Manual ETD Scintrex E3500 (Old Version)
Manual ETD Scintrex E3500 (Old Version)
DETECTOR
User Guide
The manufacturer (MFR) reserves the right to make changes to this document and the products, which it
describes without notice. The MFR shall not be liable for technical or editorial errors or omissions made
herein; nor for incidental or consequential damages resulting from the furnishing, performance, or use of this
material or of the accompanying Software or any files derived from the Software.
The product supplied by the MFR, when properly maintained, operated and calibrated in accordance with the
instructions in this User Guide, will detect specific explosives or narcotics residues at such levels and under
the conditions described in the MFR’s published specifications. The MFR disclaims responsibility for the
detection of any particular explosives compound under any particular circumstances.
The customer and/or user releases and agrees to hold harmless the MFR, its officers, servants, employees,
agents and contractors, against any and all claims, demands, actions and liabilities whatsoever, arising from
the use or operation of the product.
No part of this document may be reproduced or transmitted in any form or by any means, electronic or
mechanical, for any purpose, without the expressed written permission from Scintrex Trace Corp.
E3500, Scintrex Trace Corp, and the Scintrex Trace Corp logo are trademarks of Scintrex Trace Corp.
Other company or product names referenced may be the trademark or registered trademark of their respective
companies.
Printed in Canada
Standard Warranty
Scintrex Trace Corp warrants that the products it manufactures and sells will be free from defects in materials
and workmanship for a period of one (1) year from the date of shipment from an authorized Scintrex Trace
Corp distributor or reseller. If a product proves to be defective within the respective period, Scintrex Trace
Corp, at its option, will repair the defective product without charge for parts or labour, or will provide a
replacement in exchange for the defective product. Batteries are excluded from this warranty.
To obtain service under this warranty, customers must notify Scintrex Trace Corp of the defect before the
expiration of the warranty period and make suitable arrangements for the performance of service. Our toll-
free number within North America is 1-877-877-7147. If calling outside North America, our number is 1-
613-224-1061. Support is available during normal working hours, 8:00 a.m. to 5:00 p.m., EST. Customers
shall be responsible for packaging and shipping the defective product to the service center designated by
Scintrex Trace Corp, shipping charges prepaid, and with a copy of customer proof of purchase. Scintrex
Trace Corp shall pay for the return of the product to the customer if the shipment is to a location within the
country in which the Scintrex Trace Corp.service center is located. Customers shall be responsible for paying
all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or
inadequate maintenance and care. Scintrex Trace Corp shall not be obliged to furnish service under this
warranty:
To repair damage resulting from attempts by personnel other than certified Scintrex Trace Corp
representatives to install, repair, or service the product;
To repair damage resulting from improper use or connection to incompatible equipment;
To repair any damage or malfunction caused by the use of non-Scintrex Trace Corp supplies; or
To service a product that has been modified or integrated with other products when the effect of
such modification or integration increases the time or difficulty of servicing the product.
Except as provided in this warranty, Scintrex Trace Corp makes no warranty of any kind, expressed or
implied, including, but not limited to, the implied warranties or merchantability and fitness for a particular
purpose. In no event shall Scintrex Trace Corp be liable for indirect, special, incidental or consequential
damages.
Before operating the E3500, become familiar with all the safety symbols and instructions in this guide to
ensure that the equipment is operated and maintained safely.
Regulatory Statement
This equipment has been tested and complies with the emission limits for “Class A” and immunity
requirements of EN 61326:1997 + Amendment A1: 1998. A2: 2001, A3: 2003. Electrical equipment for
measurement, control and laboratory use – EMC requirements and verification, in accordance with EMC
directive 89/336/EEC (including amendments). For products safety, it complies to EN 61010-1:2001 in
accordance with the Low Voltage Directive 72/23/EEC (amended by 93/68/EEC).
Software Version
This User Guide covers operation of the E3500 with Operating Software, release date January 2009.
Safety Instructions
The following safety instructions must be observed whenever the E3500 is operated, serviced, or repaired.
Failure to comply with any of these instructions or with any other precaution or warning in the E3500 User
Guide is in direct violation of the standards of design, manufacture, and intended use of the equipment.
Scintrex Trace Corp assumes no responsibility for the customer's failure to comply with any of these safety
instructions.
Safety Symbols
The following safety symbols are used in this guide. The instructions and precautions associated with these
symbols must be observed whenever the E3500 is operated, serviced, or repaired.
