware upgrade only supports WIN10. It is forbidden to drag and drop files other than those officially
released by FIRST. Otherwise, it is very likely to cause irreversible consequences

PRODUCTION INFORMATION

Product name: Multi-functional Signal Generator


Brand / Model: FNIRSI / SG-003A
Manufacturer: Shenzhen FNIRSI Technology Co., Ltd.
Website: www.first.cn
Address: Building C, Weihuada Industrial Park, Dalang Street, Longhua District, Shenzhen, Guangdong, China
Function code Hexadecimal Type of Data Attributes

100 0x64 uint16_t read

101 0x65 uint16_t write

102 0x66 float read

103 0x67 float write

Table 1

Address Hexadecimal Type of Data Attributes Illustrate

40001 0X9C41 uint16_t read Firmware version

40002 0X9C42 uintl6_t RW Input signal

40003 0X9C43 uintl6_t RW Output signal

40004 0X9C44 float read Input value

40006 0X9C46 float RW Output value

40008 0X9c48 uintl6_t RW Software output mode and output switch

40009 0X9C49 uintl6_t RW Active current output value USER upper limit

40010 OX9C4A uintl6_t RW Active current output value USER lower limit

40011 OX9C4B uintl6_t RW Voltage output value USER upper limit

40012 OX9C4C uintl6_t RW Voltage output value USER lower limit

40013 OX9C4D uintl6_t RW Passive current output value USER upper limit

40014 OX9C4E uintl6_t RW Passive current output value USER lower limit

Table 2

Enter

Register upper 8 bits 01 is current; 02 is voltage; 03 is frequency; 04 is millivolt; 05 is resistance

The upper four bits are the type: 1-8 represent S, B, E, K, R, J, T, and N in turn
Register lower 8 bits
Lower four-bit mode: 1 is my, 2 is a thermocouple, 3 is WR thermocouple

Table 3
[Example Write input millivolt signal, J thermocouple type

9
01 65 42 04 62 40 AF
C

r
e
g
is
t
e
CRC high CRC lo CR
slave function r register address Register number Register number
w C lo
address code a low byte high byte low byte
byte byte w
d
d
r
e
s
s

01: Slave address set by the device; 65: Write function code, see Table 1; 9C42: Read and write input sig
nal address, see Table 2; 04: 01 is current, 02 is voltage, 03 is frequency, and 04 is
Illustrate millivolt, 05 is the resistor; 62: the upper four bits are the type: 1 is S, 2 is B, 3 is E, 4 is K, 5 is
R, 6
is J, 7 is T, 8 is N, the lower four bits mode: 1 is millivolt, 2 is a thermocouple, 3 is WR thermocouple;
40AF: CRC code (note! Send in hex)

Download Manuals
and Firmware

Manufacturer: Shenzhen FNIRSI Technology Co., Ltd.


Website: www.first.cn
Address: Building C, Weihuada Industrial Park, Dalang Street, Longhua District, Shenzhen, Guangdong, China

Documents / Resources

FNIRSI SG-003A Multi-Functional Signal Generator [pdf] Instruction Manual