Ser Man GEM
Ser Man GEM
This publication and all materials (including software) concerning the products of the IL
GEM Premier 3000 are of proprietary nature and are communicated on a strictly
confidential basis; they may not be reproduced, recorded, stored in a retrieval system,
transmitted or disclosed in any way and by any means whatsoever, whether electronic,
mechanical through photocopying or otherwise, without IL’s prior written consent.
Information contained herein is believed by IL to be accurate: in any event, no
responsibility, whether express or implied is assumed hereby by IL for or in connection
with the use thereof, or for infringement of any third party right which might arise
therefrom, or from any representation or omission contained herein. Information is
subject to change and/or updating without notice.
i
GEM Premier 3000 Service Manual
CONTENTS
General Information
Installation
Operating Programs
Fluidic/Cartridge Description
ii
GEM Premier 3000 Service Manual
Maintenance
Troubleshooting
iii
GEM Premier 3000 Service Manual
Interfacing
Parts List
iv
GEM Premier 3000 Service Manual
v
GEM Premier 3000 Service Manual
1. General Information
This manual has been created to reflect the GEM® Premier 3000 at the current
time. All descriptions refer to the current configuration of hardware and software
version (5.4.2).
This section contains an Overview of the GEM Premier 3000, a list of the
measured and calculated parameters, a description of the front components of the
instrument, and a description of the rear components of the instrument.
1.1 Overview
The GEM Premier 3000 is a portable system for use by health care
professionals to analyze whole blood samples, in any clinical setting. The
instrument provides both measured and calculated results for blood gases,
hematocrit, electrolytes, glucose, and lactate. The instrument may be
interfaced to an external IL CO-Oximeter, the IL 682™ or the GEM
OPL®, with the CO-Ox results integrated into the GEM Premier 3000
results.
The GEM Premier 3000 system is comprised of the instrument and the
cartridge. The cartridge is a self contained, disposable package that
includes all of the Process Control Solutions, sensors, and the entire
sample pathway, including the pumps, valves, and waste bag. The
instrument controls the cartridge to perform all the necessary functions for
heating, calibration, probe movement, pumping, and sample aspiration.
The instrument accepts all of the sensor data from the cartridge and
processes this data to perform and verify the calibrations and to present the
data for the analyses. The instrument will display, store, print, and
transmit the data.
The GEM Premier 3000 can use a wide variety of cartridges, from a 35
test, 4-week cartridge to a 600-test 2-week cartridge. In addition, the
instrument and cartridges can be configured to use intelligent Quality
Management (iQM) or traditional, external QC.
1-1
GEM Premier 3000 Service Manual
1-2
GEM Premier 3000 Service Manual
LCD Assembly
Printer Door
Ampoule Spinner
Handle
Access Panel
Keyboard Port
Serial # Plate (covered)
Power Cord
1-3
GEM Premier 3000 Service Manual
1.6 Symbology
1-4
GEM Premier 3000 Service Manual
2. Installation
The GEM must be placed in a location that allows adequate ventilation on the left
side and the rear of the instrument. The rear of the instrument will require
adequate clearance to allow for any connections (Ethernet, RS 232, Parallel port,
bar code reader) without putting stress on the cables. The right side of the
instrument requires 12 inches of clearance to allow insertion and removal of the
cartridge.
NOTE:
The GEM Premier 3000 has electrical filters to protect against line noise.
In locations where significant voltage or current fluctuations frequently
occur, the use of a line conditioner or an uninterruptible power supply
(UPS) is recommended
2-1
GEM Premier 3000 Service Manual
An authorized IL representative should remove the GEM Premier 3000 from its
carton. Check that the instrument and its accessories have not been damaged
during transport. If damage is noted, initiate the proper procedures to file any
claims and obtain a replacement instrument, if required.
The GEM Premier 3000 will be packaged in a plastic bag, with two desiccant
packs. There will be a separate box with the instrument accessories in it. The
complete shipping list is:
2-2
GEM Premier 3000 Service Manual
2.5 Start Up
When power is first applied, the fan will turn on and the motors will home
themselves. The screen will then show a light colored bloom, followed by a very
brief logon display. It will then display “loading GEM 3000 software for several
seconds. After that the screen will display the normal startup screen with “The
instrument is performing self diagnostics. Please wait.” After several minutes,
the screen will blank out, the motors will find home again and the display will
come on in the “Insert Cartridge” screen.
MM/DD/YYYY
DD/MM/YYYY
YYYY/MM/DD
To change the Date and/or Time, press “Configuration” and “Set Date and Time.
Any changes made in this screen will require restarting the instrument.
If the Daylight Saving Time feature is selected, the instrument will automatically
move the time forward one hour on the first Sunday in April, and it will move the
time back one hour on the last Sunday in October. If this is changed, it will
require restarting the instrument.
2-3
GEM Premier 3000 Service Manual
If the daylight Savings Time feature is not desired, it must be turned off by
pressing “Configuration”, Set Date and Time” and deselecting “Enable Daylight
Saving Time”.
Prior to inserting cartridge, verify that the desired setting for iQM is entered.
Press “configuration” and “iQM Setup”. If the cartridge is an iQM cartridge,
verify that the “iQM Mode” is selected.
2.8 Configuration
The GEM 3000 provides a significant amount of flexibility in the configuration
menu. Section 3 of the Operators Manual provides details on each of the
selections available. In the Configuration Menu, selections are available for :
Sample Setup
QC Setup
IQM Setup
Calibration Setup
Instrument Setup
Interface Setup
Security Setup
In addition, there are choices to save the configuration to a 1.44 M HD disk, and
to restore the configuration from this disk. The configuration can be saved with a
cartridge in use, but once a cartridge is inserted, the configuration disk cannot be
used to load a specific configuration onto an instrument.. This Save and Restore
2-4
GEM Premier 3000 Service Manual
feature can be very useful when installing multiple instruments in the same
facility. If one instrument is configured to suit the user’s requirements, the
configuration can be loaded onto the other instruments with the disk, saving a
significant amount of time.
The Configuration Menu also has a selection to “Restore KOPW” (Key Operator
Password). A Restore KOPW disk is provided with each instrument, in the back
of the Operators Manual. When this is selected, the Key Operator Password will
be restored to 1 2 3 4. The Date and Time can also be selected, as detailed in
Section 2.7 (above)
NOTE:
The key Operator Password is required to access all selections on the
Configuration Menu, except Restore KOPW.
The GEM can be shutdown for short periods of time to move the instrument, or to
load a software upgrade, or video disks, with a cartridge inserted. If the
instrument is shutdown with a cartridge, a popup screen will appear, indicating
that power must be restored within 60 minutes or the cartridge must be replaced.
In some instances, if the instrument is performing a 2 point cal, a C solution cal or
if there is blood on the sensors, power must be restored within 20 minutes.
If the instrument is to be shut down for an extended period of time, the cartridge
should be removed prior to shutdown.
Select “Shutdown” from the ready screen, press shutdown and confirm that the
instrument is to be turned off. There will be a prompt to wait while the
instrument is shutting down, followed by “It is now safe to turn off the
instrument.” Turning power off before this message is displayed can cause
corruption of the operating software, and will cause a delay in restarting the
instrument when power is restored.
2-5
GEM Premier 3000 Service Manual
3. Operating Programs
This section contains a brief overview of the available menu choices. Significant detail
about each of these choices is available in the Operators manual.
A B
D
E
H
I
3-1
GEM Premier 3000 Service Manual
3.2 Database
B
C
D
E
F
G
All iQM Reports are available for 12 months, then are overwritten.
3-2
GEM Premier 3000 Service Manual
3.3 Configuration
A
B
C
D
E
F
G
H
I
J
K
3-3
GEM Premier 3000 Service Manual
I Displays a prompt to insert a blank disk to save the configuration for use
on another instrument, or on the same instrument at a later date.
J This menu choice is only accessible when no cartridge is inserted.
Prompts to insert the disk created in Save Config.
K This menu choice is only accessible when no cartridge is inserted. Allows
change of date and time and selection of Daylight saving Time.
3.4 Cartridge
This menu has only one selection; “Remove Cartridge”. When this choice is
selected, it must be verified, then the cartridge may be removed.
3.5 Diagnostics
A
B
C
D
E
F
G
H
A Provides screens for status and testing of Com Port A, B, C and the
Ethernet port.
B Provides screen for status and testing of the internal printer, parallel
printer and network printer.
C Provides specific information about software version, instrument serial
number and hardware, and cartridge information.
D Provides a screen to select specific cartridges for copying IL Data to a
disk. IL Data contains no patient information.
E Initiates a 2-pt calibration.
3-4
GEM Premier 3000 Service Manual
3.6 Shutdown
This menu has only one selection “Shutdown”. When this is selected, it must be
verified, then the instrument will prepare for shutdown.
3-5
GEM Premier 3000 Service Manual
4. Fluidic/Cartridge
The GEM Premier 3000 uses a cartridge to perform all measurements. The cartridge
contains all of the fluidic components and sensors of the instrument. The sensors are
contained on a card that interfaces with the Heater Block for temperature control, and
with the analog PCB to provide the output voltages to the processor.