WARNING This procedure can result in serious injury or loss of life if not carried out in
proper compliance with all safety instructions. Ensure that all conditions necessary for
safe handling and operation are met before proceeding
Caution. Always refer to the text accompanying this symbol for instructions.
On (Power)
Off (Power)
Before Operation
Become familiar with all the safety symbols and instructions in this guide, to ensure safe operation and
maintenance of the E3500.
Inspect the equipment for any signs of damage and read the User Guide completely.
Set up and verify the performance of the E3500 as specified in Section 2 – Getting Started.
Ensure that equipment and any devices or cords connected to it are properly grounded.
While Operating
Do not operate the equipment when its covers or panels have been removed.
Do not interrupt the protective earthing connection. Any such action can lead to a potential shock
hazard that could result in serious personal injury or death.
Do not operate the equipment if an interruption to the protective earthing is suspected ensure that
the equipment remains inoperative until it is confirmed to be in a safe operating position.
Use only the type and rating of fuse specified by the manufacturer as appropriate for the equipment.
Do not use repaired or modified fuses. Avoid any situations that could short-circuit the fuse(s).
Do not attempt adjustments or perform any maintenance or repair procedure if the cover or panels
are opened while the equipment is connected to a power source. Only qualified service personnel
should perform any procedures required under these conditions.
Do not operate the equipment in the presence of flammable gasses or fumes – this is extremely
hazardous.
Do not attempt to perform any adjustment, maintenance or repair procedure that is not described in
the User Guide.
Electrical Safety
To avoid risk of injury or death, always observe the following precautions before operating the E5000.
Connect the power supply cord only to a power outlet (3-prong receptacle) equipped with a protective earth
contact. Never connect to an extension cord that is not equipped with a functioning protective earth contact.
DAMAGE: Do not operate the equipment if it bears any sign of damage to any portion of its
exterior surface such as the outer casing or panels.
E3500 SENSOR: The E3500 sensor contains a dilute Alkaline solution of potassium
hydroxide. This solution is mildly caustic and can cause skin irritation or damage sensitive
surfaces (MSDS sheets). Do not ingest or get contents on skin. In case of skin contact, flush
immediately with water. Care should be exercised to prevent rupture of the E3500 sensor
membrane. Should the membrane be damaged and leakage occur, the sensor and area
should be washed with a dilute solution of household vinegar. Used or damaged sensors
should be stored in a Ziplok or equivalent bag. The E3500 sensor cannot be discarded
without treatment (Section 5 - Troubleshooting and Maintenance).
Table of Contents
SECTION 1 – OVERVIEW
1.1 General Background
EGDN NG PN TNT
EGMN DNT Semtex Black Powder
DMNB RDX C-4 TATP
O-MNT PETN Detasheet HMX
P-MNT AN Primacord Dynamite
Vapour sampling is performed by sampling the air in and around an area where
explosives are suspected to be present. Particulate sampling is performed by
wiping the surfaces of suspect articles and transferring the particulates collected
onto a metal sample screen. The sample screen is then inserted into the E3500
for analysis. The analysis is usually completed within 20 seconds and the results
displayed on the LCD (liquid crystal display). An alarm is indicated by a red light
and an audible signal.
Under normal conditions, a fully charged NiMH battery will provide two to three
hours of continuous operation (or analysis of more than 200 samples). The
E3500 is supplied with a spare battery, and a charger capable of fully recharging
a battery in one hour. Typical warm up time is one minute.
The three phases of the E3500 explosives detection process are the sample
collection cycle, analysis, and results phase.
Explosive Particulates
Sampling
Screen Pyrolyzer
Pump:
Desorber 0.8 LPM Processing
1.5 LPM Sampling
N O2
Mode Selection Valve Luminol Cell Chamber
Luminol
Explosive Vapours Solution
PMT Signal
Desorber
Following the sample collection cycle, the Sample Pump switches into a lower
flow mode and the analysis cycle is initiated, the LCD will show Processing. The
E3500 may be moved to another location while the analysis is being performed.
The operator then presses the blue Sample Button to initiate the analysis cycle.
The NO2 is picked up in the air stream and transferred directly to the
chemiluminescence sensor. As the NO2 passes across the sensing surface of the
detector, light is generated. The light is collected by a highly sensitive
photomultiplier tube (designed for single photon counting), which converts light
into an electrical signal. The electrical signal is sent to a micro-processor that
measures the signal strength.
