Alera Operator Manual Rev 1.2
Alera Operator Manual Rev 1.2
Operator’s Manual
Company Locations
USA
Alfa Wassermann Diagnostic Technologies, LLC.
4 Henderson Drive, West Caldwell, NJ 07006
Phone: +1-973-882-8630 or 1-800-220-4488
Fax: +1-973-276-0383
web site: www.AlfaWassermannUS.com
International
Alfa Wassermann B.V.
Pompmolenlaan 24, 3447 GK Woerden, The Netherlands
Phone: +31 348 487 300
Fax: +31 348 433 000
email: [email protected]
2 DESCRIPTION ........................................................................................................ 3
2.1 Analysis Module........................................................................................... 4
2.1.1 Fluid Transport System .................................................................... 4
2.1.2 Sample/Reagent Module .................................................................. 5
2.1.3 Sample Tube Entry Port (STEP)....................................................... 6
2.1.4 Syringe Module ................................................................................ 6
2.1.5 Temperature Control System............................................................ 7
2.1.6 Reaction Wheel................................................................................ 8
2.1.7 Cuvette Hopper ................................................................................ 8
2.1.8 Optical System ................................................................................. 8
2.1.9 Liquid Level Sensing System............................................................ 9
2.1.10 System Sensors ............................................................................. 9
2.1.11 CPU Interface Board ...................................................................... 9
2.1.12 Motors .......................................................................................... 10
2.1.13 ISE Module (Optional) .................................................................. 11
2.2 Uninteruptible Power Supply/Line Conditioner ........................................... 11
2.3 Consumables............................................................................................. 12
2.3.1 Cuvettes......................................................................................... 12
2.3.2 Seglets ........................................................................................... 13
2.3.3 Sample Cups.................................................................................. 14
2.4 Reagents, Calibrators and Controls ........................................................... 15
2.4.1 Chemistry Reagents....................................................................... 15
2.4.2 Chemistry Calibrators..................................................................... 15
2.4.3 Controls.......................................................................................... 15
2.4.4 ISE Reagents and Calibrators ........................................................ 16
2.4.5 System Diluent ............................................................................... 16
2.5 Integrated Computer Workstation............................................................... 17
2.5.1 Computer ....................................................................................... 17
2.5.2 Keyboard........................................................................................ 17
2.5.3 Printer ............................................................................................ 18
2.5.4 LCD Monitor................................................................................... 19
2.6 Menus........................................................................................................ 20
2.6.1 Selecting a Menu Command .......................................................... 20
2.6.2 Navigating Through Menus ............................................................ 20
2.6.3 Main Menu ..................................................................................... 21
2.6.4 Requisitions.................................................................................... 22
2.6.5 Reports .......................................................................................... 23
2.6.6 Cup/Segment Placement................................................................ 24
2.6.7 Journals ......................................................................................... 24
2.6.8 Delete/Archive................................................................................ 25
2.6.9 Utilities ........................................................................................... 25
2.6.10 Diagnostics................................................................................... 26
3 PRINCIPLES OF OPERATION..............................................................................35
3.1 Operational Sequence Overview ................................................................35
3.1.1 Task Scheduling .............................................................................35
3.1.2 Priority of Instrument Activities........................................................36
3.1.3 Analysis Sequence .........................................................................37
3.1.4 Monitoring System Resources ........................................................37
3.1.5 Calibration of Photometric Tests .....................................................37
3.1.6 Calibration of Potentiometric (ISE) Tests ........................................38
3.1.7 Curve Correction.............................................................................39
3.1.8 Calibrator Averaging .......................................................................39
3.1.9 Response Calculation.....................................................................40
3.1.10 Result Calculation.........................................................................40
3.1.11 Routine Chemistry Equations........................................................41
3.1.12 Running Tests ..............................................................................43
3.1.13 Integrity Checks............................................................................43
3.1.14 Automatic Sample Rerun ..............................................................46
3.1.15 Result Correlation.........................................................................46
3.1.16 Reporting Results .........................................................................47
3.2 Operational Precautions .............................................................................48
3.2.1 General Precautions .......................................................................48
3.2.2 Additional ISE Module Precautions.................................................49
4 SYSTEM SETUP....................................................................................................51
4.1 Materials.....................................................................................................51
4.1.1 Materials Provided ..........................................................................51
4.1.2 ISE Parts ........................................................................................52
4.1.3 Materials Provided at Installation ....................................................52
4.2 Positioning and Connecting the Modules....................................................53
4.2.1 Site Selection..................................................................................53
4.2.2 Physical Dimensions and Space Requirements..............................54
4.2.3 Electrical Service ............................................................................55
4.2.4 Positioning of Modules....................................................................55
4.2.5 Connecting the Modules .................................................................55
4.3 Software Installation ...................................................................................56
4.3.1 Re-Loading the Operating Software................................................56
4.3.2 Re-Loading the Setup Information ..................................................57
4.4 Container Installation..................................................................................57
WARNING: The front door of the ACE Alera System must be kept closed
during operation.
Convention Purpose
Numbered Lists Describes each step that must be carried out in the
order given when performing an operation:
Symbol Purpose
?
(Information note) Designates background or additional information
that is useful in understanding the procedure or topic.
BIOBAGS Disposable plastic bags displaying a biohazard label used to line the cuvette waste
box. The bag allows removal of used cuvettes from the box without direct handling
by the operator.
CAL-A, CAL-B Solutions that are used for calibrating the electrodes in the optional ISE module.
CAL-A Cap A cap and tubing assembly that is attached to the CAL-A container before it is
Assembly placed in the container compartment. The tubing contains an in-line filter.
Cup Segment A plastic device that can hold up to 15 sample cups. Samples are loaded into the
instrument by placing the segment on posts along the outer edge of the
sample/reagent tray.
Cuvette Specially designed container of optical quality plastic. Sample and reagent are
dispensed into a cuvette by the instrument in order to perform a photometric test.
Cuvette A plastic sleeve that holds 20 cuvettes. The cartridge facilitates loading the cuvettes
Cartridge into the cuvette hopper.
Cuvette Hopper A compartment in the analysis module used to store up to 200 cuvettes. Cuvettes
are added to the hopper through a slot in the top of the compartment.
Cuvette Waste Reusable metal container lined with a BIOBAG that receives used cuvettes expelled
Box from the reaction wheel.
Diluent Cap A cap and tubing assembly that is attached to the system diluent container before it
Assembly is placed in the container compartment. The tubing contains an in-line filter.
Fluid Arm Probe Stainless steel probe used to aspirate and dispense system fluids.
Assembly
Evap-Caps Adhesive backed covers that are used to minimize evaporation of reagents stored in
the sample/reagent tray.
ISE Module An optional module that measures sodium, potassium and chloride using
ion-selective electrodes.
LIMS Laboratory Information Management System (or LIS, for Laboratory Information
System). A computerized system used for the storage, sorting, and retrieval of
laboratory data.
Open Channel A reagent bottle that is defined by the user. An open channel bottle may be filled
Bottle with a reagent obtained from a different manufacturer. Parameters must be defined
for this reagent bottle before it can be used in testing.
Reaction Wheel The circular device inside the analysis module that holds the cuvettes used in
testing. The reaction wheel holds 48 cuvettes.
Sample/Reagent A tray located within the analysis module used to hold up to five sample segments
Tray and 40 reagent bottles. Reagent bottles are placed in openings on the tray. Cup
segments and well segments are placed on posts along the outer edge of the tray.
Seglet A group of ten sample wells that are attached as a unit. A sample that is aspirated
via the STEP is dispensed into one of the wells. The wells are also used to hold
samples that are automatically diluted during testing.
STEP The STEP, or Sample Tube Entry Port, allows the user to load a patient sample into
the instrument directly from a primary blood collection tube via cap piercing
mechanism. The tube is placed into the STEP, lowered into the instrument, sampled
and returned. If the tube has a barcode label, an instrument equipped with an
optional barcode reader will read the label and associate the sample with the
corresponding requisition.
System Diluent Fluid used for rinsing the probe between reagent and sample transfers and diluting
samples.
UPS The UPS, or Uninterruptible Power Supply, provides the system with battery backup
for five minutes in the event of a power failure.
Waste Container Reusable plastic bottle used to receive waste fluid from the probe wash bath and
the ISE.
Well Segment A plastic device that holds three seglets totaling thirty sample wells. All three seglets
must be loaded onto a segment for proper operation. Well segments are used to
accept STEP samples or dilutions made by the system and are loaded onto the
instrument by placing on posts along the outer edge of the sample/reagent tray.
1 Introduction
Two types of sample segments are used by the system. Individual micro-sample cups
are used for samples that are poured or pipetted into sample cups. These cups are
placed in a 15-position cup segment, and the filled cup segment is loaded on the
system. Integral 30-position well segments are used for samples that are aspirated from
open and closed primary tubes and for automatic dilutions of samples. The combination
of cup segments and well segments used is based on the workload being processed,
and up to five segments can be placed on the system at one time.
During analysis, samples and reagents are pipetted into individual disposable cuvettes
contained on a 48-position reaction wheel. As each test is completed, the used cuvette
is ejected into a cuvette waste box and a new cuvette is loaded onto the reaction wheel
from the cuvette hopper. Since the reaction wheel contains 48 cuvettes and the cuvette
hopper contains 200 cuvettes, the system can run 48 tests at one time and 248 tests
before the cuvette hopper must be reloaded.
The analysis module includes a holographic diffraction grating spectrophotometer to
measure absorbance at 16 different wavelengths. Readings for each test are taken at
wavelengths and times that are preset for the particular test being run. The test results
for the sample are then calculated as specified by the test parameters.
An optional ISE module allows simultaneous potentiometric measurements for sodium,
potassium and chloride to be obtained. Readings for each test are taken at times that
are preset for the tests being run. On board calibrators provide periodic automatic
updates of calibration slopes.
Many different reports are available. Each report can be viewed on the workstation
screen or it can be printed. The system also provides a means to transmit report
information to a laboratory information management system (LIMS).
2 Description
The ACE Alera System consists of an analysis module with an integrated computer
workstation (Figure 2). The integrated computer workstation includes a PC-compatible
computer, LCD monitor and keyboard. A printer and uninterruptible power supply
(UPS)/line conditioner are also provided.
Analysis
Module Printer
Cuvette Hopper
Keyboard Probe
Sample/Reagent Tray
ISE
The fluid transport system transfers measured amounts of sample and reagents to a
cuvette or to the ISE module.
The fluid transport system consists of the following components:
Component Function
Fluid Arm Transfers fluids and regulates fluid temperature.
Transfer Module Controls the vertical and rotational movement of the probe
and fluid arm.
Samples can be introduced to the system directly from primary tubes or from sample
cups that are filled by the operator. Primary tubes are sampled by placing them
individually into the Sample Tube Entry Port (or STEP). When the STEP is used,
sensors detect the presence of a tube and orient it to be punctured and scanned for a
barcode label if an optional barcode reader is installed.
The system detects the level of sample within the tube and withdraws sufficient sample
from the tube to complete all tests requested.
The syringe module aspirates and dispenses precision amounts of sample, reagent
and diluent.
The syringe module consists of these components:
Component Function
Syringe Aspirates and dispenses measured quantities of sample and
reagent up to 500 µL.
Analytical Chamber
The analytical chamber uses both heating and cooling functions to maintain the strict
temperature conditions (37 ?C ± 0.3 ?C) required for sample processing.
Reagent Compartment
The temperature within the reagent compartment is monitored so that reagents are
maintained at 8 ?C ± 1 ?C.
Fluid Arm
The fluid arm prewarms fluid as it transfers it from the cold reagent compartment to the
warm analytical chamber. The temperature is regulated by means of a 20 watt heater
that is set at 42 ?C ± 0.5 ?C and bonded to the bottom of the arm. The heater is
regulated by the CPU.
The reaction wheel (Figure 5) holds a total of 48 disposable cuvettes (see page 12).
The cuvettes are automatically loaded onto the reaction wheel from the cuvette hopper.
During operation, samples and reagents are dispensed into a cuvette and mixed. The
reaction wheel (and the cuvettes) rotate, and optical absorbance measurements are
taken as the cuvette passes through the measuring station. The wavelength(s) and
times of the absorbance readings that are used in the measurement are specified in the
parameters for that test.
After the measurements for a sample are completed, the used cuvette is ejected into the
cuvette waste box and a new cuvette is inserted into that position.
Cuvettes
The cuvette hopper contains up to 200 cuvettes. The cuvettes are added to the hopper
by the operator using cartridges (see page 12) containing twenty cuvettes. The
cartridges are loaded through a slot in the top of the hopper.
A pulsed xenon lamp is the light source for the optical system. A holograph diffraction
grating spectrophotometer is used to measure absorbance at 16 different wavelengths.
Readings for each test are taken at wavelengths and times that are preset for the
particular test being run.
When aspirating fluid, the probe receives a frequency signal from a conductive metal
plate located below the reagent bottle, segment cup or well, or in the STEP. This signal
is transferred to the fluid arm. The transfer arm CPU uses this signal to determine if the
probe is in fluid or not.
Reaction Wheel Positions the reaction wheel and times the pulsing of the
xenon lamp.
Sample Cup Detects whether sample cups and seglets are in the correct
positions in the segments.
A 80C196 microcontroller acts as the master controller of the ACE Alera System. It
controls communication between the instrument and the integrated computer
workstation.
2.1.12 Motors
Motion of the various components within the analyzer is powered by various motors:
Motor Function
Syringe Module Motor Controls movement of displacement syringe.
Cuvette Hopper Motors Load and move cuvettes within the Hopper and eject used
cuvettes.
Sample Tube Entry Port Move sample tube into position for sampling and return it to
(STEP) Motors be removed from the STEP.
The ISE module is an optional integrated module that measures sodium, potassium and
chloride using ion-selective electrodes. The ISE system consists of the following
components:
Valves Control the movement of fluid and air in and out of the
measuring chamber.
2.3 Consumables
Various consumables and reagents are required for day-to-day operation. These
products can be obtained from your Alfa Wassermann Diagnostic Technologies, LLC
representative.
2.3.1 Cuvettes
Disposable cuvettes are provided in cartridges of twenty cuvettes for easy loading into
the system (Figure 6). The cuvettes have an optical pathlength of 0.67 cm and are
specially constructed from optical quality plastic.
The cuvettes are stored in the cuvette hopper, which holds up to 200 individual cuvettes.
During operation, cuvettes are transferred from the cuvette hopper to the reaction wheel,
where the reagents and sample are added to the cuvette and the ensuing reaction is
measured. As each test is completed, the used cuvette is ejected into the cuvette waste
box and a new cuvette is loaded onto the reaction wheel from the cuvette hopper.
2.3.2 Seglets
Disposable seglets (Figure 7) are used for automatic dilution of samples and for holding
aliquots of samples that are aspirated from primary tubes. Each seglet contains ten
sample wells. Three clean seglets are loaded onto a well segment (see Figure 8) by the
operator when preparing for a run.
The well segment is positioned on the posts along the outside edge of the
sample/reagent tray. At least one complete well segment containing three clean seglets
should be loaded onto the system when preparing for a run.
Figure 7. Seglet
Sample cups contain samples that are individually loaded onto the system. The sample
cups are filled by pouring or pipetting an aliquot of the sample into the cup.
A cup segment (Figure 9), contains up to 15 sample cups on the system. When
preparing for a run, the operator fills a sample cup and places the cup into one of the
numbered positions on the cup segment following the loadlist. After the cup segment has
been filled with the cups for the run, the operator places the cup segment on the posts
along the outside edge of the sample/reagent tray.
Chemistry reagents are provided in large (30 mL) or small (12 mL) bottles that fit on the
system reagent tray. A computer-coded label (Figure 10) on the bottom of each bottle
provides automatic identification of the reagent, the lot number and sequence number.