The Process Control solutions are contained in three airtight bags, with no headspace.
This eliminates any changes in values due to changes in Barometric pressure or
temperature, a key requirement for iQM.
4-1
GEM Premier 3000 Service Manual
Cartridges
The cartridge can now be inserted into the instrument, larger bar code label up.
4-2
GEM Premier 3000 Service Manual
4-3
GEM Premier 3000 Service Manual
4-4
GEM Premier 3000 Service Manual
The quantities represented by the concentration names subscripted with the word
“base” in the above table are determined from the operating software’s range table
(see section 3.3).
Interpreting Cartridge Type (fields a and G)
The operating software uses the Cartridge Type fields “a” and “G” to
determine the cartridge menu (allowed analytes), the cartridge use life
(comprising the maximum number of tests/weeks) and whether the
cartridge is a traditional or of the iQM type. The software accepts only the
following cartridge types (table entry represents a-G):
Cartridge Types
75-TEST 75-TEST 150-TEST 300-TEST 450-TEST 600-TEST
3-WKS 4-WKS 3-WKS 3-WKS 3-WKS 2-WKS
BG/Hct 3-0 3-2 4-0 5-5 6-0 N/A
P/N=24307504 P/N=24407504 P/N=24315004 P/N=24330004 P/N=24345004
BG/Hct (iQM) 3-3 3-5 4-3 5-7 6-3 N/A
P/N=24307584 P/N=24407584 P/N=24315084 P/N=24330084 P/N=24345084
BG/Hct/Lytes 3-1 3-6 4-1 5-1 6-1 7-1
P/N=24307507 P/N=24407507 P/N=24315007 P/N=24330007 P/N=24345007 P/N=24360007
BG/Hct/Lytes 3-4 3-7 4-4 5-4 6-4 7-4
(iQM) P/N=24307587 P/N=24407587 P/N=24315087 P/N=24330087 P/N=24345087 P/N=24360087
4-5
GEM Premier 3000 Service Manual
In the tables above, “Cartridge Use Life” is the number of days the
cartridge may be used after it is inserted into the instrument. “BG” and
“Lytes” stand for “Blood Gases” and “Electrolytes”, respectively.
When an analyte is not on the cartridge menu, the software internally
disables the analyte and prevents the operator from re-enabling it.
The cartridge part numbers shown in the table above are used by GEMweb
software (see Volume 3).
iQM stands for “intelligent Quality Management”, which designates
cartridges with built-in quality management processes that substitute
traditional quality controls.
Software will also accept the ETC (Electronic Test Cartridge). The ETC
has a cartridge type of 0-0.
The cartridge type 5-0 (a=5 and G=0) is reserved for experimental
cartridges.
4-6
GEM Premier 3000 Service Manual
4-7
GEM Premier 3000 Service Manual
The following table gives the base values for the concentrations of
analytes in calibration solutions based on the value of the “gh” field in the
cartridge barcode:
Analyte Concentrations Base Values
Base Name gh = 00 gh = 01 gh = 02 gh = 03 gh = 04 gh = 05 gh = 06
pHAbase 6.85 6.85 6.85 6.85 6.85 6.85 6.85
pHBbase 7.35 7.35 7.35 7.35 7.35 7.35 7.35
pOIbase 180 130 130 130 130 130 130
pCOAbase 60 60 60 60 60 60 60
pCOBbase 30 30 30 30 30 30 30
KAbase 1.60 6.60 6.60 6.60 6.60 6.60 6.60
KBbase 5.60 3.10 3.10 3.10 3.10 3.10 3.10
NaAbase 150 96 96 96 96 96 96
NaBbase 130 140 140 140 140 140 140
CaAbase 0.16 2.26 2.36 2.46 2.56 2.66 2.76
CaBbase 1.75 0.80 0.80 0.80 0.80 0.80 0.80
GluAbase undefined 100 100 100 100 100 100
LacAbase undefined 26 26 26 26 26 26
pOA0base undefined 0 0 0 0 0 0
4-8
GEM Premier 3000 Service Manual
The following table describes each of the base names used in the table
above:
Base Names Description
Base Name Description
pHAbase “A” calibration solution pH base value
pHBbase “B” calibration solution pH base value
pOIbase “B” calibration solution initial O2 partial pressure base value
expressed in mmHg
pCOAbase “A” calibration solution CO2 concentration base value
expressed in mmHg
pCOBbase “B” calibration solution CO2 concentration base value
expressed in mmHg
KAbase “A” calibration solution K concentration base value expressed
in mmol/L
KBbase “B” calibration solution K concentration base value expressed
in mmol/L
NaAbase “A” calibration solution Na concentration base value
expressed in mmol/L
NaBbase “B” calibration solution Na concentration base value expressed
in mmol/L
CaAbase “A” calibration solution Ca concentration base value expressed
in mmol/L
CaBbase “B” calibration solution Ca concentration base value expressed
in mmol/L
GluAbase “A” calibration solution glucose concentration base value
expressed in mg/dL
LacAbase “A” calibration solution lactate concentration base value
expressed in mg/dL
pOA0base “C” calibration solution O2 partial pressure base value
expressed in mmHg
Calculating Calibration Solution Concentration Constants
The operating software uses the calibration solution concentrations of the
analytes to calculate both calibration and sample results. The following
table shows how the calibration solution concentration constants are
calculated and also shows the acceptable range for each constant. The
names shown for these constants are used throughout the documentation.
To calculate a calibration solution concentration constant, a calibration
solution base value and a cartridge barcode field value are used (see
sections 0 and Error! Reference source not found., respectively).
4-9
GEM Premier 3000 Service Manual
4-10
GEM Premier 3000 Service Manual
(The exact time of the day for performing Low Oxygen calibration is
user selectable, default is 0200)
Calibration evaluation
The calibration results will be reported (no errors or an error indicated)
according to the Calibration Report settings configured by the Key
Operator
If a calibration fails, results for the analyte(s) involved will not be
reported, however, results for analytes that passed calibration will
continue to be reported
Slope (two-point calibration) and drift (one and two point calibration)
errors must be corrected prior to reporting results for the failed
analyte(s)
Calibration interruption
During the first 4 hours after cartridge insertion, a two-point calibration
cannot be interrupted for sample analysis
• Four hours after cartridge insertion, a two-point calibration
can be interrupted up to 3 consecutive times only for patient
sample analysis. Calibrations cannot be interrupted for QC
samples
• The first one-point calibration after sample analysis cannot
be interrupted
Low O2 calibration cannot be interrupted.
4-11
GEM Premier 3000 Service Manual
The cartridge has reached its time limit of 504 hours for a 3 week
cartridge, 336 hours for a 2 week cartridge or 672 hours for a 4 week
cartridge.
Blood or process control solution C has rested on the sensors for more
than 20 minutes.
The power has been interrupted for greater than one hour
In these cases, the instrument will prompt to “Remove and Discard cartridge”. In
addition, the cartridge can be removed at any time by pressing “Cartridge” on the
main screen and then pressing “Remove Cartridge” on the drop down menu. If
security is on, an authorized operator password must be entered to proceed. After
confirmation, the “Remove and Discard Cartridge” screen will be displayed.
Open the cartridge door by sliding the lock on the cartridge door forward and
opening the door. The cartridge can be removed and discarded at this time.
4-12
GEM Premier 3000 Service Manual
5. Maintenance
There is limited operator maintenance required on the GEM Premier 3000. The only
necessary maintenance is; changing printer paper, changing cartridge, emptying ampoule
breaker, and cleaning exposed surfaces. In addition, there is a Preventive Maintenance
(PM) procedure to be performed by an authorized IL service technician.
5.4 Cleaning
The instrument surfaces can be wiped clean according to the laboratory protocols.
The painted surfaces can be sprayed with a mild (10%) bleach solution, or any
approved commercially available decontaminant. To clean the LCD, spray the
decontaminant onto the wiping material, then wipe the LCD surface. Do not use
excess liquid when cleaning the LCD.
5.5 PM Procedure
Instrumentation Laboratory currently recommends a PM after the first five years
and every other year thereafter.
Preliminary
5-1
GEM Premier 3000 Service Manual
Tools
Remove cover. Visually inspect the following areas and repair if needed:
Perform Pump Spring Tension Test. Adjust pump tension to 6.2 +/- 0.1
lbs if necessary. Record pump tension as received and after setting.
Floppy Disk Ejection Test. Insert a floppy disk into the drive. Press the
ejection button to eject the floppy disk. Repeat the test three times.
Line cord: Inspect the line cord for nicks, breaks, or cuts.
General loose screw inspection: Verify that the hardware is secured on the
following assemblies and in the following areas.
Pump motor
Handle screws
Actuators
Cartridge compartment
Heater Block
Sensor flex cable: Visually inspect the sensor flex cable for fluid leak.
Display Assembly:
Check for faulty latch, loose hardware, and loose connections to the display.
Adjust the display to the upright position, and then push on the
display to tilt the unit; the latch should not slip.