The overall sampling and analysis time is about 20 seconds. During this time,
the LCD message center will show Processing.
The presence of explosives is verified when the signal strength exceeds the
alarm threshold within a time window expected for the explosives. If explosives
are detected, an “ALARM” message will be displayed (see figure 1-2). The
microprocessor calculates a number that represents the strength of the
measured signal. This number is displayed on the LCD screen as the Peak
Height = xxxx and Peak Area = xxxx values at the end of the analysis period,
Figure 1-3 on page 1-5. The displayed value is already corrected (i.e. minus
threshold value (THH), for instance if THH is 2000 and the total signal is 5000
then the displayed number will be 3000). At the end of the sampling process the
result is graphically displayed on the LCD and an alarm sound will be audible if a
trace of explosives was found. At the same time the Red LED will be lit.
Message Centre
ALARM
12.1V Battery Voltage
023548 PMT Output
WL1 Mode
a) Curve: the curve depicts the signal output from the sensor versus time.
Whenever a suspicious material is found, the graph should show two
peaks with different intensities (peak I > peak II).
b) Alarm: if a positive identification has been found, then the Red LED and
an audible sound will turn on. At the same time ALARM on READY will
appear on the LCD.
c) Battery Voltage: the number represents the strength of the battery in
volts. When the value blinks the unit will still be operational for about 30
minutes.
d) PMT output: the number represents the idle baseline value from the
sensor. The typical value for this number is between 10,000 to 100,000.
This number can be higher when the instrument is constantly being used
e) Zoom Level: the user can change the scale of the graph and zoom in
and out by pressing the up and down arrow keys. The zoom level is
indicated on the display (1-20X).
f) Mode: the mode is indicated next to the zoom level. P is for Particle
mode and V is for Vapour mode.
g) WL1: the cursor indicates the reference point that the software is using to
determine if an alarm condition exists. If the peak position is to the left of
WL1 the alarm will not sound; if the peak position is to the right of WL1
the alarm will activate.
By pressing the user can switch to the text window, where the
numerical results are displayed.
Run Number
Mode
PARTICLE RN 0189 Peak Height
PEAK HEIGHT + 0025
PEAK AREA 022354 Peak Area
PK POS 35
Peak Position
Mode Switch
Selection Sample Screen
Injector Sample Button
Vapour Inlet
Battery
LCD Keypad
Rear Panel
Access to the
E3500 Sensor
On/Off Switch
Earphone Receptacle
Volume Control
5 Amp Fuse
Battery Receptacle
The Keypad and Display Panel in Figure 1-6 indicates the operating status and
allows the Supervisor or Technician to change detection parameters.
The green READY LED illuminates when the E3500 is first turned on to verify
that the LED is functioning; and when the unit is in READY mode (waiting for the
operator to initiate a sample analysis).
The red ALARM LED illuminates when the E3500 is first turned on to verify that
the LED is functioning. The red LED ALARM is also initiated as a result of a
positive detection; or if an error message is displayed on the LCD screen.
PARTICLE RN 0189
PEAK HEIGHT + 0025
PEAK AREA 022354
PK POS 35
The E3500 and its accessories are contained in a transportable carrying case.
The case should only be opened when it is lying down flat. All items should be
accounted for and should have been received in good condition. Open the case
by unfastening the two latches at the front of the case. Remove each item from
the case and verify it against the checklist shown in Table 2-1.
Inspect the exterior of the E3500 and its accessories for any evidence of damage
resulting from shipping. If there are any discrepancies between the checklist and
the items received, or if there is any evidence of damage, contact your Scintrex
Trace Corp Sales Representative immediately.