The system must be calibrated before results for any test (except zero order kinetic
tests) can be obtained. Calibration is performed using calibrator(s) that contain a known
concentration of analyte(s) to be calibrated. The values that are obtained when running
calibrators are used by the system to calculate the results when patient samples are run.
Calibration must be repeated whenever a new lot of reagent is used and/or at intervals
that are defined for a particular test.
The number of calibration points (and therefore the number of calibrators) that are
required for a test depends upon the particular test. In some cases, a single calibrator is
needed. In other cases, up to six calibrators are used. Required calibrators are listed
automatically when test calibration is requested. Refer to page 120 for additional
information.
2.4.3 Controls
Controls are serum-based products that contain a known level of the analytes that are
being measured. Controls in several levels (normal, abnormal) of each analyte are
available. At least two levels of controls should be tested each day the instrument is run
to verify acceptable performance of the system.
Refer to pages 117 and 140 for additional information on the use of controls.
Separate reagents and calibrators are required for the ISE system:
?? Reference Electrode Solution. A concentrated solution of potassium
chloride that is used to provide a stable liquid junction potential for the ISE
reference electrode.
?? ISE Wash Solution. A solution used to flush the ISE sample port in order to
remove traces of protein or cleaning solutions.
?? ISE CAL-A. One of the two-level solutions used for calibration of sodium,
potassium and chloride measurements and single point calibrations. ISE
CAL A is stored in the system fluids compartment (Figure 29).
?? ISE CAL-B. One of the two-level solutions used for calibration of sodium,
potassium and chloride measurements.
?? Cleaning Solution. A solution used for periodic cleaning of the ISE
electrodes.
?? Conditioner. A solution used for periodic conditioning of the ISE
electrodes..
System diluent is used for rinsing the probe between reagent and sample transfers and
for diluting samples or a reaction mixture. System diluent is stored in the system fluids
compartment (see Figure 29).
2.5.1 Computer
The computer directs all system functions. It controls operation of the analysis module, it
provides an external disk drive (drive A) for loading programs and transferring
information, and it controls the user interface to the system.
2.5.2 Keyboard
F2 Choice List Displays selections available for a particular field. Note: the
prompt (F2) is displayed on the message line if selections are
available.
Home Home Returns the cursor to the first entry field of a screen.
Esc Escape Returns the display to a prior screen or menu level until the
Main menu is displayed.
Fn Function Enables the alternate functions that are printed on each key.
Esc
Key Function
Keys
LED Indicator
Lights
Ctrl Fn Alt
Key Key Key Enter
Key
Arrow
Keys
Figure 11. Function Keys and Other Special Keys on the Keyboard
2.5.3 Printer
The LCD monitor displays all user interface screens, menus, forms and messages. This
allows the operator to access system functions, enter information and view results.
Information displayed on the monitor is organized so that the same screen elements are
displayed in the same location on the screen (Figure 12). Status information such as
analyzer state, software version, date and time, and the system access level are always
displayed at the top of the screen. Menus and the various data entry and information
screens are displayed in the center portion of the screen. Finally, status indicators and
instructions concerning the highlighted command or field are displayed at the bottom of
the screen.
2.6 Menus
Menus provide the user with a means to perform the system functions.
?? Press the letter in the menu option that is highlighted and capitalized (e.g. the R
in Requisitions).
Selection of an item from a menu will display another menu in a cascading fashion, an
information screen or a data entry screen.
Press Esc to return to the prior menu. Continue to press Esc to return to the Main menu.
? Note: Menus may also be displayed when certain function keys are pressed.
Functions keys are discussed in Section 2.5.2.
The Main menu is displayed on startup. The Main menu is used to access menus for or
otherwise initiate all other system functions:
Command Purpose
Requisitions Create a patient, quality control, calibration, or correction test
request.
Exit to DOS Exits ACE Alera System software and displays the DOS prompt.
2.6.4 Requisitions
The Requisitions menu is used to create requisitions for patient, QC, calibration, and
correction samples.
Command Purpose
Patient Perform operations related to patient requisitions.
? Note: A sample cannot be loaded unless a requisition exists in the system for
that sample.
2.6.5 Reports
Quality control The last control run or statistical QC reports and plots.
Test counts The number of results obtained for each on-line photometric
test in the database since the date the counter was reset to
zero.
The Cup/Segment Placement option on the Main menu displays the Cup/Segment
Placement screen. This screen allows the operator to set up a group of up to 15 samples
in a cup segment prior to placing the segment on the system. The Cup/Segment
Placement screen will display a list of the current segments. If there are no segments
currently saved on the system, the list is blank.
2.6.7 Journals
The Journals menu allows the user to view the permanent records of various system
events.
Command Purpose
Audit Routine events such as the editing of test parameters and
the editing or deletion of results.
2.6.8 Delete/Archive
The Delete/Archive menu is used to remove a patient from the system and reset the test
counter.
Command Purpose
Delete/Archive Patients Removes and/or archives patients who have had no activity
since a specified date from the database.
Delete Test Counts Resets the test counter for on-line photometric tests to zero.
2.6.9 Utilities
Backup Copies all system data files onto a 3.5 inch disk.
2.6.10 Diagnostics
The user interface includes diagnostics to test instrument performance and function.
Command Purpose
Performance Evaluate the performance of the analyzer, loader,
chemistries, ISE and spectral response.
2.6.11 Setup
The Setup menu allows the user to customize certain aspects of the system hardware
and software in a manner that best suits the individual laboratory.
Command Purpose
Tests Modify certain specified test parameters such as test
identification and reference range checks.
Quality control Enter mean and standard deviation values for control
constituents.
Hardware Enable the ISE module and a LIMS, turn off the reagent
cooling system, set the sound level of the beeper.
The Status menu allows the user to check the status of critical areas of the system. The
Status menu is displayed by pressing the F3 key on the keyboard.
The following table provides a brief description of the commands available from the
Status menu:
Command Purpose
Reagent tray For each reagent bottle on the tray, displays remaining
volume, lot number, onboard lifetime, position and whether
or not it is required for pending requisitions on active
samples.
? Note: The status of these areas should be checked prior to daily routine
operation of the system. Expired reagent bottles must be removed from the
reagent tray.
? Note: The segment door will remain locked until Fn and F2 are pressed for
access.
The following table provides a brief description of the commands available from the
Access menu:
Command Purpose
Segment Place or remove cup or well segments.
Once access is complete, pressing ESC closes the Access menu and allows operation
to continue.
The software version represents the version of the operating software that is running on
the system.
The ACE Alera System can be configured with a multi-level password system to limit
user access to menu options. One of four levels of access (NONE, LEVEL 1, LEVEL 2,
LEVEL 3) can be defined for each menu option with the exception of Requisitions.
Each level of access (with the exception of NONE) requires the entry of a unique
password in order to use the menu options. The current user access level is displayed
on the upper right of the user interface.
The user logs on the system via the user Access selection on the Main menu. The user
Access menu has log in and log out options.
At log in, a password is entered. The user can then access all menus with that level
assignment. If the user attempts to access a menu with a higher access level, the
system displays a password prompt.
The log out selection signs the user off the system and returns the user access level to
NONE.
2.8 Specifications
ACE Alera System specifications are summarized below.
ANALYZER
General
Start-up time 15 minutes from power-up
Analysis Types Spectrophotometric, Potentiometric (optional)
Assay Types Final Point, Delta (two-point)
Slope (factor or calibrated)
Quadratic
Throughput Up to 165 results/hour
STAT Turnaround Between 1 and 15 minutes depending on test(s) ordered
Time
Calibration Frequency
Optical Monthly sample delay and optical calibration.
Photometric Test dependent (refer to package insert)
Potentiometric Automatic every three hours
Analytical 37 ?C ± 0.3 ?C
Temperature
Ambient Room 15 ?C (59 ?F) to 32 ?C (89.6 ?F)
Temperature
Humidity 20 to 80% RH (non-condensing)
Maximum Cuvette 248
Capacity
Power Rating
Analyzer 100-240 VAC, 47/63 HZ, 6 Amps Max.
ACE Alera 450 Watts Max (Analyzer only)
Nominal
Printer 120 or 240 VAC, 66 Watts Idle
340 Watts Operating
700 Watts Peak
Sample Delivery
Presentation to Closed or open tube
System Sample cups individually
Sample cups in groups of up to 15 on cup segment
Segment Capacity Five in combination (cup and/or well)
Tube Size for Direct 75 to 100 mm length; 13 to 16 mm diameter
Sampling
Stopper Styles Rubber stopper, VACUTAINER with Hemogard Closure or
stopperless.
Fill Volume Required Between 0.9 inches and 2.23 inches from top of glass tube
for Tube sampling
Serum Volume Minimum 0.5 inches above clot or separator
Required for Tube
Sampling
Sample Capacity 75-150 samples
15 sample cups/cup segment
30 samples wells/well segment
Sample ID Positive ID by optional barcode or requisition match for closed
or open tubes
Loadlist or requisition match for sample cups
Barcode Type Uniform Symbology Specification 39, Code 128 Set B and Set
C, Codabar and Interleaved 2 of 5
Barcode Placement Within the range of 0.09 inches and 2.35 inches from top of
Zone glass tubes with a maximum of 5? skew (slant)
Sample Delivery 3-200 ? L in 1 ? L increments
Capacity
Typical Volumes Used 3-20 ? L
Sample Dead Volume 90 ? L per seglet
Sample Delivery Less than 5% CV at 3 ? L
Precision
Level Detection Automatic with liquid level sensor
Sample Integrity 2 or 4 hours onboard working lifetime clock (2 or 4 can be
selected on the setup analyzer screen)
Reagent Delivery
Reagent Compartment 40 bottles
Capacity 20 with 30 mL capacity
20 with 12 mL capacity
Reagent Compartment 8 ?C + 1 ?C at the reagent sensor
Temperature
Reagent ID Positive ID by computer coded labels on bottom of bottles
Reagent Delivery 3-500 ? L in 1 ? L increments
Capacity
Reagent Delivery Less than 0.5% CV
Precision
Level Detection Automatic with liquid level sensor
Reagent Integrity Onboard working lifetime clock
Optical System
Absorbance
Measurement
Type Holographic diffraction grating
Linear Range 0.0000–2.0000 O.D. at 0.67 cm pathlength
Wavelengths 340, 378, 408, 447, 486, 505, 515, 525, 544, 554, 573, 592,
610, 629, 647, 692 nm
Lamp Pulsed Xenon
Pathlength 0.67 cm
Cuvette Volume 150-600 ? L
Analysis Cycle 10 seconds
Potentiometric System
Sample Type serum
Assay Volume 156 ? L (a minimum of 206 ? L required in cup to accommodate
probe depth and dead volume)
Analytes Measured Sodium
Potassium
Chloride
Analysis Cycle 90 seconds
Computer Processor IBM-compatible PC®
RAM Memory 32 MB
Graphics Adaptor Color SVGA compatible
External Disk Drive 3.5 in., 1.44 MB
Hard Disk Drive 20 GB
Operating System MS-DOS 6.22
Keyboard Type 83 key
LCD Monitor
Type 12.1 in. TFT Active matrix panel
Resolution SVGA 800 pixels by 600 pixels
Printer
Type Okidata B4200 Laser
Speed 19 pages per minute maximum
Resolution 600 by 1,200 dots per inch
Interface Centronics parallel compatible
Paper Type Standard letter
Capacity 250 sheets
Database Capacities
Analyzer Procedures
(Performance) 15
Bottles 250
Controls 20
Doctors 200
Dyes 15
Loader Procedures
(Performance) 2
Panels 16
Patients 1000
Requisitions 5000
Requisitions per 200
Patient
Tests 200
3 Principles of Operation
The system contains software that schedules the order in which the instrument performs
pending tasks. Each sample that is presented to the system for testing is associated with
a requisition that delineates the tests to be performed and whether the test is routine or
STAT. A STAT test will always be scheduled ahead of a routine test.
The software will schedule any calibration activities that are necessary and then samples
and controls are run on a first-in-first-out (FIFO) basis according to time of introduction to
the system.
Fluid Transfer Pipette samples, diluent, buffers and reagents into cuvettes.
The software monitors the status of the system during testing and alerts the operator to
conditions that may affect system performance.
The system checks that the required reagents, diluents and calibrators are available
whenever a test is requested. If one of the necessary components is missing, a warning
is displayed and the test is placed on hold.
The liquid level, lot number and working lifetime of each reagent are also verified. If a
new lot number of reagent is detected, a calibration is required before that reagent can
be used. If the working lifetime of a reagent has passed, the system displays a warning
and sets the bottle status to “expired,”which requires the bottle to be replaced.
The system also monitors samples, calibrators and controls for volume and the length of
time they have been in the instrument. Appropriate warnings of short volume or
exceeding time limits are displayed.
All chemistry tests except zero-order kinetic tests require a stored calibration curve. A
calibration curve is constructed by measuring and storing responses to known
concentrations of a calibrator. Sample results are calculated by interpolation and curve
fitting techniques that are based on the stored calibration curve.
A calibration run will utilize from one to six calibrators, specified in the program test
parameters. A stored calibration will remain in memory until either a new calibration is
requested or the test parameters have been modified. When a new calibration is
accepted it will replace the previously stored curve. The date and time of the calibration
will be recorded and a calibration report will automatically be printed if enabled in setup.
If a calibration fails, no new sample requests for that particular analyte will be processed
until a valid calibration is stored. Any samples already in process will be aborted.
The three ISE channels (sodium, potassium, chloride) are calibrated together and all
three must pass in order for the calibration curves to be stored. If a channel fails, the
calibration will be repeated. Any channel that cannot be calibrated after three attempts is
marked as having failed calibration and an attention message is displayed.
When a full calibration is performed for a test, a curve corrector (correction) can be
analyzed at the same time. A curve corrector may contain the analyte, and is not
necessarily a calibrator. For example, distilled water may be used as a corrector. The
response of the corrector is stored with the curve. The corrector material can be
requested any time after that.
The change in the response of the corrector between the original value stored and the
current value obtained is used to adjust the curve. The curve may be adjusted by
changing the slope or the offset. If the curve is non-linear, the corrector response and
result must be between the lowest and highest calibrator.
If a slope correction is used, each of the stored calibrator responses is adjusted by the
ratio of the stored response of the corrector to its correction response. The result is a
change in the shape and orientation of the calibration curve. The following calculation is
used:
New Cal Resp. = Cal Resp. x (Correction Resp. of Corrector/Stored Resp. of Corrector)
An offset correction adjusts each of the stored calibrator responses by the difference
between the stored response of the corrector and its correction response. This implies
that the correction has the same effect on each of the calibrators. For tests that use only
one calibrator, offset correction is not permitted. The following calculation is used:
New Cal Resp. = Cal Resp. + (Correction Resp. of Corrector – Stored Resp. of Corrector)
Should an error occur during a correction run, the correction values are not stored. The
calibration or correction must be repeated.
More than one response may be averaged for each calibrator and corrector when
calculating a curve. Up to four replicates for each response are allowed. Each calibrator
is loaded into the specified number of cuvettes. If an external blank is required, only one
is loaded per calibration. The mean response for the calibrator is calculated from the
response for each of its cuvettes. The averaged value is used to calculate or correct the
curve.
The response for a test is calculated from the optical data obtained for the readings
specified in the test parameters. The data are assembled into algebraic equations that
define the response. The basic types of equations are described below:
Equation Description
Final Point Based on the final data point.
Delta Based on the difference between the final and initial data
points.
Slope Based on a linear regression slope through all the valid data
points taken. Needs a minimum of four points.
The final result for a sample represents either the concentration or activity of the analyte.
