Welcome to the proteomics course. Today, we will talk about protein purification and
peptide isolation using chromatography methods. The chromatography techniques have
always been considered prominent for the separation of enzymes and proteins. Protein
purification by conventional chromatography is usually achieved by combining
chromatography methods such as gel filtration, ion exchange and affinity
chromatography. From complex proteome it challenging to purify a protein in a single
chromatographic step therefore sequential pre-fractionation steps involving different
modes and types of chromatographic methods are becoming necessary for proteome
level analysis.
Lecture outline- we will talk about gel filtration chromatography and ion exchange
chromatography, affinity chromatography, these methods we will be discussed in light of
protein purification. And then we will talk about few methods such as strong cation
exchange and reverse phase chromatography for peptide isolation using high
performance liquid chromatography methods.
Protein purification and peptide isolation and analysis- it can be performed by many
methods but there are certain chromatography methods which are commonly used,
such as Size exclusion chromatography (SEC), Ion exchange chromatography (IEX),
Immobilized metal ion chromatography (IMAC), Normal phase chromatography (NPC),
Reverse phase-HPLC (RP-HPLC), Hydrophobic interaction chromatography (HIC) and
Hydrophilic interaction chromatography (HILIC). These are only few chromatographic
methods but there are many more which are also used for different type of applications
So if your aim is to purify proteins, there are various techniques, which can separate
proteins and they rely on different type of principles such as
The chromatographic technique involves four major components an inlet for sample
introduction, the mobile phase, stationary phase and a detector.
Each type of chromatography requires very educated and informed choice of matrix,
bead shape, size and porosity. In addition, functional group types the charge distribution
and density as well as elusion condition such as pH, the ionic strength and gradient
shape.
So we have just talked that chromatography involves the selection of right matrix, these
matrix are very important for binding and interaction of proteins and they can determine
the kind of achievement, which we want to make from these types of chromatographic
methods.
The gel filtration chromatography it separates proteins on the basis of difference in size.
It is used to separate a protein of interest from a protein mixture that are larger or
smaller size. If the proteins are having similar sizes then gel filtration or size exclusion
chromatography is not an appropriate choice for doing the protein purification. So this
method is also used for various proteomic applications when there is need to remove
even contaminants such as low molecular weight contaminants and low molecular
weight detergents.
The gel filtration chromatography column is composed of porous beads, which are
made from polyacrylamide, dextran or agarose. Now these columns are packed with the
hydrated porous gel matrix.
The protein sample that contains a mixture of unpurified proteins of different size is then
loaded on these columns.
After these step, the fractions are collected and analyzed for protein content. So you
can expect that larger proteins will elute first and then small molecules will be eluted
later.
Size exclusion gel matrix- the matrix filling the gel filtration column consists of highly
hydrated polymeric material commonly dextran, agarose and polyacrylamide. The
protein mixture is a mixture of unpurified proteins of different sizes, which is applied on
top of the column.
Mobile phase – the proteins are moved out of the column by using a suitable mobile
phase that carries the protein out for elution. For gel filtration a salt solution of
appropriate strength is commonly used so that it will not have any effect on properties of
proteins being purified.
Sample fractions, the solution leaving the column are collected in suitably sized
fractions, initial fraction will contain only the mobile phase while later fractions will have
purified proteins.
Let me give an analogy of gel filtration with sieve. This process is very similar to
separation of small particulate matter from food grains using a sieve. The larger grains
remain behind the sieve while the smaller sand or stone particles pass through them
and are removed. In keeping with this the gel filtration chromatography is also known
molecular sieve.
Now let me show you how this process works. So first the matrix which is suitable for
the required protein separation, this packed gel matrix is loaded with the protein sample
containing a mixture of unpurified proteins of different sizes. The column is then eluted
The largest proteins eluting out first will be present in the initial fractions while the
smaller proteins which elute out later will be present in the later fractions. Once all these
fractions are collected, then it can be analyzed for their protein content by using a UV-
visible spectrophotometer. Now analyze the protein content by using UV-visible
spectrophotometer at 280 nm. Once all the absorbance values are recorded a graph of
eluant volume versus protein concentration can be plotted. The highest molecular
weight is in the beginning and lowest molecular weight towards the end.
