This document provides instructions for installing a modification on KODAK DIRECTVIEW CLASSIC/ELITE CR Systems to enable direct connection hardware and configure the necessary software. The modification involves installing circuit boards and cables, and configuring the scanner and computer IP addresses to allow them to communicate directly without a network connection.

Uploaded by

jose
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
60 views

Modification Instructions For The Kodak Directview Classic/Elite CR Systems Service Code: 5258, 5259

This document provides instructions for installing a modification on KODAK DIRECTVIEW CLASSIC/ELITE CR Systems to enable direct connection hardware and configure the necessary software. The modification involves installing circuit boards and cables, and configuring the scanner and computer IP addresses to allow them to communicate directly without a network connection.

Uploaded by

jose
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 8

{Modification/Type1}{Production}{Carestream Health Inc.

}{Confidential}

Publication No. 8H9936


01JUN10
Supersedes 8H9936
17MAY10

Modification Instructions
for the
KODAK DIRECTVIEW CLASSIC/ELITE CR Systems
Service Code: 5258, 5259
Modification No. M17
Type 1 Selective

Purpose: To install direct-connect hardware on those KODAK DIRECTVIEW CLASSIC/ELITE CR Systems


that do not have Modification No. 17 installed and to configure the software that enables the scanner
and computer to communicate with each other.

Important
• Qualified service personnel must install this modification.
• See the Carestream Health Service Portal for possible updates before doing this modification.
• When doing the procedures in this document, you must use safe work practices and wear the correct personal
protective equipment (for example, safety eyewear) according to your company’s standard operating
procedures.
Service Effects: None
Special Requirements: None
Serial Numbers: ---
Installation Time: 1 hour
Special Tools: None
Parts Status: Available from Service Parts Management.
Parts Requirements: See Page 3.

Confidential

© CARESTREAM HEALTH, INC. 2010


PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the
subject matter gained by Carestream Health, Inc. prior to publication.
No patent license is granted by this information.
Carestream Health, Inc. reserves the right to change this information without notice, and makes
no warranty, express or implied, with respect to this information. Carestream Health shall not be
liable for any loss or damage, including consequential or special damages, resulting from any
use of this information, even if loss or damage is caused by Carestream Health’s negligence or
other fault.
This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.

Table of Contents
Description Page

Checking the Modification Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3


Installing the Modification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Installing the Direct-Connect Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Configuring the Direct-Connect Software. . . . . . . . . . . . . . . . . . . . . . . . . . 4
Completing the Modification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Publication History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

2 01JUN10 – 8H9936
Checking the Modification Components
Components required for this modification include:
Part No. Description Quantity
SP2G7266 Ethernet isolation circuit board 1
--- Standoff risers 2
--- Screws, M4 2
SP6H8594 Ethernet cable, 30.48 cm (1 ft) 1

Note
Each of the parts required for this modification must be ordered separately.

Installing the Modification


Installing the Direct-Connect Hardware

Important
• Because this modification is selective, it is not required.
• Newer KODAK DIRECTVIEW CLASSIC/ELITE CR Systems that already have the required direct-connect
hardware installed will be marked as having Modification No. 17 installed.

Note
The recommended method for connecting these CR Systems is to have a separate customer network connection for
the scanner and the computer. However, removing the scanner from the network, as outlined in this procedure, may
help minimize problems with the customer network if there are performance issues with the customer network or if
an extra IP address is not available.
[1] Does your CLASSIC/ELITE CR System already have Modification No. 17 installed?
Yes No
Advance to Configuring the Direct-Connect Continue with Step 2.
Software on Page 4.

[2] Remove the left side panel from the CLASSIC/ELITE CR System.

standoff
risers

[3] Locate the 2 threaded mounting holes in the unit frame.


[4] Attach the 2 standoff risers in the threaded holes.