Calculation of the result is based on the use of calibrators or a predefined factor. The
different types of calculations can be used in combination with any of the response
calculations. Four calculation types are available in the ACE Alera System.
Type Description
Factor The response is multiplied by a predefined factor that is
stored in the test parameters.
Four Parameter The response and result have a non-linear relationship. Four
Non-Linear Least to six calibrators can be used to determine the convergence
Squares Fit of the curve.
The calculation equations used for routine chemistry tests performed on the ACE Alera
System are described below. The symbols used in the equations are defined following
the equations.
Final Point with Calibration or Factor and an Internal Reagent Blank Correction
V(P)
Result = {(AUW1 – AUW2) - ?(AUIRBW1 – AUIRBW2) x ?} x Cal Factor or Factor
V(T)
? Note: The internal reagent blank corrects for a change in the reagent
absorbance.
Final Point with Calibration or Factor and an External Sample Blank and an
Internal Reagent Blank Correction
V(P)
Result = {(AUW1 – AUW2) - ?(AUIRBW1 – AUIRBW2) x - (AESBW1 – AESBW2)? x Cal Factor or Factor
V(T)
? Note: The internal reagent blank corrects for a change in the reagent
absorbance. The external sample blank corrects for the contribution of the
sample absorbance to the final reaction absorbance.
Final Point with Calibration or Factor and an External Reagent Blank and Internal
Sample Blank Correction
V(P)
Result = {(AUW1 – AUW2) - ?(AUISBW1 – AUISBW2) x - (AUERBW1 – AUERBW2)? x Cal Factor or Factor
V(T)
? Note: The external reagent blank corrects for a change in the reagent
absorbance. The internal sample blank corrects for the contribution of the
sample absorbance to the final reaction absorbance.
? Note: The external reagent blank corrects for the drift of the reagent. It is taken
whenever a new bottle of reagent is used and when a re-calibration of the
system is performed.
Delta – Two Point with Calibration or Factor and an External Reagent Blank
Correction
Symbol Definition
A Absorbance reading.
U Unknown.
W1 Reaction wavelength.
A requisition must be created for each sample that is introduced for testing. This
requisition identifies the type of sample (patient, quality control, calibrator) and the tests
to be performed on the sample. Photometric and potentiometric tests may be ordered on
the same requisition. Each requisition is assigned a unique Accession number that is
used to track it through the system. Accession numbers may be manually assigned, or
the system may be configured to automatically assign an Accession number when a
requisition is created. Samples are then loaded into the system as a batch or individually
in preparation for testing.
? Note: The Patient First Name (if required), Patient Last Name, Patient ID,
Accession number, and at least one test to be run are required for a patient
requisition.
For photometric tests, the system automatically dispenses the appropriate reagents into
a cuvette in the reaction wheel, aspirates the required sample volume and dispenses it
into the cuvette. The system then performs any incubations that may be required as
determined by the parameters that have been preprogrammed for the test.
When the reaction is complete, each cuvette is passed through the system and it is read
at the appropriate wavelength or wavelengths for the test. The data is processed and
results are calculated.
For potentiometric (ISE) tests, the probe draws up several aliquots of the sample and
dispenses them through the electrode block to “prime”the tubing and the electrodes.
The electrode block is emptied and air bubbles are introduced into the line to aid in the
removal of the previous sample. The sample is then pumped into the measuring
chamber of the electrodes and the electrical potentials are measured several times.
The system performs checks against pre-programmed limits to detect possible ADC
saturation. The chambers are then filled with CAL-A and the electrical potentials are
again measured. At the completion of the measurement cycle all measurements are sent
to the workstation and the sample results are calculated based on the voltage difference
between the sample and the CAL-A and the stored calibration curve.
The ACE Alera System automatically verifies the validity of test results using limits that
are included in the test parameters. When an error is detected, a flag is reported instead
of a test result. This gives the operator the opportunity to evaluate the nature of the
error and take appropriate corrective action. Under certain circumstances, the system
will require that the operator perform a manual verification of a test result.
The system will automatically dilute and rerun a test on a patient sample if the results
have exceeded the linearity or calibration limit of the test or if substrate depletion has
occurred. Only one automatic rerun is permitted for a test. The system will not perform
automatic reruns on controls, calibrators, correctors or potentiometric (ISE) tests.
The chemistry flags that may be displayed or reported depend upon the type of data
being verified:
Flag Description
ANALZ Problem detected on the upper or lower photodetector output
(automatic) limits, the upper or lower limits on the averaged transmission or
intensity values or on the comparison of data to system offsets.
ANALZ Absorbance for a cuvette exceeds the maximum allowable value.
(automatic)
NEGABS Absorbance for a cuvette is less than the minimum limit (-0.05).
(automatic)
MXINIT Optical data for the early data point exceeds the maximum limit
(configurable) defined in the test parameters.
MNINIT Optical data for the early data point is less than the minimum limit
(configurable) defined in the test parameters.
DEPL Substrate depletion. Error occurs when ? A between the early data
(configurable) point and any of the first 4 readings in the reaction exceeds the
limit set in the test parameters.
MXRBDP Optical data point for the internal reagent blank exceeds the
(configurable) maximum limit.
MNRBDP Optical data point for the internal reagent blank is less than the
(configurable) minimum limit.
Response Flags
Flag Description
SIGN Reaction direction is incorrect based on the test parameters.
(configurable)
SE Standard error of linear regression in slope response calculation
(configurable) (> max. limit).
LT LC, GT HC Response is either less than the lowest or greater than the highest
(automatic) calibrator. Check is performed if Four Parameter Non-Linear Least
Squares Fit or Non-Linear Interpolation is used in results
calculation.
MAXERB Response for the external reagent blank exceeds a maximum
(configurable) limit.
MINERB Response for the external reagent blank falls below a minimum
(configurable) limit.
RSS The root-sum-square of the residuals for a quadratic least squares
(configurable) fit exceeded the maximum limit.
Result Flags
Flag Description
LIN Reported when a sample result exceeds the linearity limit for the
(configurable) method. If the sample has been automatically rerun with a dilution,
the linearity limit is scaled by the rerun dilution factor before the
check is made.
LT LC, GT HC A sample result is either less than the lowest or greater than the
(automatic) highest calibrator. Check is performed if Four Parameter Non-
Linear Least Squares Fit or Non-Linear Interpolation is used in
results calculation. If the sample has been automatically rerun with
a dilution, the range endpoints are scaled by the rerun dilution
factor before the check is made.
LT 0 Result is less than zero (negative result).
(automatic)
FORMAT Result exceeds the allowable format field size.
(automatic)
LTLL, GTHL An ISE result is either less than the lowest limit or greater than the
highest limit of electrode measurement.
Calibrator Flags
Flag Description
MATH Calibrator response is zero. A ratio cannot be determined (division
(automatic) by zero is not permitted).
CALFAC Ratio of the concentration of a calibrator or a corrector to its
(configurable) average response exceeds the range defined in the test
parameters. A non-calibrator corrector is compared to the
allowable range of the nearest calibrator.
CALDEV The expected result for a calibrator deviates from the calculated
(configurable) result obtained from the least squares fit calibration curve more
than the maximum limit set. This check is performed on Four
Parameter Non-Linear Least Squares Fit calibration curves or
Linear calibration curves.
CORDEV The correction response for a corrector deviates from its stored
(configurable) response by more that the maximum limit set.
NONMON A calibration curve is not sequentially increasing or decreasing.
(automatic)
CONVRG Improper curve fit using Four Parameter Non-Linear Least
(automatic) Squares Fit calculation.
CAL SN Correction changed the sign of a calibrator response in a
(automatic) correction run.
Flag Description
H Patient result exceeds upper range of normal limit configured in
(configurable) the test parameters.
L Patient result falls below lower range of normal limit configured in
(configurable) the test parameters.
HH Panic result. Patient value exceeds upper panic limit programmed
(configurable) in test parameters.
LL Panic result. Patient value is less than lower panic limit
(configurable) programmed in test parameters.
D Result was obtained on a diluted sample.
? Note: A panic flag (HH or LL) may be reported under the following conditions
even though another error has been detected in a run:
?? HH is reported when the patient result also exceeds the linearity limit
(LIN).
?? HH is reported when the patient result is equal to or exceeds the highest
calibrator (GT HC).
?? LL is reported when the patient result is equal to or lower than the lowest
calibrator (LT LC).
A patient sample will automatically be diluted and rerun (provided that a dilution ratio has
been specified for the test) when one of the following occurs:
?? The result exceeds the linearity limit of the test (linear test).
?? The result exceeds the calibration range for the test (non-linear test).
?? Substrate depletion occurs.
The sample is diluted with the diluent specified in the parameters for that test. The result
for the diluted sample is automatically multiplied by the dilution factor and flagged with a
D when reported to indicate the sample has been diluted.
The system will rerun a sample only once for a given test within a requisition. Automatic
reruns are not performed on controls, calibrators or correctors.
ACE Alera System test results can be correlated to the results from a different
instrument or test method. A correlation study is performed and the slope and y-intercept
from a linear regression equation are entered into the test parameters. The correlated
result is obtained by multiplying the ACE Alera System result by the slope and adding
the value of the y-intercept.
A wide variety of patient and quality control reports are available. These may be printed
at the workstation or can be transmitted to an off-line laboratory information
management system (LIMS).
The system can be configured to automatically print a report for patient, quality control
and calibration requisitions, or it can be set up so that a patient report must be reviewed
before it can be printed or transmitted to a LIMS.
?? For optimum performance, samples to be assayed for ISE tests should not be left
on the system for more than four hours.
?? Never turn off the power to the analyzer for an extended period (i.e., more than
four hours) without first going through the ISE Shut Down Procedure. Failure to
follow this procedure can result in damage to the electrodes, reference housing
and tubing. Damage can also occur if the ISE is left without CAL-A for an
extended period.
?? Perform ISE cleaning and conditioning daily. Place the ISE cleaning solution,
sodium electrode conditioner and ISE wash bottles in the reagent compartment
to perform the procedures.
?? Remove any crystals that may develop on the tubing, container caps or
electrodes. Wipe up ISE fluid spills.
?? The CAL-A, CAL-B, and reference solution containers and the ISE wash bottle
should never be refilled. Always replace them when necessary.
?? All tubing should be secured within the tubing holders and not allowed to hang
loose.
?? Always handle electrodes with care. The reference electrode must always be
stored in the transport housing filled with reference solution and never
allowed to become dry or come in contact with other fluids including water.
The reference housing should be cleaned and dried before storing. Water should
never be used to rinse the reference electrode.
?? Do not insert the stylet into any of the electrodes.
?? Whenever an electrode is removed and replaced, a two-point calibration must be
run before assaying for ISE tests.
?? Analyze serum samples within one hour of collection to minimize changes in
concentration. When serum remains in contact with red blood cells, potassium
shifts occur, resulting in an increase of potassium measured. If samples cannot
be measured within four hours, remove fluid from cells and store in a capped
tube or sample cup in the refrigerator. Prior to measuring refrigerated samples,
allow them to warm to room temperature.
?? Do not use control sera prepared with ethylene glycol or ammonium carbonate
diluents as they will give erroneous results.
?? Reference housings should be cleaned weekly in bleach, to prevent clogging of
the capillary tube with protein.
?? Make certain salt crystals or liquids do not come in contact with the CAL-B
aspirator or the CAL-B solution, as CAL-B will become contaminated, resulting in
a low potassium calibration slope and elevated potassium results.
4 System Setup
Unpacking and installation of the ACE Alera System is to be performed only by an
authorized Alfa Wassermann Diagnostic Technologies, LLC representative.
This section provides information on selecting and preparing the site for installation and
information about connecting the various components.
4.1 Materials
Article Quantity
Seglets 1 bag
Evap Caps 1 box
Cuvette Waste Box 1
BIOBAGS 1 box
Maintenance Kit 1
Maintenance Log 1
Master Log 1
The Analyzer requires approximately 0.9 m2 (10 sq ft.) of bench space. For easy
operator access, the system should be located 75 to 90 cm (30 to 35 inches) above the
floor on a clean, firm, level bench top. The bench should be free of excessive vibration
from equipment such as a centrifuge or compressor. Position the instrument as far away
as possible from any source of strong magnetic fields. Other considerations in selecting
a site include:
Accessibility – Select a location that will not hinder the operation of the system
or interfere with the operation of other equipment in the immediate work area.
The operator should also have easy access to supplies and other related
equipment.
Environment – Select a location that is free from wide fluctuations in
temperature. Do not, for example, locate the instrument next to a source of heat
or air conditioning or in direct sunlight. Verify that the following conditions are
met:
?? Ambient temperature is between 15 ?C (59 ?F) and 32 ?C (89.6 ?F).
?? Relative humidity is 20% to 80%, non-condensing.
Analyzer Printer
Height 58.4 cm (23.0 in.) Height 20.0 cm (7.9 in.)
Width 69.9 cm (27.5 in.) Width 35.5 cm (14.0 in.)
Depth 57.2 cm (22.5 in.) Depth 39.5 cm (15.6 in.)
Raised Height 95.3 cm (37.5 in.)
Weight 64 kg (145 lb.)
UPS/Line
Conditioner
Height 8.9 cm (3.5 in.)
Width 43.8 cm (17.3 in.)
Depth 49.5 cm (19.5 in.)
The ACE Alera System requires electrical service that provides power at the rated
voltage and frequency. Consult section 2.8 for absolute power requirement values.
The power source for the modules (UPS, analyzer, and printer) should be isolated,
dedicated and “clean”and free of transients and excessive voltage variations.
? Note: Consult a local electrician if you are uncertain about the adequacy of
your electrical service.
All cable connections for the modules are provided with the system.
To protect the system from a power failure, connect the power cord from the analyzer
into the uninterruptible power supply (UPS). The UPS also provides line conditioning.
Connect the UPS power cord into the electrical supply outlet. Connect the printer power
cord directly to a power outlet.
Routing of the power cables from the analysis module and printer to the UPS and routing
of the communication cable from the analyzer to the printer are shown in Figure 13.
1 With the Main menu displayed, type X to exit DOS, then turn off the power
to the system at the UPS.
2 Insert Installation Disk 1 into drive A.
? Note: It will take about five minutes for the system to boot and
display the Main menu.
The Setup Disk contains database information such as test parameters and reagent
bottle information. If the setup information has been previously customized and backed
up (copied) to a disk, the information from the backup disk should be loaded instead.
Refer to page 170 for the procedure to restore the backup disk information.
? Important: Setup data contained on the backup disk will overwrite any current
system files. As a result, chemistry calibration data, patient data and control
data that were entered after the backup disk was prepared will be lost.
Daily backup of system data onto a disk will minimize the loss of chemistry
calibration data, patient data or control data in the event of a hard disk failure.
The ISE module must first be enabled on the system and the calibration mode set. Refer
to page 181 for instructions on enabling the ISE. Refer to page 182 for instructions on
setting the ISE calibration mode.
Five prepared tubing assemblies and six separate tubing segments must be installed.
These can be obtained from your Alfa Wassermann Diagnostic Technologies, LLC
representative:
When installing the ISE tubing, remove the tubing assemblies and tubing segments from
the packaging and replace it on the instrument one at the time.
1 Orient the plastic plate so that the smaller (narrower) end is closest to the
rear of the ISE Module.
2 Loop the peristaltic pump tubing (P/N 27-11719) around the pump rollers
(Figure 14). The tubing should remain parallel and should not cross.
? Note: Do not install the pump tubing plate on the pump tubing
plate bracket at this time.
Installing the Air Inlet (3.5 inch, Light Blue Banded Silcone Rubber) Tubing
1 Select AIR: and press the insert key to open the air valve.
2 Loop the air inlet tubing (P/N 23-11730) around the air valve (Figure 15).
Pull on both ends of the tubing until it snaps into the valve.
