Bca Angio Series Data Sheet
Bca Angio Series Data Sheet
CODE MODEL
MI1014 BCA - 9RK Angio Plus 9” Rotating Anode n°2 – 19” Monochrome Angio 30fr/sec acq. Iris and parallel shutters
MI1015 BCA - 12RK Angio Plus 12” Rotating Anode n°2 – 19” Monochrome Angio 30fr/sec acq. Iris and parallel shutters
OPTIONAL
• Laser localizer
• Preview collimator
• Exporting software for 1Kx1K images recording (BMP or AVI format) on USB memory stick
• Dicom Store
• Dicom Worklist
ELECTRICAL CHARACTERISTICS
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
X– RAY TUBE
Model X20P
Anode Rotating
Anode slope 10°
Focal spot size according to IEC 336 0.3 Small Focus / 0.6 Large Focus
Anode power rating 5kW SF / 17kW LF
Max. Anode heat content 150 kJ (200 kHU)
Max. continuous anode thermal dissipation 300 W (400 HU/s)
MONOBLOC
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
IMAGE INTENSIFIER
Size 9” 12”
Model E5830SD-P4A E5796SD-P2A
Number of fields 3 (9” / 6” / 4.5”) 3 (12” / 9” / 6”)
Input nominal diameter 230 mm 310 mm
Central Resolution (typ.) 52 / 58 / 68 lp/cm 46 / 50 / 56 lp/cm
Conversion factor (cd*s/m2*µGy) 29 34
Contrast ratio (10% area) 30:1 30:1
Contrast ratio (10 mm dia.) 19:1 18:1
DQE (IEC Standard) 65% 65%
ANTI-SCATTER GRID
CCD CAMERA 1K × 1K
Matrix 1024 x 1024 pixels
Technology interline transfer CCD progressive scan with microlenses
Acquisition 1024 x 1024 x 14 bits
Contrast resolution 16384 grey levels
Signal - Noise Ratio 61 dB
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
ACQUISITION
Image acquisition 1024 x 1024 x 14 bit - 16384 grey levels
Video output DVI-A & VGA 1280 x 1024
# of images on hard disk 800.000 images 1k x 1K x 14 bit on H.D. SATA 500 GByte
Images acquisition rate 0.5/ 1 / 2 / 4 / 7.5 / 15 / 30 fps
Continuous Digital image rotation Yes
Window, Level, Brightness adj. tool Yes
Vertical image inversion Yes
Left / Right image inversion Yes
Gray scale inversion Yes
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
MONOCHROME MONITOR
Size 19” (483 mm)
Active area 14.80” x 11.85” (376 x 301 mm)
Type LCD TFT grey scale, medical
CONNECTIVITY AND CONTROL
Analog video input VGA
Digital video input DVI-D
ELECTRICAL CHARACTERISTICS
Standard power supply 110-230V
Line frequency 50-60 Hz
DISPLAY CHARACTERISTICS
Resolution 1MPixel (1280 x 1024)
Contrast 900:1
Luminance 1000 nit
Visibility angle 170° H / 170° V
Pixel pitch 0.294 mm
Aspect ratio 5:4
ALC (Ambient light compensation) YES
BLOS (Backlight Output Stabilization) YES
MECHANICAL CHARACTERISTICS
Dimensions 430 x 380 x 80 mm (LxHxW without basement)
Weight 8,6 kg
Mounting VESA 100
APPROVALS
CEI 60601-1 12/98, CEI 60601-1 04/98. CEI 60601-1 02/98
ENVIRONMENT
Operation From 0 °C to +45 °C
Humidity: 8% to 80% (not condensing)
Storage From –30°C to +70°C
Humidity: 8% to 80% (not condensing)
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
User’s interface Touch-screen 7” LCD monitor for all the operation parameters and messages concerning
possible faulty conditions.
The LCD Monitor can be rotated ± 180°
Languages Italian, English, French, German, Spanish (selection at setup)
I.I. field selection Electronic zoom selection according to the number of I.I. fields with automatic
limitation of the irradiated field.
Thermal units Real time control and visualization of monobloc thermal capacity percentage (% HU)
according to the selected parameters (kV - mA).
Continuous Fluoro up to 4’30” then there is an acoustic alarm. At the 5th minute of con-
tinuous fluoro the x-ray is stopped and the timer is re-setted.
Fluoroscopy control Multi-functioning footswitch:
• One pedal for continuous, pulsed, single shot (boost) and low dose pulsed fluoroscopy
(the selection occurs from the console)
Radiography control Two-step hand switch complete of spiral cable
Monitor positioning Two adjustable 19” Monochrome LCD monitor mounted on trolley
Safeties devices • General automatic magneto-thermic circuit breaker switch
• Safety key switch
• Emergency pushbutton
• Filament current
• Monobloc temperature
• Overload
• kV min/max. or faulty H.V.
• Stored data check
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
Continuous Fluoroscopy
Dose control (kV - mA) Automatic or manual
kV - mA correlation Automatic
Boost fluoroscopy
mA on all kV range 12 mA fixed on Small Focus
HCFLD
(High Contrast Fluoro Low Dose)
mA on all kV range 8 mA fixed on Small Focus
Digital Radiography
(One Shot)
mA on all kV range 12 mA fixed on Small Focus
Film Cinematography
mA Range 80 - 70 - 60 - 50 - 40 - 30 mA
Radiography
mA Range 25mA ÷ 100 mA
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
ENVIRONMENTAL CONDITIONS
Movement Manual.
Operator manual control of rear wheels.
swivelling front wheels.
Stationary foot-brake.
C-Arm wheels diameter • Rear: Two double wheels ø 150 x 51 mm
• Front: Double wheels ø 80 x 46 mm
Protection against cables squashing Semi-rigid rubber core-hitch on wheels.
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
TECHNICAL DATA
C-ARM Dimension in mm
B 829 mm 802 mm
C 983 mm 973 mm
D 1763 mm 1853 mm
E 448 mm 563 mm
®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®
TECHNICAL DATA
TROLLEY Dimension in mm