CHECK-
ITEM QTY DESCRIPTION PART NUMBER
LIST
1 1 Carrying Case 943 213
2 1 E3500 D350 505
3 2 NiMH Battery D100 513
Universal Battery Charger
4 1 D100 550
(90 V – 240 V)
5 1 Vapour Test Kit D1000524
6 1 Particulate Test Kit D1000526
7 12 Particulate Sample Screen D350 532
8 20 Cotton Glove 943 019
9 1 Audio Earphone 943 018
10 5 5 Amp Fuse 943 009
11 1 User Guide D350 600
12 1 E3500 Sensor (pack of 12) D350 548
13 1 RS-232 Cable 1002813
14 1 Shoulder Strap 1002866
Screwdrivers; Philips, Flat 540 006,000 359,
15 3
Head, Allen Key 1002662
Sample Accessory Kit
16 1 D350 549
(includes)
1 Vapour Injector Assembly D350 520
1 Particle Pyrolyzer D350 514
2 Needle Probe, 2 inch long 250 302
1 Nozzle Vapour 943 210
The E3500 is supplied with a spare NiMH battery, and a charger capable of fully
recharging a battery in approximately one hour. The NiMH battery is shipped in
storage mode. The customer must provide a first time charge of two hours. It is
recommended that both of the batteries be fully charged before using the E3500.
The battery charging procedure is described in Appendix A.
To connect the battery, remove the E3500, both batteries, and the charger from
the carrying case. Then insert the battery into the receptacle at the rear of the
E3500 and slide it forward, as shown in Figure 2-1 below. The battery should
snap into place to ensure a proper connection.
Make sure that the E3500 Sensor is installed right side up and the cover closed.
After the battery has been installed, and the external battery or AC power supply
has been connected, the E3500 can be turned on.
On the Rear Panel of the E3500, press the on/off power switch. It takes two to
three seconds for the E3500 to initialize, after which time the display will
indicate:
SCINTREX TRACE
Following an internal system check, the E3500 enters a initialization mode for a
period of approximately sixty seconds. The LCD will display SELF-TEST with a
timer indicating the number of seconds remaining. When SELF-TEST mode
ends, the E3500 enters READY mode shown in Figure 2-2. The green Ready
LED remains illuminated and the red Alarm LED goes out.
If the battery indicator is steady in the graphical window, turn the E3500 off and recharge
the battery or replace it with a fully charged battery.
PARTICLE RN 0189
Self Test PEAL HEIGTH + 0025
PEAK AREA 022354
PK POS 35
Prior to sampling for explosives, it is recommended that the operator verify that
the E3500 is functioning properly by testing its performance using the appropriate
test vials supplied. The Vapour Test Kit contains DMNB, while the Particulate
Test Kit contains PETN.
Performance is verified using background testing and vapour and particulate test
kits.
Background Testing
Background testing should be completed before verifying performance with the
test kit. Background testing consists of an analysis that is done in a clean
1. When the unit has entered READY mode, set the E3500 to vapour
sampling mode by pressing the Mode Selector Switch toward the front
of the unit, as shown in Figure 2-3.
2. Initiate a sample by pressing the blue Sample Button on the handle of the
E3500. Do not hold the button down - press it only once. Once pressed:
The Sample Pump will be activated.
The green Ready LED will go out.
The LCD message center shows SAMPLING.
4. At the end of the analysis, the LCD will show the graphical representation
of the result. When in the text window, if the numbers displayed (PEAK
HEIGHT or PEAK AREA) are greater than the factory-set alarm
threshold, an alarm will be initiated.
Vapour Inlet
Nozzle
Subsequent background test samples that consistently produce no alarms indicate that
the background level and alarm threshold of the E3500 are set correctly for vapour
sampling mode. It is common for the E3500 to initiate an alarm in response to the first few
samples taken when the unit is turned on after an extended shut down period. This is a
normal part of the initialization process and should be disregarded.
Remove the vapour test vial from the E3500 carrying case. With the E3500 in
READY mode, press the blue Sample Button, unscrew the lid of the test vial,
and immediately place the opening of the vial over the Vapour Inlet Nozzle for
approximately two seconds (see Figure 2-4 below).
The pump switches to a low flow mode when the sample cycle is finished.
Remove the test vial from the nozzle and screw the lid on.
Following the analysis, an alarm should be initiated. The alarm should activate
and an audible signal heard (unless the volume was set low). If the numbers
displayed (PEAK HEIGHT or PEAK AREA) are greater than the factory-set alarm
threshold, an alarm will be initiated. If an alarm is not initiated by the vapour
sampling test, refer to Section 5 - Troubleshooting and Maintenance.
After the alarm is presented, run another test to clear the residue in the system.
This is known as a blank test. Subsequent background test samples that
consistently produce no alarms indicate that the background level and alarm
thresholds of the E3500 are set correctly for particulate sampling mode. The unit
should not alarm again, if so refer to Section 5 - Troubleshooting and
Maintenance.