8H9936 – 01JUN10 3
MODIFICATION INSTRUCTIONS, TYPE 1

external
socket

IEB
connection
Ethernet isolation
circuit board cable tie

[5] Mount the Ethernet isolation circuit board onto the standoff risers.

Note
Orient the circuit board as shown in the illustration.
[6] Disconnect the existing Ethernet cable that runs between the IEB board and the external socket.
[7] Connect the existing Ethernet cable from the IEB board to the upward pointing external socket.
[8] Connect the 30.48 cm (1 ft) Ethernet cable between the remaining socket on the circuit board and the socket
leading outside of the unit.
[9] Check that the elevator and the erase lamp assembly move through a complete range of motion without
disturbing the Ethernet cables.
[10] Secure the Ethernet cables in place using a cable tie.
[11] Install the left side panel.

Configuring the Direct-Connect Software

Important
• Perform this software configuration on the CLASSIC/ELITE CR Systems to enable the scanner and computer to
communicate with each other.
• If IP addresses other than those used in this procedure are required, change the IP settings and the computer
scanner NIC settings.
• Customers that use the 192.168.0 subnet for their network may require settings other than those identified in
this procedure. For assistance in changing the settings, contact the Carestream Health Technical Support Center
(TSC).

Caution
Dangerous Voltage

[1] De-energize the computer.

Important
• Do not plug the network into the customer network drop from the back of the Local User Interface (LUI).
• You must plug the scanner network cable into the unused extra NIC connector (covered with white tape) on the
back of the computer.

4 01JUN10 – 8H9936
Model HP5700 Computer [2] Remove the white tape from the unused extra NIC
connector on the back of the computer.
[3] Plug the network cable from the scanner into the
NIC connector.
unused
POWER
PS/2 PORTS Note
• Use the same network cable that was used to plug
into the customer network drop.
COM3/unused
• A crossover cable is not required.
(ignore the
COM1/unused
painted PC label
that says COM2)
MONITOR

USB
KEYBOARD
USB MOUSE

unused
extra NIC BAR CODE
READER

TOUCHSCREEN CUSTOMER
NETWORK
H226_1027GCA
H226_1027GC

[4] Enter the network settings at the LUI to put the


scanner on a LAN with the unused extra NIC in the
computer.
[5] At the LUI, touch the Service Functions button.

Service Functions button

8H9936 – 01JUN10 5
MODIFICATION INSTRUCTIONS, TYPE 1

[6] Touch the service password: 4213


[7] Touch the green check mark to accept the
password.

[8] Touch:
• Network Configuration
• Scanner TCP/IP Settings
[9] Enter the network settings for the scanner:
• IP Address
• Gateway
• Netmask

6 01JUN10 – 8H9936
[10] Enter the network settings for the console
computer.

Note
The console computer settings will be the same
address programmed in for the computer scanner NIC
card in Step 4 on Page 5.
[11] Using the green arrow, exit all of the screens.

Note
The settings will be saved, and the scanner will reboot
automatically.

Caution
Dangerous Voltage
[12] Energize the computer.

Completing the Modification


[1] Circle 17 on the modification label.
[2] Record the following information for feedback:

Scan Installation
Service Activity Time
K-number System Code/s Serial No. Code CRM PSL (actual) Part No.
CLASSIC/ 5258, M17 SC5258M17,
ELITE CR 5259 SC5259M17

8H9936 – 01JUN10 7
Publication History

Publication Publication Changed


Date No. ECO No. Pages File Name Notes
13JAN10 8H9936 --- All 8h9936.fm New Publication
17MAY10 8H9936 --- Various 8h9936.fm Revised
01JUN10 8H9936 --- Various 8h9936.fm Revised

DIRECTVIEW is a trademark of Carestream Health, Inc.


The Kodak trademark and trade dress are used under license from Kodak.

Printed in U.S.A. • 8h9936.fm

Carestream Health, Inc.


150 Verona Street
Rochester, NY 14608

You might also like