3 Pull the tubing through the valve, leaving a small length at the top of the
valve to secure in the retainer.
4 Attach the other end to the far right hand port of the fluid manifold. There
should not be tension on the tubing when attached.
5 Press the insert key to close the air valve.
1 Select CAL-B and press insert key to open the CAL-B valve.
2 Loop the CAL-B cap assembly tubing (P/N 27-11283) around the CAL-B
cap valve (Figure 16). Pull on both ends of the tubing until it fully seats into
the valve.
3 Pull the tubing through the valve. Attach the end of the tubing to the
second inlet port from the right on the fluid manifold.
4 Pull the tubing back through the valve to remove excess slack. There
should not be tension on the tubing when attached.
5 Press the insert key to close the CAL-B valve.
Installing the Waste Transport (8.5 inch, Green Banded Silcone Rubber) Tubing
1 Attach one end of the waste transport tubing (P/N 23-11730) to the
stainless steel tube labeled WASTE.
2 Press the tubing into the rear-most pair of retaining clips on the top of the
module.
3 Attach the free end to the right rear port of the pump plate (Figure 17).
Installing the CAL-A (10.0 inch, Red Banded Silcone Rubber) Transport Tubing
1 Select CAL-A and press the insert key. The CAL-A valve opens.
2 Loop the CAL-A transport tubing (P/N 23-11730) around the CAL-A valve.
Pull on both ends of the tubing until it fully seats into the valve.
3 Attach the end of the tubing protruding from the bottom of the valve to the
second inlet port from the left on the fluid manifold (Figure 18).
4 Pull the tubing back through the valve to remove the slack between the
fluid manifold and the valve. There should not be tension on the tubing
when attached.
5 Attach the other end of the tubing to the stainless steel tube labeled
CAL-A at the top of the back wall of the module.
6 Secure the tubing in the second pair of tubing retaining clips from the rear
on the top of the module.
7 Press the insert key to close the CAL-A valve.
Installing the Sample Transport (8.5 inch, Green Banded Silcone Rubber) Tubing
1 Select SAMPLE and press the insert key. The sample valve opens.
2 Loop the sample transport tubing (P/N 23-11730) around the sample
valve. Pull on both ends of the tubing until it fully seats into the valve.
3 Attach the loose end protruding from the bottom of the valve to the first
inlet port on the left of the fluid manifold (Figure 19).
4 Pull the tubing back through the valve so that there is minimum slack
between the fluid manifold and the valve. There should not be tension on
the tubing when attached.
5 Attach the other end of the tubing to the stainless steel inlet port on the
upper left of the ISE Module marked Sample. Push the tubing
approximately 6 mm (1/4 inch) onto the port.
6 Secure the tubing in the third retaining clip from the rear on the top of the
module.
7 Press Insert to close the sample valve.
Installing the Measuring Chamber Waste (13 inch, Black Banded) Tubing
Installing the Measuring Chamber Inlet (2.5 inch, Purple Banded Silicone) Tubing
1 Move the measuring chamber forward for easier access. Attach one end
of the measuring chamber inlet tubing assembly (P/N 23-11730) to the
port on the right side of the measuring chamber.
2 Slide the measuring chamber back on its rails until it snaps in place.
3 Connect the other end of the tube to the horizontal port protruding from the
left end of the fluid manifold (Figure 21).
4 Reconnect the sample sensor cable by inserting the plastic connector into
the receptacle above the measuring chamber.
Installing the Reference Housing Outlet (12 inch, Yellow Banded) Tubing
Installing the Reference Housing Inlet (10.75 inch, Dark Blue Banded) Tubing
1 Attach the tube to the right port at the front of the pump plate (Figure 23).
2 Route the tubing toward the top of the ISE module. Secure it in the front
pair of retaining clips.
3 Connect the other end of the tubing to the unmarked port on the reference
housing when it is installed (see page 71).
1 Connect the yellow-banded tubing to the right tubing port of the Reference
reservoir (Figure 24).
2 Connect the other tube to the left front port of the peristaltic pump plate.
3 Pull upward on the pump plate, stretching the silicone tubing, and engage
the plate in the pump plate bracket. The pump tubing should remain
parallel and should not be allowed to cross.
4 Make sure that the smaller (narrower) end of the pump tubing bracket is
closest to the rear of the ISE module.
After all of the ISE tubing is installed, the reference electrode housing is installed into the
measuring chamber. The reference electrode and the three ion-selective electrodes are
then inserted into the measuring chamber.
8 Connect the clear tubing without the yellow band to the unmarked port on
the reference housing.
1 Remove the reference electrode from its box and unscrew it from the red
transport housing.
2 Make sure that the o-ring on the reference electrode is properly seated.
3 Carefully screw the reference electrode into the reference housing. Do not
overtighten.
? Note: Rinse the transport housing and save it for storage of the
reference electrode and for use during maintenance of the
reference housing.
4 Remove the sodium electrode from its box. Make sure an o-ring is present
and that is properly seated on the left side.
5 Install the electrode next to the reference electrode assembly.
6 Remove the potassium electrode from its box. Make sure an o-ring is
present and that it is properly seated on the left side.
7 Install the electrode next to the sodium electrode.
8 Remove the chloride electrode from its box. Make sure that an o-ring is
present and that it is properly seated on the left side.
9 Install the electrode next to the potassium electrode.
10 Make sure that the bottom lip of each electrode rests on the flat edge of
the measuring chamber.
11 With one hand holding the electrodes in place, close the clamp on the left
electrode holder by lifting it upward until it locks in the closed position.
? Note: Save the box that each electrode was packaged in. They
can be used for future storage of the electrodes.
12 Push the measuring chamber back until it snaps into the engaged position.
13 If the sample sensor cable was removed for any reason, reconnect the
cable.
1 Display the Pump and Valve Operation screen (see page 60).
2 Select AIR: and press the spacebar to operate the pump and valve.
3 Verify that reference solution flows from the reference container into the
lines, through the reference electrode housing, up to the ISE Reference
Reservoir and out of the ISE Reference Reservoir back into the reference
solution container.
? Note: The fluid sensor field on the screen should change from
LIQUID to AIR.
4 Press the spacebar to stop the pump.
5 Select CAL-A and press the spacebar to operate the pump and valve.
? Note: The fluid sensor field on the screen should change from
AIR to LIQUID.
9 Verify that air flows into the manifold and lines and that small drops of
reference solution are dripping into the sample path section of the
reference housing.
? Note: The fluid sensor field on the screen should change from
LIQUID to AIR.
10 Press the spacebar to stop the pump.
11 Select CAL-B and press the spacebar to operate the pump and valve.
12 When CAL-B appears on the fluid sensor, press the spacebar to stop the
pump.
? Note: The fluid sensor field on the screen should change from AIR
to LIQUID.
? Note: The drip rate should be approximately one drop every one to five
seconds. The time between successive drops should be consistent within
one second. Use a flashlight to observe drops from the reference housing
into the sample fluid flow path.
? Note: The fluid sensor field on the screen should change from LIQUID to
AIR.
1 Select CAL-A and press the spacebar to operate the pump and valve.
2 Verify that CAL-A solution is flowing into manifold, through the lines, and
into the electrode chamber.
3 Verify that no air bubbles are present.
4 Press the spacebar to stop the pump.
5 Select AIR and press the spacebar.
6 Verify that no fluid droplets are present.
7 Pump the following fluids and verify the observations.
Fluid Observation
CAL-B No air bubbles seen.
The liquid sensor must be calibrated before the ISE can be run.
The electrodes must be cleaned, conditioned and calibrated each day before tests are
run. Refer to the Maintenance section for the procedures.
Procedure Page
5.2.1 Power
Verify that the power switches to the UPS, analyzer, and printer are ON.
? Note: The power to the UPS and analyzer should remain ON at all times
except when performing maintenance or service.
When the working lifetime of a reagent bottle is reached, the reagent bottle is designated
as EXPIRED on the Reagent Tray Status screen and the bottle will no longer be used by
the system.
? Note: If the working lifetime of a reagent bottle is reached while tests are
being run on the system, any remaining tests that require that reagent will
be placed on hold.
3 Verify that:
The Calibration Status screen displays the state of each photometric test requiring
calibration.
? Note: Calibration states are NO CAL, NEW LOT, CAL PENDING, CAL
PASSED, CAL FAILED, or CAL EXPIRED.
3 Check the calibration status of each test. Note any tests that will require
calibration.
Note: Any request for a test without a valid calibration will be put on
? hold. In the example above, the DBILI calibration has expired and
DBILI test requests will be put on hold until the DBILI calibration is
successfully completed.
The On-Hold Status screen lists any tests that are on hold. A test is on hold if a
component that is required to complete the test is not available at the time the sample is
loaded.
? Note: Note any tests that are on hold so that the appropriate action
may be taken to complete the test.
3 To view the details for a sample ring section, select the section using the
arrow keys and then press ENTER:
4 Verify that the required number of well segments and cup segments are
available.
? Note: Well segments are used for samples loaded from tubes placed
in the STEP Module and for automatic dilution of samples. Cup
Segments are used for samples transferred to sample cups for
placement on the system including calibrators, controls and patient
samples. A cup segment with a status of CLEAN has positions
available to accept sample cups from the Patient Requisition Screen.
The Hopper Status screen displays the number of cuvettes remaining in the hopper, the
number of cuvettes in each of the 20 slots and the number of empty slots.
The Database Status screen lists the capacity of various system parameters by number
remaining, number used and maximum number allowed.
2 Select Hopper.
3 The message Preparing for hopper access… is briefly displayed before
instructions for inserting cartridges are displayed:
5 Grasp the cuvette cartridge at the top of one end, with the cuvettes facing
towards you (Figure 25).
7 Remove the cartridge and verify that 10 cuvettes have been removed from
the cartridge.
? Note: After the ten cuvettes are loaded, the hopper will advance to
the next available slot and the indicator will illuminate.
If the system detects that less than 10 cuvettes have been loaded, it
will display a warning message. If this occurs, respond to the screen
prompt or troubleshoot the problem.
8 Turn over the cuvette cartridge and grasp the top with the cuvettes facing
towards you.
9 Carefully slide the cartridge into the hopper until it stops.
10 Remove the cartridge and verify that the remaining ten cuvettes have
been removed from the cartridge.
11 Repeat Steps 4 through 10 until the desired number of cuvettes has been
loaded.
Well segments are used for samples that are loaded from tubes placed in the STEP and
for automatic sample dilutions on the instrument.
? Note: A well segment contains three disposable seglets, each containing ten
wells (see Figure 8 on page 13). Three clean seglets must be placed in the
well segment before it is loaded on the system.
At least one clean well segment should be loaded on the system before
starting a run. The total number of well segments required depends upon the
workload and how the samples will be loaded.
To unlock the segment door, perform the following steps:
6 Place the full well segment on the outer portion of the reagent tray,
ensuring that the well segment is securely on the posts (Figure 26).
7 Close the access cover.
8 Press F10.
9 Repeat the entire procedure to load another well segment.
Empty cup segments are used to accept sample cups that are individually loaded during
creation of a patient requisition or a control requisition.
? Note: At least one clean cup segment should be loaded on the system before
starting a run. The total number of cup segments required depends upon the
workload and how the samples are loaded.
6 Place the empty cup segment on the outer portion of the reagent tray,
ensuring that the cup segment is securely on the post (Figure 27).
Consult the package insert that accompanies each reagent for instructions concerning
the storage and preparation of the reagent before it is placed on the system.
Before placing a reagent on the system, replace the cap with an Evap-Cap:
2 Ensure the top of the bottle is dry and that no large bubbles are present
within the neck of the bottle.
? Note: Be sure that the hard plastic caps have been removed and
replaced with paper Evap-Caps.
? Note: Place the reagent bottles anywhere on the ring, with large
bottles in the outer ring and small bottles in the inner ring. The
dot-coded label on the bottom of the bottle identifies each of the
reagents on the analyzer.
If a reagent bottle does not contain sufficient volume to complete
the workload, a second bottle can be placed on the system before
the run is started. The new reagent bottle will be held in reserve
until reagent in the first bottle is depleted. The system will then
automatically switch to the second bottle.
6 Press F10.
? Note: System sensors monitor the fluid level in the waste, system diluent,
CAL-A, and CAL-B containers. Servicing the containers at the beginning of the
day avoids an interruption in workflow.
3 Open the Container Access Door on the left side of the analyzer
(Figure 29).
Cuvette Waste
Box
ISE CAL A
System Diluent
Liquid Waste
Container
Metal Release
Button
1 Check the level of system diluent. Replace the system diluent if there is
less than 1/4 inch of liquid in the container.
2 Remove the system diluent container from the compartment by tilting the
bottom of the bottle up and then sliding it out.
3 Press the metal release button on the tubing (next to the cap) and then
disconnect the tubing (Figure 31).
4 Unscrew the used system diluent container from the cap assembly. Set
the cap aside.
5 Remove the cap from a new container of system diluent. Discard the cap.
6 Place the system diluent cap assembly into the new system diluent
container and tighten it. Reinsert the tubing into the cap.
7 Tilt the system diluent container and gently slide it up and forward into the
compartment.
? Note: The label on the system diluent container and the metal
guide on the cap should be facing the liquid waste container.
Metal Release
Button
1 Check the level of ISE CAL-A. Replace CAL-A if there is less than 1/4 inch
of liquid in the container.
2 Remove the ISE CAL-A container from the compartment by tilting the
bottom of the bottle up and then sliding it out.
3 Twist the tubing connector on the tubing (next to the cap) counter-
clockwise and then disconnect the tubing (Figure 32).
4 Unscrew the used ISE CAL-A container from the cap assembly. Set the
cap aside.
5 Remove the cap from a new container of ISE CAL-A. Discard the cap.
6 Place the ISE CAL-A cap assembly into the new ISE CAL-A container and
tighten it. Reinsert the tubing into the cap
7 Tilt the ISE CAL-A container and gently slide it up and forward into the
compartment.
1 Check the level of ISE CAL-B. Replace CAL-B if there is less than 1/2 inch
of liquid in the container.
2 Remove the Reference Solution and CAL-B containers from the system.
3 Clean the ISE Module, removing any salt deposits or spills.
4 Remove any salt deposits from the cap assemblies on the ISE Reference
Solution and ISE CAL-B containers.
5 Obtain a new container of ISE CAL-B and remove the cap.
6 Carefully remove the entire foil seal (including the foil on the rim of the
container) to ensure a good seal.
Important: Be careful not to contaminate the fluid in the container.
? Clean the tool before using it to break the foil seal on the container.
7 Unscrew the ISE CAL-B cap assembly from the used ISE CAL-B container
and transfer it to the new ISE CAL-B container. Screw on the ISE CAL-B
cap assembly until it is tightened.
?
Important: Do not allow the CAL-B aspiration tube to contact
Reference Solution, dried salt deposits, or any surface. Be sure
to use the red cap assembly with ISE CAL-B.
7 Unscrew the ISE Reference Solution cap assembly from the used ISE
Reference Solution container and transfer it to the new ISE Reference
Solution container. Screw on the ISE Reference Solution cap assembly
until it is tightened.
Biohazard: Waste material from the ACE Alera System which may
be classified as biohazardous includes used cuvettes dispensed into
the cuvette waste box. Handle cautiously in accordance with
laboratory safety rules and local regulations.
Check the level of waste in the cuvette waste box . If the level of used cuvettes is
nearing the top, empty the box before starting the workload for the day.
? Note: A sensor monitors the level of used cuvettes and will indicate when
it is necessary to empty the waste box. Emptying the waste box at the
beginning of the day avoids a later interruption.
3 Open the container access door on the left side of the analyser (see
Figure 29 on page 95).