5-2
GEM Premier 3000 Service Manual
When inserting the cartridge, verify that the cartridge bar code reader
correctly reads the cartridge barcode. Use the Bar code wand to enter
either the CVP or the QC ranges. Verify that the wand reads
correctly. Use the ampoule spinner to read the CVP or the QC
material. Verify that the ampoule spinner bar code reader reads the
ampoules correctly.
Loopback Test
Ethernet Test:
Instrument Information
View the System Information screen and record the serial number and
software revision.
Final Verification:
5-3
GEM Premier 3000 Service Manual
Time and date are set - verify that time and date are set correctly to the
customer’s geographic area
Software version - verify that the software version is the same as recorded on
the Preventive Maintenance Test Record “Preliminary Check-in.” The
software version is located in the upper left corner of the display screen.
5.6 PM Checklist
The following checklist can be used to record results and document the work
performed during the PM. Procedure.
ACCOUNT NAME_________________________________________________________________
PM
Comments _____________________________________________________________________________
______________________________________________________________________________________
______________________________________________________________________________________
Tech______________________________ Date______________
5-4
GEM Premier 3000 Service Manual
Tech______________________________ Date______________
Tech______________________________ Date______________
Final Verification
Verify all items checked ______ Verify TIME and DATE _______
Printouts are present and correct ______ Verify software level is correct _______
S/W Rev Level
Verify all screws are installed ______ Inspect instrument cosmetics _______
Tech______________________________ Date______________
5-5
GEM Premier 3000 Service Manual
The majority of fasteners used on the GEM 3000 are Torx screws. There are 3
different size Torx drivers used
#15
#10
#8
#8 (6 inch long shaft –15cm) can be used to remove the three screws on
the bottom of the LCD without removing the assembly.
4 mm
2 mm
1/8 “
1/16”
The ETC is used to test the instrument prior to shipment and can be used to verify
repairs. Repairs can also be verified by use of a “live” cartridge and CVP or
liquid QC material.
There are several other parts required to utilize the ETC for testing. All of the
required parts are:
6-1
GEM Premier 3000 Service Manual
ETC 00024005050
ETC Test Harness 000995410
ETC Power Supply 000995100
ETC Adapter Board 00024005162
Premtest Disk 00024005534
Skipwarm Disk 000540177
PC or Laptop with 9-pin serial connection and Premtest disk loaded.
A Loopback test connector is available to test the three serial ports on the GEM.
The instructions to perform the Loopback test are displayed on the instrument
when “Diagnostics”, “Ports”, and “COM A, COM B, or COM C” are selected.
The test can be performed at any time. To avoid any potential problems with
future transmissions, any reports in the queue should be cleared after the
Loopback tests are completed.
To gain access to the GEM 3000, the cap (top cover) must be removed. To
remove the top cover:
Remove the two plastic, protective caps at each end of the handle
6-2
GEM Premier 3000 Service Manual
Remove the three Torx screws from the cap on the back of the instrument
and lift the top cover off.
Remove the two screws from each side of the back cover that hold the side
plates on (older instruments may only have one screw on each side).
Remove the two screws on each side of the bottom plate that hold the side
plates on (older instruments may only have one screw on each side).
Remove the two side plates. Remove the two screws, one on each side,
that secure the back panel. When these screws are removed, the back
panel can be pivoted to the back to allow complete access.
Screw on left
side of unit
NOTE: When the back panel is pivoted back, there will be stress
put on the PCB connections. Do not open the back too quickly or
force it open, to avoid damage to the cables or PC Boards. To
allow the cover to be opened further, the PCB connectors must be
removed.
6-3
GEM Premier 3000 Service Manual
Access Panel
Plug adapter Board, with ETC harness connected, into connector on Analog PCB.
Handle on adapter Board must be at top.
Handle
Adapter Board
4. Turn power OFF, inset skipwarm disk into instrument and turn power ON. Select
“Skipwarm” and “OK” on screen, then press “OK” again when prompted.
6. Verify that the two actuators on the ETC are in the vertical position, with the
marked ends up. When the instrument displays the “Insert cartridge” screen,
insert the ETC.
7. Close, but DO NOT LATCH the cartridge door. Plug the ETC harness into the
connector inside the sample probe port.
6-4
GEM Premier 3000 Service Manual
Latch the cartridge door after the connection is made. The warm up will start
automatically. The warm up will take approximately 7 minutes.
8. When the warm up is completed, the instrument will display the READY screen,
and the PC will display the voltages and the temperature, along with other
information. The acceptable ranges for the voltages at are:
36.4 to 37.6 °C
The other results displayed should read “PASS”. (The valve may display “Not
aligned”. This is acceptable, as it will be tested in the Manual Premtest.)
To simulate a sample, initiate a 2-pt Calibration from the Diagnostics drop down
menu on the instrument. When the calibration is completed, press “arterial” and
“OK”.
The instrument will simulate a sample and display the results. The acceptable
ranges for these results are:
PH 7.06 to 7.08
PCO2 60 to 63 mmHg
PO 24 to 27 mmHg
Na 138.7 to 141.0 mmol/L
K 2.79 to 2.83 mmol/L
Ca 0.56 to 0.58 mmol/L
Glu 93 to 105 mg/dL
Lac 1.4 to 1.6 mmol/L
Hct 39 to 42 %
If any values are out of range, initiate a second 2-pt calibration and repeat the test.
Manual Premtest
6-5
GEM Premier 3000 Service Manual
Disconnect G/IO cable from the CPU Board. The cable is part number 20402500.
G/IO Connector
Move the ETC harness from the PC COM port to the Serial Port B on the
instrument. (see section 1.5 – Back View)
Connect a parallel printer to the parallel connector on the instrument. (see section
1.5 – Back View)
Turn power on, press and hold the “delete” key while the instrument is starting.
The instrument will display the BIOS setup screen.
Highlight the “Basic CMOS Config” screen using the arrows on the keyboard and
press enter
Set the correct date and time, if necessary and press Esc.
Highlight “Write to CMOS and Exit” and press enter. This will initiate the
Premtest. The performance of the Manual Premtest is automatic and wiil take
approximately 21 minutes.
6-6
GEM Premier 3000 Service Manual
If the test is successful, a red status box will appear, indicating the test is complete
and there were no errors. Press enter to continue
Press enter again to access the “File” dropdown on the top left of the screen. Use
the down arrow to highlight “Quit” and press enter.
Press enter again to initiate a printout on the parallel printer. Verify that all of the
tests are labeled as “Pass”. The test is complete.
Remove the Premtest disk from the instrument, disconnect the cable to the ETC,
remove the ETC from the instrument, disconnect the adapter board and all the
ETC harness connections. DO NOT disconnect the keyboard. Reconnect the
G/IO Connection.
Insert skipwarm disk in the instrument and turn power on, press and hold the
“delete” key while the instrument is starting. The instrument will display the
BIOS setup screen.
Highlight the “Reset CMOS to Factory Default”” screen using the arrows on the
keyboard and press enter. Press “Y” to confirm.
Instrument will reboot, and display screen with selection for production or
skipwarm. Press skipwarm and OK. Press OK again when prompted. Instrument
will return to “Insert cartridge” screen. Reset time and date if necessary, and
disconnect keyboard.
The pump spring tension should be checked each time the left side cover is
removed, for any reason. The purpose of checking the tension at this time is to
verify that there is no interference with the pump slides when the cover is
replaced, caused by movement of cables and harnesses.
6-7
GEM Premier 3000 Service Manual
Indicator
1. Mount the indicator inside the cartridge compartment on the lower slide of the
pump assembly. To mount the indicator, slide the pump all the way to the left by
pushing on the rollers with one hand. With the other hand, place the hook on the
indicator over the exposed lower slide, and release the pump assembly.
The pictures below show the process required to properly mount the
indicator. These pictures were taken to clarify the mounting, without
any covers, and without the printer or heater block mounted.
Lower slide
Indicator Hook
Indicator Hook
Lower slide
6-8
GEM Premier 3000 Service Manual
2. After mounting the indicator, verify that the LED on the indicator is “ON”, by
looking through the sample probe port.
LED
3. On the test cartridge, reset the gage to 0, and insert the cartridge into the
instrument slowly. Do not insert the cartridge completely. Leave approximately
¼ inch exposed.
4. Slowly raise the cartridge door while observing the LED. Immediately when the
LED extinguishes, lower the cartridge door and remove the test cartridge.
Observe the reading of the gage.
This picture shows the cartridge door position at the point the LED
typically extinguishes.
5. Reset the Gage and repeat the test two more times. Taking 3 readings will
provide a more accurate result.
6. If the results are within the range of 6.1 to 6.3, the tension is acceptable, and does
not require adjustment.
6-9
GEM Premier 3000 Service Manual
Tension Adjustment
Adjustment of the pump spring tension requires removal of the left side cover of
the instrument (section 6.2 – Gaining Access).
Loosen the lock nut and turn the adjustment screw clockwise to increase the
tension, counterclockwise to decrease the tension (2 turns = approximately 1 lb.)
After adjusting the tension, tighten the lock nut and perform the measurement to
verify the proper result. After replacing the side cover, perform the measurement
to ensure that there is no interference with the slides.