Background Testing
Background testing should be done in a clean environment, before verifying
performance with the particulate test kit.
When the unit has entered the READY mode, set the E3500 to particulate
sampling mode by pressing the Mode Selector Switch toward the rear of the
unit (Note that each time the mode switch is activated, the instrument will
go into a self-test mode). With a clean sample screen inserted into the Injector
Inlet, (See Figure 2-6 on page 2-8) sample by pressing the Sample Button on
the handle of the E3500. Do not hold the button down - press it only once for
each sample. The LCD message center shows Processing.
The E3500 will then analyze the sample for approximately 15 seconds, after
which time (depending on the pre-set sample delay period) the green Ready
LED will illuminate. At this point the unit is ready for another sample.
Never handle sampling screens with your bare hands. The oil from human
skin, if introduced into the E3500, interferes with the detection of
explosives.
Put the cotton gloves supplied with the unit on one of your hands, to ensure that
you do not contaminate the unit or yourself. Remove the test kit from the E3500
carrying case. Remove the verification pen from the test kit. Gently apply
medium pressure on the middle of the screen with the verification pen (as shown
above). A small deposit of liquid should appear. Wait approximately 30 seconds
to allow the deposit to dry.
Insert the sample screen into the Injector Inlet illustrated below, and then press
the Sample Button to initiate an analysis cycle. The analysis should initiate an
alarm.
If the unit fails to respond with an alarm, repeat the procedure with an additional
drop of liquid from the pen. If no alarm occurs after the second analysis cycle,
refer to Section 5 – Troubleshooting and Maintenance.
Whenever the E3500 is not to be used for an extended period of time, it should
be turned off. Press the on/off power switch on the Rear Panel.
SECTION 3 – OPERATION
After the E3500 has been set up, its self-test cycle completed, and the unit’s
performance verified, sampling can begin. As discussed previously, sampling is
performed by either vapour sampling or particulate sampling.
Vapour sampling is performed by sampling the air in and around an area where
explosives are suspected to be present. Particulate sampling is performed by
wiping the surfaces of suspect articles and transferring the particulates gathered
onto a metal sample screen. The screen is then inserted into the Sample
Screen Injector Inlet of the E3500 for analysis. The sample analysis is usually
completed within 15-20 seconds and the results displayed on the LCD. An alarm
or hit, is indicated by a red light, audible signal, and a value displayed on the
LCD. If the analysis does not initiate an alarm, the next sample can be taken and
analyzed. If a sample initiates an alarm, the operator must follow the appropriate
procedures for the security environment they are working in and initiate the alarm
clearing procedure (see Section 3.7 for more details on Clearing an Alarm).
Under normal conditions, a fully charged battery will provide up to 3 hours (or
approximately 200 samples) of continuous operation. When the battery voltage
falls to approximately 12 volts, the battery voltage indicator will blink until 11.5
Volts, as shown below in Figure 3-1.
READY
11.8V
023548
P04X
Figure 3-1
When the battery indicator is blinking, the E3500 will operate for an additional
thirty minutes before the battery is fully discharged and the E3500 stops
operating.
When the battery voltage drops to 10.8 volts, the BATTERY BELOW 10.8 V –
SHUT DOWN message appears on the LCD and samples can no longer be
taken. (Figure 3-2). While one battery is being used to power the E3500, the
other battery should be charging or fully charged and ready for use. It takes
approximately one hour to fully charge a battery.
ERR-BAT RN 0950
PK HEIGHT +0200
PK AREA +02000
PK POS 023
To change the battery turn off the E3500 by pressing the on/off power switch on
the Rear Panel. Press the battery lock mechanism and slide the battery out of
the back of the unit. Slide a fully charged battery into place, and ensure that it
snaps into position. Turn the unit on by pressing the on/off power switch on the
Rear Panel. The unit will automatically enter SELF TEST mode.
The SELF TEST mode will be initiated when the user performs one of the
following operations:
Turning on the unit.
Switching from Vapour to Particle Mode and vice versa.
After editing any operational parameters (not applicable to DATE or
THRESHOLD values THA and THH).
If the unit is left idle for 1 hour.
Opening and closing the E3500 Sensor Cartridge tray.
At the end of each SELF TEST cycle the unit will automatically display the result
of the last blank sample. (See Figure 3-3).