4 Remove the waste box by pulling it out.
5 Remove the BIOBAG containing the used cuvettes from the waste box.
6 Seal the bag with one of the provided ties.
7 Dispose of the used cuvettes according to your facility regulations.
8 Insert a new BIOBAG into the waste box, folding the top of the bag down
over the sides of the box.
6 Ordering Tests
Once the daily preparation has been completed, tests can be ordered.
A requisition must be prepared for each sample on which tests are to be run. Different
types of requisitions are used for patient samples, quality control samples, and
calibrators.
If a patient record is created and saved, tests are ordered on samples from that patient
by attaching a new test requisition or by adding the tests to an existing requisition.
1 Select Patient from the Requisitions menu. A list of patients that have
already been entered into the system is displayed:
2 Press Insert. The Patient Requisition form is displayed with the LAST
NAME: field selected:
3 Type the patient’s last name and press Enter. The FIRST NAME: field is
selected.
? Note: LAST NAME: and PATIENT ID: are required fields. The
FIRST NAME: field may also be required, depending on how your
system has been configured. These fields cannot be edited once
the requisition is saved in the database.
The Tab, Enter or down arrow keys will also advance the cursor
to the next field. Use the up arrow key to return to the previous
field.
4 Type the patient’s first name (if required) and press Enter. The PATIENT
ID: field is selected.
5 Type the patient ID and press Enter. The DOB: field is selected.
6 Type the patient’s date of birth and press Enter. The SEX: field is
selected.
? Note: Enter the date of birth using the date format (MM/DD/YY or
DD/MM/YY) specified during system configuration. The AGE: field is
automatically filled in based on the entered date of birth.
If the DOB: field or any of the other remaining fields are to be left
blank, press Enter without typing any information in the field.
Press F6 at any time the cursor is in the upper section of the form to
advance the cursor to the ACCESSION: field.
? Note: If the doctor has not previously been entered in the system, a
prompt to add the new doctor is displayed.
If you prefer, press F2 to display a list of doctors that are on the
system, select the desired doctor and press Enter.
9 Type the location and press Enter. The COMMENT: field is selected.
10 Type a comment regarding the patient if desired and press Enter. The D/T
DRAWN: field is selected.
1 Select Patient from the Requisitions menu. A list of patients that have
already been entered into the system is displayed:
2 Select the desired patient (using the arrow keys) and press Enter. A list of
requisitions for the patient is displayed:
3 Press Insert. A new requisition form for the patient is displayed with the
D/T DRAWN: field selected:
The system can automatically generate multiple requisitions for the same test(s). Each
requisition has a unique Accession number and identical patient demographic
information.
1 Select Patient from the Requisitions menu. A list of patients that have
already been entered into the system is displayed.
? Note: Use a hyphen between the starting and ending values in the
number range. For example: to generate ten requisitions, type 1-10
in the ACCESSION: field.
The starting and ending Accession numbers that can be entered
must be numeric. Alphabetic characters (A through Z) can be used
when entering an individual Accession number, but not when
entering a range of Accession numbers.
Accession numbers cannot be changed once a Patient Requisition
is saved in the database.
5 Press Enter.
6 Add the desired test(s). (Refer to steps 7 and 8 of section 6.2.1).
7 Press F10 to generate the multiple requisitions.
2 Type the date the sample was drawn and press Enter. The time field is
selected.
? Note: Enter the date the sample was drawn using the date format
(MM/DD/YY or DD/MM/YY) specified during system configuration.
3 Type the time the sample was drawn and press Enter. The BY: field is
selected.
? Note: Use the 24-hour time format (e.g. 21:30 for 9:30 pm) for
entering the time.
4 Type the initials of the technician who drew the sample and press Enter.
The ACCESSION: field is selected.
5 (If needed) Type the Accession number and press Enter. The
COMMENT: field is selected.
6 Type a comment regarding the sample (if desired) and press Enter. The
TEST field is selected.
7 Type the mnemonic or abbreviation for the first test and press Enter. The
test is listed in the first TEST field, and the next TEST field is selected.
Or: press F2 to display a list of tests and test panels that are available.
Select the desired test (use the arrow keys) and press Enter:
? Note: If the test being ordered is a remote test, select the test and
then press F8. See page 113 for additional information.
Source requisitions for remote tests must be prepared first.
10 Press F7 to load the sample. Or, press F10 to save the requisition and
load the sample at a later time.
1 Select Patient from the Requisitions menu. A list of patients that have
already been entered into the system is displayed:
2 Select the desired patient (using the arrow keys) and press Enter. A list of
requisitions for the patient is displayed:
3 Select the desired requisition (using the arrow keys) and press Enter. The
requisition is displayed:
5 Type the mnemonic for the test and press Enter. The test is listed in the
TEST field, and the next TEST field is selected.
Or: Press F2 to display a list of tests and test panels that are available.
Select the desired test (use the arrow keys) and press Enter:
? Note: If the test being ordered is a remote test, select the test and
then press F8. See page 113 for additional information.
Source requisitions for remote tests must be prepared first.
7 Press F7 to choose a loading option for the sample. Or, press F10 to run
the test if the sample is already on the system.
? Note: The test will be run if the sample is already on the system
and has not been there for more than four hours.
? Note: The mnemonic for a remote test and for the corresponding source
test must be different. For example, GLU_1HR might be the mnemonic for
the one hour Glucose ordered as a remote test and GLU might be the
mnemonic for the actual Glucose test ordered (and run) on the pretreated
source sample.
The source requisition for a remote test must be prepared before the remote
test can be ordered (see Section 9.3.4 on page 201). For example, if a
Glucose Tolerance Test is ordered, the 1 Hour, 2 Hour, etc., should be
included in a requisition for the patient, and a serum sample should be run
under a second requisition so that they may be linked:
1 Create a requisition for the source test(s) (see Section 6.1 on page 103).
2 Order the source test(s).
3 Note the Accession number for the source test(s).
4 Press F10 to save the source test requisition, or press F7 to choose a
loading option for the sample(s).
After the source requisition is prepared, the GLU remote test can be included
on the requisition (see Section 6.2.2 on page 111):
1 Select the remote test (using the arrow keys) after it has been ordered.
2 Press F8. A prompt is displayed to enter the Accession number for the
source test.
3 Enter the Accession number for the source test (see Step 3 above).
1 Select Patient from the Requisitions menu. A list of patients that have
already been entered into the system is displayed:
2 Select the desired patient (using the arrow keys) and press Enter. A list of
requisitions for the patient is displayed:
3 Select the desired requisition (using the arrow keys) and press Enter. The
requisition is displayed:
7 Enter a comment.
8 Press F10. The test is deleted.
9 Press F10 again to save the requisition after deleting the test.
? Note: The First Name (if required), Last Name, Patient ID, and
Accession cannot be changed once a requisition is created and
accepted.
1 Select Patient from the Requisitions menu. A list of patients that have
already been entered into the system is displayed:
2 Select the desired patient (using the arrow keys) and press Enter. A list of
requisitions for the patient is displayed:
3 Select the desired requisition (using the arrow keys) and press Enter. The
requisition is displayed:
? Note: First Name (if required), Last Name, Patient ID, and
Accession cannot be modified.
? Note: Information for new controls or new lot numbers is entered during
system configuration.
2 Select the desired control (using the arrow keys) and press Enter. The
requisition and the current test results for the control are displayed:
3 The displayed test results must be deleted before running a new QC test.
Press F6 to clear the entire requisition, or perform one of the procedures
in the following sections.
? Note: If the sample is not on the system, pressing F10 saves the
requisition so the sample can be loaded at a later time.
? Note: If the sample is not on the system, pressing F10 saves the
requisition so the sample can be loaded at a later time.
? Note: Tests that require calibration will be noted during the calibration status
check (see page 80). A test must be calibrated whenever a new lot number of
reagents is introduced or if certain software and/or hardware changes are
made.
2 Type the mnemonic for the test to be calibrated and press Enter. The test
is listed in the first test field and the next test field is selected.
Or: Press F2 to display a list of tests and test panels that are available.
Select the desired test (use the arrow keys) and press Enter:
4 When all tests requiring calibration have been entered, press F10. The
loadlist for the calibration is displayed:
5 Type a segment number in the SEGMENT field and press Enter. The next
available cup position is selected.
7 Following the loadlist, place the calibrator(s) and any additional samples
into the designated positions on the cup segment.
? Note: When the calibrator samples are run, a report of the results
is printed (if enabled in Setup). If calibration fails for a test, that test
is put on hold and no further samples will be run for the test until a
calibration is successfully completed for the test.
2 Type the mnemonic for the test to be corrected and press Enter. The test
is listed in the first test field and the next test field is selected.
Or: Press F2 to display a list of tests and test panels that are available.
Select the desired test (use the arrow keys) and press Enter.
3 Repeat Step 2 to add additional tests to the requisition. As each test
needing correction is added to the requisition, the corrector(s) that are
required are automatically listed in the Correctors section.
4 Press F10 to display the Corrector Requisition loadlist.
5 Type a segment number in the SEGMENT: field and press Enter. The
next available cup position is selected.
6 The correctors may be run alone, or additional samples can be added to
the loadlist.
7 Following the loadlist, place the corrector(s) and any additional samples
into the designated positions on the cup segment.
? Note: When the correction samples are run, a report of the results
is printed (if enabled in Setup). If a correction fails, results will be
reported using the stored calibration curve until the correction
passes.
7 Running Tests
Samples can be loaded and run individually, or a group of samples can be loaded and
run as a batch. Samples must be run within four hours after they have been loaded.
? Note: A requisition must be created before a sample can be loaded and run.
Alternately, an aliquot of a single sample can be transferred from the tube into a sample
cup. The filled sample cup can then be loaded and run while the requisition for that
sample is displayed.
? Note: This is applicable only for systems equipped with an optional barcode
reader.
1 Place the tube in the STEP. A sample will be aspirated from the tube and
dispensed into a well segment. The ordered tests will then be run on the
sample.
? Note: A requisition must exist for the tube before it can be run.
Be sure that the barcode label is facing towards the left when
inserting the tube into the STEP.
3 Place the tube in the STEP. A sample will be aspirated from the tube and
dispensed into a well segment for processing. The ordered tests will then
be run on the sample.
4 Remove the tube from the STEP when it is returned.
4 When the access door unlocks, the segment access cover can be opened.
3 Select Fill. The Fill Cup screen is displayed listing all segments on the
sample ring.
4 Select the cup segment containing the short sample (use the arrow keys).
5 Press Enter. Information for the segment is displayed.
6 Select the short cup (use the arrow keys).
7 Press Enter. The tray is rotated to the correct position. Then, the message
Fill the cup in the instrument… is displayed.
8 Open the segment access cover and add sample to the cup indicated by
the arrow.
9 Close the cover and press F10.
10 Press Esc until the Main menu is displayed.
An expired sample cup that has tests pending may be replaced with a fresh cup.
1 Press Fn and F2 to display the Access menu.
2 Select Cup. The Cup menu is displayed.
3 Select Change. The Change Cup screen is displayed listing all segments
on the sample ring.
4 Select the cup segment containing the expired sample (use the arrow
keys).
5 Press Enter. Information for the segment is displayed.
6 Select the expired cup (use the arrow keys).
7 Press Enter. The tray is rotated to the correct position.
Then, the message Change the cup in the instrument… is displayed.
8 Open the segment access cover, remove the expired sample cup from the
position indicated by the arrow and replace it with a fresh sample cup.
9 Close the cover and press F10.
10 Press Esc until the Main menu is displayed.
3 Select Remove. The Remove Cup screen is displayed listing all segments
on the sample ring.
4 Select the cup segment containing the inactive sample (use the arrow
keys).
5 Press Enter. Information for the segment is displayed.
6 Select the inactive cup (use the arrow keys).
7 Press Enter. The tray is rotated to the correct position.
Then, the message Remove the cup from the instrument… is displayed.
8 Open the segment access cover and remove the inactive cup from the
position indicated by the arrow.
9 Close the cover and press F10.
10 Press Esc until the Main menu is displayed.
9 Press F10 again to save the edited requisition and begin processing if the
sample is already on the system.
? Note: The test will be run if the sample is already on the system
and has not been there for more than allowable time period.
When loading a batch of sample cups, an aliquot of each sample is poured into a sample
cup. Each sample cup is then placed in the designated numbered cup position on the
cup segment. After all of the sample cups have been filled and placed in their designated
positions, the filled cup segment is loaded onto the system and the tests are run.
? Note: A cup and segment placement list is created prior to loading a batch
of samples. The list shows each sample that is to be run and the numbered
position where it must be placed on the cup segment. The list can be printed.
5 Transfer the aliquot of sample to the cup and place the sample cup on the
designated cup position shown on the loadlist.
6 Press F7 to place the cup segment on the system. The messages
Pausing the Instrument… and Positioning the sample ring… are
briefly displayed before instructions for loading the cup segment are
displayed:
7 When the access door unlocks, the segment access cover can be opened.
The Rapid Tube Input (RTI) mode allows loading of multiple sample tubes into the
instrument without initiating processing. Enabling the RTI mode dedicates the probe to
aspirating samples from primary tubes placed into the STEP and dispensing these
samples into seglet wells. The instrument begins processing when the RTI mode is
disabled.
1 Press the Ctrl and F6 keys simultaneously to enable the RTI mode.
2 Place the first tube in the STEP following the instructions in section 7.1.2
on page 126.
4 When all tubes are loaded, press the Ctrl and F6 keys simultaneously to
disable the RTI mode and begin processing.
7.3 Results
Test results can be printed automatically when all tests for a requisition are completed,
or they can be printed on demand by the operator.
The patient results report contains demographic information for the patient, results for
the tests ordered on that requisition, normal ranges for those tests (if they are defined)
and a graphical representation of where the test results fall within the normal range:
D/T DRAWN: 08/13/04 13:30 BY: KMO D/T ENTERED: 08/13/04 14:39
ACCESSION: 719721 COMMENT: FASTING
APPROVED BY:
A quality control report contains identifying information for the control material, test
results, acceptable limits for those tests and a graphical representation of where the test
results fall within the limits. If a value is out of the control range, the value is identified
and space is provided for noting corrective action.
D/T: 08/13/04
09:30
CORRECTIVE ACTION:
A calibration report lists the data from the calibration runs for all tests on that requisition.
CALIBRATION REPORT
(- - - - - - - - - - - - - - - -RESPONSE- - - - - - - - - - - - - - - -) CONCENTRATION
(- - - - - - - - - - - - - - - -RESPONSE- - - - - - - - - - - - - - - -) CONCENTRATION
Results are automatically verified by the system as they are obtained. If a result is
outside the specified panic limits or if an error is detected during results calculation, the
VERIFY message is displayed (page 149) and the results must be manually verified
before the test is noted as completed.
1 Press F9 when the VERIFY message is displayed. The item number, the
reason for verification, the Accession number of the requisition, the test
name and the action to be taken are displayed.
2 Press Esc until the Main menu is displayed.
3 Select Requisition and then Patient (or Control).
4 Press F6.
5 Select the patient (or Control) requisition to be verified.
6 Examine the results of the test. Depending on the results, either delete the
test, delete the result, or press F10 to accept the result.
? Note: If the result is deleted, the test will be rerun. If the test is
deleted, it will not be rerun unless it is reordered.
Results from another instrument or other data can be entered as an off-line result, so
they can be included on the patient requisition and report.
? Note: Refer to Section 9.3.1 on page 199 for the procedure to set up an
off-line test.
Calculation test values, such as height or weight are used in derived test equations.
? Note: Refer to Section 9.3.3 on page 201 for the procedure to set up a
calculation test.
