0% found this document useful (0 votes)
92 views

Bca Angio Series Data Sheet

The document provides product data and specifications for two BCA Plus digital mobile C-arm models. Key details include tube components, image intensifier and camera specs, acquisition rates, memory capacity, and optional features.

Uploaded by

alexi mejias
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
92 views

Bca Angio Series Data Sheet

The document provides product data and specifications for two BCA Plus digital mobile C-arm models. Key details include tube components, image intensifier and camera specs, acquisition rates, memory capacity, and optional features.

Uploaded by

alexi mejias
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 10

BCA PLUS - Angio Series

Product Data Digital Mobile C-Arm


®

COMPOSITION I.I. Tube X-ray Tube # Monitor Memory Collimator

CODE MODEL
MI1014 BCA - 9RK Angio Plus 9” Rotating Anode n°2 – 19” Monochrome Angio 30fr/sec acq. Iris and parallel shutters

MI1015 BCA - 12RK Angio Plus 12” Rotating Anode n°2 – 19” Monochrome Angio 30fr/sec acq. Iris and parallel shutters

OPTIONAL

• 24x30 cm cassette holder for 9” I.I. Tube

• 30x40 cm cassette holder for 12” I.I. Tube

• Laser localizer

• Dose Area Product (DAP)

• Preview collimator

• Sterile drapes set

• Exporting software for 1Kx1K images recording (BMP or AVI format) on USB memory stick

• Dicom Store

• Dicom Worklist

• DICOM Image Storage function for USB Memory Stick recording

• DICOM Image Storage function for CD/DVD burning

• Thermostatically controlled forced ventilation

• Hybrid graphic printers

ELECTRICAL CHARACTERISTICS

• Voltage Standard: 230 Vac (±10% ) single phase


Optional: 115 Vac (±10% ) single phase
• Frequency 50/60 Hz
• Max. Absorbed current 4.5 A @ 230 Vac in fluoro mode
20 A @ 230 Vac in radiography mode
• Line compensation Automatic
• Line resistance < 0.4 Ω (supply at 230 Vac)
• Standard mains plug 16 A (for supply at 230 Vac)

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

RADIOLOGICAL DATA ROTATING ANODE TUBE


Generator nominal power (IEC 60601-2-7) 8.0 kW @ 100kV
High frequency generator 40 kHz
Max. ripple < 2%
High voltage range (Rad & Fluoro) 40 ÷ 120 kV
Current range in Continuous Fluoroscopy 0.2 ÷ 4 mA
Current range in Pulsed Fluoroscopy 8 mA - 12 mA
Current range in Radiography 25 mA ÷ 100 mA
mAs range 1 ÷ 125 mAs
Exposure time in Radiography 0.02 ÷ 3 sec
Total filtration > 3,2 mmAl

X– RAY TUBE
Model X20P
Anode Rotating
Anode slope 10°
Focal spot size according to IEC 336 0.3 Small Focus / 0.6 Large Focus
Anode power rating 5kW SF / 17kW LF
Max. Anode heat content 150 kJ (200 kHU)
Max. continuous anode thermal dissipation 300 W (400 HU/s)

MONOBLOC

Monobloc thermal capacity 650 kJ (865 kHU)


Max. continuous thermal dissipation 85 W, 115 HU/sec, 6900 HU/min

COLLIMATOR Automatic type


• Iris shutters: beam collimation from the smallest field to the selected I.I.
Iris and parallel shutters type input field
• Semi transparent shutters: n°2 parallel shutters, Cu 0.6 mm, 360° rotating
Adjustment of the collimation on the displayed image without making x-ray.
Preview collimator (Optional) Automatic positioning of the iris and parallel shutters to the selected position,
when x-ray is ON

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

IMAGE INTENSIFIER

Size 9” 12”
Model E5830SD-P4A E5796SD-P2A
Number of fields 3 (9” / 6” / 4.5”) 3 (12” / 9” / 6”)
Input nominal diameter 230 mm 310 mm
Central Resolution (typ.) 52 / 58 / 68 lp/cm 46 / 50 / 56 lp/cm
Conversion factor (cd*s/m2*µGy) 29 34
Contrast ratio (10% area) 30:1 30:1
Contrast ratio (10 mm dia.) 19:1 18:1
DQE (IEC Standard) 65% 65%