PARTICLE RN 0950
PK HEIGHT +0200
PK AREA +02000
PK POS 23
The E3500 Sensor Cartridge will need to be changed every 2000 samples. The
number of samples taken is indicated in the text window as RN (RN1500 would
mean that the E3500 sensor was used 1500 times) (Figure 3-4).
PARTICLE 1 RN 1500
PK HEIGHT +0256
PK AREA 02563
PK POS 23
NOTE: If the unit has been stored in a cool place for a long period of time, the
E3500 sensor may have a longer lifetime. Make sure that the sensor is installed
right side up.
READY
12.1V
023548
P04X
If the cartridge has been inside the unit for more than a month, an error message
ERR-CELL will be displayed on the LCD to warn the user about the status of the
cartridge. Within 1 or 2 days from the expiration date, a number 1 or 2 will be
displayed respectively as shown in Figure 3-4 (page 3-3).
ERR-CELL RN 0950
PK HEIGHT +0256
PK AREA 02563
PK POS 23
With the power on, use a flat screw driver to open the door on the side of the
unit and open the tray to change the E3500 Sensor. An ERR-PMT-D message
window will appear in the text window: (figure 3-7). Remove the sensor from the
aluminum pouch and place the sensor inside the tray with the membrane facing
up (soft side) and then close the tray.
ERR-PMT-D RN 0950
PK HEIGHT +0256
PK AREA 02563
PK POS 23
At this point the user will be asked to reset the run number. When a new E3500
Sensor is used this number should be reset to zero. After the E3500 sensor has
been replaced, the unit will go into SELF TEST mode and then will be ready for
sampling.
Place the Vapour Inlet Nozzle of the E3500 near the area to be sampled. Press
the blue Sample Button on the handle and the Sample Pump will immediately
draw an air sample into the Vapour Inlet Nozzle, The LCD will display
SAMPLING . The pump will remain on for five to fifteen seconds. This time can
be adjusted by the user for specific applications. The PROCESSING message
will be displayed on the LCD during the analysis cycle. After the pump switches
to a low flow regime, the E3500 can be moved to a new location while the
analysis is being performed. At the end of the analysis, the results will be
displayed on the LCD. The overall time for a sample to be analyzed will vary
from 15 to 20 seconds.
Take a sample from the suspect article and transfer it to a clean sample screen.
With the E3500 in READY mode, insert the sample screen into the Sample
Screen Injector Inlet of the E3500. Do not force the screen into the Sample
Screen Injector Inlet or the screen may be damaged (if inserted correctly, the
screen will slide in with little resistance). Press the blue Sample Button on the
handle and the Sample Pump should immediately draw air into the E3500. The
PROCESSING message will be displayed on the LCD during the analysis cycle.
At the end of the analysis, the results will be displayed on the LCD similar to the
default window shown in figure 1-2 on page 1-4 where the sample resulted in an
alarm.
When this occurs, the E3500 may continue to respond with alarms on
subsequent sample cycles, even when no sample has been taken. The alarm
clearing procedure is the same for nuisance alarms and alarms caused by the
presence of explosives. When a repeated alarm condition is identified, remove
the sample screen from the unit and thoroughly clean it with soap and water.
Ensure that the decontaminated screen is rinsed with alcohol or acetone solution
and thoroughly air-dried before reinserting it into the E3500.
Decontaminate ?
No Yes
3. After choosing yes the unit will proceed with the decontamination cycle,
which lasts 60 seconds.
60
STANDBY
SECTION 4 – APPLICATIONS
The longer the explosive is present, the greater the amount, or concentration, of
vapours present. Therefore, you would not have to be in the immediate vicinity of
an explosive to collect a sample.
To collect a sample keep the nozzle of the E3500 pointed in the direction of the
suspected vapour. Push the sample button on the handle of the unit and hold the
nozzle until the LCD displays PROCESSING.
Vapours escape from rooms and other enclosed areas or containers through any
available opening. Table 4-1 below indicates examples of specific areas where
vapour samples should be taken.
Collecting samples for analysis requires the operator to think like a person trying
to conceal explosives. The operator must consider where the person may have
touched the suspect article. This applies to any container. The goal is to take
samples from the most probable locations on the suspect containers that may
have been contaminated (handels, zipper).
The E3500 can detect very small traces, but it is the operator’s responsibility to
introduce these traces (when present) by using the proper sampling technique.