Adjusting screw
and lock nut
The SECO board can be identified by checking the instrument serial number in
Diagnostics, System Info. If the letter “S” follows the Serial number, the CPU is
a SECO board. If the letter “R” follows the Serial number, the CPU is a Radisys
board and the appropriate instructions must be used.
6-10
GEM Premier 3000 Service Manual
unit, next to the bar code reader port. The keyboard port is covered with a metal
plate that must be removed first.
Whenever software is loaded with the System Disk, ALL data will be lost on the
instrument. The hard drive is reformatted and started from scratch.
1. With Power off, connect the keyboard and insert the system disk in the
drive
4. Using the Arrow keys, highlight the BOOT SEQUENCE. Using the –
(minus) key toggle the sequence to A; -> C: and press ESCAPE to return
to the BIOS menu. Select WRITE TO CMOS AND EXIT, and press
ENTER.
6. The loading process will take about 6 minutes. The disk will be ejected
and the instrument will restart
7. After restarting, the GEM will continue loading the programs for
approximately three minutes, then it will display “System Halted”.
8. Turn the power off and then turn it back on. While powering up, rapidly
and continuously press DELETE, to enter setup and wait until the BIOS
setup screen is displayed.
10. Using the Arrow keys, highlight the BOOT SEQUENCE. Using the –
(minus) key toggle the sequence to C; -> A: and press ESCAPE to return
to the BIOS menu. Select WRITE TO CMOS AND EXIT, and press
ENTER.
6-11
GEM Premier 3000 Service Manual
11. When the prompt to enter the serial number appears, enter the unit serial
number and press enter for OK (make sure the Num Lock is enabled if
using the number pad on the keyboard).
13. The unit will display “follow the pointer and touch the spot!” There will
be a small “u” in the upper left corner.
14. Using a small, pointed object, press the spot inside the “u” on the GEM
LCD. The “u” will move to the upper right, lower left and lower right
after pressing the spot in each location.
15. The unit will display the maximum and minimum coordinates for X and
Y. These are used to calibrate the touch screen. Press Y to accept.
16. The display will change colors and the unit will restart to the insert
cartridge screen.
17. Disconnect the keyboard and replace the plate covering the connector.
The Radisys board can be identified by checking the instrument serial number in
Diagnostics, System Info. If the letter “R” follows the Serial number, the CPU is
a Radisys board. If the letter “S” follows the Serial number, the CPU is a SECO
board and the appropriate instructions must be used.
Whenever software is loaded with the System Disk, ALL data will be lost on the
instrument. The hard drive is reformatted and started from scratch.
1. With Power off, connect the keyboard and insert the system disk in the drive
6-12
GEM Premier 3000 Service Manual
2. While powering up, rapidly and continuously press F2, to enter setup, wait until
the BIOS setup Utility screen is displayed.
4. HARD DRIVE should be item 1 and highlighted. Using the down arrow,
highlight REMOVABLE DEVICES. Using the + key, move REMOVABLE
DEVICES to number 1. Press F10 to save and exit. Press enter to confirm.
5. The instrument will restart and ask for confirmation of the installation. Press
enter to begin. The loading process will take about 6 minutes. The disk will be
ejected and the instrument will restart
6. After restarting, the GEM will continue loading the programs for approximately
three minutes, then it will display “System Halted”.
7. Turn the power off and then turn it back on. While powering up, rapidly and
continuously press F2, to enter setup, wait until the BIOS setup Utility screen is
displayed.
9. REMOVABLE DEVICES will be item 1 and highlighted. Using the down arrow,
highlight HARD DRIVE. Using the + key, move HARD DRIVE to number 1.
10. Press F10 to save and exit and press enter to confirm. The unit will restart.
11. When the prompt to enter the serial number appears, enter the unit serial number
and press enter for OK (make sure the Num Lock is enabled if using the number
pad on the keyboard).
13. The unit will display “follow the pointer and touch the spot!” There will be a
small “u” in the upper left corner.
14. Using a small, pointed object, press the spot inside the “u” on the GEM LCD.
The “u” will move to the upper right, lower left and lower right after pressing the
spot in each location.
15. The unit will display the maximum and minimum coordinates for X and Y. These
are used to calibrate the touch screen. Press Y to accept.
16. The display will change colors and the unit will restart to the insert cartridge
screen.
6-13
GEM Premier 3000 Service Manual
17. Disconnect the keyboard and replace the plate covering the connector.
At the command line type the underlined texts then press <Enter>
Type lower case texts only and observe the spacing in the command lines 5,6 & 8.
3) ‘_ ‘ root <Enter>
¾
Upper left
6-14
GEM Premier 3000 Service Manual
To Exit
8) *[root@(none) /root]# init 5 <Enter>
The backlight can be replaced with a new one (part No. 24001801) or the foil can
be removed to restore the brightness. On the new backlights, as a replacement
part, the foil has been glued more completely and this should eliminate the
problem. If the foil is removed, the brightness will be restored, but will be very
slightly darker than a new light.
Using a #10 Torx driver, remove the two screws in the back of the LCD display
# 10 Screws
Using a #8 Torx driver, remove the 3 screws from the bottom of the LCD display.
If a long Torx driver is used, the screws can be removed without removing the
display assembly
#8 Screws
6-15
GEM Premier 3000 Service Manual
Open the LCD assembly and remove the 4 screws holding the LCD in place.
Disconnect the cable to the backlight.
LCD mounting
screws
Backlight
cable
connector
Lift LCD carefully and push down on tab (located on left side) to disconnect the
LCD input.
Turn the LCD over and press on release tab for backlight. With tab pressed, slide
backlight out of LCD.
6-16
GEM Premier 3000 Service Manual
Release tab
Replace the bulb in the holder and reverse the steps above to reassemble the
instrument
With the backlight removed, carefully extract the wires from the plastic holder
and carefully lift the bulb out of the holder. The foil may come out of the holder
with the bulb, or it may have to be removed from the holder
6-17
GEM Premier 3000 Service Manual
Bulb
Holder
Foil
Discard the foil, replace the bulb in the holder and reverse the steps above to
reassemble the instrument.
6-18
GEM Premier 3000 Service Manual
7. Troubleshooting
This section contains a list of the GEM 3000 error codes and a description of the
associated error, along with the probable cause and fix. There is also a list of
reported symptoms and fixes that has been developed from the experience gained
to date. In addition, there are some part replacement procedures.
7-1
GEM Premier 3000 Service Manual
Alarm Notification
For alarms that require operator intervention to resolve, such as the printer is out
of paper, the GEM 3000 operating software presents an alarm on the screen.
For error handling that requires immediate operator action, the operating software
displays a user dialog box, with an <OK> button, that describes the error that has
occurred and instructs the operator on what to do next. For example:
Error 1.06
No sample detected.
Test canceled.
LLP Reset
The GEM 3000 HLP (High Level Processor) software commands the GEM 3000
LLP (Low Level Processor) to perform a hardware reset when certain error
conditions occur (for example, when the LLP reports a failure to calibrate the A/D
converter at the start of a sample or calibration). When the LLP resets, it restarts
at the beginning of its code, initializes all of the LLP hardware, calibrates the A/D
converter, and verifies that the valve and arm actuators can be moved to their
home positions.
When the LLP is reset due to an error that occurred while attempting to perform a
particular procedure (such as attempting to detect calibration solution), the desired
procedure is performed a second time. If an error occurs again, the LLP is reset a
second time, and the procedure is performed a third time. If an error occurs during
the third attempt, the operating software either displays a user dialog box that
explains what has happened, or the operating software performs a system reset.
System reset
When the GEM 3000 operating software is about to perform a system reset, a
display is presented with the following message:
<type of error> X.XX
This instrument will reset in 30 seconds.
7-2
GEM Premier 3000 Service Manual
where X.XX is the number of the error that occurred (for example, 3.01) and <type
of error> is one of the following lines of text:
Cartridge error
Hardware error
Software error
Then, the GEM 3000 system resets itself, which includes resetting both its HLP
(High Level Processor) and LLP (Low Level Processor).
After the instrument is reset, the following message is printed:
SYSTEM RESET
04/07/2000 15:25:00
<type of error> X.XX
If the problem persists,
contact Technical Service.
The GEM 3000 operating software maintains the number of times it resets itself,
due to an error, in a persistent data file on its internal disk. This persistent data file
is cleared during the daily log rotation. If a system reset occurs because of an
error (any error) four times between log rotations, the GEM 3000 performs a
system halt. While the system is halting, a display is presented with the following
message:
<type of error> X.XX
The instrument will be halted.
Please wait.
and the following message is printed:
SYSTEM BEING HALTED
04/07/2000 15:25:00
<type of error> X.XX
If the problem persists,
contact Technical Service.
When the system has halted, a display is presented with the following message:
<type of error> X.XX
The instrument has been halted.
7-3
GEM Premier 3000 Service Manual
7-4
GEM Premier 3000 Service Manual
7-5
GEM Premier 3000 Service Manual
Remedy Check the connection with the LIS system and the
configuration of the LIS system then clear the alarm. If the
alarm occurs again, cycle power on the instrument.