ANTI-SCATTER GRID

Ratio 14:1 14:1

Focus 100 cm 100 cm

# of lines 60 lp/cm 60 lp/cm

CCD CAMERA 1K × 1K
Matrix 1024 x 1024 pixels
Technology interline transfer CCD progressive scan with microlenses
Acquisition 1024 x 1024 x 14 bits
Contrast resolution 16384 grey levels
Signal - Noise Ratio 61 dB

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

DIGITAL IMAGE SYSTEM MEMORY ANGIO 30 fr/sec

ACQUISITION
Image acquisition 1024 x 1024 x 14 bit - 16384 grey levels
Video output DVI-A & VGA 1280 x 1024

# of images on hard disk 800.000 images 1k x 1K x 14 bit on H.D. SATA 500 GByte
Images acquisition rate 0.5/ 1 / 2 / 4 / 7.5 / 15 / 30 fps
Continuous Digital image rotation Yes
Window, Level, Brightness adj. tool Yes
Vertical image inversion Yes
Left / Right image inversion Yes
Gray scale inversion Yes

Recursive filter - Noise reduction Yes - 4 selections: F0, F1, F2, F3


Max. Opacification during fluoroscopy acquisition Yes
Image Subtraction (DSA) Yes
Road mapping Yes
Teleinjector management Yes
POST PROCESSING
Grey scale inversion Yes
Window, Level, Brightness adj. tool Yes
Continuous Digital image rotation Yes
Overview windowing Yes, 7 images with scrolling function
Vertical image inversion Yes
Left / Right image inversion Yes
Electronic zoom (2x, 4x, 8x, 16x) Yes
Spatial filters, edge enhancement Yes - Sharpeness 3x3
Land marking Yes
Cineloop Yes - (Display of two different sequences simultaneously)
Auto play Yes
Shifting pixel Yes
Automatic mask acquisition Yes
Re-masking Yes
Image subtraction Yes
Road mapping Yes
Image Snap-Shot Yes
Preview collimator Yes - Optional
MEASURES (indicative)

Distance calculation Yes


Angles measuring Yes
UTILITY
Exporting software for 1Kx1K images recording Yes - Optional
(BMP for AVI format) on USB memory stick
DICOM Store Yes - Optional
DICOM Worklist Yes - Optional
DICOM Image Storage function for USB Memory Stick recording Yes - Optional
DICOM Image Storage function for CD/DVD burning Yes - Optional

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

MONOCHROME MONITOR
Size 19” (483 mm)
Active area 14.80” x 11.85” (376 x 301 mm)
Type LCD TFT grey scale, medical
CONNECTIVITY AND CONTROL
Analog video input VGA
Digital video input DVI-D
ELECTRICAL CHARACTERISTICS
Standard power supply 110-230V
Line frequency 50-60 Hz
DISPLAY CHARACTERISTICS
Resolution 1MPixel (1280 x 1024)
Contrast 900:1
Luminance 1000 nit
Visibility angle 170° H / 170° V
Pixel pitch 0.294 mm
Aspect ratio 5:4
ALC (Ambient light compensation) YES
BLOS (Backlight Output Stabilization) YES
MECHANICAL CHARACTERISTICS
Dimensions 430 x 380 x 80 mm (LxHxW without basement)
Weight 8,6 kg
Mounting VESA 100
APPROVALS
CEI 60601-1 12/98, CEI 60601-1 04/98. CEI 60601-1 02/98
ENVIRONMENT
Operation From 0 °C to +45 °C
Humidity: 8% to 80% (not condensing)
Storage From –30°C to +70°C
Humidity: 8% to 80% (not condensing)

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

FUNCTIONALITY DESCRIPTION DATA

User’s interface Touch-screen 7” LCD monitor for all the operation parameters and messages concerning
possible faulty conditions.
The LCD Monitor can be rotated ± 180°
Languages Italian, English, French, German, Spanish (selection at setup)

I.I. field selection Electronic zoom selection according to the number of I.I. fields with automatic
limitation of the irradiated field.
Thermal units Real time control and visualization of monobloc thermal capacity percentage (% HU)
according to the selected parameters (kV - mA).
Continuous Fluoro up to 4’30” then there is an acoustic alarm. At the 5th minute of con-
tinuous fluoro the x-ray is stopped and the timer is re-setted.
Fluoroscopy control Multi-functioning footswitch:
• One pedal for continuous, pulsed, single shot (boost) and low dose pulsed fluoroscopy
(the selection occurs from the console)
Radiography control Two-step hand switch complete of spiral cable
Monitor positioning Two adjustable 19” Monochrome LCD monitor mounted on trolley
Safeties devices • General automatic magneto-thermic circuit breaker switch
• Safety key switch
• Emergency pushbutton
• Filament current
• Monobloc temperature
• Overload
• kV min/max. or faulty H.V.
• Stored data check

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

OPERATIONAL MODES CHARACTERISTICS PERFORMANCES

Continuous Fluoroscopy
Dose control (kV - mA) Automatic or manual

mA fluoro variation range 0.2 ÷ 4 mA on Small Focus

kV - mA correlation Automatic

Safety timer • Acoustic alarm resettable after 4’30” fluoroscopy exposure.