Particulate samples are taken from suspect articles by wiping with a clean cotton
glove or wiping with a sample screen. Each sampling method has advantages
and disadvantages, however, wiping with a clean cotton glove is the preferred
method. Using a sample screen is the second recommended method.
When using either of the two methods, it is very important that all sampling of
suspect articles be done with the operator wearing clean cotton gloves on
both hands. The gloves must be discarded and a fresh pair put on whenever
the gloves become soiled, the gloves show signs of deterioration, or a sample
results in an alarm.
The glove sampling method produces the lowest rate of nuisance alarms, permits
the setting of the lowest alarm thresholds, and is the least prone to interference
from the operator’s hands.
However, because particulates must be transferred from the glove to the screen,
some could be lost in the process. If this happens, the amount deposited on the
screen could be less than that required to initiate an alarm.
The position of the glove refers to how it is held while wiping suspected areas.
The glove should be held firmly to minimize movement, to ensure that the area of
the glove or wipe that contacted the suspect article remains fixed, and to ensure
that the maximum amount of the sample is transferred to the screen.
Sampling screens require careful handling. Gently rub the surface and keep your
fingers away from the metal mesh when handling screens.
SECTION 5 – TROUBLESHOOTING
AND MAINTENANCE
5.1 TROUBLESHOOTING
At times, the E3500 may show symptoms of faults or errors. Table 5-1 lists
potential problems and related troubleshooting procedures.
PROBLEM PROCEDURE
Battery Troubleshooting
• Replace battery with fully charged battery.
• Ensure that the battery is installed securely.
• Ensure that the unit is turned on.
UNIT DOES NOT
POWER UP (WITH • Inspect the 5 Amp fuse and replace if necessary.
BATTERY INSTALLED) • Check the battery and battery contacts and clean them
if necessary.
• If the E3500 still does not power up, contact Scintrex
Trace Corp Technical Service Support.
BATTERY DOES NOT • The battery is a consumable device and has a limited
HOLD A CHARGE OR number of recharge cycles. Safely dispose of Ni-Cd
CHARGER DOES NOT battery and obtain new battery from Scintrex Trace
CHARGE BATTERY Corp Technical Service Support.
Pump Troubleshooting
AIR PUMP DOES NOT • Turn off and turn on the unit.
START WHEN THE • Contact Scintrex Trace Corp Technical Service Support.
UNIT IS TURNED ON
PMT OUTPUT HIGH • High level of oxidant in the atmosphere such as Nox,
peroxides.
• Old E3500 sensor (stored at room temperature.)
E3500 Troubleshooting
UNIT IS • Remove E3500 to a clean area.
CONTAMINATED • If Unit was contaminated in particle mode – Remove
BY EXPLOSIVES particle screen.
RESIDUE • Initiate a sample cycle and note the value.
• If the value is still too high, repeat until the value has
decreased to the background value noted at start-up.
• If the value does not decrease, contact Scintrex Trace
Corp Technical Service Support.
5.2 Maintenance
Pyrolyzer holder
assembly
Please have the following information on hand prior to calling Scintrex Trace
Corp Technical Service Support as it will be required to process your request:
The serial number of the unit
The location of the unit
A contact person on-site and their telephone number
A detailed explanation of the problem
The E3500 or any of its parts may be returned to Scintrex Trace Corp for repair
or disposal. Shipping of a complete unit or parts requires special handling.
Pack the unit in its original factory shipping crate. Parts must be well packed and
protected against any possible damage they may be exposed to during shipping.
Contact Scintrex Trace Corp. to order a replacement shipping crate if required.
Disconnect battery and remove E3500 Sensor from E3500 prior to shipping.
Protect battery terminals from electrical short circuit to prevent fire or personal
injury.
Contact Scintrex Trace Corp to coordinate shipping details such as customs and
other documentation required, pickup times, and required signatures for
insurance purposes.
After all shipping arrangements have been made and confirmed with Scintrex
Trace Corp., all returned goods should be shipped to the following address or a
DDU (Delivery Duty Unpaid) service from outside Canada.
New E3500 Sensors should be stored in a refrigerator. This simple procedure will
extend the life of the sensor.
Used E3500 Sensors cannot be discarded unless the contents are chemically
neutralized. Care should be exercised during the procedure to prevent skin
contact or contact of the contents with sensitive surfaces. Latex or vinyl gloves
and protective glasses should be worn. Should skin contact or spillage occur
flush immediately with water or a dilute solution of vinegar.