Label ERR_PRINTER_FAILURE_EXT
Alarm Message Parallel printer error
When Whenever the operating software is attempting to print
to the parallel printer.
Cause There is a hardware problem with the printer or printer
cable.
Handling Alarm notification for each failed attempt to write to
the printer if an alarm is not already posted for this
error condition.
Remedy Check parallel printer connections then reset printer
queue and try again. Cycle power to the printer if
necessary.
Label ERR_NO_PAPER_EXT
Alarm Message Parallel printer out of paper
When Whenever the operating software is attempting to print
to the parallel printer.
Cause The printer is out of paper.
Handling Alarm notification for each failed attempt to write to
the printer if an alarm is not already posted for this
error condition.
Remedy Install paper in the printer and clear the alarm. Reset the
printer queue. Cycle power to the printer if necessary.
Label ERR_PRINTER_FAILURE_NT
Alarm Message Network printer error
7-6
GEM Premier 3000 Service Manual
Label ERR_IMPROPER_SHUTDOWN
Alarm Message Improper instrument shutdown
When When the instrument restarts after it was improperly
shutdown.
Cause The “Shutdown” button was not used to shut down the
instrument before the instrument was powered off.
This can be due to an unexpected power outage or
operator error.
Handling The alarm message is logged in the system error log file
“syslog.dat”. The alarm is not displayed on the
“Alarms” screen.
Remedy Users/customers must be reminded to user proper
shutdown procedure from the Shutdown pull-down
menu.
Label ERR_IQM_PH
Alarm Message pH is permanently disabled due to iQM error
When After iQM corrective action failed to correct the iQM
error triggered following A, B, or C calibration.
Cause Sensor and/or calibration solution error.
Handling Alarm notification whenever a sensor is permanently
disabled due to unrecoverable iQM error.
7-7
GEM Premier 3000 Service Manual
Label ERR_IQM_PCO2
Alarm Message pCO2 is permanently disabled due to iQM error
When After iQM corrective action failed to correct the iQM
error triggered following A, B, or C calibration
Cause Sensor and/or calibration solution error.
Handling Alarm notification whenever a sensor is permanently
disabled due to unrecoverable iQM error.
Remedy Replace cartridge if failed sensor is essential to the
continued operation of the instrument
Label ERR_IQM_PO2
Alarm Message pO2 is permanently disabled due to iQM error
When After iQM corrective action failed to correct the iQM
error triggered following A, B, or C calibration.
Cause Sensor and/or calibration solution error.
Handling Alarm notification whenever a sensor is permanently
disabled due to unrecoverable iQM error.
Remedy Replace cartridge if failed sensor is essential to the
continued operation of the instrument.
Label ERR_IQM_NA
Alarm Message Na+ is permanently disabled due to iQM error
When After iQM corrective action failed to correct the iQM
error triggered following A or B calibration.
Cause Sensor and/or calibration solution error.
Handling Alarm notification whenever a sensor is permanently
disabled due to unrecoverable iQM error.
7-8
GEM Premier 3000 Service Manual
Label ERR_IQM_K
Alarm Message K+ is permanently disabled due to iQM error
When After iQM corrective action failed to correct the iQM
error triggered following A or B calibration.
Cause Sensor and/or calibration solution error.
Handling Alarm notification whenever a sensor is permanently
disabled due to unrecoverable iQM error.
Remedy Replace cartridge if failed sensor is essential to the
continued operation of the instrument.
Label ERR_IQM_CA
Alarm Message Ca++ is permanently disabled due to iQM error
When After iQM corrective action failed to correct the iQM
error triggered following A or B calibration.
Cause Sensor and/or calibration solution error.
Handling Alarm notification whenever a sensor is permanently
disabled due to unrecoverable iQM error.
Remedy Replace cartridge if failed sensor is essential to the
continued operation of the instrument
7-9
GEM Premier 3000 Service Manual
7-10
GEM Premier 3000 Service Manual
7-11
GEM Premier 3000 Service Manual
7-12
GEM Premier 3000 Service Manual
7-13
GEM Premier 3000 Service Manual
Label ERR_BAD_SENSOR_VALS
Log Message Na, K, Ca, pH at A/D rail for B cal
When During B calibration measurement.
Cause Any 3 of the Na, K, Ca, or pH sensor readings are at the
A/D rail and any 3 of those sensor readings were at the A/D
rail for the previous B calibration. This is probably due to
reference failure going to rail due to air bubble on reference
wire.
Handling The sensor A/D readings that were taken for the calibration
are saved to the sensor data file. Then, a system reset is
performed. No calibration report is printed, saved to the
database, or transmitted.
7-14
GEM Premier 3000 Service Manual
7-15
GEM Premier 3000 Service Manual
This section describes the GEM 3000 hardware errors that can occur (2.01-2.99).
And be displayed. The first attempt at a remedy for a persistent hardware error is
to cycle the instrument power.
7-16
GEM Premier 3000 Service Manual
Label ERR_HBLK_WARM_UP_TOUT
Log Message Power Fail Recovery temperature error
When During power fail recovery.
Cause The GEM 3000 operating software timed out waiting for
the heater block to reach a temperature of 37.0 degrees C.
Handling System reset is performed.
Label ERR_BAD_CAL_GND
Log Message Calibration ground error
When During calibration measurement.
Cause The ground A/D reading is not within +/- five bits of the
expected value.
Handling The sensor A/D readings, including ground, that were taken
for the calibration are saved to the sensor data file with a
bad ground error status. Then, a system reset is performed.
No calibration report is printed, saved to the database, or
transmitted.
7-17
GEM Premier 3000 Service Manual
Label ERR_BAD_CAL_TEMP
Log Message Calibration temperature error
When During calibration measurement.
Cause The temperature A/D reading is not 37.0 degrees C.
Handling The sensor A/D readings, including temperature, that were
taken for the calibration are saved to the sensor data file
with a bad temperature error status. Then, a system reset is
performed. No calibration report is printed, saved to the
database, or transmitted.
7-18
GEM Premier 3000 Service Manual
Label ERR_A2D_CAL
Log Message A/D converter calibration error
When During sample or calibration preparation.
Cause The A/D converter was unable to obtain valid (+/- 5 bits)
readings from the ground and/or voltage reference channels
following calibration of the A/D converter.
Handling Up to three attempts are made to calibrate the A/D
converter. After each of the first two attempts, an LLP reset
is performed. If the third attempt is unsuccessful, a system
reset is performed. No sample or calibration report is
displayed, printed, saved to the database, or transmitted.
Label ERR_A2D_READ
7-19
GEM Premier 3000 Service Manual
Label ERR_D2A_WRITE
Log Message D/A converter write error
When During LLP initialization.
Cause An error was detected and signaled by the LLP (Low Level
Processor) when it attempted to write to a D/A converter
using the synchronous serial bus.
Handling System reset is performed.
Label ERR_ARM_HOMING
Log Message Sample arm homing error
When During sample preparation or LLP initialization.
Cause The sample arm could not be returned to the home position,
or the sample arm homing sensor could not detect the
homing pin.
Handling System reset is performed.
Label ERR_VALVE_HOMING
Log Message Valve homing error
When During sample or calibration preparation, or LLP
initialization.
Cause The valve could not be returned to the home position, or the
valve homing sensor could not detect the homing pin.
7-20
GEM Premier 3000 Service Manual
Label ERR_NO_INST_SERIAL_NUM
Log Message Uninitialized instrument serial number
When During system start up.
Cause No instrument serial number was found in the instrument’s
non-volatile memory.
Handling System is halted.
Label ERR_FSCK_FAILED
Log Message File system check/repair failed
When During system start up.
Cause The file system check, performed when the GEM 3000
operating system software starts up, failed. This file system
check is performed by the Linux operating system utility
“fsck”.
Handling System is halted.
7-21
GEM Premier 3000 Service Manual
This section describes the GEM 3000 software errors that can occur (3.01
– 3.99). The remedy for a persistent software error is to reinstall the GEM
3000 application software.
For any of the software errors, the message displayed by the instrument
will be:
<type of error> X.XX
This instrument will reset in 30 seconds.
Label ERR_BAD_TIME_DATE
Log Message Unknown cartridge age
When During cartridge warm up or power fail recovery.
Cause The amount of time that power was off or the age of the
cartridge could not be determined. That is, the last time
saved is negative or greater than the system time. This may
indicate corruption of the persistent data file that contains
the last time saved.
Handling System reset is performed.
7-22
GEM Premier 3000 Service Manual
Remedy Confirm that the system time is correct. If the error occurs
following the system reset, remove the cartridge, copy the
cartridge data to a diskette, then reinstall the GEM 3000
application software.
Label ERR_LLP_HDR_READ
Log Message No header in message from LLP
When Whenever the HLP is communicating with the LLP, such
as during a sample or calibration.
Cause The GEM 3000 HLP (High Level Processor) did not find a
header in the message it received from the LLP (Low Level
Processor). The HLP needs the header to determine the
type of data contained in the message.
Handling System reset is performed.
Label ERR_LLP_HDR_MISMATCH
Log Message Unexpected header in message from LLP
When Whenever the HLP is communicating with the LLP, such
as during a sample or calibration.