In ion exchange chromatography proteins are separated based on charge difference the
proteins with overall negative charge will interact with positive charges or vice versa. So
by varying the amount of positive and negative charge amino acids and even pH
influences net charge on proteins.
So in this slide some of the common ion exchange matrixes are shown like
Carboxymethyl (CM) and Diethylaminoethyl (DEAE). So when a desired protein is
positively charged, the cation exchange chromatography should be used. When a
desired protein is negatively charged the anion exchange chromatography method
should be used.
So in ion exchange chromatography, the column is packed with a resin whether it’s
cation or anion exchanger depending upon the charge of the protein that need to be
bound to the column and purified.
So proteins are adsorbed to the ion exchange column and then it can be desorbed by
increasing the salt or altering the pH of the buffer, which can change the charge on
protein. So various anionic buffers such as acetate and phosphate are used for cation
exchange and cationic buffers such as tris chloride or ethanolamine are used for the
anion exchange.
So if you have negatively charged protein, which gets eluted first will be present in initial
fractions, while the positively charged protein that bound to the column will be eluted in
the later fractions or it will be vice versa.
So let me describe how ion exchange chromatography works step by step in following
animation.
The charge stationary phase; the column stationary phase is consisting of a positively or
negatively charged polymeric matrix, which will bind molecules of opposite charge.
Commonly used ion exchangers are included negatively charged carboxy-methyl
cellulose or CM cellulose, which is a cationic exchanger and positively charged DEAE
cellulose which is an anion exchanger.
The protein mixture- unpurified protein mixture, which consist of proteins of different net
charges is loaded on to the column. The proteins having charges opposite to that of
stationary matrix will bind to it while remaining proteins will be eluted.
Mobile phase- the proteins are eluted out of the column by using suitable mobile phase
and then samples are collected by using different sample fractions. The solution leaving
the column can be collected in suitably sized fractions for further analysis. After giving
you brief description of components, let me show you the process in animation-
The column is packed with a suitable cation or anion exchange resin depending upon
the charge of the protein that needs to be bound to the column and purified. The anion
exchange column is then loaded with the impure proteins mixture consisting of various
positively and negatively charged proteins.
The column is eluted with a buffer solution of suitable pH such that negatively charged
molecules are removed from the column while the positively charged molecules remain
bound to the anion exchange resin. The buffer solution is then changed such that the
net pH of the protein of interest is modified and no longer binds the ion exchange resin
therefore the bound protein also gets eluted out of the column in this manner. The
fractions of appropriate size must be collected and analyzed for their protein content.
The negatively charged protein which gets eluted first will be present in the initial
fractions while positively charged proteins which bound to the column are eluted in later
fractions.
So let’s now talk about another very important method, which is affinity chromatography.
It is a desirable method for protein purification to the homogeneity. Due to the selectivity
of immune recognition it is possible to purify a protein in a single step under the
favorable conditions and affinity chromatography is one of the methods for protein
purification.
Now the column is washed with a suitable mobile phase to remove the unbound protein
and protein of interest which has higher affinity for the ligand remains bound to the
derivatized column matrix and is not removed during the washing step. Now these
proteins of interest can be desorbed by applying excess ligand in the solution.
So due to the specific interaction the affinity chromatography achieves very high degree
of protein purification. It is not limited by the sample volume, which is the case of gel
filtration chromatography and because of its superiority in achieving pure protein it is
usually considered as a final step the protein purification.
There are various examples where affinity chromatography can be used, especially the
antigen – antibody pairs they are commonly used for bio-affinity pairs. The matrix
containing protein A which is used for the IgG purification so in this case the protein A
recognises the Fc region of IgG and this interaction is being used for protein purification
using affinity chromatography. Another strategy is concavalin A protein binds to the
glucose molecule so by adding a concentrated solution of glucose the glucose can
displace the column where these molecules are attached on the binding site of
concavalin A but these are only few examples there are many other examples where
There are various other examples but these are common strategies being employed for
protein purification.
The derivatized stationary phase- the stationary phase resin in affinity chromatography
consists of a covalently bound ligand that will specifically bind the protein of interest by
interacting it.