1 Select the desired patient requisition.
2 Select the calculation test (use the arrow keys).
3 Press Enter. The RESULT field is selected.
4 Enter a value for each field.
5 Press F10.
6 A prompt is displayed for entry of your initials and a comment.
7 Enter a comment.
8 Press F10.
Option Purpose
Quality Control Checks Aids in identifying trends. The system will flag results if
consecutive control values exceed a designated standard
deviation limit.
Action Purpose
Repeat Same Control(s) Do not assay or report patient results until the same
control(s) are repeated. If results from the repeated tests are
within the acceptable range of values, patient results may be
reported.
Re-calibrate If control results from the re-calibrated tests are within the
(where applicable) acceptable range of values, repeat the tests with patient
specimens.
Prepare Fresh Reagents If control results from the tests using freshly prepared
reagents are within the acceptable range of values, repeat
the tests using patient specimens.
The instrument precision, accuracy and linearity (up to three levels) for a test can be
generated by selecting options on the Diagnostics menu.
4 With the TEST: field selected, type the test mnemonic or press F2 and
select the desired test from the choice list. The NUMBER OF LEVELS:
field is selected.
5 Type the number of levels to be run (1 to 3) and press Enter. The
SAMPLES/LEVEL: field is selected.
6 Type the number of sample cups to be run for each level (1 to 60) and
press Enter. The RUNS/SAMPLE: field is selected.
7 Type the number of runs or replicates per sample (1 to 48).
12 If Assayed was selected, type a concentration and press Enter for each
level. If Unassayed was selected, proceed to the next step.
13 Press F10 when all parameters have been entered. The Precision Loadlist
screen is displayed.
14 Following the loadlist, place the samples into the designated positions of
empty cup segments.
15 Press F10, then place the first segment into the instrument when
prompted.
16 Press F10.
At the completion of the run, the statistics are calculated and a report is generated with
the following information:
?? Sample identification by segment number and cup position.
?? Sample concentration (if specified).
?? Mean, SD and % CV for samples with more than one replicate or run.
?? Grand Mean, SD and % CV on results of all samples within one level (with the
same concentration). If more than one replicate or run was performed on each
sample, the average of the results for each sample is used in the calculation.
?? Accuracy statistics reported as the percentage of the expected result
(concentration must be a value other than zero).
?? Linearity data calculated as slope, intercept, standard error and correlation
coefficient (if the concentrations were specified for more than one level).
4 With the TEST: field selected, press F2 and select the desired test. The
NUMBER OF SAMPLES: field is selected.
5 Type the number of samples to be run (1 to 60) and press Enter. The
RUNS/SAMPLE: field is selected.
6 Type the number of runs or replicates per sample (1 to 48).
10 Press F10 when all parameters have been entered. The Correlation
Loadlist screen is displayed.
12 Following the loadlist, place the samples into the designated positions of
empty cup segments.
13 Press F10, then place the first segment into the instrument when
prompted.
14 Press F10.
At the completion of the run, the statistics are calculated and a report is generated with
the following information:
?? Mean, SD and %CV for samples with more than one replicate per run.
?? Accuracy statistics reported as a percentage of the expected value for each cup
concentration. If more than one replicate or run per sample was performed, the
average of the results is used for each sample.
?? Linearity data calculated as slope, intercept, standard error and correlation
coefficient if different concentrations were specified for the cups. If more than one
replicate or run was performed per sample, the average of the results is used for
each sample.
Status Indicators
Status Indicators that displayed and the required operator action are summarized below.
PAUSED The system has been halted Complete the activity requiring
by the operator to allow access and close the access
instrument access. door. Press Esc until the
indicator is no longer displayed.
Biohazard: Federal, state and local laws and regulations require that
hazardous waste be disposed of in a specific manner. Waste material from
the ACE Alera System that may be classified as biohazardous includes used
sample cups and seglets. It is important that steps be taken to dispose of
these materials in accordance with the prevailing regulations in your location.
8 Data Management
A wide variety of reports are available on the system. These may be viewed on and
printed from the Workstation or can be transmitted to a laboratory information
management system (LIMS).
System data should be backed up on a regular basis and stored in a safe location. The
system data can then be restored if needed.
Finally, patient requisitions must be periodically deleted from the system in order to allow
additional free space to be available on the hard drive for new data. The patient
requisitions can be archived before deletion, if desired.
Report Description
Single Requisition Complete information for a specified patient requisition is
included. The information includes facility identification,
patient demographics, requisition date and time, Accession
number, comment field, test mnemonics, test results, result
flags, normal ranges (if defined), graphical display of the
result with respect to the normal range, and the reporting
units for the test.
8.1.2 QC Reports
Report Description
QC Requisition Complete information for a specified QC requisition is
included. The information includes facility identification,
control name and lot number, requisition date and time, test
mnemonic, test results, result flags, low and high acceptable
limits, graphical display of the result with respect to the
acceptable limits and the reporting units for the test. A
Corrective Action field is provided in the event of a failure.
Levey-Jennings Plots Cumulative statistics (up to 250 data points) for each analyte
are plotted. The report flags any values that violate QC
checking rules. Monthly reports can be obtained by
specifying the corresponding range of dates for the report.
Cumulative Statistics Cumulative statistics for each analyte are tabulated. This
report will display the mean and SD values entered from the
package insert for a single lot number of control until more
than 20 data points have been obtained if moving average
has been enabled.
Journal Reports System error messages and operator audit trail entries are
included. Journal reports can be prepared for routine events
such as result edits, maintenance procedures, errors
requiring operator intervention, and service procedures.
Test Parameters Prints the setup program parameters for a particular test.
Test Counts The number of tests run, since the counter was last reset,
for each on-line test.
1 Select Patient from the Reports menu. The Patient Reports menu is
displayed.
2 Select Single. The list of current patients is displayed.
3 Use the arrow keys to select the desired patient by name.
4 Press Enter. The Patient Accession list is displayed.
5 Use the arrow keys to select the desired Accession number and press
Enter. The report is generated and displayed.
6 Press F8 to obtain a hard copy of the report.
1 Select Patient from the Reports menu. The Patient Reports menu is
displayed.
2 Select Cumulative. A list of patients with at least one completed
requisition is displayed.
3 Use the arrow keys to select the desired patient by name.
4 Press Enter. A summary of all the requisitions in the database for the
patient is generated and displayed.
A merged requisition report displays all patient requisitions in the system for the selected
tests within a selected date range.
1 Select Patient from the Reports menu. The Patient Reports menu is
displayed.
2 Select Merged. A screen is displayed for entering the date and time range
for the report.
3 Type the beginning date, beginning time, ending date, and ending time for
the report.
? Note: Use the up and down arrow keys to move from one field to
the other. Use the left and right arrow keys to move within a field.
4 Press F10 when all information is entered. The message Preparing test
choice list is displayed on the screen, then the choice list is displayed.
5 Select the desired test (use the arrow keys).
6 Press Enter. A check mark is displayed next to the test.
7 Repeat steps 5 and 6 for all desired tests.
8 When all desired tests have been selected, press F10 to generate and
display the report.
9 Use the arrow keys or the page up and page down keys to view all
sections of the report.
8.2.4 Archived
Patient data that is archived on a 3.5 inch disk can be viewed and printed but not edited.
Refer to section 8.6.4 on page 174 for instructions on viewing and printing an archived
patient requisition.
If the system has been configured to require review of patient reports (refer to section
9.10.9 on page 221), each patient requisition must be reviewed before a report can be
printed.
1 Select Patient from the Reports menu. The Patient Reports menu is
displayed.
2 Select Review. A list of patients with requisitions ready for review is
displayed.
? Note: Only the most current requisition for a control can be viewed and
printed.
1 Select Quality control from the Reports menu. The QC Reports menu is
displayed.
2 Select Requisition. The Controls list is displayed.
3 Use the arrow keys to select the desired control and press Enter. The last
QC requisition report is generated and displayed.
4 Press F8 to obtain a hard copy of the QC Requisition report.
1 Select Quality control from the Reports menu. The QC Reports menu is
displayed.
2 Select Plots. The Test list is displayed.
4 Select the desired month (use the arrow keys) and press Enter. Levey-
Jennings plots for each test during the selected month are created and
can be printed.
1 Select Quality Control from the Reports menu. The QC Reports menu is
displayed.
2 Select Plots. The Test list is displayed.
3 Select the desired test from the list (use the arrow keys)
4 Press Enter. Levey-Jennings plots for the selected test are generated and
data points for the first control are displayed.
? Note: Press F5 to display data points for the next control on the
same test.
1 Select Quality control from the Reports menu. The QC Reports menu is
displayed.
2 Select Plots. The Test list is displayed.
3 Select the desired test from the list (use the arrow keys).
4 Press Enter. Levey-Jennings plots for the selected test are generated,
and the data points for the first control are displayed.
5 To add numeric results to the plotted results, press F6. To add cumulative
and filtered statistics for the control to the plotted results and numeric
results, press F6 again.
6 To filter the data by month, press F7, select the desired month, and press
F10. The data is filtered.
7 Press F8 to print a copy of the plot(s).
8.3.5 QC Statistics
The cumulative statistics report lists the estimated or established mean and standard
deviation (SD) values, the number of data points and the actual mean, actual SD and
actual coefficient of variation (% CV) of all the control results for each test run on each
control.
1 Select Quality control from the Reports menu. The QC Reports menu is
displayed.
2 Select Statistics. The Control Cumulative Statistics report is generated
and displayed.
3 Press F8 to obtain a hard copy of the report.
? Note: The information displayed can be filtered to exclude the most recent
requisitions or to include only on-line or off-line tests on the report.
1 Select Test Counts from the Reports menu. The report is generated and
displayed:
1 Select Calibration from the Reports menu. A list of tests with stored
calibration curves is displayed.
2 Select the test to be reviewed (use the arrow keys).
3 Press Enter. The Calibration report is generated and displayed.
4 Press F8 to obtain a hard copy of the report.
1 Select Raw Data from the Reports menu. A list of samples with raw data
available is displayed.
2 Select, by name, the desired sample (use the arrow keys) and press
Enter. A list of tests for the sample is displayed.
3 Select the desired test (use the arrow keys) from the list.
4 Press Enter. The Raw Data report is generated and displayed.
5 Press F8 to obtain a hard copy of the report.
The Test Parameters report displays a table of the parameters for each test stored in the
database.
3 Enter the date range for the report. Press F2 to display a list of events.
4 Highlight an event (using the arrow keys) and press Enter to select an
individual event.
4 Press F10. A status message is displayed, and then the name of each file
is displayed as it is copied to the 3.5 inch disk.
5 When the backup is complete, instructions to remove the 3.5 inch disk are
displayed.
6 Remove the 3.5 inch disk and press any key. The system is initialized and
the Utilities menu is displayed.
7 Label the 3.5 inch disk and store it in a safe place.
7 Press F10. A status message is displayed, and then the name of each file
is displayed as it is copied from the disk to the system.
8 When the restoration is complete, instructions to remove the 3.5 inch disk
are displayed.
9 Remove the backup disk and press any key. The system is initialized and
the Utilities menu is displayed.
? Note: This option allows you to archive the patient data onto a
disk without deleting the patient data on the system.
2 Type the date after which inactive patients and/or requisitions should be
archived.
6 Press F10. If patient files have never been archived on the system, a
message to confirm the creation of an archive history log is displayed.
7 Press F10. The archive log is created and instructions to insert a 3.5 inch
disk are displayed.
? Note: If patient files were previously archived on the system but the
archive log was not found, press Esc and restore the log using the
Utilities menu.
Each time an archive is performed, the archive history is updated on the hard drive (on
the archive log). The updated archive log is included with the patient files whenever an
archive is performed. If necessary, the archive history log can be restored from the most
recent archive disk.
1 Select Utilities from the Main menu.
2 Select Restore. The Restore menu is displayed.
3 Select Archive Log. If an archive log already exists on the hard drive, a
warning is displayed:
6 Press F10. A message is displayed as the archive log is copied from the
3.5 inch disk to the system.
7 When the restoration is complete, instructions to remove the 3.5 inch disk
are displayed.
8 Remove the disk and press any key. The system is initialized and the
Restore menu is displayed.
2 Type the date after which inactive patients and/or requisitions should be
archived and then deleted, and then press Enter.
4 Press F10. The patient data selected for archiving and deletion is
displayed.
7 Press F10. If patient files have never been archived on the system, a
message to confirm the creation of an archive history log is displayed.
8 Press F10. The archive log is created and instructions to insert a 3.5 inch
disk are displayed.
? Note: If patient files were previously archived on the system but the
archive log was not found, press Esc and restore the log using the
Utilities menu.
2 If you wish to delete patient data without archiving to disk, press F10. A
prompt is displayed to enter the starting date for the period of inactivity
and to specify the records that are to be deleted.
3 Type the date after which inactive patients and/or requisitions should be
deleted, and then press Enter.
6 Review the list of patients. Delete any from the purge list, if desired.
7 Press F10. The selected patient data is deleted from the system.
2 Select Delete Test Counts. The Test Counts screen is displayed listing
the test name, the number of tests run since the last reset and the date the
counter was last reset.
? Note: This indicates all test counters have been marked for
reset.
4 Press F10. All counters are reset to zero and the changes are saved.
2 Select Delete Test Counts. The Test Counts screen is displayed listing
the test name, the number of tests run since the last reset and the date the
counter was last reset.
Individual tests and requisitions for a patient can be deleted if archiving is not required. If
all tests on a requisition are to be deleted, the entire requisition must be deleted. If all
requisitions for a patient are to be deleted, the patient must be deleted.
1 Select Patient from the Requisitions menu. A list of patients that have
already been entered into the system is displayed:
2 Select the desired patient (using the arrow keys) and press Enter. A list of
requisitions for the patient is displayed:
3 Select the desired requisition (using the arrow keys) and press Delete. A
message to confirm the deletion is displayed.
4 Press Y. The requisition is deleted.
2 Select the desired patient (using the arrow keys) and press Delete. The
message This patient will be removed from the system. Are you sure
you want to do this (Y/N)? is displayed.
3 Press Y. The patient is deleted.
? Note: If the patient has at least one test without a result, the
message This patient has at least one incomplete requisition.
Do you want to delete this patient anyway (Y/N) ? is displayed.
Press Y to delete the patient.
9 System Setup
The system hardware and software can be configured to meet the needs and
requirements of an individual laboratory. Configuration is carried out using the Setup
menu.
1 Select Setup from the Main menu. The Setup menu is displayed:
2 Select ENABLE: (under ISE) and press F2. The screen displays YES and
NO options.
3 To enable the ISE, select YES and press Enter. To disable the ISE, select
NO and press Enter.
4 Press F10 to save the changes.
? Note: Once enabled, the ISE module should only be disabled if the
system is scheduled for a long-term period of inactivity. See the
Maintenance section for instructions for long-term shutdown of the
ISE.
The ISE can be run in either Standby mode or STAT Ready mode.
? Note: If the ISE is set to Standby mode, calibration is only performed when a
sample is introduced for ISE testing and the last calibration was more than
three hours ago. This reduces the amount of CAL-A and CAL-B used by the
system.
If the ISE is set to STAT Ready mode, calibration is performed every three
hours even if samples are not processed.
2 Select CALIBRATION: (under ISE) and press F2. The screen displays
STANDBY and STAT READY options.
3 Select the desired mode and press Enter.
4 Press F10 to save the changes.
The ACE Alera System can be interfaced with an external data management system that
downloads requisitions and accepts the transmission of completed requisitions
(unidirectional or bidirectional). Certain parameters must be entered when the LIMS is
enabled. Specific values for these parameters are supplied with the LIMS system.
2 Select ENABLE: (under LIMS) and press F2. The screen displays YES
and NO options.