Cause The GEM 3000 HLP (High Level Processor) did not find
the header it expected in the message it received from the
LLP (Low Level Processor). The HLP needs the header to
determine the type of data contained in the message.
Handling System reset is performed.
Label ERR_LLP_DATA_READ
Log Message No data in message from LLP
When Whenever the HLP is communicating with the LLP, such
as during a sample or calibration.
7-23
GEM Premier 3000 Service Manual
Cause The GEM 3000 HLP (High Level Processor) found only a
header in the message it received from the LLP (Low Level
Processor) but did not find any data in the message.
Handling System reset is performed.
Label ERR_STS_MSG_HDR
Log Message No status message from LLP
When Whenever the HLP sends the status command to the LLP,
such as when the HLP polls the LLP to determine if a
cartridge has been inserted or removed, or when the HLP
sends the status command to the LLP a few seconds after
sending a reset command to the LLP.
Cause The GEM 3000 HLP (High Level Processor) was waiting
to receive a status message from the LLP (Low Level
Processor), but the message that the HLP received did not
contain a status message header.
Handling System reset is performed.
Label ERR_CMP_MSG_HDR
Log Message No batch complete message from LLP
When Whenever the HLP sends a batch of commands to the LLP,
such as during a sample, calibration, or LLP initialization.
Cause The GEM 3000 HLP (High Level Processor) was waiting
to receive a batch complete message (which indicates that a
set of commands has been processed by the LLP) from the
LLP (Low Level Processor), but the message that the HLP
received did not contain a batch complete message header.
Handling System reset is performed.
Label ERR_AD_MSG_HDR
Log Message No sensor reading message from LLP
7-24
GEM Premier 3000 Service Manual
Label ERR_AS_MSG_HDR
Log Message No sensor readings message from LLP
When During sample or calibration measurement.
Cause The GEM 3000 HLP (High Level Processor) was waiting
to receive a signal channels (analyte channels, temperature,
ground, reference) sensor reading message from the LLP
(Low Level Processor), but the message that the HLP
received did not contain a signal channels sensor reading
message header.
Handling System reset is performed
Label ERR_SOLN_MSG_HDR
Log Message No solution type message from LLP
When During sample aspiration or when solution is pumped from
the cartridge during a calibration.
Cause The GEM 3000 HLP (High Level Processor) was waiting
to receive a solution type message from the LLP (Low
Level Processor), but the message that the HLP received
did not contain a solution type message header.
Handling System reset is performed.
7-25
GEM Premier 3000 Service Manual
Label ERR_REV_MSG_HDR
Log Message No revision message from LLP
When Whenever the HLP sends the revision command to the
LLP, such as during system initialization.
Cause The GEM 3000 HLP (High Level Processor) was waiting
to receive a revision message from the LLP (Low Level
Processor), but the message that the HLP received did not
contain a revision message header.
Handling System reset is performed.
Label ERR_INIT_LLP
Log Message LLP initialization failed
When During LLP initialization.
Cause The GEM 3000 HLP (High Level Processor) received an
unexpected error message from the LLP (Low Level
Processor) during LLP initialization.
Handling System reset is performed.
Label ERR_B_CAL_BAD_SOLN
Log Message No B solution message from LLP
When While B solution is being pumped from the cartridge
during a B calibration.
Cause The GEM 3000 HLP (High Level Processor) received an
incorrect solution message from the LLP (Low Level
Processor) during a B calibration. The HLP was expecting
a B solution message.
Handling System reset is performed.
7-26
GEM Premier 3000 Service Manual
Label ERR_A_CAL_BAD_SOLN
Log Message No A solution message from LLP
When While A solution is being pumped from the cartridge
during an A calibration.
Cause The GEM 3000 HLP (High Level Processor) received an
incorrect solution message from the LLP (Low Level
Processor) during an A calibration. The HLP was expecting
an A solution message.
Handling System reset is performed.
Label ERR_SAMP_BAD_SOLN
Log Message No sample solution message from LLP
When During sample aspiration.
Cause The GEM 3000 HLP (High Level Processor) received an
incorrect solution message from the LLP (Low Level
Processor) during sample processing. The HLP was
expecting a sample solution message.
Handling System reset is performed.
Label ERR_UNEXP_LLP_MSG
Log Message Unexpected message from LLP
When Whenever the HLP is communicating with the LLP, such
as during a sample or calibration.
Cause The GEM 3000 HLP (High Level Processor) received an
unexpected message from the LLP (Low Level Processor).
Handling System reset is performed.
7-27
GEM Premier 3000 Service Manual
Label ERR_LLPD_MAX_RESENDS
Log Message HLP to LLP maximum message resends
When Whenever the HLP is communicating with the LLP, such
as during a sample or calibration.
Cause The GEM 3000 LLP (Low Level Processor) is not
acknowledging (by responding with an ASCII ACK or
NAK character) a command sent from the HLP (High
Level Processor). Or, the GEM 3000 LLP has requested
the HLP to resend a command the maximum number of
times.
Handling System reset is performed.
Label ERR_CARTD_TIMEOUT
Log Message Cartridge daemon timeout
When When the GEM 3000 HLP (High Level Processor) is
waiting for the LLP (Low Level Processor) to complete a
cartridge function, such as positioning the valve or moving
the sample arm.
Cause The GEM 3000 HLP (High Level Processor) waited the
maximum amount of time for the LLP (Low Level
Processor) to complete a cartridge function.
Handling System reset is performed.
Label ERR_CARTD_UNEXP_EVENT
Log Message Unexpected cartridge daemon event
When When the GEM 3000 HLP (High Level Processor) is
waiting for the LLP (Low Level Processor) to complete a
cartridge function, such as positioning the valve or moving
the sample arm.
Cause Communication between the GEM 3000 HLP (High Level
Processor) and LLP (Low Level Processor) is not
7-28
GEM Premier 3000 Service Manual
Label ERR_UNEXP_MSG_FROM_CARTD
Log Message Frontend and cartridge out of sync
When When a GEM 3000 front end process, such as Ready,
receives an unexpected message from the cartridge daemon
(for example, during sample aspiration).
Cause Communication between a GEM 3000 front end process
and the cartridge daemon is not synchronized. The
communication failure may be due to an error detected by
the cartridge daemon and for which the system is about to
reset (for example, an unrecoverable A/D calibration error).
Handling System reset is performed.
Label ERR_TIMEOUT_FROM_CARTD
Log Message Frontend timed out on cartridge response
When When a GEM 3000 front end process, such as Ready, times
out waiting to receive a message from the cartridge daemon
(for example, during sample aspiration).
Cause Communication between a GEM 3000 front end process
and the cartridge daemon is not synchronized. The
communication failure may be due to an error detected by
the cartridge daemon and for which the system is about to
reset (for example, an unrecoverable A/D calibration error).
Handling System reset is performed.
7-29
GEM Premier 3000 Service Manual
Label ERR_DB_ERROR
Log Message Database error: %s
(where %s is the database name)
When Whenever a database operation is performed.
Cause A database operation has failed.
Handling System reset is performed.
Label ERR_FILE_IO
Log Message File I/O error: %s
(where %s is the file name)
When Whenever a file I/O operation is performed.
Cause A file I/O operation has failed.
Handling System reset is performed.
Label ERR_IPC_ERROR
Log Message Inter Process Communication error: %s
(where %s is the process name)
When Whenever communication between two processes occurs.
Cause Communication between two processes has failed.
Handling System reset is performed.
Label ERR_FSCK_ERROR
Log Message File system check detected error
When During system start up.
7-30
GEM Premier 3000 Service Manual
Cause The file system check, performed when the GEM 3000
operating system software starts up, detected an error. This
may indicate that the instrument’s internal disk is starting
to fail. The file system check is performed by the Linux
operating system utility “fsck”. When an error is detected,
“fsck” attempts to fix the area of the file system that caused
the error.
Handling The log message is logged in the system error log file
“syslog.dat”.
Label ERR_ABORT_LLP
Log Message No abort message from LLP
When When the HLP sends an abort command to the LLP, which
causes the LLP to stop execution of any command it is
currently performing and clear its command queue.
Cause The GEM 3000 HLP (High Level Processor) was waiting
to receive an abort message from the LLP (Low Level
Processor) in response to the abort command, but the
message that the HLP received did not contain an abort
message header.
Handling System reset is performed.
7-31
GEM Premier 3000 Service Manual
This section describes the GEM 3000 events that are logged in the system
error log file “syslog.dat”. (9.01 - 9.99, 0.00 - -0.99)
9.01 Event
Label EVENT_INSERT_CART
Log Message Cartridge inserted: %s
(where %s is the cartridge barcode)
When When a cartridge is inserted into the instrument.
9.02 Event
Label EVENT_POWFAIL_STATE
Log Message Cartridge power fail recovery state
When When the instrument is restarted with a cartridge inserted.
Not implemented as of V5.4.
9.03 Event
Label EVENT_REMOVE_CART
Log Message Remove cartridge state: %s
(where %s is the reason the cartridge is to be removed)
When When the instrument goes to the remove cartridge state.