• Automatic stop after 5 minutes of continuous X-ray emission

Boost fluoroscopy
mA on all kV range 12 mA fixed on Small Focus

X-ray pulse time 36 msec

X-ray pulse rate 6 Selections. 0.5 - 1 - 2 - 4 - 7.5 - 15 pulses/sec

• Acoustic alarm resettable after 4’30” fluoroscopy exposure.


Safety timer
• Automatic stop after 5 minutes of continuous X-ray emission

HCFLD
(High Contrast Fluoro Low Dose)
mA on all kV range 8 mA fixed on Small Focus

X-ray pulse time 36 msec

X-ray pulse rate 6 Selections. 0.5 - 1 - 2 - 4 - 7.5 - 15 pulses/sec

Safety timer • Acoustic alarm resettable after 4’30” fluoroscopy exposure.


• Automatic stop after 5 minutes of continuous X-ray emission

Digital Radiography
(One Shot)
mA on all kV range 12 mA fixed on Small Focus

X-ray flash time 66 msec

Film Cinematography
mA Range 80 - 70 - 60 - 50 - 40 - 30 mA

X-ray pulse rate 15 / 30 pps

Acquisition rate 0.5 - 1 - 2 - 4 - 7.5 - 15 - 30 fps

Sequence run time 12 sec

Radiography
mA Range 25mA ÷ 100 mA

Max current 100 mA

Exposure time 0.02 ÷ 3 sec

Selection x-ray parameters 3 point technique (kV, mA, time)

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

ENVIRONMENTAL CONDITIONS

• Temperature: from +10 °C to +40 °C


Environmental conditions in standard use • Humidity: 30% to 75% (not condensing)
• Pressure: from 700 to 1060hPa

• Temperature: from –25 °C to +70°C


Environmental conditions in transport and warehouse • Humidity: 10% to 90% (not condensing)
• Pressure: from 500 hPa to 1060 hPa

WEIGHT AND DIMENSIONS Description C-Arm 9” C-Arm 12” Trolley

Weight ~ 340 Kg ~ 340 Kg ~ 100 Kg

Width 779 mm 779 mm 860 mm

Depth 1940 mm 1940 mm 686 mm


Height 1763 mm 1853 mm 1776 mm

MECHANICAL CHARACTERISTICS C-Arm 9” C-Arm 12”

Vertical travel 400 mm motorized in 30 sec. 400 mm motorized in 30 sec

Horizontal travel 200 mm 200 mm

Panning movement ± 210° ± 210°

Orbital rotation 146° (+90° ÷ - 56°) 146° (+90° ÷ - 56°)

Pivotal rotation ± 10.5° ± 10.5°

Free Space 829 mm 802 mm


Depth 721 mm 721 mm

S.I.D. 983 mm 973 mm

Focus skin distance 200 mm 200 mm

Movement Manual.
Operator manual control of rear wheels.
swivelling front wheels.
Stationary foot-brake.
C-Arm wheels diameter • Rear: Two double wheels ø 150 x 51 mm
• Front: Double wheels ø 80 x 46 mm
Protection against cables squashing Semi-rigid rubber core-hitch on wheels.

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

TECHNICAL DATA

C-ARM Dimension in mm

9” I.I. Tube 12” I.I. Tube


A 148 mm 148 mm

B 829 mm 802 mm
C 983 mm 973 mm
D 1763 mm 1853 mm
E 448 mm 563 mm

®
BCA PLUS - Angio Series
Product Data Digital Mobile C-Arm
®

TECHNICAL DATA

TROLLEY Dimension in mm

PD-72-E 12/02/2013 Rev.04

Note: Data subjected to alteration without prior notice

ESSE3 Via Garibaldi 30


14022 Castelnuovo D.B. (AT)
® Tel +39 011 99 27 706
Fax +39 011 99 27 506
e-mail [email protected]
web: www.esse3-medical.com