Equipment
Process
Carefully puncture the sensor membrane located on the top of the sensor
with a sharp knife
Immerse the sensor in a liter of water (in a bowl)
Add some household vinegar
Pour the solution down the drain and flush with water
Discard the empty container
Use the Universal Battery Charger PN D100550 with the recommended Scintrex
Trace Corp NiMH battery. Charging a battery other than the recommended
battery provided may cause irreparable damage to the incompatible battery
and/or the charger. Recharge batteries at ambient temperatures of 18°C to 24°C.
Inspect the battery for any damage prior to connecting it to the charger. If any
physical damage to the battery is apparent do not attempt to charge it. Obtain a
replacement battery.
After connecting the battery to the charger the red power lamp should remain lit
and the green mode lamp should change colour to orange while the charging is
in progress the mode lamp should change back to green when the charging
process has been completed (about one hour). If the red lamp and the green
lamp both extinguish and the fan stops, this may indicate a battery which has
defective cells and can no longer be safely charged. Obtain a replacement.
When the mode lamp changes to green, after about an hour the battery is
normally charged and is safe to use. Although this is a multistage type charger, it
is good practice not to leave the battery connected for longer than one hour as
the battery may be damaged.
If a battery quickly heats up during the charging process, it may no longer charge
properly and needs to be replaced. A temperature sensor in the battery will
disconnect it in such an event. Disconnect and remove the defective battery. If
the battery does not hold a charge above 13V for the normal expected time this
also indicates a replacement battery may be necessary. The battery is a
consumable device and has a limited number of recharge cycles. To avoid loss
of run time or unexpected battery failure, replace batteries every two years.
The battery charger should be switched to the off position when not in use to
extend the life of the built in cooling fan.
APPENDIX B – COMMUNICATIONS
Communications
There are a number of techniques that can be used to communicate with the
E3500. Hyper-Terminal, LabView, Internet Browsers and Pocket PCs are all able
to communicate with the E3500.
Embedded Software
The E3500 has several frames that the user can browse through to view results,
operational parameters or settings. The user can navigate through these frames
by pressing onto the keypad located at the bottom of the LCD.
• INITIAL SCREEN
• TEXT SCREEN
• GRAPHICAL SCREEN
• LOW BATTERY •
MESSAGE •
DISPLAYED ON •
LCD •
READY
• 12.1 V
• 023548
004X
•
•
•
•
•
• If the battery indicator is blinking, the unit will be operational for
about 30 minutes.
ERR-BAT RN 0201
PK HEIGHT +0256
PK AREA 02563
PK POS 23
•
• If the battery indicator has a value less than 11.5V, the unit will not
be operational. The display will automatically switch to the text
window.
• Replace the battery with a fully charged battery.
• Recharge the battery that was removed.
• E3500 SENSOR •
ERROR •
DISPLAYED ON •
LCD READY
12.1V
023548
• P04X
•
•
•
•
• ERR-CELL •
•
•
• ERR-CELL 1 RN 2001
PK HEIGHT +0256
• PK AREA 02563
• PK POS 23
•
•
•
• ERR-PMT-D •
•
•
•
• ERR-PMT-D RN 0950
PK HEIGHT +0256
• PK AREA 02563
PK POS 23
•
•
•
• To clear this message close the tray and the door on the side of the
unit.
Hyperterminal Communication
Once completed, click OK to leave the ASCII Setup window, and then OK once
more to leave the Properties window. To save the settings, choose File->Save As
and choose a filename/location under which to save the file. The next time the
settings are required, File-> Load and choose the appropriate file.
To start an analysis, type 1. The E3500 will start sampling, and the following data
will appear in the HyperTerminal window. This data can be displayed again by
pressing L.
If more detail on the run is desired, the entire set of data points used to create
the graph shown on the primary display can be obtained by entering d (or D) in
the HyperTerminal window.
The user can also download all the results stored (895 samples) in the memory
by pressing A and Enter. To stop the process type S and Enter.
To retrieve the parameters for both modes type P, the following window will
appear:
APPENDIX C - SPECIFICATIONS
Electrical
Mechanical
DIMENSIONS (L x W x H) 51 x 11 x 14 cm
20 x 4.3 x 5.5 in
22 x 10 x 15 (in)
(L x W x H) (L x W x H)
Environmental
General Specifications