9.04 Event
Label EVENT_REMOVED_CART
Log Message Cartridge removed: %s
(where %s is the cartridge barcode)
When When a cartridge is removed from the instrument.
9.05 Event
Label EVENT_CHANGE_TIME_DATE
Log Message Time/date changed: %s
(where %s is the time and date)
When When time and date is changed on the instrument.
9.06 Event
Label EVENT_CHANGE_COLOR
7-32
GEM Premier 3000 Service Manual
9.07 Event
Label EVENT_SYS_RESTART
Log Message Instrument restart
When When the instrument is restarted. Not implemented as of
V5.4.
9.08 Event
Label EVENT_SYS_SHUTDOWN
Log Message Instrument shutdown
When When the instrument is shut down. Not implemented as of
V5.4.
9.09 Event
Label EVENT_REF_SHIF
Log Message Reference channel shifted during cal A or B
When When the instrument detects a shift in Reference sensor
during A or B calibration.
9.10 Event
Label EVENT_BLOOD_CLOT_PATTERN1
Log Message Blood clot pattern 1 detected.
When When the instrument detects the iQM clot pattern 1 in B
calibration following patient sample.
9.11 Event
Label EVENT_BLOOD_CLOT_PATTERN2
Log Message Blood clot pattern 2 detected.
When When the instrument detects the iQM clot pattern 2 in A
calibration sometime after a patient sample.
9.12 Event
Label EVENT_BLOOD_PATTERN3
Log Message Pattern 3 pO2 malfunction detected.
7-33
GEM Premier 3000 Service Manual
9.13 Event
Label EVENT_NOAIR_DETECT
Log Message Clot Pattern No-Air in B after sample detected.
When When the instrument encounters Air detection failure
during the B calibration cycle immediately following a
patient sample.
9.14 Event
Label EVENT_NOSOLU_DETECT
Log Message Clot Pattern No-Solu in B after sample detected.
When When the instrument encounters calibration solution
detection failure during the B calibration cycle immediately
following a patient sample.
9.15 Event
Label EVENT_INTERFERENCE
Log Message Sample interference detected.
When When the instrument detects Thiopental or Benzalkonium
sample interference pattern in the B calibration
immediately following a patient sample.
0.00 Event
Label CLEAR_ERR_ALL
Log Message All alarms cleared
When When all the alarms [see Section 7.1.1] have been cleared.
7-34
GEM Premier 3000 Service Manual
7-35
GEM Premier 3000 Service Manual
7.2 Symptom/Fix
This section contains a list of reported or observed symptoms and the most likely
method of repair. This has been developed from the experience of the Depot repair
department to date. Please note that these are not the only methods of repair, but
typically the most common. In some cases, there will be several fixes listed. These
are symptoms that have had multiple causes in the past, and are listed in order from
most likely to least likely. The cartridge errors listed (1.02 – 1.11) are fixes seen
when the errors are consistent on multiple cartridges. In all cases, errors 1.xx should
first have a cartridge replaced.
Error Fix
1.02 Pump Rollers, Pump Motor, DCS Bd., Analog Bd,
Heater block too tight.
1.04 Pump Roller, Pump motor, Poor contact at heater block
1.05 Pump Rollers, Pin in rollers Pump Bracket, DCS Bd.
1.06 Pump Rollers, Pin in rollers Pump Bracket.
1.08 Heater Block, pump rollers.
1.09 Pump rollers, Analog Bd., Heater Block, Pump motor.
1.11 Analog Bd., Power Supply.
2.01 Heater Block, Analog Bd.
2.03 Analog Bd., Power Supply, Grounds.
2.05 Fan, Heater Block.
2.07 Analog Bd., Power Supply, Grounds
2.08 CPU Bd.
2.10 Sample Arm loose.
2.11 Dust on Arm Sensor, CPU
2.14 Heater Block, Fan, Fan direction, Heat Sink on CPU.
2.15 Pump Roller, Pump Motor, Pump Bracket, Heater Block
tight.
3.05 Application Disk, Hard drive, DCS Bd. (need rev 4
minimum)
3.08 Application Disk, Hard drive, DCS Bd. (need rev 4
minimum)
7-36
GEM Premier 3000 Service Manual
3.15 CPU
3.16 Reload Software
3.17 Hard Drive, CPU
3.18 Application Disk
3.19 Reload software
PO2 High Heater Block, Analog Bd., DCS Bd.
Printer Failed Printer, Printer Controller Bd.
No Op. Sys Found Hard Drive
Display Dim Backlight or Remove Foil
Lac, Glu, PO2 slope Analog Bd.
Errors
K Failure Analog Bd.
Na Drift Analog Bd.
Disk Will not Eject Floppy Disk Drive.
Will not Recover from Hard Drive.
Improper Shutdown
Screen Blank LCD, Inverter Bd.
Will not Complete Start Hard Drive
Up
Resetting CPU, Backplane, Analog Bd., Heater Block.
PO2 or PCO2 PVP Heater Block.
Failure
7-37
GEM Premier 3000 Service Manual
8. Interfacing
8.1 Interface Specifications
The interface specifications for the GEM Premier 3000 are contained on a
1.44M disk, part number 24001695. These specifications are in a pdf
format and can be given to customers to address any questions that arise
regarding the instrument interface via the Ethernet or serial ports.
The Gem 3000 can be interfaced to either the IL682 or the ILOPL, to have
the CO-Oximeter data transferred to the GEM. The CO-Oximeter data
will be used to perform calculations for derived parameters, and it will be
printed, transmitted from and stored on the GEM. If a CO-Ox only
sample is analyzed and transmitted to the GEM, it will not affect the
number of tests available on the cartridge.
GEM 3000
Connect one end of the interface cable to Serial/COM port A, B, or C on
the back of the GEM.
IL682
Connect the other end of the serial cable to the “Serial 2” port on
the back of the instrument.
8-1
GEM Premier 3000 Service Manual
Standard
Baud rate – 9600
Data Bits – 8
Parity – None
Stop Bits – 1
Ack/Nack – OFF
XON/XOFF – OFF
Send Cal Data – OFF
Send QC Data – ON or OFF (as required)
Start Char – STX
Header – ON
Instrument ID – 1
Disable the Fetal Mode on the 682 (Menu 4,2,3,3,2,2) as the GEM will
perform the calculations if the Fetal % is entered. If the 682 performs the
calculation also, the data can be affected.
GEM 3000
Connect one end of the interface cable to Serial/COM port A, B, or C on
the back of the GEM.
8-2
GEM Premier 3000 Service Manual
GEM OPL
Connect other end of the cable to the RS-232 Serial Port on the
back of the OPL
8-3
GEM Premier 3000 Service Manual
9. Parts List
9.1 Items Sorted by Part Number
9-1
GEM Premier 3000 Service Manual
9-2
GEM Premier 3000 Service Manual
Cartridges
Cartridges
Supplies
005508 Print paper, 5 rolls
000825900 Capillary Sample kit, 200 tubes
24001170 Kit safety draw plastic capillary 1000pk
005330 Formatted Blank Diskette, 10/pack
9-3
GEM Premier 3000 Service Manual
CVP
Multipak, 20 ampoules x 2.5 mL x 4
24001587
levels
24001811 CVP 1, 20 ampoules x 2.5 mL
24001812 CVP 2, 20 ampoules x 2.5 mL
24001813 CVP 3, 20 ampoules x 2.5 mL
24001814 CVP 4, 20 ampoules x 2.5 mL
ContrIL 7
24001380 Multipak, 30 ampoules x 2mL x 3 levels
24001381 Level 1, 30 ampoules x 2mL
24001382 Level 2, 30 ampoules x 2 mL
24001383 Level 3, 30 ampoules x 2 mL
ContrIL 9
24001418 Multipak, 30 ampoules x 2mL x 3 levels
24001419 Level 1, 30 ampoules x 2mL
24001420 Level 2, 30 ampoules x 2mL
24001421 Level 3, 30 ampoules x 2mL
Accessories
18420238 Interface cable GEM to CO-Ox
005315 Mounting Bracket
24000130 Bar code gun (optional)
9-4
GEM Premier 3000 Service Manual
10-1
10-2
10-3
10-4
10-5
10-6
10-7
10-8
10-9
10-10
10-11
10-12
10-13
10-14
10-15
10-16
10-17
10-18
10-19
10-20
10-21
10-22
10-23
10-24
10-25
10-26
10-27
10-28
10-29
10-30
10-31
10-32
10-33
10-34
10-35
10-36
10-37
10-38
10-39
10-40
10-41
10-42
10-43
10-44
10-45
10-46
GEM Premier 3000 Service Manual
11-1
Service Product Information
Gem Premier 3000
1
Point of Care
March, 02 Page 1
11-2
Service Product Information
We have had several reports of broken printer doors during installation in the last
couple of months (3 of the last 50 installations). One or both of the tabs holding the
printer door breaks off during shipping. These tabs are part of the entire front panel
and the panel must be replaced to repair the door.
We are currently in the process of changing the mold and revising the printer door
tabs. Starting with serial number 13643, during July’s production, the molded tabs
will be tapered to relieve the stress and eliminate the breakage.