The protein mixture- it is unpurified protein mixture which consist of proteins having
different proteins and interaction specificity of ligand bound to the column matrix.
Mobile phase- following the sample loading, the unbound proteins are washed out of the
column using suitable mobile phase. Depending on protein of interest this could be
either water or sometimes salt solution.
The effluent sample fraction- the solution leaving the column is collected in suitable
sized fractions for further analysis. The unbound proteins are eluted from the column
first, followed by the bound fractions which are removed after washing with the ligand
solution.
The column is first packed with a suitable resin that has been covalently coupled to the
ligand, specific to the protein of interest. The derivatized affinity column is then loaded
with the protein mixture containing various proteins having different properties and
interaction specificity. The column is washed with a suitable mobile phase to remove all
the unbound proteins. The protein of interest which has higher affinity for the ligand
remains bound to the derivatized column matrix and is not removed during the washing.
The samples collected during washing can be analyzed assays and discarded if not
required.
After the column has been washed thoroughly the protein of interest is eluted by
passing a ligand solution which binds to the matrix bound protein and removes it from
the column. The fractions are then analyzed for their protein content by UV-visible
spectrophotometer at 280 nm. A graph of eluant volume versus protein concentration
can then be plotted.
Let’s now talk about high performance liquid chromatography (HPLC). HPLC is
separation technique that separates molecules based on their differential adsorption
and desorption between stationary phase matrix in column and mobile phase. Better
separation and resolution of the components can be achieved by HPLC.
Let’s first talk about reversed phase (RP) chromatography. The reversed phase
chromatography is based upon the hydrophobic binding interaction between the
peptides or the proteins in the mobile phase and immobilized hydrophobic ligands in the
stationary phase.
In reversed phase, the stationary phase consist of long aliphatic carbon chains, which is
highly hydrophobic in nature. Molecules are bound on the column by means of
hydrophobic interactions and are eluted out when the solvent polarity is modified. The
reversed phase chromatography is mostly used with electro-spray ionization (ESI) in
mass spectrometry due to its compatibility of its acidic aqueous and polar mobile with
ESI. It is also used for desalting the peptides before injecting for ESI.
Let’s now talk about Strong cation exchange (SCX) chromatography. SCX consist of
stationary phase matrix which is made up of negatively charged sulfonic acid groups
which bind the oppositely charged peptide molecules. These molecules can be eluted
out using a positively charged mobile phase which binds the analyte molecules more
firmly.
So we have already discussed two separate methods SCX and RP. So SCX is based
on the electrostatic attraction between the negatively charged sulfonic acid and
positively charged peptides and the elution can be caused by addition of positively
charged mobile phase. The RPC is shown on the right side is based on the hydrophobic
interactions between analyte and stationary phase. The elution can be brought about by
modifying the mobile phase polarity. The SCX can be used offline and then each
fraction can be analyzed by reverse phase HPLC followed by mass spectrometry.
Alternatively, both RP and SCX resins can be packed into a single column and by
introducing buffers in the series, the multidimensional separation can be achieved.
Now when application is to separate complex proteome and analyze using mass
spectrometry then one has to select what type of chromatography method is applicable
for their sample type. Now multi-dimensional approaches are usually coupled prior to
the mass spectrometry analysis. And as we have discussed there are certain
chromatographic methods, which are commonly available and one has to make an
So when size exclusion chromatography has been used followed by the reverse phase
so to check the compatibility of first and second dimension separation based on the size
in the SEC and the charge various type of combination methods have been tried out. So
let me first give you the abbreviations here and then what properties they separate
proteins and peptides based on and then we can talk about how these combinations
can be applied.
So in summary, in the last few years there has been an increased effort to develop the
technologies, which are capable of analyzing protein expression at the proteome scale.
Some of the chromatographic methods such as gel filtration, ion exchange and affinity
chromatography were discussed. The SCX and RP-HPLC are fundamental tools for the
isolation and analysis of peptides. The nano-liquid chromatography which makes use of
C18 capillary columns has gain popularity for the proteomic studies due to their ability to
achieve finer separation.