3 Select YES and press Enter. The BAUD RATE: field is selected.
? Note: Select NO and press Enter to disable the LIMS. Then, press
F10 to save the changes
4 Press F2. The screen displays 1200, 2400, 4800 and 9600 options.
5 Select the appropriate baud rate and press Enter. The PARITY: field is
selected.
6 Press F2. The screen displays NONE, ODD and EVEN options.
7 Select the appropriate parity and press Enter. The DATA BITS: field is
selected.
8 Type the appropriate number (7 or 8) of data bits and press Enter. The
STOP BITS: field is selected.
9 Type the appropriate number (1 or 2) of stop bits.
10 Press F10 to save the changes.
The ACE Alera System can be configured so that a patient first name is required in all
patient requisitions.
2 Select FIRST NAME REQUIRED: and press F2. The screen displays YES
and NO options.
3 To specify that the patient’s first name is required in a patient requisition,
select YES and press Enter. Otherwise, select NO and press Enter.
An audible alarm sounds when an attention message is displayed. The sound level of
the alarm can be adjusted.
2 Select LEVEL: (under BEEPER LEVEL) and press F2. The screen
displays HIGH, MEDIUM and LOW options.
3 Select the desired level and press Enter.
4 Press F10 to save the changes. The system will reinitialize and return to
the Setup menu.
The ACE Alera System can be equipped with an optional barcode reader. The default
scanner type is PSC.
2 Select TYPE: (under BARCODE SCANNER) and press F2. The screen
displays NONE, MSCAN and PSC options.
3 Select the desired option and press Enter.
4 Press F10 to save the changes. The system will reinitialize and return to
the Setup menu.
The default label symbology parameters are preprogrammed. They are applicable to
barcode labels for Code 39, Code 128 Set B and Set C, Codabar or Interleaved 2 of 5.
1 Select Label Symbology from the Main menu. The Label Symbology form
is displayed:
2 (If needed) Select the first field (using the arrow keys) for the type of
barcode label you are using.
3 Type a value in the field and press Enter. The next field is selected.
4 Repeats Steps 2 and 3 for each label parameter.
Sample lifetime is the length of time (in hours) that a sample may remain on the
analyzer awaiting testing before it is considered expired. The status of a sample is
automatically changed from active to expired if this lifetime is exceeded. If there are
pending tests for the sample, the operator is notified that the sample should be replaced.
2 Select HOURS: (under SAMPLE LIFETIME) and press F2. The screen
displays the sample lifetime options.
3 Select the desired option and press Enter.
The ACE Alera System can be configured so that leading zeros are removed from
Accession numbers that are read from barcode labels or entered by the operator.
A few of the test parameters can be customized to accommodate the needs of the
individual laboratory. For example, reference intervals and QC checking rules can be
modified in the individual test parameters.
1 Select Tests from the Setup menu. The Setup Tests list is displayed,
showing all tests that are defined in the system and the date each test was
last edited:
The parameters that can be modified within each parameter group are summarized
below:
? Note: The test mnemonic is an abbreviation for the test name. The
abbreviation can consist of up to six (6) alphanumeric characters,
but it must begin with a letter.
? Note: Predilution ratios and auto rerun dilution ratios are the only
fields which may be modified.
2 To modify the predilution ratio, select the RATIO: field of the PREDILUTE:
row.
Then, type the predilution ratio (between 2 and 101) and press Enter.
3 To modify the dilution ratio used during automatic rerun, select the RATIO:
field of the AUTO RERUN: row.
Then, type the dilution ratio (between 2 and 50) and press Enter.
This screen is used to allow reporting of results that are less than the lowest calibrator
(LT LC) or less than zero (LT 0). It can also be used to change the concentration of the
calibrator(s) used for a test.
? Note: The calibrator lot number can not be edited from this
screen since it is not test specific. Use the Setup Calibrators
screen
(page 204) to update calibrator lot numbers.
1 Select Result Type. The result type parameters are displayed for the
selected test:
2 To modify the result units, select the UNITS: field, press F2, and select the
appropriate units.
A range of normal or expected values can be assigned for each test, and separate
normal ranges can be established based on gender and/or age.
The system can automatically check test results against these normal ranges and flag
any result that is outside of the normal range. The system can also check results against
an upper and lower panic limit and display a VERIFY warning if results are outside the
limits.
The system can automatically check control value statistics. A notation is made on the
QC Requisition Report if one of the selected checks is violated. These values will also
be flagged on any Quality Control plots.
1 Select Quality Control Checks. The quality control check parameters are
displayed for the selected test:
1 Select Tests from the Setup menu. The Setup Tests list is displayed:
2 Press Insert. The types of tests that can be created are displayed:
? Note: The parameter groups that are displayed depend upon the
type of test that is to be created.
Derived Tests Tests with results that are calculated from the results of other
tests. For example, a BUN/Creatinine Ratio may be a derived by
dividing BUN results by the Creatinine results..
Calculation Tests Information other than a chemistry test result that may be
required in the calculation of a derived test result. For example,
body weight may be required.
Remote Tests Tests that are run using more than one sample. The results of
(Linking Tests) remote tests are linked so they can be included on a patient
report.
1 Select Tests from the Setup menu. The Setup Tests list is displayed.
2 Press Insert. The types of tests that can be created are displayed.
3 Select OFF-LINE from the Test Type menu. The Parameter Groups menu
for offline tests is displayed:
4 Define the test identification, result type, reference range checks, and
quality control checks. Refer to pages 192 through 197 for details.
5 Press F10 when all parameters have been entered to save the test and
return to the Setup Tests list.
6 Press Esc to return to the Setup menu.
4 Define the test identification, result type and reference range checks.
Refer to pages 192 through 196 for details.
5 Select Calculation Method. A prompt is displayed for the equation
needed to calculate the result of the derived test:
4 Define the test identification and result type. Refer to pages 192 and 195
for details.
5 When all parameters have been entered for the calculation test, press F10
to save the test and return to the Setup Tests list.
6 Press Esc to return to the Setup menu.
1 Select Tests from the Setup menu. The Setup Tests list is displayed.
2 Press Insert. The types of tests that can be created are displayed.
3 Select REMOTE from the Test Type menu. The Parameter Groups menu
for remote tests is displayed:
6 Type the mnemonic for the source test and press Enter.
Or: Press F2 to display a list of tests and test panels that are available.
Select the desired test (use the arrow keys) and press Enter:
9.4 Calibrators
Up to 250 different calibrators can be defined on the system at one time. The name of
the calibrator, the lot number, the analytes it contains, the concentration of each analyte,
and the interval between calibrations for each analyte is specified.
The number of calibrators for a test and the specific calibrators that are used are defined
in the calculation parameters (page 194).
Caution: The number of calibrators for a test and the calibration frequency are
preprogrammed by Alfa Wassermann Diagnostic Technologies, LLC and have
been used for method validation. Modification to these parameters without
manufacturer’s recommendation will invalidate performance claims.
Once calibrators for a test are defined, the lot number and analyte concentrations in a
calibrator can be conveniently modified. For example, when using a multi-constituent
calibrator (e.g., GEMCAL? ), the lot number and concentrations of all constituents can be
updated from one screen.
? Note: Calibrator concentrations for an individual test can also be edited from
the calculation method parameters screen for that test (see page 194).
1 Select Calibrators from the Setup menu. A list of calibrators that are
currently defined on the system is displayed:
2 Select the desired calibrator and press Enter. The Setup Calibrators
screen for the selected calibrator is displayed with the entered values for
the calibrator:
3 (If needed) Type a new lot number for the calibrator and press Enter.
4 Select the CONCENTRATION field for each constituent and modify the
value as desired.
5 Press F10 to save all changes.
1 Select Calibrators from the Setup menu. A list of calibrators that are
currently defined on the system is displayed.
2 Select the desired calibrator and press Delete. The selected calibrator is
removed from the list.
9.5 Controls
Up to 20 different controls can be defined on the system at one time. A particular test
(albumin, for example) can be specified in up to six defined controls.
1 Select Quality Control from the Setup menu. A list of controls that are
currently defined on the system is displayed:
5 Enter the expected values (mean and SD) for that test.
6 Repeat Steps 4 and 5 for each test for which the control will be used.
7 When all tests have been entered, press F10 to save the entries.
1 Select Quality Control from the Setup menu. The list of controls is
displayed.
2 Select the desired control and press Enter. Information for the selected
control is displayed:
3 Using the arrow keys, select the first TEST field for which expected values
are to be updated. Press Enter to select the MEAN field for that test.
4 Type the new mean value and press Enter. The SD field for that test is
selected.
5 Type the new SD value and press Enter.
6 Repeat Steps 3 and 4 for each test for which expected values are to be
updated.
7 To delete a test from the control, select the test and press Delete.
? Note: If a test is deleted, all control data for that test is lost.
1 Select Quality Control from the Setup menu. The list of controls is
displayed.
2 Select the desired control and press Enter. Information for the selected
control is displayed.
3 Using the arrow keys, select the LOT: field.
4 Type the new lot number and press Enter. A warning message is
displayed:
5 Type Y to purge accumulated data for the existing lot number and print an
end-of-lot report.
6 If needed, enter updated expected values (page 206).
1 Select Quality Control from the Setup menu. The list of controls is
displayed.
2 Select the control that is to be deleted and press Delete. A warning
message is displayed:
Open channel bottles are available from your Alfa Wassermann Diagnostic
Technologies, LLC distributor.
1 Select Bottles from the Setup menu. A list of reagent bottles that are
currently defined on the system is displayed:
1 Select Panels from the Setup menu. The Panels list is displayed.
3 Type a name for the panel in the PANEL NAME: field and press Enter.
The field for the first test to be included in the panel is selected.
1 Select Panels from the Setup menu. The Panels list is displayed.
2 Select the desired Panel and press Enter to display the Setup Panels
screen for the selected Panel.
3 Edit the Panel as required.
4 Press F10 to save the changes.
1 Select Panels from the Setup menu. The Panels list is displayed.
2 Select the desired Panel and press Delete.
3 The panel is deleted from the list of panels.
1 Select Doctors from the Setup menu. The Doctors list is displayed.
3 Type a name for the doctor in the NAME: field and press F10. The
doctor’s name is added to the list.
1 Select Doctors from the Setup menu. The Doctors list is displayed.
2 Select the desired name and press Enter to display the Setup Doctors
screen.
3 Edit the name as required.
4 Press F10 to save the changes.
1 Select Doctors from the Setup menu. The Doctors list is displayed.
2 Select the desired name and press Delete.
3 The doctor’s name is deleted from the list.
1 Select Carryover from the Setup menu. The reagents that are currently
defined on the system are displayed:
3 In the CONTAMINATES: section, type the name of the reagent that the
selected reagent affects and press Enter. Repeat for any other reagents
that the selected reagent affects.
4 In the DECONTAMINATES: section, type the name of the reagent that the
selected reagent can be used as a rinse for and press Enter. Repeat for
any other reagents that the selected reagent will decontaminate.
5 Press F10 to save the entries.
Option Purpose
Password Specify whether a password is required in order to provide
system security.
Facility ID Specify the facility name that appears in the header of reports.
Report Footer Specify the text that is displayed in the footers of patient
requisitions and merged reports.
Date Format Specify the date format (MM/DD/YY or DD/MM/YY) that is used
whenever a date is entered or reported.
Reporting Options Specify whether patient reports must be reviewed and approved
before they are printed, whether QC reports and calibration
reports are printed automatically, and whether requisitions are
automatically transmitted to a LIMS.
9.10.2 Password
The system may be configured with a multi-level password system. Up to three levels of
password protection may be defined. A unique password must be entered in order to
access the menu options on that level.
? Note: The procedure for defining the levels and their corresponding passwords
is not given here for security reasons. Contact Technical Service for instructions.
Once the multilevel password system is set, level two and three passwords can
be set using the following steps:
9.10.3 Reminders
? Note: Use the 24-hour time format (e.g., 15:30 for 3:30 p.m.).
2 Type the facility name, address, phone number, and director as desired.
When the archiving mode is enabled, patient requisitions can not be deleted from the
database until they are copied to a floppy disk.
2 Press F2 to display choices (YES and NO). Select the choice (using the
arrow keys) and press Enter.
3 (If needed) Select the AUTO ACCESSION RESET: field (using the arrow
keys) and press F2. Choices (YES and NO) are displayed. Select the
AUTO ACCESSION RESET: field (using the arrow keys) and press Enter.
The Reporting Options screen allows the user to configure the following options:
?? Review of patient requisitions before printing.
?? Automatic printing of QC and calibration reports.
?? Transmission of reports to a LIMS.
?? The basis for calculating QC cumulative statistics.
These tools are primarily for the use of Alfa Wassermann Diagnostic Technologies, LLC
technical service personnel. Do not use these tools unless directed to do so by an Alfa
Wassermann Diagnostic Technologies, LLC representative.
1 Select Utilities from the Main menu. The Utilities menu is displayed.
2 Select Date/time. The screen displays the message Placing system in
standby…
Then it displays a prompt with the date and time as it is currently entered
in the system:
5 Press F10 to save the entries and initialize the system. The Utilities menu
is displayed.
1 Select Utilities from the Main menu. The Utilities menu is displayed.
2 Select Software. The Software menu is displayed.
3 Select View to display the software version number and the date and time
it was released.
4 Press Esc until the Main menu is displayed.
The user interface text can be displayed in English (default), French, German, Italian or
Spanish.
1 Select Utilities from the Main menu. The Utilities menu is displayed.
2 Select Language. The Select Language screen is displayed.
3 Press F2 at the LANGUAGE: field and select the desired language from
the choice list.
4 Press F10 to save the selection. The system will convert the user interface
to the selected language and return to the Utilities menu.
10 Maintenance
Periodic maintenance procedures must be performed on the ACE Alera System in order
to assure consistent efficient performance or to repair or replace user-serviceable
components.
Certain procedures are to be performed at a specified scheduled frequency whereas
unscheduled maintenance procedures are performed on an as-needed basis.
All maintenance procedures performed on the instrument must be documented in a
maintenance log in order to comply with CLIA 88 requirements.
The required preventive maintenance procedures for the system are summarized below:
Monthly (MM-1) Rinse Probe and Fluid Lines with Bleach 241
3 When the access door unlocks, the segment access cover and reagent
cover can be opened.
4 If the reagent tray is in the compartment, lift it out and set it aside.
5 Inspect the compartment and camera lenses for condensation. Remove
any moisture using the lint-free cloth (Figure 36).
6 Replace the reagent tray.
7 Close the cover and press F10.
8 Press Esc until the Main menu is displayed.
Recessed Camera
Lenses
? Note: The probe will return to its home position over the wash bath.
1 Select Utilities from the Main menu to display the Utilities menu.
2 Select Probe to display the Align Probe screen.
3 Select Position Transfer Arm to ISE Port and press Enter. The probe is
positioned at the ISE port crosshairs.
4 Examine the position of the probe. If necessary, carefully bend the probe
so that the tip of the probe is positioned at the center of the crosshairs.
? Note: If you need to bend the probe, be sure to press firmly near
the top of the probe.
5 Select Test Alignment at Wash Bath and press Enter. The probe is
positioned at the wash bath and the alignment is automatically checked.
6 Observe the probe during the test. Verify that it is positioned in the middle
of the “U”located on the left side of the alignment plate.
7 Repeat the alignment and test procedure until the alignment passes.
8 Press Esc until the Main menu is displayed. The system is initialized.
1 Select Diagnostics from the Main menu to display the Diagnostics menu.
2 Select Troubleshoot to display the Troubleshoot menu.
3 Select ISE. The message Preparing for diagnostics… is displayed
followed by the Troubleshoot ISE menu.
4 Select Operate Valves/Pump to display the ISE Troubleshoot screen.
6 Select AIR: and press the spacebar to activate the pump and valve.
7 When the Fluid sensor field on the screen changes from LIQUID to AIR,
wait five seconds and press the spacebar to halt the pump.