Critical Care
Jim Carpenter
Product Line Support Manager – Critical Care
Page 1
11-3
Service Product Information
There will be two upgrade kits made available, one for 5.2 to 5.3, one for 5.0/5.1 to 5.3. The 5.2 to
5.3 kit will be distributed immediately when it is available. A sufficient quantity will be distributed to
each TSR to accomodate the accounts identified as being at revision 5.2. The second kit will be
available at a later date and can be utilized after the 5.2 upgrades are completed.
When installing 5.3 software on a 5.2 instrument, the video disks will not need to be reinstalled, as
the video will not be lost with the upgrade. If an account has instruments at 5.2 and at earlier
revisions, all of the units may be upgraded to 5.3 using the same upgrade disk. It is important to
remember that the upgrade disk(s) and the video disks used at a specific account must be left at the
account and not used at other accounts.
Point of Care
The Faxback form, for 5.2 to 5.3 upgrades, has been sent to all field personnel. Please complete and
fax this form for all Demo instruments. This will enable the use of an application disk rather than an
upgrade kit for IL instrumentation.
The 5.2 Demo disk (00540171) will not be updated. The 5.2 Demo disk will work with 5.3 software.
The Video disks (24001612 – set of 2) are labelled for 5.2 and will also work with 5.3. These disks
will not be changed at his time, but will be used with revision 5.3
IMPORTANT: For 5.2 to 5.3 upgrades, Service Representatives MUST inform customers that an
instruction label was added to p. 3.1 of the Operator’s Manual, requiring the
verification of manual configuration changes.
Jim Carpenter
Product Line Support Manager
Critical Care
Page 1
11-4
Service Product Information
If the format of YYYY/MM/DD is selected, the instrument will not be able to access
the Delta Charts, either at the instrument, or via GEMweb. When the chart is
selected, the instrument screen, or the GEMweb screen, will turn gray and remain in
that state until the instrument is powered off and on. All other time/date formats
work properly.
Accessing the Delta Charts is the only function affected. The cause has been
identified and will be corrected in version 5.4
Jim Carpenter
Product Line Support Manager
Critical Care
Page 1
11-5
Service Product Information
The GEM Premier 3000 has utilized a Radysis CPU board (P/N 0024005242) since
release. There is a new Seco CPU board (P/N 0024001597) currently being used.
The new CPU was installed in GEM instruments starting with SN 14015 until SN
14101. Instruments with SN 14102 through 14689 then had Radisys boards
installed. From SN 14690 on SECO boards have been used. The Seco board will
only work with software revision 5.1.2 and higher, the Radysis board will work with
all revisions. With Revision 5.3 and higher, select the Diagnostics Menu and
System Info. The Serial number of the instrument will be displayed, with an R or an
Critical Care
S several spaces over. This will help to determine the type of CPU in use (R =
Radysis, S = SECO).
The new Seco board is complete as received, with all components installed. There
is no need to add any parts from the old Radysis board when installing it.
In order to install a Seco board in place of a Radysis board, there are several parts
that must be changed. To facilitate this change, a kit will be been made available
with all of the required parts. The kit part number is 0024001632 and contains:
If the super floppy drive is an older version used on all GEMs below SN 13411, it
must also be changed, along with the power cable. If the super floppy has only 2
wires to the power connection, it is the old version. The proper floppy drive has a 3-
wire power connection. The part number of the floppy drive is 0024005241 and the
part number for the power cable is 0056005800.
After installing the new CPU board, connect the new GIO cable from the backplane
(J-21) to the CPU board (J-4) and the sound card (J-3). The sound card is mounted
onto the CPU board.
Page 1
11-6
Service Product Information
Purpose: The purpose of this Service Product Information Bulletin is to inform you
that we have recently become aware of a potential problem with the GEM Premier
3000. There are two cables in the GEM that could interfere with the cooling fan,
causing the fan to run slowly or not at all. The two cables are the large, black cable
that runs from the Heater Block to the Analog board and a gray cable that runs to
the DCS Board.
These cables were secured to the instrument chassis with a tie wrap to keep them
from interfering with the fan. The use of the tie wrap was discontinued when the
SECO CPU board was introduced (Serial Number 14690). After discovering that the
cables could still interfere with the fan, the use of the tie wrap was implemented
again with October 2003 production (Serial Number 15137).
Critical Care
If a GEM in this serial number range displays any temperature errors, particularly
Error 2.14, the fan should be checked for proper operation. If the fan is not blowing
air OUT, the cover should be removed and the fan checked for interference. There
is a bracket a few inches from the fan, toward the middle of the instrument. This
bracket has a single hole in it and a tie wrap, or anything suitable, can be passed
through this hole, wrapped around the two cables and secured. This will prevent the
cables from causing any problems if they shift during operation or when the
instrument is moved.
Jim Carpenter
Page 1
11-7
Service Product Information
Service Product Information 01– February, 2004
Purpose: The purpose of this Service Product Information Bulletin is to inform you
that the GEM Premier 3000 has a new 130W Power Supply. This new Power
Supply was implemented in instruments with serial number 14849 and higher.
The new supply can be utilized in older instruments by using a Power Supply Kit,
part number 0020405200. This kit contains:
These parts are used to replace the older Power Supply (00790035), the old Power
Cable (00560058), and the black, green, and white wires that run from the power
Critical Care
switch to the power supply.
At this time there is a limited quantity of the old Power Supply (00790035) available
to support the depot repair facilities. When these old supplies are no longer
available, the kit (0020405200) must be used.
Jim Carpenter
Page 1
11-8
Service Product Information
Service Product Information– February, 2004
Purpose: The purpose of this Service Product Information Bulletin is to inform you
that the GEM Premier 3000 has a new “slim line” disk drive. This new disk drive
was implemented in instruments with serial number 15456 and higher.
The new drive can be utilized in older instruments by ordering the four required parts
individually. There is not a kit available at this time. The four parts required are:
If a new, slim line drive fails, it can be replaced with just the new drive
Critical Care
(0024005413).
The existing drive, on the older GEM 3000 instruments is part number 0024005241.
A significant quantity of the old drives is available, to eliminate the need to replace
all of the parts required. It is recommended that the old drives be used in the old
instruments whenever possible
Jim Carpenter
Page 1
11-9
Service Product Information
Purpose: The purpose of this Service Product Information Bulletin is to inform you
that we have made two changes in the manufacturing process of the GEM Premier
3000 to eliminate problems recently seen with shipping of the instrument.
The first change is in the length of the double-sided tape used to mount the glass
touch screen. The tape on both sides was previously 7.375 inches long, and is now
5 inches long. This will remove the pressure point in the corners that resulted in
several touch screens being broken during shipping.
The second change is in the mounting of the Ethernet connector to the back panel.
There have been several instances where the Ethernet connector has fallen inside
the instrument during shipping. There is now a shim under the fitting that snaps the
connector in place on the back panel. This will increase the amount of pressure on
Critical Care
the fitting, securing it tightly to the panel. The connector can still be removed, if
necessary, by squeezing the tabs together, but more force will be required to
release the connector.
Both of these changes are being implemented on instrument serial number 15950.
Jim Carpenter
Page 1
11-10
Service Information Notes
Symptom: The CPU board may have a timing conflict with the memory cache.
If this occurs, the instrument will “freeze”, and it must be powered off and on to
recover the operation. This may result in loss of the cartridge. This will only
occur with the SECO CPU Board (00024001597). The SECO CPU can be
identified on the instrument by accessing “Diagnostics” and “System Info”. If the
letter following the Instrument serial number is “S”, that indicates a SECO CPU.
This symptom is most likely to occur on instruments above serial number 15750,
although it is possible that earlier serial numbers may freeze. The reported
Critical Care
incidence of the freezing has been very low at this time.
The vendor has identified the cause of the problem and developed a screening
test that has been very effective in eliminating the occurrence. All CPUs are now
screened and the good boards are marked with a green “dot”, placed on U-2.
There have been no instances of the freezing seen with a CPU that has been
screened and has the green dot on it. All instruments with serial number 17034
and above have CPUs that have been screened and have the green dot on
them. All CPU boards in stock, and all GEM SECO CPU and Cable kits
(00024001632) in Orangeburg stock have been verified to have the screened
CPU Boards.
Action Required: Check any stock of CPU Boards and Kits to determine if they
have the green dot. If there are any boards in your stock that are not screened,
contact Jim Carpenter ([email protected]) with a quantity of the unscreened
boards that you have. A quantity of screened boards will be sent to you to
replace the boards you have in stock, with an RGA to return the unscreened
boards. The unscreened boards are to be returned to Jim Carpenter in
Lexington. To facilitate the process, please remove the CPU boards from the kits
and replace them with the screened boards when they are received.
11-11
Service Information No. 01 – May 2005
The first priority is to replace all unscreened CPU Boards in your stock. After that
is complete, CPU Boards that are replaced for this specific problem will be
replenished following the same procedure. It is important that only CPU Boards,
without the green dot, and that experience this specific problem will be
replenished in this manner. Other CPU failures will continue to be addressed as
they are now.
Jim Carpenter
11-12