TB194.LISRelease of LIS-Link 2.01
TB194.LISRelease of LIS-Link 2.01
Company Confidential
Description
DYNEX LIS-Link Software provides a data communication interface between the Revelation DSX or DS-
Matrix software and an LIS system. The data transmitted to and from details test orders for the DSX or
DS2 to run and the results of the test orders after they have completed.
Revelation DSX software version 6.15 is required to send Standards and Controls to the LIMS host.
Earlier Revelation DSX versions will work with Lis-Link 2.01 but Standards and Controls will not be sent to
the host.
Matrix software has not yet been modified to send Standards and Controls. Matrix will work with Lis-Link
2.01 the same as it works with Lis-Link 1.08 and earlier.
Distribution
LIS-Link Software Version 2.01 has been developed and tested by R&D.
Purpose
Version 2.01 of LIS-Link is focused on the communication of assay data for assays using Standard and
Control result data. If the instrument returns Standard or Control results data, Lis-Link 2.01 automatically
generates data records for the Standard and Control data. These auto-generated records are sent to the
Host when the test sample results are sent to the Host.
There is no change to how the Host sends order requests to the instrument, or how sample results are
transmitted to the Host.
• An Advanced Serial transmission debugging mode can be enabled via the configuration file (please
contact Dynex if you need help enabling this feature—it is disabled by default)
This shows a typical transaction where Dynex asks for work orders, completes the orders, and returns
results to the LIMS Host. There is no change in the data flow Host Lis-Link DSX/DS2, or in how
results for test specimens are returned. What’s new here are the returned controls and standards data
values.
H|\^&|||DYNEXRD-LT||||||||1|20101208184124
L|1|N
H|\^&|||DYNEXRD-LT||||||||1|20101208184458|
P|1||Sample Batch 56-1||
O|1|Sample Batch 56-1||^^^StdsCtrls|R|20040205151351
Sometime later Dynex completes the assay and returns results to the Host.
In this case the specimen data comes first. There is no change to specimen data transfer.
H|\^&|||DYNEXRD-LT||||||||1|20101208190753
P|1||Sample Batch 56-1
O|1|Sample Batch 56-1||^^^StdsCtrls|R|20101208184500
R|1|^^^StdsCtrls|0.417 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
P|5||A1_S1_StdsCtrls_Fred_20101208190720
O|1|A1_S1_StdsCtrls_Fred_20101208190720||^^^StdsCtrls|R|20101208190725
R|1|^^^StdsCtrls|0.001 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
P|6||C1_S2_StdsCtrls_Fred_20101208190720
O|1|C1_S2_StdsCtrls_Fred_20101208190720||^^^StdsCtrls|R|20101208190725
R|1|^^^StdsCtrls|3.100 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
P|7||E1_S3_StdsCtrls_Fred_20101208190720
O|1|E1_S3_StdsCtrls_Fred_20101208190720||^^^StdsCtrls|R|20101208190725
R|1|^^^StdsCtrls|6.250 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
P|8||G1_S4_StdsCtrls_Fred_20101208190720
O|1|G1_S4_StdsCtrls_Fred_20101208190720||^^^StdsCtrls|R|20101208190725
R|1|^^^StdsCtrls|12.500 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
P|9||A2_S5_StdsCtrls_Fred_20101208190720
O|1|A2_S5_StdsCtrls_Fred_20101208190720||^^^StdsCtrls|R|20101208190725
R|1|^^^StdsCtrls|25.000 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
P|10||C2_S6_StdsCtrls_Fred_20101208190720
O|1|C2_S6_StdsCtrls_Fred_20101208190720||^^^StdsCtrls|R|20101208190725
R|1|^^^StdsCtrls|50.000 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
P|11||E2_S7_StdsCtrls_Fred_20101208190720
O|1|E2_S7_StdsCtrls_Fred_20101208190720||^^^StdsCtrls|R|20101208190725
R|1|^^^StdsCtrls|100.000 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
P|12||G2_PC1_StdsCtrls_Fred_20101208190720
O|1|G2_PC1_StdsCtrls_Fred_20101208190720||^^^StdsCtrls|R|20101208190725
R|1|^^^StdsCtrls|161.537 U/ml|U/ml||FR||F||||20101208190753|DYNEXRD-LT
L|1|N