8 Pinch the plastic tabs on either side of the measuring chamber and pull
the chamber forward.
9 Unclamp the electrodes by moving the lever on the left electrode holder
forward.
20 After 15 minutes, remove the reference housing from the bleach using
forceps or tweezers.
21 Thoroughly rinse reference housing with tap water, following with a final
rinse using distilled or deionized water.
22 Fill the service syringe with distilled or deionized water.
23 Place the service syringe tip in the reference housing opening containing
the black o-ring. Gently push water through the fluid path.
24 Place your index finger over the opening of the fluid path on the side of the
housing opposite the o-ring. Hold the reference housing with the large
opening pointing downward towards the back wall of the sink.
25 Press the tip of the service syringe into the opening and push the plunger.
A fine stream of water should be visible coming out of the large opening.
33 Wipe up any fluid that may have leaked from the tubing and close the
door.
34 Press Esc until the Main menu is displayed.
35 Select Utilities, then ISE. Select CALIBRATE and WASH PORT.
WARNING: The power to the UPS must be OFF before removing and
replacing the filters.
1 With the Main menu displayed, type X to exit DOS, then turn off the power
to the system.
2 Disconnect the analyzer power cord.
3 Remove the air filter under the left side of the analyzer.
? Note: Use your index finger to locate the corner of the filter and
pull it out.
? Note: Open the filter access door, grasp the tab on the filter, and
pull the filter out.
5 Inspect the filters for dust accumulation. If cleaning is required, wash the
filters thoroughly under running tap water, and rinse thoroughly.
? Note: A mild detergent can be used. Dry the filters with a lint-free
cloth before reinstalling. The filters must be completely dry
before they are replaced.
6 Replace the air filter under the left side of the analyzer.
? Note: Hold the filter by the corner with the tab facing towards the
front. Slide the filter into the holder, pushing it in as far as it will
go.
? Note: Open the filter access door, hold the filter by its tab, insert
the filter, and close the filter access door.
8 After both filters have been reinstalled, turn the power on at the UPS.
9 Allow the system to initialize and the temperature to equilibrate.
1 Dampen a cloth with mild detergent solution. Wipe the top, front, and sides
of the analyzer.
2 Dampen a cloth with tap water. Wipe the instrument surface to remove
any detergent.
3 Dry with a lint-free cloth.
1 Press Fn and F2 to display the Access menu. The probe/transfer arm will
move to the far left to allow better access to the touchplate (Figure 38).
2 Select Reagent.
3 Briskly wipe the touchplate with an alcohol prep pad.
4 Use a pipette to place a few drops of distilled or deionized water on the
touchplate. Rub briskly with a gauze pad or lint-free cloth.
Touchplate
Figure 41. Scraping the Touchplate Opening with an Unfolded Paper Clip
1 Select Diagnostics from the Main menu to display the Diagnostics menu.
2 Select Miscellaneous to display the Miscellaneous menu.
3 Select Prime Syringe to display the Prime Syringe Diagnostic screen is
displayed
4 Select Perform monthly maintenance. The screen displays the first step
of the maintenance routine.
5 Open the container access door.
6 Remove the system diluent container by tilting the bottom of the bottle up
and sliding it out of the compartment.
7 Push the metal release button on the tubing (next to the cap) and then
disconnect the tubing (see Figure 30 on page 96).
8 Remove the cap assembly and set the diluent aside.
9 Prepare approximately 200 mL of a 10% (v/v) solution of household
bleach in distilled or deionized water. Place the solution in an empty
system diluent container and attach the cap assembly.
10 Install the filled system diluent container in the compartment and close the
door.
11 Follow the screen prompts, pressing Enter after completion of each
requested operator action.
12 When the maintenance routine is complete, press Esc until the Main menu
is displayed. The system is initialized.
Biohazard: Waste material from the ACE Alera System which may be
classified as biohahardous includes waste fluid dispensed into the liquid
waste container. Handle cautiously in accordance with laboratory safety rules
and local regulations.
? Note: Clean one cap at a time beginning with the CAL-B cap assembly. Make
sure that the CAL-B cap assembly is reinstalled on the CAL-B container before
removing the Reference Solution cap assembly, as the CAL-B is easily
contaminated by the Reference Solution.
1 Unscrew the container from the cap assembly and place the container
aside.
2 Using a moistened cloth, remove any crystals that may have formed on
the cap assembly.
3 Wipe up any crystals that may have fallen into the solutions compartment.
4 Install new container of CAL-B.
5 Repeat with Reference Solution as described in Section 5.5.
1 Select Diagnostics from the Main menu to display the Diagnostics menu.
2 Select Hardware Calibration to display the Hardware Calibration menu.
3 Select Analyzer to display the Analyzer menu.
4 Select Table offset. A series of messages is displayed and the table
offset calibration procedure begins.
5 At the completion of the table offset calibration, the Analyzer Hardware
Calibration menu is displayed.
6 Select Sample delay and optics. A series of messages is displayed.
7 At the completion of the optical calibration, the Analyzer Hardware
Calibration menu is again displayed.
8 Press Esc until the Main menu is displayed. The system is initialized.
10 Check for fluid leaks below the peristaltic pump and around the fittings.
11 Close the compartment door.
12 Press Esc to display the Troubleshoot ISE menu.
13 Select Prime to calibrate the peristaltic pump speed to the new pump
tubing.
14 Press Esc until the Main menu is displayed. The system is initialized.
Refer to the instructions found in the manual supplied with the printer.
The LED indicator light illuminates when the toner is low. The printer will print
approximately 100 more pages before running out of toner. The toner cartridge needs to
be replaced every 2,500 pages. The drum needs to be replaced every 25,000 pages.
Refer to the instructions found in the manual supplied with the printer.
WARNING: The power to the system must be OFF before removing and
replacing the probe.
9 Grasp the black bracket at the top of the new probe and slide the probe
stem into the Teflon tube opening (at the center of the end of the fluid arm
assembly), taking care to avoid contact with the flared sealing surface.
10 Push the probe stem in as far as it will go. Hold it in place while tightening
the Allen screws in the following step.
? Note: Be sure to push the probe stem as far forward as it will go.
The level sensing contact at the top of the probe must touch the
level sensing contact of the fluid arm assembly PC board.
11 Tighten each Allen screw one turn at a time, alternating between the
screws so that the two screws are tightened together. Continue alternately
tightening each screw one turn at a time until the probe stem is snug.
Then, turn each Allen screw an additional 1/8 turn clockwise.
12 Place the cover on the fluid arm assembly. Tighten the Allen screw by
turning it clockwise.
13 Make sure the fluid arm assembly is pushed down so that the probe tip
just misses the top of the main deck.
14 Turn the power on at the UPS.
15 Allow the system to initialize. When the Main menu is displayed, align the
probe. Refer to Section 10.1.4 on page 232 for details.
1 Select Diagnostics from the Main menu to display the Diagnostics menu.
2 Select Troubleshoot to display the Troubleshoot menu.
3 Select ISE. The message Preparing for diagnostics… is displayed
followed by the Troubleshoot ISE menu.
4 Select Operate Valves/Pump to display the ISE Troubleshoot screen.
8 Pinch the plastic tabs on either side of the measuring chamber and pull
the chamber forward.
9 Unclamp the electrodes by moving the lever forward on the left electrode
holder.
10 Carefully remove the electrode to be replaced from the measuring
chamber taking care not to dislodge the small, black o-rings between the
electrodes and in the electrode holder.
11 Remove the new electrode from its box and check for the presence of an
o-ring in the left side of the electrode.
12 Install the new electrode in its labeled position in the Measuring Chamber.
? Note: Be sure the lip on the bottom of the electrode rests on the
flat edge of the measuring chamber to aid in proper positioning.
? Note: Use the fingers of your right hand to hold the electrodes in
place.
14 Close the clamp on the left electrode holder by lifting the lever upward until
it locks in the back position.
15 Slide the measuring chamber backward until it locks completely into
position.
16 Select CAL-A: and press the spacebar.
17 Verify that fluid moves into the chamber.
18 If it does not, reseat the electrodes.
19 When the Fluid sensor field changes from AIR to LIQUID, wait five
seconds and press the spacebar to halt the pump.
20 Close the container access door.
21 Press Esc until the Main menu is displayed. The system is initialized.
1 Select Diagnostics from the Main menu to display the Diagnostics menu.
2 Select Troubleshoot to display the Troubleshoot menu.
3 Select ISE. The message Preparing for diagnostics… is displayed
followed by the Troubleshoot ISE menu.
4 Open the container access door.
5 Carefully slide the pump tubing from the pump roller and slide the plastic
pump plate from its support.
6 Grasp the pump roller firmly and pull it forward to remove it from the shaft.
7 Obtain the new pump roller. Push it onto the shaft.
? Note: The pump roller is keyed to fit the shape of the shaft. It will
only slide on one way.
8 Slide the pump plate back onto its support. Stretch the tubing over the
rollers, making sure that the pump tubes are parallel to each other and do
not cross.
9 Check the installation of pump tubing as outlined in Section 10.4.1 on
page 245.
1 Select Diagnostics from the Main menu to display the Diagnostics menu.
2 Select Troubleshoot to display the Troubleshoot menu.
3 Select Tube entry to display the Tube Entry menu.
4 Select Move Motors to display the STEP Motors menu.
5 Select Close Door to close the STEP Door.
6 Remove the STEP door cover.
? Note: Remove the STEP door cover by pushing down at the rear
of the door to push the front of the door up. The door is held in
place by magnets, and it can now be removed by grasping the
front end and lifting it out.
? Note: Use a flat blade screwdriver and turn the screws counter-
clockwise to remove them.
8 Use the blade of the screwdriver to lift the tube piercing device slightly.
9 Turn the tube piercing device so that it clears the analyzer cover. Remove
it from the STEP.
10 Install the replacement tube piercing device, holding it by the flat mounting
bracket on the top and with the bevel edge facing away from the operator.
11 Turn the replacement tube piercing device slightly to clear the analyzer
cover. Avoid touching the sharp piercing tip.
12 Secure the replacement tube piercing device with the two screws.
13 Replace the STEP Door cover. Verify that the STEP door cover is
centered and that it is properly seated.
14 Check operation of the STEP door by selecting Open Door, Close Door
and Open Door.
? Note: If the STEP door does not open and close smoothly,
remove the cover, loosen the screws to the door and center it.
? Note: If the piercing device does not operate properly, repeat the
entire procedure, checking that the piercing device is secured
properly. Contact your Alfa Wassermann Diagnostic
Technologies, LLC technical representative for assistance.
CAUTION: Failure to follow the ISE shutdown procedure when turning the
analyzer power OFF for an extended period of time can result in damage
to the electrodes, reference housing and tubing.
4 Remove the CAL-B container. Remove the CAL-B cap assembly from the
container, and cap the container using an extra CAL-B (red) container cap.
5 Install an empty CAL-B container on the analyzer. Cap it using the CAL-B
(red) cap assembly.
6 Remove the Reference Solution container and remove the Reference
Solution cap assembly. Cap the container using an extra Reference
Solution (white) container cap.
8 Press F10 to drain the ISE system. When drainage is complete, the
Miscellaneous menu is displayed.
9 Fill an empty CAL-A container with distilled or deionized water and cap it
using the CAL-A cap assembly. Place the filled and capped CAL-A
container on the analyzer.
10 Remove the CAL-B container and discard any drained fluids. Fill the
CAL-B container with distilled or deionized water and place it on the
analyzer. Cap the container using the CAL-B cap assembly.
11 Remove the Reference Solution container and discard any drained fluids.
Fill the Reference Solution container with distilled or deionized water and
place it on the analyzer. Cap the container using the Reference Solution
cap assembly.
12 Pinch the plastic tabs on either side of the measuring chamber and pull
the chamber forward.
13 Unclamp the electrodes by moving the lever forward on the left electrode
holder and carefully remove the reference housing, taking care not to
dislodge the small black o-rings between the electrodes and the right
electrode holder.
14 Holding the white collar between thumb and index finger, carefully
unscrew the reference electrode from the housing and temporarily place it
on a clean, soft cloth or towel.
15 Use a transfer pipette to transfer the reference solution from the reference
housing to the ISE transport housing.
16 Screw the reference electrode into the ISE transport housing. Carefully
screw the ISE transport housing into the reference housing. Allow the
reference housing to rest in front of the module on a soft cloth or towel.
17 Check that there is an o-ring present on the left side of the dummy
electrode.
18 Taking care not to dislodge the o-ring, install the dummy electrode in the
measuring chamber next to the left electrode holder.
19 Making sure that the electrodes are seated properly, close the electrode
holder by lifting the lever upward until it locks in the back position.
20 Select Prime ISE to prime ISE module with water.
21 When the prime cycle is complete, select Prime ISE a second time.
22 When the second prime is complete, remove the containers of water from
the CAL-A and CAL-B positions and replace them with empty containers.
23 Remove the Reference Solution (white) cap assembly from the Reference
Solution container. Empty the container and then place it under the
Reference Solution (white) cap assembly to catch water as it is draining.
? Note: Rest the white cap assembly on the container but do not
tighten the cap, as this will cause the tubing to fill with water
instead of air.
24 Select Drain ISE and then press F10 to drain the water from the ISE.
25 When complete, select Drain ISE a second time.
26 When the second drain cycle is complete, carefully unscrew the ISE
transport housing with the reference electrode from the reference
electrode housing. Place the ISE transport housing and electrode in the
storage box.
27 Clean the reference housing.
28 The reference housing and the sodium, potassium and chloride electrodes
may either be left in the measuring chamber or removed and placed in
their storage boxes. If the electrodes are removed, be sure that the o-ring
seal for each electrode is present.
29 Wipe up any spills in the ISE module.
30 Slide the peristaltic pump tubing off of the rollers.
31 Remove the sample, CAL-A, CAL-B and Air Intake tubes from the pinch
valves.
32 Close the container access door.
33 Press Esc until the Main menu is displayed.
34 Select Setup to display the Setup menu.
Index
abort run, 149 cleaning and conditioning, 75
accession enabling, 181
mode, 220 installing tubing, 58
number, 109 maintainance, 233
alarm volume, 186 setting calibration mode, 182
analysis module, 4 setup, 58
analysis sequence, 37 ISE module, 11
analyzer state, 30 label symbology, 188
archiving mode, 220 LIMS, 183, 184, 190
backup system, 169 line conditioner, 3
calibration, 37 loadlist
calibration requisitions, 120 description, 24
calibrators, 15 maintenance
carryover parameters, 214 daily start up, 77, 229
controls, 15, 140 monthly, 242
correlation, 46, 145 weekly, 234
cup segment menus
description, 14 access, 29
loading, 90 journals, 24
curve correction, 39 main, 21
cuvette cartridge, 12 navigating, 20
cuvette hopper purge, 25
description, 8 reports, 23
loading, 86 requisitions, 22
cuvette waste box, 101 selecting a command, 20
cuvettes, 12 setup, 27
data management, 151 status, 28
date format, 220 utilities, 25
delete modules
patient, 179 analysis, 4
patient requisition, 178 ISE, 11
test, 114 sample/reagent, 5
dilution syringe, 6
auto, 46 monitor, 19
ratios, 193 motors, 10
dimensions, 54 operational precautions, 48
doctor list, 213 optical system, 8
electrical service, 55 panels, 211
electrodes password, 215
description, 11 patient requisition
installing, 71 creating, 103
equations, 41 deleting, 178
error messages, 148 merged, 156
flags, 44 modifying, 114, 115
fluid transport system, 4 ordering tests, 109
function keys, 17 printing archived, 174
hopper, 8 reviewing, 158
instrument access, 29 precision data, 141
ISE priority, 36